Klein Tools MM450 is an auto-ranging, true root mean square (TRMS)
multimeter that measures AC/DC voltage, AC/DC current, and resistance. It
can also measure temperature, capacitance, frequency, duty-cycle, and test
diodes and continuity. It features lead alert protection and a high visibility,
reverse contrast LCD display with an optical sensor to automatically
optimize viewability in dark or bright ambient lighting.
• Environment:
• Operating Altitude: 6562 ft. (2000m)
• Relative Humidity: <90% non-condensing
• Operating Temp: 14°F to 140°F (-10°C to 60°C)
• Storage Temp: 14°F to 140°F (-10°C to 60°C)
• Accuracy: Values stated at 65°F to 83°F (18°C to 28°C)
• Temp Coef cient: 0.1 × (Quoted Accuracy) per °C above 28°C
or below 18°C, corrections are required when ambient working
temperature is outside of Accuracy Temp range
• Accuracy: ± (% of reading + # of least significant digits)
• Drop Protection: 6.6 ft. (2m)
• Safety Rating: CAT III 600V, Class 2, Double insulation
Indoor. DO NOT expose to moisture, rain, or snow.
CAT III: Measurement category III is applicable to test and
measuring circuits connected to the distribution part of the
building’s low-voltage MAINS installation.
• Electromagnetic Environment: IEC EN 61326-1. This
equipment meets requirements for use in basic and controlled
electromagnetic environments like residential properties,
business premises, and light-industrial locations.
Overload Protection: 500mA / 600V and 10A / 600V Fuses
Frequency Range: 50 to 400Hz
Maximum Input: µA/mA setting: 400mA AC RMS / DC10A setting: 10A AC RMS / DC
• Diode Test: 1.8 mA max, open circuit voltage 3.9V DC
• Continuity Check: Audible signal <50Ω
• Sampling Frequency: 3 samples per second
• Auto-Power Off: After ~5 minutes of inactivity
• Overload: "OL" indicated on display, overload protection
600V RMS in all settings
• Polarity: "-" on display indicates negative polarity
• Display: 3 ¾ digit, 4000 Count LCD
WARNINGS
To ensure safe operation and service of the meter, follow these
instructions. Failure to observe these warnings can result in
severe injury or death.
• Before each use verify meter operation by measuring a known
voltage or current.
• Never use the meter on a circuit with voltages that exceed the
category based rating of this meter.
• Do not use the meter during electrical storms or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Probe assemblies to be used for MAINS measurements shall
meet IEC/EN 61010-031 with a voltage RATING of CAT III 600V
or better.
• Ensure meter leads are fully seated, and keep fingers away
from the metal probe contacts when making measurements.
• Do not open the meter to replace batteries while the probes
are connected.
• Use caution when working with voltages above 25V AC RMS
or 60V DC. Such voltages pose a shock hazard.
• To avoid false readings that can lead to electrical shock,
replace batteries when a low battery indicator appears.
• Do not attempt to measure resistance or continuity on a live
circuit.
• Always adhere to local and national safety codes. Use personal
protective equipment to prevent shock and arc blast injury
where hazardous live conductors are exposed.
To Power ON the meter rotate the Function Selector switch
OFF setting to any measurement setting. To Power OFF the meter rotate the
Function Selector switch
2
to the OFF setting. By default, the meter will
2
from the
automatically Power OFF after 5 minutes of inactivity. Reactivate meter by
pressing any button. To deactivate the automatic Power OFF feature, power
the meter ON with the SEL button
OFF is deactivated, the
symbol will not be visible in the display.
depressed. When automatic Power
10
"SEL" (SELECT) BUTTON (FOR SECONDARY FUNCTIONS)
The "SEL" button
accessible by the function selector switch
activates the secondary function for each application
10
2
. For current and voltage, it
toggles between AC and DC. For the other functions, it switches between °F
and °C, between Hz and % Duty-cycle, between Continuity and Resistance,
and between Capacitance and Diode-test. The default function for each
application is printed on the meter in white; the secondary function for each
location is printed on the meter in orange.
"HOLD" (DATA HOLD) BUTTON
Press the "HOLD" button
again to release the display and return to live measuring.
to hold the measurement on the display. Press
9
"RANGE" BUTTON
The meter defaults to auto-ranging measurement mode
. This
automatically determines the most appropriate measurement range for the
testing that is being conducted. To manually force the meter to measure in a
different range, use the "RANGE" button
1. Press the "RANGE" button
is deactivated on the LCD). Repeatedly press the "RANGE" button
(
to cycle through the available ranges, stopping once the desired range
7
is reached.
7
2. To return to auto-ranging mode, press and hold the "RANGE" button
for more than one second ( is reactivated).
.
7
to manually select measurement range
7
"MAX/MIN" BUTTON
When the "MAX/MIN" button
minimum and maximum value of the measurement as the meter continues
to take samples. The first press of the "MAX/MIN" button
Max value, the second press displays the Min value.
is pressed, the meter keeps track of the
8
displays the
8
To return to normal measuring mode, press and hold the "MAX/MIN"
for more than one second.
button
8
BACKLIGHT BUTTON
The meter defaults to Automatic Brightness mode, which adjusts the screen's
brightness based on the ambient lighting conditions detected by the Optical
Press the Backlight Button6 to enter manual high brightness mode.
Sensor
12
Press Backlight Button
Backlight Button
again to enter manual low brightness mode. Press the
6
once more to return to Automatic Brightness mode.
6
LEAD ALERT LEDs
When the Function Selector switch
the Lead Alert LEDs
inserted into the appropriate jacks. The lights will automatically turn off after
will illuminate to ensure that the test leads are
11
2 minutes. To disable the Lead Alert LEDs
and turn the Function Selector switch2 to any setting other than OFF.
is rotated to a measurement setting,
2
, hold the Backlight button
11
6
NOTE: Lead Alert function will reactivate by default when the meter powers OFF
Do not test if leads are improperly seated. Results could cause
intermittent display readings. To ensure proper connection, firmly
press leads into the input jack completely.
CORRECT
INCORRECT
TESTING IN CAT III / CAT IV MEASUREMENT LOCATIONS
Ensure the test lead shield is pressed firmly in place. Failure to use
the CAT III / CAT IV shield increases arc-flash risk.
5/32"
(4 mm)
TESTING IN CAT II MEASUREMENT LOCATIONS
CAT III / CAT IV shields may be removed for CAT II locations. This
will allow testing on recessed conductors such as standard wall
outlets. Take care not to lose the shields.
NOTE: The meter defaults to AC measurement. Press the "SEL"
button
to toggle between AC and DC modes. "AC "or "DC" on the
10
LCD indicates which mode is selected.
Black lead
Red lead
2. Apply test leads to the circuit to be tested to measure voltage.
The meter will auto-range to display the measurement in the
most appropriate range.
NOTE: If "–" appears on the LCD, the test leads are being applied
to the circuit in reverse polarity. Swap the position of the leads to
correct this.
NOTE: When in a voltage setting and the test leads are open,
readings of order mV may appear on the display. This is noise and
is normal. By touching the test leads together to close the circuit
the meter will measure zero volts.
NOTE: The Hazardous Voltage indicator appears for voltages >30V.
2. To measure current: Remove power from circuit, open circuit at
measurement point, connect meter in-series in the circuit using
the test leads, and apply power to circuit.
3. Measure the current. The meter will auto-range to display the
measurement in the most appropriate range.
Do not attempt to measure more than 10A.
When measuring currents greater than 6A, a measurement
time of 30 seconds followed by 10 minutes of recovery time
is recommended.
RESISTANCE MEASUREMENTS
1. Insert RED test lead into VΩ jack
into COM jack
Continuity/Resistance
,
and rotate function selector switch 2 to the
4
setting.
, and BLACK test lead
5
NOTE: The meter defaults to Continuity testing in this mode. To
enter Resistance testing mode, press the "SEL" button
The continuity icon
will disappear from the display and the meter
10
once.
with enter auto-range
2. Remove power from circuit.
3. Measure resistance by connecting test leads to circuit. The
meter will auto-range to display the measurement in the most
appropriate range.
Black lead
Red lead
NOTE: When in a Resistance setting and the test leads are open
(not connected across a resistor), or when a failed resistor is under
test, the display will indicate O.L. This is normal.
DO NOT attempt to measure resistance on a live circuit.
NOTE: The meter defaults to Continuity testing in this mode. Ensure
that the Continuity Testing icon
press the "SEL" button
2. Remove power from circuit.
3. Test for continuity by connecting conductor or circuit with test
leads. If resistance is measured less than 50Ω, an audible signal
will sound and display will show a resistance value indicating
continuity. If circuit is open, display will show "OL".
10
Black lead
Red lead
is visible on the display. If not,
once.
DO NOT attempt to measure continuity on a live circuit.
1. Insert RED test lead into VΩ jack 5 and BLACK test lead into
COM jack 4, and rotate function selector switch 2 to the
Frequency/Duty-Cycle setting.
NOTE: The meter defaults to Frequency testing in this mode. To enter
Duty-Cycle testing mode, press the "SEL" button 10 once. Ensure
that the appropriate icon (either Hz or %) appears on the display.
2. Measure by connecting test leads across the circuit.
Black lead
Red lead
TEMPERATURE
1. Insert K-type thermocouple into the VΩ 5 and COM 4 jacks
(observe the polarity markings on the thermocouple and meter), and
rotate function selector switch 2 to the Temperature setting.
NOTE: The meter defaults to Fahrenheit scale in this mode. To enter
Celsius scale, press the "SEL" button 10 once. Ensure that the
appropriate icon (either °F or °C) appears on the display.
2. To measure temperature, make contact between the thermocouple
tip and the object being measured. When thermocouple tip and
object are in thermal equilibrium, the measurement on the display
will stabilize. The meter will auto-range to display the measurement
in the most appropriate range.
K-Type
Thermocouple
Remove thermocouple before switching meter to other
measurement functions.
The thermocouple included with the original purchase is
suitable for temperatures below 356°F / 180°C only. To measure
higher temperatures, a K-type thermocouple with
the appropriate measurement range should be used.
To attach 69417
Magnetic Hanger
(sold separately):
1. Open strap and
remove clip end
(not needed).
2. Loop strap through
Attachment Slot
on back of meter.
3. Close strap.
Be sure meter is turned off and wipe with a clean, dry lint-free
Do not use abrasive cleaners or solvents.
cloth.
Remove the batteries when meter is not in use for a prolonged
period of time. Do not expose to high temperatures or
humidity. After a period of storage in extreme conditions
exceeding the limits mentioned in the General Specifications
section, allow the meter to return to normal operating
conditions before using.
13
REMOVE
CLEANING
STORAGE
FCC & IC COMPLIANCE
for FCC compliance information.
Canada ICES-003 (B) / NMB-003 (B)
WARRANTY
DISPOSAL / RECYCLE
Do not place equipment and its accessories in the trash.
Items must be properly disposed of in accordance with local
regulations. Please see www.epa.gov/recycle for additional
information.