The A-models of the PNA family (N522x/3x/4xA) will be discontinued June 2019. Their B-model replacements are available now.
For more information about the B-models, please consult the PNA
“B” configuration guide or www.keysight.com/find/pna
This configuration guide describes standard configurations,
options, accessories, upgrade kits and compatible peripherals for
the PNA Family microwave network analyzers. This guide should
be used with the Keysight Technologies, Inc. PNA Family data
sheets for a complete description of these analyzers.
PNA-X Series
N5249A 10 MHz to 8.5 GHz
N5241A 10 MHz to 13.5 GHz
N5242A 10 MHz to 26.5 GHz
N5244A 10 MHz to 43.5 GHz
N5245A 10 MHz to 50 GHz
N5247A 10 MHz to 67 GHz
PNA Series
N5221A 10 MHz to 13.5 GHz
N5222A 10 MHz to 26.5 GHz
N5224A 10 MHz to 43.5 GHz
N5225A 10 MHz to 50 GHz
N5227A 10 MHz to 67 GHz
PNA-L Series
N5239A 300 kHz to 8.5 GHz
N5231A 300 kHz to 13.5 GHz
N5232A 300 kHz to 20 GHz
N5234A 10 MHz to 43.5 GHz
N5235A 10 MHz to 50 GHz
Page 3Find us at www.keysight.com
Keysight offers the following options for all PNA Family network analyzers
Certification options
Commercial calibration certification with test data
(Option UK6)
Complete set of measurements which tests unit to manufacturer’s
published specifications. Includes calibration label, calibration
certificate, and data report. Conforms to ISO 9001.
ISO 17025 compliant calibration (Option 1A7)
Complete set of measurements which tests unit to manufacturer’s
published specifications. Includes calibration label, ISO 17025
calibration certificate, and data report, measurement uncertainties and guardbands on all customer specifications. Conforms to
ISO 17025 and ISO 9001.
ANSI Z540 compliant calibration (Option A6J)
Complete set of measurements which tests unit to manufacturer’s
published specifications. Includes pre- and post-adjustment
data with measurement uncertainty information compliant to the
ANSI/NCSL Z540 standard.
Documentation
The PNA Series instruments are equipped with an Online Help
system available within the instrument in English only. All PNA
documentation is available on the web:
www.keysight.com/find/pna
Calibration Software Licenses
Perpetual license for built-in performance test
software for Keysight inclusive cal (Option 897)
Adds built-in performance testing and calibration software
for self-maintainers. Requires additional equipment. See the
analyzer’s Service Guide for more information on equipment
required.
Perpetual license for built-in performance
test software for Standards compliant cal
(Option 898)
Adds built-in performance testing and calibration software
for self-maintainers. Requires additional equipment. See the
analyzer’s Service Guide for more information on equipment
required.
Page 4Find us at www.keysight.com
PNA Family Network Analyzer Configurations
PNA-X Series
N5249A 10 MHz to 8.5 GHz
N5241A 10 MHz to 13.5 GHz
N5242A 10 MHz to 26.5 GHz
N5244A 10 MHz to 43.5 GHz
N5245A 10 MHz to 50 GHz
PNA Series
N5221A 10 MHz to 13.5 GHz
N5222A 10 MHz to 26.5 GHz
N5224A 10 MHz to 43.5 GHz
N5225A 10 MHz to 50 GHz
N5227A 10 MHz to 67 GHz
PNA-L Series
N5239A 300 kHz to 8.5 GHz
N5231A 300 kHz to 13.5 GHz
N5232A 300 kHz to 20 GHz
N5234A 10 MHz to 43.5 GHz
N5235A 10 MHz to 50 GHz
N5247A 10 MHz to 67 GHz
Test set and power configuration options
Choose one test set and power configuration option. Option 2xx indicates two test ports. Option 4xx indicates four test ports. To add options to a product,
order the corresponding item number (N52xxA-xxx).
Internal
Description
PNA-X Series
Option 200
Option 200, H85, 285
Option 200, 219
Option 200, H85, 285, 224
Option 200, 219, 224
Option 400
Option 400, H85, 485
Option 400, 419
Option 400, H85, 485, 423
Option 400, 419, 423
1
1
Configurable
Tes t Set
1
1
Source
Attenuators
Receiver
Attenuators Bias-tees
Second
Source
Internal
Combiner
Mechanical
SwitchesAdditional Information
Not available on N5249A or N5247A
Requires Option 080
Not available on N5249A or N5247A
Requires Option 080
Option 080 recommended
Option 080 recommended
Not available on N5249A or N5247A
Option 080 recommended
Requires Option 080
Not available on N5249A or N5247A
Requires Option 080
PNA Series
Option 200
Option 201
Option 210
Option 217
Option 219
Option 400
Option 401
Option 410
Option 417
2-port metrology option
Not available on 67 GHz model
Option 080 recommended
Option 080 recommended
4-port metrology option
Option 080 recommended
Not available on 67 GHz model
Option 419
Option 080 recommended
PNA-L Series
Option 200
Option 216
Option 400Available on only N5231A/2A
Option 416
1. Order special model N524x AS instead of N524x A. Order N524xA-xxx items for other standard options. Option H85 requires Option 285 or 485 which includes
2. Adds one source at tenuator to be shared with all test ports.
2
the extended power range of Option 219 or 419; therefore these options cannot be combined.
Available on only N5231A/2A
Page 5Find us at www.keysight.com
PNA Family Network Analyzer Configurations (continued)
PNA-X Series
N5249A 10 MHz to 8.5 GHz
N5241A 10 MHz to 13.5 GHz
N5242A 10 MHz to 26.5 GHz
N5244A 10 MHz to 43.5 GHz
N5245A 10 MHz to 50 GHz
PNA Series
N5221A 10 MHz to 13.5 GHz
N5222A 10 MHz to 26.5 GHz
N5224A 10 MHz to 43.5 GHz
N5225A 10 MHz to 50 GHz
N5227A 10 MHz to 67 GHz
PNA-L Series
N5239A 300 kHz to 8.5 GHz
N5231A 300 kHz to 13.5 GHz
N5232A 300 kHz to 20 GHz
N5234A 10 MHz to 43.5 GHz
N5235A 10 MHz to 50 GHz
N5247A 10 MHz to 67 GHz
Application options
To add options to a product, order the corresponding item number (N52xxA-xxx).
For PNA-X
Measurement application
Automatic fixture removalN524xA-007 N522xA-007 N523xA-007 Requires Windows 7 OS
Vector- and scalar-calibrated converter
measurements
Embedded LO measurementsN524x A-084 N522xA-084n/aRequires Option 082 or Option 083. Also works with Options 028, 029, 086
Gain compression applicationN524x A-086 N522xA-086n/aFor measuring frequency converters, requires Option 082 or 083
Intermodulation distortion
application
Source phase controlN524x A-088 N522xA-088n/aNot available with N522xA-200 and -210
Differential and I/Q devices applicationN524x A-089 N522xA-089n/aRequires Option 080 and 400, 401, 410, 417, or 419, and Windows 7 OS
Spectrum analyzerN524x A-090 N522xA-090 N523xA-090 Requires Option 080, Windows 7 OS, and DSP version 58. Test set and power
Extended spectrum analyzer to 110 GHz
Extended spectrum analyzer above 110 GHz9N524x A-094 N522xA-094n/aRequires Option 090. Not available with N5241/49A or N5221A
Integrated true-mode stimulus applicationN524x A-460 N522xA-460n/aRequires one of Options 400, 401, 410, 417, 419 or 423
N-port capabilities
1. For N522xA and N5241/42/49A, vector-noise-corrected measurements require an ECal for use as an impedance tuner. For N5244/45/47A with Option 029,
an internal tuner is included. For calibration, Option 028 requires a power meter, and Option 029 requires either a power meter or a 346-series noise source
(Keysight 346C or 346C-K01 recommended). All options require a power meter for measuring mixers and converters.
2. Option 080 is required to configure an external source using Ex ternal Device Configuration dialog.
3. Option 082 is a subset of Option 083; therefore, they cannot be ordered together.
4. A configurable test set is required for VMC measurements (to connect a reference mixer) or for SMC+Phase measurements using the comb-generator-based
calibration. When ordered with N522x A-200, -210, -4 00 and -410, Option 083 adds phase and delay measurements only by using SMC+Phase with a
calibration mixer.
5. Option 087 can be ordered without N524x A-224 or N524xA-423, but may require external equipment such as a signal generator and a combiner. Refer to
page 13 for more details.
6. When ordering a test set, select an appropriate interface kit. Refer to page 22 Multiport Measurements section for more details.
7. When configuring N524xA as a multiport analyzer using Option 551 and a multipor t test set, the combiner feature of Option 224 or 423 is temporarily
disabled.
When configuring N524x A as a standalone analyzer, the combiner feature is enabled.
8. The Windows 7 upgrade kit for PNAs containing CPU version 6 is model N8983A. This upgrade kit contains a new solid state drive with all of the necessary
software installed. For units with older CPUs, order the N8984A with Option PC7. This upgrade kit contains a new solid state drive and a version 7 CPU. For
adding Option 090 to units with DSP version 4, add Option SP5 to the N8983A or N8984A order to get the DSP version 5.
9. Option 093 is a subset of 094; therefore they cannot be ordered together.
1
2
3, 4
5
9
6
Series
N524xA- 015 N522 xA-0 15 N5 23xA-0 15 Displays the measurement uncertainty dynamically (real-time) on the
N524x A-028 N522xA-028n/aRequires Option 080. For measuring frequency converters, requires Option
N524x A-029n/an/aRequires Option 080 and for N5241/42/49A, one of Options 219, 224, 419,
N524x A-080 N522xA-080 N523xA-080
N524x A-083 N522xA-083n/aRequires Option 080
N524x A-087 N522xA-087n/aRequires Option 080 For measuring frequency converters, requires Option
N524xA-093 N522xA-093n/aRequires Option 090. Not available with N5241/49A or N5221A
N524xA-5517N522xA-551N523xA-551 Not available with N522xA-200, -210, -400 and -410, and N523xA-200 and -400
For PNA
Series
For PNA-L
SeriesAdditional Information
8
measurement trace. Not available with N5231A,32A and 39A
082 or 083. Not available with N522x A-200, -210, -400 and -410
423 or H85. For N5244/45/47A, requires Option 224 or 423. On N5247A, noise
receivers work up to 50 GHz only. For measuring frequency converters, requires
Option 082 or 083
and 087
082 or 083. Not available with N522x A-200, -210, -400 and -410
8
configuration option with receiver attenuators is recommended
Page 6Find us at www.keysight.com
PNA Family Network Analyzer Configurations (continued)
PNA-X Series
N5249A 10 MHz to 8.5 GHz
N5241A 10 MHz to 13.5 GHz
N5242A 10 MHz to 26.5 GHz
N5244A 10 MHz to 43.5 GHz
N5245A 10 MHz to 50 GHz
PNA Series
N5221A 10 MHz to 13.5 GHz
N5222A 10 MHz to 26.5 GHz
N5224A 10 MHz to 43.5 GHz
N5225A 10 MHz to 50 GHz
N5227A 10 MHz to 67 GHz
PNA-L Series
N5239A 300 kHz to 8.5 GHz
N5231A 300 kHz to 13.5 GHz
N5232A 300 kHz to 20 GHz
N5234A 10 MHz to 43.5 GHz
N5235A 10 MHz to 50 GHz
N5247A 10 MHz to 67 GHz
Application options (continued)
To add options to a product, order the corresponding item number (N52xxA-xxx).
DescriptionFor PNA-X Series For PNA Series For PNA-L Series Additional Information
1. To configure NVNA, requires two comb generators with power supplies, Keysight
calibration kits (mechanical or ECal), power meter and sensor or USB power sensor.
Requires EXG, MXG or PSG for X-parameter extraction (PNA-X 10 MHz reference
output can be used for 10 MHz tone-spacing applications).
2. X-parameters is a trademark and registered trademark of Keysight Technologies in
the US, EU, JP, and elsewhere. The X-parameters format and underlying equations
1
3
3
U9391C/F/G n/an/a
N524xA-897N522xA-897N523xA-897
N524xA-898N522xA-898N523xA-898
are open and documented. For more information, visit http://www.Keysight.com/
find/eesof-x-parameters-info.
3. Additional hardware required. Please refer to the analyzer’s Service Guide for required
service test equipment.
4. Pulse capability requires option 021 and 025
5. Pulse capability requires option 021, 022 and 025
Page 7Find us at www.keysight.com
PNA-X Series test set and power configuration options
Test port 1
R1
Test port 2
R2
AB
Source 1
OUT 1OUT 2
Pulse
modulator
R A B C D
IF inputs
Rear panel
To receivers
LO
Pulse generators
Test port 1
R1
Test port 2
R2
35 dB
65 dB
35
dB
65 dB
AB
Source 1
OUT 1OUT 2
Pulse
modulator
R A B C D
IF inputs
Rear panel
To receivers
LO
Pulse generators
R1
Test port 2
R2
A
B
To receivers
LO
Source 2
Output 1
Source 2
Output 2
Pulse generators
Rear panel
1
2
3
4
Test port 1
Source 1
OUT 1OUT 2
Pulse
modulator
Source 2
OUT 1OUT 2
Pulse
modulator
J9J10J11J8J7J2 J1
35 dB
65 dB
35 dB
65 dB
J6 J5
RF
OUT
LO
OUT
R R1 R2A B
IF inputs
Test port 1
R1
Test port 2
R2
35 dB
65 dB
35
dB
65 dB
AB
Source 1
OUT 1OUT 2
Pulse
modulator
R A B C D
IF inputs
Rear panel
To receivers
LO
Pulse generators
1
The PNA-X is an integrated vector network analyzer featuring
a built-in S-parameter test set, one or two synthesized sources
used for device stimulus, a solid-state drive, USB interfaces,
and a 10.4” LCD touch screen display. The N5241A, N5242A, and
N5249A have 50 ohm, ruggedized 3.5 mm (m) test ports. The
N5244A and the N5245A have 50 ohm, ruggedized 2.4 mm (m)
test ports. The N5247A has 50 ohm, ruggedized 1.85 mm (m)
test ports. Included with each instrument is a mouse, keyboard
(U.S. style), and one day of on-site productivity assistance
(PS-S20-PNA).
A test set and power configuration option is mandatory, choose
one of the following:
2-port standard test set and power range
(Option 200)
The standard 2-port test set comes with six front-panel access
loops. The loops provide access to the signal path between (a)
the source output and the reference receiver, (b) the source
output and directional coupler thru arm and (c) the coupled arm
of the directional coupler and the port receiver. The standard test
set also includes a solid-state internal RF transfer switch in the
R1 reference-receiver path.
High-power configurable and extended power
range (N524xAS Option H85/285)
With extended power range and bias-tees (Option 219), the
internal bias-tees limit the maximum test port input power to
+30 dBm. The N524xAS Option H85/285 removes the bias-tees
between the source attenuators and the test port couplers. This
extends the maximum port power that the analyzer can safely
handle to +43 dBm. Selecting OptionH85/285 adds internal attenuators and cables. The N524xAS Option H85/285 and Option
219 can not be ordered together. Option H85/285 is not available
for the N5249A or the N5247A. When ordering, select N524xAS
then add items N524xAS-H85, N524xAS-285 and N524xA-200.
Extended power range and bias-tees (Option 219)
Adds to the standard test set one 65 dB for N5241/42/49A,
60 dB for N5244/45A, and 50 dB for N5247A source attenuator
(set table in 5 dB increments for N5241/42/49A and 10 dB increments for N5244/45/47A), one 35 dB receiver attenuator (settable
in 5 dB increments) for N5241/42/44/45/49A and 50 dB receiver
attenuator (settable in 10 dB increments) for N5247A, and one
bias-tee to each test port. Option 219 requires Option 200.
Add an internal second source, a combiner
and mechanical switches to 2-port analyzer
(Option 224)
Available with 2-port model only, this option adds an internal
second source, a combiner and mechanical switches. The internal second source provides an additional signal (fixed or swept)
for two-tone third-order-intercept (TOI) and intermodulation
testing of amplifiers, or it can be used as a fast swept-LO signal
for fixed-IF testing of mixers and converters. The mechanical
switches provide increased flexibility by having rear panel access
(front panel access for N5247A Source 1 and 2 Out 1) to signal
paths for advanced applications. Access to the second source
is available through two output connectors on the front panel.
Requires Options 200, 219 and 080.
1. The block diagrams shown above include hardware that must be ordered
as separate options, such as pulse generators (Option 025), pulse modulators (Options 021 and 022), and IF access (Option 020). In addition, the
combiner type and at tenuator values vary by model number. Refer to the
product data sheet for the correct block diagram for a specific model.
Page 8Find us at www.keysight.com
PNA-X Series test set and power configuration options
Test port 3
C
R3
Test port 1
R1
Test port 4
R4
Test port 2
R2
A
D
B
To receivers
LO
Pulse generators
Rear panel
1
2
3
4
Source 1
OUT 1OUT 2
Pulse
modulator
Source 2
(standard)
OUT 1OUT 2
Pulse
modulator
J9J10J11J8J7J2 J1J4 J3
35 dB35 dB35 dB
35 dB
65 dB65 dB65 dB65 dB
J6 J5
RF
OUT
LO
OUT
R A B C D
IF inputs
Test port 3
C
R3
Test port 1
R1
Test port 4
R4
Test port 2
R2
ADB
Source 2
OUT 1OUT 2
Pulse
modulator
Source 1
OUT 1OUT 2
Pulse
modulator
Pulse generators
To receivers
LO
R A B C D
IF inputs
Rear panel
Test port 3
C
R3
Test port 1
R1
Test port 4
R4
Test port 2
R2
35 dB
65 dB
35 dB
65 dB65 dB
35 dB
65 dB
A
35 dB
DB
Source 2
OUT 1OUT 2
Pulse
modulator
Source 1
OUT 1OUT 2
Pulse
modulator
Pulse generators
To receivers
LO
R A B C D
IF inputs
Rear panel
C
R3
R1
R4R2
35 dB
65 dB
35 dB
65 dB65 dB
35dB
65 dB
A
35 dB
DB
OUT 1OUT 2
Pulse
modulator
OUT 1OUT 2
Pulse
modulator
To receivers
LO
R A BCD
IFinputs
Rear panel
Source 1
Source 2
Test port 1
Test port 3
Test port 2
Pulse generators
Test port 4
1
(continued)
4-port standard test set, power range and an
internal second source (Option 400)
The standard 4-port test set comes with 12 front-panel access
loops and a built-in second source. The loops provide access to
the signal path between (a) the source output and the reference
receiver, (b) the source output and directional coupler thru arm
and (c) the coupled arm of the directional coupler and the port
receiver. The internal second source provides an additional signal
(fixed or swept) for two-tone third-order-intercept (TOI) and
intermodulation testing of amplifiers, or it can be used as a fast
swept-LO signal for fixed-IF testing of mixers and converters.
With two sources, source 1 is accessible through test ports 1
and 2, and source 2 is accessible through test ports 3 and 4.
The standard test set also includes a solid-state internal RF
transfer switch in the R1 reference-receiver path. Option 080 is
recommended.
Extended power range and bias-tees
(Option 419)
Adds to the standard test set one 65 dB for N5241/42/49A, 60 dB
for N5244/45A, and 50 dB for N5247A source attenuator (settable in 5 dB increments for N5241/42/49A and 10 dB increments
for N5244/45/47A), one 35 dB receiver attenuator (settable in
5 dB increments) for N5241/42/44/45/49A and 50 dB receiver
attenuator (settable in 10 dB increments) for N5247A, and one
bias-tee to each test port. Option 419 requires Option 400.
Add an internal combiner and mechanical
switches to 4-port analyzer (Option 423)
High-power configurable and extended power
range (N524xAS Option H85/485)
With extended power range and bias-tees (Option 419), the
internal bias-tees limit the maximum test port input power to
+30 dBm. The N524xAS Option H85/485 removes the bias-tees
between the source attenuators and the test port couplers. This
extends the maximum port power that the analyzer can safely
handle to +43 dBm. Selecting Option H85/485 adds internal attenuators and cables. The N524xAS Option H85/485 and Option
419 can not be ordered together. Option H85/485 is not available
for the N5249A or the N5247A. When ordering, select N524xAS
then add items N524xAS-H85, N524xAS-485 and N524xA-400.
Available with 4-port model only, this option adds a combiner
and mechanical switches. The mechanical switches provide
increased flexibility by having rear panel access (front panel
access for the N5247A Source 1 and 2 Out 1) to signal paths for
advanced applications. Requires Options 400, 419 and 080.
1. The block diagrams shown above include hardware that must be ordered
as separate options, such as pulse generators (Option 025), pulse modulators (Options 021 and 022), and IF access (Option 020). In addition, the
combiner type and at tenuator values vary by model number. Refer to the
product data sheet for the correct block diagram for a specific model.
Page 9Find us at www.keysight.com
PNA Series test set and power configuration options
Test port 1Test port 2
AB
R1
R2
Source
OUT 1OUT 2
Pulse
modulator
To receivers
LO
Pulse generators
RABCD
IF inputs
Rear panel
R1
R2
35 dB
65 dB65 dB
A
35 dB
B
Source
OUT 1OUT 2
Pulse
modulator
To receivers
LO
Pulse generators
RABCD
IF inputs
Rear panel
Test port 1
R1
Test port 2
R2
AB
Source
OUT 1OUT 2
Pulse
modulator
To receivers
LO
Pulse generators
RABCD
IF inputs
Rear panel
Test port 1
R1
Test port 2
R2
35 dB
65 dB65 dB
A
35 dB
B
Source
OUT 1OUT 2
Pulse
modulator
To receivers
LO
Pulse generators
RABCD
IF inputs
Rear panel
1
The PNA is an integrated vector network analyzer featuring a
built-in S-parameter test set, one or two synthesized sources
used for device stimulus, a solid-state drive, USB interfaces, and
a 10.4” LCD touch screen display. The N5221A and the N5222A
have 50 ohm, ruggedized 3.5 mm (m) test ports. The N5224A
and the N5225A have 50 ohm, ruggedized 2.4 mm (m) test ports.
The N5227A has a 50 ohm, ruggedized 1.85 mm (m) test ports.
Included with each instrument is a mouse, keyboard (U.S. style),
and one day of on-site productivity assistance (PS-S20-PNA).
A test set and power configuration option is mandatory, choose
one of the following:
2-ports, single source, base configuration (Option
200/210)
The 2-port with base configuration has no front-panel access
loops.
2-ports, single source, with extended power range
(Option 217)
The 2-port test set with extended power range comes with a
configurable test set, source and receiver attenuators at each
port. For N5221/22A, the source attenuators are 65 dB in 5 dB
steps and the receiver attenuators are 35 dB in 5 dB steps. For
N5224/25A, the source attenuators are 60 dB in 10 dB steps and
the receiver attenuators are 35 dB in 5 dB steps. Also included
is a solid-state internal RF transfer switch in the R1 referencereceiver path. This configuration is not available on N5227A.
2-ports, single source, with extended power range
and bias-tee (Option 219)
2-ports, single source, with configurable test set
(Option 201)
The 2-port configurable test set comes with six front-panel access loops. The loops provide access to the signal path between
(a) the source output and the reference receiver, (b) the source
output and directional coupler thru arm and (c) the coupled arm
of the directional coupler and the port receiver at all ports. Also
included is a solid-state internal RF transfer switch in the R1
reference-receiver path.
The 2-port test set with extended power range and bias-tee
comes with a configurable test set, source and receiver attenuators, and bias-tee at each port. For N5221/22A, the source
attenuators are 65 dB in 5 dB steps and the receiver attenuators
are 35 dB in 5 dB steps. For N5224/25A, the source attenuators
are 60 dB in 10 dB steps and the receiver attenuators are 35 dB
in 5 dB steps. For N5227A, both source and receiver attenuators
are 50 dB in 10 dB steps. Also included is a solid-state internal
RF transfer switch in the R1 reference-receiver path.
1. The block diagrams shown above include hardware that must be ordered
as separate options, such as pulse generators (Option 025), pulse modulators (Options 021 and 022), and IF access (Option 020). In addition, the
attenuator values vary by model number. Refer to the product data sheet
for the correct block diagram for a specific model.
Page 10Find us at www.keysight.com
PNA Series test set and power configuration options
Test port 3
R3
Test port 1Test port 4
R4
Test port 2
R2
R1
ACDB
Source 2
OUT 1OUT 2
Pulse
modulator
Source 1
OUT 1OUT 2
Pulse
modulator
Pulse generators
To receivers
LO
RABCD
IF inputs
Rear panel
Test port 3
C
R3
Test port 1
R1
Test port 4
R4
Test port 2
R2
35 dB
65 dB
35 dB
65 dB65 dB
35
dB
65 dB
A
35 dB
DB
Source 2
OUT 1OUT 2
Pulse
modulator
Source 1
OUT 1OUT 2
Pulse
modulator
Pulse generators
To receivers
LO
RABCD
IF inputs
Rear panel
Test port 3
C
R3
Test port 1
R1
Test port 4
R4
Test port 2
R2
ADB
Source 2
OUT 1OUT 2
Pulse
modulator
Source 1
OUT 1OUT 2
Pulse
modulator
Pulse generators
To receivers
LO
RABCD
IF inputs
Rear panel
Test port 3
C
R3
Test port 1
R1
Test port 4
R4
Test port 2
R2
35 dB
65 dB
35 dB
65 dB65 dB
35
dB
65 dB
A
35 dB
DB
Source 2
OUT 1OUT 2
Pulse
modulator
Source 1
OUT 1OUT 2
Pulse
modulator
Pulse generators
To receivers
LO
RABCD
IF inputs
Rear panel
1
(continued)
4-ports, dual source, base configuration
(Option 400/410)
The 4-port with base configuration has no front-panel access
loops.
4-ports, dual source, with configurable test set
(Option 401)
The 4-port configurable test set comes with two internal
sources, twelve front-panel access loops. The loops provide
access to the signal path between (a) the source output and the
reference receiver, (b) the source output and directional coupler
thru arm and (c) the coupled arm of the directional coupler
and the port receiver at all ports. Also included is a solid-state
internal RF transfer switch in the R1 reference-receiver path.
4-ports, dual source, with extended power range
(Option 417)
The 4-port test set with extended power range comes with
two internal sources, configurable test set, source and receiver
attenuators at each port. For N5221/22A, the source attenuators
are 65 dB in 5 dB steps and the receiver attenuators are 35 dB in
5 dB steps. For N5224/25A, the source attenuators are 60 dB in
10 dB steps and the receiver attenuators are 35 dB in 5 dB steps.
Also included is a solid-state internal RF transfer switch in the
R1 reference-receiver path. This configuration is not available on
N5 2 2 7A .
4-ports dual source, with extended power range
and bias-tee (Option 419)
The 4-port test set with extended power range and bias-tee
comes with two internal sources, configurable test set, source
and receiver attenuators, and bias-tee at each port. For
N5221/22A, the source attenuators are 65 dB in 5 dB steps and
the receiver attenuators are 35 dB in 5 dB steps. For N5224/25A,
the source attenuators are 60 dB in 10 dB steps and the receiver
attenuators are 35 dB in 5 dB steps. For N5227A, both source
and receiver attenuators are 50 dB in 10 dB steps. Also included
is a solid-state internal RF transfer switch in the R1 referencereceiver path.
1. The block diagrams shown above include hardware that must be ordered
as separate options, such as pulse generators (Option 025), pulse modulators (Options 021 and 022), and IF access (Option 020). In addition, the
attenuator values vary by model number. Refer to the product data sheet
for the correct block diagram for a specific model.
Page 11Find us at www.keysight.com