1000A DC/AC DIGITAL CLAMP METER
9 FUNCTIONS 32 RANGES
Model 2781 / 2781-T(True RMS)
? Overload protection on all ranges.
? Maximum Voltage between any
terminal and earth ground 600V rms
? Dual slope integration
FEATURES :
GENERAL SPECIFICATIONS :
ï Sensing : Average sensing (Model 2781) True RMS sensing (Model 2781-T)
ï Jaw opening size : cables f 40mm ï Display : 3 ¾ digit 3999 counts liquid
crystal LCD display. ï Over range indication : Display of “OL” on LCD at the highest position means range selection too low for the input. (Except for ranges of 1000A AC, & 600V AC / DC) ï Polarity : Symbol “-” automatic displayed for negative input.
Æ
ACCESSORIES :
Test leads (pair), Battery installed, User’s manual, K-Type Thermocouople(model 2781-T) & Carrying case.
ELECTRICAL SPECIFICATIONS : 2781 / 2781-T
Accuracy are : ±(% of reading + number of digits) at 18°C to 28°C with relative humidity below 80%R.H.
? Data-hold facility
? DCA zero button for accurate reading
? Low battery indication
? Auto power off
RESISTANCE
Range
Accuracy
Resolution
0.1 W
±(1.2%rdg+8dgts)
400 W
1 W 10 W 100 W
±(1.0%rdg+2dgts)
4 KW 40 KW 400 KW
1 KW 10 KW
± (2%rdg+5dgts)
4 MW 40 MW
Overload Protection : 250V rms
0.001Hz
0.01 Hz
0.1 Hz 1 Hz 10 Hz 100 Hz 1 KHz
Unspecified
±(0.5%rdg+3dgts)
Unspecified
FREQUENCY
10Hz
100Hz 1KHz 10KHz 100KHz 1MHz 10MHz
Range
Accuracy
Resolution
Over load protection : 250Vrms Sensitivity : 1V
(2781only)
CAPACITANCE
Range Accuracy
Resolution
0.01nF
0.1 nF
1 nF 10 nF 10 nF
±(2.5%rdg+10dgts)
±(2.0%rdg+4dgts)
Unspecified
40 nF 400 nF 4 mF 40 mF 100 mF
Overload Protection : 250V rms
TEMPERATURE (2781-T)
Range Accuracy
Resolution
o
±(2.5%+3 C)
o o
-40 C~1000 C
o o
-40 F~1820 F
o
±(2.5%+5 F)
o
1C
o
1F
Type-K thermocouple range & accuracy not specified Supplied K-type thermocouple suitable for 250°C.
DIODE & CONTINUITY TEST
Display read approx. Forward voltage of diode. Accuracy ± (3.0%rdg+3)
If the resistance is less than 50W, the beeper sounds continuously
Range
Description
Overload Protection : 250V RMS
AC CURRENT
Range Accuracy
Resolution
Overload Protection : 1200A
10 mA
40 A (2781-T)
± (2%rdg + 5dgts)
1 A
1000 A
0.1 A
400 A
ï Sampling rate : 3 times per second (Digital display) ï Auto power off : The meter is automatically powered off after idling
for 15 minutes. To awake the meter, turn the rotary function switch or push any button. ï Operating Temperature & Humidity : 0°C to 40°C; < 80% R.H. Non-condensing ï Low battery : The symbol “ ” is displayed when the batteries are weak and below the operating Voltage. Replace batteries immidiately.
ï Battery life : Approx. 60 hrs continuously use with alkaline batteries. ï Power supply : 1.5V AAA x 2 ï Dimension : 228(L) x 76(W) x 39(H) mm ï Weight : approx. 465gms. (Including batteries)
DC CURRENT
Range Accuracy
Resolution
Overload Protection : 1200A
10 mA
40 A (2781-T)
± (2%rdg + 5dgts)
1 A
1000 A
0.1 A
400 A
Overload protection : 1200Vrms Input Impedance : 10MW
DC VOLTAGE
Range Accuracy
Resolution
0.001 V
0.01 V
0.1 V
±(0.8%rdg+2dgts)
400 mV (2781-T)
4 V 40 V
400 V
±(1.0%rdg+5dgts)
1 V
1000 V
±(1.0%rdg+2dgts)
0.1 mV
Overload protection : 660Vrms Input Impedance : 10MW Frequency Response : 40Hz ~ 450Hz for 400V & below,
40Hz ~ 100Hz for 750V
AC VOLTAGE
Range Accuracy
Resolution
0.001 V
0.01 V
0.1 V
±(1.0%rdg+5dgts)
400 mV (2781-T)
4 V 40 V 400 V
±(1.8%rdg+5dgts)
1 V
750 V
±(1.5%rdg+5dgts)
0.1 mV
DUTY CYCLE
Range Accuracy
Resolution
0.1% ~ 99.9%
®
USE TRUE RMS WHEN MEASURING
The waveforms on today’s AC power lines are anything but clean. Electronic equipment such as office computers, with their switching power supplies, produce harmonics that distort power-line waveforms. These distortions make measuring AC voltage inaccurate when you use an averaging DMM.
Average voltage measurements work fine when the signal you’re measuring is a pure sine wave, but errors mount as the waveform distorts. By using true RMS measurements, however, you can measure the equivalent heating effect that a voltage produces, including the heating effects of harmonics. Table 1 shows the difference between measurements taken on averaging DMMs & those taken on true RMS DMMs. In each case, the measured signal’s peak-to-peak value is 2V. Therefore, the peak value is 1V.
For a 1-V peak sine wave, the average & RMS values are both 0.707V. But when the input signal is no longer a sine wave, differences between the RMS values & the average readig values occur. Those errors are most prominent when you are measuring square waves & pulse waveforms, which are rich in harmonics.
AC WAVEFORMS
Table 1. Average versus true RMS comparison of typical waveforms.
Waveform Actual
Pk-Pk
Sine Wave Triangle Wave Square Wave
Pulse (25% duty Cycle) Pulse (12.5% duty Cycle) Pulse (6.25% duty Cycle)
2.000 0.707 0.707 0%
2.000 0.577 0.555
2.000 1.000 1.111
2.000 0.433 0.416
2.000 0.331 0.243
2.000 0.242 0.130
True RMS Reading
Average Reading
Reading
Error
-3.8% +11.1%
-3.8%
-26.5%
-46.2%
One limitation to making true RMS measurements is crest factor, and you should consider crest factor when making AC measurements. Crest factor is the ratio of a waveform’s peak (”crest”) voltage to its RMS voltage. Table 2 shows the crest factors for ideal waveforms.
Table 2. Crest factors of typical waveforms.
Waveform
DC Square Wave
Sine Wave Triangle Wave Pulse (25% duty Cycle) Pulse (12.5% duty Cycle) Pulse (6.25% duty Cycle)
A DMM’s specifications should tell you the maximum crest factor that the meter can handle while maintaining its measurement accuracy. True RMS meters can handle higher crest factors when a waveform’s RMS voltage is in the middle of the meter’s range setting. Typically, a DMM may tolerate a crest factor of 3 near the top of its scale but it might handle a crest factor of 5 that’s in the middle of the range. Therefore, if you’re measuring waveforms with high crest factors (greater than 3), you should adjust the DMM so the measured voltage is closest to the center of the measurement range.
Another limitation of true RMS is speed. If you’re measuring relatively clean sine waves, then you can save time & money by using as averaging DMM. True RMS meters cost more than averaging meters and can take longer to produce measurements, especially when measuring millivolt-level AC signals. At those low levels, true RMS meters can take several seconds to stabilize a reading. Averaging meters won’t leave you waiting.
Crest Factor
1.000
1.000
1.414
1.732
1.732
2.646
3.873
D:/Chhaya/Coreldraw files/True RMS when measuring AC waveforms.cdr
Hz/DUTY
SELECT
1000
400
400
A
OFF
1000
A
~
V
V
~
W
RANGE REL
CAP
Hz
OFF
A
CLAMPMETER
2781
G-17, Bharat Industrial Estate, T.J.Road, Sewree (W), Mumbai - 400015. (INDIA)
Sales Direct : 24156638 Tel : 91-22-2412 4540, 2418 1649 Fax : 91-22-2414 9659 Email : kusam_meco@vsnl.net Website : www.kusamelectrical.com
Www.kusam-meco.co.in
Service Centre:
THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WA RRAN T Y OF ME RCH A N TA BILI T Y OR FITNESS FOR A PARTICULAR PURPOSE. “KUSAM-MECO” SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE WHATSOEVER.
All transaction are subject to Mumbai Jurisdiction.
Nearly every electrical engineer has a hand held digital clamp meter (Tongtester). We sometimes take them for granted, until we damage them or “burn them out”. If you incorrectly connect your clamp meter to a circuit, or if you have the clamp meter or wrong setting, you damage the meter and possibly hurt yourself. You can also get into trouble if you try to measure the voltage across a changed capacitor.
Clamp meter users frequently burn their meters by trying to measure current the same way as they measure voltage. Remember, you measure voltage across a circuit, and current through a circuit. When you use the current input, your clamp meter becomes a low impedance circuit element.
Even if you correctly insert your clamp meter in to the circuit, you can still damage you meter. Don’t try to measure current in excess of your meter’s capacity. Check the current capacity of the Clamp meter.
If you are measuring current in ind ust ria l environment. *you can easily exceed those ratings. The best way to avoid damage is to use a clampmeter with high current measuring capacity. To prevent excess current from flowing through your meter, set your meter to the correct function, say current, and its highest range for the setting. If the reading is small, change the range to the next lower range till the reading can be read with the best possible accuracy. When measuring voltage, connect the test leads before your apply power to your circuit. To be safe, start by setting your meter to its highest range first.
TAKE MEASUREMENT CAREFULLY AND YOU’LL SPARE YOUR METER AND YOURSELF, SOME PAIN
Table of Contents
Title Page
Overview.....................................................1
Unpacking Inspection 2 Specifications A. General Specifications 3 B. Electrical Specifications 4 to 6 AC Current DC Current
AC Voltage DC Voltage Resistance Capacitance Frequency & Duty cycle
Diode / Continuity Test Rules For Safe Operation 7 & 8 International Electrical Symbols 9 The Clamp Meter Structure 10 Display Symbols 12 Measurement Operation A. AC Current Measurement 13 B. DC Current Measurement 14 C. AC Voltage Measurement 15 D. DC Voltage Measurement 17 E. Measuring Resistance 18 F. Capacitance Measurement 20 G. Frequency Measurement 21 H. Testing Diodes 22 I. Testing Continuity 24 Sleep Mode 25 Maintenance :­ A. General Service 25 B. Replacing the Battery 26 Test Certificate 27 Warranty
..................................
........................
.......................
............................
...................
.........................
..........................................
....................
....................
....................
...................
.........................
..................
....................
.....................................
.................................
.................................................
...................................
..........................
............................................
......................................................28
Unpacking Inspection
Open the package case and take out the Clamp Meter from the carrying case. Check the following items carefully to see any missing or damaged part :
Item Description Qty.
1 English Operating Manual 1 piece 2 Test Lead 1 pair
In the event you find any Part missing or damaged, please contact your dealer immediately.
2
Overview
Warning
To avoid electric shock or personal injury, read the “S af ety Info rm ati on” and “Rul es for Sa fe Operation”carefully before using the Meter.
Digital Clampmeter Model - 278 (hereafter referred to as “the Meter”) is a 3¾ digits Clampmeter with steady operations, and highly reliable hand-held measuring instrument having different measurement positions. The Clampmeter not only can measure AC/DC Voltage, AC/DC Current, Resistance, Capacitance, but also has Diode Test, Frequency & duty cycle measurement & full icon Display.
Terms in this manual
Warning : identifies conditions and
actions that could result in serious injury or even death to the user.
Caution : identifies conditions and
actions that could cause damage or malfunction in the instrument.
1
!
!
1
! Frequency Response : 40~65Hz ! Overload Protection : 1100A
Range Resolution Accuracy
400 A 0.1 A
1000 A 1 A
± (2%+5)
Range Resolution Accuracy
4 V 0.001 V 40 V 0.01 V
400 V 0.1 V 750 V 1 V
± (1.0%+5)
± (1.5%+5)
Range Resolution Accuracy
400 A 0.1 A
1000 A 1 A
± (2%+5)
· Overload Protection : 1100A
· Overload protection: 820V rms
· Input impedance : 10MW
· Frequency response : 40Hz~450Hz for 400V
and below. 40Hz~100Hz for 750V
AC Voltage (Auto ranging)
DC Current
Electrical Specifications
Accuracy : ±(% reading + digits)
Operating temperature : 23°C±5°C
Relative humidity : < 75% R.H.
AC Current
4
SPECIFICATIONS
GENERAL SPECIFICATIONS :
!
Maximum voltage including transient overvoltage between any terminals and grounding : 750Vrms.
!
Display :3 digi ts
LCD for Accurate Reading.
!
Auto Polarity Display
!
Overloading : Display OL or -OL
!
Battery Deficiency :Display “ ”
!
Measurement Speed :Update 2.5 times/second.
!
Measurement :When the conductor being Deviation measured is not placed in
a correct posi ti on dur in g AC c urrent me asureme nt . It will cause ±1% reading deviation
!
Max. Jaw Size :40 mm diameter.
!
Power :2pcs of 1.5V battery (AAA)
!
Auto Power Off : 15 minutes after the last
operation was made. To
turn ON the display, turn
the rotary switch to any position or push any
button.
!
Temperature :Operating 0°C`40°C (<80%
and humidity R.H.) Storage:- 0°C~60°C
(<80% R.H.)
!
Dimensions (LxWxH):228mm X 76mm X 39mm
!
Weight :Approximate 465g (battery
Included)
!
Accessories :
¾ (3 99 9 Cou nts )
Test Leads, Carrying Case,
Manual, Battery installed.
+ -
3
DC Voltage (Auto ranging)
Overload Protection : 1100V rms Input impedance : 10MW.
Resistance (Auto Ranging)
Range Resolution Accuracy
4 V 0.001 V 40 V 0.01 V 400 V 0.1 V 1000 V 1 V
Range Resolution Accuracy
400 0.1 ±(1.2%+5)
4KW 1 W
40KW 10 W
400KW 100 W
4MW 1KW
40MW 10KW
W W
±(0.8%+2)
±(1.0%+2)
±(1.0%+2)
±(2%+5)
Overload protection : 250Vrms
Range Resolution
Accuracy
(50Hz?60Hz)
40nF 0.01 nF 400nF 0.1 nF
4mF 1 nF
40mF 10 nF
100mf 100 nF Unspecified
±(2.0%+4)
±(2.5%+10)
Overload Protection : 250V rms.
Capacitance (Auto Ranging)
5
Diode/Audible continuity
Overload protection : 250 Vrms
Display read approx. Forward voltage of diode. Accuracy : ±(3.0%rdg+3)
If the resistance is less than 50W, the beeper sounds continuously
Range
Description
Frequency & Duty Cycle (Auto Ranging)
Range Resol-
ution
Accuracy Overload
protection
10Hz 0.001Hz Unspecified 100Hz 0.01Hz 1KHz 0.1Hz 10KHz 1Hz 100KHz 10Hz 1MHz 100Hz 10MHz 1KHz Unspecified
DUTY Range : 0.1% to99.9%
Sensi­tivity
±(0.5%+3)
1V
250Vrms
6
RULES FOR SAFE OPERATION (2)
!
Disconnect circuit power and discharge all high­ voltage capacitors before testing resistance, continuity, diodes, or current.
!
Replace the battery as soon as the low battery
indicator appears. With a low battery, the
Meter might produce false readings that can
lead to electric shock and personal injury.
!
Turn the Meter power off when it is not in use
and take out the battery when not using for a
long time.
!
Constantly check the battery as it may leak
when it has not been used for some time,
replace the battery as soon as leaking
appears. A leaking battery will damage the
Meter.
8
To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, adhere to the following rules :
!
Before using the Meter inspect the case. Do not use the Meter if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic. Pay attention to the insulation around the connectors and Clamps.
!
Inspect the test leads for damaged insulation or exposed metal. Check the test leads for Continuity. Replace damaged test leads with identical electrical Specifications before using the Meter.
!
Do not apply more than the rated voltage, as marked on the Meter, between the terminals or between any terminal and grounding.
!
The rotary switch should be placed in the right position and no any changeover of range shall be made while measurement is conducted to prevent damage of the Meter.
!
When measurement is taken at an effective voltage over 60V in DC or 30V rms in AC, special care should be taken for there is danger of electric shock.
!
Use the proper terminals, function, and range for your measurements.
!
Do not use or store the Meter in an environment of high temperature, humidity, explosive, inflammable and strong magnetic field. The performance of the Meter may deteriorate after the meter gets dampened.
!
When using the test leads, keep your fingers behind the finger guards.
!
RULES FOR SAFE OPERATION
Warning
7
THE CLAMP METER STRUCTURE
1) SENSOR JAWS :
Pick up the current flowing through the conductor.
2) Jaw Opening TheTrigger :
Press the trigger to open the jaw.
3) LCD display :
3 3/4 digital LCD with indications for measurement values, unit symbol, decimal point, polarity and low battery etc.
When switch at W test position, press this button to select W or or mode. When switch at DCV, ACV test position, press this button to select V or Hz or DUTY mode. When switch at Hz test position, press this button to select Hz or DUTY mode.
4) SELECT or Hz/DUTY button :
1
4
3
6
5
2
V
1000
+
DC/AC
CLAMP METER
2754
HOLD
PEAK
OFF
A
1000
600
400
1000
400
600
600
A
V~~
4KW
Hz
DCA ZERO
600V
600V
~ MAX CAT III
COM
+
A
9
7
8
2781
10
- +
~
AC (Alternating Current).
DC (Direct Current).
Both DC & AC.
Grounding.
Double Insulated.
Deficiency of Built-In Battery.
Continuity Test.
Diode.
Fuse.
Warning ! Refer to the Operating Manual.
Caution ! Risk of Electric Shock.
~
!
International Electrical Symbols
9
No. Symbol Meaning
1 AC Indicator for AC Voltage or Current.
2 DC Indicator for DC Voltage or Current
3 The battery is low.
Warning : To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears.
4 The Meter is in auto range mode in
which the Meter automatically selects the range with the best resolution.
5 Test of diode.
6 Data hold is active.
7 The continuity buzzer is on.
8 nf, mF The unit of Capacitance
9 W : Ohm. The unit of resistance.
3
KW : kilo-ohm. 1 x 10 or 1000 ohms.
6
MW : Megaohm. 1 x 10 or 1,00,000 ohms.
- +
Auto
H
W, KW, MW
DISPLAY SYMBOLS
Auto
- +
H
V/A
DC
AC
MKW
3999
14
OL
The input value is too large for the selected range.
13
12
Indicates negative reading.
A : Amperes (amps). The unit of current.
V : Volts. The unit of voltage.
A
V
11
12
9) RANGE or REL button :
5) Data Hold button :
Press it once to hold the measured value and store the value in memory. Press again to release the hold fuction.
6) Rotary switch :
For selecting of desired range, and exiting from auto-power off mode.
7) COM socket :
Connet to negative lead (black test lead) for voltage, frequency, resistance, Capacitance, diode testing.
8) "VW Hz" socket :
Connect to positive lead (red test lead) for voltage, frequency, resistance, Capacitance, continuity, and diode measurement.
When switch at DCV, ACV or W position, press this button to enter the manual range mode. Keep this button for more than 2 seconds, return to auto-range state. When switch at DCA position, push this button to set the LCD reading to zero before measuring DCA current.
11
MEASUREMENT OPERATION
A) AC Current Measurement
W
+
DC/AC CLAMP METER
2781
HOLD
Hz/DUTY SELECT
OFF
A
1000
400
400
1000
A
~
V
~
Hz RANGE REL
600V
600V
~ MAX CAT III
COM
+
A
V
OFF
CAP
!
Warning
To avoid electric shock, never measure current
while the test leads are inserted into the input
terminals and disconnect test leads and tested
circuit connection. Never attempt an in-circuit current measure- ment
where the open-circuit voltage between the circuit
and the ground is greater than 60V Use proper function and range for the measurement.
13
!
conductor at each time.
!
When current measurement has been completed, disconnect the connection between the conductor under test and the jaw, and remove the conductor away from the transformer jaw of the Meter.
!
Place the switch in 1000A range if the measured current is unknown.
To obtain accurate reading, measure only one
Caution
The measurement range of current are : 400A / 1000A To measure current, connect the meter as follows :
1. Set the function / range switch to the A range.(400A or 1000A)
2. Clamp the sensor jaw around one of the conductors under test. Make sure that the clamp jaw is perfectly closed.
3. Read the displayed value. The measured value is shown on the display, it is
an effective value of sine wave (mean value response).
B) DC Current Measurement
W
+
DC/AC CLAMP METER
2781
HOLD
Hz/DUTY
SELECT
OFF
A
1000
400
400
1000
A
~
V
~
Hz
RANGE
REL
600V
600V
~ MAX CAT III
COM
+
A
V
OFF
CAP
14
!
Warning Make certain that all test leads are disconnected from the meter terminals.
1. Set the function / range switch to the A range. (400A or 1000A)
2. Check that the display reads 00.01. If not, then press the DCA Zero button to bring the display reading to zero.
3. Clamp the sensor jaw around one of the conductors under test. Make sure that the clamp jaw is perfectly closed
4. Read the displayed value.
!
conductor at each time.
!
When current measurement has been completed, disconnect the connection between the conductor under test and the jaw, and remove the conductor away from the transformer jaw of the Meter.
!
Place the switch in 1000A range if the measured current is unknown.
To obtain accurate reading, measure only one
Caution
Red Black
~
+
C) AC Voltage Measurement
15
!If the display indicates " I " it is required to remove
the test leads from the circuit under test as the selected range is overloaded.
To avoid harms to you or damages to the Meter from electric shock, do not attempt to measure voltage higher than 1000V DC & 750V AC although readings may be obtained
To measure AC Voltage range are : 4V, 40V, 400V, 750V.
1. Set the function/range switch to the ~V range.
2. Connect the black and red test leads to the COM and + terminals respectively.
3. Connect the test leads to the circuit being measured and read the displayed value.
!
Warning
Caution
!In each ran ge, the Meter has an input
impedance of 10 MW. This loading effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10kW, the error is negligible (0.1 or less).
!When AC voltage measurement has been
completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminal.
16
E. Measuring Resistance
!
Red Black
+
!
If the display indicates " I ", it is required to remove
the leads from the circuit under test as the selected range is overloaded.
!
In each range, the Meter has an input impedance of 10MW This loading effect can cause the
measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10kW, the error is negligible (0.1 or less).
Caution
!
When DC voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test and
remove testing leads from the input terminals.
To avoid damages to the Meter or to the devices und er t est, disconnect circuit pow er a nd discharge all the high-voltage capacitors before measuring resistance.
Warning
18
To avoid harms to you or damages to the Meter from electric shock, do not attempt to measure voltage higher than 1000V AC& 750V DC, although reading may be obtained.
The DC Voltage ranges is : 4V, 40V, 400V, 1000V To measure DC Voltage, connect the Meter as follows :
1. Set the function switch to the V range.
2. Connect the black and red test leads to the
COM and + terminals respectively.
3. Connect the test leads across with the object being measured
The measured value is shown on the display.
!
~
Red Black
+
D) DC Voltage Measurement
WARNING
17
The resistance ranges are :
400.0W, 4.000kW, 40.00kW, 400kW, 4.000MW and
40.00MW
To measure resistance, connect the Meter as follows :
1. Insert the red test lead into the VW and the black test lead into the COM terminal.
2. Set the rotary switch to W ; resistance measurement (W) is default or press SELECT button to select resistance measurement mode.
3. Connect the test leads across with the resistance
being measured.
The measured value is shown on the display.
Note :
The test leads can add 0.1W to 0.3W of error to resistance measurement.
To obtain precision reading in low-resistance, in 400W range, short-circuit the input terminals before hand and record the reading obtained (call this reading as x). (x) is the additional resistance from the test lead. Then use the equation.
Measured resistance value (y) - (x) = Precision reading of resistance.
terminal
!
For high-resistance measurement (>1MW), it is normal for the meter to take several seconds to obtain a stable reading.
!
The LCD displays OL indicating open-circuit or the tested resistor value is higher than the maximum range of the meter.
!
Resistance measurement is default to auto range mode.
!
Remove the objects being tested from the circui t whe n me as uring, to obtai n mo re accurate result.
19
The Capacitance function has 6 measurement ranges : 4nF, 40nF, 400nF, 4mF, 40mF, 100mF
To measure Capacitance connect the Meter as follows :
!
Whe n resis tance measur ement has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals.
Using Resistance measurement function in a live circuit will produce false results and may damage the instrument. In many cases the suspect component must be disconnected from the circuit to obtain an accurate reading.
Caution
Never connect high voltage to the input sockets with the switch in resistance range.
1)
2)
F. Capacitance measurement :
+
Red Black
To avoid damages to the Meter or to the equipment under test, disconnect circuit power and discharge all hig h - v o l t a g e ca p a citors befor e measu r i n g capacitance. Use the DC voltage function to confirm that the capacitor is discharged. Never attempt to input over 60V in DC or 30V rms in AC to avoid personal injury.
!
Warning
20
!
To avoid damages to the Meter or to the devices und er t est, disconnect circuit pow er a nd discharge all the high-voltage capacitors before testing diodes.
Warning
H. Testing Diodes
Red Black
+
When the frequency signal to be tested is higher than 30 V rms, the Meter cannot guarantee accuracy of the measurement. The frequency measurement range is upto 10MHz. To measure frequency; connect the meter as follows :
1. Insert the red test lead into the VW terminal and the black test lead into the COM Terminal.
2. Set the rotary switch in the Hz range.
3. Connect the test leads across with the object being measured. The measured value is shown on the display.
Caution
When H z m e a s ur e me n t h as been completed, disconnect the connection between the testing leads and the circuit under test.
22
Caution
Discharge capacitors before trying to measure it.
Set the meter in the capacitance range, using the
FUNCTION select button.
Insert the red test lead into the + terminal and
the black test lead into the COM terminal.
The measured value is displayed on the LCD.
·
·
·
·
·
·
sockets with the switch in Capacitance range.
Using Capacitance measurement function
in a Live circuit will produce false results and may damage the instrument. In many cases the suspect component must be disconnected from the circuit to obtain an accurate reading.
Never connect high voltage to the Input
G) Frequency & Duty Cycle
V
1000
+
DC/AC
CLAMP METER
2781
HOLD
PEAK
OFF
A
1000
600
400
1000
400
600
600
A
V~~
4KW
Hz
DCA ZERO
600V
600V
~ MAX CAT III
COM
+
A
Red Black
~
Warning :
To avoid harm to you or damages to the Meter, do not attempt to measure voltages higher than 60V in DC or 30V rms in AC although reading may be obtained.
!
21
Caution
!
In a circuit, a good diode should still produce a forward voltage drop reading of 0.5V to 0.8; however, the reverse voltage drop rea ding can vary depending on the resistance of the pathways between the probe tips.
!
Connect the test leads to the proper terminals as said above to avoid error display.
!
The LCD will display OL indicating either open circuit or wrong polarity connection.
!
The unit of diode is volt (V), displaying the forward voltage drop readings.
!
Remove the object being tested from the circuit when measuring, to obtain a more accurate result.
!
When dio de tes ti ng has been c ompleted, disco n n e c t the c o n n e c t i o n bet w e e n the testing leads and the circuit under test and remove testing leads from the input terminals.
Use the diode test to check diodes, transistors and other semiconductor devices. The diode test sends a current through the semiconductor junction, then measure the voltage drop across the junction. A good silicon junction drops between 0.5V and 0.8V. To test the diode out of a circuit, connect the Meter as follows :
1. Connect red test lead to the "+" terminal and
black test lead to the "COM" terminal.
2. Set switch to the diode test position " ".
3. Connect the red test to the anode side and black
test lead to the cathode side of the diode being tested.
4. Read forward voltage (V) value on display.
23
To test for continuity, connect the Meter as follows :
1. Insert the red test lead into the VW terminal and the black test lead into the COM terminal.
2. Set the rotary switch to W and press S E L E CT b u t to n t o se l e ct c o n t in u it y measurement mode.
3. The buzzer sounds if the resistance of a circuit under test is less than 50W.
Note
!
The LCD displays OL indicating the circuit being tested is open.
!
When continuity testing has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from input terminal.
I) Testing for continuity (see figure 10)
Warning
To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring continuity.
Red Black
+
24
To replace the battery :
1. Turn the Meter off and remove all the connections from the input terminals.
2. Turn the Meter’s case top down.
3. Remove the screw from the bottom case,
and separate t he bot to m ca se fro m th e front case .
4. Remove the old battery from the battery
compartment.
5. Replace the battery with 2pcs of new 1.5V (AAA)
battery.
6. R ejoin the case bott o m an d th e fro n t
case, and reinstall the screw.
!
Warning
- +
To avoid false readings, which could lead to
possible electric shock or personal injury, replace
t h e b a tt e ry a s s oo n a s t he b a tt er y
indicator “ ” appears.
Make sure the transformer jaw and the test leads
are disconnected from the circuit being tested
before opening the case bottom.
B. Replacing the Battery
26
!
Sleep Mode
To preserve battery life, the Meter automatically turns off if you do not turn the rotary switch or press any button for around 15 minutes.
The meter can be activated by turning the rotary switch or pressing any button.
MAINTENANCE
This section provides basic maintenance information
including battery replacement instruction.
25
Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevent calibration, performance test and service information.
To avoid electrical shock or damage to the Meter, do not get water inside the case.
A. General Service.
Warning
! Periodically wipe the case with a damp cloth and mild
detergent. Do not use abrasives or solvents.
! To clean the terminals with cotton bar with detergent
as dirt or moisture in the terminals can affect readings.
! Turn the Meter power off when it is not in use. ! Take out the battery when it is not using for a long
time.
! Do not use or store the Meter in a place of humidity,
high temperature, explosive, inflammable and strong
magnetic field.
27
MUMBAI
TEST CERTIFICATE
DIGITAL CLAMPMETER
This Test Certificate warrantees that the product has been inspected and tested in accordance with the published specifications.
The instrument has been calibrated by using equipment which has already been calibrated to standards traceable to national standards.
MODEL NO. ___________
KM 2781
SERIAL NO. ___________
DATE: ___________
ISO 9001
REGISTERED
QC
PASS
KUSAM-MECO
28
WARRANTY
Each “KUSAM-MECO” product is warranted to be free from defects in material and workmanship under normal use & service. The warranty period is one year (12 months) and begins from the date of despatch of goods. In case any defect occurs in functioning of the instrument, under proper use, within the warranty period, the same will be rectified by us free of charges, provided the to and fro freight charges are borne by you.
This warranty extends only to the original buyer or end­user customer of a “KUSAM-MECO” authorized dealer.
This warranty does not apply for damaged Ic’s, fuses, burnt PCB's, disposable batteries, carrying case, test leads, or to any product which in “KUSAM-MECO’s” opinion, has been misused, altered, neglected, contaminated or damaged by accident or abnormal conditions of operation or handling.
“KUSAM-MECO” authorized dealer shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of “KUSAM-MECO”.
“KUSAM-MECO’s” warranty obligation is limited, at option, free of charge repair, or replacement of a defective product which is returned to a “KUSAM­MECO” authorized service center within the warranty period.