No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Manual Part Number
V2800-90002
Edition
Print Date:October 2019
Supersedes: August 2019
Printed in USA/Malaysia
Published by:
Keysight Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hard ware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
U.S. Government Rights
The Software is “commercial
computer software,” as defined
by Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202,
the U.S. government acquires
commercial computer software
under the same terms by which
the software is customarily
provided to the public.
Accordingly, Keysight provides
the Software to U.S. government
customers under its standard
commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of
which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA
represents the exclusive authority
by which the U.S. government
may use, modify, distribute, or
disclose the Software. The EULA
and the license set forth therein,
does not require or permit,
among other things, that
Keysight: (1) Furnish technical
information related to
commercial computer software
or commercial computer
software documentation that is
not customarily provided to the
public; or (2) Relinquish to, or
otherwise provide, the
government rights in excess of
these rights customarily provided
to the public to use, modify,
reproduce, release, perform,
display, or disclose commercial
computer software or
commercial computer software
documentation. No additional
government requirements
beyond those set forth in the
EULA shall apply, except to the
extent that those terms, rights, or
licenses are explicitly required
from all providers of commercial
computer software pursuant to
the FAR and the DFARS and are
set forth specifically in writing
elsewhere in the EULA. Keysight
shall be under no obligation to
update, revise or otherwise
modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant
to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S.
government acquires no greater
than Limited Rights as defined in
FAR 27.401 or DFAR 227.7103-5
(c), as applicable in any technical
data.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
V2802A Option M81
Description 4
Verifying the Shipment 5
PXI, VSA and VSG Requirements 6
Table of Contents
Available Options
Distribution unit
PXI System Interface Accessories 6
General Specifications 7
Power Requirements
Environmental Requirements 7
Environmental Tests 7
Equipment Heating and Cooling 8
Dimensions and Space Requirements
RF Specifications at 25 Degree Celsius
Front and Rear Panel Features 10
Hardware Lock-link Setup
Stacking the Distribution Unit Using Two PXI Chassis
VSA RF Interface Cable Connections 16
VSG RF Interface Cable Connections
Service Information 20
6
6
7
8
9
for a Three Instrument Stack 13
18
13
Trouble Shooting the Distribution unit 21
RF Path Checks 21
Equipment Needed
Equipment Setup
Check Results 22
Safety and Regulatory Information 24
Introduction 24
Safety Earth Ground 24
Declaration of Co
Statement of Compliance
Before Applying Power 25
Servicing 26
Connector Care and Cleaning Precautions 27
nformity
21
21
24
24
1
Table of Contents
Electrostatic Discharge Protection 27
Instrument Markings
Regulatory Information
Battery Collection 29
Electrical Safety Compliance 29
and EMC Compliance 29
EMI
Keysight Support, Services, and Assistance 30
Service and Support Options 30
Contacting Keysight 30
Shipping Your Product to Keysight for Service or Repair
28
29
30
2
V2802A Option M81
3
Description
V2802A Option M81
Description
The Keysight V2802A Option M81 is an eight channel LO Distribution Instrument. It is
designed for use with Keysight M9381A PXI VSG and M9391A PXI VSA product
family. It facilitates the time or phase synchronization of a multi-channel source or
signal analyzer measurement system by allowing them to share the same LO signal.
The Distribution unit has the following key features:
·Frequency Range of Operation: 400 MHz to 6 GHz.
•Maximum of eight instruments for system phase lock capability.
•3.5 mm RF Ports (female connectors).
•Output LO power levels are near unity gain.
•It has no automated or monitoring controls features; it is simply an active
pass-thru instrument.
The V2802A Option M81 may also be referred to as the Distribution Unit throughout
this document. MIM0 configuration designations are used (8 x 8) and refer to (number
of sources) x (number of receivers) in the system configuration.
4
V2802A Option M81
Description
Verifying the Shipment
To verify the contents shipped with your product, refer to the “Box Content List”
included with the shipment.
Inspect the shipping container. If the container or packing material is damaged, it
should be kept until the contents of the shipment have been checked mechanically
and electrically. If there is physical damage refer to “Contacting Keysight” on page 30.
Keep the damaged shipping materials (if any) for inspection by the carrier and a
Keysight Technologies representative.
Table 1
V2802A-M81 Accessories Supplied
DescriptionPart NumberQty
Termination, 50 Ohm, SMA female1810-01188
User’s and Service Guide V2802A-M81V2800-900021
RF Cable Interface Set for 8 x 0 VSG Source System: Figure 8 on page 19
RF Cable, Upper PXI LO OUT to Distribution Unit LO INV2800-200061
RF Cable, Upper PXI LO OUT to Out 1 to 4V2800-200084
RF Cable, Lower PXI LO OUT to Out 5 to 8V2800-200094
RF Cable Interface Set for 0 x 8 VSA Analyzer System: Figure 7 on page 17
RF Cable, Upper PXI LO OUT to Distribution Unit LO INV2800-200071
RF Cable, Upper PXI LO OUT to Out 1 to 4V2800-200104
RF Cable, Lower PXI LO OUT to Out 5 to 8V2800-200114
Hardware Lock-link Material: Figure 4 on page 13
Screws, T10 M3x0.5, 8 mm (Spacers) 0515-03726
Screws, T20 M4x0.7, 10 mm (PXI) 0515-038016
Screws, T20 M4x0.7, 18 mm (Distribution Unit) 0515-04364
Lock-link Bracket, StrutsV2800-000074
Lock-link Bracket, Front SpacerV2800-200122
Lock-link Bracket, Rear SpacerV2800-200132
For your convenience, the Hardware Lock-link material has already been
assembled.
5
V2802A Option M81
PXI, VSA and VSG Requirements
PXI, VSA and VSG Requirements
It is required that the PXI modules M9381A and M9391A have Option 012
(LO sharing for phase coherency) to operate with the Distribution unit in a phase
coherency configuration.
Available Options
Distribution unit
The Distribution Unit has these available options:
Instrument Options:
• Option M81 - Eight channel LO Distribution capability
Rack Mount Accessory Options:
•V2802A Option 1CM – Rack mount kit without handles
•V2802A Option 1CN – Front handle kit
•V2802A Option 1CP – Rack mount kit with handles
PXI System Interface Accessories
The Distribution Unit uses the following Lock-links, RF cables and material for your
VSA or VSG system configuration. RF cables can be used with either configuration;
however, the Lock-Link hardware material is for one of the two configurations only.
Table 2 Interface Lock-Link and Cable Sets
System
Configuration
VSG 8 x 0
VSA 0 x 8
a. Refer to “Hardware Lock-link Setup for a Three Instrument Stack” on page 13.
b. Refer to “VSA RF Interface Cable Connections” on page 16.
c. Refer to “VSG RF Interface Cable Connections” on page 18.
Hardware
Lock-link
V2800-00007,
V2800-20012 &
V2800-20013a
RF Cable Port Number – Conn. Type
V2800-20006, 20008, 20009 – SMA/SMA
V2800-20007, 20010, 20011 –SMA/SMA
b
c
6
V2802A Option M81
General Specifications
General Specifications
Specifications for the Distribution Unit (400 MHz to 6.0 GHz) are characteristic
performance. A functional test certificate is offered only for the Distribution Unit.
Power Requirements
Verify that the required ac power is available before installing the Distribution Unit.
•100/120 Vac
220/240 Vac 50/60 Hz
•40 Watts
•The instruments can operate with mains supply voltage fluctuations up to ± 10%
of the nominal voltage.
•Air conditioning equipment (or other motor–operated equipment) should not be
placed on the same ac line that powers the Distribution Unit.
This is a Safety Class I Prod uct (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted in a socket
outlet provided with a protective earth contact. Any interruption of the protective
conductor inside or outside of the product is likely to make the product dangerous.
Intentional interruption is prohibited.
Environmental Requirements
Refer to the standard documentation for environmental requirements.
Environmental Tests
The Distribution Unit complies with all applicable safety and regulatory requirements
for the intended location of use.
•The instrument can safely operate in a relative humidity of 80% for temperatures
to 31 degrees C, decreasing linearly to 50% relative humidity at 40 degrees C.
7
V2802A Option M81
General Specifications
Equipment Heating and Cooling
If necessary, install air conditioning and heating to maintain the ambient temperature
within the appropriate range.
Ventilation Requirements: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C for every 100 watts dissipated in the cabinet. If
the total power dissipated in the cabinet is greater than 800 watts, forced convection
must be used.
Dimensions and Space Requirements
Standard installation of the measurement instrumentation should include
configuration and installation on a customer provided lab bench or table top of
adequate size and strength. For weight, dimensions and space requirements refer to
the instrument documentation that is being used.
Table 3
Instrument Dimensions
ModelWeightHeightWid thDepth
V2802A-M81 5.5 kg (12 lb)
88 mm
(3.5 in)
425 mm
(16.75 in)
500 mm
(19.7 in)
8
V2802A Option M81
General Specifications
RF Specifications at 25 Degree Celsius
It is recommended that you do not operate components near damage levels. The
power levels must be 3 dB below maximum level to ensure no damage, see Table 4.
Table 4 Performance Information
Frequency Range and Power Levels:
Port Power Damage Level15 dBm
RF Frequency Range 400 MHz to 6 GHz
Nominal Performance
Gain> 0.5 dB
Channel to Channel Gain Variation< ±1.5 dB
LO IN Match< 10 dB
OUT (1-8) Match< 10 dB
1 dB Compression> +4.5 dBm
9
V2802A Option M81
Front and Rear Panel Features
Front and Rear Panel Features
Figure 1 V2802A-M81 Front Panel
Line Switch RF Connectors (x9)
The Distribution Unit has no automated control features; all signal path
configurations are performed manually with the available front panel coaxial
accessory interface cables
Line Switch
RF ConnectorsAll of the RF connectors are 50 Ohm, 3.5 mm female
The front panel LINE switch disconnects the mains circuits from the mains supply.
The switch is located at the bottom left corner of the front panel.
connectors.
10
V2802A Option M81
Front and Rear Panel Features
Figure 2 V2802A-M81 Rear Panel
Chassis Ground Line Module
Chassis Ground
Line Module
Power Cords
A threaded terminal post for connecting the Distribution Unit to a conductive object,
cabinet or structure to ensure a common potential and reduce leakage current in a
system. Requires a (M4 x 0.7) threaded nut.
The line fuse, as well as a spare, reside within the line module. Figure 3 illustrates
where the fuses are located and how to access them.
Install the instrument so that the ON/OFF switch is readily identifiable and is easily
reached by the operator. The ON/OFF switch is the instrument disconnecting device.
It disconnects the mains circuits from the mains supply before other parts of the
instrument, or the detachable power cord can be removed from the electrical supply.
Alternately, an externally installed switch or circuit breaker which is readily
identifiable and is easily reached by the operator may be used as a disconnecting
device.
A line power cord is supplied in one of several configurations, depending on the
destination of the original shipment. Keysight can supply additional certified power
cords to meet region electrical supply and receptacle configurations. Please contact
Keysight at: www.keysight.com for assistance in power cord selection.
This is a Safety Class I Prod uct (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted in a socket
outlet provided with a protective earth contact. Any interruption of the protective
conductor inside or outside of the product is likely to make the product dangerous.
Intentional interruption is prohibited.
11
V2802A Option M81
Front and Rear Panel Features
Available Fuse
Figure 3 Line Module
Fuse F 3 A/250V (2110-0780)
For continued protection against fire hazard replace line fuse only with same type
and rating. The use of other fuses or material is prohibited.
Verify that the premise electrical voltage supply is within the range specified on the
instrument.
12
V2802A Option M81
Hardware Lock-link Setup for a Three Instrument Stack
Hardware Lock-link Setup for a Three Instrument Stack
(Not needed if equipment is being installed in a system rack cabinet)
Stacking the Distribution Unit Using Two PXI Chassis
The Lock-link material provided will be used in these instructions to mechanically
secure the Distribution Unit to two M9018A PXI chassis.
1. The lock-link kit materials needed:
•0515-0372 - Screw T10 M3x0.5, 8 mm (x6), for Struts/Spacers
•0515-0380 - Screw T20 M4x0.7, 10 mm (x16), for PXI Chassis
•0515-0436 - Screw T20 M4x0.7, 18 mm (x4), for Dist. Unit
•V2800-00007 – Lock-link, Struts (x4)
•V2800-20012 - Lock-link, Front Spacer (x2)
•V2800-20013 - Lock-link, Rear Spacer (x2)
The lock-link struts and spacers have already been assembled for your convenience
using the 0515-0372 M3 screws, as shown in Figure 4. The top of each strut has also
been labeled to aid in the correct location.
Figure 4 Lock-Link Strut and Spacer Assemblies
13
V2802A Option M81
Hardware Lock-link Setup for a Three Instrument Stack
2. Remove the two rear feet from the bottom of the Distribution Unit.
3. Place the Distribution Unit on top of the first PXI chassis.
4. Using a T20 Torx driver, attached the two struts labeled “L-TOP FRONT” and
“R-TOP FRONT” to the front side frame of the Distribution Unit using two screws
(0515-0436). Torque to 9 in-lb.
5. Align the holes in the bottom PXI Chassis with the Front Strut holes and install
two M4 10 mm screws (0515-0380) using a T20 Torx driver on both the left and
right Front sides. If may be necessary to loosen the two M3 screws (0515-0372),
Use a T10 Torx driver, on the Front Spacer so the strut can move more freely to
align the holes. Leave the screws slightly loose for now.
Figure 5 Attaching the Distribution Unit to the Lower PXI Chassis
14
V2802A Option M81
Hardware Lock-link Setup for a Three Instrument Stack
6. Lift the rear of the Distribution Unit and place the two remaining struts labeled
L-TOP REAR and R-TOP REAR so that the attached rear spacer remains held in
place by the weight of the Distribution Unit.
a. Attached the struts to the lower PXI Chassis with two M4 10 mm, screws
(0515-0380) using a T20 Torx driver on the left and right sides. It may be
necessary to loosen the M3 screw (0515-0372) using a T10 Torx driver in the
spacer above to align the holes. Leave the screws slightly loose for now.
7. Place the second PXI Chassis on top of the Distribution Unit.
a. Align the holes in the top PXI Chassis with the Strut holes and install two M4
10 mm screws (0515-0380) using a T20 Torx driver on both the left and right
sides. It may be necessary to loosen the M3 screw (0515-0372) in the spacer
above using a T10 torx driver to align the holes. Leave the screws slightly
loose for now.
8. Once all the screws are installed into the two PXI Chassis, tighten to
9 in-lbs. Ensure the M3 screws (0515-0372), used to attach the Spacers to the
Strut, are tighten to 9 in-lb as well.
Figure 6 Attaching the Distribution Unit to the Lower PXI Chassis
15
V2802A Option M81
VSA RF Interface Cable Connections
VSA RF Interface Cable Connections
Figure 7 illustrates the system configuration (0x8) and cable connections.
1. Connect the RF interconnect cables from the PXI Chassis to the Distribution Unit.
As you are connecting each cable, torque to 8 in-lb. Refer to Table 5.
Over torque will cause damage and may cause connectors to spin or become loose.
Figure 8 illustrates the system configuration (8 x 0) and cable connections.
1. Connect the RF interconnect cables from the PXI Chassis to the Distribution Unit.
As you are connecting each cable, torque to 8 in-lb. Refer to Table 6.
Over torque will cause damage to the connectors to spin or become loose.
Table 6 VSG RF Interface Cable Connections
Keysight Part
Numbers
V2800-20006LO OUT - Slot 9LO IN
V2800-20008LO IN - Slot 3OUT 5
V2800-20008LO IN - Slot 7OUT 6
V2800-20008LO IN - Slot 12OUT 7
V2800-20008LO IN - Slot 16OUT 8
V2800-20009LO IN - Slot 3OUT 1
V2800-20009LO IN - Slot 7OUT 2
V2800-20009LO IN - Slot 12OUT 3
V2800-20009LO IN - Slot 16OUT 4
From:
PXI
To: Distribution
Unit
18
V2802A Option M81
I
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
PXi
1IN2
56
34
78
I
PXi
KEYSIGHT
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
V2800-20009 x4
(Long Cable Up)
V2800-20006
V2800-20008 x4
(Short Cable Down)
KEYSIGHT
KEYSIGHT
VSG System
VSG RF Interface Cable Connections
Figure 8
VSG Front Panel RF Cables
19
V2802A Option M81
V2802A-M81 LO Distribution Network
Amp
LO IN
OUT 1
3dB
Source or Receiver
Master
LO Out
0.4 - 6 GHz
OUT 2
OUT 3
OUT 4
OUT 5
OUT 6
OUT 7
OUT 8
LO IN
LO IN
LO IN
LO IN
LO IN
LO IN
LO IN
LO IN
Source or Receiver
Slaves
Service Information
Service Information
Refer to “Contacting Keysight” on page 30.
These servicing instructions are for use by qualified personnel only. To avoid
electrical shock, do not perform any servicing unless you are qualified to do so.
Table 7 Replacement Parts
DescriptionPart Number
AC/DC Power Supply, 4 Output Ports0950-5667
RF Amplifier, 26 dBm, SMA
Termination, 50 Ohm, SMA male
Fuse 3A 250V Non-Time delay2110-0780
Bulkhead Coax Adapter, 3.5 mm (f/f)1250-3805
Cable Assembly with fan, 12 V87050-60027
V2802A Option M81 Guide
Figure 9 RF Block Diagram
0955-2712
1810-0118
V2800-90002
20
V2802A Option M81
Port 1Port 2
PNA
V2802A-M81
Port 1Port 2
PNA
Attn2 (10 dB)
3.5 f/f
RF
Cables
Attn1 (10 dB)
OUT 8
LO IN
LOAD x7
Troubl
e Shooting the Distribution unit
Troubleshooting the Distribution Unit
RF Path Checks
If you suspect a RF signal path problem, the following procedure is provided to check
all RF signal paths through the Distribution unit.
Equipment Needed
•Two RF flex cables, 3.5 mm (m/f)
•Two coax attenuators, 3.5 mm, 10 dB
•One coax connector adapter, 3.5 mm (f/f)
•Seven 50 Ohm Terminations, 3.5 mm or SMA male
•Network Analyzer with 400 MHz to 6 GHz operating capability
Equipment Setup
1. Turn on the Distribution unit and the PNA (or equivalent).
2. Set the analyzers frequency range of 400 MHz to 6 GHz and IF BW to 1 kHz.
3. Connect two flexible RF cables to Ports 1 and 2 on the analyzer.
4. Configure the analyzer and make an S21 measurement.
5. Set the source power level to -5 dBm.
6. Using appropriate coax attenuators and adapters connect the two cable ends
together.
7. Perform an S21 Normalization.
Figure 10 Equipment Setup
Normalization
S21
Measurement
S21
21
V2802A Option M81
Amp
LO IN
OUT 1
3dB
Network
Analyzer
Port 1
0.4 - 6 GHz
OUT 2
OUT 3
OUT 4
OUT 5
OUT 6
OUT 7
OUT 8
Port 2
6dB
Neg.
Sloped
10 dB
10dB
Trouble Shooting the Distribution unit
Check Results
Connect the flexible RF cables with attenuators to the Distribution unit ports as
described Table 8. Connect a 50 Ohm termination to all unconnected OUT ports.
Refer to Figure 12, “Typical Path Response,” which illustrates a typical path response
for all eight signal paths.
Table 8 Test Set RF Signal Path (S21)
Path Description
LO IN to Out 8
LO IN to Out 7
LO IN to Out 6
LO IN to Out 5
LO IN to Out 4
LO IN to Out 3
LO IN to Out 2
LO IN to Out 1
Figure 11
RF Signal Path Check
V2802A-M81 LO Distribution Network
22
V2802A Option M81
Trouble Shooting the Distribution unit
Figure 12
Typical Path Response
23
V2802A Option M81
Safety and Regulatory Information
Safety and Regulatory Information
Introduction
Review this product and related documentation to familiarize yourself with safety
markings and instructions
This product has been designed and tested in accordance with accepted industry
standards, and has been supplied in a safe condition. The documentation contains
information and warnings that must be followed by the user to ensure safe operation
and to maintain the product in a safe condition.
Safety Earth Ground
This is a Safety Class I Prod uct (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted in a socket
outlet provided with a protective earth contact. Any interruption of the protective
conductor inside or outside of the product is likely to make the product dangerous.
Intentional interruption is prohibited.
before you operate the instrument.
Always use the three prong AC power cord supplied with this product. Failure to
ensure adequate earth grounding by not using this cord may cause product damage
and the risk of electrical shock.
Declaration of Conformity
A copy of the Declaration of Conformity is available upon request, or a copy is
available on the Keysight Technologies web site at
http://regulations.keysight.com/DoC.htm
Statement of Compliance
This product has been designed and tested in accordance with accepted industry
standards, and has been supplied in a safe condition. The documentation contains
information and warnings that must be followed by the user to
and to maintain the product in a safe condition.
ensure safe operation
24
V2802A Option M81
Safety and Regulatory Information
Before Applying Power
Verify that the premises electrical supply is within the range of the instrument. The
instrument has an autoranging power supply.
If this product is not used as specified, the protection provided by the equipment
could be impaired. This product must be used in a normal condition (in which all
means for protection are intact) only.
The Mains wiring and connectors shall
premise electrical system. Failure, to ensure adequate earth grounding by not using
the correct components may cause product damage,
This product is designed for use in Installation Category II and Pollution Degree 2.
Verify that the premise electrical voltage supply is within the range specified on the
instrument.
When installing the product in a cabinet, the convection into and out of the product
must not be restricted. The ambient temperature (outside the cabinet) must be less
than the maximum operating temperature of the product by 4 °C for every 100 watts
dissipated in the cabinet. If the total power dissipated in the cabinet is greater than
800 watts, then forced convection must be used.
be compatible with the connector used in the
and serious injury.
This instrument has auto-ranging line
within the specified range and voltage fluctuations do not to exceed 10 percent of
the nominal supply voltage.
25
voltage input, be sure the supply voltage is
Servicing
V2802A Option M81
Safety and Regulatory Information
These servicing instructions are for use by qualified personnel only. To avoid
electrical shock, do not perform any servicing unless you are qualified to do so.
The opening o
voltages. Disconnect the instrument from all voltage sources before opening.
No operator serviceable parts inside. Refer servicing to qualified personnel. To
prevent electrical shock, do not remove covers.
The detachable power cord is the instrument d isconnecting device. It d isconnects
the mains circuits from the mains supply before other parts of the instrument. The
front panel switch is only a standby switch and is not a LINE switch (disconnecting
device).
The power cord is connected to internal capacitors that may remain live for 5
seconds after d isconnecting the plug from its power supply.
f covers or removal of parts is likely to expose the user to dangerous
Remove the power cord to the instrument. To clean the connectors use alcohol in a
well ventilated area. Allow all residual alcohol moisture to evaporate, and fumes to
dissipate prior to energizing the instrument.
To prevent electrical shock, d isconnect the V2802A Option M81 from mains
electrical supply before cleaning. Use a dry cloth or one slightly dampened with
water to clean the external case parts. Do not attempt to clean internally.
If flammable cleaning materials are used, the material shall not be stored, or left
open in the area of the equipment. Adequate ventilation shall be assured to prevent
the combustion of fumes, or vapors.
Electrostatic Discharge Protection
Electrostatic discharge (ESD) can damage or destroy electronic components. The
product is shipped in materials that prevent damage from static, and should only be
removed from the packaging in an anti-static area ensuring that the correct
anti-static precautions are taken.
Two types of ESD protection are listed below. Purchase acceptable ESD accessories
from your local supplier.
•Conductive table-mat and wrist-strap combination
•Conductive floor-mat and heal-strap combination
Both types, when used together, provide a significant level of ESD protection. To
ensure user safety, static-safe accessories must provide at least 1 Meg Ohm of
isolation from ground.
These techniques for a static-safe work station should not be used when working on
circuitry with a voltage potential greater than 500 volts.
27
V2802A Option M81
Instrument Markings
Instrument Markings
This section contains information that is r
regulatory agencies.
The instruction documentation symbol. The product is marked with this symbol when it is
necessary for the user to refer to the instructions in the documentation.
The AC symbol indicates the required nature of the line module input power.
This symbol indicates separate collection for electrical and electronic equipment,
mandated under EU law. All electric and
from normal waste for disposal (Reference WEEE Directive).
This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is in the STANDBY position.
This symbol indicates that the power line switch is in the OFF position.
This symbol is used to identify a terminal which is internally connected to the product frame or
chassis.
The CE mark is a registered trademark of the European Community.
ccr.keysight@keysight.com
The Keysight email address is required by EU directives applicable to our product.
equired by various government
electronic equipment are required to be separated
ICES/NMB-001
IP 2 0
The CSA mark is a registered trademark of the CSA International.
This is a symbol of an
Clause 5).
This is a marking to indicate product compliance with the Canadian Interference-Causing
Equipment Standard (ICES-001). Cet appareil ISM est conforme à la norme NMB du Canada.
Direct Current.
The instrument has been designed to meet the requirements of IP 2 0 for egress and
operational environment.
The RCM mark is a registered trademark of the
Authority.
Indicates the time period during which no hazardous or toxic substance elements are
expected to leak or deteriorate during normal use. Forty years is the expected useful life of
the product.
This symbol on all primary and secondary packaging indicates compliance to China standard
GB 18455-2001.
South Korean Certification (KC) mark; includes the marking's identifier code which follows the
format: MSIP-REM-YYY-ZZZZZZZZZZZZZZ.
Industrial Scientific and Medical Group 1 Class A product (CISPR 11,
Australian Communications and Media
28
V2802A Option M81
Regulatory Information
Regulatory Information
Battery Collection
Do not throw batteries away but collect as small chemical waste, or in accordance
with your country’s requirements. You may return the battery to Keysight
Technologies for disposal. Refer to “Contacting Keysight” on page 30 for assistance.
Safety
This instrument complies with the essential requirements of the European Low
Voltage Directive.
Acoustic Statement (European Machinery Directive)
Acoustic noise emission
LpA<70 dB
Operator position
Normal operation mode per ISO 7779
EMC Compliance
This instrument complies with the essential requirements of the European EMC Directive as
well as current editions of the following standards (dates and editions are cited in the
Declarations of Conformity at https://regulations.about.keysight.com/DoC/default.htm):
•IEC/EN 61326-1
•CISPR Pub 11 Group 1, Class A
•AS/NZS CISPR 11
•ICE
S/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada.
South Korean Class A EMC Declaration
Information to the user: If this
been conformity assessed for use in
environment, this equipment may cause radio interference.
※ This EMC statement applies to the equipment only for use in a business
environment.
이 기기는 업무용 환경에서 사용할 목적으로 적합성평가를 받은 기기로서
가정용 환경에서 사용하는 경우 전파간섭의 우려가 있습니다.
instrument
사 용자안내문
is
equipped
with a "KC" label, it has
business environments. In a residential
※ 사용자안내문은 “업무용방송통신기자재”에만적용한다.
29
V2802A Option M81
Keysight Support, Services, and Assistance
Keysight Support, Services, and Assistance
Service and Support Options
There are many other repair and calibration options available from the Keysight
Technologies support organization. Thes
agreements with varying response times. Contact Keysight for additional information
on available service agreements for this product.
Contacting Keysight
Assistance with test and measurements needs and information or finding a local
Keysight office are available at: http://www.keysight.com/find/assist
If you do not have access to the Internet, contact your field engineer.
In any correspondence or telephone conversation, refer to the Keysight product by its
model number and full serial number. With this information, the Keysight
representative can determine the warranty status of your unit.
e options cover a range of service
Shipping Your Product to Keysight for Service or Repair
IMPORTANTKeysight Technologies reserves the right to reformat or replace the internal hard disk
drive in your analyzer as part of its repair. This will erase all user information stored
on the hard disk. It is imperative, therefore, that you make a backup copy of your
critical test data located on the analyzer’s hard disk before shipping it to Keysight for
repair.
If you wish to send your instrument to Keysight Technologies for service or repair:
•To improve turn-around time, return your Distribution unit along with your
analyzer and cables to Keysight so that we may verify the operation of the
complete system.
•Include a complete description of the service requested or of the failure and a
description of any failed test and any error message.
•Remove and retain the front handles and all rack mount hardware. The analyzer
should be sent to Keysight in the same configuration as it was originally shipped.
•Ship the analyzer using the original or comparable antistatic packaging materials.
•Contact Keysight for instructions on where to ship your analyzer.
30
V2802A Option M81
Keysight Support, Services, and Assistance