KYORITSU 6315 Instruction Manual

Instruction Manual
Instruction Manual
Instruction ManualInstruction Manual
Power Quality Analyzer
Contents KEW6315
Contents ·································································································· 1
Unpacking Procedure ················································································· 5
Safety warnings ························································································· 8
Chap. 1 Instrument overview ······································································ 11
1.1 Functional overview ····································································· 11
1.2 Features ···················································································· 13
1.3 Constructional drawing ································································· 14
1.4 Steps for measurement ································································ 15
Chap.2 Instrument layout ··········································································· 16
2.1 Display (LCD)/ Keys ····································································· 16
2.2 Connector ·················································································· 17
2.3 Side face ··················································································· 18
2.4 Voltage test lead and Clamp sensor ················································ 19
Chap.3 Basic operations ··········································································· 20
3.1 Key operation ············································································· 20
3.2 Icons on the LCD ········································································ 21
3.3 Symbols on the LCD ···································································· 22
3.4 Backlight and Contrast Adjustment ·················································· 22
3.5 Screens ····················································································· 23
Inst/ Integration/ Demand ························································· 23
Vector ·················································································· 24
Waveform ············································································· 24
Harmonic Analysis ·································································· 25
Power quality ········································································ 26
Settings ················································································ 26
Chap.4 Getting started ·············································································· 27
4.1 Preparation ················································································ 27
Putting Input terminal plate on the Input terminal ·························· 27
Attaching Markers to Voltage test leads and Clamp sensors ··········· 28
- 1 - KEW6315
KEW6315 Contents
4.2 Power Supply ············································································· 29
Battery ················································································· 29
Battery Mark on the LCD/ Battery level ······································· 30
How to install batteries ···························································· 31
Power cord connection ···························································· 31
Power supply rating ································································ 32
4.3 Placing / removing SD card ··························································· 33
Inserting SD card ··································································· 34
Removing SD card ································································· 34
4.4 Voltage test leads and Clamp sensor connection ·························· 35
4.5 Start KEW6305 ······································································ 36
Start-up Screen ····································································· 36
Cautionary message ······························································· 36
4.6 Recording procedures ····························································· 37
Start of recording ··································································· 37
End of recording ···································································· 38
Start measurement with “Quick start guide” ······························· 39
Chap.5 Settings ······················································································· 47
5.1 List of setting items ······································································ 47
5.2 Basic setting ··············································································· 48
Settings of wiring system ························································· 49
Wiring connection ·································································· 51
Settings of voltage measurement ·············································· 53
VT/CT ·················································································· 54
Settings of current measurement ··············································· 56
Settings of External input terminal/ reference frequency ················· 58
5.3 Measurement setting ···································································· 59
Settings of demand measurement ············································· 59
Outline of demand measurement concept ··································· 62
Settings for Harmonic analysis ·················································· 63
Threshold setting for Power quality (Event) ································· 65
KEW6315 - 2 -
Contents KEW6315
Filter setting for Flicker measurement ········································· 69
Target power factor for Capacitance calculation ··························· 70
5.4 Recording setting ········································································ 71
Settings for recording items ······················································ 72
Saved items ·········································································· 73
Recording method ·································································· 74
Possible recording time ··························································· 76
5.5 Other settings ············································································· 77
Settings for system environment ··············································· 77
KEW6315 Setting ··································································· 79
5.6 Saved data ················································································· 82
To delete, transfer or format the recorded data ····························· 82
Type of the saved data ···························································· 87
KEW6315 settings and Data loading ·········································· 89
Chap. 6 Displayed items ············································································ 92
6.1 Instantaneous value “W” ······························································· 92
List display of the measured values ··········································· 92
Zoom display ········································································· 96
Displaying Trend graph ··························································· 97
Changing displayed items and display position ···························· 99
6.2 Integration value “Wh” ································································ 100
6.3 Demand ·················································································· 102
Showing the measured values ················································ 102
Shifts in specific period ·························································· 103
Demand change ·································································· 104
6.4 Vector ····················································································· 105
6.5 Waveform ················································································ 107
6.6 Harmonics ··············································································· 108
Displaying harmonics on the bar graph ····································· 108
Displaying the list of harmonics ················································ 112
- 3 - KEW6315
KEW6315 Contents
6.7 Power quality ············································································· 114
Factors impair power quality and symptoms ······························· 114
Displaying recorded events ····················································· 116
Displaying measured flicker values in list form ··························· 120
Displaying trend graph of Pst, 1min ········································· 121
Displaying changes of Plt ······················································· 122
Chap.7 Other functions ··········································································· 123
Chap.8 Device connection ······································································· 125
8.1 Data transfer to PC ···································································· 125
8.2 Using Bluetooth® function ···························································· 126
8.3 Signal control············································································ 126
Connection to input/ output terminals ······································· 126
8.4 Getting power from measured lines ··············································· 128
Chap. 9 PC software for setting and data analysis ········································ 129
Chap.10 Specification ············································································· 130
10.1 Safety requirements ··································································· 130
10.2 General specification ·································································· 130
10.3 Measurement specification ·························································· 133
Measured items and the number of analysis points ····················· 133
Items measured at Instantaneous measurement ························ 134
Items to be calculated ··························································· 137
Items measured at Integration measurement ····························· 140
Items measured at Demand measurement ································ 143
Items measured at Harmonics measurement ····························· 144
Items measured at Power quality measurement ························· 149
10.4 Specification of Clamp sensor ······················································ 152
Chap. 11 Troubleshooting ········································································ 157
11.1 General troubleshooting······························································ 157
11.2 Error messages and actions ························································ 158
KEW6315 - 4 -
Unpacking Procedure KEW6315
Unpacking procedure
We thank you for purchasing our Power Quality Analyzer “KEW6315”. Please
check the contents and instrument before use.
 Items listed below are included with the standard set:
1 Main unit KEW6315
:1 pce
MODEL7255 :1 set
2 Voltage test lead
*red, white, blue, black: 1 pce for each
(with alligator clips)
3 Power cord MODEL7169 :1 pce
4 USB cord MODEL7219 :1 pce
5 Quick manual 1 pce
6 CD-ROM 1 pce
7 Battery Alkaline size AA battery LR6: 6 pcs
8 SD card M-8326-02 :1 pce (2GB)
9 Carrying case MODEL9125 :1 pce
10 Input terminal plate 1 pce
8-color x 4pcs each (red, blue, yellow,
11 Cable marker
green, brown, gray, black, white)
Optional parts
12 Clamp sensor Depending on model purchased
Instruction manual for
13
1 pce
Clamp sensor
14 Magnetic carrying case MODEL9132
15 Power supply adapter MODEL8312(CAT. III 150V, CAT. II 240V)
- 5 - KEW6315
KEW6315 Unpacking Procedure
1. Main unit 2. Voltage test lead 3. Power cord 4.USB cord
5. Quick manual 6.CD-ROM 7. Battery 8. SD card 9. Carrying case
10. Input terminal plate 11. Cable marker
2GB
M-8326-02
KEW6315 - 6 -
Unpacking Procedure KEW6315
12. Clamp sensor
13. Instruction manual for Clamp sensor
14. Magnetic carrying 15. Power supply adapter
case
(depending on model purchased)
50A Type(ø24mm) M-8128
100A Type(ø24mm) M-8127
200A Type(ø40mm) M-8126
500A Type(ø40mm) M-8125
1000A Type(ø68mm) M-8124
3000A Type(ø150mm) M-8129
1000A Type(ø110mm) M-8130
10A Type(ø24mm) M-8146
10A Type(ø40mm) M-8147
10A Type(ø68mm) M-8148
1A Type(ø24mm) M-8141
1A Type(ø40mm) M-8142
1A Type(ø68mm) M-8143
 Storage
Store the items as shown below after use.
In case any of the items listed above are found to be damaged or missing or if
the printing is unclear, please contact your local KYORITSU distributor.
- 7 - KEW6315
KEW6315 Safety warnings
Safety warnings
This instrument has been designed, manufactured and tested according to IEC 61010-1: Safety requirements for Electronic Measuring apparatus, and delivered in the best condition after passing quality control tests.
This instruction manual contains warnings and safety procedures which have to be observed by the user to ensure safe operation of the instrument and to maintain it in safe condition. Therefore, read through these operating instructions before starting to use the instrument.
WARNING
- For about Instruction manual -
Read through and understand the instructions contained in this manual before using the instrument.
Keep the manual at hand to enable quick reference whenever necessary.
The instrument is to be used only in its intended applications.
Understand and follow all the safety instructions contained in the manual.
Read the enclosed Quick manual after reading this instruction manual.
As to the Clamp sensor use, refer to the instruction manual supplied with the sensor.
It is essential that the above instructions are adhered to. Failure to follow the above instructions may cause injury, instrument damage and/or damage to equipment under test. Kyoritsu assumes no responsibility for damage and injury caused by misuse or not following the instructions in the manual.
The symbol indicated on the instrument, means that the user must refer to the related parts in the manual for safe operation of the instrument. It is essential to read the instructions wherever the symbol appears in the manual.
DANGER
WARNING
CAUTION
Measurement Category
To ensure safe operation of measuring instruments, IEC 61010 establishes safety standards for various electrical environments, categorized as O to CAT.IV, and called measurement categories. Higher-numbered categories correspond to electrical environments with greater momentary energy, so a measuring instrument designed for CAT.III environments can endure greater momentary energy than one designed for CAT.II.
O : Circuits which are not directly connected to the mains power supply. CAT.II : Electrical circuits of equipment connected to an AC electrical outlet by a power cord. CAT.III : Primary electrical circuits of the equipment connected directly to the distribution panel, and feeders from the distribution panel to outlets. CAT.IV : The circuit from the service drop to the service entrance, and to the power meter and
: is reserved for conditions and actions that are likely to cause serious or fatal injury.
: is reserved for conditions and actions that can cause serious or fatal injury.
: is reserved for conditions and actions that can cause injury or instrument damage.
primary overcurrent protection device (distribution panel).
KEW6315 - 8 -
Safety warnings KEW6315
the
DANGER
 The instrument is to be used only in its intended applications or conditions. Otherwise, safety functions equipped with the instrument will not work, and instrument damage or serious personal injury may occur. Verify proper operation on a known source before taking action as a result of the indication of instrument.
 With attention to the measurement category to which the object under test belongs, do not make
measurements on a circuit in which the electrical potential exceeds the following values.
* 300V AC for CAT. IV, 600V AC for CAT. III, 1000V AC for CAT. II  Do not attempt to make measurement in the presence of flammable gasses. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion.
 Never attempt to use the instrument if its surface or your hand is wet.
- Measurement -
 Do not exceed the maximum allowable input of any measuring range.  Never open the Battery compartment cover during a measurement.
- Battery -
 Do not try to replace batteries during a measurement.  Brand and type of the batteries to be used should be harmonized.
- Power cord -
 Connect the Power cord to an outlet.  Use only the Power cord supplied with this instrument.
- Power supply connector -
 Never touch the Power supply connector although it is insulated while the instrument is operating with
batteries.
- Voltage test leads -
 Use only the ones supplied with this instrument.  Choose and use the test leads and caps that are suitable for the measurement category.  When the instrument and the test lead are combined and used together, whichever lower category
either of them belongs to will be applied. Confirm that the measured voltage rating of the test lead is
not exceeded.  Do not connect a Voltage test lead unless required for measuring the desired parameters.  Connect Voltage test leads to the instrument first, and only then connect them to the circuit under test.  Keep your fingers behind the barrier during a measurement. Barrier: provides protection against electrical shock and ensuring the minimum required air and
creepage distances.  Never disconnect the voltage test leads from the connectors of the instrument during a measurement
(while the instrument is energized).  Do not touch two lines under test with the metal tips of the test leads.  Never touch the metal tips of the test leads.
- Clamp sensor -
 Use only the ones dedicated for this instrument.  Confirm that the measured current rating of the test lead and the maximum rated voltage are not
exceeded.  Do not connect a Clamp sensor unless required for measuring the desired parameters.  Connect sensors to the instrument first, and only then connect them to the circuit under test.  Keep your fingers behind the barrier during a measurement. Barrier: provides protection against electrical shock and ensuring the minimum required air and
creepage distances.
- 9 - KEW6315
KEW6315 Safety warnings
 Never disconnect sensors from the connectors of the instrument while the instrument is in use.  Connect to the downstream side of a circuit breaker since a current capacity at the upstream side
is large.  Do not touch two lines under test with the metal tips of the test leads.
Caution
 Caution should be taken since conductors under test may be hot.  Never apply currents or voltages exceeding the maximum allowable input for the instrument for a
long time.  Do not apply currents or voltages for the Clamp sensors or Voltage test leads while the instrument
is off.  Don’t use the instrument at dusty places or to be spattered.  Don’t use the instrument under a strong electric storm or in the vicinity of energized object.  Never give strong vibrations or drop shocks.  Insert an SD card to the slot with the top side turned up. If the card is inserted up-side-down, the SD
card or the instrument may be damaged. Confirm the
 While using an SD card, do not replace or remove the card. (The symbol blinks while accessing
SD card.) Otherwise, the saved data in the card may be lost or the instrument may be damaged.
- Clamp sensor -
 Do not bend or pull the cable of the Clamp sensor.
- Treatment after use -
 Power off the instrument and disconnect the Power cord, Voltage test leads and Clamp sensors from
the instrument.  Remove the batteries if the instrument is to be stored and will not be in use for a long period.  Remove the SD card when carrying the instrument.  Never give strong vibrations or drop shocks when carrying the instrument.  Do not expose the instrument to direct sunlight, high temperatures, humidity or dew.  Use a damp cloth with neutral detergent or water for cleaning the instrument. Do not use abrasives
or solvents.  Do not store the instrument if it is wet.
Carefully read and follow the instructions: DANGER, WARNING, CAUTION and NOTE ( ) described in each section.
Meaning of symbols on the instrument:
User must refer to the explanations in the instruction manual.
Instrument with double or reinforced insulation
~~~~
KEW6315 - 10 -
AC
(Functional) Earth terminal
1.1 Functional overview
-
/ max/ min instantaneous values of
power.
Integration values also can be viewed by switching screens.
Moreover, demand values with the preset target value can
(P.
Vector and Wiring check
Vectors of voltage and current per CH are displayed
on a graph. Executing wiring check function is
possible from this screen.
See
Chap. 1 Instrument overview
to start
up setting by
for further details.
1.1 Functional overview
Start/ Stop
Choose either “Quick start guide” or “Start now recording. Can do simple and fast start­selecting “Quick start guide”.
See “4.6 Recording procedures (P.37)
Inst/ Integration/ Demand
Display the avg voltage/ active power/ apparent power/ reactive
KEW6315
current/
also be checked.
See “6.1 Inst “W” (P.92), 6.2 Integration “Wh”
6.3 Demand (P.102) ” for further details.
100),
6.4 Vector(P.105) for further details.
11 - KEW6315
KEW6315
12
Waveform
Waveforms
Waveform
Harmonic Analysis
Harmonic components of voltage and current per CH are
displayed on a graph.
See
1.1 Functional overview
graph.
See “6.5
of voltage and current per CH are displayed on a
(P.107) for further details.
6.6. Harmonics(P.108) for further details.
Setting (SET UP)
Adjust settings of KEW6315 and measurements.
See “5. Settings(P.47) for further details.
KEW6315 -
Power Quality (QUALITY) event
Display voltage swell, dip, int, transient, inrush current and flicker.
See “6.7. Power Quality(P.114) for further details.
-
1.2 Features KEW6315
1.2 Features
This is a Clamp-type Power Quality Analyzer that can be used for various wiring systems. It can be used
for simple measurements of instantaneous/ integration/ demand values, and also for analysis of harmonics
and events related to power quality and for the simulation of power factor correction with capacitor banks.
Moreover, it can display waveforms and vectors of voltage and current. Data can be saved either on the
SD card or in the internal memory, and can be transferred to PC via USB, or in real time via Bluetooth®
communication.
Safety construction
Designed to meet the international safety standard IEC 61010-1 CAT.IV 300V/ CAT.III 600V/ CAT.II 1000V.
Power quality analysis
KEW6315 is designed to meet the international standard IEC61000-4-30 Class S and can measure frequency and r.m.s. voltage with high accuracy, and also can analyze harmonics. Moreover, it can measure swell, dip, interruption, transient, inrush current and flicker, gapless, all at once.
Power measurement
KEW6315 measures active/reactive/apparent power, electrical energy, power factor, r.m.s. current, phase angle and neutral current simultaneously.
Wiring configuration
KEW6315 supports: Single-phase 2-wire (4-system), Single-phase 3-wire (2-system), Three-phase 3-wire
(2-system) and Three-phase 4-wire.
Demand measurement
Electricity consumption can be easily monitored so as not to exceed the target maximum demand values.
Waveform/ vector display
Voltage and current can be displayed by waveform or vector.
Saving data
KEW6315 is endowed with a logging function with the preset recording interval. Data can be saved by manual operation or by specifying date & time. Screen data can be saved by using the Print Screen function.
Dual power supply system
KEW6315 operates either with AC power supply or with batteries. Size AA alkaline dry-cell batteries and size AA Ni-MH rechargeable batteries can both be used. To charge size AA Ni-MH rechargeable batteries, use the charger which is manufactured by the same company as the batteries. In the event of power interruption, while operating with AC power supply, power to the instrument is automatically restored by the batteries in the instrument.
Large display
TFT color display with large screen.
Light & compact design
Clamp sensor type, compact and light weight design.
Application
Data in the SD card or the internal memory can be saved in PC via USB. Analysis of the downloaded data and instrument settings are possible by using the special software “KEW Windows for KEW6315”. Real-time communication with android devices is available via Bluetooth®.
Input/ Output function
Analog signals from thermometers or light sensors can be measured simultaneously with electrical power data via 2 analog inputs (DC voltage); when any events related to power quality occur, signals can be transmitted to alarm devices via one digital output.
- 13 - KEW6315
KEW6315
14
1.3 Constructional drawing
1.3 Constructional overview
Current input
AC Voltage input
Power cord
Digital output (1 ch) to recorder or alarm
Size AA alkaline dry-cell battery (LR6) or Size AA Ni-MH rechargeable battery
KEW6315 -
Analog input (2 ch) from thermometer
or illuminometer
USB
PC
SD card
-
1.4 Steps for measurement KEW6315
1.4 Steps for measurement
Read through the operating instructions described in “Safety
warnings” (P.8) before starting to use the instrument.
Preparation
Chap. 4: Getting started (P.27)
Cord and Sensor Connection
Clause 4.4: Voltage test leads and Clamp sensor connection (P.35)
Turning on the instrument.
Clause 4.5: Start KEW6315 (P.36)
Making basic settings.
Clause 5.2: Basic setting (P.48)
Wiring
Clause 5.2: Basic setting – Wiring (P.51)
Reading the setup data
Clause 5.6: Saved data (P.82)
Wiring check
Clause 6.4: Vector (P.105)
Advanced setting and data save method
Clause 5.3: Measurement setting (P.59)/ 5.4: Recording setting (P.71)
Confirming the measured values.
Clause 6.1: Inst value [W] (P.92)/ 6.4: Vector (P.105)
Start/ stop recording
Clause 4.6: Recording procedures (P.37)
Confirming the measured data.
Chap. 6: Displayed items (P.92)
Disconnect the cords and sensors from the measurement line, and turn off
the instrument.
Data analysis on PC
Clause 8.1: Data transfer to PC (P.125)
Chap. 9: PC software for setting and data analysis (P.129)
- 15 - KEW6315
KEW6315 2.1 Display (LCD)/ Keys
Chap.2 Instrument layout
2.1 Display (LCD)/ Keys
Display (LCD)
Keys
KEW6315 - 16 -
2.2 Connector KEW6315
2.2 Connector
AC Voltage Input Terminal
(VN, V1, V2, V3)
Current Input Terminal
(A1, A2, A3, A4)
Power Connector
AC Voltage Input
Wiring configuration
Terminal
Terminal Cover
Current Input
Single-phase 2-wire (1-system) 1P2W×1 VN, V1 A1
Single-phase 2-wire (2-system) 1P2W×2 VN, V1 A1, A2
Single-phase 2-wire (3-system) 1P2W×3 VN, V1 A1, A2, A3
Single-phase 2-wire (4-system) 1P2W×4 VN, V1 A1, A2, A3, A4
Single-phase 3-wire (1-system) 1P3W×1 VN, V1, V2 A1, A2
Single-phase 3-wire (2-system) 1P3W×2 VN, V1, V2 A1, A2, A3, A4
Three-phase 3-wire (1-system)
Three-phase 3-wire (2-system)
Three-phase 3-wire
3A 3P3W3A V1, V2, V3 A1, A2, A3
3P3W×1 VN, V1, V2 A1, A2
3P3W×2 VN, V1, V2 A1, A2, A3, A4
Terminal*
Three-phase 4-wire
3P4W×1 VN, V1, V2, V3 A1, A2, A3
* Measurements of r.m.s. values and harmonics are possible at the Current terminals, which are
not used for wiring connection.
- 17 - KEW6315
KEW6315 2.3 Side face
Analog input/ Digital output
2.3 Side face
< When the Connector cover is closed. >
SD slot cover
Analog input/ Digital output cover
< When the Connector cover is opened. >
SD card slot
USB port cover
USB port
KEW6315 - 18 -
terminals
2.4 Voltage test lead and Clamp sensor KEW6315
2.4 Voltage test lead and Clamp sensor
<Alligator clip>
* Attached to the top part of voltage test lead
<Clamp sensor>
Barrier is a mechanical safety part and provides protection against electrical shock and ensuring the
minimum required air and creepage distances. Keep your fingers and hands behind the barrier during a
measurement.
Barrier
Barrier
- 19 - KEW6315
KEW6315 3.1 Key operation
3 Basic operations
3.1 Key operation
Function Key
Execute the displayed function.
PRINT SCREEN Key
Save the displayed screen as BMP file.
DATA HOLD Key/ KEY LOCK Key
Hold the readings on the display.
* Measurement continues while the
readings are being held on the display.
Long press (at least 2 sec) disables
KEY
LOCK
all Keys to prevent operational errors. Another long press (at least 2 sec) is required to restore the disabled Keys.
LCD Key
Turn on/ off the backlight. Long press (at least 2 sec) changes the brightness and contrast.
Cursor Key
Select items or switch displays.
KEW6315 - 20 -
START/ STOP Key
Start/ stop
measurement.
Status LED
Green
Red Blink: Backlight is off.
Light up: Recording& measuring
Blink: Stand-by
W/Wh
Vector View phases.
Waveform
View inst, integration
and demand values.
View voltage/ current
waveforms.
Power Key
Power on/ off.
Menu Key
ENTER Key
Confirm the entries.
ESC Key/ RESET Key
Cancel setting changes and return to
the previous settings.
SETUP Key
Change and confirm: Basic, Measurement, Record-
ing and other settings, and also edit the saved data.
Harmonic
Analysis
Power
Quality
View harmonic voltage, current
and power energy.
View the detailed info about:
swell, dip, int, transient, inrush
current and flicker.
3.2 Icons on the LCD KEW6315
3.2 Icons on the LCD
Icon Status
KEW6315 is operating with battery. This icon varies in 4 steps according to the
battery power condition.
KEW6315 is operating with AC power.
Holding the display update.
Keys are locked.
Buzzer is off.
SD card is set and available.
Recording the data on the SD card.
Available free space in the SD card is not enough.
Failed to access to the SD card.
Internal memory is available.
* This icon is displayed when a measurement starts without SD card.
Recording the data in the internal memory.
Available free space in the internal memory is not enough.
Stand-by mode
Recording the measured data.
Capacity of recording media is full.
USB is available.
Bluetooth® is available.
- 21 - KEW6315
KEW6315 3.3 Symbols on the LCD
3.3 Symbols on the LCD
V*1
P
PF
DC1
An*2 Neutral current PA*
WP+
WP-
THD
Pst
(1min)
Phase voltage VL*1 Line voltage A Current
+
Active power
Power
factor
consumption
-
regenerating
+
lagging
-
leading
Analog input
voltage at 1ch
Active power energy
(consumption)
Active power energy
(regenerating)
Reactive
Q
power
f Frequency
DC2
3
Analog input
voltage at 2ch
Phase
angle
WS+
WS-
Apparent power energy
(consumption)
Apparent power energy
(regenerating)
+
+
-
-
lagging
leading
lagging
leading
Voltage/ Current
total distortion factor
Voltage flicker (1 min) Pst
Short term voltage flicker
*1
W screen: Displays of V and VL can be “customized” when “3P4W” is selected.
*2
W screen: “An” is displayed only when “3P4W” is selected.
*3
W screen: Displays of PA and C can be “customized”.
S
C*3
WQi+
WQc+
Plt
Apparent
power
Capacitance calculation
Reactive power energy
(lagging)
Reactive power energy
(leading)
Long term voltage flicker
3.4 Backlight and Contrast Adjustment
Hold down the “ LCD Key at least 2 sec to show the sliding bar to adjust the backlight brightness and
display contrast. Use the Cursor Key to slide the cursor on the bar for the adjustment. Press the ENTER
Key and exit from the adjustment mode. Press the ESC or LCD Key again to cancel the adjustment and exit
from the adjustment mode.
KEW6315 - 22 -
Brightness adjustment
Backlight brightness can be changed by 11 levels.
Contrast adjustment
Contrast can be changed by 11 levels.
Inst/ Integration/ Demand
-
selected items.
-
(Integration value)
3.5 Screens
KEW6315
Inst/ Integration/ Demand
Switching screens
Press the key to toggle the screens.
W (Inst value)
Wh
Customize
Select and change the items to be displayed.
Demand
Zoom and display the
4-split 8
Zoom
split
Changes of measured values are
displayed on a graph.
Trend
23 - KEW6315
KEW6315
24
Diagram of the selected wiring is
Vector
Vector
displayed.
Switching screens
Wiring diagram
Wiring check
Checked results will be displayed.
Waveform
Switching screens
KEW6315 -
-
Harmonics
-
Current
Logarithm
Phase angle
Current
KEW6315
Harmonics
Switching screens
Voltage, Linear, Overall display
List, Rate of content
Power
Zoom
RMS value
Power
25 - KEW6315
KEW6315
26
Power quality
Power quality
Switching displayed items
Settings
Switching displayed items
Event
Flicker
Toggle the screens with the Cursor Key. (right or left )
KEW6315 -
-
Putting Input terminal plate on the Input terminal KEW6315
4 Getting started
4.1 Preparation Putting Input terminal plate on the Input terminal
Six Input terminal plates are supplied with this instrument. Choose one Plate which matches the standard cord colors where the instrument is used. Put the Plate to the Input terminal observing the orientation. * Clean the Input terminal before putting the Plate and confirm it isn’t wet.
TYPE1
TYPE4
TYPE2
TYPE5
TYPE3
TYPE6
Put a proper Input terminal plate.
Input terminal plate
VN V1/A1 V2/A2 V3/A3 A4
TYPE 1 blue red green black yellow
TYPE 2 blue brown black gray yellow
TYPE 3 black yellow green red white
TYPE 4 blue black red white yellow
TYPE 5 white black red blue yellow
TYPE 6 black red yellow blue white
- 27 - KEW6315
KEW6315 Attaching Markers to Voltage test leads and Clamp sensors
Attaching Markers to Voltage test leads and Clamp sensors
Attach Markers to the both ends of the Voltage test leads and Clamp sensors harmonized with the Input terminals. * Supplied Markers are 32 pcs in total : 4pcs each color (red, blue, yellow, green, brown, gray, black, white).
Marker (32 pcs in total)
Attach Markers to the both ends of a Sensor.
Attach Markers to the both ends of a Voltage test lead.
KEW6315 - 28 -
Loading...
+ 133 hidden pages