Kyoritsu Electrical Instruments 1020R User manual

Page 1
INSTRUCTION MANUAL
y
A
A
A
A
A
A
11-15 92-2256
c
c
c
DIGITAL MULTIMETER
KEW1020R
LCD
SELECT Key
Function Switch
COM Terminal
Cap Barrier
1. Safety Warnings
WARNING
Read through and understand the instructions contained in this
manual before using the instrument.
Keep the manual at hand to enable quick reference whenever
necessary.
The instrument is to be used only in its intended applications. Understand and follow all the safety instructions contained in the
manual.
It is essential that the above instructions are adhered to. Failure
to follow the above instructions may impair the protection provided by the instrument and test leads, and may cause injury, instrument damage and/or damage to equipment under test.
HOLD / Backlight Key
REL / MAXMIN Key
Voltage Terminal
The symbol indicated on the instrument means that the user must refer to the related parts in the manual for safe operation of the instrument. It is essential to read the instructions wherever the symbol appears in the manual.
DANGER is reserved for conditions and actions that
WARNING is reserved for conditions and actions that
CAUTION is reserved for conditions and actions that
Symbols listed below are used on this instrument.
User must refer to the manual.
Instrument with double or reinforced insulation.
AC DC Ground (Earth)
This instrument complies to WEEE Directive (2002/96/EC). Please contact your local distributor at disposal.
are likely to cause serious or fatal injury.
can cause serious or fatal injury.
can cause injury or instrument damage.
Measurement Category
O Circuits which are not directly connected to the mains power supply. CAT I I Primary electrical circuits of equipment connected to an AC
CAT III Primary electrical circuits of the equipment connected directly
CAT I V The circuit from the service drop to the service entrance, and
This instrument is designed for CAT IV 300V/ CAT III 600V / CAT II 1000V. Test leads M-7066A with the supplied caps are designed for CAT IV600V/ CAT III 1000V and without the caps are for CAT II 1000V.
electrical outlet by a power cord.
to the distribution panel, and feeders from the distribution panel to outlets.
to the power meter and primary over current protection device(distribution panel).
DANGER
Never make measurements under the circumstances exceeding the
designed measurement category and the rated voltage of the instrument.
Do not attempt to make measurement in the presence of flammable
gasses. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion.
Never attempt to use the instrument if its surface or your hand is wet. Do not exceed the maximum allowable input of any measuring
range.
Never open the Battery compartment cover during a measurement. To avoid electrical shock by touching the equipment under test or its
surroundings, be sure to wear insulated protective gear.
Test leads to be used for voltage measurements shall be rated as
appropriate for Measurement Category III or IV according to IEC 61010-031 and shall have a voltage rating of 1000V or higher.
Barriers on the test leads provide protection to keep your fingers and
hands from touching an object under test. Keep your fingers and hands behind the barriers during measurement.
WARNING
Never attempt to make measurement if any abnormal conditions,
such as broken case and exposed metal parts are found on the instrument or test leads.
Verify proper operation on a known source before use or take action
as a result of the indication of the instrument.
Firmly attach the caps to the test leads when performing
measurements in CAT III or higher test environments. When KEW1020R and the test leads are combined and used together, whichever is lower category & voltage to earth either of them belong to is applied.
Do not rotate the Function Switch if the instrument and the
equipment under test are connected.
Do not install substitute parts or make any modification to the
instrument. For repair or re-calibration, return the instrument to your local KYORITSU distributor.
Use of this instrument is limited to domestic, commercial and light
industry applications. Strong electromagnetic interference or strong magnetic fields, generated by large currents, may cause malfunction of the instrument.
Firmly insert the test leads. Do not pull or twist the test leads to prevent the risk of damage. Power off the instrument after use. Remove the battery if the
instrument is to be stored and will not be in use for a long period.
Do not expose the instrument to direct sunlight, high temperature
and humidity or dewfall.
Use a cloth dipped in water or neutral detergent for cleaning the
instrument. Do not use abrasives or solvents.
NOTE
The LCD shows some digits at the Voltage range even while the test
leads are open. And, it may show some digits instead of 0 even if the test leads are shorted. However, these phenomena don’t affect measurement results.
A resistance measurement takes time to settle the reading if there is
CAUTION
a high resistance or capacitance components.
2. Specification
Accuracy (Temperature: 23 ± 5°C, Humidity: 45 - 75%)
ACV / RMS (Auto Range)
Range Display Range Accuracy(sine wave) 6V 0.000, 0.006 - 6.299V 60V 5.70 - 62.99V 600V 57.0 - 629.9V 1000V 570 – 1049V ±1.3 %rdg±3dgt (40-500Hz)
Guaranteed accuracy : 0.01V-1000V, less than 1500V peak Input impedance : approx. 10M Input protective voltage : AC/DC1200V 10 sec.
Hz Frequency - ACV measurement (Auto Range)
Range Display Range
99.99Hz 10.00 - 99.99Hz
999.9Hz 95.0 - 999.9Hz
9.999kHz 0.950 - 9.999kHz
99.99kHz 9.50 - 99.99kHz Guaranteed accuracy : 10Hz-99kHz Input protective voltage : AC/DC1200V 10 sec.
% DUTY - ACV measurement
Range Display Range
99.9 % 0.0 – 99.9 % ±1.0%rdg±3dgt (50/60Hz)
Guaranteed accuracy : 10%-90% Input protective voltage : AC/DC1200V 10 sec.
DCV (Auto Range)
Range Display Range Accurac
6.000V 0.000 - ±6.299V
60.00V ±5.70 - ±62.99V
600.0V ±57.0 - ±629.9V 1000V ±570 - ±1049V ±0.8 %rdg±3dgt
Guaranteed accuracy : 0V-±1000V Input impedance : approx. 11M(6V range) / 10M(60/600/1000V range) Input protective voltage : AC/DC1200V 10 sec
ACmV / RMS
Range Display Range Accuracy (sine wave)
600.0mV 0.0, 0.9 - 629.9mV ±2.0 %rdg±3dgt (40-500Hz)
Guaranteed accuracy : 1.2mV-600mV, less than 900mV peak Input impedance : approx. 900k Input protective voltage : AC/DC1000V 10 sec.
AC Clamp Sensor / RMS (Auto Range)
Range Display Range Accuracy (sine wave)
60.00
200.0
0.00, 0.09 - 62.99
57.0 - 209.9
+ Sensor accuracy (40-500Hz)
Direct reading from 10mV / A output Clamp sensor
Guaranteed accuracy : 0.12A-200A, less than 300A peak Input impedance : approx. 900k
±1.0 %rdg±3dgt (40-500Hz)
Accuracy
(sine wave)
±0.1%rdg±3dgt
Accuracy
(Square wave)
±0.5 %rdg±3dgt
±2.0 %rdg±3dgt
DCmV
Range Display Range Accuracy
600.0mV 0.0 - ±629.9mV ±1.5 %rdg±3dgt
Guaranteed accuracy : 0mV-±600mV, Input impedance : approx. 900k Input protective voltage : AC/DC1000V 10 sec.
DC Clamp Sensor (Auto Range)
Range Display Range Accuracy
60.00
200.0A ±57.0 - ±209.9A
0.00 - ±62.99
±1.5 %rdg±3dgt
+ Sensor accuracy
Direct reading from 10mV / A output Clamp sensor
Guaranteed accuracy : 0A-±200A, Input impedance : approx. 900k
Resistance (Auto Range)
Range Display Range Accuracy
600.0 0.0 - 629.9 ±0.5 %rdg±4dgt
6.000k 0.570 - 6.299k
60.00k 5.70 - 62.99k
600.0k 57.0 - 629.9k
6.000M 0.570 - 6.299M
40.00M 5.70 - 41.99M ±1.5 %rdg±3dgt
Guaranteed accuracy : 0-40M, Open-loop Voltage : less than 3V Measurement current : less than 1mA Input protective voltage : AC/DC1000V 10 sec.
±0.5 %rdg±2dgt
Continuity
Range Display Range Accuracy
600.0 0.0 - 629.9
Open-loop Voltage : less than 3V, Measurement current : less than 1mA Input protective voltage : AC/DC1000V 10 sec.
Buzzer threshold value : less than 90
Diode
Range Display Range Accuracy
2.000V 0.000 - 2.099V ±5 %rdg±5dgt
Guaranteed accuracy : 0V-2V, Open-loop Voltage : less than 3V Measurement current : approx.0.5mA (Vf=0.6V) Input protective voltage : AC/DC1000V 10 sec.
Capacitance (Auto Range)
Range Display Range Accuracy
60.00nF 0.00 - 62.99nF
600.0nF 57.0 - 629.9nF
6.000µF 0.570 - 6.299µF
60.00µF 5.70 - 62.99µF
600.0µF 57.0 - 629.9µF
1000µF 570 - 1049µF * Accuracy after canceling floating capacitance using REL function. Guaranteed accuracy : 0nF-1000µF Input protective voltage : AC/DC1000V 10 sec.
Measuring method : modulation Over-range indication : OL Measurement cycle : 2.5 times per second
Crest factor : less than 3 (45-65Hz)
Applicable standards :
Withstand voltage : AC5160Vrms 5sec between circuit and enclosure IP rating : IP40 (IEC60529) Insulation resistance : 100M or more /1000V between enclosure
Operating temperature and humidity range :
Storage Temperature and humidity range :
Power source : DC3V R03/LR03 (AAA) × 2 Current consumption : 3mA or less Battery life : Approx. 200 hours (ACV, continuous, no load, with R03) Dimension, Weight : 155(L)×75(W)×40(D)mm,
Accessories : Test leads (M-7066A), Instruction manual,
Option : Magnet hanger strap (M-9189)
(1000µF range of Capacitance function 0.05 times per second)
For non-sinusoidal waveforms, add ±0.5 %rdg±5dgt (Applicable functions : ACV, ACmV, AC clamp sensor)
IEC 61010-1 /61010-2-033 : CAT IV 300V / CAT III 600V / CAT II 1000V Pollution degree 2, Indoor use, Altitude up to 2000m IEC61010-031 (Test leads Model 7066A) IEC 61326 (EMC) , EN 50581 (RoHS)
and electrical circuit
0 to 40°C, 80%RH or less (no condensation)
-20 to 60°C, 80%RH or less (no condensation)
approx. 250g (including batteries and Wing-type holder)
Battery R03 (AAA) 2pcs, Wing-type holder
Test leads with alligator clips (M-7234) AC Clamp Sensor (KEW8161), AC/DC Clamp Sensor (KEW8115)
±2.0 %rdg±5dgt*
±5.0 %rdg±5dgt
3. ACV / DCV / ACmV / DCmV Measurement
DANGER
Before starting a measurement, always check and confirm the Function switch is in the appropriate measurement position.
Never make measurement on a circuit in which voltage over 1000V exists.
Keep your fingers and hands behind the barrier during
measurement.
3.1 ACV / DCV/ ACmV / DCmV measurement
(1) Set the Function switch to ACV, DCV, ACmV or DCmV position.
For frequency or DUTY measurement, set the switch to ACV and press the SELECT key.
(2) Connect the test leads to the Voltage and COM terminals.
ACV / ACmV DCV / DCmV
Press the SELECT key to toggle functions
3.2 Clamp sensor (option) measurement
(1) Set the Function switch to ACmV or DCmV position and press the
SELECT key. LCD shows “SENSOR”.
(2) Connect the clamp sensor to the Voltage and COM terminals.
AC clamp sensor
Default :ACmV
Default :DCmV
Press the SELECT key to toggle functions
NOTE
If the connection is reversed, the “ - “ mark will be displayed on the
LCD. (DCV measurement).
Press the REL key to adjust the reading on the DC clamp sensor to
“0”.
Find Quality Products Online at: sales@GlobalTestSupply.com
The sensor mode corresponds to direct reading with 10mV/A
output clamp sensor.
Default : ACV
Black
DC clamp sensor
Black
Frequency DUTY
Red
Sensor mode
(ACA)
Sensor mode
(DCA)
Red
4. Resistance / Diode / Continuity / Capacitance Measurement
WARNING
Never use the instrument on an energized circuit. Discharge the capacitor before starting a capacitance measurement.
(1) Set the Function switch to the resistance, continuity or capacitance
position. For diode measurement, press the SELECT key.
Resistance
(2) Connect the test leads to the Voltage and COM terminals.
Resistance or Capacitance
NOTE
LCD shows ”OL” when the test leads are open.
(except for capacitance measurement)
The LCD shows “OL” if the test lead connection is reversed for diode
measurement.
Measurement time on 600F/1000F range is a bit long.
(20sec max.)
5. Other Functions
REL Function
Press the REL key to enable this function and store the measured value to display the differences between the stored value and the values measured in further tests. The measurement range will be fixed when the REL function is enabled, and the measuring range will be between the initial value and the full scale value.
Press the REL Key again to release the stored value.
” appears and “AUTO” disappears when REL Key is pressed.
*To activate REL function, MAX/MIN function should be disabled.
Sleep Function
Automatically powers off the instrument in about 15 min after the last switch operation.
To exit from the Sleep mode, rotate the Function switch or press any key. To disable the Sleep function, press the HOLD/Backlight key and power on the instrument. Confirm that the LCD shows “P.OFF” about 1 sec.
One press: REL ON Another press: REL OFF
Continuity / Diode
Default : Continuity
Press the SELECT key to toggle functions
Continuity
Capacitance
Diode
Diode
Black Red
www.GlobalTestSupply.com
Max / Min value display function This function is to display the measured max and min values on the
LCD during a measurement. Press the MAX/MIN key 1 sec or longer to start recoding of max and min values. Then the LCD shows the latest max value. After that, the min and present values can be toggled and checked by pressing the MAX/MIN key (less than 1 sec). To disable this function, press the MAX/MIN key 1 sec or longer.
Max value
* To activate MAX/MIN function, REL function should be disabled.
Data hold function
Press the HOLD key (less than 1 sec). The LCD shows “H” mark and the reading will be held (Data hold mode). The max/ min values are not updated in MAX/MIN mode. Press the HOLD key again (less than 1 sec) to release the display.
Backlight function
Press the Backlight key 1 sec or longer to turn
on the backlight. Press the Backlight key another 1 sec or longer to turn it off. The light automatically turns off in 1 min.
Low battery indication
The LCD shows “ ” mark when the batteries fall below the normal operating voltage.
How to store test leads
Attach the Wing-type holder to the back of the instrument to store the test leads.
1. Attach the Wing-type holder to the back of the instrument.
2. Fit the barrier on the test lead into the groove between the
3. Then it will be easier to see the displayed readings during a
Other convenient functions
Less than 1 se
Replace the batteries with new ones when this mark appears.
instrument and the Wing-type holder.
Wing-type holder
1.
measurement.
Tilt stand
Min value
Less than 1 se
2.
Present value
Less than 1 se
3.
Magnet hanger strap
(option)
6. Battery Replacement
WARNING
Replace the battery when the " " mark- low battery voltage
warning- appears on the LCD. Otherwise, precise measurement cannot be made. If the battery is completely exhausted, the LCD goes blank without showing the " " mark.
Disconnect the test leads from the object under test and power
off the instrument before opening the Battery compartment cover for battery replacement.
Do not try to replace the battery if the surface of the instrument
is wet.
CAUTION
Do not mix old and new batteries. Brand and type of the
batteries to be used should be harmonized.
Install a batteries in correct polarity as indicated in the Battery Compartment.
(1) Set the Function Switch to "OFF" position. (2) Untighten the screw on the back of the instrument. (3) Remove the Battery compartment cover and replace the
batterie.
(4) Attach the cover to the instrument and then secure the cover
by tightening the screw.
Battery
Kyoritsu reserves the rights to change specifications or designs described in this manual without notice and without obligations.
Screw