Diaphragm liquid pump NF 25, NFB 25 About this document
This symbol indicates a potential danger.
It also indicates the possible consequences of failure
to observe the warning. The signal word (e.g. "Warning") indicates the level of danger.
Here you will see actions for avoiding the danger
Signal word
Meaning
Consequences if not observed
DANGER
warns of immedi-
Consequences are death or
WARNING
warns of potential
Death or serious injury and/or
CAUTION
warns of a potensituation
Minor injury or damage to prop-
This symbol indicates important information.
Project pumps
1. About this document
1.1. Use of the Operating and Installation Instructions
The Operating and Installation Instructions are part of the pump.
Forward the Operating and Installation Instructions to any
subsequent owners of the pump.
Customer-specific project pumps (pump models which begin with
"PL" or "PML") may differ from the Operating and Installation
Instructions.
In the case of project pumps, take note of any additionally
agreed specifications.
1.2. Symbols and markings
Warning
WARNING
and potential consequences.
Danger levels
ate danger
danger
tially dangerous
Tab. 1
Other information and symbols
This indicates a required activity (step).
1. This indicates the first step of a required activity. Additional
consecutively numbered steps follow.
serious injury and/or serious
property damage.
serious damage to property are
possible.
erty are possible.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions2
Diaphragm liquid pump NF 25, NFB 25 Use
Please contact your local KNF partner for special designs that
are not included in the technical specification (www.knf.com).
Operating parameters and
Requirements for
The pumps must not be operated in an explosive
atmosphere.
2. Use
2.1. Intended use
The pumps are intended for transferring and metering liquids and
gases.
Owner's responsibility
Only install and operate the pumps under the operating parameters
conditions
transferred medium
and conditions described in Chapter 4, Technical data.
Only completely installed pumps may be taken into service.
Before transferring or metering a medium, check whether the
medium can be transferred danger-free in the specific application
case.
Before using a medium, check the compatibility of the materials of
the pump head, pump housing, diaphragm and valves with the
medium.
The temperature of the medium must lie within the permissible
temperature range (see Chapter 4).
The transferred medium should not contain particles as these can
prevent the pump from working correctly. If this cannot be guaranteed, a filter < 50 µm with sufficiently large filter area must be used
upstream of the pump.
Filters may be ordered as accessories, see Spare Parts and
Accessories (Chapter 10).
2.2. Improper use
DANGER
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions3
Diaphragm liquid pump NF 25, NFB 25 Safety
Observe the safety precautions in Chapters
6. Installation and connection and 7. Operation.
Personnel
Working in a
Handling dangerous media
Notes
Environmental protection
Disposal
3. Safety
safety-conscious manner
The pumps are built according to the generally recognised rules of
technology and in accordance with the pertinent occupational
safety and accident prevention regulations. Nevertheless, dangers
may occur during their use which may lead to injuries to the user or
others, or to damage to the pump or other property.
Only use the pumps when they are in a good technical and proper
working order, in accordance with their intended use, observing the
safety advice within the Operating and Installation Instructions, at
all times.
Make sure that only trained and instructed personnel or specially
trained personnel work on the pumps. This especially applies to
assembly, connection and servicing work.
Make sure that all personnel have read and understood the Operating and Installation Instructions, and in particular the "Safety"
chapter.
Always ensure adherence to all pertinent accident prevention and
safety regulations when working on and operating the pump.
When transferring dangerous media, observe the safety regulations for handling such media.
Always ensure adherence to all information stickers on the pumps,
such as flow direction arrows and type plates, and keep stickers in
legible condition.
All replacement parts should be properly stored and disposed of in
accordance with the applicabl e envir onmental protection regulations. Observe the respective nat ional and int er nat io n al regul ations. This especially applies to parts contaminated with toxic
substances.
Dispose of all packaging in an environmentally-appropriate
manner. The packaging materials are recyclable.
Dispose of end-of-life equipment in an environmentally
friendly manner. Use appropriate waste collection systems
for the disposal of end-of-life equipment. Used pumps
contain valuable recyclable materials.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions4
Diaphragm liquid pump NF 25, NFB 25 Safety
EU directives/standards
Customer service and repairs
The pumps comply with the fundamental requirements of Directive
2011/65/EU (RoHS2).
The pumps comply with the safety requirements regarding electromagnetic compatibility in Directive 2004/ 1 08/ EC.
For the purposes of the Machinery Directive 2006/42/EC, pumps
are “partly completed machinery", and are therefore to be regarded
as not ready for use. Partly completed machinery may not be
commissioned until such time as it has been determined that the
machine in which the partly com pleted m achinery is to be assembled conforms to the provisions of the Machinery Directive
2006/42/EC. The essential requirements of Annex I of Directive
2006/42/EC (general principles) are applied and observed.
The following harmonised standards are met:
NF 25 DC-M
EN 55022
EN 55011
NF 25 DCB / DCB-4
EN 61000-6-2
EN 61000-6-3 (incl. EN 55022 / EN 55011)
NFB 25 DCB-B / DCB-4B
IEC/EN 61000-6-2
IEC/EN 61000-6-3
All repairs to the pump(s) must be carried out by the relevant KNF
customer service team.
Only use genuine parts from KNF for servicing work.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions5
Diaphragm liquid pump NF 25, NFB 25 Technical data
Assembly
Material1)
Pump head*
PP
Valve plate / seals
EPDM
Diaphragm
EPDM
Resonating diaphragm
EPDM
Assembly
Material1)
Pump head*
PP
Valve plate / seals
FFKM
Diaphragm
PTFE-coated
Resonating diaphragm
FFKM
Assembly
Material1)
Pump head*
PVDF
Valve plate / seals
FFKM
Diaphragm
PTFE-coated
Resonating diaphragm
FFKM
Parameter
Value
Flow rate NF 25 DC-M and DCB [ml/min]
1), 2)
250
Flow rate NF 25 DCB-4 [ml/min]
1), 2)
25-250
Flow rate NFB 25 DCB-4B [ml/min]
1), 2)
2x 50-300
Flow rate NFB 25 DCB-B [ml/min]
1), 2)
2x 300
Permissible pressure [bar g]
1
Suction head [mWG]
3
Parameter
Value
Recommended hose size ID [mm]
4
4. Technical data
Pump materials
The pump type KP stands for:
Tab. 2
1)
according to DIN ISO 1629 and 1043.1
The pump type KT stands for:
Tab. 3
1)
according to DIN ISO 1629 and 1043.1
The pump type TT stands for:
Tab. 4
* The pump head (Fig. 1) comprises a connecting plate and an intermedi-
ate plate.
1)
according to DIN ISO 1629 and 1043.1
Hydraulic ratings
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions6
Tab. 5
1)
Measured with water and using KNF standardised test equipement at
atmospheric pressure, KP pump head material
2)
Flow rates may vary from the values shown, depending on fluid viscosity, ambient temperature, pump head material and the hoses / hose connectors used.
Hydraulic connections
Tab. 6
Diaphragm liquid pump NF 25, NFB 25 Technical data
Motor voltage
12V
24V
Power consumption [W]
2.9
2.9
Max. I load [A]
0.24
0.12
Max. permissible current
Lead size [-]
AWG22
Protection class [-]
IP30
Weight
1)
[g]
64
Motor option
DCB-4
DCB 12V
DCB 24V
Supply voltage [V]
10…26.4
12
24
Power consumption [W]
I load max. [A]
0.11…0.25
0.23
0.12
Lead size [-]
AWG28
Protection class
IP40
Weight
1)
[g]
60
In the case of type NF25 DCB-4, the supply voltage is not
designed for operation with pulse
(PWM).
However, a PWM signal (frequency 1
control the control
Electrostatic sensitive components (ESD)
Specifications NF 25 DC-M
consumption [A]
Tab. 7
Specifications NF 25 DCB / DCB-4
2.9 2.8 2.9
Tab. 8
voltage.
0.31 0.14
-width modulated voltages
-20kHz) can be used to
1) The weight may differ slightly from the stated value, depending on the
version
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions7
Diaphragm liquid pump NF 25, NFB 25 Technical data
Motor option
DCB-4B
DCB-B
DCB-B
Supply voltage [V]
Power consumption [W]
I load max. [A]
0.22…0.54
0.54
0.27
Protection
class
Weight
1)
[g]
210
In the case of type NFB25 DCB-4B, the supply voltage is not
designed for operation with pulse
(PWM).
However, a PWM signal (frequency 1
control the control voltage.
Electrostatic sensitive components (ESD)
Parameter
Value
Permissible ambient temperature range [C]
+ 5 to + 40
Permissible media temperature [C]
+ 5 to + 80
Permissible kinematic viscosity of medium
≤ 150
Specifications NFB 25 DCB-B / DCB-4B
10…26.4 12 24
5.4 6.5 6.5
IP50
Tab. 9
-width modulated voltages
-20kHz) can be used to
1) The weight may differ slightly from the stated value, depending on the
version.
Other parameters
[cSt]
Tab. 10
.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions8
Diaphragm liquid pump NF 25, NFB 25 Assembly and function
1 Outlet
2
3
4
5
6
7
1 Outlet valve
2
3
4
5
6
7
(double-headed pump)
5. Assembly and function
Assembly
Inlet
Connecting plate
Drive
Connecting leads
Head plate
Intermediate plate
Inlet valve
Working chamber
Diaphragm
Eccentric
Connecting rod
Pump drive
Fig. 3 : NFB 25
Fig. 1: Micro diaphragm liquid pump NF 25
Fig. 2: Pump assembly
The diaphragm liquid pumps are based on reciprocating displacement pump technology. The elastic diaphragm (4) is moved up and
down by the eccentric (5) and the connecting rod (6). In the downward stroke it aspirates the medium to be transferred via the inlet
valve (2). In the upward stroke the diaphragm pushes the medium
out of the pump head via the outlet valve (1). The diaphragm
hermetically seals off the working chamber (3) from the pump drive
(7).
NFB 25 Version (double-headed pump, see Fig. 3)
This pump type has two NF25 pump heads that can be operated
individually or in parallel connection (see Chapter 0).
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions9
Diaphragm liquid pump NF 25, NFB 25 Installation and connection
Mounting dimensions
6. Installation and connection
Only install the pump under the operating parameters and conditions described in Chapter 4, Technical data.
Observe the safety precautions (see Chapter 3).
6.1. Installation
Before installation, store the pump at the installation location to
bring it up to ambient temperature.
Mounting dimensions (see Fig. 4 to 6)
Fig. 4: Mounting dimensions NF 25 DC-M
Fig. 5: Mounting dimensions NF 25 DCB
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions10
Diaphragm liquid pump NF 25, NFB 25 Installation and connection
Installation location
Fig. 6: Mounting dimensions NF 25 DCB-4
Fig. 6: Mounting dimensions NFB 25 DCB-B
Fig. 8: Mounting dimensions NFB 25 DCB-4B
Make sure that the installation location is dry and the pump is
protected against water in the form of rain, spray, splashes and
drips.
Protect the pump against dust.
Protect the pump against vibration and impact.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions11
Diaphragm liquid pump NF 25, NFB 25 Installation and connection
Installation position
Generally speaking, the pump can be installed in any position.
For maximum accuracy and rapid venting of one-headed
pumps, please install as shown in the illustration (Fig. 7).
Fig. 9: Optimum installation
position for one-headed pumps:
horizontal
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions12
Diaphragm liquid pump NF 25, NFB 25 Installation and connection
Note the proper polarity.
Fo
red motor cable: +
black motor cable:
Arrows on the pump head indicate the flow direction.
Keep the suction line as short as possible in order to keep the
priming process as brief as possible.
Connected
Hoses
6.2. Electrical connection
Only have the pump connected by an authorized specialist.
Only have the pump connected when the power supply is
disconnected.
When connecting the device to a power source, the relevant
norms, directives, regulations and technical standards must be
observed.
Connecting the pump
Make sure that the power supply data match the data on the
motor's type plate. The current consumption can be found on
the type plate.
Connect the motor cables. For electrical data see Chapter 4.
r DC motors:
-
6.3. Hydraulic connection
components
Only connect components to the pump that are designed to
handle the hydraulic data of the pump (see Chapter 4,
Technical data).
Only use hoses that are suitable for the maximum permissible
operating pressure of the pump (see Chapter 4).
Only use hoses that are sufficiently chemically resistant to the
liquids being transferred.
6.3.1. Connecting the pump
Remove the protective caps.
Connect the suction and pressure lines.
If the pump is used to build up pressure, make sure that all
transition joints between hose and pump are secure in order to
ensure that the hoses cannot come off.
Check that the hoses and transition joints (hose connect-
or/hose) are fitted correctly and securely.
Check that the system is leak-tight.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions13
Diaphragm liquid pump NF 25, NFB 25 Installation and connection
A double-headed pump is normally selected to reduce the
pressure surges caused by the alternation between suction
and exhaust in a diaphragm pump. In order to
effect, in addition to the pump configuration it is important to
configure the hoses correctly.
Please contact your local KNF partner for further information
(www.knf.com).
Please contact your local KNF partner for further information
(www.knf.com).
Series connection damages pump
If the two heads in the NFB versions are connected
in series, the pump will be damaged, and may leak.
Fig. 10: Hose connection options
for NFB versions
6.3.2. NFB 25 hose configuration
achieve this
For low pulsation (pressure surges) in NFB pumps:
Configure pump head hoses in parallel (see Fig. 8).
Keep hoses between pump and junction as short as possible.
Cause as little narrowing as possible with connection ele-
ments. Recommendation: Min. inside diameter 3mm.
Fig. 12: No series connection in the
case of NFB versions
Fig. 11: NFB hose configuration for low pulsation
CAUTION
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions14
Diaphragm liquid pump NF 25, NFB 25 Operation
Risk of burning
The drive heats up.
Excessive pressure and the inherent dangers thereof can be
prevented by placing a bypass line with a pressure relief valve
between the pressur
contact your local KNF partner for further information
(www.knf.com).
Risk of physical injury and damage to the pump
due to automatic start
If the pump overheats and pump operation is
stopped by the thermal switch / electronics, the
pumps will restart automatically as soon as they have
had time to cool down.
Risk of burns when transferring hot media
Pump standstill
7. Operation
Operate the pumps only under the operating parameters and
conditions described in Chapter 4, Technical data.
Make sure that the pumps are being used properly (see Chap-
ter 2.1).
Avoid improper use of the pumps (see Chapter 2.2).
Observe the safety precautions (see Chapter 3).
The pumps are components that are intended to be incorpo-
rated into another machine. Before putting them into service it
must be established that the machinery or systems in which
they are installed meet the relevant regulations.
CAUTION
Avoid contact with the pump drive.
Avoid contact with flammable materials.
e and suction side of the pump. Please
Do not touch the pump or the media transfer
system.
CAUTION
If the pump stops running, restore the system to normal at-
mospheric pressure.
For pumps with thermal switch or electronic overload protection:
WARNING
Avoid contact with flammable materials.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions15
Take steps to ensure that this cannot produce a
hazardous situation.
Diaphragm liquid pump NF 25, NFB 25 Operation
If the pump is operated with short cycles in your application,
please contact your local KNF partner for further information
(www.knf.com.)
Setting and regulating motor
Short cycle operation
Overpressure on the suction side causes medium to flow through the switched
Switching the pumps on and off
The motor speed of the pumps, and thus the flow rate, is adjusta-
speed
ble and can also be regulated to some extent.
For more details, see Chapter 4, Technical data.
Duty cycle / short cycle operation
KNF pumps are designed for continuous operation.
Short start and stop cycles may adversely affect the service life of
the brushed motors.
Switching on the pump
In order to ensure that the pump starts every time, make sure
that counterpressure is reduced to an acceptable level before
start-up. This should also be done during operation after a brief
power cut. Please contact your local KNF partner for further information (www.knf.com).
Switching off the pump
KNF recommends: If transferring aggressive liquids, the pump
should be flushed thoroughly prior to switch off (see Chapter
8.2.1), as this will help to lengthen the service life of the diaphragm.
Restore the system to normal atmospheric pressure (release
hydraulic pressure in pump).
-off pump
Take steps to ensure that this cannot produce a
CAUTION
hazardous situation.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions16
Diaphragm liquid pump NF 25, NFB 25 Operation
0.00
0.05
0.10
0.15
0.20
0.25
0.30
-4.0
-2.0
0.0
2.0
4.0
6.08.0
10.0
Suction height [mWg] Pressure [mWg]
Flow rate [l/min]
0.00
0.05
0.10
0.15
0.20
0.25
0.30
-4.0-2.00.02.04.06.08.010.0
Suction height [mWg] Pressure [mWg]
Flow rate [l/min]
0.00
0.05
0.10
0.15
0.20
0.25
0.30
012345
Control voltage Vctrl [V]
Flow rate [l/min]
0 mWg
10 mWg
Flow rate NF 25
Fig. 13: Flow rate of pump NF 25 DC-M
Fig. 14: Flow rate of pump NF 25 DCB and NF 25 DCB-4
Fig. 15: Flow rate of pump NF 25 DCB-4 depending on control voltage
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions17
Diaphragm liquid pump NF 25, NFB 25 Operation
Flow rate NFB 25
Fig. 16: Flow rate of pump NFB 25 DCB-B and NFB 25 DCB-4B (per
pump head)
Fig. 17: Flow rate of pump NFB 25 DCB-4B depending on control voltage
(per pump head)
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions18
Diaphragm liquid pump NF 25, NFB 25 Servicing
Component
Servicing interval
Pump
- Regular inspection for external
Pump head
- Clean if the flow rate decreases, the
Diaphragm, valve
- Change as soon as pumping capac-
Health hazard due to dangerous
substances in the pump
Depending on the medium transferred, caustic burns
or poisoning are possible.
With NFB pumps the necessary steps are repeated for the
second pump head
Qty
Tools
1
T6 Torx screwdriver
Information on procedure
Tools
8. Servicing
8.1. Servicing schedule
damage or leaks
pump does not work or no vacuum
is created (Chapter 8.2)
plates and seals
Tab. 11
ity decreases, preferabl y sooner
8.2. Cleaning and dismantling
WARNING
Wear protective clothing if necessary, e.g. pro-
tective gloves.
Flush the pump with a neutral liquid and pump
empty.
With NFB pumps, make sure that both pump
heads are flushed.
8.2.1. Flushing the pump
If transferring aggressive media, KNF recommends flushing
the pump with a neutral fluid under atmospheric conditions for
a few minutes before switching off, to extend the service life of
valves and diaphragm.
8.2.2. Preparations for disassembly
.
Flush the pump with a suitable neutralising liquid, and make
Pump empty.
Separate electrical connections.
Disconnect hoses from pump head.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions19
Tab. 12
sure that no dangerous substances are left in the pump.
Diaphragm liquid pump NF 25, NFB 25 Servicing
8.2.3. Disassembling the pump head
Loosen the four head screws (1) and remove the entire
head.
Remove the resonating diaphragm (3) from the connecting
plate (4).
Remove both anchor valves (6) from the intermediate plate
(7) and the connecting plate (4).
Remove the two anchor valve seals (5) from the connecting
plate (4).
Undo the diaphragm (8) by means of the head plate (2). Re-
move the washers and make sure that no washers (9) fall into the pump housing.
We recommend replacing the diaphragm (8).
8.2.4. Clean or replace parts
Clean the anchor valves (6), the anchor valve seals (5) and
the diaphragm (8) with a cloth and then blow off with compressed air or replace.
Blow off the intermediate plate (7), connecting plate (4) and
head plate (2) with compressed air.
8.2.5. Assembling the pump head
Screw diaphragm (8) with washer (9) into connecting rod.
Make sure that no washers (9) fall into the pump housing.
Use the head plate (2) to carefully tighten the diaphragm.
Insert the seals (5) in the connecting plate (4).
Insert the anchor valves (6) in the intermediate plate (7) and
the connecting plate (4).
Place the connecting plate (4) on top of the intermediate
plate (7).
A pin and hole system prevents incorrect assembly.
Insert the resonating diaphragm (3) into the connecting plate
(4).
Place the head plate (2) on top of the connecting plate (4). A
pin and hole system prevents incorrect assembly.
Insert the four head screws (1) into the through holes of the
pump head.
Place the pump head on top of the pump housing and make
sure that the flow direction of the pump head (see directional
arrow on the head plate (2)) is the same as before.
Alternately tighten the four head screws (1).
Reconnect the hoses to the pump head.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions20
Diaphragm liquid pump NF 25, NFB 25 Servicing
Escaping liquid
After assembly the pump may not be leak
to incorrect assembly, damaged or soiled seal faces,
or other reasons.
-tight due
CAUTION
Run pump for several minutes with a harmless
liquid at maximum operating pressure.
Check that pump is leak-tight.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions21
Flow rate, suction head or pressure head is too low
Possible cause
Remedy
Components in the system
Modify installation, check the cross-sections of components
Hose connections are not leak-
Secure tra nsit ion joints between hose and hose connections
Particles in the pump
Clean the pump head, install suction-side filter if required
Viscosity of the transferred
Contact KNF
Incorrect interchange of pres-
Remove pressure and suction lines and reconnect correctly
The pump parts are not re-
Replace the pump head with a compatible version
The pump does not achieve the performance stated in the technical data or on the data sheet.
connected to the suction and
pressure sides, such as hoses,
valves or filters, are causing too
much resistance
tight
with clamps or clamping elements
(see Chapter 8.2)
medium is too high
sure and suction line connections
sistant to the medium to be
transferred
Tab. 10
Fault cannot be rectified
If you are unable to identify any of the above causes, please send
the pump to KNF customer service (see address on last page).
Flush the pump to clear the pump head of any hazardous or
aggressive liquids (see Chapter 8.2.1).
Dismantle the pump.
Clean the pump (see Chapter 8.2.2).
Send the pump, with completed decontamination declaration
(see Chapter 10), to KNF customer service stating the nature
of the transferred medium.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions23
Diaphragm liquid pump NF 25, NFB 25 Accessories
for hose
for hose
Please contact your local KNF partner for further information
(www.knf.com).
10. Accessories
The following filters are available:
Type Material Mesh size
FS 60 T PVDF 70 µm 600 ml/ min UNF 1/4“- 28 165210
FS 60 X PEEK 35 µm 600 ml/ min UNF 1/4“- 28 165212
FS 25 T PVDF 70 µm 250 ml/ min
FS 25 X PEEK 35 µm 250 ml/ min
Max. flow volume
(liquid)
Tab. 16
Connections Order No.
ID 3.2 / 4 mm
ID 3.2 / 4 mm
165211
165213
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions24
KNF shall only undertake to repair the pump on condition that
the customer provides certification of the transferred media
and the cleaning of the pump (decontamination declarati
11. Decontamination declaration
on).
In order to send a product back use the decontamination
declaration, which either was delivered with the product or is
available on www.knf.com (Downloads).
Please fill in the pump type, serial number, pumped media and all
other required information. Send the signed form together with the
product to your KNF representative.
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions25
Diaphragm liquid pump NF 25, NFB 25 For your notes
KNF Flodos AG BA_NF25_NFB25_EN_07_159615
Translated from the Original Operating and Installation Instructions 26
KNF worldwide
Please find your local KNF partners at: www.knf.com