No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Manual Part Number
N8973-90014
Edition
Edition 1, April 2018
Published by:
Keysight Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
U.S. Government Rights
The Software is “commercial
computer software,” as defined
by Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202,
the U.S. government acquires
commercial computer software
under the same terms by which
the software is customarily
provided to the public.
Accordingly, Keysight provides
the Software to U.S. government
customers under its standard
commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of
which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA
represents the exclusive authority
by which the U.S. government
may use, modify, distribute, or
disclose the Software. The EULA
and the license set forth therein,
does not require or permit,
among other things, that
Keysight: (1) Furnish technical
information related to
commercial computer software
or commercial computer
software documentation that is
not customarily provided to the
public; or (2) Relinquish to, or
otherwise provide, the
government rights in excess of
these rights customarily provided
to the public to use, modify,
reproduce, release, perform,
display, or disclose commercial
computer software or
commercial computer software
documentation. No additional
government requirements
beyond those set forth in the
EULA shall apply, except to the
extent that those terms, rights, or
licenses are explicitly required
from all providers of commercial
computer software pursuant to
the FAR and the DFARS and are
set forth specifically in writing
elsewhere in the EULA. Keysight
shall be under no obligation to
update, revise or otherwise
modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant
to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S.
government acquires no greater
than Limited Rights as defined in
FAR 27.401 or DFAR 227.7103-5
(c), as applicable in any technical
data.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
Where to Find the Latest Information
Documentation is updated periodically. For the latest information about these products, including instrument software
upgrades, application information, and product information, browse to one of the following URLs, according to the name
of your product:
http://www.keysight.com/find/nfa
To receive the latest updates by email, subscribe to Keysight Email Updates at the following URL:
http://www.keysight.com/find/MyKeysight
Information on preventing instrument damage can be found at:
www.keysight.com/find/PreventingInstrumentRepair
Is your product software up-to-date?
Periodically, Keysight releases software updates to fix known defects and incorporate product enhancements. To search
for software updates for your product, go to the Keysight Technical Support website at:
http://www.keysight.com/find/sa_software
3
4
Contents
1 Quick Start
Initial Inspection 10
Verify the Contents 10
Shipping Problems? 10
Instrument Location and Rack Mounting Requirements 11
Locating the analyzer 11
Cooling and rack mounting 11
Turning on the Analyzer the First Time 12
Initializing the analyzer 12
Anti-Virus Software and Firewalls 16
Contents
Instrument Information 17
Power requirements 17
Protecting Against Overpowering 18
Instrument Maintenance 18
Protecting against electrostatic discharge 20
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements 21
Initial Connection to the Signal Analyzer 22
Guide for Using U7227-Series USB Preamplifier and the Internal Preamplifier
inside X-Series Signal Analyzer 25
This section explains how to initialize the Analyzer and view a signal.
The following topics can be found in this section:
“Initial Inspection” on page 10
“Instrument Location and Rack Mounting Requirements” on page 11
“Turning on the Analyzer the First Time” on page 12
“Anti-Virus Software and Firewalls” on page 16
“Instrument Information” on page 17
“Using the U7227-Series USB Preamplifiers in Noise Figure
Measurements” on page 21
9
Quick Start
Initial Inspection
Initial Inspection
Inspect the shipping container and the cushioning material for signs of stress.
Retain the shipping materials for future use, as you may wish to ship the
analyzer to another location or to Keysight Technologies for service.
Verify the Contents
Verify the shipping container contents using the box contents list.
Shipping Problems?
If the shipping materials are damaged or the contents of the container are
incomplete:
— Contact the nearest Keysight Technologies office.
— Keep the shipping materials for the carrier’s inspection.
— If you must return an analyzer to Keysight Technologies, use the original (or
comparable) shipping materials. See “Returning an Analyzer for Service”
on page 93.
10 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Quick Start
Instrument Location and Rack Mounting Requirements
Instrument Location and Rack Mounting Requirements
Locating the analyzer
Make sure that the fan inlet and exhaust vent areas on the sides of the analyzer
are not obstructed. The minimal required clearance is 2 inches. Airflow
restrictions cause additional airflow noise and cause the fans to speed up so
they can draw in enough air for the required cooling. This results in excessive
audible noise.
Cooling and rack mounting
Do not rack mount the analyzer side-by-side with any other instrument with
side-by-side ventilation. Make sure the exhaust air from the first instrument is
directed away from the inlet of the second unit. If the pre-heated air from the
first instrument is directed into the second instrument, it can cause excessive
operating temperatures in the second unit and can cause instrument failures.
The analyzer draws air in from the left side and exhausts air from the right side.
When installing the product in a cabinet, the convection into and out of
the product must not be restricted. The ambient temperature (outside the
cabinet) must be less than the maximum operating temperature of the
product by 4° C for every 100 watts dissipated in the cabinet. If the total
power dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 11
Quick Start
Turning on the Analyzer the First Time
Turning on the Analyzer the First Time
Initial power-on of the analyzer can be accomplished using the following
methods:
During the initial power-on process, the instrument turns off and restarts
several times. This only happens during the initial installation process.
Initializing the analyzer
StepsActionsNotes
1. Power on the analyzera. Position the analyzer so
you have easy access to
the power cord and plug
it in.
b. Press the power switch
(located next to the
lower right corner of the
display) to turn the
analyzer on.
2. Viewing the End-User
License Agreement
After a brief startup, the following window appears giving you information about the
End-User License Agreement.
See
“Instrument Location and Rack
Mounting Requirements
requirements” on page 17
The analyzer can require more than 5 minutes to
power-on.
The Keysight Technologies screen appears followed by a
screen that allows you to select Windows7 or the
Recovery option.
and “Power
for more details.
At this time, it is safe to turn off the instrument before initializing the software.
12 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Quick Start
Turning on the Analyzer the First Time
StepsActionsNotes
3. Reboot and logon
If you do not check the “Do not show this message again” check box, this message
will be displayed each time the analyzer is turned on. No application will start while
this message is displayed. Before continuing, make sure that you carefully read the
Anti-Virus message and determine what action is appropriate.
4. Disable the Anti-Virus
message
click
OK
Continue.
— Click
— Select the check box and
When the instrument restarts, the following message
window appears:
This window
appears and covers
the Launch window.
Messages similar to the following continue to appear:
5. When the installation
is complete, you
should see a display
like this:
Several required processes continue.
The application initializes
Keysight NFA X-Series Getting Started and Troubleshooting Guide 13
Quick Start
Turning on the Analyzer the First Time
StepsActionsNotes
6. Set user interface
language
a. On the instrument, press
System, User
Interface tab
b. Choose the desired
language from the
Language drop-down
menu.
.
7. Verify the installationa. On the instrument, press
System, Show
System.
b. Verify that the
purchased
application(s) appear in
the list or have an
entitlement certificate.
8. View a signalc. Press the Noise Figure 1
tab, select Spectrum
Analyzer, OK.
d. Press
Input/Output,
RF Calibrator, 50
MHz.
e. Press
f. Press
FREQ, Center
, 50, MHz.
Freq
SPAN, 50, MHz.
If you require further assistance, contact the Keysight
support team.
Online assistance:
http://
www.keysight.com/find/assist
This selects the Spectrum Analyzer mode where you can
view a reference signal.
This routes the internal 50 MHz signal to the analyzer
input.
14 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Quick Start
Turning on the Analyzer the First Time
StepsActionsNotes
The 50 MHz reference signal appears on the
display.
9. Return to the Noise Figure
measurement
— Press the Spectrum
Analyzer 1 tab, select
Noise Figure, OK.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 15
Quick Start
Anti-Virus Software and Firewalls
Anti-Virus Software and Firewalls
No anti-virus software is shipped with the analyzer. It is recommended that you
install anti-virus software if your analyzer is connected to the LAN. Check with
your IT department to see what they recommend.
The analyzer is shipped with the Windows 10 firewall enabled.
Do not modify the default network settings as this may cause problems with
the operating system of the analyzer.
16 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Quick Start
Instrument Information
Instrument Information
Power requirements
The only physical installation of your Keysight signal analyzer is a connection
to a power source. Line voltage does not need to be selected.
This analyzer does not contain customer serviceable fuses.
The instruments can operate with mains supply voltage fluctuations up to
± 10% of the nominal voltage.
This is a Safety Class 1 Product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact. Any
interruption of the protective conductor inside or outside of the product is
likely to make the product dangerous. Intentional interruption is
prohibited. (IEC 348 clauses 17.3.3c & 17.3.4)
Failure to ground the analyzer properly can result in personal injury.
Before turning on the analyzer, you must connect its protective earth
terminals to the protective conductor of the main power cable. Insert the
main power cable plug into a socket outlet that has a protective earth
contact only. DO NOT defeat the earth-grounding protection by using an
extension cable, power cable, or autotransformer without a protective
ground conductor.
This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 61010 Second Edition and IEC 664 respectively.
This instrument has autoranging line voltage input. Be sure the supply
voltage is within the specified range.
The Mains wiring and connectors shall be compatible with the connector
used in the premise electrical system. Failure, to ensure adequate earth
grounding by not using the correct components may cause product
damage, and serious injury.
AC power cord
The analyzer is equipped with a three-wire power cord, in accordance with
international safety standards. This cable grounds the analyzer cabinet when
connected to an appropriate power line outlet. The cable appropriate to the
original shipping location is included with the analyzer.
Always use the three-prong AC power cord supplied with this product.
Failure to ensure adequate earth grounding by not using this cord can
cause product damage.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 17
Quick Start
Instrument Information
If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal
condition (in which all means for protection are intact) only.
Install the instrument so that the detachable power cord is readily
identifiable and easily reached by the operator. The detachable power
cord is the instrument disconnecting device. It disconnects the mains
circuits from the mains supply before other parts of the instrument. The
front panel switch is only a standby switch and is not a LINE switch.
Alternatively, an externally installed switch or circuit breaker (which is
readily identifiable and is easily reached by the operator) may be used as
a disconnecting device.
Protecting Against Overpowering
The input circuitry of the analyzer can be damaged by applying signals that
exceed the maximum safe input level of +30 dBm average total power or +/-
0.2 Vdc (when DC coupled) or +/- 70 Vdc (when AC coupled). Refer to the
analyzer's specification guide for more details regarding the Maximum Safe
Input Level. Repairing damage to the input circuitry can be expensive.
If the analyzer will be used to measure signals which might be near the
maximum safe input level, use external attenuators and/or limiters to help
protect the analyzer input. The External Gain, amplitude Corrections, and/or
Ref Lvl Offset features may be used to compensate for the gains and losses of
external devices. External Gain and Corrections are under the Input/Output
menu and Ref Lvl Offset is under the AMPTD Y-Scale menu.
Instrument Maintenance
Cleaning the instrument
To prevent electrical shock, disconnect the signal analyzer from mains
before cleaning. Use a dry cloth or one slightly dampened with water to
clean the external case parts. Do not attempt to clean internally.
18 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Quick Start
Instrument Information
Cleaning Connectors
Cleaning connectors with alcohol shall only be done with the instrument power
cord removed, and in a well-ventilated area. Allow all residual alcohol moisture
to evaporate, and the fumes to dissipate prior to energizing the instrument.
Keep isopropyl alcohol away from heat, sparks, and flame. Store in a
tightly closed container. It is extremely flammable. In case of fire, use
alcohol foam, dry chemical, or carbon dioxide; water may be ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact with
eyes, skin, and clothing. It causes skin irritation, may cause eye damage,
and is harmful if swallowed or inhaled. It may be harmful if absorbed
through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal,
state, and local environmental regulations.
Battery Information
The analyzer uses a lithium battery located on the CPU board. This is not an
operator replaceable part. See “Returning an Analyzer for Service” on
page 93. Replaceable parts must be approved or supplied by Keysight
Technologies.
You can order the service documentation for the instrument through your
Keysight Sales and Service office.
Danger of explosion if battery is incorrectly replaced. Replace only with
the same or equivalent type recommended. Discard used batteries
according to the manufacturer’s instructions.
Do not throw batteries away but collect as small chemical waste.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 19
Quick Start
Instrument Information
Protecting against electrostatic discharge
Electrostatic discharge (ESD) can damage or destroy electronic components
(the possibility of unseen damage caused by ESD is present whenever
components are transported, stored, or used).
Test equipment and ESD
To help reduce ESD damage that can occur while using test equipment:
Do not use these first three techniques when working on circuitry with a
voltage potential greater than 500 volts.
— Before connecting any coaxial cable to an analyzer connector for the first
time each day, momentarily short the center and outer conductors of the
cable together.
— Personnel should be grounded with a 1 MΩ resistor-isolated wrist-strap
before touching the center pin of any connector and before removing any
assembly from the analyzer.
— Be sure that all instruments are properly earth-grounded to prevent
build-up of static charge.
— Perform work on all components or assemblies at a static-safe workstation.
— Keep static-generating materials at least one meter away from all
components.
— Store or transport components in static-shielding containers.
— Always handle printed circuit board assemblies by the edges. This reduces
the possibility of ESD damage to components and prevent contamination of
exposed plating.
Additional information about ESD
For more information about ESD and how to prevent ESD damage, contact the
Electrostatic Discharge Association (http://www.esda.org). The ESD standards
developed by this agency are sanctioned by the American National Standards
Institute (ANSI).
20 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Quick Start
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
Using the U7227-Series USB Preamplifiers in Noise Figure
Measurements
The U7227-Series USB Preamplifiers turn the overall X-Series noise figure
analyzers into better noise figure analyzers by reducing instrument noise
figure. Reducing instrument noise figure is necessary when the device under
test (DUT) is lossy or has low gain. This is because the instrument noise figure
contributes more to the final measurement uncertainty when the gain of the
DUT is low. Figure 1-1 shows how the measurement uncertainty of a typical
noise figure measurement depends on the DUT gain and instrument noise
figure. The data is derived using the Uncertainty Calculator in the N9069C
noise figure measurement application. It can be seen that the measurement
uncertainty increases when the DUT gain is low and instrument noise figure is
high. In optimal noise figure measurements, these regions should be avoided.
Therefore adding preamplifier correction data is not necessary.
Figure 1-1 3-D Plot of Measurement Uncertainty versus DUT Gain and Instrument Noise
Figure (DUT Noise Figure: 3 dB)
This section first introduces how to connect the U7227-Series USB
Preamplifier to the signal analyzer and then provides a rough guide about
using the USB preamplifier and the internal preamplifier inside the signal
analyzer.
“Initial Connection to the Signal Analyzer” on page 22
“Guide for Using U7227-Series USB Preamplifier and the Internal Preamplifier
inside X-Series Signal Analyzer” on page 25
Keysight NFA X-Series Getting Started and Troubleshooting Guide 21
Quick Start
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
Initial Connection to the Signal Analyzer
The U7227-Series USB Preamplifier is powered via a USB connection from the
signal analyzer. To connect it to the analyzer, follow the steps below.
Before connecting a signal to the input port of the U7227-Series USB
Preamplifier, make sure the USB preamplifier can safely accept the signal
level provided. The signal level limits are marked next to the Input
connectors of the USB preamplifier.
1. Connect the Output port of the USB preamplifier to the RF Input port of
the signal analyzer.
2. Connect the USB port (Type-B) of the USB preamplifier to the USB port
(Type-A) of the signal analyzer either in the front panel or in the rear panel
using a USB cable. You can “hot plug” the USB preamplifier into the
analyzer’s USB port at any time.
The USB connections of the U7227-Series USB Preamplifier and the
keyboard to the signal analyzer must be separated from each other; for
example, the USB preamplifier connects through the front panel while the
keyboard connects through the back panel.
Figure 1-2 shows the U7227-Series USB Preamplifier connection in the two
stages of the noise figure measurement. While the USB preamplifier is plugged
into one of the analyzer’s USB ports, the analyzer will consider it to be in the
RF input signal path and will apply the gain settings and other calibration data
of the USB preamplifier to the measurements.
Figure 1-2 U7227-Series USB Preamplifier and Analyzer Connection in the Two Stages of
Noise Figure Measurement
22 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Quick Start
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
In the noise figure measurement application, the signal analyzer will respond in
the following manner when the USB cable of the U7227-Series USB
Preamplifier is plugged into one of the analyzer’s ports:
1. The sweep will pause momentarily while the “Hardware Configuration
Updating” message appears on the signal analyzer display. See Figure
2. The USB preamplifier’s calibration data, such as gain, noise figure, and
S-parameters, will be automatically downloaded into the analyzer.
The USB preamplifier’s RF output cable, connector loss, and frequency
response are accounted for in the data that is automatically downloaded
into the analyzer’s memory. The calibration data is not user accessible.
3. The “Ready” LED is lit on the USB preamplifier.
4. The USB preamplifier’s model and serial number will be displayed. See
Figure 1-4.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 23
Quick Start
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
Figure 1-4 Signal Analyzer Display after USB Preamplifier is Ready For Use
5. The USB preamplifier is ready for use.
24 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Quick Start
1RLVH
6RXUFH
'87
86%
3HDPS
,QWHUQDO
$WWHQXDWRU
,QWHUQDO
3UHDPS
,QSXW
0L[HU
3UHVHOHFWRU
)RUIUHTXHQFLHV*+]
)RUIUHTXHQFLHV
!*+]
,QVLGHWKH6LJQDO$QDO\]HU
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
Guide for Using U7227-Series USB Preamplifier and the Internal
Preamplifier inside X-Series Signal Analyzer
This section provides a rough guide for using the U7227-Series USB
Preamplifier, the internal preamplifier, and the internal attenuator together to
make accurate noise figure measurements.
Amplitude Concerns in the Noise Figure Test System
Before talking about the guide, let us look at some signal amplitude concerns
in the test system. Figure 1-5 shows all the possible blocks in the signal path
of the noise signal in the noise figure measurement. The blocks with dotted
lines indicate that you can choose whether to use them or not. The USB
preamplifier, the internal preamplifier and input mixer inside the signal analyzer
are non-linear components and have a limit on the largest signal that can be
processed. The power limits of the noise signal are for the following places:
— before the USB preamplifier
— before the internal preamplifier
— before the input mixer
To make an accurate noise figure measurement, the noise power should be
below the maximum allowable power at all these three places. You can refer to
the technical overview of the U7227-Series USB Preamplifier and the
specification guide of the X-Series signal analyzer to check the maximum
allowable powers. For power levels before the internal preamplifier and before
the input mixer, the internal attenuator inside the signal analyzer can be
applied to reduce the power level. But at the same time, the instrument noise
figure will increase with using the internal attenuator.
The rough guide for the noise figure test system configuration in the following
section is based on the noise power calculation before the non-linear
components.
Figure 1-5 Blocks in the Signal Path of the Noise Signal in the Noise Figure Measurement
Keysight NFA X-Series Getting Started and Troubleshooting Guide 25
Quick Start
103.59 109×100 106×⁄()log×
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
Rough Guide for the Noise Figure Test System Configuration
To configure the test system for accurate noise figure measurements, the first
priority is to make the instrument noise figure (USB preamplifier + signal
analyzer) as low as possible. To reduce the instrument noise figure, the USB
preamplifier and internal preamplifier should be used whenever possible. If an
overload or compression exists in the test system, removing or turning off the
preamplifier (USB preamplifier or internal preamplifier) or adding internal
attenuation can eliminate the problem. On the signal analyzer,
— To turn on/off the internal preamplifier, press AMPTD, Signal Path tab,
Internal Preamp, On/Off.
— To set the internal attenuator, press AMPTD, Attenuation tab, and set the
attenuation.
Figure 1-6 and Figure 1-7 show the supported DUT characteristics (noise
figure and gain) of different system configurations for different frequency
ranges up to 26.5 GHz. The X-axis is for the DUT noise figure, the Y-axis is for
DUT gain, and the different colors on the figures indicate the recommended
test system configurations. This rough guide is based on the noise power
calculation before the non-linear components.
Before using Figure 1-6 and Figure 1-7 to find a proper test system
configuration, you need to pay attenuation to the following points:
— The DUT reference bandwidth used to calculate the noise power is 3.59
GHz, which is the worst case for the lower band (10 MHz to 3.6 GHz). To
maintain consistency between the lower band and higher band, the same
DUT reference bandwidth is used in the calculation for frequency ranges
above 3.6 GHz. Then the reference bandwidths (Ref_BW) for both below 3.6
GHz and above 3.6 GHz are both 3.59 GHz.
If the DUT has a narrower bandwidth, the DUT characteristics can be
increased accordingly. For example, if the DUT has a bandwidth of 100
MHz, the supported DUT gain can be increased by a factor of
, that is by 15.6 dB.
— The noise figure and gain for the U7227C USB Preamplifier are used in the
calculation. If you are using the U7227A or U7227F, the supported DUT
characteristics should be different.
— A 6 dB ENR noise source is used in the calculation. If you are using a noise
source with higher ENR, the supported DUT gain will be lower.
— For different instruments and for a wide frequency range, the data for the
USB preamplifier and the X-Series signal analyzer differs considerably, the
most precise data is used in most cases.
26 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Quick Start
AF 10=Ref_BW DUT_BW⁄()103.59109×DUT_BW⁄()log×=log×
DUT_Gain_ModifiedDUT_GainAF–=
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
Below is the procedure for using Figure 1-6 and Figure 1-7 to find the proper
system configuration. See “Examples for Using the Guide to Find the Test
System Configuration” on page 30 for examples of using this procedure.
1. Check the DUT and get the rough data for the DUT bandwidth (BW), noise
figure (NF), and gain.
2. Calculate the bandwidth adjust factor (AF) using the following equation.
1
Then subtract the AF from the DUT_Gain. The resul tant value is called
DUT_Gain_Modified.
3. Use the DUT_NF and DUT_Gain_Modified values to search in Figure 1-6 or
Figure 1-7 from bottom to top and find an approximate place for them.
4. Find the corresponding system configuration. If the system configuration
includes the internal attenuation, do the following to find the proper
attenuation level:
a. Set the attenuation to 0 dB and note the noise figure of your DUT.
b. Increase the attenuation by one step (4 dB). If the noise figure
changes too much, such as 0.3 dB, this attenuation is required for
accurate measurement.
c. Repeat step (b) until you find the proper value.
As in the calculation of the supported DUT characteristics in Figure 1-6
and Figure 1-7, the most precise data is used in most cases, it is
recommended to try the system configuration below the one you find
using DUT_NF and DUT_Gain_Modified first.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 27
1. The Ref_BW is the bandwidth used in Figure1-6 or Figure1-7 to calculate the noise power,
which is 3.59 GHz.
Quick Start
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
Figure 1-6 Supported DUT Characteristics and Recommended Test System
Configuration for Frequency Ranges below 3.6 GHz
28 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Quick Start
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
Figure 1-7 Supported DUT Characteristics and Recommended Test System
Configuration for Frequency Ranges above 3.6 GHz
For frequency ranges above 3.6 GHz, if no preamplifier is used in the signal
analyzer, the major noise power concern will be the noise power before the
mixer. In this case, the input noise outside the preselector bandwidth (less
than 80 MHz) will be filtered. This may affect the supported DUT
characteristics.
Figure 1-6 and Figure 1-7 provide a rough guide for frequency ranges up to
26.5 GHz. For the frequency ranges above, the U7227F USB Preamplifier used
together with the X-Series can support up to 50 GHz. The calculation method
for frequency ranges above 26.5 GHz is similar to the “3.6 GHz to 26.5 GHz”
frequency range. The supported DUT gain will decrease because of the
increase of the noise figure and gain for both the USB preamplifier and the
internal preamplifier. Refer to Table 1-1 and Table 1-2.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 29
Quick Start
9
×=
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
Table 1-1 Noise Figure and Gain Specification of the U7227-Series
USB Preamplifier
Specification U7227AU7U7227C7CU7227F
Frequency10 MHz to 4 GHz100 MHz to 26.5 GHz2 to 50 GHz
b
Gain (dB)
> 16 (10 to 100 MHz)
> 0.5 F + 17 (100 MHz to 4 GHz)
a
> 0.26F +16.1
(100 MHz to 26.5 GHz)
> 0.23F + 16.5
(2 to 50 GHz)
Noise Figure< 5.5 dB (10 to 100 MHz)
< 5 dB (10 MHz to 4 GHz)
< 6 dB (100 MHz to 4 GHz)
< 5 dB (4 to 6 GHz)
< 4 dB (6 to 18 GHz)
< 5 dB (18 to 26.5 GHz)
< 10 dB (2 to 4 GHz)
< 8 dB (4 to 40 GHz)
< 9 dB (40 to 44 GHz)
< 10 dB (44 to 50 GHz)
a. Specifications are tested and measured on an operating temperature of 23 °C.
b. "F" signifies frequency in GHz.
Table 1-2 Noise Figure and Gain Specification of the X-Series
Signal Analyzer
Spec
(nominal)
Gain (dB)
(.Max)
Noise Figure13 to 14.5 dB (100 kHz to 3.6 GHz)
NFA X-Series Noise Figure Analyzers
20 dB (100 kHz to 3.6 GHz)
35 dB (3.6 to 26.5 GHz)
40 dB (26.5 to 50 GHz)
9 dB (3.6 to 8.4 GHz)
10 dB (8.4 to 13.6 GHz)
11 to 18 dB (DANL+176.24 dB) (>13.6 GHz)
Examples for Using the Guide to Find the Test System Configuration
This section gives an example of using the “Rough Guide for the Noise Figure
Test System Configuration” on page 26 to find a proper test system
configuration
DUT with Frequency Ranges below 3.6 GHz
The DUT characteristic is as below:
DUT NF: 5 dB
DUT gain: 30 dB
Frequency range: 1 to 2 GHz
The DUT bandwidth is:
Hz
30 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Quick Start
9
×DUT_BW()⁄()log×5.55=
DUT_BW710
9
×=
9
×DUT_BW()⁄()log×2.9–=
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
The bandwidth adjust factor is:
The modified DUT gain is:
dB
Use DUT NF and DUT Gain Modified values to search in Figure 1-6 from
bottom to top. The test system configuration should be USB Preamp + Internal
Preamp + Internal Attenuator.
DUT with Frequency Ranges above 3.6 GHz
The DUT characteristic is as below:
DUT NF: 4 dB
DUT gain: 25 dB
Frequency range: 3 to 10 GHz
The DUT bandwidth is:
Hz
The bandwidth adjust factor is:
The modified DUT gain is:
dB
Use DUT NF and DUT Gain Modified values to search in Figure 1-7 from
bottom to top. The test system configuration should be USB Preamp only.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 31
Quick Start
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
32 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Front and Rear Panel Features
Front-Panel Features
Front-Panel Features
Item
Description
#Name
1Pouch
2Measurement KeysThese keys (in the shaded area) enable you to set the parameters used for
making measurements in the current Mode and Measurement.
3Mode/Measurement KeyThis key enables you to select the desired Mode (measurement application),
Measurement, and/or View.
4Preset KeysMode Preset - local to the current mode, global to all measurements in the
mode, affects most but not all parameters in the mode, does not affect
Input/Output or System variables
User Preset - local to the current mode, global to all measurements in the
mode, affects all parameters in the mode as well as the Input/Output
variables. Does not affect System variables.
34 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Front and Rear Panel Features
Front-Panel Features
Item
Description
#Name
5Save/Recall KeysSave - enables you to save states, traces, screen images and other items
from the analyzer to files on the analyzer’s internal storage, to removable
devices, and to directories on the network.
Recall - enables you to recall previously saved states, traces and other items
to the analyzer from files on the analyzer’s internal storage, from removable
devices, and from directories on the network.
Quick Save - repeats the previous Save at the touch of a single button.
Whatever you saved before gets saved again to the same directory, and with
a filename derived from the previous filename.
6Sweep KeysSingle/Cont - toggles between single and continuous measurement
sweeps.
Restart - restarts the measurement.
7KnobIncrements and decrements the value of the current active function.
8Enter and Arrow KeysThe Enter key terminates data entry when either no unit of measure is
needed, or you want to use the current unit.
The arrow keys:
— Increment and decrement the value of the current
measurement selection. (up/down = big increment, left/right
= small increment)
— Navigate help topics.
— Navigate or make selections within Windows dialogs.
— Navigate within forms used for setting up measurements.
— Navigate within tables.
9Headphones OutputHeadphones can be used to hear any available audio output.
10USB ConnectorsStandard USB 3.0 ports, Type A. Connect to external peripherals such as a
mouse, keyboard, DVD drive, or hard drive.
11Utility KeysThe following keys are available in the Utility section:
Numeric key pad
System
Help
Local/Cancel/(Esc)
Back-space
Delete
Control
Alt
Undo
Touch On/Off
Onscreen Keyboard
Tab
12RF InputConnector for inputting an external signal. Make sure that the total power of
all signals at the analyzer input does not exceed +30 dBm (1 watt).
Keysight NFA X-Series Getting Started and Troubleshooting Guide 35
Front and Rear Panel Features
Front-Panel Features
Item
#Name
Description
13Power Standby/ On
Turns the analyzer on. A green light indicates power on. A yellow light
indicates standby mode.
The front-panel switch is a standby switch,
not a LINE switch (disconnecting device).
The analyzer continues to draw power even
when the line switch is in standby.
The main power cord can be used as the
system disconnecting device. It
disconnects the mains circuits from the
mains supply.
36 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Front and Rear Panel Features
Display Features
Display Features
This section describes the regions of the display.
Invalid Data Indicator
The invalid data indicator is displayed whenever the data on the display
does not match the settings of the analyzer. The most common example
of this is when instrument settings have changed in the time since the data in
the traces on the display was taken. This means that the screen annotation
cannot be guaranteed to match the trace data. For example, if you change
Center Frequency, the invalid data indicator will display until a new sweep has
completed.
If any Trace is in View mode (displaying but not updating) and analyzer settings
are changed, the invalid data indicator will display as long as that trace
remains in View. Traces that are blanked do not turn on the invalid data
indicator.
Not all analyzer settings require display of the invalid data indicator when they
change, only changes that require a new acquisition will cause it to display.
For example, changing the Y-Axis scale of the analyzer does not cause the
invalid data indicator to display, unless the attenuation changes.
Also, the invalid data indicator is turned on:
— When the counter is turned on, until the completion of the first count
Keysight NFA X-Series Getting Started and Troubleshooting Guide 37
Front and Rear Panel Features
Display Features
— When a trace is imported from mass storage and the trace’s parameters do
not match the current analyzer settings
— When a trace is sent to the analyzer from a remote interface (since there is
no way to know if its settings match)
The Data Invalid Indicator has an associated status bit that can be checked
at any time to see if it is on.
Screen Tabs
Along the top of the display are tabs, one for each measurement screen you
have defined. Tap the + sign to "clone" the current measurement, which can be
changed once it is created. Tap the current screen tab (or press the
Mode/Meas hardkey) to display the following dialog:
This dialog allows you to choose a Mode, Measurement and View.
When you select a mode, the measurements that are available in the mode are
displayed in the Measurement column.
When you select the desired measurement, the views available for the
measurement are displayed under the View column.
You can have up to 16 measurement tabs, but only 6 can be viewed at one
time. If the tabs overfill the top bar, they scroll left and right using the arrows to
the left and right of the tabs.
You switch screens by touching the tabs. To view multiple screens, press the
icon on the Bottom Bar.
38 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Front and Rear Panel Features
Display Features
Meas Bar
The Meas Bar shows general measurement settings and information. The
annotations on this bar can be used to change settings. Tap anywhere in the
annotation box to access the drop-down panel that contains relevant
parameters. The following graphic shows some of the drop-down menus and
the parameters they contain.
Indicates single/continuous measurement.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 39
Front and Rear Panel Features
Display Features
Measurement Display
This area shows the measurement results in graphical and tabular form. You
can interact with this area using pinch, drag, scroll and tap gestures.
On the signal:
Horizontal pinch - changes the span of the analyzer
Horizontal drag - changes the center frequency
Vertical pinch - changes the vertical scaling
Vertical drag - changes the reference level
Markers may be moved by dragging them to the desired location
Touch and hold - simulates a right click
Swipe scroll - allows you to view information that extends beyond the window
area. When you begin to scroll, the scroll bar appears and fades once you stop
scrolling.
For demonstrations on these gestures and how they can be used, go to:
The annotation drop-downs in the window area allow you to change
parameters. The window title drop-down allows you to resize the windows in
the multi-window format. When you tap in those areas, the drop-down control
menus appear as shown in the following graphic.
40 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Front and Rear Panel Features
Display Features
Menu Panel
At the top of the menu panels are two icons:
The Preset icon accesses the following control menu:
The System icon accesses the following dialog:
Keysight NFA X-Series Getting Started and Troubleshooting Guide 41
Front and Rear Panel Features
Display Features
The hardkey dropdown panel contains the measurement controls. These are
the same as the hardkeys in the shaded area of the keypad:
You can select functions using either the keypad or the drop-down menu.
42 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Front and Rear Panel Features
Display Features
Notice that the Frequency panel has a Settings tab on the right side. Other
panels may have multiple tabs. The tabs access controls for the particular
parameter noted on the tab.
Bottom Bar
The bottom bar contains several icons that access various controls.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 43
Front and Rear Panel Features
Rear-Panel Features
Rear-Panel Features
Item
#Name
1EXT REF INInput for an external frequency reference signal:
210 MHz OUTAn output of the analyzer internal 10 MHz frequency reference signal. It is used to
lock the frequency reference of other test equipment to the analyzer.
3SNS Series Noise SourceFor use with Keysight N4000A, N4001A, N4002A Smart Noise Sources (SNS).
4Noise Source Drive +28 V
(Pulsed)
5TRIGGER 1 INAllows external triggering of measurements.
6TRIGGER 2 INAllows external triggering of measurements.
7SyncReserved for future use.
8TRIGGER 1 OUTA trigger output used to synchronize other test equipment with the analyzer.
9TRIGGER 2 OUTA trigger output used to synchronize other test equipment with the analyzer.
For use with Keysight 346A, 346B, and 346C Noise Sources.
Configurable from the Input/Output keys.
Configurable from the Input/Output keys.
Description
44 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Front and Rear Panel Features
Rear-Panel Features
Item
#Name
10Analog OutFor Option YAV:
Screen Video
Log Video
Linear Video
11Digital BusReserved for future use
12Aux IF OutCR3 Second IF Out
CRP Arbitrary IF Out
13Line power inputThe AC power connection. See the product specifications for more details.
14Removable Disk DriveStandard on current analyzers.
15VGAAllows connection of an external VGA monitor.
16DisplayPortUsed for video output. Accepts a standard mini-displayport connector, or
adaptor for connection of an external monitor.
17LAN/USBA TCP/IP Interface that is used for remote analyzer operation.
USB 2.0 port, Type A Connect to external peripherals such as a mouse, keyboard,
printer, DVD drive, or hard drive.
Description
18USB ConnectorUSB 3.0 port, Type B. USB TMC (test and measurement class) connects to an
external pc controller to control the instrument and for data transfers over a 480
Mbps link.
19USB ConnectorsStandard USB 3.0 ports, Type A. Connect to external peripherals such as a
mouse, keyboard, printer, DVD drive, or hard drive.
20PCIe X4Reserved for future use
21GPIBA General Purpose Interface Bus (GPIB, IEEE 488.1) connection that can be used
for remote analyzer operation.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 45
Front and Rear Panel Features
Front and Rear Panel Symbols
Front and Rear Panel Symbols
This symbol is used to indicate power ON (green LED).
This symbol is used to indicate power STANDBY mode
This symbol indicates the input power required is AC.
The instruction documentation symbol. The product is
The CE mark is a registered trademark of the European
The RCM mark is a registered trademark of the
This is a marking of a product in compliance with the
(yellow LED).
marked with this symbol when it is necessary for the
user to refer to instructions in the documentation.
Community.
Australian Spectrum Management Agency.
Canadian Interference-Causing Equipment Standard
(ICES-001).
This is also a symbol of an Industrial Scientific and
Medical Group 1 Class A product (CISPR 11, Clause 4).
The CSA mark is a registered trademark of the Canadian
Standards Association International.
This symbol indicates separate collection for electrical
and electronic equipment mandated under EU law as of
August 13, 2005. All electric and electronic equipment
are required to be separated from normal waste for
disposal (Reference WEEE Directive 2002/96/EC).
Indicates the time period during which no hazardous or
toxic substance elements are expected to leak or
deteriorate during normal use. Forty years is the
expected useful life of the product.
This symbol on all primary and secondary packaging
indicates compliance to China standard GB
18455-2001.
South Korean Certification (KC) mark; includes the
marking’s identifier code which follows this format:
MSIP-REM-YYY-ZZZZZZZZZZZZZZ.
This equipment is Class A suitable for professional use
and is for use in electromagnetic environments outside
of the home.
To return unwanted products, contact your local
Keysight office.
46 Keysight NFA X-Series Getting Started and Troubleshooting Guide
This chapter describes the Microsoft Windows 10, configuration and the
settings used with the Keysight instrument software. It includes information
about changing some of the system settings. And it describes the Windows
operating system configuration and the software installations that are present
on the Disk Drive when the instrument leaves the factory.
It is possible to use the front panel and touchscreen for changing operating
system configuration items, but it is easier to perform these tasks with a USB
mouse and external keyboard. For more useful shortcuts, see “Navigating
Windows Without a Mouse” on page 70.
The following topics can be found in this chapter:
“Microsoft Windows” on page 48
“Keysight Software Installed” on page 48
“User Accounts” on page 49
“Windows Configuration” on page 51
“Configuring Printers” on page 54
“Configuring LAN” on page 55
“Windows Security” on page 57
“System Maintenance” on page 59
“USB Connections” on page 60
“Disk Drive Partitioning and Use” on page 61
“Disk Drive Recovery Process” on page 63
47
Instrument Operating System
Microsoft Windows
Microsoft Windows
Your instrument has Microsoft Windows installed at the factory. Keysight has
already configured many of the settings in Microsoft Windows for optimal
behavior in your instrument. This chapter contains details about many of these
settings.
Keysight Software Installed
Keysight Noise Figure Analyzer software
The N9069C Keysight Noise Figure Measurement Application software is
installed in the analyzer. The instrument comes preconfigured with full featured
Spectrum Analyzer and IQ Analyzer (Basic) modes preinstalled.
If you install programs other than those that Keysight has tested, it could cause
problems with the instrument's applications. If this happens, you should try
uninstalling the program that has caused the problem, or try changing the
program's configuration. If this does not correct the problem, you may have to
use the Keysight Recovery system to reinstall the instrument's system software
48 Keysight NFA X-Series Getting Started and Troubleshooting Guide
User Accounts
Administrator account
Instrument Operating System
User Accounts
The instrument ships with a number of different accounts already set up. In
addition you can create your own accounts if you desire. The privileges
associated with each account determine what you can and cannot do from that
account.
Using the Administrator account you can perform the following operations:
—Install software
— Configure network and printer access
— Access all files on the instrument
— Add or change user accounts and passwords
— Change Firewall settings
— Change Windows settings (e.g., using Device Manager)
— Change the time and date
— Run any application
The default password is "Keysight4u!".
Instrument accounts
The default user account that ships from the factory is “Instrument” with the
password “measure4u”. This user is a member of the standard Users group.
Using the Instrument account, you may perform the following operations:
— Configure network and printer access (although not local printer access)
— Access files on the instrument that are accessible to the Users group
— Run applications that are accessible to the Users group
KeysightOnly user account
The instrument contains a user account called "KeysightOnly" that can be used
by Customer Support in the event that the customer has changed the
Administrator password and has forgotten the password. You must not remove
or modify the KeysightOnly account.
Service user accounts
There are user accounts defined in the instrument for servicing the instrument.
Customer creation of accounts
You can create additional user accounts and decide on the level of security
granted to any new user accounts created. For example, the level of security
can be assigned as administrator, power user, standard user, backup
operators. User names are not case sensitive but passwords are case sensitive.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 49
Instrument Operating System
User Accounts
It is Keysight’s expectation that each user’s My Documents folder is mapped to
the D: drive. This is to avoid overwriting the user’s data in the event the
Instrument Recovery must be performed. Also, this facilitates convenient
backup by copying the contents of the D: drive to external media. All user
accounts created by the factory already have My Documents mapped to the D:
drive. It is recommended to map all new users’ My Documents folders to the D:
drive.
The following measurement applications are preinstalled and licensed in the
noise figure analyzer:
1. Noise Figure
2. Spectrum Analyzer
3. IQ Analyzer (Basic)
The licenses for these measurement applications are fixed perpetual.
Fixed Perpetual
Fixed Perpetual licenses are the traditional license type (Fixed) with the same
duration (Perpetual) that have been available for all features since the
introduction of the X-Series analyzers. Fixed Perpetual licenses are identified
by the “F” in the second character and “P” in the third character of the option
designator:
Example: N9069EM0E-1FP
A license key is instrument model and serial number dependent. You can only
install the license key on the specific instrument for which it was created.
50 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Instrument Operating System
Windows Configuration
Windows Configuration
The Windows settings have been optimized for the best measurement
performance. Any modifications to these settings may degrade instrument
performance and measurement speed. In general, most Windows System
settings (typically set through the Windows Control Panel) should not be
modified. Those that can be safely modified are listed below.
To recover from problems caused by changing Windows Systems settings,
you may have to reinstall the Windows system and instrument
applications using the Instrument Recovery process.
Settings that can be changed
You may change the following Windows settings or administrative tasks
(available from the Windows Control Panel) to suit your own personal
preferences. It is recommended that you document any changes to the
instrument’s configuration in case an Instrument Recovery is performed and
the configuration is reset.
Some of these actions can only be performed with Administrator
privileges.
Table 3-1
You May Use This Feature:To Do This…
Configure Microsoft Windows Automatic Updates.
Microsoft recommends that you always get the latest critical Windows
updates to ensure that the instrument's Windows operating system is
protected. If the instrument has Internet access, the instrument default is
set to automatically check for critical Windows Updates and notify you.
Setup new user accounts.
Do not delete or modify the "KeysightOnly"
user account. Doing so may prevent Keysight
from servicing the instrument.
Add the Instrument to a network
Install and configure a printer
Set the time and date
Keysight NFA X-Series Getting Started and Troubleshooting Guide 51
Instrument Operating System
Windows Configuration
Table 3-1
You May Use This Feature:To Do This…
Settings that must not be changed
Avoid changing the settings described below (available from the Windows
Control Panel). Changes to these settings may degrade instrument
performance, screen displays, and measurement speed.
Table 3-2
If you click on “Advanced System Settings” a dialog will open called
“System Properties.” On this dialog there is an “Advanced” tab, which
opens up a dialog with a number of settings options. One of these is
“Performance”, and if you click on the “Settings” button under
“Performance”, you will see another dialog with a number of settings
options. The default is “Let Windows choose what’s best for my
computer.” You can also select “Adjust for best performance.”
You should leave the remaining selections unchanged.
Do NOT Use This Feature:To Do This…
Do not change Power Options.
If you click on “Advanced System Settings” a dialog will open called “System
Properties.”
On this dialog there is a tab called “Hardware.” You should not modify any
settings under the “Hardware” tab.
On this dialog there is also a tab called “Advanced.” You should not modify any
settings under the “Advanced” tab except as described above under “Settings
that can be changed”.
Do not remove installed Fonts
Do not change the following Display Settings:
— Screen Saver settings (under “Personalization)
— Screen resolution, 1024 x 768 (under “Adjust Resolution”)
— DPI setting (under “Set custom text size”)
Do not change any settings under “Region and Language” or the instrument
keyboard and display may not operate properly
Do not delete or modify the "KeysightOnly" user account.
52 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Instrument Operating System
Windows Configuration
In addition, Do Not:
— Add, delete, or modify disk drive partitions.
— Delete or modify Keysight registry entries.
— Change the contents of any directories containing the name "Keysight".
— Stop the IIS server
— Tamper with any virtual directories (or their contents) that came configured
with the instrument.
— Uninstall these libraries, interfaces, or programs:
— The I/O Libraries
— The .NET Framework or any Hotfixes or Service Packs for the .NET
Framework
— The “Microsoft Visual J# .NET Redistributable Package 1.1”
— Programs that begin with "Keysight"
— The Adobe Acrobat reader
—Modify:
— The I/O Library “GPIB27”, “GPIB28” interfaces shown as configured
Instrument I/O in the Connection Expert or I/O Config
Autoplay/Autorun
Since the introduction of Windows XP, the term Autoplay (sometimes also
called Autorun) has come to be associated with the feature which assists users
in selecting appropriate actions when new media and devices are detected.
The Autoplay/Autorun feature is turned off in the instrument, by default, for
heightened security, unless the Administrator account is running.
If you wish to re-enable Autoplay/Autorun, you may use the Auto Play function
in the Control Panel. However, be aware that if you do this you may be more
subject to virus attack from portable media such as USB flash drives.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 53
Instrument Operating System
Configuring Printers
Configuring Printers
Printers are configured using the Microsoft Windows Control Panel. It is easily
accessed from the Windows Start menu or from under the front panel System
key. This setup process can be done using the touch screen and front-panel
keys. See “Navigating Windows Without a Mouse” on page 70.
When setting up a new printer, you may need to load the printer driver (unless
you are using a network printer that your IT department has set up to include
the driver). The manufacturer of the printer supplies the driver software and
process. That may require that you attach an external USB disk drive. An
alternative is to connect the instrument to the LAN and download the driver
from the printer manufacturer’s internet site.
54 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Instrument Operating System
Configuring LAN
Configuring LAN
Hostname
The Computer Name, or Hostname, is preconfigured from the factory. It must
be a unique name such that it does not conflict with other equipment on your
LAN. The preconfigured Computer Name is A-N8973B-xxxxx, where xxxxx is
the last 5 digits of the instrument’s serial number.
IP Address & Gateway
The instrument is preconfigured to obtain an IP Address via DHCP. You can
change the IP Address and Gateway as you desire.
You must be logged in as an "administrator" (default password:"Keysight4u!")
to make changes.
Connecting to a Network Shared Folder
The instrument contains standard Windows networking. The time required to
authenticate is dependent on your LAN infrastructure. You may have improved
performance by mapping a network drive to the shared folder that you need to
access.
To map a network drive, click the Start icon ., System Tools, Computer.
Select the Computer Tab, and select Map Network Drive from the drop down
menu:
.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 55
Instrument Operating System
Configuring LAN
When the Map Network Drive window appears, browse to the correct folder,
and click Finish.
In Windows 10 there is no visual indication that authentication is in
progress.
56 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Instrument Operating System
Windows Security
Windows Security
Microsoft recommends the following to ensure the instrument's Windows
operating system is protected:
—Use an internet firewall.
— Get the latest critical Windows updates.
— Use up-to-date antivirus software.
Windows Firewall
The instrument is shipped with the Windows Firewall enabled. You can verify
the status of Windows Firewall by going to the Control Panel and clicking on
System and Security, Windows Firewall.
Windows Firewall Exceptions for programs and ports have been added to allow
proper operation of the instrument over a network. Modifying these settings
may cause the instrument to not operate properly.
Automatic Updates
Microsoft recommends that you always get the latest critical Windows updates
to ensure that the instrument's Windows operating system is protected. If the
instrument has internet access, the instrument default is set to automatically
check for critical Windows Updates and notify you.
You can change the configuration of the Microsoft Automatic Updates. You can
choose not to have automatic updates. However, if you do this then you should
manually update Windows periodically, by accessing Internet Explorer and
selecting Windows Update from the Tools menu.
Be aware that downloading and installing Windows Updates can be
network and CPU usage intensive (impacting the instrument
performance), and some Windows Updates automatically reboot the
instrument. It is recommended that Windows Updates be performed
when the instrument is not in normal use.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 57
Instrument Operating System
Windows Security
Virus Protection
There is no antivirus software included with your instrument. To find
information regarding recommended antivirus software, visit the NFA website:
http://www.keysight.com/find/nfa
and select FAQs under Technical Support.
Having antivirus software installed may have a slight impact on the
instrument performance.
Spyware Protection
There is no anti-spyware software installed on the instrument. This should not
be a problem if you do not use the instrument for a lot of internet browsing.
Having spyware in the instrument could have an impact on the instrument
performance.
58 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Instrument Operating System
System Maintenance
System Maintenance
Backup
It is recommended that you have a regular backup strategy. Your IT
department may already have a backup strategy in place which is suitable for
the instrument and its data. Using the Instrument Recovery system in
conjunction with a regular backup strategy should allow full recovery of the
instrument data.
Windows has a Backup utility that you can use to archive files and folders in
case of a disk drive failure. See the Windows Help and Support Center for more
information on this utility. You can also use third party backup utilities.
However, you must ensure that this third party software is compatible with the
instrument's system software.
When performing backups, we recommend that you backup the data to an
external storage device connected to the network or one of the instrument's
USB connectors. Also, you should perform backups at times when the
instrument is not being used for normal operations, as it may impact the
instrument's overall performance.
System Restore
Windows contains the capability to restore the system to a previous point in
time. System Restore is enabled with default settings as provided by Microsoft.
However, System Restore is not always 100% successful. Therefore, it is not
recommended that you rely on System Restore to protect your instrument.
System Restore has not been tested to verify successful restoring.
Disk Defragmenting
The instrument has a solid state drive. Disk defragmenting is not
recommended.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 59
Instrument Operating System
USB Connections
USB Connections
All of the USB ports are compatible with the USB 2.0/3.0 and 1.1 specification.
The three USB ports on the front panel (see graphic below) and three of the
USB ports on the rear panel are USB Series “A” ports. These are ports to which
you can connect USB mass storage devices and printers. The instrument USB
Host support includes the standard Microsoft Windows USB drivers for human
interface, mass storage, printing, scanning, and imaging devices. A complete
up-to-date list of the Windows USB class driver support is available on the
Microsoft website.
The square USB port (see graphic below) on the rear panel is a USB 3.0 Series
“B” port and is used for controlling the instrument over USB. Information to
help you program your instrument is documented in the X-Series
Programmer’s Guide. The instrument USB device driver included in the
instrument software supports the test and measurement industry standard
USBTMC-USB488 device class.
In addition, the IO Libraries CD that was included with your instrument
contains USB Host drivers that allow control of other instruments connected to
the USB ports.
Keysight Technologies does not support or warrant correct instrument
operation if additional USB drivers from third parties are installed in the
instrument. It is possible that additional drivers could break the normal USB
operation. If USB operation is broken, recovery would require reinstalling the
instrument application using the Instrument recovery process.
60 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Instrument Operating System
Disk Drive Partitioning and Use
Disk Drive Partitioning and Use
The drive is partitioned into 3 sections: C:, D: and E:
—The C: partition contains the Windows 10 operating system and software
installed by Keysight. This is an Open System which means you can install
additional software, and these should be installed on the C: drive. However,
only a limited set of software applications are tested for use with the
Keysight measurement software. The installation and/or use of other
software is not warranted and could interfere with the operation of the
measurement software. If instrument repair is ever needed, the Keysight
version of the C: drive is the only part of the instrument software that is
restored by the Instrument Recovery process. You must reload any other
software that you have added in the instrument.
—The D: partition is reserved for data storage. The User Accounts that are
configured by Keysight have their My Documents folder mapped to the D:
drive. This is for the convenience of backing-up the measurement data. You
should always back-up the data on the D: drive to an external device. This
allows you to restore the data if you ever need to replace the disk drive.
Table 3-3
—The E: partition is reserved for Keysight’s use. The primary use of the E:
drive is for housing the Calibration and Alignment data. Do not change or
overwrite the files on this drive. This could cause your instrument to not
meet specifications, or even to stop functioning correctly. Do not use this
drive for data storage. It is also recommended that you back up the
contents of this drive by using the factory calibration data backup utility.
Backing-up the factory calibration data
To back-up the factory calibration data you will need a USB storage device.
StepNotes
1. Press
System, Alignments,
Backup/Restore Align Data...
An information window appears that informs you that
the instrument must shut down to complete the
backup:
"Press OK to force shut down and
proceed. Press Cancel to exit."
Keysight NFA X-Series Getting Started and Troubleshooting Guide 61
Table 3-3
Instrument Operating System
Disk Drive Partitioning and Use
StepNotes
2. Select OK.The Alignment Data Wizard will appear:
3. Follow the on screen instructions to
backup the calibration data to the
USB drive and save the file.
The default location is an
internal flash SD card, if
present, but you can
choose another location.
62 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Instrument Operating System
Disk Drive Recovery Process
Disk Drive Recovery Process
The Instrument Recovery System can be used to repair errors on the
instrument's C: drive partition or to restore the original factory configuration of
the system software on the disk drive. The Instrument Recovery System is
stored in a separate hidden disk drive partition.
Repairing errors on the disk drive may result in loss of data or files. If you
require more information, see the “chkdsk” documentation in the Windows
Help and Support Center.
Restoring the original factory system software does not restore any of the
following items:
— Additional software that has been installed after the instrument was
shipped from the factory. (Thus, following an Instrument Recovery any
software installed after the instrument was shipped from the factory will
need to be re-installed.)
— System configurations (for example user accounts, windows configurations)
that have been made after the instrument was shipped from the factory.
(Thus, following an Instrument Recovery configuration changes will have to
be performed.)
— The Instrument Recovery overwrites the contents of the C: partition. It does
not affect the D: or E: partitions.
It is recommended that the customer use a regular back up strategy. The
customer’s IT department may already have a backup strategy in place which
is suitable for the instrument and its data. Using the Instrument Recovery
System in conjunction with a regular back up strategy should allow the
customer to fully recover the instrument's software and data.
Table 3-4 Recovery Process
StepNotes
1. Make sure the instrument is turned
off.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 63
Instrument Operating System
Disk Drive Recovery Process
Table 3-4 Recovery Process
StepNotes
2. Turn on the instrument.
— Press the down arrow key to move
the highlight to Instrument
Recovery System, then press
3. When the Instrument Recovery
System has booted, follow the
on-screen instructions to recover
the image of the C: drive.
— Press
2, then press Enter to select
the recovery.
Enter.
After the Keysight Technologies screen is
displayed,
This screen is displayed for five seconds.
4. A warning message appears.
— Press
Enter to start the recovery,
which may take up to 25 minutes to
complete.
64 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Instrument Operating System
Disk Drive Recovery Process
Table 3-4 Recovery Process
StepNotes
5. Press
Enter to exit and reboot the
instrument once this portion of the
recovery has completed.
Additional recovery steps may be required to fully recover the system to a
more current working state. This could involve restoring your own
backups of the instrument information or re-installing applications, data
and performing system customizations.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 65
Instrument Operating System
Disk Drive Recovery Process
Updating the software (required after a recovery)
To install the latest software version, the software and installation instructions
are available at:
http://www.keysight.com/find/xseries_software
Table 3-5 Installing the Software
Step
1. Exit the program
2. Log out of the default user (instrument):
— Select the Start icon , select the
Sign out.
.,
3. At the log in prompt enter:
—User Name:
—Password: Keysight4u!
4. Follow the instructions on the software update
web page.
administrator
When you see the message Programming FPGAs...Do NOT turn
off power to the instrument, be sure to do as it says and DO
NOT turn off the instrument power at this time for ANY reason. If
this process is interrupted the instrument most likely will need to
be sent back to an Keysight Service Center for servicing before it
will be usable again.
Change account settings icon
The installation process can take up to 45 minutes. Do not turn the
instrument power off or serious damage may occur. If any pop up
windows appear, click OK or Ignore to proceed.
5. When the installation has finished, select
want to restart my computer now
66 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Yes, I
, Finish.
Instrument Operating System
Disk Drive Recovery Process
Table 3-5 Installing the Software
Step
6. After the instrument restarts, the newly installed
version of the X-Series instrument software will
run.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 67
Instrument Operating System
Disk Drive Recovery Process
68 Keysight NFA X-Series Getting Started and Troubleshooting Guide
The capabilities described in this section are Microsoft Windows 10
features. The discussion provided here gives some guidelines for using
the capabilities with the instrument. You need to refer to the Windows 10
help documentation for more information. Your version of Windows may
not match these instructions exactly.
You need an external keyboard and mouse to fully use these features.
“Navigating Windows Without a Mouse” on page 70
“Remote Desktop: Using the X-Series Instruments Remotely” on page 71
“Embedded Web Server: Using the X-Series Signal Analyzers Remotely” on
page 80
“Windows Shortcuts and Miscellaneous Tasks” on page 86
69
Using Windows Tools
Navigating Windows Without a Mouse
Navigating Windows Without a Mouse
Table 4-1
Key PressesActions
Esc
Enter
Alt
Right Arrow
Left Arrow
Up Arrow
Down Arrow
Tab
Del
Alt + Tab
Alt + Enter
Exits/closes a Windows dialog box (does not exit an Application window)
Does the current "default action". If a menu item or a button is currently
highlighted, then the Enter key activates that menu item or button.
Moves focus/control to the pull down menus bar in the active Window
In pull-down menu: opens the next menu to the right, or opens a submenu
In a dialog box: selects an option button
In pull-down menu: opens the next menu to the left, or opens a submenu
In dialog box: selects an option button
In pull-down menu: Moves to next selection up in the menu
In dialog box: selects an option button
In pull-down menu: Moves to next selection down in the menu
In dialog box: selects an option button
In dialog box: moves to the next/previous field
Deletes the currently selected item
Switches between the next/previous Application
Shows the Properties of the currently selected item
Alt + Esc
Backspace
Ctrl + Left arrow
Ctrl + Right arrow
Ctrl + Tab
Ctrl + Esc
Ctrl + Alt + Delete
70 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Cycles through items in the order that they had been opened
In My Computer or Windows Explorer: move up one level
In Internet Explorer: works like the BACK arrow key
Moves to the left one word at a time
Moves to the right one word at a time
In dialog box: moves to the next/previous Tab location
Opens the Windows Start Menu
Opens a window that enables you to select the Windows Task Manager
Using Windows Tools
Remote Desktop: Using the X-Series Instruments Remotely
Remote Desktop: Using the X-Series Instruments Remotely
Windows Remote Desktop is recommended for remote control of the
instrument. It offers fully-interactive control that is almost identical to direct
face-to-face control of the instrument. You can also remotely control the
instrument using the Embedded Web Server interface. The Embedded Web
Server functionality provides a communications method that does not require
login to the instrument. However, due to its slower response time, it is only
recommended for setup and data exchanges that do not involve instrument
control.
The Remote Desktop functionality is a Microsoft Windows 10 capability.
The following discussion provides some guidelines for using this
capability with the instrument. You need to refer to the Windows 10 help
documentation for more information. As Windows evolves, these
instructions may no longer be exact.
You need an external keyboard and mouse to fully use this functionality.
Overview of Remote Desktop operation
Using the Remote Desktop functionality of the instrument allows you to control
and interact with the instrument from a remote computer as though you were
sitting in front of the instrument.
When you have configured the instrument for remote connectivity, and
configured a separate computer to act as a Remote Desktop Host, you can
send commands to the instrument from the remote computer, and you can see
the instrument display on the screen of the remote computer.
This section provides full details of how to set up the instrument for remote
connectivity, and also how to set up a computer running any version of
Microsoft Windows as a Remote Desktop Host.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 71
Using Windows Tools
Remote Desktop: Using the X-Series Instruments Remotely
Setting up Remote Desktop operation
Figure 4-1 Basic setup for remote desktop operation
To perform this operation successfully, you must have Administrator level
access to the instrument.
Table 4-2 Setting up a remote desktop connection
StepAction
1. On the instrument, open the
Windows Control Panel
2. Select System functions— From the Adjust your computers
3. Access Remote setting— In the Control Panel Home window,
4. Select the Remote option— On the Remote tab, in the Remote
5. To add users
6. Follow the on screen instructions.
— From the instrument application,
press
System, Control Panel,
settings menu, click
Remote settings
select
Desktop section, click the
appropriate checkbox.
— Click
Select Users, Add.
System.
Setting up the remote computer
The procedure depends on whether the Remote Computer to be set up is
running Windows 10, or another version of Microsoft Windows.
Remote computer running Windows 10
Windows 10 includes the Remote Desktop Connectivity Client software, so no
additional setup is required.
72 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Using Windows Tools
Remote Desktop: Using the X-Series Instruments Remotely
Remote computer running another version of Windows
You can use any version of Windows to install and run the Client software for
Remote Desktop Connectivity. However, you need to have available a Windows
installation CD-ROM, because that contains the Client software.
The following instructions relate to software provided by Microsoft
Corporation. Keysight offers no warranty regarding the operation of such
software. The procedure described here may be changed by Microsoft at
some future time.
Table 4-3 Installing the Client software
StepNotes
1. When the Welcome Screen appears,
click
Perform additional tasks
2. From the What do you want to do?
screen, click
Desktop Connection
3. Click
4. To access the installed software, click
Next.
Set up Remote
.
Start > All Programs > Accessories
>
Communications > Remote
Desktop Connection
.
The Remote Desktop Connection InstallShield
Wizard appears.
Follow the on screen instructions provided by
the Wizard.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 73
Using Windows Tools
Remote Desktop: Using the X-Series Instruments Remotely
How to locate the computer name of the instrument
To connect a remote computer to the instrument, you need to know its
Computer Name. The Computer Name can be displayed as follows:
Table 4-4 Locating the name from the Keysight application
StepNotes
On the instrument front panel, press
Show, System.
System,
A page listing various parameters appears.The
instrument’s computer name is shown in the
list next to the title Computer Name.
Table 4-5 Locating the name from the Windows desktop (with a mouse):
StepNotes
1. Click
2. Double-click
Start, Control Panel.
System
The Computer name is listed in the Computer
name, domain, and workgroup settings
section.
74 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Using Windows Tools
Remote Desktop: Using the X-Series Instruments Remotely
Running a Remote Desktop session
Initializing a Remote Desktop session
To initialize a Remote Desktop Session, you need to know the Computer
Name of the instrument.
After setting up the remote computer for Remote Desktop Connectivity, as
described in “Setting up Remote Desktop operation” on page 72, you are
ready to start a Remote Desktop session.
The default account name is Instrument and the
default password is measure4u, but these
parameters may be changed by instrument users.
Only the current User or an Administrator can remotely log into the
instrument. To see who the current user of the instrument is, press
Ctrl+Esc on the instrument until you can view the current user name on
the Start menu. If no one is currently logged into the instrument, any valid
instrument user can remotely log in.
The instrument display appears on the screen of the remote computer.
Because the instrument front-panel keys are not available when using the
instrument remotely, three alternative methods are available to perform the
functions of the front-panel keys.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 75
Using Windows Tools
Remote Desktop: Using the X-Series Instruments Remotely
Windows Remote Desktop Options
Table 4-7 Setting Remote Desktop options
StepNotes
1. On the Remote Desktop
Connection menu, click
Options.
2. Under the
that the Computer name, User
name and Domain name are set
correctly.
General tab, ensure
The Options dialog has several tabs. Generally, the
default settings are correct.
You may choose to enter the password and save it for
future sessions, by checking the
password
box.
Save my
76 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Using Windows Tools
Remote Desktop: Using the X-Series Instruments Remotely
Table 4-7 Setting Remote Desktop options
StepNotes
3. Click the
— Under
desktop size
select the size of the
window in which the
instrument display
appears. Do not select
any size smaller than
1280 x 800 pixels.
Selecting a remote
desktop size smaller
than 1280 x 800 results
in some of the items on
the instrument display
not being fully visible. In
such circumstances,
scroll bars do not
appear, so portions of
the display are not
accessible.
— Under
to 16 bits If you operate
Remote Desktop with
greater color depth (e.g.,
32-bit) your windows
may have a different
appearance than they
do on the instrument
display because
transparency becomes
enabled.
Display tab.
Remote
, you may
Colors, set this
Keysight NFA X-Series Getting Started and Troubleshooting Guide 77
Using Windows Tools
Remote Desktop: Using the X-Series Instruments Remotely
Table 4-7 Setting Remote Desktop options
StepNotes
4. Click the
tab.
— Selecting the
Drives
enables you to transfer
data between the
remote desktop and the
local PC.
—
To transfer data, click
Start on the task bar of
the remote computer,
then click
Computer
opens on the remote
computer and displays
the drives of both the
remote computer and
the local computer. You
can now copy and paste
between the two disk
drives.
5. Click the
Local Resources
Disk
checkbox
My
. Explorer
Experience tab.
To Optimize the performance of the Remote Desktop
session, choose the appropriate connection format
from the drop-down list.
78 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Table 4-8
Using Windows Tools
Remote Desktop: Using the X-Series Instruments Remotely
Ending a Remote Desktop session
There are two ways to disconnect the remote computer from the instrument,
ending the session:
StepNotes
1. Click the
or
2. When the remote desktop is full screen,
move the cursor to the bottom left of
the window:
X, then click OK.
— Click
— Click
Start, Disconnect.
Disconnect.
For full-screen, the X appears at the top center of
the window.
For non full-screen, the X appears in a red box at
the right of the window’s title bar.
You are asked to confirm that you want to
disconnect.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 79
Using Windows Tools
Embedded Web Server: Using the X-Series Signal Analyzers Remotely
Embedded Web Server: Using the X-Series Signal Analyzers
Remotely
The instrument can be controlled using either the Embedded Web Server or
Windows Remote Desktop. The Embedded Web Server is a good solution
when you do not want to log into the instrument's user account. This allows
you to view the display or control the instrument, without logging the current
user off. Multiple users can connect at the same time
The web server in the X-Series instrument updates many times per second and
it has the advantage over Remote Desktop that it does not lock out the front
panel.
Drag gestures do not work in web control, whereas they do in Remote
Desktop.
Browsers that support X-Series Embedded Web Server
(these versions or later)
IEChromeSafariFireFoxiOS SafariChrome for
1135834839
Accessing the instrument through the Internet
It is possible to access and control the instrument through the Internet and
World Wide Web, or a local internet, using the built-in Embedded Server
functionality. This section provides details of how to use this functionality.
The instrument may also be accessed and controlled using the Windows
Remote Desktop functionality (see the section “Remote Desktop: Using the
X-Series Instruments Remotely” on page 71, for details).
The instrument Embedded Server capabilities are fully compliant with the LXI
(LAN eXtensions for Instrumentation) standard.
To gain access to the instrument from the LAN, you need to know its
hostname (or IP Address). For details of how to locate this information
using the instrument Display, see “How to locate the computer name of
the instrument” on page 74
.
Android
80 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Using Windows Tools
Embedded Web Server: Using the X-Series Signal Analyzers Remotely
Table 4-9 Accessing the instrument
StepNotes
1. Enter a URL corresponding to the
instrument hostname or IP Address.
This functionality is only
fully supported when using
Internet Explorer.
In this example, the host
name is
“a-n8973b-00104”.
When the connection is
made, the welcome page
appears.
2. Click
Keysight NFA X-Series Getting Started and Troubleshooting Guide 81
More Information
This displays further settings and configuration
information.
Using Windows Tools
Embedded Web Server: Using the X-Series Signal Analyzers Remotely
Controlling the Instrument
Selecting this tab enables you to view, control and interact with the instrument
through the web server.
Table 4-10 Selecting the Control Instrument tab
StepNotes
1. Select
Control Instrument
A password entry dialog appears.
By default, this password is set at the factory as
“measure4u”. However, you may subsequently change the
password. (Select
Password
change the password.)
When the correct password has been entered, the Control
Instrument web page appears.
System, I/O Config, Web
Reset on the instrument front panel to
2. You can now enter new settings
as required.
82 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Using Windows Tools
Embedded Web Server: Using the X-Series Signal Analyzers Remotely
Enabling SCPI Control of the instrument
Selecting this tab enables you to control the instrument via SCPI.
Selecting the Get Image tab
Selecting the Get Image tab captures a screen image from the Instrument
display.
To capture a screen image using the web server, the Instrument
Application must be running.
The image is captured as a Portable Network Graphics (PNG) file, to the default
file name Screen.png. The image file can be saved to the client computer disk
drive, or copied to the Windows clipboard.
A typical screen capture image appears as follows:
Keysight NFA X-Series Getting Started and Troubleshooting Guide 83
Using Windows Tools
Embedded Web Server: Using the X-Series Signal Analyzers Remotely
Selecting the Transfer Data tab
Selecting the Transfer Data tab enables you to capture results from the
instrument’s currently active measurement. Depending on the current
measurement type, captured results consist of either Trace Data or
Measurement Results.
To capture data using the web server, the instrument application must be running.
The captured data is formatted as a Comma Separated Value (CSV) file, which
may be saved on the client computer’s disk drive, or opened with a
spreadsheet application such as Microsoft Excel, or imported into a database
application such as Microsoft Access.
A typical Transfer Data web page display is shown below:
Selecting the Save/Recall tab
Selecting this tab enables you to save and recall data from the instrument.
84 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Using Windows Tools
Embedded Web Server: Using the X-Series Signal Analyzers Remotely
Selecting the Docs/Links tab
Selecting this tab enables access links to the instrument documentation and
application software.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 85
Using Windows Tools
Windows Shortcuts and Miscellaneous Tasks
Windows Shortcuts and Miscellaneous Tasks
This section provides a list of windows shortcuts (key combinations) that are
useful when you operate the instrument without an attached mouse and
keyboard. (See also “Navigating Windows Without a Mouse” on page 70.)
Although these shortcuts are available in any Windows 10 system, they are not
commonly used when a mouse and keyboard are attached.
Windows shortcuts (key combinations)
You can use the following combinations of front panel keys to perform basic
windows tasks when using the instrument without an attached mouse and
keyboard.
Table 4-11 Windows shortcut key combinations
To do the following:Press:
Display the Windows Start Menu
Cycle through all open applications
Select the first menu of a menu bar
Move through menu headings
Open (drop down) a menu
Move through items in an expanded menu
Close the current menu selection
Cancel the current menu bar selection
Open an application’s control menu (usually the left-most menu
on the menu bar, starting with File)
In a dialog: move between tabs
In a dialog: move forward through dialog box items
In a dialog: move backward through dialog box items
In a dialog: open a list box
In a dialog list box or check box: select or deselect one item at a
time
Ctrl+Esc
Alt+Tab
Alt
Left Arrow, Right Arrow
Down Arrow
Up Arrow, Down Arrow
Esc
Alt
Alt+Down Arrow
Ctrl+Tab
Tab
Shift+Tab
Alt+Down Arrow
Shift+Up Arrow, Shift+Down
Arrow
In My Computer, expand a selected folder
In My Computer, open a folder one level up from the current
folder
86 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Enter
Bk Sp
Using Windows Tools
Windows Shortcuts and Miscellaneous Tasks
Windows taskbar auto-hide feature
The Windows taskbar should always be in the auto-hide mode when using the
instrument application. If the taskbar is not set to auto-hide, the lower part of
the instrument display is obscured by the taskbar.
If at any time the Windows taskbar is inadvertently set to the non-auto-hide
mode, you can restore the auto-hide behavior by doing the following:
Table 4-12 Restoring taskbar auto-hide mode
StepNotes
1. Click
2. Click Taskbar and Start Menu
3. Click the
4. Select the
Start > Control PanelIf not using a mouse, press Ctrl+Esc.
Taskbar tab
Auto-hide the taskbar
check box
If you are not using a mouse, use the
shortcut key combinations specified in the
Section
combinations)” on page 86
these selections.
The Taskbar and Start Menu Properties
dialog appears.
If you are not using a mouse, press Tab
repeatedly until the auto-hide option is
selected, then press
check box state.
“Windows shortcuts (key
to make
Select to toggle the
5. Click
Keysight NFA X-Series Getting Started and Troubleshooting Guide 87
OK.
This applies the change and closes the
dialog box.
Using Windows Tools
Windows Shortcuts and Miscellaneous Tasks
88 Keysight NFA X-Series Getting Started and Troubleshooting Guide
No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock do not remove covers.
89
Troubleshooting
Check the Basics
Check the Basics
— Is there power at the receptacle?
— Is the analyzer turned on? Check to see if the green LED beside the power
— If other equipment, cables, and connectors are being used with your signal
— Is the Measurement Application running? If not, there is a software launch
— Does the instrument application have the focus? If not, move focus to the
switch is on. Also, listen for internal fan noise to determine if the analyzer
cooling fans are running.
analyzer, make sure they are connected properly and operating correctly.
If the analyzer cannot completely load or run the operating system, or the
instrument application is not successfully launched, the problem could be a
corrupt disk drive. If the analyzer gets far enough along in the boot process
to run the "Instrument Recovery System", perform the “Disk Drive Recovery
Process” as described on page 60.
shortcut/icon on the desktop.
application with Alt-Tab.
— Review the measurement procedures being performed when the problem
first appeared. Are all of the settings correct?
— If the analyzer is not functioning as expected, return the analyzer to a
known state by pressing Mode Preset.
Some analyzer settings are not affected by a Preset. If you wish to reset
the analyzer settings, press System, Power On, Restore Power On Defaults.
— Is the measurement being performed, and the results that are expected,
within the specifications and capabilities of the analyzer?
Refer to the specifications guide for your analyzer.
http://www.keysight.com/find/nfa_specifications
Technical manual pdf files are available on the Keysight website:
http://www.keysight.com/find/nfa_manuals
— If the analyzer is not communicating via the LAN connection, check for the
presence of blinking yellow LEDs on the rear panel LAN connector. If the
ACT LED is not blinking, check the LAN cable and LAN integrity.
— To meet specifications, the analyzer must be aligned. Either the Auto Align
(On) feature must be selected (press System, Alignments, Auto Align, and
select Normal), or the analyzer must be manually aligned.
— Perform an Alignment. Press System, Alignments, Align Now All.
90 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Troubleshooting
Check the Basics
— If the previously performed alignments did not resolve the problem, press
System, Alignments, Restore Align Defaults. Then press System,
Alignments, Align Now All.
— If the analyzer exhibits large amplitude errors (> 10 dB) especially at
frequencies above 10 GHz, the RF preselector might not be properly
centered. Press Peak Search, AMPTD Signal Path, Presel Center. If the
signal amplitude error is corrected, the preselector characterization should
be performed. Press System, Alignments, Ad vanced, Characterize Preselector. The characterization will take several minutes and the analyzer
must not be interrupted during this time. If the analyzer is interrupted
during the characterization process, the characterization data will be
destroyed and it will be necessary to perform the entire process again.
— Is the analyzer displaying an error message? If so, refer to the Instrument
Messages Guide.
— Check if the external frequency reference is selected but not available.
Verify that it is selected by pressing Input/Output, Freq Ref In. If External is
selected, changing the setting to Sense allows the analyzer to sense the
presence of an external reference and use it only if it is available. The
frequency of the reference should be set correctly.
— If you are using a Windows program other than the instrument application,
you may notice it running slow. Place the instrument application in single
sweep/measurement.
TIPYou can get automatic electronic notification of new software releases and
other product updates/information by subscribing to the Keysight
Techno logie s Te st & Measurement E-Mail N otificatio n S ervice for your signal
analyzer at:
http://www.keysight.com/find/MyKeysight
Keysight NFA X-Series Getting Started and Troubleshooting Guide 91
Troubleshooting
Problems with Microsoft Windows 10
Problems with Microsoft Windows 10
The Microsoft Windows 10 operating system settings have been optimized for
the best performance. Modification of these settings may degrade instrument
performance and measurement speed. Those that can be safely modified are
described in “Settings that can be changed” on page 51
The X-Series Signal Analyzers operate in an open Windows environment, so
you can install software on the instrument. However, installation of
non-approved software may affect instrument performance. Keysight does not
warrant the performance with non-approved software installed.
92 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Troubleshooting
Returning an Analyzer for Service
Returning an Analyzer for Service
Calling Keysight Technologies
Keysight Technologies has offices around the world to provide you with
complete support for your analyzer. To obtain servicing information or to order
replacement parts, contact the nearest Keysight Technologies office listed
below. In any correspondence or telephone conversations, refer to your
analyzer by its product number, full serial number, and software revision.
Press System, Show System, and the product number, serial number, and
software revision information will be displayed on your analyzer screen. A serial
number label is also attached to the rear panel of the analyzer.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 93
Troubleshooting
Returning an Analyzer for Service
Read the Warranty
The warranty for your analyzer is in the front of your Specifications Guide.
Please read it and become familiar with its terms.
If your analyzer is covered by a separate maintenance agreement, please be
familiar with its terms.
Service Options
Keysight Technologies offers several optional maintenance plans to service
your analyzer after the warranty has expired. Call your Keysight Technologies
office for full details.
If you want to service the analyzer yourself after the warranty expires, you can
download the service documentation that provides all necessary
troubleshooting and maintenance information from the Keysight web page.
Performance Verification and Adjustment tests require the N7881A X-Series
Signal Analyzer Calibration Application Software. Information regarding the
N7881A X-Series Signal Analyzer Calibration Application Software can be
found at:
http://www.keysight.com/find/calibrationsoftware
94 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Troubleshooting
Returning an Analyzer for Service
Packaging the Instrument
Use original packaging or comparable. It is best to pack the unit in the original
factory packaging materials if they are available.
Analyzer damage can result from using packaging materials other than
those specified. Never use styrene pellets in any shape as packaging
materials. They do not adequately cushion the equipment or prevent it
from shifting in the carton. They cause equipment damage by generating
static electricity and by lodging in the analyzer louvers, blocking airflow.
You can repackage the analyzer with commercially available materials, as
follows:
Table 5-1
StepNotes
1. Wrap the analyzer in anti-static
plastic to reduce the possibility
of damage caused by
electrostatic discharge
2. Use a strong shipping container. The carton must be both large enough and strong
enough to accommodate the analyzer. A
double-walled, corrugated cardboard carton with 159
kg (350 lb) bursting strength is adequate. Allow at
least 3 to 4 inches on all sides of the analyzer for
packing material.
3. Surround the equipment with
three to four inches of packing
material and prevent the
equipment from moving in the
carton.
4. Seal the shipping container
securely with strong nylon
adhesive tape.
5. Mark the shipping container
“FRAGILE, HANDLE WITH CARE” to assure careful
handling.
6. Retain copies of all shipping
papers.
If packing foam is not available, the best alternative is
plastic bubble-pak. This material looks like a plastic
sheet filled with 1-1/4 inch air bubbles. Use the
pink-colored bubble which reduces static electricity.
Wrapping the equipment several times in this material
should both protect the equipment and prevent it from
moving in the carton.
Keysight NFA X-Series Getting Started and Troubleshooting Guide 95
Troubleshooting
Returning an Analyzer for Service
96 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Index
Index
A
Auto-hide Taskbar, 87
B
battery replacement, 19
C
C drive usage, 61
Computer Name, 76
computer name
how to locate it, 74
connectors, front panel, 34
contents of shipment, 10
Control Panel
taskbar, 87
controlling Windows without a
mouse, 70
cooling the instrument, 11
D
D drive data storage, 61
data storage, 61
disk drive recovery process, 63
display capture, 83
Domain Name, 76
drive C use, 61
drive D use, 61
drive recovery, 63
E
electrostatic discharge (ESD)
protecting against, 20
Embedded Web Server
Get Data Page, 84
Get Data Tab, 84
Get Image Page, 83
Get Image Tab, 83
Web Control Page, 83, 84, 85
embedded web server, 80
Enabling SCPI Control