Knick Stratos Pro A*** COND Series, Stratos Pro A211N COND, Stratos Pro A201N COND, Stratos Pro A211X COND Instruction Manual

Stratos®Pro A2... COND
Instruction Manual
Latest Product Information:
www.knick.de
Warranty
Warranty
Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender). Sensors,  ttings, and accessories: 1 year.
Subject to change without notice.
Return of Products Under Warranty
Please contact our Service Team before returning a defective device. Ship the cleaned device to the address you have been given. If the device has been in contact with process  uids, it must be decontaminated/disinfected before shipment. In that case, please attach a corresponding certi cate, for the health and safety of our service personnel.
Disposal
Please observe the applicable local or national regulations concerning the disposal of “waste electrical and electronic equipment”.
2

Documents Supplied

CD-ROM
Complete documentation:
• Instruction manuals
• Safety instructions
• Short instructions
Stratos® Pro Series
Safety Instructions
www.knick.de
Stratos® Pro A211 / A411
Short Instructions .........................3
Kurbersicht ..............................15
QuickStart ....................................27
Быстрый старт ...........................39
Inicio rápido ................................51
Início rápido ................................63
Guida rapida ................................75
Snabbstart....................................87
QuickStart ....................................99
Other languages: www.knick.de
Safety Information
In o cial EU languages and others.
• ATEX / IECEX / FM / CSA
• EC Declarations of Conformity
Short Instructions
In German, English, French, Russian, Spanish, Portuguese, Swedish, and Dutch. More languages on CD-ROM and on our website: www.knick.de
• Installation and commissioning
• Operation
• Menu structure
• Calibration
• Error messages and recommended actions
Speci c Test Report
3

Contents

Documents Supplied ..................................................................... 3
Introduction .................................................................................... 7
Intended Use ......................................................................................... 7
Safety Information ......................................................................... 8
Overview of Stratos Pro A2... COND .........................................10
Assembly ........................................................................................ 11
Package Contents ..............................................................................11
Mounting Plan, Dimensions ..........................................................12
Pipe Mounting, Protective Hood ..................................................13
Panel Mounting ..................................................................................14
Installation ..................................................................................... 15
Installation Instructions ...................................................................15
Rating Plates / Terminal Assignments ........................................16
Wiring of Stratos Pro A201/A211 COND ....................................17
Wiring Examples .................................................................................18
Connection of Memosens Sensor ................................................ 25
User Interface, Keypad ................................................................26
Display ............................................................................................27
Signal Colors (Display Backlighting) ........................................... 27
Measuring Mode...........................................................................28
Selecting the Mode / Entering Values ......................................29
Operating Modes ..........................................................................30
Menu Structure of Modes and Functions ................................. 31
HOLD Mode .........................................................................................32
Alarm ...................................................................................................... 33
Configuration ................................................................................34
Menu Structure of Configuration.................................................34
Parameter Set A/B ..............................................................................36
4
Contents
Configuration (Original for Copy) .............................................41
Sensor .....................................................................................................44
Current Output 1 ................................................................................50
Current Output 2 ................................................................................56
Temperature Compensation .......................................................... 58
Alarm Settings ..................................................................................... 62
Time and Date ..................................................................................... 64
Tag Number
Digital Sensors ..............................................................................66
Operation .............................................................................................. 66
Connecting a Digital Sensor .......................................................... 67
Sensor Replacement .........................................................................68
Calibration .....................................................................................71
Selecting a Calibration Mode ........................................................ 71
Calibration with Calibration Solution ......................................... 72
Calibration by Entry of Cell Constant ......................................... 74
Product Calibration ........................................................................... 75
Temp Probe Adjustment ................................................................. 77
Measurement ................................................................................78
Diagnostics .................................................................................... 79
Service ............................................................................................ 84
USP Function .................................................................................87
Operating States ...........................................................................89
Product Line and Accessories ....................................................90
A201/A211X: Supply Units and Connection ...........................91
Specifications ................................................................................92
Calibration Solutions ................................................................... 98
.................................................................................................64
5
Contents
Concentration Curves ................................................................100
Error Handling .............................................................................105
Error Messages ............................................................................106
Sensoface .....................................................................................108
EC Declaration of Conformity ..................................................110
Stratos Pro A2... X COND: Control Drawings .........................112
FM Control Drawing ...................................................................114
CSA Control Drawing .................................................................115
FDA 21 CFR Part 11 ....................................................................116
Electronic Signature – Passcodes ...............................................116
Audit Trail ............................................................................................116
Index .............................................................................................117
Trademarks .........................................................................................123
Passcodes .....................................................................................124
6

Introduction

Intended Use
Stratos Pro A2... COND is used for measurement of electrical conductivity and temperature in liquids. Fields of application are: biotechnology, chemical industry, environment, food processing, water/waste-water treatment. The sturdy molded enclosure can be  xed into a control panel or mounted on a wall or at a post. The protective hood, which is avail­able as accessory, provides additional protection against direct weather exposure and mechanical damage. The device has been designed for 2- and 4-electrode sensors, particu­larly for Models SE 600, SE 603, SE 604, SE 610, SE 620, SE 630 and for Memosens sensors. Plain-text messages in a large, backlit display allow intuitive opera­tion. The colored display backlighting signals alarm messages (red) or HOLD mode (orange). The “Sensocheck“ automatic monitoring of sensor and cables and the “Sensoface“ function for clear indication of the sensor condition provide excellent diagnostics. The internal logbook (TAN SW-A002) can handle up to 100 entries – up to 200 with AuditTrail (TAN SW-A003). The device provides two parameter sets which can be switched manually or via a control input for di erent process adaptations or di erent process conditions (e.g. beer and CIP). Password protection for granting access rights during operation can be con gured. Two  oating, digital control inputs (“Hold“ and “Control“) are available for external control. The device provides two current outputs (for transmission of measured value and temperature, for example).
Approvals for Measurement in Hazardous Locations:
Stratos Pro A201/A211N COND: General Safety.
Stratos Pro A201/A211X COND: Approved for operation in
hazardous locations according to IECEx / ATEX / FM / CSA.
7

Safety Information

Safety information – Be sure to read and observe the following instructions!
The device has been manufactured using state of the art technology and it complies with applicable safety regulations. When operating the device, certain conditions may nevertheless lead to danger for the operator or damage to the device.
See also separate document:
“Safety Instructions“ • (EC Declaration of Conformity, FM, CSA, ATEX (if applicable) Certi cates)
CAUTION!
Commissioning must only be performed by trained personnel autho­rized by the operating company! Whenever it is likely that protection has been impaired, the device shall be made inoperative and secured against unintended operation. The protection is likely to be impaired if, for example:
the device shows visible damage• the device fails to perform the intended measurements• after prolonged storage at temperatures above 70°C after severe transport stresses
Before recommissioning the device, a professional routine test must be performed. This test must be carried out at the manufacturer's factory.
Please note:
Before commissioning it must be proved that the device may be connected with other equipment.
8
Safety Information
Information for Installation in Hazardous Locations (Stratos Pro A201/A211X COND)
Be sure to observe the stipulations of EN 60079-10 / EN 60079-14 • or the corresponding local regulations during installation and commissioning. See also separate “Safety Instructions“ document.
Approvals for application in hazardous locations (Stratos Pro A201/A211X COND)
according to IECEx in Zone 0, 1, 20, 21• according to ATEX in Zone 0, 1, 2 , 20, 21• according to • according to FM in Class I, Div 1, 2 / Zone 1
Important Notice: The operator must indicate the type of protection!
When the device provides di erent types of protection, the operator must specify the applied type of protection during installation. To do so, use the checkboxes on the rating plate:
Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation
cCSAus in Class I Div 1, 2 / Zone 1
Terminals:
Screw terminal, suitable for single wires /  exible leads up to 2.5 mm (AWG 14). Recommended torque for the terminal screws: 0.5 ... 0.6 Nm.
2
9
Overview

Overview of Stratos Pro A2... COND

I hi
V hi
V lo
I lo
RTD (GND)
RTD
RTD (Sense)
Shield
Input +
Input –
HOLD
HOLD
CONTROL
COND
A
input
B
C
D
E
F
G
H
Current
5
input
6
HOLD
10
11
13
input
Control
Output 1
Output 2
RS 485
+ Output 1, 2 / HART
8
– Output 1 / HART
9
– Output 2
17
1
Power supply
RS 485 A
2
RS 485 B
3
GND/Shield
4
CONTROL
10
14
input
Assembly

Package Contents

Check the shipment for transport damage and completeness!
The package should contain:
Front unit, rear unit, bag containing small parts• Specific t Documentation (cf Pg 3)• CD-ROM
est report
11
10
9
8
76 5
Fig.: Assembling the enclosure
Jumper (3 x)1)
asher (1 x), for conduit 2)
W mounting: Place washer between enclosure and nut Cable tie (3 x)3) Hinge pin (1 x), insertable 4) from either side Enclosure screw (4 x)5)
1
2
3
4
Sealing insert (1 x)6) Rubber reducer (1 x)7) Cable gland (3 x)8) Filler plug (3 x)9) Hexagon nut (5 x)10) Sealing plug (2 x), for sealing 11) in case of wall mounting
11
Assembly

Mounting Plan, Dimensions

148
148
117
41
14
34
42
42
42
80
1
Cable gland (3 x)1) Knockouts for cable gland or 2)
74
6.2
½" conduit,
21.5 mm dia. (2 knockouts)
3
Conduits not included! Knockout for pipe mounting 3)
4
(4 x) Knockout for wall mounting 4) (2 x)
Fig.: Mounting plan (All dimensions in mm!)
12
21
43
2
Pipe Mounting, Protective Hood
ø40...ø60
1
2
3
4
Assembly
Hose clamp with worm gear 1) drive to DIN 3017 (2 x) Pipe-mount plate (1 x)2) For vertical or horizontal posts 3) or pipes Self-tapping screw (4 x)4)
Fig.: ZU 0274 pipe
-mount kit (All dimensions in mm!)
147199
185
91
Fig.: ZU 0737 protective hood for wall and pipe mounting
(All dimensions in mm!)
13
Assembly
1...22
Panel Mounting
<30 76 31
Circumferential sealing 1) (1 x)
crew (4 x)2)
S Position of control panel3) Span piece (4 x)4) Threaded sleeve (4 x)5)
Cutout 138 x 138 mm (DIN 43700)
1...22
Fig.: ZU 0738 panel-mount kit (All dimensions in mm!)
1
2345
14
Installation

Installation Instructions

Installation of the device must be carried out by trained experts in • accordance with this instruction manual and as per applicable local and national codes. Be sure to observe the technical speci cations and input ratings • during installation! Be sure not to notch the conductor when stripping the insulation!• The supplied current must be galvanically isolated. If not, connect • an isolator module. All parameters must be set by a system administrator prior to • commissioning!
Terminals:
suitable for single wires /  exible leads up to 2.5 mm
Additional safety precautions have to be taken for applications in hazardous locations according to IECEx, ATEX, FM, CSA! (See also separate “Safety Instructions“ document.)
2
(AWG 14)
15
Installation

Rating Plates / Terminal Assignments

Fig.: Terminal assignments of Stratos Pro A2...
Fig.: Stratos Pro A2...N rating plate at outside bottom of front
Fig.: Stratos Pro A2...X rating plate at outside bottom of front
Important Notice: The operator must indicate the type of protection!
When the device provides di erent types of protection, the operator must specify the applied type of protection during installation. To do so, use the checkboxes on the rating plate. See also “Safety Information“ chapter.
Fig.: Example of additional approval plate ( The speci cations refer to the respective device.
cCSAus, FM)
16

Wiring of Stratos Pro A201/A211 COND

Areas for placing the screwdriver to pull out the terminals
1
9
HART
10 18
Installation
Sensor connection
MK-COND module
AI hi
BV hi
CV lo
DI lo
E RTD (GND)
FRTD
G RTD (Sense)
H Shield
Terminal row 1
1 supply
2 RS 485 A
3 RS 485 B
4 GND/shield
5 + input
6 – input
7 PA (equipot.
bonding)
8 + out 1,2/HART
9 – out1/HART
Terminal row 2
10 hold
11 hold
12 n.c.
13 contr.
14 contr.
15 n.c.
16 n.c.
17 – out 2
18 n.c.
In addition: 2 HART pins (between terminal row 1 and 2)
Fig.: Terminals, device opened, back of front unit
Fig.: MK-COND module terminal assignments
17

Wiring Examples

Example 1:
Measuring task: Conductivity, temperature
Sensors (principle): 4 electrodes
I hi
U hi
U lo
B C D E F G H
A
I lo
RTD
RTD (GND)
RTD (sense)
shield
Shield
Place jumper across F and G when a 2-wire temperature probe is used!
18
Sensor(s) Cable Device
Wiring Examples
Example 2:
Measuring task: Conductivity, temperature Sensors (principle): 2 electrodes, coaxial
I hi
U hi
U lo
B C D E F G H
A
I lo
RTD
Jumper!Jumper!Jumper!
RTD (sense)
shield
Shield
RTD (GND)
Sensor(s) Cable Device
19
Wiring Examples
Example 3:
Measuring task: Conductivity, temperature
Sensors (example): SE 604 (Knick)
Cable: Schaltbau cable
I hi
U hi
U lo
B C D E F G H
A
I lo
RTD
RTD (GND)
RTD (sense)
shield
White
Pink
1 4 3 6
5 2
20
Brown
Green
Yellow
Red
Shield
Black
Gray
Blue
A
Schaltbau
A
5
6
4
1
2
3
connector
Sensor(s) Cable Device
Wiring Examples
Example 4:
Measuring task: Conductivity, temperature
Sensors (example): SE 630 (Knick)
Connection via GDM connector
I hi
U hi
U lo
Green
I lo
White
B C D E F G H
A
Brown
Yellow
2 1 3
RTD
Pink
RTD (sense)
shield
Shield
Red
RTD (GND)
Gray
Black
1
3
2
GDM connection
Sensor(s) Cable Device
21
Wiring Examples
Example 5:
Measuring task: Conductivity, temperature
Sensors (example): SE 600 / SE 603 4-EL fringe- eld sensor (Knick)
I hi
U hi
U lo
Blue
I lo
Red
Brown
B C D E F G H
A
Green
Pink
White/Green
RTD
Yellow
RTD (sense)
Shield
Green
shield
Yellow/Green
RTD (GND)
22
Sensor(s) Cable Device
Wiring Examples
Example 6:
Measuring task: Conductivity, temperature
Sensor: Memosens
Brown
Green
The Memosens sensor is connected to the RS-485 interface of the device – for an A2... Series (2-wire) device, the measuring module slot must be empty. Therefore,  rst remove the measuring module from the slot (see next page). The connected Memosens sensor is automati­cally recognized during start-up of the transmitter.
White
Yellow
Shield
Cable Device
Sensor
23
24
Connection of Memosens Sensor
Areas for placing the screwdriver to pull out the terminals
Caution!
The slot for the MK-COND module must be empty – be sure to remove the module!
1
9
HART
10 18
Connection of Memosens: Wire color
1 Power supply Brown
2 RS 485 A Green
3 RS 485 B Yellow
4 GND/shield White, transparent shield
25

User Interface, Keypad

1
2
3
4
Key Function
meas Return to last menu level
info Retrieve information
enter Con guration: Con rm entries,
Arrow keys up / down
Arrow keys left / right
Directly to measuring mode (press > 2 s)
Show error messages
next con guration step Calibration:• Continue program  ow Measuring mode: Display output current
Measuring mode: Call menu• Menu: Increase/decrease a numeral• Menu: Select
Measuring mode: Call menu• Menu: Previous/next menu group• Number entry: Move between digits
1 IrDA transmitter/receiver 2 Display 3 Keypad 4 Rating plate (bottom)
26
123 4 5 67 891011
23
1 Temperature 2 Sensocheck 3 Interval/response time 4 Sensor data
5 Not used 6 Limit values 7 Alarm 8 Service 9 Parameter sets A/B
10 Calibration 11 Digital sensor 12 Waiting time running
13 Info available 14 HOLD mode active 15 Main display 16 Secondary display 17 Proceed with enter 18 Not used 19 Diagnostics
20 Con guration mode 21 Calibration mode 22 Measuring mode 23 Sensoface 24 Measurement symbol

Display

12 13 14
24
15
16
171819202122
Signal Colors ( Display Backlighting)
Red Alarm Orange HOLD mode (Calibration, Con guration, Service) Turquoise Diagnostics Green Info Purple Sensoface message
27

Measuring Mode

After the operating voltage has been connected and the sensor identi­ ed, the device automatically goes to “Measuring“ mode. To call the measuring mode from another operating mode (e.g. Diagnostics, Service): Hold meas key depressed (> 2 s).
Sensoface indicator (sensor status)
Time
Mode indicator (measuring)
Hold meas key depressed to call the measuring mode (pressing once more switches the display)
In measuring mode the display indicates:
Measured value and time (24/12 h AM/PM) as well as temperature
in °C or °F (formats selected during con guration) By pressing the meas key in measuring mode you can view the following displays (for approx. 60 sec):
Measured value and selection of parameter set A/B
(if set to “Manual“)
Measured value and tag (point of measurement designation –
entered during con guration)
Time and date• Pressing the enter key shows the output currents. They are displayed as long as enter is held depressed, then the measured-value display will return after 3 sec.
The device must be con gured for the respective measurement task!
Active parameter set
Process variable
Temperature
enter key
28

Selecting the Mode / Entering Values

To select the operating mode:
Hold 1) meas key depressed (> 2 s) (directly to measuring mode) Press any arrow key: the selection menu appears2) Select operating mode using left / right arrow key3) Press 4) enter to con rm the selected mode
Selection menu
Selected mode (blinks)
1
To enter a value:
Select numeral: left / right arrow key5) Change numeral: up / down arrow key6) Con rm entry by pressing 7) enter
5
2
3
4
6
7
29

Operating Modes

Diagnostics
Display of calibration data, display of sensor data, performing a device self-test, viewing the logbook entries, display of hardware/software versions of the individual components. The logbook can store 100 events (00...99). They can be displayed directly on the device. The logbook can be extended to 200 entries using a TAN (Option).
HOLD
Manual activation of HOLD mode, e.g. for servicing. The signal outputs adopt a de ned state.
Calibration
Every sensor has typical characteristic values. Calibration is required to supply a correct measured value. The device checks which value the sensor delivers when measuring in a known solution. When there is a deviation, the device can be “adjusted“. In that case, the device displays the “actual“ value and internally corrects the measurement error of the sensor. During calibration the device is in HOLD mode.
During calibration the analyzer remains in the HOLD mode until it is stopped by the operator.
Con guration
The analyzer must be con gured for the respective measurement task. In the “Con guration“ mode you select the connected sensor, the measuring range to be transmitted, and the conditions for warning and alarm messages. During con guration the device is in HOLD mode.
Con guration mode is automatically exited 20 minutes after the last keystroke. The device returns to measuring mode.
Service
Maintenance functions (current source assignment, reset to factory settings, enabling of options (TAN).
), IrDA operation, passcode
30

Menu Structure of Modes and Functions

Measuring mode
(Access via
code, factory
setting:
5555)
meas
TAG display
Pressing any arrow key opens the selection menu.
elect the menu group using the left/right arrow keys.
S Press enter to open a menu. Press meas to return.
Manual activation of HOLD mode, e.g. for sensor replacement.
he signal outputs behave as configured (e.g. last measured value, 21 mA)
T
meas
after 60 s after 60 s
Display of calibration data
Display of sensor data
Self test: RAM, ROM, EEPROM, module
100 ev
Display of direct, uncorrected sensor signals
Display of software version, model designation, serial number
Calibration with calibration solution
Calibration by entry of cell constant
Product calibration
djustment of temperature probe
A
Configuring parameter set A
Configuring parameter set B
Display of measured values for validation (simulators)
Current source, output 1
urrent source, output 2
C
Activating the IrDA interface
Specifying ac
Reset to factory setting
Enabling an option via TAN
CLK display
ents with date and time
cess codes for operating modes
meas
31

HOLD Mode

The HOLD mode is a safety state during con guration and calibration. Output current is frozen (Last) or set to a  xed value (Fix). The HOLD mode is indicated by orange display backlighting.
HOLD mode, display icon:
Output Signal Response
Last: The output current is frozen at its last value. Recommended
for short con guration procedures. The process should not change
decisively during con guration. Changes are not noticed with this
setting!
Fix: The output current is set to a value that is noticeably di erent
from the process value to signal the control system that the device
is being worked at.
Output Signal During HOLD:
Output current [mA]
21
Output signal for HOLD FIX setting = 21.0 mA
Output signal for HOLD LAST setting
4
HOLD active
HOLD active
Terminating the HOLD Mode
The HOLD mode is ended by switching to measuring mode (hold meas key depressed). The display reads “Good Bye“, after that, the HOLD mode is exited. When the calibration mode is exited, a con rmation prompt ensures that the installation is ready for operation (e.g.: sensor reinstalled, located in process).
32

Alarm

External Activation of HOLD
The HOLD mode can be activated from outside by sending a signal to
the Hold input (e.g. from the process control system).
Power supply
HOLD
10
12...24 V AC/DC
input
Stratos Pro A2...
HOLD inactive 0...2 V AC/DC HOLD active 10...30 V AC/DC
Manual Activation of HOLD
The HOLD can be activated manually from the HOLD menu.
This allows checking or replacing a sensor, for example, without
provoking unintended reactions of outputs or contacts.
Press meas key to return to selection menu.
Alarm
When an error has occurred, Err xx is displayed immediately.
Only after expiry of a user-de ned delay time will the alarm be
registered and entered in the logbook.
During an alarm the display blinks, the display backlighting turns red.
Error messages can also be signaled by a 22 mA output current
(see Con guration).
2 sec after the failure event is corrected, the alarm status will be
deleted.
11
Process control system
(PCS)
33

Configuration

Menu Structure of Con guration
The device provides 2 parameter sets “A“ and “B“. By switching between the parameter sets you can adapt the device to di erent measurement situations, for example. Parameter set “B“ only permits setting of process-related parameters. The con guration steps are assigned to di erent menu groups. Using and , you can jump between the individual menu groups. Each menu g Pressing enter opens a menu item. The values are edited using  and .
ressing enter con rms/stores the settings.
P Return to measurement: Press meas.
Select menu group
roup contains menu items for setting the parameters.
Menu group Code Display
Sensor selection SNS:
Menu item 1
...
Menu item ...
Current output 1 OT1:
Select menu item
Current output 2 OT2:
Compensation COR:
Alarm mode ALA:
Setting the clock CLK:
Tag number TAG:
enter
enter
enter
enter
34
Con guration
Parameter Set A/B: Con gurable Menu Groups
(Some parameters are identical in A and B. They are con gured in
parameter set A only.)
Menu group Parameter set A Parameter set B
SENSOR Sensor selection --­OUT1 Current output 1 Current output 1 OUT2 Current output 2 Current output 2 CORRECTION Compensation Compensation ALARM Alarm mode Alarm mode PARSET Parameter set
selection CLOCK Setting the clock --­TAG Tag number ---
---
35
Con guration
Parameter Set A/B
Manual selection
Display Action Remark
To switch between parameter sets: Press meas
PARSET blinks in the lower line. Select parameter set using andkeys
Select PARSET A / PARSET B
Con rm by pressing
enter
Cancel by pressing
meas
Manual selection of parameter sets must have been preset in CONFIG mode. Default setting is a fixed parameter set A. Wrong settings change the measurement proper­ties!
36
Con guration
Con guration Choices Default
SENSOR
SNS: 2-ELECTRODE
2-EL / 4-EL CELLFACTOR 1)00.0000 -
MEAS MODE Cond
Cond MEAS RANGE
Conc Solution -01- (NaCl)
TEMP UNIT °C / °F °C
TEMPERATURE AUTO
AUTO RTD TYPE 100 PT
MAN TEMPERATURE –50...200 °C
4-ELECTRODE MEMOSENS
19.9999 c
Conc % Sal ‰ USP μS/cm
2)
x.xxx μS/cm xx.xx μS/cm xxx.x μS/cm xxxx μS/cm x.xxx mS/cm xx.xx mS/cm xxx.x mS/cm x.xxx S/m xx.xx S/m xx.xx MΩ
-02- (HCl)
-03- (NaOH)
-04- (H
-05- (HNO3)
MAN EXT via TAN)
1000 PT
8.55 NTC 30 NTC
(–58...392 °F)
2SO4
(only if enabled
2-ELECTRODE
01.0000 c
COND
xxx.x mS/cm
-01- (NaCl)
)
AUTO
100 PT
025.0 °C (077.0 °F)
37
Con guration
Con guration Choices Default
SENSOR
SNS: CIP COUNT ON/OFF OFF
SIP COUNT ON/OFF OFF
Output 1 (OUT1)
OT1: CHANNEL COND/TMP COND
OUTPUT (with Cond only) LIN / LOG LIN
LIN BEGIN 4mA xxxx 000.0 mS/cm
END 20 mA xxxx 100.0 mS/cm
LOG BEGIN 4mA Decades
END 20 mA Decades
TMP °C
TMP °F
FILTERTIME 0...120 SEC 0000 SEC
22mA-FAIL ON/OFF OFF
HOLD MODE LAST/FIX LAST
FIX HOLD-FIX 4...22 mA 021.0 mA
BEGIN 4mA 50...200 °C
END 20 mA 50...200 °C
BEGIN 4mA 58...392 °F
END 20 mA 58...392 °F
With Memosens, the cell constant is automatically loaded from the sensor. When 1)
ching from Memosens to 2-/4-electrode sensor, the cell constant is set to the
swit default value 01.0000 c and then must be entered manually. The range selection allows selecting the maximum resolution. If the upper limit 2) of this range is exceeded, the device automatically switches to the next higher range.
38
Con guration
Con guration Choices Default
Output 2 (OUT2)
OT2: CHANNEL COND/TMP TMP
... other steps like output 1
Temperature compensation (CORRECTION)
COR: TC SELECT OFF
LIN NLF NaCl HCL NH3
LIN TC LIQUID 00.00...19.99%/K 00.00%/K
I-INPUT 0...20 mA/4...20 mA 4...20 mA
°C BEGIN 4 mA 50...200 °C 000.0 °C
END 20 mA 50...200 °C 100.0 °C
°F BEGIN 4 mA 58...392 °F
END 20 mA 58...392 °F
Alarm (ALARM)
ALA: DELAYTIME 0...600 SEC 0010 SEC
SENSOCHECK ON/OFF OFF
Begin: 0 °C End: 100 °C
OFF
39
Con guration
Con guration Choices Default
Parameter set (PARSET)
PAR Select fixed parameter set
(A) or switch between A/B via control input or manu­ally in measuring mode
Real-time clock (CLOCK)
CLK: FORMAT 24 h / 12 h
24 h TIME hh/mm 00..23:00...59
12 h TIME hh/mm 00...11:00...59
DAY/MONTH 01...31/01...12
YEAR 2000...2099
Tag number (TAG)
TAG: (Input in text line) ___
PARSET FIX / CNTR INPUT / MANUAL
AM/PM:
PARSET FIX (fixed parameter set A)
40
Configuration (Original for Copy)
Default Settings of Parameter Sets
Two complete parameter sets are stored in the EEPROM. As delivered, the two sets are identical but can be edited.
Please note:
Fill in your con guration data on the following pages or use them as original for copy.
41
Con guration (Original for Copy)
Parameter Parameter set A Parameter set B
SNS: Sensor type
SNS: Cell constant
SNS: Measuring mode
SNS: Measuring range
SNS: Concentration determination
SNS: Temperature unit
SNS: Temp detection
SNS: Manual temp
SNS: RTD type
SNS: CIP counter
SNS: SIP counter
OT1: Process variable
OT1: LIN/LOG output
OT1: Current start
OT1: Current end
OT1: Filter time
OT1: 22 mA error current
OT1: HOLD mode
OT1: HOLD-FIX current
---
---
---
---
---
---
---
---
---
---
---
*)
*) These parameters cannot be adjusted in parameter set B, the values are the same as in parameter set A.
42
(Original for Copy) Con guration
Parameter Parameter set A Parameter set B
OT2: Process variable
OT2: LIN/LOG output
OT2: Current start
OT2: Current end
OT2: Filter time
OT2: 22 mA error current
OT2: HOLD mode
OT2: HOLD-FIX current
COR: TC SELECT
COR: Temp coefficient
COR: Current range
COR: Current start
COR: Current end
ALA: Delay
ALA: Sensocheck on/off
CLK: Time & Date
TAG: Tag number
*) These parameters cannot be adjusted in parameter set B, the values are the same as in parameter set A.
*)
---
*)
---
43
Con guration
Sensor
Selecting the parameters
6
1
2
3
4
5
enter
enter
enter
meas
Press any arrow key.1 Select 2 CONF using  keys,
press enter.
Select parameter set using 3  keys,
press enter.
Select 4 SENSOR menu using  keys,
press enter.
All items of this menu group are indicated by 5
the “SNS:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.
End: Press 6 meas key until the [meas] mode
indicator is displayed.
5
Select sensor type
Enter cell constant
Select measuring mode
Select range
Concentration determination
Temperature unit
Temperature detection
Select type of temp probe
Cleaning cycles
Sterilization cycles
enter
44
5
Con guration
Menu item Action Choices
Select sensor type
Select sensor type using
 keys.
Press enter to confirm.
2-ELECTRODE
4-ELECTRODE MEMOSENS
Enter cell constant
Select meas. mode
Select range
Concentration determination
Modify digit using 
keys,
select next digit using
 keys.
Press enter to confirm.
Select desired measuring
mode using  keys.
Press enter to confirm.
For cond measurement
only
Select desired range
using  keys.
Press enter to confirm.
For conc measurement
only
Select desired concentra-
tion solution using 
keys.
Press enter to confirm.
00.0000...19.9999 c
(01.0000 c)
Cond
Conc % Sal ‰ USP μS/cm
x.xxx μS/cm, xx.xx μS/cm xxx.x μS/cm, xxxx μS/cm x.xxx mS/cm, xx.xx mS/cm xxx.x mS/cm, x.xxx S/m xx.xx S/m, xx.xx MΩ
-01- (NaCl)
-02- (HCl)
-03- (NaOH)
-04- (H2SO4)
-05- (HNO3)
45
Con guration
Sensor
Select: Temperature unit, temperature detection,
type of temp probe
1
2
enter
3
Press any arrow key.1 Select 2 CONF using  keys,
press enter.
Select parameter set using 3  keys,
press enter.
Select 4 SENSOR menu using  keys,
press enter.
All items of this menu group are indicated by 5
the “SNS:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.
End: Press 6 meas key until the [meas] mode
indicator is displayed.
46
6
4
5
enter
enter
meas
Select sensor type
Enter cell constant
Select measuring mode
Select range
Concentration determination
Temperature unit
Temperature detection
Select type of temp probe
Cleaning cycles
Sterilization cycles
5
enter
5
Con guration
Menu item Action Choices
Temperature unit
Select °C or °F using 
keys.
Press enter to confirm.
°C / °F
Temp detection
(Manual temperature)
Select type of temp probe
Select mode using :
AUTO: Measured by
sensor
MAN: Direct input of
temperature, no measure-
ment (see next step)
EXT: Temperature speci-
fied via current input
(only if TAN E enabled)
Press enter to confirm.
Modify digit using
 keys,
select next digit using
 keys.
Press enter to confirm.
(not for Memosens)
Select type of tempera-
ture probe using 
keys.
Press enter to confirm.
AUTO
MAN EXT
50...200 °C (–58...+392 °F)
100 PT
1000 PT 30 NTC
8.55 NTC
47
Con guration
Sensor
Adjust: Cleaning cycles, sterilization cycles
6
1
2
3
4
5
enter
enter
enter
meas
Press any arrow key.1 Select 2 CONF using  keys,
press enter.
Select parameter set using 3  keys,
press enter.
Select 4 SENSOR menu using  keys,
press enter.
All items of this menu group are indicated by 5
the “SNS:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.
End: Press 6 meas key until the [meas] mode
indicator is displayed.
Select sensor type
Enter cell constant
Select measuring mode
Select range
Concentration determination
Temperature unit
Temperature detection
Select type of temp probe
Cleaning cycles
Sterilization cycles
5
enter
48
5
Con guration
Menu item Action Choices
CIP / SIP
Cleaning cycles On / O
Select ON or OFF using
 keys.
Activates/deactivates log-
ging in extended logbook
Press enter to confirm.
ON/OFF
Sterilization cycles On / O
Select ON or OFF using
 keys.
Activates/deactivates log-
ging in extended logbook
Press enter to confirm.
ON/OFF
Please note:
A CIP or SIP cycle is only entered into the logbook 2 hours after the start to ensure that the cycle is complete.
49
Con guration
Current Output 1
Process variable, current start, current end
6
1
2
3
4
5
enter
enter
enter
meas
Press any arrow key.1 Select 2 CONF using  keys,
press enter.
Select parameter set using 3 ,
press enter.
Select 4 OUT1 menu using  keys,
press enter.
All items of this menu group are indicated by 5
the “OT1:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.
End: Press 6 meas key until the [meas] mode
indicator is displayed.
5
Process variable LIN/LOG output Current start Current end Time averaging  lter Output current during error
message Output current during HOLD Output current for HOLD FIX
enter
50
5
Con guration
Menu item Action Choices
Process variable
Select using  keys:
Cond: Conductivity
TMP: Temperature
Press enter to confirm.
Cond/TMP
Select LIN/LOG
Current start
Current end
Select using  keys:
LIN: Linear characteristic
LOG: Logarithmic –
See right column for
selectable decades.
Press enter to confirm.
Modify digit using
 keys,
select next digit using
 keys.
Press enter to confirm.
Enter value using 
 keys.
Press enter to confirm.
Selectable decades with logarithmic setting (L
S/cm: 1.0 µS/cm, 10.0 µS/cm,
100.0 µS/cm, 1.0 mS/cm,
10.0 mS/cm, 100.0 mS/cm, 1000 mS/cm S/M: 0.001 S/m, 0.01 S/m,
0.1 S/m, 1.0 S/m, 10.0 S/m, 100 S/m
As selected for process variable/range If the adjusted range is exceeded, the device automatically switches to the next higher range ( Autorange)
As selected for process variable/range If the adjusted range is exceeded, the device automatically switches to the next higher range (Autorange)
Assignment of measured values: Current start and current end
Example 1: Range 0...200 mS/cm
[mS/cm]
200
100
0
Output current
[mA]
20 4
Example 2: Range 100...200 mS/cm Advantage: Higher resolution in range
[mS/cm]
200
100
of interest
Output current
20 4
OG):
[mA]
51
Con guration
Current Output 1
Adjust time interval of output  lter
6
1
2
3
4
5
enter
enter
enter
meas
Press any arrow key.1 Select 2 CONF using  keys,
press enter.
Select parameter set using 3  keys,
press enter.
Select 4 OUT1 menu using  keys,
press enter.
All items of this menu group are indicated by 5
the “OT1:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.
End: Press 6 meas key until the [meas] mode
indicator is displayed.
5
Process variable LIN/LOG output Current start Current end Time averaging  lter Output current during error
message Output current during HOLD Output current for HOLD FIX
enter
52
5
Con guration
Menu item Action Choices
Time averaging  lter
Time Averaging Filter (Attenuation)
To smoothen the current output, a low-pass  lter with adjustable  lter time constant can be switched on. When there is a jump at the input (100 %), the output level is at 63 % after the time interval has been reached. The time interval can be set from 0 to 120 sec. If the time interval is set to 0 sec, the current output directly follows the input.
Please note:
The  lter only acts on the current output, not on the display!
Enter value using   keys.
Press enter to confirm.
0...120 SEC (0000 SEC)
Cond
Time interval 0...120 s
53
Con guration
Current Output 1
Output current during Error and HOLD
6
1
2
3
4
5
enter
enter
enter
meas
Press any arrow key.1 Select 2 CONF using  keys,
press enter.
Select parameter set using 3  keys,
press enter.
Select 4 OUT1 menu using  keys,
press enter.
All items of this menu group are indicated by 5
the “OT1:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.
End: Press 6 meas key until the [meas] mode
indicator is displayed.
5
Process variable LIN/LOG output Current start Current end Time averaging  lter Output current during error
message Output current during HOLD Output current for HOLD FIX
enter
54
5
Con guration
Menu item Action Choices
Output current dur­ing error message
Select ON or OFF using
 keys.
Press enter to confirm.
ON/OFF
Output current during HOLD
Output current for HOLD FIX
LAST: During HOLD the
last measured value is
maintained at the output.
FIX: During HOLD a value
(to be entered) is main-
tained at the output.
Select using 
Press enter to confirm.
Only with FIX selected:
Enter current which is to
flow at the output during
HOLD
Enter value using 
 keys.
Press enter to confirm.
Output Signal During HOLD:
Output current [mA]
21
4
Output signal for HOLD FIX setting = 21.0 mA
HOLD active
LAST/FIX
04.00...22.00 mA (21.00 mA)
Output signal for HOLD LAST setting
HOLD active
55
Con guration
Current Output 2
Output current range, process variable
6
1
2
3
4
5
enter
enter
enter
meas
Press any arrow key.1 Select 2 CONF using  keys,
press enter.
Select parameter set using 3  keys,
press enter.
Select 4 OUT2 menu using  keys,
press enter.
All items of this menu group are indicated by 5
the “OT2:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.
End: Press 6 meas key until the [meas] mode
indicator is displayed.
5
Current range Process variable LIN/LOG output Current start Current end Time averaging  lter Output current during error
message Output current during HOLD Output current for HOLD FIX
enter
56
5
Con guration
Menu item Action Choices
Process variable
Select using  keys:
Cond: Conductivity
TMP: Temperature
Press enter to confirm.
. . .
All the following adjustments are made as for current output 1 (see there)!
Cond/TMP Begin: 0 °C End: 100°C
57
Con guration
Temperature Compensation
Selecting the compensation method
6
1
2
3
4
5
enter
enter
enter
meas
Press any arrow key.1 Select 2 CONF using  keys,
press enter.
Select parameter set using 3  keys,
press enter.
Select 4 CORRECTION menu using  keys,
press enter.
All items of this menu group are indicated by 5
the “COR:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.
End: Press 6 meas key until the [meas] mode
indicator is displayed.
5
Temperature compensation
enter
Current input for external temperature measurement
Current start Current end
58
5
Con guration
Menu item Action Choices
Temperature compensation
Select desired compensa-
tion using  keys:
OFF: Temperature com-
pensation switched off
LIN: Linear temperature
compensation with entry
of temperature coef-
ficient
nLF: Temperature
compensation for natural
waters to EN 27888
NaCl: Temperature
compensation for
ultrapure water with
NaCl traces
HCl: Temperature
compensation for
ultrapure water with
HCl traces
NH3:Temperature com-
pensation for ultrapure
water with NH3 traces
Press enter to confirm.
59
Con guration
Temperature Compensation
TC process medium, current input for temp measurement
6
1
2
3
4
5
enter
enter
enter
meas
Press any arrow key.1 Select 2 CONF using  keys,
press enter.
Select parameter set using 3  keys,
press enter.
Select 4 CORRECTION menu using  keys,
press enter.
All items of this menu group are indicated by 5
the “COR:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.
End: Press 6 meas key until the [meas] mode
indicator is displayed.
5
Temperature compensation
enter
Current input for external temperature measurement
Current start Current end
60
5
Con guration
Menu item Action Choices
Temp compensation, process medium
With linear compensation
only: Enter temperature
compensation of the
process medium.
Enter value using 
 keys.
Press enter to confirm.
0...19.99 %/K
Current range
Current start
Current end
Select desired range
using  keys.
Press enter to confirm.
Modify digit using ,
select next digit using
 keys.
Press enter to confirm.
Enter value using 
 keys.
Press enter to confirm.
4-20 mA / 0-20 mA
Input range: –50...200 °C / –58...392 °F
Input range: –50...200 °C/ –58...392 °F
61
Con guration
Alarm Settings
Delay, Sensocheck
6
1
2
3
4
5
enter
enter
enter
meas
Press any arrow key.1 Select 2 CONF using  keys,
press enter.
Select parameter set using 3  keys,
press enter.
Select 4 ALARM menu using  keys,
press enter.
All items of this menu group are indicated by 5
the “ALA:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.
End: Press 6 meas key until the [meas] mode
indicator is displayed.
5
Delay Sensocheck
enter
62
5
Con guration
Menu item Action Choices
Delay
Enter value using 
 keys.
Press enter to confirm.
0...600 SEC (010 SEC)
Sensocheck
Select Sensocheck
(continuous monitoring
of sensor).
Select ON or OFF using
 keys.
Press enter to confirm.
Error messages can be signaled by a 22 mA output curr
ON/OFF
ent (see Error Messages and Con guration of Output 1/Output 2). The alarm delay time delays the color change of the display backlighting to red and the 22 mA signal (if con gured).
63
Con guration
Time and Date Tag Number
6
1
2
3
4
5
enter
enter
enter
meas
Press any arrow key.1 Select 2 CONF using  keys,
press enter.
Select parameter set A using 3  keys,
press enter. 4 Press enter. Select 5 CLOCK or TAG using  keys,
press enter. All items of this menu group are indicated by 6
the “CLK:” or “TAG” code.
Press enter to select menu,
edit using arrow keys (see next page).
Confirm (and proceed) by pressing enter. End: Press 7 meas key until the [meas] mode
indicator is displayed.
5
Time format Time Day and month Year
Tag number
enter
64
Con guration
Time and Date
Control of the calibration and cleaning cycles is based on the time and date of the integrated real-time clock. In measuring mode the time is shown in the lower display. When using digital sensors, the calibration data is written in the sensor head. In addition, the logbook entries (cf Diagnostics) are provided with a time stamp.
Please note:
After prolonged power outage (> 5 days) the time display is • replaced by dashes and cannot be used for processing. Enter the correct time. There is no automatic switchover from winter to summer time!• Be sure to manually adjust the time!
Tag Number (“ TAG“)
You can enter a designation for the point of measurement (tag number) in the lower display line. Up to 32 digits are possible. Pressing meas (repeatedly) in the measuring mode indicates the tag number. Being part of the device con guration, the “TAG“ can be read out via IrDA. A standardized tag number helps, for example, to correctly re­install a device after repair.
5
Menu item Action Choices
Tag number
Select character using  keys, select next digit using  keys.
Confirm with enter
A...Z, 0...9, – + < > ? / @
The first 10 characters are seen in the display with­out scrolling.
65

Digital Sensors

Operation
Stratos Pro can be operated with digital Memosens sensors. The following display examples refer to a transmitter and a digital sensor (slight variations for other combinations).
The sensor type is selected during Con guration. The device only switches to measuring mode when the connected sensor corresponds to the type con gured (Sensoface is friendly):
Otherwise, an error message is released. The info icon is displayed. You can display the error text in the bottom line using the keys.
ensoface is sad (see table of error messages and Sensoface in the
S Appendix):
66
Digital Sensors
Connecting a Digital Sensor
Step Action/Display Remark
Before a digital sensor is connected, the error message „No sensor“ is
Connect sensor
displayed.
Wait until the sensor data are displayed.
Check sensor data
Go to measuring mode
Possible error messages Sensor defective.
Replace sensor
View sensor infor­mation using 
s, press enter to
key con rm.
Press meas, info, or enter
The hourglass in the display blinks.
Display color changes to green.
Sensoface is friendly when the sensor data are okay.
After 60 sec the device automatically returns to measuring mode (time­out).
When this error message appears, the sensor cannot be used. Sensoface is sad.
67
Digital Sensors
Sensor Replacement
A digital sensor should only be replaced during HOLD mode to pre­vent unintended reactions of the outputs or contacts. When you  rst want to calibrate the new sensor, it can also be replaced in calibration mode.
Step Action/Display Remark
Select HOLD mode Press any key to call
the selection menu, select HOLD using the keys, press
ter to con rm.
en
Disconnect and remove old sensor
Install and connect new sensor.
Wait until the sensor data are displayed.
Now the device is in HOLD mode. The HOLD mode can also be acti­vated externally via the HOLD input. During HOLD the output current is frozen at its last value or set to a fixed value.
Temporary messages which are activated dur­ing the replacement are indicated but not output to the alarm contact and not entered in the log­book.
68
Digital Sensors
Step Action/Display Remark
Check sensor data
View sensor infor­mation using 
s, press enter to
key con rm.
Check measured values
Exit HOLD Hit meas key: Return
to selection menu. Hold meas key depressed: Device switches to measur­ing mode
You can view the sensor type, serial number, and last calibration date.
The sensor replacement is entered in the extended logbook.
69
70

Calibration

Please note:
All calibration procedures must be performed by trained person-• nel. Incorrectly set parameters may go unnoticed, but change the measuring properties.
Calibration can be performed by:
Determining the cell constant with a known calibration solution• Input of cell constant (e.g. for ultrapure-water sensors)• Sampling (product calibration)• Temperature probe adjustment
Selecting a Calibration Mode
Calibration adapts the device to the individual sensor characteristics. Access to calibration can be protected with a passcode (SERVICE menu). First, you open the calibration menu and select the calibration mode:
CAL_SOL Calibration with calibration solution CAL_CELL Calibration by entry of cell constant P_CAL Product calibration (calibration with sampling) CAL_RTD Temperature probe adjustment
71
Calibration with Calibration Solution
Input of temperature-corrected value of calibration solution with simultaneous display of cell constant
Display Action Remark
Select Calibration. Press enter to proceed. Select CAL_SOL calibration method. Press enter to proceed.
Ready for calibration. Hourglass blinks.
Display (3 sec) Now the device is in HOLD mode.
72
Immerse sensor in calibration solution. Enter the temperature­corrected value of the calibration solution using the arrow keys (see table). Press enter to con rm.
The determined cell constant is displayed. The “hourglass” icon is blinking. Press enter to proceed.
Lower line: Display of cell constant and temperature
Calibration with Calibration Solution
Display Action Remark
Display of selected process variable (here: mS/cm). Now the device is in HOLD mode: Reinstall the sensor and check whether the message is OK. MEAS ends calibration, REPEAT permits repetition.
With MEAS selected: End calibration by pressing enter.
Please note:
Be sure to use known calibration solutions and the respective • temperature-corrected conductivity values (see table on calibration solution). During the calibration procedure the temperature must be kept • constant.
Display of conductiv­ity and temperature, Sensoface is active. After end of calibra­tion, the outputs remain in HOLD mode for a short time. After display of GOOD BYE, the device automatically returns to measuring mode.
73
Calibration by Entry of Cell Constant
You can directly enter the value for the cell constant of a sensor. This value must be known, e.g. determined beforehand in the laboratory. The selected process variable and the temperature are displayed.
Display Action Remark
Select Calibration. Press enter to proceed. Select CAL_CELL calibration method. Press enter to proceed.
Ready for calibration. Hourglass blinks.
Display (3 sec) Now the device is in HOLD mode.
74
Enter cell constant. Press enter to proceed.
The device shows the calculated cell constant (at 25 °C). Sensoface is active.
Use the arrow keys to select:
MEAS (end)• REPEAT
Press enter to proceed.
The selected process variable and the temper­ature are displayed.
End: HOLD is deactivated after a short time.
Product Calibration
Calibration by sampling – for product calibration, the uncompensated conductivity (μS/cm, mS/cm, S/m) is used. During product calibration the sensor remains in the process. The measurement process is only interrupted brie y.
Procedure:
The sample is measured in the lab or directly on the site using a por-1) table meter. To ensure an exact calibration, the sample temperature should correspond to the measured process temperature. During sampling the device saves the currently measured value and then returns to measuring mode. Then, the “calibration“ mode indicator blinks. In the second step you enter the measured sample value in the 2) device. From the di erence between the stored measured value and
entered sample value, the device calculates the new cell constant. If the sample is invalid, you can take over the value stored during sampling. In that case the old calibration values are stored. Afterwards, you can start a new product calibration.
Display Action Remark
Select Calibration. Press enter to proceed. Select P_CAL calibration method. Press enter to proceed.
Ready for calibration. Hourglass blinks.
Display (3 sec) Now the device is in HOLD mode.
Take sample and save value. Press enter to proceed.
Now the sample can be measured in the lab.
75
Product Calibration 77Temp Probe Adjustment
Display Action Remark
The device returns to measuring mode.
Product calibration step 2:
When the sample value has been determined, open the product cali­bration once more
The stored value is displayed (blinking) and can be overwritten with the lab value. Press enter to proceed.
Display of new cell con­stant (based on 25°C). Sensoface is active. To end calibration: Select MEAS, then
enter
After calibration is ended, the device will switch to measuring mode.
From the blinking CAL mode indicator you see that product calibration has not been terminated.
Display (3 sec) Now the device is in HOLD mode.
To repeat calibra­tion: Select REPEAT, then enter
After end of calibra­tion, the outputs re­main in HOLD mode for a short time.
76
Temp probe adjustment
Display Action Remark
Select Calibration. Press enter to proceed. Select CAL_RTD calibration method. Press enter to proceed.
Wrong settings change the measurement properties!
Measure the tempera­ture of the process medium using an external thermometer.
Enter the measured temperature value. Maximum di erence: 10 K. Press enter to proceed.
The corrected tempera­ture value is displayed. Sensoface is active. To end calibration: Select MEAS, then enter To repeat calibration: Select REPEAT, then enter
After calibration is ended, the device will switch to measuring mode.
Display (3 sec) Now the device is in HOLD mode.
Display of actual temperature (un­compensated) in the lower display.
After end of calibra­tion, the outputs re­main in HOLD mode for a short time.

Measurement

Display Remark
From the con guration or calibration menus, you can switch the device to measuring mode by pressing the meas key. In the measuring mode the main display
or AM/PM and °F:
Pressing the enter key brie y shows the output currents. By pressing the meas key you can step through the following dis- plays. When no key has been pressed for 60 sec, the device returns to the standard display.
shows the con gured process variable (Cond or temperature), the secondary display shows the time and the second con gured process variable (Cond or temperature). The [meas] mode indicator lights and the active parameter set (A/B) is indicated. A/B is not displayed with parameter set Fix A.
Please note:
After prolonged power outage (> 5 days) • the time display is replaced by dashes and cannot be used for processing. Enter the correct time.
Selecting the parameter set
(if set to “manual“ in the con guration). Select the desired parameter set using thearrow keys (PARSET A or PARSET B blinks in the lo Press enter to con rm.
wer display line).
78
Further displays (each with meas).
Display of tag number (“TAG“)1) Display of 2) time and date
Diagnostics (DIAG)

Diagnostics

In the Diagnostics mode you can access the following menus without interrupting the measurement:
CALDATA Viewing the calibration data SENSOR Viewing the sensor data SELFTEST Starting a device self-test LOGBOOK Viewing the logbook entries MONITOR Displaying currently measured values VERSION Displaying device type, software version, serial number
Access to diagnostics can be protected with a passcode (SERVICE menu).
Please note:
HOLD is not active during Diagnostics mode!
Action Key Remark
Activate Diagnostics
Select diagnostics option
End
meas End by pressing meas.
Press any arrow key to call the selection menu. (Display color changes to turquoise.) Select DIAG using  keys, press enter to confirm.
Use keys to select from: CALDATA SENSOR SELFTEST LOGBOOK MONITOR VERSION See next pages for further proceeding.
79
Diagnostics
Menu item Remark
Display of calibration data
Select CALDATA using, press enter to confirm. Use thekeys to select the desired parameter from the bottom line of the display (LAST_CAL CELLFACTOR ZERO). The selected parameter is shown in the main display.
Press meas to return to measurement.
80
Display Menu item
Device self-test
(To abort, you can press meas.)
1 Display test: Display of all segments with
changing background colors white/green/red. Press enter to proceed.
2 RAM test: Hourglass blinks, then display of --PASS--
or --FAIL--
ter to proceed.
Press en
3 EEPROM test: Hourglass blinks, then display of
--PASS-- or --FAIL-­Press en
ter to proceed.
4 FLASH test: Hourglass blinks, then display of
--PASS-- or --FAIL-­Press enter to proceed.
5 Module test: Hourglass blinks, then display of
--PASS-- or --FAIL-­Press enter or meas to return to measuring mode.
Diagnostics
81
Diagnostics
Menu item Remark
Display of logbook entries.
Select LOGBOOK using, press enter to confirm.
By using the  keys, you can scroll backwards and forwards through the logbook (entries -00-...-99-),
-00- being the last entry.
By using thekeys, you can view a logbook entry.
Press meas to return to measurement.
Extended logbook / Audit Trail (via TAN)
By using the  keys, you can scroll backwards and forwards through the extended logbook (entries -000-...-199-), -000- being the last entry.
Display: CFR
Audit Trail also records function activations (CAL CONFIG SERVICE), some Sensoface messages, and opening of the enclosure.
Display of currently measured values ( sensor monitor):
Select MONITOR using, press en Use thekeys to select the desired parameter from the bottom line of the display (R_COND G_COND RTD TEMP I-INPUT (Option)). The selected parameter is shown in the main display.
Display example:
Press meas to return to measurement.
ter to confirm.
82
Display Remark
Version
Here, you find the data you require for requesting a device-specific Option. Display of device type, software/hardware version, and serial number for all device components. Use the  keys to switch between software and hardware version. Press enter to proceed to next device component.
Diagnostics
83
Service (SERVICE) A

Service

In the Service mode you can access the following menus:
MONITOR displaying currently measured values OUT1 testing current output 1 OUT2 testing current output 2 IRDA activating and communicating via the IrDA interface CODES assigning and editing passcodes DEFAULT resetting the device to factory settings OPTION enabling options via TAN.
Please note:
HOLD is active during Service mode!
Action Key/Display Remark
Activate Service
Press any arrow key to call the selection menu. Select SERVICE using  keys, press enter to confirm.
Passcode
Display
End
84
Enter passcode “5555“ for service mode using the   keys.
Press enter to confirm.
In service mode the following icons are displayed:
[diag] mode indicator• HOLD triangle• Service (wrench)
meas End by pressing meas.
Menu item Remark
Display of currently measured values ( sensor monitor) with HOLD mode activated:
Select MONITOR using, press enter to confirm. Select variable in the bottom text line using.
The selected parameter is shown in the main display.
Display example:
As the device is in HOLD mode, you can perform validations using simulators without influencing the signal outputs.
Press meas to return to the service menu. Return to measurement: Press meas once more.
Specify current at outputs 1 and 2:
Select OUT1 or OUT2 using thekeys, press enter to confirm. Enter a valid current value for the respective output using  keys. Press enter to confirm. For checking purposes, the actual output current is shown in the bottom right corner of the display. End by pressing enter or meas.
IrDA communication:
Select IRDA using, press enter to confirm.
When IrDA communication is active, the device remains in the HOLD mode for reasons of safety. Further operation is performed via IrDA.
End communication by pressing meas.
Exception: Firmware update (must not be interrupted!)
Service
85
Service
Menu item Remark
Assigning passcodes:
In the “SERVICE - CODES“ menu you can assign pass­codes to DIAG, HOLD, CAL, CONF, and SERVICE modes (Service preset to 5555). When you have lost the Service passcode, you have to request an “ Ambulance TAN“ from the manufac­turer specifying the serial number of your device. To enter the “Ambulance TAN“, call the Service func­tion and enter passcode 7321. After correct input of the ambulance TAN the device signals “PASS“ for 4 sec and resets the Service passcode to 5555.
Reset to factory settings:
In the “SERVICE - DEFAULT“ menu you can reset the device to factory settings.
Caution!
After a reset to factory setting the device must be reconfigured completely, including the sensor parameters!
Option request:
Communicate the serial number and hardware/ software version of your device to the manufacturer. These data can be viewed in the Diagnostics/Version menu. The “transaction number“ (TAN) you will then receive is only valid for the device with the corresponding serial number.
Release of options:
Options come with a “transaction number“ ( TAN). This TAN must be entered and confirmed using enter to release the option.
86

USP Function

According to the "USP" directive (U.S.Pharmacopeia), Section 645 "Water Conductivity" the conductivity of pharmaceutical waters can be monitored online. To do so, the conductivity is measured without temperature compensation and is compared with limit values (see table on next page). The water is usable when the conductivity is below the USP limit. If the conductivity values are higher, further test steps must be performed according to the directive. To increase safety, the USP limit value can be reduced in the device. To do so, a factor (%) is speci ed.
Con guring:
SNS menu group:
When “USP function“ has been selected, the measuring range
is  xed to 00.00.....99.99 μS/cm. Temperature compensation is
switched o . Temperature is monitored.
If the USP limit is exceeded, a 22 mA signal is output.
Temperature/Conductivity Table as per USP
Temp (°C) Cond (μS/cm) Temp (°C) Cond (μS/cm)
00.6552.1 5 0.860 2.2
10 0.9652.4 15 1.0 70 2.5
20 1.1 75 2.7 25 1.3 80 2.7
30 1.4 85 2.7 35 1.590 2.7 40 1.7952.9 45 1.8 100 3.1
501.9
87
88
Operating
OUT 1
status
Measuring -
OUT 2

Operating States

Time
out
Operating states
Diag
CAL_SOL Cal solution
CAL_CELL Cell constant
P_CAL Product cal S1
P_CAL Product cal S2
CAL_RTD Temp adjustment
CONF ParSet A
CONF ParSet B
HOLD input No
60 s
No
No
No
No
No
20
min
20
min
Explanation: as configured (Last/Fix)
active
89

Product Line and Accessories

Order Code Stratos Pro A 2...
Example A 2 11X-PH -1
2-wire / 4-20 mA A 2 B,C,E
Communication
Without (HART retrofittable via TAN) 0 A HART 1 PROFIBUS PA 2 Foundation Fieldbus FF 3
Version number
Version 1
Approvals
General Safety N ATEX / IECEX Zone 2 B ATEX / IECEX / FM / CSA Zone 1 / Cl 1 Div 1 X Other approvals Z
Measuring channel
Memosens pH / Redox digital MSPH Memosens Cond digital MSCOND Memosens Oxy digital MSOXY Dual COND (2x2-electrode sensors, analog) N CC pH / ORP value (ISM digital per TAN) Cond, 2-/4-electrode Measuring module COND Conductivity, electrodeless Measuring module CONDI Oxygen (ISM digital and traces per TAN) Carbon dioxide
(ISM digital per TAN)
Options
Without 2nd current output 0 With 2nd current output 1
TAN options
HART SW-A001 (A) Logbook SW-A002 (B) Extended logbook (Audit Trail) SW-A003 (C) Trace oxygen measurement SW-A004 (D) Current input + 2 digital inputs SW-A005 (E) ISM digital SW-A006 (F)
Mounting accessories
Pipe-mount kit ZU 0274 Protective hood ZU 0737 Panel-mount kit ZU 0738
Measuring module PH F
Measuring module OXY D, F
Measuring module CO2
TAN
90

A201/A211X: Supply Units and Connection

Recommended Power Supply Units: Order No.:
Repeater power supply, IS, 24 V AC/DC, output 0/4...20 mA
Repeater power supply, IS, 90...253 V AC, output 4...20 mA
Repeater power supply, IS, 90...253 V AC, HART, output 4...20 mA
Repeater power supply, IS, 24 V AC/DC, output 4...20 mA
Repeater power supply, IS, 24 V AC/DC, HART, output 4...20 mA
Repeater power supply, non-IS, 24 V DC, output 4...20 mA
Repeater power supply, non-IS, 24 V DC, HART, output 0/4...20 mA / 0...10 V
Connection to Supply Units
+ out
- out 1
- out 2
Stratos Pro A 2...X
8
9
17
Hazardous-area requirements for cable:
Between the wires: Solid insulation ≥ 0.5 mm Test voltage ≥ 500 V
WG 20 A2
WG 21 A7
WG 21 A7 Opt. 470
WG 21 A7 Opt. 336
WG 21 A7 Opt. 336, 470
IsoAmp PWR B 10116
IsoAmp PWR A 20100
+
Supply unit 1
-
+
Supply unit 2
-
91

Specifications

COND input Input for 2-/4-elec trode sensors and Memosens sensors
E ective ranges 2-EL sensors 0.2 µS · c … 200 mS · c
Ranges Conductivity 0.000 ... 9.999 µS/cm
1,2,3)
Meas. error
Temp compensation
(reference temp 25°C) (LIN) Linear characteristic 00.00 ... 19.99 %/K
Concentration determination -01- NaCl 0.00 ... 9.99% by wt (0 ... +60°C)
4-EL sensors 0.2 µS · c … 1000 mS · c
(Conductance limited to 3500 mS)
00.00 ... 99.99 µS/cm
000.0 ... 999.9 µS/cm
0000 ... 9999 µS/cm
0.000 ... 9.999 mS/cm
00.00 ... 99.99 mS/cm
000.0 ... 999.9 mS/cm
0.000 ... 9.999 S/cm
00.00 ... 99.99 S/cm
Resistivity 00.00 … 99.99 MΩ · cm
Concentration 0.00 ... 9.99 %
Salinity 0.0 ... 45.0 ‰ (0 ... 35 °C)
Response (T90) Approx. 1 s
< 1 % meas. val. + 0.4 µS · c
*)
(OFF) Without
(NLF) Natural waters to EN 27888
(NACL) Ultrapure water with NaCl traces (0 ... 120 °C)
(HCL) Ultrapure water with HCl traces (0 ... 120 °C)
(NH3) Ultrapure water with NH3 traces (0... 120 °C)
-02- HCl 0.00 ... 9.99% by wt (–20 ... +50 °C)
-03- NaOH 0.00 ... 9.99% by wt (0 ... +100 °C)
-04- H
2SO4
-05- HNO
3
0.00 ... 9.99% by wt (–17 ... +110 °C)
0.00 ... 9.99% by wt (–17 ... +50 °C)
92
Speci cations
Sensor standardization Input of cell constant with simultaneous display of selected process
Permitted cell constant 00.0050 ... 19.9999 cm
Sensocheck Polarization detection and monitoring of cable capacitance
Delay Approx. 30 s
Sensoface Provides information on the sensor condition
Sensor monitor Direct display of measured values from sensor for validation
USP function Water monitoring in the pharmaceutical industry (USP)
Temperature input
Measuring range Pt 100/Pt 1000 50 ... +200 °C / –58 ... +392 °F
Resolution 0.1 °C / 0.1 °F
Meas. error
I input (TAN) Current input 0/4 ... 20 mA / 50 Ω for external temperature signal
Start/end of scale Con gurable –50 ... +200 °C / –58 ... +392 °F
Characteristic Linear
Measurement error
*
1,2,3)
1,3)
variable and temperature
Input of conductivity of calibration solution with simultaneous display of cell constant and temperature
Product calibration for conductivity
Temperature probe adjustment
–1
(resistance/temperature)
with additional limit value (%)
Output via HART or current output (22 mA)
Pt100/Pt1000/NTC 30 kΩ/NTC 8.55 kΩ (Betatherm)
3-wire connection, adjustable
NTC 30 kΩ 20 ... +150 °C / –4 ... +302 °F
NTC 8.55 –10 ... +130 °C / –4 ... +266 °F
< 0.5 K (< 1 K for Pt 100; <1K for NTC >100°C)
< 1% current value + 0.1 mA
93
Speci cations
HOLD input Galvanically separated (OPTO coupler)
Function Switches device to HOLD mode
Switching voltage 0 ... 2 V (AC/DC) HOLD inactive
CONTROL input Galvanically separated (OPTO coupler)
Function Selecting parameter set A/B
Switching voltage 0 ... 2 V (AC/DC) Parameter set A
Output 1 Current loop, 4 ... 20 mA,  oating, protected against inverse polarity
Supply voltage 14 ... 30 V
Process variable
*
Characteristic Linear or logarithmic
*
Overrange
*
Output  lter
Measurement error
Start/end of scale
1)
*
Minimum span LIN 5% of selected range
Output 2 Current loop, 4 ... 20 mA,  oating, protected against inverse polarity
Supply voltage 14 ... 30 V
Process variable
*
Characteristic Linear or logarithmic
*
Overrange
*
Output  lter
Measurement error
1)
10 ... 30 V (AC/DC) HOLD active
10 ... 30 V (AC/DC) Parameter set B
HART communication (see further below for speci cations)
Conductivity, resistivity, concentration, salinity, or temperature
22 mA in the case of error messages
PT1  lter, time constant 0 ... 120 s
< 0.25% current value + 0.025 mA
Con gurable within selected range
LOG 1 decade
Conductivity, resistivity, concentration, salinity, or temperature
22 mA in the case of error messages
PT1  lter, time constant 0 ... 120 s
< 0.25% current value + 0.05 mA
94
Speci cations
Start/end of scale
Minimum span LIN 5% of selected range
Real-time clock Di erent time and date formats selectable
Power reserve > 5 days
Display LC display, 7-segment with icons
Main display Character height approx. 22 mm, unit symbols approx. 14 mm
Secondary display Character height approx. 10 mm
Text line 14 characters, 14 segments
Sensoface 3 status indicators (friendly, neutral, sad face)
Mode indicators meas, cal, conf, diag
Alarm indication Display blinks, red backlighting
Keypad Keys: meas, info, 4 cursor keys, enter
HART communication HART version 6
IrDA interface Infrared interface for transmission of records and logbook,
FDA 21 CFR Part 11 Access control by editable passcodes
*
Con gurable within selected range
LOG 1 decade
Further icons for con guration and messages
Digital communication by FSK modulation of output current 1
Device identi cation, measured values, status and messages, parameter setting, calibration, records
parameter setting, calibration,  rmware update
Logbook entry and  ag via HART in the case of con guration changes
Message and logbook entry when enclosure is opened
95
Speci cations
Diagnostics functions
Calibration data Calibration date, cell constant
Device self-test Displaytest, automatic memory test (RAM, FLASH, EEPROM),
Logbook 100 events with date and time
Extended logbook (TAN) Audit Trail: 200 events with date and time
Service functions
Sensor monitor Display of direct sensor signals
Current source Current speci able for output 1 and 2 (00.00 ... 22.00 mA)
IrDA Activating the IrDA function
Passcodes Assigning passcodes for menu access
Factory setting Resetting all parameters to factory setting
TAN Enabling optionally available additional functions
Data retention Parameters, calibration data, logbook > 10 years (EEPROM)
EMC EN 61326-1 (General Requirements)
Emitted interference Class B (residential area)
Immunity to interference Industr y
Explosion protection Europe: ATEX Zone 0, 1, 2, 20, 21
Stratos Pro A201X/A211X COND USA: FM Class I Div 1,2 / Zone 1
module test
EN 61326-2-3
(pending)
Canada:
International: IECEx Zone 0, 1, 20, 21
cCSAus Class I Div 1,2 / Zone 1
96
Speci cations
Nominal operating conditions
Ambient temperature 20 ... +65 °C
Transport/Storage temperature –20 ... +70 °C
Relative humidity 10 ... 95% not condensing
Supply voltage 14 ... 30 V
Enclosure Molded enclosure made of PBT/PC, glass reinforced
Fastening Wall, pipe/post, or panel mounting
Color Gray, RAL 7001
Ingress protection IP 67
Flammability UL 94 V-0
Dimensions 148 mm x 148 mm
Control panel cutout 138 mm x 138 mm to DIN 43 700
Weight Approx. 1200 g
Cable glands 3 knockouts for M20 x 1.5 cable glands
Connections Terminals, conductor cross section max. 2.5 mm
* User-de ned 1) Acc. to EN 60746, at nominal operating conditions
2) ± 1 count 3) Plus sensor error
2 knockouts for NPT ½” or rigid metallic conduit
2
97

Calibration Solutions

Potassium Chloride Solutions
(Conductivity in mS/cm)
Temperature
[˚C] 0.01 mol/l 0.1 mol/l 1 mol/l
0
5
10 15 16 17 18 19
20 21 22 23 24 25 26 27 28 29
30 31 32 33 34 35 36
Concentration
0.776
0.896
1.020
1.147
1.173
1.199
1.225
1.251
1.278
1.305
1.332
1.359
1.386
1.413
1.441
1.468
1.496
1.524
1.552
1.581
1.609
1.638
1.667
1.696
1
7.15
8.22
9.33
10.48
10.72
10.95
11.19
11.43
11.67
11.91
12.15
12.39
12.64
12.88
13.13
13.37
13.62
13.87
14.12
14.37
14.62
14.88
15.13
15.39
15.64
100.14
102.07
104.00
105.94
107.89
109.84
111.80
113.77
115.74
65.41
74.14
83.19
92.52
94.41
96.31
98.22
Data
source: K. H. Hellwege (Editor), H. Landolt, R. Börnstein: Zahlenwerte und 1
Funktionen ..., volume 2, part. volume 6
98
Sodium Chloride Solutions
(Conductivity in mS/cm)
Temperature
[˚C] 0.01 mol/l
0 1
2
3 4
5 6 7 8 9
10 11 12 13 14 15 16 17 18 19
20 21 22 23 24 25 26 27 28 29
30 31 32 33 34 35 36
Concentration
1)
0.631
0.651
0.671
0.692
0.712
0.733
0.754
0.775
0.796
0.818
0.839
0.861
0.883
0.905
0.927
0.950
0.972
0.995
1.018
1.041
1.064
1.087
1.111
1.135
1.159
1.183
1.207
1.232
1.256
1.281
1.306
1.331
1.357
1.382
1.408
1.434
1.460
Calibration Solutions
0,1 mol/l
5.786
5.965
6.145
6.327
6.510
6.695
6.881
7.068
7.257
7.447
7.638
7.831
8.025
8.221
8.418
8.617
8.816
9.018
9.221
9.425
9.631
9.838
10.047
10.258
10.469
10.683
10.898
11.114
11.332
11.552
11.773
11.995
12.220
12.445
12.673
12.902
13.132
1)
Saturated
134.5
138.6
142.7
146.9
151.2
155.5
159.9
164.3
168.8
173.4
177.9
182.6
187.2
191.9
196.7
201.5
206.3
211.2
216.1
221.0
226.0
231.0
236.1
241.1
246.2
251.3
256.5
261.6
266.9
272.1
277.4
282.7
288.0
293.3
298.7
304.1
309.5
2)
Data source: Test solutions calculated according to DIN IEC 746-31 Data source: K. H. Hellwege (Editor), H. Landolt, R. Börnstein: Zahlenwerte und 2
Funktionen ..., volume 2, part. volume 6
99

Concentration Curves

-01- Sodium chloride solution NaCl
χ [mS/cm]
400
300
200
100
0
0
5
c [% by wt]
c [Gew%]
Conductivity versus substance concentration and process temperature for sodium chloride solution (NaCl)
[°C]
100
82
60
46
35
20
10
0
10
100
Loading...