INSTRUCTION MANUAL
DIGITAL MULTIMETER
KEW1020R
LCD
SELECT Key
Function Switch
COM Terminal
Cap Barrier
1. Safety Warnings
This instrument has been designed, manufactured and tested
according to IEC 61010: Safety requirements for electrical equipment,
and delivered in the best condition after passing quality control tests.
This instruction manual contains warnings and safety rules which have
to be observed by the user to ensure safe operation of the instrument
and to maintain it in safe condition. Therefore, read through these
operating instructions before using the instrument.
WARNING
Read through and understand the instructions contained in this
manual before using the instrument.
Keep the manual at hand to enable quick reference whenever
necessary.
The instrument is to be used only in its intended applications.
Understand and follow all the safety instructions contained in the
manual.
It is essential that the above instructions are adhered to. Failure
to follow the above instructions may impair the protection
provided by the instrument and test leads, and may cause injury,
instrument damage and/or damage to equipment under test.
HOLD / Backlight Key
REL / MAX・MIN Key
Voltage Terminal
The symbol indicated on the instrument means that the user
must refer to the related parts in the manual for safe operation of the
instrument. It is essential to read the instructions wherever the
symbol appears in the manual.
DANGER is reserved for conditions and actions that
WARNING is reserved for conditions and actions that
CAUTION is reserved for conditions and actions that
Symbols listed below are used on this instrument.
User must refer to the manual.
Instrument with double or reinforced insulation.
AC DC Ground (Earth)
This instrument complies to WEEE Directive (2002/96/EC).
Please contact your local distributor at disposal.
are likely to cause serious or fatal injury.
can cause serious or fatal injury.
can cause injury or instrument damage.
Measurement Category
O Circuits which are not directly connected to the mains power supply.
CAT I I Primary electrical circuits of equipment connected to an AC
CAT III Primary electrical circuits of the equipment connected directly
CAT I V The circuit from the service drop to the service entrance, and
This instrument is designed for CAT IV 300V/ CAT III 600V / CAT II
1000V. Test leads M-7066A with the supplied caps are designed
for CAT IV600V/ CAT III 1000V and without the caps are for CAT II
1000V.
electrical outlet by a power cord.
to the distribution panel, and feeders from the distribution
panel to outlets.
to the power meter and primary over current protection
device(distribution panel).
DANGER
Never make measurements under the circumstances exceeding the
designed measurement category and the rated voltage of the
instrument.
Do not attempt to make measurement in the presence of flammable
gasses. Otherwise, the use of the instrument may cause sparking,
which can lead to an explosion.
Never attempt to use the instrument if its surface or your hand is wet.
Do not exceed the maximum allowable input of any measuring
range.
Never open the Battery compartment cover during a measurement.
To avoid electrical shock by touching the equipment under test or its
surroundings, be sure to wear insulated protective gear.
Test leads to be used for voltage measurements shall be rated as
appropriate for Measurement Category III or IV according to IEC
61010-031 and shall have a voltage rating of 1000V or higher.
Barriers on the test leads provide protection to keep your fingers and
hands from touching an object under test. Keep your fingers and
hands behind the barriers during measurement.
WARNING
Never attempt to make measurement if any abnormal conditions,
such as broken case and exposed metal parts are found on the
instrument or test leads.
Verify proper operation on a known source before use or take action
as a result of the indication of the instrument.
Firmly attach the caps to the test leads when performing
measurements in CAT III or higher test environments. When
KEW1020R and the test leads are combined and used together,
whichever is lower category & voltage to earth either of them
belong to is applied.
Do not rotate the Function Switch if the instrument and the
equipment under test are connected.
Do not install substitute parts or make any modification to the
instrument. For repair or re-calibration, return the instrument to your
local KYORITSU distributor.
Use of this instrument is limited to domestic, commercial and light
industry applications. Strong electromagnetic interference or strong
magnetic fields, generated by large currents, may cause malfunction
of the instrument.
Firmly insert the test leads.
Do not pull or twist the test leads to prevent the risk of damage.
Power off the instrument after use. Remove the battery if the
instrument is to be stored and will not be in use for a long period.
Do not expose the instrument to direct sunlight, high temperature
and humidity or dewfall.
Use a cloth dipped in water or neutral detergent for cleaning the
instrument. Do not use abrasives or solvents.
NOTE
The LCD shows some digits at the Voltage range even while the test
leads are open. And, it may show some digits instead of 0 even if the
test leads are shorted. However, these phenomena don’t affect
measurement results.
A resistance measurement takes time to settle the reading if there is
CAUTION
a high resistance or capacitance components.
2. Specification
Accuracy (Temperature: 23 ± 5°C, Humidity: 45 - 75%)
ACV / RMS (Auto Range)
Range Display Range Accuracy(sine wave)
6V 0.000, 0.006 - 6.299V
60V 5.70 - 62.99V
600V 57.0 - 629.9V
1000V 570 – 1049V ±1.3 %rdg±3dgt (40-500Hz)
Guaranteed accuracy : 0.01V-1000V, less than 1500V peak
Input impedance : approx. 10M
Input protective voltage : AC/DC1200V 10 sec.
Hz Frequency - ACV measurement (Auto Range)
Range Display Range
99.99Hz 10.00 - 99.99Hz
999.9Hz 95.0 - 999.9Hz
9.999kHz 0.950 - 9.999kHz
99.99kHz 9.50 - 99.99kHz
Guaranteed accuracy : 10Hz-99kHz
Input protective voltage : AC/DC1200V 10 sec.
% DUTY - ACV measurement
Range Display Range
99.9 % 0.0 – 99.9 % ±1.0%rdg±3dgt (50/60Hz)
Guaranteed accuracy : 10%-90%
Input protective voltage : AC/DC1200V 10 sec.
DCV (Auto Range)
Range Display Range Accurac
6.000V 0.000 - ±6.299V
60.00V ±5.70 - ±62.99V
600.0V ±57.0 - ±629.9V
1000V ±570 - ±1049V ±0.8 %rdg±3dgt
Guaranteed accuracy : 0V-±1000V
Input impedance : approx. 11M(6V range) / 10M(60/600/1000V range)
Input protective voltage : AC/DC1200V 10 sec
ACmV / RMS
Range Display Range Accuracy (sine wave)
600.0mV 0.0, 0.9 - 629.9mV ±2.0 %rdg±3dgt (40-500Hz)
Guaranteed accuracy : 1.2mV-600mV, less than 900mV peak
Input impedance : approx. 900k
Input protective voltage : AC/DC1000V 10 sec.
AC Clamp Sensor / RMS (Auto Range)
Range Display Range Accuracy (sine wave)
60.00
200.0
0.00, 0.09 - 62.99
57.0 - 209.9
+ Sensor accuracy (40-500Hz)
Direct reading from 10mV / A output Clamp sensor
Guaranteed accuracy : 0.12A-200A, less than 300A peak
Input impedance : approx. 900k
±1.0 %rdg±3dgt (40-500Hz)
Accuracy
(sine wave)
±0.1%rdg±3dgt
Accuracy
(Square wave)
±0.5 %rdg±3dgt
±2.0 %rdg±3dgt
DCmV
Range Display Range Accuracy
600.0mV 0.0 - ±629.9mV ±1.5 %rdg±3dgt
Guaranteed accuracy : 0mV-±600mV, Input impedance : approx. 900k
Input protective voltage : AC/DC1000V 10 sec.
DC Clamp Sensor (Auto Range)
Range Display Range Accuracy
60.00
200.0A ±57.0 - ±209.9A
0.00 - ±62.99
±1.5 %rdg±3dgt
+ Sensor accuracy
Direct reading from 10mV / A output Clamp sensor
Guaranteed accuracy : 0A-±200A, Input impedance : approx. 900k
Resistance (Auto Range)
Range Display Range Accuracy
600.0 0.0 - 629.9 ±0.5 %rdg±4dgt
6.000k 0.570 - 6.299k
60.00k 5.70 - 62.99k
600.0k 57.0 - 629.9k
6.000M 0.570 - 6.299M
40.00M 5.70 - 41.99M ±1.5 %rdg±3dgt
Guaranteed accuracy : 0-40M, Open-loop Voltage : less than 3V
Measurement current : less than 1mA
Input protective voltage : AC/DC1000V 10 sec.
±0.5 %rdg±2dgt
Continuity
Range Display Range Accuracy
600.0 0.0 - 629.9
Open-loop Voltage : less than 3V, Measurement current : less than 1mA
Input protective voltage : AC/DC1000V 10 sec.
Buzzer threshold value :
less than 90
Diode
Range Display Range Accuracy
2.000V 0.000 - 2.099V ±5 %rdg±5dgt
Guaranteed accuracy : 0V-2V, Open-loop Voltage : less than 3V
Measurement current : approx.0.5mA (Vf=0.6V)
Input protective voltage : AC/DC1000V 10 sec.
Capacitance (Auto Range)
Range Display Range Accuracy
60.00nF 0.00 - 62.99nF
600.0nF 57.0 - 629.9nF
6.000µF 0.570 - 6.299µF
60.00µF 5.70 - 62.99µF
600.0µF 57.0 - 629.9µF
1000µF 570 - 1049µF
* Accuracy after canceling floating capacitance using REL function.
Guaranteed accuracy : 0nF-1000µF
Input protective voltage : AC/DC1000V 10 sec.
Measuring method : ⊿ modulation
Over-range indication : OL
Measurement cycle : 2.5 times per second
Crest factor : less than 3 (45-65Hz)
Applicable standards :
Withstand voltage : AC5160Vrms 5sec between circuit and enclosure
IP rating : IP40 (IEC60529)
Insulation resistance : 100M or more /1000V between enclosure
Operating temperature and humidity range :
Storage Temperature and humidity range :
Power source : DC3V R03/LR03 (AAA) × 2
Current consumption : 3mA or less
Battery life : Approx. 200 hours (ACV, continuous, no load, with R03)
Dimension, Weight : 155(L)×75(W)×40(D)mm,
Accessories : Test leads (M-7066A), Instruction manual,
Option : Magnet hanger strap (M-9189)
(1000µF range of Capacitance function 0.05 times per second)
For non-sinusoidal waveforms, add ±0.5 %rdg±5dgt
(Applicable functions : ACV, ACmV, AC clamp sensor)
IEC 61010-1 /61010-2-033 :
CAT IV 300V / CAT III 600V / CAT II 1000V
Pollution degree 2, Indoor use, Altitude up to 2000m
IEC61010-031 (Test leads Model 7066A)
IEC 61326 (EMC) , EN 50581 (RoHS)
and electrical circuit
0 to 40°C, 80%RH or less (no condensation)
-20 to 60°C, 80%RH or less (no condensation)
approx. 250g (including batteries and Wing-type holder)
Battery R03 (AAA) 2pcs, Wing-type holder
Test leads with alligator clips (M-7234)
AC Clamp Sensor (KEW8161), AC/DC Clamp Sensor (KEW8115)
±2.0 %rdg±5dgt*
±5.0 %rdg±5dgt
3. ACV / DCV / ACmV / DCmV Measurement
DANGER
● Before starting a measurement, always check and confirm
the Function switch is in the appropriate measurement
position.
● Never make measurement on a circuit in which voltage over
1000V exists.
● Keep your fingers and hands behind the barrier during
measurement.
3.1 ACV / DCV/ ACmV / DCmV measurement
(1) Set the Function switch to ACV, DCV, ACmV or DCmV position.
For frequency or DUTY measurement, set the switch to ACV
and press the SELECT key.
(2) Connect the test leads to the Voltage and COM terminals.
ACV / ACmV DCV / DCmV
Press the SELECT key to toggle functions
3.2 Clamp sensor (option) measurement
(1) Set the Function switch to ACmV or DCmV position and press the
SELECT key. LCD shows “SENSOR”.
(2) Connect the clamp sensor to the Voltage and COM terminals.
AC clamp sensor
Default :ACmV
Default :DCmV
Press the SELECT key to toggle functions
NOTE
If the connection is reversed, the “ - “ mark will be displayed on the
LCD. (DCV measurement).
Press the REL key to adjust the reading on the DC clamp sensor to
“0”.
Find Quality Products Online at: sales@GlobalTestSupply.com
The sensor mode corresponds to direct reading with 10mV/A
output clamp sensor.
Default : ACV
Black
DC clamp sensor
Black
Frequency DUTY
Red
Sensor mode
(ACA)
Sensor mode
(DCA)
Red
4. Resistance / Diode / Continuity / Capacitance
Measurement
WARNING
Never use the instrument on an energized circuit. Discharge the
capacitor before starting a capacitance measurement.
(1) Set the Function switch to the resistance, continuity or capacitance
position. For diode measurement, press the SELECT key.
Resistance
(2) Connect the test leads to the Voltage and COM terminals.
Resistance
or Capacitance
NOTE
LCD shows ”OL” when the test leads are open.
(except for capacitance measurement)
The LCD shows “OL” if the test lead connection is reversed for diode
measurement.
Measurement time on 600F/1000F range is a bit long.
(20sec max.)
5. Other Functions
REL Function
Press the REL key to enable this function and store the measured
value to display the differences between the stored value and the
values measured in further tests. The measurement range will be
fixed when the REL function is enabled, and the measuring range will
be between the initial value and the full scale value.
Press the REL Key again to release the stored value.
“” appears and “AUTO” disappears
when REL Key is pressed.
*To activate REL function, MAX/MIN function should be disabled.
Sleep Function
Automatically powers off the instrument in about 15 min after the last
switch operation.
To exit from the Sleep mode, rotate the Function switch or press any
key. To disable the Sleep function, press the HOLD/Backlight key
and power on the instrument.
Confirm that the LCD shows “P.OFF” about 1 sec.
One press: REL ON
Another press: REL OFF
Continuity / Diode
Default : Continuity
Press the SELECT key to toggle functions
Continuity
Capacitance
Diode
Diode
Black Red
www.GlobalTestSupply.com
Max / Min value display function
This function is to display the measured max and min values on the
LCD during a measurement. Press the MAX/MIN key 1 sec or longer
to start recoding of max and min values. Then the LCD shows the
latest max value. After that, the min and present values can be
toggled and checked by pressing the MAX/MIN key (less than 1
sec).
To disable this function, press the MAX/MIN key 1 sec or longer.
Max value
* To activate MAX/MIN function, REL function should be disabled.
Data hold function
Press the HOLD key (less than 1 sec). The LCD shows “H” mark and
the reading will be held (Data hold mode). The max/ min values are
not updated in MAX/MIN mode. Press the HOLD key
again (less than 1 sec) to release the display.
Backlight function
Press the Backlight key 1 sec or longer to turn
on the backlight. Press the Backlight key another 1 sec or longer to
turn it off. The light automatically turns off in 1 min.
Low battery indication
The LCD shows “ ” mark when the batteries fall below the
normal operating voltage.
How to store test leads
Attach the Wing-type holder to the back of the instrument to store the
test leads.
1. Attach the Wing-type holder to the back of the instrument.
2. Fit the barrier on the test lead into the groove between the
3. Then it will be easier to see the displayed readings during a
Other convenient functions
Less than 1 se
Replace the batteries
with new ones when this
mark appears.
instrument and the Wing-type holder.
Wing-type
holder
1.
measurement.
Tilt stand
Min value
Less than 1 se
2.
Present value
Less than 1 se
3.
Magnet hanger strap
(option)
6. Battery Replacement
WARNING
Replace the battery when the " " mark- low battery voltage
warning- appears on the LCD.
Otherwise, precise measurement cannot be made. If the battery
is completely exhausted, the LCD goes blank without showing
the " " mark.
Disconnect the test leads from the object under test and power
off the instrument before opening the Battery compartment
cover for battery replacement.
Do not try to replace the battery if the surface of the instrument
is wet.
CAUTION
Do not mix old and new batteries. Brand and type of the
batteries to be used should be harmonized.
Install a batteries in correct polarity as indicated in the
Battery Compartment.
(1) Set the Function Switch to "OFF" position.
(2) Untighten the screw on the back of the instrument.
(3) Remove the Battery compartment cover and replace the
batterie.
(4) Attach the cover to the instrument and then secure the cover
by tightening the screw.
Battery
Kyoritsu reserves the rights to change specifications or designs
described in this manual without notice and without obligations.
Screw