No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from
Keysight Technologies as governed by
United States and international copyright
laws.
Manual Part Number
N5990-91400
Edition
Edition 2.1, September 2018
Keysight Technologies Deutschland GmbH
Herrenberger Strasse 130,
71034 Böblingen, Germany
Technology Licenses
The hard ware and/or software described in
this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal Acquisition
Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of
Defense FAR Supplement
(“DFARS”) 227.7202, the U.S. government
acquires commercial computer software
under the same terms by which the software is customarily provided to the public.
Accordingly, Keysight provides the Software to U.S. government customers under
its standard commercial license, which is
embodied in its End User License Agreement (EULA), a copy of which can be found
at http://www.keysight.com/find/sweula.
The license set forth in the EULA represents
the exclusive authority by which the U.S.
government may use, modify, distribute, or
disclose the Software. The EULA and the
license set forth therein, does not require
or permit, among other things, that Keysight: (1) Furnish technical information
related to commercial computer software
or commercial computer software documentation that is not customarily provided
to the public; or (2) Relinquish to, or otherwise provide, the government rights in
excess of these rights customarily provided
to the public to use, mod ify, reproduce,
release, perform, display, or d isclose commercial computer software or commercial
computer software documentation. No
additional government requirements
beyond those set forth in the EULA shall
apply, except to the extent that those
terms, rights, or licenses are explicitly
required from all providers of commercial
computer software pursuant to the FAR and
the DFARS and are set forth specifically in
writing elsewhere in the EULA. Keysight
shall be under no obligation to update,
revise or otherwise modify the Software.
With respect to any technical data as
defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS 227.7102,
the U.S. government acquires no greater
than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED "AS IS," AND IS
SUBJECT TO BEING CHANGED, WITHOUT
NOTICE, IN FUTURE EDITIONS. FURTHER,
TO THE MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DISCLAIMS
ALL WARRANTIES, EITHER EXPRESS OR
IMPLIED WITH REGARD TO THIS MANUAL
AND ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT LIMITED TO
THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT SHALL
NOT BE LIABLE FOR ERRORS OR FOR
INCIDENTAL OR CONSEQUENTIAL
DAMAGES IN CONNECTION WITH THE
FURNISHING, USE, OR PERFORMANCE OF
THIS DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD KEYSIGHT
AND THE USER HAVE A SEPARATE
WRITTEN AGREEMENT WITH WARRANTY
TERMS COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH THESE
TERMS, THE WARRANTY TERMS IN THE
SEPARATE AGREEMENT WILL CONTROL.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not proceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions
are fully understood and met.
2Keysight N5990A MIPI M-PHY Receiver Test User Guide
Contents
1 Introduction
2 N5990A M-PHY Station
Document History18
First Edition (September, 2017)18
Second Edition (February, 2018)18
Third Edition (September, 2018)18
Overview19
N5990A M-PHY Station Configuration26
Test Station Selection26
System Configuration29
Separated Low Speed Generator31
Use Bit-2100 Switch32
Oscilloscope Application Supplier Company Name33
Oscilloscope Bandwidth34
Controlling Loop Parameters and Looping Over Selected Tests5
Unipro Test Mode8
Unipro Script Generation11
16Keysight N5990A MIPI M-PHY Receiver Test User Guide
Keysight N5990A MIPI M-PHY Receiver Test User Guide
1Introduction
Document History / 18
Overview / 19
1Introduction
Document History
First Edition (September, 2017)
Second Edition (February, 2018)
Third Edition (September, 2018)
The first edition of this user guide describes functionality of software
version 1.0
The second edition of this user guide describes functionality of software
version 2.0.
The third edition of this user guide describes functionality of software
version 4.01.
18Keysight N5990A MIPI M-PHY Receiver Test User Guide
Overview
Introduction1
The Keysight N5990A Test Automation software is an open and flexible
framework for automating electrical compliance tests for digital buses
such as PCI Express or USB. It is globally marketed and supported by
Keysight Technologies as N5990A.
The Keysight N5990A for MIPI (Mobile Industry Processor Interface)
M-PHY provides the automation of the MIPI M-PHY receiver testing for the
physical layer. The tests are implemented according to the requirements of
the “MIPI Alliance Conformance Test Suite for M-PHY Physical Layer” v3.1.
These tests are designed to determine if a product conforms to
specifications defined in the MIPI Alliance Specification for M-PHY, version
3.1. Additionally, Keysight offers some custom characterization tests to
provide more details on DUT behavior beyond the limits
The software supports automatic control of the J-BERT M8020A and the
J-BERT N4903 high-performance serial BERT (Bit Error Ratio Tester). It
calibrates the stress conditions and controls all test electronic equipment
for automated receiver tolerance tests. For UniPro and UFS Phy testing the
BERT equipment can be combined with the BIT-3000 DSGA to put the
DUT into UniPro Test Mode and retrieve the Frame and Error Counters.
Figures 1 to 4 show the examples of Receiver Test setups for J-BERT
M8020A (stand-alone), J-BERT M8020A + DSGA, J-BERT N4903B
(stand-alone), and J-BERT N4903B + DSGA system configurations.
Keysight N5990A MIPI M-PHY Receiver Test User Guide19
1Introduction
Figure 1MIPI M-PHY Receiver Test Setup Example (for J-BERT M8020A System
Configuration)
20Keysight N5990A MIPI M-PHY Receiver Test User Guide
Introduction1
Figure 2MIPI M-PHY Receiver Test Setup Example (for J-BERT M8010A + DSGA
System Configuration)
Keysight N5990A MIPI M-PHY Receiver Test User Guide21
1Introduction
Figure 3MIPI M-PHY Receiver Test Setup Example (for J-BERT N4903B System
Configuration)
22Keysight N5990A MIPI M-PHY Receiver Test User Guide
Introduction1
Figure 4MIPI M-PHY Receiver Test Setup Example (for J-BERT N4903B + DSGA
System Configuration)
The Keysight N5990A for MIPI M-PHY also provides the automation of the
MIPI M-PHY transmitter testing for the physical layer. In this case, the
transmitter tests are conducted with the integration of the Keysight
U7249E MIPI M-PHY Tx Test Software. The Keysight software will control a
suitable Infiniium Oscilloscope where the U7249E application is installed.
Keysight N5990A MIPI M-PHY Receiver Test User Guide23
1Introduction
NOTE
For oscilloscope Infiniium versions lower than 6,10,00709, the Tests
1,1,6 and 1,1,7 will not work.
24Keysight N5990A MIPI M-PHY Receiver Test User Guide
Keysight N5990A MIPI M-PHY Receiver Test User Guide
2N5990A M-PHY
Station
N5990A M-PHY Station Configuration / 26
Starting the N5990A MPHY User Interface / 36
Configuring the DUT / 37
2N5990A M-PHY Station
N5990A M-PHY Station Configuration
Refer to the “N5990A Test Automation Software Platform Installation Guide” for instructions on how to install and start the N5990A Test
Automation software platform. After the software has been installed, two
icons are added to the desktop as shown in Figure 5 and Figure 9. One is
for the Station Configuration and the other for the M-PHY application.
Test Station Selection
You need to start the N5990A Station Configuration prior to the M-PHY
application. You can select the application (that is, M-PHY, PCI Express 3,
HDMI …) and the set of instruments used for it. To start the software,
double-click the left mouse button on the MPHY Station Configuration
icon (see Figure 5), or alternatively go to Start > All Programs > BitifEye > MPHY > ValiFrame M-PHY Station Configuration.
Figure 5Keysight M-PHY Station Configuration Icon
The Station Selection window appears with the M-Phy Station selected
by default (see Figure 6).
26Keysight N5990A MIPI M-PHY Receiver Test User Guide
Figure 6M-PHY Station Selection Window
In the Settings section, the following options are available:
Database Option
In case you have purchased the option N5990A opt. 001, the interface to
SQL databases (and web browsers) is available. You can establish a
connection to the database application server by clearing the default
Database Offline check box and entering the IP address of the server.
N5990A M-PHY Station2
Results Viewer
Here, you can select the viewer for test results from the following options:
• Microsoft Excel
• HTML.
Sounds
A warning sound can be activated in different states of the program:
• End of Sequencer plays the selected sound at the end of a
sequence.
• Connection diagram plays the selected sound every time a
connection diagram pops up.
• Dialog Prompt plays the selected sound at each dialog prompt.
There are a few different sounds available for selection:
Keysight N5990A MIPI M-PHY Receiver Test User Guide27
2N5990A M-PHY Station
• None (deactivates the sound)
• Car brake
• Feep Feep
• Ringing
• TaDa
• Tut
You can also hear the selected sound by clicking Play before you set the
sound of your choice.
After the station has been selected, click Next to continue. The Station Configuration window is displayed as shown in Figure 7. It shows the
possible instrument combinations that can be used for M-PHY testing. It
contains the following options:
1 System Configuration
2 Separate Low Speed Generator
3 Use Bit-2100 Switch
4 Scope App Provider
5Oscilloscope bandwidth
Figure 7M-PHY Configuration Window
28Keysight N5990A MIPI M-PHY Receiver Test User Guide
System Configuration
M8020A Stand-alone
M8020A + DSGA
N5990A M-PHY Station2
• To use this setup, select M8020A from the Configuration drop-down
and leave the Separated Low Speed Generator option unchecked.
• This system configuration only requires M8020A. The instrument works
as a Generator and as an Error Detector, if necessary.
• It supports three operation modes: Loopback, Offline and Custom BER
mode, respectively.
•For UniPro Test mode, this configuration is not supported anymore due
to training pattern generation memory restrictions. This is described
below in M8020A + DSGA.
• To use this setup, select M8020A from the Configuration drop-down
and check the Separated Low Speed Generator option.
• This system configuration is recommended for the UniPro test mode. In
this case, the DSGA is used for the DUT configuration and to receive
the frame and error counters, while the M8020A is used to generate the
test pattern as well as the frame and error counter requests. If DUT
configuration must happen in the same lane, then that will be tested. It
is mandatory to use two switches to alternate the DUT inputs between
the two generators.
• The advantage of M8020A + DSGA setup with respect to the
standalone M8020A comes from the much lower DSGA data rate. Since
PWM data with M8020A is generated by bit multiplication, and
depending on test conditions, the M8020A can run at 9,984 Gbps, the
generation of the training sequence takes much longer than the DSGA,
and the pattern may not even fit into the M8020A instrument's
memory.
•The DSGA Error Detector requires that the DUT responses come in
PWM Mode, either Bursted or Continuous.
N4903B (stand-alone)
• To use this setup, select JBERT from the Configuration drop-down and
leave the Separated Low Speed Generator option unchecked.
• This system configuration only requires the N4903B JBERT. The
instrument works as Generator and as Error Detector, if necessary.
• It supports Loopback, Offline and Custom BER operation modes. The
N4903B stand-alone is not able to operate in UniPro PHY Test Mode
and has to be combined with DSGA.
Keysight N5990A MIPI M-PHY Receiver Test User Guide29
2N5990A M-PHY Station
N4903B + DSGA
• To use this setup, select JBERT from the Configuration drop-down and
check the Separated Low Speed Generator option.
• This system configuration works similarly to the M8020A + DSGA
configuration. See section “M8020A + DSGA” for more information on
the system configuration.
DSGA
• To use this setup, select DSGA from the Configuration drop-down. The
DSGA standalone configuration is only supported for transmitter
testing. It sends the training sequence to the DUT and the actual
testing is performed by the scope app.
• For any receiver or transmitter test configuration, having a DSGA
enables sending hard ware reset signals to the DUT. This allows
seamless transition from one test mode to the next without user
intervention.
30Keysight N5990A MIPI M-PHY Receiver Test User Guide
Separated Low Speed Generator
The M8020A and the N4903B instruments can be used either as
stand-alone or in combination with DSGA. Select the Separated Low Speed Generator check box to add DSGA to the system configuration.
Refer to “M8020A + DSGA”and “N4903B + DSGA” in the previous section
for more details.
N5990A M-PHY Station2
Keysight N5990A MIPI M-PHY Receiver Test User Guide31
2N5990A M-PHY Station
Use Bit-2100 Switch
The Bit-2100 Switch can be used in two cases:
1If DSGA is included in the setup (that is, the Separated Low Speed
Generator Check box is checked), then the switch will be used to
switch between the DSGA and the BERT system. In order to
alternate the input signal, send it to the DUT. Two 2:1 switch
modules are required for this purpose. If a relay switch is integrated
in DSGA, this option must not be selected.
2If DSGA is not included, the switch will be used to switch across
several DUT lanes. This will allow you to connect and test different
DUT lanes at the same time, thereby reducing the number of
connection setup changes. Two 6:1 switch modules are required for
this purpose.
In order to use a Keysight 2100 Series Switch System for automated
receiver test, the Keysight option 002 (Switch System Support, available
from Keysight as BIT-2001-0002-0), is required.
32Keysight N5990A MIPI M-PHY Receiver Test User Guide
Oscilloscope Application Supplier Company Name
Agilent Technologies was the supplier of the initial version of the MIPI
M-PHY oscilloscope application software. After the foundation of Keysight
Technologies in 2014, Keysight supplies this application. Select the proper
company name:
• Agilent (likely holds for Oscilloscopes purchased prior to July 2014).
• Keysight (likely holds for Oscilloscopes purchased after July 2014).
In case the purchase date is unknown or you are unsure about the
software supplier name, open the U7249E software on the Oscilloscope
and choose Help > About... in the menu. The About dialog displays the
supplier company name.
N5990A M-PHY Station2
Keysight N5990A MIPI M-PHY Receiver Test User Guide33
2N5990A M-PHY Station
Oscilloscope Bandwidth
Auto
If the Auto option is selected, Keysight discovers the maximum supported
oscilloscope bandwidth automatically and adjusts to the recommended
value for each gear. Firmware version 5.20 or above is required for this
purpose.
Manual
If the oscilloscope firmware is below version 5.20, it can be complicated for
Keysight to find the maximum oscilloscope bandwidth automatically. To
avoid this situation, select the Manual option to set the desired bandwidth
manually.
Once the M-PHY station is configured, you must set the Instrument Address. An example for the instrument configuration is as shown below
(see Figure 8).
Figure 8M-PHY Instrument Configuration Window
After the installation process, all instruments are configured by default in
Offline mode. In this simulation mode, the hardware does not need to be
physically connected to the test controller PC. N5990A cannot connect to
any instrument in this mode. In order to control the instruments that are
connected to the PC, the instrument address must be entered. The
address depends on the bus type used for the connection, for example,
GPIB (General Purpose Interface Bus) or LAN (Local Area Network). Most
34Keysight N5990A MIPI M-PHY Receiver Test User Guide
N5990A M-PHY Station2
of the instruments used in the M-PHY station require a VISA (Virtual
Instrument System Architecture) connection. To determine and validate
the VISA address, do the following:
1 Run the VISA Connection Expert. To access the VISA Connection
Expert, right-click the Keysight IO Libraries Suite icon in the task bar
and select Connection Expert.
2 Enter the instrument addresses in the Station Configuration wizard,
for example, by copying and pasting the address strings from the
Connection Expert entries.
3Click Apply Address before selecting the Offline check box to set
the instruments that are required to be online.
4Click Check Connections to verify that the connections for the
instruments are established successfully.
If anything is wrong with the Instrument Address, a window is displayed
with a message describing the problem.
Keysight N5990A MIPI M-PHY Receiver Test User Guide35
2N5990A M-PHY Station
Starting the N5990A MPHY User Interface
Double click the N5990A MPHY icon that appears on the desktop (see
Figure 9). Alternatively, go to Start > All Programs > BitifEye > MPHY >
ValiFrame MPHY.
Figure 9N5990A MPHY Icon
The following window is displayed (see Figure 10).
Figure 10N5990A MIPI M-PHY User Interface
You need to configure the test parameters before running any test or
calibration procedure. Click Configure DUT to pop up the Configure Product window. (See Figure 11).
36Keysight N5990A MIPI M-PHY Receiver Test User Guide
Configuring the DUT
N5990A M-PHY Station2
In the Configure DUT panel (see Figure 11 and Figure 12), you can select
DUT parameters such as Product Type, Test Type and Protocol, and also
the parameters related to the Receiver/Transmitter test configuration.
The selected parameters are later used in the calibration and test
procedures shown in the N5990A main window.
Figure 11Configure DUT Panel for Receiver
Keysight N5990A MIPI M-PHY Receiver Test User Guide37
2N5990A M-PHY Station
Figure 12Configure DUT Panel for Transmitter
38Keysight N5990A MIPI M-PHY Receiver Test User Guide
N5990A M-PHY Station2
Figure 13Configure DUT panel (for the database connection)
If you select the Database Offline option in the N5990A M-PHYStation Configuration window (see Figure 6), the Configure DUT panel appears as
shown in Figure 13. To configure the DUT with a database connection, do
the following:
1 Enter values (any) for Product Number and Serial Number.
2Click Register Product to register the database connection with the
provided values. Clicking Register Product, enables the OK button.
3Click OK to configure the DUT with the selected parameters.
The procedures that are run with the database settings are stored at
N5990A Webviewer and these can be viewed by selecting the Product Number and Serial Number (these values must be the same as the values
provided in Figure 13.
Keysight N5990A MIPI M-PHY Receiver Test User Guide39
2N5990A M-PHY Station
DUT Parameters
The DUT Parameters are listed below in Table 1.
Table 1 DUT Parameter List
Parameter NameParameter Description
Product
Product NumberThe selected name will be used to identify the product when the “Database Offline” option is selected.
Serial NumberThe selected serial number will be used to identify the product when database option is selected.
Product TypeThe product type can be selected as:
Test TypeBy default, the test type is set to Phy Test: physical layer test.
DescriptionA description of the DUT ca be added.
Num of ChannelsHere, you can select the number of DUT input channels (Receiver tests) or output channels (Transmitter
ProtocolIt can be selected as follows:
Test
User NameA user name can be added to the test information.
CommentA comment can be added to the test information.
• Receiver
• Transmitter
If Receiver is selected, the main window will d isplay the set of tests corresponding to receiver testing:
HS-RX, PWM-RX, SYS-RX and SQ-RX tests.
If Transmitter is selected, the main window will d isplay the set of tests corresponding to the transmitter
testing:
HS-TX, PWM-TX and SYS-TX tests.
tests) to be tested.
• M8020A allows to test up to 4 channels.
• N4903B allows to test up to 2 channels. When used together with the Bitifeye 2100 Switch, up to 4
channels can be tested.
• DigRFv4
• LLI (Low Latency Interface)
• UniPro (Unified Protocol)
• SSIC (Super Speed Inter Chip)
• PCIe (PCI Express)
• UFS (Universal Flash Storage
Initial Start DateTime stamp of the start of the current test session.
Last Test DateTime stamp of the last test conducted in the current session.
Compliance ModeIn this mode, the tests are cond ucted as mandated by the CTS, the test parameters used in the cal ibration
40Keysight N5990A MIPI M-PHY Receiver Test User Guide
and test procedures are shown but cannot be modified by the user.
Parameter NameParameter Description
N5990A M-PHY Station2
Expert ModeCalibrations and tests can be conducted beyond the limits and constraints of the CTS; the test
parameters used in the calibration and test procedures are shown and can be modified by the user.
For Receiver and Transmitter Test Configuration
HS GearsThe following HS-Gears are available:
• GEAR 1-A
• GEAR 1-B
• GEAR 2-A
• GEAR 2-B
• GEAR 3-A
• GEAR 3-B
• GEAR 4-A
• GEAR 4-B
The selection of gears is not limited. If "x" number of gears are selected, the HS terminated and
non-terminated tests are available "x" times, in respective test sub-groups.
For receiver test, GEAR 3-A and 3-B are only available for spec. 3.10 and 4.10. GEAR 4-A
Nominal Data RateIt sets the reference clock frequency to achieve the nominal Gear x-B data rate value as given in the
specification. In case of custom data rates, the reference clock is adjusted to achieve the HS custom data
rate.
Nominal Ref. ClockFor the selected reference clock frequency, the generated data rate is calculated as a multiple of the
reference clock.
Additional GearsPWM Gear 1 is always tested. Add itionally, PWM Gears from Gear 2 to Gear 7 can be selected.
The selection of gears is not limited. If "x" number of gears are selected, the PWM terminated and
non-terminated tests are available "x" times, in respective test sub-groups.
This option is only visible for PWM device type.
M-PHY Spec VersionSelect the version using the drop down menu:
• 3.10
• 4.10
Ref Clock FrequencyThe frequency can be chosen as follows:
Default LevelsBrings up a dialog where the voltage levels and the default values can be selected. Refer to Default
Default TimingBrings up a dialog where the timing parameters (Prepare Length, Stall, Sleep, Sync Length) can be
Default SequencesBrings up a dialog where the HS, LS, and Squelch sequence files can be selected. Refer to “Default
Receiver Test Configuration
Keysight N5990A MIPI M-PHY Receiver Test User Guide41
• 19.2 MHz
• 26 MHz
• 38.4 MHz
• 52 MHz
Levels for more details.
defined. Refer to “Default Timing” for more details.
Sequences” for more details.
2N5990A M-PHY Station
Parameter NameParameter Description
Custom HS DRIt allows addition of custom data rate values to be tested. Refer to “Custom HS DR” for more details.
Custom Data RateIt displays the custom data rates values selected in the "Custom HS DR" dialog.
Device TypeThe type of the DUT can be selected as follows:
PWM G1 DR (Min;Max)It displays the minimum and maximum PWM Gear 1 data rate. This option is only visible for PWM device
Special ParametersBrings up a dialog where ad vanced parameters such as the BER reader type and its address can be
Target BER HSIt is the target BER for the HS mode.and can be selected using the drop-down menu.
Target BER LSIt is the target BER for the LS mode and can be selected using the drop-down menu.
M8020A ISI PropertiesBrings up a dialog where the M8020 ISI can be configured. Refer to for more details to “M8020 ISI
Transmitter Test Configuration
Test Group
LA_RTSelecting the check box will add the test procedures for the respective group in the N5990A, whereas
SA_RT• RT: TX connection is resistively terminated
LA_NT
SA_NT
Number of skew Rate StatesIf the skew state rate is more than 1, add itional Tx tests for the skew characterization will be added.
L2L Skew Supported LaneSelect the lane to be tested in the L2L (lane to lane) Skew PWM Tx test and the L2L HS Tx test.
• Type I (PWM)
• Type II (SYS Burst)
type.
selected. Refer to “Special Parameters” for more details.
Properties” section.
• NT: TX connection is non-terminated (into-open)
• LA: Large Amplitude
• SA: Small Amplitude
Connection SetupIt opens a dialog that allows the connection and probing modes to be selected. Refer to “Connection
BER Reader SelectionThe BER can be measured with different methods:
SigTestCheck to use sigTest software for transmitter measurements.
42Keysight N5990A MIPI M-PHY Receiver Test User Guide
Setup” for more details.
• BERT Analyzer
• Unipro BER Reader
• Custom BER Reader
• Offline
Custom HS DR
N5990A M-PHY Station2
Click Custom HS DR in the Configure DUT Panel (see Figure 13). A
window is displayed as shown in Figure 14. It allows custom data rates to
be added to the tested data rates list.
Figure 14Custom HS Data Rate panel
Custom Data Rate
Enter the desired value into the Custom Data Rate text box and click Add.
The custom data rate is added to the list of data rates under Added Data Rates. Click Delete to remove the selected data rate from the list.
The list of procedures that depend on a data rate increases with the
corresponding number of selected custom data rates.
Nominal Data Rate
The selected Ref. Clock Frequency is adjusted to achieve the Custom
Data Rate.
Nominal Ref. Clock
For the selected Ref. Clock Frequency, the Custom Data Rate is
calculated as a multiple of the reference clock.
Ref. Clock Frequency
It can be set to 19.2 MHz, 26MHz, 38.4MHz or 52MHz.
Keysight N5990A MIPI M-PHY Receiver Test User Guide43
2N5990A M-PHY Station
Default Levels
Applied Data Rate, Multiplier Factor, Applied Frequency
These properties show the setting that will be applied during testing.
If Nominal Data Rate is selected, the Applied Data Rate is equal to the
Custom Data Rate, but the Applied Frequency can be sightly different
from the Ref.Clock Frequency selected.
If Nominal Ref. Clock is selected, the Applied Frequency is equal to the
Ref. Clock Frequency selected, but the Applied Data Rate can be sightly
different from the Custom Data Rate.
Click Default Levels in the Configure DUT Panel to display the Set M-PHY
Default Levels window (see Figure 15). It allows the levels (Ampl itude and
Offset) for the Data, DSGA and Reference Clock channels to be set.
Figure 15Set M-PHY Default Levels panel
44Keysight N5990A MIPI M-PHY Receiver Test User Guide
Default Timing
NOTE
N5990A M-PHY Station2
Click Default Timing to display the Set M-PHY Default Timings window
(see Figure 16).
Figure 16Set M-PHY Default Timings panel
This dialog lists all the M-PHY timing parameters. Each parameter value
can be selected and modified. Click Set Values to Default to reset all the
modifications and set all the timing parameters to their default value.
Default Sequences
Click Default Sequences (see Figure 13) to display the Select M-PHY Sequences panel (see Figure 17). In this panel, click Browse to select the
files for HS Sequence (Burst), HS Sequence (Cont), LS Sequence (Burst)
and Squelch Sequence test groups.
HS implies High Speed
LS implies Low Speed
Cont implies Continuous
Sync Pattern
The D10.5 and D26.5 symbols are used as default for sync pattern.
Keysight N5990A MIPI M-PHY Receiver Test User Guide45
2N5990A M-PHY Station
NOTE
Word Size
You can select the word size as 8 bit, 10 bit, 16 bit and 20 bit from the
drop-down menu.
Re-Init Sequence after Reset DUT
If you select the Re-Init Sequence check box, the pattern generator
sequencer will restart for every test step. It will bring up the link while the
signal impairments are being applied, which can be harder on the DUT. If
you do not select the Re-Init Sequence check box, the sequencer will
bring up the link only once during the Init.
Wait for Manual Break
This option is available when you select the Re-Init Sequence check box. A
dialog to manually restart the DUT, is displayed for every test step. This is
useful for DUT's that come out of test mode when loop-back is being used.
Switch with
Under certain conditions, the switch from DSGA to BERT makes some
devices come out of test mode. Using the Switch with property, you can
retain the devices in test mode, by modifying how the switching is
performed. The following selections are possible:
• Test amplitude: The levels are not changed when switching
• Nominal amplitude: before switching, the BERT is set to the default
200/100 mV levels, and after switching the original amplitude is
restored.
• 0 Amplitude: The amplitude in both BERT and DSGA is set to 0 before
switching and then restored to the original value.
Generate Unipro Scripts
Click Generate UniPro Scripts to generate the scripts required for the
UniPro Test Mode (see
details).
Appendix A, “Unipro Script Generation for more
The step to generate UniPro Scripts is necessary only for this test mode
and requires a license for option 167.
46Keysight N5990A MIPI M-PHY Receiver Test User Guide
Wildcards
N5990A M-PHY Station2
Figure 17Select M-PHY Sequences panel
A wildcard is an expression or symbol that is used especially to represent
any value when selecting specific file names or directories. This section
refers to the wildcard’s availability in the N5990A M-PHY for the sequence
file names selected in the Select M-PHY Sequences panel (see Figure 17).
The default sequence names do not use wildcards and are used for all
tests and setup procedures with out applying a change. If a wildcard
sequence name is provided, the sequence name is replaced with different
names during run time depending on the test specifics. It is your
responsibility to provide such sequence files in the M-PHY pattern
directory (by default C:\ProgramData\BitifEye\N5990A\Pattern\MPhy).
The supported wildcards are as listed below in Table 2 (none are
mandatory).
Table 2 Supported Wildcards
WildcardVal ues
[Type]RX, TX
[LsGear]PWM1, PWM2, PWM3,..PWM7
[HsGear]Gear1A, Gear1B,...Gear3B
[Mode]Cont, Burst
[Amplitude]Large, Small
[Channel]Data0, Data1, Data2, Data3, and DataAll
[Terminated]RT, NT
Keysight N5990A MIPI M-PHY Receiver Test User Guide47
2N5990A M-PHY Station
NOTE
Example
If a HS-Sequence file is provided as “MphyCompliance[HsGear].seq” and Gear-1A and Gear-2A are selected as HS-Gears in the Configure DUT
panel:
1 The Gear-1A tests run with a sequence file named
MPhyComplianceGear1A.seq.
2 The Gear-2A tests run with a sequence file named
MPhyComplianceGear2A.seq.
Figure 18An Example of A Wildcard Sequence
For the GearX-A tests, the sequence file name appears as
MphyComplianceGearXA.seq. If the GearX-A appears with "-" (dash),
N5990A does not recognize the sequence file and displays an error.
Special Parameters
Click Special Parameters (See Figure 13), to display the Special Parameters window (see Figure 19).
48Keysight N5990A MIPI M-PHY Receiver Test User Guide
N5990A M-PHY Station2
Figure 19Special Parameters Dialog
BER Settings
Select BER Reader
For automated receiver testing, it is necessary to determine whether the
DUT receives the data properly. This can be achieved by reading pass / fail
information from the device. The Bit Error Ratio (BER) is measured and
read. N5990A supports four different BER Reader implementations:
1Bert Analyzer
2 Offline BER Reader
3 Unipro BER Reader
4 Custom BER Reader
Keysight N5990A MIPI M-PHY Receiver Test User Guide49
2N5990A M-PHY Station
Bert Analyzer
The DUT is configured in Loopback Mode, so it will loop back the received
test pattern. The Error Detector (ED) then compares the pattern returned
by the DUT with the generated pattern to detect bit errors and compute
the BER.
The pattern must match and be in phase. This is ensured by a common
reference clock. The same pattern is loaded to the generators and the ED.
When the received bits are not synchronized with the pattern of the ED,
the computed BER will be very high. In this case, the synchronization
algorithm in the ED is restarted.
It is important to note that the ED of the BERT compares the whole
pattern, including the Sleep, Stall and Prepare states. This makes
Loopback Mode testing a not very ideal solution for bursted pattern test.
For example, if the Prepare Length that the DUT sends back is different to
that coming from the BERT generator, the computer BER will be high and
the test will fail, even if the returned test pattern matches the original
pattern.
Offline BER Reader
For each step of the test procedure, N5990A shows pop-up dialogs (see
Figure 20, to Figure 22) requesting you to reset and initialize the device
and decide whether the device is working properly. This method is
applicable to devices that allow a visual check, for example, a Display
Serial Interface (DSI) device connected to a display. It is also possible to
connect the DUT to the scope and verify if the output data is valid with
help of the serial decoder. Using an offline BER reader will result in a
semi-automated test where at each test point, you have to enter the pass /
fail information.
Figure 20MIPI Offline BER Reader Pop-up Dialog 1
50Keysight N5990A MIPI M-PHY Receiver Test User Guide
Figure 21MIPI Offline BER Reader Pop-up Dialog 2
N5990A M-PHY Station2
Figure 22M-PHY Test DUT (Offline BER Reader) Pop-up Dialog-3
UniPro BER Reader
The DUT is configured to Test Mode and Frame and Error counter requests
are interleaved with the test pattern.
N5990A then decodes the responses captured with the test equipment
and calculates the BER.
For more details about UniPro Test Mode refer to “Unipro Test Mode”.
Keysight N5990A MIPI M-PHY Receiver Test User Guide51
2N5990A M-PHY Station
Custom BER Reader
The usage of a Custom BER Reader enables fully automated testing for all
transmission modes (HS and LS) without the need for a BER ED or Unipro
BER reader support. This method requires the implementation of a class
supporting the IBerReader interface by you, providing access to the DUT's
pass/fail information. For example, by reading the DUT's internal error
counter registers. The definition of the IBerReader interface is available
from Keysight.
Address
It is the BER reader's address. It may be an IP address or a file directory.
Get Errors from BERT
This option is enabled if Custom BER Reader is chosen. It allows to read
the errors from the BERT Analyser like in Loopback, but controls the rest of
the test execution.
Settings
This option is visible when Unipro BER Reader is chosen. Click Settings to
open the Analyzer Settings dialog as shown in Figure 23.
Figure 23Analyzer Settings
In Unipro Test Mode, either the M8020A or the DSGA can be used as
Analyzer.
In case of DSGA, it is necessary to specify in the PWM Data Rate property,
a rough estimate of the data rate the DUT is responding with. The
Compare Mode can be selected as Differential or Single-Ended. The
Single-Ended mode allows you to connect the Tx- to the DSGA and Tx+ to
the scope and verify that the DUT is working, by using the scope decoder.
Calibrations Files Directory
Browse
52Keysight N5990A MIPI M-PHY Receiver Test User Guide
Click Browse to select a directory to save and load calibrations.
Default
Click Default to set the default calibrations files directory to: C\
ProgramData\BitifEye\N5990A\Calibrations\MPHY.
BERT Analyzer Settings
Use CDR for Analyzer
It is available when you select the J-BERT N4903B or M8020A
configuration in the Station Configuration (see Figure 7).
Loop Band width
It allows you to configure the loop bandwidth of the analyzer CDR. Note
that the value introduced here will be used for HS-G1 tests (see Figure ).
For other gears, the bandwidth is automatically calculated as a multiple
(x2 for G2, x3 for G3 and x4 for G4). To directly modify the CDR Loop BW
value for each gear, use the CDR Loop Bandwid th - Top Level property
that is accessible in the Property panel of the main windows when
selecting the desired Terminated Mode group in the test procedure tree
(see Figure 25).
N5990A M-PHY Station2
Figure 24CDR Loop Bandwidth for G1
Keysight N5990A MIPI M-PHY Receiver Test User Guide53
2N5990A M-PHY Station
Figure 25Modify CDR Loop Bandwidth for each gear
54Keysight N5990A MIPI M-PHY Receiver Test User Guide
Loop Order
This option is available only for M8020A system configuration. It can be
selected as First or Second order.
Fine Adjust
This option is available only for JBERT system configuration.
Transition Density
The transition density of the CDR can be set from 0% to 100%.
Peaking
The CDR peaking can be selected as Low, Med ium or High.
Input Termination
It can be selected as Balanced or Unbalanced.
Infiniium Settings
For calibrations, Infiniium can be used to add transfer functions to the
oscilloscope channels. Select the Use Infiniium check box to enable this
option.
N5990A M-PHY Station2
Other Settings
Reduced Amplitude at Init
If you select the check box for Reduced Amplitude at Init,the amplitude of
the generator is set to the value defined by you at the beginning of each
test procedure, while the generators are being initialized. This is used to
prevent glitches in the generated signal when the outputs of the BERT are
enabled.
Setup procedure in each test subgroup
If you select the check box for Setup procedure in each test subgroup,
the Setup Procedure Pattern is available for each test sub group in the
N5990A User Interface, allowing you to configure the DUT before tests are
started.
Use De-emphasis
If you select the check box for Use De-emphasisis, the generator will send
de-emphasized signals.
Keysight N5990A MIPI M-PHY Receiver Test User Guide55
2N5990A M-PHY Station
NOTE
Use D-Box
For J-BERT N4903B configuration, the de-emphasis can be generated
either with a De-emphasis Box or with the second channel of the J-BERT.
Select Use D-Box to use the external De-Emphasis Box.
Use external Ref Clock
It is available only when you select the J-BERT N4903B or M2080A
configuration systems in the Station Configuration (see Figure 15). The
reference clock generated by the J-BERT is not continuous when the
generators are stopped. This can happen when loading a new pattern to it
and the jitter sources are enabled. In this case, use an external reference
clock source by selecting the Use External Clock check box. The reference
clock source can either be an additional instrument or the reference clock
output from the DUT.
The available options are:
• PLL (only HS tests): The External PLL mode is used to lock the
generator to an external clock and the provided clock must not be
modulated. A clock multiplication with x/y is possible, with x/y = 1, 2, 3,
to 255 for N4903B and x/y = 2 to 162 for M8020A. It can not be used
for PWM tests because of the hardware multiplier limitation.
• Direct: In External clock mode, all output signals follow the external
clock and it's modulation. For J-BERT N4903B system configuration,
the modulation range is from 6.75 Gbps to 12.5 Gbps and below 6.75
GHz, SJ and SSC are not available. However, the external clock can
optionally be divided by 1, 2, 4, 8, or 16. For M8020A generator, the
modulation range is from 8.1 GHz to 16.207 GHz.
• External 10Mhz: The generator will be lock to an external 10MHz
reference clock.
For more details about the external clock refer to the “J-BERT 4903B” and
“J-BERT M8020A” manuals, respectively.
The N5990A allows an external reference clock source to be used but it is
not responsible for its setup.
Manual break at the sequence start
If selected, the sequence will start with a manual break.
Lane Test Mode
The Unipro PHY test mode allows to test multi-lane simultaneously.
56Keysight N5990A MIPI M-PHY Receiver Test User Guide
N5990A M-PHY Station2
If the Simultaneously Lane Mode is selected, the test tree doesn't show
separate test items for each lane. Instead, each test is shown a single time
and is referenced as Data All (see Figure 26.
Note that the time is halved, but two sets of TTCs and ISI traces are
required (see Figure 27).
Figure 26Test Tree for Individually Lane Mode (up) and for Simultaneously Lane
Mode (down)
Keysight N5990A MIPI M-PHY Receiver Test User Guide57
2N5990A M-PHY Station
M8020 ISI Properties
58Keysight N5990A MIPI M-PHY Receiver Test User Guide
Figure 27Connection Diagram for Simultaneously Lane Test Mode
This dialog allows to configure the ISI generation when M8020 is selected
as system configuration in the Station Configuration window (see
Figure 7).
N5990A M-PHY Station2
Figure 28M8020 ISI dialog
The preset can be selected as:
• Physical Trace: In this case the ISI is not generated internally with the
M8020. The ISI is added by including physical traces in the setup.
• MPHY G3A Ch1, MPHY G3A Ch2, MPHY G3B Ch1, MPHY G3B Ch2,
MIPI Short, MIPI Standard and MIPI Long are pre-defined presets of the
M8020.
• Custom: The preset is defined by selecting the following properties:
• Mode: Two pairs Frequency/Insertion Loss must be defined to
characterize the preset.
•If One Point mode is selected, the first point is pre-defined as
0Hz/0dB.
•In Two Point mode, the two Frequency/Insertion Loss points
can be defined.
• Slope: It depends on the points defined. It must be between -6,00dB
and -0,01dB.
• Insertion Loss Offset: It is the insertion loss at 0Hz.
• Frequency 1: First frequency point.
• Insertion Loss 1: Insertion loss applied to the first frequency point.
• Frequency 2: Second frequency point.
Keysight N5990A MIPI M-PHY Receiver Test User Guide59
2N5990A M-PHY Station
Connection Setup
• Insertion Loss 2: Insertion loss applied to the second frequency
point.
• Automatic: The Embedded ISI fixture is calibrated to a target value,
discounting the loss of the cables and of the Switch (if present). Using
the Automatic mode will result in a new calibration, named Embedded Fixture ISI Calibration.
These properties can be set for each Gear.
Click the Connection Setup button (see Figure 12) to open the
Connection Setup panel (see Figure 29).
• Active Probe (Active Termination Adapter)[Manual]
• Direct Connect
2 Connection Type: It can be selected as:
• Single-Ended
60Keysight N5990A MIPI M-PHY Receiver Test User Guide
N5990A M-PHY Station2
• Differential
3 TXDP and TXDN: It contains two blocks as given below
• Single-Ended
• Differential
Depending on the selection of Connection Type, the relative block
(Single-Ended or Differential) is enabled and it allows you to select
channels with the Primary Lane and Secondary Lane (see Figure 29).
Keysight N5990A MIPI M-PHY Receiver Test User Guide61
Keysight N5990A MIPI M-PHY Receiver Test User Guide
3Calibrations and Test
Procedures
Example of MIPI M-PHY Calibration / Test Procedure / 65
Connection Diagram / 67
Result Description / 71
MIPI M-PHY Parameters / 72
During the execution of all calibration and test procedures, the results are
displayed automatically in a data table as well as graphically. The viewer
can be either an MS-Excel or a HTML worksheet; this can be chosen in the Station Configuration (see Figure 6). Once a specific calibration or test
procedure is finished, the MS-Excel/HTML worksheet is closed. To
re-open it at any time, double click on the respective procedure in the
respective tree.
To save calibration data worksheets in a workbook, go to File > Save Results as Workbook... at any time. It is recommended you carry out this
step at least at the end of each N5990A run.
If the calibration and test procedures are conducted several times during
the same N5990A run, the result worksheets are combined in the
workbook.
If you conduct a test procedure without prior execution of calibration
procedures in the same test run, only the test results will be saved to the
workbook. As a safety feature, all calibration and test results are saved by
default to the N5990A “Tmp” directory (refer to the “Keysight N5990A Test
Automation Software Platform Installation Guide”). The sub-folder
Results/M-PHY Station contains the Excel files of the final results
measured at each calibration and test procedure. In addition to the
calibration data worksheets, the calibration data files are generated. These
3Calibrations and Test Procedures
files are saved by default to the N5990A calibrations folder. If these
calibrations are run again, the data file will be overwritten. In order to save
the calibration data files at each configuration, you must copy the files
from the directory: C:\ProgramData\BitifEye\N5990A\Calibrations\MPHY
and save them manually in any folder before re-running the calibrations.
64Keysight N5990A MIPI M-PHY Receiver Test User Guide
Example of MIPI M-PHY Calibration / Test Procedure
All calibration and test procedures are included in the respective groups in
a way similar to how they are organized in the CTS. For most procedures,
some specific parameters can be set in Expert Mode by the user. In
Figure 30, the Level Calibration Into Open is highlighted as an example.
The respective parameters are shown on the right side of the N5990A User
Interface. This is achieved by clicking on the calibration/test name. To start
one or more procedures, do the following:
1 Select the check box for the corresponding procedure. The Start
button is enabled and colored in green.
2Click Start to run the procedure.
3 Once all the procedures are run, click Save to store the N5990A
configuration as a single ".vfp" file. You can also click Load to recall
an N5990A configuration. In this way, you can run the DUT again
without configuring it.
Calibrations and Test Procedures3
Figure 30Example of MIPI M-PHY Calibration or Test Procedure
Keysight N5990A MIPI M-PHY Receiver Test User Guide65
3Calibrations and Test Procedures
CAUTION
Before executing the calibration or test procedures, ensure that the
Station Configuration is configured properly with all necessary
instruments such as the Infiniium oscilloscope set to Online. All
calibrations can be run in Offline mode, that is, without any instrument
connected. The Offline mode is intended for product demonstrations
with simulated data. CALIBRATIONS RUN IN OFFLINE MODE DO NOT
GENERATE VALID CALIBRATION DATA.
66Keysight N5990A MIPI M-PHY Receiver Test User Guide
Connection Diagram
Calibrations and Test Procedures3
To display the connection diagram, right-click the desired test or
calibration and select Show Connection (see Figure 31). Alternatively, the
connection diagram is displayed automatically when you start the selecetd
procedure.
Figure 31Show Connection Diagrams
Multi-Channel without Switch
When the setup does not include a switch for multi-lane purposes, it is
necessary to connect each DUT lane to a different data channel output of
the pattern generator. In this case, the J-BERT M8020A system allows to
connect 1,2,3 or 4 data channels and the J-BERT N4903B 1 or 2 data
channels. Figure 32 shows the data channels for the M8020A and
Figure 33 for the N4903B.
Keysight N5990A MIPI M-PHY Receiver Test User Guide67
3Calibrations and Test Procedures
Figure 32Data Channels for M8020A
Figure 33Data Channels for N4903B
Multichannel with Switch
When the setup includes a switch to connect several DUT lanes, only the
Output Data 0 of the data generator is used and must be connected to the
matrix input. In the connection diagram the switch is represented as a red
box between the generator data output and the Rx DUT input (see
Figure 34).
68Keysight N5990A MIPI M-PHY Receiver Test User Guide
Calibrations and Test Procedures3
Figure 34Connection Diagram with Switch
Click Show Switch Connections (see Figure 34) to display the switch
connections in detail (see Figure 35).
Keysight N5990A MIPI M-PHY Receiver Test User Guide69
3Calibrations and Test Procedures
NOTE
Figure 35Connection Diagram for Switch in detail
70Keysight N5990A MIPI M-PHY Receiver Test User Guide
If the "Use Switch" option is selected in the Station Configuration (see
Figure 7), but also the "Separate Low Speed Generator" option is
checked, the switch will be included in the setup but not for multi-lane
purposes (in this case is used to switch between generators). Therefore
the connections for more than one lane must to be done as explained in
the MultiChannel without Switch section.
If the "Use Switch" option is selected and the "Separated Low Speed
Generator" not, the switch will be used for multi-lane purposes and the
connections for more than one lane must to be done as explained in the
Multichannel with Switch section.
Result Description
SmileyDescription
Indicates that the procedure passed successfully at the previous run and the resul ts are available.
Indicates that the procedure was passed in offline mode previously and the results are available.
Indicates that the procedure passed successfully at the present run.
Indicates that the procedure was aborted/disturbed somehow and failed at the previous run.
Calibrations and Test Procedures3
Once the selected procedures are run, the smiley at the individual
procedure indicates the result (Pass / Fail / Incomplete) by displaying it's
face in specific ways as given below (see Table 3).
Table 3 Smiley's Result Description table
Indicates that the procedure was aborted/disturbed somehow and failed at the present run.
Indicates that the procedure failed at the previous run.
Indicates that the procedure failed at the present run.
Generally this kind of smiley displays two results such as the first half ind icates that the result of the present run and the
second half shows the result of the previous run. In this example, the first hal f indicates that the procedure passed
successfully at the present run and the second hal f means that it was not completely run at the previous run.
Keysight N5990A MIPI M-PHY Receiver Test User Guide71
3Calibrations and Test Procedures
MIPI M-PHY Parameters
Sequencer Parameters
The MIPI M-PHY parameters are of three types:
1 Sequencer parameters
2 Group parameters
3 Procedure parameters
The Sequencer parameters control the flow of the test sequencer, not the
behavior of individual procedures. They are identical across all versions of
N5990A. One of them, Repetitions, is available for all procedures and
groups in the procedure tree. The others are only available for procedures.
Like all other parameters the sequencer parameters are shown on the right
side of the N5990A user interface and you can change them (see
Figure 36).
Figure 36MPHY Sequencer Parameters
All sequencer parameters are listed in alphabetical order in Table 4.
Table 4 MPHY Sequencer Parameters
72Keysight N5990A MIPI M-PHY Receiver Test User Guide
Parameter NameParameter Description
Calibrations and Test Procedures3
Procedure Error Case BehaviorProceed With Next Procedure - If an error occurs in the current test or calibration procedure, continue by
running the next procedure in the sequence.
Abort Sequence - Abort the execution of the sequence.
Procedure Failed Case BehaviorProceed With Next Procedure - If the current test or calibration procedure fails, continue by running the
next procedure in the sequence.
Abort Sequence - Abort the execution of the sequence.
RepetitionsThe number of times the group or procedure is going to be repeated. If the value is '0', it runs only once.
Keysight N5990A MIPI M-PHY Receiver Test User Guide73
3Calibrations and Test Procedures
Group Parameters
Group parameters are used for several related calibration or test
procedures. They are shown on the right side of the N5990A user interface
when the selected entry of the procedure tree on the left is a group instead
of an individual procedure.
The MIPI M-PHY Receiver Test Software has some group parameters (in
addition to Repetitions on the top-level entry of the procedure tree (see
Figure 37). These will be common for all N5990A procedures.
Figure 37MPHY Group Parameters
Table 5 describes the group parameters.
Table 5 Group Parameters
Parameter NameParameter Description
0 V Amplitude at InitThe value is True/False. If is set to true, the amplitude of the generator will be set to 0 Volts during the
Reduced Amplitude at InitThe voltage value set during the Init phase of each test.
Re-Init SequenceThe value is True/False. The defaul t value is false. If the value is set to True, the test sequence starts
74Keysight N5990A MIPI M-PHY Receiver Test User Guide
Init phase of each test.
again from the beginning in each test step to ensure that the DUT is in a defined state. This results in an
increased test duration
Parameter NameParameter Description
Calibrations and Test Procedures3
Wait for Manual BreakThe value is True/False. If the value is set to True, a d ialog asking the user to manually restart the DUT will
be shown for every test step. This is useful for DUT's that comes out of test mode when loopback is being
used.
Switch ValuesThis property allows to modify how the switching is performed. The following selections are possible:
• testStepValue: the levels are not changed when switching
• nominalValue: before switching, the BERT is set to the default 200/100 mV levels, and after switching
the original amplitude is restored.
• zeroAmp: the amplitude in both BERT and DSGA is set to 0 before switching and then restored to the
original value.
Show DUT Configure DialogThe value is True or False. The default value is False. If the Init step fails (where all parameters are at
default values), a dialog pops up to check the configuration of the DUT.
Sample RateSpecify the sampling rate to use for all tests. If it is set to default, then the sample rate will be set to the
appropriate value depend ing on the tested data rate.
Trig Threshold ModeWhen this option is set to "Auto", the appl ication will automatically determine the threshold value. When
this option is set to "Manual", then the value of the "Trigger Level" option will be used as the trigger
threshold.
Trigger LevelSpecify the value of the trigger level used for triggering the test signal when running the M-PHY tests. The
value of this option will be used ONLY when the "Threshold Mode" option is set to "Manual". By default, it
is set to 0V.
Keysight N5990A MIPI M-PHY Receiver Test User Guide75
3Calibrations and Test Procedures
Procedure Parameters
Procedure Parameters are parameters that do not fall into one of the
previously described categories. They are shown on the right side of the
N5990A user interface when the selected entry of the procedure tree on
the left is an individual procedure. They only change the behavior of that
single procedure. Procedures often have parameters with the same name,
but set settings always apply on the selected procedure, and the meaning
may be slightly different. These parameters are listed in Table 6.
Table 6 Procedure Parameters
Parameter NameParameter Description
Accumulated Voltage Eye OpeningIt is the accumulated vol tage set in the data generator to get the desired Eye Opening.
Additional Jitter FrequenciesThese are additional jitter frequency points beside the frequency range, where the actual jitter amplitude is
Additional StepsThese are additional steps to be performed for the test.
AmplitudeIt is the differential voltage amplitude of the tested/calibrated data signal.
Amplitude RangeIt is the range of the amplitude values to be tested. It is defined with four values separated by semicolons:
Analysis MethodIt exists for Low Frequency RJ Calibration and it can be selected as “Spectral” or “Tailfit”. Refer to the Scope
BER TargetThe target BER used in the calibration/procedure.
Check PatternIt can be “True”/”False”. If “True”, the selected BER reader is used to verify that the DUT is in loopback mode
CMI AmplitudesIt is the range of the common-mode interface values to be tested over the jitter frequency range. They must be
CMI FrequencyIt is the common mode interference frequency value.
Common Mode Voltage LevelsThese are the common mode voltage levels to be tested.
Custom Data RateIt displays the custom data rates added by the user, using the “Custom HS DR” button (see
Data RateIt is the data rate for the selected GEAR.
Data Rate Deviation over Nominal
Value
Deterministic JitterIt is the deterministic jitter amplitude value.
measured.
<Start Value>;<End Value>;<Number of Steps>;<Scale Type>. The scale type can be Linear or Logarithmic.
manual for more details as it is related to the EZJit Plus software.
and transmitting the test pattern properly.
separated by semicolons.
Figure 20).
It is the deviation value of the data rate across the nominal value.
Differential Voltage Start ValueIt is the first and maximum differential voltage to be calibrated.
Eye Width TargetIt is the eye width to be calibrated.
76Keysight N5990A MIPI M-PHY Receiver Test User Guide
Parameter NameParameter Description
Eye Width Max. VariationIt is the allowed variation of the target eye width.
Eye Height TargetIt is the eye height to be calibrated.
Filter Delay Chan1/2The filter delay can be enabled on the oscilloscope.
Frequency DeviationIt is the deviation that has occurred in the frequency value.
Calibrations and Test Procedures3
Frequency RangeIt is the range of the jitter frequency values to be tested. It is defined with four val ues separated by semicolons:
<Start Value>;<End Value>;<Number of Steps>;<Scale Type>. The scale type can be Linear or Logarithmic.
HS Sequence FileIt is the sequence file to be used during the calibration/test.
HS Sync lengthIt is the number of sync pattern symbols before the data burst. It must be specified in SI (1SI = 10UI).
Initial Jitter Amplitude StepIt is the initial step size value of jitter amplitude.
IBerReader Init ModeIt refers to the configuration parameters of the BER reader for the system initialization mode with the
parameters such as data channel, data rate and termination mode (Terminated / Unterminated).
ISI (p-p)It is the amount of ISI (Inter Symbol Interface) added during the test procedure.
ISI trace / J20It is the amount of ISI (Inter Symbol Interface) added d uring the test procedure. It is only available for J-BERT
configurations with J20 option.
Jitter Calibration FileIt is the sequence file for the jitter calibration.
Jitter FrequenciesThese are the frequency points to be tested/calibrated.
Jitter Frequency RangeIt is the range of the jitter frequency values to be tested. It is defined with four values separated by semicolons:
<Start Value>;<End Value>;<Number of Steps>;<Scale Type>. The scale type can be Linear or Logarithmic.
Jitter Amplitude Min StepIt is the minimum step size value that is used to find the DUT limit for each jitter frequency.
Jitter Increase AccuracyThis parameter is available in the Jitter Tolerance test (2.1.7) when the 'Perform Jitter Limit Test" property is
set to true. It defines the windows size of the binary search that is used to find the maximum tolerated jitter.
Lane Under Test Termination ModelFor M8020A, system configuration is necessary to indicate the termination model such as 50Ohm or 100Ohm.
If the selected mode is not right, the M8020A outputs will be turned off. When the 100 Ohm Termination board
is not available, you can set this parameter to “50 Ohm” and connect SMA cables directly to the scope.
Level PairsThey are pairs of single ended amplitudes (differential and common mode voltage) separated by “|”.
LFSJ FrequenciesThese are the frequencies to be tested during Jitter Tolerance Test.
Max. Calibration Voltage AmplitudeIt is the start voltage value to find the target eye height.
Max Interference AmplitudeIt is the maximum value of the interference amplitude to be tested.
Max Interference ValueIt is the higher limit for the tested RF (Rad io Frequency) Interference. If the minimum and maximum
Max Jitter ValueIt is the maximum jitter value to be calibrated.
Keysight N5990A MIPI M-PHY Receiver Test User Guide77
interference values are equal, only one interference level is tested.
3Calibrations and Test Procedures
Parameter NameParameter Description
Max Tested HFSJ AmplitudeIt is the maximum value of the HFSJ to be tested.
Max Tested LFSJ AmplitudeIt is the maximum value of the LFSJ to be tested.
Max Tested ValueIt is the maximum skew value to be tested.
Max. Time Span Ch1/2It is the time span set to the oscilloscope channels.
Minimum Accumulated Differential
Voltage
Frequency OffsetIt is the frequency deviation relative to the nominal bit rate.
Min Interference AmplitudeIt is the minimum value of the interference amplitude to be tested.
Min Interference ValueIt is the lower limit for the tested RF Interference. If the minimum and maximum interference values are equal,
Min Jitter ValueIt is the minimum jitter value to be calibrated.
Min SpecIt is the minimum value of the parameter that the DUT must tolerate according to the specification.
Min Tested ValueIt is the minimum skew value to be tested.
Min User-Defined Interference
Amplitude
Min User-Defined SJ AmplitudeIt is a user-defined value for the Jitter Sensitivity test. It determines the minimum SJ ampl itude value that
Number of Histogram HitsIt is the number of histogram hits using the oscilloscope.
Number of PulsesThe default value is 1, that is, only one DIFN (negative differential voltage) pulse is generated.
Offset RangeIt is the range of the offset values to be tested. It is defined with four values separated by semicolons: <Start
Oscilloscope Band widthIt allows modification of the oscilloscope band width in cal ibrations. As it impacts the measured jitter, the
Pattern Padding ModeIt can be selected as “Multiply Pattern” and “Fill Pattern”.
If the measured accumulated differential voltage is smaller than the given differential voltage, the calibration
stops.
only one interference level is tested.
It is a user-defined value for the Common Mode Interference test. It determines the minimum Interference
value that must meet the target BER to pass the test.
must meet the target BER to pass the test.
Value>;<End Value>;<Number of Steps>;<Scale Type>. The scale type can be Linear or Logarithmic.
default value is recommended.
Perform Jitter Limit TestThis property is set to to true, in the Jitter Tolerance Test (2.1.7), to search for the maximum amount of jitter
PJ AmplitudeIt is the value of the PJ (Period Jitter) amplitude.
PJ FrequencyIt is the frequency value of PJ.
(HS/LS) Prepare LengthIt defines the PREPARE sub-state. It specifies the number of DIFP (Differential Positive Pulse) states before the
78Keysight N5990A MIPI M-PHY Receiver Test User Guide
that the DUT can tolerate. The SJ component is increased until the BER test fails.
If is set to false (by defaul t), the total jitter is set to the worst case scenario defined in the CTS.
HS/LS-burst starts. It must be specified in SI (1SI = 10UI).
Parameter NameParameter Description
Calibrations and Test Procedures3
Prepare Length Capabil ityIt is a DUT attribute that determines the minimum PREPARE time supported. It has a permitted range from 0 to
Pulse DistanceIt is the distance between DIFN pulses. If only one pulse is appl ied, it gives the distance between the following
Pulse WidthIt is the width of the DIFN pulse. It's default value is 20 ns.
PWM Data RateIt is the data rate for the PWM mode.
PWM Receive RatioIt is the PWM Receive Ration provided for the test.
Ref Clock AdjustmentIt can be adjusted as “NominalRefClock” and “NominalDataRate”.
Ref Clock FrequencyIt is the frequency value of the reference clock.
Reference Clock Skew Step SizeIt is the step size value of the reference clock skew d uring the Jitter Tolerance test.
Reference Clock Skew Step
Duration
Re-Init Sequence after Reset DUTIf the value is set to True, the test sequence starts again from the beginning every time the DUT is reset.
RJ (p-p) AmplitudeIt is the amplitude value of the RJ added to the test signal during calibration.
Scope RJrmsDuring RJ calibrations, the oscilloscope adds an add itional amount of jitter. This jitter can be calculated by
Semi Automated TestThis property is used for HS, PWM and SYS Termination tests. If false, the software automatically measures the
15.
DIFP (Positive Differential Voltage) state next to the DIFN state.
It is the step duration value of the reference clock skew d uring the Jitter Tolerance test.
the scope using EZJit Wizard. If that calculated value is set to the “Scope RJrms” parameter, it will be
subtracted from the RJ measurement.
termination time. If true, the automation find s the spot where the burst starts but lets you place the markers
manually. Then, with the position of the markers, the termination time is calculated. This is useful in cases
where the software is unable to calculate this time automatically because of, for example, reflections in the
signal that impair the algorithm.
SequenceIt is the sequence file used d uring the test to the DUT.
Set OffsetIt is the signal offset set for the data channel.
Set Single Ended AmplitudeIt is the single ended amplitude value set for the data channel.
Short Term Deterministic JitterIt is the short term deterministic jitter amplitude value.
Short Term Random Jitter / STRJ
Amplitude (p-p)
Short Term Total JitterIt is the short term total jitter amplitude value.
Show Real Time EyeIt can be selected as “True”/”False” to switch between real time eye and single waveform. The measurement
Sleep LengthIt is the provided sleep length value (it is the power saving state between LS Bursts).
Keysight N5990A MIPI M-PHY Receiver Test User Guide79
It is the amount of short term random jitter amplitude added to the signal.
time increases by viewing the real time eye.
3Calibrations and Test Procedures
Parameter NameParameter Description
Sleep Time CapabilityIt is the sleep length required by the DUT.
Skew ToleranceIt is a limit value. If the skew between the lanes is below this limit value, the calibration process is ended.
Squelch Test SequenceIt is the pattern generator sequence for squelch tests.
Stall LengthIt defines the STALL state. It specifies the number of DIFN states after the data burst. It must be specified in UI.
Stall Time CapabilityIt is the STALL time required by the DUT. It has a permitted range of 1 to 255 SI (1SI = 10UI).
Steps Interference AmplitudeIt is the number of steps performed between the Minimum and Maximum interference ampl itude points.
Step SizeThis is the value to be increased/decreased at each step of the calibration/test procedure.
(HS / LS) Sync lengthIt defines the length of the SYNC sequence. It has a permitted range of 0 to 15 SI (1SI = 10UI).
(HS / LS) Sync rangeIt is the range for the SYNC sequence, It can be set as Fine or Coarse.
T Activate capabilityIt is the Activate time required by the DUT to exit the hibernate state and enter into the sleep state.
Target BERThe default target BER value is 1E-10 and if the measured BER is smaller than the target value, the test is
Test SequenceIt is the sequence file used d uring the test to the DUT.
Tested Accumulated Differential
Voltage
Tested Accumulated Differential
Range
Tested Frequency OffsetIt isthe frequency deviation relative to the nominal bit rate.
Tested OffsetIt is the list of common-mode voltage values tested.
Tested RangeIt is the range used to test the frequency offset value. It is defined with four values separated by semicolons:
Total JitterIt is the total jitter amplitude value.
TransitionsIt is the number of transitions used in each trace calibration.
Use Infiniisim
Transfer Function Chan1
Transfer Function Chan2
considered as passed.
It is the accumulated d ifferential voltage value to be tested. The differential voltage amplitude is set
according to the calibration in order to match the desired accumulated differential voltage.
It is the accumulated differential voltage range to be tested. It is defined with four values separated by
semicolons: <Start Value>;<End Value>;<Number of Steps>;<Scale Type>. The scale type can be Linear or
Logarithmic.
<Start Value>;<End Value>;<Number of Steps>;<Scale Type>. The scale type can be Linear or Logarithmic.
The parameter, “Use Infiniisim” can be set to “True” or “False”. When a replica channel is not available or you
want to use the Infiniisim Transfer Function instead of a replica channel, set the “Use InfiniiSim” parameter to
“True” and copy the Transfer Function file to the scope, either under “C:\Users\Public\Publ ic Documents\
Infiniium\Filters\M-PHY” or “C\Documents and Settings\All Users\Documents\Infiniium\Filters\M-PHY”.
80Keysight N5990A MIPI M-PHY Receiver Test User Guide
Keysight N5990A MIPI M-PHY Receiver Test User Guide
4Calibrations
Calibration Overview / 82
Reference Clock Calibration / 83
Level Calibration Terminated / Into Open / 90
ISI (w/Switch) Calibration / 99
Embedded Fixture ISI (w/Switch) Calibration / 104
Short Term RJ Calibration / 109
Low Frequency RJ Calibration / 113
RJ Calibration / 117
High Frequency SJ Calibration / 121
SJ Calibration / 125
Eye Opening Calibration with Jitter/ Eye Opening SigTest Cal / 129
Accumulated Voltage Calibration Terminated / 134
ISI Calibration (J20) / 138
Inter Pair Skew Calibration / 143
4Calibrations
Calibration Overview
Before any test procedure can be run, the MIPI M-PHY test system must
be calibrated. The calibration plane is given by the DUT input ports. The
receiver test signal characteristics such as the M-PHY signal generator
output voltage level, timing and jitter parameters are typically affected by
the signal transmission between the M-PHY signal generator output ports
and the DUT input ports. Transmission elements such as cables, power
dividers, splitters, and TTC (transition time converters) impact the signal
generator's output signal, for example, by reducing signal levels as a result
of losses and slowing down transition times. Thus, for any signal output
parameter selected by you (set value), the jitter and the signal received at
the DUT input ports (actual value) deviates from the set value. The
additional deviations can be caused by effects such as offset errors,
hysteresis, and non-linear behavior of the M-PHY signal generator.
calibration procedures compensate the deviations of the relevant signal
output parameter actual values from the set values over the required
parameter range.
All calibration procedures required for the MIPI M-PHY receiver testing are
included in the software. calibration procedures are implemented such
that the process is conducted as fast as possible and is also automated as
much as possible, for example, by minimizing the number of
reconfigurations of the hardware connections. Most of the calibration
procedures are run individually for each clock and data lane, for instance,
the LP and HS level calibrations and the inter-pair skew calibration.
Calibrations for all hardware configurations are explained in the following
sections.
82Keysight N5990A MIPI M-PHY Receiver Test User Guide
Reference Clock Calibration
Calibrations4
The Reference Clock Calibration is available for all hardware
configurations (see Figure 38).
Figure 38Reference Clock Calibration
Purpose
This procedure calibrates the output levels (Amplitude and Offset) of the
reference clock.
Keysight N5990A MIPI M-PHY Receiver Test User Guide83
4Calibrations
Procedure
The output used to generate the reference clock is connected to the
oscilloscope. The data generator does a sweep of the offset according to
the range defined by the “Offset Range” parameter. For each offset value,
a sweep of the voltage amplitude is also performed according to the range
defined by the “Amplitude Range” parameter. Then, for each
offset-amplitude pair set in the data generator, the actual values of offset
and differential voltage amplitude are measured with the oscilloscope and
stored.
The results are saved in two separate calibration data files, one for the
offset and one for the amplitude values.
Connection Diagram
Figure 39 and Figure 40 show the connection diagrams for different
system configurations.
Figure 39Connection Diagram for Reference Clock Calibration (for M8020A)
84Keysight N5990A MIPI M-PHY Receiver Test User Guide
Parameters
Calibrations4
Figure 40Connection Diagram for Reference Clock Calibration (for N4903B)
The parameters used in Expert Mode for this calibration, are as follows:
•Oscilloscope bandwidth
• Amplitude Range
• Offset Range
• Use Infiniisim
• Transfer Function Chan1
• Filter Delay Chan1
• Max. Time Span Chan1
These parameters are listed in Table 6.
Dependencies
No procedure is required for this calibration.
Results
An example of a HTML Viewer for the Reference Clock Calibration
procedure containing two results is as shown below (see Figure 41 and
Figure 42). The results comprise of the following:
• A calibration data graph.
• The common parameter list (in Expert Mode).
• A data table for the offset and amplitude that are being calibrated.
(Refer to Table 7 for the parameter description).
Keysight N5990A MIPI M-PHY Receiver Test User Guide85
4Calibrations
86Keysight N5990A MIPI M-PHY Receiver Test User Guide
Calibrations4
Figure 41Example HTML Viewer for Offset in Reference Clock Calibration
Keysight N5990A MIPI M-PHY Receiver Test User Guide87
4Calibrations
88Keysight N5990A MIPI M-PHY Receiver Test User Guide
Calibrations4
Figure 42Example HTML Viewer for Amplitude in Reference Clock Calibration
Table 7 Reference Clock Calibration data table
Parameter NameParameter Description
Result“Pass”/”Fail”. If the set-up is wrong, the calibration procedure fails due to the non-monotonic values
measured.
Set OffsetIt is the set offset value for a given step.
Measured OffsetIt is the effective offset value as measured using a DSO with the histogram measurement.
Set AmplitudeIt is the set amplitude value for a given step.
Measured AmplitudeIt is the effective amplitude value as measured using a DSO with the histogram measurement.
Keysight N5990A MIPI M-PHY Receiver Test User Guide89
4Calibrations
Level Calibration Terminated / Into Open
This calibration is available for all hardware configurations.
Figure 43Level Calibration Terminated
Purpose
This procedure is used to calibrate the amplitude and offset of the signal
generators when those are terminated or into open.
This calibration should run once at each selected lane (see Figure 43).
For the M8020A/N4903B + DSGA setup, the calibration for Data0 includes
the switch. If more than one channel is selected, the other data outputs
are calibrated without switch.
90Keysight N5990A MIPI M-PHY Receiver Test User Guide
Procedure
NOTE
Connection Diagram
Calibrations4
The normal and the complement output signals of the reference board are
connected to the DSO channels with direct SMA connection in the
Terminated case and through a fixture without termination in the Into
Open case.
The data generator sends a slow clock pattern. It does a sweep of the
offset according to the range defined by the “Offset Range” parameter. For
each offset value, a sweep of the voltage amplitude is also performed
according to the range defined by the “Amplitude Range” parameter.
Then, for each offset-amplitude pair set in the data generator, the actual
values of offset and differential voltage amplitude are measured with
oscilloscope and stored.
The results are saved in two separate calibration data files, one for the
offset and one for the amplitude values.
Figure 44 to Figure 47 show the connection diagrams for the different
system configurations in Terminated mode and Figure 48 to Figure 51 in
Non-terminated mode.
The connection diagrams shown are for Gear1. They are the same for
Gear2, Gear3 and Gear4 but vary in the TTC value (G1: 250ps; G2: 125ps;
G3 and G4: 60ps).
Figure 44Connection Diagram for Level Calibration (for M8020A)
Keysight N5990A MIPI M-PHY Receiver Test User Guide91
Figure 46Connection Diagram for Level Calibration (for N4903B)
92Keysight N5990A MIPI M-PHY Receiver Test User Guide
Calibrations4
Figure 47Connection Diagram for Level Calibration, Into Open Data 0 (for N4903B
+DSGA)
Figure 48Connection Diagram for Level Calibration, Into Open Data 0 (for M8020A)
Keysight N5990A MIPI M-PHY Receiver Test User Guide93
4Calibrations
Figure 49Connection Diagram for Level Calibration, Into Open Data 0 (for M8020A +
DSGA)
Figure 50Connection Diagram for Level Calibration, Into Open Data 0 (for N4903B)
94Keysight N5990A MIPI M-PHY Receiver Test User Guide
Parameters
Calibrations4
Figure 51Connection Diagram for Level Calibration, Into Open Data 0 (for N4903B +
DSGA)
The parameters used in Expert Mode for this calibration, are as follows:
•Oscilloscope Bandwidth
• Amplitude Range
• Offset Range
• Use Infiniisim
• Transfer Function Chan
• Filter Delay Chan1
• Max. Time Span Chan1
These parameters are listed in Table 6.
Dependencies
No calibration is required for this procedure.
Keysight N5990A MIPI M-PHY Receiver Test User Guide95
4Calibrations
Results
An example of a HTML Viewer for the Levels Calibration Terminated / Into
Open procedure containing two results is as shown (see Figure 52 and
Figure 53. The results comprise of the following:
• A calibration data graph.
• The common parameter list (in Expert Mode).
• A data table for the offset and amplitude that are being calibrated
(Refer to Table 8).
96Keysight N5990A MIPI M-PHY Receiver Test User Guide
Calibrations4
Figure 52Example HTML Viewer for Levels Calibration Offset Terminated or Into
Open
Figure 53Example HTML Viewer for Levels Calibration Amplitude Terminated or Into
Open
Table 8 Levels Calibration Amplitude Terminated / Into Open data table
Keysight N5990A MIPI M-PHY Receiver Test User Guide97
4Calibrations
Parameter NameParameter Description
Result"Pass"/"Fail"- If the set-up is wrong, the calibration procedure fails due to the non-monotonic values
Set OffsetIt is the set offset value for a given step.
Measured OffsetIt is the effective offset value as measured using a DSO with the histogram measurement.
Set AmplitudeIt is the set amplitude value for a given step.
Measured AmplitudeIt is the effective amplitude value as measured using a DSO with the histogram measurement.
measured.
98Keysight N5990A MIPI M-PHY Receiver Test User Guide
ISI (w/Switch) Calibration
Calibrations4
This procedure is available for all hardware configurations.
Figure 54ISI Calibration
Purpose
The receiver test setup must include an ISI Compliance Channel, capable
of introducing a channel loss as defined in the M-PHY specification. This
procedure calibrates the ISI of the trace used for that purpose.
This calibration should run once at each data rate (see Figure 54).
Keysight N5990A MIPI M-PHY Receiver Test User Guide99
4Calibrations
NOTE
When the DSGA is used, two calibrations are required at each data rate:
one for Data0 and another one for the other channels. The calibration for
Data0 includes the switch. For the other channels, Data1 is calibrated
without switch.
Procedure
For Gear 1 & Gear2 calibrations, the Keysight SATA II ISI channel
(N4915-60001) can be used. The 3 Gb/s trace length (535 mm is
recommended) is compatible with the M-PHY specification requirements.
For Gear 3 and Gear 4 calibrations, there is no a specific ISI trace defined.
The channel must meet the requirement of the Reference CH1 channel
defined in the M-PHY specification (last version 4,10). Instead of using a
physical ISI channel, embedded ISI of M8020A can be used, if available.
The data generator sends a continuous CJTPAT pattern at the selected
HS-Gear data rate. The ISI is then measured and stored.
Connection Diagram
Figure 55 to Figure 58 show the connection diagrams for the different
system configurations.
The connection diagrams shown are for Gear1. They are the same for
Gear2 and Gear 3, but vary in the TTC value (G1: 250ps; G2: 125ps; G3
and G4: 60ps).
Figure 55Connection Diagram for ISI Calibration (for M8020A)
100Keysight N5990A MIPI M-PHY Receiver Test User Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.