Kobold KAL-L Specifications

Thermal Air Flow Switch
measuring
monitoring
analyzing
KAL-L
Negligible Pressure Loss
No Moving Parts
Adjustable Response Time
Compensated for Temperature
Easy to Install
KOBOLD companies worldwide:
ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, EGYPT, FRANCE, GERMANY, HUNGARY, INDIA, INDONESIA, ITALY, MALAYSIA, MEXICO, NETHERLANDS, PERU, POLAND, REPUBLIC OF KOREA, ROMANIA, SINGAPORE, SPAIN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED KINGDOM, USA, VIETNAM
01/05- 2014
KOBOLD Instruments, Inc. 1801 Parkway View Drive Pittsburgh, PA 15205
Main Ofce:
1.800.998.1020
Thermal Air Flow Switch Model KAL-L
Description
The KOBOLD KAL-L flow switch allows rapid detection of flow rate changes in non-hazardous gases. Fast reaction times are guaranteed by the electronic design. Through use of two RTD's, one to sense flow and the other to detect ambient temperature, the switch can compensate for thermal changes in its surroundings. This “smart” behavior minimizes erroneous switching due to changes caused by things such as the weather. Almost any conceivable pipe size can be accommodated by the insertion style design. Knowledge of the flow velocity in your system is all that is needed to make the KAL-L work. Typical application areas include: air conditioning systems, ventilation systems, and other non-hazardous gas flow monitoring systems.
Specifications
Power Supply: 24 VAC/VDC - 15%, +10% Max. Power: 4 VA Ambient Temperature: 14...140 °F Media Temperature: -10...250 °F Max. Pressure: 120 PSIG Max. Warm-up Time: 30 seconds Switching Range: 3.3...66 ft/sec at 20 °C, 14.5 PSIG
(Restricted Span for other Conditions)
Switching Accuracy: ± 10% of Reading Repeatability: ± 1% of Reading Temperature Gradient: 30 °K (°C)/min, @ 8m/s, 90 °C Response Time: 1...60 s, Adjustable Flow Rate Indication: 8-Digit LED Bargraph Switch Point Adjustment: via Potentiometer Output Indicator: 2-Color LED Switch Output: Relay, SPDT Switch Capacity: Max. 250 VAC / 1000 VA / 4A
(Option M12: Max. 24 VAC/VDC, 3A)
Protection: IP 65 Housing Material: Polyamide Sensor Material: Ni-plated Brass
Order Details (Example: KAL-8115 M12)
Electrical Connection
Dimensions (mm)
Fitting
1/2" NPT KAL-8115
2-Bolt Flange KAL-8115FL
2
Order
Number
Options
Suffix "C"
½" Conduit Hub
Suffix "M12"
Quick Disconnect w/6' Cable
www.koboldusa.com
No responsibility taken for errors;
subject to change without prior notice.
Loading...