Kobold DOG-4 Operating Instructions Manual

Page 1
Operating Instructions
for
Oscillation Flowmeter
Model: DOG-4...
Transducer and Electronic Options A/B…O/P
Page 2
DOG-4
page 2 DOG-4 K08/0818
1. Contents
1.Contents ....................................................................................................... 2
2.Note .............................................................................................................. 4
3.Information on operating the device in potentially explosive environments
(Ex, A0/D0/F0/H0/K0/N0/P0 option) ........................................................... 4
3.1 General requirements .......................................................................... 5
3.2 Startup, installation .............................................................................. 6
3.3 Application ........................................................................................... 6
3.4 Repair, maintenance (for applications in Ex areas) ............................. 7
3.1 Disposal .............................................................................................. 7
4.Instrument Inspection ................................................................................... 8
5.Regulation Use ............................................................................................. 8
6.Operating Principle ....................................................................................... 9
7.Mechanical Connection .............................................................................. 10
8.Identification (Ex version) ........................................................................... 11
9.Electrical connections ................................................................................. 12
9.1 General ............................................................................................. 12
9.2 Measuring cable in Ex areas ............................................................. 12
9.3 Connection to the measuring sensor ................................................. 12
9.4 Terminal assignment of the DOG-4… reader/transducer .................. 13
9.5 Terminal assignment of the electronic option
G0/H0/I0/K0/M0/N0/O0/P0 ................................................................ 15
9.6 Connections of DOG-4 Transmitter with KOBOLD Evaluation
Electronics ZED-X and ZOK-Zx ......................................................... 17
10.Startup ........................................................................................................ 20
11.Setup and operation of the flow controller/flow computer (option
G0/H0/I0/K0/M0/N0/L0/O0/P0) ................................................................ 20
12.Transmitter error message ......................................................................... 21
12.1Status display not illuminated ............................................................ 21
12.2OPERATING illuminated green but no output signal ......................... 21
12.3RANGE illuminated yellow................................................................. 21
12.4ERROR illuminated red ..................................................................... 21
13.Maintenance ............................................................................................... 22
13.1Replacing the sensor ......................................................................... 22
13.2Disassembly and cleaning the Flowmeter ......................................... 22
14.Technical Data ............................................................................................ 24
14.1Electronic options .............................................................................. 25
15.Order Details .............................................................................................. 29
16.Pressure Loss/Flow .................................................................................... 31
17.Dimensions ................................................................................................. 32
18.EU Declaration of Conformance ................................................................. 34
19.ATEX-Certificate ......................................................................................... 35
20.IECEx-Certificate ........................................................................................ 41
Page 3
DOG-4
DOG-4 K08/0818 page 3
Manufactured and sold by:
Kobold Messring GmbH
Nordring 22-24
D-65719 Hofheim
Tel.: +49(0)6192-2990
Fax: +49(0)6192-23398
E-Mail: info.de@kobold.com
Internet: www.kobold.com
Page 4
DOG-4
page 4 DOG-4 K08/0818
2. Note
Please read these operating instructions before unpacking and putting the unit into operation. Follow the instructions precisely as described herein. The devices are only to be used, maintained and serviced by persons familiar with these operating instructions and in accordance with local regulations applying to Health & Safety and prevention of accidents.
as per machine-guidelines 2006/42/EC
CE mark acc. ATEX-guideline 2014/34/EU When used in machines, the DOG-4 should be used only when the machines fulfil
the EC-machine guidelines.
as per PED 2014/68/EU
Model Connection PN
no dangerous gases
(diagr. 7, 2)
dangerous gases
(diagr. 6, 1)
DOG-4
DN 25 10 article 4, § 3 I DN 40 until DN 50 10 I II DN 80 until DN 150 10 I II DN 200 10 I II
Funded by the Federal Ministry of Economics and Technology based on a decision of the German Bundestag.
3. Information on operating the device in potentially explosive environments (Ex, A0/D0/F0/H0/K0/N0/P0 option)
The devices can be used in the following way:
1. DOG-4S Flowmeter: In zones 0, 1 and 2 (gas Ex, category 1G, 2G, 3G) in explosion groups IIA, IIB and IIC
2. DOG-4A Reader/Transducer Reader: Outside the EX area Here the following ambient temperatures must be adhered to
-20 °C Ta 60 °C The DOG-4 is suitable for use with gases of the explosion group IIC and temperature
class T4. Further important details can be found in the EC Type Examination Certificate.
Page 5
DOG-4
DOG-4 K08/0818 page 5
3.1 General requirements
In the event of failure to comply with this information or unauthorized tampering with the device, the manufacturer's liability will no longer apply. Moreover, the guarantee for the device and accessory parts will become void.
Comply with the information in these operating instructions and adhere to the
conditions of use and permissible data printed on each device / type plates.
Follow the generally accepted rules of technology when selecting and
operating a device.
Take appropriate measures to prevent unintentional activation or inadmissible
impairments.
The devices may be used only in the correct way and for the intended
purpose in a normal industrial environment. Use for unsuitable purposes, will render all guarantees and liability of the manufacturer void!
Ensure that only devices with ignition protection suitable for the operating
zones are installed.
All electronic operating resources connected must be suitable for the
respective use.
The operator is obligated to provide lightning protection according to local
regulations.
Page 6
DOG-4
page 6 DOG-4 K08/0818
3.2 Startup, installation
The devices are intended for installation in a higher-level system. The intervals for cleaning the operating resources (dust deposits) are specified depending on the degree of IP protection. It is extremely important to ensure that only devices with suitable ignition protection for the zones/categories are installed! It is essential that the installation regulations applicable at the national level, e.g. EN 60079-14, are adhered to during installation. Other important factors:
In the event of adverse environmental conditions, it is necessary to ensure
that the devices are accordingly protected.
Follow the operating instructions for the respective device and adhere to any
special conditions described there.
The device may only be used for the purpose for which it was intended. It is essential to avoid electrostatic charges. Any possible metal parts in the device/lines (e.g. shielding) must be
incorporated in the potential equalization PA in compliance with the user’s country regulations.
Parts that have jammed (e.g. as the result of frost or corrosion) may not be
loosened by force in hazardous atmospheres.
Operate the device only in the completely assembled state and enclosed in
undamaged housing. Operating with a damaged housing is prohibited.
At ambient temperatures of less than -5°C the connecting lines must be
securely laid.
Do not allow the outside of the device to come into contact with strongly
corrosive media.
Do not subject the system to excessive vibrations, bending or torsion. The devices may not – or only with the manufacturer's permission, and then
only using special measures – be used in systems with electric corrosion protection. Parasitic currents must not be fed in via the shielding.
Installations in Ex areas may be carried out only in compliance with the local
installation regulations.
Installation and maintenance only in ex-free atmospheres in compliance with
the user’s national regulations. Additional precautionary measures must be taken if there is a possibility that hydrogen sulfide, ethylene oxide and/or carbon monoxide could be present. These compounds have very low ignition power! Only non-arcing tools may be used for these compounds as well as all compounds included in explosion group IIC – if explosive atmospheres are still to be expected!
3.3 Application
The devices may be used only in the correct way and for the intended purpose in a normal industrial environment. Use for unsuitable purposes, will render all guarantees and liability of the manufacturer void! See Chapter 4, 5, 7 and 9.
Page 7
DOG-4
DOG-4 K08/0818 page 7
3.4 Repair, maintenance (for applications in Ex areas)
Definition of terms as defined in IEC 60079-17: Repair and maintenance: A combination of all activities carried out to maintain an
item or restore the object to a state in which it is able to meet the requirements of the relevant specification and ensure the execution of its required functions. Inspection: An activity comprising the careful examination of an item carried out either without demounting or, if necessary, with partial demounting through such steps as measurement in order to reliably determines the condition of the item. Visual inspection: An inspection in which visible faults, e.g. missing bolts, are detected without the use of equipment or tools. Close inspection An inspection which comprises the aspects of a visual inspection in addition to identifying defects, e.g. loose bolts, apparent only by access with the use of equipment, e.g. steps (when necessary) and tools. Close inspections do not normally require an open enclosure or the de-energizing of the equipment. Detailed inspection An inspection which encompasses those aspects covered by a close inspection and, in addition, identifies those defects, e.g. loose connections which are only be apparent by opening the enclosure and/or using tools and test equipment where necessary.
Maintenance work may be performed only by qualified personnel with the
equivalent of or qualification according to TRBS 1203 (German Technical Rules for Industrial Safety).
Only accessory parts which fulfill all European regulations and national laws
may be used in hazardous areas.
The replacement of components may only take place using original spare
parts approved for use in Ex areas as well.
The devices in Ex areas must be cleaned regularly. The intervals must be
specified by the operator according to the environmental load.
After maintenance and/or repair, replace all barriers and notices removed in
the process in their original positions.
Uninstall the device if any faults are detected. The customer may not repair
internal parts. Send the device to the manufacturer for inspection.
Task
Visual inspection
monthly
Check
every 6 months
Detailed inspection
every 12 months
Visual inspection of the device for integrity, remove dust residues
Inspection of the entire system Responsibility of the operator
3.1 Disposal
Disposal of the packaging and the used parts must be carried out according to the regulations of the country in which the device is installed.
Page 8
DOG-4
page 8 DOG-4 K08/0818
4. Instrument Inspection
Instruments are inspected before shipping and sent out in perfect condition. Should damage to a device be visible, we recommend a thorough inspection of the delivery packaging. In case of damage, please inform your parcel service / forwarding agent immediately, since they are responsible for damages during transit.
Scope of delivery:
The standard delivery includes:
DOG-4S Flowmeter including flange housing, measuring head, thermal wire
sensor and connection box.
Measuring cable from the Flowmeter to the measuring converter (optional). DOG-4A/B…O/P reader/transducer. Operating instructions.
5. Regulation Use
DOG-4 type devices are used for measuring and monitoring the rate of flow. Only clean media may be measured, against which the materials used are resistant. Dirt particles and other impurities can impair the measurement results, in spite of the fact that the continuous change in direction of the flow ensures a self-cleaning effect. The measuring sensor may be used only with the corresponding transducer.
Connectioon box
Measuring cable
Reader/ Transducer
Flowmeter
Flange housing
Measuring head
Direction
Page 9
DOG-4
DOG-4 K08/0818 page 9
6. Operating Principle
This device is a oscillating beam device and works without any movable parts. An orifice plate in the base creates flow resistance, which forces a partial flow to be directed into the measurement head. The gas in the measuring head begins to oscillate. The frequency of oscillation is proportional to the flow of volume. As the ratio between the flow through the measuring head and the flow through the measuring housing is constant, the frequency of oscillation is directly proportional to the entire flow of volume through the device. A platinum sensor determines the oscillation in the measuring head. The measuring converter normalises the measured oscillation to 0 Hz (without flow) and to 150 Hz for the end of the measuring range.
View from underneath
Platinum sensor
View from above
Page 10
DOG-4
page 10 DOG-4 K08/0818
7. Mechanical Connection
The measuring sensor should be installed in a horizontal position with the measuring head facing upwards. A slight inclination in the direction of flow is permitted. The arrow on the flange housing must point in the direction of flow. The recommended minimum inlet path is 10 times the diameter of the pipe and the maximum outflow zone 5 times the diameter of the pipe. To prevent any possible rotational flow, it is recommended that a flow straightener be fitted upstream of the counter. For damp media, slightly inclined installation in the direction of flow is recommended in order to ensure the removal of condensate from the oscillator.
Direction of flow

It is also recommended to install a condensate separator, preferably in a vertical pipe in front of the flow meter (see figure below). The condensate can thus be removed before entering the flow meter. Should the temperature difference between ambient and medium be +20 °C or greater, the pipeline and the flow meter should be thermally insulated to prevent condensation.
Page 11
DOG-4
DOG-4 K08/0818 page 11
8. Identification (Ex version)
Flowmeter (DOG-4S) type plate (flange housing, measurement tube)
There is a warning label on the plastic housing “WARNING – DANGER Of ELECTROSTATIC DISCHARGE – SEE OPERATING INSTRUCTIONS”
Transducer (DOG-4A) type plate (see left)
Page 12
DOG-4
page 12 DOG-4 K08/0818
9. Electrical connections
9.1 General
Install the measuring sensor near the transducer (max. 100 m cable length,
depending on the electrical interference zone).
The measuring cable must be laid well away from strong sources of electrical
interference and not parallel to power cables.
The measuring cable of several DOG-4s must not be laid over long distances
next to one another or bundled together.
The DOG-4A transducer must be installed outside of the Ex zone. Lay the electrical cabling according to the following wiring diagram. Each transducer is matched specifically to the respective measuring sensor
and must not be swapped over.
The pipelines and the flange housing must be grounded.
9.2 Measuring cable in Ex areas
An Ölflex EP (without shield) or Ölflex EBCY (with shield) may be used as measuring cable between the sensor and transmitter. Alternatively a cable with comparable properties may be used.
Ölflex EB Li=0.65 mH/km Cisy=110 nF/km Ölflex EBCY Li=0.65 mH/km Cisy=135 nF/km Ciasy=185 NF/km The length of the cable may not exceed 100 m. The maximum permitted cable
inductance is Limax=65 µH and the maximum capacitance Cimax=32 nF.
9.3 Connection to the measuring sensor
To connect the measuring cable, first unscrew the lid of the connection box and remove the lid. Feed the cable through the cable gland and connect it to the connecting terminal (independent of polarity). When using a shielded cable, the shield must be connected to the grounding screw.
Page 13
DOG-4
DOG-4 K08/0818 page 13
9.4 Terminal assignment of the DOG-4… reader/transducer
230 V
AC
110 VAC 24 VAC 24 V
DC
Output
Sensor
L = 230 V/110 V/24 V outer conductor + = supply DC+ N= 230 V/110 V/24 V neutral conductor - = supply DC-
E = emitter optical coupler C = collector optical coupler
I-= sensor conductor I+= sensor conductor
Supply
black
1 2 3 4
L N + -
Output orange
1 2 3 4
E C
Sensor
blue
1 2 3 4
+ -
for AC for DC
Page 14
DOG-4
page 14 DOG-4 K08/0818
U
HIGH
= ; U
LOW
< 0.5 V
Example: Vs = 24 VDC: U
HIGH
=17 VDC@ R=10 kOhm, U
HIGH
=23 VDC@R=80 kOhm,
R
R+4.4 k
Optical coupler output, external supply max. 30 VDC, I
max
= 50 mA
Page 15
DOG-4
DOG-4 K08/0818 page 15
9.5 Terminal assignment of the electronic option G0/H0/I0/K0/M0/N0/O0/P0
With the electronic option G0/H0/M0/N0, the reader/transducer is located together with the flow controller/flow computer prewired in the plastic wall housing. The connection terminal strip is located underneath the terminal cover. This must be removed for the electrical connection.
Note:
Additional pressure and temperature sensors are required for the electronic options M0/N0/O0/P0 (flow computer). These sensors are not included in the scope of delivery.
Page 16
DOG-4
page 16 DOG-4 K08/0818
Wiring diagram electronic option DOG-4 G0/H0/I0/K0
Wiring diagram electronic option DOG-4 L0
Wiring diagram electronic option DOG-4 M0/N0/00/P0
   
Note:
There is installation space for 2 additional Zener barriers in the housing if the pressure/temperature sensors do not have the required certification. Establishing contact in Ex areas can take place via terminals 17 to 20. The required cabling is the responsibility of the installer.
Page 17
DOG-4
DOG-4 K08/0818 page 17
9.6 Connections of DOG-4 Transmitter with KOBOLD Evaluation Electronics ZED-X and ZOK-Zx
DOG-4 connection with ZED-Zx
1,2 kOhm
Page 18
DOG-4
page 18 DOG-4 K08/0818
DOG-4 connection with ZOK-ZxP
Notes:
* The sensor supply from ZOK-ZxP must be set to 8 VDC. From the software menu of ZOK-Zx, the ‘Sensor type’ must be set to ‘PNP’
(For details please see the operating manual of ZOK-Zx electronics).
Page 19
DOG-4
DOG-4 K08/0818 page 19
Connection example for DOG-4 with ZOK-ZxK
The sensor type must be set to "PNP" in the ZOK-ZX software menu (see operating manual ZOK-Zx electronics).
Page 20
DOG-4
page 20 DOG-4 K08/0818
10. Startup
During startup, ensure that the shut-off valves upstream and downstream of the device are opened slowly. Avoid increasing the flow velocity to prevent the discharge rate from damaging the sensor. The output frequency of the DOG-4 is proportional to the measuring range.
In the event of a sensor fault (breakdown or short circuit) the transistor output is permanently connected. During self-diagnosis, the transistor output is switched to the high-resistance state.
11. Setup and operation of the flow controller/flow computer (option G0/H0/I0/K0/M0/N0/L0/O0/P0)
The integrated flow controller/flow computer are preset and calibrated at the factory. See the separate operating instructions for the DOG-4G0/H0 and the DOG-4M0/N0 for details.
Page 21
DOG-4
DOG-4 K08/0818 page 21
12. Transmitter error message
12.1 Status display not illuminated
Check terminal “230 VAC” of the terminal assignment.  Check supply voltage
12.2 OPERATING illuminated green but no output signal
Check terminal “Output” of the terminal assignment
12.3 RANGE illuminated yellow
Measuring range overflows or underflows, decrease or increase flow
12.4 ERROR illuminated red
Short circuit or interruption of the sensor circuit from
the sensor to the transmitter.
Check terminal “Sensor” of the terminal assignment Defective platinum sensor (target value: 40-60 Ohm)
Page 22
DOG-4
page 22 DOG-4 K08/0818
13. Maintenance
13.1 Replacing the sensor
To replace the sensor, turn off the flow of medium and discharge the pressure in the
lines. For versions with ball valves, it is sufficient to simply shut these. Demounting (see diagram) is carried out in the following order:
1) Open the connection box (4 x screws)
2) Unclamp the sensor from the strip terminal (2 screws)
3) If necessary (due to lack of space or tight cable arrangement), also unclamp from the strip terminal, loosen the cable gland and pull out the cable
4) Loosen the connection box from the plate (2 x screws)
5) Slowly release the sensor screws and carefully pull out the sensor. Take care that no sensor parts remain in the device and also that no foreign parts fall into the device (2 screws).
Reassembly is in the reverse order. Replac the O-rings with the new ones supplied with the sensor. Uniformly tighten the sensor screws crosswise.
13.2 Disassembly and cleaning the Flowmeter
In the event that the internal section of the oscillator is dirtied by the medium or foreign particles fall into this during replacement of the sensor, the device must be demounted and cleaned (see figure at left).
Before the device can be demounted, the connection box and the sensor must be unscrewed. For this, see point 13.1. Then unscrew the screws in the upper plate, separate the parts and clean the flow channel of the oscillator on both sides.
Gasket above
Gasket below
2) Order of disassembly 2x Number of screws NM Requires tightening torque
1) 4x – 1.2-1.5 Nm
2) 4x – 0.3 – 0.5 Nm
5) 4x – 1 – 1.2 Nm
4) 2x – 3 – 4Nm
3) 1x – 4 – 5Nm
Page 23
DOG-4
DOG-4 K08/0818 page 23
Do not use any sharp objects or aggressive cleaning agents. These could damage the oscillator, causing measurement inaccuracies or breakdown of the device. Check the inflow and discharge channel and clean if necessary. Reassembly takes place in the reverse order. The two gaskets must be replaced and are included in the repair set. Tighten the screws uniformly with a torque of 24 to 27 Nm. The figure at the lower left shows the required order when tightening. Otherwise the device may leak or be damaged.
Flow channel top
Flow channel bottom
Page 24
DOG-4
page 24 DOG-4 K08/0818
14. Technical Data
Measuring accuracy: 1,5% of meas. value (at Qt-100%*) 5% of measured value (at 1%-Qt*) *The lower limit Qt depends on the density. Qt = 8% at density 1 kg/m³ Qt = 4% at density 2 kg/m³ Qt = 2% at density 4 kg/m³ Qt = 1% at density 8 kg/m³ Repeatability: 0,1% of measured value Media temperature: -20 ...+120 °C (non ATEX version)
-20 ...+60 °C (ATEX version) Ambient temperature: -25 …+80 °C (non ATEX version)
-25…+60 °C (ATEX version) Operating pressure: DOG-42xxx25…DOG-42xxx50…: PN 40 DOG-42xxx80…DOG-42xxx1F…: PN 16 Span: 1:100 Sensor: platinum sensor Protection: IP 65
Materials (Transmitter)
Housing: stainless steel 1.4404/316L Orifice: stainless steel 1.4404/316L Measuring head: polyphenylene sulfide (PPS) Sensor: platinum Gaskets: Klingersil® C-4265, NBR Ball valves: stainless steel
Page 25
DOG-4
DOG-4 K08/0818 page 25
14.1 Electronic options
Electronics DOG-…A0/B0 (Transducer with/without ATEX certification)
Power supply A/B: 230 V
AC
± 10 %, 50…60 Hz
Power supply C/D: 110 V
AC
± 10 %, 50…60 Hz
Power supply E/F: 24 V
AC
± 20 %, 50…60 Hz
Power supply R: 24 V
DC
± 20 % Input: Platinum sensor (Allowed distance: max. 50 m to transmitter) Output: Opto coupler, frequency linear to flow V
CE
: 12-24 V (recommended), max. 30 V I
C
: max. 50 mA
P
tot
: 100 mW at 25 °C Derating: 0.91 mW/°C Ambient temperature: -25…+60 °C Protection: IP20 Ex-version(A): ATEX Transducer: II (1)G [Ex ia Ga] IIC Sensor: II 1 G Ex ia IIC T4 Ga IECEx Transducer: [Ex ia Ga] IIC Sensor: Ex ia IIC T4 Ga Mounting: DIN Rail Dimensions: Width: 45 mm Heigth: 105.6 mm Depth: 113.6 mm Weight: approx. 200 g
Page 26
DOG-4
page 26 DOG-4 K08/0818
Electronics DOG-…G0/H0 (Transducer without/with ATEX certification and Flow rate/Unit counter, with current/pulse output)
Display: alphanumeric LCD, UV-resistant, with displayed functions: Compensated flow rate (7 digits, 17 mm high) Compensated total (7 digits, 17 mm high) resettable Accumulated total (11 digits, 8 mm high) not resettable Units: Flow: m3, cf, scf, Nm
3
Time units: /sec, /min, /hr, /day Total: m3 Accumulated total: m3 Decimal places: Flow: 0, 1, 2 or 3 Total: 0, 1, 2 or 3 Accumulated total: according to selection for total Backlightning: yes Signal input: Flow: DOG-4 sensor Power supply: G/H: 230 V
AC
± 10 %, 50…60 Hz I0/K0: 110 VAC ± 10 %, 50…60 Hz L0: 24 VDC ± 20 %
Electrical connection: 4 x M16 x 1.5 cable gland Housing material: ABS with PC cover Weight: approx. 1800 g Analogue output: 4…20 mA (active) 10-Bit resolution, 3-wire Pulse output: PNP, 24 V
DC
active max. 50 mA, scaled according to linearised accumulated total (e.g. pulse every 12 liters) Pulse duration: user defined 0.008 s…2 s max. frequency: 64 HZ Protection: IP65 Mounting: Wall mounting Data protection: EEPROM backup, backup of running totals every minute, Data retention at least 10 years Communication: Modbus RTU RS485 2-wire (optional, other Modbus versions on request)
Page 27
DOG-4
DOG-4 K08/0818 page 27
Electronic Options DOG-…M0/N0/O0/P0 (Transducer without/with ATEX certification and Flow computer)
Display: Alphanumeric LCD, UV-resistant with Displayed functions: Compensated Flow rate (7 digits, 17 mm high) Compensated total: (7 digits, 17 mm high) resettable Accumulated total (11 digits, 8 mm high) not resettable Actual line temperature (6 digits) Actual line pressure (6 digits) Units: Flow: m3, cf, scf, Nm3 Time units: /sec, /min, /hr, /day Total: m3 Accumulated total: m3 Temperature: °C, °F, or K Pressure: mbar, bar PSI Decimals: Flow: 0, 1, 2 or 3 Total: 0, 1, 2 or 3 Accumulated total: according to selection for total Temperature/Pressure: 1 Backlightning: yes Signal input: Flow: DOG-4 sensor
Temperature: PT100, 2- or 3-wire Pressure: 0(4)…20 mA (passive), 14-Bit resolution, 2- or 3-wire Power supply: M/N: 230 V
AC
± 10 %, 50…60 Hz O/P: 110 VAC ± 10 %, 50…60 Hz Electrical Connection: 5 x M16 x 1.5 cable gland Housing material: ABS with PC cover Weight: approx. 1800 g Analogue output: 4…20 mA (active), 10-Bit resolution, 3-wire Pulse output: PNP, 24 VDC active max. 50 mA, scaled according to linearised accumulated total (e.g. pulse every 12 liters) Pulse duration: user defined
0.001 s…10 s max. frequency 500 Hz Protection: IP65 Mounting: Wall mounting Data protection: EEPROM backup, backup of running totals every minute, Data retention at least 10 years Communication: Modbus RTU RS485 2-wire (optional, other Modbus versions on request)
Page 28
DOG-4
page 28 DOG-4 K08/0818
Display
Note: Temperature and pressure sensors are not included in scope of delivery.
Page 29
DOG-4
DOG-4 K08/0818 page 29
15. Order Details
Order details for DOG-4 (Example: DOG-4200 S 50 0 A0 0)
Measuring
range air
[m
3
/h]
Model
Material
st.steel
Pressure
rating
[PN]
Connection
flange
[size/type]
Ball valve
Electronics
Options
0,12...12
DOG-42S0S25..
10…40 bar
DN25
0 = without ball valve 1 = with ball valve
B0 = Frequency output, 230 V
AC
A0 = as ‘B’, with ATEX/IECEx G0 = Unit counter, pulse
output, analogue output, 230 V
AC
H0 = as ‚G‘, with ATEX/IECEx M0 = Flow computer, pulse
output, analogue output, 230 V
AC
N0 = as ‚M‘, with ATEX/IECEx C0 = as ‘B0’, 110 V
AC
D0 = as ‘A0’, 110 VAC E0 = as ‘B0’, 24 V
AC
F0 = as ‘A0’, 24 V
AC
R0 = as ‘B0’, 24 VD
C
I0 = as ‘G0’, 110 V
AC
K0 = as ‘I0’, 110 V
AC
L0 = as ‘G0’, 24 VD
C
O0 = as ‘M0’, 110 V
AC
P0 = as ‘N0’, 110 V
AC
Y0 = Special (specify in clear text)
0 = without Y = Special option
(specify in clear text)
DOG-42S0S40..
DN40
DOG-42S0S50..
DN50
DOG-42S0B25..
Klasse 300
ANSI 1“
DOG-42S0B40..
ANSI 1½“
DOG-42S0B50..
ANSI 2“
0.2…20
DOG-4200S25..
10…40 bar DN25
DOG-4200S40..
DN40
DOG-4200S50..
DN50
DOG-4200B25..
Class 300 ANSI 1“
DOG-4200B40..
ANSI 1½“
DOG-4200B50..
ANSI 2“
0.35…35
DOG-4250S25..
10…40 bar DN25
DOG-4250B25..
Class 300 ANSI 1“
0.7…70
DOG-42A0S25..
10…40 bar DN25
DOG-42A5B25..
Class 300 ANSI 1“
0.9…90
DOG-42A5S40..
10…40 bar DN40
DOG-42A5B40..
Class 300 ANSI 1½“
1.1…110
DOG-42B0S50..
10…40 bar DN50
DOG-42B0B50..
Class 300 ANSI 2“
1.4…140
DOG-42B5F80..
16 bar DN80
DOG-42B5S80..
40 bar DN80
DOG-42B5A80..
Class 150 ANSI 3“
DOG-42BFB80..
Class 300 ANSI 3“
2…200
DOG-42C0S40..
10…40 bar DN40
DOG-42C0B40..
Class 300 ANSI 1½“
2.5…250
DOG-42C5S50..
10…40 bar DN50
DOG-42C5B50..
Class 300 ANSI 2“
2.7…270
DOG-42D0F1H..
16 bar DN100
DOG-42D0S1H..
40 bar DN100
DOG-42D0A1H..
Class 150 ANSI 4“
DOG-42D0B1H..
Class 300 ANSI 4“
4.5…450
DOG-42D5F80..
16 bar DN80
DOG-42D5S80..
40 bar DN80
DOG-42D5A80..
Class 150 ANSI 3“
DOG-42D5B80..
Class 300 ANSI 3“
6.0…600
DOG-42E0F1F..
16 bar DN150
DOG-42E0S1F..
40 bar DN150
DOG-42E0A1F..
Class 150 ANSI 6“
DOG-42E0B1F..
Class 300 ANSI 6“
6.5…650
DOG-42E5F1H..
16 bar DN100
DOG-42E5S1H..
40 bar DN100
DOG-42E5A1H..
Class 150 ANSI 4“
DOG-42E5B1H..
Class 300 ANSI 4“
Page 30
DOG-4
page 30 DOG-4 K08/0818
Order details for DOG-4 (Example: DOG-4200 S 50 0 A0 0) (continuation)
Measuring
range air
[m
3
/h]
Model Material st. steel
Pressure
rating
[PN]
Connection
flange
[size/type]
Ball valve
Electronics
Options
8.0…800
DOG-42F0F80..
16 bar DN80
0 = without ball valve 1 = with ball valve
B0 = Frequency output, 230 V
AC
A0 = as ‘B’, with ATEX/IECEx G0 = Unit counter, pulse
output, analogue output, 230 V
AC
H0 = as ‚G‘, with ATEX/IECEx M0 = Flow computer, pulse
output, analogue output, 230 V
AC
N0 = as ‚M‘, with ATEX/IECEx C0 = as ‘B0’, 110 V
AC
D0 = as ‘A0’, 110 VAC E0 = as ‘B0’, 24 V
AC
F0 = as ‘A0’, 24 V
AC
R0 = as ‘B0’, 24 VD
C
I0 = as ‘G0’, 110 V
AC
K0 = as ‘I0’, 110 V
AC
L0 = as ‘G0’, 24 VD
C
O0 = as ‘M0’, 110 V
AC
P0 = as ‘N0’, 110 V
AC
Y0 = Special (specify in clear text)
0 = without Y = Special
option (specify in clear text)
DOG-42F0S80..
40 bar DN80
DOG-42F0A80..
Class 150 ANSI 3”
DOG-42F0B80..
Class 300 ANSI 3”
10…1000
DOG-42F5F1H..
16 bar DN 100
DOG-42F5S1H..
40 bar DN 100
DOG-42F5A1H..
Class 150 ANSI 4“
DOG-42F5B1H..
Class 300 ANSI 4“
12…1200
DOG-42G0F1F..
16 bar DN 150
DOG-42G0S1F..
40 bar DN 150
DOG-42G0A1F..
Class 150 ANSI 6“
DOG-42G0B1F..
Class 300 ANSI 6“
DOG-42G0E2H..
10 bar DN 200
DOG-42G0F2H..
16 bar DN 200
DOG-42G0S2H..
40 bar DN 200
DOG-42G0A2H..
Class 150 ANSI 8“
DOG-42G0B2H..
Class 300 ANSI 8“
25…2500
DOG-42G5E2H..
10 bar DN 200
DOG-42G5F2H..
16 bar DN 200
DOG-42G5S2H..
40 bar DN 200
DOG-42G5A2H..
Class 150 ANSI 8“
DOG-42G5B2H..
Class 300 ANSI 8“
30…3000
DOG-42H0F1F..
16 bar DN 150
DOG-42H0S1F..
40 bar DN 150
DOG-42H0A1F..
Class 150 ANSI 6“
DOG-42H0B1F..
Class 300 ANSI 6“
60…60001)
DOG-42H5E2H..
10 bar DN200
DOG-42H5F2H..
16 bar DN 200
DOG-42H5S2H..
40 bar DN 200
DOG-42H5A2H..
Class 150 ANSI 8“
DOG-42H5B2H..
Class 300 ANSI 8“
Special
DOG-42YYYYY.. Sonder Sonder
1)
Calibrated up to 4000 m3/h. Higher flow rate calibration on request.
Page 31
DOG-4
DOG-4 K08/0818 page 31
16. Pressure Loss/Flow
The diagram applies for gases with a density of air at NPT (0 °C and 1013.25 mbar). The pressure loss is always proportional to the density of the gas. For example, the pressure loss doubles at 100% higher operating pressure.
Calculating the Actual Density
The actual density can be calculated with the following formula:
D0 = density at 1 bar abs. and 0 °C (= 273 K) T = temperature in K (= °C + 273 for example 20 °C = 273 + 20 = 293 K) T0 = 273 K P = operating pressure in bar (absolute pressure)
Calculating the Norm Flow
QN = norm flow at 1.013 bar abs. and 0 °C Q = operating flow P = operating pressure in bar (absolute pressure) T = operating temperature in K
Pressure drop [mbar]
Page 32
DOG-4
page 32 DOG-4 K08/0818
17. Dimensions
Dimensions and Weights DOG-4 (without ball valve)
Dimensions and Weights DOG-4 (with ball valve)
Dimensional details without ball valve
DN
[mm] L (Length)
[mm]
H
(Height)
[mm]
øD
(outer ø)
[mm]
øK
(pitch circle)
[mm]
øL
(hole ø)
[mm]
Ød1
(sealing surface)
[mm]
No. of
screws
Screw size Weight
[kg]
25 300 150 115 85 14 68 4 M12 8,1 40 300 158 150 110 18 88 4 M16 10 50 300 164 165 125 18 102 4 M16 11,6
80 300 178 200 160 18 138 8 M16 14,4 100 320 191 220 180 18 58 8 M16 16,6 150 320 218 285 240 22 212 8 M20 24,8 200 320 243 340 295 22 268 8 M20 35,8
Page 33
DOG-4
DOG-4 K08/0818 page 33
Dimensions Electronics DOG-…A/B
Dimensions Electronics DOG-…G/H/M/N
Accessories (optional)
Replacement sensor Sealing for oscillator Recalibration set for transmitter
Dimensional details with ball valve
DN
[mm] L (Length)
[mm]
H
(Height
)
[mm]
øD
(outer ø)
[mm]
øK
(pitch circle)
[mm]
øL
(hole ø)
[mm]
Ød1
(sealing surface)
[mm]
No. of
screws
Screw size Weight
[kg]
25 300 166 115 85 14 68 4 M12 8,5 40 300 174 150 110 18 88 4 M16 10,4 50 300 180 165 125 18 102 4 M16 12
80 300 194 200 160 18 138 8 M16 14,8 100 320 207 220 180 18 58 8 M16 16,9 150 320 234 285 240 22 212 8 M20 25,3 200 320 259 340 295 22 268 8 M20 36,3
Page 34
DOG-4
page 34 DOG-4 K08/0818
18. EU Declaration of Conformance
We, KOBOLD Messring GmbH, Hofheim-Ts, Germany, declare under our sole responsibility that the product:
Oscillation Flowmeter Model: DOG-4 to which this declaration relates is in conformity with the standards noted below:
EN 61326:2013-07 Electrical equipment for measurement, control and laboratory
use - EMC requirements - Part 1: General requirements
EN 13480-1:2013-11 Metallic industrial piping - Part 1: General EN 50581:2012 Technical documentation for the assessment of electrical and
electronic products with respect to the restriction of hazardous substances
zusätzlich DOG-4…A/H/N/D/F/P: EN 60079-0:2014-06 Explosive atmospheres - Part 0: Equipment –
General requirements EN 60079-11:2012 Explosive atmospheres - Part 11: Equipment protection by intrinsic safety "i" EN 60079-26:2007 Explosive atmospheres - Part 26: Equipment with equipment protection level (EPL) Ga
Also the following EU guidelines are fulfilled:
2014/30/EU EMC Directive 2014/35/EU Low Voltage Directive 2011/65/EU RoHS (category 9) 2014/68/EU PED
Category III (IV) Diagram 1, vessel, group 1 dangerous fluids Module D, marking CE0575 Notified body: DNV GL Certificate No. PEDD000000R
additional DOG-4…A/H/N/D/F/P: 2014/34/EU Equipment and protective systems intended for use in
potentially explosive atmospheres (ATEX 100a)
Production quality assurance
Certification no.: BVS 15 ATEX ZQS / E 110
Notified body: DEKRA EXAM Identification no.: 0158
Hofheim, 31. July 2018 H. Peters M. Wenzel General Manager Proxy Holder
Page 35
DOG-4
DOG-4 K08/0818 page 35
19. ATEX-Certificate
Page 36
DOG-4
page 36 DOG-4 K08/0818
Page 37
DOG-4
DOG-4 K08/0818 page 37
Page 38
DOG-4
page 38 DOG-4 K08/0818
Page 39
DOG-4
DOG-4 K08/0818 page 39
Page 40
DOG-4
page 40 DOG-4 K08/0818
Page 41
DOG-4
DOG-4 K08/0818 page 41
20. IECEx-Certificate
Page 42
DOG-4
page 42 DOG-4 K08/0818
Page 43
DOG-4
DOG-4 K08/0818 page 43
Page 44
DOG-4
page 44 DOG-4 K08/0818
Loading...