Kobastar K1, XK3118T1 User Manual

1
KOBASTAR K1 USER MANUAL
2
TABLE OF CONTENT
1.0 BRIEF INTRODUCTION………………..…………………………………………..………………...……………………………….3
2.0 INSTALLATION………….…….………………………………………………………….…………….………..………………………3
2.1 INDICATOR INTERFACE………….………………..…………………………………………..……3
2.1.1 K1 BACKSIDE……..…..…………………………………………………………..…………3
2.1.2 Loadcell with connection…..……..………………….…………………………….…4
2.1.3 Serial Communication Wire Connection..…….………………………………..4
3.0 OPERATION………………..…………………………………………………………………………………………………………….4
3.1 AUTO ZERO WHEN TURN ON AND OFF…………………………………………………………4
3.2 MANUALLY ZERO...……………………………………………………………………………………….4
3.3 TARE………….………………………………………………………………………………………………….5
3.4 OPERATION FOR TOTAL DISPLAY AND FUNCTION……………………....……………..…5
5.0 CALIBRATION.……………………………………..……………………………………………………………………………………6
6.0 OTHER PARAMETER SETTING AND FUNCTION……………..……….…………………..……………………………..6
7.0 ERROR WARNING.…………………………………………………………………………………………………………………….9
3
1.0 BRIEF INTRODUCTION
Weighing Indicator XK3118T1 adopts high anti-jamming Single-chip microprocessor and high precision A/D conversion technology. It is with stainless steel frame (XK3118T1B is with plastic frame) and widely applied in electronic pplatform scale, electronic floor scale and other static weighing systems that equipped with 1-6pcs load cells. AC/DC are both available fort his indicator.
1.1 TECHNICAL PARAMETERS
Accuracy class: III, n=3000
A/D conversion method 20bit
Input sensitivity > 1.5uV/e
Load cell excitation: DC 5V, connect with 1-6 pcs loadcells with 350 ohm each
Loadcell connection mode: 6 wire (long wire auto compensation)
Division: 1/2/5/10/20/50 optional
Display: 6 bits LED, 10 indicating signs;
Specialized adapter power supply: AC 110-220V, 50Hz-60Hz(special adapter)
6V/4AH built-in maintaince free Lead-Acid battery
Working temperature 0-+40 °C , Humidity=85%(RH)
Storage temperature: -20~+60°C, Humidity=95%(RH)
2.0 INSTALLATION
2.1 INDICATOR INTERFACE
2.2 K1 BACK SIDE
4
2.1.2 Loadcell with Connection
Specification
- Excitation
- Feedback
+ Excitation
+ Feedback
- Signal
+ Signal
Shield
! Connection between load cell and indicator must be reliable, shield wire must be connected
to ground reliably. Connection or disconnection are not allowed when the indicator is on, which may damage the indicator or load cells.
! Static protection must be properly adopted as the load cell and indicator are all static sensitive
equipments. Welding or other strong electricity operation should be strictly forbidden. During
thunderstorm season, proper lightening protection should be tanken care of to protect the load
cells and indicators from damaging by lightening and to ensure the personal safety and the safely
running of the weighing and related equipments.
2.1.3 Serial Port Cummunication Wire Connection
Serial Port Chart
5
3.0 OPERASYON
3.1 CİHAZ AÇMA/KAPATMA VE OTOMATİK SIFIRLAMA
The indicator power can be controlled by the on-off key on the front faceplate . The indicator will perform self-check after turned on. If the scale was found departure from the calibrate zero however still within the range of turn on auto zero then the indicator will display “0” and the in dicator light for “zero digit” will be on. If the scale was found departure from the calibrate zero and out of the range of turn on auto zero then the indicator will display the current read
If the range setting of the turn on zero is”- -“ wihch means zero of last turn off, then zero operation will not be carried on and the zero of last turn off will be automatically readin and the current weight will be displayed.
3.2 ELLE SIFIRLAMA
Indıcator will back to zero when pressing “ZERO” key if the gross weight is within the range of manually zero and stable. Manually zero is not valid under the “NET” displaying mode;
3.3 DARA ALMA
TARE operation can be carried out if both gross weight and net weight are above zero and stable.
The indicator will display ”0” after pressing the “TARE” key. The tared weight is current gross weight. And the indicator will enter “NET” displaying mode and the “NET” indicator lights will be on;
The indicator will exit “NET” displaying mode after pressing “TARE” when the gross weight is “0” and under “NET” displaying mode
3.4 TOPLAM ALMA, TOPLAMI GÖSTERME VE TOPLAMI TEMİZLEME
1. In case, the net weight is bigger than the minimum measurable value (5 divison) and stable,
When pressing “TOTAL”, the current net weight will be added and the “TOTAL” indicator light will be
on and total value will be displayed , which will be changed to adding times [n ***]
After 3 seconds and exit the total displaying status after another 3 seconds. Next operation will be
only valid when the net weight is smaller than the minimum measurable value.
2. If keep pressing “TOTAL” until buzzer alarms under the weighing status “FUNCTION” and “TOTAL lights will be on and you can check the total value by pressing “” to see the adding times and again
” to exit.
3. If you press ““ after keep pressing “ TOTAL” and enter the total displaying status there will be an
notice [CLRAr-] which means to clear the total value or not. Press ““ to clear and “FUNCTION” to
exit and keep the value.
6
OPERATION DIAGRAM
4.0 CALIBRATION
After self checking of turn on, it will enter normal weighing display, warm up for 15-30 minutes, open the calibration cover in the back side of the indicator; touch the calibration switch to enter calibration status, the detailed steps as below:
Steps Operation Display Specification
CAL Means enter calibration status, press “” button to
next step
1 ”to select
division “”to confirm
E 01 Set division125102050
2 ”to select
decimal point “”to confirm
dC 0 Set decimal point position press up and down arrow
to switch decimal point00.00.00、0.000。
3 Set full capacity F030.00 Set Full Capacity:
Press ”” the flash digit will move towards right Press ”” to increase the number of flash digit
Press “” to confirm and enter next step. For example: 3000
4 After stable, press to
confirm
noLoAd
Zero calibration:
Make su r e indicator under no-load
status, press “”
after date stable to confirm Zero
5 Input the weight of
loading weights
AdLoAd
After display for 2 seconds, auto switch to load input
003.000
Linearity Calibration:
Add the weights to the scale, input the value of the weights in indicator, it is better to close to the full capacity.
Operation is same as full capacity setting.
5 seconds after the stable indicator light on Press “” to confirm and finish the calibration
6 Return back to
weighing status
【 End 】 【 3.000】
After calibration, 2 second after display “End”, the current weight will be displayed immediately.
“ Fn” tuşu ile kalibrasyonu tamamlamadan çıkabilirsiniz.
Keep Pressing
“Total”
Display Total
Value
Press ”” to see
the adding times
Press ”” to exit
Press “” to
celar total value
Press “” to
confirm clear
Press “FUNCTION”
to cancel
7
5.0 OTHER PARAMETER SETTING AND FUNCTION
Steps
Operations Display Note
1 Keep Pressing
FUNCTION” to Enter
“ to Switch
“ to Confirm
[FN **] “Functıon” Setting:
[ Lb]: One key switch (kg/lb), (Not available under net weighing status) [ANL]: Animal Scale, One key total and lock display. [ --]
2
“ to Switch
“ to Confirm
[PS **] Power Save Setting:
[oFF]: Power save mode off [oN]: Open power save mode. The power save mode will be entered 5 minutes after weight stable. The indicator will only display date circultly in last digit [onP]: Enchanced power save mode which will automatically turn off the indicator after 5 minutes power save mode.
3
“ to Switch
“ to Confirm
[ br****] Baud Rate Setting: 600~9600bps Optional
4
“ to Switch
“ to Confirm
[Co *] Comunicate Mode Setting: 1~6 Optional, Detailed format followed
5
High Setting
[H*****] High Setting:
Press “ “ the flash digit will move towards right Press “ “ to increase the number of flash digit Press “ “ to confirm and enter next step such as 2000
6
Low Setting
[ L*****] Low Setting:
Press ““ the flash digit will move towards right Press “ “ to increase the number of flash digit Press “ “ to confirm and enter next step such as 1000
If weight is higher than High setting the “HI” light on left side of the indicator will be on If weight is lower than low setting the “LO” light on left side of the indicator will be on If the weight is between High and Low setting the the “OK” light will be on
After enter calibration and display CAL, press “” button, indicator displays Zero then press “” enter to other parameter calibration setting, the detailed steps as below:
Step Operation Display Specification
Zero
Means enter other parameter calibration status, press“↙”button to next step
1
to
switch
to
confirm
Zot *.*
Zero trace rangeZero Trace04d
When the object weight is within the set division, the indicator will return back to zero automatically
2
to
switch
to
confirm
nt **
Manual set zero rangeManual Set Zero):0
241020100 % of full capacity;
8
3
to
switch
to
confirm
At **
Power on auto set zero rangeAuto Set Zero
0241020100 % of full capacity
4
to
switch
to
confirm
FL ***
Filter setFilterStb】:stable algorithmStable) SEn】:sensitive algorithmSensitive); 3】:strong vibration resistance
algorithm 4】:high accuracy algorithm
COMMUNICATION MODE FORMAT
Serial Number of each frame Note
1
8
Reversely send the Net Weight date. For example if the net weight is 23.45 kg. ASCII code 54.3200 will be sent. And if the net weight is -
23.45kg, ASCII code 54.320- will be sent
2 8 Reversely send Gross Weight date. The format is same as
serial 1
3
14
Positively send the Net Weight date with unit. For example if the netweight is 23.45kg, ASCII code = 0023.45 (kg) will be sent. End with Hex number 0D,0A
4 14 Positively send the Net weight date with unit. The format is
same as serial 3
5
No
Confirm
Order Response Mode: Order mode 02 “Order” 03 (Hex) There are 5 pices order, ASCII code ‘A’~’E’. Take gross weight
23.45 kg, net weight 13.45kg and tare 10.00 kg for example ‘A’: Read gross weight, indicator back: GW:0023.45(kg) ‘B’: Read net weight, indicator back: NW:0013.45(kg) ‘C’: Read tare, indicator back: TW:0010.00(kg) ‘D’: Manually zero, indicator back: ‘D’ ‘E’: Tare operation, indicator back: ‘E’ All order back add 02 at the beginning and 03 at the end (HEX)
6 Net and total weight can be automatically output in Total
operation and the date can be printed if connected with serial printer.
9
6.0 ERROR NOTICE
Display Note Err 01 Exceed the Zero Range Err 02 Not Meet the Requirement of Total Err 03 Weight Overloaded Err 04 Weight not stable during Calibration Err 05 Load Calibratin Error. Too low load or
Calibration code too small or AD everse
Err 09 Data read verify Error, Data Memory Damaged Err 10 Boot verift Error, SCM damaged
K1 INBOX
No Name PCS
1 Indicator 1 2 Adapter 1 3 Communication Plug (9 Pin) 1 4 Manual 1 5 Certificate 1
10
www.kobastar.com
KOBASTAR CENTER OFFICE
Adress : Fevzi Çakmak Mah. Ayyıldız Cad. No:16/F Karatay/Konya / TURKEY Phone : +90 332 249 38 15
+90 332 249 38 16
Fax :+90 332 249 38 17 E-mail : bilgi@kobastarloadcell.com