KIRWAN AURA 28-2000, AURA 28-2600 Operating Instructions And Specifications

Kirwan Surgical Products LLC
OPERATING INSTRUCTIONS AND SPECIFICATIONS FOR
THE
®
28-2000 - 70W Coagulator
28-2600 - 70W Coagulator w/Irrigation
IFU-70WGEN-5 Rev. K (11-2018)
TABLE OF CONTENTS
English 1
Français 13
Deutsch 25
Italiano 37
Español 49
Português 61
IFU-70WGEN-5 Rev. K (11-2018)
English
APPLICATION
®
This manual applies to the AURA
70 watt bipolar electrosurgical generators (Model Numbers
28-2000 and 28-2600) produced by Kirwan Surgical Products LLC
INDICATIONS
The AURA
®
70 watt bipolar electrosurgical coagulator is a general purpose solid state bipolar coagulator used to supply the RF signal to electrosurgical handpieces used on soft body tissues where a wide range of tissue types, patient conditions, and load impedances are encountered.
WARNINGS AND CAUTIONS
This device is restricted to use by a qualified, licensed practitioner. Federal (USA) law restricts this device to sale by or on the order of a physician.
Read this entire manual prior to use of this product.
Do not remove cover. Refer servicing to qualified service personnel.
Do not use in the presence of flammable gasses.
Use of this device in the presence of patients with pacemakers or active implants is not recommended. Approved qualified advice should be obtained prior to any such use.
If the AURA® 70 is used simultaneously with any physiological monitoring equipment, monitoring electrodes should be placed as far from surgical electrodes as possible. Needle monitoring electrodes are not recommended. Monitoring systems incorporating high frequency current limiting devices are recommended.
Connect the generator power cord to a properly grounded receptacle. Do not use adapters or extension cords.
Connect only bipolar accessories to the bipolar receptacles. Improper connection of accessories may result in inadvertent accessory activation or other potentially hazardous conditions.
Use only accessories whose voltage rating is greater than the maximum output voltage of the AURA
70 (210 V
). Accessories, including cables, should be routinely inspected for any
peak
®
damage prior to use. Refer to the instructions for use of each individual accessory as needed.
Temporarily unused active accessories shall be stored in a location that is isolated from the patient. Accessory cables shall be positioned in such as way as to avoid contact with the patient or other leads.
IFU-70WGEN-5 Rev. K (11-2018) 1
The lowest possible output setting necessary to achieve the desired effect should be used at all times.
Special precautions regarding electromagnetic compatibility (EMC) should be exercised prior to installing and putting the AURA® 70 into service. Refer to technical service manual IFU­70WSERV-5 for further information.
The AURA® 70 complies with standard EMC requirements when used with bipolar handpieces up to 10.5 inches (26.7cm) in length and bipolar cables up to 12 feet (3.7m) in length produced by Kirwan Surgical Products LLC (refer to product catalog L-GENCAT-6 or www.ksp.com for additional accessory information). The use of longer handpieces and/or cables, or those produced by other manufacturers, could adversely affect the EMC performance of the unit.
The AURA® 70 produces high frequency energy that may adversely affect the operation of other electronic equipment in its vicinity.
Portable or mobile communications equipment used in the vicinity of the AURA® 70 could potentially cause slight disturbances in the function of the device.
Any use of this device for tasks other than for which it is indicated, may result in damaged equipment or possible injury to the user or patient.
ESSENTIAL PERFORMANCE
The AURA® 70 shall not change output power settings during use. Unintended HF energization is not allowed.
ON-SCREEN ERROR MESSAGES
System Operation
Color Message
“Hot! Decrease
Yellow
Power” No
Stopped Cause Action
Internal temperature of the unit exceeds 40°C Decrease power setting.
“Pump Stopped.
Yellow
Yellow
Release pedal for reset” No “Pump Error. Refer for service” No
Maximum pump motor current limit is reached Release pedal. Problem with irrigation pump Contact service.
Contact service.
Red
“Error. Refer for service. Code XX.” Yes Internal malfunction
Reference "XX" error code.
IFU-70WGEN-5 Rev. K (11-2018) 2
CONTROLS AND INDICATORS
Figure 1
Pump Tube Slot - Accepts standard Kirwan irrigation cords Pump Module – Accepts silicone pump tube on standard Kirwan irrigation cords HF Out LED – Lit when bipolar output is activated Bipolar Output Receptacles – Accept standard 4mm banana pins Flow Rate Display – Displays flow rate setting of pump module in cubic centimeters per
minute (cc/min)
Flow Rate Adjust – Adjusts flow rate of pump module up or down Irrigation ON/OFF – Enables irrigation pump module ON in tandem with bipolar output Active/Inactive Display – Indicates status of bipolar output Power Level Display, Watts – Displays output power level setting in watts Power Level Adjust – Adjusts output power level up or down Time Display – Displays total elapsed time bipolar output is activated during procedure Reset – Resets time display to 00:00 Options – Accesses controls for screen brightness, speaker volume, and tube priming Power Level Display, Neuro – Displays output power level setting in Neuro units
IFU-70WGEN-5 Rev. K (11-2018) 3
Figure 2
Power On/Off Switch Fuse Holder Mains Power Cord Connector Equipotentiality Connection Slot for IV Pole – Located at top right corner of heat sink Footswitch Connector
SETUP
Connect the power cord to the power cord connector on the rear panel of the unit. Ensure the main power switch is in the off (0) position prior to connecting the cord.
Connect the footswitch to the footswitch connector on the rear panel of the unit.
Connect the banana pins of the bipolar cord into the bipolar output connectors.
If irrigation is to be used, refer to the figure below and load the silicone pump tube as follows:
Form the silicone pump tube into a semi-circle and loop it around the roller wheel of the
pump module, proximal end (closest to the spike) on top.
Pull the plastic connectors on either end of the pump tube along the slots toward the rear
of the unit, until the flanges on the connectors can be secured into place in the recesses located at the midpoint of the slots. Avoid over-stretching the tube.
Insert the spike at the proximal end of the irrigation line into the appropriate container.
Figure 3 – Loading Silicone Pump Tube
IFU-70WGEN-5 Rev. K (11-2018) 4
Connect the appropriate bipolar handpiece to the instrument end of the bipolar cord. Refer to the specific instructions for use provided with the accessory before using.
OPERATION
Turn the unit on via the main power switch on the rear panel. After a brief diagnostic routine, the unit should come up in the state shown in Figure 1 (flow and power will be set to 00).
To adjust the volume level of the audible tone:
Touch the OPTIONS button on the main screen Touch the SOUND VOLUME adjustment button and slide it left or right as needed, or
touch the right or left arrows to increase or decrease the volume.
Touch the EXIT button to return to the main screen
Adjust the output power level to achieve the desired effect by touching the power level adjustment arrows on the display screen. The power level will be displayed simultaneously in both watts and Neuro units.
Depress the footswitch pedal to activate the bipolar output to the accessory handpiece. An audible tone will sound indicating that the bipolar output is activated. Units will also provide visual indication via a blue “Active” indicator on the screen.
To reset the elapsed time display to 00:00, press the RESET button.
IRRIGATION OPERATION (Model 28-2600 only)
To prime the irrigation line with irrigating solution:
Touch the OPTIONS button on the main screen Touch the PRIME button. The pump motor will run continuously while this button is
touched. Observe distal end of handpiece for presence of irrigant.
Touch the EXIT button to return to the main screen
Adjust the flow rate as necessary by touching the flow rate adjustment arrows on the display screen. Note - Flow rates indicated on-screen are approximations. Variations in tubing durometer, solution container types, etc., may cause variations in actual flow output rates.
Touch the Irrigation ON/OFF button on the screen in order to allow flow of irrigating solution to the accessory handpiece via footswitch activation.
To return to activation of the bipolar output only, touch the Irrigation ON/OFF button again.
IFU-70WGEN-5 Rev. K (11-2018) 5
SPECIFICATIONS
Input: 90-240VAC, 50/60Hz, max 200VA Class: I Applied Part: Type BF, Defibrillation-proof Output: 512KHz, max 70W at 100-300 ohms load Rated load: 200 ohms Maximum output voltage: 210 V Fuses: 2 x T2.5A 250V Weight: 15 lbs. (6.8 kg) Footprint: 11 ½ in W x 8 in H x 11 in D (height does not include IV pole) (29.0 cm x 20.5 cm x 28.0 cm)
EXPLANATION OF EXTERNAL MARKINGS
peak
Consult the Instructions for Use (label) Warning! Dangerous voltage (front panel)
Defibrillator-proof type BF applied part (front panel) High Frequency current output (front panel)
Caution! Consult Accompanying Documents (rear panel) Footswitch connection (rear panel)
Non-ionizing radiation (rear panel) Date of Manufacture (rear panel) Equipotentiality (rear panel) Any accessories and all replaced parts must be disposed
of and recycled in accordance with local environmental regulations regarding the disposal of used equipment and waste (rear panel)
IFU-70WGEN-5 Rev. K (11-2018)
CE Mark
6
MAINTENANCE
CLEANING INSTRUCTIONS
Use a mild detergent and damp cloth to clean the generator cover, power cord, and footswitch. In the presence of significant amounts of splashed liquids, protecting the footswitch with a disposable plastic bag may be desirable in order to simplify cleaning and ensure optimum performance of the switch over the life of the product.
Use only a soft damp cloth to clean the touch screen display. No chemicals should be used. Clean the display only when the unit is turned off to avoid any inadvertent ac tivation.
The generator and all components must be completely dry before use. Do not sterilize the generator.
PREVENTIVE MAINTENANCE
At least once per year, the following inspections should be performed: Visually check the condition of the AC power cord and footswitch cord (fraying of cables,
loosening of connector ends, damaged insulation or contact pins, etc.). See the service parts list below if replacements are required.
Test the actual output power of the unit compared to the front panel setting, and the RF leakage current to ground, via an ESU analyzer or other appropriate method. The measured output power should not vary from the set power by more than 20%. The RF leakage current should not exceed 30mA. If the unit fails to comply with either of the se requirements, it should be r eturned to Kirwan Surgical Products LLC as per the last page of this document for further evaluation.
SERVICE/ACCESSORY PARTS LIST
Description Part Number Fuse, T2.5AL/250V 28-2000-70 Detachable Power Cord, Hospital Grade, 12 ft (3.7m) 26-2535 Square Pedal Footswitch w/ 15 ft (4.6m) Cord & 4 Pin Connector 28-2000-61 IV Pole 28-2600-40 Testing & Service Manual IFU-70WSERV-5
All other components of this product are NOT field-replaceable. Contact Kirwan Surgical Products LLC as per the last page of this document if servicing of other components is required.
IFU-70WGEN-5 Rev. K (11-2018) 7
TROUBLESHOOTING GUIDE
Symptom Possible Cause(s) Correction
Unit does not power up Power cord not plugged in correctly Check power cord connection. Blown fuse(s) Check/replace fuse(s) Bad power cord Check/replace power cord Fault in internal power supply Refer to qualified personnel for
servicing
No RF output Bipolar cord not plugged in correctly Check bipolar cord-ESU connection Instrument not plugged into bipolar
cord correctly
Footswitch cord not plugged in
correctly
Faulty footswitch Check/replace footswitch Bad bipolar cord Check/replace bipolar cord Eschar build-up on instrument tip Check/clean instrument tip
Fault in internal circuitry Refer to qualified personnel for
No irrigation flow Irrigation tube not installed correctly Check irrigation tubing installation Leak or blockage in irrigation tubing Check/replace bipolar irrigation cord Debris in pump roller assembly Refer to qualified personnel for
Fault in pump motor assembly Refer to qualified personnel for
LCD Display Screen goes blank
RF power drops unexpectedly
Eschar build-up on instrument tip Check/clean instrument tip Touch screen not
functioning
Fault in screen and/or other internal
LCD Display too dim Incorrect brightness/contrast
Fault in internal circuitry Refer to qualified personnel for
Audible tone too low Incorrect volume adjustment Adjust volume control Fault in internal circuitry Refer to qualified personnel for
Excessive Electrostatic Discharge (ESD) from operator to screen
Transient Voltage on Mains Line Release footswitch pedal and depress
Software error Cycle Power switch off and on again.
circuitry
adjustment
Check instrument-bipolar cord connection
Check footswitch cord connection.
servicing
servicing
servicing Cycle Power switch off and on again.
Refer to qualified personnel for servicing if necessary
again
Refer to qualified personnel for servicing if necessary
Refer to qualified personnel for servicing
Adjust brightness/contrast control
servicing
servicing
IFU-70WGEN-5 Rev. K (11-2018) 8
OUTPUT WAVEFORMS
80
70
60
Power vs load
50
40
Power (W)
30
20
10
0
0 100 200 300 400 500 600 700 800 900 1000
Load (ohms)
P10 P20 P30 P40 P50 P60 P70
Output versus Setting
70
60
50
40
30
Actual Output, W
20
10
0
0 10203040506070
Output Setting, W
IFU-70WGEN-5 Rev. K (11-2018) 9
CONVERSION CHART – WATTS vs. “Neuro” UNITS
Watts "Neuro" Units Watts "Neuro" Units
1 9 36 122 2 17 37 124 3 20 38 126 4 23 39 129 5 25 40 131 6 27 41 133 7 30 42 134 8 32 43 136
9 35 44 138 10 37 45 140 11 39 46 142 12 41 47 144 13 43 48 146 14 45 49 148 15 47 50 150 16 49 51 152 17 51 52 154 18 55 53 156 19 59 54 158 20 63 55 159 21 67 56 161 22 71 57 163 23 76 58 165 24 81 59 167 25 89 60 169 26 92 61 171 27 95 62 173 28 98 63 175 29 100 64 177 30 104 65 179 31 107 66 181 32 110 67 183 33 113 68 184 34 116 69 186 35 119 70 188
IFU-70WGEN-5 Rev. K (11-2018) 10
WARRANTY INFORMATION
WARRANTY
This product is guaranteed to be free from defects in material and workmanship in normal use in surgery. If the product breaks or fails to function in normal use due to defects in materials or workmanship, within one year of purchase, Kirwan Surgical Products LLC will repair or replace such equipment at no cost to the customer after an evaluation has been made.
The above warranty excludes modified or altered products, and applies only to the original buyer and is in lieu of all other warranties either expressed or implied.
LIMITATIONS OF CLAIMS
No claim of any kind, whether as to materials delivered or for non-delivery of materials shall be greater in amount than the purchase price of the materials in respect of which damages are claimed, and in addition, SELLER shall have no liability whatsoever for consequential damages, including, but not limited to, damages for injuries to persons or property. Claims for non-delivery of shortages must be submitted in writing within 30 days of date of the invoice. Claims for damaged material must be submitted within 30 days of the date of shipment. Deductions for late claims shall be disallowed.
OPEN SOFTWARE
This product includes software code developed by third parties, including software code subject to the GNU General Public License ("GPL") or GNU Lesser General Public License ("LGPL"). As applicable, the terms of the GPL and LGPL, and information on obtaining access to the GPL
Code and LGPL Code used in this product, are available to you by e-mailing a request to sales@ksp.com, or by calling the telephone numbers listed at the end of this document.
The GPL Code and LGPL Code used in this product is distributed WITHOUT ANY WARRANTY and is subject to the copyrights of one or more authors. For details, see the GPL Code and LGPL Code for this product and the terms of the GPL and LGPL.
IFU-70WGEN-5 Rev. K (11-2018) 11
RETURN OF UNUSED EQUIPMENT POLICY
1. There is no restocking charge for unopened products returned within 30 days as a result of order
error, if a replacement is ordered. Otherwise, there will be a 15% restocking charge based on the original purchase price.
2. To return unused equipment, in unopened packages, the BUYER must request a RETURN
AUTHORIZATION NUMBER by calling customer service at the telephone numbers provided below. All returns must be accompanied by paperwork listing products to be returned, a copy of the purchase order and invoice, and return authorization number.
3. All returns must be freight prepaid and sent to: Kirwan Surgical Products LLC 180 Enterprise Drive Marshfield, Massachusetts 02050 USA
Phone: (781) 834-9500 1-888-547-9267 (Outside of MA) FAX: (781) 834-0022 Web: www.ksp.com
For Regulatory Issues Only Medical Product Service GmbH Borngasse 20 35619 Braunfels, Germany
IFU-70WGEN-5 Rev. K (11-2018) 12
Français
APPLICATION
Ce manuel s'applique aux générateurs électrochirurgicaux bipolaires 70 watts AURA® (numéros de modèle 28-2000 et 28-2600) de Kirwan Surgical Products LLC.
INDICATIONS
Le coagulateur électrochirurgical bipolaire 70 watts AURA® est un coagulateur bipolaire à semiconducteurs général qui est utilisé pour fournir le signal RF aux pièces à main électrochirurgicales utilisées sur les tissus mous corporels les plus divers dans les situations où l'on rencontre un large éventail de conditions patient et d'impédances de charge.
AVERTISSEMENTS ET MISES EN GARDE
L'utilisation de cet appareil est réservée à un médecin diplômé qualifié. La loi f édérale (des États­Unis) limite la vente de ce dispositif à un médecin ou sur prescription de ce dernier.
Lire intégralement le présent manuel avant d'utiliser ce produit.
Ne pas retirer le capot. Confier toute réparation à un personnel de maintenance qualifié.
Ne pas utiliser en présence de gaz inflammables.
L'utilisation de cet appareil en présence de patients porteurs d'un stimulateur cardiaque ou d'un implant actif est déconseillée. Toute utilisation dans de telles conditions doit faire l'objet d'une autorisation qualifiée.
Si l'AURA® 70 est utilisé simultanément avec un quelconque équipement de surveillance physiologique, les électrodes de surveillance doivent être placées le plus loin possible des électrodes chirurgicales. L'utilisation d'électrodes-aiguilles de surveillance est déconseillée. L'utilisation de systèmes de surveillance incorporant des dispositifs de limitation des courants haute fréquence est recommandée.
Brancher le cordon d'alimentation du générateur sur une prise correctement reliée à la terre. Ne pas utiliser d'adaptateurs ou de rallonges.
Brancher uniquement des accessoires bipolaires sur les prises bipolaires. Le mauvais branchement des accessoires peut entraîner l'activation intempestive de l'accessoire ou une autre situation potentiellement dangereuse.
Utiliser uniquement des accessoires dont la tension nominale est supérieure à la tension de sortie maximale de l'AURA
70 (210 V
). Inspecter régulièrement l'état des accessoires, y compris
crête
®
des câbles, avant utilisation. Consulter le mode d'emploi de chaque accessoire au besoin.
Les accessoires actifs temporairement inutilisés doivent être rangés dans un endroit isolé du patient. Les câbles des accessoires devront être positionnés de manière à éviter tout contact avec le patient ou d'autres câbles.
IFU-70WGEN-5 Rev. K (11-2018) 13
Le réglage de sortie le plus bas possible nécessaire pour obtenir l'effet souhaité devra être utilisé à chaque fois.
Prendre des précautions spéciales concernant la compatibilité électromagnétique (CEM) avant l'installation et la mise en service de l'AURA® 70. Pour plus d'informations, se reporter au manuel technique IFU-70WSERV-5.
L'AURA® 70 est conforme aux exigences CEM standard lorsqu'il est utilisé avec des pièces à main bipolaires d'une longueur de 10,5 po (26,7 cm) maximum et des câbles bipolaires d'une longueur de 12 po (3,7 m) fabriqués par Kirwan Surgical Products LLC (pour plus d'informations sur les autres accessoires, se reporter au catalogue de produits L-GENCAT-6 ou consulter l'adresse www.ksp.com). L'utilisation de pièces à main et/ou de câbles plus longs ou d'autres fabricants risque de nuire à la performance CEM de l'appareil.
L'AURA® 70 produit une énergie haute fréquence qui risque d'altérer le fonctionnement des autres appareils électroniques se trouvant à proximité.
Les appareils de communication portables ou mobiles utilisés à proximité de l'AURA® 70 risquent de perturber légèrement le fonctionnement de l'appareil.
Toute utilisation de l'appareil à d'autres fins que ses indications prévues risque de détériorer l'appareil ou de blesser l'utilisateur ou le patient.
PERFORMANCE ESSENTIELLE
Le réglage de la puissance de sortie de l'AURA® 70 ne doit pas être modifié en cours d'utilisation. L'émission d'énergie HF involontaire n'est pas autorisée.
MESSAGES D'ERREUR
Couleur Message
Jaune
« Hot! Decrease Power »
« Pump Stopped.
Jaune
Release pedal for reset »
Jaune
Rouge
« Pump Error. Refer for service »
« Error. Refer for service. Code XX. »
Arrêt du système
Non
Non
Non
Oui
Cause Action
La température interne de l'appareil dépasse 40 °C
Diminuer le réglage de la puissance.
Limite d'intensité maximale du moteur de
Relâcher la pédale. pompe atteinte Problème au niveau de la pompe d'irrigation
Dysfonctionnement interne
Contacter le SAV.
Contacter le SAV et
indiquer le code
d'erreur « XX ».
IFU-70WGEN-5 Rev. K (11-2018) 14
COMMANDES ET VOYANTS
Figure 2
Fente pour tubulure de pompe - Accepte les cordons d'irrigation Kirwan standard. Module de pompe - Accepte une tubulure de pompe en silicone sur les cordons d'irrigation Kirwan
standard.
Voyant HF Out - S'allume quand la sortie bipolaire est activée. Prises de sortie bipolaires - Accepte des broches bananes standard de 4 mm. Affichage du débit - Affiche le réglage de débit du module de pompe en centimètres cubes par
minute (cc/min).
Réglage du débit - Permet de diminuer ou d'augmenter le débit du module de pompe. Irrigation ON/OFF - Permet d'activer le module de pompe d'irrigation en tandem avec la sortie
bipolaire.
Affichage Active/Inactive - Indique l'état de la sortie bipolaire. Affichage du niveau de puissance (en watts) - Affiche le réglage de la puissance de sortie en watts. Réglage du niveau de puissance - Permet de diminuer ou d'augmenter le niveau de puissance. Chronomètre - Affiche le temps total d'activation de la sortie bipolaire durant la procédure. Reset - Permet la remise à zéro du chronomètre. Options - Permet d'accéder aux commandes de luminosité de l'écran, de volume du haut-parleur et
d'amorçage de la tubulure.
Affichage du niveau de puissance (en Neuro) - Affiche le réglage de la puissance de sortie en unités
Neuro.
IFU-70WGEN-5 Rev. K (11-2018) 15
Figure 2
Interrupteur marche/arrêt. Porte-fusibles. Connecteur du cordon d'alimentation secteur. Borne d'équipotentialité. Fente pour potence i.v. - Située en haut à droite du dissipateur thermique. Connecteur de pédale.
INSTALLATION Brancher le cordon d'alimentation sur le connecteur du cordon (panneau arrière de l'appareil).
S'assurer que l'interrupteur marche/arrêt est en position (0) avant de brancher le cordon.
Brancher la pédale sur le connecteur de la pédale (panneau arrière de l'appareil).
Brancher les broches bananes du cordon bipolaire sur les connecteurs de sortie bipolaires.
Si l'irrigation doit être utilisée, se reporter à la figure ci-dessous et installer la tubulure de pompe en silicone de la manière suivante :
Former un demi-cercle avec la tubulure de pompe en silicone autour de la molette du
module de pompe, à l'extrémité proximale (côté trocart) sur le dessus.
Tirer sur les connecteurs en plastique à l'une ou l'autre extrémité de la tubulure le long des
fentes vers l'arrière de l'appareil, jusqu'à ce que les brides des connecteurs puissent être fixées en position dans les renfoncements situés à mi-distance des fentes. Éviter de tirer excessivement sur la tubulure.
Insérer le trocart à l'extrémité proximale de la ligne d'irrigation dans le récipient approprié.
Figure 3 - Mise en place de la tubulure de pompe en silicone
IFU-70WGEN-5 Rev. K (11-2018) 16
Brancher la pièce à main bipolaire appropriée sur l'extrémité instrument du cordon bipolaire. Avan t utilisation, lire le mode d'emploi fourni avec l'accessoire.
UTILISATION
Mettre l'appareil sous tension via l'interrupteur marche/arrêt à l'arrière de l'appareil. Après une courte séquence diagnostique, l'appareil devrait se présenter comme sur la figure 1 (débit et puissance à 00).
Pour régler le volume de la tonalité :
Appuyer sur le bouton OPTIONS de l'écran principal. Appuyer sur le bouton de réglage SOUND VOLUME et le glisser à gauche ou à droite
selon le cas, ou appuyer sur les flèches droite ou gauche pour augmenter ou diminuer le volume.
Appuyer sur le bouton EXIT pour retourner à l'écran principal.
Ajuster le niveau de la puissance de sortie pour obtenir l'effet souhaité en appuyant sur les flèches de réglage de la puissance à l'écran. Le niveau de puissance s'affiche sim ultanément en watts et en unités Neuro.
Appuyer sur la pédale pour activer la sortie bipolaire vers la pièce à main (accessoire). Une tonalité retentit, indiquant l'activation de la sortie bipolaire. L'activation est également confirmée par un indicateur visuel - un voyant Active bleu qui s'allume à l'écran.
Pour remettre le chronomètre à zéro, appuyer sur le bouton RESET.
UTILISATION DE L'IRRIGATION (modèle 28-2600 uniquement)
Pour amorcer la ligne d'irrigation avec la solution d'irrigation :
Appuyer sur le bouton OPTIONS de l'écran principal. Appuyer sur le bouton PRIME. Le moteur de la pompe tourne en continu tant que
l'utilisateur appuie sur ce bouton. Observer l'extrémité distale de la pièce à main pour vérifier la présence de solution d'irrigation.
Appuyer sur le bouton EXIT pour retourner à l'écran principal.
Ajuster le débit au besoin en appuyant sur les flèches de réglage du débit à l'écran. Remarque - Les débits indiqués sur l'écran sont approximatifs. Les différences de durométrie de tubulure, de types de récipient à solution, etc., peuvent causer des variations du débit réel.
Appuyer sur le bouton Irrigation ON/OFF de l'écran pour permettre la circulation de la solution d'irrigation jusqu'à la pièce à main (accessoire) en activant la pédale.
Pour reprendre l'activation de la sortie bipolaire uniquement, appuyer à nouveau sur le bouton Irrigation ON/OFF.
IFU-70WGEN-5 Rev. K (11-2018) 17
CARACTÉRISTIQUES TECHNIQUES
Entrée : 90-240 V~, 50/60 Hz, 200 VA max. Classe : I Pièce appliquée : type BF, à l'épreuve des défibrillateurs Sortie : 512 kHz, 70 W max. à une charge comprise entre 100 et 300 ohms Charge nominale : 200 ohms Tension de sortie maximale : 210 V Fusibles : 2 x T2.5A 250 V Poids : 15 lb (6,8 kg) Encombrement : 11 ½ po (largeur) x 8 po (hauteur) x 11 po (profondeur) (la hauteur n'inclut pas
la potence i.v.)
(29,0 cm x 20,5 cm x 28,0 cm)
EXPLICATION DES SYMBOLES
crête
Consulter le mode d'emploi
(le
label
)
Avertissement ! Tension dangereuse (panneau avant) Pièce appliquée de type BF à l'épreuve des défibrillateurs
(panneau avant) Sortie de courant haute fréquence (panneau avant)
Attention ! Consulter la documentation jointe (panneau arrière) Connexion de la pédale (panneau arrière)
Rayonnement non ionisant (panneau arrière)
Date de fabrication (panneau arrière)
Équipotentialité (panneau arrière)
Tous les accessoires et toutes les pièces remplacées doivent être mises au rebut et recyclées conformément aux réglementations environnementales locales concernant la mise au rebut des appareils usagés et des déchets (panneau arrière)
IFU-70WGEN-5 Rev. K (11-2018)
CE Mark
18
MAINTENANCE
INSTRUCTIONS DE NETTOYAGE
Utiliser un détergent doux et un chiffon humide pour nettoyer le capot du générateur, le cordon d'alimentation et la pédale. En présence d'une grande quantité de projections de liquide, il est conseillé de protéger la pédale à l'aide d'un sac en plastique jetable pour simplifier le nettoyage et garantir la performance optimale de l'interrupteur tout au long de la durée de vie du produit.
Utiliser uniquement un chiffon humide doux pour nettoyer l'écran tactile. L'utilisation de produits chimiques est interdite. Nettoyer l'écran uniquement lorsque l'unité est hors tension pour éviter toute activation intempestive.
Le générateur et tous les composants doivent être complètement secs avant usage. Ne pas stériliser le générateur.
MAINTENANCE PRÉVENTIVE
Effectuer les inspections suivantes au moins une fois par an : Vérifier visuellement l'état du cordon d'alimentation secteur et du cordon de la pédale (effilochage
des câbles, extrémités de connecteur desserrées, isolant ou broches de contact abîmés, etc.). Voir la liste des pièces détachées ci-dessous en cas de remplacement.
Comparer la puissance de sortie réelle de l'appareil avec le réglage du panneau avant et tester le courant de fuite RF à la terre, via un analyseur d'unité électrochirurgicale ou une autre méthode appropriée. La puissance de sortie mesurée ne doit pas varier de la puissance définie de plus de 20 %. Le courant de fuite RF ne doit pas dépasser 30 mA. Si l'appareil ne se conforme pas à ces exigences, il devra être renvoyé à Kirwan Surgical Products LLC comme indiqué à la dernière page de ce document pour des évaluations supplémentaires.
LISTE DES PIÈCES DÉTACHÉES/ACCESSOIRES
Description Numéro de référence Fusible, T2.5AL/250 V 28-2000-70 Cordon d'alimentation détachable, qualité hospitalière, 12 pieds (3,7 m) 26-2535 Pédale carrée avec cordon de 15 pieds (4,6 m) et connecteur à 4 broches 28-2000-61 Potence i.v. 28-2600-40 Manuel de test et de maintenance IFU-70WSERV-5
AUCUN autre composant de ce produit n'est remplaçable sur site. Pour la réparation d'autres composants, contacter Kirwan Surgical Products LLC comme indiqué à la dernière page de ce document.
IFU-70WGEN-5 Rev. K (11-2018) 19
GUIDE DE DÉPANNAGE
Symptôme Cause(s) possible(s) Mesure corrective
L'appareil ne se met pas sous tension Fusible(s) grillé(s) Vérifier/remplacer le(s) fusible(s)
Cordon d'alimentation défectueux Vérifier/remplacer le cordon d'alimentation Panne d'alimentation interne Contacter le SAV
Pas de sortie RF Cordon bipolaire mal branché Vérifier le branchement du cordon bipolaire
Instrument mal branché sur le cordon
Cordon de la pédale mal branché Vérifier le branchement du cordon de la
Pédale défectueuse Vérifier/remplacer la pédale. Cordon bipolaire défectueux Vérifier/remplacer le cordon bipolaire. Accumulation de tissus nécrosés sur
Erreur dans le circuit interne Contacter le SAV
Absence d'irrigation Tubulure d'irrigation mal installée Vérifier l'installation de la tubulure
Fuite ou blocage dans la tubulure
Particules étrangères dans l'ensemble
Panne du moteur de pompe Contacter le SAV
Cordon d'alimentation mal branché Vérifier le branchement du cordon
d'alimentation
sur l'unité électrochirurgicale. Vérifier le branchement du cordon
bipolaire
l'extrémité de l'instrument
d'irrigation
des galets de la pompe
d'alimentation sur l'instrument
pédale
Vérifier/nettoyer l'extrémité de l'instrument.
d'irrigation Vérifier/remplacer le cordon d'irrigation bipolaire Contacter le SAV
L'écran LCD s'efface. Décharge électrostatique excessive de
l'opérateur à l'écran
Chute de puissance RF intempestive
Accumulation de tissus nécrosés sur
Écran tactile en panne Erreur logicielle Mettre hors tension, puis remettre sous
Panne d'écran et/ou au niveau d'un
Écran LCD trop sombre Luminosité/contraste mal réglés Régler la luminosité/le contraste Erreur dans le circuit interne Contacter le SAV
Tonalité trop basse Volume mal réglé Régler le volume Erreur dans le circuit interne Contacter le SAV
Tension transitoire sur la ligne secteur Relâcher la pédale et l'enfoncer à nouveau
l'extrémité de l'instrument
autre circuit interne
Mettre hors tension, puis remettre sous tension. Contacter le SAV au besoin
Vérifier/nettoyer l'extrémité de l'instrument.
tension. Contacter le SAV au besoin
Contacter le SAV
IFU-70WGEN-5 Rev. K (11-2018) 20
,
FORMES D'ONDE DE SORTIE
80
70
60
Puissance/charge
50
40
30
Puissance (W)
20
10
0
0 100 200 300 400 500 600 700 800 900 1000
Charge (ohms)
P10 P20 P30 P40 P50 P60 P70
Sortie par rapport au réglage
70
60
50
W
40
30
Sortie réelle
20
10
0
0 10203040506070
Réglage de sortie, W
IFU-70WGEN-5 Rev. K (11-2018) 21