Krypton Multi Installation and Set Up Guide
Kuntze Instruments
3316 Leechburg Road | Lower Burrell | Pennsylvania | 15068
724-339-8510 | support@kuntze.com | www.kuntzeinstruments.com
Cleaning Sensors
IMPORTANT: Sensors must be cleaned by hand before being
inserted into their respective chamber!
1. How to clean pH sensor:
- Remove sensor from assembly and cable
- Wipe off any pollution with paper tissue, rinse with water
- NEVER use abrasive substances on glass bulb!
- Use mild acid (dilute HCl) to remove stains, use alcohol to
remove fats
- Rinse with water and reinstall
2. How to clean DIS sensor:
- Remove sensor from assembly and cable
- Wipe off any pollution with paper tissue, rinse with water
- Clean metal bands with powder detergent (ex. Comet), wipe
gently
- Rinse with water and reinstall
Calibrating Sensors
pH Sensor
1. Menu > Cal > Cal pH
2. Calibrate sensor with pH 7 buffer solution followed by pH
4 buffer solution
3. Save results
DIS Sensor
1. IMPORTANT: Clean sensor before proceeding with
calibration!
2. Using water sample taken from Sampling Reference
Valve, perform DPD test
3. Menu > Cal > Cal DIS > enter DPD reference value > Ok
> Save
Inserting Sensors
1. Insert pH sensor into the Flow Cell Chamber for the pH
Sensor (7) and screw on the pH cable
2. Insert Disinfection sensor into the Flow Cell Chamber for
the Disinfection Sensor (9) and screw on the Disinfection
Cable
Check Your Flow
Make sure the Water Sampling Reference Valve is in the closed
position.
System needs proper flow to operate. “Digital input 1 – no
water” will show on display screen if flow is not strong enough.
Flow minimum: 5 psi Flow maximum: 87 psi
General Steps
1. Mount the system upright on the wall
2. Insert clean sensors
3. Power on system
4. Check flow
5. Calibrate sensors
6. If needed, set user specific settings –
see operations manual
a. ASR
b. 4 – 20 mA
c. Relay Alarms
IMPORTANT: Secure lock pin after opening and
closing your system to protect from water
damage!
Power Requirements
System comes with US power cord
Line Voltage:
85-265 V AC, 50-60 Hz
Power Consumption:
10 VA
***Picture depicts system with three chambers.
1 – Neon Multi Instrument
2 – Lock Pin
3 – Argon Stabiflow
4 – FTG Temperature Cable
5
– Optional third measurement
sensor
6 – pH Cable
7 – Flow Cell Chamber for pH
Sensor
9 – Flow Cell Chamber
for Disinfection
Sensor
10 – Water Inflow Valve
11 – Water Outflow
Valve
12 – Water Sampling
Reference Valve
13 – Mounting Board

Krypton Multi Installation and Set Up Guide
Kuntze Instruments
3316 Leechburg Road | Lower Burrell | Pennsylvania | 15068
724-339-8510 | support@kuntze.com | www.kuntzeinstruments.com
All-purpose pH sensor: 1mm ceramic junction, Tepox gel, S8 plug (swivel PG 13.5), 120mm
Sensor: 2 gold rings, ceramic junction, Tepox gel, M12 plug (swivel PG 13.5), for Free Chlorine,
Chlorine dioxide, and Ozone
Flow monitor, temperature sensor and ground for Argon Stabiflow
Connection cable for pH or ORP sensor
Connection cable for disinfection sensor; Type of cable: 5-core, 5 x 0.25 Somme, screen; Cable
length: 0.8m; Type of plug: M12 (female, other end with wire sleeves; Temperature range: -20…+70C
To see exploded view of Argon Stabiflow, open manual to section 20.3.3 Operations and Maintenance
Inlet, outlet and sample cock
3x ball valve 1/4"
2x hose attachment 1/4" 6/8 1x hose barb 1/4"
3x O-ring 6x3 FKM 3x O-ring 11x3 FKM
Membrane springs and valve pin, 1 membrane, 1 spring pressure side, 1 spring chamber side, 1 valve
pin
Filter 5x stainless Steel 500um filter 5x o-ring FKM 26x1.5
1x o-ring FKM 12x2; 1x o-ring FKM 73x2
1x o-ring FKM 26x2; 1x o-ring FKM 26x1,5
1x o-ring FKM 5x1,8; 3x o-ring FKM 6x3
1x o-ring FKM 19x2,5; 1x o-ring FKM 7,6x2,4
1x o-ring FKM 108x3; 3x o-ring FKM 11x3
Spare parts inflow; 1x hex spud; 2x O-ring 26x2; 1x O-ring 5x1.8; 1x filter support; 1x ball