KNF N 810 FT.18, N 820 FT.18, N 840 FT.18, N 840.1.2 FT. 18, N 810.3 FT.18 User guide

...
Page 1
Operating Instructions
Read and observe these Operating Instructions!
Laboport®
Chemically-resistant Laboratory Pumps
and Systems
N 810 FT.18 N 820 FT.18 N 840 FT.18 N 840.1.2 FT.18
N 810.3 FT.18 N 820.3 FT.18 N 840.3 FT.18 N 842.3 FT.18
SR SH SC SCC
KNF Neuberger GmbH Alter Weg 3 D-79112 Freiburg Germany Phone ++49 / (0)7664 / 5909-0 Fax ++49 / (0)7664 / 5909-99 E-Mail: www.knf.de
KNF 121211-121345 07/13 Translation of original Operating Instructions, english
info@knf.de
Page 2
Page 3
LABOPORT Pumps and Systems Content
Content Page
1. About this document 4
1.1. Using the Operating Instructions 4
1.2. Symbols and Markings 4
2. Use 5
2.1. Proper Use 5
2.2. Improper Use 5
3. Safety 6
4. System overview 8
5. Technical Data 10
5.1. Pumps 10
5.2. Electrical Supply Unit 17
5.3. Vacuum Controller 17
5.4. Vacuum Systems 18
6. Design and Function 19
6.1. Pump 19
6.2. Separator 20
6.3. High Performance Condenser 21
6.4. Vacuum Controller 22
7. Installation, mounting and connection 23
7.1. Connect pump 23
7.2. Baseplate 24
7.3. Separator 24
7.4. High Performance Condenser 25
7.5. Electrical Supply Unit and
Vacuum Controller 25
7.6. Gas Ballast 28
7.7. Mounting of Systems 28
7.7.1. System SR 28
7.7.2. System SH 29
7.7.3. System SC 30
7.7.4. System SCC 31
7.7.5. Retrofitting from System SC to
System SCC 32
9.2.3. Emptying and Cleaning the Separator 37
9.2.4. Emptying and Cleaning the Condenser 37
9.3. Changing Diaphragm and Valve
Plates/Sealings 38
10. Troubleshooting 42
10.1. Pump/System without Vacuum
Controller 42
10.2. System with one Vacuum
Controller 43
10.3. System with two Vacuum
Controllers 45
10.4. Fault cannot be rectified 45
11. Ordering Information 46
11.1. Pumps and Spare Parts 46
11.2. Accessories for Pump Systems 46
12. Decontamination Declaration 47
8. Operation 33
8.1. Pump 33
8.1.1. Preparing for Start-up 33
8.1.2. Starting 33
8.2. Switching Pump/System on and off 35
8.3. Notices related to System Operation 35
8.3.1. Separators 35
8.3.2. High Performance Condenser 35
8.3.3. Vacuum Controller 36
8.3.4. Gas Ballast 36
9. Servicing 37
9.1. Servicing Schedule 37
9.2. Cleaning 37
9.2.1. Flushing Pump 37
9.2.2. Cleaning Pump 37
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 3
Page 4
About this document LABOPORT Pumps and Systems
warning are specified here. The signal word, e.g.
1. About this document
1.1. Using the Operating Instructions
The Operating Instructions are part of the pump/system.
Carefully study the Operating Instructions before using a pump
or system.
Always keep the Operating Instructions handy in the work
area.
Pass on the Operating Instructions to the next owner.
Project systems and pumps
Customer-specific project systems or pumps (systems and pump models which begin with “PJ” or “PM”) may differ from the Operat­ing Instructions.
For project systems and pumps, also observe the agreed upon specifications.
1.2. Symbols and Markings
Warning
A danger warning is located here.
Possible consequences of a failure to observe the
WARNING
Danger levels
Signal word Meaning Consequences if not observed
DANGER
WARNING
CAUTION
Tab. 1
Other information and symbols
Warning, indicates the danger level.
Measures for avoiding the danger and its conse-
quences are specified here.
warns of immedi­ate danger
warns of possible danger
warns of a possibly dangerous situa­tion
Death or serious injuries and/or serious damage are the consequence.
Death or serious injuries and/or serious damage are possible.
Minor injuries or damage are possible.
4 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
An activity to be carried out (a step) is specified here.
1. The first step of an activity to be carried out is specified here. Additional, consecutively numbered steps follow.
This symbol refers to important information.
Page 5
LABOPORT Pumps and Systems Use
2. Use
2.1. Proper Use
The pump/system is exclusively intended for transferring gases and vapors.
Owner's responsibility
Operating parameters and
conditions
Requirements for
transferred medium
High performance condenser
Accessories
Only install and operate the pump/system under the operating parameters and conditions described in chapter 5, Technical data.
Make sure that the installation location is dry and the pump/system is protected against rain, splash, hose and drip water.
Before using a medium, check the compatibility of the materials of the pump head, diaphragm and valves with the medium.
Before using a medium, check whether the medium can be trans­ferred danger-free in the specific application case.
Only transfer gases which remain stable under the pressures and temperatures occurring in the pump.
The high performance condenser must be installed on the outlet side of the pump; if it is installed on the inlet side there is a danger of implosion.
Observe the correct usage of the gas- and cooling liquid­connections on the high performance condenser. Inlet and outlet connections for the gas are not interchangeable.
Laboratory equipment or additional components connected to a pump/system have to be suitable for use with the pneumatic capa­bilities of the pump (see chapter 5.1, page 10).
2.2. Improper Use
The pump/system may not be operated in an explosive atmos­phere.
The pump/system is not suitable for transferring dusts.
The pump/system is not suitable for transferring liquids.
For LABOPORT systems with vacuum controller: The vacuum system must not be used if the entry of air or gas into the vacuum system during venting (pump vent valve) could result in the crea­tion of reactive, explosive or otherwise hazardous mixtures (e.g. with the medium).
The pump/the system must not be used to create vacuum and overpressure simultaneously.
An overpressure must not be applied to the suction side of the pump/ the system.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 5
Page 6
Safety LABOPORT Pumps and Systems
3. Safety
Note the safety precautions in chapters 7. Installation, mount­ing and connection, and 8. Operation.
The pump/system is built according to the generally recognized rules of technology and in accordance with the occupational safety and accident prevention regulations. Nevertheless, dangers can result during their use which lead to injuries to the user or others, or to damage to the pump/system or other property.
Only use the pump/system when it is in a good technical and proper working order, in accordance with its intended use, observ­ing the safety advice within the operating instructions, at all times.
Personnel
Working in a safety-
conscious manner
Fig. 1: Notice sticker
Handling dangerous media
Handling flammable media
Make sure that only trained and instructed personnel or specially trained personnel work on the pump/system. This especially ap­plies to assembly, connection and servicing work.
Make sure that the personnel has read and understood the operat­ing instructions, and in particular the "Safety" chapter.
Observe the accident prevention and safety regulations when performing any work on the pump/system and during operation.
Do not expose any part of your body to the vacuum.
Open housing parts with notice sticker (see fig. 1) only after separating mains plug from power source.
When transferring dangerous media, observe the safety regula­tions when handling these media.
Be aware that the pump/system is not designed to be explosion­proof.
Make sure the temperature of the medium is always sufficiently below the ignition temperature of the medium, to avoid ignition or explosion. This also applies for unusual operational situations.
Note that the temperature of the medium increases when the pump compresses the medium.
Hence, make sure the temperature of the medium is sufficiently below the ignition temperature of the medium, even when it is compressed to the maximum permissible operating pressure of the pump. The maximum permissible operating pressure of the pump is stated in the technical specifications (see chapter 5.1, page 10).
If necessary, consider any external sources of energy, such as radiation, that may add heat to the medium.
In case of doubt, consult the KNF customer service.
Environmental protection
6 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Store all replacement parts in a protected manner and dispose of them properly in accordance with the applicable environmental protection regulations. Observe the respective national and inter­national regulations. This especially applies to parts contaminated with toxic substances.
Page 7
LABOPORT Pumps and Systems Safety
Standards
Customer service and
repairs
The pumps conform to the Directive 2011/65/EU (RoHS2).
The pumps conform to the safety regulations of the EC Directive 2004/108/EC concerning Electromagnetic Compatibility and the EC Directive 2006/42/EC concerning Machinery. The requirements of the following harmonised standards are fulfilled:
DIN EN 61010-1
DIN EN 61326-1
DIN EN 61000-3-2/3
The pumps correspond to IEC 664:
the overvoltage category II
the pollution degree 2
Only have repairs to the pump/system carried out by the KNF Customer Service responsible.
Only authorized personnel should open those parts of the housing that contain live electrical parts.
Use only genuine parts from KNF for servicing work.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 7
Page 8
System overview LABOPORT Pumps and Systems
4. System overview
1 Electrical supply unit 2 Vacuum controller 3 Vent valve at vacuum
controller
4 On/off switch of vacuum
controller 5 Pump vent valve 6 Vacuum valve 7 High performance conden-
ser 8 Hose connector 9 Angled nozzle for coolant 10 Spring clamp 11 Flask for condenser 12 Support for high perfor-
mance condenser 13 Locating pin for pump 14 Separator 15 Holder for Separator 16 Support for vacuum control-
ler 17 Baseplate 18 Vacuum pump 19 On/off switch of pump
Fig. 2: Full expanded LABOPORT System SCC..., for example pump model N 840.3 FT.18 (symbolic)
8 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 9
LABOPORT Pumps and Systems System overview
Fig. 3: LABOPORT Systems
Module SR... SH... SC... SCC...
System
Pump X X X X Baseplate X X X X 2 separators X 1 separator X X X High performance
X X X
condenser Electrical supply unit X X 1 vacuum controller X 2 vacuum controllers X
Tab. 2
Each pump and all systems can be expanded modularly up to a full system SCC (see chapter 7.7, page 28).
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 9
Page 10
Technical Data LABOPORT Pumps and Systems
5. Technical Data
5.1. Pumps
All pumps are secured against overheating with thermal switches and are equipped with a mains fuse.
Pump materials (for all pump types) *
Pump head PTFE Diaphragm PTFE coated Valve FFPM
Tab. 3 *according to DIN ISO 1629 and 1043.1
Refer to the type plate for the pump’s electrical configuration.
N 810 FT.18
Pneumatic performance
Max. permissible operating
1.0
pressure [bar g] Ultimate vacuum [mbar abs.]
Delivery rate at atm. pressure
100
max. 10
[l/min]*
Pneumatic connection
Hose connection [mm] ID 10
Ambient and media temperature
Permissible
+ 5 °C to + 40 °C
ambient temperature Permissible media temperatu-
+ 5 °C to + 40 °C
re
Other parameters
Weight [kg] 5.9
Dimensions: L x H x W [mm] 256 x 187 x 146
Maximum permissible ambient relative humidity
80 % for temperatures up to 31 °C, decreasing linearly to 50 % at 40 °C
Maximum altitude of site
2000
[m above sea level]
Electrical Data
Voltage [V] 100 115 230 Frequency [Hz] 50/60 60 50 Maximum current consump-
1.4 1.3 0.6
tion [A] Power consumption pump [W] 110 110 100 Maximum permitted mains
+/- 10 % +/- 10 % +/- 10 %
voltage fluctuation Fuse pump
2.5 2.5 1.25
(2x) T [A] Protection class motor IP44
Tab. 4 * Liters in standard state (1,013 mbar)
10 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 11
LABOPORT Pumps and Systems Technical Data
N 820 FT.18
Pneumatic performance
Max. permissible operating
1.0
pressure [bar g] Ultimate vacuum [mbar abs.]
Delivery rate at atm. pressure
100
max. 20
[l/min]*
Pneumatic connection
Hose connection [mm] ID 10
Ambient and media temperature
Permissible
+ 5 °C to + 40 °C
ambient temperature Permissible media temperatu-
+ 5 °C to + 40 °C
re
Other parameters
Weight [kg] 7.1
Dimensions: L x H x W [mm] 268 x 207 x 159
Maximum permissible ambient relative humidity
80 % for temperatures up to 31 °C, decreasing linearly to 50 % at 40 °C
Maximum altitude of site
2000
[m above sea level]
Electrical Data
Voltage [V] 100 115 230 Frequency [Hz] 50/60 60 50 Maximum current consump-
2.2 1.9 0.9
tion [A] Power consumption pump [W] 145 145 130 Maximum permitted mains
+/- 10 % +/- 10 % +/- 10 %
voltage fluctuation Fuse pump
4.0 3.15 2.0
(2x) T [A] Protection class motor IP44
Tab. 5 * Liters in standard state (1,013 mbar)
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 11
Page 12
Technical Data LABOPORT Pumps and Systems
N 840 FT.18
Pneumatic performance
Max. permissible operating
1.0
pressure [bar g] Ultimate vacuum [mbar abs.]
Delivery rate at atm. pressure
100
max. 34
[l/min]*
Pneumatic connection
Hose connection [mm] ID 10
Ambient and media temperature
Permissible
+ 5 °C to + 40 °C
ambient temperature Permissible media temperatu-
+ 5 °C to + 40 °C
re
Other parameters
Weight [kg] 10.3
Dimensions: L x H x W [mm] 297x 226 x 171
Maximum permissible ambient relative humidity
80 % for temperatures up to 31 °C, decreasing linearly to 50 % at 40 °C
Maximum altitude of site
2000
[m above sea level]
Electrical Data
Voltage [V] 100 115 230 Frequency [Hz] 50/60 60 50 Maximum current consump-
4.4 3.2 1.5
tion [A] Power consumption pump [W] 200 220 180 Maximum permitted mains
+/- 10 % +/- 10 % +/- 10 %
voltage fluctuation Fuse pump
6.3 6.3 3.15
(2x) T [A] Protection class motor IP44
Tab. 6 * Liters in standard state (1,013 mbar)N 810.3 FT.18
12 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 13
LABOPORT Pumps and Systems Technical Data
N 810.3 FT.18
Pneumatic performance
Max. permissible operating
1.0
pressure [bar g] Ultimate vacuum [mbar abs.]
Delivery rate at atm. pressure
8
max. 10
[l/min]*
Pneumatic connection
Hose connection [mm] ID 10
Ambient and media temperature
Permissible
+ 5 °C to + 40 °C
ambient temperature Permissible media temperatu-
+ 5 °C to + 40 °C
re
Other parameters
Weight [kg] 6.9
Dimensions: L x H x W [mm] 281 x 187 x140
Maximum permissible ambient relative humidity
80 % for temperatures up to 31 °C, decreasing linearly to 50 % at 40 °C
Maximum altitude of site
2000
[m above sea level]
Electrical Data
Voltage [V] 100 115 230 Frequency [Hz] 50/60 60 50 Maximum current consump-
1.4 1.3 0.6
tion [A] Power consumption pump [W] 100 110 90 Maximum permitted mains
+/- 10 % +/- 10 % +/- 10 %
voltage fluctuation Fuse pump
2.5 2.5 1.25
(2x) T [A] Protection class motor IP44
Tab. 7 *Liters in standard state (1,013 mbar)
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 13
Page 14
Technical Data LABOPORT Pumps and Systems
N 820.3 FT.18
Pneumatic performance
Max. permissible operating
1.0
pressure [bar g] Ultimate vacuum [mbar abs.]
Delivery rate at atm. pressure
8
max. 20
[l/min]*
Pneumatic connection
Hose connection [mm] ID 10
Ambient and media temperature
Permissible
+ 5 °C to + 40 °C
ambient temperature Permissible media temperatu-
+ 5 °C to + 40 °C
re
Other parameters
Weight [kg] 9.3
Dimensions: L x H x W [mm] 312 x 207 x 144
Maximum permissible ambient relative humidity
80 % for temperatures up to 31 °C, decreasing linearly to 50 % at 40 °C
Maximum altitude of site
2000
[m above sea level]
Electrical Data
Voltage [V] 100 115 230 Frequency [Hz] 50/60 60 50 Maximum current consump-
1.8 1.2 0.7
tion [A] Power consumption pump [W] 130 130 120 Maximum permitted mains
+/- 10 % +/- 10 % +/- 10 %
voltage fluctuation Fuse pump
3.15 2.5 1.6
(2x) T [A] Protection class motor IP44
Tab. 8 *Liters in standard state (1,013 mbar)
14 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 15
LABOPORT Pumps and Systems Technical Data
N 840.3 FT.18
Pneumatic performance
Max. permissible operating
1.0
pressure [bar g] Ultimate vacuum [mbar abs.]
Delivery rate at atm. pressure
8
max. 34
[l/min]*
Pneumatic connection
Hose connection [mm] ID 10
Ambient and media temperature
Permissible
+ 5 °C to + 40 °C
ambient temperature Permissible media temperatu-
+ 5 °C to + 40 °C
re
Other parameters
Weight [kg] 12.6
Dimensions: L x H x W [mm] 341 x 226 x 166
Maximum permissible ambient relative humidity
80 % for temperatures up to 31 °C, decreasing linearly to 50 % at 40 °C
Maximum altitude of site
2000
[m above sea level]
Electrical Data
Voltage [V] 100 115 230 Frequency [Hz] 50/60 60 50 Maximum current consump-
4.4 3.2 1.5
tion [A] Power consumption pump [W] 220 250 245 Maximum permitted mains
+/- 10 % +/- 10 % +/- 10 %
voltage fluctuation Fuse pump
6.3 6.3 3.15
(2x) T [A] Protection class motor IP44
Tab. 9 *Liters in standard state (1,013 mbar)
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 15
Page 16
Technical Data LABOPORT Pumps and Systems
N 840.1.2 FT.18
Pneumatic performance
Max. permissible operating
1.0
pressure [bar g] Ultimate vacuum [mbar abs.]
Delivery rate at atm. pressure
90
max. 60
[l/min]*
Pneumatic connection
Hose connection [mm] ID 10
Ambient and media temperature
Permissible
+ 5 °C to + 40 °C
ambient temperature Permissible media temperatu-
+ 5 °C to + 40 °C
re
Other parameters
Weight [kg] 12.6
Dimensions: L x H x W [mm] 341 x 226 x 160
Maximum permissible ambient relative humidity
80 % for temperatures up to 31 °C, decreasing linearly to 50 % at 40 °C
Maximum altitude of site
2000
[m above sea level]
Electrical Data
Voltage [V] 100 115 230 Frequency [Hz] 50/60 60 50 Maximum current consump-
5.1 4.2 1.9
tion [A] Power consumption pump [W] 275 280 270 Maximum permitted mains
+/- 10 % +/- 10 % +/- 10 %
voltage fluctuation Fuse pump
6.3 6.3 3.15
(2x) T [A] Protection class motor IP44
Tab. 10 *Liters in standard state (1,013 mbar)
16 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 17
LABOPORT Pumps and Systems Technical Data
N 842.3 FT.18
Pneumatic performance
Max. permissible operating
1.0
pressure [bar g] Ultimate vacuum [mbar abs.]
Delivery rate at atm. pressure
2
max. 34
[l/min]*
Pneumatic connection
Hose connection [mm] ID 10
Ambient and media temperature
Permissible
+ 5 °C to + 40 °C
ambient temperature Permissible media temperatu-
+ 5 °C to + 40 °C
re
Other parameters
Weight [kg] 13.4
Dimensions: L x H x W [mm] 341 x 223 x 167
Maximum permissible ambient relative humidity
80 % for temperatures up to 31 °C, decreasing linearly to 50 % at 40 °C
Maximum altitude of site
2000
[m above sea level]
Electrical Data
Voltage [V] 100 115 230 Frequency [Hz] 50/60 60 50 Maximum current consump-
4.4 3.2 1.5
tion [A] Power consumption pump [W] 260 290 245
Fuse pump
6.3 6.3 3.15
(2x) T [A] Protection class motor IP44
Tab. 11 *Liters in standard state (1,013 mbar)
5.2. Electrical Supply Unit
Electrical Data
Voltage [V] 100 115 230 Frequency [Hz] 50/60 60 50 Power consumption* [W] 270 300 295 Fuse pump
6.3 6.3 3.15
(2x) T [A]
Tab. 12
* includes the power consumption of the whole system (inclusive pump and vacuum controller); for pumps N 840.1.2 FT.18 and N 842.3 FT.18 power consumption raises by approx. 50 W.
5.3. Vacuum Controller
See Operating Manual of the vacuum controller.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 17
Page 18
Technical Data LABOPORT Pumps and Systems
5.4. Vacuum Systems
Ambient temperature
Permissible
+ 5 °C to + 40 °C
ambient temperature
Operating parameters of coolant (only for high performance condensers)
Permissible pressure [bar g] 3 Permissible temperature - 15 °C to + 20 °C Condenser connectors For tube 8mm ID
Tab. 13
18 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 19
LABOPORT Pumps and Systems Design and Function
6. Design and Function
6.1. Pump
Design
1 Connection piece 2 Pneumatic connection 3 Pump head 4 Outlet (pressure side) 5 Inlet (suction side) 6 Power switch
1 Outlet valve 2 Inlet valve 3 Transfer chamber 4 Diaphragm 5 Eccentric 6 Connecting rod 7 Pump drive
Fig. 4: Diaphragm pump (shown: pump N 840.3 FT.18)
Function Diaphragm pump
Fig. 5: Pump head
Diaphragm pumps transfer, compress (depending on pump ver­sion) and evacuate gases and vapors.
The elastic diaphragm (4) is moved up and down by the eccentric (5) and the connecting rod (6). In the downward stroke it aspirates the gas to be transferred via the inlet valve (2). In the upward stroke, the diaphragm presses the medium out of the pump head via the outlet valve (1). The transfer chamber (3) is hermetically separated from the pump drive (7) by the diaphragm.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 19
Page 20
Design and Function LABOPORT Pumps and Systems
1
2
3
4
6.2. Separator
Design
1 Tubing 2 Hose nozzle 3 Separator container 4 Holder for separator
Fig. 6: Separator (shown: LABOPORT system SR with pump N 840.3 FT.18 as example)
Function
Condensable components in the gas can be separated on the pressure side of the pump. On the suction side the separator collects particular matter and droplets. This protects the pump from contamination.
The separator is made of a specially treated glass and features implosion protection.
20 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 21
LABOPORT Pumps and Systems Design and Function
1
2
3
3
4
5
6
7
6.3. High Performance Condenser
Design
1 Hose nozzle 2 High Performance Conden-
ser
3 Angled nozzle for coolant 4 Tubing 5 Spring clamp 6 Flask 7 Support
Fig. 7: High performance condenser (shown: LABOPORT system SH with pump N 840.3 FT.18 as example)
Function
The high performance condenser at pump outlet enables conden­sable components in the vapour to be separated out.
The condensate is collected in a glass flask. The flask is attached to the condenser flange with a clamp. The condensation tempera­ture is maintained by running cold water or recirculated coolant through the condenser.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 21
Page 22
Design and Function LABOPORT Pumps and Systems
1 2 2 3 1
5
5
4
6.4. Vacuum Controller
1 Tubing 2 Power switch 3 Support 4 Electrical supply unit 5 Vacuum controller
Fig. 8: Vacuum controller (shown: LABOPORT system SCC with pump N 840.3 FT.18 as example)
Function
In laboratory systems without a vacuum controller the pump runs continuously and works against its vacuum. If there is a vacuum controller, it will switch the pump off once the selected target vacuum has been reached.
With two vacuum controllers it becomes possible to simultaneously run two different and independent processes with just one pump.
Once one vacuum controller’s set point has been reached, that vacuum controller will close the valve it is controlling. If two vacuum controllers are used, the pump will continue to work until the second controller’s set point has been reached. Then the second vacuum controller will close the valve it is controlling. The pump switches off and the pump vent valve is opened.
If one of the two vacuum values falls below the hysteresis range, the pump will switch back on.
Refer to the vacuum controller’s operating instructions for more information.
22 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 23
LABOPORT Pumps and Systems Installation, mounting and connection
7. Installation, mounting and connection
Only install and operate the pumps/systems under the operating parameters and conditions described in chapter 5, Technical data.
Observe the safety precautions (see chapter 3, page 6).
Before installation, store the pump/the accessories at the
installation location to bring it up to room temperature.
Dimensions
Cooling air supply
Installation location
Connected components
Pump exhaust
See chapter 5, Technical data, for the dimensions of system.
Install the pump/system so that the motor fan can intake suffi-
cient cooling air.
Make sure that the installation location is dry and the
pump/system is protected against rain, splash, hose and drip water.
Choose a safe location (flat surface) for the pump/system.
Protect the pump/system from dust.
Protect the pump/system from vibrations and jolts.
7.1. Connect pump
Only connect components to the pump which are designed for
the pneumatic data of the pump (see chapter 5.1, page 10).
If the pump is used as a vacuum pump, safely discharge the
pump exhaust at the pump’s pneumatic outlet.
A marking on the pump head shows the direction of flow.
1. Remove the protective plugs from the pneumatic connectors of the pump.
2. Connect the suction line and pressure line.
3. Lay the suction and pressure line at a downward angle to prevent condensate from running into the pump.
4. Insert the power cable’s plug into a properly installed shockproof socket.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 23
Page 24
Installation, mounting and connection LABOPORT Pumps and Systems
7.7.1 (system SR), 7.7.2 (system SH), 7.7.3 (system SC), 7.7.4
7.2. Baseplate
Condition
Fig. 9: Installation of the mounts at
the pump
Conditions
Tools and material
Pump disconnected from mains
1. Unscrew the pump`s rubber feet (1) anti-clockwise.
2. Install the mounts (2) onto the pump.
3. Stand the pump (fitted with the mounts) onto the baseplate
4. Aligne the mounts on the pump and the baseplate.
5. Slide the locating pin (see fig. 2/13) through the mount holes
on the baseplate and on the pump until they are fully engaged.
6. Check that the pump is securely fastened to the baseplate via both mounts.
7.3. Separator
Baseplate mounted (see chapter 7.2)
Pump disconnected from mains
Qty Material 1 Phillips-head screwdriver No. 3
Tab. 14
1. Slide the container for the separator (see fig. 6/4) into the baseplate’s upward-pointing slotted bar.
2. Fix the holders by tighten the screw in the bottom of them.
3. Place the glass separators into the holders.
Laying hoses: depending on system configuration see chapter
(system SCC) or 7.7.5 (retrofitting from system SC to system SCC).
24 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 25
LABOPORT Pumps and Systems Installation, mounting and connection
7.4. High Performance Condenser
The high performance condenser is delivered installed on a support.
Conditions
Touls and material
Baseplate mounted (see chapter 7.2, page 24)
Pump disconnected from mains and de-energized
Qty Material 1 Allen key 5 mm
Tab. 15
Installing the high performance condenser
Danger of implosion if installed improperly
There is a danger of implosion if the high performance condenser is installed on the pump
WARNING
1. Slide the support of high performance condenser (see fig. 7/7)
into the baseplate’s upward-pointing slotted bar.
inlet. Connect the high performance condenser to the
pump outlet only.
Conditions
Tools and material
2. Fix the support to the baseplate by tightening the allen screw.
3. Align the high performance condenser so that the hose connection at the inlet is lower than the pump outlet.
Laying hoses: depending on system configuration see chapter
7.7.2 (system SH), 7.7.3 (system SC), 7.7.4 (system SCC) or
7.7.5 (retrofitting from system SC to system SCC).
7.5. Electrical Supply Unit and Vacuum Controller
The vacuum controller(s) is (are) delivered with the ventilation valve and pump vent valve installed on a support.
When retrofitting from system SC to system SCC, the second vacuum controller is delivered complete with the ventilation valve installed on a new support. Install the existing vacuum controller on the new support. For more information about retrofitting from system SC to system SCC, refer to chapter
7.7.5, page 32.
Baseplate mounted (see chapter 7.2, page 24)
Pump disconnected from mains
Qty Material 1 Allen key 4 mm 1 Allen key 5 mm
Tab. 16
Mounting the electrical supply unit
1. Slide the electrical supply unit (see fig. 8/4) with both tightening nuts into the baseplate’s lateral slotted bar.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 25
Page 26
Installation, mounting and connection LABOPORT Pumps and Systems
2. Tighten the allen screws at the bottom of the electrical supply unit.
Mounting the vacuum controller
1. Slide the support’s attachment part (see fig. 8/3) into the baseplate’s lateral, upward-pointing slotted bar.
2. Tighten the allen screw at the bottom of the support.
Wire and connect with each other the electrical supply unit and the vacuum controller
Figures 11 and 12 provide an overview of the electrical connections for LABOPORT systems SC and SCC.
1. Connect signal socket of the vacuum controller with the elet­rical supply unit socket marked VC 1 (fig. 10/5) using the con­nection cable.
2. When using two vacuum controllers, connect the second vacuum controller’s signal socket to the electrical supply unit’s VC2 connection (6).
Fig. 10: Connections of the electri-
cal supply unit
3. Insert the vacuum controller power cord plug into the designat­ed electrical supply unit socket (2).
4. When using two vacuum controllers, insert the vacuum valves’ plugs into the electrical supply unit’s VV1 (4) und VV2 (3) connections.
5. Insert the pump vent valve plug into the electrical supply unit socket marked PRV (7).
6. Insert the vacuum pump power cord plug into the designated electrical supply unit socket (1).
7. Tuck all of the wires into the electrical supply unit and secure the cover.
8. Make sure power switches on the pump and on the vacuum controller(s) are switched off.
9. Insert the electrical supply unit`s system power cord plug into a shockproof socket.
26 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 27
LABOPORT Pumps and Systems Installation, mounting and connection
(optional)
Pump vent
to vacuum pump
System power connect
Electrical supply unit
valve
Vacuum controller (rear)
Fig. 11: Electrical connection LABOPORT System SC
Vacuum controller 1 (rear)
to vacuum pump
System power connect
Vacuum valve 1
Vacuum valve 2
Pump vent valve
Vacuum controller 2
Coolant valve
Electrical supply unit
Fig. 12: Electrical connection LABOPORT System SCC
Laying hoses: depending on system configuration see chapter
7.7.3 (system SC), 7.7.4 (system SCC) or 7.7.5 (retrofitting from system SC to system SCC).
Mounting coolant valve for high performance condenser (optional)
With a coolant valve it becomes possible to stop the flow of coolant if the system’s control operation is interrupted or terminated.
Danger of the high performance condenser bursting
Make sure that the coolant valve is mounted
between the coolant supply and the coolant inlet
WARNING
port of high performance condenser.
1. Connect the coolant valve to the coolant supply.
2. Insert the coolant valve’s plug into the vacuum
controller’s VV1 socket (see fig. 10/4).
See fig. 16, page 29 for coolant inlet and outlet.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 27
Page 28
Installation, mounting and connection LABOPORT Pumps and Systems
7.6. Gas Ballast
Only for two-headed pumps.
Condition
1 2
4
Fig. 13: Mounting the gas ballast
Motor disconnected from mains and de-energized
Mounting
1. Disconnect the tubes from the inlet (4) and the outlet (3) of the
3
pump.
2. Remove the pump from the baseplate.
3. Open the connection (2) of the pump heads.
4. Screw out the connecting piece (1) of the vacuum side head.
5. Screw the gas ballast into the pump head.
6. Remount the connection (2) between the both pump heads.
7.7. Mounting of Systems
The displayed system tube configuration is for two-headed pumps only.
Contact KNF Service for information about laying tubes for one-headed pumps and pumps with aluminum heads (“A” in the type designation).
7.7.1. System SR
Tools and material
Fig. 14: Tubing system SR
The system SR consists of:
baseplate
two separators
Qty Material 1 Phillips-head screwdriver No. 3 2 Tubes (see fig. 14)
Resistant to the medium employed
Inside diameter 10 mm
Length: approx. 150 mm
Tab. 17
1. Attach pump to the baseplate (see chapter 7.2, page 24).
2. Install the separators (chapter 7.3, page 24).
3. Lay tubes for separators (see fig. 14).
28 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 29
LABOPORT Pumps and Systems Installation, mounting and connection
7.7.2. System SH
The system SH consists of:
Baseplate
1 separator
1 high performance condenser
Tools and material
Qty Material 1 Phillips-head screwdriver No. 3 1 Allen key 5 mm 2 Tubes (see fig. 15)
Resistant to the medium employed
Inside diameter 10 mm
Length: (1) approx. 150 mm, (2) approx. 220 mm
Tab. 18
Danger of the high performance condenser bursting
Make sure that the high performance conden-
ser´s upper gas outlet is not blocked.
WARNING
Fig. 15: Tubing system SH
Fig. 16: Coolant supply of high
performance condenser
Improper laying of hoses will result in damage to the pump
Correctly assign gas and coolant hose
DANGER
connections.
Do not reverse the gas connections’ inlets and
outlets.
1. Attach pump to the baseplate (see chapter 7.2, page 24).
2. Install the separator (see chapter 7.3, page 24).
3. Install the high performance condenser (see chapter 7.4, page
25).
4. Lay hoses for system (see fig. 15).
The condenser-connectors for the coolant require connection tubing with an inside diameter of 8 mm. Coolant inlet and outlet see fig. 16.
When using a coolant valve:
Danger of the high performance condenser bursting
WARNING
Make sure that the coolant valve is mounted
between the coolant supply and the coolant inlet port of high performance condenser.
1 Coolant inlet 2 Coolant outlet
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 29
Page 30
Installation, mounting and connection LABOPORT Pumps and Systems
) approx. 300 mm,
7.7.3. System SC
The system SC consists of:
Baseplate
1 separator
1 high performance condenser
Electrical supply unit with 1 vacuum controller
Tools and material
Qty Material 1 Phillips-head screwdriver No. 3 1 Allen key 4 mm 1 Allen key 5 mm 4 Tubes (see fig. 17)
Resistant to the medium employed
Inside diameter 10 mm
Length: (1,4) approx. 220 mm, (2
(3) approx. 150 mm
Tab. 19
1. Attach pump to the baseplate (see chapter 7.2, page 20).
Fig. 17: Tubing system SC
2. Install the separator (see chapter 7.3, page 20).
3. Install the high performance condenser (see chapter 7.4, page
21).
4. Install the electrical supply unit and the vacuum controller (see chapter 7.5, page 21).
5. Lay hoses for system (see fig. 17).
Coolant inlet and outlet see fig. 16.
When using a coolant valve:
Danger of the high performance condenser bursting
Make sure that the coolant valve is mounted
between the coolant supply and the coolant inlet
WARNING
port of high performance condenser.
6. Wire and connect with each other the electrical supply unit and the vacuum controller (see chapter 7.5, page 21).
30 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 31
LABOPORT Pumps and Systems Installation, mounting and connection
7.7.4. System SCC
The system SCC consists of:
Baseplate
1 separator
1 high performance condenser
Electrical supply unit with 2 vacuum controllers
Tools and material
Qty Material 1 Phillips-head screwdriver No. 3 1 Allen key 4 mm 1 Allen key 5 mm 8 Tubes (see fig. 18)
Resistant to the medium employed
Inside diameter 10 mm
Length: (1) approx. 320 mm, (2) approx. 260 mm,
(3,4) approx. 120 mm, (5) approx. 220 mm, (6) approx. 135 mm, (7) approx. 175 mm, (8) approx. 350 mm
Tab. 20
1. Attach pump to the baseplate (see chapter 7.2, page 20).
Fig. 18: Tubing system SCC
2. Install the separator (see chapter 7.3, page 20).
3. Install the high performance condenser (see chapter 7.4, page
21).
4. Install the electrical supply unit and both vacuum controllers (see chapter 7.5, page 21).
5. Lay hoses for system (see fig. 18).
Coolant inlet and outlet see fig. 16.
When using a coolant valve:
Danger of the high performance condenser bursting
Make sure that the coolant valve is mounted
between the coolant supply and the coolant inlet
WARNING
port of high performance condenser.
6. Wire and connect with each other the electrical supply unit and the vacuum controller (see chapter 7.5, page 21).
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 31
Page 32
Installation, mounting and connection LABOPORT Pumps and Systems
7.7.5. Retrofitting from System SC to System SCC
The second vacuum controller is delivered complete with the vent valve installed on a new support (length: 410 mm). Install the existing vacuum controller on the new support.
Conditions
Tools and material
Pump removed from operation (see chapter 8.2, page 35)
Pump disconnected from mains
Qty Material 1 Allen key 4 mm 1 Allen key 5 mm See chapter 7.7.4 for the required hoses
Tab. 21
Removing the existing vacuum controller
The following item numbers refer to fig. 2, page 8.
1. Disconnect the tubes from the pump vent valve (5) as well as from the controller (2) vent valve (3).
2. Pull out all plugs from the electrical supply unit (1).
3. Remove the vacuum controller from the support (16) after loosening the clamp assembly.
4. Remove the lateral support from the main support after loosen­ing the clamp assembly.
5. Loosen the allen screw at the foot of the main support.
Fig. 19: Tubing system SCC
6. Slide the support out of the baseplate (17) mounting slot as­sembly.
Mount new vacuum controller
The following item numbers refer to fig. 2, page 8.
1. Slide the removed vacuum controller (2) into the slot on the new support (16) and fix it in place with the clamp connection.
2. Slide the support’s attachment part into the baseplate’s (17) lateral upward-pointing slotted bar.
3. Tighten the allen screw at the foot of the support.
4. Mount the lateral support onto the main support.
5. Slide two additional vacuum valves(6) onto the lateral support mounting slot and fix the device with help of the clamping as­sembly.
6. Lay hoses for system (see fig. 19).
7. Wire and connect with each other the electrical supply unit and the vacuum controller (see chapter 7.5, page 25).
32 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 33
LABOPORT Pumps and Systems Operation
8. Operation
8.1. Pump
8.1.1. Preparing for Start-up
Before switching on the pump/system, observe the following points:
Operational requirements Pump All hoses attached properly
Pump/ System
System Clamp connections are tight
Tab. 22
8.1.2. Starting
Only operate the pump under the operating parameters and
conditions described in chapter 5, Technical data.
Make sure the pumps are used properly (see chapter 2.1,
page 5).
Fan openings not blocked
Specifications of the power supply correspond with
the data on the pump’s/the electrical supply unit’s type plate.
The pump outlet is not closed or constricted.
When using two vacuum controllers: The media
are compatible with each other (when running two different processes simultaneously)
When using a gas ballast: No explosive or
poisonous mixtures may be produced when ventilating the vacuum system through the air inlet.
All cables attached properly
Make sure the pumps are not used improperly (see chapter
2.2, page 5).
Observe the safety precautions (see chapter 3, page 6).
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 33
Page 34
Operation LABOPORT Pumps and Systems
Only throttle or regulate the air or gas quantity in
When the operation of the pump is interrupted by the
the
event of power failure, remove the pump’s mains
Hazard of the pump head bursting due to excessive pressure increase
Do not exceed max. permissible operating
WARNING
pressure (see chapter 5.1, page 10).
Monitor pressure during operation.
If the pressure exceeds the maximum permissi-
ble operating pressure, immediately shut down pump and eliminate fault (see chapter 10, page
42).
the suction line to prevent the maximum permis­sible operating pressure from being exceeded.
If the air or gas quantity in the pressure line is
throttled or regulated, make sure that the maxi­mum permissible operating pressure of the pump is not exceeded.
Excessive pressure (with all of the related hazards) can be prevented by placing a bypass line with a pressure-relief valve between the pressure and suctions sides of the pump. For further information, contact our technical adviser.
Pump standstill
With the pump at a standstill, open pressure and suction lines
to normal atmospheric pressure.
Automatic starting can cause personal injury and pump damage
WARNING
thermal switch, the pump will restart automatically after cooling down.
After triggering of the thermal protection or in
34 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
plug from the socket so that the pump cannot start uncontrollably.
Attempt work on the pump or system only if the
pump/system is separated from mains power.
Page 35
LABOPORT Pumps and Systems Operation
on. This also applies in operation following a brief power
water connection or recirculated cooling system.
8.2. Switching Pump/System on and off
Switching pump on
The pump may not start up against overpressure during switch­failure. If a pump starts against pressure, it may block. This activates the thermal switch, and the pump switches off.
Make sure that no pressure is present in the lines during
switch-on.
Switch on pump with mains switch (see fig. 2/19)
Switching off the pump/removing from operation
When transferring aggressive media, flush the pump prior to
switch-off to increase the service life of the diaphragm (see chapter 9.2.1, page 37).
Switch off pump with mains switch (see fig. 2/19).
Open pressure and suction lines to normal atmospheric pres-
sure.
Tools and material
Disconnect the power source.
8.3. Notices related to System Operation
8.3.1. Separators
You can move the holders for the separator vessels after loosening the attachment screw.
8.3.2. High Performance Condenser
Danger of high performance condenser bursting
Make sure that the high performance
condenser’s upper gas outlet is not blocked.
CAUTION
A high performance condenser may be used only with a cold-
Adjustment of the high performance condenser altitude level
Qty Material 1 Allen key 5 mm
Tab. 23
1. Loosen the both allen screws at the foot of the high perfor­mance condenser.
2. Remove the high performance condenser’s clamp assembly in
3. Retighten the allen screws.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 35
the support mounting slot.
Page 36
Operation LABOPORT Pumps and Systems
Make sure that no reactive or explosive mixtures
of reactive, explosive or
The gas ballast is adjusted with the button
open
8.3.3. Vacuum Controller
Personal injury caused by poisoning or explosion and damage to the pump.
WARNING
will be produced when ventilating the vacuum system through the air inlet.
Make sure that the media are compatible with
each other (when running two different processes simultaneously).
Use the mains switch to switch the vacuum controller on and off.
Each vacuum controller can be set individually and will work independently of the other.
Refer to the vacuum controller’s operating instructions for more information.
8.3.4. Gas Ballast
closed
Fig. 20: Gas ballast operating
knob
Personal injury caused by poisoning or explosion and damage to the pump.
Ensure that the creation
WARNING
otherwise hazardous mixtures during the supply of air is prevented.
If the gas ballast valve is open the maximum achievable vacu­um level is reduced. (see fig. 20).
36 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 37
LABOPORT Pumps and Systems Servicing
9. Servicing
9.1. Servicing Schedule
Component Servicing interval Pump/system Regular inspection for external damage or
leaks
Diaphragm and valve plates/sealings
Tab. 24
9.2. Cleaning
When cleaning, make sure that no liquids enter the inside of the housing.
9.2.1. Flushing Pump
Before switching off the pump, flush it with air (if necessary for
safety reasons: with an inert gas) for about five minutes under atmospheric conditions (ambient pressure).
9.2.2. Cleaning Pump
Replace at the latest, when pump output decreases
Only clean pump with a damp cloth and non-flammable clean-
ing agents.
9.2.3. Emptying and Cleaning the Separator
1. Unscrew the connecting nozzles (fig. 2/8, page 8).
2. Remove separator (fig. 2/14) from the holder (fig. 2/15) and dispose of contents according to applicable regulations in your area. Then rinse out the separator.
3. Replace separator into the holder.
4. Screw the connecting nozzles back on.
9.2.4. Emptying and Cleaning the Condenser
1. Carefully remove the spring clamp (fig. 2/10) while supporting the flask (fig. 2/11).
2. Empty and clean the flask – observing safety precautions.
3. Re-connect the flask to the condenser and replace the spring clamp.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 37
Page 38
Servicing LABOPORT Pumps and Systems
9.3. Changing Diaphragm and Valve Plates/Sealings
Conditions
Tools and material
Information on procedure
Pump is switched off and mains plug is removed from the
socket
Pump is clean and free of hazardous materials
Tubes removed from pump’s pneumatic inlet and outlet
Pump is removed from baseplate (if pump is integrated into a
system)
Qty Material 1 Phillips-head screwdriver No. 2 1 Service Set (see chapter 11.1, page 46) 1 Felt-tip pen
Tab. 25
Always replace diaphragm and valve plates/sealings together
to maintain the pump performance.
With multi-head pumps, parts of the individual pump heads can be confused.
Replace the diaphragm and valve plates/sealings of the indi-
vidual pump heads consecutively.
Health hazard due to dangerous substances in the pump!
WARNING
Depending on the substance transferred, caustic burns or poisoning are possible.
Wear protective clothing if necessary, e.g.
protective gloves.
Flush pump before replacing the diaphragm and
valve plates/sealings (see chapter 9.2.1, page
37).
38 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 39
LABOPORT Pumps and Systems Servicing
mber of diaphragm spacers
Removing pump head
Pump N 842.3 FT.18 has a round shape of head, not a hexagonal.
1. For two-headed pumps: On the pneumatic head connection
(3), loosen the union nuts (2) by hand. Then slightly loosen the angle-fitting (1) in the pump head (4) by turning it anticlock­wise, so that the connecting tube can be pulled out.
2. Mark the position of top plate (fig. 24/5), head plate (fig. 24/6),
intermediate plate (fig. 24/8) and adapter relatively to each other by a drawing line (for two-headed pumps: at both pump heads) with a felt-tip pen (1). This helps to avoid incorrect as­sembly later.
Fig. 21: Removing pump head
be loosened in the next work step.
3. Loosen the outer screws (5) on the pump head/pump heads.
In version N 842.3 FT.18, twelve (instead of six) screws must
4. Carefully remove pump head / pump heads.
Change diaphragm
Fig. 22: Changing structured
diaphragm
Replace the diaphragms of two-headed pumps consecutively in order to ensure that the same nu is used as before.
1. For two-headed pumps: Push down one diaphragm until other
diaphragm is pushed upwards to its highest position.
2. Carefully unscrew the upper diaphragm (1) anti-clockwise
using both hands.
3. Replace spacer thick (2) and spacers thin (3) onto the screw
thread of the new diaphragm (same number and order).
4. Screw in the new diaphragm and tighten it by hand.
5. With a two-headed pump: Complete steps 1 through 4 for the
second pump head.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 39
Page 40
Servicing LABOPORT Pumps and Systems
Change valve plates/sealings
Replacing the valve plates/sealings of two-headed pumps consecutively.
In version N 842.3 FT.18, three (instead of one) screws must be loosened in the first work step.
1. Loosen screw(s) (1) in the center of the top plate (2). With a
two-headed pump: loosen the screw(s) from just one pump head.
2. Remove top plate (2) and head plate (3) from intermediate
plate (5). Valve plates/sealings are visible.
3. Remove old valve plate/sealings.
Fig. 23: Changing valve
plates/sealings
1
2 3
4 5
6 7
8
Fig. 24: Refitting pump head
4. Clean intermediate plate (5) carefully (if there should be depos-
its in the recesses in the intermediate plate).
5. Insert new valve plates/sealings (4) in the recesses in the
intermediate plate (5).
6. For two-headed pumps: Carry out the steps 1 to 5 for the
second pump head.
7. Dispose of the old diaphragm(s) and valve plates/sealings
properly.
Refitting pump head
1. Apply pressure all around the edge of the diaphragm. With a
two-headed pump: Apply pressure to diaphragm on only one pump.
2. Place the intermediate plate (8) with valve plates/sealings on
the adapter in accordance with the felt-tip pen marking.
3. Place the head plate (6) on the intermediate plate (8) in the
position indicated by the guide pin (7).
4. Place the top plate (5) on the head plate (6) in accordance with
the felt-tip pen marking.
5. Gently tighten screws (4) in diagonal order.
In version N 842.3 FT.18, three (instead of one) screws must be tightened in the next work step.
6. Insert screw(s) (1) with disk springs (2, 3) in the center of the
top plate (5). In doing so, make sure that the disk springs are arranged properly (see fig. 24).
7. Screw in the screw/screws (1) in the centre of the pump top
plate (5) until it is flush with the top plate (they are flush with the top plate); then screw one final half turn to tighten.
8. For two-headed pumps: Carry out steps 1 to 7 for the second
pump head.
9. For two-headed pumps: Refit the pneumatic head connection:
Place tube onto the connecting part of the angle fitting, turn angle fitting to a straight position and tighten the union nut.
40 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 41
LABOPORT Pumps and Systems Servicing
Final steps
1. Remount the pump to the baseplate (if applicable).
2. Reconnect suction and pressure line to the pump.
3. Reconnect the pump to the electricity supply.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 41
Page 42
Troubleshooting LABOPORT Pumps and Systems
Disconnect the pump power supply before working
10. Troubleshooting
Extreme danger from electrical shock!
on the pump.
DANGER
Check the pump (see Tab. 26 to 35).
10.1. Pump/System without Vacuum Controller
Pump does not transfer
Cause Fault remedy No voltage in the power source Check room fuse and switch on if necessary.
Make sure the pump is de-energized and secure.
Thermal switch has operated following to over-heating.
Connections or lines blocked. Check connections and lines.
External valve is closed or filter is clogged.
Condensate has collected in pump head.
Diaphragm or valve plates/ sealings are worn.
Flow rate, pressure or vacuum too low
The system/the pump does not achieve the output specified in the Technical data or the data sheet. Cause Fault remedy Condensate has collected in
pump head.
There is gauge pressure on pressure side and at the same time vacuum or a pressure above atmospheric pressure on suction side.
Pneumatic lines or connection parts have an insufficient cross section or they are throttled.
Leaks occur on connections, lines or pump head.
Connections or lines completely or partially jammed.
Head parts are soiled. Clean head components.
Disconnect pump from mains. Allow pump to cool. Trace cause of over-heating and eliminate it.
Remove blockage. Check external valves and filters.
Detach the condensate source from the pump. Flush pump (see chapter 9.2.1, page 37).
Replace diaphragm and valve plates/sealings (see chapter
9.3, page 38).
Tab. 26
Detach the condensate source from the pump. Flush pump (see chapter 9.2.1, page 37).
Change the pressure conditions.
Disconnect pump from system to determine output values. Eliminate throttling (e.g. valve) if necessary. Use lines or connection parts with larger cross section if
necessary.
Check that tubes sit correctly on hose nozzles. Replace leaky tubes. Eliminate leaks.
Check connections and lines. Remove the jamming parts and particles.
Diaphragm or valve plates/sealings are worn.
42 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Replace diaphragm and valve plates/sealings,
(see chapter 9.3, page 38).
Page 43
LABOPORT Pumps and Systems Troubleshooting
Flow rate, pressure or vacuum too low
The system/the pump does not achieve the output specified in the Technical data or the data sheet. Cause Fault remedy After diaphragm and valve
plates/sealings have been replaced.
Check that the spacers have been replaced onto the dia-
phragm screw thread.
Check head connection and hose connections. Possibly carefully tighten the outer screws (fig. 21/5, page
39) of the top plate crosswise.
Tab. 27
Pump is switched on, but does not run, the on/off-switch on the pump is not lit
Cause Fault remedy Pump is not connected with the
Connect pump to mains power.
power source. No voltage in the power source Fuse in the pump is defective.
Check room fuse and switch on if necessary. Remove pump’s mains plug from the socket.
Loosen marked lid on underside of the pump. Select and replace suitable fuse (see chapter 5.1, page 10).
Tab. 28
Pump is switched on, but does not run, the on/off-switch on the pump is lit
Cause Fault remedy The thermal switch has opened
due to overheating.
Remove pump’s mains plug from the socket. Allow pump to cool. Trace cause of over-heating and eliminate it.
Tab. 29
10.2. System with one Vacuum Controller
See also chapter 10.1, pump/system without vacuum controller, table 26 and 27.
Sufficient vacuum is not reached
Cause Fault remedy Solid particles in the pump vent
valve Condensate in pump head
Clean pump vent valve.
Let the pumpe run to the ultimate vacuum (Drying Mode, see
Operating Manual of Vacuum Controller).
Tab. 30
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 43
Page 44
Troubleshooting LABOPORT Pumps and Systems
Vacuum Controller is switched on and set on "Regulation Mode", the desired vacuum level is not reached, pump does not run, the on/off-switch on the pump is not lit
Cause Fault remedy Pump is not switched on. Incorrect electrical connection of
controller
Fuse in the pump is defective.
Switch on pump. Switch off system and separate from power supply.
Check cabling. Remove pump’s mains plug from the socket. Loosen marked lid on underside of the pump. Select and replace suitable fuse (see chapter 5.1, page 10).
Tab. 31
Vacuum Controller is switched on and set on "Regulation Mode", the desired vacum level is not reached, pump does not run, the on/off-switch on the pump is lit
Cause Fault remedy The thermal switch has opened
due to overheating.
Remove pump’s mains plug from the socket. Allow pump to cool. Trace cause of over-heating and eliminate it.
Tab. 32
Vacuum Controller is switched on, no LED-indication
Cause Fault remedy In the electrical supply unit the
vacuum controller plug is plugged into the socket for the
Make sure that the vacuum controller’s mains plug is plugged
into the socket (on the electrical supply unit) for vacuum controllers (see fig. 10, page 26, position 2).
pump. The fuse in the vacuum control-
ler is defective.
Switch off system and separate from power supply. Select an appropriate fuse (refer to the vacuum controller’s
operating instructions).
Replace fuse above the circuit closer/circuit breaker.
The fuse in the electrical supply unit is defect.
Switch off system and separate from power supply. Select suitable fuse (see chapter 5.2, page 17). Replace fuse on the front side, below the mains connection.
Tab. 33
Vacuum Controller shows unrealistic values
Cause Fault remedy The factory-set pressure
Contact KNF (see page 47).
equalization has changed. Sensor is defective Contact KNF (see page 47).
Tab. 34
44 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 45
LABOPORT Pumps and Systems Troubleshooting
10.3. System with two Vacuum Controllers
See also chapter 10.2, system with one vacuum controller.
System does not work, despite one Vacuum Controller (both controllers) being set on "Reg­ulation Mode"
Cause Fault remedy In the electrical supply unit the
connecting wires for the vacuum controllers are plugged in incor­rectly.
In the electrical supply unit the connecting wires for the vacuum valve are plugged incorrectly.
Exchange the plugs between vacuum controller 1 and vacu-
um controller 2.
Exchange the plugs between vacuum valve 1 and vacuum
valve 2.
Tab. 35
10.4. Fault cannot be rectified
If you are unable to determine any of the specified causes, send the pump to KNF Customer Service (see last page for the ad­dress).
1. Flush the pump to free the pump head of dangerous or ag-
gressive gases (see chapter 9.2.1, page 37).
2. Remove the pump.
3. Clean the pump (see chapter 9.2.2, page 37).
4. Send the pump to KNF with a filled out decontamination decla-
ration (see chapter 12) and specification of the medium trans­ferred.
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 45
Page 46
Ordering Information LABOPORT Pumps and Systems
11. Ordering Information
11.1. Pumps and Spare Parts
A service set contains: 1 diaphragm and 2 valve plates/sealings for one-headed
pumps
2 diaphragms and 4 valve plates/sealings for two-headed
pumps
Pump type Order-No. pump Order-No. Service Set N 810 FT.18 057986 058077 N 820 FT.18 057901 058078 N 840 FT.18 057911 058079 N 810.3 FT.18 057500 057357 N 820.3 FT.18 057501 057358 N 840.3 FT.18 057502 057359 N 840.1.2 FT.18 057571 057359 N 842.3 FT.18 057634 057359
Tab. 36
11.2. Accessories for Pump Systems
Accessory
Baseplate and mounting Baseplate and mounting Separator complete including tray High performance condenser (in-
cluding support) Gas ballast Gas ballast Vacuum controller (including support
and electrical supply unit) Vacuum controller (including sup-
port) * Two vacuum controllers (complete
including support and electrical supply unit)
Vacuum chemical tubing for pneu­matic connection** (yard ware***)
Tab. 37 * For retrofitting from system SC to SCC ** Material: Norprene A60G *** Please specify the required length (in whole meters).
Type
N 810 FT.18
N 820 FT.18
N 840 FT.18
N 810.3 FT.18
N 820.3 FT.18
N 840.3 FT.18
NP 810/820 X X X X 028129 NP 840 X X X X 028128 NR 800 X X X X X X X X 026225 NH 800
NG 810 X 028476 NG 820/840 X X X 028477 NC 800A
NC 800B
NC 800A/B
X X X X X X X X 026231
X X X X X X X X 046380
X X X X X X X X 046381
X X X X X X X X 046382
X X X X X X X X 028187
N 840.1.2 FT.18
Order-No.
N 842.3 FT.18
46 Translation of original Operating Instructions, english, KNF 121211-121345 07/13
Page 47
LABOPORT Pumps and Systems Decontamination Declaration
12. Decontamination Declaration
The condition for the repair of a pump by KNF is the certifica­tion of the customer on the transferred media and on the cleaning of the pump (decontamination declaration).
Copy this page.
Enter the pump model, the Serial No. and the transferred
media in the form below and sent the signed form together with the flushed and cleaned pump to KNF Customer Service (see last page for the address).
Customer decontamination declaration for repair order
We herewith confirm that the following media have been pumped with the pump listed below, and that the pump has been flushed and cleaned.
Pump model
Serial No.
Fed media
The pump contains neither aggressive, biological, radioactive, poisonous nor other dangerous media.
Company
Translation of original Operating Instructions, english, KNF 121211-121345 07/13 47
Date/Signature
Page 48
KNF worldwide
Benelux
Netherlands
KNF Verder B.V.
Utrechtseweg 4a NL-3451 GG Vleuten Tel. 0031 (0)30 677 92 40 Fax 0031 (0)30 677 92 47 E-mail: info@knf-verder.nl
www.knf-verder.nl
Benelux
Belgium, Luxembourg
KNF Verder N.V.
Kontichsesteenweg 17 B-2630 Aartselaar Tel. 0032 (0)3 8719624 Fax 0032 (0)3 8719628 E-mail: info@knf.be
www.knf.be
China
KNF Neuberger Trading
(Shanghai) Co., Ltd No. 36 Lane 1000 Zhang Heng Road Shanghai 201203, P.R. China Tel. 0086 (0)21 685 965 66 Fax 0086 (0)21 339 006 26 E-mail: info@knf.com.cn
www.knf.com.cn
Germany
KNF Neuberger GmbH
Alter Weg 3 D-79112 Freiburg Tel. 0049 (0)7664 5909-0 Fax 0049 (0)7664 5909-99 E-mail: info@knf.de
www.knf.de
France, Morocco, Algeria
KNF Neuberger
4, Bld. d’Alsace Z.I. F-68128 Village-Neuf Tel. 0033 (0)389 70 35 00 Fax 0033 (0)389 69 92 52 E-mail: info@knf.fr
www.knf.fr
Great Britain
KNF Neuberger U.K. Ltd.
Avenue 2 Station Lane Industrial Estate Witney Oxon OX28 4FA Tel. 0044 (0)1993 77 83 73 Fax 0044 (0)1993 77 51 48 E-mail: info@knf.co.uk
www.knf.co.uk
India
KNF Pumps + Systems (India) Pvt. Ltd.
RAJIV GANDHI INFOTECH PARK Phase 1 Ganga Estate, Survey No. 152/2/2 Above AXIS BANK Hinjewadi Pune 411 057 Tel. 0091 (0)20 640 13 923 0091 (0)20 640 08 923 Fax 0091 (0)20 229 33 923 E-mail: info@knfpumps.in
www.knfpumps.in
Italy
KNF ITALIA S.r.l.
Via Flumendosa, 10 I-20132 Milano Tel. 0039 02 27 20 38 60 Fax 0039 02 27 20 38 48 E-mail: info@knf.it
www.knf.it
Japan
KNF Japan Co.Ltd.
Across Shinkawa Bldg. Annex 3F 1-16-14 Shinkawa, Chuo-ku, Tokyo, Japan 104-0033 Tel. 0081 (0)3 3551-7931 Fax 0081 (0)3 3551-7932 E-mail: info@knf.co.jp
www.knf.co.jp
Korea
KNF Neuberger Ltd.
Woosan Bldg.RM#202, 336-4, Hwikyung-Dong Dongdaemun-Ku., 130-090, Seoul Tel. 0082 (0)2 959-0255/6 Fax 0082 (0)2 959-0254 E-mail: knf@knfkorea.com
www.knfkorea.com
Sweden, Denmark, Finland, Norway
KNF Neuberger AB
Mejerivägen 4, P.O. Box 44060 SE-10073 Stockholm Tel. 0046 (0) 87445113 Fax 0046 (0) 87445117 E-mail: info@knf.se
www.knf.se
Switzerland
Sales
KNF Neuberger AG
Stockenstrasse 6 CH-8362 Balterswil Tel. 0041 (0)71 973 993 0 Fax 0041 (0)71 973 993 1 E-mail: knf@knf.ch
www.knf.ch
Taiwan
KNF Neuberger Ltd.
9-2 FL., No., 24, Lane 123, Section 6, Ming Chuan East Road Taipei City, Taiwan Tel. 00886-2-2794-1011 Fax 00886-2-8792-1648 E-mail: knftwn@knftwn.com.tw
www.knftwn.com.tw
USA, Canada, South America
KNF NEUBERGER, INC.
Two Black Forest Road Trenton, New Jersey 08691-1810 Tel. 001 (609) 890 86 00 Fax 001 (609) 890 83 23 E-mail: knfusa@knf.com
www.knfusa.com
South America Direct Phone: 001 609 649 1010 E-mail: gb@knf.com
KNF product centres
Product centre for gas pumps:
Germany
KNF Neuberger GmbH
Alter Weg 3 D-79112 Freiburg Phone 0049(0)7664 5909-0 Fax 0049(0)7664 5909-99 E-mail: info@knf.de
www.knf.de
Product centre for fluid pumps:
Switzerland
KNF FLODOS AG
Wassermatte 2 CH-6210 Sursee Phone 0041(0)41 925 00 25 Fax 0041(0)41 925 00 35 E-mail: info@knf-flodos.ch
www.knf-flodos.ch
Product centre for micro pumps:
Switzerland
KNF Micro AG
Zelglimatte 1b CH-6260 Reiden Tel. 0041(0)62 787 88 88 Fax 0041(0)62 787 88 99 E-mail: info@knf-micro.ch
www.knf-micro.ch