Kobold RCD Specifications

Differential Pressure Flowmeters
for Water or Water-like Liquids
measuring
monitoring
analyzing
RCD
Measuring Ranges:
0.2...0.88 to 100...600 GPM (Water)
Accuracy: ± 3% of Full Scale
Max. Pressure: 580 PSIG
1/2" to 3" Line Sizes
316L Stainless Steel
Digital Indicator, Transmitters,
and Switches Available
KOBOLD companies worldwide:
ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, EGYPT, FRANCE, GERMANY, HUNGARY, INDIA, INDONESIA, ITALY, MALAYSIA, MEXICO, NETHERLANDS, PERU, POLAND, REPUBLIC OF KOREA, ROMANIA, SINGAPORE, SPAIN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED KINGDOM, USA, VIETNAM
01/03-28-2018
KOBOLD Instruments, Inc. 1801 Parkway View Drive Pittsburgh, PA 15205
Main Ofce:
1.800.998.1020
1.412.788.4890 info@koboldusa.com www.koboldusa.com
Differential Pressure Flowmeters Model RCD
Description
The RCD series differential pressure flowmeter measures flows of water and water-like liquids using the orifice principle. The orifice, located in the flow body, generates a differential pressure which varies with the rate of flow. The differential pressure is applied to a measuring bellows. This is then linked to a Hall effect sensor, which is input to a compact or full featured electronic display and controller. This unique design, using a single measuring bellows, enables the accuracy to be unaffected by fluctuations in pressure. The meter movement is a single fulcrum lever design, with a minimal number of wear components, that is simple and reliable. It is available with brass or 316L stainless steel bodies ranging in size from 1/2" NPT to 3" NPT. The various electronic displays and control units offer a variety of options such as digital or bargraph displays, adjustable switches, and flow transmitters.
Areas of Application
Machinery Manufacturing
Chemical Industry
Heavy Industry
Process Equipment
Housing
Gehäuse
Flow
Durchfluss
Force
P
1
P
2
P
1
P
2
V
V
1
2
Bewegung
Magnet
Magnet
Bellows
Faltenbalg
Technical Details
Measuring Accuracy: 3 % of Full Scale Repeatability: 1 % of Full Scale Process Temperature: 32...176 °F Ambient Temperature: Max. 176 °F Max. Pressure: 360 PSIG
(RCD-11 Connection: 3" NPT) 580 PSIG (All Others)
Pressure Loss: Approx. 5 PSI Minimum Pressure: 10 PSIG Protection: IP 65
Materials
Display Case: Cast Aluminum or Stainless Steel Body: RCD-x1..: Brass RCD-x2..: 316L Stainless Steel Differential Pressure Housing: RCD-x1..: Brass RCD-x2..: 316L Stainless Steel Pressure Meas. Cell: 316L Stainless Steel Orifice: 316L Stainless Steel Seals: RCD-x1..: NBR RCD-x2..: FKM
Orifice
> V1 => P
V
2
> P
1
2
Venturidüse
Display/Electronics:
Compact Electronics: RCD-C
Display: 3-Digit LED Analog Output: 4...20 mA Adjustable, Max. 500 Ω Switching Outputs: 1 (2) Semiconductor PNP or NPN, Factory Set Contact Operation: N/C / N/O Programmable Setting: Via 2 Buttons Supply: 24 VDC ± 20%, 3-wire, Approx. 100 mA Electrical Conn.: Plug Connector M12 x 1
ADI-1 Electronics: RCD-K
Display: Bar Graph and 5-Digit Digital Display Analog Output: 4...20 mA, 0-10 V
DC
2 Switching Outputs: Relay/Changeover Contact, Max. 250 VAC/5 A Resistive Load, Max. 30 VDC / 5 A Setting: Via 4 Buttons Supply: 100 ... 240 V 18 ... 30VAC / 10 ... 40 V
±
10 % or
AC
DC
Electr. Connection: Pluggable Terminal Block Via Cable Gland
2
www.koboldusa.com
No responsibility taken for errors;
subject to change without prior notice.