Kobold EDM User Manual

Page 1
SAVE THESE INSTRUCTIONS
Operations Guide for
EDM Electronic Water Meters
Install your meter in-line either horizontally or vertically or at the end of the hose adjacent to the nozzle. Installation to metal connections is not recommended. Install as follows:
1. Plan to install turbine with a minimum straight pipe length as follows: – Upstream from the turbine, allow a mini-
mum straight pipe length of 10 times the internal diameter of the turbine.
TABLE OF CONTENTS
English ......................................................... 1
Español ........................................................ 6
Deutsch ...................................................... 11
Italiano ....................................................... 17
Français ..................................................... 22
ENGLISH
–Downstream from the turbine, allow a
minimum straight pipe length of 5 times the internal diameter of the turbine.
2.
For Soc Fittings use only primer and sol­vents approved for PVC gluing. For NPT Fittings wrap all connections with 3 to 4 wraps of PTFE tape. Make sure the tape does not intrude into the flow path.
3. Attach meter with arrow pointed in the direction of flow.
For NPT Fittings - Hand tighten the meter at
4. the housing ends. Do no use a wrench or similar tool to tighten. This can damage the housing.
INSTALLATION
IMPORTANT NOTICE
Use EDM Series meters with water and other chemicals compatible with wetted components (see Specifications Section). Do not use to meter fuel or chemicals. EDM Series meters measure in gallons or litres. Refer to the Cali­bration Section for details.
These meters are not legal for trade applica­tions.
EDM Series meters are very sensitive to electric noise if operated within one to two inches of some electric motors or other sources of elec­tronic noise.
09/06
Verify Meter Accuracy
Before using, check the meter’s accuracy and verify calibration.
1. Make sure there is no air in the system by starting the flow until it runs steadily. Then, stop the flow using a valve or nozzle.
2. If desired, hold down DISPLAY for 3 sec­onds to zero the meter’s Batch Total. When zeros appear, release the button.
3. Meter an exact known volume into an ac­curate container. For best results, meter with one continuous full stream.
Manual Number 92175901
1
Page 2
4. Check the readout. If the amount metered is accurate, field calibration is not neces­sary. If not, refer to the Calibration Section for further instructions.
OPERATION
Batch and Cumulative Totals
The meter maintains two totals. The Cumula­tive Total provides continuous measurement and cannot be manually reset. The Batch Total can be reset to measure flow during a single use. The Cumulative Total is labeled with TO­TAL 1 LOCKED indicating that this total is locked and cannot be manually zeroed. Batch Total is labeled with TOTAL 2.
When the Cumulative Total reaches a maxi­mum reading of 999,999, it will automatically reset to zero.
Press the DISPLAY button briefly to switch between the batch, cumulative total, and flowrate.
NOTE: Totalization counts total units without
differentiating between gallons, litres or field calibrated units.
Flowrate Feature
When this feature is activated, the word FLOW­RATE displays to the left on the bottom line.
When FLOWRATE is displayed, the numbers on the middle line reflect the rate of flow, for example, the current gallons per minute (GPM) or litres per minute (LPM).
Display FLOWRATE
To use this feature, press and release DISPLAY until FLOWRATE appears to the left of the bottom line.
Activate the Meter
Press DISPLAY briefly to display the Batch Total. Hold the DISPLAY button down for 3 seconds to reset the Batch Total to zero.
2
The meter is programmed to turn off auto­matically if not used for 4 minutes.
Factory and Field Calibration Curves
All calibration information is visible to the user as words in the upper part of the display, above the numeric digits.
All units are configured with a “factory” calibra­tion curve. Both gallons and litres are available (“GAL” or “LTR” will be displayed). Use the CALIBRATE and DISPLAY buttons to switch between gallons and litres. This curve is NOT user adjustable: the word “PRESET” is dis­played to show this. (The factory calibration is stored permanently in the computer’s memory.)
The “field” calibration curve may be set by the user, and can be changed or modified at any time using the calibration procedure described below in the Calibration Section. Totals or flowrate derived from the field calibration are visible when the field calibration setting is se­lected (“CAL B” will be visible on the top line).
Selecting a Different Calibration Setting
You can switch between GAL and LTR modes at will without “corrupting” totalizer contents. For example, the computer can totalize 10.00 gallons. If the user switches to LTR mode, the display will immediately change to “37.85 the same amount in units of litres). GAL / LTR switching also works in FLOWRATE mode.
To select a different calibration setting, first press and hold the CALIBRATE button. Con­tinue to hold it while also pressing and releasing the DISPLAY button. (You may then also re­lease the CALIBRATE button.) The flag indica­tors in the top line of the display will change to show the newly selected calibration setting. Calibration settings change in this order: GAL, LTR, CAL B, GAL, etc. While fluid is flowing, only the GAL and LTR selections may be made. However, when NO fluid flow is occurring, any setting may be selected.
Field Calibration
Factory calibration settings are custom-pro­grammed into each flowmeter during produc­tion, using water at 70°F (21°C). Readings us­ing the standard factory calibration curves may not be accurate in some situations – for ex­ample, under extreme temperature conditions.
Page 3
You can field calibrate the meter if you are using fluids other than water.
For improved accuracy under such conditions, the Kobold flow computer allows for “field” calibration, that is, user entry of custom calibra­tion parameters. A “single point” calibration may yield acceptable accuracy in the middle of the flow range, but 5 or more calibration points may yield a higher level of accuracy, especially at the lower end of the flow range. Up to 15 custom calibration points can be entered.
The use of a uniformly dependable, accurate calibration container is highly recommended for the most accurate results. Due to high flowrate, it is strongly recommended that Field Calibration be completed with a combination of volume and weight using fine resolution scales.
Before Beginning Field Calibration
For the most accurate results, dispense at a flowrate which best simulates your actual oper­ating conditions. Avoid “dribbling” more fluid or repeatedly starting and stopping the flow – these actions will result in less accurate calibra­tions.
EDM Series Meters
1/2 inch meter 1 GPM (3.8 LPM) 3/4 inch meter 2 GPM (7.5 LPM) 1 inch meter 5 GPM (18.8 LPM) 1-1/2 inch meter 10 GPM (37.85 LPM) 2 inch meter 20 GPM (75 LPM)
For best results, the meter should be installed and purged of air before field calibration.
Dispense/Display Field Calibration Procedures
1. Hold down CALIBRATE while pressing and releasing DISPLAY until the field calibra­tion curve appears (“CAL B” message will be displayed). Release both buttons.
2. To calibrate, press and hold the CALI­BRATE button. While continuing to hold CALIBRATE, also press and hold the DIS­PLAY button. Hold both buttons for about 3 seconds until you see a blinking “dd CAL” message. Once the “dd CAL” message appears, release both buttons. You are now in field calibration mode.
3. Once the buttons have been released from Step 2, the display will show the blinking message “run 01”. If you want to exit the calibration now before dispensing any fluid, go to Step 11.
4. If you want to continue with the calibration, but have not dispensed any fluid yet, make your final preparations to your pumping system, but don’t start pumping yet.
5. Start your pumping system so that fluid flows through the meter. The display will stop blinking and show the “run 01” mes­sage. Dispense into a container that allows you to judge the amount of fluid pumped. When you have pumped the desired amount (for example, 10 gallons), stop the fluid flow quickly.
6. Once the flow has stopped, briefly press and release both buttons. At this point the computer display will change to “0000.00” with the left-hand digit blinking.
7. Enter the volume (amount) of fluid that you dispensed (for example, if your 10-gallon container is full, enter “10.0” for gallons or “37.85” for litres). To enter numbers, use the CALIBRATE button to change the value of the digit that is blinking and use the DISPLAY button to shift the “blink” to the next digit.
8. Once the correct number is entered, briefly press and release both buttons. The dis­play will now change to a blinking “run 02” message. You have installed the new cal­curve point. You are ready to end calibra­tion (Step 10) or enter another new calibra­tion point (Step 9).
9. To enter another calibration point, go back and repeat Steps 3 through 8. It is possible to set up to 15 cal-curve points, and the “run ##” message will increment each time you repeat the calibration process (run 01, run 02, run 03, etc., up to run 15).
10. To end calibration, press and hold both buttons for about 3 seconds until you see the “CAL End” message. After you release the buttons the computer will resume nor­mal operations with the new cal point(s) active.
3
Page 4
11. If you HAVE NOT dispensed any fluid, you can exit calibration without changing the cal curve. If the message “run 01” is show­ing and you have not dispensed any fluid, hold both buttons for about 3 seconds until you see a “CAL End” message. After you release the buttons, the computer will re­sume normal operation and the old curve (if you entered one in the past) is still intact.
MAINTENANCE
Proper handling and care will extend the life and service of the meter.
Turbine Rotor
The meter is virtually maintenance-free. How­ever, it is important the rotor moves freely. Keep the meter clean and free of contaminants.
If the rotor does not turn freely, apply a pen­etrating lubricant on the rotor, shaft, and bear­ings. Remove any debris or deposits from the rotor using a soft brush or small probe. Be care­ful not to damage the turbine rotor or supports.
! ! ! CAUTION ! ! !
Blowing compressed air through the tur­bine assembly could damage the rotor.
5. Reseat batteries, if necessary, and posi­tion the faceplate on the turbine housing. To avoid moisture damage, make sure the O-ring is fully seated. Tighten the four screws on the faceplate.
SPECIFICATIONS
Inlet and Outlet:
EDM-8501S 1/2" Schedule 40, Soc EDM-8502S 3/4" Schedule 40, Soc EDM-8503S 1" Schedule 40, Soc EDM-8504S 1-1/2" Schedule 40, Soc EDM-8505S 2" Schedule 40, Soc EDM-8501N 1/2" NPT EDM-8502N 3/4" NPT EDM-8503N 1" NPT EDM-8504N 1-1/2" NPT EDM-8505N 2" NPT
Design Type: Turbine Wetted Components:
Housing: PVC Journal Bearings: Ceramic Shaft: Tungsten Carbide Rotor and Supports: PVDF Retaining Washer: Stainless Steel
Fitting Types: Slip - Schedule 40 Soc or NPT Max. Working Pressure: 150 PSIG @ 73°F
Battery Replacement
The meter is powered by two 3-volt lithium batteries which may be replaced while the meter is installed. When batteries are removed or lose power, the batch and cumulative totals reset to zero but the field and factory calibra­tions are retained.
If the meter display becomes dim or blank, replace the batteries as follows:
1. Remove the four Phillips-head screws from the face of the meter and lift the faceplate from the turbine.
2. Remove the old batteries and clean any corrosion from the terminals.
3. Install new batteries. Make sure the posi­tive post is in the correct position.
4. When the batteries are replaced, the face­plate will power ON. Check the display to ensure normal functions have resumed before assembling again.
4
U.S. Measurement
Unit of Measure: Gallon Flow Range:
1/2 inch 1 - 10 GPM 3/4 inch 2 - 20 GPM 1 inch 5 - 50 GPM 1-1/2 inch 10 - 100 GPM 2 inch 20 - 200 GPM
Accuracy: ± 3.0% (Accuracy can be
improved with field calibration)
Operating Temperature: Storage Temperature: –40° to +158° F Product Weight:
1/2 inch .31 lbs. 3/4 inch .38 lbs. 1 inch .51 lbs. 1-1/2 inch .81 lbs. 2 inch 1.14 lbs.
* +32° to +140° F
* Do not allow fluid to freeze inside meter.
Page 5
Dimensions - inches (W x H x L):
1/2 inch 2 x 2.5 x 4.3 3/4 inch 2 x 2.7 x 5.3 1 inch 2 x 3 x 6 1-1/2 inch 2.3 x 3.6 x 7 2 inch 2.8 x 4.1 x 7.5
Computer Kits:
12550902 1/2 inch, Computer Assy Kit 12551102 3/4 inch, Computer Assy Kit 12551302 1 inch, Computer Assy Kit 12551502 1-1/2 inch, Computer Assy Kit 12551702 2 inch, Computer Assy Kit
Metric Measurement
Unit of Measure: Litre Flow Range:
1/2 inch 3.8 - 38 LPM 3/4 inch 7.6 - 76 LPM 1 inch 19 - 190 LPM 1-1/2 inch 38 - 380 LPM 2 inch 76 - 760 LPM
Accuracy: ± 3.0% (Accuracy can be
improved with field calibration)
Operating Temperature:* 0° to +60° C Storage Temperature: –40° to +70° C Product Weight:
1/2 inch .141 kg 3/4 inch .173 kg 1 inch .232 kg 1-1/2 inch .368 kg 2 inch .518 kg
Dimensions - cm (W x H x L):
1/2 inch 5.1 x 6.4 x 10.9 3/4 inch 5.1 x 6.9 x 13.5 1 inch 5.1 x 7.6 x 15.2 1-1/2 inch 5.8 x 9.1 x 17.8 2 inch 7.1 x 10.4 x 19.1
* Do not allow fluid to freeze inside meter.
SERVICE
For warranty consideration, contact your local distributor. If you need further assistance, con­tact the Kobold Customer Service Department:
412-788-2830
You will need to:
• Provide information from the decal on your meter.
• Receive a Return Authorization number.
• Flush any fluid from the meter before ship­ping to the factory.
• If possible leave customer installed fittings or ample length of bare pipe for reinstalla­tion.
! ! ! CAUTION ! ! !
Do not return the meter without specific authority from the Kobold Customer Service Department. Due to strict regulations gov­erning transportation, handling, and disposal of hazardous or flammable liquids, Kobold will not accept meters for rework unless they are completely free of liquid residue.
PARTS
The following replacement parts and acces­sories are available for the EDM Series meters:
Part No. Description
113520-1 Battery Replacement Kit 901002-52 O-Ring 125508-01 1/2 inch, Soc Turbine Assy Kit 125508-02 1/2 inch NPT, PVC Turbine Assy Kit 125510-01 3/4 inch, Soc Turbine Assy Kit 125510-02 3/4 inch NPT, PVC Turbine Assy Kit 125512-01 1 inch, Soc Turbine Assy Kit 125512-02 1 inch NPT, PVC Turbine Assy Kit 125514-01 1-1/2 inch, Soc Turbine Assy Kit 125514-02 1-1/2 inch NPT, PVC Turbine Assy Kit 125516-01 2 inch, Soc Turbine Assy Kit 125516-02 2 inch NPT, PVC Turbine Assy Kit
WEEE DIRECTIVE
The Waste Electrical and Electronic Equipment (WEEE) directive (2002/96/EC) was approved by the European Parliament and the Council of the European Union in
2003. This symbol indicates that this product contains electrical and electronic equipment that may
include batteries, printed circuit boards, liquid crystal displays or other components that may be subject to local disposal regulations at your location. Please understand those regulations and dispose of this product in a responsible manner.
5
Page 6
Manual Number 92175901
09/06
Kobold Instruments, Inc. 1801 Parkway View Drive Pittsburgh, PA 15205 Phone: 412/ 788-2830 Fax: 412/788-4890 www.koboldusa.com
Loading...