K jump Health TEMPONITOR User Manual

Clinic Patch Thermometer
(Temponitor)
USER INSTRUCTION
Model KD-2100
C
15:38
TIME
37.0
Manufacturer :
K-Jump Health Co., Ltd. No. 56 Wu Kung 5 Rd., Wu Ku Industrial Park, Taipei Hsien, Taiwan 248
th
Contents Precaution About the Device (2)
Contents...............................................1
Precaution............................................2
About the device....................................2
Battery installation/replacement.............3
Setting the device..................................4
Preparation before measurement............5
Important information.............................6
Start measurement.................................7
Memory reading.....................................8
Using the flash light................................8
Turn off the device..................................8
Maintenance..........................................9
Error message......................................10
Specification........................................ 11
Intended for Use
This device is intended to be used to monitor the body temperature by applying the temperature patch to body skin or axillary.
Notice
1. Use of this thermometer is not to be a substitute for consulting with a physician.
2. Person with pacemaker implemented
should consult the doctor before using this device.
3. Intensive supervision is necessary when using this device on infant, children and compromised person.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1)this device may not cause harmful interference, and (2)this device must accept any interference received, including interference that may cause undesired operation.
1
FCC Interference Statement
FCC ID : RYZTEMPONITOR
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference with not occur in a particular installation. If this equipment does not cause harmful interference to radio or TV reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocated the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into and outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Caution: To assure continued FCC compliance
Any changes or modifications not expressly approved by the grantee of this device could be void the user's authority to operate the device.
About the Device (1)
The Patch
KD-PT01 Under Arm Patch
Temperature
KD-PT02 Body Skin Patch
Transmission PartTransmission Part
Sensor
Transmission Part
Temperature Sensor
2
KD-PT03 Ear Plug Patch
KD-2101 Temperature Reader
C
15:38
Memory Key
Start Key
Light Key
TIME
37.0
The LCD
Date/Time
DATE
(M/D)
TIME
Temperature
Battery installation / replacement
1. Remove battery cover by pulling the battery cover.
2. Install/replace four new 1.5 V Size- AAA batteries into the
battery compartment.
Note that be sure the batteries should have the correct polarity.
3. Close the battery cover.
3
Temperature Sensor
Transmission Part
LCD Display
On/Off Key
Memory
MEMORY
Temperature Unit
Low Battery
Light
Setting the Device
Preparation before Measurement
Important Information Start Measurement
Setting Date and Time
After installation/replacement of the batteries, the date and time are re-freshed to default value; a setting is necessary by following steps:
1. Under power-off mode,
M key
keep pressing the "M" key until it shows a flashing "2003" which is the setting mode of year.
2. Every clicking of the "M" key to enter the next mode said month date
DATE
(M/D)
DATE
(M/D)
hour minute to the
end.
3. Every clicking of the of "O/I" key will add one to the value for your setting.
4. After the setting, the
TIME
TIME
O/I key
O/I key (+ 1)
(+ 1)
device is always back to "power-off" mode.
Setting °C or °F
The KD 2101 provides user to select the unit in either °C or °F by following steps:
1. Click the "O/I" key to turn on the device.
2. The LCD will show all
TIME
segments first and enter to default unit mode; now, you have 5-second period for
(Keep pressing O/I key for 2 sec)
the setting; keep pressing the O/I key and the unit will switch to either °C or
TIME
°F you want.
4
About the sensor
The sensors is constructed by two parts: transmitter and temperature sensor. The transmitter is used for inducing power as well as data transmission; the temperature sensor is used for body temperature detecting.
Applying the Sensor
For fixing the sensor, a dressing (or surgical) adhesive tape will be needed. A high quality tape without side-effect is recommended since long period (usually 24~48 hr) adhesion.
1. Apply the sensor to ear canal, armpit or abdominal area below navel(depended on different types).
2. For armpit and body skin, fixing the sensor by medical adhesive tape(s). Be sure the firmness.
3. Subject is allowed to dress cloth rather than naked.
4. For infant or baby, the patch under diaper will benefit the measurement result since warmer measurement environment.
Insert the ear tape patch into ear canal.
Adhesive tape
sensor
Adhesive tape
sensor
5
Period for Heat Equilibrium
After applying the sensor, a 5~10 minutes waiting for temperature equilibrium between the sensor and your skin surface will be necessary before taking the first measurement.
Skin surface temperature
Room temperature
5~10 min.
Start data reading
Room Temperature Affects Skin Temperature
Skin is major organ for maintenance of body temperature; either hot or cold environment will cause skin temperature to rise or to fall. It is recommended that subject is kept in a comfortable room temperature (usually 22~28°C (71.6~82.4 °F) for taking the measurement.
Notice
When using the skin sensor or armpit sensor, wipe body skin with a dry towel .
6
1. Turn on the device, the device will perform system checking automatically until "ready for detecting" as figure:
TIME TIME
2. Place the reader on the upright direction of the transmitter around 3~4 cm (1,5 inches) distance.
w
i
th
in
(1
3cm
.
5inch)
Start key
3. Click the "start key", the device starts scanning the patch temperature with flashing "---".
TIME TIME
4. The result is then shown on the LCD with 2 beep sounds. Measurement is completed.
TIME TIME
5. For another reading, just repeat the steps mentioned above.
7