Klein Tools NCVT1XTKIT User manual

I NSTRUCTIONS – Non-Contact Voltage and GFCI Receptacle Electrical Test Kit
NCVT-1XTKIT
4
ENGLISH
Power On / Off Button2.Power On / Low Battery Indicator
1.
Voltage Detection LEDs (Inside Tip)4.Non-Contact Tip
3
3.
Pocket Clip
5.
2x AAA Batteries (Included)
8.
Keep hand and fingers below this line while using tester
9.
O-ring Seal
6.
Battery Cap
7.
NOTE: No user-serviceable parts inside.
9 2
1
5
6
• Non-contact detection of 70 to 1000V AC in cables, cords, circuit breakers, lighting  xtures, switches, outlets, and wires
• When voltage is detected, a high intensity, bright red LED illuminates and a beeping sound is generated
Warning – Risk of electric shock / Advertencia: riesgo de choque eléctrico / Avertissement – Risque d'électrocution
Risk of danger. Important information: It is important that users of this tester read, understand, and follow all warnings, cautions, safety information, and instructions in this manual before operating or servicing this tester. Failure to follow instructions could result in death or serious injury.
Double insulated / Doble aislamiento / Double isolation Read instructions / Lea las instrucciones / Lire les instructions
Conformité Européenne:
Conforms with European Economic Area directives. Cumple con las normas del Área Económica Europea. Conforme aux directives de l’Espace économique Européen.
UKCA:
UK Conformity Assessment Conformidad evaluada porelReino Unido Évaluation de la conformité du Royaume-Uni
This product has been independently tested by Intertek and meets applicable published standards. Este producto ha sido probado de manera independiente por Intertek y cumple con las normas publicadas vigentes. Ce produit a été testé de manière indépendante par Intertek et répond aux exigences des normes applicables.
For measurements performed at the source of low-voltage installation and outside lines.
CAT
Para mediciones realizadas en la fuente de la instalación de bajo voltaje y líneas externas.
IV
Pour des mesures prises à la source d’une installation à faible tension et des lignes extérieures.
5001748
DURABILITY / DURABILIDAD / DURABILITÉ
Ingress Protection
Protección contra el ingreso Protection contre les infiltrations
Drop Protection
IP67
Protección ante caídas Protection contre les chutes
Safety Rating
9.8 ft. Clasificación de seguridad
(3 m)
Cote de sécurité
CAT IV 1000V
RT210
FIG. 1
Indicator
Indicador
Voyant
Open Ground
Conexión a tierra abierta
8
Mise à la terre non connectée
Open Neutral
Neutro abierto Neutre ouvert
Open Hot
Vivo abierto Phase ouverte
Hot/Ground Reversed
Vivo/Tierra invertidos Phase/mise à la terre inversées
Hot/Neutral Reversed
Vivo/Neutro invertidos Phase/neutre inversés
Correct
Correcto Correct
Indicator Illuminated
Indicador iluminado Voyant allumé
Fault
Falla
Anomalie
Ground contact is not connected
El contacto a tierra no está conectado Le contact de mise à la terre n’est pas connecté
Neutral contact is not connected
El contacto neutro no está conectado Le contact neutre n’est pas connecté
Hot contact is not connected
El contacto vivo no está conectado Le contact de phase n’est pas connecté
Hot and ground connections are reversed
Las conexiones viva y de tierra están invertidas Les connexions de phase et de mise à la terre sont inversées
Hot and neutral connections are reversed
Las conexiones viva y neutra están invertidas Les connexions de phase et de neutre sont inversées
Receptacle is wired correctly
El receptáculo está cableado correctamente La prise est câblée correctement
Indicator Not Illuminated
Indicador no iluminado Voyant éteint
Explanation
Explicación Explication
• Designed to detect the most common wiring problems in standard and GFCI receptacles: Open ground, reverse polarity, open hot, open neutral,
Neutral (White)
Neutro (blanco)
Neutre (blanc)
Ground (Green) Tierra (verde) / Mise à la terre (vert)
Hot (Black)
Vivo (negro) Phase (noir)
CAT II
5001748
BASED ON: 1330999 Rev 04/22 B
300V
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
ENGLISH
Risk of electric shock and burn. Contact with live circuits could result in death or serious
injury.
Use caution with voltages above 25V AC as a shock hazard may exist.
injury.
injury.
A blinking LED or a steadily illuminated LED in the tip and audible beeps indicate presence
Use caution with voltages above 25V AC as a shock hazard may exist.
Use caution with voltages above 25V AC as a shock hazard may exist.
of voltage. If no indication, voltage could still be present.
A blinking LED or a steadily illuminated LED in the tip and audible beeps indicate presence
A blinking LED or a steadily illuminated LED in the tip and audible beeps indicate presence
Before and after each use, verify operation by testing a known working circuit that is within
of voltage. If no indication, voltage could still be present.
of voltage. If no indication, voltage could still be present.
the rating of this unit.
Before and after each use, verify operation by testing a known working circuit that is within
Before and after each use, verify operation by testing a known working circuit that is within
Never assume neutral or ground wires are de-energized. Neutrals in multi-wire branch
the rating of this unit.
the rating of this unit.
circuits may be energized when disconnected and must be retested before handling.
Never assume neutral or ground wires are de-energized. Neutrals in multi-wire branch
Never assume neutral or ground wires are de-energized. Neutrals in multi-wire branch
The tester
circuits may be energized when disconnected and must be retested before handling.
circuits may be energized when disconnected and must be retested before handling.
WILL NOT
detect voltage if:
The wire is shielded.
detect voltage if:
detect voltage if:
The operator is not grounded or is otherwise isolated from an effective earth ground.
The voltage is DC.
The operator is not grounded or is otherwise isolated from an effective earth ground.
The operator is not grounded or is otherwise isolated from an effective earth ground.
The tester
The voltage is DC.
The voltage is DC.
MAY NOT
detect voltage if:
The user is not holding the tester.
detect voltage if:
detect voltage if:
The user is insulated from the tester with a glove or other materials.
The user is not holding the tester.
The user is not holding the tester.
The wire is partially buried or in a grounded metal conduit.
The user is insulated from the tester with a glove or other materials.
The user is insulated from the tester with a glove or other materials.
The tester is at a distance from the voltage source.
The wire is partially buried or in a grounded metal conduit.
The wire is partially buried or in a grounded metal conduit.
The field created by the voltage source is being blocked, dampened, or otherwise
The tester is at a distance from the voltage source.
The tester is at a distance from the voltage source.
interfered with.
The frequency of the voltage is not a perfect sine wave between 50 and 500Hz.
The tester is outside of operating conditions (listed in Specifications section).
The frequency of the voltage is not a perfect sine wave between 50 and 500Hz.
The frequency of the voltage is not a perfect sine wave between 50 and 500Hz.
Operation may be affected by differences in socket design and insulation thickness and
The tester is outside of operating conditions (listed in Specifications section).
The tester is outside of operating conditions (listed in Specifications section).
type; tester may not be compatible with some types of standard or tamper resistant (TR)
Operation may be affected by differences in socket design and insulation thickness and
Operation may be affected by differences in socket design and insulation thickness and
electrical outlets.
type; tester may not be compatible with some types of standard or tamper resistant (TR)
type; tester may not be compatible with some types of standard or tamper resistant (TR)
In bright light conditions, the LED visual indicators will be less visible.
When powered-ON, one of the “power-ON” LED’s will be illuminated, and a green light
In bright light conditions, the LED visual indicators will be less visible.
In bright light conditions, the LED visual indicators will be less visible.
will illuminate the tip.
When powered-ON, one of the “power-ON” LED’s will be illuminated, and a green light
When powered-ON, one of the “power-ON” LED’s will be illuminated, and a green light
DO NOT USE TESTER UNLESS ONE OF THE “POWER-ON” LED
s
IS
ILLUMINATED.
Do not use if tester appears damaged or is not operating properly. If in doubt, replace the
tester.
Do not use if tester appears damaged or is not operating properly. If in doubt, replace the
Do not use if tester appears damaged or is not operating properly. If in doubt, replace the
Do not apply more than the rated voltage as marked on the tester (1000V).
Do not apply to uninsulated hazardous live conductors.
Do not apply more than the rated voltage as marked on the tester (1000V).
Do not apply more than the rated voltage as marked on the tester (1000V).
Detection above 70V AC is specified under "normal" conditions as detailed below. The tester
may detect at a different threshold at different conditions, or may not detect at all unless:
Detection above 70V AC is specified under "normal" conditions as detailed below. The tester
Detection above 70V AC is specified under "normal" conditions as detailed below. The tester
The tip of the tester is within 0.25" (6 mm) of an AC voltage source radiating unimpeded.
The user is holding the body of the tester with his or her bare hand.
The tip of the tester is within 0.25" (6 mm) of an AC voltage source radiating unimpeded.
The tip of the tester is within 0.25" (6 mm) of an AC voltage source radiating unimpeded.
The user is standing on or connected to earth ground.
The user is holding the body of the tester with his or her bare hand.
The user is holding the body of the tester with his or her bare hand.
The air humidity is nominal (50% relative humidity – non-condensing).
The user is standing on or connected to earth ground.
The user is standing on or connected to earth ground.
The tester is held still.
The air humidity is nominal (50% relative humidity – non-condensing).
The air humidity is nominal (50% relative humidity – non-condensing).
Always wear approved eye protection.
Comply with local and national safety requirements.
Always wear approved eye protection.
Always wear approved eye protection.
If this product is used in a manner not specified by the manufacturer, protection provided
Comply with local and national safety requirements.
Comply with local and national safety requirements.
by the product may be affected.
If this product is used in a manner not specified by the manufacturer, protection provided
If this product is used in a manner not specified by the manufacturer, protection provided
DO NOT
attempt to repair this tester. It contains no serviceable parts.
DO NOT
attempt to repair this tester. It contains no serviceable parts.
attempt to repair this tester. It contains no serviceable parts.
expose tester to extremes in temperature or high humidity.
attempt to repair this tester. It contains no serviceable parts.
attempt to repair this tester. It contains no serviceable parts.
attempt to repair this tester. It contains no serviceable parts.
attempt to repair this tester. It contains no serviceable parts.
attempt to repair this tester. It contains no serviceable parts.
attempt to repair this tester. It contains no serviceable parts.
NCVT-1XT RT210
GENERAL SPECIFICATIONS
The Klein Tools NCVT-1XT is a durable 70 to 1000V AC non-contact voltage tester with IP67 rating and a 3m (9.8') drop protection.
• Environment: Indoor
• Measurement Range: 70 to 1000V AC
• Frequency Range: 50 to 500 Hz
• Batteries: 2x AAA 1.5V Alkaline
• Operating and Storage Altitude: Up to 6562 ft. (2000 m)
• Operating and Storage Temp: 14° to 122°F (-10° to 50°C)
• Relative Humidity: <95% non-condensing
• Dimensions: 5.34" x 1.08" x 0.97" (135.6 x 27.4 x 22.1 mm)
• Weight: 1.65 oz (46.7 g) with batteries
• Pollution degree: 2
• Safety Rating: CAT IV 1000V AC
• Drop Protection: 9.8 ft. (3 m)
• Ingress Protection: IP67
• Standards: EN 61010-1:2010+A1: 2019,
EN IEC 61010-2-030:2021+A11:2021.
Conforms to UL STD. 61010-1, 61010-2-030 Certi ed to CSA STD. C22.2 No. 61010-1, 61010-2-030.
5001748
Specifications subject to change.
WARNINGS
To ensure safe operation and service of the tester, follow these instructions. Failure to observe these warnings can result in severe injury or death.
FUNCTION BUTTONS
NCV POWER ON/OFF BUTTON
1
Press and release the NCV Power button 1. The tester will emit a
single beep, the Power On Indicator 2 and green LED 3 in the tip will illuminate. To power-OFF the tester, press and release the Power On / Off button 1. The tester will emit a double beep.
NOTE: The tester will automatically power-OFF following 4 minutes of inactivity to conserve battery life.
OPERATING INSTRUCTIONS
CHECKING FOR THE PRESENCE OF AC VOLTAGE
1. Press the Power On / Off button. After performing a self-test, a green LED3 will illuminate the tip when no voltage is detected.
2. Prior to use, test on a known live circuit to verify tester functionality.
3. Place tip of the tester
near AC voltage. If voltage is present,
4
the unit will emit audible beeps and the red LED3 in the tip will illuminate.
SILENT OPERATION
To activate silent operation, power-ON by pressing and holding the NCV Power-ON button 1 for more than 2 seconds. Power­ON in this manner each time silent operation is desired.
MAINTENANCE
BATTERY REPLACEMENT
When the battery is low, the unit will power-on with a blinking red LED2. When this happens, the batteries must be replaced:
1. Unscrew the battery cap7and remove/recycle spent
batteries.
2. Install two new AAA batteries. Note proper polarity.
3. Screw battery cap tightly to ensure a tight seal with the O-Ring6.
CLEANING
Be sure tester is turned off and wipe with a clean, dry lint­free cloth.
Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when not in use for a prolonged period of time. Do not expose to high temperatures or humidity. After a period of storage in extreme conditions exceeding the limits mentioned in the General Specifications section, allow the tester to return to normal operating conditions before using.
FCC AND IC COMPLIANCE
S ee this product’s page at for FCC compliance information.
Canada ICES-003 (B) / NMB-003 (B)
CAUTION
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
FUNCTION BUTTONS
NCV POWER ON/OFF BUTTON
1
Press and release the NCV Power button 1. The tester will emit a single beep, the Power On Indicator 2 and green LED 3 in the tip will illuminate. To power-OFF the tester, press and release the Power On / Off button 1. The tester will emit a double beep.
NOTE: The tester will automatically power-OFF following 4
• Environment: Indoor
• Operating and Storage Altitude: Up to 6562 ft. (2000 m)
• Operating Temperature: 32° to 140°F (0° to 40°C)
• Storage Temperature: 14° to 122°F (-10° to 50°C)
• Relative Humidity: <85% non-condensing
• Mains supply voltage  uctuations: ±10%
• Nominal Voltage: 110 - 125V AC at 50/60Hz in 3-wire outlet
• Standards: Conforms to UL Std 61010-1, 61010-2-030,1436.
minutes of inactivity to conserve battery life.
OPERATING INSTRUCTIONS
CHECKING FOR THE PRESENCE OF AC VOLTAGE
1. Press the Power On / Off button. After performing a self-test, a green LED3 will illuminate the tip when no voltage is detected.
2. Prior to use, test on a known live circuit to verify tester functionality.
3. Place tip of the tester
near AC voltage. If voltage is present,
4
the unit will emit audible beeps and the red LED3 in the tip will illuminate.
SILENT OPERATION
To activate silent operation, power-ON by pressing and holding the NCV Power-ON button 1 for more than 2 seconds. Power­ON in this manner each time silent operation is desired.
MAINTENANCE
BATTERY REPLACEMENT
When the battery is low, the unit will power-on with a blinking red LED2. When this happens, the batteries must be replaced:
1. Unscrew the battery cap7and remove/recycle spent
batteries.
2. Install two new AAA batteries. Note proper polarity.
3. Screw battery cap tightly to ensure a tight seal with the O-Ring6.
CLEANING
Be sure tester is turned off and wipe with a clean, dry lint­free cloth.
Do not use abrasive cleaners or solvents.
Conditions indicated: wiring correct, open ground, reverse polarity, open hot, open neutral and hot/
ground reversed. Conditions NOT indicated: quality of ground, multiple hot wires, combinations of defects, reversal of
grounded and grounding conductors. All appliances or equipment on the circuit being tested should be unplugged to help reduce the possibility of erroneous readings.
STANDARD RECEPTACLES
1. Verify tester operation by testing on a known live and correctly wired receptacle.
2. Plug tester into receptacle.
3. Compare the illuminated lights on the tester to the key code printed on the tester.
4. If the tester indicates that the receptacle is not wired correctly, consult a qualified electrician.
GFCI RECEPTACLES
1. Check the GFCI receptacle user manual for information on how the specific receptacle operates prior to
2. Insert the tester into the receptacle under test to check for correct wiring (See FIG. 1). Lights on the
3. Press the "TEST" button on the GFCI receptacle.
STORAGE
Remove the batteries when not in use for a prolonged period of time. Do not expose to high temperatures or humidity. After
4. Press and hold the test button on the tester for 7 seconds.
a period of storage in extreme conditions exceeding the limits mentioned in the General Specifications section, allow the tester to return to normal operating conditions before using.
FCC AND IC COMPLIANCE
Se e this product’s page at for FCC compliance information.
Canada ICES-003 (B) / NMB-003 (B)
Wipe with a clean, dry lint-free cloth.
Do not place equipment and its accessories in the trash. Items must be properly disposed of in accordance with local regulations. Please see for additional information.
GENERAL SPECIFICATIONS
• Weight: 1.2 oz (33 g)
• Pollution degree: 2
• Safety Rating: CAT II 300V
• Drop Protection: 6.6 ft. (2 m)
Certi ed to
5001748
CSA Std C22.2 #61010-1, 61010-2-030,160
.
Specifications subject to change.
WARNINGS
WIRING CONFIGURATION TESTING
using this tester.
tester should illuminate.
Did the GFCI trip and the lights on the tester go dark?
YES:
Reset the GFCI by pressing the reset button. Proceed to step 4.
NO:
The GFCI is not operating properly or the receptacle is miswired. Consult a qualified electrician.
Did the GFCI trip and the lights on the tester go dark?
YES:
Reset the GFCI by pressing the reset button. The GFCI appears to be operating properly.
NO:
The GFCI is not operating properly or the receptacle is miswired. Consult a qualified electrician.
CLEANING
Do not use abrasive cleaners or solvents.
DISPOSAL / RECYCLE