KNF N 035 AVE, N 035 AN.9 E, N 035 ATE, N 035 AV.9 E, N 035 SNE Operating And Installation Instructions

...
Valid from : Aug. 2009
Diaphragm Vacuum Pumps and Compressors
N 035 ANE N 035 AN.9 E N 035 ATE N 035 AVE N 035 AV.9 E N 035 SNE N 035 STE N 035 ST.9 E N 035 SVE
Operating and installation instructions
Read and observe these operating and installation instructions!
N 035.1 ANE N 035.1 AN.9 E N 035.1 ATE N 035.1 AT.9 E N 035.1 AVE N 035.2 ANE N 035.2 AN.9 E N 035.2 ATE N 035.2 AVE N 035.3 ANE N 035.3 ATE N 035.3 AVE
N 035.1 SNE N 035.1 STE N 035.1 SVE N 035.2 SNE N 035.2 STE N 035.2 SVE N 035.3 SNE N 035.3 STE N 035.3 SVE
KNF Neuberger GmbH Alter Weg 3 D-79112 Freiburg
Germany Phone +49-(0)7664 / 5909-0
Fax +49-(0)7664 / 5909-99 E-mail: info@knf.de
www.knf.de
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions Keep for future use!
Contents Page
1.
About this document................................................................. 2
2. Use........................................................................................... 3
3. Safety ....................................................................................... 4
4. Technical Data ......................................................................... 5
5. Design and Function ................................................................ 9
6. Installation and connection..................................................... 12
7. Operation................................................................................ 19
8. Servicing................................................................................. 21
9. Troubleshooting...................................................................... 28
10. Spare parts and accessories.................................................. 30
11. Decontamination Declaration................................................. 32
Diaphragm Pump N 035 E About this document
1. About this document
1.1. Using the Operating and installation instruc­tions
The operating and installation instructions are part of the pump. Pass on the operating and installation instructions to the next
owner.
Project pumps
Customer-specific project pumps (pump models which begin with “PJ” or “PM”) may differ from the operating and installation instruc­tions.
For project pumps, also observe the agreed upon specifica-
tions.
1.2. Symbols and Markings
Warning
A danger warning is located here. Possible consequences of a failure to observe the
warning are specified here. The signal word, e.g.
WARNING
Danger levels
Signal word Meaning Consequences if not observed
DANGER
WARNING
CAUTION
Tab. 1
Other information and symbols
Warning, indicates the danger level. Measures for avoiding the danger and its con-
sequences are specified here.
warns of immedi­ate danger
warns of possible danger
warns of a possibly dangerous situa­tion
Death or serious injuries and/or serious damage are the consequence.
Death or serious injuries and/or serious damage are possible.
Minor injuries or damage are possible.
An activity to be carried out (a step) is specified here.
1. The first step of an activity to be carried out is specified here.
Additional, consecutively numbered steps follow.
This symbol refers to important information.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 2
Diaphragm Pump N 035 E Use
2. Use
2.1. Proper use
The pumps are exclusively intended for transferring gases and vapors.
Owner's responsibility
Operating parameters and
conditions
Requirements for
transferred medium
Only install and operate the pumps under the operating parameters and conditions described in Chapter 4, Technical data.
Before using a medium, check whether the medium can be trans­ferred danger-free in the specific application case.
Before using a medium, check the compatibility of the materials of the pump head, structured diaphragm and valves with the medium.
If combustible media are used: Hazard of fires and explosions due to excessively high media temperature:
Be aware that the pumps are not designed to be explosion-
proof.
Make sure the temperature of the medium is always sufficiently
below the ignition temperature of the medium, to avoid ignition or explosion. This also applies for unusual operational situa­tions.
Note that the temperature of the medium increases when the
pump compresses the medium.
Hence, make sure the temperature of the medium is suffi-
ciently below the ignition temperature of the medium, even when it is compressed to the maximum permissible operating pressure of the pump. The maximum permissible operating pressure of the pump is stated in the technical specifications (Chapter 4).
If necessary, consider any external sources of energy, such as
radiation, that may add heat to the medium. In case of doubt, consult the KNF customer service. Only transfer gases which remain stable under the pressures and
temperatures occurring in the pump.
2.2. Improper use
The pumps may not be operated in an explosive atmosphere. The pumps are not suitable for transferring dusts. The pumps are not suitable for transferring liquids.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 3
Diaphragm Pump N 035 E Safety
3. Safety
Note the safety precautions in sections 6. Installation and connection, and 7. Operation.
The pumps are built according to the generally recognized rules of technology and in accordance with the occupational safety and accident prevention regulations. Nevertheless, dangers can result during their use which lead to injuries to the user or others, or to damage to the pump or other property.
Only use the pumps when they are in a good technical and proper working order, in accordance with their intended use, observing the safety advice within the operating and installation instructions, at all times.
Personnel
Working in a safety-
conscious manner
Handling dangerous media
Environmental protection
Standards
Make sure that only trained and instructed personnel or specially trained personnel work on the pumps. This especially applies to assembly, connection and servicing work.
Make sure that the personnel has read and understood the operat­ing and installation instructions, and in particular the "Safety" chapter.
Observe the accident prevention and safety regulations when performing any work on the pump and during operation.
When transferring dangerous media, observe the safety regula­tions when handling these media.
Store all replacement parts in a protected manner and dispose of them properly in accordance with the applicable environmental protection regulations. Observe the respective national and inter­national regulations. This especially applies to parts contaminated with toxic substances.
The pumps conform to the safety regulations of the EC Low Volt­age Directive 2006/95 EC, and the EC Directive 2004/108 EC concerning Electromagnetic Compatibility.
The following harmonized standards have been used:
IP20 IP44 EN 55014 EN 55014 EN 61000-3-2/3 EN 61000-3-2/3 EN 60335-1 EN 60204-1
Tab. 2
Customer service and
repairs
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 4
Only have repairs to the pump carried out by the KNF Customer Service responsible.
Use only genuine parts from KNF for servicing work.
Diaphragm Pump N 035 E Technical Data
4. Technical Data
Pump materials
Material* Pump type Pump head Diaphragm Valve Gasket
N 035 ANE N 035 AN.9 E N 035.1 ANE N 035.1 AN.9 E N 035.2 ANE N 035.2 AN.9 E N 035.3 ANE N 035 ATE N 035.1 ATE N 035.1 AT.9 E N 035.2 ATE N 035.3 ATE N 035 AVE N 035 AV.9 E N 035.1 AVE N 035.2 AVE N 035.3 AVE N 035 SNE N 035.1 SNE N 035.2 SNE N 035.3 SNE N 035 STE N 035 ST.9 E N 035.1 STE N 035.2 STE N 035.3 STE N 035 SVE N 035.1 SVE N 035.2 SVE N 035.3 SVE
Aluminium CR Stainless Steel CR
Aluminium PTFE-coated Stainless Steel FPM
Aluminium FPM Stainless Steel FPM
Stainless Steel CR CR -
Stainless Steel PTFE-coated PTFE -
Stainless Steel FPM FPM -
Tab. 3 *according to DIN ISO 1629 und 1043.1
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 5
Diaphragm Pump N 035 E Technical Data
Pneumatic values
Pump type Delivery rate* (l/min)
at atm. pressure
N 035 ANE IP20 N 035 AN.9 E IP20 N 035 ANE IP44 N 035 AN.9 E IP44
30
N 035 ANE IP44 3 phase N 035 SNE IP20 N 035 SNE IP44 N 035 ATE IP20 N 035 ATE IP44 N 035 STE IP20 N 035 STE IP44
27
N 035 ST.9 E IP44 N 035 STE IP44 3 phase N 035 AVE IP20 N 035 AVE IP44 N 035 AV.9 E IP44 N 035 AVE IP44 3 phase
30
N 035 AV.9 E IP44 3 phase N 035 SVE IP20 N 035 SVE IP44 N 035.1 ANE IP 20 N 035.1 AN.9 E IP 20 N 035.1 ANE IP 44 N 035.1 AN.9 E IP 44 N 035.1 ANE IP 44 3 phase N 035.1 AVE IP 20 N 035.1 AVE IP 44 N 035.1 AVE IP 44 3 phase
55 - 100
N 035.1 SNE IP 20 N 035.1 SNE IP 44 N 035.1 SNE IP 44 3 phase N 035.1 SVE IP 20 N 035.1 SVE IP 44 N 035.1 SVE IP 44 3 phase N 035.1 ATE IP 20 N 035.1 ATE IP 44 N 035.1 AT.9 E IP 44 N 035.1 ATE IP 44 3 phase
49.5 - 100
N 035.1 STE IP 20 N 035.1 STE IP 44
Tab. 4 (part 1) *Liters in standard state (1,013 mbar)
Max. permissible operating pressure (bar)
4
4
2
Ultimate vacuum (mbar abs.)
100
100
100
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 6
Diaphragm Pump N 035 E Technical Data
Pump type
N 035.2 ANE IP 20 N 035.2 ANE IP 44 N 035.2 AN.9 E IP 44 N 035.2 ANE IP 44 3 phase N 035.2 SNE IP 20 N 035.2 SNE IP 44 N 035.2 ATE IP 20 N 035.2 ATE IP 44 N 035.2 STE IP 20 N 035.2 STE IP 44 N 035.2 AVE IP 20 N 035.2 AVE IP 44 N 035.2 AVE IP 44 3 phase N 035.2 SVE IP 20 N 035.2 SVE IP 44 N 035.3 ANE IP 20 N 035.3 ANE IP 44 N 035.3 ANE IP 44 3 phase N 035.3 AVE IP 20 N 035.3 AVE IP 44 N 035.3 SNE IP 20 N 035.3 SNE IP 44 N 035.3 SVE IP 20 N 035.3 SVE IP 44 N 035.3 SVE IP 44 3 phase N 035.3 ATE IP 44 N 035.3 STE IP 44
Delivery rate* (l/min) at atm. pressure
55 4 -
49.5 4 -
55 2 -
30 - 13
27 - 20
Tab. 4 (part 2) *Liters in standard state (1,013 mbar)
Max. permissible operat­ing pressure (bar)
Ultimate vacuum (mbar abs.)
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 7
Diaphragm Pump N 035 E Technical Data
Electrical data
Parameter Value one-
headed pumps
Voltage / Frequency of
230 V / 50 Hz 230 V / 50 Hz
Value two­headed pumps
AC motor Voltage / Frequency of 3 phase motor
230/400 V 50 Hz
230/400 V
50 Hz Power IP20 version 220 W 300 W Power IP44 version capacitor
230 W 320 W
motor Power IP 44 version 3 phase
250 W 280 W
motor Operating current
1 A 1.55 A
IP20 version Operating current
1.7 A 1.90 A
IP44 version Operating current IP44 ver-
1.7 A/1.0 A 2.1 A / 1.05 A
sion threephase motor Protection class* IP20 / IP44 IP 20 / IP 44
Tab. 5 * see type plate
The pumps are fitted as standard with a thermal-switch to protect against overloading.
Other parameters
Parameter Values Permissible
ambient temperature
+ 5 °C to + 40 °C
Permissible media temperature + 5 °C to + 40 °C Gas-tightness of pump head
ca. 6 x 10-3 mbar l/s (leak rate)* for all pumps except .9 versions (not tested)
Gas-tightness of pump head
< 6 x 10-3 mbar l/s (leak rate)* for N 035_ _.9 E
Tab. 6
* After opening pump head or replacing the diaphragm and reed valves (or
valve plate) the gas tightness is no longer guaranteed. A leak test is able to verify that the original standard of gas-tightness has been achieved.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 8
Diaphragm Pump N 035 E Design a nd Function
5. Design and Function
Design N 035 A__E (IP20)
1 Pneumatic pump outlet 2 Pneumatic pump inlet 3 Electrical Connection 4 Motor
1 Pneumatic pump outlet 2 Pneumatic pump inlet 3 Terminal box (electrical
connetion)
4 Motor
Fig. 1: Diaphragm Pump N 035 ANE (IP20)
Design N 035 A__E (IP44)
Fig. 2: Diaphragm Pump N 035 ANE (IP44)
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 9
Diaphragm Pump N 035 E Design a nd Function
Fig. 3: Pneumatic connection of two-headed pumps
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 10
Diaphragm Pump N 035 E Design a nd Function
Function diaphragm pump
1 Outlet valve 2 Inlet valve 3 Transfer chamber 4 Diaphragm 5 Eccentric 6 Connecting rod 7 Pump drive
Fig. 4: Pump head
The pump transfers, compresses (depending on pump version) and evacuates gases and vapors.
The elastic diaphragm (4) is moved up and down by the eccentric (5) and the connecting rod (6). In the downward stroke it aspirates the gas to be transferred via the inlet valve (2). In the upward stroke, the diaphragm presses the medium out of the pump head via the outlet valve (1). The transfer chamber (3) is hermetically separated from the pump drive (7) by the diaphragm.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 11
Diaphragm Pump N 035 E Installation and connection
6. Installation and connection
Only install and operate the pumps under the operating parameters and conditions described in chapter 4, Technical data.
Observe the safety precautions (see chapter 3).
6.1. Installation
Before installation, store the pump at the installation location to
bring it up to room temperature.
Mounting dimensions
Mounting dimensions (see figs. 5 to 16).
Fig. 5: Mounting dimensions N 035 A_E (IP20) including .9 versions
(All dimensional tolerances conform to DIN ISO 2768-1, Tolerance Class V)
Fig. 6: Mounting dimensions N 035 A_E (IP44) including .9 versions
(All dimensional tolerances conform to DIN ISO 2768-1, Tolerance Class V)
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 12
Diaphragm Pump N 035 E Installation and connection
Fig. 7: Mounting dimensions N 035 S_E (IP20)
(All dimensional tolerances conform to DIN ISO 2768-1, Tolerance Class V)
Fig. 8: Mounting dimensions N 035 S_E (IP44) including .9 versions
(All dimensional tolerances conform to DIN ISO 2768-1, Tolerance Class V)
Fig. 9: Mounting dimensions N 035.1 A_E (IP20) and N 035.3 A_E (IP20)
including .9 versions (All dimensional tolerances conform to DIN ISO 2768-1, Tolerance Class V)
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 13
Diaphragm Pump N 035 E Installation and connection
Fig. 10: Mounting dimensions N 035.1 A_E (IP44) and N 035.3 A_E (IP44)
including .9 versions (illustration without pneumatic connection; all dimensional tolerances conform to DIN ISO 2768-1, Tolerance Class V)
Fig. 11: Mounting dimensions N 035.2 A_E (IP20)
(All dimensional tolerances conform to DIN ISO 2768-1, Tolerance Class V)
Fig. 12: Mounting dimensions N 035.2 A_E (IP44) including .9 versions
(All dimensional tolerances conform to DIN ISO 2768-1, Tolerance Class V)
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 14
Diaphragm Pump N 035 E Installation and connection
Fig. 13: Mounting dimensions N 035.1 S_E (IP20) and N 035.3 S_E (IP20)
(illustration without pneumatic connection; all dimensional toler­ances conform to DIN ISO 2768-1, Tolerance Class V)
Fig. 14: Mounting dimensions N 035.1 S_E (IP44) and N 035.3 S_E (IP44)
(All dimensional tolerances conform to DIN ISO 2768-1, Tolerance Class V)
Fig. 15: Mounting dimensions N 035.2 S_E (IP20)
(illustration without pneumatic connection; all dimensional toler­ances conform to DIN ISO 2768-1,Tolerance Class V)
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 15
Diaphragm Pump N 035 E Installation and connection
Fig. 16: Mounting dimensions N 035.2 S_E (IP44)
(All dimensional tolerances conform to DIN ISO 2768-1, Tolerance Class V)
Cooling air supply
Install the pump so that the motor fan can intake sufficient
cooling air.
Installation location
Make sure that the installation location is dry and the pump is
protected against rain, splash, hose and drip water.
Install the pump at the highest point in the system to prevent
condensate from collecting in the pump head.
Protect the pump from dust. Protect the pump from vibrations and jolts.
6.2. Electrical connection
Extreme danger from electrical shock Only have the pump connected by an authorized
specialist.
DANGER
When connecting the device to a power source, the relevant
standards, directives, regulations, and technical standards must be observed.
In the electrical installation, arrangements (complying with EN
60335-1) must be made for disconnecting the pump motor from the electrical supply.
Only have the pump connected when the power
supply is disconnected.
KNF recommends that a fuse is installed in the motor supply
circuit (overcurrent release).
For operating current see type plate or data sheet.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 16
Diaphragm Pump N 035 E Installation and connection
Connecting pump
1. Compare the supply data with the data on the motor-plate. For operating current see type plate.
The voltage must not vary by more than + 10% and - 10% from that shown on the type-plate.
2. For IP 44 versions: open terminal box cover.
3. All pumps except versions with 3 phase motor: Connect the mains cables to the connections L1 and N of the pump motor.
4. Connection of pumps with 3 phase motor according to figs. 17 or 18.
5. Connect the earth (ground) wire to the motor.
6. For IP 44 versions: close the terminal cover box.
Fig. 17: Y-Connection (high voltage)
Fig. 18: -Connection (low voltage)
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 17
Diaphragm Pump N 035 E Installation and connection
6.3. Pneumatic connection
Connected components
Pump exhaust
Only connect components to the pump which are designed for
the pneumatic data of the pump (see section 4).
If the pump is used as a vacuum pump, safely discharge the
pump exhaust at the pump’s pneumatic outlet.
Connecting pump
A marking on the pump head shows the direction of flow. For two-headed pumps fig. 3 shows the pneumatic connections.
1. Remove the protective plugs from the hose connection threads.
2. The assecorries silencer, filter, and hose connectors (where applicable) are screwed into the port threads.
If the pump is used as a vacuum pump (not permitted with series N 035.2), mount the silencer at the pressure side if necessary. If the pump is used a compressor (not permitted with series N 035.1 and N 035.3), mount the filter at the suc­tion side if necessary.
3. Connect the suction line and pressure line (thread size G ¼)
4. Lay the suction and pressure line at a downward angle to prevent condensate from running into the pump.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 18
Diaphragm Pump N 035 E Operation
7. Operation
Only operate the pump under the operating parameters and
conditions described in chapter 4, Technical data.
Make sure the pumps are used properly (see section 2.1). Make sure the pumps are not used improperly (see section
2.2).
Observe the safety precautions (see chapter 3).
Hazard of the pump head bursting due to excessive pressure increase
Do not exceed max. permissible operating
WARNING
pressure (see section 4).
Monitor pressure during operation. If the pressure exceeds the maximum permissi-
ble operating pressure, immediately shut down pump and eliminate fault (see chapter 9. Trou­bleshooting).
Only throttle or regulate the air or gas quantity in
the suction line to prevent the maximum permis­sible operating pressure from being exceeded.
Pump standstill
If the air or gas quantity in the pressure line is
throttled or regulated, make sure that the maxi­mum permissible operating pressure of the pump is not exceeded.
Excessive pressure (with all of the related hazards) can be prevented by placing a bypass line with a pressure-relief valve between the pressure and suctions sides of the pump. For further information, contact our technical adviser.
With the pump at a standstill, open pressure and suction lines
to normal atmospheric pressure.
For pumps with thermo switch (special design):
Automatic starting can cause personal injury and pump damage
When the operation of the pump is interrupted by
WARNING
the thermal switch, the pump will restart automati­cally after cooling down.
Take all necessary care to prevent this leading
to a dangerous situation.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 19
Diaphragm Pump N 035 E Operation
Switching pump on
The pump may not start up against pressure or vacuum during switch-on. This also applies in operation following a brief power failure.
Make sure that no pressure is present in the lines during
switch-on.
Switching off the pump
KNF recommends: When transferring aggressive media, flush
the pump prior to switch-off to increase the service life of the diaphragm (see section 8.2.1).
Open pressure and suction lines to normal atmospheric pres-
sure.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 20
Diaphragm Pump N 035 E Servicing
8. Servicing
8.1. Servicing Schedule
Component Servicing interval Pump Regular inspection for external damage or
leaks
Diaphragm and valve plates or reed valves
Silencer/filter (ac­cessory)
Tab. 7
8.2. Cleaning
When cleaning, make sure that no liquids enter the inside of the housing.
8.2.1. Flushing Pump
Replace at the latest, when pump output decreases
Change if it is dirty
When transferring aggressive media, flush the pump under atmos­pheric conditions some minutes with air (or, if necessary for safety reasons, with an inert gas) prior to switch-off to increase the ser­vice life of the diaphragm.
8.2.2. Cleaning Pump
Only use solvents for cleaning if the head materials cannot be
attacked (check the resistance of the material!).
If compressed air is available, blow out the components.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 21
Diaphragm Pump N 035 E Servicing
8.3. Changing Diaphragm and Valves
8.3.1. Pumps with aluminium head
N 035 ANE N 035.1 ANE N 035.2 ANE N 035.3 ANE N 035 AN.9 E N 035.1 AN.9 E N 035.2 AN.9 E N 035.3 ATE N 035 ATE N 035.1 ATE N 035.2 ATE N 035.3 AVE N 035 AVE N 035.1 AT.9 E N 035.2 AVE N 035 AV.9 E N 035.1 AVE
Conditions
Spare parts
Tools
Pump is switched off and mains plug is removed from the
socket
Pump is clean and free of hazardous materials
Spare part* Position** Quantity per pump head Diaphragm (F) 1 Countersunk
(D) 1
screw*** Reed valve (M,P) 2 Gasket (V) 1
Tab. 8 * According to Spare parts list, chapter 10
** According to Fig. 19
***Not for .9 versions
Quantity Tools/Material 1 Allen key 3 mm 1 Allen key 4 mm 1 Allen key 5 mm 1 Screwdriver blade width 6.5 1 Screwdriver blade width 4.0 1 Fork wrench 16 mm (only for two-headed pumps) 1 Pencil 1 Adjustable pin–wrench for two-hole nuts or KNF
wrench for retainer plate (see accessory, section 10) (only for .9 versions)
Tab. 9
Information on procedure
With multi-head pumps, parts of the individual pump heads can be confused.
Replace the diaphragm and reed valves of the individual pump
heads consecutively.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 22
Diaphragm Pump N 035 E Servicing
Health hazard due to dangerous substances in the pump!
Depending on the substance transferred, caustic
WARNING
burns or poisoning are possible. Wear protective clothing if necessary, e.g.
protective gloves.
Flush pump before replacing the diaphragm and
reed valves (see section 8.2.1).
*
Fig. 19: Pump parts for versions with aluminium head
*not for .9 versions
1. For two-headed pumps: At one pump head open the union nut of pneumatic head con­nection and pull off the tube.
2. Mark the position of the diaphragm head C in relation of the housing A with a pencil.
3. Loosen the four allen screws B and remove the diaphragm head C.
4. For all pumps except .9 versions: Unscrew the countersunk screw D, remove the retainer plate E and the diaphragm F.
5. For pumps N 035_ _.9 E: To undo the retainer plate E use the wrench for retainer plate to turn it anti-clockwise; remove re­tainer plate and diaphragm F.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 23
Diaphragm Pump N 035 E Servicing
6. Loosen the four screws G and remove the cover plate H.
7. Turn the counterweight J so that the connection rod K is in the mid-position; fit the new diaphragm F.
8. For all pumps except .9 versions: Place the retainer plate E on the diaphragm F and tighten the new countersunk screw D (torque: 5.0 Nm).
The self-locking screw D can only be used once.
9. For pumps N 035_ _.9 E: Place the retainer plate E on the diaphragm F. Screw on the retainer plate E with the wrench for retainer plate uniformly and diagonally (torque: 5.0 Nm).
10. Change lower reed valve:
Undo the cheese head screw N and exchange the reed
valve M.
11. Change upper reed valve:
Loosen the allen screws S, remove the cover plate T and
the gasket V.
Undo the cheese head screw U and exchange the reed
valve P; tighten the cheese head screw U.
Replace the cover plate T with a new gasket V and tighten
the allen screws S.
12. Place the diaphragm head C on the diaphragm F according to the marks made previously and tighten the screws B uniformly and diagonally (torque: 10.0 Nm).
13. Turn the counterweight J to check that the pump run freely, replace the cover plate H and secure it with the four screws G.
14. For two-headed pumps: Carry out steps 2 to 13 for the second pump head.
15. For two-headed pumps: Reattach the tube of pneumatic head connection onto the hose connector and tighten the union nut.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 24
Diaphragm Pump N 035 E Servicing
8.3.2. Pumps with stainless steel head
N 035 SNE N 035.1 SNE N 035.2 SNE N 035.3 SNE N 035 STE N 035.1 STE N 035.2 STE N 035.3 STE
Conditions
Spare parts
Tools
Information on procedure
N 035 ST.9 E
N 035.1 SVE N 035.2 SVE N 035.3 SVE
N 035 SVE
Motor disconnected from mains and de-energized Pump is clean and free of hazardous materials
Spare part* Position** Quantity per
pump head Diaphragm (F) 1 Countersunk
(D) 1
screw*** Valve plate (Z) 1
Tab. 10 * According to Spare parts list, chapter 10
** According to Fig. 20
***Not for .9 versions
Quantity Tools/Material 1 Allen key 4 mm 1 Allen key 5 mm 1 Screwdriver blade width 6.5 1 Pencil 1 Adjustable pin–wrench for two-hole nuts or KNF
wrench for retainer plate (see accessory, section 10) (only for .9 versions)
Tab. 11
With multi-head pumps, parts of the individual pump heads can be confused.
Replace the diaphragm and valve plate of the individual pump
heads consecutively.
Health hazard due to dangerous substances in the pump!
Depending on the substance transferred, caustic
WARNING
burns or poisoning are possible. Wear protective clothing if necessary, e.g.
protective gloves.
Flush pump before replacing the diaphragm and
the valve plate (see section 8.2.1).
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 25
Diaphragm Pump N 035 E Servicing
*
Fig. 20: Pump parts for versions with stainless steel head
*not for .9 versions
1. For pumps N 035.1 S_E and N 035.3 S_E: Pull the pneumatic head connection hose off one pump head.
2. For pumps N 035.2 S_E: On the pneumatic connection, loosen the hose clip on one pump head and pull the hose off.
3. Mark the position of the head plate W and intermediate plate X in relation of the housing A with a pencil.
4. Loosen the four allen screws Y and remove the head plate head W, valve plate Z and intermediate plate X.
5. For all pumps except .9 versions: Unscrew the countersunk screw D, remove the retainer plate E and the diaphragm F.
6. For pumps N 035_ _.9 E: To undo the retainer plate E use the wrench for retainer plate to turn it anti-clockwise; remove re­tainer plate and diaphragm F.
7. Loosen the four screws G and remove the cover plate H.
8. Turn the counterweight J so that the connection rod K is in the mid-position; fit the new diaphragm F.
9. For all pumps except .9 versions: Place the ratainer plate E on the diaphragm F and tighten the new countersunk screw D (torque: 5.0 Nm).
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 26
Diaphragm Pump N 035 E Servicing
The self-locking screw D can only be used once.
10. For pumps N 035_ _.9 E: Place the retainer plate E on the diaphragm F. Screw on the retainer plate E with the wrench for retainer plate uniformly and diagonally (torque: 5.0 Nm).
11. Place the intermediate plate X on the top of the diaphragm F so that it corresponds to the marks on the housing.
12. Place the new valve plate Z on the intermediate plate X.
13. Place the head plate head W on the diaphragm F according to the marks made previously and tighten the screws Y uniformly and diagonally (torque: 10.0 Nm).
14. Turn the counterweight J to check that the pump run freely, replace the cover plate H and secure it with the four screws G.
15. For two-headed pumps: Carry out steps 3 to 14 for the second pump head.
16. For two-headed pumps: Pull the pneumatic head connection hose back onto the hose connector.
17. For pump type N 035.2 S_E: Retighten the hose clip on the pneumatic head connection.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 27
Diaphragm Pump N 035 E Troubleshooting
9. Troubleshooting
Extreme danger from electrical shock! Disconnect the pump power supply before working
on the pump.
DANGER
Check the pump (see Tab. 12 and 13).
Pump does not transfer
Cause Fault remedy No voltage in the power source
Connections or lines blocked.  Check connections and lines.
External valve is closed or filter is clogged.
Condensate has collected in pump head.
Diaphragm or reed valves (valve plate) are worn.
Flow rate, pressure or vacuum too low
Check room fuse and switch on if necessary.
Remove blockage. Check external valves and filters.
Flush pump (see Section 8.2.1). Install pump at highest point in system. Replace diaphragm and reed valves (valve plate), (see
Section 8.3).
Tab. 12
Make sure the pump is de-energized and secure.
The pump does not achieve the output specified in the Technical data or the data sheet. Cause Fault remedy Condensate has collected in
pump head. There is gauge pressure on
pressure side and at the same time vacuum or a pressure above atmospheric pressure on suction side.
Pneumatic lines or connection parts have an insufficient cross section.
Leaks occur on connections, lines or pump head.
Connections or lines completely or partially jammed.
Head parts are soiled.  Clean head components. Diaphragm or reed valves
(valve plate) are worn.
Flush pump (see Section 8.2.1). Install pump at highest point in system.
Change the pressure conditions.
Disconnect pump from system to determine output values. Eliminate throttling (e.g. valve) if necessary. Use lines or connection parts with larger cross section if
necessary.
Eliminate leaks.
Check connections and lines. Remove the jamming parts and particles.
Replace diaphragm and reed valves (valve plate),
(see Section 8.3).
Tab. 13
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 28
Diaphragm Pump N 035 E Troubleshooting
Fault cannot be rectified
If you are unable to determine any of the specified causes, send the pump to KNF Customer Service (see first page for the ad­dress).
1. Flush the pump to free the pump head of dangerous or ag­gressive gases (see Section 8.2.1).
2. Remove the pump.
3. Clean the pump (see Section 8.2.2).
4. Send the pump to KNF with a filled out decontamination decla­ration (see chapter 11) and specification of the medium trans­ferred.
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 29
Diaphragm Pump N 035 E Spare parts and accessories
10. Spare parts and accessories
Spare parts
N 035 ANE, N 035.1 ANE, N 035.2 ANE, N 035.3 ANE
Spare part Position* Order No. Diaphragm (F) Countersunk screw** (D) Reed valve (M, P) Gasket (V)
Tab. 14 * According to Fig. 19
N 035 AVE, N 035.1 AVE, N 035.2 AVE, N 035.3 AVE
Spare part Position* Order No. Diaphragm (F) Countersunk screw** (D) Reed valve (M, P) Gasket (V)
Tab. 15 * According to Fig. 19
001312 110711 001328 001326
** Not for .9 versions
001405 110711 001328 011796
** Not for .9 versions
N 035 ATE, N 035.1 ATE, N 035.2 ATE, N 035.3 ATE
Spare part Position* Order No. Diaphragm (F) Countersunk screw** (D) Reed valve (M, P) Gasket (V)
Tab. 16 * According to Fig. 19
001406 110711 001328 011796
** Not for .9 versions
N 035 SNE, N 035.1 SNE, N 035.2 SNE, N 035.3 SNE
Spare part Position* Order No. Diaphragm (F) Countersunk screw** (D) Valve plate (Z)
Tab. 17 * According to Fig. 20
001312 110711 001308
** Not for .9 versions
N 035 SVE, N 035.1 SVE, N 035.2 SVE, N 035.3 SVE
Spare part Position* Order No. Diaphragm (F) Countersunk screw** (D) Valve plate (Z)
Tab. 18 * According to Fig. 20
001405 110711 001404
** Not for .9 versions
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 30
Diaphragm Pump N 035 E Spare parts and accessories
N 035 STE, N 035.1 STE, N 035.2 STE, N 035.3 STE
Spare part Position* Order No. Diaphragm (F) Countersunk screw** (D) Valve plate (Z) 001407
Tab. 19 * According to Fig. 20
001406 110711
** Not for .9 versions
Accessories
Accessory for pump type Order No. Silencer/filter (G ¼) all 000352 Pressure relief valve
N 035 ANE 047601
4 bar Fine control valve with pressure
gauge, pressure side Fine control valve with vacuum
gauge, suction side
N 035 ANE N 035.2 ANE
N 035 ANE N 035.1 ANE
000482
000354
N 035.3 ANE
Hose connector G1/4 for tube ID 9
Hose connector stainless steel G1/4
all
all
000362
020234
for tube ID 9 Wrench for retainer plate .9 versions 018812
Tab. 20
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 31
Diaphragm Pump N 035 E Decontamination Declaration
11. Decontamination Declaration
The condition for the repair of a pump by KNF is the certifica­tion of the customer on the transferred media and on the cleaning of the pump (decontamination declaration).
Copy this page.
KNF Neuberger GmbH Alter Weg 3 D-79112 Freiburg
Germany Tel. +49-(0)7664-5909-0
Fax +49-(0)7664- 5909-99 E-mail: info@knf.de
www.knf.de
Enter the pump model, the Serial No. and the transferred
media in the form below and sent the signed form together with the flushed and cleaned pump to KNF Customer Service.
Customer decontamination declaration for repair order
We herewith confirm that the following media have been pumped with the pump listed below, and that the pump has been flushed and cleaned.
Pump model
Serial No.
Fed media
The pump contains neither aggressive, biological, radioactive, poisonous nor other dangerous media.
Company Date/Signature
KNF N 035 E e 08/09 Translation of the original Operating and installation instructions 32
Loading...