Knauer Azura DAD 6.1L, Azura MWD 2.1L, Azura DAD 2.1L User Manual

Detector DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual
Detektor DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch
V6700
HPLC

Table of Contents

Note For your own safety, read the manual and always observe the
warnings and safety information on the device and in the manual!
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Device Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Side View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Solvents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Scope of Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Safety for Users . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Definition of Personal and Material Damage . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Decontamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Symbols and Signs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Contact Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Installation Site and Space Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Operating Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Space Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Inserting the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Inserting the Flow Cell with Fiber Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Connecting the Capillaries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Using PEEK Fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Connecting the Leak Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Connecting the Device to the Computer . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Configuring the LAN Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Connecting the Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Configuring the Router . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Integrating the LAN into a Company Network . . . . . . . . . . . . . . . . . . . . . 25
Controlling Several Systems Separately in a LAN . . . . . . . . . . . . . . . . . . . . 25
Pin Header Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Wiring the Pin Header . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Integrator Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Integrating the Detector into a System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Pre-Installed Capillaries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Initial Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Switch-On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Operation with Chromatography Software . . . . . . . . . . . . . . . . . . . . . . . . . 32
Control with Control Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Control with Mobile Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Meaning of the LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Re-Setting the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Optimizing the Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Warm-Up Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Selecting the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Flow Cell Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Path Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Pressure Stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Remote Operation (Fiber Optics) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Selecting the Wavelength . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Spectral Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Time Constant and Data Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Integration time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Subtraction of the Baseline Chromatogram . . . . . . . . . . . . . . . . . . . . . . . . . 41
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Functionality Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Installation Qualification (IQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Operation Qualification (OQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
LAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Possible Problems and Solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
System Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Maintenance and Care. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Contact the Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Maintenance Contract . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Maintenance Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Cleaning and Caring for the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Checking the Fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Putting the Device out of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Disconnect from Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Cleaning the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Rinsing the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Replacing the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Replacing the Lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Removing the Deuterium or Halogen Lamp . . . . . . . . . . . . . . . . . . . . . . . . 53
Installing the Deuterium or Halogen Lamp . . . . . . . . . . . . . . . . . . . . . . . . . 55
Removing a Leak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
DAD 6.1L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
DAD 2.1L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
MWD 2.1L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Conditions for Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Legal Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Warranty Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Transport Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
HPLC Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Repeat Orders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Accessories and Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Flow Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Flow Cells with Fiber Optics Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Fiber Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Table of Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
To whom it may concern In case you prefer a French language user manual for this prod-
uct, submit your request including the corresponding serial number via email or fax to KNAUER:
support@knauer.net+49 30 8015010
Thank you.
A qui que ce soit Si jamais vous préfériez un manuel en francais pour ce poduit
contacter KNAUER par email ou par fax avec le no. de série:
support@knauer.net+49 30 8015010
Merci beaucoup.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
6Intended Use
8
7
2
3 2
2
4
1
6
5

Intended Use

Note Only use the device for applications that fall within the range of
the intended use. Otherwise, the protective and safety equip­ment of the device could fail.
Description The detector is suitable for analytical, semipreparative, and pre-
parative applications.
Scope of application The device can be used in the following areas:
Biochemistry analysisChemical analysisFood analysisPharmaceutical analysisEnvironmental analysis

Device Overview

Legend
1 Leak tray
2 Capillary guide
3 Lamp cover
4 Serial number
5 Leak Management
6 Switch/Standby button
7 Status LED
8 Flow cell
Front View
On the front of the device, you find the flow cell and a lamp cover for replacing the lamp.
Fig. 1 Front view (example DAD 6.1L)
Rear View
On the back of the detector, the connectors for LAN and power plug are located as well as the serial number of the device and the power switch (on/off).
RS-232, service interface Please note that the service interface is solely used for repair and
maintenance tasks performed by service technicians.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Legend
2345
6
71
1
1 Device information
2 Integrator outlets
3 LAN connector
4 Service Interface
5 Pin header
6 Fan
7 Connection and
power switch
7Intended Use
Fig. 2 Rear view (example DAD 6.1L)
Side View
DAD 6.1L On the side of the detector, you find the connector for the oper-
Legend
1 USB port for the
Control Unit
ating panel Control Unit (see See “Control with Mobile Control” on page 32.).
Note The USB ports of the DAD 2.1L and the MWD 2.1L are not
active. These detectors cannot be operated with the Control Unit.
Fig. 3 USB port for the Control Unit

Features

In an HPLC system, the detector serves to detect substances in liquids and show their concentration. The sensitivity of the detector depends on the installed flow cell.
Optical system The detectors measure the light absorption of the sample in the
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
ultraviolet and visual spectra. The entire spectrum permeates the sample and afterwards is split up by a grate. The split light falls onto a geometric array of 1024 (DAD 6.1L) or 256 (DAD 2.1L, MWD 2.1L) separate photodiodes – the photodiode array.
Legend
3
5
1
2
4
6
1 Deuterium lamp
2 Halogen lamp
3 Flow cell
4 Diode array sensor
5 Diffraction grating
6 Mirror
8Intended Use
The optical system of the AZURA DAD 6.1L detector is depicted in figure 4. An active temperature control with 0.5°C sensitivity ensures baseline stability for minimal baseline drift.
Fig. 4 Optical system DAD 6.1L
Lamps The lamps can be exchanged quickly and easily without the
need of a service technician.
DAD 6.1L Two light sources, a deuterium lamp and a halogen lamp, cover
the entire UV-Vis wavelength range of remove space 190–1020 nm.
DAD 2.1L, MWD 2.1L A deuterium lamp covers the UV wavelength range of 190–500
nm.
Lamp mirror
DAD 6.1L
Flow cell Various types of flow cells are available to the user and allow a
KNAUER LightGuide
flow cell cartridge
Data rate The detector measures with a maximum data rate of 100 Hz.
Maximum light intensity is provided by effectively mixing the light sources from the deuterium and halogen lamps with a mir­ror using Polka Dot technology.
variety of applications. Flow cell cartridges with standard and KNAUER LightGuide technology are available.
KNAUER LightGuide flow cell cartridges combine a maximum light transmission (using total reflection) with a minimum cell volume to offer an ideal signal-to-noise ratio.
Adapter for fiber optics are additionally available. Fiber optics allow separating the flow cell from the detector. This way, work­ing with dangerous, explosive, or toxic materials becomes safer.
During the measurement, approx. 100 spectra per second are recorded.
Leak Management Integrated leak management drains liquids resulting from leaks.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
9Intended Use
GLP data The Mobile Control and the different software products (e.g.
ClarityChrom
®
or OpenLAB®) can be used to display or read out the GLP data (e.g. operating hours, lamp operating hours, num­ber of lamp ignitions, etc.) of the detector. For the DAD 6.1L, the Control Unit is available for reading out the GLP data. You find a detailed description on how to display or read out GLP data in the respective user manuals.
Optional accessories The following optional accessories are available:
The Mobile Control app for control on Android and
Windows 8
The Control Unit for operating the DAD 6.1L

Solvents

Even small quantities of other substances, such as additives, modifiers, or salts can influence the durability of the materials.
If there is any doubt, contact the Technical Support of the man­ufacturer.
Suitable solvents
Acetate buffer solutionsAcetone at 4 °C–25 °C (39.2 °F–77.0 °F)Acetonitrile
2
1
BenzeneCarbon dioxide (liquid 99.999 % CO
)
2
ChloroformDilute acetic acid (e.g. 0.1–1 %) at 25 °C/77.0 °FDilute ammonia solutionDilute sodium hydroxide (1 M)Ethyl acetateEthanolFormiate buffer solutionIsopropanolMethanolPhosphate buffer solutions (0.5 M)ToluolWater
1. valid for the specified temperature range
2. not recommended in combination with PEEK small parts and PEEK capillary
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
10 Intended Use
Less suitable solvents
Diethylamine (0.1 %) (DEA)Dilute phosphoric acidDimethyl sulfoxide (DMSO)
Methylene chloride
1
Slightly volatile solventsTetrahydrofuran (THF)
1
Triethylamine (0.1 %) (TEA)Trifluoroacetic acid (0.1 %) (TFA)
1. not recommended in combination with PEEK small parts and PEEK capillaries
Not suitable solvents
Concentrated mineral and organic acidsConcentrated bases
Halogenated hydrocarbons, e.g. Freon
®
Perfluorinated solvents, e.g. Fluorinert® FC-75, FC-40  Perfluorinated polyether, e.g. Fomblin
®
Solvents containing particles
Solvents not suitable for degassers
AzidesBenzeneCarbon dioxide (liquid 99.999 % CO
)
2
Concentrated mineral and organic acidsConcentrated basesDilute sodium hydroxide (1 M)
Halogenated hydrocarbons, e.g. Freon
®
HexafluoroisopropanolHexanes (60 % n-Hexane)Hydro fluoro solvents
Perfluorinated solvents, e.g. FluorinertPerfluorinated polyether, e.g. Fomblin
®
FC-75, FC-40
®
Solvents containing particles
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
11 Scope of Delivery

Scope of Delivery

Note Only use spare parts and accessories made by KNAUER or a
company authorized by KNAUER.
DAD 6.1L Detector
Power cable
User manual English/German
Installation Qualification documents
Accessory kit DAD 6.1L
AZURA
DAD 2.1L Detector
Power cable
User manual English/German
Installation Qualification documents
Accessory kit DAD 2.1L
AZURA
MWD 2.1L Detector
Power cable
User manual English/German
Installation Qualification documents
Accessory kit MWD 2.1L
AZURA
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
12 Safety for Users

Safety for Users

Professional group The user manual addresses persons who are qualified as chemi-
cal laboratory technicians or have completed comparable voca­tional training.
The following knowledge is required:
Fundamental knowledge of liquid chromatographyKnowledge regarding substances that are suitable only to a
Knowledge regarding the health risks of chemicalsParticipation during an installation of a device or a training by
If you do not belong to this or a comparable professional group, you may not perform the work described in this user manual under any circumstances. In this case, please contact your supe­rior.
Safety equipment When working with the device, take measures according to lab
regulations and wear protective clothing:
limited extent for use in liquid chromatography
the company KNAUER or an authorized company.
What must be taken into
account?
Safety glasses with side protectionProtective glovesLab coat
All safety instructions in the user manualThe environmental, installation, and connection specifica-
tions in the user manual
National and international regulations pertaining to labora-
tory work
Original spare parts, tools, and solvents made or recom-
mended by KNAUER
Good Laboratory Practice (GLP)Accident prevention regulations published by the accident
insurance companies for laboratory work
Filtration of substances under analysisUse of inline filtersOnce they have been used, never re-use capillaries in other
areas of the HPLC system.
Only use a given PEEK fitting for one specific port and never
re-use it for other ports. Always install new PEEK fittings on each separate port.
Follow KNAUER or manufacturer's instructions on caring for
the columns
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
13 Safety for Users
More safety-relevant information is listed in alphabetical order in the following table:
Topic Explanations
Flammability Organic solvents are highly flammable.
Since capillaries can detach from their screw fittings and allow solvent to escape, it is prohibited to have any open flames near the analytical system.
Solvent tray Risk of electrical shock or short circuit if
liquids get into the device's interior. For this reason, place all bottles in a solvent tray.
Solvent lines Install capillaries and tubing in such a way
that liquids cannot get into the interior in case of a leak.
Leaks Regularly check if any system compo-
nents are leaking.
Where is use of the
device prohibited?
Power cable Defective power cables are not to be used
to connect the device and the power sup­ply system.
Self-ignition point
Only use eluents that have a self-ignition point higher than 150 °C under normal ambient conditions.
Power strip If several devices are connected to one
power strip, always consider the maxi­mum power consumption of each device.
Power supply Only connect devices to voltage sources,
whose voltage equals the device's volt­age.
Toxicity Organic eluents are toxic above a certain
concentration. Ensure that work areas are always well-ventilated! Wear protective gloves and safety glasses when working on the device!
Never use the system in potentially explosive atmospheres with­out appropriate protective equipment. For further information, contact the Technical Support of KNAUER.
Taking the device out of
operation
At any time, take the device completely out of operation by either switching off the power switch or by pulling the power plug.
Opening the device The device may be opened by the KNAUER Technical Support or
any company authorized by KNAUER only.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
14 Safety for Users

Definition of Personal and Material Damage

Possible dangers related to the device are divided into personal and material damage in this user manual.
Category Explanations
DANGER! Lethal or very serious injuries can occur.
WARNING! Serious or moderate injuries can occur.
CAUTION! Moderate injuries can occur. Device defects
can occur.

Decontamination

Contamination of devices with toxic, infectious or radioactive substances poses a hazard for all persons during operation, repair, sale, and disposal of a device.
DANGER! Danger if getting in contact with toxic, infectious or radio-active substances. Before disposing off or sending away contami­nated devices, commission an expert with the decontamination.
All contaminated devices must be properly decontaminated by a specialist company or the operating company before they can be recommissioned, repaired, sold, or disposed of.
All materials or fluids used for decontamination must be col­lected separately and disposed of properly.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
15 Symbols and Signs

Symbols and Signs

The following symbols and signs can be found on the device, in the chromatography software or in the user manual:
Symbol Meaning
Warning signs High-voltage hazard
Electric shock hazard
Electrostatic discharge hazard, damages to system, device, or components may occur.
General warning sign, moderate injuries may occur and also damages to system, device, or components.
UV-light hazard, eye injuries may occur.
Burn hazard, hot surface
Hazardous substances.
Obey maximum load for leak tray during transportation, installation and operation.
Mandatory signs Wear protective gloves to avoid skin irrita-
tions.
CE mark A device or system marked with CE fulfills
the product specific requirements of Euro­pean directives. This is confirmed in a Declaration of Conformity.
Testing seals in Canada and the USA at nationally recognized testing centers (NRTL). The certified device or system has successfully passed the quality and secu­rity tests.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
16 Installation

Installation

This chapter describes all preparatory steps prior to the start-up. If you encounter difficulties during installation, contact the Technical Support.

Contact Data

Technical Support Phone +49 30 809727-111
Fax +49 30 8015010
E-mail support@knauer.net

Transport

Carefully prepare the device for transport or storage. If you want to return your device to KNAUER for repairs, enclose the Service Request Form which can be downloaded from our website.
Device data For a secure transport, note the weight and dimensions of the
device (see Technical Data).
CAUTION! Damage to the device by lifting on protruding housing parts! Lift the device on the side of the housing only.
Lifiting Clasp the device at its side panels and lift it out of the packag-
ing. Do not hold onto front cover or leak tray.
Only if the requirements for ambient conditions of the operating environment are met, can the intended use be ensured. You will find the ambient conditions under Technical Data.
CAUTION! Defect of the device due to overheating possible! Protect the device against exposure to direct sun­light. Keep clear at least 15 cm at the rear and 5–10 cm at each side for air circulation.
Installation Site and Space Requirements
Position the device on a level and even surface.Sunlight: Protect the device against direct exposure to sun-
light.
AC system: Set up the device at a location not exposed to air
drafts.
Vibration: Do not set up the device in the vicinity of other
machines that cause floor vibrations.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
17 Installation
1
Lateral distance to other devices:
At least 5 cm if another device is set up on one sideAt least 10 cm if further devices are set up on both sidesLeave the power plug on the rear of the device accessible
to be able to disconnect the device from the mains.
Avoid sources of high frequencies near the device. High-fre-
quency sources may compromise measuring values.

Operating Environment

Earth quake regions If the device is set up in a region where earth quakes are com-
mon, secure the device at the two fixing points points are situated on both sides of the device.
1 . The fixing
Fig. 5 Fixing points on the device
Space Requirements
Side clearance to other devices:
At least 5 cm, if there is another device on one side.At least 10 cm, if there are devices set up on both sides.
At least 15 cm on the rear panel for the fan.

Setup

Prior to setting up the device, find a suitable place according to the requirements and remove the packaging. The requirements and a description can be found in the following section.
Unpacking
Prerequisite Check packaging for damage caused during transportation.
Tool Utility knife
CAUTION! Damage to the device by lifting on protruding housing parts! Lift the device on the side of the housing only.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
18 Installation
Process
1. Set-up the package in such a way that you can read the label. Using the utility knife, cut the adhesive tape and open the packaging.
2. Remove the foam insert. Take out the accessory kit and the manual.
3. Open the accessory kit and check the scope of delivery. In case any parts are missing, contact the Technical Support.
4. Clasp the device from below, lift it out of the packaging and place it on its feet. Do not hold onto the front cover.
5. Check the device for signs of damage that occurred during transport. In case you notice any damage, contact the Technical Support.
6. Place the device in its site of operation and remove protec­tive foil.
Next steps Store packaging and keep the included packing list for repeat
orders.
Note The test cell is used during operation qualifications and must be
stored.

Inserting the Flow Cell

Prerequisite No flow cell or test cell is installed.
The device is switched off.
Procedure Insert the flow cell.
Note Pay attention to the compatibility of the flow cells.
WARNING! Irritation of retina through UV light! Concen­trated UV light can leak out from the flow cell or the fiber optic connectors. Switch off the device or the lamps.
Procedure Process Figure
1. Place the flow cell into the opening and slide to the back until it clicks into place inside the device.
Fig. 6 Inserting the Flow Cell
Next steps Connect the capillaries.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
19 Installation
1
1
1
3
Inserting the Flow Cell with Fiber Optics
You may also connect the flow cell with fiber optic cables. The fiber optics are connected via an adapter.
UV light will cause the fiber optics to become blind with time (solarization), making them no longer suitable for use.
Observe the following regarding the use of UV fiber optics:
Do not touch the ends of the fiber optics with your fingers, as
this could falsify the measurement.
Handle the fiber optics with care, avoid impacts or hard
actions.
Move the fiber optics carefully without using pressure or
bending it.
Prerequisites No flow cell or test cell is installed.
Auxiliary material Screw-type cap fittings for the fiber optics
WARNING! Irritation of retina through UV light! Concen­trated UV light can leak out from the flow cell or the fiber optic connectors. Switch off the device or the lamps.
Procedure Process Figure
1. Remove the cap fittings
1 from the fiber optics
adapter.
2. Place the fiber optics adapter into the open­ing and slide to the back until it clicks into place inside the device.
3. Remove the screw-type cap fittings fiber optics.
4. Manually, screw the fiber optics
fiber optics adapter.
5. Remove the cap fittings from the flow cell.
6. Manually, screw the fit­tings
3 of the fiber
optics to the flow cell.
1 from the
2 to the
Fig. 7 Fiber optics adapter
Fig. 8 Screw-type cap fitting
on the fiber optics
Fig. 9 Fitting of the fiber
optics on the flow cell
Next steps Connect the capillaries.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
20 Installation
132
4
13
2
4
4
4

Connecting the Capillaries

Capillaries connect the detector with the column and guides li­quids. For the flow cell connectors, we recommend PEEK capil­laries and PEEK connectors.
CAUTION! Damage to the PEEK threads of the flow cell through metal fittings possible. Attach capillaries with PEEK fittings only. Observe torque of 0.5 Nm maximum.
Prerequisite The flow cell is inserted.
Screw-type cap fittings and cap fittings are removed.
Procedure Attach the capillaries with screw fitting.
Procedure Process Figure
1. Push the capillary through the fitting
2. Slide the lock ring over the tubing. Note that the tapered end of
the lock ring has to point to the seal
ring
4.
3. Attach the seal ring
4. Fasten the capillary at the flow cell
tight.
Next steps Bring the device into operation.
1
2.
3
4.
4 finger-
Fig. 10 Capillary connectors
Fig. 11 Flow cell with capillary
Using PEEK Fittings
PEEK fittings withstand a maximum pressure of 400 bar. Note the torque of 0.5 Nm.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
21 Installation
1
2
3
4
5

Connecting the Leak Management

The leak management consists of the leak sensor and the drain­age system. The drainage system ensures that escaping liquids flow into a waste bottle. If there is too much liquid, the red LED starts flashing. Both the device and the data acquisition via chro­matography software are stopped.
Prerequisite The front cover has been removed.
Procedure
Process Figure
1. Carefully push the fun-
1 into the center
nel opening of the capillary guide 2.
Fig. 12 Funnel and capillary
2. Push the long ending of the first nozzle
the hose
3. Connect the nozzle and the funnel.
4. Push the other end of the hose onto the noz-
zle
5 of the leak tray.
4 into
3.
Fig. 13 Hose and nozzle
guide
Fig. 14 Hose connected to
device
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
22 Installation
6
1423
Procedure Process Figure
5. For the bottom device, push the short end of
the nozzle opening in the collec­tion point of the leak
tray.
6. Connect the hose to the nozzle and lead the second ending to the waste bottle.
7. Place the waste bottle below the bottom device.
Next steps Attach the front cover.
6 into the
Fig. 15 Leak tray with nozzle
Legend
1 Integrator outlets
2 LAN connector
3 Service interface
4 Pin header

Control

The detector can be controlled externally in two ways:
Via pin headerAs part of a LAN, via the LAN connector of the router
All connectors for external control are located on the rear side of the detector
The DAD 6.1L can also be controlled via the Control Unit. The Control Unit connector is located on the side of the device.
.
Fig. 16 Connectors on rear side
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
DAD 6.1L
5
1 USB port for the
Control Unit
23 Installation
Fig. 17 Control Unit port on DAD 6.1L
Connecting the Device to the Computer
This section describes how to set up an HPLC system in a local area network (LAN) and how a network administrator can inte­grate this LAN into your company network. The description applies to the operating system Windows® and all conventional routers.
Note To set up a LAN, we recommend to use a router. That means
the following steps are required:
Process 1. On the computer, go to the control panel and check the LAN
properties.
2. Hook up the router to the devices and the computer.
3. On the computer, configure the router to set up the network.
4. Install the chromatography software from the data storage device.
5. Switch on the device and run the chromatography software.
Configuring the LAN Settings
The LAN uses only one server (which is normally the router) from that the devices automatically receive their IP address.
Prerequisite In Windows
screen saver must be deactived.
In case you use an USB-to-COM box, the option "Allow the
computer to turn off ths device to save power" in the device­manager must be deactivated for all USB hosts.
Only for Windows 7: For the network adapter, the option
"Allow the computer to turn off this device to save power" in the Device Manager must be deactivated.
®
, power saving, hibernation, standby, and
Procedure 1. In Windows 7 choose Start
Sharing Center.
2. Double-click on LAN Connection.
3. Click on the button Properties.
4. Select Internet Protocol version 4 (TCP/IPv4).
5. Click on the button Properties.
6. Check the settings in the tab General. The correct settings for
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
the DHCP client are:
Control Panel  Network and
24 Installation
12
3
4
5
a) Obtain IP address automatically b) Obtain DNS server address automatically
7. Click on the button OK.
Connecting the Cables
A router has several LAN ports and one WAN port that can be used to integrate the LAN into a wide area network (WAN), e.g. a company network or the Internet. In contrast, the LAN ports serve to set up a network from devices and a computer. To avoid interference, we recommend operating the HPLC system separately from the company network.
You will find patch cables for each device and the router in the accessories kit. To connect the router to a WAN, an additional patch cable is required, which is not supplied within the scope of delivery.
Legend
1 Modules
2 Router
4 WAN port
5 Workstation
3 LAN ports
Fig. 18 Cabling system LAN
Prerequisite The computer has been switched off.
There is a patch cable for each device and the computer.
Procedure 1. Use the patch cable to connect the router and the computer.
Repeat this step to connect all devices.
2. Use the power supply to connect the router to the mains power system.
Configuring the Router
The router is preset at the factory. You will find a label at the bottom side of the router, on which IP address, user name, and password are printed. These information help to open the router configuration.
Procedure 1. To open the router configuration, start your Internet browser
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
and enter the IP address (not for all routers).
2. Enter user name and password.
25 Installation
3. Configure the router as DHCP server.
4. In the router configuration, check the IP address range and make changes if necessary.
Result Once the router has assigned IP addresses to all devices, the
chromatography software can be used to remotely control the system.
Integrating the LAN into a Company Network
A network administrator can integrate the LAN into your com­pany network. In this case you use the WAN port of the router.
Prerequisite There is a patch cable for the connection.
Procedure 1. Check that the IP address range of the router and of the com-
pany network do not overlap.
2. In case of an overlap, change the IP address range of the router.
3. Use the patch cable to connect the router WAN port to the company network.
4. Restart all devices, including the computer.
Controlling Several Systems Separately in a LAN
Devices connected to a LAN communicate through ports, which are part of the IP address. If more than one HPLC system is con­nected to the same LAN and you plan on controlling them sepa­rately, you can use different ports to avoid interference. Therefore, the port number for each device must be changed and this same number must be entered into the device configu­ration of the chromatography software. We recommend to use the same port number for all devices in the same system.
Note The port is set to 10001 at the factory. You must use the same
numbers in the device configuration of the chromatography software as in the device, otherwise the connection fails.
Procedure 1. Find out port number and change it on the device.
2. Enter the port number in the chromatography software.
Result The connection is established.
Pin Header Connectors
For test purposes or in some other cases, it can make sense to manually enter these signals.
Sending control signals (Events) to external devicesOpening and closing contactsActivating 500 ms pulses
The following remote signals can be received and sent:
For receiving start, control, and error signals from external
devices
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
26 Installation
For sending start, control and error signals to external devices
Fig. 19 Relation between display and Events terminal strip
Explanations for Pin Header Connector
Connection Function
Relay contact
EV 1
(Event 1)
The contact is on a floating basis. Its setting depends on the settings in the Control Unit, the Mobile Control, or software. Steady-rate signal:
passive = open relay contactactive = closed relay contact
Pulse:
Closed relay contact for at least
1000 ms Permissible load of the relay contact: 1 A/ 24 V DC
EV 2 (Event 2) TTL output
Levels:
passive 5 Vactive 0 V
Pulse:
0 V for at least 1000 ms
Error OUT TTL output
Levels:
passive 5 Vactive 0 V
Output is active until the Error con­dition has been eliminated.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
27 Installation
Connection Function
Error IN TTL input
Start IN TTL input
Low active
Secure switching threshold at least 10 mA
After receiving a signal (short-circuit to ground) from an external device, an error message appears and the device stops.
Low active
Secure switching threshold at least 10 mA
After receiving a signal (short-circuit to ground) from an external device, the device starts. If controlled with soft­ware, an electronic trigger is send through the LAN.
Autozero TTL input
Low active
Secure switching threshold at least 10 mA
A signal (short-circuit to ground) sets the measuring signal to zero.
+5 V Provides a voltage of 5 V with respect
to GND. This makes it possible to sup­ply a load that is switched by an EVENT. Protection: 5 V-50 mA
GND Reference point of the voltage at the
signal inputs.
+24 V Valve Event-controlled switching of 24 V
against GND Protection: 24 V-200 mA
Valve OUT Output is active until the valve condi-
tion has been eliminated.
GND Reference point of the voltage at the
signal inputs.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
28 Installation
1
2
3
Wiring the Pin Header
To control one device through another, you use the pin header. The single ports are used to exchange control signals.
To use remote control, you have to connect cables to the pin header. Therefore, connect the cables firstly to the spring strip and push the spring contacts of the spring strip secondly onto the pins of the pin header.
Prerequisites The device has been turned off.
The power plug has been pulled.
Tools Depressor tool
CAUTION! Short-circuit hazard. Turn off the device before connecting it to the pin header. Pull the power plug.
CAUTION! Electrostatic discharge can destroy the electronics! Wear protective bracelet against electrostatic dis­charge and ground.
Procedure Process Figure
1. Place the spring strip on a suitable surface.
2. Push the depressor tool into the opening on the
spring strip.
3. Continue pushing the depressor tool down and
lead the cable front end of the spring strip.
4. Remove the depressor tool.
5. Check whether the cables are tightly attached.
6. Plug the spring strip onto the pin header.
2 into the
3
1
Fig. 20 Spring strip
Next steps Finish the installation and perform the initial startup.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
29 Installation
Integrator Connector
The integrator connector sends measuring signals from the detector.
non-bipolar4 channel0 to 5 V DAC 16 bitscalableAdjustable to offset

Power Supply

Note The nominal capacity of the connected devices must be maxi-
mum 50 % of the power supply to account for larger inrush cur­rents when switching on the modules.
The maximum power input is 75 watts.
Prerequisites The electrical power supply at the installation site must be
connected directly to the nearest main power line.
The power must be free from ripple, residual current, voltage
peaks and electromagnetic interference.
The connectors for the mains voltage are grounded accord-
ingly.
The device receives sufficient power with reserve capacity.
The device is intended for use with AC power networks of 100 – 240 V.
Power cable Use only the enclosed power cable to connect the device to the
mains. Replace defective power cables only with accessories from KNAUER. Only use power cables admitted for use in your country.
In case of queries, contact the Technical Support.
CAUTION! Electric hazard if the power adapter is turned on while connecting or interrupting the power con­nection. Switch off the power adapter beforehand.
Power plug Make sure that the power plug on the rear of the device is
always accessible, so that the device can be disconnected from the power supply.
Next steps Finish the installation and connect the device to the mains.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Procedure Stainless steel fittings are tightened with 5 Nm, PEEK fittings are
Capillary layout
30 Installation
Integrating the Detector into a Sys­tem
To integrate the pump into a system, note the ambient condi­tions found in the sections Operating Environment and Techni­cal Data as well as the ambient conditions of other devices to be integrated into that system.
The detector is integrated into the HPLC flow system by con­necting the capillary to the flow cell and the HPCL system.
tightened with 0.5 Nm.
Fig. 21 Installation of the detector
Pre-Installed Capillaries
Pre-installed stainless steel and PEEK capillaries are color-coded according to their inner diameter.
Color Material Inner diameter
red marker Stainless steel 0.1 mm
blue marker Stainless steel 0.25 mm
black marker Stainless steel 0.4 mm
red stripes PEEK 0.1 mm
blue stripes PEEK 0.25 mm
orange stripes PEEK 0.5 mm
Note PEEK capillaries are not suitable for use with pure Acetonitrile.
Acetonitrile can cause capillaries to crack or rupture.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0

Operation

1
2
31 Operation
In this chapter, you find information relevant for operating the detector.

Initial Start-Up

Use this checklist to determine whether the detector is ready for the initial start-up:
Devices is positioned in the correct location.The power plug has been pulled.
If the device is part of an HPLC system, you should also note the following:
The network connection to the router is established.The chromatography software has been installed by KNAUER
or a company authorized by KNAUER.
A flow cell has been installed.The capillaries have been connected.

Switch-On

Prerequisite The installation has been completed.
The flow cell or test cell has been inserted.
CAUTION! Possible damage to the device caused by con­densed water! Allow device to acclimate for 3 h, before it is con­nected to power supply and taken into operation.
Procedure Process Figure
1. On the rear of the device, plug the power cable into the power
connector
2. Plug in the power sup­ply.
3. On the rear of the device, switch on the detector using the
power switch
2.
Fig. 22 Power switch with
1.
female connector at the rear panel
Result The detector starts its self-validation. If the self-validation has
been successfully completed, the LED on the right and in the center light up green.
Next steps Bring the detector into service.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
32 Operation

Control

There are several options for controlling the device:
with chromatography softwarewith the Control Unitwith an app (Mobile Control)
Note It is not possible to use to control methods simultaneously. If the
device is connected to the software, it cannot be controlled via Control Unit or Mobile Control.
Operation with Chromatography Software
To control the device with software, you have to establish a con­nection between the LAN port and a computer.
DAD 6.1L
Devices can be controlled with e.g. OpenLAB version A.04.05 or higher, or with ClarityChrom version
or higher. You find a detailed description on chromatography software in
the corresponding user manual.
®
EZChrom edition
®
5.0.3
Control with Control Unit
The Control Unit is an optional accessory. To control the DAD
6.1L using the Control Unit, connect the cable of the Control Unit to the device. You find a detailed description on the Con­trol Unit in its accompanying user manual.
Control with Mobile Control
The Mobile Control is an app which can be installed on your computer or tablet computer. To control the device using the Mobile Control, connect the computer or tablet computer to a wireless LAN router. Data transfer between the device and the app is actualized through wireless LAN. You find a detailed description on the Mobile Control in its accompanying user manual.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Legend
123 4
1 Left LED
2 Center LED
3 Right LED
4 Switch/Standby but-
ton
33 Operation

Meaning of the LEDs

There are three LEDs and a switch on the front of the device.
Fig. 23 LEDs and switch/standby button on the front of the
device
The LEDs can have different colors depending on the operating conditions.
Standby To start the standby, keep the standby button pressed for
5 seconds.
Note Malfunctioning system after repeated standby possible. After
repeatedly using standby, restart the device using the power switch to reset the device's data storage.
Color Status Operation
Left LED red Error Message Check the sys-
tem
Shortly press
the switch to deactivate the error message.
green 3D data are
acquired.
Center LED does not light The lamp has
been switched off.
Right LED green The device has
flashes green The lamp/lamps
are initializing or the validation is progressing.
green The deuterium
lamp is active.
been switched on.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Wait until the
lamp is running or the valida­tion is finished.
34 Operation
Color Status Operation
flashes green The device not
ready for opera­tion.
blue The device is in
standby
Wait until the
device is ready for operation.
Press the
standby but­ton to end the standby.

Re-Setting the Device

The DAD 6.1L can be reset to default settings using the Control Unit.
Parameter Setting
Network LAN DHCP, port 10001
Lamps D2 ON, Halogen OFF
Time constant 2 s
Channels Channel 1: WL = 254 nm, BW = 8 nm
Reference correction Reference channel enabled for channel
1, WL = 360 nm, BW = 30 nm
Analog out Offset 1 = 0.00 mV, Scale 1 = 1 AU/V
Flow cell Test cell
Event check All events deactivated (o)
Date/Time Current date/time
Wake-up Current date/time
Leak sensor ON, Sensitivity = low
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
35 Optimizing the Detector

Optimizing the Detector

In this chapter, you find information on how to optimize your detector.

Location

In order to ensure thermo stability and to prevent drift effects, note the following aspects:
Protect the detector against strong ventilationProtect the detector against direct sunlightNote the space requirementsAvoid vibrations

Warm-Up Time

The recommended warm-up time for the detectors is 30 min­utes. This recommendation is applicable after turning the detec­tor on as well as and after turning the lamp/s on.
Fig. 24 Cold start temperature profile

Selecting the Flow Cell

Several different flow cells are available for the detectors (see “Flow Cells” on page 68). Note that the detector is only shipped with a test cell. A flow cell must be ordered separately.
Signal sensitivity, peak broadening and response can all be affected by the choice of flow cell. Important variables to take into account are the volume and path length as well as the flow cell material, required working pressure range, flow cell connec­tion, and remote operation.
Flow Cell Volume
Depending on your instrument setup, column and sample(s), one flow cell volume may be more appropriate than another. If the flow cell is too large, two adjacent peaks may get mixed in the cell. If the flow cell is too small, the noise may be higher and
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
36 Optimizing the Detector
the signal may be too small due to less light reaching the photo diodes.
Ideal flow cell volume therefore is a compromise between peak broadening and sensitivity (see figure 25).
Fig. 25 Volume comparison
A good rule of thumb is that flow cell volume should not be more than 1/3 of the peak volume of your separated sample. To determine the volume of your peaks, take the peak width as reported in the integration results, multiply it by the flow rate and divide it by 3.
Volumes of 2 μl, 6 μl and 10 μl are available for the detectors. Narrow-bore columns (~ 2,1 mm ID) are suitable for flow cells with smaller volumes. Columns with a larger inner diameter ( 3,0 mm ID) are less affected by the volume of the flow cell.
The flow rate should also be taken into consideration. A lower flow rate increases the axial and longitudinal diffusion and adds to a broadened flow profile which may lead to a peak broaden­ing.
Path Length
As described by the Beer-Lambert law (see figure 26), the path length of a flow cell affects the intensity of light that is detected.
Fig. 26 Beer-Lambert law
A: Measured absorption at a wavelength
T: Transmittance, defined as the quotient of the light
intensity (I) after passing through the sample and the initial light intensity (I
) before passing through the
0
sample.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
37 Optimizing the Detector
: Molar absorptivity coefficient (wavelength- and tem-
perature-dependent)
b: Path length
c: Analyte concentration (temperature-dependent)
For the same volume of sample, the apparent concentration of the sample will appear to be higher if the path length is longer. Optical path lengths of 3 mm, 10 mm, and 50 mm are available for the detectors. A longer path length increases the signal intensity as more light is absorbed and thereby may enhance the sensitivity of a method (see figure 27).
Fig. 27 Path length comparison
Material
The flow cell materials must be chemically compatible with the solvents and sample you are working with. Stainless steel and metal-free flow cells are available. The metal-free variants are bio-inert.
Pressure Stability
The different flow cells can withstand various maximum pres­sures. The upper pressure limits of the flow cells are 30 bar, 100 bar, or 300 bar. The flow cell should not be subjected to the maximum pressure for a long period of time.
Connection
Undesired effects, like the loss of resolution in the chromato­gram, may be prevented by ensuring a correct connection to the flow cell and the removal of any dead volume.
Remote Operation (Fiber Optics)
The detectors can be operated with a spatially separated flow cell which is connected with fiber optics. This provides enhanced security for hazardous, explosive, or toxic work pro­cesses. Sensitivity is typically reduced to half that of an equiva­lent standard cell when working with standard length fiber optic cables (750 mm). Overall, sensitivity is inversely proportional to the length of the fiber optic cables.
Temperature changes around the cables may cause drift.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
38 Optimizing the Detector

Selecting the Wavelength

Signal wavelength Wavelength selection can influence the sensitivity, selectivity
and linearity of a measurement. The best wavelength for a given measurement (signal wavelength) is that which gives the maxi­mum absorbtion above the UV cutoff of the mobile phase. In cases where there are multiple components with different absor­bance maxima, a compromise wavelength must be chosen where all components absorb.
Reference wavelength In order to minimize baseline drift due to refractive index effects,
a reference wavelength can be set (see figure 28). The reference should be set in the same spectral region as the signal wave­length (UV or VIS) but at a wavelength at which the analyte has no absorbance.
Default reference wave-
length
By default, the reference wavelength 360 nm is activated (see “Re-Setting the Device” on page 34). This is a suitable value for most applications.
Fig. 28 Comparison with and without reference wavelength
When selecting the signal and reference wavelengths, the respective bandwidths must also be selected (see the following section).

Bandwidth

The bandwidth defines the total number of wavelengths actually registered by the photodiode when a specific wavelength is set.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
39 Optimizing the Detector
1
10
------
For example, a wavelength set at 254 nm with a bandwidth of 4 nm results in an average absorbtion of 252–256 nm.
Fig. 29 Bandwidth
The selection of the bandwidth is a balance between sensitivity and selectivity. Narrow bandwidths increase selectivity, whereas broad bandwidths increase sensitivity.
Default bandwidth By default (see “Re-Setting the Device” on page 34), the band-
width for the signal wavelength is set to 8 nm and for the refer­ence wavelength the bandwidth is set to 30 nm.

Spectral Range

When measuring with chromatography software, the spectral range selected for a given measurement influences the integra­tion time and consequently the sensitivity of the measurement, as well as the disk space. The disk space is necessary to save gen­erated data.
A narrow spectral range increases sensitivity and reduces disk space. The range, however, should be wide enough to ensure the detection of all components. Also, the spectral range always must include the signal wavelength and the reference wave­length.

Time Constant and Data Rate

Response time The time constant influences the response time of the detector.
The response time determines how quickly the detector responds to a change in signal. A good rule of thumb for selec­tion of the detector time constant is that it should be no larger
than of the baseline peak width of the first peak of interest (in seconds). Increasing the time constant allows more averag-
ing of the signal (also known as digital filtering) and results in less baseline noise (see figure 30). However, increasing the time constant too much may result in broad peaks, reduced peak heights and asymmetric peak shapes. Therefore, a compromise has to be found.
Time constant In general, the best time constant is the reciprocal of the data
rate. If increased sensitivity is desired, or if the baseline noise is
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
40 Optimizing the Detector
interfering with integration, the time constant should be increased. If resolution is compromised, it should be decreased.
It is recommended to set the time constant and data rate in rela­tion to peak width.
Peak width Time constant Data rate
[min] [s] [Hz]
<0.003 0.01 100
>0.007 0.02 50
>0.017 0.05 20
>0.033 0.1 10
>0.067 0.2 5
>0.167 0.5 2
>0.333 1 1
Data rate The data rate (or sampling rate) is the number of data points per
second (Hz) at which the detector transmits data to the com­puter.
Default data rate The default data rate setting for the detectors is 1 Hz (see “Re-
Setting the Device” on page 34). The maximum data rate (digi­tal signal) is 100 Hz. Slower data rates store average data points. A 50 Hz data rate averages 2 points. A 10 Hz data rate averages 10 points. The analog data rate is fixed at 12.5 Hz.
Optimizing the
data rate
The optimal data rate depends on your application. Too few points across a peak (short data rate) decrease detail and com­promise reproducibility. Too many points (high data rate) intro­duce noise into the system and the resulting files can become very large. Some general considerations are listed below:
Each peak should be defined by 20–30 data points. For chro-
matograms with co-eluting peaks or low signal-to-noise ratios, 40–50 data points per peak are recommended.
If all peaks are relatively wide, select a slower data rate.If any peaks of interest are less than a few seconds, select a
faster data rate.
If the data rate is too slow, the start and end points of the peaks are not accurately determined. If the data rate is too fast, data
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
41 Optimizing the Detector
files may occupy excessive disk space and post-run analyses may require more processing time.
Fig. 30 Noise with different data rates

Integration time

Signal level The integration time influences the intensity of the signal and
therefore the sensitivity of the measurement. The larger the inte­gration time, the higher the intensity of the signal until the max­imum sensor counts are reached. The integration time is automatically calculated by the software before the start of a measurement. The calculations are in relation to the spectral range (see section “Spectral Range” on page 39).
When a narrower spectral range is selected, the signal intensity will be increased. This increase is limited however by the data rate.

Subtraction of the Baseline Chromatogram

The baseline subtraction can eliminate the effects of drift that result from solvent, gradient, or flow programming. The base­line profile is subtracted from the measured chromatogram. This results in a mathematically reprocessed chromatogram with an ideally flat baseline.

General

The performance of the detector is largely dependent on the
performance of the HPLC system.
Noise can be related to pump stability, the detector cell
cleanliness, lamp quality, mobile phase composition and other factors.
Drift is usually related to long-term changes in the environ-
ment, such as detector warm-up or fluctuations in tempera­ture and mobile phase composition.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
42 Functionality Tests

Functionality Tests

Installation Qualification (IQ)

Installation report Certification on the functionality of the device. During installa-
tion of the device, an installation report (IQ document) is cre­ated upon request in coordination with the Technical Support of KNAUER.
This installation report needs to be completed in full and signed by both parties. It serves as proof of the properly executed installation and functionality of the device.

Operation Qualification (OQ)

Extensive
functionality test
Test Intervals Run the extensive functionality test at the following time inter-
Execution The test can carried out either by KNAUER's Technical Support
Extensive test of the device’s functionality. A successfully exe­cuted OQ ensures that the device functions properly.
vals:
Average use Device test
1 to 5 days/week: Every 6
months
More than 5 days/week or 24 hours/day: Every 3
months
Operation with buffer solutions or other salt solutions:
or by a technical service authorized by KNAUER. The OQ documentation required for executing the OQ is with
costs (once) and can be ordered separately from KNAUER.
Every 3 months
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
43 Troubleshooting

Troubleshooting

First Measures 1. Check all cabling.
2. Check all screw fittings.
3. Check whether air has gotten into the supply lines.
4. Check device for leaks.
5. Pay attention to system messages.
LAN
In case no LAN connection between the computer and the devices can be established, go through the following steps. Check after each step if the problem is solved. If the problem cannot be located, call the Technical Support.
1. Check the status of the LAN connection in the Win­dows task bar:
Connected
Connection not established
If no connection was established, test the following:
Is the router switched on?Is the patch cable connected correctly to the router
and the computer?
2. Check the router settings:
Is the router set to DCHP server?Is the IP address range sufficient for all the connected
devices?
3. Check all connections:
Are the patch cables connected to the LAN ports and
not the WAN port?
Are all cable connections between devices and router
correct?
Are the cables plugged in tightly?
4. If the router is integrated into a company network, pull out the patch cable from the WAN port.
Can the devices communicate with the computer,
even though the router is disconnected from the com­pany network?
5. In case you own a Control Unit, check the settings in the menu Setup > Network.
Is LAN-DHCP set for controlling?Did the device receive an IP address?
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
44 Troubleshooting
6. Turn off all devices, router, and computer. Firstly, turn on the router and secondly turn on the devices and the computer.
Has this been successful?
7. Replace the patch cable to the device that no connec­tion could be established with.
Has this been successful?
8. Ensure that the IP port of the device corresponds to the IP port set in the chromatography software.

Possible Problems and Solutions

Problem Solution
Baseline drift Maintain constant temperature condi-
tions during the measurement.
Device cannot be switched on
Device cannot be calibrated
Baseline noise 1. Inspect the flow-cell assembly.
The relationship of the signal to the light path reference is very low
Further Measures Install the maintenance software (service tool).
Save device information and send to KNAUER.Inform the KNAUER Technical Support.
Inspect the power cable to ensure that it is plugged into the power supply.
1. Insert the test cell.
2. Inspect the calibration with a weak absorbing solvent.
2. Exchange the defective flow cell.
3. Inspect the service life of the lamp on the display.
4. Reduce the air in the flow cell by using a degasser.
1. Flush the flow cell.
2. Replace the lamps

System Messages

If other system messages are displayed besides those listed below, please turn the device off and then on. Inform the Tech­nical Support of the manufacturer in case the system message repeats itself.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
45 Troubleshooting
The system messages are in alphabetical order:
System message Error List and Solutions
A “Ambient tempera-
ture sensor failed”
C “Cannot initialize
LAN”
“Instrument not vali­dated”
“Cannot proceed: D2 lamp heating”
“Cannot proceed: D2 lamp off”
“Cannot proceed: Low light”
Main board temperature sensor not detected or does not respond.
Check cables and connections in local area network.
Data aquisition or single scan cannot be executed when the instrument is not validated.
Manual validation cannot be executed during lamp ignition. Wait for the deuterium lamp to finish the initializa­tion process, then continue.
Manual validation can not be exe­cuted while deuterium lamp is off. Switch on the lamp. If the error occurs again, restart the device. If the error message re-appears, replace the lamp.
Validation failed due to integration time exeeding limits. If the recom­mended operating time of the lamp has been exceeded. Replace the deu­terium lamp. If the flow cell is dirty. Clean the flow cell. Finally, if the optical system is faulty. Inform the KNAUER Technical Sup­port.
D “D2 lamp failed” Manual validation cannot be executed
when no deuterium lamp is present.
“Communication
RS-232 communication disrupted.
buffer overflow”
“Communication timeout”
Timeout: RS-232 (5 s), Leak sensor (0.5 s) or hardware component (lamp power supply, EEPROM, I2C with GUI) communication failure
“CRC failure” Correct command, disruption during
communication (not currently active).
”D2 lamp operation failed”
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Restart the device. If the error occurs again, replace the lamp.
46 Troubleshooting
System message Error List and Solutions
”D2 lamp does not start”
Switch off lamp on touchscreen and restart. Inform the KNAUER Technical Support in case the system message repeats itself. Replace the lamp unit. Upon device start up or manual vali­dation attempt, lamp does not ignite.
”Data acquisition active”
No entries are possible. First stop acquiring measurement data, after­wards you can make a new entry.
E "Error input acti-
vated"
Error outside of the device. Check the external devices and cable connec­tions. Check the system to localize and resolve the source of the external error.
H ”HAL lamp does not
start” (DAD 6.1L only)
Switch on the lamp. If the error occurs again, restart the device. If the lamp is off after restarting, replace the lamp.
I “Instrument busy” Attempt to scan while 3D data acqui-
sition is active, leak sensor or shutter is busy processing previous command. Wait until the device has complete processing.
“Instrument in standby mode”
“I2C lamp(s) opera­tion failed”
“I2C operation failed”
Command not allowed for standby mode.
Lamp power supply I2C communica­tion failure.
Leak sensor or lamp power supply firmware upload or EEPROM I2C com­munication failure.
“Instrument in standalone mode”
”Instrument remote controlled”
“Insufficient access for operation”
Command cannot be executed in local modus (only in remote modus)
This entry is not executable. Quit soft­ware.
Lamp power supply data cannot be edited (service mode should be acti­vated)
”Invalid command” Check the cable connections. Change
the entry. Wrong command sent to device. Check the validity of the command.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
47 Troubleshooting
System message Error List and Solutions
“Invalid parame­ter(s)”
Correct command, incorrect parame­ter. Check the validity of the parame­ters
L “Lamp cover open” Protective lamp cover not correctly
attached or micro switch malfunction. Attach the lamp cover.
“Lamp not installed” GLP data and lamp operation cannot
be performed as this lamp (D2 or hal­ogen) is not installed.
“Lamps off” Command cannot be carried out
while the lamps are turned off.
“Lamp supply is not available”
“Lamp supply tem­perature limit
Lamp supply not installed or does not respond.
Upper temperature limit for lamp power supply has been reached.
exceeded”
“Lamp unit tem­perature limit
Upper temperature limit for lamp unit has been reached.
exceeded”
“Lamp unit tem­perature limit exceeded”
“Leak sensor failed” Leak sensor not detected or does not
"Leak was detected" Switch off the device. Remove the leak
N “No D2 lamp
detected”
“Not available for this communication mode”
R “Recommended D2
lamp life exceeded”
“Recommended HAL lamp life exceeded”
Main board temperature sensor not detected or does not respond.
respond. Restart the device. If the leak sensor is still not present, contact the Technical Support of the manufacturer.
and restart the device.
Make sure that the deuterium lamp has been installed correctly.
3D data aquisition or single scan is not available over RS-232.
The recommended 2000 hours for the deuterium lamp has been exceeded. Replace the lamp.
The recommended 1000 hours for the halogen lamp has been exceeded. Replace the lamp.
S “Shutter position
Shutter filter motor malfunction.
failed”
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
48 Troubleshooting
System message Error List and Solutions
“Spectrum buffer overflow”
“Spectrum output busy or not ready”
T “Temperature con-
trol failed”
”Time already exists”
W “Wake up time
already passed”
“WL/BW out of spectral range”
”WL validation failed”
Internal 3D data buffer is exhausted due to LAN connection failure.
3D data acquisition cannot be started. Wait until the data transfer is com­plete before continuing.
Maximum temperature on lamp unit exceeded.
Correct the time entry.
The wake up entry is set in the past. Enter a new time.
Selected wavelength and bandwidth exceed the spectral range. Extend the selected spectral (scan) range or change the selected wavelength and / or bandwidth.
The wavelength precision validation has failed. Restart the validation. Hol­mium oxide and/or H
, H lines out of
specification.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
49 Maintenance and Care

Maintenance and Care

In this chapter, you find all steps necessary for maintenance, care, and storage. Additionally, you find instructions for mainte­nance tasks that you may perform yourself. If maintenance tasks are necessary that are not described here, contact your dealer or the Technical Support.
Organic solvents are toxic above a certain concentration. Ensure that work areas are always well-ventilated! When performing maintenance tasks on the device, always wear safety glasses with side protection, protective gloves, and a lab coat.
All wetter components of a device, e.g. flow cells of detectors or pump heads and pressure sensors for pumps, have to be flushed with isopropanol first and water afterwards before being main­tained, disassembled or disposed.
Opening the device The device may only be opened by the KNAUER Technical Sup-
port or any company authorized by KNAUER.
WARNING! Irritation of retina through UV light! Concen­trated UV light can leak out from the flow cell or the fiber optic connectors.
Switch off the device or the lamps.
CAUTION! Performing maintenance tasks on a switched on device can cause damage to the device. Switch off the device and pull the power plug.
DANGER! Danger of electric shock! Voltage-carrying parts inside the device. Before opening the housing, be sure to switch off the device and pull the power plug.
Users may perform the following maintenance tasks themselves:
Regularly checking the operating hours of the lamp.Checking the flow cell installation.Replacing the flow cellReplacing the lamp
Proper maintenance of your HPLC device will ensure successful analyses and reproducible results.
Find the order numbers of the required spare parts in the chap­ter Accessories and Spare Parts.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
50 Maintenance and Care

Contact the Technical Support

If you have any technical questions regarding KNAUER hardware or software, please use one of the contact options below:
Technical Support hotline:
European hotline Languages: Available by telephone
in German or English:8 am to 5 pm (CET) Phone:+49 30 809727-111 Telefax:+49 30 8015010
E-mail contact: support@knauer.net

Maintenance Contract

The following maintenance work on the device may only be per­formed by KNAUER or a company authorized by KNAUER and is covered by a separate maintenance contract:
• Opening the device or removing housing parts.

Maintenance Intervals

Operating hours The Control Unit, the Mobile Control and the different software
®
products (e.g. ClarityChrom play or read out the operating time of the detector. You find a detailed description on how to display or read out GLP data in the respective user manuals.
Operating hours Measures
1000 Replace halogen lamp (DAD 6.1L
2000 Replace deuterium lamp
or OpenLAB®) enable you to dis-
only)

Cleaning and Caring for the Device

CAUTION! Intruding liquids can cause damage to the device! Place solvent bottles next to the device or in a solvent tray. Moisten the cleaning cloth only slightly.
All smooth surfaces of the device can be cleaned with a mild, commercially available cleaning solution, or with isopropanol.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
51 Maintenance and Care

Checking the Fittings

Check if all fittings are tight. If fittings are not tight, re-tighten fittings.
Capillary fittings Torque
Stainless steel fittings 5 Nm
PEEK fittings 0.5 Nm

Putting the Device out of Operation

The detector is designed for use with different solvents. If the detector is out of operation for several weeks, residual solvent can cause damage. Therefore, it is recommended to:
flush the flow cell, the capillary, and the detectorremove the used solvent completelyfill the flow cell, the capillary, and the detector with isopropa-
nol
Seal the open connectors of the flow cell with cap fittings.
Prerequisite The detector head has been purged.
Tools Cap fittings and/or screw-type cap fittings
Procedure Process
1. Remove the solvent tubing and seal all open connectors with cap fittings.
2. Disconnect the detector from the HPLC flow system and seal the open connectors of the flow cell with cap fittings.
Next steps Choose a storage location according to the ambient conditions
listed in the respective chapter of this user manual.

Storage

Note that prior to storing all tubes and capillary should be emp­tied or filled with an appropriate flushing solution (e.g. isopro­panol). To prevent algae formation, do not use pure water. Seal all inlets and outlets with cap fittings.
Note the ambient conditions for storage (See “Ambient condi­tions” on page 58, 59, 60).
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
52 Maintenance and Care
Disconnect from Power Supply
Prerequisite The device has been switched off.
Procedure
Next steps Disconnect all power connectors. Remove the flow cell and pack
Process
1. Pull the power plug from the socket and afterwards remove from the device.
2. Pack the power cable with the device.
the device for transport or storage.

Cleaning the Flow Cell

Increased noise of the baseline and reduced sensitivity can be a result of a dirty flow cell. Often it is sufficient to rinse the flow cell to restore optimal sensitivity.
Rinsing the Flow Cell
CAUTION! Contamination of the flow cell caused by oil drops! Do not use compressed air for drying.
The following solvents are recommended for rinsing:
dilute HCl (1 mol/L)1 mol/L NaOH aq.EthanolAcetone
Tool Syringe
Procedure 1. Fill the syringe with eluent.
2. Inject it into the inlet of the flow cell and allow it to act for 5 minutes.
3. Fill the syringe with water and flush again.
4. Remove the flow cell from the detector and use a nitrogen stream to dry it.
Next steps Check if the baseline noise has disappeared.

Replacing the Flow Cell

UV light will cause the flow cells to become blind with time (solarization), making them no longer suitable for use. The man­ufacturer recommends replacing the flow cell after about 6000 operating hours.
Prerequisite The capillaries are disconnected.
The device is switched off.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
53 Maintenance and Care
1
4
3
1
2
WARNING! Irritation of retina through UV light! Concen­trated UV light can leak out from the flow cell or the fiber optic connectors. Switch off the device or the lamps.
Procedure Steps Figure
1. Press the release
1 down.
lever
2. Remove the flow cell.
3. Insert the new flow cell until it locks into place.
Next steps Connect the capillaries.
Fig. 31 Releasing the flow cell
DAD 6.1L, DAD 2.1L,
MWD 2.1L
DAD 6.1L The recommended operating time of the lamp is 1000 hours.
Legend
1 Deuterium lamp con-
nector
2 Halogen lamp connec-
tor (DAD 6.1L only)
3 Halogen lamp (DAD
6.1L only)
4 Deuterium lamp

Replacing the Lamp

Replace the lamp if it malfunctions or its intensity is low.
The recommended operating time of the lamp is 2000 hours.
Fig. 32 Overview deuterium and halogen lamp
Removing the Deuterium or Halogen Lamp
Prerequisites The device is switched off.
The lamp has cooled down.
Tool Screwdriver Allen wrench, 2.5 mm
Note If the device is still in operation, the integrated security switch
turns off the lamp automatically on opening the lamp cover. An
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
54 Maintenance and Care
1
2
3
4
5
6
error message is displayed. In addition, the red LED lights up and the center LED does not light.
DANGER! Risk of death due to high voltages. Do not change lamps if the lamp area is not volt­age-free. Check the status of the lamps in the software and on the LEDs.
WARNING! Burn hazard from hot lamp! Switch off the lamp meanwhile the device remains switched on for further 15 minutes, oth­erwise the lamp can not cool down. Afterwards, switch off the device and pull the power plug.
Procedure Process Figure
Deuterium lamp
1. Loosen the screws
2 of the lamp
and cover.
2. Remove the lamp cover.
3. Loosen the lock ring 3 or 4 of the lamp plug and pull out the plug.
4. Using the screw­driver, loosen the
two screws
5 or 6.
1
Fig. 33 Lamp cover
Fig. 34 Lock ring of the lamp plug
5. Remove the lamp.
Fig. 35 Screws on the lamp socket
Next steps Insert a new lamp.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
55 Maintenance and Care
1
2
3
7
6
4
5
Installing the Deuterium or Halogen Lamp
Removing the lamp cover switches off the device.
Prerequisites The device is switched off.
The lamp cover has been removed.The old lamp has been removed.
Tool Screwdriver Allen wrench, 2.5 mm
CAUTION! Damage to the lamp and inaccurate measuring results due to residue possible! Do not touch the glass body with your hands! Wear gloves. Use a clean, soft cloth.
Procedure Process Figure
1. Hold the lamp at the lamp socket and guide the glass body into the lamp pod
1.
2. The bolt deuterium lamp sits in a notch in the lamp socket.
Note
Insert the halogen lamp
2 of the
3 into the lamp pod at
a slight angle.
3. Using the screw­driver, tighten the
two screws on the lamp socket.
4. Connect the plug or 6 and tighten
the lock ring.
4 or 7
5
Fig. 36 Lamp pod of the D2 lamp
Fig. 37 Lamp pod of the halogen
lamp
Fig. 38 Installing the deuterium
lamp
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
56 Maintenance and Care
8
9
Procedure Process Figure
5. Attach the lamp cover and fix with
the screws
8 and
9.
Fig. 39 Lamp cover
Next steps Bring the device into operation.

Removing a Leak

Prerequisite If the leak tray is filled with liquid, the leak has to be removed.
Auxiliary material Cloth for drying the leak sensor
Procedure Process
1. Remove the leak
2. Dry the leak tray.
3. Acknowledge the error message via Mobile Control, Con­trol Unit, the software of the standby switch (without soft­ware control).
Next steps Bring the device into operation.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
57 Technical Data

Technical Data

DAD 6.1L

Detection Detector type Diode array detector
Detection channels 4 analog and 8 digital channels
Number of diodes 1024
Diode resolution 0.8 nm/diode
Light source Deuterium (D
GLP chip and halogen lamp
Wavelength range 190–1020 nm
Optical bandwidth < 3.5 nm
Wavelength accu­racy
Wavelength verifica­tion
Wavelength preci­sion
Noise ±3.5 μAU at 254 nm
Drift 300 μAU/h at 254 nm
Linearity > 2.0 AU
Time constants 0 / 0.01 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 /
± 1 nm at 254 nm
Internal holmium filter
≤ 0.5 nm
(ASTM E1657-94)
(ASTM E1657-94)
5.0 / 10.0 s
) lamp with integrated
2
Integration time Automatic
Communication Maximum data rate 100 Hz (LAN)
Inputs Error (IN), Start (IN), Autozero
Outputs Events 1–2, Error (OUT), + 5 V,
Valve + 24 V, Valve (OUT)
Analog outputs 4 × 0-5 V scalable, 16 bit, offset
adjustable
Control Digital: LAN-DHCP, connector
Manual: Control Unit, Mobile Control (optional)
Programming Wake-up
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
58 Technical Data
Technical parameters GLP function Detailed report including lamp recog-
nition, operating hours, lamp operat­ing hours, number of lamp ignitions.
Display Control Unit
Mobile Control
Ambient conditions
General Power supply Input 100–240 V
Temperature range 4–40 °C, 39.2–104 °F
Humidity below 90 %, non-condensing
Output 50–60 Hz
Maximum power consumption
Dimensions 361 mm x 158 mm x 523 mm
Weight 13.8 kg
Leak sensor Yes
Degree of protection IP 20
Height above sea level
75 Watt
(Width × Height × Depth)
maximum 2000 meters

DAD 2.1L

Detection Detector type Diode array detector
Detection channels 4 analog and 8 digital channels
Number of diodes 256
Diode resolution 2.2 nm/diode
Light source Deuterium (D
GLP chip
Wavelength range 190-500 nm
Optical bandwidth < 7 nm
Wavelength accu­racy
Wavelength verifica­tion
Wavelength preci­sion
Noise
± 1 nm at 254 nm
Internal holmium filter
≤ 0.5 nm
±1.0 x 10 (ASTM E1657-94)
) lamp with integrated
2
-5
AU at 254 nm
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
59 Technical Data
Drift
Linearity > 1.8 AU
Time constants 0 / 0.01 / 0.02 / 0.05 / 0.1 / 0.2 / 0.5
Integration time Automatic
Communication Maximum data rate 100 Hz (LAN)
Inputs Error (IN), Start (IN), Autozero
Outputs Events 1–2, Error (OUT), + 5 V,
Analog outputs 4 × 0-5 V scalable, 16 bit, offset
Control Digital: LAN-DHCP, connector
Programming Wake-up
5.0 x 10 (ASTM E1657-94)
/ 1.0 / 2.0 / 5.0 / 10.0 s
Valve + 24 V, Valve (OUT)
adjustable
Manual: Mobile Control (optional)
-4
AU at 254 nm
Technical parameters GLP function Detailed report including lamp recog-
nition, operating hours, lamp operat­ing hours, number of lamp ignitions.
Display Mobile Control
Ambient conditions
General Power supply Input 100–240 V
Temperature range 4–40 °C, 39.2–104 °F
Humidity below 90 %, non-condensing
Output 50–60 Hz
Maximum power consumption
Dimensions 361 mm x 158 mm x 523 mm
Weight 13.8 kg
Leak sensor Yes
Degree of protection IP 20
Height above sea level
75 Watt
(Width × Height × Depth)
maximum 2000 meters

MWD 2.1L

Detection Detector type Variable multi-wavelength UV/VIS
detector
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
60 Technical Data
Detection channels 4 analog and 8 digital channels
Light source Deuterium (D
) lamp with integrated
2
GLP chip
Wavelength range 190-500 nm
Optical bandwidth < 7 nm
Wavelength accu-
≤ 1 nm at 254 nm
racy
Wavelength verifica-
Internal holmium filter
tion
Wavelength preci-
< 0.5 nm
sion
Noise
±1.0 x 10
-5
AU at 254 nm
(ASTM E1657-94)
Drift
5.0 x 10
-4
AU at 254 nm
(ASTM E1657-94)
Linearity > 1.8 AU
Time constants 0 / 0.01 / 0.02 / 0.05 / 0.1 / 0.2 / 0.5
/ 1.0 / 2.0 / 5.0 / 10.0 s
Integration time Automatic
Communication Maximum data rate 100 Hz (LAN)
Inputs Error (IN), Start (IN), Autozero
Outputs Events 1–2, Error (OUT), + 5 V,
Valve + 24 V, Valve (OUT)
Analog outputs 4 × 0-5 V scalable, 16 bit, offset
adjustable
Control Digital: LAN-DHCP, connector
Manual: Mobile Control (optional)
Programming Wake-up
Technical parameters GLP function Detailed report including lamp recog-
nition, operating hours, lamp operat­ing hours, number of lamp ignitions.
Display Mobile Control
Ambient conditions
Temperature range 4–40 °C, 39.2–104 °F
Humidity below 90 %, non-condensing
General Power supply Input 100–240 V
Output 50–60 Hz
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
61 Technical Data
Maximum power consumption
Dimensions 361 mm x 158 mm x 523 mm
Weight 13.8 kg
Leak sensor Yes
Degree of protection IP 20
Height above sea level
75 Watt
(Width × Height × Depth)
maximum 2000 meters

Conditions for Specifications

The technical data has been determined in accordance with the ASTM standard E1657-94.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0

Disposal

62 Disposal
Hand in old devices or disassembled old components at a certi­fied waste facility, where they will be disposed of properly.
AVV marking in Ger-
many
WEEE registration KNAUER as a company is registered by the WEEE number DE
Solvents and other oper-
ating materials
According to the German "Abfallverzeichnisverordnung" (AVV) (January, 2001), old devices manufactured by KNAUER are marked as waste electrical and electronic equipment: 160214
34642789 in the German "Elektroaltgeräteregister" (EAR). The number belongs to category 8.
All distributors and importers are responsible for the disposal of old devices, as defined by the WEEE directive . End-users can send their old devices manufactured by KNAUER back to the dis­tributor, the importer, or the company free of charge, but would be charged for the disposal.
All solvents and other operating materials must be collected sep­arately and disposed of properly.
All wetted components of a device, e. g. flow cells of detectors or pump heads and pressure sensors for pumps, have to be flushed with isopropanol first and water afterwards before being maintained, disassembled or disposed.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
63 Legal Information

Legal Information

Warranty Conditions

The factory warranty for the device is valid for 12 months after the date of dispatch. All warranty claims shall expire in the event that any unauthorized changes are made to the device.
During the warranty period, any components with material or design-related defects will be replaced or repaired by the manu­facturer free of charge.
This warranty excludes the following:
Accidental or willful damageDamage or errors caused by third parties that are not con-
tractually related to the manufacturer at the time the damage occurs
Wear parts, fuses, glass parts, columns, light sources, cuvettes
and other optical components
Damage caused by negligence or improper operation of the
device and damage caused by clogged capillary
Packaging and transport damage
In the event of device malfunctions, directly contact the manu­facturer.
Wissenschaftliche Gerätebau Dr. Ing. Herbert Knauer GmbH Hegauer Weg 38 14163 Berlin, Germany Phone:+49 30 809727-111 Telefax:+49 30 8015010 e-mail:info@knauer.net Internet:www.knauer.net

Transport Damage

The packaging of our devices provides the best possible protec­tion against transport damage. Check the devices for signs of transport damage. In case you notice any damage, contact the Technical Support and the forwarder company within three workdays.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
64 HPLC Glossary

HPLC Glossary

In the following chapter you find abbreviations and terminology that is used in HPLC.
Term Definition
Absorption A substance retaining light of certain wave-
lengths when being shone at.
Adsorption The stationary phase attracting the molecules
of separated substances.
Analytical Qualitative analysis of samples in HPLC (see:
preparative)
Backflush­ing
Calibration Process for correcting measuring values by the
Capillary Thin metal or PEEK pipe that connects the
Carrier material
Chromato­gram
Control Unit
Column The pipe with two end fittings which is perme-
Correction factor
Flushing of columns and pre-columns by reversing the flow direction. This enables strongly retarding substances to be separated.
value that a measuring device deviates from the standard.
components and devices inside of an HPLC sys­tem.
Solid particles which carry the stationary phase (packing material).
The record of a detector signal, depending on the flow volume of the mobile phase or time.
Hand-held device with a touchscreen to drive single AZURA devices locally.
able for solvent. The pipe is filled with packing material.
Factor that arithmetically corrects device­related deviations from measuring values.
Dead vol­ume
Degasser Degasser module for fluids
Detector Device measuring the composition or the
Flow cell Measuring cell
GLP Quality assurance system for laboratories
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Volume of capillary and system components between mixer, injector and column as well as between column and detector.
quantity of a substance
(Good Laboratory Practice)
65 HPLC Glossary
Term Definition
Gradient Time-dependent composition of solvent
Ground Protective measure for electro-conductive
HPLC High Pressure Liquid Chromatography (HPLC)
isocratic Mode of sample separation where the compo-
(mobile phase) on low-pressure or high-pres­sure side of system
housing parts
sition of a solvent remains constant
KNAUER LightGuide
KNAUER LightGuide flow cell cartridges com­bine a maximum light transmission (using total reflection) with a minimum cell volume to offer an ideal signal-to-noise ratio.
Luer-Lock cannula
Mobile Control
Mobile phase
A standardized connector between syringes and cannulas
Mobile application to control and program the whole HPLC system via wireless LAN.
Flowing agent that transports the substances to be separated or isolated through the column (solvent).
Packing material
Solid particles which carry the stationary phase (carrier material).
Peak The deflection of a detector for an analyte in a
chromatogram.
pH Measure for the acidic or alkaline character of
an aqueous solution.
pH elec-
pH sensor
trode
Preparative Isolating the maximum amount of a substance
in a short amount of time in a required purity.
Response time
The time required, after a stepwise change in the composition in the detector cell, for the output signal to reach a predetermined per­centage (67%) of the new equilibration value. For example, if the time constant is 1 s, it will take 1 s to attain 67% if the value of the new signal.
Retention time
The time required from injection of a substance until the maximum concentration of a sub­stance becomes visible.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
66 HPLC Glossary
Term Definition
Sample A mixture of different components which are
to be separated via chromatography. The com­ponents are moved by the mobile phase and dissolved from the column.
Sample loop
A loop which is separate from a chromato­graphic system and which the sample is injected to. After a switch is actuated, the sol­vent flow passes the loop and the sample is flushed onto the column.
Solvent Flowing agent that transports the substances
to be separated or isolated through the column (mobile phase)
Stationary phase
The stationary substance in a chromatographic system on which the substance to be separated settles. It is the liquid or the liquid film on the surface of the packing material.
Tempera­ture com­pensation
The settings under Temperature compensation equalize the changing measuring values with the ambient temperature.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
67 Repeat Orders

Repeat Orders

Note For repeat orders of spare parts use the enclosed packing list.
Contact the Technical Support in case there are any questions on spare parts or accessories.
Further Information Find further information regarding spare parts and accessories at
www.knauer.net.

Modules

Name Order number
DAD 6.1L Diode Array Detector DAD 6.1L without
flow cell
Test cell AMLX8
DAD 2.1L Diode Array Detector DAD 2.1L without
flow cell
Test cell AMLX8
MWD 2.1L Diode Array Detector MWD 2.1L without
flow cell
Test cell AMLX8
ADC11
ADC01
ADB01

Accessories and Spare Parts

Name Order number
User Manual User Manual EN/DE V6700
Operation
Qualification
DAD 6.1L DAD 2.1L MWD 2.1L
VOQDAD61L VOQDAD21L VOQMWD21L
Installation
Qualification
Lamps High-intensity deuterium lamp AZL01
Drainage system Corrugated hose 16 cm, PE gray A9846-1
Capillary guide Capillary guide top P6424
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
DAD 6.1L DAD 2.1L MWD 2.1L
Halogen lamp AZL02
Funnel P6431
Exhaust P6432
Capillary guide side P6425
VIQDAD61L VIQDAD21L VIQMWD21L
68 Repeat Orders
Name Order number
Control Unit Control Unit CU 2.1, incl. user manual (for
Accessories kit AZURA accessories kit FZA02
Tools Tool Kit AZURA A1033
KNAUER LightGuide
Flow Cell Cartridge
KNAUER LightGuide
High Sensitivity
Flow Cell Cartridge
AZD00
DAD 6.1L only)
DAD 6.1L, DAD 2.1L, MWD 2.1L accesso­ries kit
FDC
Flow Cells
Technical Data Order number
Path length Connection Volume Material Max. flow rate Max. pressure
Path length Connection Volume Material Max. flow rate Max. pressure
10 mm 1/16“ 2 μl (0.8 μl dispersion vol.) PEEK, quartz, teflon 5 ml/min 100 bar
50 mm 1/16“ 6 μl (2 μl dispersion vol.) PEEK, quartz, teflon 5 ml/min 100 bar
AMC19
AMD59
Flow Cells with Fiber Optics Connectors
Technical Data Order number
Path length Connection Volume Material
Max. flow rate Max. pressure
Path length Connection Volume Material
Max. flow rate Max. pressure
10 mm 1/16“ 10 μl stainless steel, quartz, PEEK 20 ml/min 300 bar
3 mm 1/16“ 2 μl stainless steel, quartz, PEEK 50 ml/min 300 bar
A4130
A4131
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
69 Repeat Orders
Technical Data Order number
Path length Connection Volume Material Max. flow rate Max. pressure
Path length Connection Volume Material
Max. flow rate Max. pressure
Path length Connection Volume Material
Max. flow rate Max. pressure
Path length Connection Volume Material
Max. flow rate Max. pressure
3 mm 1/16“ 2 μl PEEK, quartz, PTFE 50 ml/min 30 bar
0.5 /1.25 /2 mm 1/8“
1.7/4.3/6.8 μl stainless steel, quartz, PTFE 1000 ml/min 200 bar
0.5 /1.25 /2 mm 1/8“
1.7/4.3/6.8 μl stainless steel, quartz, PTFE 1000 ml/min 100 bar
0.5 /1.25 /2 mm 1/4“
1.7/4.3/6 μl stainless steel, quartz, PTFE 10000 ml/min 200 bar
A4132
A4133
A4134
A4135
Path length Connection Volume Material
Max. flow rate Max. pressure
Path length Connection Volume Material Max. flow rate Max. pressure
Fiber Optics Adapter Necessary for connecting the fiber optics AMKX8
0.5 mm 1/16“
1.7 μl stainless steel, quartz, PEEK 250 ml/min 200 bar
0.5 mm 1/16“
1.7 μl PEEK, quartz, PTFE 250 ml/min 100 bar
A4136
A4137
Fiber Optics
Name Order number
2 x fiber optic cables 750 mm A0740 2 x fiber optic cables, customized A0743
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
70 Table of Figures

Table of Figures

Fig. 1: Front view (example DAD 6.1L) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Fig. 2: Rear view (example DAD 6.1L) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Fig. 3: USB port for the Control Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Fig. 4: Optical system DAD 6.1L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Fig. 5: Fixing points on the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Fig. 6: Inserting the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Fig. 7: Fiber optics adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Fig. 8: Screw-type cap fitting on the fiber optics . . . . . . . . . . . . . . . . . . . . 19
Fig. 9: Fitting of the fiber optics on the flow cell . . . . . . . . . . . . . . . . . . . . 19
Fig. 10: Capillary connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Fig. 11: Flow cell with capillary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Fig. 12: Funnel and capillary guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Fig. 13: Hose and nozzle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Fig. 14: Hose connected to device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Fig. 15: Leak tray with nozzle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Fig. 16: Connectors on rear side . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Fig. 17: Control Unit port on DAD 6.1L . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Fig. 18: Cabling system LAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Fig. 19: Relation between display and Events terminal strip . . . . . . . . . . . . . 26
Fig. 20: Spring strip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Fig. 21: Installation of the detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Fig. 22: Power switch with female connector at the rear panel . . . . . . . . . . . 31
Fig. 23: LEDs and switch/standby button on the front of the device . . . . . . 33
Fig. 24: Cold start temperature profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Fig. 25: Volume comparison . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Fig. 26: Beer-Lambert law . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Fig. 27: Path length comparison . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Fig. 28: Comparison with and without reference wavelength . . . . . . . . . . . 38
Fig. 29: Bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Fig. 30: Noise with different data rates . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Fig. 31: Releasing the flow cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Fig. 32: Overview deuterium and halogen lamp . . . . . . . . . . . . . . . . . . . . . 53
Fig. 33: Lamp cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Fig. 34: Lock ring of the lamp plug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Fig. 35: Screws on the lamp socket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Fig. 36: Lamp pod of the D2 lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Fig. 37: Lamp pod of the halogen lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Fig. 38: Installing the deuterium lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Fig. 39: Lamp cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0

Index

71 Index
A
Absorption 38, 39, 64 Accessories 11 Adapter 19 Additives 9 Adsorption 64 Ambient conditions 16, 51, 57, 58, 59, 60 Analog port 29 analytical 64 AVV marking 62
B
Backflushing 64 Bandwidth 38 Beer-Lambert law 36
C
Calibration 64 Capillary 64
clamping ring 20 connector 20 pre-installed 30
torque 20 Carrier material 64 Chromatogram 64 ClarityChrom 32 ClarityChrom® 50 Cleaning the flow cell 52 Column 64 Connect capillary 20 Contact data 50 Control 22, 32
chromatography software 32
Control Unit 22, 32
LAN 22 Control Unit 22, 32, 50, 64 Control with
Mobile Control Correction factor 64 Customer support 50 Cut off wavelength 38
32
D
D2 lamp
replace Data rate 8, 39 Dead volume 37, 64 Decontamination 14
53
Degasser 10, 64 Detector 64
optimize 35 rear view 7
technical data 57 Device Overview 6 Disposal 62 Drift 37
E
Electrical connections
Events terminal strip
25
F
Features 7 Fiber optics 69
adapter 19
flow cell 19 Fittings 51 Flow Cell
replacing Flow cell 8, 64
clean 52
insert 18
material 37
pressure stability 37
select 35
volume 35 Flow cells
fiber optics Front View 6 Functionality Test
Operation Qualification (OQ) Functionality Tests 42
Installation Qualification (IQ) 42
52
68
G
General 41 GLP 64 GLP data 9 Gradient 65 Ground 65 GROUND, see pin header 27
H
Halogen lamp
replace HPLC 65
53
42
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
72 Index
I
Initial start-up 31 Installation qualification 42 Installation site 16 Integration time 41 Integrator, see analog port 29 Intended Use 6 Isocratic 65
K
KNAUER LightGuide flow cell cartridge 8
L
Lamp
replace Lamp mirror 8 Lamps 8 LAN 22
connection problems 43
port 25
router 24
settings 23
setup 24 Leak 13
remove 56 Leak Management 8 Leak management
connect Leak sensor 56 LED 33
color 33 Light intensity 36 Luer-lock cannula 65
53
21
M
Mains
disconnect Maintenance 49
maintenance contract 50 Maintenance intervals 50 Maximum pressure 37 Mobile Control 32, 50, 65 Mobile phase 65 Modifiers 9
17
O
OpenLAB 32 OpenLAB® 50 Operating environment 16 Operating hours 50 Operation 31
Operation qualification 42 Optical system 7 Optimization
select the wavelength Optimizing the detector 35 Optional accessories 9 Order numbers 67
38
P
Packing list 18 Packing material 65 Path Length 36 Path length 37 Peak 65 Peak broadening 35 PEEK 9 PEEK Fittings 20 pH 65 pH electrode 65 Pin header
Autozero
Start-IN port 27 Port (LAN) 25 Power consumption 58, 59, 61 Power plug 17 Power supply 13, 29
connect 29
power cable 13
power strip 13 preparative 65 Pressure
maximum Pressure Stability 37 Professional group 12
27
37
R
Rear View 6 Reference wavelength 38, 39 Refractive index 38 Remote operation 37 Repeat orders 67
fiber optics 69
flow cells 68 Re-Setting 34 Response time 65 Retention time 65 Router (LAN) 24
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
73 Index
S
Safety 12 Safety equipment 12 Salts 9 Sample 66 Sample loop 66 Selectivity 39 Self-calibration 31 Sensitivity 37, 39 Service 50 Setup 17 Side View 7 Signal sensitivity 35 Signal wavelength 38, 39 Software 50 Solvent 66
flammability 13 line 13 self-ignition point 13 toxicity 13 tray 13
Solvents 9
suitable 9 Space Requirements 17 Space requirements 16 Spare parts 11, 67 Standby 33 Standby button 33 Start IN, see pin header 27 Stationary phase 66 Storage 49, 51 Subtraction of the baseline 41 Switch-On 31 Symbols and Signs 15 System integration 30 System Messages 44
T
Technical Data 57
DAD 2.1L 58 DAD 6.1L 57 MWD 2.1L 59
Technical data
detector Technical Support 50 Temperature compensation 66 Tests
Intervals Time constant 39 Transport damage 63 Troubleshooting 43
LAN 43
system messages 44
57
42
U
Unpacking 17 USB port 7
W
Warm-Up Time 35 Warranty 63 Wavelength
cut-off
select 38
38
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Declaration of Conformity
Manufacturer Wissenschaftliche Gerätebau
Dr. Ing. Herbert Knauer GmbH
Hegauer Weg 38
14163 Berlin, Germany
Product: Detector DAD 6.1L, DAD 2.1L,
MWD 2.1L
Order number EDC11 EDC01 EDB01
Designation 1
Designation 2
The product complies with the following requirements and product specifications:
Europe
DAD 6.1L DAD 2.1L MWD 2.1L
Diode array detector DAD 6.1L Diode array detector DAD 2.1L Variable multi wavelength
 DIRECTIVE 2006/42/EC OF THE EUROPEAN PARLIAMENT
AND OF THE COUNCIL of 17 May 2006 on machinery, and amending Directive 95/16/EC (recast)
 IEC 60799 (June 1998) Electrical accessories - Cord sets and
interconnection cord sets
 IEC 61010-1 (2010 + Corrigendum: 2011) Safety require-
ments for electrical equipment for measurement, control and laboratory use
 IEC 61010-2-081 (2001 + A1:2003) Safety requirements
for electrical equipment for measurement, control and laboratory use - Part 2-081: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes
UV/VIS detector MWD 2.1L
 RoHS directives 2002/95/EC (2003) and 2011/65/EU on
the restriction of the use of certain hazardous substances in electrical and electronic equipment
 WEEE directive 2002/96/EC (2003) and 2012/19/EU on
waste electrical and electronic equipment
Canada
 UL 61010 - 1: 2012 (2012-05-11) Safety requirements for
electrical equipment for measurement, control, and labora­tory use - Part 1: General requirements
USA
 CAN/CSA - C22.2 NO. 61010-1-12 (2012) - Safety require-
ments for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
The product was tested with a typical configuration. Berlin, 2014-08-15
 IEC 61010-2-010 (2003) Safety requirements for electrical
equipment for measurement, control, and laboratory use ­Part 2-010: Particular requirements for laboratory equip­ment for the heating of materials
 Low voltage directive (2006/95/EC)
 IEC 61000-3-2 (2005 + A1:2008 + A2:2009) Electromag-
netic compatibility (EMC) Part 3-2
 EMC standard (2004/108/EC)
 IEC 61326-1 (2012) Electrical equipment for measurement,
control and laboratory use – EMC requirements – Part 1: General requirements
 Directives for an environmentally sound use of electrical and
electronic equipment
Alexandra Knauer (CEO and owner)
The mark of conformity has been applied to the rear panel of the device.
3 Inhaltsverzeichnis

Inhaltsverzeichnis

Hinweis Lesen Sie unbedingt zu Ihrer eigenen Sicherheit das Handbuch
und beachten Sie immer die auf dem Gerät und im Handbuch angegebenen Warn- und Sicherheitshinweise!
Bestimmungsgemäße Verwendung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Geräteübersicht . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Vorderansicht . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Rückansicht . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Seitenansicht . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Leistungsübersicht . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Eluenten . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Lieferumfang . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Sicherheit für Anwender . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Definition möglicher Personen- oder Sachschäden . . . . . . . . . . . . . . . . . . . . . 14
Dekontamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Symbole und Kennzeichen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Kontakt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Betriebsumgebung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Aufstellort und Platzbedarf . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Platzbedarf . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Aufbau . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Auspacken . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Durchflusszelle einsetzen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Durchflusszelle mit Lichtwellenleiter einbauen . . . . . . . . . . . . . . . . . . . . . . 19
Kapillare anbringen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Einsatz von PEEK-Verschraubungen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Leckmanagement anschließen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Steuerung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Gerät an den Computer anschließen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
LAN-Eigenschaften einstellen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Geräte zum LAN verkabeln . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Router einstellen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
LAN in das Firmennetzwerk integrieren . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Mehrere Systeme in einem LAN separat steuern . . . . . . . . . . . . . . . . . . . . 26
Anschlüsse der Stiftleiste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Stiftleiste verkabeln . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Integrator-Anschluss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Stromanschluss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Den Detektor in ein System integrieren . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Werkseitig installierte Kapillaren . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
4 Inhaltsverzeichnis
Betrieb . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Erstinbetriebnahme . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Einschalten . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Bedienung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Bedienung mit Chromatografie-Software . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Bedienung mit Control Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Bedienung mit Mobile Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Bedeutung der LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Gerät zurücksetzen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Detektor optimieren . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Einsatzort . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Aufwärmzeit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Durchflusszelle auswählen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Volumen der Durchflusszelle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Schichtdicke . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Druckstabilität . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Verbindungsart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Remote-Betrieb (LWL) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Wellenlänge auswählen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Bandbreite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Spektralbereich . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Zeitkonstante & Datenrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Integrationszeit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Basislinie subtrahieren . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Sonstiges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Funktionstests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Installation Qualification (IQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Operation Qualification (OQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Fehlerbehebung. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
LAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Mögliche Probleme und Abhilfen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Systemmeldungen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Wartung und Pflege . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Kontakt zur Technischen Kundenbetreuung . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Wartungsvertrag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Wartungsintervalle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Gerät reinigen und pflegen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Verschraubungen kontrollieren . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Außer Betrieb setzen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Lagerung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Stromanschluss trennen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Durchflusszelle reinigen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Durchflusszelle spülen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
5 Inhaltsverzeichnis
Durchflusszelle austauschen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Lampe austauschen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Deuterium- oder Halogenlampe ausbauen . . . . . . . . . . . . . . . . . . . . . . . . . 57
Deuterium- oder Halogenlampe einbauen . . . . . . . . . . . . . . . . . . . . . . . . . 58
Leck beseitigen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Technische Daten. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
DAD 6.1L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
DAD 2.1L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
MWD 2.1L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Spezifikationsbedingungen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Entsorgung. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Rechtliche Hinweise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Gewährleistungsbedingungen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Transportschäden . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
HPLC-Glossar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Nachbestellungen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Geräte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Durchflusszellen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Durchflusszellen mit Lichtwellenleiteranschlüsse . . . . . . . . . . . . . . . . . . . . . 72
Lichtwellenleiter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Abbildungsverzeichnis. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Stichwortverzeichnis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
An alle, die es betrifft Wenn Sie ein französischsprachiges Benutzerhandbuch zu diesem
Produkt wünschen, senden Sie ihr Anliegen und die entsprechende Seriennummer per E-Mail oder Fax an KNAUER:
support@knauer.net+49 30 8015010
Vielen Dank.
A qui que ce soit Si jamais vous préfériez un manuel en francais pour ce poduit cont-
acter KNAUER par email ou par fax avec le no. de série:
support@knauer.net+49 30 8015010
Merci beaucoup.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
6 Bestimmungsgemäße Verwendung
8
7
2
3 2
2
4
1
6
5

Bestimmungsgemäße Verwendung

Hinweis Das Gerät ausschließlich in Bereichen des bestimmungsgemä-
ßen Betriebs einsetzen. Andernfalls können die Schutz- und Sicherheitseinrichtungen des Geräts versagen.
Beschreibung Der Detektor ist für analytische, semipräparative und präparative
Anwendungen geeignet.
Einsatzbereiche Das Gerät kann in folgenden Bereichen eingesetzt werden:
Biochemische AnalytikChemische AnalytikLebensmittelanalytikPharmazeutische AnalytikUmweltanalytik
Legende
1 Leckwanne
2 Kapillarführung
3 Lampenklappe
4 Seriennummer
5 Leckmanagement
6 Schalter/Standby-Taste
7 Status LED
8 Durchflusszelle

Geräteübersicht

Vorderansicht
Auf der Vorderseite des Geräts befindet sich die Durchflusszelle und die Lampenklappe zum Austausch der Lampen.
Abb. 1 Vorderansicht (Beispiel DAD 6.1L)
Rückansicht
RS-232, Service-Schnitt-
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
stelle
Auf der Rückseite des Detektors finden Sie neben den Anschlüs­sen für die LAN-Verbindung und den Netzstecker auch die Seri­ennummer sowie den Netzschalter zum Ein- und Ausschalten des Gerätes.
Die Service-Schnittstelle ist ausschließlich für Wartungs- und Reparaturarbeiten durch einen Service-Techniker zu verwenden.
Legende
2345
6
71
1
1 Geräteinformatio-
nen
2 Integratorausgänge
3 Anschluss LAN
4 Schnittstelle für den
Service
5 Stiftleiste
6 Lüfter
7 Netzanschluss mit
Netzschalter
7 Bestimmungsgemäße Verwendung
Abb. 2 Rückansicht (Beispiel DAD 6.1L)
Seitenansicht
DAD 6.1L An der Seite des Detektors befindet sich der Anschluss, mit dem
Hinweis Der USB-Anschluss für den DAD 2.1L und den MWD 2.1L ist
Legende
1 USB-Anschluss für
die Control Unit
die Bedieneinheit Control Unit (siehe ’Bedienung mit Mobile Control’ auf Seite 33.) verbunden wird.
nicht aktiv. Die Detektoren können nicht mit der Control Unit gesteuert werden.
Abb. 3 USB-Anschluss für die Control Unit

Leistungsübersicht

Der Detektor dient in einem HPLC-System dazu, Stoffe in Flüs­sigkeiten nachzuweisen und ihre Konzentration anzugeben. Die Empfindlichkeit des Detektors hängt von der eingebauten Durchflusszelle ab.
Optisches System Die Detektoren messen eine Lichtabsorption durch die Probe im
ultravioletten und visuellen Bereich. Das gesamte Spektrum wird durch die Probe gesendet und danach durch ein Gitter aufge­spalten. Das aufgespaltene Licht fällt auf eine geometrische Anordnung von 1024 (DAD 6.1L) oder 256 (DAD 2.1L, MWD
2.1L) einzelnen Photodioden, dem Photodiodenarray.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
Legende
3
5
1
2
4
6
1 Deuteriumlampe
2 Halogenlampe
3 Durchflusszelle
4 Diodenarray-Sensor
5 Diffraktionsgitter
6 Spiegel
8 Bestimmungsgemäße Verwendung
Das optisches System des AZURA DAD 6.1L Detektors ist in Abbildung 4 dargestellt. Eine aktive Temperaturkontrolle mit 0,5°C Empfindlichkeit sichert gute Basislinienstabilität für eine minimale Basisliniendrift.
Abb. 4 Optisches System DAD 6.1L
Lampen Beide Lampen können ohne Service Techniker schnell und ein-
fach von vorne gewechselt werden.
DAD 6.1L Zwei Lichtquellen, eine Deuterium- und eine Halogenlampe
decken den gesamten UV-Vis Wellenlängenbereich von 190–1020 nm ab.
DAD 2.1L, MWD 2.1L Eine Deuteriumlampe deckt den UV Wellenlängenbereich von
190–500 nm ab.
Lampenspiegel
DAD 6.1L
Durchflusszelle Eine Vielzahl unterschiedlicher Durchflusszellen für verschiedene
KNAUER LightGuide
Messzellkartuschen
Eine effektive Lichtmischung der Deuterium- und Halogenlam­pen durch einen Spiegel mit Polka Dot Technologie ermöglicht eine maximale Lichtintensität über den gesamten Wellenlängen­bereich.
Anwendungen stehen zur Verfügung. Messzellkartuschen mit Standard und KNAUER LightGuide Technologie sind erhältlich.
KNAUER LightGuide Messzellkartuschen kombinieren eine maxi­mierte Lichtausbeutel (durch Totalreflexion) mit einem minima­len Zellvolumina um ein optimales S/N-Verhältnis zu gewährleisten.
Adapter für Lichtwellenleiter sind zusätzlich erhältlich. Lichtwel­lenleiter erlaubten die räumliche Trennung von Messzelle und Detektor. Dies erhöht die Sicherheit bei Arbeitsprozessen mit gefährlichen, explosiven oder toxischen Materialien.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
9 Bestimmungsgemäße Verwendung
Datenrate Der Detektor misst mit einer maximalen Datenrate von 100 Hz.
Dabei werden pro Sekunde ca.100 Spektren aufgenommen.
Leckmanagement Eingebautes Leckmanagement zum Ableiten von austretenden
Flüssigkeiten.
GLP-Daten Mit Hilfe der Mobile Control sowie der verschiedenen Software-
pakete (ClarityChrom
®
oder OpenLAB
®
) ist es möglich, die GLP­Daten wie Betriebsstunden, Lampen-Betriebsstunden, Anzahl der Lampenzündungen, etc. aus dem Detektor anzuzeigen oder auszulesen. Dem DAD 6.1L steht zusätzlich die Control Unit zum Auslesen der GLP-Daten zur Verfügung. Eine detaillierte Beschreibung zum Anzeigen oder Auslesen der GLP-Daten ist im jeweiligen Benutzerhandbuch zu finden.
Weiteres Zubehör Als weiteres Zubehör sind folgende Artikel erhältlich:
Die Mobile Control App, für Android und Windows 8, zur
Bedienung
Die Control Unit zur Bedienung des DAD 6.1L

Eluenten

Schon die Zugabe kleiner Mengen anderer Substanzen wie Additive, Modifier oder Salze können die Beständigkeit der Materialien beeinflussen.
In Zweifelsfällen kontaktieren Sie die technische Kundenbetreuung.
Geeignete Eluenten
Acetatpufferlösungen
Aceton bei 4 °C–25 °C (39,2 °C–77,0 °F)
Acetonitril
2
BenzolChloroformEssigsäureethylesterEthanolFormiatpufferlösungenIsopropanolKohlendioxid (flüssiges 99,999 % CO
MethanolPhosphatpufferlösungen (0,5 M)Toluolverdünnte ammoniakalische Lösungverdünnte Essigsäure (z. B. 0,1–1 %) bei 25 °Cverdünnte Natronlauge (1 M)Wasser
1
)
2
1) gültig im angegebenen Temperaturbereich
2) nicht in Kombination mit PEEK-Kleinteilen oder PEEK-Kapillaren zu empfehlen
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
10 Bestimmungsgemäße Verwendung
Bedingt geeignete Eluenten
Diethylamin (0,1 %) (DEA)Dimethylsulfoxid (DMSO)leicht flüchtige Eluenten
MethylenchloridTetrahydrofuran (THF)
1
1
verdünnte PhosphorsäureTriethylamin (0,1 %) (TEA)Trifluoressigsäurelösung (0,1 %) (TFA)
1) nicht in Kombination mit PEEK-Kleinteilen oder PEEK-Kapillaren zu empfehlen
Nicht geeignete Eluenten
Halogenkohlenwasserstoffe, z. B. Freonkonzentrierte mineralische und organische Säurenkonzentrierte LaugenPartikelhaltige Eluenten
Perfluorierte Eluenten, z. B. FluorinertPerfluorierte Polyether, z. B. Fomblin
®
®
FC-75, FC-40
®
Für Degasser nicht geeignete Eluenten
AzideBenzolFlurkohlenwasserstoffe
Halogenkohlenwasserstoffe, z. B. Freon
®
HeptanHexafluoroisopropanol (HFIP)HexanFlusssäure-LösungenKohlendioxid (flüssiges 99,999 % CO
)
2
konzentrierte mineralische und organische Säurenkonzentrierte LaugenPartikelhaltige Eluenten
®
Perfluorierte Eluenten, z. B. FluorinertPerfluorierte Polyether, z. B. Fomblin
FC-75, FC-40
®
Salzsäureverdünnte Natronlauge (1 M)
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
11 Lieferumfang

Lieferumfang

Hinweis Ausschließlich Ersatzteile und Zubehör vom Hersteller oder einer
DAD 6.1L Detektor
Beipack DAD 6.1L
DAD 2.1L Detektor
vom Hersteller autorisierten Firma verwenden.
Netzkabel
Benutzerhandbuch deutsch und englisch
Installation Qualification-Dokument
AZURA
Netzkabel
Benutzerhandbuch deutsch und englisch
Installation Qualification-Dokument
Beipack DAD 2.1L
AZURA
MWD 2.1L Detektor
Netzkabel
Benutzerhandbuch deutsch und englisch
Installation Qualification-Dokument
Beipack MWD 2.1L
AZURA
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
12 Sicherheit für Anwender

Sicherheit für Anwender

Berufsgruppe Das Benutzerhandbuch richtet sich an Personen, die mindestens
eine Ausbildung zum Chemielaboranten oder einen vergleichba­ren Ausbildungsweg abgeschlossen haben.
Folgende Kenntnisse werden vorausgesetzt:
Grundlagenkenntnisse der FlüssigchromatografieKenntnisse über Substanzen, die nur bedingt in der Flüs-
sigchromatografie eingesetzt werden dürfen
Kenntnisse über die gesundheitlichen Risiken beim Umgang
mit Chemikalien
Teilnahme an der Installation eines Geräts oder einer Schu-
lung durch die Firma KNAUER oder einer von KNAUER autori­sierten Firma
Gehören Sie nicht zu dieser oder einer vergleichbaren Berufs­gruppe, dürfen Sie die in diesem Benutzerhandbuch beschriebe­nen Arbeiten auf keinen Fall ausführen. Informieren Sie in diesem Fall Ihre Vorgesetzte oder Ihren Vorgesetzten.
Schutzausrüstung Bei allen Arbeiten an dem Gerät sind die im Labor notwendigen
Schutzmaßnahmen zu beachten und folgende Schutzkleidung zu tragen:
Schutzbrille mit zusätzlichem SeitenschutzSchutzhandschuheLaborkittel
Was ist zu beachten? Alle Sicherheitshinweise im Benutzerhandbuch
Die Umgebungs-, Aufstell- und Anschlussbestimmungen im
Benutzerhandbuch
Bei der Arbeit mit Lösungsmitteln den Raum immer gut lüf-
ten
Nationale und internationale Vorschriften für das Arbeiten im
Labor
Vom Hersteller empfohlene oder vorgeschriebene Originaler-
satzteile, Werkzeuge und Eluenten
Good Laboratory Practice (GLP)Unfallverhütungsvorschriften der Unfallkrankenkassen für
Labortätigkeiten
Aufreinigung der zu analysierenden SubstanzenEinsatz von Inline-FilternKeine gebrauchten Kapillaren an anderer Stelle im HPLC-Sys-
tem einsetzen
PEEK-Verschraubungen nur für ein- und denselben Port ver-
wenden oder grundsätzlich neue PEEK-Verschraubungen ein­setzen
Hinweise von KNAUER oder anderer Hersteller zur Säulen-
pflege beachten
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
13 Sicherheit für Anwender
Weitere für Ihre Sicherheit wichtige Themen sind in der folgen­den Tabelle alphabetisch sortiert:
Thema Erläuterungen
Entflammbarkeit Organische Eluenten sind leicht ent-
flammbar. Keine offenen Flammen in der Nähe des Geräts betreiben, da Kapillaren sich aus der Verschraubung lösen kön­nen, und dann eventuell leicht entflamm­barer Eluent austritt.
Flaschenwanne Es besteht die Gefahr eines Stromschlags,
falls Eluenten oder andere Flüssigkeiten in das Innere des Geräts gelangen. Deshalb immer eine Flaschenwanne verwenden.
Flüssigkeits- leitungen
Kapillare und Schläuche so verlegen, dass beim Auftreten von Lecks austretende Flüssigkeiten nicht in darunter angeord­nete Geräte eindringen können.
Lecks Regelmäßige Sichtkontrolle des Anwen-
ders auf Undichtigkeit im System wird empfohlen.
Netzkabel Beschädigte Netzkabel dürfen nicht für
den Anschluss der Geräte an das Strom­netz benutzt werden.
Selbstent- zündungs- temperatur
Ausschließlich Eluenten verwenden, die unter normalen Raumbedingungen eine Selbstentzündungstemperatur höher als 150 °C haben.
Steckdosenleiste Beim Anschluss von mehreren Geräten an
eine einzige Steckdosenleiste immer die maximal zulässige Stromaufnahme der Geräte beachten.
Stromversorgung Geräte dürfen nur an zugelassene Span-
nungsquellen angeschlossen werden, deren Spannung mit der zulässigen Span­nung des Geräts übereinstimmt.
Toxizität Organische Eluenten sind ab einer
bestimmten Konzentration toxisch. Arbeitsraum immer gut belüften! Beim Arbeiten am Gerät Schutzhandschuhe und Schutzbrille tragen!
Wo darf das Gerät nicht
eingesetzt werden?
Das Gerät darf ohne besonderen und zusätzlichen Explosions­schutz nicht in explosionsgefährdeten Bereichen betrieben wer­den. Weitere Informationen erhalten Sie von der technischen Kundenbetreuung von KNAUER.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
14 Sicherheit für Anwender
Gerät sicher außer
Betrieb nehmen
Gerät öffnen Gerät ausschließlich von der technischen Kundenbetreuung von
Das Gerät lässt sich jederzeit durch Ausschalten am Netzschalter oder durch Lösen des Netzanschlusses vollständig außer Betrieb nehmen.
KNAUER oder einer von KNAUER autorisierten Firma öffnen las­sen.

Definition möglicher Personen- oder Sachschäden

Möglichen Gefahren, die von einem Gerät ausgehen können, werden in dem vorliegenden Benutzerhandbuch in Personen­oder Sachschäden unterschieden.
Kategorie Erläuterungen
GEFAHR! Lebensgefahr oder sehr schwere Verletzun-
gen sind möglich.
WARNUNG! Schwere bis mittlere Verletzungen sind
möglich.
VORSICHT! Leichte bis sehr leichte Verletzungen sind
möglich. Ein Defekt des Geräts ist möglich.

Dekontamination

Die Kontamination von Geräten mit toxischen, infektiösen oder radioaktiven Substanzen sind sowohl in Betrieb, bei der Repara­tur, beim Verkauf als auch bei der Entsorgung eines Gerätes eine Gefahr für alle Personen.
GEFAHR! Gefahr durch den Kontakt mit toxischen, infektiö­sen oder radioaktiven Substanzen. Bevor Geräte entsorgt oder zur Reparatur ver­schickt werden, müssen sie fachgerecht dekonta­miniert werden.
Alle kontaminierten Geräte müssen von einer Fachfirma oder selbständig fachgerecht dekontaminiert werden, bevor diese wieder in Betrieb genommen, zur Reparatur, zum Verkauf oder in die Entsorgung gegeben werden.
Alle zur Dekontamination verwendeten Materialien oder Flüssig­keiten müssen getrennt gesammelt und fachgerecht entsorgt werden.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
15 Symbole und Kennzeichen

Symbole und Kennzeichen

Die folgenden Symbole und Kennzeichen befinden sich am Gerät, in der Chromatografie-Software oder im Benutzerhand­buch.
Symbol Bedeutung
Warnzeichen Gefährdung durch Hochspannung.
Gefährdung durch Stromschlag.
Gefährdung durch potentiell toxische Substanzen.
Mögliche Gefährdung der Augen durch UV-Licht.
Mögliche Verbrennungen beim Berühren von heißen Bauteilen.
Hinweis auf eine Situation, die, wenn sie nicht vermieden wird, zu leichten Verlet­zungen führen kann. Hinweis auf eine Situation, die, wenn sie nicht vermieden wird, zu Sachschäden führen kann.
Möglicher Sachschaden durch elektrosta­tische Entladung am System, am Gerät oder an bestimmten Bauteilen.
Maximale Gewichtsbelastung der Leck­wanne während des Transports, der Installation und im Betrieb beachten.
Gebotszeichen Schutzhandschuhe schützen vor Haut-
schäden.
Konformitätszeichen Das mit dem CE-Zeichen gekennzeich-
nete Gerät oder System erfüllt die pro­duktspezifisch geltenden europäischen Richtlinien. Dies wird in der Konformitäts­erklärung bestätigt.
Das Prüfsiegel einer in Kanada und den USA national anerkannten Prüfstelle (NRTL). Das zertifizierte Gerät oder Sys­tem hat die Prüfungen auf Qualität und Sicherheit erfolgreich bestanden.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0

Installation

16 Installation
Das Kapitel Installation beschreibt diejenigen Handlungen, die Sie vor der Inbetriebnahme beachten sollen. Wenn bei Ihnen während der Installation Schwierigkeiten auftreten, setzen Sie sich mit der Technischen Kundenbetreuung in Verbindung

Kontakt

Technischen Kunden-
betreuung
Gerätedaten Berücksichtigen Sie für einen sicheren Transport das Gewicht
Heben Das Gerät seitlich am Gehäuse umfassen und aus der Verpa-
Telefon +49 30 809727-111
Fax +49 30 8015010
E-Mail support@knauer.net

Transport

Bereiten Sie das Gerät sorgfältig auf den Transport oder die Lagerung vor. Wenn Sie das Gerät zur Reparatur an KNAUER verschicken wollen, legen Sie den Servicebegleitschein bei, der zum Download auf der Website bereitsteht.
und die Maße des Geräts (siehe Technische Daten).
VORSICHT! Schäden am Gerät durch Heben an hervorstehen­den Geräteteilen möglich! Gerät ausschließlich seitlich am Gehäuse hochhe­ben.
ckung heben.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
17 Installation
1

Betriebsumgebung

Der bestimmungsgemäße Betrieb ist nur gewährleistet, wenn Sie sich an die Vorgaben für die Umgebungsbedingungen des Einsatzorts halten. Die Umgebungsbedingungen finden Sie im Kapitel Technische Daten.
VORSICHT! Gerätedefekt durch Überhitzung möglich! Gerät vor Sonneneinstrahlung schützen. Auf der Rückseite des Geräts mindestens 15 cm und auf der Seite der Lüftungsschlitze mindes­tens 5–10 cm Platz für die Luftzirkulation lassen.
Aufstellort und Platzbedarf
Das Gerät auf eine ebene und gerade Fläche stellen.Sonneneinstrahlung: Das Gerät vor direkter Sonneneinstrah-
lung schützen.
Klimaanlage: Das Gerät frei von Zugluft aufstellen.Vibration: Das Gerät nicht neben Maschinen platzieren, die
Bodenvibrationen verursachen.
Seitlicher Abstand zu weiteren Geräten:
Mindestens 5 cm, wenn auf einer Seite ein weiteres Gerät
aufgestellt wird
Mindestens 10 cm, wenn auf beiden Seiten ein weiteres
Gerät aufgestellt wird
Netzstecker auf der Geräte-Rückseite frei zugänglich hal-
ten, damit das Gerät vom Stromnetz getrennt werden kann.
Gerät von Hochfrequenzquellen fern halten. Hochfrequenz-
quellen können die Messwerte beeinflussen.
Erdbebengebiet In Erdbebengebieten das Gerät an den Fixierungspunkten
befestigt. Die Fixierungspunkte befinden sich rechts oder links am Gerät.
Abb. 5 Fixierungspunkte am Gerät
1
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
18 Installation
Platzbedarf
Seitlicher Abstand zu weiteren Geräten:
mindestens 5 cm, wenn auf einer Seite ein weiteres Gerät
aufgestellt ist.
mindestens 10 cm, wenn auf beiden Seiten ein weiteres
Gerät aufgestellt ist.
Mindestabstand 15 cm zum Lüfter auf der Geräterückseite.

Aufbau

Bevor Sie das Gerät aufbauen, sollen Sie einen geeigneten Ein­satzort gemäß der Vorgaben auswählen und die Verpackung entfernen. Die Vorgaben und eine Handlungsbeschreibung fin­den Sie im nachfolgenden Abschnitt.
Auspacken
Voraussetzung Der Karton wurde auf Transportschäden geprüft.
Werkzeug Cuttermesser
Ablauf
1. Die Verpackung so aufstellen, dass die Schrift am Etikett richtig herum steht. Das Klebeband mit einem Cutter­messer durchtrennen und die Verpackung öffnen.
2. Die Schaumstoffauflage abheben. Die Zubehörteile und das Benutzerhandbuch herausnehmen.
3. Die Zubehörteile aus der Tüte nehmen und den Lieferum­fang prüfen. Im Fall einer unvollständigen Lieferung die Technische Kundenbetreuung kontaktieren.
4. Das Gerät von unten umfassen, aus der Verpackung heben und auf den Füßen abstellen. Dabei nicht an der vorderen Abdeckung festhalten.
5. Das Gerät auf Transportschäden prüfen. Im Fall einer Beschädigung die Technische Kundenbetreuung kontaktie­ren.
6. Das Gerät am Einsatzort platzieren und die Schutzfolien abziehen.
VORSICHT! Schäden am Gerät durch Heben an hervorstehen­den Geräteteilen möglich! Gerät ausschließlich seitlich am Gehäuse hochhe­ben.
Nächste Schritte Lagern Sie Karton und Verpackung und bewahren Sie die beilie-
gende Packliste für spätere Nachbestellungen auf.
Hinweis Die Testzelle wird für Funktionstests benötigt und muss sorgfäl-
tig aufbewahrt werden.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
19 Installation

Durchflusszelle einsetzen

Voraussetzung Es ist keine Durchflusszelle oder eine Testzelle eingesetzt.
Das Gerät ist ausgeschaltet.
Vorgehensweise Die Durchflusszelle einsetzen
Hinweis Achten Sie darauf, dass kompatible Durchflusszellen eingesetzt
werden.
WARNUNG! Reizung der menschlichen Netzhaut durch UV­Licht. Gebündeltes UV-Licht kann an der Durch­flusszelle oder am Lichtwellenleiter austreten. Das Gerät oder die Lampen ausschalten!
Vorgehensweise Ablauf Abbildung
1. Die Durchflusszelle in die Öffnung setzen und nach hinten schieben bis sie einrastet.
Nächste Schritte Schließen Sie die Kapillaren an.
Durchflusszelle mit Lichtwellenleiter einbauen
Sie können die Durchflusszelle auch mit Lichtwellenleitern ver­binden. Die Lichtwellenleiter mit der Durchflusszelle werden an einen Adapter angeschlossen.
UV-Licht lässt Lichtwellenleiter im Laufe der Zeit solarisieren, sodass diese für den Einsatz nicht mehr geeignet sind.
Beim Benutzen von Lichtwellenleitern folgende Punkte beach­ten:
Die Enden des Lichtwellenleiters nicht mit den Fingern berüh-
ren, da dies zu fehlerhaften Messungen führen könnte.
Stöße und hartes Aufschlagen des Lichtwellenleiters vermei-
den.
Lichtwellenleiter vorsichtig ohne Druck und ohne Verbiegen
bewegen.
Abb. 6 Durchflusszelle einset-
zen
Voraussetzungen Es ist keine Durchflusszelle oder Testzelle eingesetzt.
Hilfsmittel Blindverschraubungen für die Lichtwellenleiter.
WARNUNG! Reizung der menschlichen Netzhaut durch UV­Licht. Gebündeltes UV-Licht kann an der Durch­flusszelle oder am Lichtwellenleiter austreten. Das Gerät oder die Lampen ausschalten!
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
20 Installation
1
1
1
3
Vorgehensweise
Ablauf Abbildung
1. Die Endstücke LWL-Adapter abziehen.
2. Den LWL-Adapter in die Öffnung setzen und nach hinten schieben bis er einrastet.
3. Die Blindverschraubun­gen
1 von den Enden
des Lichtwellenleiter abziehen.
4. Die Verschraubungen der Lichtwellenleiter von Hand am LWL-
Adapter ben.
5. Die Endstücke der Durchflusszelle abzie­hen.
6. Die Verschraubungen
1 am
Abb. 7 LWL-Adapter
Abb. 8 Blindverschraubung der
Lichtwellenleiter
2 anschrau-
3 der Lichtwellenleiter
an der Durchflusszelle von Hand anschrauben.
Abb. 9 Verschraubungen der
Lichtwellenleiter an der Durchflusszelle
Nächste Schritte Schließen Sie die Kapillaren an.

Kapillare anbringen

Die Kapillaren verbinden den Detektor mit der Säule und leiten die Flüssigkeiten. Für die Anschlüsse der Durchflusszelle empfeh­len wir PEEK-Kapillare und PEEK-Verschraubungen.
VORSICHT! Schäden an den PEEK-Gewinden der Durchfluss­zelle durch Metall-Verschraubungen möglich. Kapillaren ausschließlich mit PEEK-Verschraubun­gen anbringen. Drehmoment von maximal 0,5 Nm beachten.
Voraussetzung Die Durchflusszelle ist eingesetzt.
Blindverschraubungen und Endstücke sind entfernt.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
21 Installation
132
4
13
2
4
4
4
Vorgehensweise Die Kapillare mit Verschraubung anbringen.
Vorgehensweise
Ablauf Abbildung
1. Kapillare Verschraubung
schieben.
2. Den Klemmring den Schlauch schie-
ben. Bei der Ausrich­tung des Klemmrings darauf achten, dass die angeschrägte Seite in
Richtung Dichtring zeigt.
3. Den Dichtring cken.
4. Kapillare an der Durch­flusszelle verschrauben.
1 durch die
2
3 über
4
4aufste-
4 handfest
Abb. 10 Kapillarverschraubun-
gen
Abb. 11 Durchflusszelle mit
Nächste Schritte Nehmen Sie das Gerät in Betrieb.
Einsatz von PEEK-Verschraubungen
PEEK-Verschraubungen können bis zu einem Druck von 400 bar verwendet werden. Beachten Sie den Drehmoment von 0,5 Nm.

Leckmanagement anschließen

Das Leckmanagement setzt sich aus einem Lecksensor und einem Drainagesystem zusammen. Das Drainagesystem sorgt dafür, dass ausgetretene Flüssigkeiten automatisch in einen Abfallbehälter fließen. Wenn sich zu viel Flüssigkeit ansammelt, blinkt die rote LED am Gerät. Das Gerät und die Datenaufnahme über die Chromatograpfiesoftware werden gestoppt.
Kapillare
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
22 Installation
12
3
4
5
6
Voraussetzung Die Frontabdeckung wurde abgenommen.
Vorgehensweise
Ablauf Abbildung
1. Den Trichter mittlere Öffnung der
Kapillarführung cken.
2. Die Schlauchstutzen 4 an der langen Seite in den Drainageschlauch
1 in die
2 ste-
Abb. 12 Trichter mit Kapillarfüh-
rung
3 stecken.
Abb. 13 Drainageschlauch mit
Schlauchstutzen
3. Die Schlauchstutzen an den Trichter stecken.
4. Den Drainageschlauch mit dem Schlauchstut-
zen
5 an der Leck-
wanne verbinden.
Abb. 14 Drainageschlauch am
Gerät befestigen
5. Am untersten Gerät den Abfallstutzen
befestigen.
6. Den Abfallschlauch am Abfallstutzen befestigen und mit dem Abfallbe­hälter verbinden.
7. Den Abfallbehälter unterhalb der Geräte aufstellen.
Nächste Schritte Befestigen Sie die Frontabdeckung.
6
Abb. 15 Leckwanne mit Abfall-
stutzen
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
Legende
1423
5
1 Integratorausgänge
2 Anschluss LAN
3 Schnittstelle für den
Service
23 Installation

Steuerung

Der Detektors wird auf 2 Arten extern gesteuert:
Über die StiftleisteInnerhalb eines lokalen Netzwerks über den LAN-Anschluss
an den Router.
Sämtliche Anschlüsse für die externe Steuerung befinden sich auf der Rückseite des Detektors
Eine lokale Ansteuerung des DAD 6.1L über die Control Unit ist ebenfalls möglich. Der Anschluss für die Control Unit befindet sich an der Seite des Geräts.
.
4 Stiftleiste
DAD 6.1L
1 USB-Anschluss für
die Control Unit
Abb. 16 Anschlüsse Rückseite
Abb. 17 Anschluss Control Unit am DAD 6.1L
Gerät an den Computer anschließen
Dieses Kapitel beschreibt, wie ein HPLC-System zu einem lokalen Netzwerk (LAN) aufgebaut wird und wie das LAN durch einen Netzwerkadministrator zum Datenaustausch an ein Firmennetz­werk angeschlossen werden kann. Die Beschreibung gilt für das Betriebssystem Windows® und alle gängigen Router.
Hinweis Um ein LAN aufzubauen, wird die Verwendung eines Routers
empfohlen. Das heißt, dass folgende Schritte erforderlich sind:
Ablauf 1. Am Computer in der Systemsteuerung die LAN-Eigenschaften
prüfen.
2. Den Router mit den Geräten und dem PC verkabeln.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
3. Am Computer den Router für das Netzwerk einrichten.
24 Installation
4. Die Chromatografiesoftware vom entsprechenden Datenträ­ger installieren.
5. Die Geräte einschalten und Chromatografiesoftware starten.
LAN-Eigenschaften einstellen
Im LAN wird ausschließlich ein Server (im Regelfall der Router) verwendet, von dem die Geräte automatisch ihre IP-Adresse im Netzwerk beziehen.
Voraussetzung In Windows
Standby-Funktion und Bildschirmschoner ausgeschaltet.
Wenn eine “USB to COM”-Box verwendet wird, muss im
Gerätemanager die Einstellung “Computer kann das Gerät ausschalten, um Energie zu sparen” für alle USB-Hosts deakti­viert werden.
Nur Windows 7: Für den Netzwerkadapter im Gerätemana-
ger die Einstellung “Computer kann das Gerät ausschalten, um Energie zu sparen” deaktivieren.
®
sind Energiesparfunktionen, Ruhezustand,
Vorgehensweise 1. In Windows 7 Start
gabecenter auswählen.
2. Auf LAN-Verbindung doppelklicken.
3. Die Schaltfläche Eigenschaften anklicken.
4. Internetprotokoll Version 4 (TCP/IPv4) auswählen.
5. Die Schaltfläche Eigenschaften anklicken.
6. In der Registerkarte Allgemein die Einstellungen prüfen. Die korrekten Einstellungen des DHCP-Clients sind: a) IP-Adresse automatisch beziehen b) DNS-Serveradresse automatisch beziehen
7. Die Schaltfläche OK anklicken.
Geräte zum LAN verkabeln
Der Router hat mehrere LAN-Anschlüsse und einen WAN-/Inter­netanschluss, über den der Router an ein Wide Area Network (WAN) angeschlossen werden kann, wie z. B. ein Firmennetz­werk oder das Internet. Die LAN-Anschlüsse dagegen dienen zum Aufbau eines Netzwerks aus Geräten und Computer. Um Störungen zu vermeiden wird empfohlen, das HPLC-System außerhalb des Firmennetzwerks zu betreiben.
Für jedes Gerät und für den Router wird ein Patch-Kabel mitge­liefert. Um den Router an das Netzwerk anzuschließen, wird ein
Systemsteuerung  Netzwerk- und Frei-
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
25 Installation
12
3
4
5
zusätzliches Patch-Kabel benötigt, das nicht im Lieferumfang enthalten ist.
Legende
1 Geräte
2 Router
4 WAN-/Internetanschluss
5 Workstation
3 LAN-Anschlüsse
Abb. 18 Kabelplan LAN
Voraussetzung Der Computer wurde ausgeschaltet.
Für die Geräte und den Computer ist je ein Patch-Kabel vor-
handen.
Vorgehensweise 1. Mit dem Patch-Kabel den Router und den Computer verbin-
den. Diesen Schritt wiederholen, um die Geräte anzuschlie­ßen.
2. Mit dem Netzteil den Router an das Stromnetz anschließen.
Vorgehensweise 1. Um die Routerkonfiguration zu öffnen, im Browser die IP-
Router einstellen
Der Router wird mit werkseitigen Voreinstellungen ausgeliefert. Auf der Unterseite des Routers ist ein Aufkleber angebracht, auf dem die IP-Adresse, Benutzername und Passwort zu finden sind, mit denen man die Routerkonfiguration öffnen kann.
Adresse des Routers eingeben (gilt nicht für alle Router).
2. Den Benutzernamen und das Passwort eingeben.
3. Den Router als DHCP-Server einstellen.
4. In der Routerkonfiguration den IP-Adressbereich prüfen und ggf. ändern.
Ergebnis Sobald der Router allen Geräten eine IP-Adresse zugewiesen hat,
übernimmt die Chromatografiesoftware die Steuerung des HPLC-Systems.
LAN in das Firmennetzwerk integrieren
Der Router kann durch den Netzwerkadministrator an das Fir­mennetzwerk angeschlossen werden. Dazu wird der WAN-/ Internetanschluss des Routers verwendet.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
26 Installation
Voraussetzung Das Patch-Kabel ist vorhanden.
Vorgehensweise 1. Prüfen, dass es keine Überschneidung zwischen den IP-Adres-
sen des Routers und des Firmennetzwerks gibt.
2. Im Fall einer Überschneidung in der Routerkonfiguration den IP-Adressbereich ändern.
3. Mit dem Patch-Kabel den WAN-/Internetanschluss des Rou­ters mit dem Firmennetzwerk verbinden.
4. Alle Geräte einschließlich des Computers neu starten.
Mehrere Systeme in einem LAN separat steuern
Die Kommunikation in LANs läuft über sogenannte Ports, die Teil der Netzwerkadresse sind. Wenn in einem LAN mehrere HPLC-Systeme vernetzt sind, die separat gesteuert werden sol­len, können dafür unterschiedliche Ports verwendet werden, um Störungen zu vermeiden. Dafür muss die Portnummer an jedem Gerät geändert und die gleiche Portnummer in der Gerätekonfi­guration der Chromatografiesoftware eingegeben werden. Es empfiehlt sich, für alle Geräte eines Systems dieselbe Portnum­mer zu verwenden.
Hinweis Der Port ist bei allen Geräten werkseitig auf 10001 eingestellt.
Die Portnummern in der Konfiguration der Geräte in der Chro­matografiesoftware und am Gerät müssen identisch sein, ansonsten kann keine Verbindung hergestellt werden.
Vorgehensweise 1. Die Portnummer bestimmen und am Gerät ändern.
2. Die Portnummer in der Chromatografiesoftware eingeben.
Ergebnis Die Verbindung wird hergestellt.
Anschlüsse der Stiftleiste
Bei bestimmten Anlässen oder zu Testzwecken kann die manu­elle Eingabe von Signalen zweckmäßig sein:
Senden von Steuersignalen (Events) an externe GeräteÖffnen und Schließen von KontaktenAktivierung von 500 ms-Impulsen
Folgende Remote Signale können empfangen und gesendet werden:
Empfang von Start- und Error-Signalen externer Geräte
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
27 Installation
Senden von Steuer- und Error-Signalen an externe Geräte
Abb. 19 Beziehung zwischen Anzeige und Anschlussleiste
Erläuterungen zur Stiftleiste
Anschluss Funktion
Relaiskontakt
EV 1
(Event 1)
Der Kontakt ist potentialfrei. Die Ein­stellung ist abhängig von den Einste­lungen in der Control Unit, der Mobile Control oder der Software. Dauersignal:
passiv = Relaiskontakt geöffnetaktiv = Relaiskontakt geschlossen
Impuls:
Relaiskontakt geschlossen für
min. 1000 ms Zulässige Belastung des Relaiskontakts: 1 A/ 24 V DC
EV 2 (Event 2) TTL-Ausgang
Pegel:
passiv 5 Vaktiv 0 V
Impuls:
0 V für min. 1000 ms
Error OUT TTL-Ausgang
Pegel:
passiv 5 Vaktiv 0 V
Der Ausgang ist aktiv, bis die Error­Bedingung entfällt.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
28 Installation
Anschluss Funktion
Error IN TTL-Eingang
Start IN TTL-Eingang
Low-aktiv
Sichere Schaltschwelle min. 10 mA Bei einem Signal (Kurzschluss nach
GND) von einem externen Gerät, erscheint eine Fehlermeldung.Das Gerät stoppt.
Low-aktiv
Sichere Schaltschwelle min. 10 mA Bei einem Signal (Kurzschluss nach
GND) von einem externen Gerät star­tet das Gerät. Bei Softwarekontrolle wird ein elektronischer Trigger per LAN verschickt.
Autozero TTL-Eingang
Low-aktiv
Sichere Schaltschwelle min. 10 mA Ein Signal (Kurzschluss nach GND)
stellt das aktuelle Messsignal auf Null.
+5 V Stellt eine Spannung von 5 V gegen
GND zur Verfügung. Damit kann ein mit Event geschalteter Verbraucher versorgt werden. Absicherung: 5 V–50 mA
GND Bezugspunkt der Spannung an den
Signaleingängen.
+24V Valve Event-gesteuerte Spannung von 24 V
gegen GND. Absicherung: 24 V–200 mA
Valve OUT Ausgang ist aktiv, bis die Valve-Bedin-
gung entfällt.
GND Bezugspunkt der Spannung an den
Signaleingängen.
Stiftleiste verkabeln
Um ein Gerät durch ein anderes Gerät anzusteuern, wird ein Industriesteckverbinder verwendet. Über die einzelnen Anschlüsse werden Steuersignale übertragen.
Um Geräte fernzusteuern, müssen Sie die Kabel an die Stiftleiste anschließen. Die Stifte der Stiftleiste stecken Sie abschließend auf die Federkontakte der Federleiste, über die eine Verbindung zur Platine hergestellt wird.
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
29 Installation
1
2
3
Voraussetzung Das Gerät wurde ausgeschaltet.
Der Netzstecker wurde gezogen.
Werkzeug Hebeldrücker
VORSICHT! Beschädigung des Geräts durch Kurzschluss mög­lich! Gerät vor dem Anschließen an die Stiftleiste aus­schalten. Netzstecker ziehen.
VORSICHT! Zerstörung der Elektronik durch elektrostatische Entladung!
Vorgehensweise Ablauf Abbildung
Geerdetes Armband tragen.
1. Die Stiftleiste Unterlage legen.
2. Den Hebeldrücker die Öffnung auf der Ober­seite stecken und nach unten drücken.
3. Den Hebeldrücker gedrückt halten und die
Kabel
2 in die Vorder-
seite einführen. Danach den Hebeldrücker heraus­ziehen.
4. Prüfen, ob die Kabel fest verbunden sind.
5. Die Stiftleiste auf die Federleiste drücken.
Nächste Schritte Beenden Sie die Installation. Nehmen Sie danach das Gerät in
Betrieb.
3 auf eine
1 in
Abb. 20 Stiftleiste
Integrator-Anschluss
Der Integrator-Anschluss gibt Messsignale des Detektors aus.
nicht bipolar4 Kanal0 bis 5 V DAC 16 bitskalierbarOffset einstellbar
KNAUER DAD 6.1L, DAD 2.1L, MWD 2.1L Benutzerhandbuch V6700, Version 1.0
Loading...