○ This instrument has been designed, manufactured and tested
according to IEC 61010: Safety requirements for Electronic
Measuring apparatus, and delivered in the best condition after
passing quality control tests. This instruction manual contains
warnings and safety rules which must be observed by the user
to ensure safe operation of the instrument and to maintain it in
safe condition. Therefore, read through these operating
instructions before using the instrument.
# WARNING
● Read through and understand instructions contained in this
manual before starting to use the instrument.
● Keep the manual at hand to enable quick reference whenever
necessary.
● The instrument is to be used only in its intended applications.
● Understand and follow all the safety instructions contained in
the manual. It is essential that the above instructions are
adhered to.
Failure to follow the above instructions may cause injury,
instrument damage and/or damage to equipment under test.
○ The symbol # indicated on the instrument means that the user
must refer to the related parts in the manual for safe operation
of the instrument. It is essential to read the instructions wherever
the symbol # appears in the manual.
#DANGER is reserved for conditions and actions that are
likely to cause serious or fatal injury.
#WARNING i s reserved for conditions and actions that can
cause serious or fatal injury.
#CAUTION is reserved for conditions and actions that can
● Do not use this instrument to test the circuits in which 300V
or higher potentials exist in Measurement category IV (CAT
IV) environment or 600V or higher potentials exist in CAT III
environment.
● Do not attempt to make measurements in the presence of
ammable gasses. Otherwise, the use of the instrument may
cause sparking, which can lead to an explosion.
● Never attempt to use the instrument if its surface or your
hand is wet.
● Be careful not to short-circuit the power line with the metal
part of the test leads when measuring a voltage. It may
cause personal injury.
● Do not exceed the maximum allowable input of any measuring
ranges.
● Do not press the Test button when connecting the test leads
to the instrument.
● Never open the Battery compartment cover during a
measurement.
● To prevent possible electrical shock, do not touch the circuit
under test during an insulation resistance measurement or
right after a measurement.
#DANGER
● Never attempt to make any measurements if any abnormal
conditions such as broken case and exposed metal parts are
noted.
● Do not rotate the Function switch with the test leads
connected to the object under test.
● Do not install substitute parts or make any modications to
the instrument. Return the instrument to your local
KYORITSU distributor for repair or re-calibration.
● Do not try to replace the batteries if the surface of the
instrument is wet.
● Insert the plug into the terminal rmly when using test leads.
● Ensure that the instrument is switched off before opening the
Battery compartment cover for battery replacement.
● Before starting a measurement, conrm that the Function
switch is at an appropriate position.
● Set the Function switch to “OFF” position after use. Remove
the batteries if the instrument is to be stored and will not be
in use for a long period.
● Do not expose the instrument to direct sunlight, high temperatures and humidity or dew.
● Use a damp cloth with alcohol for cleaning the instrument
and the areas around the measuring terminals.
● When this instrument is wet, please store it after it dries.
Symbols
Danger of possible electric shock
Instrument with double or reinforced insulation
User must refer to the manual
#
DC
AC
Earth terminal
Crossed-out wheel bin symbol (according to WEEE
Directive: 2002/96/EC) indicating that this electrical
product may not be treated as household waste, but that
it must be collected and treated separately.
To ensure safe operation of measuring instruments, IEC 61010
establishes safety standards for various electrical environments,
categorized as O to CAT IV, and called measurement categories.
Higher-numbered categories correspond to electrical environments
with greater momentary energy, so a measuring instrument
designed for CAT III environments can endure greater momentary
energy than one designed for CAT II.
O : Circuits which are not directly connected to the mains
power supply.
CAT II : Electrical circuits of equipment connected to an AC
electrical outlet by a power cord.
CAT III : Primary electrical circuits of the equipment connected
directly to the distribution panel, and feeders from the
distribution panel to outlets.
CAT IV : The circuit from the service drop to the service
entrance, and to the power meter and primary
overcurrent protection device (distribution panel).
O: Device which is
not directly
connected to the
mains power supply
KEW3121B (2500V/100GΩ) and KEW3122B (5000V/200GΩ)
are battery-operated high voltage insulation testers.
● Designed to meet the following safety standards:
IEC 61010-1,-2-030 (CAT III 600V/CAT IV 300V Pollution
degree 2)
IEC 61010-031 (Requirements for hand-held probes)
● Suited for heavy duty electrical maintenance and servicing
ondustrial installations, cables, transformers, generators and
switchgear where high voltage insulation tests are required.
● Dual scales for low an d high ranges which change automatically.
Colour coded scales for easy reading plus LED's that illuminate
in matching colour.
● Hard carrying case furnished as standard accessory. Houses
both instrument and test leads in compact form. Made of
plastic, it is highly water resistant.
● Rated output voltage is maintained down to 0.1GΩ for
KEW3121B, 0.2GΩ for KEW 3122B. This permits accurate
measurements of low insulation resistance.
●
Audible high voltage warning at insulation resistance measurement
● Optional adapter MODEL8324 is available for connection to
recorder and enables cable insulation monitoring.(Option)
IEC 61010-1,-2-030
Measurement CAT III 600V Pollution degree2
Measurement CAT IV 300V Pollution degree2
IEC 61010-031 Standard for hand-held probes
MODEL7165A(CAT IV 600V)
MODEL7264(CAT IV 600V)
MODEL7265(CAT IV 600V)
* When KEW3121B/3122B and the test lead
IEC 61326-1,-2-2 EMC standard
IEC 60529 IP40
EN 50581 RoHS Directive
● Measuring range and accuracy
ModelKEW3121BKEW3122B
Rated voltage2500V5000V
Measuring range
Temperature:
23ºC±5ºC
Temperature:
-10ºC to +40ºC
Output voltage
are combined and used together, whichever
is lower category either of them belongs to is
applied.
0 - 2GΩ/ 1 - 100GΩ
(Automatic change)
First effective
measuring range
Accuracy±5% of indicated value
Second effective
measuring range
Accuracy
First effective
measuring range
Accuracy±10% of indicated value
Second effective
measuring range
Accuracy±20% of indicated value or 1.0% of scale length
0.1 - 50GΩ0.2 - 100GΩ
Measuring ranges other than listed above , 0 and ∞
±10% of indicated value or 0.5% of scale length
0.1 - 50GΩ0.2 - 100GΩ
Measuring ranges other than listed above , 0 and ∞
●Accessories : Line probe: MODEL 7165A
(Including MODEL8255 Straight type prod
with molded parts)
Earth cord: MODEL 7264
Guard cord: MODEL 7265
LR14(Alkaline battery size C)x 8pcs
Instruction manual
Hard Case: MODEL 9179
Pickel Type Prod: MODEL 8019
Straight Type Prod: MODEL 8254
●Optional : Adaptor for recorder:MODEL8324
accessories 3m Line probe with alligator clip: MODEL7168A
15m Line probe with alligator clip: MODEL7253
3 Earth terminal
4 LED for high & low range indication
5 PRESS TO TEST button
6 Function switch
7 Meter movement zero adjust
8 Guard cord (green)
9 Earth cord (black)
10 Line probe
● Be careful. High voltage may present across LINE and
EARTH terminals of the instrument when the PRESS TO
TEST button is operated. Make sure to connect the Earth
terminals of the circuit under test and of this instrument to
earth. The buzzer sounds continuously while measuring
insulation resistance.
5-1 Mechanical Zero Adjustment
Set the Function switch to “OFF” position. Adjust the Meter
movement zero adjust located at the center of the Front
panel with a screwdriver so that the pointer of the instrument
reads “∞“ symbol on the scale.
5-2 Battery Check
Set the Function switch to “BATT. CHECK” position and press
the PRESS TO TEST button. The battery voltage is enough
when the the pointer stays in BATT. GOOD area or to the
right of this area.
If not, replace the batteries with new ones.
NG
# DANGER
OK
●
Do not hold down or lock the PRESS TO TEST button when
checking the battery voltage. It may lead to shorter battery life.
● Use a high voltage detector and conrm that there is no
electrical charge in the circuit under test.
● Wear a pair of insulated gloves for high voltage.
● If the Function switch is at the insulation resistance range
and the Test button is being pressed down, high voltages are
generated and applied to the test leads and the circuit under
test continuously. Do not touch the circuit or the test leads.
● The Battery compartment cover must be closed and screwed
during a measurement.
● Do not make measurement when thunder rumbling.
● Connect the Earth Cord (black) to the earth terminal of the
circuit under test.
This instrument measures insulation resistances and checks
whether the insulations of electrical equipments or circuits are in
good condition or not. So, please check the voltage that can be
applied to the object under test before making a measurement.
Note:
● The insulation resistance values of the objects under test may
not be stable, and the readings may be unstable.
● Bleep sound may be heard during an insulation resistance
measurement, but it is not malfunction.
●It takes time to measure a capacitive load.
● At insulation resistance measurement, positive (+) voltage is
outputted from the Earth terminal and negative (-) voltage is
outputted from the Line terminal.
Connect the Earth cord to the Earth (ground) terminal.
It is recommended to connect the positive(+) pole to the earth
side when measuring insulation resistance against the ground
or when a part of the equipment under test is earthed.
With this connection, smaller measured value can be obtained
(1) Check the voltage which can be applied to the circuit under
test, and set the Function switch to a desired insulation
resistance range.
(2) Connect the Earth cord (black) to the Earth terminal of the
circuit under test.
(3) Place the tip of the Line probe (red) to the circuit under test.
Then press the “PRESS TO TEST” button. The buzzer sounds
intermittently during a measurement.
(4) When the green LED illuminates, read insulation resistance
on the outer scale (for high range). Use the inner scale where
the red LED illuminates.
# Caution
Always turn off the
Breaker for the circuit
under test.
(5) After a test, release the PRESS TO TEST button and wait for
the same amount of time that it took for the measurement
performed. Keep the Line probe and the circuit tested
connected until electrical charges stored in the circuit under
test are completely discharged.
# DANGER
● Do not touch the circuit under test immediately after testing.
Capacitances stored in the circuit may cause electric shock.
● Leave the test leads connected to the circuit and never touch
the circuit until electrical discharge completes.
(6) Set the Function switch to “OFF” position, and disconnect the
When measuring the insulation resistance of a cable, leakage
current owing on the surface of cable jacket and the current
owing inside the insulator are mixed and may cause error in
insulation resistance value. In order to prevent such error,
wind a conductive wire around the point where leakage
current ows. Then connect it to the Guard terminal. This is
to move out the surface leakage resistance of the cable
insulation to measure only the volume resistance of insulator.
Make sure to use the Guard cord supplied with this instrument
to connect the instrument to Guard terminal.
* It is possible to move out the surface leakage resistance of the
insulation and measure only the volume resistance by using the
Guard terminal. This is helpful when performing tests in humid
● Do not open the Battery compartment cover if the surface of
the instrument is wet.
● Never open the Battery compartment cover during a
measurement. To avoid electric shock, power off the
instrument and disconnect the test leads from the instrument
when replacing batteries.
● Always attach and close the Battery compartment cover
when making a measurement.
#CAUTION
● Do not mix new and old batteries or brands of batteries.
● Make sure to install batteries in correct polarity as marked
inside.
(1) Set the Function switch to “OFF” position, and disconnect the
test leads from the instrument.
(2) Loosen the Battery compartment cover-xing screws, and
remove the Battery compartment cover. Remove all eight
batteries and replace them with new ones.
(3) After replacing batteries, be sure to tighten up the screw for
the Battery compartment cover.
Make sure to install
batteries in correct
polarity as marked
This instrument is managed by our company's quality standard
and is delivered in the best condition after passed the inspection.
But in the dry time of winter static electricity sometimes builds up
on the meter cover due to the characteristic of plastic.
When static electricity builds up on the meter cover and affects
the meter reading, use a cloth dampened with off-the-shelf antistatic agent or detergent to wipe the meter cover surface.
# CAUTION
● When the pointer deects by touching the surface of this
tester or zero adjustment cannot be made, do not try to make
measurements.
● Anti-static agent has been applied to the surface of the Meter
cover of this instrument. Clean the surface of the cover gently
when it gets dirty.
● To avoid possible deforming or discoloring, do not use
organic solvents.
Attach MODEL8255 to the test leads to use the instrument in
CAT II or higher environments. MODEL8254 and 8019 have
exposed large metal parts, therefore, they may short circuit or
damage the objects under test, cause re or pose fatal or
serious injuries to the users.
(1) Tip metal parts
MODEL8255: Standard Prod (straight type, with molded
parts)
MODEL8254: Straight Type Prod
MODEL8019: Pickel Type Prod
To be used to hook the instrument.
(2) Replacement
Turn the Line probe counterclockwise to remove the attached
tip metal. Put the tip metal you want to use to the hexagon
socket and turn it clockwise together with the tip of probe,
Waste Electrical and Electronic Equipment (WEEE), Directive
2002/96/EC
This Product complies with the WEEE Directive (2002/96/EC)
marking requirement. The afxed product label (see below)
indicates that you must not discard this electrical/electronic
product in domestic household waste.
Product Category
With reference to the equipment types in the WEEE directive
Annex 1, this product is classied as a “Monitoring and Control
instrumentation” product.
Disposing lead-storage batteries
When you throw away the batteries, be sure to cover their
positive and negative terminals and always observe local laws
and regulations.
Insufcient insulation of the terminals may cause explosion or
re because electrical energies remain in lead-storage batteries