Thermo Scientific
EMCPro PLUS
Advanced EMC test system for compliance
testing to 6 IEC/EN standards
The Thermo Scientific EMCPro PLUS
test system features resident
capabilities for EMC CE Mark
compliance testing to 6 IEC/EN
standards, and fully addresses new
requirements for a 100 kHz burst
rate per IEC 61000-4-4, Edition 2 (EFT)
and 80% dip per IEC 61000-4-11,
Edition 2 (PQF™).
Product Specifications
Features
• Portable, mid-range EMC test system
• Resident capabilities for compliance
testing to 6 IEC/EN standards
•
Addresses ANSI/IEEE, ITU, ETSI
and UL standards
• Surge testing to 6.0kV with Combination,
Telecom, and Ring Waves
• Monitors surge voltage and current
at the output terminals
• Monitors output of the coupling unit
and automatically switches connections
according to coupling mode
• Highest test levels, widest selection
of tests and lowest in-use costs
• Upgradable as standards change
A true, total immunity EMC test system
The Thermo Scientific EMCPro PLUS is the
answer to manufacturers’ demand for a
midrange, multi-capability EMC immunity
tester
. It operates via our easy to use
Windows
front panel, and is easily configured to meet
immunity standards required for CE Marking
and compliance requirements.
ideal for companies who require flexibility,
versatility, and the highest test level-to-cost
ratio instrument on the market.
test beyond levels dictated by the EMC
Directive, including demanding national and
international standards such as ANSI/IEEE,
ITU, ETSI, and UL, as well as company- and
market-driven test programs, which ensure
quality and reliability in the field.
®
- based PC software or from the
Portable and low cost, EMCPro
It also accommodates companies that
®
PLUS is
Users can configure custom test protocols
to meet specific test requirements using the
full line of options and accessories, including
mains and I/O line coupler/decouplers,
magnetic field monitors, coils, and more.
When test requirements change, or standards
evolve, upgrading is a simple matter of
adding appropriate options or accessories.
Breaking the 4.4kV voltage barrier
for combination W
and Ring Wave Surge testing
The EMCPro PLUS features surge testing
to 6kV with the combination, telecom and
ring waves. It is the only combination tester
on the market to offer the combination
wave with one of two additional built-in
surge waveforms.
The EMCPro PLUS, and its many
technically advanced, cost effective features
will serve you well for many years to come.
ave, Telecom

Thermo Scientific EMCPro PLUS
Model PRO-BASE
EMCPro PLUS Base Unit
General Specifications
System Voltage 90-240VAC, 50/60Hz
Integrated EUT Mains Coupler/Decoupler
AC Voltage 1 phase, 50 - 250 VAC. 50/60 Hz
AC Current 16 A maximum**
DC Voltage 100 VDC maximum
DC Current 10 A maximum
Frequency 50/60 Hz
EUT Connectors Nema, British, Schuko
Control Interface
Interface RS232 Fiber-optic
Safety Features
External Interlock for users
Interlock for CCL connector
External stop input
Environmental Operating Conditions
Temperature +15°C to +40°C
Humidity 10-75%, non-condensing
Altitude 8000 ft. maximum
Physical Specifications
Height 22.9 cm (8.7 in)
Width 43.4 cm (17.1 in)
Depth 64.8 cm (25.5 in)
Weight 39 kg (85 lb)
CE Marking
Safety and EMC Directives
Model PRO-EFT
EFT per IEC 61000-4-4 Edition 2, EN 61000-4-4 and ANSI C62.41
General Specifications
Voltage Waveform 5/50 ns ±30%
Peak Voltage 250 V - 4.4k V ±5%
Burst Period 300 ms ±10%
Burst Duration 15 ms ±20%, for pulse frequencies up to 5 kHz,
0.75 ms above 5 kHz
Frequency 1-100 kHz, in 0.5k Hz steps, ±10%
DC Blocking Capacitor 10 nF (internal)
Options Available
Model CM-3CD-16/32 16 or 32 Amp, 3 phase EFT and
surge coupler/decoupler
Model CCL Capacitive coupling clamp
Model CCLC Coupling clamp cover
Model EFT-ATTN EFT attenuator for oscilloscope monitoring
Model PRO-RING**
Ring Wave Surge per ANSI C62.41 Cat. A, B, UL 864, IEC 61000-4-12
and EN 61000-4-12
General Specifications
Voltage Waveform 100 kHz damped cosine
Peak Voltage 250 - 6.0k V ±5%
Repetition Rate Fewer than 4 per minute at 6.0 kV,
faster at lower voltages
Open-Circuit Voltage Rise Time: 0.5 µs ±30%
Short-Circuit Current Vp/Ip: 12 ±3 or 30 ±8 software selectable
Options Available
Model CM-3CD-16/32 16 or 32 Amp, 3 phase EFT and
surge coupler/decoupler
** PRO-TELECOM and PRO-RING can not be installed in same unit.
Model PRO-SURGE
Surge for compliant testing per IEC 61000-4-5, EN 61000-4-5,
ANSI C62.41 Category B and
General Specifications
Voltage Waveform 1.2/50 µs
Peak Voltage 250V - 6.0 kV ±5%, 12 mode
Peak Current 125 A - 3.0k A ±10%
Additional 10 Resistor Software selectable
Repetition Rate Up to 4 per minute
Open-circuit Voltage Front time: 1.2 µs ±30%
Short-circuit Current Front time: 8.0 µs ±20%
Duration* 20 µs ±20%
Undershoot 30%
Line Sync Accuracy ±10%, 50 - 277 VAC
Options Available
Model CM-3CD-16/32 16 or 32 Amp, 3 phase EFT and
Model CM-I/OCD External 8-line coupler/decoupler
Model CM-I/OCD-HS High speed I/OCD option for testing
* Durations are reduced in 12 mode and when coupling multiple lines to PE
UL 1449
250V - 6.0 kV ±5%, 2 mode
Duration: 50 µs ±20%
Undershoot: 30%
surge coupler/decoupler
for I/O signal lines
data rates to 100 kHz
1
Model PRO-TELECOM**
Surge Telecom compliant testing per IEC 61000-4-5, EN 61000-4-5,
FCC Part 68, ITU K.17, K.20, K.21 and ETSI
General Specifications
Voltage Waveform 10/700 µs (9/720 µs FCC Part 68)
Peak Voltage 250 V - 6.0 kV ±5%
Peak Current 6.25 - 150 A +10/-0%, 40 mode
Repetition Rate Up to 4 per minute
Open-circuit Voltage Front time: 7.0 µs to 11.7 µs
Duration: 576 µs to 840 µs
Short-circuit Current Front time: 3.5 µs to 6.5 µs
Duration: 256 µs to 384 µs
Options Available
Model CM-TELCD External coupler for telecom lines
** PRO-TELECOM and PRO-RING can not be installed in same unit.
Surge Waveform Monitoring
General Specifications
Lines Monitored Monitors are automatically switched
Open-circuit Voltage 1000:1 ±10%
Short-circuit Current Attenuation200:1 ±7%
to match generator coupling mode
Model PRO-HPOWER
Power Frequency Magnetic Field for compliant testing per IEC
61000-4-8 and EN 61000-4-8
General Specifications
Field Frequency 50 Hz/60 Hz
Field Amplitude 0.5 - 4 A/m, in 0.25 A steps, ±10%
(with CM-HCOIL) up to 100 A/m with
optional external HPOWER-EXT
AC Source Internal
Resolution 0.25 A minimum
Coil Factor 0.65 to 1.00
Coil Resistance 0.05 maximum
Options Available
Model CM-HMON Measurement probe for power
frequency magnetic fields
Model CM-HCOIL 1 m x 1 m magnetic field coil
Model HPOWER - EXT External generator for power
frequency magnetic field to 30 A/m