Kobold **C30R Series, **C30M Series, **C3 Series, **C34N Series, **C34P Series Operating Instructions Manual

Page 1
Operating Instruction
Supplement
for
Compact Electronics
Model:..C30R;..C30M
..C34P;..C34N
Page 2
...C3...
Seite 2 ...C3... K01/0916
1. Contents
1. Contents ........................................................................................................ 2
2. Note .............................................................................................................. 3
3. Electrical Connection .................................................................................... 3
4. Programming ................................................................................................ 5
5. Maintenance ............................................................................................... 12
6. Technical Information .................................................................................. 12
Manufactured and sold by:
Kobold Messring GmbH
Nordring 22-24
D-65719 Hofheim
Tel.: +49(0)6192-2990
Fax: +49(0)6192-23398
E-Mail: info.de@kobold.com
Internet: www.kobold.com
Page 3
...C3...
...C3... K01/0916 Seite 3
2.Note
This programming instruction is only a supplement to the operating instruction for the sensor.
3. Electrical Connection
3.1. General
Attention! Make sure that the voltages in your plant correspond with the instrument voltages
Make sure that the supply wires are de-energized.Connect the supply voltage and evaluation of both output signals to the plug
connector PINs as shown in section 3.2 and 3.3.
We recommend the use of wires with cross sectional area of min. 0,25 mm²
Attention! Incorrect wiring will lead to damage of the unit’s electronics.
Page 4
...C3...
Seite 4 ...C3... K01/0916
3.2. Compact Electronics: (..C30R, ..C30M)
3.3. Compact Electronics: (..C34P, ..C34N)
Terminals 3 and 5 are linked internally and can therefore be used either for the output signal or the power supply.
+ Vs
GND
GND
1
2
4
3
5
Switc h o u t 2
Switc h o u t 1
+ Vs
GN D
GN D
1
2
4
3
5
(0)4 -20 mA
Switc h o ut 1
Page 5
...C3...
...C3... K01/0916 Seite 5
4. Programming
Connect the compact electronics according to the previous wiring diagram and apply the specified voltage.
The measuring range (upper range value) is displayed for 3 seconds after switch-on.
4.1. Key Function
Standard mode (measuring mode)
Set-up mode
4.2. Settings
The following values can be changed in the compact electronics:
Scale range Factory setting
Switching point (SPo, SP1, SP2)
0 - 999 0.00
Hysteresis (HYS)
-199 - 0 -0.00
Window point (duo)
Switching point ...999 --- (de-activated)
Filter (Filt)
1/2/4/8/16/32/64 1
Contact type (Con, Co1, Co2) N/O contact (no), N/C (nc)
or frequency (Fr)**
no (N/O contact)
Start current (S-C)*
000 - 999 000
End current (E-C)*
000 - 999 Upper range value
Start current selection (SCS)
0-- (0 mA), 4-- (4 mA) 4 mA
Change code (CCo)
000 - 999 000
* Lower and upper range values of the flow rate are based on 0/4-20 mA. ** only for sensors with impulse output (e.g. DPE)
: pressing 3 sec.
set-up mode
: switchpoint/window point
Standard mode
3 sec
or 20 sec
no key pressing
Anytime:
: Next level
: Set value
Page 6
...C3...
Seite 6 ...C3... K01/0916
4.3. Value Setting
You can select Value setting in the main menu item (for example: Switching point, "SPo") by pressing the "" key. The structure shown below illustrates the universal routine for changing individual parameters.
from the main menu item
1. Digit setting
2. Digit setting
3. Digit setting
Decimal point setting
Choosed value storing or new setting
to the next main menu item
Storing
Page 7
...C3...
...C3... K01/0916 Seite 7
4.4. Set-up Mode
Changing code
N/O
Storing
Contact 1 function
N/O
N/C
Frequency*
Storing
Filter
7 Levels
Value setting
Hysteresis
Switching point 2
Swichting point 1
Code=
3 sec
Code entering
Value setting
Value setting
Value setting
Value setting
Contact 2 function
Value setting
*Only for sensors with impulse output
Page 8
...C3...
Seite 8 ...C3... K01/0916
Code entering
Filter
Windowpoint
Hysteresis
Switching point
3 sec
Code=
Start current
7sec
7sec
Contact function
Value setting
Value setting
Value setting
Value setting
Value setting
7 levels
N/O
N/C
Frequency*
Value setting
Value setting
Storing
End current
*Onl
y
for sensors with impulse output
Page 9
...C3...
...C3... K01/0916 Seite 9
Changing code
Value setting
Storing
7sec
4-20
m
A
0-20
m
A
Analog output
choosing
4.5. Main Menu Items
Switching point
The switching point is entered in menu item "SPo, SP1, SP2". A value in the range 000 to 999 may be chosen. A decimal point position is also assigned to this value. The position of the decimal point can be set after the first, second or last position (no decimal point). If the indicated value exceeds the set switching point, then the electronics switches and energizes the LED. If the hysteresis is equal to zero and the window point is de-activated, the electronic switches back whenever the indicated value falls below the switching point.
Hysteresis
After the switching point, hysteresis can be entered as a negative value in the menu "HYS". The default hysteresis value is zero. However this can lead to confused switching, when the measuring signal fluctuates around the switching point or window point. This problem can be cured by increasing the hysteresis. The hysteresis is based on the switching point and the window point (switching point minus hysteresis; window point plus hysteresis).
Example: switching point 100 l/min; hysteresis: -2.5 l/min The electronics switches when 100 l/min is exceeded and switches
back when the flow rate falls below 97.5 l/min.
Window point (duo point)
A window point "duo" (duo point) can be defined in addition to the switching point. The window point must be greater than the switching point. The measured value can be monitored within a set range by means of the window point and switching point. The switching point marks the lower end of the range of values and the window point the upper end.
Page 10
...C3...
Seite 10 ...C3... K01/0916
If the window point (duo point) is lower than or equal to the switching point, an error message is displayed (Er4), its value is then deleted and thus its function is disabled (applies to window point and switching point setting).
Value setting is similar to switching point setting. The window point serves to monitor the measured value within a set range.
Example: Switching point: 100 l/min; window point: 150 l/min; hysteresis: -1 l/min The electronics switches when 100 l/min is exceeded. When the measured value stays within the limits 99 l/min (100-1) and 151 l/min (150+1), then the electronics also remains in the activated switch state (LED on). Should the measured value exceed 151 l/min or drop below 99 l/min, then the electronics switches back.
Switching performance
The switching performance of the electronics is illustrated in the diagram below. The contact closes (N/O contact) when the switching point is exceeded or when the value drops below the window point. It opens when the window point plus hysteresis is exceeded or when the value drops below the switching point minus hysteresis. The switch state of the electronics is indicated by an LED.
Time / t
Display barC)
Hysteresis
LED on
Time / t
Display bar (°C)
Switchpoint
Hysteresis
LED on
Hysteresis
Window point
LED on
Flow
Flow
Page 11
...C3...
...C3... K01/0916 Seite 11
Filter The filter function "Filt" generates the sliding average value from the measured values. The following values are available (see section 6.2 Settings):
1 / 2 / 4 / 8 / 16 / 32 / 64
The filter value determines the dynamic behaviour of the indicated value: high values result in a slow display response. The filter is disabled if a filter value of "1" is selected. In other words, the indicated value is equal to the unfiltered measured value.
The integrated step detector responds to a measured-value step change greater than approximately 6.25% of the upper range value. When a measured-value step change is detected, the actual measured value is displayed immediately.
Contact Model The function of the transistor switching output is set in menu item "Con, Co1 or Co2". The switching function switches from no - N/O contact to nc - N/C to Fr – frequency (Con and Co1 only and
for sensors with impulse
output) and back. N/O contact means: contact closes when switching point is exceeded N/C means: contact opens when switching point is exceeded Frequency means: frequency output synchronized with vane frequency
Current output
The current output is selected in menu items "S-C" Start current indicated value < > 0(4) mA "E-C" End current indicated value < > 20 mA "SCS" Start current selection (0-20 mA or 4-20 mA). The indicated value at which 0(4) mA flow is entered in menu item Start current. The indicated value at which 20 mA flow is entered in menu item End current.
Change Code
The change code option "CCo" secures the unit against unauthorised tempering. If the code is different from 000, the user must input the code immediately after entering the adjustment mode.
Page 12
...C3...
Seite 12 ...C3... K01/0916
5. Maintenance
Work on the electronics should only be carried out by the supplier, otherwise the guarantee is nullified.
6. Technical Information
Display: 3-segment LED Display case: stainless steel Analogue output: (0)4 - 20 mA scalable (...C34 only) Transistor output: 1 (2) semiconductor PNP or NPN, set at the factory. Frequency output: 1 transistor output vane frequency (see sensor) Max. switching current: 300 mA Function: N/C / N/O contact programmable Setting: with 2 buttons Supply: 24 V
DC ±20%, 3-wire technology
Electrical connection: plug connector M12x1
Loading...