Kobold MAS Series, MAS-1000, MAS-2000, MAS-1100, MAS-2100 Instruction Manual

KOBOLD MAS Series
Thermal Mass
Flowmeters
March 1997
r2_12-13
Part Number:
IM-82
(412) 788-2830 · (800) 998-1020 · Fax (412) 788-4890
KOBOLD Instruments Inc
MAS Instruction Manual
The purpose of this instruction manual is to cover the following Kobold Instruments Inc products:
Model MAS-1000 Flowmeter with Po
Model MAS-2000 Flowmeter with Polyamide body and no display.
Model MAS-3000 Flowmeter with stainless steel flow body and display.
Model MAS-4000 Flowmeter with stainless steel flow body and no display.
Model MAS-1100 Flowmeter with aluminum body and display.
Model MAS-2100 Flowmeter with aluminum body and no display.
Kobold Instruments’ MAS Series Flowmeters measure the mass flow rate of gases in ranges from 0-10 standard cubic centimeters per minute (SCCM) to 0-500 standard liters per minute (SLM). For most models, accuracy is 1.5% of full scale over a wide temperature and pressure range, and time response is 2 seconds to within 2% of final flow. Certain models are rated at 1% or 5% of full scale.
lyamide body and display.
1
INTRODUCTION
1.1
Purpose
The MAS is ideal for a complete range of gas flow applications including general process control, laboratories, instrument OEM’s, gas panels, and flow calibration.
Figure 1-1
Tiltable Display is Viewed
from any Angle
Page 1
MAS Instruction Manual
CAUTION! Any application whatsoever related to humanCAUTION! Any application whatsoever related to human
CAUTION! Any application whatsoever related to human
CAUTION! Any application whatsoever related to humanCAUTION! Any application whatsoever related to human respiration must have the written consent of Koboldrespiration must have the written consent of Kobold
respiration must have the written consent of Kobold
respiration must have the written consent of Koboldrespiration must have the written consent of Kobold Instruments Inc.Instruments Inc.
Instruments Inc.
Instruments Inc.Instruments Inc.
The versatile MAS Flowmeter digitally displays the mass flow rate directly in engineering units or percent of full scale. The
display is tiltable over 180° for easy viewing and can be removed
for remote mounting on a front panel.
The Kobold MAS directly monitors the mass flow rate of the gas. This means it measures molecular flow – the measurement quality of direct concern in most applications, such as human respiration, chemical processes, combustion, and heating or cooling. No temperature or pressure corrections are required, as in the case of most other flow monitoring devices like rotameters, turbine meters, and orifice plates.
Figure 1-2
MAS Operation and Features
Direct Monitoring of Mass Flow.
No temperature or pressure corrections required.
Unexcelled Performance.
● ±
1.5% accuracy.
0.5% repeatability.
2 second response time
Tiltable Display.
9 positions provide easy viewing from any angle.
Connector.
9-Pin "D" Sub-type. Has 0-5 VDC or 4-20 mA (optional) output signal linearly proportional to mass flow rate.
Input Power Jack.
Accepts 12 to15 VDC or 24 VDC input power from Kobold power supply or customer supplied power.
Inlet Screen.
Filters out particulates. Easily removable.
Digital Display.
Gives mass flow rate in engineering units.
Zero and Span Potentiometers.
Adjustable from outside of enclosure.
Patented Sensor Tube.
Straight, large-I.D. sensor tube is easy to clean.
Patented Laminar Flow Bypass.
Provides a variety of flow ranges. Unique cutouts make range changes easy.
Flow Body.
Corrosion resistant plastic or stainless steel.
FKM
rings standard.
"O"-
Inlet/Outlet Fittings.
Available with Female NPT or Swagelok® style compression fittings.
Page 2
Mounting Holes.
The MAS mounts in any position for convenience of installation.
The MAS Flowmeter is a transducer requiring a 12-15 VDC (24 VDC optional) external power source. A 0-5 VDC or optional 4-20 mA output singal linearly proportional to gas mass flow rate is provided for recording, data-logging, or control. A 9-Pin “D” sub­connector is provided for power input and signal output. The MAS is available in several basic configurations with either (female) or
1
/4 to 1/2 inch O.D. tube compression inlet/outlet fittings,
1
/4 inch NPT
with or without the digital display, or the optional power supply.
Gas enters the MAS flow body and divides into two flow paths. Most of the flow goes through the laminar flow bypass. This creates a pressure drop that forces a small fraction of the flow through the sensor tube.
TWO COILS
MAS Instruction Manual
1.2
Principle of
Operation
SENSOR TUBE
m
m
1
P
1
m
2
P
2
LAMINAR FLOW BYPASS
The patented* straight sensor tube is mounted parallel to the bypass flow path. Since both paths are perfectly laminar, the ratio of the total flow (
˙
m
) to the sensed flow (
˙
m
) is constant. Two
1
resistance temperature detector (RTD) coils around the sensor tube direct a constant amount of heat into the gas stream.
RTD COILS
, T
R
1
1
·
m
1
R2, T
2
H
0
Figure 1-3
Two Flow Paths
CONSTANT HEAT, H
FIRST LAW OF THERMODYNAMICS
(HEAT IN = HEAT OUT) H = m·
*U.S. Patent No. 4,487,062.
(T2 – T1) + H
1 Cp
H–H
m· =
Cp∆T
Figure 1-4
Measuring Sensor Flow
0
0
Page 3
MAS Instruction Manual
A
A
“GATES”
SECT. A-A
Figure 1-5
Changing Flow Ranges is Easy with MAS’s Patented Bypass
In actual operation, the gas mass flow carries heat from the upstream coil to the downstream coil. The resulting temperature difference T
is detected by the RTD coils and gives the output
2–T1
signal. Since the molecules of the gas carry away the heat, the output signal is directly and linearly proportional to gas mass flow.
MAS's patented laminar flow bypass makes changing of flow ranges easy with the proper calibration facilities. Each of the two bypasses in the optional Laminar Flow Bypass Set has a combination of rectangular slots along its circumference as shown in Figure 1-5 below.
1.3 Specifications
To change the flow range of your MAS, follow the instructions provided with the Kobold Model EL Laminar Flow Bypass Set and cut away the “gate(s)” leading to the right combination of laminar flow paths in one of the two bypasses. This procedure requires proper calibration facilities and minimal skill in electronics.
FLOW RANGES _________________________________________________________________
_________________________________________________________________
_________________________________________________________________
Flow ranges specified are for an equivalent flow of nitrogen at 760 mm
Hg and 21°C (70°F). Other ranges are available.
GASES
Most gases; check compatibility with wetted materials; specify when ordering.
Code SCCM Code S LM
01 0-10 07 0-1 02 0-20 08 0-2 03 0-50 09 0-5 04 0-100 10 0-10 05 0-200 11 0-20 06 0-500 12 0-30
13 0-40
DISPLAY
Page 4
31/2 digit LCD (0.5 inches tall); removable for remote mounting.
OUTPUT SIGNAL
Linear 0-5 VDC standard, 1000 ohms min. load impedance; 4-20 mA optional, 50 to 500 ohms loop resistance.
POWER REQUIREMENTS
12-15 VDC nominal; 100 mA max (24 VDC optional, specify when ordering).
ACCURACY
±1.5% of full scale including linearity over 15 to 25 °C and 5 to 60
psia (0.35 to 4.2 kg/cm
2
); with special calibration ±1% of full scale
accuracy at a specific temperature and pressure is available.
REPEATABILITY
±0.5% of full scale.
TEMPERATURE COEFFICIENT
0.15% of full scale per °C, or better.
MAS Instruction Manual
PRESSURE COEFFICIENT
0.01% of full scale per psi (0.07 kg/cm2), or better.
RESPONSE TIME
800 ms time constant; 2 seconds (typical) to within ±2% of final
value over 25 to 100% of full scale.
PRESSURE DROP ____________________________________________________________
Typical Maximum Pressure Drop
_____________________________________________________________
Flow Rates (cm of Water)
100 SCCM .025
1 SLM .454 1 0 SL M 6.00 2 0 S L M 23.83 3 0 S L M 45.60
___________________________________________________________
4 0 S L M 83.36
GAS PRESSURE
150 psi (10 kg/cm2) gauge max; 20 psi (1.4 kg/cm2) gauge optimum.
LEAK INTEGRITY
–4
1×10
SCCS of helium max to outside environment.
GAS AND AMBIENT TEMPERATURE
0 to 50 °C.
Page 5
MAS Instruction Manual
1.15
(29.21)
3.05
(77.47)
4.16
(105.66)
(WITHOUT
DISPLAY)
5.38
(136.65)
3.50 (88.90)
1
/4" NPT
6.68 (169.67)
1
/4" SWAGELOCK
®
ALL DIMENSIONS IN INCHES
0.58
(14.73)
1.16
(29.46)
0.56
(14.22)
6-32 X 0.15 DP.
(SELF TAPPING)
0.75 (19.05)
0.38
(9.65)
0.13 (3.30)
2.75
(69.85)
3.00
(76.20)
0.125 DIA. THRU PANEL-2 PLACES;
FOR TWO 4-40 FLAT HEAD SCREWS
0.25 DIA. THRU; FOR 4 WIRES
0.88 (22.35)
1.00 (25.4) PANEL MOUNTING HOLES
DIMENSIONS IN INCHES
(DIMENSIONS IN MILLIMETERS IN PARENTHESES)
SIDE VIEW
END VIEW
BOTTOM VIEW
Dimensions
WETTED MATERIALS
5% glass-filled Polyamide 6/6; 316 stainless steel; FKM “O”-
rings standard, FFKM and silicone O-rings optional.
WEIGHT
2 lb. (0.9 kg) net; 3 lb. (1.4 kg) shipping.
Page 6
MAS Instruction Manual
Many options are available for your Kobold MAS. Please consult Kobold’s current Price List for the respective prices.
V4 (Optional) 4-20 mA output signal C( ) Cable assembly, specify cable type (round or ribbon) and
length T2 1-5 Channel Power Supply, in NEMA Box, 115 VAC T4 1-5 Channel Power Supply, in NEMA Box, 230 VAC T5 Power Supply, for one channel, 110 VAC T6 Power Supply, for one channel, 230 VAC UV Upstream valve DV Downstream valve M S Replacement mass flow sensor, includes “O”-rings and
mounting flanges OV
Complete set of FKM “O”-rings T P Tip plate for electrical enclosure RD–( ) Remote display
1.4
Options
* U.S. and foreign patents pending. ® Registered trademarks:
Swagelok, VCO, VCR–Crawford Fitting Co.
Page 7
MAS Instruction Manual
2 INSTALLATION
2.1 Receipt of Your MAS
2.2 Return Shipment
When packing carton for damage incurred in shipment. If the packing carton is damaged, the local carrier should be notified at once regarding their liability. Kobold’s Customer Service Department should also be notified immediately.
Customer Service Phone Number:
For our customers other than the USA, Canada and Mexico please call the distributor from whom you purchased your MAS or call our head office in Germany for the Kobold location nearest you:
Remove the packing slip from its envelope and verify that the carton contains all parts listed. Inspect the carton and packing material thoroughly to ensure that no spare parts or accessories are mistakenly discarded. In case of shortages, please contact Customer Service at the above phone number, or they may be contacted in writing at the address listed in the next section.
Please do not return any equipment without a Return Material Authorization, which is obtained from the Customer Service Department. Information describing the problem, corrective action or work to be performed at the factory, the purchase order number that the equipment was purchased under, and the name of the person to contact must be included with the returned equipment. The use of a Return Material Authorization is for your benefit. It makes the proper service and return of your equipment much easier to accomplish. Return shipping address:
the equipment is received, carefully check the outside
USA (412) 788-2830, or Fax (412) 788-4890
Germany 49(0)61-92-29-90 or Fax 49(0)61-92-23-398
Non-North American customers may contact the KOBOLD Messring GmbH Customer Service Department Technical Assistance at:
NOTE: Equipment returned for repair that is found to be completely operational will be subject to the current “no problem found” billing rate.
Page 8
USA KOBOLD INSTRUMENTS INC.
1801 Parkway View Drive Pittsburgh, PA 15205
Europe KOBOLD MESSRING GmbH
Nordring 22-24 65719 Hofheim/Ts. Germany
CAUTION! The maximum pressure and temperature in theCAUTION! The maximum pressure and temperature in the
CAUTION! The maximum pressure and temperature in the
CAUTION! The maximum pressure and temperature in theCAUTION! The maximum pressure and temperature in the flow line in which your MAS is to be installed must notflow line in which your MAS is to be installed must not
flow line in which your MAS is to be installed must not
flow line in which your MAS is to be installed must notflow line in which your MAS is to be installed must not exceed 150 psig (10 kg/cmexceed 150 psig (10 kg/cm
exceed 150 psig (10 kg/cm
exceed 150 psig (10 kg/cmexceed 150 psig (10 kg/cm respectively.respectively.
respectively.
respectively.respectively.
22
2
22
gauge) or 150 gauge) or 150
gauge) or 150°
gauge) or 150 gauge) or 150
F (65F (65
F (65°
F (65F (65
C),C),
C),
C),C),
MAS Instruction Manual
In order to ensure a successful installation, inlet and outlet tubing or piping should be in a clean state prior to plumbing your MAS to the system. MAS is applicable to
clean gas onlyclean gas only
clean gas only because
clean gas onlyclean gas only particulates and other foreign matter may clog the sensor tube and laminar flow element over a period of time. If the gas contains particulate matter, install a high-efficiency, 50 to 100 micron, in­line filter upstream of the MAS.
Do not locate the MAS Flowmeter in areas subject to sudden temperature changes, moisture, drafts, or near equipment radiating significant amounts of heat. Allow adequate space for cable connectors and wiring. If your MAS is to be mounted in other than a horizontal position, the zero will need adjustment. See Section
5.2, R
CAUTION! Be sure the arrow on the side of the transducerCAUTION! Be sure the arrow on the side of the transducer
CAUTION! Be sure the arrow on the side of the transducer
CAUTION! Be sure the arrow on the side of the transducerCAUTION! Be sure the arrow on the side of the transducer points in the direction of flow.points in the direction of flow.
points in the direction of flow.
points in the direction of flow.points in the direction of flow.
ECALIBRATION OVER THE SAME FLOW RANGE.
The MAS may be mounted to a chassis with two 6-32 self-tapping screws. See Section 1.3, S
PECIFICATIONS, for hole dimensions.
2.3
Mechanical
Installation
Your MAS is supplied with either
1
/4 , 3/8 or 1/2 inch compression inlet and outlet fittings. These
or
1
/4 inch female NPT (standard)
fittings should not be removed unless your MAS is being cleaned
®
or calibrated for a new flow range. VCO
1
available on special order.
/4-inch pipe requires a good quality
or VCR® fittings are
paste pipe thread sealant and should be installed in the inlet and
1
outlet fittings 1
CAUTION! Over-tightening will crack the fittings andCAUTION! Over-tightening will crack the fittings and
CAUTION! Over-tightening will crack the fittings and
CAUTION! Over-tightening will crack the fittings andCAUTION! Over-tightening will crack the fittings and shift calibration.shift calibration.
shift calibration.
shift calibration.shift calibration.
/2 turns beyond hand-tight.
For the first installation of compression fittings, simply insert the tubing into the fitting. Make sure that the tubing rests firmly on the shoulder of the fitting and that the nut is hand-tight. Scribe the nut at the six o’clock position. While holding the fitting body steady
1
with a back-up wrench, tighten the nut 1
/4 turns, watching the scribe mark make one complete revolution and continue to the nine o’clock position. After this, the fitting can be reconnected by snugging with a wrench. Do not fail to use a back-up wrench or the inlet fitting may be damaged.
2.4
Plumbing
Connections
Page 9
MAS Instruction Manual
Figure 2-1
Plumbing Requirements for Model’s 1100/2100
FLOW
5.0"
(127 mm)
1
/2" NPT PIPE
(10 DIAMETERS)
EXAMPLE: Cross Section
CORRECT
1
/2" NPT pipe for
1
/2" NPT flow body
CAUTION! Do not mix or interchange parts of tube fittingsCAUTION! Do not mix or interchange parts of tube fittings
CAUTION! Do not mix or interchange parts of tube fittings
CAUTION! Do not mix or interchange parts of tube fittingsCAUTION! Do not mix or interchange parts of tube fittings made by different manufacturers.made by different manufacturers.
made by different manufacturers.
made by different manufacturers.made by different manufacturers.
INCORRECT Avoid use of compression fitting on inlet of flow body. The reduction of internal diameter of the fitting results in a “jet”, creating noise and affecting calibration. DO NOT reduce down right at the inlet with
1
/4" pipe.
2.50"
(63.5 mm)
1
/2" NPT PIPE
(5 DIAMETERS)
Finally, check the system’s entire flow path thoroughly for leaks before proceeding to Section 3, O
CAUTION! All instruments are leak-tested prior to shipping.CAUTION! All instruments are leak-tested prior to shipping.
CAUTION! All instruments are leak-tested prior to shipping.
CAUTION! All instruments are leak-tested prior to shipping.CAUTION! All instruments are leak-tested prior to shipping. To check your installation, test the fittings only. Do not useTo check your installation, test the fittings only. Do not use
To check your installation, test the fittings only. Do not use
To check your installation, test the fittings only. Do not useTo check your installation, test the fittings only. Do not use liquid leak detectors such as Snoop® to search for leaksliquid leak detectors such as Snoop® to search for leaks
liquid leak detectors such as Snoop® to search for leaks
liquid leak detectors such as Snoop® to search for leaksliquid leak detectors such as Snoop® to search for leaks inside or outside the MAS. Instead, monitor pressureinside or outside the MAS. Instead, monitor pressure
inside or outside the MAS. Instead, monitor pressure
inside or outside the MAS. Instead, monitor pressureinside or outside the MAS. Instead, monitor pressure decay.decay.
decay.
decay.decay.
Page 10
PERATION.
MAS Instruction Manual
MAS flowmeters require a single +12 to +15 VDC power supply capable of providing a minimum current of 100 mA. The MAS can also be configured for +24 VDC power at 100 mA.
Operating power input is via either the DC power jack or the 9-pin “D” connector on the side of the enclosure. Kobold offers the Model MAS-5000 single channel power supplies for single transducer applications and the Model MAS-5100 for powering up to eight transducers through the "D" connector.
CAUTION! Do not supply + DC power at the “D” connectorCAUTION! Do not supply + DC power at the “D” connector
CAUTION! Do not supply + DC power at the “D” connector
CAUTION! Do not supply + DC power at the “D” connectorCAUTION! Do not supply + DC power at the “D” connector while using an MAS-5000 power supply at the DC powerwhile using an MAS-5000 power supply at the DC power
while using an MAS-5000 power supply at the DC power
while using an MAS-5000 power supply at the DC powerwhile using an MAS-5000 power supply at the DC power jack. Do not plug power connector into DB9 connector.jack. Do not plug power connector into DB9 connector.
jack. Do not plug power connector into DB9 connector.
jack. Do not plug power connector into DB9 connector.jack. Do not plug power connector into DB9 connector. Damage to electronics will result.Damage to electronics will result.
Damage to electronics will result.
Damage to electronics will result.Damage to electronics will result.
The standard MAS is provided with a 9-pin “D” sub type connector located on the side of the MAS enclosure as shown in Figure 2-2. The pin numbers for this “D” connector are also shown in Figure 2­2, and the pin assignments are given in Table 2-1 on the next page. The output signal is obtained from the 9-pin “D” connector. A 0 to 5 VDC output signal linearly proportional to gas mass flow rate is standard. A 4-20 mA current loop signal is optionally available. The mating connector is included.
2.5
Electrical
Connections
2.5.1
9-Pin “D”-Connector
Pin Assignments
When the MAS is configured for a remote display, connections are made via the 9-pin “D” connector. Power connections for the display and transducer are shared in this mode unless the accessory MAS-5000 power supply is used.
“D” CONNECTOR
7 8 9
6
1 2 3 4 5
DC POWER JACK
Figure 2-2
“D” Connector and DC
Power Jack Location and
Number Assignments
Page 11
MAS Instruction Manual
Figure 2-3
Printed Circuit Board Input/Output Solder Pad Assignments (wires shown are typical)
Table 2-1
The following connection points can be made through the “D” connector or, in OEM applications, made through the circuit board solder pad connections.
The display pad connections are shown for applications requiring remote mounting of the digital display. The letters appear on the display circuit board and are shown in Figure 2-3. ___________________________________________________________
____________________________________________________________
Pin No. Function Display Pad
1 No Connection N/A 2 Signal Common N/A 3 0 to +5 VDC Flow Signal N/A 4 + Power Supply (12 or 24 VDC)
*1, *2
(A) 5 Remote Display Flow Signal (D) 6 Remote Display Reference (C) 7 Power Common (B) 8 4 to 20 mA Return (Common) N/A
____________________________________________________________
9 4 to 20 mA Output N/A
NOTE: the numbers on the connector plug may not agree with the numbering system as it appears on our Figure 2-2 (on previous page). It is important to make sure that the proper wires are in the proper location rather than the proper number. Most connectors utilize a standard numbering scheme but there are a few that do not.
The remote display connects through the 9-pin “D” connector only. The pads A-D in the top right of the main circuit board are for integral display mounting only.
*1 Power supply voltage must be specified at time of order. Operating a 12 VDC meter at
*2 Do not supply + DC power at the “D” connector while using a power supply at the DC
Page 12
24 VDC will cause damage. Running a 24 VDC meter at 12 VDC will result in faulty operation.
power jack. Both supplies may be damaged.
MAS Instruction Manual
As shown in Figure 2-4, the digital display on your MAS may be removed and mounted remotely on a front panel. The MAS transducer is mounted at any convenient location in your system.
Remote installation is not recommended unless you possess the proper soldering tools and skills to accomplish the job. Remote installation of the digital display is accomplished by following these steps:
1. Remove the digital display from your MAS by following the procedure described in Section 2.5.3, OEM E
TIONS.
LECTRICAL C ONNEC-
2. Gain access to the printed circuit board (PCB) also by following the procedure in Section 2.5.3, OEM E
LECTRICAL C ONNECTIONS.
3. Very carefully unsolder the four short leads from the PCB to the display.
4. Gain access to the display circuit board by removing the two snap rings (No. 36) and opening the “clam shell” display enclo­sure. Remove the four short wires by carefully unsoldering. Solder the longer four wires (26AWG, 100 ft. max shielded cable recommended) to the digital display and solder the other ends to the appropriate pin numbers of the 9-pin “D” connector as shown in Figure 2-3 and Table 2-1 (on previous page).
2.5.2
Remote Installation of
Digital Display
5. Reassemble the MAS, replacing the display base (No. 35) with the Model TP Plain Top Cover (No. 28).
6. Make the necessary plumbing connections to the unit.
7. Mount the digital display as shown in Figure 2-4 referring to the panel mounting hole dimensions in Section 1.3, S
PECIFICATIONS.
Figure 2-4
Remote Installation of
Digital Display
Page 13
MAS Instruction Manual
2.5.3 OEM Electrical Connections
The OEM version MAS flowmeter has an electrical port (hole) on its side for electrical input/outputs. This port is just above the input power jack (see Figure 2-2). Wires entering the MAS via this port are soldered to the printed circuit board as shown in Figure 2-
3. The solder pad assignments are given in Table 2-1.
How to Gain Access to the MAS’s Printed Circuit Board (PCB): To gain access to the PCB to make the solder connections, please refer to the exploded view in Appendix A and follow these simple steps.
1. If your MAS has the digital display:
(a) First, remove the display by carefully rotating the display until it hits the top plate. Slowly continue to rotate until this lever arm action snaps out the two yokes holding the display (Nos. 30). Use extra caution during this operation as excessive force will break the delicate wire connections. Carefully move the display assembly to expose the two screws securing the display base (No. 35). Do not exert excessive force on the display while rotating as doing so could crack the LCD display.
(b) Next, remove the two screws (Nos. 21) in the display base (No. 35) and the two screws (also Nos. 21) in the back of the enclosure (No. 10).
2.5.4 Using Kobold’s Single, Dual, and Flo-Box™ Electronics
(c) The top, front, and back sides of the enclosure can now be removed, (the front slides out towards you and perpendicular to the flow path) exposing the PCB.
2. If your MAS does not have the digital display:
(a) Remove the label (No. 29) from the plain top cover (No. 28) to expose the two screws (Nos. 21).
(b) Then follow steps 1(b) and 1(c) above.
To reassemble, just reverse the process.
For applications requiring flow totalization or alarms simply order the optional Model EC-( ) cable to connect to the rear panel of Kobold Instruments’ Single-Channel, Dual-Channel, or Flo­Box (1-5 Channels) Electronics. The Kobold electronics will provide you with a selectable digital readout for each channel, input power, high, low, or window alarms, and optional flow totalization.
Page 14
MAS Instruction Manual
Quick operating instructions are given on the first page of this manual.
Following are important notes and comments regarding the Quick Operating Instructions.
NOTE 1 R
EFERENCING THE FLOW RATE TO OTHER TEMPERATURE AND
PRESSURE CONDITIONS: The gas flow rate output of your MAS is referenced to “standard”
conditions of 21°C (70°F) and 760 mm of mercury (1 atmosphere),
unless you have specified otherwise. Be sure you know the reference conditions of your MAS, because it may make a difference if you are comparing the output of the MAS with another type of flow meter. For example, the output reading of the MAS will be approximately
7% lower if it is referenced to 0°C rather than 21°C. Appendix B
shows how to convert the flow rate output of your MAS to other standard conditions and how to find the flow rate referenced to the actual temperature and pressure conditions in the pipe where your MAS is located.
NOTE 2 A
CCURACY:
The standard accuracy of the MAS is ±1.5% of full scale. The ±1.5%
of full scale accuracy means the 0-5 VDC output signal is accurate
to within ±0.1 VDC, and the 4-20 mA output is accurate to within ±0.4 mA. This means, for example, that the output signal for zero flow can be as much as ± 0.1 VDC or ±0.4 mA. Please note if you get
an output signal at zero flow (as long as it is within either of these two ranges) it does not mean your MAS is malfunctioning. For MAS’s with the digital readout, the accuracy is simply 1.5% times the full scale flow rate listed on your MAS’s front label. For example, if your full scale is 10 SLM, the digital readout will be
accurate to ±0.2 SLM, and the reading at zero flow may be as much as ±0.2 SLM and still be within the stated accuracy specification.
3
OPERATION
3.1
Quick Operating
Instructions
3.2
Notes to Operating
Instructions
NOTE 3 O
VERRANGING:
If the flow rate exceeds the full scale range listed on your MAS’s front label, the output signal and digital display (if you have it) will read a higher value.
The MAS has not been calibrated for overranged flows and probably will be both non-linear and inaccurate. If the supply voltage is only 12 VDC, the overranged reading may only exceed the full scale reading by 10% maximum. If the supply voltage is higher, such as with the 24 VDC option, then the output can exceed full scale by as much as 50%, or more. If you have the digital display, the display can not exceed the four digits 1999. If the flow rate exceeds 1999 the
Page 15
MAS Instruction Manual
right most digits will blank and only the left-hand “1” will appear on the display.
Overrange conditions are indicated by the display and/or output going to a high level, above the full scale range. After the overrange condition has been removed, it may take several seconds for the MAS to recover and resume normal operation. This will not harm the instrument.
NOTE 4 O
PTIONAL 4-20 mA OUTPUT S IGNAL:
The 4-20 mA output signal current flows from the 4-20 mA output pin on the “D” connector through the load (50 to 500 ohms) to ground (see Section 2.5.1, 9-P
IN “D” CONNECTOR P IN A SSIGNMENTS).
Figures 3-1 and 3-2 illustrate single and multiple installations with current loop outputs.
NOTE 5 Z
ERO AND SPAN ADJUSTMENTS:
The zero and span potentiometers are accessed through marked ports on the right side of your MAS. If your zero output is more than
±1.5% of full scale, you may adjust the zero potentiometer when
you are absolutely certain that you have zero flow.
Since the output does not indicate negative numbers, it is necessary to adjust down from a slightly positive reading. Slowly rotate the zero pot clockwise until a positive reading is indicated. To complete the zero adjustment, slowly turn the pot counterclockwise until zero is reached.
Normally, span adjustments are not made unless you areNormally, span adjustments are not made unless you are
Normally, span adjustments are not made unless you are
Normally, span adjustments are not made unless you areNormally, span adjustments are not made unless you are calibrating your MAS, as described in Section 5. calibrating your MAS, as described in Section 5.
calibrating your MAS, as described in Section 5. The span
calibrating your MAS, as described in Section 5. calibrating your MAS, as described in Section 5. adjustment should not be used unless you have a known precise non-zero flow rate that you wish to match.
NOTE 6 A Unless specified otherwise, your MAS has been calibrated for
installation with the flow direction in the horizontal plane (±15°)
with the enclosure facing upward. If your actual installation orientation is different, you will have to make a small zero adjustment.
Page 16
TTITUDE:
Loading...
+ 37 hidden pages