Keysight Technologies N5511A User Manual

Page 1
Keysight N5511A Phase Noise Test System
User’s Guide
Page 2
Notices
© Keysight Technologies, Inc. 2019
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Trademark Acknowledgments
Manual Part Number
N5511-90002
Edition
Edition 1, September 2019
Printed in USA/Malaysia
Published by: Keysight Technologies
1400 Fountaingrove Parkway Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional
government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Page 3

Where to Find the Latest Information

Documentation is updated periodically. For the latest information about these products, including instrument software upgrades, application information, and product information, browse to one of the following URLs, according to the name of your product:
http://www.keysight.com/find/n5511a
To receive the latest updates by email, subscribe to Keysight Email Updates at the following URL:
http://www.keysight.com/find/MyKeysight
Information on preventing instrument damage can be found at:
www.keysight.com/find/PreventingInstrumentRepair
Is your product software up-to-date?
Periodically, Keysight releases software updates to fix known defects and incorporate product enhancements. To search for software updates for your product, go to the Keysight Technical Support website at:
http://www.keysight.com/find/techsupport
3
Page 4
4
Page 5

Table of Contents

1. Getting Started
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Documentation Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Additional Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
System Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2. Introduction
Introducing the GUI. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Designing to Meet Your Needs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Beginning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
N5511A Operation: A Guided Tour . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
How to begin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Powering the System On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
To power on a racked system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Starting the Measurement Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Verify License Key is Installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Powering the System Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Contents
3. Phase Noise Basics
What is Phase Noise? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Ideal vs Real Word Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Phase terms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4. Expanding Your Measurement Experience
GUI Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Viewing Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Display Preferences. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Omitting Spurs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Displaying the Parameter Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Exporting Measurement Results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Saving Measurement State. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Recall Measurement State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Measurement Preferences. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Toolbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
5. Absolute Measurement Fundamentals
The Phase-Lock-Loop Technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Single Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Dual Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
The Phase-Lock Loop Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
The Capture and Drift tracking ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Peak tune range (PTR). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
N5511A Phase Noise Test System User’s Guide 5
Page 6
Contents
Selecting the VCO Source. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
What Sets the Measurement Noise Floor? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
The System Noise Floor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
The Noise Level of the Reference Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Selecting a Reference (Single Channel). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Using a Similar Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Using a Signal Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Selecting a Reference (Dual Channel) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Tuning Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Estimating the Tuning Constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Tracking Frequency Drift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Evaluating beat note drift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Changing the PTR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
The Tuning Qualifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Minimizing Injection Locking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Adding Isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Increasing the PLL Bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Inserting a Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
An attenuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
An amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Evaluating Noise Above the Small Angle Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Determining the Phase-Lock-Loop bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Measured Beat Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Single-sided Spur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Double-sided Spur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
6. Absolute Measurement Examples
Example Overviews . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Input Ports. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Single Channel Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Configuring Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Measurement Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Dual Channel (EFC). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Configuring Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Measurement Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Dual Channel (DCFM). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Configuring Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
Measurement Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
6 N5511A Phase Noise Test System User’s Guide
Page 7
OCXO Dual Channel Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Configuring Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Measurement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
7. Residual and Additive Noise Measurement Fundamentals
What is Residual Noise? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
The noise mechanisms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Fundamental Measurement Technique. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
Assumptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
Frequency translation devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
Calibrating the Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
User entry of phase detector constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Measured Beat Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
Measured
Double-Sided Spur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182
Single-Sided Spur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
Measurement Difficulties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186
System connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186
± DC Peak Voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180
Contents
8. Residual Measurement Examples
Amplifier Measurement Example - Single Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Configuring Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
Measurement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
Amplifier Measurement Example - Dual Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
Configuring Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
Measurement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
9. FM Discriminator Fundamentals
The Frequency Discriminator Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218
Basic theory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218
The discriminator transfer response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 219
10. FM Discriminator Measurement Examples
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224
FM Discriminator Single Channel Measurement using Double-Sided Spur Calibration . . . . . . . . . . . . . . . . . . . . 225
Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225
Determining the discriminator (delay line) length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 226
Define the measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 227
Setup considerations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233
Making the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236
N5511A Phase Noise Test System User’s Guide 7
Page 8
Contents
When the measurement is complete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237
FM Discriminator Single Channel Measurement using FM Rate and Deviation Calibration . . . . . . . . . . . . . . . . . 240
Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240
Determining the discriminator (delay line) length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240
Define the measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 242
Setup considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 247
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248
Making the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 251
When the measurement is complete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 252
11. AM Noise Measurement Fundamentals
AM-Noise Measurement Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 256
Basic noise measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 256
Phase noise measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 256
Amplitude Noise Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 257
AM noise measurement block diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 257
Using an External AM Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 258
Calibration and Measurement General Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 260
User entry of phase detector constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 261
Double-Sided Spur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 263
Single-Sided Spur. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 267
12. AM Noise Measurement Examples
AM Noise Measurement Examples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 272
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 272
Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 272
Configuring Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 273
Measurement Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 274
13. Baseband Noise Measurement Example
Baseband Noise Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284
Baseband Noise with Test Set Measurement Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285
Define Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285
New Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285
14. Evaluating Your Measurement Results
Evaluating the Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 290
Looking for obvious problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 290
Comparing against expected data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 291
Gathering More Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 293
Repeating the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 293
Doing more research . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 293
Outputting the Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 294
Using a printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 294
Graph of Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 295
Marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 295
8 N5511A Phase Noise Test System User’s Guide
Page 9
Omit Spurs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 297
Parameter summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300
Problem Solving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 302
Discontinuity in the graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 302
Higher noise level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 303
Spurs on the graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 304
15. Advanced Software Features
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 308
Phase-Lock-Loop Suppression . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 309
PLL suppression parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 310
Ignore-Out-Of-Lock Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 312
PLL Suppression Verification Process. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 313
Accuracy degradation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 321
Blanking Frequency and Amplitude Information on the Phase Noise Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 322
Security level procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 322
16. Reference Graphs and Tables
Approximate System Noise Floor vs. R Port Signal Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 328
Phase Noise Floor and Region of Validity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 329
Phase Noise Level of Various Keysight Sources. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 330
Increase in Measured Noise as Ref Source Approaches DUT Noise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 331
Approximate Sensitivity of Delay Line Discriminator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 332
AM Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 333
Voltage Controlled Source Tuning Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 334
Tune Range of VCO for Center Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 335
Peak Tuning Range Required by Noise Level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 336
Phase Lock Loop Bandwidth vs. Peak Tuning Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 337
Noise Floor Limits Due to Peak Tuning Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 338
Tuning Characteristics of Various VCO Source Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 339
Contents
17. System Specifications
System Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 342
Power requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 344
18. System Interconnections
N5511A System Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 346
Making Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 347
Before connecting the cables to any device: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 347
Proper Connector Torque . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 347
Connecting a Display to your System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 348
N5511A Two Channel Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 349
System Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 354
19. Preventive Maintenance
N5511A Phase Noise Test System User’s Guide 9
Page 10
Contents
Using, Inspecting, and Cleaning RF Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 358
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 358
RF Cable and Connector Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 358
Before connecting the cables to any device: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 359
Proper Connector Torque . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 359
Connector Wear and Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 359
2.92 Connector Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 359
Cleaning Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 360
General Procedures and Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 361
Connector Removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 362
Instrument Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 364
Standard instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 364
Half-Rack-Width Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 365
Benchtop Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 366
Instrument Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 367
Standard rack instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 367
Half-Rack-Width instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 368
Benchtop instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 368
10 N5511A Phase Noise Test System User’s Guide
Page 11
Contents
N5511A Phase Noise Test System User’s Guide 11
Page 12
Contents
12 N5511A Phase Noise Test System User’s Guide
Page 13
Keysight N5511A Phase Noise Test System
User’s Guide

1 Getting Started

“Introduction” on page 14
“Documentation Map” on page 15
“Additional Documentation” on page 16
“System Overview” on page 17
13
Page 14

Introduction

Getting Started Introduction
This guide introduces you to the Keysight N5511A Phase Noise Test System software and hardware. It provides procedures for configuring the N5510 Phase Noise Measurement software, executing measurements, evaluating results, and using the advanced software features. It also covers phase noise basics and measurement fundamentals to get you started.
Use Table 1-1 on page 15 as a guide to:
— Learning about the N5511A phase noise test system
— Learning about phase noise basics and measurement fundamentals
— Using the N5511A system to make specific phase noise measurements.
In this guide you’ll also find information on system connections and specifications, and procedures for re-installing phase-noise-specific hardware and software in the system PC.
Installation information for your system is provided in the Keysight N5511A Phase Noise Test System Getting Started Guide.
14 N5511A Phase Noise Test System User’s Guide
Page 15
Getting Started Documentation Map

Documentation Map

Table 1-1 N5511A user’s guide map
Learning about the N5511A System
Chapter 1, “Getting Started”
Chapter 2, “Introduction” Chapter 3, “Phase Noise Basics”
Chapter 4, “Expanding Your Measurement Experience”
Learning Phase Noise Basics & Measurement Fundamentals
Chapter 5, “Absolute Measurement Fundamentals”
Chapter 7, “Residual and Additive Noise Measurement Fundamentals”
Chapter 9, “FM Discriminator Fundamentals”
Chapter 11, “AM Noise Measurement Fundamentals”
Using the N5511A for Specific Phase Noise Measurements
Chapter 6, “Absolute Measurement Examples”
Chapter 8, “Residual Measurement Examples”
Chapter 10, “FM Discriminator Measurement Examples”
Chapter 12, “AM Noise Measurement Examples”
Chapter 13, “Baseband Noise Measurement Example”
Chapter 14, “Evaluating Your Measurement Results”
Chapter 15, “Advanced Software Features”
Chapter 17, “System Specifications”
Chapter 18, “System Interconnections”
Chapter 19, “Preventive Maintenance”
Chapter 16, “Reference Graphs and Tables”
N5511A Phase Noise Test System User’s Guide 15
Page 16
Getting Started Additional Documentation

Additional Documentation

The N5511A system documentation includes:
— Keysight N5511A Phase Noise Test System Installation Guide
— Keysight N5511A Series Phase Noise Test Systems SCPI Command
Reference
16 N5511A Phase Noise Test System User’s Guide
Page 17
Getting Started System Overview

System Overview

The Keysight N5511A Phase Noise Test System provides flexible sets of measurements on one-port devices such as voltage controlled oscillators (VCOs), dielectric resonator oscillators (DROs), crystal oscillators, and synthesizers, and on two-port devices such as amplifiers and frequency converters. The N5511A system measures absolute and residual phase noise, and AM noise for CW and pulsed signals. It operates in the frequency range of 50 kHz to 40 GHz.
The N5511A Phase Noise Test System combines standard instruments, phase noise measurement components, and PC software for maximum flexibility and re-use of assets. The system PC operates under Windows 10 Professional and controls the system through the N5510 measurement software. The N5510 software enables many stand-alone instruments to work in the system. This stand-alone instrument architecture easily configures for various measurement techniques, including the absolute phase noise PLL/reference-source technique, and delay-line and FM-discriminator methods.
The N5511A system is available as a benchtop model. Due to the system’s flexibility, the hardware in the system varies greatly with the options selected. You may be installing instruments you already own in the system as well. A typical N5511A system includes these components:
— N5511A PXIe chassis
— M9037A Controller with removable SSD drive with Windows 10 Professional
— Keysight N5510 Phase Noise Measurement software
— M9550A Phase Detector 1 or 2
— M9551A Data Converter
— M9300A Frequency Reference
— Microwave Power Splitter
Customer provided monitor with display port cable, keyboard, mouse, and RF source(s).
Additional instruments may include a spectrum analyzer, oscilloscope, RF counter, and power meter.
For detailed information on the instruments in your Keysight N5511A Phase Noise Test System, refer to the individual instrument user guides.
N5511A Phase Noise Test System User’s Guide 17
Page 18
Getting Started System Overview
Figure 1-1 Keysight N5511A benchtop system
Figure 1-1 shows the N5511A Phase Noise Test System.
The N5511A can replace earlier Keysight E5505A phase noise systems. The N5511A system uses a LAN or USB/GPIB port to communicate with the assets in the system. However, the N5511A system and N5510 software are backwards compatible with earlier E5505A systems and instruments. You may easily integrate existing assets into your N5511A system. Figure 1-2 and Table
1-2 show the N5511A and earlier-model equivalents.
GPIB communication is done by using an 82357B GPIB to USB interface adapter.
18 N5511A Phase Noise Test System User’s Guide
Page 19
Getting Started System Overview
Figure 1-2 Keysight E5505A system comparison to N5511A system
N5511A Phase Noise Test System User’s Guide 19
Page 20
Getting Started System Overview
Table 1-2 Equivalent system/instrument model numbers
System or Instrument Number Old Number
Phase Noise Test System N5511A E5505A
50 kHz - 40 GHz phase detector M9550A N5500A-201 50kHz - 26.5 GHz
Baseband Test Set
FFT/ Data Converter DC - 160 MHz M9551A E5505A-RHK
PC with data converter card and swept analyzer
20 N5511A Phase Noise Test System User’s Guide
Page 21
Keysight N5511A Phase Noise Test System
User’s Guide

2 Introduction

“Introducing the GUI” on page 22
“Designing to Meet Your Needs” on page 24
“N5511A Operation: A Guided Tour” on page 25
“Powering the System On” on page 26
“Starting the Measurement Software” on page 27
“Powering the System Off” on page 30
21
Page 22
Introduction Introducing the GUI

Introducing the GUI

The graphical user interface (GUI) gives the user instant access to all measurement functions, making it easy to configure a system and define or initiate measurements. The most frequently used functions are displayed as icons on a toolbar, allowing quick and easy access to the measurement information.
The forms-based graphical interaction helps you define your measurement quickly and easily. Each form tab is labeled with its content, preventing you from getting lost in the defining process.
The system provides three default segment tables. To obtain a quick look at your data, select the “fast” quality level. If it is important to have more frequency resolution to separate spurious signals, use the “normal” and “high resolution” quality levels. If you need to customize the offset range beyond the defaults provided, tailor the measurement segment tables to meet your needs and save them as a custom selection.
You can place up to nine markers on the data trace that can be plotted with the measured data.
Other features include:
— Plotting data without spurs
— Tabular listing of spurs
— Plotting in alternate bandwidths
— Parameter summary
— Color printouts to any supported color printer
Figure 2-1 shows an example of the GUI.
22 N5511A Phase Noise Test System User’s Guide
Page 23
Introduction Introducing the GUI
Figure 2-1 N5510 graphical user interface (GUI)
N5511A Phase Noise Test System User’s Guide 23
Page 24
Introduction Designing to Meet Your Needs

Designing to Meet Your Needs

The N5511A Phase Noise Test System is a high performance measurement tool that enables you to fully evaluate the noise characteristics of your electronic instruments and components with unprecedented speed and ease. The phase noise measurement system provides you with the flexibility needed to meet today’s broad range of noise measurement requirements.
In order to use the phase noise system effectively, it is important that you have a good understanding of the noise measurement you are making. This manual is designed to help you gain that understanding and quickly progress from a beginning user of the phase noise system to a proficient user of the system’s basic measurement capabilities.
If you have just received your system or need help with connecting the hardware or loading software, refer to your Keysight N5511A Phase Noise Test System Installation Guide now. Once you have completed the installation procedures, return to “N5511A Operation: A Guided
Tour” to begin learning how to make noise measurements with the system.

Beginning

The section “N5511A Operation: A Guided Tour” contains a step-by-step procedure for completing a phase noise measurement. This measurement demonstration introduces system operating fundamentals for whatever type of device you plan to measure.
Once you are familiar with the information in this chapter, you should be prepared to start Chapter 4, “Expanding Your Measurement Experience”. After you have completed that chapter, refer to Chapter 14, “Evaluating Your
Measurement Results” for help in analyzing and verifying your test results.
24 N5511A Phase Noise Test System User’s Guide
Page 25
Introduction N5511A Operation: A Guided Tour

N5511A Operation: A Guided Tour

This measurement demonstration introduces you to the system’s operation by guiding you through an actual phase noise measurement.
You will be measuring the phase noise of the Keysight N5500A Phase Noise Test Set’s low noise amplifier. (The measurement made in this demonstration is the same measurement that is made to verify the system’s operation.)
As you step through the measurement procedures, you will soon discover that the phase noise measurement system offers enormous flexibility for measuring the noise characteristics of your signal sources and two-port devices.

Required equipment

The equipment shipped with this system is all that is required to complete this demonstration. (Refer to the N5511A Phase Noise Test System Installation Guide if you need information about setting up the hardware or installing the software.)

How to begin

Follow the setup procedures beginning on the next page. The phase noise measurement system displays a setup diagram that shows you the front panel cable connections to make for this measurement.
If you need additional information about connecting instruments, refer to Chapter 18,
“System Interconnections”.
N5511A Phase Noise Test System User’s Guide 25
Page 26
Introduction Powering the System On

Powering the System On

Connect your system to an appropriate AC power source using the power cord provided.
The N5511A system is shipped with an AC power cord appropriate for your location.
Before applying power, make sure the AC power input and the location of the system meet the requirements given in the Getting Started guide for your system. Failure to do so may result in damage to the system or personal injury.
Warm-up Time: The downconverter and RF source instruments contain ovenized oscillators which must warm up for 30 minutes to produce accurate measurements.
Standby Mode: The RF source uses a standby mode to keep the ovenized oscillator warm when the instrument is connected (plugged in) to AC power, even when the power switch is in the off position. To completely shut down the instrument, you must disconnect it from the AC power supply.
The N5511A Benchtop system consists of an N5511A Phase Noise Test System with one or two test sets installed. You must connect a monitor, keyboard, and mouse before powering on the system.
Press the system power switch.
Figure 2-2 Power on the N5511A System

To power on a racked system

1. Press the system power switch (front, top right of the rack) to the on position.
2. Verify that all instrument power switches are on.
3. Allow the system to warm up for 30 minutes.
26 N5511A Phase Noise Test System User’s Guide
Page 27
Introduction Starting the Measurement Software

Starting the Measurement Software

The N5510 software is pre-installed on the N5511A Phase Noise system.
Keysight Technologies, Inc. has not provided internet security software for this N5511A Phase Noise Test System. Connecting the PC to a Local Area Network (LAN), without first installing internet security software (firewall, virus protection, etc) puts both your PC and data at risk. If you decide to connect the N5511A to a LAN, without first installing internet security software, you do so at your own risk.
Keysight recommends turning on Windows updates and installing updates when available from Microsoft.
Choose the N5510 software icon to launch the user interface.
Figure 2-3 Splash Screen
N5511A Phase Noise Test System User’s Guide 27
Page 28
Introduction Starting the Measurement Software
1. To start the program, double-click on the N5510 icon on the desktop shortcut (shown above), or navigate to the N5510 User Interface through the Windows start menu. Click Start > All Programs > Keysight N5510 >
N5510 User Interface.
2. When the program starts, the main N5510 measurement window appears
(see Figure 2-4). It shows the phase noise graph.
Figure 2-4 Main N5510 user interface window
28 N5511A Phase Noise Test System User’s Guide
Page 29
Introduction Starting the Measurement Software

Verify License Key is Installed

The N5511A will have the license key already installed, but if you ever need to install the license key, use the following procedure.
3. Use the Keysight License Manager to see the license keys installed. Start > All Programs > Keysight License Manager > Keysight License Manager
4. Verify the licenses are installed.
Figure 2-5 Keysight License Manager
N5511A Phase Noise Test System User’s Guide 29
Page 30
Introduction Powering the System Off

Powering the System Off

1. On the N5510 software menu, select File\Exit, Start icon, then shut down. Always shut down the N5510 software before powering off the N5511A system.
2. Use the Start menu to shut down the PC. Press the power switch on each instrument to the off position.
If you receive error messages during the power on or off procedures, or during operation, use the Windows event log for detailed information on the errors.
30 N5511A Phase Noise Test System User’s Guide
Page 31
Keysight N5511A Phase Noise Test System
User’s Guide

3 Phase Noise Basics

“What is Phase Noise?” on page 32
“Ideal vs Real Word Signals” on page 33
“Phase terms” on page 34
31
Page 32
Phase Noise Basics What is Phase Noise?

What is Phase Noise?

Frequency stability can be defined as the degree to which an oscillating source produces the same frequency throughout a specified period of time. Every RF and microwave source exhibit some amount of frequency instability.
This stability can be broken down into two components:
—long-term stability
—short-term stability
Figure 3-1 Frequency Stability
Long-term stability describes the frequency variations that occur over long time periods, expressed in parts per million per hour, day, month, or year.
Short-term stability contains all elements causing frequency changes about the nominal frequency of less than a few seconds duration. The chapter deals with short-term stability.
32 N5511A Phase Noise Test System User’s Guide
Page 33
Phase Noise Basics What is Phase Noise?

Ideal vs Real Word Signals

Mathematically, an ideal sine wave can be described by
Where A
=nominal amplitude, fo=nominal frequency. In the time domain, this
o
signal is a perfect sinusoidal waveform, and in the frequency domain, it is represented by a single spectral line. See Figure 3-2.
Figure 3-2 Single Spectral Line
In practice however, there are always small, unwanted amplitude and phase fluctuations present on the signal. An actual signal is better modeled by
Where E(t) = Amplitude fluctuations, and = randomly fluctuating phase term, or phase noise. This randomly fluctuating phase term could be observed on an ideal RF analyzer (one which has no sideband noise of its own) as seen in
Figure 3-3. The signal is now represented by a spread of spectral lines - both
above and below the nominal signal frequency in the form of modulation sidebands due to the random amplitude and phase fluctuations.
Figure 3-3 Spread of Spectral Lines
N5511A Phase Noise Test System User’s Guide 33
Page 34
Phase Noise Basics What is Phase Noise?

Phase terms

There are two types of fluctuating phase terms:
—spurious signals
— phase noise
Spurious signals
The first are discrete signals appearing as distinct components in the spectral density plot. These signals, commonly called spurious, can be related to known phenomena in the signal source such as power line frequency, vibration frequencies, or mixer products.
Phase noise
The second type of phase instability is random in nature and is commonly called phase noise. The sources of random sideband noise in an oscillator include thermal noise, shot noise, and flicker noise. Many terms exist to quantify the characteristic randomness of phase noise. Essentially, all methods measure the frequency or phase deviation of the source under test in the frequency or time domain. Since frequency and phase are related to each other, all of these terms are also related.
Spectral density
One fundamental description of phase instability or phase noise is spectral density of phase fluctuations on a per-Hertz basis. The term spectral density describes the energy distribution as a continuous function, expressed in units of variance per unit bandwidth. We can the convert rms phase fluctuations into a spectral density by dividing by the bandwidth of the noise sideskirts:
Where BW (bandwidth is negligible with respect to any changes in .
Because phase modulation is a symmetric process (both sidebands are identical), we need only consider one of the noise side skirts. We use the
right-hand side noise side skirt and call that . is directly related to by a simple approximation which has generally negligible error if the modulation sidebands are such that the total phase deviation are much less than 1 radian
<< radian).
(Δφ
pk
34 N5511A Phase Noise Test System User’s Guide
Page 35
Phase Noise Basics What is Phase Noise?
L(f)
Another useful measure of noise energy is L(f), which is then directly related to
by a simple approximation which has generally negligible error if the modulation sidebands are such that the total phase deviation are much less than 1 radian (Δφ
Figure 3-4 CW signal sidebands viewed in the frequency domain
<< radian).
pk
L(f) is an indirect measurement of noise energy easily related to the RF power spectrum observed on an RF analyzer. Figure 3-5 shows that the National Institute of Standards and Technology (NIST) defines L(f) as the ratio of the power--at an offset (f) Hertz away from the carrier. The phase modulation sideband is based on a per Hertz of bandwidth spectral density and or offset frequency in one phase modulation sideband, on a per Hertz of bandwidth spectral density and (f) equals the Fourier frequency or offset frequency.
= single sideband (SSB) phase noise to carrier ration (per Hertz)
N5511A Phase Noise Test System User’s Guide 35
Page 36
Phase Noise Basics What is Phase Noise?
Figure 3-5 Deriving L(f) from a RF analyzer display
L(f) is usually presented logarithmically as a spectral density plot of the phase modulation sidebands in the frequency domain, expressed in dB relative to the carrier per Hz (dBc/Hz) as shown in Figure 3-6. This chapter, except where noted otherwise, uses the logarithmic form of L(f) as follows:
Figure 3-6 L(f) Described Logarithmically as a Function of Offset Frequency
36 N5511A Phase Noise Test System User’s Guide
Page 37
Phase Noise Basics What is Phase Noise?
Caution must be exercised when L(f) is calculated from the spectral density of the phase fluctuations because the calculation of L(f) is dependent on
the small angle criterion. Figure 3-7, the measured phase noise of a free running VCO described in units of L(f), illustrates the erroneous results that can occur if the instantaneous phase modulation exceeds a small angle line. Approaching the carrier L(f) obviously increases in error as it indicates a relative level of +45 dBc/Hz at a 1 Hz offset (45 dB more noise power at a 1 Hz offset in a 1 Hz bandwidth than in the total power of the signal); which is of course invalid.
Figure 3-7 shows a 10 dB/decade line drawn over the plot, indicating a peak
phase deviation of 0.2 radians integrated over any one decade of offset frequency. At approximately 0.2 radians the power in the higher order sidebands of the phase modulation is still insignificant compared to the power in the first order sideband which insures that the calculation of L(f) remains
valid. Above the line the plot of L(f) becomes increasingly invalid, and must be used to represent the phase noise of the signal.
Figure 3-7 Region of validity of L(f)
N5511A Phase Noise Test System User’s Guide 37
Page 38
Phase Noise Basics What is Phase Noise?
38 N5511A Phase Noise Test System User’s Guide
Page 39
Keysight N5511A Phase Noise Test System
User’s Guide

4 Expanding Your Measurement Experience

“GUI Features” on page 40
“Viewing Markers” on page 40
“Display Preferences” on page 41
“Omitting Spurs” on page 43
“Displaying the Parameter Summary” on page 45
“Exporting Measurement Results” on page 47
“Saving Measurement State” on page 51
“Recall Measurement State” on page 51
“Measurement Preferences” on page 52
“Toolbar” on page 53
39
Page 40
Expanding Your Measurement Experience GUI Features

GUI Features

Viewing Markers

The marker function allows you to display the exact frequency and amplitude of any point on the results graph.
To access the marker function, on the View menu, click Markers. In the dialog box containing Marker buttons at the bottom of the application, up to nine markers may be added. To add a marker, click Add Marker at the bottom of the display and to remove a highlighted marker, click the Delete Marker button at the bottom of the display. Markers are added to the latest measured trace on the display.
Figure 4-1 View Markers
Figure 4-2 Add/Delete Markers
40 N5511A Phase Noise Test System User’s Guide
Page 41
Expanding Your Measurement Experience GUI Features

Display Preferences

N5510 User Interface display colors are highly customizable. Navigate to View, allowing the ability for the user to customize the colors of various items on the display.
Changing the noise color will affect the noise trace of the next measurement, not the currently displayed trace.
Figure 4-3 View Display Preferences
Display Parameters. The following menu will appear
N5511A Phase Noise Test System User’s Guide 41
Page 42
Expanding Your Measurement Experience GUI Features
Figure 4-4 Display Preferences Window
42 N5511A Phase Noise Test System User’s Guide
Page 43
Expanding Your Measurement Experience GUI Features

Omitting Spurs

The Omit Spurs function plots the currently loaded results without displaying any spurs that may be present. The ability to omit or view spurs is conditional on the 'Mark Spurs' option under the Type and Range tab of the measurement definition being checked.
1. On the View menu, click Display Preferences.
2. In the Display Preferences dialog box, uncheck Spurs and click OK. See
Figure 4-5.
Figure 4-5 Uncheck spurs
N5511A Phase Noise Test System User’s Guide 43
Page 44
Expanding Your Measurement Experience GUI Features
3. The graph is displayed without spurs. See Figure 4-6.
Figure 4-6 Graph displayed without spurs
4. To re-display the spurs, check Spurs in the Display Preferences dialog
box. This feature can also be accessed by right-clicking on the plot and unselecting View Spurs. Figure 4-7 shows the graph displayed with spurs.
Figure 4-7 Graph displayed with spurs
44 N5511A Phase Noise Test System User’s Guide
Page 45
Expanding Your Measurement Experience GUI Features

Displaying the Parameter Summary

The Parameter Summary function allows you to quickly review the measurement parameter entries that were used for this measurement. The parameter summary data is included when you print the graph.
1. On the View menu, click Parameter Summary. See Figure 4-8.
Figure 4-8 Navigate to Parameter Summary
N5511A Phase Noise Test System User’s Guide 45
Page 46
Expanding Your Measurement Experience GUI Features
2. The Parameter Summary Notepad dialog box appears. The data can be
printed or changed using standard Notepad functionality. See Figure 4-9.
Figure 4-9 Parameter summary notepad
46 N5511A Phase Noise Test System User’s Guide
Page 47
Expanding Your Measurement Experience GUI Features

Exporting Measurement Results

The Export Measurement Results function exports data in one of three types:
— Exporting Trace Data
— Exporting Spur Data
— Exporting X-Y Data
1. To export measurement results, on the File menu, point to Export Results,
then click on either Trace Data, Spur Data, or X-Y Data. See Figure 4-10.
Figure 4-10 Export results choices
N5511A Phase Noise Test System User’s Guide 47
Page 48
Expanding Your Measurement Experience GUI Features
Exporting Trace Data
1. On the File menu, point to Export Results, then click on Trace Data. See
Figure 4-11.
Figure 4-11 Tra ce data results
48 N5511A Phase Noise Test System User’s Guide
Page 49
Expanding Your Measurement Experience GUI Features
Exporting spur data
1. On the File menu, point to Export Results, then click on Spur Data. See
Figure 4-12.
Figure 4-12 Spur data results
N5511A Phase Noise Test System User’s Guide 49
Page 50
Expanding Your Measurement Experience GUI Features
Exporting X-Y Data
1. On the File menu, point to Export Results, then click on X-Y Data. See
Figure 4-13.
Figure 4-13 X-Y data results
50 N5511A Phase Noise Test System User’s Guide
Page 51
Expanding Your Measurement Experience GUI Features

Saving Measurement State

The current measurement state can be saved as a .pnx file. This will save all of the measurement parameters as well as the trace that is on the plot. Click File, Save as… Select Use Title as Filename to name the file using the title of the graph.
Figure 4-14 Saving Measurement State

Recall Measurement State

A .pnx file can be opened to recall a previously saved state. Navigate to File, Open and browse to the state file to recall. Note, N5510 Phase Noise Measurement Software can recall legacy files in a .pnm format as well. Multiple .pnx files can be recalled, resulting in overlaying traces with the active measurement parameters being those of the last recalled file.
N5511A Phase Noise Test System User’s Guide 51
Page 52
Expanding Your Measurement Experience GUI Features

Measurement Preferences

Figure 4-15 Measurement Preferences
Pause At Connection Diagram: check to skip Connection Diagram when a New Measurement is initiated.
Don't Automatically Clear Graph: when restarting a measurement, the default behavior is for the current trace to be cleared. To retain the trace, uncheck this option.
Don't Automatically change Previous Graph Color: when retaining a trace, the application defaults the previous trace to a different color to not conflict with the new trace. To keep the previous trace color, uncheck this option.
Don't Mark Spur: spurs will not be marked as spurs.
52 N5511A Phase Noise Test System User’s Guide
Page 53
Expanding Your Measurement Experience GUI Features

Toolbar

Figure 4-16 Toolbar selections
N5511A Phase Noise Test System User’s Guide 53
Page 54
Expanding Your Measurement Experience GUI Features
54 N5511A Phase Noise Test System User’s Guide
Page 55
Keysight N5511A Phase Noise Measurement System
User’s Guide

5 Absolute Measurement Fundamentals

“The Phase-Lock-Loop Technique” on page 56
“The Phase-Lock Loop Circuit” on page 60
“What Sets the Measurement Noise Floor?” on page 63
“Selecting a Reference (Single Channel)” on page 65
“Estimating the Tuning Constant” on page 71
“Tracking Frequency Drift” on page 72
“Changing the PTR” on page 74
“Minimizing Injection Locking” on page 76
“Inserting a Device” on page 79
“Evaluating Noise Above the Small Angle Line” on page 82
“Calibration” on page 86
55
Page 56
Absolute Measurement Fundamentals The Phase-Lock-Loop Technique

The Phase-Lock-Loop Technique

Single Channel

The phase lock loop measurement technique for a single channel setup requires two signal sources; the device-under-test and a reference source. This measurement type requires that one of the two sources is a voltage-controlled-oscillator (VCO).
In a PLL configuration, the reference source is locked in quadrature with the DUT, which means that the signal from the references is 90 degrees out of phase with the DUT. The signals from the DUT and the reference serve as the RF and LO inputs to the phase detector. The carrier is canceled by the phase detector, leaving only noise components in the resultant measurement.
Figure 5-1 Single Channel Phase Lock Loop
A closer look:
Figure 5-2 shows a high-level walk-through of the theory behind the single
channel PLL measurement technique.
Figure 5-2 Single Channel PLL Theory
56 N5511A Phase Noise Test System User’s Guide
Page 57
Absolute Measurement Fundamentals The Phase-Lock-Loop Technique
Let the input signal from the DUT be denoted by
Let the signal from the reference be denoted by -recall the reference is locked in quadrature.
Let the noise floor of the system be denoted by
Consider the sum and difference trigonometric formula:
1. The output of the phase detector therefore is:
2. The output of the phase detector therefore is:
Consider the small angle theorem sin(A)=A. The filtered output will be represented as:
3. Noise contributed from the channel internally is added to the overall
output. The FFT output can be represented as a sum:
When using the PLL technique in a single channel configuration, the overall output will include the phase noise of reference and the noise floor of the system. If the phase noise of the DUT is better than the phase noise of the reference, the phase noise of the reference will limit the dynamic range. If the DUT is better than the noise floor of the system, the measurement will also not result in the true performance of the device under test. A method to significantly increase dynamic range is to perform a dual-channel measurement instead of single channel.
N5511A Phase Noise Test System User’s Guide 57
Page 58
Absolute Measurement Fundamentals The Phase-Lock-Loop Technique

Dual Channel

N5511A has the option for a second phase detector module, which allows for dual-channel cross correlation. In a dual channel setup, the DUT signal is split to provide the input signal to each of the phase detector modules. Two separate reference sources are required to provide a reference to each of the phase detectors. The result of this setup is having two separate single channel measurements with a common DUT signal, allowing for uncorrelated noise of the two references as well as the noise contributions from the detectors to be removed by cross-correlation.
Figure 5-3 Dual Channel Phase Lock Loop
A closer look:
Figure 5-4 shows a high-level walk-through of the theory behind the dual
channel PLL measurement technique.
Figure 5-4 Dual Channel PLL Theory
58 N5511A Phase Noise Test System User’s Guide
Page 59
Absolute Measurement Fundamentals The Phase-Lock-Loop Technique
Following the logic used in a single channel measurement, the output of each of the channels in a dual-channel, measurement can be represented as follows:
The two outputs are cross-correlated, resulting in the phase noise of the DUT. A dual-channel measurement removes the uncorrelated noise of the references and the phase detectors, therefore eliminating the limiting factors in a single channel measurement that could limit the dynamic range of the measurement.
N5511A Phase Noise Test System User’s Guide 59
Page 60
Absolute Measurement Fundamentals The Phase-Lock Loop Circuit

The Phase-Lock Loop Circuit

The Capture and Drift tracking ranges

Like other PLL circuits, the phase lock loop created for the measurement has a Capture Range and a drift tracking range. The Capture Range is equal to 5% of the system's peak tuning range, and the drift tracking range is equal to 24% of the system's peak tuning range.
The system's peak tuning range is derived from the tuning characteristics of the VCO source used for the measurement. Figure 5-5 illustrates the relationship that typically exists between the VCO's peak-to-peak tuning range and the tuning range of the system. The system's drift tracking range is limited to a small portion of the peak tuning range to minimize the possibility of measurement accuracy degradation caused by non-linearity across the VCO's
tuning range.

Peak tune range (PTR)

The peak tuning range is determined using two parameters:
— VCO tuning sensitivity (Hz/Volt)
— Total voltage tuning range (Volts)
PTR = (VCO Tuning Sensitivity) X (Total Voltage Tuning Range)
PTR = (100 Hz/V) X (10 V) = 1000 Hz
Figure 5-5 Capture and Drift-Tracking Range with Tuning Range of VCO
As an example:
A Peak Tuning Range of 1000 Hz provides the following ranges:
Capture Range = 0.05 X 1000 Hz = 50 Hz
Drift Tracking Range = 0.24 X 1000 Hz = 240 Hz
60 N5511A Phase Noise Test System User’s Guide
Page 61
Absolute Measurement Fundamentals The Phase-Lock Loop Circuit
Tuning Requirements
The peak tuning range required for a measurement depends on the frequency stability of the sources used. The signals from the sources are mixed in each of the channel's phase detectors to create a beat note. In order for the loop to acquire lock, the center frequencies of the sources must be close enough together to create a beat note that is within the system's Capture Range. Once the loop is locked, the frequency of the beat note must remain within the drift tracking range for the duration of the measurement. In Figure 5-6, the ranges calculated in the previous example are marked to show their relationship to the beat note frequency.
Figure 5-6 Capture and Drift-Tracking Ranges and Beat Note Frequency
If the beat note does not remain within the drift tracking range during the measurement, the out of lock detector is set and the System stops the measurement. If this happens, you need to increase the system's drift tracking range by increasing the system's peak tuning range (if possible) or by selecting a VCO source with a greater tuning range.
N5511A Phase Noise Test System User’s Guide 61
Page 62
Absolute Measurement Fundamentals The Phase-Lock Loop Circuit

Selecting the VCO Source

Although you must select a VCO source that provides a sufficient tuning range to permit the system to track the beat note, keep in mind that a wide tuning range typically means a higher noise level on the VCO source signal. When the VCO source for your measurement is also the reference source, this trade-off can make reference source selection the most critical aspect of your measurement setup.
Specifying your VCO Source
When you set up your PLL measurement, you need to know four things about the tuning characteristics of the VCO source you are using. The System determines the VCO source's peak tuning range from these four parameters.
— Tuning Constant, estimated tuning sensitivity (Hz/V)
— Center Voltage of Tuning Range, (V)
— Tune Range of VCO, (±V)
— Input Resistance of Tuning Port, (ohms) if the tuning constant is not to be
measured.
The measurement examples in the next chapter that recommend a specific VCO source provides you with the tuning parameters for the specified source.
62 N5511A Phase Noise Test System User’s Guide
Page 63
Absolute Measurement Fundamentals What Sets the Measurement Noise Floor?

What Sets the Measurement Noise Floor?

The noise floor for your measurement is set by two things:
— The noise floor of the phase detector and low-noise amplifier (LNA)
— The noise level of the reference source you are using

The System Noise Floor

The noise floor of the system is directly related to the amplitude of the input signal at the R input port of the system’s phase detector. Table 5- 1 shows the amplitude ranges for the L and R ports.
Table 5-1 Amplitude ranges for L and R ports
Phase Detector
50 kHz to 1.6 GHz
Ref Input
(L Port)
+ 15 dBm
to
+ 23 dBm
a. Phase noise test set Options 001 and 201 with no attenuation. b. Phase noise test set Option 001 with no attenuation.
Signal Input
(R Port)
0 dBm
to
+ 23 dBm
1.2 to 26.5 GHz
Ref Input
(L Port)
+ 7 dBm
to
+ 10 dBm
If the L port (Reference Input) signal is within the amplitude range shown in
Table 5-1, the signal level at the R (Signal Input) port sets the noise floor for
the system.
a
Signal Input
(R Port)
0 dBm
to
+ 5 dBm
50 kHz to 26.5 GHz
AM Noise
0 dBm
to
20 dBm
b
N5511A Phase Noise Test System User’s Guide 63
Page 64
Absolute Measurement Fundamentals What Sets the Measurement Noise Floor?
Figure 5-7 shows the relationship between the R (signal) input level and the
system noise floor.
Figure 5-7 Relationship between the R input level and system noise floor

The Noise Level of the Reference Source

Unless it is below the system’s noise floor, the noise level of the source you are using as the reference source sets the noise floor for the measurement. When you set up your measurement, you want to use a reference source with a noise level that is at or below the level of the source you are going to measure.
Figure 5-8 demonstrates that as the noise level of the reference source
approaches the noise level of the DUT, the level measured by the System (which is the sum of all noise sources affecting the system) is increased above the actual noise level of the DUT.
Figure 5-8 Reference source noise approaches DUT noise
64 N5511A Phase Noise Test System User’s Guide
Page 65
Absolute Measurement Fundamentals Selecting a Reference (Single Channel)

Selecting a Reference (Single Channel)

Selecting an appropriate reference source is critical when you are making a phase noise measurement using the phase lock loop technique. The key to selecting a reference source is to compare the noise level of the reference with the expected noise level of the DUT. In general, the lower the reference source’s noise level is below the expected noise level of the DUT the better. (Keep in mind that you only need to be concerned about the reference source’s noise level within the frequency offset range over which you plan to measure the DUT.)
As shown by the graph in Figure 5-9, the further the reference source’s noise level is below the noise level of the DUT, the less the reference source’s noise contributes to the measurement results.
Figure 5-9 DUT noise approaches reference noise

Using a Similar Device

The test system performs best when you are able to use a device similar to the DUT as the reference source for your PLL measurement. Of course one of the devices must be capable of being voltage tuned by the system to do this.
To select a similar device for use as the reference source, you must establish that the noise level of the reference source device is adequate to measure your DUT. The Three Source Comparison technique enables you to establish the actual noise levels of three comparable devices when two devices are available in addition to the DUT.
If only one device is available in addition to the DUT, you can perform the Phase Noise Using a Phase Locked Loop Measurement using these two devices and know that the noise level of each of the devices is at least as good as the measured results. (The measured results represent the sum of the noise of both devices.)

Using a Signal Generator

When using a signal generator as a reference source, it is important that the generator’s noise characteristics are adequate for measuring your device.
N5511A Phase Noise Test System User’s Guide 65
Page 66
Absolute Measurement Fundamentals Selecting a Reference (Dual Channel)

Selecting a Reference (Dual Channel)

Selecting references for a dual channel measurement can be less critical than in a single channel setup. The one true requirement is that the references are capable of being voltage tuned. With cross-correlation, PNTS removes any noise that doesn't originate from the DUT and achieves an ultimate sensitivity established by thermal phase noise at -177 dBm/Hz. However, the greater the amount of uncorrelated noise present in the system, the more time it takes for the system to remove it. This translates to low-performance reference sources causing a longer measurement time if their phase noise is significantly worse than the device being tested.
If the references are identical or better in performance than the device under test, the N5511A measurement sensitivity starts at the DUT phase noise performance level. This means the cross-correlation process starts out at this sensitivity and this system sensitivity only improves as cross-correlations are processed. This can be quantified by saying that for a 10 times increase in the number of cross-correlations, there is a 5 dB reduction in uncorrelated noise (and 5 dB improvement in PNTS system sensitivity). Therefore, in order to minimize the time required to measure the DUT, references should be as good or even better than the DUT.
User devices today often exceed the best signal generator's (or internal references in some phase noise systems) phase noise performance. In scenarios like this, the flexibility of N5511A allows for copies of the DUT to be used as references in order to measure high performance devices without taking a toll on time from cross-correlating out noise from low-performing references. Figure 5-10, Figure 5-11, and Figure 5-11 show examples of DUTs being measured using copies of the same high performance device as references.
66 N5511A Phase Noise Test System User’s Guide
Page 67
Absolute Measurement Fundamentals Selecting a Reference (Dual Channel)
Example 1: Using a 100 MHz high-performance DUT as both REF1 and REF2 significantly reduces the number of cross-correlations - resulting in a dramatic reduction in measurement time (~40 second to get to a -184 dBc/Hz correlated device noise floor.
Figure 5-10 Using a 100 MHz High-Performance DUT
Example 2: Using a 9.6 GHz high-performance DUT as both REF1 and REF2 significantly reduces the number of cross-correlations - resulting in a dramatic reduction in measurement time (~25 second to get to a -171 dBc/Hz correlated device phase noise floor).
Figure 5-11 Using a 9.6 GHz High-Performance DUT
N5511A Phase Noise Test System User’s Guide 67
Page 68
Absolute Measurement Fundamentals Selecting a Reference (Dual Channel)
Example 3: Using a 10MHz high-performance DUT as both REF1 and REF2 significantly reduces the number of cross-correlations - resulting in a dramatic reduction in measurement time (~20 minutes to get to a -60 dBc/Hz correlated device phase noise floor at a .01 Hz offset).
Figure 5-12 Using a 10 MHz High-Performance DUT
68 N5511A Phase Noise Test System User’s Guide
Page 69
Absolute Measurement Fundamentals Tuning Requirements

Tuning Requirements

Often the reference source you select also serves as the VCO source for the PLL measurement. (The VCO source can be either the DUT or the reference source.) To configure a PLL measurement, you need to know the following tuning information about the VCO source you are using.
— Tuning Constant (Hz/V) (within a factor of 2)
— Tuning Voltage Range (V)
— Center Voltage of Tuning Range (V)
— Input Resistance of Tuning Port (W)
The primary consideration when evaluating a potential VCO source for your measurement is whether it provides the test system with sufficient capture and drift tracking ranges to maintain lock throughout the measurement. To make this determination, you must estimate what the drift range of the sources you are using will be over the measurement period (thirty minutes maximum). (Details on the relationship between the capture and drift tracking ranges and the tuning range of the VCO source are provided in Table 5-2. This information helps you evaluate your VCO source based on the estimated drift of your sources.)
Table 5-2 lists the tuning parameters for several VCO options.
Table 5-2 Tuning Characteristics of Various VCO Source Options
VCO Source Carrier
Freq.
Agilent 8257D Measure
EFC
DCFM
Keysight 8662/3A
EFC υ
DCFM FM Deviation 0 10 1 K (8662) Compute
Keysight 8642A/B FM Deviation 0 10 600 Compute
Keysight 8644B FM Deviation 0 10 600 Compute
υ
0
υ
0
0
Tuning Constant (Hz/V)
7 E - 8 x υ
FM Deviation 0 10 50
5 E – 9 x υ
0
0
Center Voltage (V)
0 10 1E + 6 Compute
010 1E + 6Measure
Voltage Tun ing Range (± V)
Input Resistance (W)
600
600 (8663) Compute
Tuning Calibration Method
Compute
Other Signal Generator
DCFM Calibrated for ±1V FM Deviation 0 10 Rin Compute
N5511A Phase Noise Test System User’s Guide 69
Page 70
Absolute Measurement Fundamentals Tuning Requirements
Table 5-2 Tuning Characteristics of Various VCO Source Options
VCO Source Carrier
Freq.
Other User VCO Source Estimated within a
Tuning Constant (Hz/V)
factor of 2
Center Voltage (V)
–10 to +10
Voltage Tun ing Range (± V)
See Figure
5-13
Input Resistance (W)
1 E + 6 Measure
Tuning Calibration Method
Figure 5-13 Voltage tuning range limits relative to center voltage of the VCO tuning curve
70 N5511A Phase Noise Test System User’s Guide
Page 71
Absolute Measurement Fundamentals Estimating the Tuning Constant

Estimating the Tuning Constant

The VCO tuning constant is the tuning sensitivity of the VCO source in Hz/V. The required accuracy of the entered tuning constant value depends on the VCO tuning constant calibration method specified for the measurement. The calibration method is selected in the Calibration Process menu. Table 5-3 lists the calibration method choices and the tuning constant accuracy required for each.
Table 5-3 VCO tuning constant calibration method
VCO Tuning Constant Calibration Method (selected in calibration screen)
Use the current tuning constant
(must be accurate from a previous measurement of the same source).
Measure the VCO tuning constant Within a factor of 2 of actual value.
Required Tuning Constant Accuracy (entered in parameter screen)
Within a factor of 2 of actual value.
(Enter 1 E + 6 for Input Resistance.)
(Enter 1 E + 6 for Input Resistance.)
N5511A Phase Noise Test System User’s Guide 71
Page 72
Absolute Measurement Fundamentals Tracking Frequency Drift

Tracking Frequency Drift

The system’s frequency drift tracking capability for the phase lock loop measurement is directly related to the tuning range of the VCO source being used. The system’s drift tracking range is approximately 24% of the peak tuning range (PTR) of the VCO.
PTR= VCO Tuning Constant X Voltage Tuning Range
This is the frequency range within which the beat note signal created by the test set’s phase detector must remain throughout the measurement period. In addition, the beat note signal must remain within the system’s Capture Range (5% of the PTR) during the time it takes the system to calibrate and lock the phase lock loop.
The stability of the beat note is a function of the combined frequency stability of the sources being used for the measurement. If beat note drift prevents the beat note from remaining within the Capture Range long enough for the system to attain phase lock, the computer informs you by displaying a message. If the beat note drifts beyond the drift tracking range during the measurement, the computer stops the measurement and inform you that the system has lost lock.
72 N5511A Phase Noise Test System User’s Guide
Page 73
Absolute Measurement Fundamentals Tracking Frequency Drift

Evaluating beat note drift

The Checking the beat note section included in each phase lock loop measurement example in this chapter provides a procedure for adjusting the beat note to within the Capture Range set for the measurement. If you have not done so already, verify that the beat note signal can be tuned to within the Capture Range and that it will remain within the range.
Continue to observe the beat note and verify that it will not drift beyond the drift tracking range (24% of the PTR) during the measurement period. The length of the measurement period is primarily a function of the frequency offset range specified for the measurement (Start to Stop Frequency).
Action
If beat note drift exceeds the limits of the Capture or drift tracking ranges set for your measurement, the system is not able to complete the measurement. You have two possible alternatives.
1. Minimize beat note drift.
By Allowing sources to warm-up sufficiently.
By Selecting a different reference source with less drift.
2. Increase the capture and drift tracking Ranges.
By Selecting a measurement example in this chapter that specifies a
drift rate compatible with the beat note drift rate you have observed.
By Increasing the peak tuning range for the measurement. (Further
information about increasing the PTR is provided in Changing the PTR.)
N5511A Phase Noise Test System User’s Guide 73
Page 74
Absolute Measurement Fundamentals Changing the PTR

Changing the PTR

The peak tuning range (PTR) for the phase lock loop measurement is set by the tune range entered for the VCO and the VCO’s tuning constant. (If the calibration technique is set to measure the VCO tuning constant, the measured value is used to determine the system’s PTR.)
PTR= VCO Tuning Constant X Voltage Tuning Range
From the PTR, the phase noise software derives the capture and drift tracking Ranges for the measurement. These ranges set the frequency stability requirements for the sources being used.
The PTR also determines the phase lock loop (PLL) bandwidth for the measurement. An important attribute of the PLL bandwidth is that it suppresses the close-in noise which would otherwise prevent the system from locking the loop.
Figure 5-14 Peak tuning range
74 N5511A Phase Noise Test System User’s Guide
Page 75
Absolute Measurement Fundamentals Changing the PTR

The Tuning Qualifications

Changing the PTR is accomplished by changing the tune range of VCO value or the VCO tuning constant value or both. There are several ways this can be done. However, when considering these or any other options for changing the PTR, it is important to remember that the VCO source must always meet the following tuning qualifications.
— The tuning response of the VCO source must always remain monotonic.
— The VCO source’s output level must remain constant across its tuning
range.
As long as these qualifications are met, and the software does not indicate any difficulty in establishing its calibration criteria, an increase in PTR will not degrade the system’s measurement accuracy.
The following methods may be considered for increasing or decreasing the PTR.
Voltage controlled oscillators
1. Select a different VCO source that has the tuning capabilities needed for
the measurement.
2. Increase the tune range of the VCO source.
Be careful not to exceed the input voltage limitations of the Tune Port on the VCO source.
Increasing the tune range of the VCO is only valid as long as the VCO source is able to continuously meet the previously mentioned tuning qualifications.
N5511A Phase Noise Test System User’s Guide 75
Page 76
Absolute Measurement Fundamentals Minimizing Injection Locking

Minimizing Injection Locking

Injection locking occurs when a signal feeds back into an oscillator through its output path. This can cause the oscillator to become locked to the injected signal rather than to the reference signal for the phase locked loop.
Injection locking is possible whenever the buffering at the output of an oscillator is not sufficient to prevent a signal from entering. If the injection locking occurs at an offset frequency that is not well within the PLL bandwidth set for the measurement, it can cause the system to lose phase lock.

Adding Isolation

The best way to prevent injection locking is to isolate the output of the source being injection locked (typically the DUT) by increasing the buffering at its output. This can be accomplished by inserting a low noise amplifier and/or an attenuator between the output of the source being injection locked and the test set. (Refer to “Inserting a Device” in this section.
In N5511A, one can troubleshoot isolation issues through an oscilloscope connected to the Monitor outputs of the phase detector modules.
Figure 5-15 shows the beat notes from an absolute phase noise measurement
of a 10 MHz OCXO. Notice the impurity of the signal present at the output of the phase detector. This reflects isolation issues. By adding isolation in each channel, in this case by using amplifiers, the issue is improved.
Figure 5-15 Beat Notes from an Absolute Phase Noise Measurement of a 10 MHz OCXO
76 N5511A Phase Noise Test System User’s Guide
Page 77
Absolute Measurement Fundamentals Minimizing Injection Locking

Increasing the PLL Bandwidth

If the injection locking bandwidth is less or equal to the PLL bandwidth, it may be possible to increase the PLL bandwidth sufficiently to complete the measurement. The PLL bandwidth is increased by increasing the peak tuning range (PTR) for the measurement.
The PTR for the measurement is set by the tuning characteristics of the VCO source you are using. Figure 5-16 shows that increasing the PLL bandwidth can require a substantially larger increase in the PTR. For information on the limitations of increasing the PTR, refer to “Changing
the PTR” in this section.
To estimate the PTR needed to prevent injection locking from causing the system to lose lock:
1. Determine the injection locking bandwidth. Tune the beat note toward
0 Hz using the procedure described in the Checking the beat note section of each phase lock loop measurement example in this chapter. When the injection locking occurs, the beat note disappears. The injection locking bandwidth is the frequency of the beat note just prior to where the injection locking occurs as the beat note is tuned toward 0 Hz.
2. Multiply the injection locking bandwidth by 2 to determine the minimum
PLL bandwidth required to prevent the injection locking from causing the system to lose lock. (To prevent accuracy degradation, it may be necessary to increase the PLL bandwidth to 4 X the injection locking bandwidth. The computer informs you during the measurement if the possibility of accuracy degradation exists.)
3. Locate the required PLL bandwidth in Figure 5-16 to determine the PTR
required for the measurement. (For details on increasing the PTR, refer to
“Changing the PTR” in this section.
N5511A Phase Noise Test System User’s Guide 77
Page 78
Absolute Measurement Fundamentals Minimizing Injection Locking
Figure 5-16 Peak tuning range (PTR) Required by injection locking.
78 N5511A Phase Noise Test System User’s Guide
Page 79
Absolute Measurement Fundamentals Inserting a Device

Inserting a Device

An attenuator

You may find that some of your measurement setups require an in-line device such as an attenuator in one of the signal source paths. (For example, you may find it necessary to insert an attenuator at the output of a DUT to prevent it from being injection-locked to the reference source.) The primary consideration when inserting an attenuator is that the signal source has sufficient output amplitude to maintain the required signal level at the test set’s phase detector input port. The signal level required for the measurement depends on the noise floor level needed to measure the DUT.
Figure 5-17 shows the relationship between the signal level at the R port and
the measurement noise floor.
Figure 5-17 Measurement noise floor relative to R-Port signal level
This is an important consideration in a single channel configuration. For a dual channel configuration, equal attenuation can be placed in both channels. This flexibility allows for cross-correlation to remove the effects of the attenuators and recover the SNR prior to the signal of the DUT being split.
N5511A Phase Noise Test System User’s Guide 79
Page 80
Absolute Measurement Fundamentals Inserting a Device

An amplifier

If a source is not able to provide a sufficient output level, or if additional isolation is needed at the output, it may be necessary to insert a low phase-noise RF amplifier at the output of the source.
Note, however, that the noise of the inserted amplifier is also summed into the measured noise level along with the noise of the source in a single channel measurement.
Use the following equation to estimate what the measurement noise floor is as a result of the added noise of an inserted amplifier: Figure 5-18 shows an example.
L(f) out = –174 dB + Amplifier Noise Figure – Power into Amplifier – 3dB
Figure 5-18 Measurement noise floor as a result of an added attenuator
80 N5511A Phase Noise Test System User’s Guide
Page 81
Absolute Measurement Fundamentals Inserting a Device
The N5511A's dual channel cross-correlation capability offers a unique advantage when performing measurements requiring amplifiers. In a dual channel configuration, amplifiers of identical gain can be placed in each of the separate channel paths. See Figure 5-19.
Figure 5-19 Dual Channel Configuration
When a measurement is configured in such manner, the noise contribution from the amplifiers is removed by cross-correlation due to their noise being uncorrelated. This enables the resulting noise floor measured to be the true performance of the DUT. This technique also means that the performance of the amplifiers need not be a critical factor; however, one must be considerate of the impact that higher noise amplifiers will have on the number of cross-correlations required to reach the correlated noise floor of the DUT.
N5511A Phase Noise Test System User’s Guide 81
Page 82
Absolute Measurement Fundamentals Evaluating Noise Above the Small Angle Line

Evaluating Noise Above the Small Angle Line

If the average noise level on the input signals exceeds approximately 0.1 radians RMS integrated outside of the Phase Lock Loop (PLL) bandwidth, it can prevent the system from attaining phase lock.
The following procedure allows you to evaluate the beat note created between the two sources being measured. The intent is to verify that the PLL bandwidth is adequate to prevent the noise on the two sources from causing the system to lose lock.
If the computer is displaying the hardware Connect Diagram you are ready to begin this procedure. (If it is not, begin a New Measurement and proceed until the hardware Connect Diagram appears on the display.)

Determining the Phase-Lock-Loop bandwidth

1. Determine the Peak Tuning Range (PTR) of your VCO by multiplying the
VCO Tuning Constant by the Tune Range of VCO value entered. (If the phase noise software has measured the VCO Tuning Constant, use the measured value.)
PTR = VCO Tuning Constant X Voltage Tuning
For Example:
2. Estimate the Phase Lock Loop (PLL) bandwidth for the measurement
using the PTR of your VCO and the graph in Figure 5-20.
Observing the beat note
If the beat note frequency is below 100 kHz it appears on the analyzer’s display in both the frequency domain and the time domain. If the beat note does not appear on the RF analyzer, then the beat note is either greater than 100 kHz or it does not exist.
If incrementing the frequency of one of the sources does not produce a beat note within 100 kHz, you need to verify the presence of an output signal from each source before proceeding.
82 N5511A Phase Noise Test System User’s Guide
Page 83
Absolute Measurement Fundamentals Evaluating Noise Above the Small Angle Line
Figure 5-20 Phase lock loop bandwidth provided by the peak tuning range
1. Once the beat note is displayed, Auto Tune the analyzer.
2. Set the span width on the signal analyzer to approximately 4 x PLL
bandwidth. Adjust the beat note to position it near the center of the display.
If you are not able to tune the beat note to 2 X PLL bandwidth (center of display) due to frequency drift, refer to Tracking Frequency Drift in this section for information about measuring drifting signals. If you are able to locate the beat note, but it distorts and then disappears as you adjust it towards 0 Hz, then your sources are injection locking to each other. Set the beat note to the lowest frequency possible before injection locking occurs and then refer to “Minimizing
Injection Locking” on page 76 for recommended actions.
a. Turn on trace averaging.
b. Perform Peak Search.
3. Set a Delta Marker.
On the analyzer, offset the marker by the PLL bandwidth. Read the offset frequency and noise level indicated at the bottom of the display.
4. Compare the average noise level at the PLL bandwidth offset to the small
angle criterion level shown on the graph in Figure 5-21. The average noise level of the signal must remain below the small angle line at all offset frequencies beyond the PLL bandwidth. (The small angle line applies only to the level of the average noise. Spur levels that exceed the small angle line do not degrade measurement accuracy provided they do not exceed — 40 dBc.)
N5511A Phase Noise Test System User’s Guide 83
Page 84
Absolute Measurement Fundamentals Evaluating Noise Above the Small Angle Line
Figure 5-21 Graph of small angle line and spur limit
5. Continue moving the marker to the right to verify that the average noise
level remains below the small angle line.
6. Increase the span by a factor of ten by selecting FREQ and SPAN. Continue
comparing the noise level to the graph.
7. Continue to increase the span width and compare the noise level out to
100 kHz. (If the noise level exceeds the small angle line at any offset frequency beyond the PLL bandwidth, note the offset frequency and level of the noise. Use the graph in Figure 5-22 to determine the Peak Tuning Range (PTR) necessary to provide a sufficient PLL bandwidth to make the measurement.
Figure 5-22 Requirements for noise exceeding small angle limit
84 N5511A Phase Noise Test System User’s Guide
Page 85
Absolute Measurement Fundamentals Evaluating Noise Above the Small Angle Line
Measurement options
If the observed level exceeded the small angle line at any point beyond the PLL bandwidth set for the measurement, you need to consider one of the following measurement options.
1. Evaluate your source using the noise data provided by the RF analyzer in
the procedure you just performed.
2. Increase the PTR if possible, to provide a sufficient PLL bandwidth to
suppress the noise. (For information on increasing the PTR, refer to Changing the PTR in this section.)
3. Reduce the noise level of the signal sources.
4. Use the Discriminator technique to measure the phase noise level of your
source.
N5511A Phase Noise Test System User’s Guide 85
Page 86

Calibration

Overview

Absolute Measurement Fundamentals Calibration
User calibrations are used to establish a reference constant for relative measurements made on the system. For absolute measurements, the amplitude of the carrier needs to be determined before the measurement since the carrier is removed from the measurement when quadrature is established. In absolute measurements various parameter of the reference need to be measured. The type of calibration to be used is determined by the system configuration and equipment availability. User calibrations need to be run every time there is a change to the system or DUT parameters.
For absolute measurements, the N5511A supports four different options for calibration, see Figure 5-23:
— Use current phase detector constant
— Derive detector constant from measured beat note
— Derive detector constant from single-sided spur
— Derive detector constant from double-sided spur
Figure 5-23 Phase Detector Calibration Options
86 N5511A Phase Noise Test System User’s Guide
Page 87
Absolute Measurement Fundamentals Calibration

Measured Beat Note

The measured beat note calibration method is the most common user calibration and is the least complex. This method does not require an additional source to set a calibration tone and is therefore the best option when hardware is limited. The beat note frequency for each channel is set by the relative frequency difference between the DUT and the reference in the respective channel. If the DUT and reference are very accurate sources set at the same frequency, the resulting beat note will be very close to 0 Hz. The advantage is this is a simple method of calibration. The disadvantage is it
Advantages
— Simple method of calibration
Disadvantages
— Requires two RF sources, separated by 0.1 Hz to 50 MHz at the phase
detector. The calibration source output power must be manually adjusted to the same level as the power splitter output it replaces (requires a power meter).
Searching for the beat note will require that you adjust the center frequency of one of the sources above and below the frequency of the other source until the beat note appears on the oscilloscope's display. If incrementing the frequency of one of the sources does not produce a beat note, you will need to verify the presence of an output signal from each source before proceeding.
Theory
Recall the diagram for a single channel setup, Figure 5-24. A "beat note" is established by shifting the frequency of the reference by 10% of the peak tune range. This delta frequency will show up as a spur in the IF at the amplitude of the carrier minus any losses.
The slope is measured in the linear region of the sinusoid and using the "small angle" theorem Vpeak is determined. This value is then used to calculate the dBc values when the phase noise is measured.
Beat note user calibration works identically in the dual channel setup, with the only difference being that the calibration is performed separately by the instrument for each of the phase detector modules.
N5511A Phase Noise Test System User’s Guide 87
Page 88
Absolute Measurement Fundamentals Calibration
Figure 5-24 Beat Note Single Channel
88 N5511A Phase Noise Test System User’s Guide
Page 89
Absolute Measurement Fundamentals Calibration
Checking the Beat Note
While the connect diagram is still displayed, use an oscilloscope (connected to the Monitor port on the test set) or a counter to check the beat note being created between the reference source and your DUT. The objective of checking the beat note is to ensure that the center frequencies of the two sources are close enough in frequency to create a beat note that is within the capture range of the system. The phase lock loop (PLL) capture range is 5% of the peak tuning range of the VCO source you are using. (The peak tuning range for your VCO can be estimated by multiplying the VCO tuning constant by the tune range of VCO. Refer to Chapter 14, “Evaluating Your Measurement Results” if you are not familiar with the relationship between the PLL capture range and the peak tuning range of the VCO.
If the center frequencies of the sources are not close enough to create a beat note within the capture range, the system will not be able to complete its measurement.
The beat note frequency is set by the relative frequency difference between the two sources. If you have two very accurate sources set at the same frequency, the resulting beat note will be very close to 0 Hz. Searching for the beat note will require that you adjust the center frequency of one of the sources above and below the frequency of the other source until the beat note appears on the oscilloscope's display. If incrementing the frequency of one of the sources does not produce a beat note, you will need to verify the presence of an output signal from each source before proceeding.
Figure 5-25 Oscilloscope Display of Beat Note from Test Set Monitor Port
N5511A Phase Noise Test System User’s Guide 89
Page 90
Absolute Measurement Fundamentals Calibration
Estimate the system's capture range (using the VCO source parameters entered for this measurement). The estimated VCO tuning constant must be accurate within a factor of 2. A procedure for “Estimating the Tuning Constant” is located in this chapter.
If you are able to locate the beat note, but it distorts and then disappears as you adjust it towards 0 Hz, your sources are injection locking to each other. Set the beat note to the lowest frequency possible before injection locking occurs and then refer to the “Minimizing Injection Locking” section of this chapter for recommended actions.
If you are not able to tune the beat note to within the capture range due to frequency drift, refer to the “Tracking Frequency Drift” section of this chapter for information about measuring drifting signals.
90 N5511A Phase Noise Test System User’s Guide
Page 91
Absolute Measurement Fundamentals Calibration

Single-sided Spur

Another common calibration method is using a single-sided spur. Single-sided spur method and double-sided spur method are the two most accurate calibration methods. Figure 5-26 shows the setup for a measurement using the single channel single-sided spur calibration.Figure 5-27 shows the setup for a measurement using the dual channel single-sided spur calibration.
Figure 5-26 Single Channel SSB Cal
Figure 5-27 Dual Channel SSB Cal
N5511A Phase Noise Test System User’s Guide 91
Page 92
Absolute Measurement Fundamentals Calibration
Requirements:
— This calibration method requires a third source to generate a single sided
spur, in the case of a single channel measurement. For a dual channel measurement, a fourth source will be required.
— An external power combiner (or directional coupler) to add the calibration
spur to the frequency carrier under test. The calibration spur must have an amplitude -100 dB and -20 dB relative to the carrier amplitude. The offset frequency of the spur must be 20 Hz and 20 MHz.
— A spectrum analyzer or other means to measure the single sided spur
relative to the carrier signal
The equipment setup for this calibration option is similar to the others except that an additional source and a power splitter have been added so that the spur can be summed with the input carrier frequency.
Advantages
— Calibration is done under actual measurement conditions so all
non-linearities and harmonics of the phase detector are calibrated out.
Disadvantages
— Requires an extra RF sources that can be set between 10 Hz and up to
50 MHz (depending on the baseband analyzer used) from the carrier source frequency.
— Requires an RF spectrum analyzer for manual measurement of the
signal-to-spur ratio and the spur offset frequency.
For a single-sided spur user calibration, a spur is combined with the DUT carrier. The relative amplitude of the spur is set to a convenient level. An analyzer is typically used to measure the relative amplitude as well as offset of the spur. The dBc value and offset is entered in the Cal tab of the Define Measurement menu shown in Figure 5-28.
92 N5511A Phase Noise Test System User’s Guide
Page 93
Absolute Measurement Fundamentals Calibration
Figure 5-28 Define Measurement Menu - Cal: Known Spurs Parameters
N5511A Phase Noise Test System User’s Guide 93
Page 94
Absolute Measurement Fundamentals Calibration
The model of a single sided spur is a narrow-band PM signal summed with an amplitude modulated signal of the same depth. The summation constructively adds in voltage to the upper sideband in this case and destructively adds to the lower sideband. This has the effect of increasing the upper sideband by 6 dB and eliminating the lower sideband.
Figure 5-29 Single Sided Spur
When the phase modulation is detected in the phase detector, the AM component is rejected. This detection process reveals the lower sideband but reduces the upper sideband by 6 dB. The software takes this into account during the calibration process.
Figure 5-30 AM and Carrier Removal
As a result, the measured cal tone by the system will be 6 dB lower compared to the measured amplitude by the analyzer when setting the SSB tone. Figure
5-31 shows an example of spur set to -40 dBc at a 10 kHz offset, showing how
the system will detect a -46 dBc spur.
Figure 5-31 Measured Cal Tone
94 N5511A Phase Noise Test System User’s Guide
Page 95
Absolute Measurement Fundamentals Calibration

Double-sided Spur

The double-sided spur method, along with single-side spur is the most accurate user calibration. This calibration method conveniently needs no extra RF source, if the DUT is capable of being phase or amplitude modulated at the carrier frequency. When performing a double-sided spur user calibration, the calibration is done under actual measurement conditions so all non-linearities and harmonics of the phase detector are calibrated out. The offset frequency or modulation frequency must be between 10 Hz and the maximum (See the table from the “Measured Beat Note” section on page 87. The resultant sideband spurs from the phase modulation must have amplitudes that are
-100 dB and -20 dB relative to the carrier amplitude.
Advantages
— Calibration is done under actual measurement conditions so all
non-linearities and harmonics of the phase detector are calibrated out.
— No additional RF source is needed
Disadvantages
— Requires a phase modulator which operates at the desired carrier
frequency.
— Requires RF spectrum analyzer for manual measurement of
or preferably a modulation analyzer.
The double-sided spur user calibration method connection setup is the standard absolute measurement setup shown in Figure 5-1 on page 56 and in
Figure 5-3 on page 58. Figure 5-32 shows an example of a phase modulated
carrier with the upper side band measuring -40 dBc at a 10 kHz offset.
ΦM sidebands
N5511A Phase Noise Test System User’s Guide 95
Page 96
Absolute Measurement Fundamentals Calibration
Figure 5-32 Double-Sided Spur Method
When measuring phase noise, the carrier is modulated with a known PM depth. This value relative to the carrier as well as the modulating rate is entered in the software under the Cal tab.
In the example in Figure 5-33, the system measures a tone at -40 dBc
Figure 5-33 Known Spur Parameters
The system is phase locked and the modulating tone measured. In this case, the tone is a double-sideband PM tone so there is no AM to reject, so the entered value in the application is used to determine the amplitude of the carrier.
96 N5511A Phase Noise Test System User’s Guide
Page 97
Keysight N5511A Phase Noise Measurement System
User’s Guide

6 Absolute Measurement Examples

“Example Overviews” on page 98
“Input Ports” on page 99
“Single Channel Measurement” on page 100
“Dual Channel (EFC)” on page 117
“Dual Channel (DCFM)” on page 134
“OCXO Dual Channel Measurement” on page 150
97
Page 98
Absolute Measurement Examples Example Overviews

Example Overviews

This chapter contains 6 different measurement examples to show how to perform an absolute phase noise measurement using the N5511A Phase Noise Test System. The guide demonstrates single channel and dual channel absolute phase noise measurements. The following are the examples:
— Single Channel Measurement
— Dual Channel (using EFC)
— Dual Channel (using DCFM)
— Single channel measurement of an E8257D PSG UNY using beat note
user calibration method and EFC tuning mode.
— Dual channel measurement of an E8257D PSG UNY at 1 GHz using
beat note user calibration method and EFC tuning mode. The example features the use of two other E8257D PSG UNY as references.
— Dual channel measurement of an E8257D PSG UNY at 10 GHz using
beat note user calibration method and DCFM tuning mode. The example features the use of two other E8257D PSG UNY as references.
— OCXO Dual Channel Measurement
— Dual channel measurement of a 10 MHz OCXO using beat note user
calibration method and EFC tuning mode. Example features the two copies of DUT as references.
98 N5511A Phase Noise Test System User’s Guide
Page 99

Input Ports

Absolute Measurement Examples Input Ports
The N5511A Phase Noise Tests System phase detector frequency specification are the following.
Carrier frequency range
RF input power LO input power
Low Frequency Inputs (SMA) All Options
50 kHz to 3 GHz 1.2 GHz to 26.5 GHz 1.2 GHz to 40 GHz 50 kHz to 40 GHz
0 dBm to +23 dBm +15 dBm to +23 dBm
High Frequency Inputs (Type K) Option 526
0 dBm to +15 dBm +7 dBm to +15 dBm
For optimal performance, however, when performing a measurement, N5510 software defaults to detector selections that do not match the hardware capabilities. When performing a measurement, the software default detector selections use the following guidelines:
Low Frequency Inputs 50 kHz to 1.6 GHz
High Frequency Inputs 1.6 GHz and above
This can be overwritten by manually selecting the detector to be utilized in the measurement setup, as long as the input signal is within the hardware specifications of the detector input.
High Frequency Inputs (Type K) Option 540
0 dBm to +15 dBm +7 dBm to +15 dBm
AM Noise Input (SMA) All Options
0 dBm to +30 dBm N/A
N5511A Phase Noise Test System User’s Guide 99
Page 100
Absolute Measurement Examples Single Channel Measurement

Single Channel Measurement

A single channel measurement will be a very familiar experience on the N5511A PNTS for those that have used the E5500 system before. In a single channel measurement, the phase noise of the reference source will contribute to the overall phase noise of the measurement -therefore, it is very important to have a reference source that has much lower (better) phase noise than the phase noise of the DUT. The basic idea is we will use one PSG signal generator as a reference source and one PSG as the device under test (DUT).

Required equipment

— Two SMA 3.5 mm cables - ideal lengths: approximately 20 inch
—One BNC-to-BNC cable
— Two signal sources: one source acting as a DUT and the other source acting
as a reference (REF) with external frequency control (EFC) capability
— One Keysight N5511A (PNTS) Phase Noise Test Set
— One oscilloscope
— One SMB-BNC cable, or appropriate cable for connection from M9550A
monitor port to oscilloscope
Reference Source: This setup calls for a tunable reference source with the same center frequency as the DUT. In order for the noise measurement results to accurately represent the noise of the DUT, the noise level of the reference source should be below the expected noise level of the DUT.
100 N5511A Phase Noise Test System User’s Guide
Loading...