K-Patents PR-14038 Instructions Manual

Cooling cover PR-14038
Document/Revision No. IM-EN-CCPR43/1.04
E󰀨ective: April 23, 2019
Instructions
Cooling cover instrucon manual
Document/Revision No. IM-EN-CCPR43/1.04 Eecve: April 23, 2019
1 Introducon 1
1 Introducon
The cooling cover adds water cooling to a PR-43 refractometer when the refractometer temperature threatens to rise too high for its electrical components. A cooling cover is needed when the ambient temperature is over 45°C (113°F) or the refractometer internal temperature alarm has been triggered, which happens when the internal tem­perature rises over 65°C (149°F). It is recommended to use cooling cover when internal temperature rises above 60°C (140°F). The cooling cover is installed between the cover plate and the sensor cover.
Figure 1.1 Cooling cover
1.1 Disposal
When wishing to dispose of an obsolete instrument, please observe local and naonal regulaons and requirements for the disposal of electrical and electronic equipment. The cooling cover can be recycled with other metallic waste of the same type.
2 Cooling cover instrucon manual
ESD
ESD
P4-034
Wear wrist or foot grounding equipment
2 Installaon
2.1 Installing the cooling cover
The cooling cover delivery contains the following parts:
• Cooling cover
• 2 x screw M4x25 Torx DIN7895
• O-ring PR-14002-FPM
• desiccant (dryer)
• 2 x M5x20 pulling screw
Important: Before opening refractometer the cover and touching the electronic parts,
ensure that the working table and the person are grounded. Wear wrist or foot ground­ing equipment.
Figure 2.1 Grounding the work environment
Document/Revision No. IM-EN-CCPR43/1.04 Eecve: April 23, 2019