KNF NF 25 Series, NFB 25 Series, NFB 25 DCB-B, NF 25 DC-M, NFB 25 DCB-4B Operating And Installation Instructions

...
DIAPHRAGM LIQUID PUMP NF 25
, NFB 25
Read and observe these Operating and Installation Instructions!
An additional letter prefixing the NF model code is a country with no
NF 25 KP DC-M 12V
DC-M / DCB / DCB-4 / DCB-B / DCB-4B
25
NF / NFB
Supply voltage [Ch. 4]
- / PMLxxxx / PLxxxx [Ch. 1]
KP / KT / TT [Ch. 4]
-specific designation,
technical relevance.
KNF Flodos AG Wassermatte 2 6210 Sursee, Switzerland
Tel +41 (0)41 925 00 25 Fax +41 (0)41 925 00 35
www.knf-flodos.ch info@knf-flodos.ch
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions Keep for future reference!
Contents Page
1.
About this document ................................................................. 2
2. Use ........................................................................................... 3
3. Safety ....................................................................................... 4
4. Technical data .......................................................................... 6
5. Assembly and function ............................................................. 9
6. Installation and connection ..................................................... 10
7. Operation ................................................................................ 15
8. Servicing ................................................................................. 19
9. Troubleshooting ...................................................................... 22
10. Accessories ............................................................................ 24
11. Decontamination declaration .................................................. 25
Diaphragm liquid pump NF 25, NFB 25 About this document
This symbol indicates a potential danger.
It also indicates the possible consequences of failure to observe the warning. The signal word (e.g. "Warn­ing") indicates the level of danger.
Here you will see actions for avoiding the danger
Signal word
Meaning
Consequences if not observed
DANGER
warns of immedi-
Consequences are death or
WARNING
warns of potential
Death or serious injury and/or
CAUTION
warns of a poten­situation
Minor injury or damage to prop-
This symbol indicates important information.
Project pumps

1. About this document

1.1. Use of the Operating and Installation Instruc­tions
The Operating and Installation Instructions are part of the pump. Forward the Operating and Installation Instructions to any
subsequent owners of the pump.
Customer-specific project pumps (pump models which begin with "PL" or "PML") may differ from the Operating and Installation Instructions.
In the case of project pumps, take note of any additionally
agreed specifications.

1.2. Symbols and markings

Warning
WARNING
and potential consequences.
Danger levels
ate danger
danger
tially dangerous
Tab. 1
Other information and symbols
This indicates a required activity (step).
1. This indicates the first step of a required activity. Additional
consecutively numbered steps follow.
serious injury and/or serious property damage.
serious damage to property are possible.
erty are possible.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 2
Diaphragm liquid pump NF 25, NFB 25 Use
Please contact your local KNF partner for special designs that are not included in the technical specification (www.knf.com).
Operating parameters and
Requirements for
The pumps must not be operated in an explosive atmosphere.

2. Use

2.1. Intended use

The pumps are intended for transferring and metering liquids and gases.
Owner's responsibility
Only install and operate the pumps under the operating parameters
conditions
transferred medium
and conditions described in Chapter 4, Technical data. Only completely installed pumps may be taken into service. Before transferring or metering a medium, check whether the
medium can be transferred danger-free in the specific application case.
Before using a medium, check the compatibility of the materials of the pump head, pump housing, diaphragm and valves with the medium.
The temperature of the medium must lie within the permissible temperature range (see Chapter 4).
The transferred medium should not contain particles as these can prevent the pump from working correctly. If this cannot be guaran­teed, a filter < 50 µm with sufficiently large filter area must be used upstream of the pump.
Filters may be ordered as accessories, see Spare Parts and
Accessories (Chapter 10).

2.2. Improper use

DANGER
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 3
Diaphragm liquid pump NF 25, NFB 25 Safety
Observe the safety precautions in Chapters
6. Installation and connection and 7. Operation.
Personnel
Working in a
Handling dangerous media
Notes
Environmental protection
Disposal

3. Safety

safety-conscious manner
The pumps are built according to the generally recognised rules of technology and in accordance with the pertinent occupational safety and accident prevention regulations. Nevertheless, dangers may occur during their use which may lead to injuries to the user or others, or to damage to the pump or other property. Only use the pumps when they are in a good technical and proper working order, in accordance with their intended use, observing the safety advice within the Operating and Installation Instructions, at all times.
Make sure that only trained and instructed personnel or specially trained personnel work on the pumps. This especially applies to assembly, connection and servicing work.
Make sure that all personnel have read and understood the Oper­ating and Installation Instructions, and in particular the "Safety" chapter.
Always ensure adherence to all pertinent accident prevention and safety regulations when working on and operating the pump.
When transferring dangerous media, observe the safety regula­tions for handling such media.
Always ensure adherence to all information stickers on the pumps, such as flow direction arrows and type plates, and keep stickers in legible condition.
All replacement parts should be properly stored and disposed of in accordance with the applicabl e envir onmental protection regula­tions. Observe the respective nat ional and int er nat io n al regul a­tions. This especially applies to parts contaminated with toxic substances.
Dispose of all packaging in an environmentally-appropriate manner. The packaging materials are recyclable.
Dispose of end-of-life equipment in an environmentally friendly manner. Use appropriate waste collection systems for the disposal of end-of-life equipment. Used pumps contain valuable recyclable materials.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 4
Diaphragm liquid pump NF 25, NFB 25 Safety
EU directives/standards
Customer service and repairs
The pumps comply with the fundamental requirements of Directive 2011/65/EU (RoHS2).
The pumps comply with the safety requirements regarding elec­tromagnetic compatibility in Directive 2004/ 1 08/ EC.
For the purposes of the Machinery Directive 2006/42/EC, pumps are “partly completed machinery", and are therefore to be regarded as not ready for use. Partly completed machinery may not be commissioned until such time as it has been determined that the machine in which the partly com pleted m achinery is to be assem­bled conforms to the provisions of the Machinery Directive 2006/42/EC. The essential requirements of Annex I of Directive 2006/42/EC (general principles) are applied and observed.
The following harmonised standards are met:
NF 25 DC-M
EN 55022  EN 55011
NF 25 DCB / DCB-4
EN 61000-6-2  EN 61000-6-3 (incl. EN 55022 / EN 55011)
NFB 25 DCB-B / DCB-4B
IEC/EN 61000-6-2  IEC/EN 61000-6-3
All repairs to the pump(s) must be carried out by the relevant KNF customer service team.
Only use genuine parts from KNF for servicing work.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 5
Diaphragm liquid pump NF 25, NFB 25 Technical data
Assembly
Material1)
Pump head*
PP
Valve plate / seals
EPDM
Diaphragm
EPDM
Resonating diaphragm
EPDM
Assembly
Material1)
Pump head*
PP
Valve plate / seals
FFKM
Diaphragm
PTFE-coated
Resonating diaphragm
FFKM
Assembly
Material1)
Pump head*
PVDF
Valve plate / seals
FFKM
Diaphragm
PTFE-coated
Resonating diaphragm
FFKM
Parameter
Value
Flow rate NF 25 DC-M and DCB [ml/min]
1), 2)
250
Flow rate NF 25 DCB-4 [ml/min]
1), 2)
25-250
Flow rate NFB 25 DCB-4B [ml/min]
1), 2)
2x 50-300
Flow rate NFB 25 DCB-B [ml/min]
1), 2)
2x 300
Permissible pressure [bar g]
1
Suction head [mWG]
3
Parameter
Value
Recommended hose size ID [mm]
4

4. Technical data

Pump materials The pump type KP stands for:
Tab. 2
1)
according to DIN ISO 1629 and 1043.1
The pump type KT stands for:
Tab. 3
1)
according to DIN ISO 1629 and 1043.1
The pump type TT stands for:
Tab. 4 * The pump head (Fig. 1) comprises a connecting plate and an intermedi-
ate plate.
1)
according to DIN ISO 1629 and 1043.1
Hydraulic ratings
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 6
Tab. 5
1)
Measured with water and using KNF standardised test equipement at
atmospheric pressure, KP pump head material
2)
Flow rates may vary from the values shown, depending on fluid viscosi­ty, ambient temperature, pump head material and the hoses / hose con­nectors used.
Hydraulic connections
Tab. 6
Diaphragm liquid pump NF 25, NFB 25 Technical data
Motor voltage
12V
24V
Power consumption [W]
2.9
2.9
Max. I load [A]
0.24
0.12
Max. permissible current Lead size [-]
AWG22
Protection class [-]
IP30
Weight
1)
[g]
64
Motor option
DCB-4
DCB 12V
DCB 24V
Supply voltage [V]
10…26.4
12
24
Power consump­tion [W]
I load max. [A]
0.11…0.25
0.23
0.12
Lead size [-]
AWG28
Protection class
IP40
Weight
1)
[g]
60
In the case of type NF25 DCB-4, the supply voltage is not designed for operation with pulse (PWM).
However, a PWM signal (frequency 1 control the control
Electrostatic sensitive components (ESD)
Specifications NF 25 DC-M
consumption [A]
Tab. 7
Specifications NF 25 DCB / DCB-4
2.9 2.8 2.9
Tab. 8
voltage.
0.31 0.14
-width modulated voltages
-20kHz) can be used to
1) The weight may differ slightly from the stated value, depending on the version
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 7
Diaphragm liquid pump NF 25, NFB 25 Technical data
Motor option
DCB-4B
DCB-B
DCB-B
Supply volt­age [V]
Power con­sumption [W]
I load max. [A]
0.22…0.54
0.54
0.27
Protection class
Weight
1)
[g]
210
In the case of type NFB25 DCB-4B, the supply voltage is not designed for operation with pulse (PWM).
However, a PWM signal (frequency 1 control the control voltage.
Electrostatic sensitive components (ESD)
Parameter
Value
Permissible ambient temperature range [C]
+ 5 to + 40
Permissible media temperature [C]
+ 5 to + 80
Permissible kinematic viscosity of medium
150
Specifications NFB 25 DCB-B / DCB-4B
10…26.4 12 24
5.4 6.5 6.5
IP50
Tab. 9
-width modulated voltages
-20kHz) can be used to
1) The weight may differ slightly from the stated value, depending on the version.
Other parameters
[cSt]
Tab. 10
.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 8
Diaphragm liquid pump NF 25, NFB 25 Assembly and function
1 Outlet 2 3 4 5 6 7
1 Outlet valve 2 3 4 5 6 7
(double-headed pump)

5. Assembly and function

Assembly
Inlet Connecting plate Drive Connecting leads Head plate Intermediate plate
Inlet valve Working chamber Diaphragm Eccentric Connecting rod Pump drive
Fig. 3 : NFB 25
Fig. 1: Micro diaphragm liquid pump NF 25
Fig. 2: Pump assembly
The diaphragm liquid pumps are based on reciprocating displace­ment pump technology. The elastic diaphragm (4) is moved up and down by the eccentric (5) and the connecting rod (6). In the down­ward stroke it aspirates the medium to be transferred via the inlet valve (2). In the upward stroke the diaphragm pushes the medium out of the pump head via the outlet valve (1). The diaphragm hermetically seals off the working chamber (3) from the pump drive (7).
NFB 25 Version (double-headed pump, see Fig. 3) This pump type has two NF25 pump heads that can be operated
individually or in parallel connection (see Chapter 0).
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 9
Diaphragm liquid pump NF 25, NFB 25 Installation and connection
Mounting dimensions

6. Installation and connection

Only install the pump under the operating parameters and condi­tions described in Chapter 4, Technical data.
Observe the safety precautions (see Chapter 3).

6.1. Installation

Before installation, store the pump at the installation location to
bring it up to ambient temperature.
Mounting dimensions (see Fig. 4 to 6)
Fig. 4: Mounting dimensions NF 25 DC-M
Fig. 5: Mounting dimensions NF 25 DCB
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 10
Diaphragm liquid pump NF 25, NFB 25 Installation and connection
Installation location
Fig. 6: Mounting dimensions NF 25 DCB-4
Fig. 6: Mounting dimensions NFB 25 DCB-B
Fig. 8: Mounting dimensions NFB 25 DCB-4B
Make sure that the installation location is dry and the pump is
protected against water in the form of rain, spray, splashes and drips.
Protect the pump against dust. Protect the pump against vibration and impact.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 11
Diaphragm liquid pump NF 25, NFB 25 Installation and connection
Installation position
Generally speaking, the pump can be installed in any position.
For maximum accuracy and rapid venting of one-headed pumps, please install as shown in the illustration (Fig. 7).
Fig. 9: Optimum installation
position for one-headed pumps:
horizontal
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 12
Diaphragm liquid pump NF 25, NFB 25 Installation and connection
Note the proper polarity. Fo
red motor cable: + black motor cable:
Arrows on the pump head indicate the flow direction.
Keep the suction line as short as possible in order to keep the priming process as brief as possible.
Connected
Hoses

6.2. Electrical connection

Only have the pump connected by an authorized specialist. Only have the pump connected when the power supply is
disconnected.
When connecting the device to a power source, the relevant
norms, directives, regulations and technical standards must be observed.
Connecting the pump
Make sure that the power supply data match the data on the
motor's type plate. The current consumption can be found on the type plate.
Connect the motor cables. For electrical data see Chapter 4.
r DC motors:
-

6.3. Hydraulic connection

components
Only connect components to the pump that are designed to
handle the hydraulic data of the pump (see Chapter 4, Technical data).
Only use hoses that are suitable for the maximum permissible
operating pressure of the pump (see Chapter 4).
Only use hoses that are sufficiently chemically resistant to the
liquids being transferred.

6.3.1. Connecting the pump

Remove the protective caps. Connect the suction and pressure lines.
If the pump is used to build up pressure, make sure that all
transition joints between hose and pump are secure in order to ensure that the hoses cannot come off.
Check that the hoses and transition joints (hose connect-
or/hose) are fitted correctly and securely.
Check that the system is leak-tight.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 13
Diaphragm liquid pump NF 25, NFB 25 Installation and connection
A double-headed pump is normally selected to reduce the pressure surges caused by the alternation between suction and exhaust in a diaphragm pump. In order to effect, in addition to the pump configuration it is important to configure the hoses correctly.
Please contact your local KNF partner for further information (www.knf.com).
Please contact your local KNF partner for further information (www.knf.com).
Series connection damages pump
If the two heads in the NFB versions are connected in series, the pump will be damaged, and may leak.
Fig. 10: Hose connection options for NFB versions

6.3.2. NFB 25 hose configuration

achieve this
For low pulsation (pressure surges) in NFB pumps:
Configure pump head hoses in parallel (see Fig. 8). Keep hoses between pump and junction as short as possible. Cause as little narrowing as possible with connection ele-
ments. Recommendation: Min. inside diameter 3mm.
Fig. 12: No series connection in the case of NFB versions
Fig. 11: NFB hose configuration for low pulsation
CAUTION
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 14
Diaphragm liquid pump NF 25, NFB 25 Operation
Risk of burning
The drive heats up.
Excessive pressure and the inherent dangers thereof can be prevented by placing a bypass line with a pressure relief valve between the pressur contact your local KNF partner for further information (www.knf.com).
Risk of physical injury and damage to the pump due to automatic start
If the pump overheats and pump operation is stopped by the thermal switch / electronics, the pumps will restart automatically as soon as they have had time to cool down.
Risk of burns when transferring hot media
Pump standstill

7. Operation

Operate the pumps only under the operating parameters and
conditions described in Chapter 4, Technical data.
Make sure that the pumps are being used properly (see Chap-
ter 2.1).
Avoid improper use of the pumps (see Chapter 2.2). Observe the safety precautions (see Chapter 3). The pumps are components that are intended to be incorpo-
rated into another machine. Before putting them into service it must be established that the machinery or systems in which they are installed meet the relevant regulations.
CAUTION
Avoid contact with the pump drive. Avoid contact with flammable materials.
e and suction side of the pump. Please
Do not touch the pump or the media transfer
system.
CAUTION
If the pump stops running, restore the system to normal at-
mospheric pressure.
For pumps with thermal switch or electronic overload protection:
WARNING
Avoid contact with flammable materials.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 15
Take steps to ensure that this cannot produce a
hazardous situation.
Diaphragm liquid pump NF 25, NFB 25 Operation
If the pump is operated with short cycles in your application, please contact your local KNF partner for further information (www.knf.com.)
Setting and regulating motor
Short cycle operation
Overpressure on the suction side causes medi­um to flow through the switched
Switching the pumps on and off
The motor speed of the pumps, and thus the flow rate, is adjusta-
speed
ble and can also be regulated to some extent. For more details, see Chapter 4, Technical data.
Duty cycle / short cycle operation
KNF pumps are designed for continuous operation. Short start and stop cycles may adversely affect the service life of
the brushed motors.
Switching on the pump
In order to ensure that the pump starts every time, make sure
that counterpressure is reduced to an acceptable level before start-up. This should also be done during operation after a brief power cut. Please contact your local KNF partner for further in­formation (www.knf.com).
Switching off the pump
KNF recommends: If transferring aggressive liquids, the pump
should be flushed thoroughly prior to switch off (see Chapter
8.2.1), as this will help to lengthen the service life of the dia­phragm.
Restore the system to normal atmospheric pressure (release
hydraulic pressure in pump).
-off pump
Take steps to ensure that this cannot produce a
CAUTION
hazardous situation.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 16
Diaphragm liquid pump NF 25, NFB 25 Operation
0.00
0.05
0.10
0.15
0.20
0.25
0.30
-4.0
-2.0
0.0
2.0
4.0
6.0 8.0
10.0
Suction height [mWg] Pressure [mWg]
Flow rate [l/min]
0.00
0.05
0.10
0.15
0.20
0.25
0.30
-4.0 -2.0 0.0 2.0 4.0 6.0 8.0 10.0
Suction height [mWg] Pressure [mWg]
Flow rate [l/min]
0.00
0.05
0.10
0.15
0.20
0.25
0.30
0 1 2 3 4 5
Control voltage Vctrl [V]
Flow rate [l/min]
0 mWg 10 mWg
Flow rate NF 25
Fig. 13: Flow rate of pump NF 25 DC-M
Fig. 14: Flow rate of pump NF 25 DCB and NF 25 DCB-4
Fig. 15: Flow rate of pump NF 25 DCB-4 depending on control voltage
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 17
Diaphragm liquid pump NF 25, NFB 25 Operation
Flow rate NFB 25
Fig. 16: Flow rate of pump NFB 25 DCB-B and NFB 25 DCB-4B (per pump head)
Fig. 17: Flow rate of pump NFB 25 DCB-4B depending on control voltage (per pump head)
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 18
Diaphragm liquid pump NF 25, NFB 25 Servicing
Component
Servicing interval
Pump
- Regular inspection for external
Pump head
- Clean if the flow rate decreases, the Diaphragm, valve
- Change as soon as pumping capac-
Health hazard due to dangerous substances in the pump
Depending on the medium transferred, caustic burns or poisoning are possible.
With NFB pumps the necessary steps are repeated for the second pump head
Qty
Tools
1
T6 Torx screwdriver
Information on procedure
Tools

8. Servicing

8.1. Servicing schedule

damage or leaks
pump does not work or no vacuum is created (Chapter 8.2)
plates and seals
Tab. 11
ity decreases, preferabl y sooner

8.2. Cleaning and dismantling

WARNING
Wear protective clothing if necessary, e.g. pro-
tective gloves.
Flush the pump with a neutral liquid and pump
empty.
With NFB pumps, make sure that both pump
heads are flushed.

8.2.1. Flushing the pump

If transferring aggressive media, KNF recommends flushing
the pump with a neutral fluid under atmospheric conditions for a few minutes before switching off, to extend the service life of valves and diaphragm.

8.2.2. Preparations for disassembly

.
Flush the pump with a suitable neutralising liquid, and make
Pump empty. Separate electrical connections. Disconnect hoses from pump head.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 19
Tab. 12
sure that no dangerous substances are left in the pump.
Diaphragm liquid pump NF 25, NFB 25 Servicing

8.2.3. Disassembling the pump head

Loosen the four head screws (1) and remove the entire
head.
Remove the resonating diaphragm (3) from the connecting
plate (4).
Remove both anchor valves (6) from the intermediate plate
(7) and the connecting plate (4).
Remove the two anchor valve seals (5) from the connecting
plate (4).
Undo the diaphragm (8) by means of the head plate (2). Re-
move the washers and make sure that no washers (9) fall in­to the pump housing.
We recommend replacing the diaphragm (8).

8.2.4. Clean or replace parts

Clean the anchor valves (6), the anchor valve seals (5) and
the diaphragm (8) with a cloth and then blow off with com­pressed air or replace.
Fig. 18
1 Head screws 2 Head plate 3 Resonating diaphragm 4 Connecting plate 5 Anchor valve seals 6 Anchor valves 7 Intermediate plate 8 Diaphragm 9 Washer
Blow off the intermediate plate (7), connecting plate (4) and
head plate (2) with compressed air.

8.2.5. Assembling the pump head

Screw diaphragm (8) with washer (9) into connecting rod.
Make sure that no washers (9) fall into the pump housing. Use the head plate (2) to carefully tighten the diaphragm. Insert the seals (5) in the connecting plate (4). Insert the anchor valves (6) in the intermediate plate (7) and
the connecting plate (4). Place the connecting plate (4) on top of the intermediate
plate (7).
A pin and hole system prevents incorrect assembly. Insert the resonating diaphragm (3) into the connecting plate
(4). Place the head plate (2) on top of the connecting plate (4). A
pin and hole system prevents incorrect assembly. Insert the four head screws (1) into the through holes of the
pump head. Place the pump head on top of the pump housing and make
sure that the flow direction of the pump head (see directional
arrow on the head plate (2)) is the same as before.
Alternately tighten the four head screws (1). Reconnect the hoses to the pump head.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 20
Diaphragm liquid pump NF 25, NFB 25 Servicing
Escaping liquid
After assembly the pump may not be leak to incorrect assembly, damaged or soiled seal faces, or other reasons.
-tight due
CAUTION
Run pump for several minutes with a harmless
liquid at maximum operating pressure.
Check that pump is leak-tight.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 21
Diaphragm liquid pump NF 25, NFB 25 Troubleshooting
Pump does not work
Possible cause
Remedy
Pump not connected to power
Connect pump to power supply Power supply is not sw itched on
Switch on power supply
Thermal switch or pump elec-
Disconnect pump from the mains
Connections or hoses are
Check connections and hoses External valve is closed or filter is clogged
Check external valves and filters
Diaphragm, anchor valves or
Replace diaphragm, anchor valves and seals
Pump is not priming
Possible cause
Remedy
Suction side of pump not con-
Connect the suction side of the pump
Liquid in the container is too low
Fill container
Hose connections are not leak-
Secure transition joints between hose and connections with
Peripheral valve is closed or
Open the peripheral valve
Pump head is filled with gas.
Reduce pressure at pressure side
Particles in the pump
Clean the pump head (see Chapter 8.2)
The pump parts are not re-
Replace the pump head with a compatible version
Incorrect interchange of pres-
Remove pressure and suction lines and reconnect correctly

9. Troubleshooting

Disconnect the pump from the power supply before working on
the pump.
Make sure that the pump is de-energised.
supply
tronics have tripped
Allow pump to cool Identify and eliminate cause of overheating/overload
blocked
seals are worn.
nected
tight
filter is clogged
The system is unable to handle the pressure on the pressure side
Remove blockage
(see Chapter 8.3)
Tab. 13
clamps or other clamping elements
Clean filter
sistant to the medium to be transferred
sure and suction line connec­tions
Tab. 14
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 22
Diaphragm liquid pump NF 25, NFB 25 Troubleshooting
Flow rate, suction head or pressure head is too low
Possible cause
Remedy
Components in the system
Modify installation, check the cross-sections of components
Hose connections are not leak-
Secure tra nsit ion joints between hose and hose connections
Particles in the pump
Clean the pump head, install suction-side filter if required Viscosity of the transferred
Contact KNF
Incorrect interchange of pres-
Remove pressure and suction lines and reconnect correctly
The pump parts are not re-
Replace the pump head with a compatible version
The pump does not achieve the performance stated in the technical data or on the data sheet.
connected to the suction and pressure sides, such as hoses, valves or filters, are causing too much resistance
tight
with clamps or clamping elements
(see Chapter 8.2)
medium is too high
sure and suction line connec­tions
sistant to the medium to be transferred
Tab. 10
Fault cannot be rectified
If you are unable to identify any of the above causes, please send the pump to KNF customer service (see address on last page).
Flush the pump to clear the pump head of any hazardous or
aggressive liquids (see Chapter 8.2.1). Dismantle the pump. Clean the pump (see Chapter 8.2.2). Send the pump, with completed decontamination declaration
(see Chapter 10), to KNF customer service stating the nature
of the transferred medium.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 23
Diaphragm liquid pump NF 25, NFB 25 Accessories
for hose
for hose
Please contact your local KNF partner for further information (www.knf.com).

10. Accessories

The following filters are available:
Type Material Mesh size
FS 60 T PVDF 70 µm 600 ml/ min UNF 1/4“- 28 165210
FS 60 X PEEK 35 µm 600 ml/ min UNF 1/4“- 28 165212
FS 25 T PVDF 70 µm 250 ml/ min
FS 25 X PEEK 35 µm 250 ml/ min
Max. flow volume (liquid)
Tab. 16
Connections Order No.
ID 3.2 / 4 mm
ID 3.2 / 4 mm
165211
165213
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 24
Diaphragm liquid pump NF 25, NFB 25 Decontamination declaration
KNF shall only undertake to repair the pump on condition that the customer provides certification of the transferred media and the cleaning of the pump (decontamination declarati

11. Decontamination declaration

on).
In order to send a product back use the decontamination
declaration, which either was delivered with the product or is
available on www.knf.com (Downloads).
Please fill in the pump type, serial number, pumped media and all other required information. Send the signed form together with the product to your KNF representative.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615 Translated from the Original Operating and Installation Instructions 25
Diaphragm liquid pump NF 25, NFB 25 For your notes
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions 26