Kinetec Spectra User Manual

Manuel d'utilisationManuel d'utilisation
Avant toute utilisation lire ce document. AbilityOne Kinetec se réserve le droit de toutes modifications techniques.
User manualUser manual
Before use, please read this document. AbilityOne Kinetec reserves the right to effect technical modifications.
BedienungsanleitungBedienungsanleitung
Vor Benutzung unbedingt dieses Dokument lesen. AbilityOne Kinetec behält sich das Recht vor, jegliche technische Änderung durchzuführen.
Istruzioni per l’usoIstruzioni per l’uso
Prima di mettere in funzione l’apparecchio leggere con attenzione il presente documento. La AbilityOne Kinetec si riserva il diritto di apportare modifiche tecniche.
Manual de empleoManual de empleo
Antes de cualquier utilización, lea este documento. AbilityOne Kinetec se reserva el derecho a cualquier modificación técnica.
F
GB
D
I
E
Your contact:
11
GB
SUMMARYSUMMARY
Page Page
• Definition__________________________2
• Description ________________________3
• Safety____________________________4
• Electrical connection __________________5
• Use of the hand control
- Select your language _______________________6
- Locking the hand control setting ________________7
- START/STOP/REVERSE function_________________7
- Quick Start _____________________________8
- Possible values for each parameter ______________8
- How to adjust the basic parameters of the movement ___9
- How to modify programs PROGRAM MODE __________10-12
- Using Programs __________________________13
- Reading the values of a program________________14
- How to use the WARM UP key _________________ 15
- How to use the MODULATION key _______________ 16
- How to use the BYPASS key __________________ 17
- Session Time____________________________ 18
• Patient Pad kit_______________________ 19
• Setting up the patient___________________ 20
• Options ___________________________ 21
• Product information
- Maintenance____________________________ 22
- Troubleshooting__________________________ 22
- Cleaning_______________________________ 23
- Technical specifications _____________________ 23
- Symbols used ___________________________ 24
- Warranty ______________________________ 24
- Maintenance sheet ________________________ 25
SAFETY RECOMMENDATIONSSAFETY RECOMMENDATIONS
• The physician defines the protocol and ensures that it is correctly implemented (adjustments, session time and frequency of use).
• We recommend that you lock the hand control when you give it to the patient.
• For optimum safety, always give the hand control to the patient before starting the system.
• Explosion hazard: KINETEC Spectra is not designed for use in the presence of flammable anaesthetics.
• Check that the electrical socket is in good condition and is suitable for the splint power supply cord. The latter complies with current standards and has a grounding socket. The plug may be connected to any standard socket. The socket must however have a grounded pin. To connect the power supply, only use the original cable supplied with the machine. Check that the cables remain free around the device so that they do not get damaged.
Manuel N°467896275 - Updated 21/02/03
22
GB
DefinitionDefinition
The KINETEC Spectra is a Knee PASSIVE mobilization device enabling the extension and flexion movement from -10° to 120°.
• Indications• Indications
Knee replacement surgery.
Fractures (patellar, tibia plateau, femoral,...).
Arthrolysis
Hip surgery, including hip replacement, hip pinning, osteotomy,...).
Ligament repairs.
Arthroscopic surgery (menisectomies, patellectomies,...)
Burns, joints sepsis,...
• Clinical Benefits• Clinical Benefits
Breaks the cycle of trauma, inflammation and the loss of range of motion.
Prevents joint stiffness.
Speeds the recovery of post-operative range of motion.
Maintains the quality of the joint surface.
Reduces pain and edema.
Promotes joint cartilage healing.
Reduces hospitalization time
Reduces the need for pain medication.
Provides immediate post-operative continuous passive motion.
Digital ROM readout on the patient hand control for positive reinforcement.
Maintains desired positions for stretching and muscular rest.
• Contraindications• Contraindications
Bone Cancer, Warped joint surfaces, Spastic paralyses, Unstable fractures, Uncontrolled infection. The machine are not adapted for patients height more 2 m(6'7") or under 1,40m (4'7").
33
GB
DescriptionDescription
KINETEC Spectra consists of the following components: 1 • Lower limb support.
2 • Thigh support. 3 • Foot support and hand control location for transport. 4 • Hand control. 5 • Thigh support setting lock. 6 • Lower limb support setting lock . 7 • Foot support positioning setting lock. 8 • Transport handle. 9 • ON/OFF switch and fuses.
10 • Liquid-crystal display (2 lines of 16 characters). 11 • SESSION TIME display key. 12 • BYPASS mode key. 13 • MODULATION key. 14 • WARM UP key. 15 • EXTENSION setting key. 16 • FLEXION setting key. 17 • Increase / decrease keys. 18 • START key. 19 • STOP key. 20 • FORCE key. 21 • SPEED key. 22 • PAUSE key. 23 • TIMER key. 24 • PROGRAM access key.
112233
4455667788
99
1010
1414
1313
1212
1111
1515
1616
1717
1818
1919
2020
2121
2222
2323
2424
44
GB
SafetySafety
The physician defines the protocol and ensures that it is correctly implemented (adjustments, session duration and frequency of use). The patient must know the start/stop/reverse function on the control handle. Hand control must be accessible to patient at all times. (See page 7). KINETEC Spectra complies with Directive 93/42/CEE.
EXPLOSION HAZARD:EXPLOSION HAZARD:
KINETEC Spectra is not designed for use in the presence of flammable anesthetics.KINETEC Spectra is not designed for use in the presence of flammable anesthetics.
In case of electromagnetic interference with other devices move the device. KINETEC Spectra is in compliance with standards in force (IEC 601.1.2), electromagnetic compatibility standard for medical devices.
55
GB
Kinetec
30 STOP 15 90
Please wait
KINETEC
Spectra V01.0
Electrical connectionElectrical connection
KINETEC Spectra is a type B class I device. Before connecting the device to the power supply, check that the mains voltage matches that shown on the identification plate (100-240 V~ 50-60Hz). Connect the hand control (4) Connect the power supply cable (25).
IMPORTANTIMPORTANT Check that the electrical socket is in good condition and is suitable for the splint power supply cord. The latter complies with current standards and has a grounding socket. The plug may be connected to any standard socket. The socket must however have a grounded pin. To connect the power supply, only use the original cable supplied with the machine. Check that the cables remain free around the device so that they do not get damaged.
Starting the unitStarting the unit
Switch on (9). While the unit begins an auto diagnostic, the display shows the following :
Your KINETEC Spectra is ready to be used.
25254499
66
GB
Use of the hand controlUse of the hand control
Select your languageSelect your language
Beginning Keys to press Display Remarks
Switch the unit ON Check if the hand control is not blocked
(See page 7).
Press the 2 keys in the same time The display indicate the language selected.
To change the language or
The English language is selected. English
French German Italian Spanish
To validate the new language.
To exit and confirm the new language, switch OFF and Switch ON the unit.
Kinetec
30 STOP 15 90
Please wait
KINETEC
Spectra V01.0
Language:
english
Language:
English
Ok
Switch on/off
77
GB
Use of the hand controlUse of the hand control
Locking the hand control settingLocking the hand control setting
The hand control allows the patient to control the machine as appropriate. Simultaneously press the and keys to lock the hand control, The display reads "LOCK", you can not change the parameters, if you try the display reads "LOCK". To unlock the hand control, simultaneously press the and keys, The display reads "UNLOCK".
We recommend that you lock the hand control when you give it to the patient.We recommend that you lock the hand control when you give it to the patient.
Comment: The hand control is unlocked when you switch the unit ON/OFF.
START/STOP/REVERSE functionSTART/STOP/REVERSE function
As with all KINETEC systems, KINETEC Spectra is equipped with a START/STOP/REVERSE function.
When the unit is running, the display reads RUN Press the key of the hand control. The movement stops. The display reads STOP. Press the key of the hand control. The movement starts in the opposite direction and the display reads RUN.
Caution:Caution:
For optimum safety, always give the hand control to the patient before starting the system.For optimum safety, always give the hand control to the patient before starting the system.
88
GB
Use of the hand controlUse of the hand control
Quick StartQuick Start
Set up the patient and proceed as below:
Beginning Keys to press Display Remarks
Switch the unit ON
Or
Check if the hand control is not blocked (See page 7).
If the Warm Up mode is selected, switch OFF this mode by pressing on the key
Start the session with the original parameters of the movement (default setting).
The value change at the speed of the movement.
Possible values for each parameter:Possible values for each parameter:
Possible values Default setting
• Treatement Mode
Normal
• Extension limit -10 to 115° 30°
• Flexion limit -5° to 120° 90°
• Speed
1 to 5 (from 30° to 160° per minute)
2
• Force 1 to 6 6
• Extension pause 0 to 900 seconds (15minutes) 0
• Flexion pause 0 to 900 seconds (15minutes) 0
• Timer No timer(00H00) to 24H00 0
Kinetec
30 STOP 15 90
Please wait
KINETEC
Spectra V01.0
WARMUP
30 STOP 15 90
KINETEC
30 run 45 90
99
GB
Use of the hand controlUse of the hand control
How to adjust the basic parameters of the movementHow to adjust the basic parameters of the movement
Beginning Keys to press Display Remarks
To stop the unit
Check if the hand control is not blocked (See page 7).
To display the extension or flexion limit of the movement
or The value blinks.
To change the limit if necessary Or The new value blinks.
To validate the new value, press another key or wait more than 3 seconds.
While the value blinks press the or key to change if necessary.
Or Or to display pause in extension or flexion limit
Successive presses on this key selects the pause at the extension or flexion limit.
To change the pause if necessary Or The new pause value blinks. To validate the new value press another key
or wait more than 3 seconds. The display shows the selected mode.
The unit is ready to start with the new parameters.
KINETEC
30 STOP 15 90
KINETEC
40 Flex 15 110
KINETEC
30 Ext 15 90
KINETEC
40 Flex 15 90
KINETEC
SPEED 2
No timer
KINETEC
PAUSE HIGH 0
Pause low 0
FORCE _ _ _ _ _ _
KINETEC
PAUSE low 15
KINETEC
40 STOP 15 110
1010
GB
Use of the hand controlUse of the hand control
How to modify programs PROGRAM MODEHow to modify programs PROGRAM MODE
Beginning Keys to press Display Remarks
To switch off the unitTo switch off the unit
Check if the hand control is not blocked (See page 7).
Press the two keys at the same time and switch the unit ON
Welcome text during 3 seconds.
Then The program number blinks.
To change the program if necessary or The new program number blinks.
To choose the treatment mode
The display indicates the selected treatment mode, the program number blinks again.
Or
To display the extension or flexion limit of the movement
or The value blinks.
KINETEC
Spectra V01.0
Program 2
EMPTY
Program 10
25 kinetec 110
Program 10
25 warmup 110
Program 10
25 kinetec 110
Program 10
30 warmup 100
Program 10
30 warmup 110
1111
GB
Use of the hand controlUse of the hand control
How to modify programs PROGRAM MODE (continued)How to modify programs PROGRAM MODE (continued)
Beginning Keys to press Display Remarks
To change the limit of movement if necessary or The new value blinks.
To validate the new value, press another key or wait more than 3 seconds While the value blinks press
the or key to change if necessary.
To record the program 10
Then
The program 10 has been recorded and the display indicates the next program so you can change another program.
OR To cancel the program
The program 10 has been cancelled and the display indicates the next program so you can change another program.
To exit program mode, switch OFF and switch ON the unit. To use the modified program see page 10.
Program 10
30 warmup 120
Program 10
SPEEd: 2
No timer
Pause low 0
FORCE _ _ _ _ _ _
Program 10
Save:+ clear:-
Program 10
Save
Program 11
EMPTY
Program 10
clearing
Program 11
EMPTY
KINETEC
Spectra V01.0
1212
GB
Use of the hand controlUse of the hand control
Comments Comments:
• When a program has been deleted, the display shows
Program table:Program table:
• Program number
Treatment Mode Flexion limit Extension limit Speed Force Flexion Pause Extension Pause Timer
11 22 33 44 55 66 77 88
99 1010 1111 1212 1313 1414 1515 1616
Program 11
EMPTY
1313
GB
Use of the hand controlUse of the hand control
Using ProgramsUsing Programs
The KINETEC Spectra allows you to store up to 16 programs, including the type of treatment, ROM, speed, load, pauses and timer. The original parameter values of the program are empty.
These values can be modified and recorded at any time (see ‘How to enter a program’ page 10).
Beginning Keys to press Display Remarks
To stop the unit
Check if the hand control is not blocked (See page 7).
To access the program mode The program number blinks.
To change the program if necessary or The new program number blinks.
To exit and validate the selected program
The current parameters have been recorded in program 3.
To exit without validation of selected program Back to the starting parameters.
Start the unit
The value change at the speed of the movement.
Comments:Comments:
• The values shown in the ‘Display’ column are examples. They actually depend on the stored programs.
• The current movement parameters can be changed while using that program but no data will be stored in the original program. See the programming mode (p 10) to modify programs.
KINETEC
40 STOP 15 110
PROGRAM 1
70 WARMUP 110
PROGRAM 3
25 warmup 90
KINETEC
Please wait
warmup
25 run 20 90
KINETEC
40 STOP 15 110
1414
GB
Use of the hand controlUse of the hand control
Reading the values of a program: example SPEEDReading the values of a program: example SPEED
Beginning Keys to press Display Remarks
To stop the unit
Check if the hand control is not blocked (See page 7).
To access the program mode The program number blinks.
To change the program if necessary or To change the program if necessary.
To read the speed value Displays the speed value.
After 15 seconds or after pressing on another key
To exit and validate the selected program
The current parameters have been recorded in program 3.
Start the unit
The value change at the speed of the movement.
Comments:Comments:
• The values shown in the ‘Display’ column are examples. They actually depend on the stored programs.
• The current movement parameters can be changed while using that program but no data will be stored in the original program. See the programming mode (p 10) to modify programs.
KINETEC
40 STOP 15 110
Program 1
70 WARMUP 110
PROGRAM 3
25 warmup 90
PROGRAM 3
Speed 5
PROGRAM 3
70 WARMUP 110
KINETEC
40 STOP 15 110
KINETEC
40 run 15 110
1515
GB
Use of the hand control Use of the hand control
How to use the WARM UP keyHow to use the WARM UP key
Beginning Keys to press Display Remarks
To stop the unit
Check if the hand control is not blocked (See page 7).
To select the Warm Up mode
The display shows the default settings (see page 8). To change the movement value if necessary see page 9.
Start the movement
The value change at the speed of the movement.
Warm Up rules: The Kinetec Spectra starts at 70% of the full ROM, increasing 5% of the range
every other cycle until the pre-set ROM is reached.
CommentsComments: The calculation mode used enables reaching the pre-set ROM in approximately seven full cycles.
ExampleExample: For a Warm Up treatment with a pre-set ROM from 0° to 100°. The first cycle starts from 15° to 85° to 15° and increase 5% each cycle.
KINETEC
40 STOP 15 110
WARM UP
30 STOP 15 90
PLEASE WAIT
WARM UP
15 RUN 15 110
100°
50°
Cycle 1 (70%)
Cycle 2 (75%)
Cycle 3 (80%)
Cycle 4 (85%)
Cycle 5 (90%)
Cycle 6 (95%)
Cycle 7 (100%)
Pre-Set ROM
1616
GB
Use of the hand control Use of the hand control
How to define the patient tolerance • At the start of a sessionHow to define the patient tolerance • At the start of a session
This function, MODULATION Mode is a way to set within the tolerance of a patient at the beginning of a session. Set up the patient and proceed as below:
Beginning Keys to press Display Remarks
Switch the unit ON
Or
Check if the hand control is not blocked (See page 7).
If the Warm Up mode is selected, switch OFF this mode by pressing on the key
To select the MODULATION Mode
The display indicates the keys to be used to drive the machine, this message is displayed 3 seconds.
To select the pain level ( for flexion, for extension)
or
The unit is moving to the selected way. It is waiting for you to select new limits.
To set the pain level when reached, immediately press or The new limit of the movement is recorded
To start the session with the new movement limits
The angle display changes with current movement.
Kinetec
30 STOP 15 90
Please wait
KINETEC
Spectra V01.0
WARMUP
30 STOP 15 90
MODUL.: use + or -
30 STOP 15 90
MODUL.: use + or -
30manual 50 90
MODUL.: use + or -
30manual 50 50
KINETEC
30 run 45 50
1717
GB
Use of the hand controlUse of the hand control
How to define the patient tolerance • During the sessionHow to define the patient tolerance • During the session
This function, The BY-PASS MODE is a way to address the pain threshold of a patient during a session.
IMPORTANT: Can be used only when the machine RUN.IMPORTANT: Can be used only when the machine RUN.
Beginning Keys to press Display Remarks
The unit is running
Or
Check if the hand control is not blocked (See page 7).
To select the BYPASS mode
The display indicates the keys used to run the machine. This message is displayed 3 seconds.
To select the NEW pain level or
The unit is moving to the selected way. It is waiting for you to select new limits.
To set the NEW pain level when reached, immediately press
or The new limit of the movement is recorded.
Continue the session with the new movement limits.
The angle display changes with current movement.
Kinetec
30 RUN 15 90
WARMUP
30 RUN 15 90
PYPASS: use + or -
30 RUN 15 90
BYPASS: use + or -
30bypass 100 90
bypass.: use + or -
30manual 50 50
WARMUP
30 run 45 50
1818
GB
TIME 02h25
30 STOP 15 90
Use of the hand controlUse of the hand control
Session TimeSession Time
This function shows the running time (in minutes) of the session (motor functioning).
• It directly accessible by the key , the display shows
This counter is reset each time the unit is switched ON.
1919
GB
Use of the Use of the Kinetec Patient Pad KitKinetec Patient Pad Kit
The KINETEC Patient Pad Kit is designed for rapid fitting, optimal hygiene and maximum patient comfort.
- For using and positioning the straps, please refer to here under. Make sure that the self-adhesive parts (26) are visible.
- Place the sponge side next to the skin.
FOR OPTIMAL HYGIENE, A NEW SET OF PADS SHOULD BE USED FOR EACH PATIENT.FOR OPTIMAL HYGIENE, A NEW SET OF PADS SHOULD BE USED FOR EACH PATIENT.
Each cover is provided with a label to record the patient’s nameEach cover is provided with a label to record the patient’s name
CLEANING:
- Sterilization of the pads ( if necessary): Sterilized at 134 °C during 18 minutes.
- Disinfecting of the pads: Washing at 30°C with use of a disinfecting solution during the rising cycle. Example of product that can be used: Solution " Baclinge" at 0.125 % or "Souplanios" at 0,125% from ANIOS Laboratory. A complete list of distributors in your country is available on request.
The KINETEC Spectra is delivered with a complete set. Components:
- 4 straps
- 1 foot support
- 1 cover
2626
2020
GB
Setting up the patientSetting up the patient
Place the KINETEC Spectra machine in a position that will be comfortable for the patient.
- Measure in cm or inches the length of the patient’s femur (L); adjust the thigh support to this measurement using knobs (5)
- To install the patient on the KINETEC Spectra machine.
- Push the foot plate (3) up to the patient and tighten the knobs (6).
- Adjust the plantar flexion (40°) or the dorsal flexion (30°) of the foot, with the knobs (7).
- Adjust the internal (30°) or external (30°) flexion of the foot, with the knob (27). IMPORTANTIMPORTANT
Adjust the axis of the patient’s hip (28) with the axis rotation (29) of the KINETEC Spectra machine,
and the axis of the patient’s knee (30) with the axis rotation (31) of the KINETEC Spectra machine.
LL5533662828
2929
3030
3131
2727
77
2121
GB
OptionsOptions
CartCart
2222
GB
Product informationProduct information
MaintenanceMaintenance
After 2,000 hours of operation, KINETEC Spectra requires a few lubrication and maintenance operations (lubrication of the joints, pointer stops and ball screws). The need for maintenance is indicated by display of the message SERV. MOTOR when the system is switched on. Despite that warning, you can continue to use your KINETEC Spectra by pressing START, but you should contact your nearest KINETEC technician to have the maintenance operations conducted as soon as possible.
• A motor running time counter is available. Simultaneously press keys and , the displays shows
this counter can be resetting by pressing the key .
TroubleshootingTroubleshooting
A spare parts list and technical catalog are available to you on request from your KINETEC distributor. If, after connecting the power supply cable to the power supply and switching on KINETEC Spectra:
• The display does not indicate any information:
- Check that the electrical socket is live using another device.
- Replace the fuse(s) (32) of the connector with fuses of the same type and calibre: 2 fuses T 750 mA 250V (6.3 x 32) (KINETEC order: 4610007434).
- If the display still does not indicate any information, contact your nearest KINETEC technician.
• Your KINETEC Spectra does not work and the display indicates 50 STOP 25 115, Press START again. Your KINETEC Spectra still does not function: Contact your nearest KINETEC technician.
• Your KINETEC Spectra does not function and the display indicates:
ANGULAR POSI.: angle measurement function failure, or NO MOVEMENT: no movement, or BAD WAY: motor rotation failure, or LOAD MAXI: abnormal consumption, or POWER SUPPLY: power failure;
Contact your nearest KINETEC technician if the same message is displayed after having switched the device off, then on, and started it by pressing START.
RESET TIME 924H
Reset: limit low
3232
2323
GB
Product informationProduct information
CleaningCleaning
Before conducting any cleaning operation, SWITCH the unit OFF and disconnect the power supply. Use a DISINFECTANT (PROPANOL/ISOPROPANOL or ALDEHYDE-based solution). Spray the disinfectant on the SURFACES (plastic shells and metal components). In order to ensure optimal hygiene, you are advised to clean the covers for each new patient. All the consumables enable hazard-free disposal.
Technical specificationsTechnical specifications
Product:Product:
Weight: 11Kg (24 pounds) Splint dimensions: 94cm (37 inches) x 33cm (13 inches) x 33cm (13 inches) Angular limits: -10° to 120° Speed: from 30 to 160° per minute for average femur length Patient height: full leg: 71 to 104 cm (28 to 41 inches)
Tibia: 38 to 58 cm (15 to 23 inches) Femur: 33 to 46 cm (13 to 18 inches)
Electricity:Electricity: Power supply: 100-240 V~ Frequency: 50-60 Hz Power consumption: 50 VA Device of type B class I IP 20. Fuse T 750mA 250V 6.3x32mm (KINETEC order: 4610007434)
EnvironmentEnvironment
- Storage/transport conditions: Temperature: -40 to 70°C / -40 to 160°F
Relative humidity: up to 90%
- Operating conditions: Room temperature: 10 to 40°C / 50 to 105°F
Relative humidity: up to 80%
2424
GB
Product informationProduct information
Symbols usedSymbols used
TYPE B device
(protection against electric shocks)
Speed Flexion limit
Caution (consult the accompanying documents)
Timer Warm Up
STOP (power off) Force Modulation
ON (power on)
Pause Bypass
Start movement Increase Session time
Stop movement Decrease
~
Alternative current
Program access Extension limit
WarrantyWarranty
The KINETEC warranty is strictly limited to the replacement free of charge or repair in the plant of the component or components found to be defective. KINETEC guarantees its joint passive mobilization systems for 1 year against all defects of manufacture from the date of purchase by the consumer. KINETEC is the only organization able to assess the application of the warranty to its systems. The warranty will be considered null and void if the device has been used abnormally or under conditions of use other than those indicated in the user's manual. The warranty will also be considered null and void in the event of deterioration or an accident due to negligence, inappropriate surveillance or inappropriate maintenance, or due to transformation of the equipment or an attempt to repair the equipment.
2525
GB
Product informationProduct information
Maintenance sheetMaintenance sheet
Serial number: _______________
Purchasing date: ____________________
Date: _________________________________________ Operations done: ___________________________________ Technician: _____________________________________ ______________________________________________ Running time (see page 22): __________________________ ______________________________________________ Exchange parts: __________________________________ ______________________________________________
_____________________________________________ ______________________________________________
Date: _________________________________________ Operations done: ___________________________________ Technician: _____________________________________ ______________________________________________ Running hours (see page 22): _________________________ ______________________________________________ Exchange parts: __________________________________ ______________________________________________
_____________________________________________ ______________________________________________