Voltage setting for insulation resistance measurement
Page 4
1. Safe testing
Electricity is dangerous and can cause injury and death. Always
treat it with the greatest of respect and care. If you are not quite sure
how to proceed, then stop, take advice from a qualified person.
This instruction manual contains warning and safety rules which
must be observed by the user to ensure safe operation of the Tester
and retain it in safe condition. Therefore, read through these
operating instructions before using the Tester.
IMPORTANT:
The Tester must only be used by a competent and trained person
and operated in strict accordance with these instructions.
KYORITSU will not accept liability for any damage or injury caused
by misuse or non-compliance with the instructions or with the safety
procedures.
It is essential to read and to understand the safety rules contained
in these instructions and with the safety procedures.
The symbol # indicated on the Tester means that the user must
refer to the related sections in the manual for safe operation of the
Test e r.
Be sure to carefully read instructions following each symbol # in
this manual.
#DANGER : is reserved for conditions and actions that are likely
to cause serious or fatal injury.
#WARNING: is reserved for conditions and actions that can
cause serious or fatal injury.
# CAUTION : is reserved for conditions and actions that can
cause injury or instrument damage.
#DANGER
● The Tester can be connected only to the commercial power of
100V – 253V, 50Hz.
● For safety reasons, only use the Test Leads designed to be
used with this Tester and recommended by KYORITSU.
● Use only grounded mains outlets to supply the Tester.
● Do not touch the device under test while testing is in progress.
3
Page 5
# DANGER
●
Since a high voltage 250V or 500V is being output continuously,
when measuring insulation resistance, the user may get an
electrical shock. Any capacitors in the appliance under test may
become charged during testing and may contain hazardous
voltages do not touch them.
● When testing, always be sure to keep your fingers behind the
safety protective finger guard on the test leads.
●
Disconnect the Tester from the power supply when measurement
is finished.
● Do not leave the Tester with connected to the power supply.
# WARNING
●
Never open the instrument case – because dangerous voltages
are present. Only fully trained and competent electrical engineers
should open the case.
● If abnormal conditions of any sort are noted (such as a faulty
display, unexpected readings, broken case, cracked test leads,
etc.) do not use the Tester and return it to your distributor for
inspection and repair.
● Never attempt to use the Tester if the Tester or your hands are
wet.
● Stop using the test lead if the outer jacket is damaged and the
inner white jacket is exposed.
#CAUTION
● When using Test Leads with the alligator clip, be sure to check
the alligator clip is firmly connected to the metal part of the
device under test. Otherwise, inaccurate measurements or
arcing at the contacts may occur.
● The rated measuring voltage for the insulation test is 250V or
500V DC. If this voltage seems too high for the appliance
under test contact the appliance manufacturer for advice.
● When testing a faulty appliance, it may trip the circuit breaker
of main power supply during test and may cause interruption of
service. Be careful when the same main power supply is used
for PCs.
● We are not liable for loss of data on PC during testing with the
Tester. The appliance under test is powered on during most
tests, but please turn it to the OFF position after testing.
● Use a very slightly damp cloth for cleaning the Tester. Do not
use abrasives or solvents.
4
Page 6
Measurement Category
To ensure safe operation of measuring instruments, IEC 61010
establishes safety standards for various electrical environments,
categorized as O to CAT IV, and called measurement categories.
Higher-numbered categories correspond to electrical environments
with greater momentary energy, so a measuring instrument
designed for CAT III environments can endure greater momentary
energy than one designed for CAT II.
O : Circuits which are not directly connected to the mains
power supply.
CAT II : Electrical circuits of equipment connected to an AC
electrical outlet by a power cord.
CAT III : Primary electrical circuits of the equipment connected
directly to the distribution panel, and feeders from the
distribution panel to outlets.
CAT IV : The circuit from the service drop to the service entrance,
and to the power meter and primary over-current
protection device (distribution panel).
O: Device which is
not directly
connected to the
mains power supply
5
Page 7
2. Product summary and explanation
2.1 Product summary
KEW6205 (Tester) is a hand-held por table appliance tester and can
test electrical safety of Class I and Class II appliances. The Tester
performs test and indicates PASS/ FAIL result complying with the
criteria of judgement defined in AS/NZS3760.
As a guide, AS/NZS3760 defines these two categories as follows:
Class I : Appliances which have a functional insulation throughout
and an earth connected case. These are often described
as earthed appliances.
Class II : Appliances which have both functional and additional
insulation where any metal parts cannot become
Liveunder fault conditions.
2.2 Test Function
●CL I:
・
Protective conductor resistance test (Test current 200mA DC nominal)
・Insulation test (250V or 500V)
・Run leakage current test
・Load current test
●CL II:
・Insulation test (250V or 500V)
・Run leakage current test
・Load current test
●Extension Lead test
・
Protective conductor resistance test (Test current 200mA DC nominal)
・Insulation test (250V or 500V)
・L-N Short test
・Polarity test
・Run leakage current test
●RCD test
2.3 Features
●
Check for whether the appliance is switched ON.
●
Selection for 250V or 500V on the insulation resistance test.
●
Null function for the protective conductor resistance test.
●
Auto-testing & PASS/ FAIL indication with backlight
●
Auto-power-of f
6
Page 8
2.4 Tester layout
Connector
Display(LCD)
Switches
Connector
2.4.1 Function Switches
SwitchDetails
Short press: Starts CLI test.
CLI
switch
CLII
switch
Extension
switch
NULL
switch
(Continuity test: 1 sec.)
Long press: Starts CLI test.
(Continuity test: 5 sec.)
Starts CLII test.
Short press: Starts Extension lead test.
(Continuity test: 1 sec.)
Long press: Starts Extension lead test.
(Continuity test: 5 sec.)
Performs NULL.
7
Page 9
SwitchDetails
Short press:
Power
switch
Long press: Turns on/ off the Tester.
250V
/500V
Light
switch
Switches output voltage for insulation
resistance test between 250 V and 500
V. (Default setting: 250 V)
Short press: Turns on/ off LCD backlight.
Long press: Enters into clock mode.
Switches to RCD stand-by screen.
RCD
switch
Another press on the stand-by screen
starts RCD test. Long press on the
screen switches test current between 10
mA and 30 mA. (Default setting: 30 mA)
Memory
switch
Saves test results.
Works as ESC switc h. Returns
to the stand-by screen when
a test is halte d byFAILresult.
Recall
switch
Print
switch
Recalls the saved data.
Transfers the displayed test result to a
pr inter.
8
Page 10
2.4.2 LCD indications
<All symbols to be displayed on the LCD>
Display symbols
NULL value has been set.
Rated voltage for insulation resistance
test
Leakage current test under actual
working condition
Rated current for RCD test
Measurement in progress
Protective conductor resistance
Insulation resistance
Insulation resistance between L-N
(Extension Lead Test)
Leakage current
Load current
9
Page 11
Mains voltage
Polarity test result
Indicates items failed the test.
RCD test range
Prompts to check and turn on a DUT.
Indicates the connection for RCD test
is incorrect.
Indicates the resistance between L-L
(N-N) is less than 10 Ω and polarity is
correct.
Open circuit warning –resistance
between L-L (N-N) is 10 Ω or higher at
polarity test.
Crossed L-N polarity warning resistance between L-N (N-L) is 10 Ω
or lower at polarity test.
Short-circuited L-N warning –
resistance between L-N is 100 kΩ or
lower at insulation test. (Extension
Le a d Test )
Fault voltage of 50 V or higher is
detected at RCD test.
Test result is PASS.
Test result is FAIL.
RCD tripped at fault current detection.
Warning – voltage of 30 V or higher is
being applied to the IEC connector
while a function other than RCD is
selected.
Blinks while saving data.
Data no. given for the saved data.
10
Page 12
2.4.3 Connector
(1)
(5)
(6)
(4)
(3)
(2)
(1) Test s o cke tInsert the mains plug of DUT to this
socket for the polarity test of protective
conductor resistance, insulation
resistance and leakage current test.
(2) Terminal for
mains lead
(3) Rpe Test Lead
terminal
This terminal is connected to a mains
supply via MODEL7277.
Connect the test lead MODEL7129A to
this terminal for the measurement of
protective conductor resistance, and clip
the metal parts of DUT with the alligator
clip.
(4) Terminal for
Extension leads
adaptor
It corresponds to L, N, E of test socket,
and the extension leads adaptor
connected with the cord reel to be
plugged to it.
(5) Terminal for USB
cable
Connect the optional USB cable to this
terminal to transfer the internal data to a
PC.
(6) Terminal for
printer cable
This terminal is connected to a printer
with the optional printer cable
MODEL7275.
11
Page 13
2.5 Accessories
(1) Mains lead MODEL7277
This mains lead can be connected to the mains supply so that the
Tester supply power to the DUT.
(2) Test Lead with safety alligator clip (MODEL7129A) and Probe
with Blade type Prod(MODEL7161A). The probe and alligator
clip are interchangeable.
Please use it according to a measurement use.
Alligator Clip
Probe with Blade Type Prod
*1
Protective finger guard is a part providing protection against
Protective
finger guard*
Protective
finger guard*
1
1
electrical shock and ensuring the minimum required air and
creepage distances.
(3) Extension leads adaptor MODEL7276
This is for connecting the Tester and an extension lead (cord
reel).
IEC Connector
(4) Soft case MODEL9193
(5) Six size AA alkaline batteries(LR6)
(6) Instruction manual
(7) Shoulder strap with Buckle
12
Page 14
2.6 Optional accessories
(1) Printer cable MODEL7275
(2) USB cable with "KEW Report(Software)" MODEL8263-USB
13
Page 15
3. Specification
3.1 General specification, measuring range and accuracy
Mains voltage indication:
Display range30 - 270 V
Hi/ Lo indication< 30 V: No voltage indication
> 270 V: >270 V
Resolution1 V
Accuracy±5 V
* Voltage between L-N of mains lead terminal is measured and
displayed at voltage test.
Protective conductor resistance test
Measuring range
Over-range indication
Resolution0.01Ω
Open-circuit voltage 5 V ±0.4 V DC
Measuring current200 mA DC (nominal value)
Accuracy±3%rdg ±5dgt
* Resistances exceeding 3 Ω cannot be canceled by NULL function.
Insulation resistance test
Rated voltage 250V500V
Measuring
range
Over-range
indication
Resolution0.01 MΩ
Rated voltage 250V DC +20%, -10%
Short-circuit
current
Rated current
Accuracy±2%rdg±3dgt
0.00 - 20.00 Ω (NULL value included.) *
> 21.00 Ω
0.00 - 20.00 MΩ
> 21.00 MΩ
500V DC+20%, -10%
@ 1MΩ
1.5 mA DC or less
1 -1.2 mA @ 0.25 MΩ
@ 1MΩ
1 -1.2 mA @0.5 MΩ
14
Page 16
Leakage current test under actual working condition:
(Load current is also measured.)
ItemLoad currentLeakage current
Mains voltage
100 - 253 V/ 50 Hz
range
Measuring range 0.10 - 10.00 A rms0.10 - 20.00 mA rms
Display range0.00 - 12.00 A0.00 - 21.00 mA
Over-range
> 12.00 A> 21.00 mA
indication
Resolution0.01A0.01mA
Accuracy±10%rdg ±5dgt±3%rdg ±5dgt
Max. rating2.5 kVA max./
---
15 sec. continuous
Insulation test between L-N at Extension Lead Test
Open-circuit voltage 4.5 V ±0.5 V DC
Criteria of judgement 100 kΩ ±20 kΩ or less: LCD shows
<0 .1 0 MΩand the measurement is
halted.
RCD test:
Four tests with the pre-set rated current are performed in the
following sequence: ×1(0°)→×1(180 °) →×5(0°)→×5(180°).
Rated voltage230 V -15% +10% (50Hz)
Rated current10 mA/ 30 mA
Function×1×5
Test duration0.0 ms - 500.0 ms 0.0 ms - 40.0 ms
EnergizationFS ±3%
Test current accuracy ±3%
Operating time
accuracy
±2 ms (≦40 ms)
±8 ms (>40 ms)
15
Page 17
3.2 General specification
Reference conditionsSpecifications are based on the
following conditions, except
where otherwise stated: -
(1) Ambient temperature: 23±5˚C
(2) Relative humidity: 45 - 75%
(3) Attitude: Horizontal
(4) AC power supply: 230 V, 50Hz
(5) Altitude: 2000m or less
Battery typeSix size AA alkaline batteries
(LR6)
Operating temperature and
humidity range
Storage temperature and
humidity range
0ºC to +40ºC, relative humidity:
85% or less (no condensation)
-20ºC to +60ºC, relative humidity:
85% or less (no condensation)
Rate voltage and frequency Rated voltage: 230 V -15% +10%
Possible number of measurements where battery voltage
Approx. 3000 times
(CLI Rpe: 0 Ω, Rins: 1.1 MΩ)
is within the effective range
(measuring ever y 30 sec.)
Applicable standards:
Instrument operationAS/NZS3760:2010
SafetyIEC/EN61010-1 CAT II 300V-instrument
IEC/ EN61010-2-030
IEC/ EN61010- 031
MODEL7129A(CAT III 600V) *
EMC EN61326-1, 2-2
Environmental standard EN50581
* When the instrument and the test lead are combined and used
together, whichever lower category either of them belongs to will
be applied.
16
Page 18
Symbols used on the Tester:
Electrical circuits of equipment connected to an AC
CAT II
electrical outlet by a power cord.
Double or reinforced insulation
User must refer to the explanations in the instruction
#
manual.
Earth
ATT(Anti-Trip Technology)
The Tester uses mains voltage across L-N for PRCD
testing.It enables PRCD testing without tripping the RCD
of mains power.
Crossed-out wheel bin symbol (according to WEEE
Directive: 2002/96/EC) indicating that this electrical
product may not be treated as household waste, but that it
must be collected and treated separately.
3.3 Threshold and display
Protective
Function
Class l
Class ll
Extension
Lead
RCD
RC D Ty p eType l (10mA)Type ll (30mA)
Function×1×5×1×5
conductor
resistance
RPE ≦ 1ΩRINS ≧ 1MΩ ILEAK ≦ 5mA
RPE ≦ 1ΩRINS ≧ 1MΩ ILEAK ≦ 1mA
40ms300ms40ms
Insulation
resistance
RINS ≧ 1MΩ ILEAK ≦ 1mA
Leakage
current
17
Page 19
4. Preparation before a measurement
4.1 Visual inspection
Before starting a measurement, the user should undertake visual
checks on the mains lead, case and that the correct type and rated
fuse is fitted to the appliance under test. There should also be no
evidence of damage of a nature that may impair the electrical safety
of the appliance.
4.2 Batter y Voltage Check
(1) Insert batteries with reference to 11.Batter y/ Fuse replacement
in this manual.
(2) Hold down Power switch 1 sec. to turn on the Tester.
(3) Check the battery indicator displayed in the upper left corner of
the LCD.
If just the last one segment of the indicator remains, the battery
voltage is extremely low. Refer to 11. Battery/ Fuse replacement
and replace batteries to perform further measurements.
If an empty battery indicator is displayed, the battery voltage is
lower than the working voltage limit. Replace batteries immediately
to keep accurate measurement.
Battery recommendation:
Use of alkaline battery is recommended. Use of other type of
batteries may cause improper indication of battery level.
4.3 Setting
4.3.1 Null setting
AS/NZS3760 pass level for Protective conductor resistance is 1 Ω,
which is a low value. So even the resistance of Test Leads will affect
the measurement result.
The Tester can cancel the resistance of test lead by pressing
The procedure of Null setting is shown below. The Null function is
held in memory even when the Tester is turned off, so theres no
need to Null the lead resistance every time. However, when replacing
fuses or test leads, it is recommended to do a Null setting again.
NULL
.
Note:
Null setting is possible at both Class l Test and Extension Lead
Test. However, only one Null value can be held in memory. For
example, when the Null setting is carried out at Class l Test, the set
value will also be used for Extension Lead Test (unless it is reset).
(1) Insert Test Lead with probe (MODEL7161A) into the Rpe Test
Lead terminal of the Tester, and contact the tip of the Test Lead
with the earth contacts of the socket on the Tester.
(2) Press NULL switch on any screen.
18
Page 20
(3) The screen switches to NULL screen, and NULL measurement
starts. The LCD shows blinking MEAS mark and measured
value during a measurement. The measured value is saved and
subtracted from the further measured values.
(4) If the measured value is less than 3 Ω, previous NULL value is
cleared and a new value is saved. When 3 Ω or higher is
measured, previous NULL value is just cleared (NULL mark
disappears).
4.3.2 Voltage setting for insulation resistance measurement
(How to change between 250V and 500V)
Two different rated voltages, 250 V and 500 V, are available for
insulation resistance test. Press 250V/500V switch on the stand-by
screen or while measured result is being displayed at CL I/ CL II/
Extension lead test. Default setting is 250 V.
Sets to 500 V.
19
Sets to 250 V.
Page 21
5. Measuring method
5.1 Class I Test
The purpose of the test carried out for Class I appliances is to check
the resistance of ear th continuity from exposed metal parts and the
plug is below a certain level and the insulation resistance between
live and neutral connected together and earth is 1MΩ or more.
To perform Class I test, connect the mains plug of the appliance to
the test socket (1) described in clause 2.4.3. Connector and the Rpe
Test Lead terminal (3). To perform RUN leak test, connect Mains
lead MODEL7277 to Mains lead terminal (2).
Connect to exposed metal
parts but not rotating parts or
heating elements.
Power ON
the DUT.
RUN leak test:
Connect Tester to outlet
Press CLI switch for Class I test.
1. Continuity test
Earth continuity test between the
exposed metal parts and plug
2. Insulation testBetween LINE, NEUTRAL and
EARTH.
3. Leakage
current test
RUN Leak test:
Connect Tester to outlet:
(Load current is also measured.)
20
Page 22
CL I Test Flow
(1sec .)
Press to halt a
measurement in
progress.
Tester performs next
test when CL I switch
is pressed while a test
is halted with “FAIL”
indication.
Stand-by
CLⅠ
Voltage value is displayed
while the Tester is connected
to an outlet.
Press CLI to start a test.
Short press: 1 sec of continuity test
Long press: 5 sec of continuity test
Continuity test
Rpe<1Ω =PASS Rpe>1Ω = FAIL
(Red backlight blinks.)
2. Insulation resistance test 1. Continuity test
DUT ON Check
DUT is ON.
DUT is OFF.
Insulation resistance test
21
DUT is powered on, or
CL I is pressed.
Page 23
Rins≧1MΩ = PASS Rins<1MΩ = FAIL
(Red backlight blinks.)
Check screen for RUN leakage test.
Buzzer sounds and blue backlight blinks.
3. Leakage current test
CLⅠ
RUN Leak/ Load current test
ILEAK≦5mA = PASS
(Green backlight blinks.)
22
ILEAK>5mA= FAIL
(Red backlight blinks.)
Page 24
# WARNING
● When conducting a leakage test the appliance will automatically
power on and will operate in its normal manner. It is imperative
that the appliance is secured safely before the test is carried
out. Extra care needs to be taken with appliances which have
heating elements and rotating parts.
● Firmly insert the plug of the appliance to the socket of the
Tester. Plugs may get hot if Leakage current test is performed
with improper connection.
● Do not connect/remove the plugs during Leakage current test.
It may cause a reading error.
● Do not use the Tester on the device which has a power
consumption of 2.5 kVA or more.
#CAUTION
● Where DUT ON Check function detects that DUT is off, blinking
OFF?appears in the LCD and the test is stopped. Turn on
the DUT; then the Tester automatically resumes the test.
Depending on DUTs, even they are turned on, blinkingOFF?
warning sometimes doesnt disappear and the Tester cannot
continue the test automatically. In such a case, press CL I
switch to continue the test.
● Follow the procedure described in 4.3.1 and undertake the
NULL setting before a measurement.
● The alligator clip must make good contact with the enclosure of
the appliance.
● Do not touch the appliance under test while testing is in
progress. Since a high voltage of 250V/500V will be present
and the user may get an electrical shock.
● When the terminal is open or the resistance value exceeds the
measuring range, > (over range display) appears on the
LCD.
5.2 Class ll Test
The Class ll a ppliances have the i ndication ofDOUBLE INSULATION
or the symbol.
Class II test is performed to confirm insulation resistance and
leakage current of the appliance are within the allowable range
specified in the related standard.
To perform Class II test, connect the mains plug of the appliance to
the test socket (1) described in clause 2.4.3. Connector and the Rpe
Test Lead terminal (3). To perform RUN leak test,connect Mains
lead MODEL7277 to Mains lead terminal (2).
Use the following setups, depending upon the type of appliance.
23
Page 25
Connect to exposed metal parts but not
rotating parts or heating elements.
Switch ON
the power.
Press CLII switch to
perform Class II test.
RUN leak test:
Connect Tester to outlet
1. Insulation
test
2. Leakage
current test
Between LINE, NEUTR AL
and EARTH.
RUN Leak test:
Connect Tester to outlet.
(Load current is also measured.)
24
Page 26
CL II Test Flow
(1sec .)
Press to halt a
measurement in
progress.
Tester performs next
test when CL II switch
is pressed while a test
is halted with “FAIL”
indication.
1. Insulation resistance test
Insulation resistance test
Stand-by state
Connection Check
DUT ON Check
DUT is ON.
Voltage value is displayed
while the Tester is connected
to an outlet.
CLⅡ
"CL I" is displayed if Rpe terminal is
integrated in the power cord of DUT.
Press to proceed to
the next; continuity test
CLⅡ
is regarded as PASS.
DUT is powered on, or
CL II is pressed.
DUT is OFF.
25
Page 27
Rins≧1MΩ = PASS Rins<1 MΩ = FAIL
(Red backlight blinks.)
Check screen for RUN leakage test.
Buzzer sounds and blue backlight blinks.
2. Leakage current test
CLⅡ
RUN Leak/ Load current test
ILEAK≦1mA = PASS ILEAK>1mA = FAIL
(Green backlight blinks.)
26
(Red backlight blinks.)
Page 28
# CAUTION
● Where DUT ON Check function detects the DUT is off, blinking
OFF?appears in the LCD and the test is stopped. Turn on
the DUT; then the Tester automatically resumes the test.
Depending on DUTs, even they are turned on, blinking
OFF?warning sometimes doesnt disappear and the Tester
cannot continue the test automatically. In such a case, press
CL ll switch to continue the test.
● When the terminal is open or the resistance value exceeds the
measuring range, > mark (over range display) appears on
the LCD.
● Do not touch the appliance under test while testing is in
progress. Since a high voltage of 250V or 500V will be present
and the user may get an electrical shock.
#WARNING
● When conducting a leakage test the appliance will
automatically power on and will operate in its normal manner.
It is imperative that the appliance is secured safely before the
test is carried out. Extra care needs to be taken with
appliances which have heating elements and rotating parts.
● Firmly insert the plug of the appliance to the socket of the
Tester. Plugs may get hot if Leakage current test is performed
with improper connection.
● Do not connect/remove the plugs during Leakage current
test. It may cause a reading error.
● Do not use the Tester on the device which has a power
consumption of 2.5 kVA or more.
27
Page 29
5.3 Extension Leads Test
This test is for extension leads, and checks for;
● Protective conductor resistance between accessible conductive
parts and connection of protective ear th.
● Insulation resistance between L/N and PE.
● Insulation test between L-N
● Polarity check of the Line and Neutral terminal of plug and socket
● Run leakage current test (Run leakage current test on Extension Leads Test doesnt measure load current.)
Press the Extension switch to
Extension Leads test.
1. Continuity
test
2. Insulation testBetween LINE, NEUTRAL and
3. Polarity
check
4. Leakage
current test
For PE terminal of Extension Leads
EARTH, and between LINE and
NEUTR AL
For LINE and NEUTRAL
RUN Leak test:
Connect Tester to outlet:
28
Page 30
Extension Lead test flow
2. Insulation resistance test 1. Continuity test
(1sec .)
Press to halt a
measurement in
progress.
Tester performs next
test when Extension
switch is pressed
while a test is halted
with “FAIL” indication.
Stand-by
EXTENSION
Press Extension to start a test.
Short press: 1 sec of continuity test
Long press: 5 sec of continuity test
Continuity test
Rpe≦1Ω= PASS Rpe>1Ω=FAIL
(Red backlight blinks.)
Insulation resistance test between:
(1) L&N - E, (2) L-N
L&N-E≧1MΩ= PASS
L&N-E <1MΩ= FAIL
(Red backlight blinks.)
L-N <0.1MΩ = FAIL
(Red backlight blinks.)
29
Page 31
3. Polarity check
Polarity check
Polarity of L and N is correct.
(Pass)
4. Leakage current test
ILEAK≦1mA = PASS
(Green backlight blinks.)
Polarity of L and N is
reversed. (Cross)
(Red backlight blinks.)
Check screen for RUN leakage test.
Buzzer sounds and blue backlight blinks.
EXTENSION
RUN Leak test
(Red backlight blinks.)
Break in L or N
(Open)
(Red backlight blinks.)
ILEAK>1mA = FAIL
30
Page 32
# CAUTION
● Follow the procedure described in 4.3.1 and do Null setting
before a measurement, but use the short MODEL7276 lead
instead of the MODEL 7129A test lead, by plugging the
MODEL 7276 into the Extension Lead Adaptor terminal and
the AU socket on the front of the unit.
● When the terminal is open or the resistance value exceeds
the measuring range, > (over range display) appears on
the LCD.
● Do not touch the device under test while testing is in progress.
Since a high voltage of 250V or 500V will be present, the user
may get an electrical shock.
31
Page 33
5.4 RCD TEST
(1) PRCD test
This test is to test and confirm a PRCD (Portable Residual
Current Device) trips within the rated time by applying specified
current. The Tester incorporates a circuit to test PRCD with
Rated Tripping Current (IΔn) of 10 mA or 30 mA. PRCD trip time
is measured in sequence: ×1(0°)→×1(18 0°)→×5(0°)→×5(180°).
After each test, reset the tripped PRCD to proceed to the next
test. PASS result is given where the PRCD tripped within the
rated time through all the tests.
Press the RCD switch
PRCD
TEST
RESET
to perform RCD test.
# CAUTION
● The Tester can test RCDs rated at 230V+10%, -15%. Input
the rated voltage of the RCD to be tested.
32
Page 34
RCD Test Flow
Stand-by
(1sec.)
Press to halt a
measurement in
progress.
RCD
RCD test stand-by screen
30mA
RCD
×1(0°) test
×1( 0°) test result
Reset the tripped RCD.
Long press switches I∆n:
10 mA and 30 mA
RCD
10mA
33
Page 35
×1(180°) test
×1( 180°) test result
Reset the tripped RCD.
×5(0°) test
×5( 0°) test result
Reset the tripped RCD.
×5(180°) test
Test results are all "PASS":
"PASS" is displayed.
(Green backlight blinks.)
If any one of test results is "FAIL":
"FAIL" is displayed.
(Red backlight blinks.)
34
Page 36
● Check the connection if symbol is displayed in the LCD.
Connection may be incorrect and a test doesnt start even RCD
switch is pressed.
● If " IΔn" setting is greater than the rated residual current of the
RCD, the RCD will trip and "no" may be displayed on LCD.
● Special conditions of RCDs of a particular design, for example
S- type, should be taken into consideration.
● ATT (Anti-Trip Technology)
The KEW6205 uses mains voltage across L-N for PRCD testing. It enables PRCD testing without tripping the RCD of mains power.
It works with the connection for "(1) PRCD test" as shown above;
doesn't work with the connection for "(2) PRCD test via socket outlet".
(2) PRCD test via socket outlet
Make connection as the following illustration indicates when
testing a PRCD connected to a socket outlet. For this test, use
MODEL7276 for connection.
PRCD
MODEL7276
Press the RCD switch to
perform RCD test.
● If a fault voltage exceeds 50 V while applying a test current to test a bu ilt - in type RCD, the LC D sho w sUf. Hi and the test is halted.
● If a voltage exists between the protective conductor and earth, it
may influence the measurements.
● If a voltage exists between neutral and earth, it may influence the
measurements, therefore, the connection between neutral point
of the distribution system and earth should be checked before
testing.
● The potential fields of other earthing installations may influence
the measurement.
● Resistance of earthed electrodes in the circuit under test with a
Probe - auxiliary earthed electrode - should be 200Ω or less.
● Equipment following the RCD, e.g. capacitors or rotating
machinery, may cause a significant lengthening of the measured
trip time.
35
Page 37
6. Memory/ Recall function
6.1 Memory function
Test results on each function can be saved with this function. Press
the Memory switch to save the displayed result. (Max: 999). Hold
down the Memory switch for a while to edit the data no., site no. and
device no. Data can be saved according to the following sequence.
MEMORY
(Short press)*2
POWER
Screen when a test completes.
[No. edit screen ]
Press Memory switch to increase
the number by 1.
Press RECALL switch to decrease
the number by 1.
MEMORY(Long press)*2
Data No. and result are displayed.
"MEM" starts blinking.
PRINT(Enter)
Blink
MEMORY
POWER(ESC) *1
Save the measured values.
"MEM" blinks 2 sec. and
then stops blinking and
stays on 2 sec.
Blink
4sec. later
Returns to the
first screen.
POWER(ESC)
POWER(ESC)
PRINT(Enter)
[ Data No. edit screen ]
Blink
PRINT(Enter)
[ Site No. edit screen ]
Blink
PRINT(Enter)
[ Device No. edit screen ]
Blink
※1 All the changes made through edit screens will be canceled
when the Power switch is pressed and the edit screen is closed.
※2 "FULL" appears when the number of stored results reaches to
the max.
Press the POWER (ESC) switch to return to the screen showing a
measured result.
36
Page 38
6.2 Recall function
This function is to recall and display the saved data on LCD. To see
the date, site no. or device no., recall the saved data on the LCD and
then press the RCD switch. The data recalled with this function can
also be printed out. Saved data can be recalled according to the
following sequence.
On any screen
POWER(ESC)
MEMORY(▲)
RECALL
Data No. and result are
displayed.
* The latest saved data
is displayed.
RCD
PRINTPRINT
Output to printer
RECALL(▼)
Press the MEMORY switch to move to the data no.
containing data in ascending order.
Press the RECALL switch to move to the data no.
containing data in descending order.
Saved date, site no.
and device no. are
displayed.
When there's no saved data:
37
Page 39
6.3 Delete the saved data
Hold down the Memory and Recall switch simultaneously while the
saved data is displayed on the LCD by using Recall function. (Data
wont be cleared at this moment.) Select to delete just one result or
all results. Data can be deleted according to the following sequence.
On any screen
POWER(ESC)
POWER(ESC)
RECALL
The latest saved data is displayed.
MEMORY
RECALL
Confirmation message: delete one saved result.
PRINT(Enter)
A beep sounds.
The next data no. containing
data is displayed.
POWER(ESC)
Returns to the first screen.
Simultaneous
press
MEMORY(▲)
RECALL(▼)
POWER(ESC)
PRINT(Enter)
A beep sounds.
38
Page 40
7. Print function
Test results displayed on the LCD can be transferred to a printer by
pressing the PRINT switch. Use the optional printer cable MODEL7275
to connect a printer and the instrument. Long press of the PRINT
switch can toggle and change the site no. or device no. for print.
- Compatible printer: TLP2824PLUS (Zebra)
The table below shows the communication specs. (RS-232C)
Baud rate9600bps
ParityNon
Stop bit1-bit
Data length8-bit
HandshakeXon/Xoff
Printer
MODEL7275
39
Print example
Page 41
Data can be printed out according to the following sequence.
Screen showing the measured/ recalled result
POWER(ESC)
PRINT(Long press)
[ Site No. edit screen ]
Blink
PRINT(Enter
[ Device No. edit screen ]
Blink
PRINT(Enter)
The displayed device
no. stops blinking.
Data output to printer
[Site No./ Device No. edit screen ]
Press MEMORY switch to increase the
number by 1.
Press RECALL switch to decrease the
number by 1.
)
Returns to the first screen.
40
Page 42
8. USB communication
This instrument has USB communication to transfer the internal
data to a PC. Data can be edited by using the special application
KEW Report. Use the optional USB cable MODEL8263 to connect
the instrument and a PC. The table below shows the USB
communication specs.
Help and information on how to install KEW Report, please see the
instruction manual supplied with MODEL8263-USB (optional item)
or HELP in the KEW Repor t.
Communication method USB Ver1.1
USB ID No.Vender ID:12EC(Hex)
Product ID:6205(Hex)
Serial no.: 0 + 7-digit number
Communication speed12 Mbps
41
Page 43
9. Clock mode
A long press of backlight switch on any screen shows the current
time. To adjust the time, press the PRINT switch while the time is
being displayed.
The time can be adjusted according to the following sequence.
On any screen
POWER(ESC)
PRINT(ENTER)
POWER(ESC)
PRINT(ENTER)
POWER(ESC)
PRINT(ENTER)
POWER(ESC)
LIGHT(Long press)
[Show the current date and time]
PRINT(ENTER)
[Edit mode - Date]
Blink Blink
RCD
[Edit mode - Month]
Blink Blink
RCD
[Edit mode - Year] [Edit mode - Hour]
RCD
Blink Blink
RCD
[Edit mode - Minute]
Blink
Blink
RCD
Blink
Blink
PRINT(ENTER)
POWER(ESC)
PRINT(ENTER)
POWER(ESC)
42
Page 44
10. Backlight
Press the backlight switch to turn on the backlight. Press the switch
again to turn off the backlight. The backlight automatically turns off
if there is no activity for about 2 min.
11. Battery / Fuse replacement
#DANGER
Never open the Battery cover during a measurement. Dispose
the used batteries according to the rules, which is defined by
each community.
#WARNING
To avoid possible electric shock, remove test leads before
opening the battery cover. After replacing batteries, be sure to
tighten up the screws for battery cover.
#CAUTION
Do not mix new and old batteries. Install batteries in correct
polarity as marked inside the battery compartment.
11.1 Battery Replacement
(1) Disconnect the test probe from the Tester.
(2) Open the battery cover by unscrewing the captive screw to
reveal battery compartment.
(3) Always replace all six batteries with new ones at the same time.
Six size A A alkaline batteries (LR6)
(4) Screw the batter y cover back on before using the Tester.
43
Page 45
11.2 Fuse Replacement
(1) Disconnect the test probe from the Tester.
(2) Unscrew the captive and open the battery compartment cover to
replace the fuse.
Fuse type: 10A/250V(F)
Fast acting type ceramic fuse Φ 5 x 20mm.
(3) Screw the battery cover back on
before using the Tester.
+
Fuse
+
+
-
12. Shoulder strap attachment
(1) Turn the knob counterclockwise until it stops. Then the knob
comes out.
Knob
(2) Attach the buckle of the shoulder strap and slide until it clicks.
Slide until it clicks.
44
Page 46
(3) Pass the strap belt down through the buckle from the top, and
up. Adjust the strap for length and secure.
13. Maintenance
Use a very slightly damp cloth for cleaning the Tester. Do not use
abrasives or solvents.
45
Page 47
Page 48
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or designs
described in this manual without notice and without obligations.
02-18 92-2304
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.