No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation into
a foreign language) without prior agreement
and written consent from Keysight Technologies, Inc. as governed by United States and
international copyright laws.
Manual Part Number
N9322-90002
Edition
First Edition, July 2014
Printed in China
Keysight Technologies, Inc.
No 116 Tianfu 4th Street
Chengdu 610041, China
Software Revision
This guide is valid for revision A.04.20 of the
Keysight N9322C Spectrum Analyzer
firmware.
Warranty
The material contained in this document is provided “as is,” and is subject
to being changed, without notice, in
future editions. Further, to the maximum extent permitted by applicable
law, Keysight disclaims all warranties,
either express or implied, with regard
to this manual and any information contained herein, includ ing but not limited
to the implied warranties of merchantability and fitness for a particular purpose. Keysight shall not be liable for
errors or for incidental or consequential
damages in connection with the furnishing, use, or performance of this
document or of any information contained herein. Should Keysight and the
user have a separate written agreement
with warranty terms covering the material in this document that conflict with
these terms, the warranty terms in the
separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software
and technical data rights granted to the federal government include only those rights
customarily provided to end user customers.
Keysight provides this customary commercial
license in Software and technical data pursuant to FAR 12.211 (Technical Data) and
12.212 (Computer Software) and, for the
Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Computer
Software Documentation).
Safety Notices
CAUTION
A CAUTION notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in personal
injury or death. Do not proceed
beyond a WARNING notice until the
indicated cond itions are fully
understood and met.
1Overview1
N9322C at a Glance2
Front Panel Overview4
Display Annotations7
Rear Panel Overview9
Front and rear panel symbols10
2Getting Started11
Check the Shipment and Order List12
Options12
Safety Notice13
Power Requirements14
Power On and Check16
Environmental Requirements18
Contents
Electrostatic Discharge Protection21
BSA PC software22
Helpful Tips23
Perform Calibration23
Using an External Reference24
Enable an Options24
Firmware Upgrade25
IO Configuration25
Timed Power On/Off26
Power Preset Last26
Data Securities27
Probe Power Output27
1
Contents
3Functions and Measurements29
Making a Basic Measurement30
Using the Front Panel30
Presetting the Spectrum Analyzer31
Viewing a Signal33
Measuring Multiple Signals35
Comparing Signals on the Same Screen Using Marker Delta35
Comparing Signals not on the Same Screen Using Marker Delta37
Resolving Signals of Equal Amplitude39
Resolving Small Signals Hidden by Large Signals42
Measuring a Low-Level Signal44
Reducing Input Attenuation44
Decreasing the Resolution Bandwidth46
Trace Averaging47
Improving Frequency Resolution and Accuracy49
Tracking Drifting Signals51
Making Distortion Measurements53
Identifying Analyzer Generated Distortion53
Third-Order Intermodulation Distortion56
Stimulus Response Transmission59
Measuring Stop Band Attenuation of a Lowpass Filter61
Channel Scanner64
Top/Bottom N Channel Scanner64
List N Channel Scanner66
Channel Scanner Setup68
Spectrum Monitor69
2
Reflection Measurement71
Preparation71
Measuring Cable Reflection72
Measuring Distance to Fault72
Making a Power Measurement with USB Power Sensor74
Making an Average Power Measurement75
Making a Peak Power Measurement79
Power Meter Settings81
Demodulate the AM/FM signal84
Demodulating an AM Signal84
Demodulating an FM Signal85
Analysis the Modulated Signals87
AM/FM Modulation Analysis87
ASK/FSK Modulation Analysis90
Contents
Measuring Channel Power93
Files Operation95
4Key Reference101
Amplitude102
BW106
Display109
Enter110
ESC <-110
Frequency111
Marker114
Meas119
3
Contents
MODE130
Peak Search148
System151
Span154
Channel Power119
Occupied BW121
ACPR123
Spectrum Emission Mask124
Tracking Generator130
Power Meter133
Spectrum Monitor136
Reflection Measurement138
AM/FM Modulation Analysis140
ASK/FSK Modulation Analysis144
Sweep155
Limit159
Trace162
5Instrument Messages167
Overview168
Command Errors169
Execution Conflict171
Device-Specific Errors173
6Troubleshooting177
Check the basics178
Contact Keysight Technologies180
4
7Menu Maps181
Amplitude Menu182
BW/Avg Menu183
Det/Display Menu184
File Menu (1 of 2)185
File Menu (2 of 2)186
Frequency Menu187
Marker Menu188
Marker-> Menu189
Measure Menu (1 of 2)190
Measure Menu (2 of 2)191
MODE - Tracking Generator192
MODE - Power Meter193
Contents
MODE - AM/FM Modulation Analysis194
MODE - ASK/FSK Modulation Analysis195
Name Editor Menu196
Peak Search Menu197
Preset/System Menu198
SPAN Menu199
Sweep/Trig Menu200
Trace Menu201
Index203
5
Contents
6
Overview
1Overview
Keysight N9322C at a Glance2
Front Panel Overview4
Rear Panel Overview9
Front and rear panel symbols10
This chapter provides a description of the Keysight N9322C spectrum
analyzer and an introduction to the buttons, features, and functions of
the front and rear instrument panels.
1
Overview
N9322C at a Glance
N9322C at a Glance
Features
The Keysight N9322C spectrum analyzer is a portable, swept
spectrum
7.0 GHz. It can be widely used in an electronic
manufacturing environment and in functional/final/QA test
systems.
The Keysight N9322C spectrum analyzer primary features
and functions are described below:
analyzerwithafrequencyrangeof9kHzto
• Power Measurement Suite
The built- in one- button power measurement suite offers
channel power, ACP, OBW, and TOI measurements, and
contains a Pass/Fail testing capability with a mask for
out- of- channel emissions measurement.
• Precision Frequency Reference (Option PFR)
The optinal OCXO frequency reference provides the precise
frquency measurement capability.
• Spectrum Monitor (Option MNT)
Provides the capability to analyze the stability of a signal
over time. The analyzer can be used to monitor the signal
capturing performance or intermittent event over extended
periods of time.
• High Sensitive Measurement (Option P07)
Includes an optional pre- amplifier for signals in the
frequency range up to 7 GHz, enabling more sensitive
measurements. This feature is a great help in analysis of low
level signals.
• High Accuracy Power Measurement
Supports Keysight U2000 series (option PWM) and U2020
series (option PWP) power sensors for high accuracy power
measurement as a power meter.
2
Overview
N9322C at a Glance
• Modulation Analysis Function
provides optional AM/FM and ASK/FSK modulation analysis
function. (AM/FM: Option- AMA ASK/FSK: Option- DMA)
• Tracking Generator (Option TG7)
provides an RF source for scalar network analysis.
• Baseband Channel (Option BB1)
Provides superior DANL and SSB between 9 kHz to 10 MHz.
• Cable & Antenna Test (Option RM7; Requires option TG7)
Provides VSWR, return loss, cable loss and distance- to- fault
function for reflection measurement.
• Task Planner (Option TPN)
Provides task planner function to integrate different
measurements for test automation.
• Time-gated Spectrum Analysis (Option TMG)
Measures any one of several signals separated in time and
xxclude interfering signals.
• Channel Scanner (Option SCN)
provide the channel scan funtion in spectrum monitoring,
coverage test, and band clearance.
Optimized Usability
The analyzer provides the enhanced usability as below:
• Socket/Telnet remote control via USB, and LAN port.
• The BSA PC Software on help kit CD provides further editing
and data analysis functions.
• Keypad can be locked and unlocked with password to
forbid undesired keypad operation.
• The GPIB interface (option G01) make the unit easily
compatible with your GPIB test system.
• Security feartures (Option SEC) allows user to erase all
customized files and data in analyzer for security.
3
Overview
Front Panel Overview
Front Panel Overview
BWTrace/
Marker
8
VWX
596
MNO
DEF
·
40MHz25dBm
6
Preset
Disp
UserDet
ESC
Enter
Local
PROBE POWER
System
Limit
Sweep
Save
Shift
File
RF IN 50
50VDC Max
33dBm (2W) Max
12
Peak
Search
MeasMode
YZ_
PQR
3
GHI
13
2
3
4
1
N9322C 9kHz - 7.0 GHz
Spectrum Analayz er
19
18
17
16
Remot e
Standby
On
15
5
Frequency
Span
Amplitude
7
STU
4
JKL
102
ABC
TG SOURCE CAL OUT
Rev Pwr 50VDC Max
+30dBm (1W) Max
14
1Screen 6.5 inch TFT color screen shows information of the
current function, including the signal traces, status
indicators, and instrument messages. Labels for softkeys are
located on the right- hand side of the screen.
2Softkeys are the unlabeled keys next to the screen. They
activate functions displayed to the left of each key.
3Amplitude activates the reference level function and
accesses the amplitude softkeys, with which you set
functions that affect data on the vertical axis.
7
8
9
10
11
4SPAN sets the frequency range symmetrically about the
center frequency. The frequency- span readout describes the
total displayed frequency range.
5Frequency activates the center- frequency function, and
accesses the menu of frequency functions.
6Function Keys relate directly to the following main functions:
• BW activates the resolution bandwidth function and
accesses the softkeys that control the bandwidth functions
and averaging.
• Trace/Det accesses the softkeys that allow you to set up
the trace settings and detector functions.
4
Front Panel Overview
• Disp - User accesses the submenu to recall the frequently
used user- defined state, or control what is displayed on
the analyzer, including the display line, graticule and
annotation after pressing Shift.
• Mode selects the measurement mode according to the
activated options.
• Marker accesses the marker control keys that select the
type and number of markers and turns them on and off.
• Limit - Sweepaccesses the softkeys that allow you to set
the sweep time, select the sweep mode and trigger mode,
or to set the limit functions after pressing Shift.
• Peak Search places a marker on the signal peak.
• Measaccesses the measurements mode selection submenu
in spectrum analyzer mode or the measurement submenu
in other optional modes.
7Utility Keys relate directly to the following utilities:
• Preset resets the analyzer to a known state
• System accesses the submenu to configure a series of
system settings, such as ports, option and languages.
• Shift switches alternate upper blue function of the
hardkeys.
• Save - File accesses the file operation submenu or activate
the quick- save function after pressing Shift.
Overview
8Arrow Keys The up and down arrow keys shift the selected
item when you press [MODE] hardkey; you can also change
the mode by rotating the knob.
9Knob The knob increases or decreases a value, a numeric
digit, or scrolls up and down to select an item in a list.
10Data Control Keys includes the numeric keypad, ENTER key
and backspace key, change the numeric value of an active
function such as center frequency, start frequency, resolution
bandwidth, and marker position. if the analyzer is in the
remote mode, pressing ENTER key returns the analyzer to the
local mode and enables front- panel control.
5
Overview
Front Panel Overview
11RF IN connector recieves the signal input. The maximum
damage level is average continuous power +33 dBm or DC
voltage 50 VDC. The impedance is 50 W. (N- type female).
12PROBE POWER connectorprovides power for high- impedance
AC probes or other accessories (+15 V, –12 V, 150 mA
maximum).
13CAL OUTconnectorprovides an amplitude reference signal
output of 40 MHz at –25 dBm (BNC female).
14TG SOURCE connectorN- type female, is the source output for
the built- in tracking generator. The maximum damage level
is average continuous power +30 dBm or DC voltage 50 VDC.
The impedance is 50 W. (for Option TG7)
15Standby Switch switches on all functions of the analyzer. To
switch the analyzer off, press the switch for at least 2
seconds. This deactivates all the functions but retains power
to internal circuits so long as the analyzer is connected to
line power.
16On LED (green) lights when the analyzer is switched on.
17Standby LED lights when the analyzer is connected to the
line power.
18Remote LED lights when the analyzer is remotely controlled
by a PC via the USB host interface on the rear panel.
19USB Device Connector provides a connection between external
USB devices and the analyzer, such as a USB memory
device.
6
Display Annotations
5
4
3
2
1
26
25
24
23
22
21
Overview
Front Panel Overview
6
7
8
10
9
11
12
13
1920
18
17
16
14
15
ItemDescriptionNotes (Associated function key)
1Average Type[BW] > Average Type
2Reference Offset[Amplitude] > Ref Offset
3Amplitude scale[Amplitude] > Scale Type
4Reference level[Amplitude]
5Display Line[Disp]
> Ref Level
> Display Line
6Time and date display[System] > Time/Date
7RF attenuation[Amplitude] > Attenuation
8Marker frequency[Marker]
9Marker amplitude[Marker]
10IndicatorS - Shift
R - Remote Control
11USB disk indicatorThe USB memory device is in use.
12Key menu titleDependent on key selection.
7
Overview
Front Panel Overview
13Softkey menuRefer to “Key Reference” for details.
Dependent on current function key selection.
14Frequency span[SPAN]
15Sweep time[Sweep] > Sweep Time
16Fequency offset[Frequency] > Freq Offset
17Video bandwidth[BW] > Video BW
18Display status lineDisplay status and instrument messages.
19Resolution bandwidth[BW] > Res BW
20Center Frequency[Frequency] > Center Freq
21Active function blockThe functino in use
22Low frequency channel[Frequency] > LowFrqChannel
23Sweep time rule[Sweep] > Sweep Setup > Swp Time Rule
24Sweep type[Sweep] > Sweep Setup > Sweep Type
25 - rightTrigger/Sweep
F - free run trigger
V - video trigger
E - external trigger
C - continuous sweep
S - single sweep
25 - leftInternal preamp[Amplitude] > Int Preamp
26Trace mode
W - clear write
M - maximum hold
m - minimum hold
V - view
S - store blank
Detect mode
N - Normal
A - Average
P - Positive peak
S - Sample
N - Neg Peak
[Sweep] > Trigge
[Trace]
[Trace] > Detector
8
Rear Panel Overview
1
SERIAL LABEL
ATTACH HERE
HIPOT LABEL
ATTACH HERE
LAN
TRIG IN10MHz
TTL
REF IN
10MHz
REF OUT
10
Overview
Rear Panel Overview
9
OCXO
REF OUT
10MHz
PC
100
~100-240
W MAX
V
Hz
50/60
8
2
3
5
4
6
7
1Kensington Lock lock the instrument and keep its safety.
2USB Host connector provides a connection between the
analyzer and an PC for remote control.
3LAN port A TCP/IP Interface used for remote operation.
4EXT TRG IN (TTL) connector accepts an external voltage input,
the positive edge of which triggers the analyzer sweep
function. (BNC female)
5REF IN connector accepts an external timebase with a
frequency of 10 MHz, amplitude of –5 to +10 dBm. (BNC
female)
6REF OUT connector provides a frequency of 10 MHz, amplitude
of –10 dBm reference output. (BNC female)
7REF OUT connector(Option PFR) is an optional connector for
precision 10 MHz reference from OCXO.
8AC Power Receptacle accepts a three- pin line power plug.
9Power switch isolates the analyzer from the AC line power.
After switch on, the analyzer enters into standby mode and
the orange standby LED on the front panel lights.
10GPIB connector(Option G01) is an optional interface. This GPIB
connector supports remote operation.
9
Overview
Front and rear panel symbols
Front and Rear Panel Symbols
The instruction manual symbol: indicates that the user must refer to
specific instructions in the manual.
The CE mark: a registered trademark of the European Community.
ISM1-A
C
US
ICES/NMB-001
Shows that this is an Industrial Scientific and Medical Group 1
Class A product. (CISPR 11, Clause 4)
The CSA mark: a registered trademark of the Canadian Standards
Association International.
The ISM device complies with Canadian Interference- Causing
Equipment Standard- 001.
Cet appareil ISM est conforme à la norme NMB- 001 du Canada.
All Level 1, 2 or 3 electrical equipment offered for sale in
Australia and New Zealand by Responsible Suppliers must be
marked with the Regulatory Compliance Mark (RCM Mark).
marks the “on/standby” position of the switch.
indicates that the instrument requires AC power input.
indicates this product complies with the WEEE Directive(2002/96/EC)
marking requirements and you must not discard this equipment in
domestic household waste. Do not dispose in domestic household
waste.
indicates the time period during which no hazardous or toxic
substance elements are expected to leak or deteriorate during
normal use. Forty years is the expected useful life of the product.
KC mark: as Korea Certification. This equipment is Class A suitable
for professional use and is for use in electromagnetic environments
outside of the home.
10
2Getting Started
Check the Shipment and Order List12
Safety Notice14
Power Requirements15
Power On and Check17
Environmental Requirements19
Electrostatic Discharge Protection22
Helpful Tips23
Getting Started
This chapter helps you in preparing the spectrum analyzer for use and
provides the information to start using the spectrum analyzer correctly.
11
Getting Started
Check the Shipment and Order List
Check the Shipment and Order List
After receiving the shipment, first check the shipment and
your order list. Inspect the shipping container and the
cushioning material for signs of stress. Retain the shipping
materials for future use, as you may wish to ship the
analyzer to another location or to Keysight Technologies for
service. Verify that the contents of the shipping container
are complete.
Shipping Problems?
If the shipping materials are damaged or the contents of the
container are incomplete, please contact the nearest Keysight
Technologies office to arrange for repair or replacement. You
will not need to wait for a claim settlement. If you must
return an analyzer to Keysight Technologies, use the original
(or comparable) shipping materials.
Options
12
Verify if that the shipment includes your ordered options by
checking the option label on the shipping container.
Unless specified otherwise, all options are available when
you order a spectrum analyzer; some options are also
available as kits that you can order and install/activate after
you receive the analyzer. Order kits through your local
Keysight Sales and Service Office.
At the time of analyzer purchase, options can be ordered
using your product number and the number of the option
you are ordering. For example, if you are ordering option
TG7, you would order N9322C-TG7.
If you are ordering an option after the purchase of your
analyzer, you will need to add a K to the product number
and then specify which option you are ordering. For
example, N9322CK- TG7.
Safety Notice
Getting Started
Safety Notice
Read the following warnings and cautions carefully before
powering on the spectrum analyzer to ensure personal and
instrument safety.
WARNING
WARNING
WARNING
CAUTION
CAUTION
Always use a well-grounded, three-pin AC power cord to connect to
power source. Personal injury may occur if there is any interruption of
the AC power cord. Intentional interruption is prohibited. If this product
is to be energized via an external auto transformer for voltage
reduction, make sure that its common terminal is connected to a
neutral (earthed pole) of the power supply.
Personal injury may result if the spectrum analyzer covers are removed.
There are no operator-serviceable parts inside. To avoid electrical
shock, refer servicing to qualified personnel.
Electrical shock may result if the spectrum analyzer is connected to the
power supply while cleaning. Do not attempt to clean internally.
Prevent damage to the instrument and ensure protection of the input
mixer by limiting average continuous power input to +33 dBm, DC voltage
to 50 VDC with >
if these precautions are not followed.
To install the spectrum analyzer in other racks, note that they may
promote shock hazards, overheating, dusting contamination, and inferior
system performance. Consult your Keysight customer engineer about
installation, warranty, and support details.
10 dB input attenuation. Instrument damage may result
13
Getting Started
Power Requirements
Power Requirements
AC Power Cord
The spectrum analyzer has an auto- ranging line voltage
input. The AC power supply must meet the following
requirements:
Voltage:100 to 240 VAC
Frequency:50 to 60 Hz
Power:Maximum 100 W
The analyzer is equipped with a three- wire power cord, in
accordance with international safety standards. This cable
grounds the analyzer cabinet when connected to an
appropriate power line outlet. The cable appropriate to the
original shipping location is included with the analyzer.
14
Various AC power cables are available that are unique to
specific geographic areas. You can order additional AC power
cables for use in different areas. The table AC Power Cords
lists the available AC power cables, the plug configurations,
and identifies the geographic area in which each cable is
appropriate.
The detachable power cord is the product disconnecting
device. It disconnects the mains circuits from the mains
supply before other parts of the product. The front panel
switch is only a standby switch and do not disconnect
instrument from LINE power.
AC Power Cords
Getting Started
Power Requirements
Plug TypeCable Part
Number
8121-1703BS 1363/AOption 900
250V 10A
8120-0696AS 3112:2000Option 901
250V 10A
8120-1692IEC 83 C4Option 902
250V 16A
8120-1521CNS 10917-2
125V 10A
8120-2296SEV 1011Option 906
250V 10A
8120-4600SABS 164-1Option 917
230V 15A
8120-4754JIS C8303Option 918
125V 15A
8120-5181SI 32Option 919
a
Plug
Description
/NEMA 5-15P
For use in
Country & Region
United Kingdom, Hong Kong,
Singapore, Malaysia
Australia, New Zealand
Continental Europe, Korea,
Indonesia, Italy, Russia
Option 903
United States, Canada, Taiwan,
Mexico
Switzerland
South Africa, India
Japan
Israel
250V 16A
8120-8377GB 1002Option 922
China
250V 10A
a. Plug description describes the plug only. The part number is for the complete cable assembly.
15
Getting Started
Power On and Check
Power On and Check
1Connect the AC power cord into the instrument. Insert the
2Press the AC line switch on the rear panel. The standby
plug into a power socket provided with a protective ground.
Set the tilt adjustors for your preference.
Figure 1Plug in and angle adjustment
LED on the front panel will light and the spectrum analyzer
is in standby mode (AC power applied).
16
NOTE
3Press the standby switch on the front panel. The On LED
will light, and the spectrum analyzer boots up.
Self- initialization takes about 25seconds; the spectrum
analyzer then defaults to the menu mode. After power on,
let the spectrum analyzer warm up for 45 minutes for
stabilization.
The front panel switch is a standby switch only; it is not a power
switch. To completely disconnect the spectrum analyzer from the AC
line power, shut off the power switch on the rear panel.
Check for Instrument Messages
The spectrum analyzer has two categories of instrument
messages: error and warning messages. A error message is
triggered by operation errors, for example, parameter setting
conflicts or data input that is out of the range of a
parameter. An warning message may be triggered by
hardware defects which could result in damage to
instrument.
Here are some tips to check the instrument messages.
1Check the display to see if any messages display in the
status bar. Press [System] > System Info > Error History to review
each message. Refer to Chapter 5, “Instrument Messages” for
detailed system message descriptions.
2When you have reviewed and resolved all of the error
messages, press [System] > System Info > Error History > Clear
error queue to delete the messages.
3Cycle the power to the analyzer and re-check to see if the
instrument messages are still there.
Getting Started
Power On and Check
4If the error messages cannot be resolved, please contact the
Keysight Customer Contact Center for assistance or service.
17
Getting Started
Environmental Requirements
Environmental Requirements
Keysight Technologies has designed this product for use in
Installation Category II, Pollution Degree 2, per IEC 61010- 1.
Keysight has designed the spectrum analyzer for use under
the following conditions:
• Indoor use
• Altitude < 3,000 meters
• Operating temperature range: +5 to +45
Storage temperature range: –20 to +70
• Relative humidity range 15% to 95 % at 40
o
C;
o
C
o
C
NOTE
Ventilation
Cleaning Tips
This equipment is Class A suitable for professional use and is for use in
electromagnetic environments outside of the home.
Ventilation holes are located on the rear panel and one side
of the spectrum analyzer cover. Do not allow these holes to
be obstructed, as they allow air flow through the spectrum
analyzer.
When installing the spectrum analyzer in a cabinet, do not
restrict the convection of the analyzer. The ambient
temperature outside the cabinet must be less than the
maximum operating temperature of the spectrum analyzer by
o
C for every 100 watts dissipated within the cabinet.
4
To prevent electrical shock, disconnect the spectrum analyzer
from line power before cleaning. Use a dry cloth or one
slightly dampened with water to clean the external case
parts. Do not attempt to clean internally.
18
Rack Mount
Getting Started
Environmental Requirements
It is recommended to use the Keysight rackmount kit (option
1CM) to install the spectrum analyzer into a rack.
Do not attempt to rack mount the spectrum analyzer by the
front panel handles only. This rackmount kit will allow
mounting of the spectrum analyzer with or without handles.
Refer to the following instructions when installing the
rackmount kit on the spectrum analyzer.
1Remove feet, key- locks and tilt stands.
2Remove side trim strips and one middle screw per side.
19
Getting Started
Environmental Requirements
3Attach rackmount flange and front handle assembly with 3
screws per side.
20
CAUTION
4Attach the spectrum analyzer to the rack using the
rackmount flanges with two dress screws per side.
Installing the spectrum analyzers into other racks may promote shock
hazards, overheating, dust contamination, and inferior system
performance. Consult your Keysight customer engineer about installation,
warranty, and support details.
Electrostatic Discharge Protection
Electrostatic discharge (ESD) damages or destroys electronic
components (the possibility of unseen damage caused by
ESD is present whenever transported, stored, or while the
instrument is in use).
This product contains components that are easily damaged
by electrostatic discharge. To help reduce the possibility of
ESD damage that can occur while using test equipment,
follow these guidelines:
1Before connecting any coaxial cable to the spectrum analyzer
connector for the first time each day, momentarily short the
center and outer conductors of the cable together to
eliminate any potential electrostatic charges that may exist.
2Personnel should be grounded with an approved type, 1 MΩ
resistor- isolated ESD wrist- strap before touching the center
pin of any connector, and before removing any assembly
from the spectrum analyzer.
Getting Started
Electrostatic Discharge Protection
3Be sure that all instruments are properly earth grounded to
prevent build- up of static charges.
21
Getting Started
BSA PC software
BSA PC software
Keysight HSA PC software is an easy- to- use, PC- based
remote control tool for the N9322C spectrum analyzer.
It is able to be discretely used as a spectrum monitor to
display and control the trace scans simultaneously with the
analyzer, or a file manager to send/get files between the
anaylzer and PC. It also provides some data analysis
function for your further use.
22
For the further descrption of the BSA PC software, please
refer to the online help embedded in this software.
Helpful Tips
Perform Calibration
Getting Started
Helpful Tips
The following contains information to help in using and
maintaining the instrument for optimum operation, including
alignment, external reference, firmware update and option
activation.
The N9322C provides three manual calibration function to
calibrate the time base and amplitude. The analyzer should
warm up for 30 minutes before calibration.
Time Base Calibration
Perform a time base calibration to guarantee the frequency
accuracy. The analyzer should warm up for approximately
30 minutes before calibration.
When the calibration is triggered, the current measurement
is interrupted and a gauge displays on the LCD. The gauge
simply indicates calibration is in process. When it is
finished, the interrupted measurement restarts.
Please refer to the operation procedures as below:
1Input a 10 MHz, 0 dBn signal to EXT TRIG IN.
2Press [System] > Service > Calibration > Time Base by Ext to
initiate the calibration.
Time base calibration takes only a short time when the inner
temperature is stable. When the inner temperature is
increasing, calibration takes a long period of time or will
fail. If the input reference signal is abnormal, the calibration
cycle will take a long and unpredictable time to exit, and the
LCD displays an error message.
23
Getting Started
Helpful Tips
Amplitude Calibration
The analyzer privides the internal amplitude calibration
function. Please refer to the procedures below to perform an
amplitude calibration:
1 Press [System] > More > Service > Calibration > Amp Alignment
2 2 Connect the CAL OUT and RF IN connector with a
N- BNC cable. Then Press [Enter] to continue.
The calibration will start immediately. The calibration will be
completed in serveral minutes.
Using an External Reference
To use an external 10 MHz source as the reference
frequency, connect the external reference source to the REF IN connector on the rear panel. An EREF indicator will
display in the left side of the display. The signal level must
be in the range of –5 to +10 dBm. When the reference signal
is ready, the instrument switches the reference from time
base to the external reference.
24
Enable an Options
Option license key information is required to enable product
options. Contact your nearest Keysight office for purchasing
a license. Refer to the procedures below to activate the
options you have purchased.
1Press [System] > More > Service > Add Option
2Enter the option number to be enabled. Press [Enter] to
confirm your input.
3Enter the license key in the onscreen window. Press [Enter]
to confirm your input and terminate the license key input.
4Press Activate License. The option
The analyzer provides the trial license with limited time
usage(15/270/365 days) of the wanted options.
willbeenabled immediately.
Firmware Upgrade
Press [System] > System Info > Show system to view the
firmware revision. If you call Keysight Technologies
regarding your analyzer, it is helpful to have this revision
and the analyzer serial number available.
Follow this procedure to finish the firmware update:
2Plug the USB stick into the connector. Press [System] >
Service > Upgrade Firmware
The analyzer will perform the update process automatically.
The upgrade procedure will take several minutes. Once the
upgrade is completed, please follow the instruction to reboot.
Getting Started
Helpful Tips
CAUTION
Any interruption during the update process will result in update failure
and system data lost. Do not remove the USB storage device until the
update is finished.
IO Configuration
The N9322C spectrum analyzer provides three types of IO
connnection: USB, LAN and optional GPIB interface. Press
[System] > More > Port Setting to set the corresponding
interface as your need.
USB
Select USB to enable the USB connection for remote control.
LAN
The N9322C supports LAN port connection for remote
control. Press [System] > {More} > {Port Setting} > {IP Config} to
set the IP parameters for the network connection.
25
Getting Started
Helpful Tips
Press {IP Address Static} to manually set the IP address,
gateway and subnet mask with the proper LAN information,
or just press {IP Address DCHP} to set the IP address in LAN
dynamically according DCHP.
Press {Apply} to enable the LAN connection according to the
IP configuration you set.
GPIB
Press [System] > {More} > {Port Setting} > {GPIB Address} to set
the GPIB address for the analyzer. This softkey is only vaild
{Timed Pwr Off} sets the time switch to power on/off the
N9322C in a user- defined time and date. This function
requires the power supply connected.
Press {Repeat Mode Once/Everyday} to set the N9322C boot
up/off in the pre- saved time everyday. The pre- saved date is
invalid in this mode.
26
Power Preset Last
Press [System] > {Pwr On/Preset} > {Power On Last} to activate
this function. For the standard N9322C, this operation only
save the last state if the analyzer is turned off by the front
panel power switch.
Data Securities
The N9322C offers the optional memory erase function for
data security. Press [System] > {More} > {Securities} > {Erase Memory} to erase all the user data in internal memory. Press
Enter as a terminator to start the erase process immediately.
The memory erase process takes about 15 minutes. During
the time, there must be a constant power supply to ensure
the successful erase. If the erase process is interrupted,
please reboot the instrument and erase memory again.
Probe Power Output
The Probe Power provides power for high- impedance AC
probes or other accessories (+15 V, –12V, 150 mA maximum).
The Porbe Power is set to off as default. Press [System] > {More} > {Port Setting} > {Probe Power On} to switch on the
porbe power output.
Getting Started
Helpful Tips
27
Getting Started
South Korea Class A EMC Declaration
South Korea Class A EMC Declaration
This equipment is Class A suitable for professional use and
is for use in electromagnetic environments outside of the
home.
28
Functions and Measurements
3Functions and Measurements
Making a Basic Measurement30
Measuring Multiple Signals35
Measuring a Low-Level Signal44
Improving Frequency Resolution and Accuracy49
Tracking Drifting Signals51
Making Distortion Measurements53
Stimulus Response Transmission59
Measuring Stop Band Attenuation of a Lowpass Filter61
Channel Scanner64
Spectrum Monitor69
Reflection Measurement71
Making a Power Measurement with USB Power Sensor74
Demodulate the AM/FM signal84
Analysis the Modulated Signals87
Measuring Channel Power93
Files Operation95
This chapter provides information on the analyzer functions and
specific measurements capabilities of the spectrum analyzer.
29
Functions and Measurements
Making a Basic Measurement
Making a Basic Measurement
This section provides information on basic operations. It
assumes that you are familiar with the front and top panel
buttons and keys, and display annotations of your analyzer.
If you are not, please refer to “Front Panel Overview” on
page 4 and “Rear Panel Overview” on page 9.
For more details on making measurements, please refer to
“Key Reference” on page 101”.
Using the Front Panel
This section provides you with the information on using the
front panel of the spectrum analyzer.
Entering Data
When setting the measurement parameters, there are several
ways to enter or modify the value of the active function:
KnobIncrements or decrements the current value.
Arrow KeysIncrements or decrements the current value by a step unit.
30
Numeric KeysEnters a specific value. Then press the desired terminator
(either a unit softkey, or [Enter] hardkey).
Unit SoftkeysTerminate (enter) a value with a unit softkey from the menu.
Enter KeyTerminates an entry when no unit of measure is required, or
the instrument uses the default unit.
ESC KeyTo delete
the current input digit prior to entering the value.
Using Softkeys
Softkeys are used to modify the analyzer function parameter
settings. Some examples of softkey types are:
ToggleTurn on or off an instrument state.
SubmenuDisplays a secondary menu of softkeys, {More}.
ChoiceSelecting from a list of standard values or filenames.
AdjustHighlights the softkey and sets the active function.
Functions and Measurements
Making a Basic Measurement
Presetting the Spectrum Analyzer
Preset function provides a known instrument status for
making measurements. There are two types of presets,
factory and user:
Factory PresetWhen this preset type is selected, it restores the analyzer to
its factory- defined state. A set of known instrument
parameter settings defined by the factory.
User PresetRestores the analyzer to a user- defined state. A set of user
defined instrument parameter settings saved for assisting the
user in quickly returning to known a instrument
measurement setup.
Press System > Pwr on/Off Preset > Preset Type to select the
preset type.
When Preset Type is set to Factory, pressing Preset triggers a
factory preset condition. The instrument will immediately
return to the factory default instrument parameter setting.
When Preset Type is set to User, pressing Preset triggers a user
preset condition. The instrument will immediately return to
the user defined instrument parameter setting.
When Preset Type is set to Last, pressing Preset triggers a last
power- off condition. The instrument will immediately return
to the parameter setting when it is power off last time.
Setting up a User Preset
To quickly return to instrument settings that are user
defined, perform the following steps to save the instrument
state as the user- defined preset:
1Set the instrument parameters to the values and settings
necessary for the user preset state. This would include the
frequency, span, amplitude, BW, and measurement type and
any other setup details desired.
31
Functions and Measurements
Making a Basic Measurement
2Press System > Pwr on/off Preset> Save User, to save the
Attenuation20 dB(Auto)Average TypeLog Pwr
Center frequency3.5 GHzAverage Number100 (Off)
Reference offset0 dBPeak Excursion6 dB
Start Frequency0 HzPeak Threshold-140 dBm
Stop Frequency7 GHzPeak Search TypeMax
Hi-SensitivityOffContinuous PkOff
Span7 GHzGraticule On
Reference level0 dBmY ScaleOn
Log scale/Division10 dBCal OutOff
Int PreampOffProbe PowerOff
Trace 1Clear WriteTr i g gerFr e e Run
Trace 2, 3, 4BlankMarker ModeNormal
current instrument settings as the ‘user preset’ state. The
user preset will not affect the default factory preset settings.
User preset settings can be changed and saved at any time.
Figure 3-1 Factory Default Preset Setting
32
Sweep Time328.5 ms(Auto)Marker FunctionNormal
SweepContinuousMarker TableOff
Limit LineOffMeasureMeas Off
LimitOffFile CatalogInt
Impedance50 ΩFile SortBy Date
DetectorPos PeakGated Sweep Off
Display lineOffGate Delay12 μs
Video bandwidth3 MHz(Auto)Gate Length84 μs
VBW/RBW ratio1.000 (Auto)Gate SourceExternal
Resolution
bandwidth
3 MHz(Auto)Sweep Time RuleSpeed
Viewing a Signal
Refer to the procedures below to view a signal.
1Press System > Pow on/off Preset > Preset Type > Factory to enable
the factory- defined preset state.
2Press Preset to restore the analyzer to its factory- defined
state.
3Connect the 10 MHz REF OUT on the rear panel to the
front- panel RF IN.
Setting the Reference Level and Center Frequency
1Press Amplitude > 10 > dBm to set 10 dBm reference level.
2Press Frequency > 30 > MHz to set 30 MHz to center frequency.
Setting Frequency Span
Press Span > 50 > MHz to set 50 MHz frequency span.
Functions and Measurements
Making a Basic Measurement
NOTE
Changing the reference level changes the amplitude value of the top
graticule line. Changing the center frequency changes the horizontal
placement of the signal on the display. Increasing the span will
increase the frequency range that appears horizontally across the
display.
Reading Frequency and Amplitude
1Press Peak Search to place a marker (labeled 1) on the
10 MHz peak.
33
Functions and Measurements
Making a Basic Measurement
Note that the frequency and amplitude of the marker appear
both in the active function block, and in the upper- right
corner of the screen.
Figure 3-1 10 MHz Internal Reference Signal
Marker Annotation
Marker
10.000000 MHz
0.43 dBm
34
NOTE
2Use the knob, the arrow keys, or the softkeys in the Peak
Search menu to move the marker. The marker information
will be displayed in the upper- right corner of the screen.
Changing Reference Level
1Press Amplitude and note that reference level (Ref Level) is
now the active function.
2Press Marker
> Marker To > Mkr-> Ref Lvl.
Changing the reference level changes the amplitude value of the top
graticule line.
Measuring Multiple Signals
This section provides the information on how to measure
multiple signals.
Comparing Signals on the Same Screen Using Marker Delta
The delta marker function allows the user to compare two
signals when both appear on the screen at the same time.
In the following example, harmonics of the 10 MHz reference
signal available are used to measure frequency and
amplitude differences between two signals on the same
screen. Delta marker is used to demonstrate this
comparison.
1Preset the analyzer:
Press Preset (With Preset Type of Factory)
2Connect the rear panel REF OUT to the front panel RF IN.
3Set the analyzer center frequency, span and reference level
to view the 10 MHz signal and its harmonics up to 50 MHz:
Press Frequency > 30 > MHz
Functions and Measurements
Measuring Multiple Signals
Press Span > 50 > MHz
Press Ampulitude > 10 > dBm
4Place a marker at the highest peak on the display (10 MHz):
Press Peak Search
The marker should be on the 10 MHz reference signal. Use
the Next Right Pk and Next Left Pk softkeys to move the marker
from peak to peak.
5Anchor the first marker and activate a second marker:
Press Marker > Mode > Delta
The label on the first marker now reads 1R, indicating that it
is marking the reference point.
35
Functions and Measurements
Measuring Multiple Signals
6Move the second marker to another signal peak using the
front- panel knob or by using Peak Search.
Press Peak Search > Next Peak or
Press Peak Search > Next Right Pk or Next Left Pk. Continue
pressing the Next Pk softkeys until the marker is on the
correct signal peak.
The amplitude and frequency differences between the
markers are displayed in the active function block.
Figure 3-2 Delta pair marker with signals on the same screen
36
NOTE
To increase the resolution of the marker readings, turn on the
frequency count function. For more information, refer to
Frequency Resolution and Accuracy” on page 49
“Improving
.
Functions and Measurements
M
Measuring Multiple Signals
Comparing Signals not on the Same Screen
Using Marker Delta
The analyzer will compare the frequency and amplitude
differences between two signals which are not displayed on
the screen at the same time. (This technique is useful for
harmonic distortion tests.)
In this example, the analyzer’s 10 MHz signal is used to
measure frequency and amplitude differences between a
signal on screen, and another signal off screen. Delta marker
is used to demonstrate this comparison.
1Preset the analyzer:
Press Preset (With Preset Type set to Factory)
2Connect the rear panelREF OUTto the front panelRFIN.
3Set the center frequency, span and reference level to view
only the 50 MHz signal:
Press Frequency > 50 > MHz
Press Span > 25 > MHz
Press Amplitude > 10 > dBm
4Place a marker on the 50 MHz peak and then set the center
frequency step size equal to the marker frequency (10 MHz):
Press Peak Search
Press Marker > Marker To > Mkr -> CF Step
5Activate the marker delta function:
Press Marker > Mode > Delta
6Increase the center frequency by 10 MHz:
Press Frequency,
The first marker moves to the left edge of the screen, at the
amplitude of the first signal peak.
F
37
Functions and Measurements
Measuring Multiple Signals
Figure 3-3 shows the reference annotation for the delta
marker (1R) at the left side of the display, indicating that the
marker set at the 50 MHz reference signal is at a lower
frequency than the frequency range currently displayed.
Figure 3-3 Delta Marker with Reference Signal Off-Screen
The delta marker appears on the peak of the 100 MHz
component. The delta marker annotation displays the
amplitude and frequency difference between the 50 and
100 MHz signal peaks.
38
7Press Marker > Mode > Off to turn the markers off.
Resolving Signals of Equal Amplitude
In this example a decrease in the resolution bandwidth
(RBW) is used in combination with a decrease in video
bandwidth (VBW) to resolve two signals of equal amplitude
with a frequency separation of 100 kHz.
Figure 3-4 Setup for obtaining two signals
Functions and Measurements
Measuring Multiple Signals
Signal Generator
N9310A RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby
On
Spectrum Analyzer
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
Remote
Standby
On
7
4
102
TG SOURCE CAL OUT
RF IN
BW/
Auto
Sweep/
Avg
Tune
Trig
Local
View/
Det/
File/
Marker
Print
Trace
Display
Save
Peak
Marker
Meas
MODE
Search
5896
Back
3
Enter
·
50VDC MAX
30dBm 1W MAX
RF IN 50
50MHz10dBm
PROBE POWER
CAT Ⅱ
Frequency
Amplitude FM
Enter
7
4
102
RF OUT
Signal Generator
FUNCTION S
Utility
Preset
AM I/Q
Sweep
Local
File
LF Out
TriggerPulseM
MOD
9
8
On/Off
5
6
3
RF
·
On/Off
LF OUT RF OUT 50
REVERSE PWR
4W MAX30VDC
N9310A RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby
On
Frequency
Amplitude FM
Enter
7
4
102
RF OUT
FUNCTIONS
Utility
Preset
AM I/Q
Sweep
Local
File
LF Out
TriggerPulseM
MOD
9
8
On/Off
5
6
3
RF
·
On/Off
LF OUT RF OU T 5 0
REVERSE PWR
4W MAX 30VDC
Directional Coupler
Notice that the final RBW selection to resolve the signals is
the same width as the signal separation while the VBW is
slightly narrower than the RBW.
1Connect two sources to the analyzer input as shown above.
2Set one source to 300 MHz. Set the frequency of the other
source to
300.1MHz. Set both source amplitudes to –20 dBm.
3Setup the analyzer to view the signals:
Press Preset (With Preset Type of Factory)
Press Frequency > 300 > MHz
Press BW > 300 > kHz
Press Span > 2 > MHz
A single signal peak is visible. See
Figure 3-5 for example.
39
Functions and Measurements
Measuring Multiple Signals
If the signal peak is not present on the display, increase the
frequency span out to 20 MHz, turn signal tracking on,
decrease the span back to 2 MHz and turn signal tracking
off:
Press Span > 20 > MHz
Press Peak Search
Press Span > 2 > MHz
Figure 3-5 Unresolved Signals of Equal Amplitude
40
4Change the resolution bandwidth (RBW) to 100 kHz so that
the RBW setting is less than or equal to the frequency
separation of the two signals:
Press BW > 100 > kHz
Notice that the peak of the signal has become flattened
indicating that two signals are present.
5Decrease the video bandwidth to 3 kHz:
Press BW > Video BW > 3 > kHz
Two signals are now visible as shown in
front- panel knob or arrow keys to further reduce the
resolution bandwidth and better resolve the signals.
Figure 3-6. Use the
Functions and Measurements
Measuring Multiple Signals
Decreasing the resolution bandwidth improves the resolution
of the individual signals and increases the sweep time.
Figure 3-6 Resolving Signals of Equal Amplitude
For fastest measurement times, use the widest possible
resolution bandwidth. Under factory preset conditions, the
resolution bandwidth is coupled to the span.
41
Functions and Measurements
Measuring Multiple Signals
Resolving Small Signals Hidden by Large Signals
1Connect two sources to the RF IN as shown in
2Set one source to 300 MHz at –10 dBm. Set the other source
3Set the analyzer as follows:
4Set the 300 MHz signal to the reference level:
This example uses narrow resolution bandwidths to resolve
two RF signals that have a frequency separation of 50 kHz
and an amplitude difference of 60 dB.
Figure 3-4.
to 300.05 MHz at –70 dBm.
Press Preset. (With Preset Type of Factory)
Press Frequency > 300 > MHz
Press BW > 30 > kHz
Press Span > 500 > kHz
Press Peak Search
Press Marker > Mode > Mkr -> Ref Lvl
42
NOTE
The 30 kHz filter shape factor of 15:1 has a bandwidth of 450 kHz at the 60
dB point. The half-bandwidth (225 kHz) is NOT narrower than the
frequency separation of 50 kHz, so the input signals can not be resolved.
Figure 3-7 Unresolved small signal from large signal
Functions and Measurements
Measuring Multiple Signals
5Reduce the resolution bandwidth filter to view the smaller
signal. The smaller signal will be hidden by the larger signal
when the bandwidth settings are wider. Reducing the RBW
setting will allow less of the larger signal to pass through
the analyzer and the smaller signals peak will then rise out
of the noise floor. Place a delta marker on the smaller
signal:
Press BW > 1 > kHz
Press Peak Search
Press Marker > Mode > Delta
Press 50 > kHz
Figure 3-8 Resolved small signal from large signal
NOTE
The 1 kHz filter shape factor of 15:1 has a bandwidth of 15 kHz at the
60 dB point. The half-bandwidth (7.5 kHz) is narrower than the
frequency separation of 50 kHz, so the input signals can be resolved.
43
Functions and Measurements
Measuring a Low-Level Signal
Measuring a Low-Level Signal
This section provides information on measuring low-level
signals and distinguishing them from spectrum noise.
Reducing Input Attenuation
The ability to measure a low- level signal is limited by
internally generated noise of the spectrum analyzer. The
analyzers input attenuator affects the level of a signal
passing through the analyzer. If a signal power level is close
to the noise floor, reducing the analyzer input attenuation
will help raise the signal so that it can be seen rising out of
the noise.
CAUTION
Ensure that the total power of all input signals at the analyzer RF input
does not exceed +30 dBm (1 Watt).
1Preset the analyzer
Press Preset (With Preset Type of Factory)
2Set the source frequency to 300 MHz, amplitude to –70 dBm.
Connect the source RF OUT to the analyzer RF IN.
Figure 3-9 Setup for obtaining one signal
Signal Generator
N9310A RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby
On
Frequency
Amplitude
Enter
7
8
4
5
102
·
RF OUT
FUNCTION S
Utility
Preset
AM I/Q
File
Sweep
FM
Local
LF Out
TriggerPulseM
MOD
9
On/Off
6
3
RF
On/Off
LF OUT RF OUT 50
REVERSE PWR
4W MAX30VDC
Spectrum Analyzer
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
Remote
Standby
On
7
4
102
TG SOURCE CAL OUT
RF IN
BW/
Auto
Sweep/
Avg
Tune
Trig
Local
File/
View/
Det/
Marker
Print
Trace
Display
Save
Peak
Marker
MeasMODE
Search
5896
Back
3
Enter
·
50VDC MAX
30dBm 1W MAX
RF IN 50
50MHz10dBm
PROBE POWER
CAT Ⅱ
44
Functions and Measurements
Measuring a Low-Level Signal
3Set the center frequency, span and reference level:
Press Frequency > 300 > MHz
Press Span > 2 > MHz
Press Amplitude > 40 > –dBm.
4Move the desired peak to the center of the display:
Press Peak Search
Press Marker > Marker To > Mkr -> CF
5Reduce the span to 500 kHz, if necessary re- center the peak:
Press Span > 500 > kHz
6Set the attenuation to 20 dB:
Press Amplitude > Attenuation > 20 > dB
Figure 3-10 A signal closer to the noise level
Note that increasing the attenuation moves the noise floor
closer to the signal level.
7To allow more of the signal power to pass through the
analyzer, decrease the attenuation to 0 dB.
45
Functions and Measurements
Measuring a Low-Level Signal
A lower attenuation value will mean that more of the signal
strength will be displayed on screen:
Press Amplitude > Attenuation > 0 > dB
Figure 3-11 Measuring a low-level signal using 0 dB Attenuation
46
Decreasing the Resolution Bandwidth
Resolution bandwidth settings affect the level of internal
noise but have little affect on the displayed level of
continuous wave (CW) signals. Decreasing the RBW by a
decade (factor of ten) reduces the noise floor by 10 dB.
1Refer to the procedure
page 44
2Decrease the resolution bandwidth:
Press BW and
The low- level signal appears more clearly due to the noise
level being reduced by the decrease in RBW (see
and follow steps 1, 2 and 3.
FM
Figure 3-12).
“Reducing Input Attenuation” on
NOTE
Functions and Measurements
Measuring a Low-Level Signal
Figure 3-12 Decreasing Resolution Bandwidth
A “#” mark appears next to the Res BW annotation in the
lower left corner of the screen, indicating that the resolution
bandwidth is uncoupled.
Uncoupled indicates that the function is in manual control
mode, not auto control mode. Manual control mode allows
the user to change the parameter value for that function
without affecting any other settings.
The analyzer allows you to change the RBW in a 1-3-10 sequence by
the data control keys. The RBWs below 1 kHz are digital and have a
selectivity ratio of 5:1 while RBWs at 1 kHz and higher have a 15:1
selectivity ratio. The maximum RBW is 3 MHz and minimum is 10 Hz.
Trace Averaging
Averaging is a digital process in which each sweep of the
trace returns measurement values for each point in the
trace. These values are then mathematically averaged with
the previous sweep trace data which has been stored in the
analyzer. The amount of averaging is selected by choosing
the number of trace sweeps to be included in the process.
The averaging function uses the most recent trace sweep
values so that the display shows any signal changes.
47
Functions and Measurements
Measuring a Low-Level Signal
NOTE
Selectingaveraging,whentheanalyzerisauto- coupled,
changes the detection mode to Sample, smoothing the
displayed noise level.
This is a trace processing function and is not the same as using the
average detector (as described on
page 47).
1Refer to the procedure
page 44
2To turn averaging on, toggle the softkey menu labeled
Average:
Press BW > Average (On)
3Set the number of averages to 20, using the number keypad,
up and down arrows, or the knob:
Press 20, Enter
The averaging process smooths the viewed trace, low level
signals become more visible (see
the average number will restart the averaging process.
Figure 3-13 Trace Averaging
of this chapter and follow steps 1, 2 and 3.
“Reducing Input Attenuation” on
Figure 3-13). Changes to
48
Functions and Measurements
Improving Frequency Resolution and Accuracy
Improving Frequency Resolution and Accuracy
This section provides information on using the frequency
counter function to improve frequency resolution and
accuracy.
1Press Preset (With Preset Type of Factory)
2Connect a cable from the front panel CAL OUT to RF IN;
Press System > Service > Calibration > CAL OUT ON to toggle on
and enable the 50 MHz amplitude reference signal.
3Press Frequency > Auto Tune.
The analyzer will detect the signal peak and locate it to the
center ofthedisplay (Referto“Auto Tune” on page 111).
4Turn the frequency counter on:
Marker > Function > Counter.
Press
5Move the marker, with the front- panel knob, half- way down
the skirt of the signal response.
NOTE
Figure 3-14Using Frequency Counter
Counted result
The frequency and amplitude of the marker appears in the active function
area (this is not the counted result). The counted result appears in the
upper-right corner of the display to the right-side of C1.
49
Functions and Measurements
Improving Frequency Resolution and Accuracy
The marker readout in the active frequency function changes
while the counted frequency result (upper-right corner of
display) does not. For an accurate count, the marker need
not be placed at the exact peak of the signal response.
NOTE
NOTE
The Frequency counter properly functions only on stable, CW signals
or discrete spectral components. The marker power level must be
greater than 40 dB above the displayed noise level.
6To turn off the marker counter:
Press Marker
> Function > Off.
When using the frequency counter function, the ratio of the resolution
bandwidth to the span must be greater than 0.02.
50
Tracking Drifting Signals
This section provides information on measuring and tracking
signals that drift in frequency.
Measuring a Source’s Frequency Drift
The analyzer will measure source stability. The maximum
amplitude level and the frequency drift of an input signal
trace can be displayed and held by using the maximum hold
function. Using the maximum hold function you can measure
and determine how much of the frequency spectrum a signal
occupies. For more information, refer to “Max Hold” on
page 162.
Use signal tracking to return a signal drifting in frequency
to the center of the display. The drifting is captured by the
analyzer using the maximum hold function.
1Connect the signal generator to the analyzer RF IN.
Functions and Measurements
Tracking Drifting Signals
2Output a signal with the frequency of 300 MHz and
amplitude of –20 dBm.
3Set the analyzer center frequency, span and reference level.
Press Preset. (With Preset Type of Factory)
Press Frequency > 300 > MHz
Press Span > 10 > MHz
Press Amplitude > –10 > dBm
4Place a marker on the peak of the signal:
Press Peak Search
Press Marker > Marker To > Marker to Center
Press Span > 1 > MHz
The signal is in the center of the display.
6Measure the excursion of the signal with maximum hold:
Press Trace/Det > Max Hold
51
Functions and Measurements
Tracking Drifting Signals
7Activate trace 2 and change the mode to continuous sweep:
8Slowly increase the frequency of the signal generator. Your
As the signal varies, maximum hold maintains the maximum
responses of the input signal. Annotation on the left side of
the screen indicates the trace mode. For example, M1 S2 S3 S4, indicates trace 1 is in maximum-hold mode, trace 2, trace
3, and trace 4 are in store- blank mode.
Press Trace/Det > Select Trace > Trace 2
Press Clear Write
Trace 1 remains in maximum hold mode to show any drift
in the signal.
analyzer display should look similar to
Figure 3-15 Viewing a Drifting Signal With Max Hold and Clear
Write
Figure 3-15.
52
Making Distortion Measurements
This section provides information on measuring and
identifying signal distortion.
Identifying Analyzer Generated Distortion
High- level input signals may cause analyzer distortion
products that could mask the real distortion measured on
the input signal. Use trace and the RF input attenuator to
determine which signals, if any, are internally generated
distortion products.
In this example, we use the RF output of a signal generator
to determine whether the harmonic distortion products are
internally generated by the analyzer.
1Connect the signal generator to the analyzer RF IN.
2Set the source frequency to 200 MHz, amplitude to 0 dBm.
3Set the analyzer center frequency and span:
Press Preset
Press Frequency > 400 > MHz
Press Span > 500 > MHz
Functions and Measurements
Making Distortion Measurements
Figure 3-16Harmonic Distortion
53
Functions and Measurements
Making Distortion Measurements
4Change the center frequency to the value of the first
5Change the span to 50 MHz and re- center the signal:
6Set the attenuation to 0 dB:
7To determine whether the analyzer generates harmonic
The signal produces harmonic products (spaced 200 MHz
from the original 200 MHz signal) in the analyzer input
mixer as shown in
harmonic:
Press Peak Search > Next Peak
Press
Marker > Marker To > Mkr -> CF
Press Span > 50 > MHz
Press Peak Search
Press Marker > Marker To > Mkr -> CF
Press Amplitude > Attenuation > 0 > dB
distortion products, first display the trace data in trace 2 as
follows:
Press Trace/Det > Select Trace > Trace 2
Press Clear Write
Figure 3-16.
54
8Allow trace 2 to update (minimum two sweeps), then store
the data from trace 2 and place a delta marker on the
harmonic of trace 2:
Press Trace/Det > View
Press Peak Search
Press Marker > Mode > Delta
The analyzer display shows the stored data in trace 2 and
the measured data in trace 1. The MarkerΔ indicator reads
the difference in amplitude between the reference and active
markers.
9Increase the RF attenuation to 10 dB:
Press Amplitude > Attenuation > 10 > dB
Functions and Measurements
Making Distortion Measurements
Notice the MarkerD amplitude readout. This is the difference
of the distortion product amplitude between 0 dB and 10 dB
input attenuation settings. If the MarkerΔ absolute amplitude
is approximately 1 dB for an input attenuator change, the
analyzer is generating, at least in part, the distortion.
The MarkerΔ amplitude readout comes from two sources:
2) The reduced contribution of the analyzer circuits to the
harmonic measurement causes the MarkerΔ to be negative.
Large MarkerΔ readout indicates significant measurement
errors. Set the input attenuator to minimize the absolute
value of MarkerΔ.
55
Functions and Measurements
Making Distortion Measurements
Third-Order Intermodulation Distortion
1Connect the equipment as shown in figure below. This
Two- tone, third- order intermodulation distortion is a
common specification in communication systems. When two
signals are present in a non- linear system, they may interact
and create third- order intermodulation distortion (TOI)
products that are located close to the original signals.
System components such as amplifiers and mixers contribute
to the generation of these distortion products.
This example will test a device for third-order
intermodulation through the use of markers. Two sources
are used, one set to 300 MHz and the other to 301 MHz.
combination of signal generators, low pass filters, and
directional coupler (used as a combiner) results in a
two-tone source with very low intermodulation distortion.
Although the distortion from this setup may be better than
the specified performance of the analyzer, it is useful for
determining the TOI performance of the source/analyzer
combination. After the performance of the source/analyzer
combination has been verified, the device- under- test (DUT)
(for example, an amplifier) would be inserted between the
directional coupler output and the analyzer input.
56
Signal Generator
N9310A RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby
On
Frequenc y
Amplitude FM
Enter
7
8
4
5
102
·
RF OUT
AM I/Q
9
6
3
FUNCTIONS
Sweep
MOD
On/Off
RF
On/Off
LF OUT RF OUT 50
Utility
File
TriggerPulseM
REVERSE PWR
4W MAX 30VDC
Preset
Local
LF Out
300 MHz LOW
PASS FILTER
Signal Generator
N9310A RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby
On
300 MHz LOW
PASS FILTER
FUNCTIONS
Utility
Preset
AM I/Q
Frequenc y
Amplitude FM
Local
File
Sweep
LF Out
TriggerPulseM
Enter
MOD
9
7
8
On/Off
4
6
5
102
3
RF
·
On/Off
LF OUT RF OUT 50
REVERSE PWR
4W MAX 30VDC
Directional Coupler
Spectrum Analyzer
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
Remote
Standby
On
7
4
102
TG SOURCE CAL OUT
RF IN
BW/
Auto
Sweep/
Avg
Tune
Trig
Local
View/
Det/
File/
Marker
Print
Trace
Display
Save
Peak
Marker
Meas
MODE
Search
5896
Back
3
Enter
·
50VDC MAX
30dBm 1W MAX
RF IN 50
50MHz10dBm
PROBE POWER
CAT Ⅱ
Functions and Measurements
MFM
Making Distortion Measurements
NOTE
The coupler should have a high degree of isolation between the two
input ports so the sources do not intermodulate.
2Set one source (signal generator) to 300 MHz and the other
source to 301 MHz, for a frequency separation of 1 MHz. Set
the sources equal in amplitude as measured by the analyzer
(in this example, they are set to –5 dBm).
3Set the analyzer center frequency and span:
Press Preset (With Preset Type of Factory)
Press Frequency > 300.5 > MHz
Press Span > 5 > MHz
4Reduce the RBW until the distortion products are visible:
Press BW and
F
5Move the signal to the reference level:
Press Peak Search
Press Marker > Marker To > Mkr -> Ref Lvl
6Reduce the RBW until the distortion products are visible:
Press BW and
7Activate the second marker and place it on the peak of the
distortion product(beside the test signal) using the Next Peak:
Press Marker > Mode > Delta
Press Peak Search > Next Peak
8Measure the other distortion product:
Marker > Normal
Press
Press Peak Search > Next Peak
57
Functions and Measurements
Making Distortion Measurements
9Measure the difference between this test signal and the
second distortion product (see
Press Marker > Mode > Delta
Press Peak Search > Next Peak
Figure 3-17Measuring the Distortion Product
Figure 3-17):
58
Stimulus Response Transmission
The procedure below describes using the built- in tracking
generator of the analyzer to measure the rejection of a low
pass filter. A type of transmission measurement.
1To measure the rejection of a low pass filter, connect the
equipment as shown in
filter as the DUT.
Figure 3-18Transmission Measurement Test Setup
Functions and Measurements
Stimulus Response Transmission
Figure 3-18. A 370 MHz low pass
CAUTION
NOTE
Remote
Standby
On
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
7
4
102
·
TG SOURCE CAL OUT
Auto
BW/
Sweep/
Trig
Tune
Avg
Local
Det/
File/
View/
Marker
Print
Display
Trace
Save
Peak
Marker
MeasMODE
Search
5896
Back
3
Enter
50VDC MAX
30dBm 1W MAX
RF IN 50
50MHz10dBm
PROBE POWER
CATⅡ
DUT
2Press Preset. (With Preset Type of Factory)
3Set the start and stop frequencies and resolution bandwidth:
4Turn on the tracking generator and if necessary, set the
output power to - 10 dBm:
Press Mode > Tracking Generator > Amplitude (On) > -10 > dBm
Excessive signal input may damage the DUT. Do not exceed the
maximum power that the device under test can tolerate.
To reduce ripples caused by source return loss, use 6 dB or greater
output attenuation. Tracking generator output attenuation is normally
a function of the source power selected. However, the output
attenuation may be controlled in the Tracking Generator menu.
59
Functions and Measurements
Stimulus Response Transmission
5To increase measurement sensitivity and smooth the noise:
6Connect the cable from the tracking generator output to the
7Reconnect the DUT to the analyzer and change the
8Measure the rejection of the low pass filter:
Press BW > 30 > kHz
Press BW >Video BW > 300 > kHz
A
decreaseindisplayedamplitudeiscausedbytracking
.
error
analyzer input. Store the frequency response in trace 3 and
normalize:
Press Meas > Normalize > Store Ref (1
→ 4) > Normalize (On)
normalized reference position:
Press Meas > Normalize > Norm Ref Posn > 8
Press Marker > 370 > MHz, Mode > Delta > 160> MHz
The marker readout displays the rejection of the filter at 160
MHz above the cutoff frequency of the low pass filter. See
Figure 3-19 as below.
Figure 3-19 Measure the Rejection Range
60
Functions and Measurements
Measuring Stop Band Attenuation of a Lowpass Filter
Measuring Stop Band Attenuation of a Lowpass Filter
When measuring filter performance, it is useful to look at
the stimulus response over a wide frequency range. Setting
the analyzer y- axis (amplitude) to Log scale type presents
the measurement data in a usable format for determining
filter characteristics.
The following example uses the tracking generator to
measure the stop band attenuation of a 370 MHz low pass
filter.
1To measure the response of a low pass filter, connect the
equipment as shown in
370 MHz low pass filter.
Figure 3-20Transmission Measurement Test Setup
Figure 3-20. This example uses a
Remote
Standby
On
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
7
4
102
TG SOURCE CAL OUT
5896
·
Auto
BW/
Sweep/
Trig
Tune
Avg
Local
Det/
File/
View/
Marker
Print
Display
Trace
Save
Peak
Marker
MeasMODE
Search
Back
3
Enter
50VDC MAX
30dBm 1W MAX
RF IN 50
50MHz10 dBm
PROBE POWER
CAT Ⅱ
DUT
2Press Preset. (With Preset Type of Factory)
3Set the start and stop frequencies and resolution bandwidth:
Press Frequency > Start Freq > 100 > MHz
Press Frequency > Stop Freq > 1 > GHz
Press Amplitude > Scale Type > Log
61
Functions and Measurements
Measuring Stop Band Attenuation of a Lowpass Filter
4Set the resolution bandwidth to 10 kHz:
Press BW > 10 > KHz
CAUTION
Excessive signal input may damage the DUT. Do not exceed the
maximum power that the device under test can tolerate.
5Turn on the tracking generator and, if necessary, set the
11In this example, the attenuation over this frequency range is
- 17.98 dB/octave (one octave above the cutoff frequency).
Functions and Measurements
Measuring Stop Band Attenuation of a Lowpass Filter
12Use the front- panel knob to place the marker at the highest
peak in the stop band to determine the minimum stop band
attenuation. In this example, the peak occurs with Delta of
563 MHz. The attenuation is - 60.93 dB. See
Figure 3-21 Minimum Stop Band Attenuation
Figure 3-21.
63
Functions and Measurements
Channel Scanner
Channel Scanner
Top/Bottom N Channel Scanner
The channel scanner - option SCN is a tool for coverage test,
band clearance, and spectrum monitoring. It measures the
channel power of multiple transmitted channels. The power
is displayed via either the graphic bar chart or time bar
chart. It allows you to set maximum 20 channels for your
test. You can continuously log the scanned channel data as
CSV file for post processing on your PC.
Top/Bottom N channel scan is used for searching the
top/bottom channels from specified channel range according
channel power.
Figure 3-22 Top N Channel Scanner Measurement
64
Please refer to the procedures below to perform a Top N
channel scanner measurement.
1 Connect the antenna to RF IN connector.
2 Turn on the channel scanner function.
Press Meas > Channel Scanner
Functions and Measurements
Channel Scanner
3 Press Meas > Scan Mode > Top N to set the scan mode to
TOP N channel mode. Input the channel number you need
to monitoring. Press Enter as a terminator.
4 Press Range Edit to access the submenu to edit the scan
range.
Toggle Edit Method to underline Std. It allows you to set the
range according to the embedded communication standard,
such as CDMA and GSM which is used specifically in
different countries. Press Signal Std to select the standard.
Press Recall as a terminator. Press Begin Chn and End Chn to
set the begin and end channel for scanning. Press Apply to
apply all the range settings.
If you toggle Edit Method to underline Custom, it allows you
to set the customer-defined scan as your need.
Press Start Freq to set the scan start frequency.
Press Freq Step Size to set the frequency interval for
adjacent channels.
Press Bandwidth to set the bandwidth of each frequency
channel which is used to calculate the channel power.
Press Number of Chn to set the number of frequency
channels which will be scanned.
Press Apply as a terminator to apply all the settings above.
5 Press Scan Start to initiate the channel scan function.
6 Press Logging Start to start recording the scan process
which includes the channel power and related channel
bandwidth value to a CSV file. Press Logging Stop to end
the logging. The CSVfile will be saved immediately.
65
Functions and Measurements
Channel Scanner
List N Channel Scanner
The List N channel scan is used for listing the frequency
channels from specific channel standard or customized
frequency bandwidth.
Please refer to the procedures below to perform a List N
channel scanner measurement.
1 Connect the antenna to RF IN connector.
2 Turn on the channel scanner function.
Press Meas > Channel Scanner}
3 Press Meas > Scan Mode > List to set the scan mode to List
N channel mode. Input the channel number you need to
monitoring. Press Enter as a terminator.
4 Press List Edit to access the submenu to edit the scan
range.
Press Insert to insert a channel and access the channel
setting submenu to customize the channel parameters.
Press Edit Method Std. It allows you to set the range
according to the embedded communication standard, such
as CDMA and GSM which is used specifically in different
countries.
Press Signal Std to load a standard file as your need.
Please refer to
Press Chn ID to select the specific channel according to the
signal standard.
“Loading a File” on page 98 for details.
66
Or press Edit Method Custom. It allows you to set the
customer- defined scan as your need.
Press Freq to set the center frequency of the current
channel.
Press Bandwidth to set the bandwidth of each frequency
channel which is used to calculate the channel power.
The first channel editing is completed. Press Return > Insert
to add more channels as your need.
If there is a need to further edit the channel set previously, press Index, rotate the knob to highlight the channel
as your need for further editing.
Press Apply as a terminator to apply all the settings above.
NOTE
Functions and Measurements
Channel Scanner
5 Press Scan Start to initiate the channel scan function.
6 Press Logging Start to start recording the scan process
which includes the channel power and related channel
bandwidth value to a CSV file. Press Logging Stop to end
the logging. The CSV file will be saved immediately.
Please make sure the file catalog is set to INT or USB unless the logging
file can not be saved successfully. When the logging file exceeds the
maximum volume of the storage catalog, the logging will automatically
pause and the file will be saved immediately.
Display
In channel scanner mode, the [Disp] hardkey is used to
adjust the channel scanner display to review the customized
scan result.
Press {Freeze} to pause the display refresh but the scan is
still running.
You can sort the channel by ID or power, in ascending or
decending order. Or you can press {Disp Mode} to toggle the
display mode between bar mode and time mode.
Figure 3-23 Time display mode in List N channel scanner
67
Functions and Measurements
Channel Scanner
Channel Scanner Setup
To customized the channel scan process, you can press Meas
> Channel Scanner > Setup to set the measurement interval and
rule in the submenu.
The Meas Interval is a value of time/distance during which the
every two scan process is triggered. When it is set to off, the
channel will be scanned with default settings.
The Interval Type is toggled between time and distance. When
it is set to Time, the minimum interval time is 100 ms. When
it is set to Distance, it allows you to measure the channel
with the specific distance on the road test. The minimum
interval distance is 1 meter.
The Measure Rule and Search Rule is toggled between Speed
and Accuracy.
For each scan process, the analyzer will first search and find
out the available channels according to the selected scan
mode. Then, a precise measurement will perform to get the
scan result.
68
The accuracy mode in measurement rule is set as default to
get the precise measurement for the searching- out channels.
The speed mode will speed the measurement but reduce the
measurement accuracy. It is set to default in search rule to
search out the available channels approximately.
The To p Cyc l e number is the searching times for each
measurement. It allows you to measure the specific
searching- out channels continuously.
Spectrum Monitor
Functions and Measurements
Spectrum Monitor
The scrolling three- dimensional display is noted for its
ability to track the frequency and power behavior over the
time, particularly intermittent signals. Spectrum Monitor may
be used to analyze the stability of a signal over the time, or
to identify intermittent interference signals in
communications systems.
The X- axis represents frequency, and amplitude is
represented by color; red for a strong signal and blue for
noise floor. The Y-axis represents time, with the trace from
the newest sweep displayed at the bottom of the screen.
Earlier traces move up toward the top with each new sweep.
Two coupled markers allow the user to place a marker on
any trace in the Spectrum Monitor and view the normal
spectrum for the time of that sweep. The time interval
between sweeps can be adjusted, and up to 1500 traces can
be displayed and saved.
Figure 3-24 Spectrum Monitor Submenu
Please refer to the procedures below to perform a spectrum
monitor measurement.
1 Connect the omni antenna to RF IN connector.
2 Set the center frequency and span.
Press Frequency > 1.8052 > GHz
69
Functions and Measurements
Spectrum Monitor
Press SPAN > 5 > MHz
3 Turn on the spectrum monitor function.
Press Meas > Spectrum Monitor.
4 Set the update interval time between two frames.
Press Update Int. On, rotating the knob or pressing numeric
keypads to set the update interval time.
5 Turn on Marker 1 and Marker 2.
Press Marker 1 > State On, Marker 2 > State On.
6 Press Frequency, rotate knob to change the frequency of
marker as you want.
7 Press Meas > File Logging > Start Save to save the spectrogram
data. After pressing Stop Save, the spectrogram will be
saved into a trace or CSV data file for your future use.
Please refer to
saved spectrogram trace file.
“Loading a File” on page 98 to load the
Spectrum Monitor Audio Alert
The spectrum monitor provides an audio alert with a
frequency dependant on the different control parameters.
This alert needs two markers to specify the measure
channel. The Band Power of the channel specified by two
markers is measured without the audio alert. The Alert Limit
is set to specify the lowest band power that makes the
buzzer beep.
Please refer to the procedures below to turn on the audio
alert.
1 Press Meas > Spectrum Monitor to access into the spectro-
gram mode.
2 Press Marker > More > Audio Alert to enable the audio alert.
70
Reflection Measurement
Press MODE > Reflection Measurement to access the cable and
antenna measurement submenu. There are three different
measurement mode for your selection:
• Reflection measurement
• One port cable loss
• Distance to fault
Preparation
Before making a measurement in a specific mode, the HSA
requires a calibration to make sure the measurement
accuracy. A new calibration must be performed if any
components used in the calibration setup are changed. For
example, a short coaxial cable used to connect the tester to
the DUT
Press Frequency and Span to set the measurement frequency
range. It is best to keep the calibration frequency range as
close as possible to the actual sweep frequencies necessary
for the measurements. Calibrating over a larger frequency
range than the actual measurement range will reduce the
accuracy of the test results.
Functions and Measurements
Reflection Measurement
Calibrate with a T-combo calibrator
T-combo calibratorThis is a T-shaped mechanical calibrator.
It consists of a calibrated open, short and 50 Ω load. Using
the T- combo calibrator will result in a more accurate
calibration than when using the an electronic calibrator.
Using a T- combo calibrator is a three step process requiring
the user to manually change the open- short-load
components. The default order of the calibration process is
open-short-load.
1 Press Frequency > Calibrate to bring up the dialog box for
calibration instruction. Follow the on- screen instruction to
connect the Open component of the T- combo calibrator to
71
Functions and Measurements
Reflection Measurement
Measuring Cable Reflection
the RF OUT connector of the analyzer. Press ENTER to start
the open calibration.
2 Follow the instruction to connect the short component of
the T- combo calibrator to the RF OUT connector of the
tester. Press ENTER to start the short calibration.
3 Connect the Load component of the T- combo calibrator to
the RF OUT connector.
Press ENTER to start the load calibration.
A Calibrated indicator is shown on the upper left corner of
the screen when the calibration is completed. Then, the
analyzer is ready for your further cable measurement.
Press Mode > Reflection Measurement > Reflection Measurement to
select the Reflection Measurement Mode. It allows you to set
the Y axis of the graticule as return loss, VSWR or reflection
coefficient value.
72
Measuring Distance to Fault
For this measurement, an internally generated signal emits
from the RF OUT port of the tester and is transmitted to the
cable and antenna system (DUT). The tester receives and
processes the signal power reflected from faults and
imperfections. In performing a DTF measurement, the tester
uses frequency domain reflectometry. The transmitted and
reflected signals contain information about the distance to
the faults. This information is used to determine the
physical distance to the faults. The tester displays the
physical distance to the probable faults, corrected for cable
loss and the velocity propagation factor of the cable.
Functions and Measurements
Reflection Measurement
Perform a DTF measurement
To make a DTF-return loss measurement over a frequency
range of 50 to 400 MHz, perform the following steps:
1 Press the MODE > Reflection Measurement > Distantce To Fault to
select the DTF mode.
2 Press FREQ > Start >50 > MHz to set the start frequency to
50 MHz.
3 Press Stop > 400 > MHz to set the stop frequency to 400
MHz.
4 Press Calibrate to bring up the calibration process guidance.
Follow the instructions to perform a calibration. For more
information about calibration, refer to
page 71
5 Remove the calibrator and connect the cable under test
(DUT) to the RF OUT connector on the top panel of the
tester.
The DTF measurement starts and the results are displayed
on the screen. A peak in the horizontal trace indicates a
defective location.
.
“Preparation” on
73
Functions and Measurements
Making a Power Measurement with USB Power Sensor
Making a Power Measurement with USB Power Sensor
The N9322C spectrum analyzer supports the U2000 and
U2020 series USB power sensors.
The U2000 Series USB power sensors do not need manual
calibration and zero routines performed. Calibration and
zeroing are performed without removing the power sensor
from the source, through internal zeroing. With internal
zeroing of U2000 Series USB power sensors, there is no need
to disconnect the sensor or power- off the DUT. The U2000
Series do not require 50 MHz reference signal calibration,
allowing the factory calibration to ensure measurement
accuracy. For best accuracy, users are recommended to
perform external zeroing for input signals below -30 dBm for
best accuracy.
CAUTION
NOTE
The maximum power for the RF IN port and the RF OUT port of the
analyzer is +30 dBm (1 W). The maximum power for the U2000 Series
Power Sensor port is +25 dBm (316 mW). When directly coupled to a
base station, the test set can be damaged by excessive power applied
to any of these three ports.
To prevent damage in most situations when directly connect an
analyzer to a base station, use a high power attenuator between the
analyzer and the BTS.
If you suspect other signals may be present, it is recommended that
you turn off all the other channels and measure average power only
on the signal of interest. Another option is to measure channel power
(which is less accurate), that filters out all other channels (signals).
74
Functions and Measurements
Making a Power Measurement with USB Power Sensor
Making an Average Power Measurement
The average power measurements provide a key metric in
transmitter performance.
Base station transmit power must be set accurately to
achieve optimal coverage in wireless networks. If the
transmit power is set too high due to inaccurate power
measurements, undesired interference can occur. If the
transmit power is set too low, coverage gaps or holes may
occur. Either case affects system capacity and may translate
into decreased revenue for service providers. Average power
can be measured for the channel of interest while the base
station is active. All other channels should be inactive.
Average power is a broadband measurement. If other signals
are present the analyzer will also measure their power
contributions.
Figure 3-25 Connection with base station
To make an average power measurement, you connect the
power sensor and cable, zero and calibrate the meter, before
making a measurement. Zeroing of the power meter will
occur automatically:
• Every time you use the Power Meter.
• When a 5 degree C. change in temperature occurs.
• Whenever you change the power sensor.
• Every 24 hours.
75
Functions and Measurements
Making a Power Measurement with USB Power Sensor
• Before measuring low level signals—for example, 10 dB
above the lowest specified power for your power sensor.
Calibrating the power meter every time you cycle the power
on and off.
In most situations, you can press Zero to complete the two
steps (zero and cal) together.
Follow the steps below to make a basic average power
measurement.
1Press Preset. (With Preset Type of Factory)
2Press Mode > Power Meter to access the power meter mode.
3Zero and calibrate the meter. Press Zero to make a Zero
operation of the power sensor followed by a calibration
operation.
4Connect the power sensor to the USB port of N9322C. The
analyzer will recognize the power sensor automatically.
76
CAUTION
To prevent damage to your sensor, do not exceed the power levels below.
U2000/1/2A +25 dBm (320 mW) average, 20 VDC
+33 dBm (2 W) peak, <10 μs
For more information on the maximum power of U2000 series power
sensors, please refer to the Operating and Service Guide of Keysight
U2000 Series USB Power Sensors.
5Connect the external attenuation, if required, with the power
sensor used.
6Connect the analyzer to the base station. Connect the power
sensor to the signal to be measured. A typical BTS
connection is after the output of the power amplifier and
duplexer. The connection requires a cable and two Type- N
barrel connectors, in addition to the power sensor and
external attenuator.
Functions and Measurements
Making a Power Measurement with USB Power Sensor
7Press Meas Setup > Disp Range to set the display range of the
measurement. when it is set to On, the display range will be
adjusted automatically according to the signal power
dynamic range. When it is set to Off, press To p and Bottom to
set the display range manually.
8Press General Setup > Freq to set the power frequency for the
DUT. The current frequency will display in the bottom line.
9Press Power Range > Auto Power to toggles the measurement
range between On and Off. When it is set to On, the power
range will be adjusted automatically. Use auto range when
you are not sure of the power level you are about to
measure. When it is set to Off, press Range to select the
upper range or lower range. The lower range uses the more
sensitive path and the upper range uses the attenuated path.
Table 1Power Sensor Range
SensorLower rangeUpper range
U2000/1/2/4A
U2000/1/2H
U2000/1B
–60 dBm to –7 dBm–7 dBm to +20 dBm
–50 dBm to +3 dBm+3 dBm to +30 dBm
–30 dBm to +23 dBm+23 dBm to +44 dBm
77
Functions and Measurements
Making a Power Measurement with USB Power Sensor
10Press Meas > Meas Seup > Disp Mode to set the display in
meter. chart or trace display. the measurement result will
displays on the screen as shown.
Figure 3-26 Power Measurement
78
Making a Power Measurement with USB Power Sensor
Making a Peak Power Measurement
The Keysight U2020 X-series USB peak and average power
sensors enable peak power measurements that have the same
accuracy as measurements obtained using traditonal peak
power meters, while providing the advantages of being a
very compact and portable form factor.
Follow the steps below to make a peak power measurement.
1Press Mode > Power Meter to access the power meter mode.
2Connect the power sensor to the USB port of N9322C. The
sensor will automatically initialize, perform zeroing (While
remianing connectored to the device under test), and
calibrate without connecting to an external source.
NOTE
You can manually perform zeroing and calibration by perssing Zeroing
> Zeroing and Calibration at any time after the sensor is connected
and initialized.
3Press General Setup > Freq to set the frequency of the signal
under test.
As a wideband device, the power sensor has a frequency
response. The power sensor utilizes a complex correction
algorithm to remove power measurement error due to
linearity error, frquency response, and temperature drift to
generate a more accurate measurement.
Functions and Measurements
79
Functions and Measurements
Making a Power Measurement with USB Power Sensor
The analyzer provide the customer preset function with
U2020 X- series power sensor. Press Meas > Preset and the
predefined protocol list will pop up. Select the wanted item
and press Preset as a terminator. The related settings will be
set automatically.
Figure 3-27 Customized Preset the U2020 Power Sensor
80
4Press Meas Disp > Disp Mode > Trace to find the measurement
results as
Figure 3-28 Peak Power Measurement
Marker Readings
Figure 3-28 below.
Gate ResultsPulse Parameter Results
Power Meter Settings
Please find the sections below to find the general power
meter measurement settings.
Setting the Power Meter Top and Bottom End-Points
Setting the end points (Disk Range) close to expected
measurement value changes the sensitivity of the Power
Meter scale resolution. This is an advantage in viewing small
changes in power. However, this will not affect the overall
range of the sensor. This is only available for Meter and
Chart display mode. For Trace display, please press Meas > Peak Setup > Trace Setup to set the X and Y scale.
1Press Meas > Meas Disp > Dis Range to access the end-points
menu.
2Press To p and enter the maximum scale value using the
numeric keypad. Then press dBm to complete the setup.
3Press Bottom and enter the minimum scale value desired
using the numeric keypad. Then press dBm to complete the
setup.
Functions and Measurements
Making a Power Measurement with USB Power Sensor
Setting Power Meter Resolution
There are four levels of power meter resolution. Higher
resolutions provide more accuracy but slow the measurement
speed. The resolution is only available for U2000 series
power meter and free- run trigger mode in U2020 series
power meter..
Press Meas > Meas Disp > Resolution. Select 1,2,3 or 4. The four
options represent different resolution as follow:
• [1] = 1
• [2] = 0.1
• [3] = 0.01
• [4] = 0.001
81
Functions and Measurements
Making a Power Measurement with USB Power Sensor
Set the Step Detection
Press Meas > General Setup > Average > Step Detect to set the
step Detect On and Off. When it is set to on, the filter will
re- initialize upon detection of a step increase or decrease in
the measured power to reduce the filter settling time after a
significant step in the measured power, Step detection can
be set in both manual and automatic filter modes.
Set the Measurement Gate
Press Meas > Peak Setup > Gate Setup to set the start time and
gate length for a measurement gate. A measurement gate
allows measurements to be performed on particular sections
of the input signal. The gate is defined by a start time
relative to the trigger instant and a duration. Signal samples
acquired during the time interval specified by the gate are
used for the measurements in that gate. A system of up to
four independent gates is provided.
82
Set the Trigger
Press Meas > Peak Setup > Trig/Acq to set the trigger
parameters listed below for the measurement gate.
AcqnSet the single, free run or continuous trigger mode.
Select to enable trace and auto trigger delay for the single and
continuous trigger modes. For the free run mode, only auto trigger
delay can be enabled.
SourceSet the trigger source to internal channel or an
external source.
ModeEnable auto level or manually set the trigger level if
the internal trigger source is selected.
Levelsets the trigger level manually when the Mode is set
to manual.
Functions and Measurements
Making a Power Measurement with USB Power Sensor
DelaySet the delay time to be applied between the trigger
event and all the gate start times. This allows you to
time- shift all the gates by the same amount with one setting
change.
SlopeSelect the positive or negative slope type to
determine if the trigger event is recognized on the rising or
falling edge of a signal respectively.
HoldoffSet the holdoff time to disable the trigger
mechanism after a trigger event occurs.
HysteresisSet the hysteresis to help generate a more stable
trigger by preventing triggering unless the RF power level
achieves the trigger level and the additional hysteresis value.
It can be applied to both rising and falling edge trigger
generation.
Hysteresis is only available for the internal trigger source
and manual trigger level.
83
Functions and Measurements
Demodulate the AM/FM signal
Demodulate the AM/FM signal
The N9322C provide the AM/FM demodulation function.
Demodulating an AM Signal
The demodulation function allows you to demodulate and
hear signal information displayed on the analyzer.
Simply place a marker on a signal of interest, set the
analyzer in zero span, activate AM demodulation, turn the
speaker on, and then listen.
1Press Preset. (With Preset Type of Factory)
2Connect an antenna to the RF IN connector.
3Select a frequency range on the analyzer. For example, the
frequency range for AM broadcasts in the United States is
550 to 1650 kHz:
5Set the frequency of interest to center frequency:
Press Marker > Marker To > Mkr -> Center.
6Change the resolution bandwidth to 100 kHz:
Press BW > 100 > kHz.
7Press Span > Zero Span
8Press Sweep > Sweep Time > 10 s to set the sweep time.
8Set the detector type and turn on AM demodulation:
Press Trace/Det > Detector > Sample.
Press Marker > Function > Demod
Press Demod Setting > Demod Type AM
Demodulating an FM Signal
This section demonstrates how to demodulate and listen to
an FM signal using the built- in FM demodulator.
Using the built in FM demodulator you can tune to an FM
signal and view the results of the detector output as
displayed in the time- domain.
1Press Preset. (With Preset Type of Factory)
2Use a signal source or connect an antenna to recieve FM
signal. In this example the signal is modulated at 300 MHz
with FM deviation of 10 kHz and FM rate of 1 kHz.
NOTE
If you are using a broadcast FM signal, the FM channels are broadcasting
between 87.7 MHz to 107.7 MHz.
3Set the center frequency to the center of the FM signal (in
this case 300 MHz):
Press Frequency > 300 > MHz.
Functions and Measurements
Demodulate the AM/FM signal
4Set the analyzer to zero span for time- domain analysis:
Press Span > Zero Span.
Press Sweep > Sweep Time > 5 > s.
5Set the resolution bandwidth to capture the full bandwidth
of the FM signal.
Press BW > 100 > kHz.
6Turn on the optional pre- amplifier to detect small signals:
Press Amplitude > Preamp On.
7Turn on the FM demodulator:
Press Marker > Function > Demod
Press Demod Setting > Demod Type FM
85
Functions and Measurements
Demodulate the AM/FM signal
Listen to the FM signal.
Press Demod Setting > Speaker Vol, rotate the knob to adjust
the speaker volume.
The headphone jack is availabe for the audio signal output
alternatively.
86
Analysis the Modulated Signals
The N9322C provides the optional AM/FM (Option AMA) and
ASK/FSK (Option- DMA) modulation analysis function for
analysis the modulated signals. These functions can be
activated by license key.
AM/FM Modulation Analysis
Optional AM/FM demodulation analysis provides modulation
metrics, including carrier power, modulation rate, AM
depth/FM deviation, SINAD and carrier frequency offset.
User definable limits provide Pass/Fail indicators of four
types:
Maximum carrier power, Maximum AM modulation index or
FM deviation, Minimum AM modulation index or FM
deviation, and Minimum carrier frequency offset.
The user can save the waveforms with metrics for reporting
as well as the set- up parameters for future measurements or
analysis. In the following example, you can use any source
available (such as AM/FM radio) as long as it has AM/FM
signal generation capability.
Functions and Measurements
Analysis the Modulated Signals
AM Modulation Analysis
Please refer to the procedures below for the AM modulation
analysis:
1 Connect the signal generator to RF IN connector and turn
on the signal generator’s AM and RF output.
2 Turn on the AM demodulation analysis function.
Press MODE > Modulation Analysis > AM.
3 Press Frequency > Carrier Freq > 100 > MHz to set the carrier
frequency.
4 Set the attenuation state to Auto (default):
Press Amplitude > Attenuation Auto
87
Functions and Measurements
Analysis the Mod ulated Signals
5 Press Shift > Disp > Y Scale> Scale Coupling On to set the Y
scale auto coupling.
6 Set the AM depth limit and turn on the limit function:
Press Shift > Limit > AM Depth UP > 51 > % > AM Depth Low > 49
> % > Limits on
The red value in the result indicates the demodulated AM
depth exceeds the limits set as above.
Figure 3-29 AM demodulation analysis
88
FM Modulation Analysis
Please refer to the procedures below for the FM modulation
analysis:
1 Connect the signal generator to RF IN connector and turn
on the signal generator’s FM and RF output.
2 Turn on the FM demodulation analysis function.
Press MODE > Modulation Analysis > AM.
3 Press Frequency > Carrier Freq > 100 > MHz to set the carrier
frequency.
4 Set the attenuation state to Auto (default):
Press Amplitude > Attenuation Auto
Functions and Measurements
Analysis the Modulated Signals
5 Press Shift > Disp > Y Scale> Scale Coupling On to set the Y
scale auto coupling.
6 Set the FM deviation limit and turn on the limit function:
The red value in the result indicates the demodulated FM
deviation exceeds the limits set as above.
Figure 3-30 FM modulation analysis
89
Functions and Measurements
Analysis the Mod ulated Signals
ASK/FSK Modulation Analysis
The ASK/FSK modulation analysis function supports four
display modes:
• Symbol
• Wave form
• ASK/FSK Error
• Eye Diagram.
User definable limits provide Pass/Fail indicators of four
types:
• Maximum carrier power
• Maximum ASK modulation depth/FSK frequency deviation
• Minimum ASK modulation depth/FSK frequency deviation
• Maximum carrier frequency offset.
The metrics includes carrier power, ASK/FSK error, ASK
depth/FSK frequency deviation, and ASK index etc. The
waveform with metrics and setup parameters can be saved
for reports and future Measurements.
90
In the following example, you can use any source available
(such as your remote key of the car) as long as it has
ASK/FSK generation capability. For ASK/FSK signal, its
Filter is Nyquist and Alpha is 0.35.
ASK Modulation Analysis
Please refer to the procedures below to analysis the ASK
demodulation.
1 Connect the signal generator to RF IN connector and turn
on the signal generator’s ASK modulation and RF output.
2 Turn on the ASK demodulation analysis function.
Press MODE > Modulation Analysis > ASK.
3 Press Frequency > Carrier Freq > 100 > MHz to set the carrier
frequency.
4 Press Meas > Symbol Rate > 10 > MHz to set the symbol rate.
Functions and Measurements
Analysis the Modulated Signals
5 Press Filter Setup > Ref Filter > Nyquist to set the filter type.
6 Set the attenuation state to Auto (default):
Press Amplitude > Attenuation Auto
7 Press Shift > Disp > Y Scale> Scale Coupling On to set the Y
scale auto coupling.
8 Setting the ASK depth limit and turn on the limit