King Instrument 7450 Series Installation Instructions Manual

7450 Series
Installation Instructions
FLOW METER LIMITED WARRANTY
Meters are warranted against defects in materials and workmanship to the original user for a period of thir­teen (13) months from the date of factory shipment, provided the meter is installed, operated and maintained in accordance with King Instrument Company’s instructions and recommendations.
KING INSTRUMENT COMPANY, INC. MAKES NO WARRANTY AS TO THE FITNESS OF ITS PRODUCTS FOR SPECIFIC APPLICATIONS.
This warranty is valid for the original end-user only and does not apply to products that have been damaged or modied. This warranty is non-transferrable and is limited to replacement or repair. The liability of King In­strument Company arising out of its supply of the products, or their use, shall not in any case exceed the cost of correcting defects in the products as set forth above.
THIS WARRANTY IS A LIMITED WARRANTY AND SHALL BE IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. THERE ARE NO OTHER WARRANTIES WHICH EXIST BEYOND THE DESCRIPTION OR FACE HEREOF.
IN NO EVENT SHALL KING INSTRUMENT COMPANY BE LIABLE FOR LOSS OF PROFITS, INDIRECT, CONSEQUEN­TIAL OR INCIDENTAL DAMAGES.
Products should be returned, prepaid, to King Instrument Company, Inc. with proof of purchase. Call factory for Return Merchandise Authorization (RMA) number and return instructions.
THIS IS IMPORTANT INFORMATION. READ IT CAREFULLY BEFORE BEGINNING WORK.
1) Inspect meter for damage that may have occurred during shipping. Report any damage to the container to the freight carrier immediately.
2) Make sure your pressure, temperature, uid and other requirements are compatible with the meter (including o-rings.)
3) Select a suitable location for installation to prevent excess stress on the meter which may result from:
a) Misaligned pipe.
b) The weight of related plumbing.
c) “Water Hammer” which is most likely to occur when ow is suddenly stopped as with quick closing solenoid operated valves. (If necessary, a surge chamber should be installed. This will also be useful in pressure start-up situations.)
d) Thermal expansion of liquid in a stagnated or valve isolated system.
e) It is recommended to install valving which will allow the meter to be drained. Meter should be drained when not in use or prior to maintenance.
f) Instantaneous pressurization which will stress the meter and could result in tube failure.
NOTE: In closed thermal transfer or cooling systems, install the meter
in the cool side of the line to minimize meter expansion and contrac­tion and possible uid leaks at the threaded connections.
4) Handle the meter carefully during installation.
a) Use an appropriate amount of teon tape on external pipe threads before making connections. Do not use paste or stick type thread sealing products.
5) Install the meter vertically with the inlet port at the bottom.
6) Meters with stainless steel ttings will support several feet of pipe as long as signicant vibration or stress resulting from misaligned pipe are not factors.
7) Meters used in gas service should have suitable valves plumbed in at the inlet and outlet of the meter. These valves should be no more than 1-1/2 pipe diameters from the meter ports. The valve at the out­let should be used to create back pressure as required to prevent oat bounce. It should be set initially and then left alone. The inlet valve should be used for throttling purposes. Depending on the installation, valves may not be essential, but they are most useful in many instal­lations. Remember: To get a correct reading of ow in gas service, it is necessary to know the pressure right at the outlet of the meter (before the valve).
8) Pressure and temperature maximums must never be exceeded.
(714) 891-0008 • www.kinginstrumentco.comWhen it comes to ow...we’re instrumental.
1
Pressure and temperature ratings are based on a study of the engineering data
for particular materials used in construction and on the design of individual models. This information is supplemented by destructive test results. Meters with stainless enclosures must never be operated without shields securely in place. Meters exposed to difficult environments such as those created by certain chemicals, excessive vibration or other stress inducing factors could fail at or below the suggested maximums. Never operate meters above pressure and temperature maximums. It is strongly recommended that all meter installations utilize an appropriate pressure relief valve and/or rupture disc. The pressure settings and locations of these devices should be such that meters cannot be over pressurized. Meter failure could result in damage to equipment and serious personal injury. Always use suitable safety gear, including OSHA approved eye protection when working around meters in service. We are happy to pass along chemical compatibility information that has been published by the manufacturer's of raw materials used in our products; however, this information should not be construed as a recommendation made by King Instrument Company, Inc. for a specific application.
7450 meters that require repair should be sent to the factory. Please call for a Return Merchandise Authorization (RMA) number and return instructions.
2) Make sure your pressure, temperature, fluid and other
a) Misaligned pipe.
b) The weight of related plumbing.
c) "Water Hammer" which is most likely to occur when flow is suddenly stopped as with quick closing solenoid operated valves. (If necessary, a surge chamber should be installed. This will also be useful in pressure start-up situations.)
d) Thermal expansion of liquid in a stagnated or valve isolated system.
f) Instantaneous pressurization which will stress the meter and could result in tube failure.
a) Use an appropriate amount of teflon tape on external pipe threads before making connections. Do not use paste or stick type thread sealing products.
EPR 225 °F
Buna-N
Viton
Kalrez
275 °F 350 °F 400 °F
Non-valve
Inlet / Outlet Valve
4,000 psig
1,500 psig
Maximum Non-Shock Pressure and Temperature O-Ring Temperature
Ambient temperature
33° F - 125° F
Viton and Kalrez are registered trademarks of DuPont Dow Elastomers.
O-Ring Material Maximum
Temperature
Meters are not specifically recommended for service other than water or air. The user must determine meter suitability for use with other fluids.
e) It is recommended to install valving which will allow the meter to be drained. Meter should be drained when not in use or prior to maintenance.
-Extra caution must be exercised when meters are used in high pressure gas cylinder applications. Pressure regulators should be installed at the cylinder and at the inlet of the meter.
-Serious property damage and great personal injury could occur as the result of a meter misused or used in an unsuitable application.
Carefully remove the flowmeter from the piping system. Remove the front and back shield. Remove the end fittings using a
7
8
" wrench. Using a back off the compression screw by turning counter clockwise. This will release the compression of the glass tube. Carefully remove the screws that hold the outlet block. Remove the outlet block and internals. Remove glass tube.
Valve Models: Remove valve from the bottom block using a wrench. Disassembly of valve is not recommended. Outer valve o-ring should be
replaced during meter maintenance and cleaning.
To reassemble, position the glass tube on the bottom block and center onto bottom gasket. Insert float, if on a guide rod make sure the bottom end of the guide rod is inserted into the center hole of the bottom float stop. Re-install the outlet block. Using a
3
16
" hex key turn compression screw counterclockwise while keeping the glass tube centered between the bottom and top gaskets. Re-install end fittings, valve and shields.
All components are now fully accessible for cleaning. Components can be cleaned with a mild soap solution. This will be an effective cleaner of rust stains. a bottle brush may be helpful in cleaning the inside of the glass tube. Caution must be used so that materials of construction are not damaged by cleaning solutions. Hard water deposits can be removed with a 5% acetic acid solution (vinegar). Before the meter is reassembled, inspect all parts for damage. O-rings should be replaced during meter maintenance and cleaning.
Pressure and temperature ratings are based on a study of the engineering data
for particular materials used in construction and on the design of individual models. This information is supplemented by destructive test results. Meters with stainless enclosures must never be operated without shields securely in place. Meters exposed to difficult environments such as those created by certain chemicals, excessive vibration or other stress inducing factors could fail at or below the suggested maximums. Never operate meters above pressure and temperature maximums. It is strongly recommended that all meter installations utilize an appropriate pressure relief valve and/or rupture disc. The pressure settings and locations of these devices should be such that meters cannot be over pressurized. Meter failure could result in damage to equipment and serious personal injury. Always use suitable safety gear, including OSHA approved eye protection when working around meters in service. We are happy to pass along chemical compatibility information that has been published by the manufacturer's of raw materials used in our products; however, this information should not be construed as a recommendation made by King Instrument Company, Inc. for a specific application.
7450 meters that require repair should be sent to the factory. Please call for a Return Merchandise Authorization (RMA) number and return instructions.
EPR 225 °F
Buna-N
Viton
Kalrez
275 °F 350 °F 400 °F
O-Ring Material Maximum
Temperature
Meters are not specifically recommended for service other than
water or air. The user must determine meter suitability for use with other fluids.
-Extra caution must be exercised when meters are used in high pressure gas cylinder applications. Pressure regulators should be installed at the cylinder and at the inlet of the meter.
-Serious property damage and great personal injury could occur as the result of a meter misused or used in an unsuitable application.
Carefully remove the flowmeter from the piping system. Remove the front and back shield. Remove the end fittings using a
7
8
" wrench. Using a
3
16
" hex key, back off the compression screw by turning counter clockwise. This will release the compression of the glass tube. Carefully remove the screws that hold the outlet block. Remove the outlet block and internals. Remove glass tube.
Valve Models: Remove valve from the bottom block using a
7
8
" open ended
wrench. Disassembly of valve is not recommended. Outer valve o-ring should be
replaced during meter maintenance and cleaning.
To reassemble, position the glass tube on the bottom block and center onto bottom gasket. Insert float, if on a guide rod make sure the bottom end of the guide rod is inserted into the center hole of the bottom float stop. Re-install the outlet block. Using a
3
16
" hex key turn compression screw counterclockwise while keeping the glass tube centered between the bottom and top gaskets. Re-install end fittings, valve and shields.
All components are now fully accessible for cleaning. Components can be cleaned with a mild soap solution. This will be an effective cleaner of rust stains. a bottle brush may be helpful in cleaning the inside of the glass tube. Caution must be used so that materials of construction are not damaged by cleaning solutions. Hard water deposits can be removed with a 5% acetic acid solution (vinegar). Before the meter is reassembled, inspect all parts for damage. O-rings should be replaced during meter maintenance and cleaning.
Pressure and temperature ratings are based on a study of the engineering data
for particular materials used in construction and on the design of individual
models. This information is supplemented by destructive test results. Meters with
stainless enclosures must never be operated without shields securely in place. Meters exposed to difficult environments such as those created by certain chemicals, excessive vibration or other stress inducing factors could fail at or below the suggested maximums. Never operate meters above pressure and
temperature maximums. It is strongly recommended that all meter installations utilize an appropriate pressure relief valve and/or rupture disc. The pressure
settings and locations of these devices should be such that meters cannot be
over pressurized. Meter failure could result in damage to equipment and serious personal injury. Always use suitable safety gear, including OSHA approved eye protection when working around meters in service. We are happy to pass along chemical compatibility information that has been published by the manufacturer's
of raw materials used in our products; however, this information should not be construed as a recommendation made by King Instrument Company, Inc. for a
specific application.
7450 meters that require repair should be sent to the factory. Please call for a Return Merchandise Authorization (RMA) number and return instructions.
275 °F 350 °F 400 °F
Temperature
Meters are not specifically recommended for service other than
water or air. The user must determine meter suitability for use with other fluids.
-Extra caution must be exercised when meters are used in high pressure gas cylinder applications. Pressure regulators should be installed at the cylinder and at the inlet of the meter.
-Serious property damage and great personal injury could occur as the result of a meter misused or used in an unsuitable application.
OUTLET END FITTING
O-RING
COMPRESSION FITTING
COMPRESSION SEAT
INLET SEAL INLET FLOAT STOP /
INLET END FITTING
SIDE PLATE /
SIDE PLATE
SCALE SIDE PLATE
OUTLET SEAL OUTLET FLOAT STOP
FLOAT STOP /
FLOAT
GLASS METER TUBE
SHIELD (CLEAR)
SHIELD (WHITE)
END FITTING
END FITTING NUT SCALE BRACKET
SCALE BRACKET SCALE PLATE
SCALE PLATE
OUTLET END FITTING
O-RING
COMPRESSION FITTING
COMPRESSION SEAT
INLET SEAL INLET FLOAT STOP /
INLET END FITTING
SIDE PLATE /
SIDE PLATE
SCALE SIDE PLATE
OUTLET SEAL OUTLET FLOAT STOP
OUTLET FLOAT STOP /
FLOAT
GLASS METER TUBE
SHIELD (CLEAR)
SHIELD (WHITE)
END FITTING
END FITTING NUT SCALE BRACKET
SCALE BRACKET SCALE PLATE
SCALE PLATE
BLOCK
EXTENSION ASSEMBLY
BLOCK
SHIELD SCREW
HOLDER
EXTENSION ASSEMBLY
SCREW
SCREW
BLOCK
EXTENSION ASSEMBLY
BLOCK
SHIELD SCREW
HOLDER
EXTENSION ASSEMBLY
SCREW
SCREW
Carefully remove the flowmeter from the piping system. Remove the front and back shield. Remove the end fittings using a
7
8
" wrench. Using a
3
16
" hex key,
back off the compression screw by turning counter clockwise. This will release
the compression of the glass tube. Carefully remove the screws that hold the outlet block. Remove the outlet block and internals. Remove glass tube.
Valve Models: Remove valve from the bottom block using a
7
8
" open ended
wrench. Disassembly of valve is not recommended. Outer valve o-ring should be
replaced during meter maintenance and cleaning.
To reassemble, position the glass tube on the bottom block and center onto
bottom gasket. Insert float, if on a guide rod make sure the bottom end of the
guide rod is inserted into the center hole of the bottom float stop. Re-install the outlet block. Using a
3
16
" hex key turn compression screw counterclockwise while keeping the glass tube centered between the bottom and top gaskets. Re-install end fittings, valve and shields.
All components are now fully accessible for cleaning. Components can be cleaned with a mild soap solution. This will be an effective cleaner of rust stains. a bottle brush may be helpful in cleaning the inside of the glass tube. Caution must be used so that materials of construction are not damaged by cleaning
solutions. Hard water deposits can be removed with a 5% acetic acid solution (vinegar). Before the meter is reassembled, inspect all parts for damage. O-rings
should be replaced during meter maintenance and cleaning.
Pressure and temperature ratings are based on a study of the engineering data for particular materials used in construction and on the design of individual models. This information is supplemented by destructive test results. Meters with stainless enclosures must never be operated without shields securely in place. Meters exposed to difficult environments such as those created by certain chemicals, excessive vibration or other stress inducing factors could fail at or below the suggested maximums. Never operate meters above pressure and temperature maximums. It is strongly recommended that all meter installations utilize an appropriate pressure relief valve and/or rupture disc. The pressure settings and locations of these devices should be such that meters cannot be over pressurized. Meter failure could result in damage to equipment and serious personal injury. Always use suitable safety gear, including OSHA approved eye protection when working around meters in service. We are happy to pass along chemical
Carefully remove the flowmeter from piping system. Remove the 4 screws on each side holding the side plates on. Remove the side plates. Carefully remove the glass meter tube from the end fittings. Be sure to not let the internals fall out. Use caution when removing the glass meter tube. Do not allow float to fall out. Float damage may result in inaccuracy. All necessary instrument components are now fully accessible for cleaning with a bottle brush and appropriate mild soap solution*. Before the meter is reassembled, inspect all parts for damage. O-rings should be replaced during meter maintenance and cleaning.
To reassemble, install the glass meter tube back onto the end fittings. Reinstall the side plates. Tighten the 4 screws on each side. Reinstall the instrument into the plumbing system after removing the old teflon tape (with a wire brush) and replacing with fresh teflon tape.
*Do not use cleaning agents that will damage float, tube or o-rings.
Meters should be cleaned with a mild soap solution. This will be an effective cleaner of rust stains. Caution must be used so that materials of construction are not damaged by cleaning solutions. Hard water deposits can be removed with 5% acetic acid solution (vinegar).
7460 meters that require repair should be sent to the factory. Please call for a Return Merchandise Authorization (RMA) number and return instructions.
-O-rings should be replaced if meter is disassembled after it has been in service.
-Serious property damage and great personal injury could occur as the result of a meter misused or used in an unsuitable application.
CAUTION:
CLEANING:
REPAIR:
WARNING:
Highest immunity to
viscous fluids with
medium capacity.
Highest immunity to
viscous fluids with
medium capacity.
Pressure and temperature ratings are based on a study of the engineering data for particular materials used in construction and on the design of individual models. This information is supplemented by destructive test results. Meters with stainless enclosures must never be operated without shields securely in place. Meters exposed to difficult environments such as those created by certain chemicals, excessive vibration or other stress inducing factors could fail at or below the suggested maximums. Never operate meters above pressure and temperature maximums. It is strongly recommended that all meter installations utilize an appropriate pressure relief valve and/or rupture disc. The pressure settings and locations of these devices should be such that meters cannot be over pressurized. Meter failure could result in damage to equipment and serious personal injury. Always use suitable safety gear, including OSHA approved eye protection when working around meters in service. We are happy to pass along chemical
Carefully remove the flowmeter from piping system. Remove the 4 screws on each side holding the side plates on. Remove the side plates. Carefully remove the glass meter tube from the end fittings. Be sure to not let the internals fall out. Use caution when removing the glass meter tube. Do not allow float to fall out. Float damage may result in inaccuracy. All necessary instrument components are now fully accessible for cleaning with a bottle brush and appropriate mild soap solution*. Before the meter is reassembled, inspect all parts for damage. O-rings should be replaced during meter maintenance and cleaning.
To reassemble, install the glass meter tube back onto the end fittings. Reinstall the side plates. Tighten the 4 screws on each side. Reinstall the instrument into the plumbing system after removing the old teflon tape (with a wire brush) and replacing with fresh teflon tape.
*Do not use cleaning agents that will damage float, tube or o-rings.
Meters should be cleaned with a mild soap solution. This will be an effective cleaner of rust stains. Caution must be used so that materials of construction are not damaged by cleaning solutions. Hard water deposits can be removed with 5% acetic acid solution (vinegar).
7460 meters that require repair should be sent to the factory. Please call for a Return Merchandise Authorization (RMA) number and return instructions.
-O-rings should be replaced if meter is disassembled after it has been in service.
-Serious property damage and great personal injury could occur as the result of a meter misused or used in an unsuitable application.
CAUTION:
CLEANING:
REPAIR:
WARNING:
Highest immunity to
viscous fluids with
medium capacity.
Highest immunity to
viscous fluids with
medium capacity.
Pressure and temperature ratings are based on a study of the engineering data for particular materials used in construction and on the design of individual models. This information is supplemented by destructive test results. Meters with stainless enclosures must never be operated without shields securely in place. Meters exposed to difficult environments such as those created by certain chemicals, excessive vibration or other stress inducing factors could fail at or below the suggested maximums. Never operate meters above pressure and temperature maximums. It is strongly recommended that all meter installations utilize an appropriate pressure relief valve and/or rupture disc. The pressure settings and locations of these devices should be such that meters cannot be over pressurized. Meter failure could result in damage to equipment and serious personal injury. Always use suitable safety gear, including OSHA approved eye protection when working around meters in service. We are happy to pass along chemical
Carefully remove the flowmeter from piping system. Remove the 4 screws on each side holding the side plates on. Remove the side plates. Carefully remove the glass meter tube from the end fittings. Be sure to not let the internals fall out. Use caution when removing the glass meter tube. Do not allow float to fall out. Float damage may result in inaccuracy. All necessary instrument components are now fully accessible for cleaning with a bottle brush and appropriate mild soap solution*. Before the meter is reassembled, inspect all parts for damage. O-rings should be replaced during meter maintenance and cleaning.
To reassemble, install the glass meter tube back onto the end fittings. Reinstall the side plates. Tighten the 4 screws on each side. Reinstall the instrument into the plumbing system after removing the old teflon tape (with a wire brush) and replacing with fresh teflon tape.
*Do not use cleaning agents that will damage float, tube or o-rings.
Meters should be cleaned with a mild soap solution. This will be an effective cleaner of rust stains. Caution must be used so that materials of construction are not damaged by cleaning solutions. Hard water deposits can be removed with 5% acetic acid solution (vinegar).
7460 meters that require repair should be sent to the factory. Please call for a Return Merchandise Authorization (RMA) number and return instructions.
-O-rings should be replaced if meter is disassembled after it has been in service.
-Serious property damage and great personal injury could occur as the result of a meter misused or used in an unsuitable application.
CAUTION:
CLEANING:
REPAIR:
WARNING:
Highest immunity to
viscous fluids with
medium capacity.
Highest immunity to
viscous fluids with
medium capacity.
Pressure and temperature ratings are based on a study of the engineering data for particular materials used in construction and on the design of individual models. This information is supplemented by destructive test results. Meters with stainless enclosures must never be operated without shields securely in place. Meters exposed to difficult environments such as those created by certain chemicals, excessive vibration or other stress inducing factors could fail at or below the suggested maximums. Never operate meters above pressure and temperature maximums. It is strongly recommended that all meter installations utilize an appropriate pressure relief valve and/or rupture disc. The pressure settings and locations of these devices should be such that meters cannot be over pressurized. Meter failure could result in damage to equipment and serious personal injury. Always use suitable safety gear, including OSHA approved eye protection when working around meters in service. We are happy to pass along chemical
Carefully remove the flowmeter from piping system. Remove the 4 screws on each side holding the side plates on. Remove the side plates. Carefully remove the glass meter tube from the end fittings. Be sure to not let the internals fall out. Use caution when removing the glass meter tube. Do not allow float to fall out. Float damage may result in inaccuracy. All necessary instrument components are now fully accessible for cleaning with a bottle brush and appropriate mild soap solution*. Before the meter is reassembled, inspect all parts for damage. O-rings should be replaced during meter maintenance and cleaning.
To reassemble, install the glass meter tube back onto the end fittings. Reinstall the side plates. Tighten the 4 screws on each side. Reinstall the instrument into the plumbing system after removing the old teflon tape (with a wire brush) and replacing with fresh teflon tape.
*Do not use cleaning agents that will damage float, tube or o-rings.
Meters should be cleaned with a mild soap solution. This will be an effective cleaner of rust stains. Caution must be used so that materials of construction are not damaged by cleaning solutions. Hard water deposits can be removed with 5% acetic acid solution (vinegar).
7460 meters that require repair should be sent to the factory. Please call for a Return Merchandise Authorization (RMA) number and return instructions.
-O-rings should be replaced if meter is disassembled after it has been in service.
-Serious property damage and great personal injury could occur as the result of a meter misused or used in an unsuitable application.
CAUTION:
CLEANING:
REPAIR:
WARNING:
GV FLOAT (Rib Guided)
Highest immunity to
viscous fluids with medium capacity.
Highest immunity to
viscous fluids with medium capacity.
Pressure and temperature ratings are based on a study of the engineering data for particular materials used in construction and on the design of individual models. This information is supplemented by destructive test results. Meters with stainless enclosures must never be operated without shields securely in place. Meters exposed to difficult environments such as those created by certain chemicals, excessive vibration or other stress inducing factors could fail at or below the suggested maximums. Never operate meters above pressure and temperature maximums. It is strongly recommended that all meter installations utilize an appropriate pressure relief valve and/or rupture disc. The pressure settings and locations of these devices should be such that meters cannot be over pressurized. Meter failure could result in damage to equipment and serious personal injury. Always use suitable safety gear, including OSHA approved eye protection when working around meters in service. We are happy to pass along chemical
Carefully remove the flowmeter from piping system. Remove the 4 screws on each side holding the side plates on. Remove the side plates. Carefully remove the glass meter tube from the end fittings. Be sure to not let the internals fall out. Use caution when removing the glass meter tube. Do not allow float to fall out. Float damage may result in inaccuracy. All necessary instrument components are now fully accessible for cleaning with a bottle brush and appropriate mild soap solution*. Before the meter is reassembled, inspect all parts for damage. O-rings should be replaced during meter maintenance and cleaning.
To reassemble, install the glass meter tube back onto the end fittings. Reinstall the side plates. Tighten the 4 screws on each side. Reinstall the instrument into the plumbing system after removing the old teflon tape (with a wire brush) and replacing with fresh teflon tape.
*Do not use cleaning agents that will damage float, tube or o-rings.
Meters should be cleaned with a mild soap solution. This will be an effective cleaner of rust stains. Caution must be used so that materials of construction are not damaged by cleaning solutions. Hard water deposits can be removed with 5% acetic acid solution (vinegar).
7460 meters that require repair should be sent to the factory. Please call for a Return Merchandise Authorization (RMA) number and return instructions.
-O-rings should be replaced if meter is disassembled after it has been in service.
-Serious property damage and great personal injury could occur as the result of a meter misused or used in an unsuitable application.
CAUTION:
CLEANING:
REPAIR:
WARNING:
Highest immunity to
viscous fluids with
medium capacity.
Highest immunity to
viscous fluids with
medium capacity.
7450 Series
Maximum Non-Shock Pressure and Temperature
Temperature Pressure
Ambient Temp. 33° F -125° F
200° F 200 psig
CAUTION
• 7450 Series meters have o-ring seals. Use with incompatible uid­may cause o-rings to swell which may cause glass tube to fail.
• Extra caution must be exercised when meters are used in high pressure gas cylinder applications. Pressure regulators should be installed at the cylinder and at the inlet of the meter.
• Serious property damage and great personal injury could occur as the result of a meter misused or used in an unsuitable application.
CLEANING
Carefully remove the owmeter from the piping system. Remove the front and back shield. Remove the end ttings using a
3
ing a
/16” hex key, back o the compression screw by turning counter
7
/8” wrench. Us-
clockwise. This will release the compression of the glass tube. Carefully remove the screws that hold the outlet block. Remove the outlet block and internals. Remove glass tube.
Valve Models: Remove valve from the bottom block using a ended wrench. Disassembly of valve is not recommended. Outer valve o-ring should be replaced during meter maintenance and cleaning.
All components are now fully accessible for cleaning. Components can be cleaned with a mild soap solution. This will be an eective cleaner of rust stains. A bottle brush may be helpful in cleaning the inside of the glass tube. Caution must be used so that materials of construction are not damaged by cleaning solutions. Hard water deposits can be removed with a 5% acetic acid solution (vinegar). Before the meter is reassembled, inspect all parts for damage. O-rings should be replaced during meter maintenance and cleaning.
To reassemble, position the glass tube on the bottom block and center onto bottom gasket. Insert oat, if on a guide rod make sure the bottom end of the guide rod is inserted into the center hole of the bottom oat stop. Re-install the outlet block. Using a compression screw counterclockwise while keeping the glass tube centered between the bottom and top gaskets. Re-install end ttings, valve and shields.
7
/8” open
3
/16” hex key turn
Installation Instructions
REPAIR
7450 meters that require repair should be sent to the factory. Please call for a Return Merchandise Authorization (RMA) number and return instructions.
WARNING:
Pressure and temperature ratings are based on a study of the engi­neering data for particular materials used in construction and on the design of individual models. This information is supplemented by destructive test results. Meters with stainless enclosures must never be operated without shields securely in place. Meters exposed to dicult environments such as those created by certain chemicals, excessive vibration or other stress inducing factors could fail at or below the suggested maximums. Never operate meters above pressure and temperature maximums. It is strongly recommended that all meter installations utilize an appropriate pressure relief valve and/or rupture disc. The pressure settings and locations of these devices should be such that meters cannot be over pressurized. Meter failure could result in damage to equipment and serious personal injury. Always use suitable safety gear, including OSHA approved eye protection when working around meters in service. We are happy to pass along chemical compatibility information that has been published by the manufacturer’s of raw materials used in our products; however, this information should not be construed as a recommendation made by King Instrument Company, Inc. for a specic application.
FLOAT TYPES AND ORIENTATIONS
GV FLOAT LP FLOAT
SL FLOAT
(714) 891-0008 • www.kinginstrumentco.comWhen it comes to ow...we’re instrumental.
SP FLOAT
2
Loading...
+ 2 hidden pages