Keysight Technologies N9322C User Manual

Keysight N9322C Spectrum Analyzer
Notice: This document contains references to Agilent. Please note that Agilent’s Test and M
easurement business has become Keysight
Technologies.
User’s Guide

Notices

© Keysight Technologies, Inc. 2012-2014
No part of this manual may be reproduced in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technolo­gies, Inc. as governed by United States and international copyright laws.
Manual Part Number
N9322-90002
Edition
First Edition, July 2014
Printed in China
Keysight Technologies, Inc. No 116 Tianfu 4th Street Chengdu 610041, China
Software Revision
This guide is valid for revision A.04.20 of the Keysight N9322C Spectrum Analyzer firmware.
Warranty
The material contained in this docu­ment is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maxi­mum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied, with regard to this manual and any information con­tained herein, includ ing but not limited to the implied warranties of merchant­ability and fitness for a particular pur­pose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the fur­nishing, use, or performance of this document or of any information con­tained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the mate­rial in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to the fed­eral government include only those rights customarily provided to end user customers. Keysight provides this customary commercial license in Software and technical data pursu­ant to FAR 12.211 (Technical Data) and
12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Computer Software Documentation).
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated cond itions are fully understood and met.
1Overview1
N9322C at a Glance 2
Front Panel Overview 4
Display Annotations 7
Rear Panel Overview 9
Front and rear panel symbols 10
2 Getting Started 11
Check the Shipment and Order List 12
Options 12
Safety Notice 13
Power Requirements 14
Power On and Check 16
Environmental Requirements 18

Contents

Electrostatic Discharge Protection 21
BSA PC software 22
Helpful Tips 23
Perform Calibration 23
Using an External Reference 24
Enable an Options 24
Firmware Upgrade 25
IO Configuration 25
Timed Power On/Off 26
Power Preset Last 26
Data Securities 27
Probe Power Output 27
1
Contents
3 Functions and Measurements 29
Making a Basic Measurement 30
Using the Front Panel 30
Presetting the Spectrum Analyzer 31
Viewing a Signal 33
Measuring Multiple Signals 35
Comparing Signals on the Same Screen Using Marker Delta 35
Comparing Signals not on the Same Screen Using Marker Delta 37
Resolving Signals of Equal Amplitude 39
Resolving Small Signals Hidden by Large Signals 42
Measuring a Low-Level Signal 44
Reducing Input Attenuation 44
Decreasing the Resolution Bandwidth 46
Trace Averaging 47
Improving Frequency Resolution and Accuracy 49
Tracking Drifting Signals 51
Making Distortion Measurements 53
Identifying Analyzer Generated Distortion 53
Third-Order Intermodulation Distortion 56
Stimulus Response Transmission 59
Measuring Stop Band Attenuation of a Lowpass Filter 61
Channel Scanner 64
Top/Bottom N Channel Scanner 64
List N Channel Scanner 66
Channel Scanner Setup 68
Spectrum Monitor 69
2
Reflection Measurement 71
Preparation 71
Measuring Cable Reflection 72
Measuring Distance to Fault 72
Making a Power Measurement with USB Power Sensor 74
Making an Average Power Measurement 75
Making a Peak Power Measurement 79
Power Meter Settings 81
Demodulate the AM/FM signal 84
Demodulating an AM Signal 84
Demodulating an FM Signal 85
Analysis the Modulated Signals 87
AM/FM Modulation Analysis 87
ASK/FSK Modulation Analysis 90
Contents
Measuring Channel Power 93
Files Operation 95
4 Key Reference 101
Amplitude 102
BW 106
Display 109
Enter 110
ESC <- 110
Frequency 111
Marker 114
Meas 119
3
Contents
MODE 130
Peak Search 148
System 151
Span 154
Channel Power 119
Occupied BW 121
ACPR 123
Spectrum Emission Mask 124
Tracking Generator 130
Power Meter 133
Spectrum Monitor 136
Reflection Measurement 138
AM/FM Modulation Analysis 140
ASK/FSK Modulation Analysis 144
Sweep 155
Limit 159
Trace 162
5 Instrument Messages 167
Overview 168
Command Errors 169
Execution Conflict 171
Device-Specific Errors 173
6 Troubleshooting 177
Check the basics 178
Contact Keysight Technologies 180
4
7 Menu Maps 181
Amplitude Menu 182
BW/Avg Menu 183
Det/Display Menu 184
File Menu (1 of 2) 185
File Menu (2 of 2) 186
Frequency Menu 187
Marker Menu 188
Marker-> Menu 189
Measure Menu (1 of 2) 190
Measure Menu (2 of 2) 191
MODE - Tracking Generator 192
MODE - Power Meter 193
Contents
MODE - AM/FM Modulation Analysis 194
MODE - ASK/FSK Modulation Analysis 195
Name Editor Menu 196
Peak Search Menu 197
Preset/System Menu 198
SPAN Menu 199
Sweep/Trig Menu 200
Trace Menu 201
Index 203
5
Contents
6

Overview

1 Overview
Keysight N9322C at a Glance 2
Front Panel Overview 4
Rear Panel Overview 9
Front and rear panel symbols 10
This chapter provides a description of the Keysight N9322C spectrum analyzer and an introduction to the buttons, features, and functions of the front and rear instrument panels.
1
Overview

N9322C at a Glance

N9322C at a Glance

Features

The Keysight N9322C spectrum analyzer is a portable, swept spectrum
7.0 GHz. It can be widely used in an electronic manufacturing environment and in functional/final/QA test systems.
The Keysight N9322C spectrum analyzer primary features and functions are described below:
analyzer with a frequency range of 9 kHz to
• Power Measurement Suite
The built- in one- button power measurement suite offers channel power, ACP, OBW, and TOI measurements, and contains a Pass/Fail testing capability with a mask for out- of- channel emissions measurement.
• Precision Frequency Reference (Option PFR)
The optinal OCXO frequency reference provides the precise frquency measurement capability.
• Spectrum Monitor (Option MNT)
Provides the capability to analyze the stability of a signal over time. The analyzer can be used to monitor the signal capturing performance or intermittent event over extended periods of time.
• High Sensitive Measurement (Option P07)
Includes an optional pre- amplifier for signals in the frequency range up to 7 GHz, enabling more sensitive measurements. This feature is a great help in analysis of low level signals.
• High Accuracy Power Measurement
Supports Keysight U2000 series (option PWM) and U2020 series (option PWP) power sensors for high accuracy power measurement as a power meter.
2
Overview
N9322C at a Glance
• Modulation Analysis Function
provides optional AM/FM and ASK/FSK modulation analysis function. (AM/FM: Option- AMA ASK/FSK: Option- DMA)
• Tracking Generator (Option TG7)
provides an RF source for scalar network analysis.
• Baseband Channel (Option BB1)
Provides superior DANL and SSB between 9 kHz to 10 MHz.
• Cable & Antenna Test (Option RM7; Requires option TG7)
Provides VSWR, return loss, cable loss and distance- to- fault function for reflection measurement.
• Task Planner (Option TPN)
Provides task planner function to integrate different measurements for test automation.
• Time-gated Spectrum Analysis (Option TMG)
Measures any one of several signals separated in time and xxclude interfering signals.
• Channel Scanner (Option SCN)
provide the channel scan funtion in spectrum monitoring, coverage test, and band clearance.

Optimized Usability

The analyzer provides the enhanced usability as below:
Socket/Telnet remote control via USB, and LAN port.
The BSA PC Software on help kit CD provides further editing
and data analysis functions.
Keypad can be locked and unlocked with password to
forbid undesired keypad operation.
The GPIB interface (option G01) make the unit easily
compatible with your GPIB test system.
Security feartures (Option SEC) allows user to erase all
customized files and data in analyzer for security.
3
Overview

Front Panel Overview

Front Panel Overview
BW Trace/
Marker
8
VWX
596
MNO
DEF
·
40MHz 25dBm
6
Preset
Disp
UserDet
ESC
Enter
Local
PROBE POWER
System
Limit
Sweep
Save
Shift
File
RF IN 50
50VDC Max 33dBm (2W) Max
12
Peak Search
MeasMode
YZ_
PQR
3
GHI
13
2
3
4
1
N9322C 9kHz - 7.0 GHz
Spectrum Analayz er
19
18
17
16
Remot e
Standby On
15
5
Frequency
Span
Amplitude
7
STU
4
JKL
102
ABC
TG SOURCE CAL OUT
Rev Pwr 50VDC Max +30dBm (1W) Max
14
1Screen 6.5 inch TFT color screen shows information of the
current function, including the signal traces, status indicators, and instrument messages. Labels for softkeys are located on the right- hand side of the screen.
2Softkeys are the unlabeled keys next to the screen. They
activate functions displayed to the left of each key.
3 Amplitude activates the reference level function and
accesses the amplitude softkeys, with which you set functions that affect data on the vertical axis.
7
8
9
10
11
4SPAN sets the frequency range symmetrically about the
center frequency. The frequency- span readout describes the total displayed frequency range.
5Frequency activates the center- frequency function, and
accesses the menu of frequency functions.
6 Function Keys relate directly to the following main functions:
BW activates the resolution bandwidth function and
accesses the softkeys that control the bandwidth functions and averaging.
Trace/Det accesses the softkeys that allow you to set up
the trace settings and detector functions.
4
Front Panel Overview
Disp - User accesses the submenu to recall the frequently
used user- defined state, or control what is displayed on the analyzer, including the display line, graticule and annotation after pressing Shift.
Mode selects the measurement mode according to the
activated options.
Marker accesses the marker control keys that select the
type and number of markers and turns them on and off.
Limit - Sweep accesses the softkeys that allow you to set
the sweep time, select the sweep mode and trigger mode, or to set the limit functions after pressing Shift.
Peak Search places a marker on the signal peak.
Meas accesses the measurements mode selection submenu
in spectrum analyzer mode or the measurement submenu in other optional modes.
7 Utility Keys relate directly to the following utilities:
Preset resets the analyzer to a known state
System accesses the submenu to configure a series of
system settings, such as ports, option and languages.
Shift switches alternate upper blue function of the
hardkeys.
Save - File accesses the file operation submenu or activate the quick- save function after pressing Shift.
Overview
8 Arrow Keys The up and down arrow keys shift the selected
item when you press [MODE] hardkey; you can also change the mode by rotating the knob.
9 Knob The knob increases or decreases a value, a numeric
digit, or scrolls up and down to select an item in a list.
10 Data Control Keys includes the numeric keypad, ENTER key
and backspace key, change the numeric value of an active function such as center frequency, start frequency, resolution bandwidth, and marker position. if the analyzer is in the remote mode, pressing ENTER key returns the analyzer to the local mode and enables front- panel control.
5
Overview
Front Panel Overview
11 RF IN connector recieves the signal input. The maximum
damage level is average continuous power +33 dBm or DC voltage 50 VDC. The impedance is 50 W. (N- type female).
12 PROBE POWER connector provides power for high- impedance
AC probes or other accessories (+15 V, –12 V, 150 mA maximum).
13 CAL OUT connector provides an amplitude reference signal
output of 40 MHz at –25 dBm (BNC female).
14 TG SOURCE connector N- type female, is the source output for
the built- in tracking generator. The maximum damage level is average continuous power +30 dBm or DC voltage 50 VDC. The impedance is 50 W. (for Option TG7)
15 Standby Switch switches on all functions of the analyzer. To
switch the analyzer off, press the switch for at least 2 seconds. This deactivates all the functions but retains power to internal circuits so long as the analyzer is connected to line power.
16 On LED (green) lights when the analyzer is switched on.
17 Standby LED lights when the analyzer is connected to the
line power.
18 Remote LED lights when the analyzer is remotely controlled
by a PC via the USB host interface on the rear panel.
19 USB Device Connector provides a connection between external
USB devices and the analyzer, such as a USB memory device.
6

Display Annotations

5
4
3
2
1
26
25
24
23
22
21
Overview
Front Panel Overview
6
7
8
10
9
11
12
13
1920
18
17
16
14
15
Item Description Notes (Associated function key)
1 Average Type [BW] > Average Type
2 Reference Offset [Amplitude] > Ref Offset
3 Amplitude scale [Amplitude] > Scale Type
4 Reference level [Amplitude]
5Display Line [Disp]
> Ref Level
> Display Line
6 Time and date display [System] > Time/Date
7 RF attenuation [Amplitude] > Attenuation
8 Marker frequency [Marker]
9 Marker amplitude [Marker]
10 Indicator S - Shift
R - Remote Control
11 USB disk indicator The USB memory device is in use.
12 Key menu title Dependent on key selection.
7
Overview
Front Panel Overview
13 Softkey menu Refer to “Key Reference” for details.
Dependent on current function key selection.
14 Frequency span [SPAN]
15 Sweep time [Sweep] > Sweep Time
16 Fequency offset [Frequency] > Freq Offset
17 Video bandwidth [BW] > Video BW
18 Display status line Display status and instrument messages.
19 Resolution bandwidth [BW] > Res BW
20 Center Frequency [Frequency] > Center Freq
21 Active function block The functino in use
22 Low frequency channel [Frequency] > LowFrqChannel
23 Sweep time rule [Sweep] > Sweep Setup > Swp Time Rule
24 Sweep type [Sweep] > Sweep Setup > Sweep Type
25 - right Trigger/Sweep
F - free run trigger V - video trigger E - external trigger C - continuous sweep S - single sweep
25 - left Internal preamp [Amplitude] > Int Preamp
26 Trace mode
W - clear write M - maximum hold m - minimum hold V - view S - store blank
Detect mode
N - Normal A - Average P - Positive peak S - Sample N - Neg Peak
[Sweep] > Trigge
[Trace]
[Trace] > Detector
8

Rear Panel Overview

1
SERIAL LABEL ATTACH HERE
HIPOT LABEL ATTACH HERE
LAN
TRIG IN 10MHz
T T L
REF IN 10MHz
REF OUT
10
Overview
Rear Panel Overview
9
OCXO
REF OUT 10MHz
PC
100
~100-240
W MAX
V
Hz
50/60
8
2
3
5
4
6
7
1 Kensington Lock lock the instrument and keep its safety.
2 USB Host connector provides a connection between the
analyzer and an PC for remote control.
3LAN port A TCP/IP Interface used for remote operation.
4 EXT TRG IN (TTL) connector accepts an external voltage input,
the positive edge of which triggers the analyzer sweep function. (BNC female)
5REF IN connector accepts an external timebase with a
frequency of 10 MHz, amplitude of –5 to +10 dBm. (BNC female)
6 REF OUT connector provides a frequency of 10 MHz, amplitude
of –10 dBm reference output. (BNC female)
7 REF OUT connector (Option PFR) is an optional connector for
precision 10 MHz reference from OCXO.
8AC Power Receptacle accepts a three- pin line power plug.
9Power switch isolates the analyzer from the AC line power.
After switch on, the analyzer enters into standby mode and the orange standby LED on the front panel lights.
10 GPIB connector (Option G01) is an optional interface. This GPIB
connector supports remote operation.
9
Overview

Front and rear panel symbols

Front and Rear Panel Symbols
The instruction manual symbol: indicates that the user must refer to
specific instructions in the manual.
The CE mark: a registered trademark of the European Community.
ISM1-A
C
US
ICES/NMB-001
Shows that this is an Industrial Scientific and Medical Group 1 Class A product. (CISPR 11, Clause 4)
The CSA mark: a registered trademark of the Canadian Standards Association International.
The ISM device complies with Canadian Interference- Causing Equipment Standard- 001. Cet appareil ISM est conforme à la norme NMB- 001 du Canada.
All Level 1, 2 or 3 electrical equipment offered for sale in Australia and New Zealand by Responsible Suppliers must be marked with the Regulatory Compliance Mark (RCM Mark).
marks the “on/standby” position of the switch.
indicates that the instrument requires AC power input.
indicates this product complies with the WEEE Directive(2002/96/EC) marking requirements and you must not discard this equipment in domestic household waste. Do not dispose in domestic household waste.
indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
KC mark: as Korea Certification. This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.
10
2 Getting Started
Check the Shipment and Order List 12
Safety Notice 14
Power Requirements 15
Power On and Check 17
Environmental Requirements 19
Electrostatic Discharge Protection 22
Helpful Tips 23

Getting Started

This chapter helps you in preparing the spectrum analyzer for use and provides the information to start using the spectrum analyzer correctly.
11
Getting Started

Check the Shipment and Order List

Check the Shipment and Order List
After receiving the shipment, first check the shipment and your order list. Inspect the shipping container and the cushioning material for signs of stress. Retain the shipping materials for future use, as you may wish to ship the analyzer to another location or to Keysight Technologies for service. Verify that the contents of the shipping container are complete.
Shipping Problems?
If the shipping materials are damaged or the contents of the container are incomplete, please contact the nearest Keysight Technologies office to arrange for repair or replacement. You will not need to wait for a claim settlement. If you must return an analyzer to Keysight Technologies, use the original (or comparable) shipping materials.

Options

12
Verify if that the shipment includes your ordered options by checking the option label on the shipping container.
Unless specified otherwise, all options are available when you order a spectrum analyzer; some options are also available as kits that you can order and install/activate after you receive the analyzer. Order kits through your local Keysight Sales and Service Office.
At the time of analyzer purchase, options can be ordered using your product number and the number of the option you are ordering. For example, if you are ordering option TG7, you would order N9322C-TG7.
If you are ordering an option after the purchase of your analyzer, you will need to add a K to the product number and then specify which option you are ordering. For example, N9322CK- TG7.

Safety Notice

Getting Started
Safety Notice
Read the following warnings and cautions carefully before powering on the spectrum analyzer to ensure personal and instrument safety.
WARNING
WARNING
WARNING
CAUTION
CAUTION
Always use a well-grounded, three-pin AC power cord to connect to power source. Personal injury may occur if there is any interruption of the AC power cord. Intentional interruption is prohibited. If this product is to be energized via an external auto transformer for voltage reduction, make sure that its common terminal is connected to a neutral (earthed pole) of the power supply.
Personal injury may result if the spectrum analyzer covers are removed. There are no operator-serviceable parts inside. To avoid electrical shock, refer servicing to qualified personnel.
Electrical shock may result if the spectrum analyzer is connected to the power supply while cleaning. Do not attempt to clean internally.
Prevent damage to the instrument and ensure protection of the input mixer by limiting average continuous power input to +33 dBm, DC voltage to 50 VDC with > if these precautions are not followed.
To install the spectrum analyzer in other racks, note that they may promote shock hazards, overheating, dusting contamination, and inferior system performance. Consult your Keysight customer engineer about installation, warranty, and support details.
10 dB input attenuation. Instrument damage may result
13
Getting Started

Power Requirements

Power Requirements

AC Power Cord

The spectrum analyzer has an auto- ranging line voltage input. The AC power supply must meet the following requirements:
Voltage: 100 to 240 VAC
Frequency: 50 to 60 Hz
Power: Maximum 100 W
The analyzer is equipped with a three- wire power cord, in accordance with international safety standards. This cable grounds the analyzer cabinet when connected to an appropriate power line outlet. The cable appropriate to the original shipping location is included with the analyzer.
14
Various AC power cables are available that are unique to specific geographic areas. You can order additional AC power cables for use in different areas. The table AC Power Cords lists the available AC power cables, the plug configurations, and identifies the geographic area in which each cable is appropriate.
The detachable power cord is the product disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the product. The front panel switch is only a standby switch and do not disconnect instrument from LINE power.
AC Power Cords
Getting Started
Power Requirements
Plug Type Cable Part
Number
8121-1703 BS 1363/A Option 900
250V 10A
8120-0696 AS 3112:2000 Option 901
250V 10A
8120-1692 IEC 83 C4 Option 902
250V 16A
8120-1521 CNS 10917-2
125V 10A
8120-2296 SEV 1011 Option 906
250V 10A
8120-4600 SABS 164-1 Option 917
230V 15A
8120-4754 JIS C8303 Option 918
125V 15A
8120-5181 SI 32 Option 919
a
Plug Description
/NEMA 5-15P
For use in Country & Region
United Kingdom, Hong Kong, Singapore, Malaysia
Australia, New Zealand
Continental Europe, Korea, Indonesia, Italy, Russia
Option 903
United States, Canada, Taiwan, Mexico
Switzerland
South Africa, India
Japan
Israel
250V 16A
8120-8377 GB 1002 Option 922
China
250V 10A
a. Plug description describes the plug only. The part number is for the complete cable assembly.
15
Getting Started

Power On and Check

Power On and Check
1 Connect the AC power cord into the instrument. Insert the
2 Press the AC line switch on the rear panel. The standby
plug into a power socket provided with a protective ground. Set the tilt adjustors for your preference.
Figure 1 Plug in and angle adjustment
LED on the front panel will light and the spectrum analyzer is in standby mode (AC power applied).
16
NOTE
3 Press the standby switch on the front panel. The On LED
will light, and the spectrum analyzer boots up.
Self- initialization takes about 25 seconds; the spectrum analyzer then defaults to the menu mode. After power on, let the spectrum analyzer warm up for 45 minutes for stabilization.
The front panel switch is a standby switch only; it is not a power switch. To completely disconnect the spectrum analyzer from the AC line power, shut off the power switch on the rear panel.

Check for Instrument Messages

The spectrum analyzer has two categories of instrument messages: error and warning messages. A error message is triggered by operation errors, for example, parameter setting conflicts or data input that is out of the range of a parameter. An warning message may be triggered by hardware defects which could result in damage to instrument.
Here are some tips to check the instrument messages.
1 Check the display to see if any messages display in the
status bar. Press [System] > System Info > Error History to review each message. Refer to Chapter 5, “Instrument Messages” for detailed system message descriptions.
2 When you have reviewed and resolved all of the error
messages, press [System] > System Info > Error History > Clear
error queue to delete the messages.
3 Cycle the power to the analyzer and re-check to see if the
instrument messages are still there.
Getting Started
Power On and Check
4 If the error messages cannot be resolved, please contact the
Keysight Customer Contact Center for assistance or service.
17
Getting Started

Environmental Requirements

Environmental Requirements
Keysight Technologies has designed this product for use in Installation Category II, Pollution Degree 2, per IEC 61010- 1. Keysight has designed the spectrum analyzer for use under the following conditions:
Indoor use
Altitude < 3,000 meters
Operating temperature range: +5 to +45
Storage temperature range: –20 to +70
Relative humidity range 15% to 95 % at 40
o
C;
o
C
o
C
NOTE

Ventilation

Cleaning Tips

This equipment is Class A suitable for professional use and is for use in
electromagnetic environments outside of the home.
Ventilation holes are located on the rear panel and one side of the spectrum analyzer cover. Do not allow these holes to be obstructed, as they allow air flow through the spectrum analyzer.
When installing the spectrum analyzer in a cabinet, do not restrict the convection of the analyzer. The ambient temperature outside the cabinet must be less than the maximum operating temperature of the spectrum analyzer by
o
C for every 100 watts dissipated within the cabinet.
4
To prevent electrical shock, disconnect the spectrum analyzer from line power before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
18

Rack Mount

Getting Started
Environmental Requirements
It is recommended to use the Keysight rackmount kit (option 1CM) to install the spectrum analyzer into a rack.
Do not attempt to rack mount the spectrum analyzer by the front panel handles only. This rackmount kit will allow mounting of the spectrum analyzer with or without handles.
Refer to the following instructions when installing the rackmount kit on the spectrum analyzer.
1 Remove feet, key- locks and tilt stands.
2 Remove side trim strips and one middle screw per side.
19
Getting Started
Environmental Requirements
3 Attach rackmount flange and front handle assembly with 3
screws per side.
20
CAUTION
4 Attach the spectrum analyzer to the rack using the
rackmount flanges with two dress screws per side.
Installing the spectrum analyzers into other racks may promote shock hazards, overheating, dust contamination, and inferior system performance. Consult your Keysight customer engineer about installation, warranty, and support details.

Electrostatic Discharge Protection

Electrostatic discharge (ESD) damages or destroys electronic components (the possibility of unseen damage caused by ESD is present whenever transported, stored, or while the instrument is in use).
This product contains components that are easily damaged by electrostatic discharge. To help reduce the possibility of ESD damage that can occur while using test equipment, follow these guidelines:
1 Before connecting any coaxial cable to the spectrum analyzer
connector for the first time each day, momentarily short the center and outer conductors of the cable together to eliminate any potential electrostatic charges that may exist.
2 Personnel should be grounded with an approved type, 1 MΩ
resistor- isolated ESD wrist- strap before touching the center pin of any connector, and before removing any assembly from the spectrum analyzer.
Getting Started
Electrostatic Discharge Protection
3 Be sure that all instruments are properly earth grounded to
prevent build- up of static charges.
21
Getting Started

BSA PC software

BSA PC software
Keysight HSA PC software is an easy- to- use, PC- based remote control tool for the N9322C spectrum analyzer.
It is able to be discretely used as a spectrum monitor to display and control the trace scans simultaneously with the analyzer, or a file manager to send/get files between the anaylzer and PC. It also provides some data analysis function for your further use.
22
For the further descrption of the BSA PC software, please refer to the online help embedded in this software.

Helpful Tips

Perform Calibration

Getting Started
Helpful Tips
The following contains information to help in using and maintaining the instrument for optimum operation, including alignment, external reference, firmware update and option activation.
The N9322C provides three manual calibration function to calibrate the time base and amplitude. The analyzer should warm up for 30 minutes before calibration.
Time Base Calibration
Perform a time base calibration to guarantee the frequency accuracy. The analyzer should warm up for approximately 30 minutes before calibration.
When the calibration is triggered, the current measurement is interrupted and a gauge displays on the LCD. The gauge simply indicates calibration is in process. When it is finished, the interrupted measurement restarts.
Please refer to the operation procedures as below:
1 Input a 10 MHz, 0 dBn signal to EXT TRIG IN.
2 Press [System] > Service > Calibration > Time Base by Ext to
initiate the calibration.
Time base calibration takes only a short time when the inner temperature is stable. When the inner temperature is increasing, calibration takes a long period of time or will fail. If the input reference signal is abnormal, the calibration cycle will take a long and unpredictable time to exit, and the LCD displays an error message.
23
Getting Started
Helpful Tips
Amplitude Calibration
The analyzer privides the internal amplitude calibration function. Please refer to the procedures below to perform an amplitude calibration:
1 Press [System] > More > Service > Calibration > Amp Alignment
2 2 Connect the CAL OUT and RF IN connector with a
N- BNC cable. Then Press [Enter] to continue.
The calibration will start immediately. The calibration will be completed in serveral minutes.

Using an External Reference

To use an external 10 MHz source as the reference frequency, connect the external reference source to the REF IN connector on the rear panel. An EREF indicator will display in the left side of the display. The signal level must be in the range of –5 to +10 dBm. When the reference signal is ready, the instrument switches the reference from time base to the external reference.
24

Enable an Options

Option license key information is required to enable product options. Contact your nearest Keysight office for purchasing a license. Refer to the procedures below to activate the options you have purchased.
1 Press [System] > More > Service > Add Option
2 Enter the option number to be enabled. Press [Enter] to
confirm your input.
3 Enter the license key in the onscreen window. Press [Enter]
to confirm your input and terminate the license key input.
4 Press Activate License. The option
The analyzer provides the trial license with limited time usage(15/270/365 days) of the wanted options.
will be enabled immediately.

Firmware Upgrade

Press [System] > System Info > Show system to view the firmware revision. If you call Keysight Technologies regarding your analyzer, it is helpful to have this revision and the analyzer serial number available.
Follow this procedure to finish the firmware update:
2 Plug the USB stick into the connector. Press [System] >
Service > Upgrade Firmware
The analyzer will perform the update process automatically. The upgrade procedure will take several minutes. Once the upgrade is completed, please follow the instruction to reboot.
Getting Started
Helpful Tips
CAUTION
Any interruption during the update process will result in update failure and system data lost. Do not remove the USB storage device until the update is finished.

IO Configuration

The N9322C spectrum analyzer provides three types of IO connnection: USB, LAN and optional GPIB interface. Press [System] > More > Port Setting to set the corresponding interface as your need.
USB
Select USB to enable the USB connection for remote control.
LAN
The N9322C supports LAN port connection for remote control. Press [System] > {More} > {Port Setting} > {IP Config} to set the IP parameters for the network connection.
25
Getting Started
Helpful Tips
Press {IP Address Static} to manually set the IP address, gateway and subnet mask with the proper LAN information, or just press {IP Address DCHP} to set the IP address in LAN dynamically according DCHP.
Press {Apply} to enable the LAN connection according to the IP configuration you set.
GPIB
Press [System] > {More} > {Port Setting} > {GPIB Address} to set
the GPIB address for the analyzer. This softkey is only vaild
after the option G01 is installed.

Timed Power On/Off

Pressing [System] > {Power On/Off Preset} > {Timed Pwr On} or
{Timed Pwr Off} sets the time switch to power on/off the
N9322C in a user- defined time and date. This function
requires the power supply connected.
Press {Repeat Mode Once/Everyday} to set the N9322C boot up/off in the pre- saved time everyday. The pre- saved date is invalid in this mode.
26

Power Preset Last

Press [System] > {Pwr On/Preset} > {Power On Last} to activate this function. For the standard N9322C, this operation only save the last state if the analyzer is turned off by the front panel power switch.

Data Securities

The N9322C offers the optional memory erase function for data security. Press [System] > {More} > {Securities} > {Erase Memory} to erase all the user data in internal memory. Press Enter as a terminator to start the erase process immediately.
The memory erase process takes about 15 minutes. During the time, there must be a constant power supply to ensure the successful erase. If the erase process is interrupted, please reboot the instrument and erase memory again.

Probe Power Output

The Probe Power provides power for high- impedance AC probes or other accessories (+15 V, –12V, 150 mA maximum).
The Porbe Power is set to off as default. Press [System] > {More} > {Port Setting} > {Probe Power On} to switch on the porbe power output.
Getting Started
Helpful Tips
27
Getting Started

South Korea Class A EMC Declaration

South Korea Class A EMC Declaration
This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.
28

Functions and Measurements

3 Functions and Measurements
Making a Basic Measurement 30
Measuring Multiple Signals 35
Measuring a Low-Level Signal 44
Improving Frequency Resolution and Accuracy 49
Tracking Drifting Signals 51
Making Distortion Measurements 53
Stimulus Response Transmission 59
Measuring Stop Band Attenuation of a Lowpass Filter 61
Channel Scanner 64
Spectrum Monitor 69
Reflection Measurement 71
Making a Power Measurement with USB Power Sensor 74
Demodulate the AM/FM signal 84
Analysis the Modulated Signals 87
Measuring Channel Power 93
Files Operation 95
This chapter provides information on the analyzer functions and specific measurements capabilities of the spectrum analyzer.
29
Functions and Measurements

Making a Basic Measurement

Making a Basic Measurement
This section provides information on basic operations. It assumes that you are familiar with the front and top panel buttons and keys, and display annotations of your analyzer. If you are not, please refer to “Front Panel Overview” on
page 4 and “Rear Panel Overview” on page 9.
For more details on making measurements, please refer to
“Key Reference” on page 101”.

Using the Front Panel

This section provides you with the information on using the front panel of the spectrum analyzer.
Entering Data
When setting the measurement parameters, there are several ways to enter or modify the value of the active function:
Knob Increments or decrements the current value.
Arrow Keys Increments or decrements the current value by a step unit.
30
Numeric Keys Enters a specific value. Then press the desired terminator
(either a unit softkey, or [Enter] hardkey).
Unit Softkeys Terminate (enter) a value with a unit softkey from the menu.
Enter Key Terminates an entry when no unit of measure is required, or
the instrument uses the default unit.
ESC Key To delete
the current input digit prior to entering the value.
Using Softkeys
Softkeys are used to modify the analyzer function parameter settings. Some examples of softkey types are:
Toggle Turn on or off an instrument state.
Submenu Displays a secondary menu of softkeys, {More}.
Choice Selecting from a list of standard values or filenames.
Adjust Highlights the softkey and sets the active function.
Functions and Measurements
Making a Basic Measurement

Presetting the Spectrum Analyzer

Preset function provides a known instrument status for making measurements. There are two types of presets, factory and user:
Factory Preset When this preset type is selected, it restores the analyzer to
its factory- defined state. A set of known instrument parameter settings defined by the factory.
User Preset Restores the analyzer to a user- defined state. A set of user
defined instrument parameter settings saved for assisting the user in quickly returning to known a instrument measurement setup.
Press System > Pwr on/Off Preset > Preset Type to select the preset type.
When Preset Type is set to Factory, pressing Preset triggers a factory preset condition. The instrument will immediately return to the factory default instrument parameter setting.
When Preset Type is set to User, pressing Preset triggers a user preset condition. The instrument will immediately return to the user defined instrument parameter setting.
When Preset Type is set to Last, pressing Preset triggers a last power- off condition. The instrument will immediately return to the parameter setting when it is power off last time.
Setting up a User Preset
To quickly return to instrument settings that are user defined, perform the following steps to save the instrument state as the user- defined preset:
1 Set the instrument parameters to the values and settings
necessary for the user preset state. This would include the frequency, span, amplitude, BW, and measurement type and any other setup details desired.
31
Functions and Measurements
Making a Basic Measurement
2 Press System > Pwr on/off Preset > Save User, to save the
Attenuation 20 dB(Auto) Average Type Log Pwr
Center frequency 3.5 GHz Average Number 100 (Off)
Reference offset 0 dB Peak Excursion 6 dB
Start Frequency 0 Hz Peak Threshold -140 dBm
Stop Frequency 7 GHz Peak Search Type Max
Hi-Sensitivity Off Continuous Pk Off
Span 7 GHz Graticule On
Reference level 0 dBm Y Scale On
Log scale/Division 10 dB Cal Out Off
Int Preamp Off Probe Power Off
Trace 1 Clear Write Tr i g ger Fr e e Run
Trace 2, 3, 4 Blank Marker Mode Normal
current instrument settings as the ‘user preset’ state. The user preset will not affect the default factory preset settings. User preset settings can be changed and saved at any time.
Figure 3-1 Factory Default Preset Setting
32
Sweep Time 328.5 ms(Auto) Marker Function Normal
Sweep Continuous Marker Table Off
Limit Line Off Measure Meas Off
Limit Off File Catalog Int
Impedance 50 Ω File Sort By Date
Detector Pos Peak Gated Sweep Off
Display line Off Gate Delay 12 μs
Video bandwidth 3 MHz(Auto) Gate Length 84 μs
VBW/RBW ratio 1.000 (Auto) Gate Source External
Resolution bandwidth
3 MHz(Auto) Sweep Time Rule Speed

Viewing a Signal

Refer to the procedures below to view a signal.
1 Press System > Pow on/off Preset > Preset Type > Factory to enable
the factory- defined preset state.
2 Press Preset to restore the analyzer to its factory- defined
state.
3 Connect the 10 MHz REF OUT on the rear panel to the
front- panel RF IN.
Setting the Reference Level and Center Frequency
1 Press Amplitude > 10 > dBm to set 10 dBm reference level.
2 Press Frequency > 30 > MHz to set 30 MHz to center frequency.
Setting Frequency Span
Press Span > 50 > MHz to set 50 MHz frequency span.
Functions and Measurements
Making a Basic Measurement
NOTE
Changing the reference level changes the amplitude value of the top graticule line. Changing the center frequency changes the horizontal placement of the signal on the display. Increasing the span will increase the frequency range that appears horizontally across the display.
Reading Frequency and Amplitude
1 Press Peak Search to place a marker (labeled 1) on the
10 MHz peak.
33
Functions and Measurements
Making a Basic Measurement
Note that the frequency and amplitude of the marker appear both in the active function block, and in the upper- right corner of the screen.
Figure 3-1 10 MHz Internal Reference Signal
Marker Annotation
Marker
10.000000 MHz
0.43 dBm
34
NOTE
2 Use the knob, the arrow keys, or the softkeys in the Peak
Search menu to move the marker. The marker information
will be displayed in the upper- right corner of the screen.
Changing Reference Level
1 Press Amplitude and note that reference level (Ref Level) is
now the active function.
2 Press Marker
> Marker To > Mkr-> Ref Lvl.
Changing the reference level changes the amplitude value of the top graticule line.

Measuring Multiple Signals

This section provides the information on how to measure multiple signals.

Comparing Signals on the Same Screen Using Marker Delta

The delta marker function allows the user to compare two signals when both appear on the screen at the same time.
In the following example, harmonics of the 10 MHz reference signal available are used to measure frequency and amplitude differences between two signals on the same screen. Delta marker is used to demonstrate this comparison.
1 Preset the analyzer:
Press Preset (With Preset Type of Factory)
2 Connect the rear panel REF OUT to the front panel RF IN.
3 Set the analyzer center frequency, span and reference level
to view the 10 MHz signal and its harmonics up to 50 MHz:
Press Frequency > 30 > MHz
Functions and Measurements
Measuring Multiple Signals
Press Span > 50 > MHz
Press Ampulitude > 10 > dBm
4 Place a marker at the highest peak on the display (10 MHz):
Press Peak Search
The marker should be on the 10 MHz reference signal. Use the Next Right Pk and Next Left Pk softkeys to move the marker from peak to peak.
5 Anchor the first marker and activate a second marker:
Press Marker > Mode > Delta
The label on the first marker now reads 1R, indicating that it is marking the reference point.
35
Functions and Measurements
Measuring Multiple Signals
6 Move the second marker to another signal peak using the
front- panel knob or by using Peak Search.
Press Peak Search > Next Peak or
Press Peak Search > Next Right Pk or Next Left Pk. Continue pressing the Next Pk softkeys until the marker is on the correct signal peak.
The amplitude and frequency differences between the markers are displayed in the active function block.
Figure 3-2 Delta pair marker with signals on the same screen
36
NOTE
To increase the resolution of the marker readings, turn on the frequency count function. For more information, refer to
Frequency Resolution and Accuracy” on page 49
“Improving
.
Functions and Measurements
M
Measuring Multiple Signals

Comparing Signals not on the Same Screen Using Marker Delta

The analyzer will compare the frequency and amplitude differences between two signals which are not displayed on the screen at the same time. (This technique is useful for harmonic distortion tests.)
In this example, the analyzer’s 10 MHz signal is used to measure frequency and amplitude differences between a signal on screen, and another signal off screen. Delta marker is used to demonstrate this comparison.
1 Preset the analyzer:
Press Preset (With Preset Type set to Factory)
2 Connect the rear panel REF OUT to the front panel RF IN.
3 Set the center frequency, span and reference level to view
only the 50 MHz signal:
Press Frequency > 50 > MHz
Press Span > 25 > MHz
Press Amplitude > 10 > dBm
4 Place a marker on the 50 MHz peak and then set the center
frequency step size equal to the marker frequency (10 MHz):
Press Peak Search
Press Marker > Marker To > Mkr -> CF Step
5 Activate the marker delta function:
Press Marker > Mode > Delta
6 Increase the center frequency by 10 MHz:
Press Frequency,
The first marker moves to the left edge of the screen, at the amplitude of the first signal peak.
F
37
Functions and Measurements
Measuring Multiple Signals
Figure 3-3 shows the reference annotation for the delta
marker (1R) at the left side of the display, indicating that the marker set at the 50 MHz reference signal is at a lower frequency than the frequency range currently displayed.
Figure 3-3 Delta Marker with Reference Signal Off-Screen
The delta marker appears on the peak of the 100 MHz component. The delta marker annotation displays the amplitude and frequency difference between the 50 and 100 MHz signal peaks.
38
7 Press Marker > Mode > Off to turn the markers off.

Resolving Signals of Equal Amplitude

In this example a decrease in the resolution bandwidth (RBW) is used in combination with a decrease in video bandwidth (VBW) to resolve two signals of equal amplitude with a frequency separation of 100 kHz.
Figure 3-4 Setup for obtaining two signals
Functions and Measurements
Measuring Multiple Signals
Signal Generator
N9310A RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby
On
Spectrum Analyzer
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
Remote
Standby On
7
4
102
TG SOURCE CAL OUT
RF IN
BW/
Auto
Sweep/
Avg
Tune
Trig
Local
View/
Det/
File/
Marker
Print
Trace
Display
Save Peak
Marker
Meas
MODE
Search
5896
Back
3
Enter
·
50VDC MAX 30dBm 1W MAX
RF IN 50
50MHz 10dBm
PROBE POWER
CAT
Frequency
Amplitude FM
Enter
7
4
102
RF OUT
Signal Generator
FUNCTION S
Utility
Preset
AM I/Q
Sweep
Local
File
LF Out
TriggerPulseM
MOD
9
8
On/Off
5
6
3
RF
·
On/Off
LF OUT RF OUT 50
REVERSE PWR 4W MAX 30VDC
N9310A RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby
On
Frequency
Amplitude FM
Enter
7
4
102
RF OUT
FUNCTIONS
Utility
Preset
AM I/Q
Sweep
Local
File
LF Out
TriggerPulseM
MOD
9
8
On/Off
5
6
3
RF
·
On/Off
LF OUT RF OU T 5 0
REVERSE PWR 4W MAX 30VDC
Directional Coupler
Notice that the final RBW selection to resolve the signals is the same width as the signal separation while the VBW is slightly narrower than the RBW.
1 Connect two sources to the analyzer input as shown above.
2 Set one source to 300 MHz. Set the frequency of the other
source to
300.1 MHz. Set both source amplitudes to –20 dBm.
3 Setup the analyzer to view the signals:
Press Preset (With Preset Type of Factory)
Press Frequency > 300 > MHz
Press BW > 300 > kHz
Press Span > 2 > MHz
A single signal peak is visible. See
Figure 3-5 for example.
39
Functions and Measurements
Measuring Multiple Signals
If the signal peak is not present on the display, increase the frequency span out to 20 MHz, turn signal tracking on, decrease the span back to 2 MHz and turn signal tracking off:
Press Span > 20 > MHz
Press Peak Search
Press Span > 2 > MHz
Figure 3-5 Unresolved Signals of Equal Amplitude
40
4 Change the resolution bandwidth (RBW) to 100 kHz so that
the RBW setting is less than or equal to the frequency separation of the two signals:
Press BW > 100 > kHz
Notice that the peak of the signal has become flattened indicating that two signals are present.
5 Decrease the video bandwidth to 3 kHz:
Press BW > Video BW > 3 > kHz
Two signals are now visible as shown in front- panel knob or arrow keys to further reduce the resolution bandwidth and better resolve the signals.
Figure 3-6. Use the
Functions and Measurements
Measuring Multiple Signals
Decreasing the resolution bandwidth improves the resolution of the individual signals and increases the sweep time.
Figure 3-6 Resolving Signals of Equal Amplitude
For fastest measurement times, use the widest possible resolution bandwidth. Under factory preset conditions, the resolution bandwidth is coupled to the span.
41
Functions and Measurements
Measuring Multiple Signals

Resolving Small Signals Hidden by Large Signals

1 Connect two sources to the RF IN as shown in
2 Set one source to 300 MHz at –10 dBm. Set the other source
3 Set the analyzer as follows:
4 Set the 300 MHz signal to the reference level:
This example uses narrow resolution bandwidths to resolve two RF signals that have a frequency separation of 50 kHz and an amplitude difference of 60 dB.
Figure 3-4.
to 300.05 MHz at –70 dBm.
Press Preset. (With Preset Type of Factory)
Press Frequency > 300 > MHz
Press BW > 30 > kHz
Press Span > 500 > kHz
Press Peak Search
Press Marker > Mode > Mkr -> Ref Lvl
42
NOTE
The 30 kHz filter shape factor of 15:1 has a bandwidth of 450 kHz at the 60 dB point. The half-bandwidth (225 kHz) is NOT narrower than the frequency separation of 50 kHz, so the input signals can not be resolved.
Figure 3-7 Unresolved small signal from large signal
Functions and Measurements
Measuring Multiple Signals
5 Reduce the resolution bandwidth filter to view the smaller
signal. The smaller signal will be hidden by the larger signal when the bandwidth settings are wider. Reducing the RBW setting will allow less of the larger signal to pass through the analyzer and the smaller signals peak will then rise out of the noise floor. Place a delta marker on the smaller signal:
Press BW > 1 > kHz
Press Peak Search
Press Marker > Mode > Delta
Press 50 > kHz
Figure 3-8 Resolved small signal from large signal
NOTE
The 1 kHz filter shape factor of 15:1 has a bandwidth of 15 kHz at the 60 dB point. The half-bandwidth (7.5 kHz) is narrower than the frequency separation of 50 kHz, so the input signals can be resolved.
43
Functions and Measurements

Measuring a Low-Level Signal

Measuring a Low-Level Signal
This section provides information on measuring low-level signals and distinguishing them from spectrum noise.

Reducing Input Attenuation

The ability to measure a low- level signal is limited by internally generated noise of the spectrum analyzer. The analyzers input attenuator affects the level of a signal passing through the analyzer. If a signal power level is close to the noise floor, reducing the analyzer input attenuation will help raise the signal so that it can be seen rising out of the noise.
CAUTION
Ensure that the total power of all input signals at the analyzer RF input does not exceed +30 dBm (1 Watt).
1 Preset the analyzer
Press Preset (With Preset Type of Factory)
2 Set the source frequency to 300 MHz, amplitude to –70 dBm.
Connect the source RF OUT to the analyzer RF IN.
Figure 3-9 Setup for obtaining one signal
Signal Generator
N9310A RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby On
Frequency
Amplitude
Enter
7
8
4
5
102
·
RF OUT
FUNCTION S
Utility
Preset
AM I/Q
File
Sweep
FM
Local
LF Out
TriggerPulseM
MOD
9
On/Off
6
3
RF On/Off
LF OUT RF OUT 50
REVERSE PWR 4W MAX 30VDC
Spectrum Analyzer
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
Remote
Standby On
7
4
102
TG SOURCE CAL OUT
RF IN
BW/
Auto
Sweep/
Avg
Tune
Trig
Local File/
View/
Det/
Marker
Print
Trace
Display
Save Peak
Marker
MeasMODE
Search
5896
Back
3
Enter
·
50VDC MAX 30dBm 1W MAX
RF IN 50
50MHz 10dBm
PROBE POWER
CAT
44
Functions and Measurements
Measuring a Low-Level Signal
3 Set the center frequency, span and reference level:
Press Frequency > 300 > MHz
Press Span > 2 > MHz
Press Amplitude > 40 > –dBm.
4 Move the desired peak to the center of the display:
Press Peak Search
Press Marker > Marker To > Mkr -> CF
5 Reduce the span to 500 kHz, if necessary re- center the peak:
Press Span > 500 > kHz
6 Set the attenuation to 20 dB:
Press Amplitude > Attenuation > 20 > dB
Figure 3-10 A signal closer to the noise level
Note that increasing the attenuation moves the noise floor closer to the signal level.
7 To allow more of the signal power to pass through the
analyzer, decrease the attenuation to 0 dB.
45
Functions and Measurements
Measuring a Low-Level Signal
A lower attenuation value will mean that more of the signal strength will be displayed on screen:
Press Amplitude > Attenuation > 0 > dB
Figure 3-11 Measuring a low-level signal using 0 dB Attenuation
46

Decreasing the Resolution Bandwidth

Resolution bandwidth settings affect the level of internal noise but have little affect on the displayed level of continuous wave (CW) signals. Decreasing the RBW by a decade (factor of ten) reduces the noise floor by 10 dB.
1 Refer to the procedure
page 44
2 Decrease the resolution bandwidth:
Press BW and
The low- level signal appears more clearly due to the noise level being reduced by the decrease in RBW (see
and follow steps 1, 2 and 3.
FM
Figure 3-12).
“Reducing Input Attenuation” on
NOTE
Functions and Measurements
Measuring a Low-Level Signal
Figure 3-12 Decreasing Resolution Bandwidth
A “#” mark appears next to the Res BW annotation in the lower left corner of the screen, indicating that the resolution bandwidth is uncoupled.
Uncoupled indicates that the function is in manual control mode, not auto control mode. Manual control mode allows the user to change the parameter value for that function without affecting any other settings.
The analyzer allows you to change the RBW in a 1-3-10 sequence by the data control keys. The RBWs below 1 kHz are digital and have a selectivity ratio of 5:1 while RBWs at 1 kHz and higher have a 15:1 selectivity ratio. The maximum RBW is 3 MHz and minimum is 10 Hz.

Trace Averaging

Averaging is a digital process in which each sweep of the trace returns measurement values for each point in the trace. These values are then mathematically averaged with the previous sweep trace data which has been stored in the analyzer. The amount of averaging is selected by choosing the number of trace sweeps to be included in the process. The averaging function uses the most recent trace sweep values so that the display shows any signal changes.
47
Functions and Measurements
Measuring a Low-Level Signal
NOTE
Selecting averaging, when the analyzer is auto- coupled, changes the detection mode to Sample, smoothing the displayed noise level.
This is a trace processing function and is not the same as using the average detector (as described on
page 47).
1 Refer to the procedure
page 44
2 To turn averaging on, toggle the softkey menu labeled
Average:
Press BW > Average (On)
3 Set the number of averages to 20, using the number keypad,
up and down arrows, or the knob:
Press 20, Enter
The averaging process smooths the viewed trace, low level signals become more visible (see the average number will restart the averaging process.
Figure 3-13 Trace Averaging
of this chapter and follow steps 1, 2 and 3.
“Reducing Input Attenuation” on
Figure 3-13). Changes to
48
Functions and Measurements

Improving Frequency Resolution and Accuracy

Improving Frequency Resolution and Accuracy
This section provides information on using the frequency counter function to improve frequency resolution and accuracy.
1 Press Preset (With Preset Type of Factory)
2 Connect a cable from the front panel CAL OUT to RF IN;
Press System > Service > Calibration > CAL OUT ON to toggle on and enable the 50 MHz amplitude reference signal.
3 Press Frequency > Auto Tune.
The analyzer will detect the signal peak and locate it to the center of the display (Refer to “Auto Tune” on page 111).
4 Turn the frequency counter on:
Marker > Function > Counter.
Press
5 Move the marker, with the front- panel knob, half- way down
the skirt of the signal response.
NOTE
Figure 3-14Using Frequency Counter
Counted result
The frequency and amplitude of the marker appears in the active function area (this is not the counted result). The counted result appears in the upper-right corner of the display to the right-side of C1.
49
Functions and Measurements
Improving Frequency Resolution and Accuracy
The marker readout in the active frequency function changes while the counted frequency result (upper-right corner of display) does not. For an accurate count, the marker need not be placed at the exact peak of the signal response.
NOTE
NOTE
The Frequency counter properly functions only on stable, CW signals or discrete spectral components. The marker power level must be greater than 40 dB above the displayed noise level.
6 To turn off the marker counter:
Press Marker
> Function > Off.
When using the frequency counter function, the ratio of the resolution bandwidth to the span must be greater than 0.02.
50

Tracking Drifting Signals

This section provides information on measuring and tracking signals that drift in frequency.

Measuring a Source’s Frequency Drift

The analyzer will measure source stability. The maximum amplitude level and the frequency drift of an input signal trace can be displayed and held by using the maximum hold function. Using the maximum hold function you can measure and determine how much of the frequency spectrum a signal occupies. For more information, refer to “Max Hold” on
page 162.
Use signal tracking to return a signal drifting in frequency to the center of the display. The drifting is captured by the analyzer using the maximum hold function.
1 Connect the signal generator to the analyzer RF IN.
Functions and Measurements
Tracking Drifting Signals
2 Output a signal with the frequency of 300 MHz and
amplitude of –20 dBm.
3 Set the analyzer center frequency, span and reference level.
Press Preset. (With Preset Type of Factory)
Press Frequency > 300 > MHz
Press Span > 10 > MHz
Press Amplitude > –10 > dBm
4 Place a marker on the peak of the signal:
Press Peak Search
Press Marker > Marker To > Marker to Center
Press Span > 1 > MHz
The signal is in the center of the display.
6 Measure the excursion of the signal with maximum hold:
Press Trace/Det > Max Hold
51
Functions and Measurements
Tracking Drifting Signals
7 Activate trace 2 and change the mode to continuous sweep:
8 Slowly increase the frequency of the signal generator. Your
As the signal varies, maximum hold maintains the maximum responses of the input signal. Annotation on the left side of the screen indicates the trace mode. For example, M1 S2 S3 S4, indicates trace 1 is in maximum-hold mode, trace 2, trace 3, and trace 4 are in store- blank mode.
Press Trace/Det > Select Trace > Trace 2
Press Clear Write
Trace 1 remains in maximum hold mode to show any drift in the signal.
analyzer display should look similar to
Figure 3-15 Viewing a Drifting Signal With Max Hold and Clear
Write
Figure 3-15.
52

Making Distortion Measurements

This section provides information on measuring and identifying signal distortion.

Identifying Analyzer Generated Distortion

High- level input signals may cause analyzer distortion products that could mask the real distortion measured on the input signal. Use trace and the RF input attenuator to determine which signals, if any, are internally generated distortion products.
In this example, we use the RF output of a signal generator to determine whether the harmonic distortion products are internally generated by the analyzer.
1 Connect the signal generator to the analyzer RF IN.
2 Set the source frequency to 200 MHz, amplitude to 0 dBm.
3 Set the analyzer center frequency and span:
Press Preset
Press Frequency > 400 > MHz
Press Span > 500 > MHz
Functions and Measurements
Making Distortion Measurements
Figure 3-16Harmonic Distortion
53
Functions and Measurements
Making Distortion Measurements
4 Change the center frequency to the value of the first
5 Change the span to 50 MHz and re- center the signal:
6 Set the attenuation to 0 dB:
7 To determine whether the analyzer generates harmonic
The signal produces harmonic products (spaced 200 MHz from the original 200 MHz signal) in the analyzer input mixer as shown in
harmonic:
Press Peak Search > Next Peak
Press
Marker > Marker To > Mkr -> CF
Press Span > 50 > MHz
Press Peak Search
Press Marker > Marker To > Mkr -> CF
Press Amplitude > Attenuation > 0 > dB
distortion products, first display the trace data in trace 2 as follows:
Press Trace/Det > Select Trace > Trace 2
Press Clear Write
Figure 3-16.
54
8 Allow trace 2 to update (minimum two sweeps), then store
the data from trace 2 and place a delta marker on the harmonic of trace 2:
Press Trace/Det > View
Press Peak Search
Press Marker > Mode > Delta
The analyzer display shows the stored data in trace 2 and the measured data in trace 1. The MarkerΔ indicator reads the difference in amplitude between the reference and active markers.
9 Increase the RF attenuation to 10 dB:
Press Amplitude > Attenuation > 10 > dB
Functions and Measurements
Making Distortion Measurements
Notice the MarkerD amplitude readout. This is the difference of the distortion product amplitude between 0 dB and 10 dB input attenuation settings. If the MarkerΔ absolute amplitude is approximately 1 dB for an input attenuator change, the analyzer is generating, at least in part, the distortion.
The MarkerΔ amplitude readout comes from two sources:
1) Increased input attenuation causes poorer signal- to- noise
ratio. This causes the MarkerΔ to be positive.
2) The reduced contribution of the analyzer circuits to the
harmonic measurement causes the MarkerΔ to be negative.
Large MarkerΔ readout indicates significant measurement errors. Set the input attenuator to minimize the absolute value of MarkerΔ.
55
Functions and Measurements
Making Distortion Measurements

Third-Order Intermodulation Distortion

1 Connect the equipment as shown in figure below. This
Two- tone, third- order intermodulation distortion is a common specification in communication systems. When two signals are present in a non- linear system, they may interact and create third- order intermodulation distortion (TOI) products that are located close to the original signals. System components such as amplifiers and mixers contribute to the generation of these distortion products.
This example will test a device for third-order intermodulation through the use of markers. Two sources are used, one set to 300 MHz and the other to 301 MHz.
combination of signal generators, low pass filters, and directional coupler (used as a combiner) results in a two-tone source with very low intermodulation distortion. Although the distortion from this setup may be better than the specified performance of the analyzer, it is useful for determining the TOI performance of the source/analyzer combination. After the performance of the source/analyzer combination has been verified, the device- under- test (DUT) (for example, an amplifier) would be inserted between the directional coupler output and the analyzer input.
56
Signal Generator
N9310A RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby On
Frequenc y
Amplitude FM
Enter
7
8
4
5
102
·
RF OUT
AM I/Q
9
6
3
FUNCTIONS
Sweep
MOD On/Off
RF On/Off
LF OUT RF OUT 50
Utility
File
TriggerPulseM
REVERSE PWR 4W MAX 30VDC
Preset
Local
LF Out
300 MHz LOW PASS FILTER
Signal Generator
N9310A RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby On
300 MHz LOW PASS FILTER
FUNCTIONS
Utility
Preset
AM I/Q
Frequenc y
Amplitude FM
Local
File
Sweep
LF Out
TriggerPulseM
Enter
MOD
9
7
8
On/Off
4
6
5
102
3
RF
·
On/Off
LF OUT RF OUT 50
REVERSE PWR 4W MAX 30VDC
Directional Coupler
Spectrum Analyzer
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
Remote
Standby On
7
4
102
TG SOURCE CAL OUT
RF IN
BW/
Auto
Sweep/
Avg
Tune
Trig
Local
View/
Det/
File/
Marker
Print
Trace
Display
Save Peak
Marker
Meas
MODE
Search
5896
Back
3
Enter
·
50VDC MAX 30dBm 1W MAX
RF IN 50
50MHz 10dBm
PROBE POWER
CAT
Functions and Measurements
MFM
Making Distortion Measurements
NOTE
The coupler should have a high degree of isolation between the two input ports so the sources do not intermodulate.
2 Set one source (signal generator) to 300 MHz and the other
source to 301 MHz, for a frequency separation of 1 MHz. Set the sources equal in amplitude as measured by the analyzer (in this example, they are set to –5 dBm).
3 Set the analyzer center frequency and span:
Press Preset (With Preset Type of Factory)
Press Frequency > 300.5 > MHz
Press Span > 5 > MHz
4 Reduce the RBW until the distortion products are visible:
Press BW and
F
5 Move the signal to the reference level:
Press Peak Search
Press Marker > Marker To > Mkr -> Ref Lvl
6 Reduce the RBW until the distortion products are visible:
Press BW and
7 Activate the second marker and place it on the peak of the
distortion product(beside the test signal) using the Next Peak:
Press Marker > Mode > Delta
Press Peak Search > Next Peak
8 Measure the other distortion product:
Marker > Normal
Press
Press Peak Search > Next Peak
57
Functions and Measurements
Making Distortion Measurements
9 Measure the difference between this test signal and the
second distortion product (see
Press Marker > Mode > Delta
Press Peak Search > Next Peak
Figure 3-17Measuring the Distortion Product
Figure 3-17):
58

Stimulus Response Transmission

The procedure below describes using the built- in tracking generator of the analyzer to measure the rejection of a low pass filter. A type of transmission measurement.
1 To measure the rejection of a low pass filter, connect the
equipment as shown in filter as the DUT.
Figure 3-18Transmission Measurement Test Setup
Functions and Measurements
Stimulus Response Transmission
Figure 3-18. A 370 MHz low pass
CAUTION
NOTE
Remote
Standby
On
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
7
4
102
·
TG SOURCE CAL OUT
Auto
BW/
Sweep/ Trig
Tune
Avg
Local
Det/
File/
View/
Marker
Print
Display
Trace
Save
Peak
Marker
MeasMODE
Search
5896
Back
3
Enter
50VDC MAX 30dBm 1W MAX
RF IN 50
50MHz 10dBm
PROBE POWER
CAT
DUT
2 Press Preset. (With Preset Type of Factory)
3 Set the start and stop frequencies and resolution bandwidth:
Press Frequency > Start Freq > 100 > MHz Press Frequency> Stop Freq > 1 > GHz Press Bw > 3 > MHz
4 Turn on the tracking generator and if necessary, set the
output power to - 10 dBm: Press Mode > Tracking Generator > Amplitude (On) > -10 > dBm
Excessive signal input may damage the DUT. Do not exceed the maximum power that the device under test can tolerate.
To reduce ripples caused by source return loss, use 6 dB or greater output attenuation. Tracking generator output attenuation is normally a function of the source power selected. However, the output attenuation may be controlled in the Tracking Generator menu.
59
Functions and Measurements
Stimulus Response Transmission
5 To increase measurement sensitivity and smooth the noise:
6 Connect the cable from the tracking generator output to the
7 Reconnect the DUT to the analyzer and change the
8 Measure the rejection of the low pass filter:
Press BW > 30 > kHz
Press BW > Video BW > 300 > kHz
A
decrease in displayed amplitude is caused by tracking
.
error
analyzer input. Store the frequency response in trace 3 and normalize:
Press Meas > Normalize > Store Ref (1
4) > Normalize (On)
normalized reference position:
Press Meas > Normalize > Norm Ref Posn > 8
Press Marker > 370 > MHz, Mode > Delta > 160> MHz
The marker readout displays the rejection of the filter at 160 MHz above the cutoff frequency of the low pass filter. See
Figure 3-19 as below.
Figure 3-19 Measure the Rejection Range
60
Functions and Measurements

Measuring Stop Band Attenuation of a Lowpass Filter

Measuring Stop Band Attenuation of a Lowpass Filter
When measuring filter performance, it is useful to look at the stimulus response over a wide frequency range. Setting the analyzer y- axis (amplitude) to Log scale type presents the measurement data in a usable format for determining filter characteristics.
The following example uses the tracking generator to measure the stop band attenuation of a 370 MHz low pass filter.
1 To measure the response of a low pass filter, connect the
equipment as shown in 370 MHz low pass filter.
Figure 3-20Transmission Measurement Test Setup
Figure 3-20. This example uses a
Remote
Standby On
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
7
4
102
TG SOURCE CAL OUT
5896
·
Auto
BW/
Sweep/ Trig
Tune
Avg
Local
Det/
File/
View/
Marker
Print
Display
Trace
Save Peak
Marker
MeasMODE
Search
Back
3
Enter
50VDC MAX 30dBm 1W MAX
RF IN 50
50MHz 10 dBm
PROBE POWER
CAT
DUT
2 Press Preset. (With Preset Type of Factory)
3 Set the start and stop frequencies and resolution bandwidth:
Press Frequency > Start Freq > 100 > MHz
Press Frequency > Stop Freq > 1 > GHz
Press Amplitude > Scale Type > Log
61
Functions and Measurements
Measuring Stop Band Attenuation of a Lowpass Filter
4 Set the resolution bandwidth to 10 kHz:
Press BW > 10 > KHz
CAUTION
Excessive signal input may damage the DUT. Do not exceed the maximum power that the device under test can tolerate.
5 Turn on the tracking generator and, if necessary, set the
output power to - 10 dBm
Press Mode > Tracking Generator > Amplitude (On) > -10 > dBm
6 Connect the test cable (but not the DUT) from the tracking
generator output to the analyzer input. Store the frequency response into trace 3 and normalize:
Press Meas > Normalize > Store Ref (1
7 Reconnect the DUT to the analyzer. Note that the units of
the reference level have changed to dB, indicating that this is now a relative measurement.
8 To change the normalized reference position:
Press Meas > Normalize > Norm Ref Posn > 8
9 Place the reference marker at the specified cutoff frequency:
Press Marker > Mode > Delta Pair (Ref) > 370 > MHz
10 Place the second marker at 37 MHz:
:
4) > Normalize (On)
62
Press Delta Pair (Delta) > 37 > MHz
11 In this example, the attenuation over this frequency range is
- 17.98 dB/octave (one octave above the cutoff frequency).
Functions and Measurements
Measuring Stop Band Attenuation of a Lowpass Filter
12 Use the front- panel knob to place the marker at the highest
peak in the stop band to determine the minimum stop band attenuation. In this example, the peak occurs with Delta of 563 MHz. The attenuation is - 60.93 dB. See
Figure 3-21 Minimum Stop Band Attenuation
Figure 3-21.
63
Functions and Measurements

Channel Scanner

Channel Scanner

Top/Bottom N Channel Scanner

The channel scanner - option SCN is a tool for coverage test, band clearance, and spectrum monitoring. It measures the channel power of multiple transmitted channels. The power is displayed via either the graphic bar chart or time bar chart. It allows you to set maximum 20 channels for your test. You can continuously log the scanned channel data as CSV file for post processing on your PC.
Top/Bottom N channel scan is used for searching the top/bottom channels from specified channel range according channel power.
Figure 3-22 Top N Channel Scanner Measurement
64
Please refer to the procedures below to perform a Top N channel scanner measurement.
1 Connect the antenna to RF IN connector.
2 Turn on the channel scanner function.
Press Meas > Channel Scanner
Functions and Measurements
Channel Scanner
3 Press Meas > Scan Mode > Top N to set the scan mode to
TOP N channel mode. Input the channel number you need to monitoring. Press Enter as a terminator.
4 Press Range Edit to access the submenu to edit the scan
range.
Toggle Edit Method to underline Std. It allows you to set the range according to the embedded communication standard, such as CDMA and GSM which is used specifically in different countries. Press Signal Std to select the standard. Press Recall as a terminator. Press Begin Chn and End Chn to set the begin and end channel for scanning. Press Apply to apply all the range settings.
If you toggle Edit Method to underline Custom, it allows you to set the customer-defined scan as your need.
Press Start Freq to set the scan start frequency.
Press Freq Step Size to set the frequency interval for adjacent channels.
Press Bandwidth to set the bandwidth of each frequency channel which is used to calculate the channel power.
Press Number of Chn to set the number of frequency channels which will be scanned.
Press Apply as a terminator to apply all the settings above.
5 Press Scan Start to initiate the channel scan function.
6 Press Logging Start to start recording the scan process
which includes the channel power and related channel bandwidth value to a CSV file. Press Logging Stop to end the logging. The CSVfile will be saved immediately.
65
Functions and Measurements
Channel Scanner

List N Channel Scanner

The List N channel scan is used for listing the frequency channels from specific channel standard or customized frequency bandwidth.
Please refer to the procedures below to perform a List N channel scanner measurement.
1 Connect the antenna to RF IN connector.
2 Turn on the channel scanner function.
Press Meas > Channel Scanner}
3 Press Meas > Scan Mode > List to set the scan mode to List
N channel mode. Input the channel number you need to monitoring. Press Enter as a terminator.
4 Press List Edit to access the submenu to edit the scan
range. Press Insert to insert a channel and access the channel setting submenu to customize the channel parameters.
Press Edit Method Std. It allows you to set the range according to the embedded communication standard, such as CDMA and GSM which is used specifically in different countries. Press Signal Std to load a standard file as your need. Please refer to Press Chn ID to select the specific channel according to the signal standard.
“Loading a File” on page 98 for details.
66
Or press Edit Method Custom. It allows you to set the customer- defined scan as your need.
Press Freq to set the center frequency of the current channel.
Press Bandwidth to set the bandwidth of each frequency channel which is used to calculate the channel power.
The first channel editing is completed. Press Return > Insert to add more channels as your need.
If there is a need to further edit the channel set previ­ously, press Index, rotate the knob to highlight the channel as your need for further editing.
Press Apply as a terminator to apply all the settings above.
NOTE
Functions and Measurements
Channel Scanner
5 Press Scan Start to initiate the channel scan function. 6 Press Logging Start to start recording the scan process
which includes the channel power and related channel bandwidth value to a CSV file. Press Logging Stop to end the logging. The CSV file will be saved immediately.
Please make sure the file catalog is set to INT or USB unless the logging file can not be saved successfully. When the logging file exceeds the maximum volume of the storage catalog, the logging will automatically pause and the file will be saved immediately.
Display
In channel scanner mode, the [Disp] hardkey is used to adjust the channel scanner display to review the customized scan result.
Press {Freeze} to pause the display refresh but the scan is still running.
You can sort the channel by ID or power, in ascending or decending order. Or you can press {Disp Mode} to toggle the display mode between bar mode and time mode.
Figure 3-23 Time display mode in List N channel scanner
67
Functions and Measurements
Channel Scanner

Channel Scanner Setup

To customized the channel scan process, you can press Meas > Channel Scanner > Setup to set the measurement interval and
rule in the submenu.
The Meas Interval is a value of time/distance during which the every two scan process is triggered. When it is set to off, the channel will be scanned with default settings.
The Interval Type is toggled between time and distance. When it is set to Time, the minimum interval time is 100 ms. When it is set to Distance, it allows you to measure the channel with the specific distance on the road test. The minimum interval distance is 1 meter.
The Measure Rule and Search Rule is toggled between Speed and Accuracy.
For each scan process, the analyzer will first search and find out the available channels according to the selected scan mode. Then, a precise measurement will perform to get the scan result.
68
The accuracy mode in measurement rule is set as default to get the precise measurement for the searching- out channels.
The speed mode will speed the measurement but reduce the measurement accuracy. It is set to default in search rule to search out the available channels approximately.
The To p Cyc l e number is the searching times for each measurement. It allows you to measure the specific searching- out channels continuously.

Spectrum Monitor

Functions and Measurements
Spectrum Monitor
The scrolling three- dimensional display is noted for its ability to track the frequency and power behavior over the time, particularly intermittent signals. Spectrum Monitor may be used to analyze the stability of a signal over the time, or to identify intermittent interference signals in communications systems.
The X- axis represents frequency, and amplitude is represented by color; red for a strong signal and blue for noise floor. The Y-axis represents time, with the trace from the newest sweep displayed at the bottom of the screen. Earlier traces move up toward the top with each new sweep. Two coupled markers allow the user to place a marker on any trace in the Spectrum Monitor and view the normal spectrum for the time of that sweep. The time interval between sweeps can be adjusted, and up to 1500 traces can be displayed and saved.
Figure 3-24 Spectrum Monitor Submenu
Please refer to the procedures below to perform a spectrum monitor measurement.
1 Connect the omni antenna to RF IN connector.
2 Set the center frequency and span.
Press Frequency > 1.8052 > GHz
69
Functions and Measurements
Spectrum Monitor
Press SPAN > 5 > MHz
3 Turn on the spectrum monitor function.
Press Meas > Spectrum Monitor.
4 Set the update interval time between two frames.
Press Update Int. On, rotating the knob or pressing numeric keypads to set the update interval time.
5 Turn on Marker 1 and Marker 2.
Press Marker 1 > State On, Marker 2 > State On.
6 Press Frequency, rotate knob to change the frequency of
marker as you want.
7 Press Meas > File Logging > Start Save to save the spectrogram
data. After pressing Stop Save, the spectrogram will be saved into a trace or CSV data file for your future use.
Please refer to saved spectrogram trace file.
“Loading a File” on page 98 to load the
Spectrum Monitor Audio Alert
The spectrum monitor provides an audio alert with a frequency dependant on the different control parameters. This alert needs two markers to specify the measure channel. The Band Power of the channel specified by two markers is measured without the audio alert. The Alert Limit is set to specify the lowest band power that makes the buzzer beep.
Please refer to the procedures below to turn on the audio alert.
1 Press Meas > Spectrum Monitor to access into the spectro-
gram mode.
2 Press Marker > More > Audio Alert to enable the audio alert.
70

Reflection Measurement

Press MODE > Reflection Measurement to access the cable and antenna measurement submenu. There are three different measurement mode for your selection:
Reflection measurement
One port cable loss
Distance to fault

Preparation

Before making a measurement in a specific mode, the HSA requires a calibration to make sure the measurement accuracy. A new calibration must be performed if any components used in the calibration setup are changed. For example, a short coaxial cable used to connect the tester to the DUT
Press Frequency and Span to set the measurement frequency range. It is best to keep the calibration frequency range as close as possible to the actual sweep frequencies necessary for the measurements. Calibrating over a larger frequency range than the actual measurement range will reduce the accuracy of the test results.
Functions and Measurements
Reflection Measurement
Calibrate with a T-combo calibrator
T-combo calibrator This is a T-shaped mechanical calibrator. It consists of a calibrated open, short and 50 Ω load. Using the T- combo calibrator will result in a more accurate calibration than when using the an electronic calibrator.
Using a T- combo calibrator is a three step process requiring the user to manually change the open- short-load components. The default order of the calibration process is open-short-load.
1 Press Frequency > Calibrate to bring up the dialog box for
calibration instruction. Follow the on- screen instruction to connect the Open component of the T- combo calibrator to
71
Functions and Measurements
Reflection Measurement

Measuring Cable Reflection

the RF OUT connector of the analyzer. Press ENTER to start the open calibration.
2 Follow the instruction to connect the short component of
the T- combo calibrator to the RF OUT connector of the tester. Press ENTER to start the short calibration.
3 Connect the Load component of the T- combo calibrator to
the RF OUT connector. Press ENTER to start the load calibration.
A Calibrated indicator is shown on the upper left corner of the screen when the calibration is completed. Then, the analyzer is ready for your further cable measurement.
Press Mode > Reflection Measurement > Reflection Measurement to select the Reflection Measurement Mode. It allows you to set the Y axis of the graticule as return loss, VSWR or reflection coefficient value.
72

Measuring Distance to Fault

For this measurement, an internally generated signal emits from the RF OUT port of the tester and is transmitted to the cable and antenna system (DUT). The tester receives and processes the signal power reflected from faults and imperfections. In performing a DTF measurement, the tester uses frequency domain reflectometry. The transmitted and reflected signals contain information about the distance to the faults. This information is used to determine the physical distance to the faults. The tester displays the physical distance to the probable faults, corrected for cable loss and the velocity propagation factor of the cable.
Functions and Measurements
Reflection Measurement
Perform a DTF measurement
To make a DTF-return loss measurement over a frequency range of 50 to 400 MHz, perform the following steps:
1 Press the MODE > Reflection Measurement > Distantce To Fault to
select the DTF mode.
2 Press FREQ > Start > 50 > MHz to set the start frequency to
50 MHz.
3 Press Stop > 400 > MHz to set the stop frequency to 400
MHz.
4 Press Calibrate to bring up the calibration process guidance.
Follow the instructions to perform a calibration. For more information about calibration, refer to
page 71
5 Remove the calibrator and connect the cable under test
(DUT) to the RF OUT connector on the top panel of the tester.
The DTF measurement starts and the results are displayed on the screen. A peak in the horizontal trace indicates a defective location.
.
“Preparation” on
73
Functions and Measurements

Making a Power Measurement with USB Power Sensor

Making a Power Measurement with USB Power Sensor
The N9322C spectrum analyzer supports the U2000 and U2020 series USB power sensors.
The U2000 Series USB power sensors do not need manual calibration and zero routines performed. Calibration and zeroing are performed without removing the power sensor from the source, through internal zeroing. With internal zeroing of U2000 Series USB power sensors, there is no need to disconnect the sensor or power- off the DUT. The U2000 Series do not require 50 MHz reference signal calibration, allowing the factory calibration to ensure measurement accuracy. For best accuracy, users are recommended to perform external zeroing for input signals below -30 dBm for best accuracy.
CAUTION
NOTE
The maximum power for the RF IN port and the RF OUT port of the analyzer is +30 dBm (1 W). The maximum power for the U2000 Series Power Sensor port is +25 dBm (316 mW). When directly coupled to a base station, the test set can be damaged by excessive power applied to any of these three ports.
To prevent damage in most situations when directly connect an analyzer to a base station, use a high power attenuator between the analyzer and the BTS.
If you suspect other signals may be present, it is recommended that you turn off all the other channels and measure average power only on the signal of interest. Another option is to measure channel power (which is less accurate), that filters out all other channels (signals).
74
Functions and Measurements
Making a Power Measurement with USB Power Sensor

Making an Average Power Measurement

The average power measurements provide a key metric in transmitter performance.
Base station transmit power must be set accurately to achieve optimal coverage in wireless networks. If the transmit power is set too high due to inaccurate power measurements, undesired interference can occur. If the transmit power is set too low, coverage gaps or holes may occur. Either case affects system capacity and may translate into decreased revenue for service providers. Average power can be measured for the channel of interest while the base station is active. All other channels should be inactive. Average power is a broadband measurement. If other signals are present the analyzer will also measure their power contributions.
Figure 3-25 Connection with base station
To make an average power measurement, you connect the power sensor and cable, zero and calibrate the meter, before making a measurement. Zeroing of the power meter will occur automatically:
Every time you use the Power Meter.
When a 5 degree C. change in temperature occurs.
Whenever you change the power sensor.
Every 24 hours.
75
Functions and Measurements
Making a Power Measurement with USB Power Sensor
Before measuring low level signals—for example, 10 dB above the lowest specified power for your power sensor.
Calibrating the power meter every time you cycle the power on and off.
In most situations, you can press Zero to complete the two steps (zero and cal) together.
Follow the steps below to make a basic average power measurement.
1 Press Preset. (With Preset Type of Factory)
2 Press Mode > Power Meter to access the power meter mode.
3 Zero and calibrate the meter. Press Zero to make a Zero
operation of the power sensor followed by a calibration operation.
4 Connect the power sensor to the USB port of N9322C. The
analyzer will recognize the power sensor automatically.
76
CAUTION
To prevent damage to your sensor, do not exceed the power levels below. U2000/1/2A +25 dBm (320 mW) average, 20 VDC
+33 dBm (2 W) peak, <10 μs For more information on the maximum power of U2000 series power sensors, please refer to the Operating and Service Guide of Keysight U2000 Series USB Power Sensors.
5 Connect the external attenuation, if required, with the power
sensor used.
6 Connect the analyzer to the base station. Connect the power
sensor to the signal to be measured. A typical BTS connection is after the output of the power amplifier and duplexer. The connection requires a cable and two Type- N barrel connectors, in addition to the power sensor and external attenuator.
Functions and Measurements
Making a Power Measurement with USB Power Sensor
7 Press Meas Setup > Disp Range to set the display range of the
measurement. when it is set to On, the display range will be adjusted automatically according to the signal power dynamic range. When it is set to Off, press To p and Bottom to set the display range manually.
8 Press General Setup > Freq to set the power frequency for the
DUT. The current frequency will display in the bottom line.
9 Press Power Range > Auto Power to toggles the measurement
range between On and Off. When it is set to On, the power range will be adjusted automatically. Use auto range when you are not sure of the power level you are about to measure. When it is set to Off, press Range to select the upper range or lower range. The lower range uses the more sensitive path and the upper range uses the attenuated path.
Table 1 Power Sensor Range
Sensor Lower range Upper range
U2000/1/2/4A
U2000/1/2H
U2000/1B
–60 dBm to –7 dBm –7 dBm to +20 dBm
–50 dBm to +3 dBm +3 dBm to +30 dBm
–30 dBm to +23 dBm +23 dBm to +44 dBm
77
Functions and Measurements
Making a Power Measurement with USB Power Sensor
10 Press Meas > Meas Seup > Disp Mode to set the display in
meter. chart or trace display. the measurement result will displays on the screen as shown.
Figure 3-26 Power Measurement
78
Making a Power Measurement with USB Power Sensor

Making a Peak Power Measurement

The Keysight U2020 X-series USB peak and average power sensors enable peak power measurements that have the same accuracy as measurements obtained using traditonal peak power meters, while providing the advantages of being a very compact and portable form factor.
Follow the steps below to make a peak power measurement.
1 Press Mode > Power Meter to access the power meter mode.
2 Connect the power sensor to the USB port of N9322C. The
sensor will automatically initialize, perform zeroing (While remianing connectored to the device under test), and calibrate without connecting to an external source.
NOTE
You can manually perform zeroing and calibration by perssing Zeroing > Zeroing and Calibration at any time after the sensor is connected
and initialized.
3 Press General Setup > Freq to set the frequency of the signal
under test.
As a wideband device, the power sensor has a frequency response. The power sensor utilizes a complex correction algorithm to remove power measurement error due to linearity error, frquency response, and temperature drift to generate a more accurate measurement.
Functions and Measurements
79
Functions and Measurements
Making a Power Measurement with USB Power Sensor
The analyzer provide the customer preset function with U2020 X- series power sensor. Press Meas > Preset and the predefined protocol list will pop up. Select the wanted item and press Preset as a terminator. The related settings will be set automatically.
Figure 3-27 Customized Preset the U2020 Power Sensor
80
4 Press Meas Disp > Disp Mode > Trace to find the measurement
results as
Figure 3-28 Peak Power Measurement
Marker Readings
Figure 3-28 below.
Gate Results Pulse Parameter Results

Power Meter Settings

Please find the sections below to find the general power meter measurement settings.
Setting the Power Meter Top and Bottom End-Points
Setting the end points (Disk Range) close to expected measurement value changes the sensitivity of the Power Meter scale resolution. This is an advantage in viewing small changes in power. However, this will not affect the overall range of the sensor. This is only available for Meter and Chart display mode. For Trace display, please press Meas > Peak Setup > Trace Setup to set the X and Y scale.
1 Press Meas > Meas Disp > Dis Range to access the end-points
menu.
2 Press To p and enter the maximum scale value using the
numeric keypad. Then press dBm to complete the setup.
3 Press Bottom and enter the minimum scale value desired
using the numeric keypad. Then press dBm to complete the setup.
Functions and Measurements
Making a Power Measurement with USB Power Sensor
Setting Power Meter Resolution
There are four levels of power meter resolution. Higher resolutions provide more accuracy but slow the measurement speed. The resolution is only available for U2000 series power meter and free- run trigger mode in U2020 series power meter..
Press Meas > Meas Disp > Resolution. Select 1,2,3 or 4. The four options represent different resolution as follow:
[1] = 1
[2] = 0.1
[3] = 0.01
[4] = 0.001
81
Functions and Measurements
Making a Power Measurement with USB Power Sensor
Set the Step Detection
Press Meas > General Setup > Average > Step Detect to set the
step Detect On and Off. When it is set to on, the filter will
re- initialize upon detection of a step increase or decrease in
the measured power to reduce the filter settling time after a
significant step in the measured power, Step detection can
be set in both manual and automatic filter modes.
Set the Measurement Gate
Press Meas > Peak Setup > Gate Setup to set the start time and gate length for a measurement gate. A measurement gate allows measurements to be performed on particular sections of the input signal. The gate is defined by a start time relative to the trigger instant and a duration. Signal samples acquired during the time interval specified by the gate are used for the measurements in that gate. A system of up to four independent gates is provided.
82
Set the Trigger
Press Meas > Peak Setup > Trig/Acq to set the trigger parameters listed below for the measurement gate.
Acqn Set the single, free run or continuous trigger mode.
Select to enable trace and auto trigger delay for the single and continuous trigger modes. For the free run mode, only auto trigger delay can be enabled.
Source Set the trigger source to internal channel or an
external source.
Mode Enable auto level or manually set the trigger level if
the internal trigger source is selected.
Level sets the trigger level manually when the Mode is set
to manual.
Functions and Measurements
Making a Power Measurement with USB Power Sensor
Delay Set the delay time to be applied between the trigger
event and all the gate start times. This allows you to time- shift all the gates by the same amount with one setting change.
Slope Select the positive or negative slope type to
determine if the trigger event is recognized on the rising or falling edge of a signal respectively.
Holdoff Set the holdoff time to disable the trigger
mechanism after a trigger event occurs.
Hysteresis Set the hysteresis to help generate a more stable
trigger by preventing triggering unless the RF power level achieves the trigger level and the additional hysteresis value. It can be applied to both rising and falling edge trigger generation.
Hysteresis is only available for the internal trigger source and manual trigger level.
83
Functions and Measurements

Demodulate the AM/FM signal

Demodulate the AM/FM signal
The N9322C provide the AM/FM demodulation function.

Demodulating an AM Signal

The demodulation function allows you to demodulate and hear signal information displayed on the analyzer.
Simply place a marker on a signal of interest, set the analyzer in zero span, activate AM demodulation, turn the speaker on, and then listen.
1 Press Preset. (With Preset Type of Factory)
2 Connect an antenna to the RF IN connector.
3 Select a frequency range on the analyzer. For example, the
frequency range for AM broadcasts in the United States is 550 to 1650 kHz:
84
Press Frequency > Start Freq > 550 > kHz > Stop Freq > 1650> kHz.
4 Place a marker on the signal of interest:
Press Peak Search > Next Right Pk or Next Left Pk
5 Set the frequency of interest to center frequency:
Press Marker > Marker To > Mkr -> Center.
6 Change the resolution bandwidth to 100 kHz:
Press BW > 100 > kHz.
7 Press Span > Zero Span
8 Press Sweep > Sweep Time > 10 s to set the sweep time.
8 Set the detector type and turn on AM demodulation:
Press Trace/Det > Detector > Sample.
Press Marker > Function > Demod
Press Demod Setting > Demod Type AM

Demodulating an FM Signal

This section demonstrates how to demodulate and listen to an FM signal using the built- in FM demodulator.
Using the built in FM demodulator you can tune to an FM signal and view the results of the detector output as displayed in the time- domain.
1 Press Preset. (With Preset Type of Factory)
2 Use a signal source or connect an antenna to recieve FM
signal. In this example the signal is modulated at 300 MHz with FM deviation of 10 kHz and FM rate of 1 kHz.
NOTE
If you are using a broadcast FM signal, the FM channels are broadcasting between 87.7 MHz to 107.7 MHz.
3 Set the center frequency to the center of the FM signal (in
this case 300 MHz):
Press Frequency > 300 > MHz.
Functions and Measurements
Demodulate the AM/FM signal
4 Set the analyzer to zero span for time- domain analysis:
Press Span > Zero Span.
Press Sweep > Sweep Time > 5 > s.
5 Set the resolution bandwidth to capture the full bandwidth
of the FM signal.
Press BW > 100 > kHz.
6 Turn on the optional pre- amplifier to detect small signals:
Press Amplitude > Preamp On.
7 Turn on the FM demodulator:
Press Marker > Function > Demod
Press Demod Setting > Demod Type FM
85
Functions and Measurements
Demodulate the AM/FM signal
Listen to the FM signal.
Press Demod Setting > Speaker Vol, rotate the knob to adjust the speaker volume.
The headphone jack is availabe for the audio signal output alternatively.
86

Analysis the Modulated Signals

The N9322C provides the optional AM/FM (Option AMA) and ASK/FSK (Option- DMA) modulation analysis function for analysis the modulated signals. These functions can be activated by license key.

AM/FM Modulation Analysis

Optional AM/FM demodulation analysis provides modulation metrics, including carrier power, modulation rate, AM depth/FM deviation, SINAD and carrier frequency offset. User definable limits provide Pass/Fail indicators of four types:
Maximum carrier power, Maximum AM modulation index or FM deviation, Minimum AM modulation index or FM deviation, and Minimum carrier frequency offset.
The user can save the waveforms with metrics for reporting as well as the set- up parameters for future measurements or analysis. In the following example, you can use any source available (such as AM/FM radio) as long as it has AM/FM signal generation capability.
Functions and Measurements
Analysis the Modulated Signals
AM Modulation Analysis
Please refer to the procedures below for the AM modulation analysis:
1 Connect the signal generator to RF IN connector and turn
on the signal generator’s AM and RF output.
2 Turn on the AM demodulation analysis function.
Press MODE > Modulation Analysis > AM.
3 Press Frequency > Carrier Freq > 100 > MHz to set the carrier
frequency.
4 Set the attenuation state to Auto (default):
Press Amplitude > Attenuation Auto
87
Functions and Measurements
Analysis the Mod ulated Signals
5 Press Shift > Disp > Y Scale > Scale Coupling On to set the Y
scale auto coupling.
6 Set the AM depth limit and turn on the limit function:
Press Shift > Limit > AM Depth UP > 51 > % > AM Depth Low > 49
> % > Limits on
The red value in the result indicates the demodulated AM depth exceeds the limits set as above.
Figure 3-29 AM demodulation analysis
88
FM Modulation Analysis
Please refer to the procedures below for the FM modulation analysis:
1 Connect the signal generator to RF IN connector and turn
on the signal generator’s FM and RF output.
2 Turn on the FM demodulation analysis function.
Press MODE > Modulation Analysis > AM.
3 Press Frequency > Carrier Freq > 100 > MHz to set the carrier
frequency.
4 Set the attenuation state to Auto (default):
Press Amplitude > Attenuation Auto
Functions and Measurements
Analysis the Modulated Signals
5 Press Shift > Disp > Y Scale > Scale Coupling On to set the Y
scale auto coupling.
6 Set the FM deviation limit and turn on the limit function:
Press Shift > Limits > FreqDev UP > 11 > % > FreqDev Low > 9.9
> % > Limits on
The red value in the result indicates the demodulated FM deviation exceeds the limits set as above.
Figure 3-30 FM modulation analysis
89
Functions and Measurements
Analysis the Mod ulated Signals

ASK/FSK Modulation Analysis

The ASK/FSK modulation analysis function supports four display modes:
Symbol
Wave form
ASK/FSK Error
Eye Diagram.
User definable limits provide Pass/Fail indicators of four types:
Maximum carrier power
Maximum ASK modulation depth/FSK frequency deviation
Minimum ASK modulation depth/FSK frequency deviation
Maximum carrier frequency offset.
The metrics includes carrier power, ASK/FSK error, ASK depth/FSK frequency deviation, and ASK index etc. The waveform with metrics and setup parameters can be saved for reports and future Measurements.
90
In the following example, you can use any source available (such as your remote key of the car) as long as it has ASK/FSK generation capability. For ASK/FSK signal, its Filter is Nyquist and Alpha is 0.35.
ASK Modulation Analysis
Please refer to the procedures below to analysis the ASK demodulation.
1 Connect the signal generator to RF IN connector and turn
on the signal generator’s ASK modulation and RF output.
2 Turn on the ASK demodulation analysis function.
Press MODE > Modulation Analysis > ASK.
3 Press Frequency > Carrier Freq > 100 > MHz to set the carrier
frequency.
4 Press Meas > Symbol Rate > 10 > MHz to set the symbol rate.
Functions and Measurements
Analysis the Modulated Signals
5 Press Filter Setup > Ref Filter > Nyquist to set the filter type.
6 Set the attenuation state to Auto (default):
Press Amplitude > Attenuation Auto
7 Press Shift > Disp > Y Scale > Scale Coupling On to set the Y
scale auto coupling.
8 Setting the ASK depth limit and turn on the limit
function:
Press Shift > Limits > ASK Depth UP > 71 > % > ASK Depth Low >
70 > % > Limits on
9 Press Shift > Disp > Display > Eye Diagram to view the eye
diagram.
Figure 3-31 ASK Eye Diagram
91
Functions and Measurements
Analysis the Mod ulated Signals
FSK Demodulation Analysis
Please refer to the procedures below to analysis the FSK demodulation.
1 Connect the signal generator to RF IN connector and turn
on the signal generator’s FSK modulation and RF output.
2 Turn on the ASK demodulation analysis function.
Press MODE > Modulation Analysis > FSK.
3 Press Frequency > Carrier Freq > 100 > MHz to set the carrier
frequency.
4 Press Meas > Symbol Rate > 10 > ksps to set the symbol rate.
5 Press Filter Setup > Ref Filter > Nyquist to set the filter type.
6 Set the attenuation state to Auto (default):
Press Amplitude > Attenuation Auto
7 Setting the FSK deviation limit and turn on the limit:
Press Shift > Limit > FreqDev UP > 85 > kHz > FreqDev Low > 80
> kHz > Limits on
The figure below shows the FSK symbol and the measurement result.
Figure 3-32 FSK Symbol View
92
Loading...