OperatingSystem :Dual Integration
Display :Field effect 3・1/2 digit liquid crystal
display with maximum count of 1999
Rangeselection :Manual
OverrangeIndication :1is displayed on the highest digit
except for 1000A AC range.
ResponseTime :Approx. 3 seconds
SampleRate :Approx. 3 times per second
DataHold :Available on all ranges
Locationforuse :Indoor use, Altitude up to 2000m
Temperature& Humidity :23℃
±
5℃, relative humidity up to 85%
forSpecifiedAccuracy without condensation
StorageTemperature :-20〜60℃, relative humidity up to 80%
&Humidity without condensation
OperatingTemperature :0〜40℃,relative humidity up to 85%
&Humidity without condensation
PowerSource :One 6 F22 or equivalent battery
LowBatteryWarning :
symbol is shown on the display.
CurrentConsumption :
Approx. 5mA max.(use Back Light:8mA max.)
Standard(Safety) :IEC61010-1
IEC61010-2-032
CAT.Ⅲ,AC300V,Pollution degree 2
(EMC) :EN 61326
・EN 55022
・
EN 61000-4-2 (performance criterion B)
・
EN 61000-4-3 (performance criterion B)
OverloadProtection :1500A AC max for 1 minute
WithstandVoltage :3700V AC for 1 minute between
electrical circuit and housing cases
InsulationResistance :50MΩ or greater at 1000V between
electrical circuits and housing cases or
metal parts of jaws
ConductorSize :Approx. 68mm diameter max.
Dimensions/Weight :250(L)
×
130(W)×50(D)mm.
Approx. 600g (including battery)
Accessories :6F22 battery
Carrying case
Instruction manual
OptionalAccessories :Model 7073 two-way output cord
PeakHold :Response time selectable from approx.
10ms and 100ms.
1/√2 of the peak current is displayed,
which means an RMS reading will be
obtained when current having sinusoidal
waveform in measured.
Peak hold reading varies by approximately
±
1% max of full scale in the first minute at
23℃
±
5
℃ and relative humidity of 75%
without condensation.