PC Requirements for USB VNA control ............................................................................ 25
Literature Information ........................................................................................................ 25
Web Resources ................................................................................................................. 25
Find us at www.keysight.com Page 2
P9370A
300 kHz to 4.5 GHz, 2-port
P9371A
300 kHz to 6.5 GHz, 2-port
P9372A
300 kHz to 9 GHz, 2-port
P9373A
300 kHz to 14 GHz, 2-port
P9374A
300 kHz to 20 GHz, 2-port
P9375A
300 kHz to 26.5 GHz, 2-port
P5000A
9 kHz to 4.5 GHz, 2-port
P5001A
9 kHz to 6.5 GHz, 2-port
P5002A
9 kHz to 9 GHz, 2-port
P5003A
9 kHz to 14 GHz, 2-port
P5004A
9 kHz to 20 GHz, 2-port
Keysight Streamline Series: Exceptional Performance in a Small Package
Balance deadlines, productivity, budget and bench space with the Streamline Series. You’ll move
confidently across every stage of your product’s development lifecycle by leveraging accurate and
repeatable measurements, automated code capability, and a consistent, intuitive user experience. With
comprehensive Keysight Services including calibration, education and consulting, these instruments
enhance your solution to help you accelerate technology adoption and lower costs.
This configuration guide describes standard configurations, options, accessories, upgrade kits and
compatible peripherals for the Keysight Streamline Series USB vector network analyzer (VNA).
P937xA Series
P500xA Series
.
Find us at www.keysight.com Page 3
Keysight Streamline Series Vector Network Analyzer Configurations
P937xA Series
P9370A
300 kHz to 4.5 GHz, 2-port
3.5 mm (f)
P9371A
300 kHz to 6.5 GHz, 2-port
3.5 mm (f)
P9372A
300 kHz to 9 GHz, 2-port
3.5 mm (f)
P9373A
300 kHz to 14 GHz, 2-port
3.5 mm (f)
P9374A
300 kHz to 20 GHz, 2-port
3.5 mm (f)
P9375A
300 kHz to 26.5 GHz, 2-port
3.5 mm (f)
P500xA Series
P5000A
9 kHz to 4.5 GHz, 2-port
3.5 mm (f)
P5001A
9 kHz to 6.5 GHz, 2-port
3.5 mm (f)
P5002A
9 kHz to 9 GHz, 2-port
3.5 mm (f)
P5003A
9 kHz to 14 GHz, 2-port
3.5 mm (f)
P5004A
9 kHz to 20 GHz, 2-port
3.5 mm (f)
P500xA Series
Option 200
Not available for the P937xA Series
This option enables the internal pulse modulator
measurements.
Add spectrum analysis
hardware
S97090A application software is required for
spectrum analysis.
Model Description Test port connectors
Test set options
Choose one of the frequency models, and one test set option for the P500xA Series. Option 2xx indicates
two test ports. To add options to a product, order the corresponding item number (P500xA-xxx).
Description 2-port
Hardware options
Description
Add pulse modulation
hardware
Additional information
・
For P937xA
Series
N/A P500xA-021
N/A P500xA-090
For P500xA
Series
Additional information
on the analyzer’s source. S97025A application
software is required for pulsed-RF
Find us at www.keysight.com Page 4
Application software1
Automatic fixture removal
S97007A
S97007A
Time domain analysis
S97010A
S97010A
Basic pulsed-RF
measurements
Requires hardware option 021.
Noise figure measurements
with vector correction
Standard receivers are used.
Provides ability to independently set
class.
Embedded-LO capability
N/A
S97084A
Gain-compression
measurements
Spectrum analysis
N/A
S97090A
Requires hardware option 090.
Provides multiport calibrated
instruments.
ISO 17025 compliant
calibration
ANSI Z540 compliant
calibration
Commercial calibration
certificate with test data
Description
Scalar mixer/converter
measurements
Multiport calibrated
measurements
Calibration options
For P937xA
Series
For P500xA
Series
N/A S97025A
N/A S97029A
S97082A S97082A
N/A S97086A
S97551A S97551A
Additional information
the frequency of internal sources and
receivers, and to configure external
sources. Provides SMC measurement
measurements using up to two
2,3,4
Description
1. Supported software license types: node-locked perpetual (1FP), node-locked 12 month (1FL). 1FP is the only
type that can be ordered as part of the instrument. The other types have to be ordered as separate items and
installed from the web aster the receipt of the instruments.
2. When configuring a multiport VNA using multiple USB VNAs, the Y1710A is recommended for each additional
multiport interconnection.
3. The P937xA and P500xA can not be combined in multiport configurations.
4. A 4-port ECal module (i.e. N4431/32/33) is recommended for multiport measurements (n > 4).
Find us at www.keysight.com Page 5
For P937xA
Series
For P500xA
Series
P937xA-1A7 P500xA-1A7
P937xA-A6J P500xA-A6J
P937xA-UK6 N/A
Additional information
Accessories
Includes the tools for SMA and SMB
connector removal
Order Option 001 for 1-slot USB
instruments.
Includes interconnect cables for
additional USB instrument.
Includes latch kit for connecting two
USB instruments.
Hard transit case
Y1710A
Y1710A
Includes PXI module to monitor
2 flexible control cables.
Description
Accessory and tool kit Y1281A Y1281A
Rack mount kit Y1700A-001 Y1700A-001
Multiple USB instruments
configuration kit
PXI adapter module N/A Y1731A-001
For P937xA
Series
For P500xA
Series
Y1701A-001 Y1701A-002
Y1701A-101 Y1701A-101
Additional information
configurations using two USB
instruments. Add one kit for each
control signals with the P500xA and
Find us at www.keysight.com Page 6
Measurement Application Software
Solid black series name indicates the feature is available on that series, while gray series name with
strikethrough indicates the feature ins not available on that series. For example:
P500xA Series: Available on P500xA series
P937xA Series: Not available on P937xA Series
Automatic fixture removal (S97007A)
P937xA Series P500xA Series
Many devices do not have coaxial connectors and are put in fixtures in order to measure them in a
coaxial environment. Accurately removing the effects of the fixture is required to get a good measurement
of the device under test (DUT). This application adds a powerful application wizard to guide you through
characterizing a fixture and removing it from the measurement. Devices can be single-ended or
differential. Files can be saved in a variety of formats for later use in ENA, PNA, PXI VNA, USB VNA,
ADS, and PLTS.
Time domain analysis (S97010A)
P937xA Series P500xA Series
This application enables the analyzer to view reflection and transmission responses in time or distance.
Use time domain to tune filters, gate out the response of fixtures and cables, characterize the impedance
of transmission lines and more.
Basic Pulsed-RF measurements (S97025A)
P937xA Series P500xA Series
This application enables internal pulse generators that can be used to control the internal pulse
modulators, and it provides an integrated pulse application that uses the wideband-detection method. The
software requires hardware option 021 pulse modulator hardware.
The pulse application provides an easy way to set up point-in-pulse measurements with pulse width as
narrow as 1 us, and pulse-profile measurements with 40 ns minimum timing resolution. Using the built-in
pulse modulators, the USB VNAs are complete pulsed-RF measurement solutions, eliminating the need
for external test sets and pulse generators. S97025A also controls external pulse generators and
modulators and can synchronize to external master pulses. The Y1731A PXI Adapter Module is
recommended if using external master pulses or external pulse modulators.
Find us at www.keysight.com Page 7
Noise figure measurements with vector correction (S97029A)
P937xA Series P500xA Series
This software application enables high-accuracy noise figure and noise-power measurements of
amplifiers utilizing Keysight’s unique vector-source-correction technique that uses a source-impedance
tuner to remove the effects of imperfect system-source match. This approach yields accuracy that
surpasses that provided by the Y-factor method and other cold-source implementations, especially for infixture, on-wafer, and automated-test environments. A scalar-calibrated method is also available that
offers less accuracy but is faster and does not require an impedance tuner.
The instrument’s standard receivers are used for noise figure measurements with the S97029A. An
external preamplifier and filter(s) is required for devices with < 30 dB of excess noise (gain plus noise
figure in dB). S97029A controls Keysight N469x Series
1
ECal modules configured as impedance tuners
for use with the USB VNA.
For calibration, a standard mechanical cal kit or ECal module is required for the S-parameter portion of
the cal (an ECal used as a tuner cannot be shared for calibration). To calibrate a standard receiver for
noise figure measurements, a power meter is required. All calibration accessories must be ordered
separately.
Scalar mixer/converter measurements (S97082A)
P937xA Series P500xA Series
With a simple setup and calibration, this application delivers the highest accuracy for scalar conversionloss/gain measurements by combining one-port and power-meter calibrations to remove mismatch errors.
S97082A provides an intuitive and easy-to-use user interface for setting up mixer and converter
measurements, with single or dual conversion stages. It can control external signal generators for use as
LO signals. Supported external sources include the Keysight ESG, PSG, EXG, and MXG Series, as well
as other SCPI-controlled signal generators.
S97082A is compatible with S97084A, which enables measurements of converters with internal LOs.
Embedded LO capability (S97084A)
P937xA Series P500xA Series
This application tunes the analyzer’s receivers to the output frequency of the converter under test without
the need for access to internal LOs or a common reference signal. S97084A is intended to work with
S97082A measurement applications.
1. 8509x, N443x or N755xA Series ECal modules are not supported.
Find us at www.keysight.com Page 8
Gain compression measurements (S97086A)
P937xA Series P500xA Series
The gain compression application (GCA) provides input power, output power, gain, and phase at the
compression point of an amplifier over a specified frequency range. GCA’s SMART Sweep is very fast
and easy-to-use. GCA also includes a guided calibration that corrects for absolute power levels,
frequency response, and mismatch errors.
Spectrum analysis (S97090A)
P937xA Series P500xA Series
The spectrum analyzer (SA) application adds high-performance microwave spectrum analysis to USB
VNAs. With fast stepped-FFT sweeps resulting from optimized data processing, the SA application
provides quick spurious searches over broad frequency ranges. Simultaneous spectrum measurements
can be done using test and reference receivers. This multi-channel SA can be used with the internal
swept-signal generators for efficient measurements of spurious signals emanating from mixers and
frequency converters. The SA application employs source-power and receiver-response calibration as
well as fixture de-embedding, providing in-fixture and on-wafer spectrum measurements with the highest
level of accuracy. Optional external attenuators should be connected with the VNA’s test ports to avoid
receiver compression when measuring large signals.
The S97090A software requires option 090 spectrum analysis hardware.
Multiport calibrated measurements (S97551A)
P937xA Series P500xA Series
Keysight USB VNA offers a capability to extend the number of ports for your measurements by using up
to two instruments. When multiple instruments are used, they are configured for multiport operation. The
instruments may be identified by the firmware as one VNA by an external PC. At least one VNA
instrument connected to your PC must have one S97551A license to maintain N-port capabilities with
multiple instruments. The frequency of the multiport array is determined by the lowest frequency
instrument configured in the array. For example, a 4-port analyzer configuration using a P5000A (4.5
GHz) and a P5004A (20 GHz) would have a maximum frequency of 4.5 GHz when performing 4-port
measurements. This behavior extends to other measurement application software (S97xxxA). In a
multiport configuration, only one instrument must have a valid license for these capabilities to function in
multiport mode.
Each instrument is connected into the array with Keysight cables. A Y1701A Multiple USB instruments
configuration kit should be ordered for connection of two P937xA (with Y1701A-001) or two P500xA
instruments (with Y1701A-002). A single Y1281A accessory and tool kit should be ordered for easier
cable connections.
The P937xA and P500xA can not be combined to configure multiport VNAs due to hardware difference.
Find us at www.keysight.com Page 9
Figure 1: 4-port configuration using two P937xA VNAs
Material measurements
P937xA Series P500xA Series
The Keysight N1500A materials measurement suite streamlines the process of measuring complex
permittivity and permeability with a vector network analyzer. Various type of measurements, such as
transmission line and free space, arch reflectivity, resonant cavity, and coaxial probe are available as
options. The easy-to-use software guides the user through setup and measurement, instantly converting
S-parameter network analyzer data into the data format of your choice and displaying the results within
seconds. Results can be charted in a variety of formats: ε
’, εr”, tand, μr’, μr”, tandμ, and Cole-Cole.
r
N1930B Physical Layer Test System (PLTS) Software
P937xA Series P500xA Series
The PLTS software platform has become an industry standard for calibration, measurement, and analysis
of linear passive interconnects such as cables, connectors, backplanes, and printed circuit boards.
Utilizing either a vector network analyzer (VNA) or a time domain reflectometer (TDR), fast and accurate
measurements can be obtained without in-depth knowledge of microwave measurement techniques.
Refer to www.keysight.com/find/plts or the technical overview (literature number 5989-6841EN) for more
details.
Find us at www.keysight.com Page 10
Upgrading Your System
P9370AU-
N/A
N/A
N/A
N/A
N/A
No
P9370AU-
P9371AU-
N/A
N/A
N/A
N/A
No
P9370AU-
P9371AU-
P9372AU-
N/A
N/A
N/A
No
P9370AU-
P9371AU-
P9372AU-
P9373AU-
N/A
N/A
No
P9370AU-
P9371AU-
P9372AU-
P9373AU-
P9374AU-
N/A
No
Upgrade kits are available to add options after initial purchase. To upgrade the USB VNAs, order the
corresponding item number. To add application software, order the appropriate standalone mode
numbers (S97xxxA).
A complete list of RF and microwave test accessories is available on our Web site:
www.keysight.com/find/mta
Accessories are available in these connector types: 50 Ω Type-N, 3.5 mm, 7 mm, 2.4 mm, 2.92 mm, 1.85
mm, 1.0 mm, and waveguide. Test port cables and a calibration kit should be added for a complete
measurement system. A verification kit is used to verify corrected system performance.
Cable and adapter sets
Keysight offers cables in the following types:
Single cables in semi-rigid and flexible
•
Cable sets in semi-rigid and flexible
•
There are also adapter sets available that protect the test port and convert the port to the desired
connector interface. These kits contain:
One male adapter
•
One female adapter
•
To attain the best mechanical rigidity for device connection, use a single cable and the appropriate
special adapter set. To attain the greatest flexibility for device connection, use a cable set.
Find us at www.keysight.com Page 12
Calibration kits
Coaxial measurements
Mechanical calibration kits include standards, such as opens, shorts and loads, which are measured by
the network analyzer for increased measurement accuracy.
Choose a calibration kit for each connector type to be used.
Economy, includes:
Open standards (male and female)
•
Short standards (male and female)
•
Fixed-termination standards (male and female)
•
Standard, includes the devices in the economy kit and adds:
Sliding load standards (male and female) or a series of offset shorts
•
Precision, includes the devices in the economy kit and adds:
50 Ω airline(s) for TRL calibration
•
TRL adapters
•
Electronic calibration (ECal) kits replace mechanical calibration standards with one solid-state calibration
module that is controlled by the network analyzer via USB, to present many different impedances to the
test ports. A full two-port calibration can be performed quickly with a single connection. This technique
reduces operator errors and connector wear and abrasion.
For more information about ECal modules, refer to the technical overview 5963-3743E (N4690, 85090,
N4430 and N7550 Series).
Waveguide measurements
For waveguide measurements, Keysight offers mechanical calibration kits that include:
Waveguide-to-coax adapters (X, P, K, R, Q, U, V)
•
Precision waveguide section
•
Flush short circuit
•
Fixed terminations
•
Straight section
•
Find us at www.keysight.com Page 13
For devices with 3.5 mm or SMA connectors
Mechanical calibration kits
85033E economy: DC to 9 GHz. Includes:
85033-60016 3.5 mm (m) load
•
85033-60017 3.5 mm (f) load
•
85033-60018 3.5 mm (m) open
•
85033-60019 3.5 mm (f) open
•
85033-60020 3.5 mm (m) short
•
85033-60021 3.5 mm (f) short
•
8710-1761 torque wrench
•
Option 85033E-100 adds:
85027-60005 3.5 mm (f) to 3.5 mm (f) adapter
•
Option 85033E-200 adds:
85027-60007 3.5 mm (m) to 3.5 mm (m) adapter
•
Option 85033E-300 adds:
85027-60006 3.5 mm (m) to 3.5 mm (f) adapter
•
Option 85033E-400 adds:
1250-1744 3.5 mm (f) to 50 Ω Type-N (m) adapter
•
1250-1743 3.5 mm (m) to 50 Ω Type-N (m) adapter
•
1250-1745 3.5 mm (f) to 50 Ω Type-N (f) adapter
•
1250-1750 3.5 mm (m) to 50 Ω Type-N (f) adapter
•
Option 85033E-500 adds:
1250-1746 3.5 mm (m) to 7 mm adapter (two included)
•
1250-1747 3.5 mm (f) to 7 mm adapter (two included)
•
85052B standard: DC to 26.5 GHz. Includes:
00902-60003 3.5 mm (m) fixed load
•
00902-60004 3.5 mm (f) fixed load
•
00911-60019 3.5 mm (m) sliding load
•
00911-60020 3.5 mm (f) sliding load
•
85052-60006 3.5 mm (m) short
•
85052-60007 3.5 mm (f) short
•
85052-60008 3.5 mm (m) open
•
85052-60009 3.5 mm (f) open
•
85052-60012 3.5 mm (f) to 3.5 mm (f) adapter
•
85052-60013 3.5 mm (f) to 3.5 mm (m) adapter
•
85052-60014 3.5 mm (m) to 3.5 mm (m) adapter
•
Find us at www.keysight.com Page 14
85052C precision TRL: DC to 26.5 GHz. Includes:
00902-60003 3.5 mm (m) fixed load
•
00902-60004 3.5 mm (f) fixed load
•
85052-60006 3.5 mm (m) short
•
85052-60007 3.5 mm (f) short
•
85052-60008 3.5 mm (m) open
•
85052-60009 3.5 mm (f) open
•
85052-60032 3.5 mm (f) to 3.5 mm (f) adapter
•
85052-60033 3.5 mm (m) to 3.5 mm (m) adapter
•
85052-60034 3.5 mm (f) to 3.5 mm (m) adapter
•
85052-60035 3.5 mm short TRL line
•
85052-60036 3.5 mm long TRL line
•
85052D economy: DC to 26.5 GHz. Includes:
00902-60003 3.5 mm (m) fixed load
•
00902-60004 3.5 mm (f) fixed load
•
85052-60006 3.5 mm (m) short
•
85052-60007 3.5 mm (f) short
•
85052-60008 3.5 mm (m) open
•
85052-60009 3.5 mm (f) open
•
85052-60012 3.5 mm (f) to 3.5 mm (f) adapter
•
85052-60013 3.5 mm (f) to 3.5 mm (m) adapter
•
85052-60014 3.5 mm (m) to 3.5 mm (m) adapter
•
Electronic calibration (ECal) modules
85093C RF ECal: 300 kHz to 9 GHz, 2-ports. Standard module includes:
Option 00F: Both 3.5 mm connectors are female
•
Option 00M: Both 3.5 mm connectors are male
•
Option M0F: One female and one male connector, both 3.5 mm
•
Option 00A adds:
•
o
85052-60012 3.5 mm (f) to 3.5 mm (f) adapter
o
85052-60014 3.5 mm (m) to 3.5 mm (m) adapter
Option 150: Replaces standard storage container with a wooden box
•
Find us at www.keysight.com Page 15
85093C mixed-connector options
3.5 mm
101
102
Type-N 50 Ω
203
204
7-16
205
206
Four 3.5 mm (f)
010
Four Type-N 50 Ω (f)
020
3.5 mm (f)
101
201
301
401
3.5 mm (m)
102
202
302
402
Type-N 50 Ω (f)
103
203
303
403
Type-N 50 Ω (m)
104
204
304
404
7-16 (f)
105
205
305
405
7-16 (m)
106
206
306
406
Four 3.5 mm (f)
010
3.5 mm (f)
101
201
301
401
3.5 mm (m)
102
202
302
402
Port A Option Port B Option
Connector type Female Male Connector type Female Male
N4431B ECal: 300 kHz to 13.5 GHz, 4-ports.
Connector Type Port A Option Port B Option Port C Option Port D Option
Option 150: Replaces standard storage container with a wooden box
N4433A ECal: 300 kHz to 20 GHz, 4-ports.
Connector Type Port A Option Port B Option Port C Option Port D Option
Option 150: Replaces standard storage container with a wooden box
N4691D Microwave ECal, DC or 300 kHz to 26.5 GHz, 2-ports. Includes:
Option F0F: Both 3.5 mm connectors are female
•
Option M0F: One female and one male connector, both 3.5 mm
•
Option M0M: Both 3.5 mm connectors are male
•
Option 0DC: DC to 26.5 GHz
•
Option 003: 300 kHz to 26.5 GHz
•
Option 00A adds:
•
o
85052-60012 3.5 mm (f) to 3.5 mm (f) adapter
o
85052-60014 3.5 mm (m) to 3.5 mm (m) adapter
Find us at www.keysight.com Page 16
N755XA Series Economy ECal, 2-ports:
N7550A DC to 4 GHz, 2-ports
•
N7551A DC to 6.5 GHz, 2-ports
•
N7552A DC to 9 GHz, 2-ports
•
N7553A DC to 14 GHz, 2-ports
•
N7554A DC to 18 GHz, 2-ports
•
N7555A DC to 26.5 GHz, 2-ports
•
N755xA Series includes:
Option 3FF: Both 3.5 mm connectors are female
•
Option 3MF: One female and one male connector, both 3.5 mm
•
Option 3MM: Both 3.5 mm connectors are male
•
Option 150: Plastic storage box
•
N7550X-151: 3.5 mm or 2.92 mm torque wrench
•
Cables
1
85131C
3.5 mm (f) to PSC-3.5 mm (f), 81 cm, 32 inches
•
85131D
single, semi-rigid:
1
set, semi-rigid:
85131-60009 One 3.5 mm (f) to 3.5 mm (m), 53 cm, 21 inches
•
85131-60010 One 3.5 mm (f) to PSC-3.5 mm (f), 53 cm, 21 inches
•
85131E
3.5 mm (f) to PSC-3.5 mm (f), 96.5 cm, 38 inches
•
85131F
85131-60012 One 3.5 mm (f) to 3.5 mm (m), 62.2 cm, 24.5 inches
•
85131-60013 One 3.5 mm (f) to PSC-3.5 mm (f), 62.2 cm, 24.5 inches
•
85131H
3.5 mm (f) to 3.5 mm (m), 62.2 cm, 24.5 inches
•
85134E
PSC-3.5 mm (f) to 2.4 mm (f), 96 cm, 38 inches
•
85134F
85134-60004 One 2.4 mm (f) to PSC-3.5 mm (f), 53 cm, 21 inches
•
85134-60003 One 2.4 mm (f) to PSC-3.5 mm (m), 53 cm, 21 inches
•
1
single, flexible:
1
set, flexible:
1
single, flexible:
1
single, flexible:
1
set, flexible:
1. Special rugged female connector specifically for connecting to the network analyzer test port, but does not
mate with a standard male connector.
Find us at www.keysight.com Page 17
85134H1 single, flexible:
2.4 mm (f) to PSC-3.5 mm (m), 53 cm, 21 inches
•
N4419AK20 single, flexible:
3.5 mm (m) to 3.5 mm (f), 91.4 cm, 36 inches
•
Z5623A-K20 single, flexible:
3.5 mm (m) to 3.5 mm (m), 91.4 cm, 36 inches
•
Adapter sets
85130D 3.5 mm
•
1
to 3.5 mm
For devices with Type-N connectors
Mechanical calibration kits
85032F standard, DC to 9 GHz Includes:
85032-60017 Type-N (m) fixed load
•
85032-60018 Type-N (f) fixed load
•
85032-60013 Type-N (m) open
•
85032-60014 Type-N (f) open
•
85032-60016 Type-N (m) short
•
85032-60015 Type-N (f) short
•
Option 85032F-100 adds:
85032-60021 Type-N (f) to Type-N (f) adapter
•
Option 85032F-200 adds:
85032-60019 Type-N (m) to Type-N (m) adapter
•
Option 85032F-300 adds:
85032-60020 Type-N (m) to Type-N (f) adapter
•
Option 85032F-500 adds:
85054-60001 Type-N (f) to 7 mm adapter (two included)
•
85054-60009 Type-N (m) to 7 mm adapter (two included)
•
1. Special rugged female connector specifically for connecting to the network analyzer test port, but does not
mate with a standard male connector.
Option M0F: one female and one male connector, both Type-N
•
Option 00A adds:
•
o
85054-60037 Type-N (f) to Type-N (f) adapter
o
85054-60038 Type-N (m) to Type-N (m) adapter
N4432A ECal: 300 kHz to 18 GHz, 4-ports.
Connector Type
Port A Option Port B Option
Port C Option
Option 150: Replaces standard storage container with a wooden box
N4690D Microwave ECal, DC or 300 kHz to 18 GHz, 2-ports. Includes:
Option F0F: Both Type-N connectors are female
•
Option M0F: One female and one male connector, both Type-N connectors
•
Option M0M: Both Type-N connectors are male
•
Option 0DC: DC to 18 GHz
•
Option 003: 300 kHz to 18 GHz
•
Option 00A adds:
•
o
85054-60037 Type-N (f) to Type-N (f) adapter
o
85054-60038 Type-N (m) to Type-N (m) adapter
Port D Option
Find us at www.keysight.com Page 20
N755XA Series Economy ECal, 2-ports:
N7550A DC to 4 GHz, 2-ports
•
N7551A DC to 6.5 GHz, 2-ports
•
N7552A DC to 9 GHz, 2-ports
•
N7553A DC to 14 GHz, 2-ports
•
N7554A DC to 18 GHz, 2-ports
•
N755xA Series includes:
Option NFF: Both Type-N connectors are female
•
Option NMF: One female and one male connector, both Type-N
•
Option NMM: Both Type-N connectors are male
•
Option 150: Plastic storage box
•
N7550X-152: Type-N torque wrench
•
Cables
N6314A 50 Ω Type-N RF cable, DC to 12.4 GHz Includes:
8120-8862 one 610 mm (24 in) cable with male connectors
•
N6315A 50 Ω Type-N RF cable, DC to 12.4 GHz Includes:
8121-0027 one 610 mm (24 in) cable with both female and male connectors
•
Adapter sets
11853A 50 Ω Type-N accessory kit. Includes:
1250-1472 Type-N (f) to Type-N (f) adapter (two included)
•
1250-1475 Type-N (m) to Type-N (m) adapter (two included)
•
11511A Type-N (f) short
•
11512A Type-N (m) short
•
11878A Type-N to 3.5 mm adapter kit. Includes:
1250-1744 3.5 mm (f) to 50 Ω Type-N (m) adapter
•
1250-1743 3.5 mm (m) to 50 Ω Type-N (m) adapter
•
1250-1745 3.5 mm (f) to 50 Ω Type-N (f) adapter
•
-1250-1750 3.5 mm (m) to 50 Ω Type-N (f) adapter
•
11524A 7 mm to Type-N (f) adapter
11525A 7 mm to Type-N (m) adapter
85130C 3.5 mm to Type-N Includes:
85054-60029 3.5 mm to Type-N (f)
•
85054-60030 3.5 mm to Type-N (m)
•
Find us at www.keysight.com Page 21
For devices with 7 mm connectors
Mechanical calibration kits
85050B standard: DC to 18 GHz. Includes:
00909-60008 7 mm coax termination
•
85050-60006 7 mm fixed broadband load
•
85050-80007 7 mm short
•
85050-80010 7 mm open
•
85050-80011 7 mm sliding load
•
85050C precision TRL: DC to 18 GHz. Includes:
00909-60008 7 mm coax termination
•
85050-60003 7 mm to 7 mm airline
•
85050-60005 7 mm to 7 mm TRL adapter
•
85050-60006 7 mm fixed broadband load
•
85050-80008 7 mm short
•
85050-80009 7 mm short with collect
•
85050-80010 7 mm open
•
85050D economy: DC to 18 GHz. Includes:
85050-60006 7 mm fixed broadband load
•
85050-80007 7 mm short
•
85050-80010 7 mm open
•
Electronic calibration (ECal) modules
N4696D Microwave ECal: DC or 300 kHz to 18 GHz, 2-ports, 7 mm connectors. Includes:
Option 0DC: DC to 18 GHz
•
Option 003: 300 kHz to 18 GHz
•
For devices with waveguide
Mechanical calibration kits
X-Band
X11644A standard, WR-90: 8.2 to 12.4 GHz. Includes:
00896-60008 X-band standard section
•
00910-60003 X-band termination
•
11644-20018 X-band short
•
11644-20021 X-band shim
•
X281C adapter (included in calibration kit): WR-90 to 7 mm
•
Find us at www.keysight.com Page 22
P Band
P11644A standard, WR-62: 12.4 to 18 GHz. Includes:
00896-60007 P-band standard section
•
00910-60002 P-band termination
•
11644-20017 P-band short
•
11644-20020 P-band shim
•
P281C adapter (included in calibration kit): WR-62 to 7 mm
•
K Band
K11644A standard, WR-42: 18 to 26.5 GHz. Includes:
00896-60006 K-band standard section
•
00910-60001 K-band termination
•
11644-20016 K-band short
•
11644-20019 K-band shim
•
K281C adapter (included in calibration kit): WR-42 to 3.5 mm (f) Option 012 WR-42 to 3.5 mm (m)
R Band
R11644A standard, WR-28: 26.5 to 40 GHz. Includes:
00914-20028 R-band termination
•
11644-20005 R-band short
•
11644-20003 R-band shim
•
11644-60001 R-band 10 cm straight waveguide
•
11644-60016 R-band 5 cm straight waveguide
•
Q Band
Q11644A standard, WR-22: 33 to 50 GHz. Includes:
11644-60005 Q-band termination
•
11644-20004 Q-band short
•
11644-20001 Q-band shim
•
11644-60002 Q-band 10 cm straight waveguide
•
11644-60017 Q-band 5 cm straight waveguide
•
U Band
U11644A standard, WR-19: 40 to 60 GHz. Includes:
11644-60006 U-band termination
•
11644-20004 U-band short
•
11644-20002 U-band shim
•
11644-60003 U-band 10 cm straight waveguide
•
11644-60018 U-band 5 cm straight waveguide
•
Find us at www.keysight.com Page 23
Verification kits
All Keysight verification kits include:
Precision Z0 airline or match thru
•
Mismatched airline or mismatch thru
•
Fixed attenuators (except 85059V)
•
Traceable measured data and uncertainties
•
85051B 45 MHz to 18 GHz 7 mm kit
Includes attenuators, airline and mismatch airline with data on a USB drive for use in confirming
calibrated system performance, traceable to national standards. Test procedure is provided in the service
manual.
85053B 300 kHz to 26.5 GHz 3.5 mm kit
Includes attenuators, airline and mismatch airline with data on a USB drive for use in confirming
calibrated system performance, traceable to national standards. Test procedure is provided in the service
manual.
85055A 300 kHz to 18 GHz Type-N kit
Includes attenuators, airline and mismatch airline with data on a USB drive for use in confirming
calibrated system performance, traceable to national standards. Test procedure is provided in the service
manual.
85057B 45 MHz to 50 GHz 2.4 mm kit
Includes attenuators, airline and mismatch airline with data on a USB drive for use in confirming
calibrated system performance, traceable to national standards. Test procedure is provided in the service
manual.
85058V 45 MHz to 67 GHz 1.85 mm kit
Includes attenuators, match thru and mismatch thru with data on a USB drive for use in confirming
calibrated system performance, traceable to national standards. Test procedure is provided in the service
manual.
Electric Calibration (ECal) Modules – Technical Overview, 5963-7343E
Web Resources
www.keysight.com/find/usb-vna
www.keysight.com/find/na
www.keysight.com/find/ecal
For more information on Keysight Technologies’ products, applications or services,
please contact your local Keysight office. The complete list is available at:
www.keysight.com/find/contactus
Find us at www.keysight.com Page 25