Knick Stratos 2201 pH, Stratos 2211 pH, Stratos 2201 X pH, Stratos 2211 X pH Instruction Manual

Instruction Manual
Stratos 2201 (X) pH Stratos 2211 (X) pH with HARTr Communication
Software Version: 3.x
TA-194.100-KNE03 150101
Elektronische Meßgeräte GmbH & Co. Beuckestr. 22, D–14163 Berlin Postfach 37 04 15, D–14134 Berlin Phone: +49-30-801 91-0 Fax: +49-30-801 91-200 Internet: http://www.knick.de E-mail: knick@knick.de
Warranty
Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender). Accessories: 1 year
Note
Changes 3
Changes for software release 3
T emperature probe adjustment
The Stratos can be adjusted to a connected temperature probe. This is particularly useful for Pt 100 temperature probes. (See Pg. 32)
This function should only be used by experts. Incorrectly set parameters may go unnoticed, but change the measuring properties.
Changes 4
Warning
Warning
Information 5
Safety Precautions
Be sure to read and observe the following requirements!
Never use the Stratos 2201/2211 pH for measurement in areas that are subject to an explo­sion hazard during operation. Be sure that the supply voltage does not exceed 30 Vdc.
Stratos 2201/221 1 X pH is approved for operation in hazardous locations. Before connecting the Transmitter to a power supply unit, make sure that this is an associ­ated apparatus.
Whenever it is likely that the protection has been impaired, the instrument shall be made inoperative and secured against unintended operation. The protection is likely to be impaired if, for example:
the instrument shows visible damagethe instrument fails to perform the intended measurementsafter prolonged storage at temperatures above 70 °Cafter severe transport stresses
Before recommissioning the instrument, a professional routine test in accordance with EN 61010-1 must be performed. This test should be carried out at our factory.
The instrument shall not be used in a manner not specified by this manual.
cal
Note
Warning
conf
cal
Information 6
Information on this Instruction Manual
ITALICS are used for texts which appear in the Stratos 2201/2211 (X) pH display.
Bold print is used to represent keys, e.g. cal.
Keys for which the functions are explained are frequently shown in the left-hand column.
Notes provide important information that should be strictly followed when using the unit.
Warning means that the instructions given must always be followed for your own safety. Failure to follow these instructions may result in injuries.
Mode Codes
After pressing conf or cal you can enter one of the following codes to access the desig- nated mode:
conf, 0000: Error info conf, 1200: Configuration conf, 5555: Current source
cal, 0000: Cal info cal, 1015: Temperature probe adjustment cal, 1100: Calibration cal, 2222: T est mode (electrode potential display)
Contents 7
Contents
Safety Precautions 5. . . . . . . . . . . . . . . . . . . . . .
Information on this Instruction Manual 6. . .
Mode Codes 6. . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 Assembly 8. . . . . . . . . . . . . . . . . . . . . . . . . . .
Package Contents and Unpacking 8. . . .
Assembly 8. . . . . . . . . . . . . . . . . . . . . . . . . .
2 Installation, Connection
and Commissioning 12. . . . . . . . . . . . . . . . .
Proper Use 12. . . . . . . . . . . . . . . . . . . . . . . .
Overview of the Stratos 12. . . . . . . . . . . . . .
T erminal Assignment 13. . . . . . . . . . . . . . . .
Installation and Commissioning 13. . . . . . .
Wiring Diagrams (pH) 14. . . . . . . . . . . . . . .
Wiring Diagrams (ORP) 19. . . . . . . . . . . . .
3 Operation 20. . . . . . . . . . . . . . . . . . . . . . . . . . .
User Interface 20. . . . . . . . . . . . . . . . . . . . . .
Display 21. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Keypad 21. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Safety Functions 22. . . . . . . . . . . . . . . . . . . .
Outputs 23. . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuration 24. . . . . . . . . . . . . . . . . . . . . . .
Calibration 26. . . . . . . . . . . . . . . . . . . . . . . . .
Measurement 33. . . . . . . . . . . . . . . . . . . . . .
4 Diagnostics, Maintenance and Cleaning 34
Sensoface
, Sensocheck
34. . . . . . . . . .
Error Messages 35. . . . . . . . . . . . . . . . . . . .
Calibration Error Messages 37. . . . . . . . . .
Diagnostics Functions 38. . . . . . . . . . . . . . .
Maintenance and Cleaning 39. . . . . . . . . . .
5 Annex 40. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Product Line 40. . . . . . . . . . . . . . . . . . . . . . .
Specifications 41. . . . . . . . . . . . . . . . . . . . . .
Certificate of Conformity 44. . . . . . . . . . . . .
Buffer Charts 47. . . . . . . . . . . . . . . . . . . . . . .
Technical Terms 49. . . . . . . . . . . . . . . . . . . . . . . . .
Index 51. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Assembly 8
1 Assembly
Package Contents and Unpacking
Unpack the instrument carefully and check the ship­ment for transport damage and completeness. The package contains:
Front unit of Stratos 2201/2211 (X) pHLower caseShort instruction sheetThis instruction manual
Bag containing:
2 sealing plugs 1 hinge pin5 hexagon nuts 3 cable ties3 Pg cable glands 3 filler plugs1 rubber reducer 3 gaskets1 Pg plug
11 1 washer
4 enclosure screws
12
1 jumper
Assembly
Fig. 1 Assembling the case
(For sealing in case
of wall mounting)
(Can be inserted from either side)
11
For conduit mounting: Place washer between enclosure and nut.
11
12
Assembly 9
80 [3.15]
32 [1.26]
72 [2.83]
6.2 [0.24]
approx. 14
84 [3.31]
42 [1.65]
Pg 13.5 (3 pcs.)
144 [5.67]
144 [5.67]
21
43 [1.69]
105 [4.13]
27
(not included in supply)
Control panel 1 – 22 mm
max. 25
78 [3.07]
27
Control panel cutout 138 x 138 mm (DIN 43700)
Panel-mount kit consisting of:
4 screws 4 span pieces 4 threaded sleeves 1 seal
➀ ➁ ➂ ➃
Holes for post mounting
Holes for wall mounting
(4 x)
(2 x)
For Pg 13.5 threaded cable glands (use only plastic glands)
[0.55]
[1.06]
[0.83]
[0.98]
[1.06]
[5.43 x 5.43]
[0.04 – 0.87]
Note: All dimensions
in mm [inches]
Fig. 2 Dimension drawing for Stratos, mounting diagram and ZU 0275 panel-mount kit
Assembly 10
horizontal post/
40–60 mm dia.
ZU 0276 protective hood (if required)
For vertical or pipe mounting
Pipe-mount kit consisting of:
4 self-tapping screws 1 post mounting plate 2 hose clamps with
➀ ➁ ➂
worm gear drive to DIN 3017
[1.57–2.36]
Fig. 3 ZU 0274 pipe-mount kit
165 [6.5]
173 [6.81]
132 [5.2]
Fig. 4 ZU 0276 protective hood for wall and pipe mounting
Assembly 11
280 [11.03]
7 [0.28]
Recommended stripping lengths for
multi-core cables
Recommended stripping length for
coaxial cable
Pulling out the terminals using a
screwdriver (also see
➆)
Cable laying in the Stratos
Connecting lines for loop current Cover for electrode and temperature
probe terminals
Areas for placing the screwdriver to
pull out the terminals
Connection of handheld terminal
Be sure not to
notch the cable cores when stripping the insulation!
Dimensions in mm [inches].
Coaxial cable
13 mm [0.5]
7 mm [0.28]
Fig. 5 Installation information Stratos 2201/2211 (X) pH
Warning
Warning
Capabilities, Connection 12
2 Installation, Connection and Commissioning
Proper Use
Stratos 2201/2211 (X) pH is used for pH and temper­ature measurement in industry, the environment, food processing and waste-water field. It can be either field-mounted or fixed into a control panel.
Never
use the Stratos 2201/221 1 pH for measurement in areas that are subject to an explo­sion hazard during operation.
Stratos 2201/2211 X pH is ap­proved for operation in hazardous locations.
Overview of the Stratos
pH / mV Sensocheck
Temperature Pt 100/1000/ NTC 30 kW
Stratos 2201/2211 (X) pH
mA
Fig. 6 System functions of Stratos 2201/2211 (X) pH
Inputs for glass and reference electrodeInput for temperature probeCurrent loop 4 – 20 mA,
transports power to and output signal from the Stratos,
with Stratos 2211 (X) pH also for HARTr commu­nication
Equipotential bonding (only with Stratos
2201/2211 X pH)
Warning
Warning
Warning
Warning
Warning
Warning
Capabilities, Connection 13
T erminal Assignment
+ 11 10
4 to 20 mA
+
glass el.
9 8 7 6 5 4 2 1
í
output input
shield
aux el.
n.c.
HART
observe grounding conditions
reference el.
Fig. 7 Terminal assignment of Stratos 2201/2211 (X) pH
Installation and Commissioning
Installation and commissioning of the Stratos may only be carried out in accordance with this instruction manual and per applicable local and national codes. Be sure to ob­serve the technical specifications and input ratings.
Before connecting the Stratos 2201/221 1 pH to a power supply unit, make sure that this is not ca­pable of outputting more than 30 Vdc.
Do not use alternating current or mains power supply!
Stratos 2201/2211 X pH may only be connected to an explosion-proof power supply unit (for input ratings refer to annex of Certificate of Con­formity).
When commissioning, a complete configuration must be carried out.
For easier installation, the terminal strips are of a plug-in design. The terminals are suitable for single
wires and flexible leads up to 2.5 mm
2
(AWG 14)
(see Pg. 11). See Pg. 14 and following for connection examples.
Do not use cable clamp (shield) for reference electrode connection when auxiliary electrode (solution ground) is in use.
Capabilities, Connection 14
Wiring Diagrams (pH)
reference el.
glass el. +
í
í
shield
aux. el.
8 7 4
2 1
54 21
5
Temperature probe
Combination electrode
Fit jumper between terminals 4 + 5.
Fig. 8a pH measurement with combination electrode and temperature probe,
Sensocheck
limited to the glass electrode only
54 21
í
Temperature probe
Combination electrode
shield
Fit jumper between terminals 4 + 5.
reference el.
glass el. +
í
aux. el.
8 7 4
2 1
5
Fig. 8b pH measurement with combination electrode and temperature probe,
Sensocheck
limited to the glass electrode only
Capabilities, Connection 15
Combination electrode
shield
54 21
Specify temperature!
Fit jumper between terminals 4 + 5.
reference el.
glass el. +
í
aux. el.
8 7 4
2 1
5
Fig. 9 pH measurement with combination electrode without temperature probe,
Sensocheck
limited to the glass electrode only
Capabilities, Connection 16
54 21
í
shield
reference el.
glass el. +
í
aux. el.
8 5 4
2 1
7
Temperature probe
Glass electrode
Reference electrode
Fit jumper between terminals 4 + 5.
Fig. 10 pH measurement with separate glass and reference electrode and temperature probe,
Sensocheck
limited to the glass electrode only
54 21
shield
reference el.
glass el. +
í
aux. el.
8 5 4
2 1
7
Glass electrode
Reference electrode
Specify temperature!
Fit jumper between terminals 4 + 5.
Fig. 11 pH measurement with separate glass and reference electrode without temperature probe,
Sensocheck
limited to the glass electrode only
Capabilities, Connection 17
reference el.
glass el. +
í
í
shield
aux. el.
8 5 4
2 1
54 21
7
Temperature probe
Combination electrode
Aux. electrode
Insulate!
Fig. 12 pH measurement with combination and aux. electrode and temperature probe,
Sensocheck
for glass and reference electrode
reference el.
glass el. +
í
shield
aux. el.
8 5 4
2 1
54 21
7
Combination electrode
Aux. electrode
Specify temperature!
Insulate!
Fig. 13 pH measurement with combination and aux. electrode without temperature probe,
Sensocheck
for glass and reference electrode
Capabilities, Connection 18
54 21
í
shield
reference el.
glass el. +
í
aux. el.
8 5 4
2 1
7
Temperature probe
Glass electrode
Reference electrode
Aux. electrode
Fig. 14 pH measurement with separate glass, reference and aux. electrode and temperature probe,
Sensocheck
for glass and reference electrode
54 21
shield
reference el.
glass el. +
í
aux. el.
8 5 4
2 1
7
Glass electrode
Reference electrode
Aux. electrode
Specify temperature!
Fig. 15 pH measurement with separate glass, reference and aux. electrode without temperature probe,
Sensocheck
for glass and reference electrode
Capabilities, Connection 19
Wiring Diagrams (ORP)
54 21
í
shield
reference el.
glass el. +
í
aux. el.
8 7 4
2 1
5
Insulate!
Temperature probe
combination ORP electrode
Fig. 16 ORP measurement with separate combination ORP electrode and temperature probe
54 21
shield
reference el.
glass el. +
í
aux. el.
8 7 4
2 1
5
Insulate!
combination ORP electrode
Specify temperature!
Fig. 17 ORP measurement with combination ORP electrode without temperature probe
Operation 20
3 Operation
User Interface
meas cal
conf
Stratos
cal
conf
Y
"
enter
Alarm LED
Keypad
Instrument designation
Rating plate
Status fields:
- Measuring mode
- Calibration mode
- Alarm (Contact only for Model 2401)
-Wash contact active (only Model 2401 pH)
- Configuration mode
Fig. 18 Front view of Stratos
cal
conf
"
Y
enter
cal
enter
enter
conf
Y
"
+
Operation 21
Display
Mode code
entry
Interval/
Response time
Hold state
active
Continue with
enter
Unit value symbols
Tempera-
ture
Loop
current
Alarm
setting
Senso­check
Calibration
Manual temp
specification
Sensor
data
Sensoface
Wait
Display/
Measured variable
Bar for instrument status
Fig. 19 Stratos display
Keypad
Start, end calibration
Start, end configuration
Select digit position (selected position flashes)
Change digit
Prompt in display: continue in program sequence, Configuration:
Confirm entries, next configuration step, Measuring mode:
Display output
current
Cal info, display asymmetry poten­tial (zero) and slope (see Pg. 33)
Error info, display last output error message (see Pg. 33)
Start manual unit self-test GainCheck
(see Pg. 22)
Y
"
+
Operation 22
Safety Functions
Sensoface electrode monitoring
Sensoface
provides information on the electrode state. The asym­metry potential, slope and response time during calibration are eva­luated. Sensocheck
continuously monitors the glass and reference electrode.
For more detailed information, see chapter “Diagnos- tic, Maintenance and Cleaning (Pg. 34).
GainCheck
manual self-test
Simultaneously pressing Y and " starts the manual unit self-test.
A display test is carried out, the software version is displayed and the memory and measured value transmission are checked.
Automatic self-test
The automatic unit self-test checks the memory and the measured-value transmission. It runs automati­cally in the background at fixed intervals.
Operation 23
Outputs
Loop current
The current loop transports power to and output sig­nals from the Stratos. The current is controlled by the measured variable selected in the configuration. The current beginning and end can be set to any desired value. To check connected peripherals (e.g. limit switches, controllers), the loop current can be manually speci­fied (see Pg. 38).
HART communication
Stratos 2211 (X) pH can be remote controlled via HART communication. It can be configured using a handheld terminal or from the control room. Mea­sured values, messages and unit identification can be downloaded at any time. This allows easy integra­tion also in fully automatic process sequences.
For more detailed information, refer to HART
Com-
mand Specification.
Alarm
During an error message the alarm LED flashes. Alarm response time is permanently set to 10 s.
Error messages can also be signaled with a 22 mA signal via the loop current (see Configuration, Pg. 25).
conf
Operation 24
Configuration
Here the basic settings of the unit are carried out. Symbols show which parameter is being configured.
Activate with conf change parameter with Y and ", confirm/continue with enter, end configuration with conf
Mode code “1200”
During configuration the loop cur­rent is frozen .
When the configuration mode is exited, the unit re­mains in the Hold state for safety reasons. This pre­vents undesirable reactions of the connected periph­erals (e.g. limit switches, controllers) due to incorrect settings. The measured value and Hold are dis­played alternately. Now you can check whether the measured value is plausible and specifically end the Hold state with enter. After 20 sec. (measured value stabilization) the Hold state is ended.
Operation 25
Configuration parameters
Before attempting any changes refer to the parameter setup list shown below. This table presents the possible options and the factory settings.
Picto­graph
Parameter Choices Factory
Setting
Meas. variable (when changed, complete configuration required)
pH 0.00 to 14.00 ORP –1,500 to 1,500 mV
pH
Temperature display / detection
°C / °F automatic detection
Auto °C
°C / °F
manual specification °C/ °F automatic detection during measurement manual specification during calibration
T emperature probe Pt 100 / Pt 1000 /
NTC 30 kW
Pt 100
Current beginning, current end
4 mA, 0 to 14 pH, 20 mA, 0 to 14 pH
4 mA, 0 pH 20 mA, 14 pH
Hold state Last: Last current value
Fix: Loop current specified
Last
22 mA signal for error message On / OFF OFF Sensocheck
(sensor diagnostics)
On / OFF OFF
Calibration mode: Automatic with Calimatic
BUF –00– Knick techn. buffers BUF –01– Mettler Toledo techn. buffers BUF –02– Merck/Riedel de Haën BUF –03– Ciba (94) BUF –04– Mettler Toledo (USA) BUF –05– NIST standard buffers BUF –06– HACH BUF –07– WTW
BUF –01–
Manual MAN Manual buffer entry
DAT Data entry of premeasured values
Calibration timer interval 0,000 to 9,999 hours 0000 (OFF)
Note
cal
Note
cal
enter
Note
Operation 26
Calibration
You can conduct either a one or a two-point calibra­tion. The calibration can be carried out with the Cali­matic
automatic buffer recognition, with manual buffer input or by entering pre-measured electrode data.
With manual temperature specifica­tion, the temperature of the buffer solutions must be entered in the configuration prior to calibration.
Activate with cal confirm/continue with enter, abort with cal enter
During calibration the loop current is frozen. The instrument is in the Hold state.
When the calibration mode is exited, the Stratos re­mains in the Hold state for safety reasons. This pre­vents undesirable reactions of the connected periph­erals (e.g. limit switches, controllers) due to incorrect settings. The measured value and Hold are dis­played alternately. Now you can check whether the measured value is plausible and specifically end the Hold state with enter or repeat calibration with cal. If you end the Hold state, the Stratos will return to measuring mode after 20 sec. (measured value sta­bilization).
Automatic calibration with Calimatic
BUF
and automatic calibration temp detection
The instrument can only operate properly when the buffer solutions used correspond to the configured buffer set. Other buffer solutions, even those with the same nominal values, may demonstrate a different temperature behavior. This leads to measurement errors.
Activate calibration by pressing the cal key. Using the Y , " keys enter mode code 1100and then press enter.
Remove electrode and temperature probe and immerse them in the first buffer solution; it does not matter which buffer solution is taken first.
Start calibration with enter.
While the hour glass flashes, the electrode and temperature probe remain in the first buffer solution.
The response time of the electrode and the temperature probe is con­siderably reduced if the electrode is first moved about in the buffer solu­tion and then held still. Stirring pro­vides stable values faster.
Note
Note
cal
Operation 27
Buffer recognition
Nominal buffer value is displayed
Stability check: measured mV value is displayed
The stability check can be aborted with cal. However, accuracy of the calibration will be compromised.
Calibration with the first buffer is completed. Remove the electrode and temperature probe from the first buffer solution and rinse off both thoroughly.
If you would like to carry out a one-point calibra-
tion, end the calibration now with cal. The Stratos then shows the newly determined asymmetry po­tential in the lower display and the old slope in the main display.
If you would like to carry out a two-point calibra-
tion, immerse the electrode and the temperature probe in the second buffer solution. Now start the calibration again with enter. The calibration pro­cess runs again as for the first buffer.
At the end of calibration the slope and asymmetry potential (based on 25 °C) of the electrode are dis­played. Calibration is ended with enter. The Stratos remains in the Hold state. You can now reinstall the electrode and the temperature probe and end the Hold state with enter. After 20 sec. (measured value stabilization) the Stratos re­turns to measuring mode.
Automatic calibration with Calimatic
BUF
and manual specification of calibration temp
The instrument can only operate properly when the buffer solutions used correspond to the configured buffer set. Other buffer solutions, even those with the same nominal values, may demonstrate a different temperature behavior. This leads to measurement errors.
Activate calibration by pressing the cal key. Using the Y , " keys enter mode code 1100and then press enter.
Remove the electrode and immerse it in the first buffer solution; it does not matter which buffer solution is taken first. Enter the calibration temperature using the Y and " keys.
enter
Note
Note
Operation 28
Start calibration with enter.
While the hour glass flashes, the electrode remains in the first buffer solution.
The response time of the electrode is considerably reduced if the elec­trode is first moved about in the buffer solution and then held still. Stirring provides stable values faster.
Buffer recognition Nominal buffer value is displayed
Stability check: measured mV value is displayed
The stability check can be aborted with cal. However, accuracy of the calibration will be compromised.
Calibration with the first buffer is completed. Remove the electrode from the first buffer solution and rinse it off thoroughly.
If you would like to carry out a one-point calibra-
tion, end the calibration now with cal. The Stratos then shows the newly determined asymmetry po­tential in the lower display and the old slope in the main display.
If you would like to carry out a two-point calibra-
tion, immerse the electrode in the second buffer solution. Now start the calibration again with enĆ ter. The calibration process runs again as for the first buffer.
At the end of calibration the slope and asymmetry potential (based on 25 °C) of the electrode are dis­played. Calibration is ended with enter. The Stratos remains in the Hold state. You can now reinstall the electrode and end the Hold state with enter. After 20 sec. (measured value stabilization) the Stratos returns to measuring mode.
cal
Note
Note
Operation 29
Manual calibration MAN with automatic calibration temp detection
(if selected in Conf mode) For calibration with manual buffer specification, you
must enter the pH value of the buffer solution used in the instrument for the proper temperature. This enables calibration with any desired buffer solu­tion.
Activate calibration by pressing the cal key. Using the Y , " keys enter mode code 1100and then press enter.
Remove electrode and temperature probe and immerse them in the first buffer solution; it does not matter which buffer solution is taken first. Confirm with enter.
Set the pH value of your buffer solution for the proper temperature with Y and ". Start calibration with enter.
The response time of electrode and temperature probe is considerably reduced if the electrode is first moved about in the buffer solution and then held still.
Stability check: measured mV value is displayed.
The stability check can be aborted with cal. However, the calibration accuracy will be compromised.
Calibration with the first buffer is completed. Remove the electrode and temperature probe from the first buffer solution and rinse off both thoroughly.
If you would like to carry out a one-point calibra-
tion, press cal to end the calibration now . The Stratos then shows the newly determined asym­metry potential (zero point) in the main display and the old slope in the lower display.
If you would like to carry out a two-point calibra-
tion, immerse the electrode in the second buffer solution. Enter the pH value of the second buffer solution as has been done with the first buffer. Now start the calibration again with enter. The calibration process runs again as for the first buffer.
At the end of calibration the slope and asymmetry potential (based on 25 °C) of the electrode are dis­played. Press enter to end calibra­tion. The Stratos remains in the Hold state. You can now reinstall the electrode and temperature probe and end the Hold state with enter. After 20 sec (measured value stabilization) the Stratos re­turns to measuring mode.
cal
Note
Note
Operation 30
Manual calibration MAN with manual specification of calibration temp
(if selected in Conf mode) For calibration with manual buffer specification, you
must enter the pH value of the buffer solution used in the instrument for the proper temperature. This enables calibration with any desired buffer solu­tion.
Activate calibration by pressing the cal key. Using the Y , " keys enter mode code 1100and then press enter.
Remove the electrode and immerse it in the first buffer solution; it does not matter which buffer solution is taken first. Enter the calibration temperature using the Y and " keys and confirm with enter.
Set the pH value of your buffer solution for the proper temperature with Y and ". Start calibration with enter.
The response time of electrode is considerably reduced if the elec­trode is first moved about in the buffer solution and then held still.
Stability check: measured mV value is displayed.
The stability check can be aborted with cal. However, the calibration accuracy will be compromised.
Calibration with the first buffer is completed. Remove the electrode from the first buffer solution and rinse it off thoroughly.
If you would like to carry out a one-point calibra-
tion, press cal to end the calibration now . The Stratos then shows the newly determined asym­metry potential (zero point) in the main display and the old slope in the lower display.
If you would like to carry out a two-point calibra-
tion, immerse the electrode in the second buffer solution. Enter the pH value of the second buffer solution as has been done with the first buffer. Now start the calibration again with enter. The calibration process runs again as for the first buffer.
At the end of calibration the slope and asymmetry potential (based on 25 °C) of the electrode are dis­played. Press enter to end calibra­tion. The Stratos remains in the Hold state. You can now reinstall the electrode and end the Hold state with enter. After 20 sec. (measured value stabilization) the Stratos returns to measuring mode.
cal
Operation 31
Data entry of premeasured electrodes DAT
You can directly enter the slope and asymmetry po­tential of an electrode. The values must be known, e.g. determined beforehand in the laboratory.
Activate calibration by pressing the cal key. Using the Y , " keys enter mode code 1100and then press enter.
Enter asymmetry potential, confirm with enter
Enter slope, confirm with enter
At the end of calibration the slope and asymmetry potential (based on 25 °C) of the electrode are dis­played. Press enter to end calibra­tion. The unit remains in the Hold state. You can now reinstall the electrode and the temperature probe and end the Hold state with enter. After 20 sec. (measured value stabilization) the unit returns to measuring mode.
Convert slope [%] to slope [mV/pH] at 25 °C
%
78 80 82 84 86 88 90 92 94 96 98 100 102
mV/pH 46.2 47.4 48.5 49.7 50.9 52.1 53.3 54.5 55.6 56.8 58.0 59.2 60.4
Note
cal
enter
Operation 32
Adjustment of temperature probe
This function should only be used by experts. Incorrectly set parame­ters may go unnoticed, but change the measuring properties. Especially for Pt 100 temperature probe, it is advisable to perform an adjustment.
Activate calibration by pressing the cal key. Using the Y , " keys enter mode code 1015and then press enter.
Measure the temperature of the process medium using an external thermometer.
Using the Y , " keys enter the de­termined temperature value in the main display. If you take over the temperature value shown in the lower display , the correction is with­out effect.
Press enter to confirm the temper­ature value.
The Stratos remains in the Hold state. You can end the Hold state with enter. After 20 sec (measured value stabilization) the Stratos re­turns to measuring mode.
Operation 33
Measurement
Measuring mode
In the measuring mode the main display shows the configured measured variable (pH or mV) and the secondary display the temperature.
Cal info
With cal and mode code 0000 you can activate the cal info. Cal info shows the current calibration data for approx. 20 sec. The 20 sec can be reduced by pressing enter. During cal info the unit is not
in Hold
state.
Error info
With conf and mode code 0000 you can activate the error info. Error info shows the most recent error message for approx. 20 sec. After that the message will be deleted. The 20 sec can be reduced by pres­sing enter. During error info the unit is not
in Hold
state.
Manual temperature specification
The
indicator signals that the temperature will be manually specified. The measuring temperature can be set in the configuration, the calibration tempera­ture in calibration.
Hold state
The unit will enter the Hold state under the following conditions:
For calibration: Mode code 1015
Mode code 1100 Mode code 2222
configuration: Mode code 1200
Mode code 5555
The loop current is frozen at Last or Fix (configura­tion Pg. 25). If the calibration or configuration mode is exited, the unit remains in the Hold state for safety reasons. This prevents undesirable reactions of the connected peripherals (e.g. limit switches, control­lers) due to incorrect settings. The measured value and Hold are displayed alternately. Now you can check whether the measured value is plausible and specifically end the Hold state with enter. After 20 sec. (measured value stabilization) the unit returns to measuring mode.
Note
Note
Troubleshooting, Cleaning 34
4 Diagnostics, Maintenance and Cleaning
Sensoface, Sensocheck
Sensoface provides information
on the electrode state. The slope, asymmetry potential and response time during calibration are eva­luated. Sensocheck
continuously monitors the glass and reference electrode. With Sensocheck
turned off no appears.
Three Smileys provide information on wear and re­quired maintenance. However, the unit can still de­termine the measured variable and output it via the loop current.
The worsening of a Sensoface
cri­terion leads to the devaluation or of the Sensoface indicator.
An improvement
of the Sensoface indicator
can only take place after calibration or removal of an electrode defect. is only displayed when Senso-
check
has been activated.
The condition for accurate informa­tion is proper calibration.
Sensoface
displays during calibration
Display Problem Status
Electrode re­sponse time
The electrode adjusts slowly . You should consider replacing it. It may be possible to achieve an improvement by cleaning or, in the case of an electrode stored dry, by “watering”.
The electrode adjusts very slowly to the measured value. Correct mea­surement is no longer ensured. The electrode should be replaced.
Asymmetry potential and
Asymmetry potential (zero point) and slope of the electrode are still okay, however the electrode should be replaced soon.
slope
Asymmetry potential (zero point) and/or slope of the electrode have reached values which no longer ensure proper calibration. It is advisable to replace the electrode.
enter
conf
Troubleshooting, Cleaning 35
Sensoface
displays during measurement
Display Problem Status
Calibration
Over 80 % of the calibration interval has already past.
timer
The calibration interval has been exceeded.
Electrode defect
Check the electrode and its connections (also see error messages 33 and 34).
Error Messages
When one of the following error messages is output, the unit can no longer correctly determine the mea­sured variable or output it via the loop current.
During an error message the alarm LED flashes. The alarm response time is permanently set to 10 sec.
During alarm condition the unit does not switch into Hold state. The current loop will remain active and still represents the currently displayed reading. If 22 mA function is configured (see page 25), error messages will also be indicated with a 22 mA signal.
Error info
With conf and mode code 0000 you can activate the error info. Er­ror info shows the most recent error message for approx. 20 sec. After that the message will be deleted. The 20 sec can be reduced by pressing enter. During error info the unit is not
in Hold state.
Error number
Display (flashing)
Problem Possible causes
Err 01 pH electrode - Electrode defective
- Too little electrolyte in electrode
- Electrode not connected
- Break in electrode cable
- Incorrect electrode connected
- Measured pH value less than 0 or greater than +14
Err 02 Redox electrode - Electrode defective
- Electrode not connected
- Break in electrode cable
- Measured electrode voltage less than –1500 mV or greater than +1500 mV
Troubleshooting, Cleaning 36
Error number
Possible causesProblemDisplay
(flashing)
Err 03 T emperature probe - Incorrect temperature probe connected or configured
- Open or short circuit in temperature probe
- Measured temperature less than –20 °C or greater than +150 °C (NTC 30 kW: +130 °C)
Err 21 Loop current - Measured value below configured current beginning
- Check configuration current beginning (see Pg. 25)
Err 22 Loop current - Measured value above configured current end
- Check configuration current end (see Pg. 25)
Err 23 Loop current - Configured current span too large or too small
(Difference between current beginning and end)
Err 33 Glass electrode - Glass electrode defective
- Connection cable or electrode cap defective
- Connection terminals or electrode cap dirty
Err 34 Reference elec-
trode
- Reference electrode defective
- Connection cable or electrode cap defective
- Connection terminals or electrode cap dirty
- Jumper between terminal 4 and 5 missing (see Figs. 8a – 11 on Pg. 14 and the following)
Err 98 System error -Configuration or calibration data defective; completely
reconfigure and recalibrate the unit
- Measured value transmission defective
- Memory error in unit program (PROM defective)
Err 99 Factory settings -EEPROM or RAM defective,
- Error in factory settings
This error message normally should not occur, as the data are protected from loss by multiple safety functions. Should this error message nevertheless occur, there is
no remedy. The unit must be repaired and recalibrated at
the factory.
Troubleshooting, Cleaning 37
Calibration Error Messages
(only during calibration)
Display
Problem Possible causes Asymmetry poten-
tial (zero) out of range 〈±60 mV)
- Electrode worn out
- Buffer solutions contaminated
- Buffer does not belong to configured buffer set
- Temperature probe not immersed in buffer solution (for automatic temperature compensation)
- Wrong buffer temperature set (for manual temperature specifica­tion)
- Electrode with nominal zero point < pH 6 or > pH 8 is used
Electrode slope out of range
(80 – 103 %)
- Electrode worn out
- Buffer solutions contaminated
- Buffer does not belong to configured buffer set
- Temperature probe not immersed in buffer solution (for automatic temperature compensation)
- Wrong buffer temperature set (for manual temperature specifica­tion)
- Electrode used has different nominal slope
Problems during recognition of the buffer solution
- Same or similar buffer solution was used for both calibration steps
- Buffer solution used does not belong to buffer set currently config­ured in the unit
- During manual calibration the buffer solutions were not used in the specified order
- Buffer solutions contaminated
- Electrode defective
- Electrode not connected
- Electrode cable defective
- Wrong buffer temperature set (for manual temperature specifica­tion)
Calibration was canceled after approx. 2 minutes, because the elec­trode drift was too large.
- Electrode defective or dirty
- No electrolyte in the electrode
- Electrode cable insufficiently shielded or defective
- Strong electric fields influence the measurement
- Major temperature fluctuation of the buffer solution
- No buffer solution or extremely diluted
Warning
Troubleshooting, Cleaning 38
Diagnostics Functions
Cal info
Pressing cal and entering mode code 0000 is go­ing to activate the cal info. Cal info shows the current calibration data for approx. 20 sec. During cal info the unit is not
in Hold state.
Error info
Pressing conf and entering mode code 0000 is going to activate the error info. Error info shows the most recent error message for approx. 20 sec. After that the message will be deleted. The 20 sec can be reduced by pressing enter. During error infothe in­strument is not
in Hold state.
Display electrode potential
During electrode maintenance it is useful to directly indicate the electrode potential. This allows, for ex­ample, to check electrode response after cleaning.
Pressing cal and entering mode code 2222 will display the electrode potential. The instrument is in Hold state.
Display loop current
Pressing enter in measuring mode displays the loop current for a few seconds.
Current source
To check the connected peripherals (e.g. limit switches, controllers), the loop current can be manu­ally specified.
In the current source function the loop current no longer follows the measured value! It is manually spe­cified. Therefore, it must be ensured that the connected devices (control room, controllers, indicators) do not interpret the current value as a measured value!
Pressing conf and entering mode code 5555 you enter the current source mode. Specify the loop cur­rent using ", Y and enter. The present loop cur­rent is shown in the lower display. Pressing conf exits the current source mode again.
GainCheck
manual unit self-test
To start press Y and " simultaneously. A display test is carried out, the software version is
displayed and the memory and measured-value transmission checked.
Automatic unit self-test
The automatic unit self-test checks the memory and the measured-value transmission. It runs automati­cally in the background at fixed intervals.
Troubleshooting, Cleaning 39
Maintenance and Cleaning
Maintenance
The Stratos 2201/221 1 (X) pH contains no user re­pairable components. If problems persist even after reviewing section 4, please contact the factory.
Cleaning
To remove dust, dirt and spots, the external surfaces of the unit may be wiped with a damp, lint-free cloth. A mild household cleaner may also be used if neces­sary.
Specifications 40
5 Annex
Product Line
Units
Ref. No.
Stratos 2201 pH 2201 pH Stratos 2201 X pH for application
in hazardous areas
2201 X pH
Stratos 2211 pH with HART
communication
2211 pH
Stratos 2211 X pH with HART
communication, for
application in hazardous areas
2211 X pH
Mounting Accessories
Ref. No.
Pipe-mount kit ZU 0274 Panel-mount kit ZU 0275 Protective hood ZU 0276
Further Accessories
Ref. No.
HART
test socket, integrated in Pg cable gland (for Stratos 2211 (X) pH only)
ZU 0287
SMEK input socket (for sensor cable with one coax line)
ZU 0322
Suggested Power Supplies
Ref. No.
Power supply / isolator for 24 Vac/dc
WG 20 A2
Repeater power supply for 90 - 253 Vac
WG 21 A7
for 24 Vac/dc WG 21 A7
Opt. 336
with HART
communication
WG 21 A7
Opt. 470
Loop-powered supply with HART
communication
WG 25 A7
Specifications 41
Specifications
Stratos 2201 pH / 2211 pH pH/mV input
Ranges pH value 0.00 to +14.00
mV value –1500 to +1500 mV
Glass
input resistance > 0.5
*
1012 W
input current (20 °C)
1)
< 2*10
–12
A
Reference
input resistance > 1
*
1010 W
input current (20 °C)
1)
< 1*10
–10
A
Meas. Error (+ 1 count)
pH value < 0.02 mV < 1 mV
Electrode monitoring
Sensocheck: Monitoring of glass and reference electrode (can be switched off)
Alarm Limits
Determination during calibration
Electrode Standardiza­tion
*)
– Calimatic automatic calibration
with the buffer sets:
-00- Knick technical buffers
-01- (correspond to Mettler Toledo techn. buffers)
2.00/4.01/7.00/9.21
-02- Merck/Riedel de Haën
2.00/4.00/7.00/9.00/12.00
-03- Ciba (94)
2.06/4.00/7.00/10.00
-04- Mettler Toledo (USA)
4.00/7.00/10.01
-05- Standard buffers NIST
4.006/6.865/9.180
-06- HACH
4.00/7.00/10.18
-07- WTW techn. buffers
2.00/4.01/7.00/10.00
Stratos 2201 X pH / 2211 X pH pH/mV input EEx ia IIC
Ranges pH value 0.00 to +14.00
mV value –1500 to +1500 mV
Glass
input resistance > 0.5
*
1012 W
input current (20 °C)
1)
< 2*10
–12
A
Reference
input resistance > 1
*
1010 W
input current (20 °C)
1)
< 1*10
–10
A
Meas. Error (+ 1 count)
pH value < 0.02 mV < 1 mV
Electrode monitoring
Sensocheck: Monitoring of glass and reference electrode (can be switched off)
Alarm Limits
Determination during calibration
Electrode Standardiza­tion
*)
– Calimatic automatic calibration
with the buffer sets:
-00- Knick technical buffers
-01- (correspond to Mettler Toledo techn. buffers)
2.00/4.01/7.00/9.21
-02- Merck/Riedel de Haën
2.00/4.00/7.00/9.00/12.00
-03- Ciba (94)
2.06/4.00/7.00/10.00
-04- Mettler Toledo (USA)
4.00/7.00/10.01
-05- Standard buffers NIST
4.006/6.865/9.180
-06- HACH
4.00/7.00/10.18
-07- WTW techn. buffers
2.00/4.01/7.00/10.00
Specifications 42
Stratos 2201 pH / 2211 pH
– Manual entry of individual
buffer values (MAN)
– Data entry for pre-measured
electrodes (DAT)
– Temperature probe adjustment
Calibration Timer
0 to 9999 h
Calibration Ranges
asymmetry potential ± 60 mV slope 80 to 103 % For values outside this range, display message (Sensoface
)
Temp Input
Pt 100 / Pt 1000 / NTC 30 kW
Ranges
NTC –20.0 to +130.0 °C
–4 to +266 °F
Pt –20.0 to +150.0 °C
–4 to 302 °F
Meas. Error
< 0.5 K
2)
± 1 count
T emp Com­pensation
automatic with Pt 100 / Pt 1000 / NTC 30 kW or manual
Display
LC display, alarm LED
Loop Current
4 to 20 mA, floating 22 mA for error message
*)
supply voltage 12 to 30 V
Current Error
< 0.3 % of current value ± 50 mA
Current Source
3.80 mA to 22.00 mA
Start/End of Scale
*)
as desired within pH or mV ranges
Spans
*)
pH value 2.00 to 14.00 mV value 200 to 3000 mV
Stratos 2201 X pH / 2211 X pH
– Manual entry of individual
buffer values (MAN)
– Data entry for pre-measured
electrodes (DAT)
– Temperature probe adjustment
Calibration Timer
0 to 9999 h
Calibration Ranges
asymmetry potential ± 60 mV slope 80 to 103 % For values outside this range, display message (Sensoface
)
Temp Input
Pt 100 / Pt 1000 / NTC 30 kW EEx ia IIC
Ranges
NTC –20.0 to +130.0 °C
–4 to +266 °F
Pt –20.0 to +150.0 °C
–4 to 302 °F
Meas. Error
< 0.5 K
2)
± 1 count
T emp Com­pensation
automatic with Pt 100 / Pt 1000 / NTC 30 kW or manual
Display
LC display, alarm LED
Loop Current
EEx ib IIC
4 to 20 mA, floating 22 mA for error message
*)
supply voltage 12 to 30 V I
max
= 100 mA, P
max
= 0.8 W
Current Error
< 0.3 % of current value ± 50 mA
Current Source
3.80 mA to 22.00 mA
Start/End of Scale
*)
as desired within pH or mV ranges
Spans
*)
pH value 2.00 to 14.00 mV value 200 to 3000 mV
Specifications 43
Stratos 2201 pH / 2211 pH HART
Com-
munication
(Model 2211 only)
digital communication via FSK modu­lation of loop current, point-to-point connection, reading of device identification, mea­sured values, status and messages reading and writing of parameters
Explosion Protection
Data Retention
> 10 years (EEPROM)
EMC Emitted interference:
EN 61 326 Class B Immunity to interference: EN 61 326, EN 61 326/A1
Temperature Operating/environmental temp
–20 to +55 °C Transport and storage temp –20 to +70 °C
Enclosure Material: thermoplastic polyester , re-
inforced (polybutylene terephthalate) Protection: IP 65 Color: bluish gray RAL 7031
Cable Glands 3 Pg 13.5 threaded cable glands
(not mounted), up to 5 Pg threaded cable glands possible (only use plastic glands)
Dimensions See Dimension drawings, Pg. 9 ff
Weight
approx. 1 kg
*) user defined 1) doubles every 10 K
2) Pt 100: 1K
Stratos 2201 X pH / 2211 X pH HART
Com-
munication
(Model 2211X only)
digital communication via FSK modu­lation of loop current, point-to-point connection, reading of device identification, mea­sured values, status and messages reading and writing of parameters
Explosion Protection
II 2 G EEx ib [ia] IIC T6 CE 0032 TÜV 99 ATEX 1446
Data Retention
> 10 years (EEPROM)
EMC Emitted interference:
EN 61 326 Class B Immunity to interference: EN 61 326, EN 61 326/A1
Temperature Operating/environmental temp
–20 to +55 °C Transport and storage temp –20 to +70 °C
Enclosure Material: thermoplastic polyester , re-
inforced (polybutylene terephthalate) Protection: IP 65 Color: bluish gray RAL 7031
Cable Glands 3 Pg 13.5 threaded cable glands
(not mounted), up to 5 Pg threaded cable glands possible (only use plastic glands)
Dimensions See Dimension drawings, Pg. 9 ff
Weight
approx. 1 kg
*) user defined 1) doubles every 10 K
2) Pt 100: 1K
Specifications 44
EC-T ype-Examination Certificate
Specifications 45
Specifications 46
Specifications 47
Declaration of conformity
Buffer Charts 48
Buffer Charts
-00- Knick technical buffers
-01- (correspond to Mettler-Toledo technical buffers)
°CpH
0
2.03 4.01 7.12 9.52
5 2.02 4.01 7.09 9.45 10 2.01 4.00 7.06 9.38 15 2.00 4.00 7.04 9.32 20 2.00 4.00 7.02 9.26
25 2.00 4.01 7.00 9.21
30 1.99 4.01 6.99 9.16 35 1.99 4.02 6.98 9.11 40 1.98 4.03 6.97 9.06 45 1.98 4.04 6.97 9.03 50 1.98 4.06 6.97 8.99 55 1.98 4.08 6.98 8.96 60 1.98 4.10 6.98 8.93 65 1.99 4.13 6.99 8.90 70 1.99 4.16 7.00 8.88 75 2.00 4.19 7.02 8.85 80 2.00 4.22 7.04 8.83 85 2.00 4.26 7.06 8.81 90 2.00 4.30 7.09 8.79 95 2.00 4.35 7.12 8.77
-02- Merck-Titrisols, Riedel Fixanals °CpH
0
2.01 4.05 7.13 9.24 12.58
5 2.01 4.04 7.07 9.16 12.41 10 2.01 4.02 7.05 9.11 12.26 15 2.00 4.01 7.02 9.05 12.10
20 2.00 4.00 7.00 9.00 12.00
25 2.00 4.01 6.98 8.95 11.88 30 2.00 4.01 6.98 8.91 11.72 35 2.00 4.01 6.96 8.88 11.67 40 2.00 4.01 6.95 8.85 11.54 45 2.00 4.01 6.95 8.82 11.44 50 2.00 4.00 6.95 8.79 11.33 55 2.00 4.00 6.95 8.76 11.19 60 2.00 4.00 6.96 8.73 11.04 65 2.00 4.00 6.96 8.72 10.97 70 2.01 4.00 6.96 8.70 10.90 75 2.01 4.00 6.96 8.68 10.80 80 2.01 4.00 6.97 8.66 10.70 85 2.01 4.00 6.98 8.65 10.59 90 2.01 4.00 7.00 8.64 10.48 95
2.01 4.00 7.02 8.64 10.37
-03- Ciba (94) buffers Nominal values: 2.06, 4.00, 7.00, 10.00 °CpH
0
2.04 4.00 7.10 10.30
5 2.09 4.02 7.08 10.21 10 2.07 4.00 7.05 10.14 15 2.08 4.00 7.02 10.06 20 2.09 4.01 6.98 9.99 25 2.08 4.02 6.98 9.95 30 2.06 4.00 6.96 9.89 35 2.06 4.01 6.95 9.85 40 2.07 4.02 6.94 9.81 45 2.06 4.03 6.93 9.77 50 2.06 4.04 6.93 9.73 55 2.05 4.05 6.91 9.68 60 2.08 4.10 6.93 9.66 65 2.07* 4.10* 6.92* 9.61* 70 2.07 4.11 6.92 9.57 75 2.04* 4.13* 6.92* 9.54* 80 2.02 4.15 6.93 9.52 85 2.03* 4.17* 6.95* 9.47* 90 2.04 4.20 6.97 9.43 95 2.05* 4.22* 6.99* 9.38*
* extrapolated
-04- Mettler Toledo (USA) °CpH
0
4.00 7.12 10.32
5 4.00 7.09 10.25 10 4.00 7.06 10.18 15 4.00 7.04 10.12 20 4.00 7.02 10.06
25 4.00 7.00 10.01
30 4.01 6.99 9.97 35 4.02 6.98 9.93 40 4.03 6.98 9.89 45 4.04 6.97 9.86 50 4.06 6.97 9.83 55 4.06
*
6.97
*
9.83
*
60 4.06
*
6.97
*
9.83
*
65 4.06
*
6.97
*
9.83
*
70 4.06
*
6.97
*
9.83
*
75 4.06
*
6.97
*
9.83
*
80 4.06
*
6.97
*
9.83
*
85 4.06
*
6.97
*
9.83
*
90 4.06
*
6.97
*
9.83
*
95 4.06
*
6.97
*
9.83
*
* extrapolated
Buffer Charts 49
-05- NIST standard-buffers °CpH
0
4.010 6.984 9.464
5 4.004 6.951 9.395 10 4.000 6.923 9.332 15 3.999 6.900 9.276 20 4.001 6.881 9.225
25 4.006 6.865 9.180
30 4.012 6.853 9.139 35 4.021 6.844 9.102 40 4.031 6.838 9.068 45 4.043 6.834 9.038 50 4.057 6.833 9.011 55 4.071 6.834 8.985 60 4.087 6.836 8.962 65 4.109 6.841 8.942 70 4.126 6.845 8.921 75 4.145 6.852 8.903 80 4.164 6.859 8.885 85 4.185 6.868 8.868 90 4.205 6.877 8.850 95
4.227 6.886 8.833
-06- HACH buffers Nominal values: 4.00, 7.00, 10.18 °CpH
0
4.00 7.14 10.30
5 4.00 7.10 10.23 10 4.00 7.04 10.11 15 4.00 7.04 10.11 20 4.00 7.02 10.05 25 4.01 7.00 10.00 30 4.01 6.99 9.96 35 4.02 6.98 9.92 40 4.03 6.98 9.88 45 4.05 6.98 9.85 50 4.06 6.98 9.82 55 4.07 6.98 9.79 60 4.09 6.99 9.76 65 4.09
*)
6.99
*)
9.76
*)
70 4.09
*)
6.99
*)
9.76
*)
75 4.09
*)
6.99
*)
9.76
*)
80 4.09
*)
6.99
*)
9.76
*)
85 4.09
*)
6.99
*)
9.76
*)
90 4.09
*)
6.99
*)
9.76
*)
95 4.09
*)
6.99
*)
9.76
*)
* values complemented Buffer values up to 60 °C as specified by Bergmann & Beving Process
AB.
-07- WTW technical buffers °C ST 1 ST 2 ST 3 ST 5
0
2.03 4.01 7.12 10.65
5 2.02 4.01 7.09 10.52 10 2.01 4.00 7.06 10.39 15 2.00 4.00 7.04 10.26 20 2.00 4.00 7.02 10.13
25 2.00 4.01 7.00 10.00
30 1.99 4.01 6.99 9.87 35 1.99 4.02 6.98 9.74 40 1.98 4.03 6.97 9.61 45 1.98 4.04 6.97 9.48 50 1.98 4.06 6.97 9.35 55 1.98 4.08 6.98 60 1.98 4.10 6.98 65 1.99 4.13 6.99 70 2.00 4.16 7.00 75 2.00 4.19 7.02 80 2.00 4.22 7.04 85 2.00 4.26 7.06 90 2.00 4.30 7.09 95
2.00 4.35 7.12
Asymmetry potential (zero point)
Buffer set
Buffer solution
cal
Calibration
Calibration buffer set
Calimatic
Combination electrode
conf
Electrode slope
Electrode zero point
GainCheck
HART
T echnical Terms 50
Technical Terms
The voltage which a pH electrode gives off at a pH of 7. The asymme­try potential is different for every electrode and changes with age and wear.
Contains selected buffer solutions which can be used for automatic calibration with the Knick Cali­matic
. The buffer set must be se-
lected prior to calibration.
Solution with an exactly defined pH value for calibrating a pH meter.
Key for activating and ending cal­ibration.
Adjustment of the pH meter to the current electrode characteristics. The asymmetry potential (zero point) and slope are adjusted. Ei­ther a one-or two-point calibration can be carried out. With one-point calibration only the asymmetry po­tential is adjusted.
See Buffer set.
Automatic buffer recognition. Be­fore the first calibration, the buffer set used must be activated once.
The patented Calimatic
then auto­matically recognizes the buffer solu­tion used during calibration.
Combination of glass and reference electrode in one body.
Key for starting and ending configu­ration.
Is indicated in % of the theoretical slope (59.2 mV/pH at 25 °C). The electrode slope is different for every electrode and changes with age and wear.
See Asymmetry potential.
Unit self-test which runs automati­cally in the background at fixed in­tervals. The memory and mea­sured-value transmission are checked. Y ou can also start the GainCheck
manually (see Pg. 22). Then a display test is also con­ducted and the software version displayed.
Highway Addressable Remote Transducer , digital communication
via FSK modulation of the loop cur­rent
Mode code
One-point calibration
pH electrode system
Response time
Sensocheck
Sensoface
Slope
Two-point calibration
Zero point
T echnical Terms 51
Preset four-digit number to select certain modes.
Calibration with which only the elec­trode asymmetry potential (zero point) is taken into consideration. The previous slope value is re­tained. Only one buffer solution is required for a one-point calibration.
A pH electrode system consist of a glass and a reference electrode. If they are combined in one body, they are referred to as a combina­tion electrode.
Time from the start of a calibration step to the stabilization of the elec­trode potential.
Sensocheck
continuously moni­tors the glass and reference elec­trodes.
provides information on the status of the electrode. The zero point, slope and response time are eva­luated. The glass and reference electrodes are continuously moni­tored.
See Electrode slope.
Calibration with which the electrode asymmetry potential (zero point) and slope are determined. Two buffer solutions are required for two-point calibration.
See Assymmetry potential.
Index 52
Index
, 34
22 mA signal for alarm, 23, 35
configuring, 25
A
Alarm, 23
response time, 35 Alarm LED, 35 Alarm via loop current, 23, 35
configuring, 25 Assembly, 8 Asymmetry potential
definition, 50
display , 38 Automatic calibration
with automatic cal temp detec-
tion, 26
with manual cal temp, 27
B
Buffer charts, 48 Buffer set
definition, 50
selecting, 25 Buffer solution, definition, 50
C
Cal info, 33, 38
Calibration, 26
automatic
with automatic cal temp
detection, 26
with manual cal temp, 27
Calimatic
with automatic cal temp
detection, 26
with manual cal temp, 27 data entry, 31 definition, 50 manual
with automatic cal temp
detection, 29
with manual cal temp, 30 temp probe adjustment, 32
Calibration buffer set, definition, 50
Calibration data, display, 38 Calibration error messages, 37 Calibration mode, configuring, 25 Calibration timer interval, configur-
ing, 25 Calimatic, definition, 50 Certificate of Conformity, 44 Cleaning, 39 Combination electrode, definition,
50 Configuration, 24
Connecting
handheld terminal, 11
lines, 11 Connecting cable, fixing, 11 Current source, 38
D
Declaration of Conformity, 47 Diagnostics functions, 38 Dimension drawings, 9 Display , 21
E
Electrode monitoring, Sensoface, 22, 34
Electrode potential, display, 38 Electrode slope
definition, 50
display , 38 Electrode zero point, definition, 50 Error info, 33, 35, 38 Error message, last, 33, 35, 38 Error message via loop current,
23, 35
configuring, 25 Error messages, 35–38
G
GainCheck, 22, 38
definition, 50
Index 53
H
HART, definition, 50 HART communication, 23 Hold mode, 33
Input voltage, display, 38 Installation, 13
K
Keypad, 21
L
Loop current, 23
configuring, 25 display , instantaneous, 38 frozen, 33 Hold state, 25
M
Maintenance, 39 Manual calibration
with automatic cal temp detec­tion, 29 with manual cal temp, 30
Manual temperature specification, 33
Measured variable, configuring, 25
Measuring mode, 33
Messages, Sensoface, 34 Mode code, 6
definition, 51 Mounting diagram, 9 mV measurement, 33
O
One–point calibration, 27, 28, 29, 30
definition, 51 ORP measurement, 19 Outputs, 23
P
Packing list, 8 pH electrode system, definition,
51 pH measurement, 33 Pipe mount kit, 10 Product line, 40 Protective hood, 10
R
Response time, definition, 51
S
Safety precautions, 5
Self–test
automatic, 38 manual, 22, 38
Sensocheck, 34
definition, 51 on or off, 25
Sensoface, 22
definition, 51 diagnostics, 34 messages, 34
Slope
definition, 51
display , 38 Smiley, 34 Software version, display, 22, 38 Specifications, 41 Stripping lengths, 11
T
Technical terms, 50 Temp probe adjustment, 32 T emperature detection, configur-
ing, 25 Terminals, pulling out, 11 Two–point calibration, 29, 30
definition, 51 Type Examination Certificate, 44
U
Unit self–test, automatic, 22 User Interface, 20
Index 54
W
Wiring
ORP measurement, 19 pH measurement, 14
Z
Zero point, definition, 51
Loading...