Rapid Digital Thermometer
(Flexible tip,Waterproof)
0197
Thermometric
Sensor
Thermometer tip
LCD
ON/OFF
Button
Battery
Compartment
Congratulations on your purchase of this product.
Please read the instructions carefully before using
the thermometer for the first time, and keep these
instructions in a safe place. This product is intended
for the measurement of human body temperature.
Operating Instructions
Before using, please disinfect the probe. (See Cleaning and disinfection
below.) To switch on, press the ON/OFF button next to the display; a
short beep will sound, indicating that the thermometer is operational.
At the same time the thermometer runs a self-check test, during which
all the digital segments appear on the LCD. When the letters “Lo”
and a flashing “°C” or “°F” display, the thermometer is now ready
for use. If the ambient temperature is below 32°C or 89.6°F, then
“lo°C” or “Lo°F” will appear on the LCD and if it is more than 42°C
or 107.6°F, then “Hi°C” or “Hi°F” will appear on the LCD. During the
reading, the current temperature is displayed continuously and the
“°C” or “°F” symbol flashes. The measurement is completed when a
constant temperature value has been reached. The temperature value is
considered constant when the temperature rises less than 0.1 C within 4
seconds. As soon as the constant temperature value is reached, a beep
will sound four times, and the “°C” or “°F”symbol will stop flashing. The
highest temperature measured appears on the LCD. However, please
note that this thermometer is a maximum thermometer, i.e. the displayed
temperature can increase slightly if measurement continues after the
beep. This is particularly the case with axillary (armpit) measurements,
should a temperature value be recorded which approximates the core
body temperature. In this instance please note the description under
“Methods of measuring temperature”. When the measurement is
completed, switch the thermometer off by pressing the ON/OFF button.
After the temperature has been displayed, the thermometer will shut off
automatically after 10 minutes.
Memory function
Switch the thermometer off, then press the ON/OFF button for about
2 seconds. The last measured value with “°C” or “°F” will appear
automatically on the LCD. This display is visible as long as the ON/
OFF button is held down. The reading is only over-written when a new
temperature value is recorded.
Methods of measuring temperature
It is important to remember that the body temperature reading depends
on the site where it is measured. For this reason, the measurement site
must always be specified in order to ensure that a correct temperature
reading is recorded.
In the rectum (rectal)
This is the most accurate method from a medical point of view, because
it comes closest to the core body temperature. The thermometer tip is
inserted carefully into the rectum for a maximum of 2 cm. The usual
measuring time is approximately 9 to 11 seconds.
Under the arm (axillary)
Placing the thermometer in the armpit provides a measurement of
surface temperature that can fluctuate by around 0.5°C to 1.5°C from
rectal temperature readings in adults. The usual measuring time for
this method is approximately 23 to 29 seconds. It should be noted,
however, that an exact reading cannot be obtained if, for example, the
armpits have been allowed to cool. If this is the case, we recommend
extending the measuring time by around 5 minutes in order to obtain
the most precise possible reading that corresponds as closely as
possible to the core body temperature.
In the mouth (oral)
There are different heat zones in the mouth. As a general rule, the oral
temperature is 0.3°C to 0.8°C lower than the rectal temperature. To
ensure that reading is as accurate as possible, place the thermometer
tip to the left or right of the root of the tongue. The thermometer tip must
have constant contact with the tissue during the reading and be placed
under the tongue in one of the two heat pocket at the back. Keep
the mouth closed during the reading and breathe evenly through the
nose. Do not eat or drink anything before the measurement. The usual
measuring time is approximately 15 to 19 seconds.
Note: Rectal testing is the most precise method for identifying the basal
temperature. For most accurate results, you can extend the measuring
time by 3 minutes after the beep.
Cleaning and disinfection
The best way to clean the thermometer tip is by applying a disinfectant
(e.g. 70% ethyl alcohol) with a damp cloth. This waterproof
thermometer can be immersed in liquid or lukewarm water for thorough
cleaning and disinfection.
Please see other side of this pamphlet for important information.
Safety precautions
•Do not allow the device to come into contact with hot water.
•Do not expose to high temperatures or direct sunlight.
•Do not drop the thermometer. It is neither shock-proof nor
impact-resistant.
•Do not bend or open the device (except the battery compartment).
•Do not clean with thinners, petrol or benzenel. Only clean with water
or disinfectant.
•Do not immerse in liquid for longer than 30 minutes.
•The thermometer contains small parts (battery, battery compartment)
which can be swallowed by children. For this reason, do not leave
the thermometer unattended in the hands of children.
•Avoid bending the tip by more than 45 degrees.
•If the ambient temperature is over 35°C or 95°F, dip the thermometer
tip in cold water for approx. 5 to 10 seconds prior to measuring
the temperature.
•Persistent fever, particularly in children, has to be treated by a doctor.
Contact a doctor right away if there are signs of a persistent fever.
Battery replacement
The battery is empty and needs replacing when the “ “ or “ “ battery
symbol appears on the right of the LCD. Remove the battery cover and
replace it with a battery (preferably non-mercury) of the same type.
Please note: The “+” sign up and “–” sign down.
Product disposal
Please ensure environmental protection. Batteries do not belong in
domestic waste. Please turn them in at a collection point or a municipal
recycle material center as special waste.
This symbol on products and/or accompanying documents means that
consumed electronic products must not be mixed with conventional
domestic waste. Take these products to the corresponding collection
points for correct treatment and recycling, where they will be accepted
free of charge. For more information on the closest collection point,
please check with your local authorities.
Technical data
Type: maximum thermometer
Measurement range: (32.0~42.0)°C/(89.6~107.6)°F
Measurement accuracy:
+/-0.1°C/0.2°F(35.5 C~42.0°C/95.9°F~107.6°F)
+/-0.2°C/0.4°F (for the remaining measurement range)
Storage/transportation temperature: (-25~55)°C, <_95%RH
Ambient temperature during use: (5~35)°C, <_80%RH
Battery type:
Alkallmanganese battery, type LR41, 1.5V, service life
minimum 100 hours under continuous operation.
Weight: Approx. 13g
Explanation of symbols
or battery is empty
Product disposal instructions for electronic devices
The battery in this product complies with the
requirements stated in European Directives
2006/66/EEC.
Lo°C or Lo°F: temperature under 32°C or 89.6°F
Hi°C or Hi°F: temperature over 42°C or 107.6°F
Type B equipment
Legal requirements and guidelines
This product complies with the European Directive for Medical Devices
93/42/EEC and carries the CE mark. The device also complies with
the specifications of the European Standard for clinical thermometersPart 3: Performance of compact electrical thermometers (nonpredictive
and predictive) with maximum device. The CE marking confirms that
this is a medical device with a measuring function in the sense of the
Medical Devices Act, which has undergone a conformity assessment
procedure. A Notified Body(TUV Produkt Service GmbH) confirms that
this product fulfils all the appropriate statutory regulations.
Calibration check
This thermometer is initially calibrated at the time of manufacture. If this
thermometer is used according to the operation instruction, periodic
re-adjustment is not required.
A calibration check has to be carried out immediately if there are
indications that the product does not keep the defined error limits or
the calibration properties could have been affected by an intervention
or by any other means. Please also observe any national statutory
regulations. The calibration check can be carried out by authorized
service providers. Test instructions for a calibration check can be
provided to authorized services providers on request.
Warranty
This product is covered under warranty for 1 year after the original sale
date. Damage resulting from incorrect use or abuse is not covered by
the warranty. Battery and packaging are excluded from the warranty.
Claims beyond this, including claims for damages, are excluded. If
you believe that the thermometer is not working, please first check
the battery. If replacing the battery does not fix the issue, please call
877.334.6876 or email service@blufire.com to get authorized return
instructions from the manufacturer.