K-Systems AV1, AV5 User Manual

Anti Vibration Table - AV1 and AV5
Manual part no: 33095
K-SYSTEMS
USER MANUAL
Version 05
1
AV1 User Manual
Table of contents
1 Unpacking and inspection ............................................................................................................ 3
2 General description ...................................................................................................................... 3
3 Assembling the AV1/AV5 ........................................................................................................... 4
3.1 Weight adjustment ................................................................................................................. 6
3.1.1 Removing the middle spring .......................................................................................... 7
3.1.2 Adding the Weight Plate ................................................................................................ 8
4 Trouble Shooting.......................................................................................................................... 9
5 Maintenance ............................................................................................................................... 10
6 Technical data AV1 ................................................................................................................... 10
7 Warranty – Liability issues ........................................................................................................ 11
7.1 Limited Warranty: .............................................................................................................. 11
7.2 Liability: .............................................................................................................................. 11
7.3 Replacement: ....................................................................................................................... 12
8 Technical assistance ................................................................................................................... 12
2
AV1 User Manual
1 Unpacking and inspection
Unpack the equipment. Check the packing and picking list to insure all accessories are there. Examine the packing material carefully for separately packed items. Check the table for external surface damage. Please be aware that the springs must not be touched at any time. Please use gloves when you assemble the unit.
2 General description
The K-SYSTEMS Anti Vibration tables are designed to meet the requirements of ICSI Procedures. With an isolated microscope platform and a complex anti-vibration system, the table is preventing vibrations from reaching the microscope platform and dampening unavoidable vibrations in the shortest possible time.
The inverted microscope is placed on an individually suspended floating platform. The platform must be levelled and isolated – meaning not touching – the rest of the table plate. The vibrations are reduced by friction between the platform and the springs holding the platform.
3
3 Assembling the AV1/AV5
For assembling the AV1 /AV5 please avoid touching the springs.
Place the anti vibration table in its required position on the floor. Preferably on a hard floor.
Level the table plate horizontally by adjusting the feet on the table legs. To ensure the best result, place the gauge/bulls eye on the table plate and make sure it is leveled.
Gauge
AV1 User Manual
Feet
NOTE: The table can be modified to optimally work with equipment of different weight. It is important that you know the exact weight of the equipment before proceeding. See chapter 3.1.
Put on the gloves. Remove the foam pads from the springs – without touching the springs with your bare hands.
Place the floating plate upon the springs, leave equal distance from all sides to the hole of the table plate for centering the floating plate. It is very important that there is no contact between the floating plate and the table!
Place the inverted microscope on the floating plate.