Kyoritsu Electrical Instruments 1019R User manual

INSTRUCTION MANUAL
LCD
Push hole
Slit
Be sure that the test
CARD TYPE AUTO RANGE
DIGITAL MULTIMETER
KEW1019R
Function Switch
Hard Case
Cap
Hard Case
Hard Case Cover
Test Leads
Barrier
Cord Holder
Cap Holder
RELΔ Key
Cap
Features
True-RMS type
1. Safety Warnings
This instrument has been designed, manufactured and tested according to IEC 61010: Safety requirements for Electronic Measuring apparatus, and delivered in the best condition after passing quality control tests. This instruction manual contains warnings and safety rules which have to be observed by the user to ensure safe operation of the instrument and to maintain it in safe condition. Therefore, read through these operating instructions before using the instrument.
WARNING
Read through and understand the instructions contained in this manual before using the instrument.
Keep the manual at hand to enable quick reference whenever necessary.
The instrument is to be used only in its intended
applications.
Understand and follow all the safety instructions
contained in the manual.
It is essential that the above instructions are adhered to.
Failure to follow the above instructions may impair the protection provided by the instrument and test leads, and may cause injury, instrument damage and/or damage to equipment under test.
The symbol indicated on the instrument means that the user must refer to the related parts in the manual for safe operation of the instrument. It is essential to read the instructions wherever the symbol appears in the manual.
DANGER is reserved for conditions and actions that
are likely to cause serious or fatal injury.
WARNING is reserved for conditions and actions that
can cause serious or fatal injury.
CAUTION is reserved for conditions and actions that
can cause injury or instrument damage.
Symbols listed below are used on this instrument.
User must refer to the manual.
Instrument with double or reinforced insulation.
AC DC Ground (Earth)
This instrument complies to WEEE Directive
(2002/96/EC). Please contact your local distributor at disposal.
Measurement Category
O
Circuits which are not directly connected to the mains
power supply.
CAT II
Primary electrical circuits of equipment connected
to an AC electrical outlet by a power cord.
CAT III
connected directly to the distribution panel, and feeders from the distribution panel to outlets.
CAT IV The circuit from the service drop to the service
entrance, and to the power meter and primary over current protection device(distribution panel).
This instrument is basically designed for CAT II 600V, but the cap for CAT III 300V is supplied with.
Primary electrical circuits of the equipment
DANGER
Never make measurements under the circumstances exceeding the designed measurement category and the rated voltage of the instrument.
Do not attempt to make measurement in the presence of flammable gasses. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion.
Never attempt to use the instrument if its surface or your hand is wet.
Do not exceed the maximum allowable input of any measuring range.
Never open the case during a measurement.
To avoid electrical shock by touching the equipment under
test or its surroundings, be sure to wear insulated protective gear.
Barriers on the test leads provide protection to keep your fingers and hands from touching an object under test. Keep your fingers and hands behind the barriers during measurement.
WARNING
Never attempt to make measurement if any abnormal conditions, such as broken case and exposed metal parts are found on the instrument or test leads.
Verify proper operation on a known source before use or take action as a result of the indication of the instrument.
Firmly attach the caps to the test leads when performing measurements in CAT III test environments.
Do not rotate the function switch if the instrument
and the equipment under test are connected.
Do not install substitute parts or make any modification to the instrument. For repair or re-calibration, return the instrument to your local KYORITSU distributor.
CAUTION
Use of this instrument is limited to domestic, commercial and light industry applications. Strong electromagnetic interference or strong magnetic fields, generated by large currents, may cause malfunction of the instrument.
Set the function switch to an appropriate position before starting measurement.
This instrument isnt dust & water proofed. Keep away from dust and water.
Do not pull or twist the test leads to prevent the risk of damage.
Power off the instrument after use. Remove the battery if the instrument is to be stored and will not be in use for a long period.
Do not expose the instrument to direct sunlight, high temperature and humidity or dewfall.
Use a cloth dipped in water or neutral detergent for cleaning the instrument. Do not use abrasives or solvents.
2. Specification
Temperature: 23 ± 5°C, Humidity: 45 - 75% ACV
Range Display Range 6V 0.000,
0.006-6.299V
60V 5.70-62.99V ±1.6 %rdg±5dgt (50/60Hz) 600V 57.0-629.9V
Guaranteed accuracy : 0.010V-600.0V CF≦ 3 (50/60Hz), less than 900V peak For non-sinusoidal waveforms, add ±0.5 %rdg±5dgt
DCV
Range Display Range Accuracy 600mV 0.0-±629.9mV 6V ±0.570-±6.299V 60V ±5.70-±62.99V 600V ±57.0-±629.9V ±1.0%rdg±5dgt
Guaranteed accuracy : 0.0mV-±600.0V ACV/DCV input impedance : approx. 10M
Resistance / Continuity (Auto Range)
Range Display Range Accuracy 600 0.0-629.9 6k 0.570-6.299 k 60k 5.70-62.99 k 600k 57.0-629.9 k 6M 0.570-6.299 M 40M 5.70-41.99 M ±2.5%rdg±5dgt
Continuity 0.0-629.9
Guaranteed accuracy : 0.0 -40.00M Open-loop voltage :less than 3V Input protective voltage : AC/DC600V 10 sec
Capacitance (Auto Range) Range 6nF 0.000-6.299nF ±3.5%rdg±50dgt 60nF 5.70-62.99nF ±3.5%rdg±10dgt 600nF 57.0-629.9nF 6µF 0.570-6.299µF 60µF 5.70-62.99µF 600µF 57.0-629.9µF ±4.5%rdg±5dgt Guaranteed accuracy : 0.000nF-600.0µF Input protective voltage : AC/DC600V 10 sec
Measuring method : Σ method
Over-range indication : OL
Measurement cycle : 2.5 times per second
Applicable standards :
IEC 61010-1/ 61010-031/ 61010-2-033
In the radio-frequency electromagnetic field of 3V/m, EN 50581 (RoHS)
Withstand voltage :
Insulation resistance : 100M or more /1000V between enclosure and
Operating temperature and humidity range :
Storage Temperature and humidity range :
Display Range Accuracy
Pollution degree 2, Indoor use, Altitude up to 2000m IEC 61326 (EMC)
accuracy is within five times the rated accuracy.
AC3470Vrms 5sec
electrical circuit
0 to 40°C, 80%RH or less (no condensation)
-20 to 60°C, 80%RH or less (no condensation)
Accuracy (sine wave) ±1.3 %rdg±5dgt (50/60Hz) ±1.7 %rdg±5dgt (45-500Hz)
±2.0 %rdg±5dgt (45-500Hz)
Buzzer threshold value 60 or less.
(600µF range of Capacitance
function 0.2 times per second)
CAT III 300V / CAT II 600V
between circuit and enclosure
(Auto Range)
(Auto Range)
±0.8%rdg±5dgt
±1.0%rdg±5dgt
±3.5%rdg±5dgt
Power source : DC3V CR2032×1
Current consumption : 2mA or less
Battery life (ACV, continuous, no load, with CR2032) :
Approx. 120 hours
Dimension, Weight : 126(L)×85(W)×18(D)mm,
a
pprox. 135g (including battery and hard case)
Accessories : Battery CR2032 1pce Instruction manual 1pce Hard case (M-9188) 1pce
3. ACV/DCV Measurement
DANGER
Never make measurement on a circuit in which
voltage over 600V exists.
Keep your fingers and hands behind the barrier
during measurement.
ACV
DCV
Black(-) Red(+)
NOTE
Beep sounds when the instrument switches the ranges from 60V to 600V.
If the connection is reversed, the mark will be displayed on the LCD. (DCV measurement).
The LCD shows some digits at the ACV or the DCV range even while the test leads are open. And, it may show some digits instead of 0 even if the test leads are shorted. However, these phenomena dont affect measurement results.
4.Resistance(Continuity)Measurement
WARNING
Never use the instrument on an energized circuit.
Resistance Continuity
Find Quality Products Online at: sales@GlobalTestSupply.com
appears on the LCD.
Beep 60 or less.
NOTE
LCD shows OL when the test leads are open.
Even if the test leads are shorted, the indicated values
may not be0. But this is because of the resistance of test leads and not a failure. High resistance measurement and capacitive components may fluctuate readings.
5. Capacitance Measurement
WARNING
Never use the instrument on an energized circuit. Discharge the capacitor before measurement.
Press the RELΔ key before starting a measurement and adjust the displayed value to 0.000nF.
Capacitance
NOTE
Measurement time at 600µF range is a bit long since the reading is updated once in approx 5 sec.
6. Other Functions
REL Function Press the RELΔ key to enable this function and store the measured value to display the differences between the stored value and the values measured in further tests. (at any functions other than Continuity) The measure­ment range will be fixed when the REL function is enabled, and the measuring range will be between the initial value and the full scale value. (except for Capacitance) Press the RELΔ Key again to release the stored value.
Δ symbol appears
when RELΔ Key is pressed.
Low battery indication
symbol appears
at 2.3±0.1V or less.
Sleep Function Automatically powers off the instrument in about 15 min after the last switch operation. Buzzer beeps five times one minute before entering into the Sleep mode, and also one time just before entering into the mode.
www.GlobalTestSupply.com
Push RELΔ key. ” Δ “ appears on the LCD.
This instrument can make measurements while one test
lead is left in place. So you can perform a test with checking the readings.
7. Battery Replacement
WARNING
● Replace the battery when the " " mark- low battery voltage warning- appears on the LCD. Otherwise, precise measurement cannot be made. If the battery is completely exhausted, the LCD goes blank without showing the " " mark.
Do not try to replace the battery if the surface of the
instrument is wet.
Power off the instrument before opening the case for
battery replacement.
CAUTION
Install a battery in correct polarity as indicated in the
Battery Compartment. (1) Set the Function Switch to "OFF" position.
(2) Remove the Hard case.
1. Remove the test leads from the holder
2. Open and hold the Hard case cover about 90-degree, and then push the instrument through the hole on the back side of the Hard case.
1. 2.
(3) Loosen the screw on the back side of the instrument
and remove the case.
(4) Replace the battery with a new one (CR2032)
observing the correct polarity.
.
about 90-degree
leads should be in the guide slot well.
(5) Be sure that the test leads should be in the guide slot
well, then, install the case and tighten the screw.
(6) Attach the Hard case.
2.
1. Pass the test leads through the slit first, and put the main unit in the Hard case.
Slide and attach the Hard case cover.
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or designs described in this manual without notice and without obligations.
12-14 92-2188B