Kinetec Spectra Essential User Manual

Spectra Essential
User manual
Before use, please read this document. Kinetec reserves the right to effect technical modifications.
EN
Série 1 – 2 – 3 - 4
Notice Originale
Distributedby: Manufacturedby:
Spectra Essential EN 1/12
USER MANUAL
Summary
Page
• Definition _____________________________________________________________________ 2
• Warning and safety instructions ________________________________________________ 2
• Compliance __________________________________________________________________ 3
• Unpacking and packing ______________________________________________________ 3
• Installing the device __________________________________________________________ 3
• Description ___________________________________________________________________ 4
• Electrical connection _________________________________________________________ 5
• Procedure to start the machine ________________________________________________ 5
• Using the hand control
- Changing the display language _____________________________________________ 5
- START/STOP/REVERSE function _______________________________________________ 5
- Procedure to stop the machine _____________________________________________ 5
- Locking-Unlocking the hand control setting __________________________________ 6
- Setting the movement parameters __________________________________________ 6
- Possible values for each parameter __________________________________________ 6
Using the Plastic Comfort Case kit ________________________________________________ 7
Using the Kinetec Patient Pad kit _________________________________________________ 7
Setting up the patient ___________________________________________________________ 8
Options ________________________________________________________________________ 8
• Product information
- Maintenance ______________________________________________________________ 9
- Troubleshooting guide ______________________________________________________ 9
- Cleaning ___________________________________________________________________ 9
- Disposal and recycling ______________________________________________________ 9
- Technical specifications ____________________________________________________ 10
- Symbols used _______________________________________________________________ 10
- Warranty ___________________________________________________________________ 10
- Guidance and manufacturer’s declaration __________________________________ 11
Spectra Essential EN 2/12
Definition
The device KINETEC
®
Spectra Essential is a PASSIVE KNEE mobilisation device enabling extension and flexion
movements from -10° to 120°.
Indications
Arthroplasties of the knee and hip joints.
• Osteosynthesised femoral or tibial fractures.
• Patellar fractures.
• Arthrolysis and palliative surgery (cartilage lesions, removal of osteomas, etc.).
• Osteotomies of the pelvis or femur.
• Ligament repairs (LCI, LCE, LLI, LLE).
• Freeing the knee extensor mechanism (Judet's operation).
• Synovectomies, Meniscectomies, Patellectomies, Arthroscopies.
Clinical Benefits
• Effectively breaks the vicious circle: trauma, immobility, effusion and atrophy.
• Prevents joint stiffness of the knee and hip.
• Speeds the recovery of post-operative range of motion.
• Maintains the quality of the joint surface.
• Promotes joint cartilage healing.
• Prevents venous thrombosis.
• Provides immediate post-operative continuous passive motion.
• Reduces hospitalization time.
• Reduces the need for pain medication
Contraindications
Rheumatoid arthritis in the inflammatory phase, gout, algodystrophy in the inflammatory phase (hyper painful), para­osteo-arthroplasty, unhealed infected wounds, established phlebitis, bone cancer, myositis ossificans of the quadriceps, arthrodesis of the hip, infectious arthritis, deformed joint surfaces, paralysed limbs (atonic or spastic), non-stabilised fractures. The machine is not suitable for patients over 1.95 m (6'7") or under 1.45m (4'7") tall.
Warning and safety instructions
WARNING: The machine must be installed and commissioned according to the information provided in
this manual.
WARNING: If you need any assistance in the assembly, use or maintenance of the device, please
contact your Kinetec distributor.
WARNING: The practitioner determines the protocol and ensures its proper implementation (settings,
session duration and frequency of use). WARNING: Run a cycle with the device unloaded before installing the patient on the machine. WARNING: For optimum safety, always give the hand control to the patient before starting the system.
The patient must know the start/stop/reverse function on the hand control, see page 5. WARNING: To avoid the parameters being changed, lock the machine’s hand control before giving it
to the patient, see page 6. WARNING: Danger, risk of explosion: Do not use the machine with anaesthetic gas or in an environment
that is rich in oxygen. WARNING: For Type B Class I devices, and to avoid all risks of electric shock, the machine should only be
connected to a power supply that has protective earthing, see page 5. WARNING: Before using this machine, always check that the electrical socket is in good condition and is
suitable for the splint power supply cord. Only use the original cable supplied with the
machine. Check that the cables remain free around the device so that they do not get
damaged. WARNING: Before using this machine, always check that the machine is not damaged, in particular the
protective housings. WARNING: In case of electromagnetic interference with other devices move the device. WARNING: Please do not touch the fixed or moving parts while the unit is running: pinching or crushing
risk. Keep children and pets away from the machine. WARNING: Modifying the machine in any way is strictly forbidden. WARNING: Always check the motion parameters displayed on the hand control before starting the
device. WARNING: Only the accessories, spare parts and supplies described in this manual should be used with
this machine. WARNING: Do not connect the device to other devices not described in this manual. WARNING: If unforeseen events or malfunctions occur, please contact your KINETEC distributor. WARNING: Wireless communications devices, such as domestic wireless devices in networks, mobile
phones, wireless telephones and their base stations and walkie-talkies, may affect the
machine. You are recommended to keep at least a distance d between these devices and
the machine. See the table on page 12.
Spectra Essential EN 3/12
Compliance:
The device KINETEC
®
Spectra Essential complies with the standards of Directive 93/42/EEC, and bears the EC mark. The device KINETEC
®
Spectra Essential complies with the standards in force (IEC 60601-1-2) concerning the electromagnetic compatibility of medical devices and IEC 60601-1 concerning electrical safety. The device KINETEC
®
Spectra Essential meets the requirements of the Machinery Directive No. 2006/42/EC.
Unpacking and packing
Unpacking
When you unpack the machine, don't forget that you may need to pack it up again. We recommend that you keep the packaging materials, boxes and plastic bags.
Recommendations for plastic bags: do not put them over the head as there is a risk of suffocation, and keep them out of the reach of children. Be careful with small-sized pieces: they could be swallowed by a child.
Be careful with cords and cables: risk of strangulation.
Before using your machine you must move the foot plate into its working position. (See page 8).
Your machine is ready to be connected to the power supply. (See page 5).
Packing
To prevent any problems when the machine is transported, only pack it using its original packaging.
- Set the leg support to 42cm
- Stop the unit at 5° of flexion.
- Move the foot plate back into its packing position.
Installing the device
The device KINETEC
®
Spectra Essential machine is designed to be used in hospitals, clinics, doctors' offices or in private homes (rental).
The machine must be installed on a flat surface that is wide enough to accommodate the entire device plus the other leg. We recommend using the machine with a physiotherapy table, a healthcare bed, a bed or a bench. We do not recommend the use of an air mattress.
Spectra Essential EN 4/12
2
1
3
4
5
6
7
8
9
10
12
13
11
15
14
16
17
Description
The device KINETEC
®
Spectra Essential machine consists of the following components:
1 • Lower limb support. 2 • Thigh support. 3 • Articulated foot plate and hand control location for
transport. 4 • Hand control. 5 • Thigh support setting lock. 6 • Lower limb support setting lock. 7 • Foot plate position setting lock. 8 • Transport handle. 9 • ON/OFF switch and fuses.
10 • Liquid-crystal display
(2 lines of 16 characters). 11 • Increase / decrease keys. 12 • EXTENSION setting key. 13 • FLEXION setting key. 14 • STOP key. 15 • START key. 16 • PAUSE key. 17 • SPEED key.
Display Details:
A • 16-character line, used to display various messages
when starting up the machine; then the display does
not change while the machine is being used. B • 16-character line, used to display various messages
when starting up the machine; then it displays the
operational parameters. C • 3-character area showing the extension limit. D • 4-character area showing various messages: RUN,
STOP, EXT, FLEX. E • 3-character area showing the real-time angle of the
knee; this value changes with the current
movement. F • 3-character area showing the flexion limit.
Spectra Essential EN 5/12
EXT FLEX 30 STOP 15 70
Electrical connection: safety first.
2 versions of electrical protection are available for the device KINETEC
®
Spectra Essential machine. The identification label shows which version a machine is equipped with: Type BF, Class II devices, for home use, bear the following symbols:
Type B, Class I devices, for use in a professional environment, bear the following symbols: See page 10 for the meaning of the symbols.
Before connecting the device to the power supply, check that the mains voltage matches that shown on the identification plate (100-240V~ 50/60Hz). Connect the hand control (4). Connect the power supply cable (18).
IMPORTANT
For Type B Class I devices, and to avoid all risks of electric shock, the machine should only be connected to a power supply that has protective earthing. To connect the power supply, only use the original cable supplied with the machine. Check that the cables remain free around the device so that they do not get damaged. Check that the machine is not damaged, in particular the protective housings.
Procedure to start the machine
Press the ON / OFF switch (9). The display comes on, the machine carries out a self-test, and then the display shows: Your device KINETEC
®
Spectra Essential is ready to be used.
Changing the display language
Beginning
Keys to
press
Display Remarks
Switch the unit ON
KINETEC
40 STOP 50 110
Check if the hand control is not locked (See page 6).
Press simultaneously on the 2 keys
and
LANGUAGE
ENGLISH
The display indicates the language selected.
To change the language
Or
LANGUAGE
FRENCH
The French language is selected. Available languages: English, French, German, Italian and Spanish.
To confirm the new language
OK
SWITCH ON/OFF
Switch the machine off and then on again to apply the changed display language.
START / STOP / REVERSE function
As with all KINETEC systems, KINETEC
®
Spectra Essential machines are equipped with a
START/STOP/REVERSE function.
Press the key of the hand control. The movement stops,
Press the key of the hand control. The movement starts in the opposite direction.
CAUTION
For optimum safety, always give the hand control to the patient before starting the system.
Procedure to stop the machine:
To stop the machine's movement: press the key
To switch the unit off: press the ON / OFF switch (9) (see page 5)
4
18
9
Spectra Essential EN 6/12
Locking the hand control setting
The hand control allows the patient to control the machine as appropriate.
Simultaneously press the and keys to lock the hand control,
The display reads LOCK, you cannot change the parameters; if you try the message
LOCK appears. To unlock the hand control, simultaneously press the same keys. The display will show UNLOCK.
We recommend that you lock the hand control when you give it to the patient.
Note: the hand control locking is preserved when you switch the unit ON/OFF.
Setting the movement parameters
Select the parameter to be set:
Extension limit or flexion limit or speed or
pause at the extension or flexion limit; the setting to change will flash.
Press the or buttons to modify the setting; the new setting will flash.
To confirm the new setting, press another function button or wait approximately 5 seconds for automatic confirmation.
Movement parameters can be set either when the machine is stopped or when it is in operation.
Note: pressing the button repeatedly allows you to select pausing at the extension or flexion limit.
Possible values for each parameter:
Possible values Default setting
• Extension limit -10 to 115° 30°
• Flexion limit -5° to 120° 70°
• Speed
1 to 5 (from 45° to 155° per
minute)
2
• Extension pause 0 to 900 seconds (15 minutes) 0
• Flexion pause 0 to 900 seconds (15 minutes) 0
Spectra Essential EN 7/12
a
b
c
d
e
f
g
Using the Plastic Comfort Case kit
Plastic Comfort Cases are specially designed to improve comfort and hygiene for the patient. They have clips, fixed directly on the tubes of the machine's thigh and lower limb support segments, and straps with protection stops to precisely and quickly adjust to the patient's leg dimensions.
Cleaning
To ensure optimal hygiene, clean the supports after each patient use. Use a DISINFECTANT product (alcohol-free or <5% alcohol solution) in spray (plastic cases and metal components).
We recommend changing the cases every 500 hours of operation.
(See the Maintenance chapter for the running time counter).
Replacement parts
a 4670024048 Complete foot support b 4635010561 Foot support strap kit c 4635010157 Tibia case only d 4670024329 Tibia case with straps e 4635010165 Femur case only f 4670024337 Femur case with straps g 4650001876 Single strap Part number to order a complete kit:
- fastening with clips: 4670024345
- fastening without clips: 4670023701 (if your machine is not fitted with clips).
Using the Kinetec Patient Pad kit
The KINETEC
®
Patient Pad Kit is designed for rapid fitting, optimal hygiene and maximum patient comfort.
- For using and positioning the straps, please refer to here under. Make sure that the self-adhesive parts (19) are visible.
- Place the sponge side next to the skin.
FOR OPTIMAL HYGIENE, A NEW SET OF PADS SHOULD BE USED FOR EACH PATIENT.
Each cover is provided with a label to record the patient’s name
CLEANING:
- Sterilizing the pads (if necessary): Sterilize at 134 °C for 18 minutes.
- Disinfecting the straps: Wash at 30°C, using a disinfectant solution during the rinse cycle. Example of product that can be used: Solution "Bac linge" at 0.125 % or "Souplanios" at 0.125% from ANIOS Laboratories. A complete list of distributors in your country is available on request.
The device KINETEC
®
Spectra is delivered with a complete set, comprising:
- 4 straps (4650001107)
- 1 foot support (4650001420)
- 1 cover (4650001090)
Part number to order the complete set: 4650001868
19
Spectra Essential EN 8/12
20
7
L
5
3
6
21
22
24
23
7
Setting up the patient
See page 3, Installing the Device chapter, for the positioning conditions. Place the unit in a position that will be comfortable for the patient.
- Measure the length of the patient’s femur in cm or inches (L); adjust the thigh support to this measurement using knobs (5).
- Install the patient on the machine.
- Bring the foot plate (3) into contact with the patient's foot, then tighten both buttons (6).
- Adjust the plantar flexion (40°) or the dorsal flexion (30°) of the foot, with the knobs (7).
- Adjust the internal (30°) or external (30°) flexion of the foot, with the knob (20).
CAUTION
Adjust the axis of the patient’s hip (21) with the
“THEORETICAL” axis of rotation (22) of the unit,
and the axis of the patient’s knee (23) with the axis of articulation (24) of the unit.
Options
Trolley for all CPM Cart for bed use Seat adaptor Part number to order: Part number to order: Part number to order: 4655001053 4665003297 4670024098
Transport box Paediatric foot plate Part number to order: Part number to order: 4640001927 4670023777
Spectra Essential EN 9/12
25
Maintenance
After 2,000 hours of operation, or once a year, the device KINETEC® Spectra Essential requires lubrication and maintenance operations (lubrication of the joints, pointer stops and ball screws). The need for maintenance is indicated by display of the message SERVICE TIME Mx when the system is switched on. Despite this indication, you can continue to use your machine by pressing [START], but you should contact your nearest KINETEC technician to have the maintenance operations carried out as soon as possible. An after-sales service inspection sheet and the technical catalogue are available on request from your KINETEC distributor.
WARNING: Before using this machine, always check that the electrical socket is in good condition and is
suitable for the splint power supply cord. Only use the original cable supplied with the machine. Check that the cables remain free around the device so that they do not get damaged.
WARNING: Before using this machine, always check that the machine is not damaged, in particular the
protective housings.
When the machine is no longer in working condition, please return it to us, together with its accessories, for destruction.
A motor running time counter is available by simultaneously pressing keys and ;
the displays shows RESET TIME 215H (this is an example).
This counter can be reset by pressing the key .
Troubleshooting guide
A spare parts list and technical catalogue are available on request from your KINETEC distributor. If, after connecting the power supply cable to the power supply and switching on the machine:
• The display does not indicate any information:
- Check that the electrical socket is live using another device or voltmeter.
- Replace the fuse(s) (25) of the connector with fuses of the same type and calibre: 2 fuses T 750 mA 250V (6.3 x 32) (KINETEC order: 4610007434).
- Check that the hand control is connected properly.
- If the display still does not indicate any information, contact your nearest KINETEC technician.
• Your machine does not work and the display indicates 50 STOP 25 115, Press [START] again. If your machine still does not work, contact your nearest KINETEC technician.
• Your machine does not work and the display indicates:
"SERVICE D1": angle measurement function failure, or "SERVICE D2": no movement, or "SERVICE D3": abnormal consumption, or "PUSH STOP/START": power failure or disconnected motor. switch the machine off then on, contact your nearest KINETEC technician if the same message is displayed
Cleaning
Before carrying out any cleaning operation, SWITCH OFF the unit and disconnect the power supply. In order to ensure optimal hygiene, you are advised to clean the machine for each new patient. Cleaning should be carried out in the environmental conditions specified in the “Technical Specifications” section below. Use a DISINFECTANT product (alcohol-free or <5% alcohol solution) in spray. In order to ensure optimal hygiene, you are advised to clean the covers for each new patient. All the consumables enable hazard-free disposal.
Disposal and recycling
aPackaging: The packaging must be separated into plastic and paper / cardboard components and taken to
special recycling sites.
b • Kinetec patient pad kit: Clean with a disinfectant product then take it to special recycling sites. c • Unit: It contains electronic components, cables, aluminium, steel and plastic parts. When the machine is no longer
operational, disassemble it, separate it into different types of material and take these to authorised recycling centres or return the machine to Kinetec for destruction. Or contact the local authorities to determine the appropriate method of disposal for parts and accessories that are potentially hazardous to the environment.
Spectra Essential EN 10/12
Technical specifications
Product:
Lifespan of the machine: 12 years Weight: 12Kg (26 pounds) Splint dimensions: 95cm (37 inches) x 33cm (13 inches) x 33cm (13 inches) Angular limits: -10° to 120° Speeds: from 45 to 155° per minute Patient sizes: Full leg: 71 to 99 cm (28 to 39 inches) Tibia: 38 to 53 cm (15 to 21 inches) Femur: 33 to 46 cm (13 to 18 inches) Maximum weight of the user: 135 kg (297 pounds) Acoustic pressure: <70dB
Electricity:
Power supply: 100-240V~ Frequency: 50-60 Hz Power consumption: 50 VA Class: Device of Type BF Class II or Type B Class I Protection classification: IP 20 (Protected against solid objects greater than 12.5mm, but not protected against liquids) Fuse: T 750mA 250V 6.3 x 32mm (Kinetec order: 4610007434)
Environment:
Storage/transport conditions: Temperature: -25 to 70°C / -13 to 158°F. Relative humidity: up to 93% without condensation. Operating conditions: Temperature: 5 to 40°C / 41 to 104°F. Relative humidity: 15% to 93% without condensation. Atmospheric pressure: 700 hPa to 1060 hPa.
Symbols used
Warning or CAUTI ON (consult the accompanying documentation)
Speed
Flexion limit
STOP (power off)
Pause
Extension limit
ON (power on)
Increase
Start movement
Right way up when box is stored
Decrease
Stop movement
Contains electric and electronic components; do not throw away with household refuse.
Temperature Limit during storage or transport
~
Alternating current
Follow the instructions for use
Fragile
Keep dry during storage or transport
Follow the instructions for use
Class II device
TYPE BF device (protection against electric shocks)
TYPE B device (protection against electric shocks)
Warranty
The KINETEC warranty is strictly limited to the replacement, free of charge, or to factory repairs of part(s) recognised as defective. KINETEC guarantees its continuous passive motion systems for 2 years against all defects of manufacture from the date of purchase by the consumer. KINETEC is the only organization able to assess the application of the warranty to its systems. The warranty will be considered null and void if the device has been used abnormally or under conditions of use other than those indicated in the user's manual. The warranty will also be considered null and void in the event of deterioration or an accident due to negligence, inappropriate surveillance or inappropriate maintenance, or due to transformation of the equipment or an attempt to repair the equipment.
Spectra Essential EN 11/12
Guidance and manufacturer’s declaration
Guidance and manufacturer’s declaration - Electromagnetic emissions
The device KINETEC® SPECTRA Essential is intended for use in the electromagnetic environment specified below. The customer or the user of the device KINETEC
®
SPECTRA Essential should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
Radio frequency emissions CISPR 11 Group 1
The device KINETEC
®
SPECTRA Essential uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Radio frequency emissions - CISPR 11 Class B
The device KINETEC
®
SPECTRA Essential is suitable for use in all establishments including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions - IEC 61000-3-2 Class A Voltage fluctuations / Flicker emissions - IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration - Electromagnetic immunity
The device KINETEC® SPECTRA Essential is intended for use in the electromagnetic environment specified below. The customer or the user of the device KINETEC
®
SPECTRA Essential should ensure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD) IEC 61000-4-2
±2 kV, ±4 kV, ±6 kV
contact
±2 kV, ±4 kV, ±8 kV
air
±2 kV, ±4 kV, ±6 kV
contact
±2 kV, ±4 kV, ±8 kV
air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient / burst IEC 61000-4-4
±2 kV
for power supply lines
±2 kV
for power supply lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-4-5
±0,5 kV, ±1 kV
between lines
±0.5 kV, ±1 kV, ±2 kV
between line and earth
±0,5 kV, ±1 kV
between lines
±0.5 kV, ±1 kV, ±2 kV
between line and
earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage interruptions IEC 61000-4-11
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
When the interruption occurs, the device KINETEC
®
SPECTRA Essential is reset. After turning on, push START to
begin the session.
Voltage dips and voltage variations on power supply input lines IEC 61000-4-11
< 5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in UT)
for 5 cycles
70% U
T
(30% dip in UT)
for 25 cycles
< 5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in UT)
for 5 cycles
70% U
T
(30% dip in UT)
for 25 cycles
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
device KINETEC
®
SPECTRA Essential requires continued operation during power supply interruptions, we recommend powering the device KINETEC
®
SPECTRA Essential using an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field - EC 61000-4-8
3A/m 3A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE: UT is the AC mains voltage prior to application of the test level.
Spectra Essential EN 12/12
Guidance and manufacturer’s declaration
Guidance and manufacturer’s declaration - Electromagnetic immunity
The device KINETEC® SPECTRA Essential is intended for use in the electromagnetic environment specified below. The customer or the user
of the device KINETEC
®
SPECTRA Essential should ensure that it is used in such an environment.
Immunity test
Test level according to
EIC 60601
Compliance
level
Electromagnetic environment –
guidance
Mobile and portable RF communication devices should not be used closer to any part of the device KINETEC
®
SPECTRA
Essential machine, including its cables, than the recommended separation distance, calculated based on the equation applicable to the emitter's frequency.
Recommended separation distance
Conducted RF interference IEC 61000-4-6
3 Veff
from 150 kHz to 80 MHz
3 V
d = 1.2
P
Radiated RF interference IEC 61000-4-3
3 V/m
from 80 MHz to 2.5 GHz
3 V/m
d = 1.2
P from 80 MHz to 800 MHz
d = 2.3
P from 800 MHz to 2.5 GHz
where P is the emitter's maximum output power characteristic in watts (W), according to the emitter's manufacturer, and d is the recommended separation distance in metres (m).
The field intensities of fixed RF emitters, determined by an on­site electromagnetic investigation
a
, should be below the
compliance level in each frequency range
b
.
There may be interference near appliances bearing the following symbol:
NOTE 1 At 80 and 800 MHz, the highest frequency range is applicable. NOTE 2 These directives cannot be applied in every situation. Electromagnetic propagation is affected by absorption and reflection by structures, objects and people.
a
The field intensity of fixed emitters such as base stations for radio-telephones (cellular/cordless) and land mobile radios, amateur
radio, AM/FM radio broadcasts and TV broadcasts cannot be predicted exactly in theory. To evaluate the electromagnetic environment due to fixed RF emitters, an on-site electromagnetic investigation should be considered. If the field intensity measured where the device KINETEC
®
SPECTRA Essential machine is used exceeds the aforementioned applicable RF compliance level, the
device KINETEC
®
SPECTRA Essential machine should be monitored to check that it is working normally. If abnormal results are
observed, additional measures may be necessary, such as reorienting or repositioning the device KINETEC
®
SPECTRA Essential.
b
Over the frequency range 150 kHz to 80MHz, field intensities should be less than 3V/m.
Recommended separation distances between mobile and portable RF communication devices and the device KINETEC® SPECTRA Essential machine
The device KINETEC® SPECTRA Essential machine is designed to be used in an electromagnetic environment in which radiated RF interference is controlled. The customer or user of the device KINETEC
®
SPECTRA Essential machine can help prevent electromagnetic interference by maintaining a minimum distance between mobile and portable RF communication devices (emitters) and the device KINETEC
®
SPECTRA Essential machine, as recommended below, according to the communication device's maximum output power.
Maximum assigned output power
for the emitter
W
Separation distance according to the emitter's frequency
m
from 150 kHz to 80 MHz
d = 1.2
P
80 MHz to 800 MHz
d = 1.2 P
from 800 MHz to 2.5 GHz
d = 2.3 P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73 1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23 For emitters whose assigned maximum emitted power is not given above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the emitter frequency, where P is the emitter's maximum emission power characteristic in watts (W), according to the latter's manufacturer. NOTE 1 At 80 and 800 MHz, the separation distance for the highest frequency range is applicable. NOTE 2 These directives cannot be applied in every situation. Electromagnetic propagation is affected by absorption and reflection by structures, objects and people.
Distributedby: Manufacturedby: