customerservice@kleintools.com
Emissivity is the relative ability of a surface to emit energy by radition. Each
type of surface (metal, brick, wood, etc.) has a different emissivity level
which must be accounted for when taking a measurement with an infrared
The IR2000A has a variable emissivity from 0.10 to 1.00 which allows
accurate measurement of most types of materials. Shiny bright surfaces
(i.e. chrome, new copper, white boards) have a much lower emissivity
than flat black materials. The emissivity of the IR2000A should be set
manually according to the chart below in order to obtain the most accurate
measurements. The chart is for guidance only, as the emissivity of objects
varies depending on surface finish, measurement wave length, field of
view, temperature, and the shape of the object.
Commercial sheet aluminum
Find a comprehensive list of emissivity values at
www.kleintools.com/content/instructions
the battery compartment by pulling the compartment lid at the
indentations, away from the trigger (see Thermometer Attributes section).
and recycle exhausted batteries.
2 x AAA alkaline batteries observing the polarity indicated
compartment lid and make sure it locks into place.
± 4°F (± 2°C) or 2% of reading
± (4°F + |Temp Measured in °F| x 0.1) or
± (2°C + |Temp Measured in °C| x 0.05)
e.g., If -20°F is measured,
accuracy ≈ ± (4 + 20 x 0.1)°F ≈ ± 6°F
± 1.8°F (± 1°C) or 1% of reading
ENGLISH
The distance-to-spot ratio is a representation of how focused the IR sensor
is on the measurement area. The closer you are to the surface you are
measuring, the smaller the area of measurement is.
The IR2000A has a 12:1 distance-to-spot ratio
(see illustration below).
www.kleintools.com/warranty
Clean the instrument by using a damp cloth. Do not use abrasive cleaners
Take care to keep the sensor lens clean at all times. Clean the lens using a soft
cloth or cotton swab with water or rubbing alcohol only, and allow the lens to
Remove the batteries when instrument is not in use for a prolonged period of
time. Do not expose to high temperatures or humidity. After a period of storage
in extreme conditions exceeding the limits mentioned in the Specifications
section, allow the instrument to return to normal operating conditions before
Do not place equipment and its accessories in the trash. Items
must be properly disposed of in accordance with local regulations.
IR2000A
Material Emissivity
Temperature - IR
Ambient
Temperature
Target
Temperature Accuracy
.5" (13 mm) 1" (25 mm) 2" (51 mm)
6" (152 mm)
12" (305 mm)
24" (610 mm)
48" (1219 mm)
4" (102 mm)
Temperature - Thermocouple
Ambient
Temperature Range Accuracy
The Klein Tools IR2000A is a professional series infrared thermometer. It
features a wide measurement range, a tight distance-to-spot ratio, a K-type
thermocouple input, dual lasers, and several calculation modes to assist
you in making the best possible measurements.
4.7" x 1.87" x 6.76" (119 x 47 x 172 mm)
2 x AAA alkaline batteries
21 CFR, Chapter 1, Subchapter J
To ensure safe operation and service of the tester, follow these instructions.
Avoid using the instrument around strong electromagnetic fields.
This instrument is designed for professionals who know the hazards
associated with their trade. While this instrument offers no foreseeable
as the environment you're working in, can be hazardous. These are a few
common safety practices for those working around temperature critical
to touch or enter, verify your readings are reasonably accurate
To take measurements with your IR2000A, aim the instrument at the object
to be measured and pull the trigger. The unit has an auto off after releasing
the trigger of 60 seconds.
The IR2000A features a dual laser to assist in targeting the area to be
to Spot Ratio for more information). The approximation will be mostly
accurate at a minimum distance of 12" (305 mm) from the IR2000A.
to alternate between Fahrenheit and Celsius. You can select either scale
The IR2000A will hold the last temperature measured for 60 seconds after
the trigger is released. To recall this value or associated calculations press
the MODE key to activate the display and show the held temperature.
auto-scan mode, the unit will automatically power off after 60 minutes.
Toggling Laser & Backlight
To toggle the laser, hold down the trigger and press the DOWN Key. To
toggle the backlight, hold down the trigger and press the UP Key.
The IR2000A can measure the temperature from a K-Type thermocouple
temperature will automatically be displayed. To see the minimum or
The last measurement taken is being shown.
The thermometer is currently taking measurements.
Auto-scan mode is active.
The IR2000A has several modes of operation. Press the MODE button
to cycle through modes. The following chart shows the mode name, the
screen identifier, and the function of the auxiliary buttons in the mode.
Allows adjustment of the emissivity. See Emissivity section..
taken while the trigger is held down.
the measured temperature is less than the alarm threshold.
The temperature being measured has exceeded the allowable
The temperature being measured is greater than the high
alarm (HAL) setting or lower than the low alarm (LAL) setting.
a minimum of 30 minutes between large ambient temperature
The ambient temperature has exceeded the allowable range.
Any other error requires the thermometer to be reset. Turn off
the thermometer, remove the batteries, wait one minute, then
ENGLISH
Laser Warning or Caution
Operating Altitude:
Relative Humidity:
Operating Temperature:
Storage Temperature:
Measurement Range:
Emissivity:
Optical Resolution (Distance: Spot):
Dimensions:
Weight:
Battery Type:
Battery Life:
Display Resolution:
Calibration:
Laser:
Max Output:
Wavelength:
Compliance:
Drop Protection:
Certi cations:
Follow safe disposal
procedures
Battery
Dangerous levels
Mode Screen DOWN Key UP Key
tomar una medición con un termómetro infrarrojo.
superficies brillantes y lustrosas (como el cromo, cobre nuevo y placas
guía, dado que la emisividad de los objetos varía en función del acabado
temperatura y la forma del objeto.
www.kleintools.com/content/instructions
el compartimento de batería tirando de la tapa por las indentaciones
y recicle las baterías agotadas.
2 baterías alcalinas AAA observando la polaridad indicada en
± 4°F (± 2°C) o 2% de la lectura
+ |Temperatura medida en °F| x 0,1) o
+ |Temperatura medida en °C| x 0,05)
Ejemplo: si se mide un valor de -20
la precisión es ≈ ± (4 + 20 x 0,1) °F ≈ ± 6
± 1,8°F (± 1°C) o 1% de la lectura
ESPAÑOL
enfocado el sensor infrarrojo en el área de medición. Cuanto más próximo
se halla uno sobre la superficie que se va a medir, menor es el área de
www.kleintools.com/warranty
se deben desechar correctamente de acuerdo con las regulaciones
IR2000A
Material Emisividad
Temperatura - IR
Temperatura
ambiente
Temperatura
objetivo Precisión
.5" (13 mm) 1" (25 mm) 2" (51 mm)
6" (152 mm)
12" (305 mm)
24" (610 mm)
48" (1219 mm)
4" (102 mm)
Temperatura - Termopar
Temperatura
ambiente Rango Precisión
ajustada, entrada para termopar tipo K, doble láser y varios modos de
cálculo que lo ayudarán a realizar las mejores mediciones posibles.
32°F - 122 °F (0°C - 50 °C)
-76°F - 1022°F (-60°C - 550°C)
4,7" × 1,87" × 6,76" (119 mm × 47 mm × 172 mm)
en general, 180horas; 140horas de uso
continuo como mínimo (baterías alcalinas, sin láser ni retroiluminación)
21 CFR, Capítulo 1, Subcapítulo J
estas instrucciones. El incumplimiento de estas advertencias puede
Actúe con prudencia respecto de las mediciones en materiales
conocen los riesgos asociados a sus prácticas. Si bien este instrumento
algunas prácticas de seguridad comunes para quienes trabajan en entornos
Antes de utilizar el instrumento para determinar si un área es segura
capta los datos de temperatura.
4. Compartimento de baterías:
se extrae para su apertura y almacena
cambia el modo del visor.
cambia la escala de temperatura, disminuye el valor
sirve para ingresar al modo de autoescaneo
emisividad o alterna y pasa a retroiluminación según el modo.
apagado automático a los 60segundos de liberado el gatillo.
circular del objetivo enfocada por el sensor infrarrojo. Esta área se ampliará
a medida que la superficie objetivo se aleje de la pistola (para obtener más
y Celsius. Puede seleccionar cualquiera de las escalas durante la medición,
continuamente sin necesidad de apretar el gatillo. Se puede mantener el
gatillo presionado para activar el láser, pero no es necesario.
60segundos una vez liberado el gatillo. Para recuperar este valor o los
cálculos asociados, presione la tecla MODE (Modo), que activará el visor y
automáticamente después de 60minutos.
Alternar entre láser y retroiluminación
tipo K. Presione la tecla MODE (Modo) para ingresar al modo de sonda.
auxiliares en el modo específico.
temperatura medida sea mayor que el umbral de alarma.
temperatura medida sea menor que el umbral de alarma.
alta (HAL) configurada o menor que la alarma baja (LAL)
cambios rápidos de temperatura ambiente.
error. Apague el termómetro, retire las baterías, espere un
ESPAÑOL
Altitud de funcionamiento:
Humedad relativa:
Temperatura de operación:
Temperatura de almacenamiento:
Rango de medición:
Emisividad
Resolución óptica (Distancia: Objetivo):
Dimensiones:
Peso:
Tipo de batería:
Vida útil de la batería:
Resolución del visor:
Calibración:
Láser:
Potencia máxima:
Longitud de onda:
Conformidad con la normativa:
Protección ante caídas:
Certi caciones:
Láser Advertencia o precaución
Respete los
procedimientos de
eliminación segura
Batería
Niveles peligrosos
Modo Pantalla Tecla
DOWN
Tecla UP