Knick 765 Instruction Manual

Page 1
Instruction Manual
765 Laboratory pH Meter
Page 2
TA-167.000-KNE04 051009 Software version: 4.x
Subject to change without notice
Elektronische Meßgeräte GmbH & Co.
Tel: +49 (0) 30-80191-0 Fax: +49 (0) 30-80191-200 www.knick.de [email protected]
Warranty
Defects occurring within 3 years from delivery date shall be remedied free of charge at our works (carriage and insurance paid by sender). Accessories: 1 year
Page 3
The 765 Laboratory pH Meter I
Contents
Information on this instruction manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II
Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III
EC Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV
1 The 765 Laboratory pH Meter
Package contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Short device description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Meter design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Power-on and start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Parameter level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Calibration level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Diagnostics level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Trueline
®
recorder output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Serial interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Standard settings for ZU 0244 Lab Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Command set for the serial interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Product line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Buffer tables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Page 4
II Information
765 Laboratory pH Meter
Information on this instruction manual
Warnings and notes
Instructions marked with this sign must be strictly observed for reasons of your own safety! Failure to follow these instructions may result in injuries.
Notes provide important information that should be strictly followed when using the device.
Typical representations
Display example
A gray representation of the display text indicates a flashing display.
Keys whose functions are described.
Markings in the text
Keys are represented by bold-faced text,
e.g. , , , , , , ... , .
on/standby
Page 5
Information III
Safety information
Be sure to read and observe the following instructions!
Before connecting the device to the power supply, make sure that the voltage corre­sponds with the rating given on the rating plate of the device.
Opening the device exposes live parts. Therefore, it shall not be opened. If a repair should be required, return the de
vice to our factory.
If opening the device is inevitable, it shall first be disconnected from all voltage sources.
Make sure that the mains supply has been disconnected.
Repair or adjustment of an opened device under voltage shall be carried out only by a skilled person who is aw
are of the hazards involved.
Remember that the voltage across accessible parts of the open device may be dan­gerous to life.
Whenever it is likely that the protection has been impaired, the device shall be made inoper
ativ
e and secured against unintended operation.
The protection is likely to be impaired if, for example:
• the device shows visible damage
• the device fails to perform the intended measurements
• after prolonged storage at temperatures above 70°C
• after severe transport stresses
Before recommissioning the device, a professional routine test in accordance with EN 61010-1 m
ust be perf
ormed. This test should be carried out at our factory.
Page 6
]
Dokument-Nr. / Document
No. /
|
eCSOet
Zn
Aurbewahruns / Keepins / Garde en döpot
lNo.
document
I
Jürgen
cammin
(KB)
ililililil111ilililililililililil1il111il1ililililililililr1
I
wir, oie /
We, / Nous,
Niederspannungs-Richtlinie
/
Low-voltage directive
/
Directive basse
tension
Harmonisierte
Normen
/ Harmonised
Standards
/
Normes harmonisÖes
I
f ulv-Ricfrttinie
/
EMC directive
/
I
Directive CEM
I
Norm / Standard
/
Norme
I
Ausstellungsort,
-datum
/
Place and
date of
issue i
Lieu et
date d'Ömission
ililililllillililililililililililililililililililililililililililililililililllllllllllllllllllllllllll
E
G
-Konfo
rm itätserkl äru
n
g EC Declaration of Gonformity D6claration de Gonformit6 CE
Knick
)
Knick Elektronische
Messgeräte
GmbH & Co.
KG
Beuckestr.
22
D-14163 Berlin
Jahr
der An0ringung der CE-Kennzerchnung
/
1 995
Year In whrch the CE
mark ng was affixed
/
L'annee
d'apposition du
marquage CE
Knick Elektronische
Messgeräte GmbH & Go.
KG
Beuckestr.
22, D-14163 Berlin
erklären
in alleiniger
Verantwortung, daß dieses
Produkt / diese Produkte,
declare under
our sole responsibility
that the
product
/
products,
d6clarons sous
notre seule
responsabilitÖ
que
le
produit
/ les
produits,
Labor-pH-Meter
765, Opt.
...
]
Produktbezeichnung
/
Product
identification /
Designation
du
produit
I
auf
welche(s) sich
diese Erklärung
bezieht,
mit allen
wesentlichen
Anforderungen der
folgenden
Richtlinien des
Rates übereinstimmen:
I
to
which this declaration
relates
is/are
in
conformity
with all essential
requirements
of the Council
Directives
relating
to:
J
auqueliauxquels
se röfere
cette döclaration
esVsont
conforme(s)
aux exigences
essentielles
de
la Directives du Conseil
relatives ä:
2006/95/EG
EN 61010-1:2001
20041'l08lEG EN 61326-1:
2006
EN 61326-2-3:
2006
Berlin.
17.08.2009
Knick
Elektronische
Messgeräte
GmbH & Co.
KG
ppa. Bernhard
Kusig
(Vice
President
Wolfgang
Page 7
Operation 1
The 765 Laboratory pH Meter
Package contents
After unpacking, please check the shipment for completeness. The package should contain:
• 765 Laboratory pH Meter
• Power cord
• This instruction manual
Short device description
• The Model 765 is used for pH and ORP measure­ments in the laboratory.
• Temperature compensation is automatic using a Pt 100 or Pt 1000 temperature probe or manual by entering the temperature.
• Buffer solutions of eleven different, selectable buffer sets are provided for calibration. Buffer rec­ognition is automatic via Calimatic
®
.
• The calibration timer alerts you when the preset interval between two calibrations has expired.
• Sensoface
®
electrode monitoring checks the con­nected electrode and provides information on the electrode condition.
• Knick Fullcheck
®
device self-test checks the
device functions at the press of a key.
• Records of parameter setting, calibration and diagnostics are particularly helpful for QM docu­mentation to ISO 9000 and GLP. The records can be directly output to a standard printer via the inte­grated serial interface.
Page 8
765 Laboratory pH Meter
2 Operation
2 Operation
Meter design
1, 2 Electrode connection 3, 4 Temperature probe connection 5 DS connection 6, 7 Recorder output 8 Interface connection 9 Power input 10 Sensoface
®
display 11, 13 Displays 12, 14 Measurement symbols
Page 9
Operation 3
General information
Keypad
Pressing turns the pH meter on or switches to standby mode. Standby mode is indi­cated by a lighted Sensoface
®
status display or two measurement symbols. At power-on, the pH meter automatically performs a short self test and then goes to measuring mode.
Pressing opens the Calibration level. On the Calibration level the pH meter is adapted to the electrode. You can conduct either a one or a two­point calibration.
Pressing opens the Parameter level. On the Parameter level all variable device parameters are set.
• In the VIEW menu you can view all parameters.
• In the EDIT menu you can also edit the parame-
ters.
Pressing opens the Diagnostics level. On the Diagnostics level you can get information on the electrode condition, view the current electrode data, and perform a complete device self test.
• The Sensoface
®
menu lists the individual param-
eters of Sensoface
®
automatic electrode monitor-
ing with the corresponding evaluations.
• The Electrode Data menu shows the values for
zero and slope of the electrode determined during last calibration, the remaining time until the next due calibration determined by the calibration timer and the data of the electrode statistics.
• With the Knick Fullcheck
®
menu, a complete
device self-test is performed.
on/standby
cal
par
diag
Page 10
765 Laboratory pH Meter
4 Operation
Pressing stores an entered parameter. If you have not made any changes, pressing selects the next parameter (instead of ).
Pressing exits a function level and returns you to measuring mode.
Pressing in measuring mode prints out the current pH, mV and temperature values with date and time. Pressing on one of the function levels prints out a complete record of the data stored.
• Pressing or in measuring mode selects the
measured variable for the left or right display, respectively. The corresponding measurement symbol appears on the right side of the display.
• On the function levels these keys are used to
select parameters.
• When entering numerical parameters, they are
used to increment or decrement a numeral.
• Pressing in the EDIT menu of the Parameter
level selects the parameter you want to edit.
• Pressing this key in the VIEW menu or on the
Diagnostics level stops automatic line scrolling.
enter
meas
print
Page 11
Operation 5
Menu structure
Sensoface® automatic electrode monitoring
Sensoface® automatic electrode monitoring pro­vides information on the electrode condition. Zero point, slope and response time are evaluated dur­ing calibration.
Impedance, dry-out and the calibration timer are evaluated during measurement and in standby mode.
A summary of the individual results is expressed by three face symbols.
The electrode is in good condition.
The electrode is still usable, but should be cali­brated.
The electrode urgently requires calibration. If the display does not change after calibration, replace the electrode.
Page 12
765 Laboratory pH Meter
6 Operation
More detailed information on the indicated elec­trode condition and the individual parameter evalu­ations is given in chapter ”Diagnostics level” (see Pg. 21).
Sensoface
®
was specially designed for monitoring pH electrodes. Since ORP electrodes have com­pletely different characteristics, they provoke invalid Sensoface
®
displays. Therefore, you should
disable Sensoface® for ORP measurements.
Sensoface® display can also be degraded when you use a pH simulator.
Power-on and start-up
Mains supply The pH meter is designed for 230 VAC supply
(Option 363: 115 V AC). Connect the power input of the pH meter to a mains outlet using the included power cord.
If the pH meter is disconnected from mains, clock and calibration timer are not affected (reserve power approx. 1 year). Settings, calibration and diagnostics data remain permanently stored.
Electrode connection The pH meter can be combined with commercial
electrodes: with nominal zero at pH 7 and with
• standard coaxial plug to DIN 19 262 and/or
•4 mm banana plug.
For special electrodes with differing specifications, we offer Option 346 (nominal zero and slope of electrode user defined).
Page 13
Operation 7
Connection Socket
Combination electrode 1 Single sensing electrode 1 Single reference electrode 2 Temp probe of combination electrodes 3 (Electrodes with integrated temp probe) Separate temperature probe 3, 4
If no temperature probe is connected, the meter uses the manually selected temperature. The deci­mal point of the temperature display flashes.
If you use double platinum electrodes for voltamet­rical indication of titrations:
• Connect the banana plugs of the double platinum
electrode to sockets 1 and 2.
• To connect the 10 µA polarization current to the
double platinum electrode, connect socket 5 with the inner pole of socket 1 (e.g. using a banana plug with cross hole).
For the RS 232 interface, a shielded cable must be used (e.g. ZU 0152, ZU 0153).
Page 14
765 Laboratory pH Meter
8 Operation
Standby mode If the meter is connected to a mains outlet but not
switched on, it is in standby mode. This indicated by a lighted Sensoface
®
display.
If Sensoface® display is turned off, two measure­ment symbols are lighting.
Clock and calibration timer are running in standby mode. Knick Drycheck
®
and impedance monitoring are active. Settings, calibration and diagnostics data remain permanently stored. The interface is deactivated.
Pressing switches the meter to mea­suring mode.
At power-on, the meter performs a short check:
• Simultaneous lighting-up of all display segments,
measurement symbols and Sensoface
®
displays
• Display of model name PH 765
• Display of software version
• Display of nominal buffer values of selected buffer
set
• Memory test
To stop the short check, press .
Do not disconnect the meter from mains if you do not use it. Only when the meter remains in standby mode can the electrode be monitored and the bat­tery for the clock be charged.
on/standby
Page 15
Operation 9
Measuring mode In measuring mode both displays indicate the
respectively selected measured variable. The vari­able for the left display is selected using , for the right display using . The corresponding mea­surement symbol is shown on the right side of the display. You can choose any of the following vari­ables:
• pH value
• Electrode potential [mV]
• Temperature [°C]
• Time
A flashing point on the temperature display indi­cates that no temperature probe is connected. The meter now uses the manually selected tempera­ture.
If you have connected a printer, pressing in the measuring mode gives you a printout of the cur­rently measured values with date and time.
print
Page 16
765 Laboratory pH Meter
10 Operation
Parameter level
On the Parameter level all variable device parame­ters are set. Parameters are set in dialog mode using different menus.
Activating parameter setting Press to exit measuring mode and access the
main menu of the Parameter level. To exit Parame­ter level, press . It is exited automatically when all parameter setting steps have run through.
Main menu In the main menu of the Parameter level you select
the desired submenu using or . Press­ing confirms your choice and gives access to the corresponding submenu.
The VIEW menu automatically displays all parame­ters . Settings cannot be changed, however.
• To hold the display, press .
• To scroll one line forwards or backwards, press
or , respectively.
In the EDIT menu you can view and edit all param­eters.
• Select the parameter you want to edit using or
.
• Press to start parameter editing. To indicate
that the parameter can now be edited, the right display flashes.
• While the right display is flashing, you can edit the
parameter using and . When entering numerical values, select the position using and then count up or down using or .
• Confirm with . The selected value will be
stored and the next parameter displayed.
• You can now select the next parameter using
or or press to exit the Parameter level.
par
Page 17
Operation 11
Typical setting procedure The meter is factory set to calibration buffer set 01
(Mettler Toledo technical trade buffers). However, you want to calibrate using the technical buffer solutions to DIN 19 267, calibration buffer set 03. Therefore, you must select buffer set 03 on the Parameter level.
Press to access the Parameter level.
Press to select the EDIT menu.
Press to open the EDIT menu.
Select buffer set parameter BUFF using .
Now press to change the buffer set. The right display will flash.
Select buffer set 03, technical buffers to DIN 19 267, using .
par
enter
Page 18
765 Laboratory pH Meter
12 Operation
Press to confirm your choice. Buffer set 03 will be stored and the next parameter displayed.
Press to exit Parameter level.
Printout of parameter record If you have connected the pH meter to a printer, you
can print out all stored parameter settings. Press to open the Parameter level. Press to print out all settings. The meter will automatically return to measuring mode.
Parameters In the following, the individual parameters and their
possible settings will be explained.
Manual temperature (–50 to +150 °C)
If no temperature probe is connected, the meter uses the manually selected temperature. In this case, the decimal point of the temperature display will flash in measuring mode.
Sensoface
®
(off / on)
Display of Sensoface® electrode monitoring can be turned on or off. If Sensoface® display is turned off, standby mode is indicated by two lighted measure­ment symbols. Display of Sensoface
®
parameters
on the Diagnostics level is not affected.
For information on Sensoface®, refer to chapter ”Sensoface® menu” (Pg. 22).
enter
meas
print
Page 19
Operation 13
Sensoface® was specially designed for monitoring pH electrodes. Since ORP electrodes have com­pletely different characteristics, they provoke invalid Sensoface
®
displays. Therefore, you should disable Sensoface® for ORP measurements. Sen­soface
®
display can also be degraded when you
use a pH simulator.
Displaymatic® (off / on)
Displaymatic® can be turned on or off. Display­matic
®
ensures a stable measured value display in case of rapid changes of the pH value. This allows for easy reading of the value. It is achieved by sup­pressing the two running digits behind the decimal point. When the value changes slowly, only the last digit will be suppressed. When the value has stabi­lized, all digits will be displayed.
Only for meters with Option 346:
Setting the nominal electrode zero point and slope:
Nominal zero (pH 0.00 to 14.00)
For special electrodes with differing nominal zero. The entered zero point is taken as reference for Calimatic
®
. The exact zero point will be determined
during calibration.
Nominal slope (25.0 to 61.0 mV/pH)
For special electrodes with differing nominal slope. The entered slope is taken as reference for Cali­matic
®
. The exact slope will be determined during calibration.
Isothermal potential (–500 to +500 mV)
The isothermal potential entered here is automati­cally taken into account for calibration and pH determination.
Page 20
765 Laboratory pH Meter
14 Operation
Calibration buffer set (00 to 10)
Select the buffer set with the buffer solutions you are using. The meter has stored eleven different calibration buffer sets.
(see Buffer sets Pg. 49).
The meter can only operate properly when the buffer solutions used correspond to the selected, activated buffer set.
Other buffer solutions, even those with the same nominal values, may demonstrate a different tem­perature behavior.
This leads to measurement errors.
BUFF –00– Knick technical buffers
pH 2.00 4.01 7.00 9.21 (25°C
BUFF –01– Mettler-Toledo technical buffers (correspond to Ingold technical buffers)
pH 2.00 4.01 7.00 9.21 (25°C)
BUFF –02– Merck/Riedel
pH 2.00 4.00 7.00 9.00 12.00 (20 °C)
BUFF –03– DIN19267
pH 1.09 3.06 4.65 6.79 9.23 12.75 (25°C)
BUFF –04– DIN19266 and NIST (NBS)
pH 1.679 4.006 6.865 9.180 12.454 (25°C)
BUFF –05– Merck/Riedel
pH 1.00 3.00 6.00 8.00 10.00 13.00 (20°C)
BUFF –06– Merck
pH 4.66 6.88 9.22 (20°C)
BUFF –07– Ciba(94)
pH 2.06 4.00 7.00 10.00
BUFF –08– customer specific
pH 2.00 4.62 7.00 8.20 10.00 12.00 (20°C)
BUFF –09– customer specific
pH 2.00 4.01 7.00 9.21 12.00 (25 °C/20 °C)
BUFF -10– Mettler Toledo (USA)
pH 4.00 7.00 10.01 (25 °C)
Page 21
Operation 15
First calibration (no / yes)
With YES, all statistics data will be erased with the next calibration, and the new electrode data will be stored as First Calibration data. After calibration, this parameter will automatically reset to NO.
You should always perform this step when replac­ing the electrode so that a new statistic is started for the new electrode. For further information on elec­trode statistics, refer to chapter ”Diagnostics level” (Pg. 25).
Calibration timer (0 to 2000 h)
With the calibration timer, you can preset a time interval for the next due calibration. After approx. 80 % of the preset interval have expired, the timer sets the Sensoface
®
display from to . After
the total interval has expired, the display is set to
. The timer is reset by calibration. To turn off the
timer, enter 0 as interval time.
Recorder output (pH / mV / °C / PRNT)
The recorder output can output pH value, mea­sured voltage or temperature.
With PRNT selected, the recorder output serves as input for activating a print command. With a simple contact, such as a foot switch, the currently mea­sured values can be printed. For further informa­tion, refer to chapter ”Trueline
®
recorder output”
(Pg. 30).
Do not apply an external voltage to the recorder output. The meter might be damaged.
Output voltage:
OUT PH 100 mV/pH OUT mV 1 mV/mV OUT ° C 10 mV/°C
Page 22
765 Laboratory pH Meter
16 Operation
Baud rate (600 / 1200 / 2400 / 4800 / 9600)
Here, you can select the interface transmission rate.
Data format (8 NO / 7 EV / 7 OD)
You can choose between:
Transmission protocol (NO / XON)
The interface can operate either without transmis­sion protocol or with XON/XOFF protocol.
Interface (PRNT / PC)
For direct printer control, select PRNT. You can directly print out measured values and records. To connect the meter to a computer (PC), select PC. The pH meter will be completely computer control­lable. All measured values and parameters will be retrievable via computer.
Print interval timer (0.1 to 999.9 min)
This step only appears if you have set the interface for printer control. The print interval timer presets an interval for printout of currently measured values with time and date. To turn off the timer, enter 0 as interval time.
Data word length Parity
8 bits none 7 bits even 7 bits odd
Page 23
Operation 17
Time
Date
Year
End of VIEW menu.
End of EDIT menu.
Page 24
765 Laboratory pH Meter
18 Operation
Calibration level
Calibration adjusts the pH meter to the electrode in use. It is adjusted for zero and slope of the elec­trode.
You only have to select the used buffer set once on the Parameter level. The patented Calimatic
®
then automatically recognizes the buffer solution in use, calculates zero point and slope of the electrode and adjusts correspondingly. In addition, the Model 765 measures the electrode response time during cali­bration.
Activating calibration Press to exit measuring mode and access the
Calibration level. You can stop calibration at any time by pressing .
The meter can only operate properly when the buffer solutions used correspond to the buffer set selected on the Parameter level. Other buffer solu­tions, even those with the same nominal values, may demonstrate a different temperature behavior. This leads to measurement errors.
Immerse the electrode and temperature probe in the first buffer solution. This can be any of the selected solutions. Press or to start cal­ibration. If you do not want to calibrate, select CAL1 NO using and press to exit Calibration level.
The response time of the electrode and tempera­ture probe is considerably reduced if the electrode is first moved about in the buffer solution and then held still. The values will stabilize more quickly.
cal
Page 25
Operation 19
During buffer recognition CAL 1 is flashing. The right display reads the measured temperature. When you use manual temperature compensation, the entered temperature is displayed. This is indi­cated by a flashing decimal point.
The nominal value of the identified buffer solution is displayed for approx. 5 sec. Electrode and temper­ature probe still remain in the first buffer solution.
Stability of the electrode is checked. The left-hand display reads the measured mV value. The right­hand display shows a running clock measuring the electrode response time. For each revolution (4 sec), a mark is added to the tally beside the clock. Stability check can be stopped by pressing
. However, this reduces calibration accuracy.
Calibration with the first buffer is terminated. Take the electrode and temperature probe out of the solution and rinse them thoroughly.
• If you only want to perform a one-point calibra-
tion, press to stop calibration, or select CAL2 NO using and press to exit the Calibration level. The meter will display the new, calculated zero point and the old slope and return to measuring mode.
• If you want to perform a two-point calibration,
immerse the electrode and temperature probe in the second buffer solution. Press or
to restart calibration. The calibration process
runs again as for the first buffer.
Buffer recognition, CAL 2 flashing, display of tem­perature
Page 26
765 Laboratory pH Meter
20 Operation
Display of nominal buffer value for approx. 5 sec
Check of electrode stability and response time
Display of zero point determined for the electrode
Display of slope determined for the electrode
End of calibration
Printing calibration record If you have connected the meter to a printer, you
can print out a record of the last calibration.
Press to open the Calibration level. Press to print out the complete record of the last calibration. The meter will automatically return to measuring mode.
print
Page 27
Operation 21
Diagnostics level
On the Diagnostics level the complete pH measure­ment equipment is checked. This also serves for quality management to ISO 9000.
Activating diagnostics Press to exit measuring mode and access
the main menu of the Diagnostics level. To exit Diagnostics level, press . It is exited auto­matically when all diagnostics functions have been performed.
Main menu In the main menu you select the desired function of
the Diagnostics level using or . Pressing
gives access to the corresponding sub-
menu.
The Sensoface
®
menu shows the states of the cri-
teria that control Sensoface
®
display.
The Electrode Data menu shows the electrode data calculated during last calibration, the calibration timer count and the electrode statistics.
The Knick Fullcheck
®
menu provides a complete device self test to check the operability of the indi­vidual device components.
All submenus are run automatically but can be influ­enced via keypad.
• To stop the sequence, press . To restart it,
press once more.
• To scroll one step forwards or backwards, press
or , respectively.
diag
Page 28
765 Laboratory pH Meter
22 Operation
Sensoface® menu In the Sensoface
®
menu you see the criteria that control Sensoface® display. Each criterion is dis­played separately with the corresponding evalua­tion. This gives you important information on the condition of you electrode. Sensoface
®
alerts you in time for possible errors or aging of your elec­trode.
The worsening of a Sensoface
®
criterion leads to the devaluation of the Sensoface® display. This devaluation is permanent. An improvement can only take place after a calibration.
Here, the Sensoface
®
display provides information on zero and slope of the electrode. For the exact zero and slope values, refer to the Electrode Data menu.
Zero and slope of the electrode are okay.
Zero and slope of the electrode are still okay. How­ever, the electrode should be replaced soon.
Zero and/or slope of the electrode have reached values which no longer ensure proper calibration. It is advisable to replace the electrode.
The values for zero and slope are determined dur­ing calibration only. Therefore, correct information depends on a proper calibration. Make sure that you always use correct and pure buffer solutions.
Page 29
Operation 23
The Sensoface® display provides information on the electrode’s response time. This is the time the electrode requires for delivering a stable measured value. The value determined during calibration. Wear and aging, but also wrong treatment (such as dry storage) can reduce the gel layer of the glass membrane. This results in a longer response time. The electrode gets sluggish.
The electrode promptly adjusts to the measured value.
The electrode adjusts slowly. You should consider replacing it. It may be possible to achieve an improvement by cleaning, or in the case of an elec­trode stored dry, by ”watering”.
The electrode adjusts very slowly to the measured value. Correct measurement is no longer ensured. The electrode should be replaced.
Page 30
765 Laboratory pH Meter
24 Operation
On the Parameter level, the calibration timer allows to preset a time interval for the next due calibration. The calibration timer keeps running in standby mode and with mains supply disconnected.
The calibration interval is still running.
Over 80 % of the calibration interval have run out.
The calibration interval has been exceeded.
®
menu.
Sensoface® was specially designed for monitoring pH electrodes. Since ORP electrodes have com­pletely different characteristics, they provoke in­valid Sensoface
®
displays. Therefore, you should disable Sensoface® for ORP measurements. Sensoface
®
display can also be degraded when
you use a pH simulator.
Page 31
Operation 25
Electrode Data menu The Electrode Data menu shows the values for
zero and slope of the electrode calculated during the last calibration, the cal timer count, and the electrode statistics.
Display of electrode zero determined during last calibration.
Display of electrode slope determined during last calibration.
Display of time in hours until the next calibration is requested by the timer.
The electrode statistics include the electrode zero, slope and response time of the last three calibra­tions with corresponding date and time. In addition, the data of the first calibration (see Pg. 15) per­formed on this electrode are provided. This allows you to evaluate electrode behavior over its whole life span. Since the data are very comprehensive, they should be printed out or read out via computer.
Date of last calibration
Date of second last calibration
Date of third last calibration
Date of first calibration
Page 32
765 Laboratory pH Meter
26 Operation
Time of last calibration
Time of second last calibration
Time of third last calibration
Time of first calibration
Electrode zero determined during last calibration
Electrode zero determined during second last cali­bration
Electrode zero determined during third last calibra­tion
Electrode zero determined during first calibration
Slope determined during last calibration
Slope determined during second last calibration
Slope determined during third last calibration
Slope determined during first calibration
Page 33
Operation 27
Electrode response time of last calibration
Electrode response time of second last calibration
Electrode response time of third last calibration
Electrode response time of first calibration
End of Electrode Data menu.
Knick Fullcheck® menu With the Knick Fullcheck
®
menu, a complete device self-test is performed. The complete measuring cir­cuitry, measured value processing, memories, dis­play and keypad are checked and each result is displayed. The tests run automatically. Only during keypad testing you have to press the requested keys. Longer testing periods are indicated by a little running clock on the right display.
Successful testing is confirmed by OK on the right display.
RAM test
EPROM test
EEPROM test
Amplifier linearity test: Using an integrated, high­precision reference, the complete measuring cir­cuitry from input amplifier to recorder output is checked. The displayed mV values lie across the recorder output. The electrode is internally discon­nected from the amplifier.
Page 34
765 Laboratory pH Meter
28 Operation
Linearity test at -1500 mV
Linearity test at -750 mV
Linearity test at 0 mV
Linearity test at +750 mV
Linearity test at +1500 mV
Test of impedance measuring circuit
Amplifier linearity test is terminated. The device takes a few seconds to bumplessly reconnect the electrode to the amplifier.
Display test: All Sensoface
®
indicators, all seg­ments of the two LED displays and all measure­ment symbols light up.
Check whether really everything is lighting.
During keypad testing, you are prompted to press the corresponding key.
Press .
Press .
Press .
Press .
Press .
Press .
Page 35
Operation 29
Press .
Press .
Press .
Press .
End of device test
Printing diagnostics record If you have connected the meter to a printer, you
can print out a record of the diagnostics. Press to open the Diagnostics level.
Press to print out the complete diagnostics record.
print
Page 36
765 Laboratory pH Meter
30 Operation
Trueline® recorder output
The Trueline® recorder output was developed for the 765 Laboratory pH Meter. Trueline
®
supplies an analog output signal without disturbing voltage steps, but being calibrated to pH. Galvanic output isolation is standard. Connected recorders and data acquisition systems thus do not have to be floating.
On the Parameter level, the output can be set for three different input variables:
If pH has been selected as output variable, the out­put voltage zero is at pH 7. For meters with Option 346 (Nominal electrode zero and slope user­defined), the output voltage zero is at the selected nominal zero point.
If the recorder output has been set for printer con­trol, a voltage of approx. 1.5 V lies across its output. By short-circuiting (current approx. 1.5 mA), e.g. using a foot switch, you can print out the currently measured values.
Do not apply an external voltage to the recorder output. The meter might be damaged.
Input variable Output voltage
pH value 100 mV/pH Voltage 1 mV/mV Te mperature 10 mV/°C
Page 37
Operation 31
Serial interface
The Model 765 comes with an RS 232 interface. The interface is user-definable for direct control of a printer with serial input or for direct connection to a computer. The pH meter is completely computer controllable and all values and parameters can be read out.
Interface parameters The RS 232 interface is user-definable for all com-
mon baud rates and data protocols.
Settings are made on the Parameter level.
• Baud rate: 600 / 1200 / 2400 / 4800 / 9600 Bd
• Data format:
• Protocol: No protocol
XON/XOFF bidirectional handshake If not ready to accept data, the meter transmits XOFF < 13 > H, if ready to accept data, XON < 11 > H
Pin assignment The pH meter has a 9-pin D SUB connector (con-
nector with pin contacts). The metallic connector shell is connected to signal ground via a capacitor and provides EMI shielding.
Data word length Parity Stop bit
7 bits even 1 7 bits odd 1 8 bits none 1
Contact Signal Input/Output
2 (RD) Received data Input 3 (TX) Transmitted data Output 5 (SGND) Signal ground 4,6,8 Jumpered
Page 38
765 Laboratory pH Meter
32 Operation
Interface cable Knick offers as accessory:
• ZU 0245 interface cable for connecting the
pH meter to the ZU 0244 Lab Printer.
• ZU 0152 interface cable for connecting the
pH meter to a computer (PC):
Standard settings for ZU 0244 Lab Printer
Settings on Laboratory pH Meter
Command set for the serial interface
The pH meter’s command set is divided into read and write commands.
• Read commands start with ”R”. They read out
data from the pH meter. Read commands always return a response. Device function is not affected.
• Write commands start with ”W”. They send com-
mands and parameters to the pH meter. A write command modifies device settings or parameters. The meter does not return a response. Acknowl­edgement of write commands can be enabled with the ”WPMSR1” command. Then the meter will return ”CR” (carriage return <0D> H) after each write command.
D-SUB 9-pin socket D-SUB 9-pin socket
2 (RD) 3 3 (TD) 2 5 (SGND) 5 4 4 6 6 8 8
Connector shell (SHIELD) Connector shell
Parameter Value to be set Setting
Baud rate 4800 Bd 4800 Data format 7 data bits,
parity even
7 EV
Protocol XON/XOFF XON Interface Printer PRNT
Page 39
Operation 33
Message terminator Message terminators for read and write
• Read: The pH meter terminates the string by ”CR”
(carriage return <0D> H).
• Write: The pH meter expects ”CR” or ”LF” (line
feed <0A> H) or any combination of these as mes­sage terminator.
Numerical parameters Format of a numerical parameter
• Mantissa:
preceded by +, –, blank or without sign up to 14 valid digits incl. sign Decimal point or comma floating or without
• Exponent:
”E” indicates exponent beginning + or – sign 1 to 3 digits Exponent may be omitted.
Example: +1234567E–123
Exponent: Numerical digits Sign Exponent beginning Mantissa: Numerical digits Sign
Reading measured values or results
Com­mand
Response Unit Description
RV0 ± xx.xx [pH] pH value RV1 ± xxxxE–3 [V] Electrode
potential
RV2 ± xxx.x [°C] Pt100/Pt1000
temperature
R
VTRT xxxx [hhmm] Time: hours/
minutes
RVDRT xxxxxx [ddmmyy] Date: day/month/
year
RVTCA xxxx [h] Calibration timer
count
RVZA ± xx.xx [pH] Current
electrode zero point
RVSA ± xxxxE–3 [V/pH] Current
electrode slope
Page 40
765 Laboratory pH Meter
34 Operation
Reading error messages
List of error messages (xx):
Reading operating states
Reading Sensoface
®
states
List of Sensoface® messages (x):
Com­mand
Response Description
RSF1 xx First error message RSFA xx;xx;xx;... All active
error messages
List of error messages (xx):
01 pH range –2.00 to +16.00 exceeded 02 mV range -2000 to +2000 mV exceeded 03 °C range –50. to +150 °C exceeded 04 Zero range pH 6.00 to 8.00 exceeded 05 Slope range 47.0 to 67.0 mV/pH exceeded 09 Buffer cannot be identified 10 Identical buffers 20 Interface error 90 System failure
Com­mand
Re­sponse
Description
RSP xx Operating states
00 Measuring mode 01 Parameter mode 02 Calibration mode 08 Diagnostics mode
Com­mand
Re­sponse
Description
RSES x
Sensoface® display
RSEZS x
Sensoface
®
: zero/slope
RSETR x
Sensoface
®
: electrode response
RSERG x
Sensoface
®
: electrode impedance
RSERT x
Sensoface
®
: Drycheck
RSETC x
Sensoface
®
: Calibration timer
0
1
2
Page 41
Operation 35
Query keypad
Reading results of Knick Fullcheck® self test
Reading calibration data
Com­mand
Re­sponse
Description
RSK xx Query keypad
00
key
01
key
02
key
03
key
04
key
05
key
06
key
07
key
08
key
09
key
Command Re-
sponse
Unit Description
RSTET xxxx [hhmm] Last Fullcheck/
time RSTED xxxxxx [ddmmyy] Last Fullcheck/date RSTERR x RAM test RSTERP x EPROM test RSTERE x EEPROM test RSTERA x Amplifier test RSTERI x Impedance test RSTERDI x Display test RSTERKY x Keypad test
List of test results
0 ok (display test executed) 1 Test not executed 2 Defective
Command Response Unit Description
RSCPT xxxx [hhmm] Last calibration/
time
RSCPD xxxxxx [ddmmyy] Last calibration/
date RSCPS xx Buffer set RSCP1NB ± xx.xx [pH] Buffer 1:
nominal value RSCP11 ± xxxxE–3 [V] Buffer 1: electrode
potential RSCP12 ± xxx.x [°C] Buffer 1:
temperature RSCP1TR xxxx [s] Buffer 1: response
time
Page 42
765 Laboratory pH Meter
36 Operation
Reading electrode statistics
RSCP2NB ± xx.xx [pH] Buffer 2: nominal
value RSCP21 ± xxxxE–3 [V] Buffer 2: electrode
potential RSCP22 ± xxx.x [°C] Buffer 2:
temperature RSCP2TR xxxx [s] Buffer 2: response
time
Command Response Unit Description
Com­mand
Response Unit Description
Record 3 (last calibration): RSSTT3 xxxx [hhmm] Calibration/time RSSTD3 xxxxxx [ddmmyy] Calibration/date RSSTZ3 ± xx.xx [pH] Electrode zero point RSSTS3 ± xxxxE–3 [V/pH] Electrode slope RSSTTR3 xxxx [s] Electrode response
Record 2 (second last calibration): RSSTT2 xxxx [hhmm] Calibration/time RSSTD2 xxxxxx [ddmmyy] Calibration/date RSSTZ2 ± xx.xx [pH] Electrode zero point RSSTS2 ± xxxxE–3 [V/pH] Electrode slope RSSTTR2 xxxx [s] Electrode response
Record 1 (third last calibration): RSSTT1 xxxx [hhmm] Calibration/time RSSTD1 xxxxxx [ddmmyy] Calibration/date RSSTZ1 ± xx.xx [pH] Electrode zero point RSSTS1 ± xxxxE–3 [V/pH] Electrode slope RSSTTR1 xxxx [s] Electrode response
Record 0 (first calibration): RSSTT0 xxxx [hhmm] Calibration/time RSSTD0 xxxxxx [ddmmyy] Calibration/date RSSTZ0 ± xx.xx [pH] Electrode zero point RSSTS0 ± xxxxE–3 [V/pH] Electrode slope RSSTTR0 xxxx [s]
Electrode response
Page 43
Operation 37
Reading parameters
Command Response Unit Description
RPTMMV ± xxx.x [°C] Manual temperature RPCATI xxxx [h] Calibration timer
interval RPINPTI xxx.x [min] Print timer interval RPMSR x Response to
write command on/off
0 Off 1 On
RPDIE x
Sensoface
®
on/off 0 Off 1 On
RPDIDA x
Displaymatic
®
on/off 0 Off 1 On
RPAINA x Assignment for
recorder output 0 pH 1 mV 2 Te mperature 8 Input for
printer control
RPCASA xx Calibration buffer set
00 Knick technical
buffers 01 Mettler Toledo
technical buffers 02 Merck Titrisol buffers 03 Techn. buffers to
DIN 19 276 04 Standard buffers to
DIN 19 266 05 Merck Titrisol buffers 06 Merck buffers ready
to use 07 Ciba (94) 08 customer specific 09 customer specific 10
Mettler Toledo (USA)
Page 44
765 Laboratory pH Meter
38 Operation
Only for meters with Option 346:
Only for meters with Option 346: Nominal electrode zero and slope user-defined
Com­mand
Response Unit Description
RPCA0NZ ± xx.xx [pH] Nomminal zero RPCA0NS ± xxxxE–3 [V/pH] Nominal slope RPCA0U ± xxxxE–3 [V] V iso isothermal
potential
Page 45
Operation 39
Writing parameters
Command Parameter Unit Description
WPTMMV Num. par. [°C] Manual temperature WPCATI Num. par. [h] Calibration timer
interval
WPINPTI Num. par. [min] Print timer interval WPMSR x Response to
write command on/off 0 Off 1 On
WPDIE x
Sensoface
®
on/off 0 Off 1 On
WPDIDA x
Displaymatic
®
on/off 0 Off 1 On
WPAINA x Assignment of
recorder output 0 pH 1 mV 2 Te mperature 8 Input for
printer control
WPCASA xx Calibration buffer set
00 Knick techn. buffers 01 Mettler Toledo
technical buffers 02 Merck Titrisol buffers 03 Techn. buffers to
DIN 19 276 04 Standard buffers to
DIN 19 266 05 Merck Titrisol buffers 06 Merck buffers
ready to use 07 Ciba (94) 08 customer specific 09 customer specific 10
Mettler Toledo (USA)
Page 46
765 Laboratory pH Meter
40 Operation
Only for meters with Option 346:
Nominal electrode zero and slope are set as fol­lows:
Control commands
Reading device description
Command Parameter Unit Description
WPCA0NZ Num. par. [pH] Nominal
zero point
WPCA0NS Num. par. [V/pH] Nominal
slope
WPCA0U Num. par. [V]
V iso isothermal potential
Command Description
WCIU Initialize device WCTEA Perform Fullcheck WCRTT [hhmm] Set time WCRTD [ddmmyy] Set date WCOM00 Switch to measuring mode WCCASTI Set first calibration
(only for Option 347) WCCAA1 Start calibration with 1st buffer WCCAA2 Start calibration with 2nd buffer WCDISLA0 Left display, indicate pH WCDISLA1 Left display, indicate mV WCDISLA2 Left display, indicate temperature WCDISLATRT Left display, indicate time WCDISRA0 Right display, indicate pH WCDISRA1 Right display, indicate mV WCDISRA2 Right display, indicate
temperature WCDISRATRT
Right display, indicate time
Com­mand
Response Description
RDMF KNICK Manufacturer RDUN 765 Model name RDUS xxxxxx Serial number RDUV xx;xx Software/hardware version RDUP xxx;xxx;xxx
Options
Page 47
Operation 41
3 Troubleshooting
Error messages
Range exceeded If a measured value is out of range, an error mes-
sage is displayed instead of the value measured.
The measured pH value is
< –2 or > +16.
Possible causes:
• Electrode defective
• Too little electrolyte in electrode
• Electrode not connected
• Break in electrode cable
• Incorrect electrode connected, e.g. ORP electrode
The measured electrode potential is
< –1999 mV or > +1999 mV.
Possible causes:
• Electrode defective
• Electrode not connected
• Break in electrode cable
The measured temperature is
< –50 °C or > +150 °C
Possible causes:
• Temperature probe defective
Page 48
765 Laboratory pH Meter
42 Operation
Calibration error messages When errors occur during calibration or when the
determined electrode data are out of range, an error message will be displayed.
The meter cannot identify the buffer solution in use. This message is only momentarily displayed during calibration.
Possible causes:
• Buffer solution does not belong to the selected buffer set
• Electrode defective
• Electrode not connected
• Break in electrode cable
• Wrong buffer temperature selected (for manual temperature compensation)
The meter identified two identical buffer solutions. This message is only momentarily displayed during calibration.
Possible causes:
• Same or similar buffer solution used for second calibration step
• Buffer solutions contaminated
• Electrode defective
• Electrode not connected
• Break in electrode cable
Calibration was canceled after approx. 2 minutes, because the electrode drift was too large. This mes­sage is only momentarily displayed during calibra­tion.
Possible causes:
• Electrode defective
• No electrolyte in the electrode
• Electrode cable not sufficiently shielded or defective
• Strong electric fields influence the measurement
• Major temperature fluctuation of the buffer solution
• No buffer solution or extremely diluted
Page 49
Operation 43
The electrode zero determined during calibration is out of range. Electrode zero is < pH 6 or > pH 8
(for Option 346: at ± 1 pH unit from nominal zero)
This message is displayed in measuring mode after a calibration. It can only be reset by a new calibra­tion.
Possible causes:
• Electrode "worn out"
• Buffer solutions contaminated
• Temperature probe not immersed in buffer solution (for automatic temperature compensation)
• Wrong buffer temperature selected (for manual temperature compensation)
• Used electrode has different nominal zero point
The electrode slope determined during calibration is out of range. The electrode slope is < 47.0 mV/pH or > 61.0 mV/pH (for Option 346: at ± 7 mV/pH from nominal slope). This message is displayed in mea­suring mode after a calibration. It can only be reset by a new calibration.
Possible causes:
• Electrode ”worn out”
• Buffer solutions contaminated
• Temperature probe not immersed in buffer solution (for automatic temperature compensation)
• Wrong buffer temperature selected (for manual temperature compensation)
• Used electrode has different nominal slope
Page 50
765 Laboratory pH Meter
44 Operation
Interface error message When errors occur during transmission via inter-
face, an error message is displayed.
The meter has received an invalid interface com­mand.
Possible causes:
• Syntax error in interface command
• Too many characters in one string
• No valid message terminator
• Wrong transmission rate (baud rate) selected
• Wrong data word length or parity selected
• Wrong transmission protocol (handshake) selected
• Interference during transmission
System error message When a system error is found during the self test,
an error message is displayed.
Error in the configuration or calibration data or in the factory settings
Possible causes:
• Clock defective
• EPROM, EEPROM or RAM defective
• Error in factory settings
This error message normally should not occur, as the data are protected from loss by multiple safety functions. Should this error message nevertheless occur, there is no remedy. The meter must be reca­librated at the factory.
Opening the meter exposes live parts. Therefore, it shall not be opened. If a repair should be required, return the meter to our factory.
Page 51
Appendix 45
Appendix
Product line
Ref. No.
Device
pH meter with power cord, without electrode
765
Accessories
Attachable stand, for ZU 6955 immersion stirrer and three sensors, directly connected to pH meter
ZU 6954
Immersion stirrer, powered via ZU 6956 power pack
ZU 6955
Power pack for immersion stirrer ZU 6956
Lab Printer ZU 0244
Interface cable for connecting the Model 765 to a printer (ZU 0244)
ZU 0245
Interface cable for connecting the Model 765 to a computer (special EMC cable)
ZU 0152
Sensors and buffer solutions
Combination pH/Pt 1000 electrode (glass, with integrated Pt 1000 temp probe 0 to 80 °C, pH 0 to 14)
SE 100
Combination pH electrode (glass, without temperature probe, 0 to 100 °C, pH 0 to 14)
SE 103
Pt 1000 temperature probe, for fast response temperature measurements (stainless steel, -10 to +100 °C)
ZU 6959
Calibration buffer set with Knick technical buffer solutions (250 ml each of pH 4, pH 7, pH 9 and KCl solution)
ZU 0261
Calibration buffer set with standard buffer solutions to DIN 19 266 and NIST (NBS) (250 ml each of pH 4, pH 7 and KCl solution)
ZU 6941
Options
Power supply 115 V AC 363
Nominal electrode zero point and slope user-defined
346
Page 52
765 Laboratory pH Meter
46 Appendix
Specifications
Ranges pH: –2.00 to +16.00
mV: –1999 to +1999 °C: –50.0 to +150.0
Display Alphanumeric, 2 x 4 digits,
14-segment LED, Character height 13 mm Parameters: pH/mV/°C/time 3 status indicators for electrode con­dition
Measuring cycle Approx. 1/sec
Accuracy
*
* ± 1 count
pH: < 0.01
mV: < 0.1 % ± 0.3 mV
°C: < 0.3 K Input DIN 19 262 Input resistance
> 1 * 10
12
Ohms
Input current
< 1 * 10
–12
A (20°C)
**
** 45°C: factor 10
Temperature coefficient
< 0.1 count/K
Electrode standardization
Calimatic
®
automatic calibration and buffer recognition (German patent 29 37 227) For buffer tables, see Apppendix Calibration ranges:
Zero: pH 6 to 8 Slope: 47 to 61 mV/pH
(25°C)
(Option 346)
Nominal zero / slope / V iso
***
*** User-defined
Zero: pH 0 to 14 Slope: 25 to 61 mV/pH V iso: –500 to +500 mV
Page 53
Appendix 47
Electrode monitoring
Sensoface
®
: Evaluation of zero point, slope, response time and impedance of electrode, and of calibration interval. Electrode condition displayed as good/average/poor, can be disabled
Knick Drycheck
®
: Monitors electrode impedance for dry-out, glass breakage and defec­tive connections, also in standby mode Calibration timer monitors a preset calibration interval
Knick Fullcheck
®
device self-test
Test of complete measurement elec­tronics including recorder output, segment and keypad test in Diag­nostics mode, automatic short check at power-on
Records Parameter record
Calibration record Diagnostics record Records for QM documentation to
ISO 9000 and GLP
*
, retrievable via interface (printer) or display
* Good Laboratory Practice
Displaymatic
®
Digit suppression according to signal change, can be disabled
Te mperature compensation
Pt 100 / Pt 1000, automatic selection
Manual: –50.0 to +150 °C
Dead stop current -10 µA
Tr ueline
®
recorder output
Galvanically isolated, pH compensated
mV: 1 mV/mV pH: 100 mV/pH °C: 10 mV/°C Isolation voltage: 40 V DC,
20 V AC
Page 54
765 Laboratory pH Meter
48 Appendix
Remote interface RS 232 without control lines, galvani-
cally isolated, user defined as printer or computer interface
Baud rate: 600/1200/2400/4800/
9600
*
* User-defined
Data formats: 7 bits even/odd
*
8 bits without parity * Protocol: None, XON/XOFF * Stop bits: 1 Isolation
voltage:
40 V DC, 20 V AC
Printer control for standard printer with serial port,
printing at keystroke or via print interval timer 0.1 to 999.9 min
Clock Clock with date, self-contained Calibration data
storage
Automatic storage of calibration data, self-contained
Data retention Parameters,
statistics and factory set­tings: >10 years
(EEPROM) Clock: Reserve power>1year
(battery-backed)
Electrode statistics
Storage of zero and slope of first cali­bration, as well as data of last three calibrations with time and date, self­contained, output via interface to printer or computer.
EMC 2004/108/EC
Emitted interference: Class B Immunity to interference: Industry
Standards: DIN EN 61326 -1 (VDE 0843 Part 20-1): 2006-10 DIN EN 61326-2-3 (VDE 0843 Part 20-2-3): 2007-05
Low-voltage directive
2006/95/EC Standards: EN 61010-1: 2001
Operating temp 0 to +45 °C Storage and
transport temp
–20 to +70 °C
Page 55
Appendix 49
Buffer tables
Buffer set 00 Knick technical buffers
Power supply 230 V –15 % +10 %, 48 to 62 Hz,
< 10 VA, Class II Option 363: 115 V AC
Enclosure Glass reinforced polyamide 12, stain-
less steel cover, IP 54 protection, ready for connection of ZU 6954 attachment
stand Dimensions 244 x 95 x 255 mm (w x h x d) Weight
Approx. 2 kg
°C pH
0 2.03 4.01 7.12 9.52 5 2.02 4.01 7.09 9.45 10 2.01 4.00 7.06 9.38 15 2.00 4.00 7.04 9.32 20 2.00 4.00 7.02 9.26
25 2.00 4.01 7.00 9.21
30 1.99 4.01 6.99 9.16 35 1.99 4.02 6.98 9.11 40 1.98 4.03 6.97 9.06 45 1.98 4.04 6.97 9.03 50 1.98 4.06 6.97 8.99 55 1.98 4.08 6.98 8.96 60 1.98 4.10 6.98 8.93 65 1.99 4.13 6.99 8.90 70 1.99 4.16 7.00 8.88 75 2.00 4.19 7.02 8.85 80 2.00 4.22 7.04 8.83 85 2.00 4.26 7.06 8.81 90 2.00 4.30 7.09 8.79 95 2.00 4.35 7.12 8.77
Page 56
765 Laboratory pH Meter
50 Appendix
Buffer set 01 Mettler-Toledo technical buffers
(correspond to Ingold technical buffers)
Buffer set 02 Merck Titrisol buffers and ready-to-use buffer
solutions, Riedel Fixanal buffers and ready-to-use buffer solutions
°C pH
0 2.03 4.01 7.12 9.52 5 2.02 4.01 7.09 9.45 10 2.01 4.00 7.06 9.38 15 2.00 4.00 7.04 9.32 20 2.00 4.00 7.02 9.26
25 2.00 4.01 7.00 9.21
30 1.99 4.01 6.99 9.16 35 1.99 4.02 6.98 9.11 40 1.98 4.03 6.97 9.06 45 1.98 4.04 6.97 9.03 50 1.98 4.06 6.97 8.99 55 1.98 4.08 6.98 8.96 60 1.98 4.10 6.98 8.93 65 1.99 4.13 6.99 8.90 70 1.99 4.16 7.00 8.88 75 2.00 4.19 7.02 8.85 80 2.00 4.22 7.04 8.83 85 2.00 4.26 7.06 8.81 90 2.00 4.30 7.09 8.79 95 2.00 4.35 7.12 8.77
°C pH
0 2.01 4.05 7.13 9.24 12.58 5 2.01 4.04 7.07 9.16 12.41 10 2.01 4.02 7.05 9.11 12.26 15 2.00 4.01 7.02 9.05 12.10
20 2.00 4.00 7.00 9.00 12.00
25 2.00 4.01 6.98 8.95 11.88 30 2.00 4.01 6.98 8.91 11.72 35 2.00 4.01 6.96 8.88 11.67 40 2.00 4.01 6.95 8.85 11.54 45 2.00 4.01 6.95 8.82 11.44 50 2.00 4.00 6.95 8.79 11.33 55 2.00 4.00 6.95 8.76 11.19 60 2.00 4.00 6.96 8.73 11.04 65 2.00 4.00 6.96 8.72 10.97 70 2.01 4.00 6.96 8.70 10.90 75 2.01 4.00 6.96 8.68 10.80 80 2.01 4.00 6.97 8.66 10.70 85 2.01 4.00 6.98 8.65 10.59 90 2.01 4.00 7.00 8.64 10.48 95 2.01 4.00 7.02 8.64 10.37
Page 57
Appendix 51
Buffer set 03 Technical buffer solutions to DIN 19 267
* Extrapolated
Buffer set 04 Standard buffer solutions to DIN 19 266 and NIST
(NBS)
* Extrapolated
°C pH
0 1.08 3.14* 4.67 6.89 9.48 13.95* 5 1.08 3.12* 4.67 6.87 9.43 13.63* 10 1.09 3.10 4.66 6.84 9.37 13.37 15 1.09 3.08 4.66 6.82 9.32 13.16 20 1.09 3.07 4.65 6.80 9.27 12.96
25 1.09 3.06 4.65 6.79 9.23 12.75
30 1.10 3.05 4.65 6.78 9.18 12.61 35 1.10 3.05 4.65 6.77 9.13 12.45 40 1.10 3.04 4.66 6.76 9.09 12.29 45 1.10 3.04 4.67 6.76 9.04 12.09 50 1.11 3.04 4.68 6.76 9.00 11.98 55 1.11 3.04 4.69 6.76 8.96 11.79 60 1.11 3.04 4.70 6.76 8.92 11.69 65 1.11 3.04 4.71 6.76 8.90 11.56 70 1.11 3.04 4.72 6.76 8.88 11.43 75 1.11 3.04 4.73 6.77 8.86 11.31 80 1.12 3.05 4.75 6.78 8.85 11.19 85 1.12 3.06 4.77 6.79 8.83 11.09 90 1.13 3.07 4.79 6.80 8.82 10.99 95 1.13* 3.08* 4.82* 6.81* 8.81* 10.89*
°C pH
0 1.666 4.010 6.948 9.464 13.423 5 1.668 4.004 6.951 9.395 13.207 10 1.670 4.000 6.923 9.332 13.003 15 1.672 3.999 6.900 9.276 12.810 20 1.675 4.001 6.881 9.225 12.627 25 1.679 4.006 6.865 9.180 12.454 30 1.683 4.012 6.853 9.139 12.289 35 1.688 4.021 6.844 9.102 12.133 40 1.694 4.031 6.838 9.068 11.984 45 1.700 4.043 6.834 9.038 11.841 50 1.707 4.057 6.833 9.011 11.705 55 1.715 4.071 6.834 8.985 11.574 60 1.723 4.087 6.836 8.962 11.449 65 1.733 4.109 6.841 8.942 11.330* 70 1.743 4.126 6.845 8.921 11.210* 75 1.755 4.145 6.852 8.903 11.100* 80 1.766 4.164 6.859 8.885 10.990* 85 1.779 4.185 6.868 8.868 10.890* 90 1.792 4.205 6.877 8.850 10.790* 95 1.806 4.227 6.886 8.833 10.690*
Page 58
765 Laboratory pH Meter
52 Appendix
Buffer set 05 Merck Titrisol buffers and ready-to-use buffer
solutions, Riedel Fixanal buffers and ready-to-use buffer solutions
Buffer set 06 Merck ready-to-use buffer solutions
* Values to DIN 19 267 (not defined by Merck)
°C pH
0 0.96 3.05 6.04 8.15 10.26 13.80 5 0.99 3.05 6.02 8.10 10.17 13.59 10 0.99 3.03 6.01 8.07 10.11 13.37 15 0.99 3.01 6.00 8.04 10.05 13.18
20 1.00 3.00 6.00 8.00 10.00 13.00
25 1.01 3.00 6.02 7.96 9.94 12.83 30 1.01 3.00 6.03 7.94 9.89 12.67 35 1.01 3.00 6.03 7.92 9.84 12.59 40 1.01 2.98 6.04 7.90 9.82 12.41 45 1.01 2.98 6.05 7.88 9.78 12.28 50 1.01 2.97 6.06 7.85 9.74 12.15 55 1.02 2.97 6.08 7.84 9.71 11.95 60 1.02 2.97 6.10 7.83 9.67 11.75 65 1.02 2.97 6.11 7.82 9.65 11.68 70 1.02 2.97 6.12 7.80 9.62 11.61 75 1.02 2.97 6.14 7.79 9.59 11.50 80 1.02 2.97 6.17 7.78 9.55 11.39 85 1.02 2.97 6.20 7.77 9.52 11.27 90 1.02 2.96 6.24 7.75 9.49 11.15 95 1.02 2.96 6.28 7.74 9.46 11.03
°C pH
0 4.68 6.98 9.46 5 4.68 6.95 9.40 10 4.67 6.92 9.33 15 4.67 6.90 9.28
20 4.66 6.88 9.22
25 4.66 6.86 9.18 30 4.66 6.86 9.14 35 4.66 6.85 9.10 40 4.67 6.84 9.07 45 4.68 6.84 9.04 50 4.68 6.84 9.01 55 4.69 6.84 8.99 60 4.70 6.84 8.96 65 4.71* 6.84 8.95 70 4.72* 6.84 8.93 75 4.74* 6.85 8.91 80 4.75* 6.86 8.89 85 4.77* 6.87 8.87 90 4.79* 6.88 8.85 95 4.81* 6.89 8.83
Page 59
Appendix 53
Buffer set 07 Ciba (94)
Nominal value: 2.06, 4.00, 7.00, 10.00
* Extrapolated
5
Interpolated
Buffer set 10 Mettler-Toledo (USA)
* Extrapolated
°C pH
0 2.04 4.00 7.10 10.30 5 2.09 4.02 7.08 10.21 10 2.07 4.00 7.05 10.14 15 2.08 4.00 7.02 10.06 20 2.09 4.01 6.98 9.99 25 2.08 4.02 6.98 9.95 30 2.06 4.00 6.96 9.89 35 2.06 4.01 6.95 9.85 40 2.07 4.02 6.94 9.81 45 2.06 4.03 6.93 9.77 50 2.06 4.04 6.93 9.73 55 2.05 4.05 6.91 9.68 60 2.08 4.10 6.93 9.66 65 2.07
5
4.10
5
6.92
5
9.61
5
70 2.07 4.11 6.92 9.57 75 2.04
5
4.13
5
6.92
5
9.54
5
80 2.02 4.15 6.93 9.52 85 2.03 4.17
5
6.95 9.47
5
90 2.04 4.20 6.97 9.43 95 2.05* 4.225* 6.99* 9.385*
°C pH
0 4.00 7.12 10.32 5 4.00 7.09 10.25 10 4.00 7.06 10.18 15 4.00 7.04 10.12 20 4.00 7.02 10.06
25 4.00 7.00 10.01
30 4.01 6.99 9.97 35 4.02 6.98 9.93 40 4.03 6.98 9.89 45 4.04 6.97 9.86 50 4.06 6.97 9.83 55 4.06* 6.97* 9.83* 60 4.06* 6.97* 9.83* 65 4.06* 6.97* 9.83* 70 4.06* 6.97* 9.83* 75 4.06* 6.97* 9.83* 80 4.06* 6.97* 9.83* 85 4.06* 6.97* 9.83* 90 4.06* 6.97* 9.83* 95 4.06* 6.97* 9.83*
Page 60
765 Laboratory pH Meter
54 Appendix
Page 61
Glossary 55
Glossary
Buffer set Contains selected buffer solutions which can be
used for automatic calibration with the Knick Cali­matic
®
. The buffer set must be selected prior to the
first calibration.
Buffer solution Solution with an exactly defined pH value for cali-
brating a pH meter.
cal Key for activating the Calibration level.
Cal timer Counts the time passed since the last calibration.
Calibration Adjustment of the pH meter to the current electrode
characteristics. Zero and slope are adjusted.
Calibration buffer set See buffer set.
Calibration level On the Calibration level the meter is adjusted for
the connected electrode (calibration). You can con­duct either a one or a two-point calibration.
Calibration record Printout of all relevant data of the last calibration for
documentation to GLP. Start printout by pressing
and .
Calimatic
®
Automatic buffer recognition. Prior to first calibra­tion, the used buffer set must be activated once on the Parameter level. The patented Calimatic
®
then automatically recognizes the buffer solution used during calibration.
Combination electrode Combination of glass and reference electrode in
one body.
diag Key for activating the Diagnostics level.
Diagnostics level Display of criteria for Sensoface
®
display, all rele­vant electrode data, electrode statistics and activa­tion of Fullcheck
®
device self test.
Page 62
765 Laboratory pH Meter
56 Glossary
Diagnostics record Printout of criteria for Sensoface
®
display, all rele­vant electrode data, electrode statistics and results of Fullcheck
®
device self test for documentation to
GLP. Start printout by pressing and .
Electrode slope Specified in mV/pH. It is different for each electrode
and changes with age and wear.
Electrode statistics The electrode statistics provide the electrode data
of the last three calibrations and the first calibration.
Electrode zero point Voltage output by a pH electrode at pH 7 (for devi-
ating nominal zero point at correspondingly differ­ent pH value). It is different for each electrode and changes with age and wear.
enter Key for confirming entries.
First calibration During first calibration, the electrode data are
stored as reference values for electrode statistics.
GLP Good Laboratory Practice: Guidelines for perfor-
mance and documentation of measurements in the laboratory.
Isothermal potential The isothermal potential is the potential at which
the calibration curves for different temperatures in­tersect. Ideally, this is at 0 mV.
Knick Drycheck
®
Knick Drycheck® monitors the electrodes and alerts for dry out, glass breakage and defective connec­tions.
Knick Fullcheck
®
Device self test, checks complete measuring cir­cuitry, signal processing, memories, display, and keypad.
meas This key allows return to measuring mode from all
other levels.
Measuring mode When no function level is activated, the meter is in
measuring mode. The two displays indicate the re­spectively assigned variable.
Page 63
Glossary 57
Nominal electrode zero point
Nominal value of electrode zero point. For commer­cial electrodes, the nominal electrode zero point is at pH 7.
Special electrodes can have a different zero point. If you want to use such an electrode, your meter must be equipped with Option 346.
Nominal electrode slope Nominal value of electrode slope. For commercial
eletrodes, the nominal slope is 59.2 mV/pH at 25 °C.
Special electrodes, such as Pfaudler or Antimony probes, can have a different slope. If you want to use such an electrode, your meter must be equipped with Option 346.
One-point calibration Calibration that only takes account of the electrode
zero point. The old slope value is retained. Only one buffer solution is required for a one-point cali­bration.
par Key for activating the Parameter level.
Parameter level The Parameter level is divided into two submenus.
VIEW menu and EDIT menu. The VIEW menu al­lows to display all parameters without editing them. The EDIT menu allows to view and edit all parame­ters.
Parameter record Printout of all stored parameter settings for docu-
mentation to GLP. Start printout by pressing and .
pH electrode A pH electrode system consists of a glass and a ref-
erence electrode. If they are combined in one body, they are referred to as combination electrode.
Print interval timer The print interval timer allows to preset an interval
for printout of currently measured values with time and date.
Response time Time from the start of a calibration step to the sta-
bilization of the electrode potential.
Page 64
765 Laboratory pH Meter
58 Glossary
Sensoface
®
Automatic electrode monitoring. The Sensoface® display provides information on the electrode con­dition. It evaluates zero point, slope, response time, impedance and dry-out of the electrode and the cal timer count.
Slope See Electrode slope.
Trueline
®
The Trueline® recorder output provides a pH-cali­brated analog output signal, without disturbing volt­age steps.
Two-point calibration Calibration that takes account of zero point and
slope of the electrode. Two buffer solutions are re­quired for two-point calibration.
Zero point See electrode zero point.
Page 65
Index
B
Buffer set 14 Buffer tables 49
C
Calibration error messages 42 Calibration level 18 Calibration record 20 Calibration timer
Setting 15 Commissioning 6 Connection
Double platinum electrode 7
Electrode 7
Mains 6
Polarization current 7
D
Device self-test 27 Diagnostics
Diagnostics level 21
Electrode Data menu 21
Knick Fullcheck menu 21
Sensoface menu 21, 22 Diagnostics record 29 Displaymatic 13 Double platinum electrode 7
E
EDIT menu 10 Electrode connection 6 Electrode Data menu 25 Electrode statistics 25 Error messages 41
List of 34
F
First calibration
Activation 15
I
Interface cable 32
K
Keypad 3
L
Lab printer, ZU 0244
Standard settings 32
M
Mains supply 6 Measuring mode 9 Menu structure 5
N
Nominal electrode slope (setting for Option 346) 13 Nominal electrode zero (setting for Option 346) 13
O
One-point calibration 19 Operation
Meter design 2
Option 346 13
P
Package contents 1 Parameter record 12 Parameter setting
Activation 10 EDIT menu 10 Example 11
VIEW menu 10 Polarization current 7 Print interval timer 16 Printer, ZU 0244
Standard settings 32 Printing
Calibration record 20
Diagnostics record 29
Measured values 9
Parameter record 12 Product line 45
R
Recorder output 30
Setting 15 Remote interface
Cable 32
Commands 32
Page 66
Parameters 31 Pin assignment 31 Printer 32
RS 232 interface 31
S
Safety information III Sensoface 5 Sensoface menu 21, 22 Sensor monitoring 5 Setting
Remote interface 16
Short check (at power-on) 8 Short description 1 Specifications 46 Standby mode 8
T
Trueline recorder output 30 Two-point calibration 19
V
VIEW menu 10