No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Manual Part Number
N5180-90056
Edition
Edition 1, August 2018
Supersedes: August 2016
Published by:
Keysight Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
U.S. Government Rights
The Software is “commercial
computer software,” as defined
by Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202,
the U.S. government acquires
commercial computer software
under the same terms by which
the software is customarily
provided to the public.
Accordingly, Keysight provides
the Software to U.S. government
customers under its standard
commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of
which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA
represents the exclusive authority
by which the U.S. government
may use, modify, distribute, or
disclose the Software. The EULA
and the license set forth therein,
does not require or permit,
among other things, that
Keysight: (1) Furnish technical
information related to
commercial computer software
or commercial computer
software documentation that is
not customarily provided to the
public; or (2) Relinquish to, or
otherwise provide, the
government rights in excess of
these rights customarily provided
to the public to use, modify,
reproduce, release, perform,
display, or disclose commercial
computer software or
commercial computer software
documentation. No additional
government requirements
beyond those set forth in the
EULA shall apply, except to the
extent that those terms, rights, or
licenses are explicitly required
from all providers of commercial
computer software pursuant to
the FAR and the DFARS and are
set forth specifically in writing
elsewhere in the EULA. Keysight
shall be under no obligation to
update, revise or otherwise
modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant
to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S.
government acquires no greater
than Limited Rights as defined in
FAR 27.401 or DFAR 227.7103-5
(c), as applicable in any technical
data.
Safety Notices
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION
notice until the indicated
conditions are fully understood
and met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in personal injury or death.
Do not proceed beyond a
WARNING notice until the
indicated conditions are fully
understood and met.
Where to Find the Latest Information
Documentation is updated periodically. For the latest information about these products, including instrument software
upgrades, application information, and product information, browse to one of the following URLs, according to the name
of your product:
http://www.keysight.com/find/X-Series_SG
To receive the latest updates by email, subscribe to Keysight Email Updates at the following URL:
http://www.keysight.com/find/MyKeysight
Information on preventing instrument damage can be found at:
www.keysight.com/find/PreventingInstrumentRepair
Is your product software up-to-date?
Periodically, Keysight releases software updates to fix known defects and incorporate product enhancements. To search
for software updates for your product, go to the Keysight Technical Support website at:
—Real-Time Fading (Option 660)
—Custom Digital Modulation (
Option 431)
—Multitone and Two-Tone Waveforms (Option 430)
—Troubleshooting
—Working in a Secure Environment
Programming Guide
SCPI Reference
—Getting Started with Remote Operation
—Using IO Interfaces
—Programming Examples
—Programming the Status Register System
—Creating and Downloading Files
—Creating and Downloading User–Data Files
—SCPI Basics
—Basic Function Commands
—System Commands
—Analog Modulation Commands
—Arb Commands
—Avionics VOR/ILS Commands
—Bit Error Rate Test (BERT) Commands
—Digital Signal Interface Module Commands
—Real-Time Commands
xv
Programming
Compatibility Guide
—Provides a listing of SCPI commands and programming codes for
signal generator models that are supported by the Keysight EXG
and MXG X- Series signal generators.
Service Guide
—Troubleshooting
—Replaceable Parts
—Assembly Replacement
—Post–Repair Procedures
—Safety and Regulatory Information
—Instrument History
Key Help
a
—Key function description
—Related SCPI commands
a. Press the Help key, and then the key for which you wish help.
xvi
N5171B/72B/73B EXG and N5181B/82B/83B MXG
1Signal Generator Overview
To avoid damaging or degrading the performance of the instrument, do not exceed 33 dBm (2W)
maximum (27 dBm (0.5W) for N5173N/83B) of reverse power levels at the RF input. See also Tips
for Preventing Signal Generator Damage on www.keysight.com.
For more details on hardware, firmware, software, and documentation features and options, refer to
the data sheet shipped with the signal generator and available from the Keysight Technologies
website at http://www.keysight.com/find/X-Series_SG.
Modes of Operation
Depending on the model and installed options, the Keysight X-Series signal generator provides up
to four basic modes of operation: continuous wave (CW), swept signal, analog modulation, and
digital modulation.
Continuous Wave
In this mode, the signal generator produces a continuous wave signal. The signal generator is set to
a single frequency and power level. Both the analog and vector models can produce a CW signal.
Swept Signal
In this mode, the signal generator sweeps over a range of frequencies and/or power levels. Both the
analog and vector models provide list and step sweep functionality.
Analog Modulation
In this mode, the signal generator modulates a CW signal with an analog signal. The analog
modulation types available depend on the installed options.
Option UNT provides AM, FM, and ΦM modulations. Some of these modulations can be used
together.
Option 303 provides a multifunction generator that consists of seven waveform generators.
Option UNW provides standard and narrow pulse modulation capability.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 3
Signal Generator Overview
Modes of Operation
Digital Modulation (Vector Models with Option 65x Only)
In this mode, the signal generator modulates a CW signal with an arbitrary I/Q waveform. I/Q
modulation is only available on vector models. An internal baseband generator (Option 65x) adds
the following digital modulation formats:
— Custom Arb Waveform Generator mode can produce a single–modulated carrier or multiple–
modulated carriers. Each modulated carrier waveform must be calculated and generated before
it can be output; this signal generation occurs on the internal baseband generator. Once a
waveform has been created, it can be stored and recalled, which enables repeatable playback of
test signals. To learn more, refer to “Using the Arbitrary Waveform Generator” on page 379.
— Custom Real-Time Waveform Generator mode can produce a single–modulated carrier or
multiple–modulated carriers. Each modulated carrier waveform must be calculated and
generated before it can be output; this signal generation occurs on the internal baseband
generator. Once a waveform has been created, it can be stored and recalled, which enables
repeatable playback of test signals. To learn more, refer to “Using the Arbitrary Waveform
Generator” on page 379.
— Multitone mode produces up to 64 continuous wave signals (or tones). Like the Two Tone mode,
the frequency spacing between the signals and the amplitudes are adjustable. To learn more,
refer to “Creating a Custom Multitone Waveform” on page 412.
— Two–tone mode produces two separate continuous wave signals (or tones). The frequency
spacing between the signals and the amplitudes are adjustable. To learn more, refer to Chapter
17, “Multitone and Two-Tone Waveforms (Option 430).”.
— Dual ARB mode is used to control the playback sequence of waveform segments that have been
written into the ARB memory located on the internal baseband generator. These waveforms can
be generated by the internal baseband generator using the Custom Arb Waveform Generator
mode, or downloaded through a remote interface into the ARB memory. To learn more, refer to
“Dual ARB Player” on page 187.
4 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Front Panel Overview
22. Power Switch and LEDs
21. More and LED
20. Return
18. Knob
19. Incr Set
15. Page
Down
16. I Input
(vector
models only)
17. Q Input (vector
models only)
12. RF Output
(N5171B,
N5172B,
N5181B,
N5182B)
13. RF On/Off and LED
14. Mod On/Off and LED
7. MENUS
11. Preset and
User Preset
9. Local
Cancel/(Esc)
5. Arrows and Select
6. Page Up
8. Trigger
4. Numeric
Keypad
3. Softkeys
2. Display
1. Host USB
10. Help
Signal Generator Overview
Front Panel Overview
1. Host USB
ConnectorType A
USB Protocol2.0
Use this universal serial bus (USB) to connect a USB Flash Drive (UFD) for data transfer. You can
connect or disconnect a USB device without shutting down or restarting the signal generator. The
instrument also has a rear panel device USB connector (see page 14) used to remotely control the
instrument.
2. Display
The LCD screen provides information on the current function. Information can include status
indicators, frequency and amplitude settings, and error messages. Labels for the softkeys are
located on the right hand side of the display. See also, “Front Panel Display” on page 9.
3. Softkeys
A softkey activates the function indicated by the displayed label to the left of the key.
4. Numeric Keypad
The numeric keypad comprises the 0 through 9 hardkeys, a decimal point hardkey, a minus sign
hardkey, and a backspace hardkey. See “Entering and Editing Numbers and Text” on page 41.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 5
Signal Generator Overview
See page62
See page56
See page43
See page159
See page169
Active only on
vector models.
See page43
See page46
See page25
See page165
See
page185
See page68
See page70
Front Panel Overview
5. Arrows and Select
The Select and arrow hardkeys enable you to select items on the signal generator’s display for
editing. See “Entering and Editing Numbers and Text” on page 41.
6. Page Up
In a table editor, use this hardkey to display a previous page. See “Example: Using a Table Editor”
on page 42. When text does not fit on one page in the display area, use this key in conjunction with
the PageDown key (page 7) to scroll text.
7. MENUS
These hardkeys open softkey menus that enable you to configure instrument functions or access
information.
8. Trigger
When trigger mode is set to Trigger Key, this hardkey initiates an immediate trigger event for a
function such as a list or step sweep.
9. Local Cancel/(Esc)
This hardkey deactivates remote operation and returns the signal generator to front panel control,
cancels an active function entry, and cancels long operations (such an IQ calibration).
10. Help
Use this key to display a description of any hardkey or softkey. See “Viewing Key Descriptions” on
page 40.
6 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Signal Generator Overview
Front Panel Overview
11. Preset and User Preset
These hardkeys set the signal generator to a known state (factory or user–defined). See “Presetting
the Signal Generator” on page 40.
12. RF Output (N5171B, N5172B, N5181B, N5182B)
ConnectorStandard:
Option 1EM:
Impedance:
Damage Levels50 Vdc, 2 W maximum RF power
female Type–N
Rear panel output
50 Ω
12. RF Output (N5173B, N5183B)
ConnectorOption 513/520:
Option 532/540:
Option 1ED Option
1EM:
Impedance:
Damage Levels0 Vdc, 0.5 W maximum RF power
male Precision APC-3.5
male Precision 2.4 mm
female Type–N
Rear panel output
50 Ω
13. RF On/Off and LED
This hardkey toggles the operating state of the RF signal present at the RF OUTPUT connector. The
RF On/Off LED lights when RF output is enabled.
14. Mod On/Off and LED
This hardkey enables or disables the modulation of the output carrier signal by an active
modulation format. This hardkey does not set up or activate a format (see “Modulating the Carrier
Signal” on page 54).
The MOD ON/OFF LED lights when modulation of the output is enabled.
15. Page Down
In a table editor, use this hardkey to display the next page. See “Example: Using a Table Editor” on
page 42. When text does not fit on one page in the display area, use this key in conjunction with the
PageUp key (page 6) to scroll text.
Rotating the knob increases or decreases a numeric value, or moves the highlight to the next digit,
character, or item in a list. See also, “Front Panel Knob Resolution” on page 28.
19. Incr Set
This hardkey enables you to set the increment value of the currently active function. The increment
value also affects how much each turn of the knob changes an active function’s value, according to
the knob’s current ratio setting (see “Front Panel Knob Resolution” on page 28).
20. Return
This hardkey enables you to retrace key presses. In a menu with more than one level, the Return key
returns to the prior menu page.
21. More and LED
When a menu contains more softkey labels than can be displayed, the More LED lights and a More
message displays below the labels. To display the next group of labels, press the More hardkey.
22. Power Switch and LEDs
This switch selects the standby mode or the power on mode. In the standby position, the yellow
LED lights and all signal generator functions deactivate. The signal generator remains connected to
the line power, and some power is consumed by some internal circuits. In the on position, the green
LED lights and the signal generator functions activate.
8 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Signal Generator Overview
5. Error Message Area
6. Text Area
7. Softkey Label Area
4. Amplitude Area
3. Annunciators
2. Frequency Area
1. Active Function Area
Scroll Bar
If there is more
text than can be
displayed on one
screen, a scroll
bar appears here.
Use the Page Up
and Page Down
keys to scroll
through the text.
Front Panel Display
Front Panel Display
1. Active Function Area
This area displays the currently active function. For example, if frequency is the active function, the
current frequency setting appears. If the currently active function has an increment value
associated with it, that value also appears.
2. Frequency Area
This area displays the current frequency setting.
3. Annunciators
Annunciators show the status of some of the signal generator functions, and indicate error
conditions. An annunciator position may be used by more than one annunciator; in this case, only
one of the functions sharing a given position can be active at a given time.
This annunciator appears when...
ΦMPhase modulation is on. If you turn frequency modulation on, the FM annunciator replaces ΦM.
ARBThe ARB generator is on. ARB is running and not waiting on a trigger.
ALC OFFThe ALC circuit is disabled. The UNLEVEL annunciator appears in the same position if the ALC is enabled and
is unable to maintain the output level.
AMAmplitude modulation is on.
ARMEDA sweep has been initiated and the signal generator is waiting for the sweep trigger event.
ATTNHOLDThe attenuator hold function is on. When this function is on, the attenuator is held at its current setting.
AWGNReal Time I/Q Baseband additive white Gaussian noise is on.
BBG DACA DAC overflow is occurring, adjust the runtime scaling adjust until the BBG DAC annunciator turns off.
CHANCORRThe internal channel correction is enabled.
DETHTRThe ALC detector heater is not up to temperature. To meet ALC specifications the heater must be at
DIGBUSThe digital bus is in use.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 9
Another annunciator, UNLOCK, appears in the same position and has priority over the BBG DAC annunciator
(see UNLOCK, below).
temperature.
Signal Generator Overview
Front Panel Display
This annunciator appears when...
DIGMODCustom Arb waveform generator is on.
ERRAn error message is placed in the error queue. This annunciator does not turn off until you either view all of the
error messages or clear the error queue (see “Reading Error Messages” on page68).
EXTREFAn external frequency reference is applied.
FMFrequency modulation is on. If you turn phase modulation on, the ΦM annunciator replaces FM.
I/QI/Q vector modulation is on.
LThe signal generator is in listener mode and is receiving information or commands over the GPIB, USB, or VXI–
11/Sockets (LAN) interface.
M–TONEMultitone waveform generator is on.
MULTA frequency multiplier is set (see “Setting a Frequency Multiplier” on page108).
OFFSAn output offset is set (see “Setting an Output Offset” on page106).
PNPhase noise interference is on.
PULSEPulse modulation is on.
RThe signal generator is remotely controlled over the GPIB, USB, or VXI–11/Sockets (LAN) interface. When the
signal generator is in remote mode, the keypad is locked out. To unlock the keypad, press Local (see page 6).
REFAn output reference is set (see “Setting an Output Reference” on page107).
RF OFFThe signal generator’s RF Output is not enabled.
SThe signal generator has generated a service request (SRQ) over the GPIB, USB, or VXI–11/Sockets (LAN)
interface.
SWEEPThe signal generator is currently sweeping in list or step mode.
SWMANThe signal generator is in manual sweep mode.
TThe signal generator is in talker mode and is transmitting information over the GPIB, USB, or VXI–11/Sockets
(LAN) interface.
T–TONETwo–Tone waveform generator is on.
UNLEVELThe signal generator is unable to maintain the correct output level. This is not necessarily an indication of
instrument failure; unleveled conditions can occur during normal operation. Another annunciator, ALC OFF,
appears in the same position when the ALC circuit is disabled (see ALC OFF, above).
UNLOCKAny of the phase locked loops cannot maintain phase lock. To determine which loop is unlocked, examine the
error messages (see “Reading Error Messages” on page68).
WATRGThe current modulation mode is waiting on the Arb trigger.
WINITThe signal generator is waiting for you to initiate a single sweep.
4. Amplitude Area
This area displays the current output power level setting (If the RF Output is off, this area is greyed
out).
5. Error Message Area
This area displays abbreviated error messages. If multiple messages occur, only the most recent
message remains displayed. See “Reading Error Messages” on page 68.
6. Text Area
This area displays signal generator status information, such as the modulation status, and other
information such as sweep lists and file catalogs. This area also enables you to perform functions
such as managing information (entering information, and displaying or deleting files).
10 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Signal Generator Overview
Front Panel Display
7. Softkey Label Area
This area displays labels that define the function of the softkeys located immediately to the right of
the display. Softkey labels change, depending on the function selected.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 11
Signal Generator Overview
4. SWEEP OUT
2. EXT 1 &
EXT 2
5. PULSE
6. TRIG 1 & 2
7. REF IN
9. GPIB
10. LAN
11. Device USB
8. 10 MHz OUT
1. AC Power Receptacle
Option 1EM
only
See page7
Digital Modulation Connectors (Vector Models
Only) on page 15
The AC power cord receptacle accepts a three–pronged AC power cord that is supplied with the
signal generator. For details on line setting requirements and the power cord, see the Getting
Started Guide.
To avoid the loss of data, GPIB settings, or current user instrument states that have not been
permanently saved to nonvolatile memory, the signal generator should always be powered down
either via the instrument’s front panel power button or the appropriate SCPI command. Signal
generators installed in rack systems and powered down with the system rack power switch rather
than the instrument’s front panel switch display a Error -310 due to the instrument not being
powered down correctly.
12 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Connectorfemale BNCImpedance nominally 50 Ω
SignalAn externally supplied ±1 V
Damage Levels5 V
and 10 V
rms
p
signal that produces the indicated depth.
p
3. LF OUT
Connectorfemale BNC Impedance 50 Ω
SignalVoltage range: 0 to +5 V
Offset: -5 V to +5 V, nominal
For more information, see page
p
165.
4. SWEEP OUT
Signal Generator Overview
Connectorfemale BNC
Can drive 2 kΩ.
SignalVoltage range: 0 to +10 V, regardless of sweep width
In swept mode: beginning of sweep = 0 V; end of sweep = +10 V
In CW mode: no output
This is a multiple use connector. For signal routing selections, see pages
Impedance <1 Ω
47 and 169.
5. PULSE
Connectorfemale BNCImpedance nominally 50 Ω
SignalExternally supplied: +1 V = on; 0 V = off
Damage Levels≤ −0.3 and ≥+5.3 V
6. TRIG 1 & 2
Connectorfemale BNCImpedance high Z
SignalAn externally supplied TTL or CMOS signal for triggering operations, such as point to point in
manual sweep mode or an LF sweep in external sweep mode.
Triggering can occur on either the positive or negative edge.
Damage Levels≤ −0.5 and ≥+5.5 V
7. REF IN
Connectorfemale BNCImpedance nominally 50 Ω
SignalAn externally supplied −3.5 to +20 dBm signal from a timebase reference that is within ±1 ppm.
In its factory default mode, the signal generator can detect a valid reference signal at this
connector and automatically switch from internal to external reference operation. See “Presetting
the Signal Generator” on page 40. With Option 1ER (flexible reference input), you must explicitly
tell the signal generator the external reference frequency you wish to use; enter the information
through the front panel or over the remote interface.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 13
Connectorfemale BNCImpedance nominally 50 Ω
SignalA nominal signal level greater than 4 dBm.
9. GPIB
This connector enables communication with compatible devices such as external controllers, and is
one of three connectors available to remotely control the signal generator (see also 10. LAN and
11. Device USB).
10. LAN
The signal generator supports local area network (LAN) based communication through this
connector, which enables a LAN–connected computer to remotely program the signal generator.
The LAN interface supports auto–MDIX. The signal generator is limited to 100 meters on a single
cable (100Base–T). For more information on the LAN, refer to the Programming Guide.
11. Device USB
ConnectorType B
USB ProtocolVersion 2.0
Use this universal serial bus (USB) connector to connect a PC to remotely control the signal
generator.
12. Host USB
ConnectorType A
USB Protocol2.0
Use this universal serial bus (USB) to connect a USB Flash Drive (UFD) for data transfer. You can
connect or disconnect a USB device without shutting down or restarting the signal generator.
13. SD Card
Holds the Secure Digital (SD) nonvolatile memory card.
14 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Digital Modulation Connectors (Vector Models Only)
I OUT, Q OUT, OUT, OUT
OUT and OUT, require Option 1EL.
ConnectorType: female BNC Impedance: 50 Ω
DC–coupled
Signal
I OUTThe analog, in–phase component of I/Q modulation from the internal baseband generator.
Q OUTThe analog, quadrature–phase component of I/Q modulation from the internal baseband
generator.
OUT
OUT
Damage Levels> 1 VrmsDC Origin
Output Signal Levels into a 50 Ω Load
Used in conjunction with the I OUT connector to provide a balanced
Used in conjunction with the Q OUT connector to provide a balanced
Offset
—0.5 V
— 0.69 V
alpha = 0.5
— 0.71 V
alpha = 0.35
—Typically 1 V
, typical, corresponds to one unit length of the I/Q vector
pk
(2.84 dB), typical, maximum crest factor for peaks for π/4 DQPSK,
pk
(3.08 dB), typical, maximum crest factor for peaks for π/4 DQPSK,
pk
maximum
p–p
typically <10 mV
a
baseband stimulus.
a
baseband stimulus.
a. Balanced signals are signals present in two separate conductors that are symmetrical relative
to ground, and are opposite in polarity (180 degrees out of phase).
BB TRIG 1 & BB TRIG 2
Connectorfemale BNCImpedance nominally 50 Ω
SignalReserved for arbitrary and real-time baseband generators I/O, such as markers or trigger inputs.
EVENT 1
Connectorfemale BNC Impedance: nominally 50 Ω
SignalA pulse that can be used to trigger the start of a data pattern, frame, or timeslot.
Adjustable to ± one timeslot; resolution = one bit
Markers
Each Arb–based waveform point has a marker on/off condition associated with it.
Marker 1 level = +3.3 V CMOS high (positive polarity selected); –3.3 V CMOS low (negative
polarity selected).
Output on this connector occurs whenever Marker 1 is on in an Arb–based waveform (see
“Using Waveform Markers” on page203).
Damage Levels< −4 and > +8 V
Keysight EXG and MXG X-Series Signal Generators User’s Guide 15
Connectorfemale BNC
SignalA TTL/CMOS low to TTL/CMOS high, or TTL/CMOS high to TTL/CMOS low edge trigger.
The input to this connector triggers the internal digital modulation pattern generator to start a
single pattern output or to stop and re–synchronize a pattern that is being continuously output.
To synchronize the trigger with the data bit clock, the trigger edge is latched, then sampled
during the falling edge of the internal data bit clock.
This is the external trigger for all ARB waveform generator triggers.
Minimum Trigger Input Pulse
Width
Minimum Trigger Delay (trigger edge to first bit of frame) = 1.5 to 2.5 bit clock periods
Damage Levels< −4 and > +8 V
(high or low) = 10 ns
DIGITAL BUS I/O
This is a proprietary bus used by Keysight Technologies signal creation software. This connector is
not operational for general purpose use. Signals are present only when a signal creation software
option is installed (for details, refer to http://www.keysight.com/find/signalcreation).
The X-Series’ Digital BUS I/O connector can be used for enabling
operation with the Keysight Technologies N5106A PXB MIMO Receiver
Teste r.
16 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Signal Generator Overview
18
1
19
36
The AUX I/O connector is a shielded .050 series
View looking into the rear panel
The AUX I/O mating connector manufacturer’s part number is
3M® 10136-3000 (wire mount plug).
This female 36-pin connector is available only on instruments with an internal baseband generator
(Option 653, 655, 656, 657). On signal generators without one of these options, this connector is
not present.
The AUX I/O connector allows the X-Series signal generator to interface with external equipment
by sending and/or receiving supplementary (auxiliary) signaling information. This information is
non-RF related signaling such as:
— output markers to an external device from Arbitrary waveform playback sent to external
equipment to trigger or respond to waveform changes.
— output of signal markers to an external device from real-time signal generation personalities.
Signals such as frame markers, pulse-per-second, and even-second, for example, may be
supported, depending on the signal generation personality (CDMA, 3GPP, GNSS, LTE, etc.).
— input signals from external devices under test to cause the signal generator to modify
characteristics of a signal being generated, depending on the signal generation personality
(CDMA, 3GPP, LTE, etc.).
Table 1-1 on page 19 describes the inputs and outputs accessible through the AUX I/O connector.
The specific functions controllable by auxiliary signaling vary significantly from one real-time signal
generation personality to another. Refer to the documentation for each real-time signal generation
personality for additional information.
The AUX I/O connector supports standard 3.3V TTL signaling levels.
Signals support data rates up to 50 MHz with minimum rise and fall times
of 3ns. Any pins that are not connected will have a weak pull-up to 3.3V.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 17
Signal Generator Overview
Markers (pins 1-4)
Each Arb–based waveform point has a marker on/off condition associated with it.
Each real-time signal can be routed to the output marker signals using SCPI commands or the real-time personalities.
Marker level = +3.3 V high (positive polarity selected); 0V low (negative polarity selected).
Event 1 (pin 1)
Pin 1 outputs a pulse that can be used to trigger the start of a data pattern, frame, or timeslot.
Adjustable to ± one timeslot; resolution = one bit
Data Clock Out (pin 7)
Pin 7 is used with an internal baseband generator. This pin relays a CMOS bit clock signal for synchronizing serial data.
Data In (pin 23)
Pin 23 accepts an externally supplied CMOS-compatible signal data input used with digital modulation applications. The
expected input is a CMOS signal where a CMOS high is equivalent to a data 1 and a CMOS low is equivalent to a data 0.
The maximum input data rate is 50 Mb/s. The data must be valid on the DATA CLOCK
falling edges.
Symbol Sync In (pin 25)
Pin 25 accepts an externally supplied symbol sync signal for use with digital modulation applications.
Data Clock In (pin 29)
Pin 29 accepts an externally supplied CMOS-compatible signal data-clock input used with digital modulation applications.
The expected input is a MCOS bit clock signal where the rising edge is aligned with the beginning data bit. The falling edge
is used to clock the DATA and SYMBOL SYNC signals.
The maximum clock rate is 50 MHz.
Event 2 (pin 31)
Pin 31 outputs data enable signal for gating external equipment. The output is applicable when the external data is clocked
into internally generated timeslots. Data is enabled when the signal is low.
a. Settings shown are for the Error Out signal configuration of the AUX I/O connector (BERT > I/O
Setup > Aux I/O Out). Press the Help hardkey, then either Reference Out or PN9 Out for the
respective signal configuration.
20 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Rear Panel Overview (N5173B & N5183B)
4. SWEEP OUT
2. EXT 1 & EXT 2
5. PULSE
6. TRIG 1 & 2
7. REF IN
9. GPIB
10. LAN
11. Device USB
8. 10 MHz OUT
1. AC Power Receptacle
Option 1EM
only
See page7
12. Host USB
13. SD Card
3. LF OUT
14. ALC INPUT
15. Z AXIS OUTPUT
1. AC Power Receptacle
Signal Generator Overview
Rear Panel Overview (N5173B & N5183B)
The AC power cord receptacle accepts a three–pronged AC power cord that is supplied with the
signal generator. For details on line setting requirements and the power cord, see the Getting
Started Guide.
To avoid the loss of data, GPIB settings, or current user instrument states
that have not been permanently saved to nonvolatile memory, the signal
generator should always be powered down either via the instrument’s
front panel power button or the appropriate SCPI command. Signal
generators installed in rack systems and powered down with the system
rack power switch rather than the instrument’s front panel switch display a
Error -310 due to the instrument not being powered down correctly.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 21
Signal Generator Overview
Rear Panel Overview (N5173B & N5183B)
2. EXT 1 & EXT 2
Connectorfemale BNCImpedance nominally 50 Ω
SignalAn externally supplied ±1 V
Damage Levels5 V
and 10 V
rms
p
signal that produces the indicated depth.
p
3. LF OUT
Connectorfemale BNC Impedance 50 Ω
SignalVoltage range: 0 to +5 V
Offset: -5 V to +5 V, nominal
For more information, see page <PagenumOnlyCallout>165.
p
4. SWEEP OUT
Connectorfemale BNC
Can drive 2 kΩ.
SignalVoltage range: 0 to +10 V, regardless of sweep width
In swept mode: beginning of sweep = 0 V; end of sweep = +10 V
In CW mode: no output
This is a multiple use connector. For signal routing selections, see pages
Impedance <1 Ω
47 and 169.
5. PULSE
Connectorfemale BNCImpedance nominally 50 Ω
SignalExternally supplied: +1 V = on; 0 V = off
Damage Levels≤ −0.3 and ≥+5.3 V
6. TRIG 1 & 2
Connectorfemale BNCImpedance high Z
SignalAn externally supplied TTL or CMOS signal for triggering operations, such as point to point in
manual sweep mode or an LF sweep in external sweep mode.
Triggering can occur on either the positive or negative edge.
Damage Levels≤ −0.5 and ≥+5.5 V
7. REF IN
Connectorfemale BNCImpedance nominally 50 Ω
SignalAn externally supplied −3.5 to +20 dBm signal from a timebase reference that is within ±1 ppm.
In its factory default mode, the signal generator can detect a valid reference signal at this
connector and automatically switch from internal to external reference operation. See “Presetting
the Signal Generator” on page 40. With Option 1ER (flexible reference input), you must explicitly
tell the signal generator the external reference frequency you wish to use; enter the information
through the front panel or over the remote interface.
22 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Signal Generator Overview
Rear Panel Overview (N5173B & N5183B)
8. 10 MHz OUT
Connectorfemale BNCImpedance nominally 50 Ω
SignalA nominal signal level greater than 4 dBm.
9. GPIB
This connector enables communication with compatible devices such as external controllers, and is
one of three connectors available to remotely control the signal generator (see also 10. LAN and
11. Device USB).
10. LAN
The signal generator supports local area network (LAN) based communication through this
connector, which enables a LAN–connected computer to remotely program the signal generator.
The LAN interface supports auto–MDIX. The signal generator is limited to 100 meters on a single
cable (100Base–T). For more information on the LAN, refer to the Programming Guide.
11. Device USB
ConnectorType B
USB ProtocolVersion 2.0
Use this universal serial bus (USB) connector to connect a PC to remotely control the signal
generator.
12. Host USB
ConnectorType A
USB Protocol2.0
Use this universal serial bus (USB) to connect a USB Flash Drive (UFD) for data transfer. You can
connect or disconnect a USB device without shutting down or restarting the signal generator.
13. SD Card
Holds the Secure Digital (SD) nonvolatile memory card.
14. ALC INPUT
This input connector is used for negative external detector leveling.
Connectorfemale BNCImpedance nominally 100 kΩ
Signal-0.2 mV to -0.5 V
Damage Levels -12 to 1 V
Keysight EXG and MXG X-Series Signal Generators User’s Guide 23
Signal Generator Overview
Rear Panel Overview (N5173B & N5183B)
15. Z AXIS OUTPUT
This female BNC connector supplies a +5 V (nominal) level during retrace and band-switch intervals
of a step or list sweep. During step or list sweep, this female BNC connector supplies a -5 V
(nominal) level when the RF frequency is at a marker frequency and intensity marker mode is on.
This signal is derived from an operational amplifier output so the load impedance should be greater
than or equal to 5 kohms.1
Connectorfemale BNCImpedance nominally 50 Ω
SignalA nominal signal level greater than 4 dBm.
24 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Keysight X-Series Signal Generators
Display Settingson page 27
Front Panel Knob Resolutionon page 28
Power On and Preseton page 28
Enabling an Optionon page 37
GPIB Address and Remote Languageon page 31
Configuring the LAN Interfaceon page 32
Enabling LAN Services: “Browser,” “Sockets,” and “VXI–11”on page 32
Configuring the Remote Languageson page 33
User Preferences
Remote Operation Preferences
Setting Time and Dateon page 29
Upgrading Firmwareon page 31
Hardware Assembly Installation and Removal Softkeyson page 38
Reference Oscillator Tuneon page 30
N5171B/72B/73B EXG and N5181B/82B/83B MXG
User’s Guide
2Preferences & Enabling Options
The Utility menu provides access to both user and remote operation preferences, and to the menus
in which you can enable instrument options.
25
Preferences & Enabling Options
User Preferences
User Preferences
From the Utility menu, you can set the following user preferences:
— Display Settings, below
— Power On and Preset on page28
— Front Panel Knob Resolution on page28
26 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Display Settings
Range: 0 to 100
Range: 1 to 12 hours, in 1 hour increments
Light Only: turns the display light off, leaving the text visible at a lo w inten sity.
Light & Text: turns the display light and the text off.
If the display remains unchanged for long periods of time, use this mode to
prevent the text from burning the display.
When on, commands executed through
the remote control bus update the signal
generator display accordingly.
Bright without color
Utility > Display
See also, Using Secure Display (Option 006)on page 438.
For details on each key, use key help
as described on page40.
page438
Dark text on a light
background.
Light text on a dark
background.
Preferences & Enabling Options
User Preferences
X-Series signal generators are shipped from the factory with default
display settings. Automated Test Environment (ATE) users may benefit
from display settings other than the default settings. When the signal
generator display is not required to be active for long periods of time,
consider using the Screen Saver Mode to extend the life of the display.
With the brightness set to minimum, the display may be too dark to see
the softkeys. If this happens, use the figure above to locate the brightness
softkey and adjust the value so that you can see the display.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 27
Preferences & Enabling Options
Select the GPIB language desired after a preset.
See also, the Programming Guide and the SCPI Command Reference.
Utility > Power On/Preset
Restores persistent settings
(those unaffected by a power
cycle*, preset, or recall)
to their factory defaults.
Available only when 8648 is either the selected preset language or
the selected remote language (see page31).
*
Caution
To avoid the loss of data, GPIB settings, or current user instrument states that have not been permanently saved to
nonvolatile memory, the X-Series signal generator should always be powered down either via the instrument’s front
panel power button or the appropriate SCPI command. X-Series signal generators installed in rack systems and
powered down with the system rack power switch rather than the instrument’s front panel switch display a Error -310
due to the instrument not being powered down correctly.
Note
To define a user preset, set the instrument up as desired and press User Preset > Save User Preset.
Makes the increment value of the current function the active entry.
Utility >
Instrument Adjustments
The increment value and the step/knob ratio determine how much each
turn of the knob changes the active function value.
For example, if the increment value of the active function is 10 dB and the
step/knob ratio is 50 to 1, each turn of the knob changes the active
function by 0.2 dB (1/50th of 10 dB).
To change the amount for each turn of the knob, modify the increment
value, the step/knob ratio, or both.
For details on each key, use key help
as described on page40.
page30.
page29.
Executes the ALC Modulator Bias Adjustment. When using
the ALC in open loop mode, this compensates for open loop
power drift due to temperatur e and humidity.
SCPI Commands:
:CALibration:ALC:MODulator:BIAS
User Preferences
Power On and Preset
Front Panel Knob Resolution
28 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Preferences & Enabling Options
Utility >
Instrument Adjustments
>
Original time/date reference point
Time
New time/date reference point
User Preferences
Setting Time and Date
The signal generator’s firmware tracks the time and date, and uses the latest time and date as its
time/date reference point.
Changing the time or date can adversely affect the signal generator’s ability to use
time.
— based licenses, even if a time based license is not installed.
Figure 2-1
Setting the Time or Date Forward
If you set the time or date forward, be aware that you are using up any installed time–based
licenses, and that you are resetting the signal generator’s time/date reference point. When you
set a new time or date that is later than the signal generator’s current reference point, that date
becomes the new reference point. If you then set the date back, you run the risk described in the
next section.
Setting the Time or Date Backward
When you set the time back, the signal generator notes that the time has moved back from the
reference point. If you set the time back more than a few hours, you disable the signal generator’s
ability to use time–based licenses, even if there is no license installed at the time that you set the
time back. In this case, you can re-enable the signal generator’s ability to use time–based licenses
by moving the clock forward to the original time or simply waiting that length of time.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 29
Preferences & Enabling Options
Tunes the internal VCTXCO oscillator frequency.
The user value offsets the factory tuned value (the value is added to the
factory calibrated DAC value). The tune value of 0 sets the factory
calibrated value.
The range; –8192 to 8192, can be set by using the front panel keypad,
knob or remote command.
For details on each key, use key help
as described on page40.
See also the SCPI Command Reference.
Utility > Instrument Adjustments
User Preferences
Reference Oscillator Tune
30 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Preferences & Enabling Options
page32
page32
NOTES
USB is also available. It is not shown in the menu because it requires no configurati on.
For details on using the instrument remotely, see the Programming Guide.
Select the desired language.
This setting is not persistent and is cleared by performing an
instrument Preset. In most cases, it is best to use Utility > Power On/Preset > Language for a permanent language change. See
page33.
For details on each key, use key help as described on page40.
Select On to view the associated SCPI command for a key press.
Select On to place a copy of every subsequent SCPI command
executed in the log.
Select Off to stop logging. Selecting Off do es not clear the log.
Press the Clear SCPI Log softkey to clear the log.
Upgrading Firmware
Upgrading Firmware
For information on new firmware releases, go to http://www.keysight.com/find/upgradeassistant.
Remote Operation Preferences
For details on operating the signal generator remotely, refer to the Programming Guide.
GPIB Address and Remote Language
Keysight EXG and MXG X-Series Signal Generators User’s Guide 31
Preferences & Enabling Options
NOTES
Use a 100Base–T LAN cable to connect the
signal generator to the LAN.
Use a crossover cable to connect the signal
generator directly to a PC.
For details on using the instrument remotely,
refer to the Programming Guide and to
www.keysight.com and search on FAQs:
Hardware Configurations and Installation for the
Keysight MXG.
Utility > I/O Config
Values are listed in the
Programming Guide
For details on each key, use key help as described on page40.
page32.
Enable remote (browser)
access to the
instrument’s file system.
Use a browser to control
the signal generator.
Utility > I/O Config
For details on each key, use key help
as described on page40.
License Manager Server (On)
allows updates of the
instrument licenses, disable
for additional instrument
security.
For more information refer to
the Programming Guide.
Remote Operation Preferences
Configuring the LAN Interface
Enabling LAN Services: “Browser,” “Sockets,” and “VXI–11”
32 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Configuring the Remote Languages
Utility > I/O Config
For details on each key, use key help
as described on page40.
Select the desired Remote language.
Refer to the SCPI Command Reference.
Figure 2-2 N5171B/72B/81B/82B
Preferences & Enabling Options
Remote Operation Preferences
Keysight EXG and MXG X-Series Signal Generators User’s Guide 33
Preferences & Enabling Options
Utility > I/O Config
For details on each key, use key help as described on page40.
Select the desired Remote language.
Refer to the SCPI Command Reference.
Remote Operation Preferences
Figure 2-3 N5173B/83B
34 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Configuring the Preset Languages
Utility > Power On/Preset
For details on each key, use key help
as described on page40.
Select the desired Remote language.
page28
Refer to the SCPI Command Reference.
Figure 2-4 N5171B/72B/81B/82B
Preferences & Enabling Options
Remote Operation Preferences
Keysight EXG and MXG X-Series Signal Generators User’s Guide 35
Preferences & Enabling Options
Utility > Power On/Preset
For details on each key, use key help as described onpage40.
Select the desired Remote language.
page28
Refer to the SCPI Command Reference.
Remote Operation Preferences
Figure 2-5 N5173B/83B
36 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Preferences & Enabling Options
Utility >
Instrument Info
Service Software Licenses
Waveform licenses from some
Instrument options appear
here. A check mark means that
an option is enabled.
For details on each key, use key help
as described on page40.
Signal Studio applications appear here.
appear here.
Enabling an Option
Enabling an Option
There are two ways to enable an option:
— Use the License Manager software utility:
1. Run the utility and follow the prompts.
2. Download the utility from www.keysight.com/find/LicenseManager and select license (.lic)
files from an external USB Flash Drive (UFD).
— Use SCPI commands, as described in the Programming Guide.
Viewing Options and Licenses
Keysight EXG and MXG X-Series Signal Generators User’s Guide 37
Preferences & Enabling Options
Utility > More 2 of 2 >
For details on each key, use key help
as described on page40.
Verify output attenuator
operation using a power
meter at the RF Output.
Select either Enhanced Factory
Calibration or Factory Calibration
to calibrate your instrument.
The Enhanced Factory
Calibration requires a spectrum
analyzer in addition to a power
meter.
The Factory Calibration uses only
a power meter.
Whether a softkey
is available depends
on the model of signal
generator.
Whether a softkey is
available depends on
the model of signal
generator.
Hardware Assembly Installation and Removal Softkeys
Hardware Assembly Installation and Removal Softkeys
38 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Keysight X-Series Signal Generators
N5171B/72B/73B EXG and N5181B/82B/83B MXG
User’s Guide
3Basic Operation
This chapter introduces fundamental front panel operation. For information on remote operation,
refer to the Programming Guide.
— Presetting the Signal Generator on page 40
— Viewing Key Descriptions on page 40
— Entering and Editing Numbers and Text on page 41
— Setting Frequency and Power (Amplitude) on page 43
— Setting ALC Bandwidth Control on page 45
— Configuring a Swept Output on page 46
— Modulating the Carrier Signal on page 54
— Working with Files on page 56
— Reading Error Messages on page 68
39
Basic Operation
To return the signal generator to a known state, press either Preset or User Preset. Preset is the factory preset; User Preset is a custom preset** (see also, page28).
To reset persistent settings (those unaffected by preset, user preset, or power cycle*),
press: Utility > Power On/Preset > Restore System Defaults.
*Caution
T o avoid the l oss of data, GP IB settings, or current user instrument stat es that ha ve not been per manently save d to nonvol atile memory ,
the instrument should always be powered down either via the instr ument's front pan el power but ton or the appr opriate SC PI command.
instrument's installed in rack systems and powered down with the system rack power switch rather than the instrument's front panel
switch display a Error -310 due to the instrument not being powered down correctly.
**You can create more than one user preset by giving each saved state file a different name (see Figure 3-9 on page66).
The Help hardkey enables you to display a description of any hardkey or softkey.
To display help text:
1. Press Help.
2. Press the desired key.
The help displays and the key’s normal function does not execute.
Presetting the Signal Generator
Presetting the Signal Generator
Viewing Key Descriptions
40 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Entering and Editing Numbers and Text
Up/down arrow keys increase/decrease a selected (highlighted) numeric value, and move the cursor vertically.
To specify a negative value, enter the
negative sign either before or after
Backspace moves the cursor to the left,
deleting characters as it goes.
Left/right arrow keys move the cursor horizontally.
Page up/down keys move tables of data up and down within the display area.
Use the number keys and decimal point to enter numeric data.
Note
: Rotating the knob increases or
decreases a numeric value, changes a
highlighted digit or character, or steps
through lists or items in a row.
the numeric value (this key is a toggle).
For details on each key, see page40.
Use the Select hardkey to choose part of an entry, as when entering alpha
characters. In some menus, the
Select key also acts as a terminator, and is
equivalent to the
Enter softkey.
See also “Front Panel Knob Resolution” on page28.
to move the cursor
within the active value
rather than within the
alpha table, turn the
alpha table off.
Add/edit comments for saved
instrument state files (see page62).
Note: File names are limited to 25 characters.
Entering Numbers and Moving the Cursor
Basic Operation
Entering and Editing Numbers and Text
Entering Alpha Characters
Data entry softkeys appear in various
menus. If their meaning is not clear in
context, use the help key (described on
page 40) to display an explanation. Use the
softkey next to the alpha table for help on
the table.
Selecting data that accepts alpha
characters, displays one of the menus
shown at right.
Use the arrow keys or knob to highlight the
desired letter, then press the Select
hardkey (or the softkey next to the alpha table). To correct errors, use Bk Sp or Clear Text.
To terminate the entry, press the Enter softkey.
A subset of this menu appears for hexadecimal characters. The character menu displays only the
letters A through F (use the numeric keypad for other values).
Keysight EXG and MXG X-Series Signal Generators User’s Guide 41
Basic Operation
Cursor
Table Editor Softkeys
Table Editor Name
Table Items
Displays the active item as you edit it.
Highlighting indicates the selected item. To make this the active (editable)
item, either press Select, or simply enter the desired value.
Table items are also
called data fields.
Used to load, navigate, modify , an d
store table item values. For details
on each key, use the key help:
Press the Help hardkey and then
the desired key.
Active Function Area
Current / Total Number of Pages
Indicates that
another menu
is available; to
display the
second menu,
press More.
(vector models only)
Entering and Editing Numbers and Text
Example: Using a Table Editor
Table editors simplify configuration tasks. The following procedure describes basic table editor
functionality using the List Mode Values table editor.
1. Preset the signal generator: Press Preset.
2. Open the table editor: Press Sweep > More > Configure List Sweep.
The signal generator displays the editor shown in the following figure.
3. Highlight the desired item: use the arrow keys or the knob to move the cursor.
4. (Optional) Display the selected item in the active function area: Press Select.
5. Modify the value:
— If the value is displayed in the active function area, use the knob, arrow keys, or numeric
keypad to modify the value.
— If the value is not displayed in the active function area, use the numeric keypad to enter the
desired value (which then appears in the active function area).
6. Terminate the entry:
— If available, press the desired units.
— If units are not displayed, press either Enter (if available) or Select.
The modified item is displayed in the table.
42 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Setting Frequency and Power (Amplitude)
page106
To display the next menu, press
More.
For details on each key, use key help
as described on page40.
page44
Refer to the SCPI Command Reference.
Opens the
Atten/ALC
Control
menu.
Enables the optimize signal-to-noise ratio state (Optimize S/N) On or Off.
The optimize signal–to–noise softkey changes the attenuator and alc setting to provide
optimal signal–to–noise performance; it does not change the RF output power.
Caution: When the optimize signal-to-noise ratio state is enabled, some increased levels of
harmonic distortion can occur. This increased harmonic distortion could degrade ACPR and
EVM.
Note: This mode is mutually exclusive with attenuator hold (Atten Hold), and any modulation
type. A settings conflict error will be generated if attentuator hold or any modulation is
activated when optimize signal–to–noise is active (On).
page45
This softkey sets a user determined
maximum power level on the
instrument that is persistent.
Available on vector
models with BBG.
Enables the instrument
to select between two
power levels with a
hardware marker or
using an external BNC.
page82
In Amplitude mode, this menu is
automatically displayed when entering
a numeric value with the front panel
keypad.
In Frequency mode, this menu is
automatically displayed when entering
a numeric value with the front panel
keypad.
Sets the current relative
phase of the RF output
signal as the zero
reference.
All subsequent
adjustments are relative
to this adjustment.
dBuVemf terminates
the value as dBuV
electromotive force.
Emf is the output
voltage with no load.
This unit is external
only. Internal
representation as well
as increment and
decrement values are
in dBm.
Note: dBuV, mV, and
nVemf behave similarly
(i.e. Values are external
only. Internal
representation as well
as increment and
decrement values are
in dBm).
Enables the mode to improve non-harmonics performance (low spurs mode).
Note: When this mode is enabled, the instrument switching speed and behavior of
the source’s settled signal are affected.
Modifies the attenuator and ALC settings
for optimal performance. It does not
change the RF output power.
Available on
vector models.
Figure 3-1 Frequency and Amplitude Softkeys
Basic Operation
Setting Frequency and Power (Amplitude)
Keysight EXG and MXG X-Series Signal Generators User’s Guide 43
Basic Operation
For details on each key, use key help as described on page40.
Range: 1 to 50 MHz
For guaranteed operation below 2 MHz, select Narrow bandwidth; in Wide bandwidth,
operation using frequencies below 2 MHz is not guaranteed.
Freq > More
Figure3-2 illustrates the better close in phase noise of the wide bandwidth mode. If the
external reference you are using is noisy or spurious, use the narrow bandwidth mode.
With Wide bandwidth selected, if you select a frequency less than 2 MHz, the signal
generator displays a warning message.
uses the signal generator’s internal reference oscillator.
uses the REF IN connector to get it’s Frequency Reference. Note: An error
screen is displayed if the reference is not connected or is tuned to the wrong
frequency.
uses the signal generator’s internal reference unless a signal is present at
the rear panel REF IN connector. If an external signal is present, that signal is
used as the reference.
uses the baseband generator clock (BBG) as the reference oscillator source.
Available on vector models with BBG.
Setting Frequency and Power (Amplitude)
Example: Configuring a 700 MHz, −20 dBm Continuous Wave Output
1. Preset the signal generator.
The signal generator displays its maximum specified frequency and minimum power level (the
front panel display areas are shown on page 9).
2. Set the frequency to 700 MHz: Press Freq > 700 > MHz.
The signal generator displays 700 MHz in both the FREQUENCY area of the display and the
active entry area.
3. Set the amplitude to −20 dBm: Press Amptd > –20 > dBm.
The display changes to −20 dBm in the AMPLITUDE area of the display, and the amplitude
value becomes the active entry. Amplitude remains the active function until you press another
function key.
4. Turn on the RF Output: Press RF On/Off.
The RF Output LED lights, and a 700 MHz, −20 dBm CW signal is available at the RF OUTPUT
connector.
Using an External Reference Oscillator
When using an external reference, you can select either narrow or wide bandwidth mode.
44 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Figure 3-2 Using an External Reference Oscillator
To display the next menu, press
More.
For details on each key, use key help
as described on page40.
Refer to the SCPI Command Reference.
Enables the
automatic
ALC
bandwidth
mode (Auto).
Basic Operation
Setting ALC Bandwidth Control
Setting ALC Bandwidth Control
Figure 3-3 Amplitude Softkeys
Keysight EXG and MXG X-Series Signal Generators User’s Guide 45
Basic Operation
Sweep without waiting for
a trigger at each point.
Point Trigger pauses fo r the dw ell
time prior to the first sweep. See
page110.
Apply a TTL/CMOS signal to the
Trigger In connector.
Periodically issue a trigger event to
whatever selects it as a source.
Using timer trigger with single
sweep results in a delay prior to the
first sweep. See page110.
Trigger on a remote command.
Down sweeps from
stop to start
frequency/amplitude.
Up sweeps from start
to stop
frequency/amplitude.
page49
The selected sweep type determines the displayed parameter.
Selecting step sweep also displays the step spacing (Lin or Log).
On vector models:
Available when
Sweep Type = List
page53
During a sweep, the swept parameter (frequency, amplitude, or both) turns grey and changes as the parameter sweeps.
Progress Bar: Note that very fast sweeps
can appear to sweep randomly or
backward.
page47
For details on each key, use key help as described on page40.
page47
Configuring a Swept Output
Configuring a Swept Output
The signal generator has two methods of sweeping through a set of frequency and amplitude
points:
Step sweep (page 47) provides a linear or logarithmic progression from one selected frequency,
amplitude, or both, to another, pausing at linearly or logarithmically spaced points (steps) along
the sweep. The sweep can progress forward, backward, or manually.
List sweep (page 49) enables you to enter frequencies and amplitudes at unequal intervals, in
nonlinear ascending, descending, or random order. List sweep also enables you to copy the current
step sweep values, include an Arb waveform in a sweep (on a vector instrument), and save list
sweep data in the file catalog (page 61).
Figure 3-4 Sweep Softkeys
46 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Operation
These softkeys are
only available for
the vector models
with a BBG.
Routes non
Step or List
Sweep signals
(i.e. SCPI
command
development.
Source Settled
is not affected
by Trig Out
Polarity Neg
Pos softkey
selection.
Routes Step or
List Sweep
For details on each key, use key help as described on page40.
Opens a menu for
routing Markers 1
through 4 to the
rear panel
baseband BNC
connectors.
Routes a signal to the Trig Out (Trig 1 or Trig 2) rear panel
connectors as follows:
None: No signals are routed to Trig 1 or Trig 2 BNCs
Sweep Trigger Out: Sweep Trigger Out signal (trigger
handshake)
Source Settled: Source Settled signal
Pulse Sync: Pulse Modulation Synchronization signal
Pulse Video: Pulse Modulation Video signal
LXI: LXI Trigger Output
Sweep Run: Sweep Run signal (sweep is running)
Other Trigger BNC: The signal on the other Trigger BNC
Pulse BNC: The signal on the PULSE IN BNC
Swept Func Done: Swept Function Sweep Done signal
Configuring a Swept Output
Routing Signals
Sweep > More > More > Route Connectors
Step Sweep
Step sweep provides a linear or logarithmic progression from one selected frequency, or amplitude,
or both, to another, pausing at linearly or logarithmically spaced points (steps) along the sweep.
The sweep can progress forward, backward, or be changed manually.
Figure 3-5 Signal Routing Softkeys
Keysight EXG and MXG X-Series Signal Generators User’s Guide 47
Basic Operation
Dwell Time = the time that the signal is settled and you can make
a measurement before the sweep moves to the next point.
(Point to point time is the sum of the value set for the dwell plus
processing time, switching time, and settling time.)
Lin = steps equally spaced over the sweep; the output
changes linearly.
Log = step spacing increases logarithmically over the
sweep; the output changes exponentially.
Step Sweep and List Sweep dwell times are set
For details on each key, use key help
as described on page40.
Sets the step size for a linear step sweep in frequency (the
difference between frequency points).
NOTE: Setting the step size will determine the number of
points in the step sweep based on the current start and stop
frequencies. Due to the integer number of step points, the
step size may be adjusted in order to yield a true linear
sweep between the start and stop frequencies.
Enables protection for the mechanical attenuator by automatically turning on
Atten Hold during frequency and/or power step sweeps.
This may cause unleveled RF output to occur for certain sweep configurations.
Disabling this attenuator protection will allow the sweep to optimally set both
the automatic leveling control (ALC) and output attenuation at each sweep
point.
When the attenuator protection is disabled, the step dwell time will be set to a
minimum of 50 ms as a precaution.
Example: Configuring a Continuous, Linear Step Sweep
Output:A signal that continuously sweeps from 500 to 600 MHz and from −20 to 0 dBm, with a dwell time of
500 ms at each of six equally spaced points
1. Preset the instrument and open the Sweep/List menu: Press Preset > SWEEP.
Because continuous is the default sweep repeat selection, and linear is the default step
spacing selection, you do not need to set these parameters.
2. Open the step sweep menu: Press Configure Step Sweep.
3. Set the following parameters:
Start frequency 500 MHz:
Stop frequency 600 MHz:
Amplitude at the beginning of the sweep, −20 dBm:
Amplitude at the end of the sweep, 0 dBm:
6 sweep points:
Dwell time at each point, 500 milliseconds:
4. Sweep both frequency and amplitude: Press Return > Return > Sweep > Freq Off On > Amptd
Off On.
48 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Press
Freq Start > 500 > MHz
Press Freq Stop > 600 > MHz
Press
Amptd Start > −20 > dBm
Press Amptd Stop > 0 > dBm.
Press
# Points > 6 > Enter
Press More > Step Dwell > 500 >
msec
Basic Operation
see page50
The selected sweep determines which dwell time the signal generator uses. Step Swe ep dwell
time is the same at each point; List Sweep dwell time can be different at each point.
Dwell Time = the time that the signal is settled and you can make a measurement before the
sweep moves to the next point.
Point–to–Point Time = the sum of the value set for the dwell plus processing time, switching
time, and settling time.
Each line defines the
corresponding point in
the sweep. For example,
line 1 defines point 1.
page47
Available only on vector models,
and
only when Sweep Type = List.
Displays the selected
sweep type parameters
(see page50)
Vector models only
page47
For details on each key, use key help
as described on page40.
Configuring a Swept Output
A continuous sweep begins, from the start frequency/amplitude to the stop
frequency/amplitude. The SWEEP annunciator displays, and sweep progress is shown in the
frequency display, the amplitude display, and the progress bar.
5. Turn the RF output on: Press RF On/Off.
The RF LED lights, and the continuous sweep is available at the RF Output connector.
List Sweep
List sweep enables you to enter frequencies and amplitudes at unequal intervals in nonlinear
ascending, descending, or random order. List sweep also enables you to copy the current step
sweep values, include a waveform in a sweep (on a vector instrument), and save list sweep data in
the file catalog (page 61). Dwell time is editable at each point. For fastest switching speeds, use list
sweep.
Figure 3-7 List Sweep Configuration Softkeys and Display
Keysight EXG and MXG X-Series Signal Generators User’s Guide 49
Basic Operation
Vector Models:
Presetting the list clears any
previously selected waveforms.
For information on selecting a list
sweep waveform, see “Example:
Editing List Sweep Points” on
page50.
Waveforms are available
only on vector models.
Configuring a Swept Output
Example: Configuring a List Sweep Using Step Sweep Data
1. Set up the desired step sweep, but do not turn the sweep on. This example uses the step
sweep configured on page 48.
2. In the SWEEP menu, change the sweep type to list:
Press SWEEP > Sweep Type List Step to highlight List.
The display shows sweep list parameters, as shown below.
3. Open the List Sweep menu: Press More > Configure List Sweep.
4. Clear any previously set values from the menu and load the points defined in the step sweep
into the list: Press More > Preset List > Preset with Step Sweep > Confirm Preset.
The display updates with the values loaded from the step sweep, as shown.
5. Sweep frequency and amplitude: Press SWEEP (hardkey) > Sweep > Freq Off On > Amptd Off
On.
Setting the sweep turns on the sweep function; a continuous sweep begins. On the display,
the SWEEP annunciator appears, and the progress bar shows the progression of the sweep.
6. If not already on, turn the RF output on: Press RF On/Off.
The RF Output LED lights, and a continuous sweep is available at the RF OUTPUT connector.
Example: Editing List Sweep Points
If you are not familiar with table editors, refer to page 42.
1. Create the desired list sweep. This example uses the list sweep created in the previous
example.
50 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Operation
Either select a waveform,
page40
or
select no modulation.
Configuring a Swept Output
2. If sweep is on, turn it off. Editing list sweep parameters with sweep on can generate an error.
3. Ensure that the sweep type is set to list: Press SWEEP > Sweep Type List Step to highlight
List.
4. In the List Mode Values table editor, change the point 1 dwell time (defined in row 1) to 100
ms:
a. Press More > Configure List Sweep.
b. Highlight the point 1 dwell time.
c. Press 100 > msec.
The next item in the table (the frequency value for point 2) highlights.
5. Change the selected frequency value to 445 MHz: Press 445 > MHz.
6. Add a new point between points 4 and 5: Highlight any entry in row 4 and press Insert Row.
This places a copy of row 4 below row 4, creating a new point 5, and renumbers subsequent
rows.
7. Shift frequency values down one row, beginning at point 5: Highlight the frequency entry in
row 5, then press More > Insert Item.
This places a copy of the highlighted frequency value in row 6, shifting the original frequency
values for rows 6 and 7 down one row. The new row 8 contains only a frequency value (the
power and dwell time entries do not shift down).
8. Change the still active frequency value in row 5 to 590 MHz: Press 590 > MHz. The power in
row 5 is now the active parameter.
9. Insert a new power value (−2.5 dBm) for point 5, and shift down the original power values for
points 5 and 6 by one row: Press Insert Item > −2.5 > dBm.
10.To complete the entry for point 8, insert a duplicate of the point 7 dwell time by shifting a copy
of the existing value down: Highlight the dwell time in row 7 and press Insert Item.
11.For an analog instrument, go to step 14. For a vector instrument, continue with step 12.
12.Select a waveform for point 2:
a. Highlight the waveform entry for point 2 and press the More > Select Waveform.
The signal generator displays the available waveforms, as shown in the following example.
b. Highlight the desired waveform (in this example, SI NE_TEST) and press either the
Select
hardkey or the Select Waveform softkey.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 51
Basic Operation
The empty entry is equivalent to
choosing CW (no modulation).
Configuring a Swept Output
13.As desired, repeat step 12 for the remaining points for which you want to select a waveform.
The following figure shows an example of how this might look.
14.Turn sweep on:
Press Return > Return > Return > Sweep > Freq Off On > Amptd Off On > Waveform Off On.
15.If it is not already on, turn the RF output on:
Press RF On/Off.
The SWEEP annunciator appears on the display, indicating that the signal generator is
sweeping, and the progress bar shows the progression of the sweep.
If the instrument is in manual sweep (page 53), the active row (row 6 in the
figure above) is the selected (manual) point, and the signal generator
outputs the settings for that selection when the RF output is on.
Example: Using a Single Sweep
1. Set up either a step sweep (page 48) or a list sweep (page 50).
2. In the List/Sweep menu, set the sweep repeat to single:
Press Sweep Repeat Single Cont to highlight Single.
Sweep does not occur until you trigger it.
Note that the WINIT annunciator appears on the display, indicating that the sweep is waiting
to be initiated.
3. If not already on, turn the RF output on: Press RF On/Off.
4. Initiate the sweep: Press Single Sweep.
A single repetition of the configured sweep is available at the RF Output connector.
As the signal generator sweeps, the SWEEP annunciator replaces WINIT on the display, and the
progress bar shows the progression of the sweep.
At the end of the sweep, there is no progress bar, and the WINIT annunciator replaces SWEEP.
52 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Operation
The parameters of the selected sweep point (point 3
of 6 in this example) define the signal av ailab le at the
RF Output connector.
When you enter a manual point,
the progress bar moves to and
stops at the selected point.
The SWMAN annunciator
indicates that the sweep
is in manual mode.
When you turn manual mode on, the
current sweep point becomes the
manual point.
The currently sweeping
parameter stops and
displays the selected point.
Configuring a Swept Output
Example: Manual Control of Sweep
1. Set up either a step sweep (page 48) or a list sweep (page 50).
2. In the Sweep/List menu, select a parameter to sweep: Press Sweep > parameter > Return.
3. Select manual mode: Press More > Manual Mode Off On.
When you select manual mode, the current sweep point becomes the selected manual point.
4. If it is not already on, turn the RF output on: Press RF On/Off.
5. Select the desired point to output: Press Manual Point > number > Enter.
The progress bar changes to indicate the selected point.
6. Use the knob or arrow keys to move from point to point. As you select each point, the RF
output changes to the settings in that selection.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 53
Basic Operation
Annunciator indicates active AM modulation
AM format on
A lit LED indicates that any active
modulation format can modulate
the carrier.
Modulating the Carrier Signal
Modulating the Carrier Signal
To modulate the carrier signal, you must have both
— an active modulation format
and
— modulation of the RF output enabled
Example
1. Preset the signal generator.
2. Turn on AM: Press AM > AM Off On (requires Option UNT).
You can turn on the modulation format before or after setting signal parameters.
The modulation format generates, but does not yet modulate the carrier signal.
Once the signal generates, an annunciator showing the name of the format appears, indicating
that a modulation format is active.
3. Enable modulation of the RF output: Press the Mod On/Off key until the LED lights.
If you enable modulation without an active modulation format, the carrier signal does not
modulate until you subsequently turn on a modulation format.
4. To make the modulated carrier available at the RF output connector, press the RF On/Off key
until the LED lights.
To turn modulation
off, press the Mod On/Off key until the LED turns off.
When the Mod On/Off key is off, the carrier signal is not modulated, even
with an active modulation format.
See also:“Analog Modulation (Option UNT)” on page159
“Pulse Modulation (Options UNW and 320)” on page169
“I/Q Modulation” on page245
54 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Simultaneous Modulation
The Keysight X-Series signal generators are capable of simultaneous
modulation. All modulation types (AM, FM, fM, Pulse, and I/Q) may be
simultaneously enabled, but there are some exceptions. Refer to Table 3-1.
Table 3-1 Simultaneous Modulation Type Combinations
Basic Operation
Modulating the Carrier Signal
a
AM
AM--xxxx
FM
fM
Pulsexxx--x
I/Qxxxx--
c
x
c
x
FMfM
--not
applicable
not
applicable
--xx
b
Pulse
xx
I/Q
a. Linear AM and Exponential AM cannot be enabled simultaneously. Refer to
Chapter 6.
b. Pulse modulation requires Option UNW. Refer to Chapter 7.
c. FM and fM cannot be enabled simultaneously.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 55
Basic Operation
Working with Files
Working with Files
— File Softkeys on page 57
— Viewing a List of Stored Files on page 58
— Storing a File on page 60
— Loading (Recalling) a Stored File on page 61
— Moving a File from One Media to Another on page 62
— Working with Instrument State Files on page 62
— Selecting the Default Storage Media on page 67
The signal generator recognizes several types of files, such as instrument state files, license files,
and list sweep files. Files can be stored either in the signal generator’s internal storage or on the
USB media. This section provides an overview of how to navigate the signal generator’s file menus,
and how to view, store, load, and move files.
The Keysight MXG and EXG nonvolatile internal memory is allocated according to a Microsoft
compatible file allocation table (FAT) file system. Refer to the Programming Guide.
See also: Storing, Loading, and Playing a Waveform Segment on page 189.
56 Keysight EXG and MXG X-Series Signal Generators User’s Guide
File Softkeys
Instrument operating parameters (see page62).
Sweep data from the List Mode Values table editor.
User flatness calibration corrections.
Note: Available file ty pes depend on the installed options.
For details on each key, use key help as described on page40.
Deletions require
confirmation.
Available only with USB connected
(i.e. Without external USB connected,
softkeys are greyed out).
page67
This key changes, depending on
the selected file. See
Shows the current director y
page62
page62
Requires
confirmation
The display indicates when the current directory is the default storage path.
For information on setting the default storage path, see page67.
When you connect USB media to the instrument, the si gnal genera tor displays the USB Media menu and the message External USB Storage attached. When you
disconnect the USB media, the message External USB Storage detached displays. When you open the External Media menu without US B m ed i a co nn ecte d, the
signal generator displays the message External Media Not Detected.
The signal generator does not
format USB media or change file
permissions. Use a computer to
perform these operations.
Available only when the
current directory is not the
top level directory.
Display internal or USB
files, depending on the
selected storage media.
page58
Displays IQ Files
Displays FSK
Files
Displays FIR
Files
Displays a list of
Pulse Train files.
Displays Bit
Files
page437
Basic Operation
Working with Files
Keysight EXG and MXG X-Series Signal Generators User’s Guide 57
Basic Operation
For details on each key, use key help as described on page40.
Waveform files and their associated marker
and header information.
Note: Available file type s depend on the installed options.
The selected file catalog and
the storage media
Working with Files
ARB File Softkeys
Viewing a List of Stored Files
The information in this section is provided with the assumption that default storage media is set to
Auto, as described on page 67.
Viewing a List of Files Stored in the Signal Generator
1. If USB media is connected, disconnect it. The signal generator’s storage media switches to
internal, so you can now use the file catalogs to see files stored in the signal generator.
2. Select the desired file catalog: Press File > Catalog Type > desired catalog (in this example,
All). The selected files appear in alphabetical order by file name, as shown in the following
figure.
58 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Operation
Use the Page Up and
Page Down hardkeys to scroll
through the contents of the
directory.
Working with Files
Viewing a list of Files Stored on USB Media
With USB media connected, you can view files on USB media using either the file catalogs, which
can display only a selected type of file, or the USB File Manager, which displays all files.
Using the File Catalogs:
— With the USB media connected, select the desired file catalog: press > Catalog Type >
desired catalog. The selected files appear in alphabetical order by file name.
Using the USB File Manager:
— With USB media connected, open the USB File Manager: press File > More >
USB File Manager.
The instrument displays the default directory on the USB Media, as shown in the following
figure. Note that when you attach USB media, the display goes directly to this menu.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 59
Basic Operation
Use this menu to enter the file name ,
as described on page41.
For details on each key, use key help as described on page40.
File Type
List
State
Waveform
User Flatness
User Preset
Pulse Train
Save From
Sweep menu
Save menu
Mode menu
Amplitude menu
User Preset menu
Pulse Train menu
File Length (including extension)
Internal Media: 25 characters
USB Media: 39 characters
Working with Files
Storing a File
Several menus enable you to store instrument parameters. For example, you can store instrument
states, lists, and waveforms.
— An instrument state file contains instrument settings. For this type of file, use the Save hardkey
shown in Figure 3-8 on page 63.
— For other types of data, use the Load/Store softkey (shown below) that is available through the
menu used to create the file.
60 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Operation
File Type
List
State
Waveform
User Flatness
User Preset
Pulse Train
Recall From
Sweep menu
Save menu
Mode menu
Amplitude menu
User Preset menu
Pulse Train menu
This key changes, depending
on the selected file.
If the signal generator does not recognize the file,
you must select how the file is to be used.
File Type
List
State
Waveform
User Flatness
User Preset
License
Pulse Train
Pressing Select with file highlighted...
loads list and starts sweep
load instrument state
loads and plays waveform
loads and applies user flatness
loads and executes user preset
installs purchased license
loads and applies pulse train
For details on each key, use key help as described on page40.
Copy & Select User Flatness
Copy & Play Waveform
Copy & Play Sequence
Enter Directory
Recall Instrument State
Load List
Load Pulse Train
Install licenses
No action (USB disconnected)
Use As
File > Catalog Type > <type> > More > USB File Manager
or
File > More > USB File Manager
or
Insert the USB media
Working with Files
Loading (Recalling) a Stored File
There are several ways to load (recall) a stored file.
— For an instrument state file, use the Recall hardkey shown in Figure 3-8 on page 63.
— For other types of data, use the Load/Store softkey (shown below) that is available through the
menu used to create the file.
Loading a File From USB Media
To load a file from USB media, use the USB file manager shown below.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 61
Basic Operation
Selecting a waveform or
an unknown file type
displays this softkey.
File > Catalog Type > <type> > More > USB File Manager
or
File > More > USB File Manager
or
Insert the USB Flash Drive (UFD)
For details on each key, use key help as described on page40.
This key changes,
depending on the selected
file. See page61
Whether a menu is available
depends on the selected file.
Requires confirmation
Catalog Type Menu
page57
Requires confirmation
Moves up one directory level
These
softkeys are
These
softkeys
If the file type is recognized
the “Filename.xxx copied
to Internal Storage” is
displayed.
Working with Files
Moving a File from One Media to Another
Use the USB Media Manager to move files between USB and internal media.
Working with Instrument State Files
— Saving an Instrument State on page 64
— Saving a User Preset on page 64
— Recalling an Instrument State on page 64
— Recalling an Instrument State and Associated Waveform File on page 65
— Recalling an Instrument State and Associated List File on page 65
— Moving or Copying a Stored Instrument State on page 66
62 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Figure 3-8 Save and Recall Softkeys
When saved to the signal generator, instrument settings
(states) save to instrument state memory*. Instrument
state memory is divided into 10 sequences (0 through 9);
each sequence comprises 100 registers (00 through 99).
Delete softkeys in the Save and Recall menus enable you
to delete the contents of a specific register, or th e contents
of all sequences in the state file catalog.
The signal generator requires that you confirm a deletion.
*
Caution
T
o avoid the loss of data, GPIB settings, or current user
instrument states that have not been permanently saved
to nonvolatile memory, the instrument should always be
powered down either via the instrument's front panel
power button or the appropriate SCPI command.
instrument's installed in rack systems and powered down
with the system rack power switch, rather than the
instrument's front panel switch display a Error -310 due to
the instrument not being powered down correctly.
For details on each key, use key help
as described on page40.
Table 3-2
Basic Operation
Working with Files
The following information is not stored in a state file:
System Security LevelSweep listsHostnameRemote
Step increment values
Language
System Security Level DisplayPulse Train listsIP AddressFTP ServerARB Files
System Security Level StateDisplay State On/OffSubnet MaskManual DHCPMAC
Web Server (HTTP)FilesDefault
Gateway
VXI–11 SCPIUser Power
Correction
Sockets SCPI (TELNET)I/Q Calibration Data
Keysight EXG and MXG X-Series Signal Generators User’s Guide 63
Basic Operation
Working with Files
Saving an Instrument State
1. Preset the signal generator and set the following:
• Frequency: 800 MHz• Amplitude: 0 dBm• RF: on
2. (Optional, vector models only) Associate a waveform file with these settings:
a. Press Mode > Dual ARB > Select Waveform.
b. Highlight the desired file and press Select Waveform. If the file is not listed, you must first
move it from internal or external media to BBG media, see page 189.
3. Select the desired memory sequence (for this example, 1): Press Save > Select Seq > 1 >
Enter.
4. Select the desired register (in this example, 01): Press Select Reg > 1 > Save Reg.
If a waveform is currently selected, saving the instrument state also saves a pointer to the
waveform file name.
5. Add a descriptive comment to sequence 1 register 01:
Press Add Comment to Seq[1] Reg[01], enter the comment and press Enter. The comment
appears in the Saved States list when you press Recall. If the instrument state has an
associated waveform, entering the waveform name in the comment makes it easy to identify
which instrument state applies to which waveform.
Saving a User Preset
A user preset is a special type of instrument state file.
1. Preset the signal generator and set as desired.
2. Press User Preset > Save User Preset.
This saves a state file named USER_PRESET, which the signal generator recognizes as
containing user preset information.
You can set up several preset conditions under different names:
1. After you save a user preset, rename it to something other than USER_PRESET (see page 66).
2. Save as many user presets as you wish, renaming the USER_PRESET file each time.
3. Give the desired file the name USER_PRESET.
Recalling an Instrument State
1. Preset the signal generator.
2. Press Recall.
The Select Seq softkey shows the last sequence used, and the display lists any states stored
in the registers in that sequence; RECALL Reg is the active entry.
3. Select the desired instrument state:
64 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Operation
Working with Files
If the desired state is listed in the currently selected sequence, press desired number > Enter.
If not, press Press Select Seq > desired number > Enter > RECALL Reg > desired number > Enter.
Recalling an Instrument State and Associated Waveform File
1. Ensure that the desired waveform file exists, and that it is in BBG media (page 189).
If the waveform file is not in BBG media, this procedure generates an error.
Recalling an instrument state with an associated waveform file recalls only the waveform
name. It does not recreate the waveform file if it was deleted, or load the file into BBG media if
it is in internal or USB media.
2. Recall the desired instrument state (see previous example).
3. View the waveform file name recalled with the instrument state: press Mode > Dual ARB.
The name is displayed as the selected waveform.
4. Turn on the waveform file: Press Mode > Dual ARB > ARB Off On.
Recalling an Instrument State and Associated List File
Recalling an instrument state recalls only the list sweep setup. It does not recall the frequency
and/or amplitude values. Because you must load the list file from the file catalog, when you store a
list file, be sure to give it a descriptive name (up to 25 characters).
1. Recall the desired instrument state (see previous example).
2. Load the desired list file:
a. Press Sweep > More > Configure List Sweep > More > Load/Store.
b. Highlight the desired file and press Load From Selected File > Confirm Load From File.
Editing The Comment on an Instrument Comment
Use the following steps to change a comment on an instrument state saved using the Save key.
This is not the file name that appears in the State catalog (which is the file’s memory location).
1. Press Save
2. Highlight the desired register
3. Press Edit Comment In Seq[n] Reg [nn].
4. Press Re–SAVE Seq[n] Reg[nn].
This overwrites previously saved instrument state settings with the new comment.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 65
Basic Operation
SequenceRegister
The signal generator recognizes only the file named USER_PRESET as user preset information (page64).
Working with Files
Moving or Copying a Stored Instrument State
Figure 3-9 Instrument State File Catalog
A user–created state file’s default name is its memory location (sequence and register).
To move the file, rename it to the desired sequence and register; you can not give a file the same
name as an existing file. If you rename a state file to something other than a valid sequence/register
name, the file does not appear in either the Save or Recall menu.
66 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Operation
File > More >
Current
selection
Memory in the instrument is used, and USB media is ignored.
Media attached to the front panel USB connector is used, and no user
data is stored to internal stor age.
This is the recommended selection. When USB media is attached to the
front panel, it is the selected media; when USB media is not attach ed,
internal storage is used.
For details on each key, use key help as described on page40.
Select the directory on the USB media to be used for all file operations
when USB media is attached and enabled.
Working with Files
Selecting the Default Storage Media
You can configure the signal generator to store user files to either the internal storage or to external
USB media. To automatically switch between USB media and internal storage, depending on
whether USB media is attached, select Automatically Use USB Media If Present. To avoid storing
any confidential information in the instrument, select Use Only USB Media. To avoid storing any
confidential information to USB media, select Use Only Internal Storage.
This selection is unaffected by a power cycle or preset.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 67
Basic Operation
Error Number
Error Message
Error Description
(May be truncated on the
−222 Data out of range; value clipped to lower limit.
Indicates that the user has entered a deviation, depth, or internal
source frequency that is beyond the specific limits
Explanation is provided in the Error Message file (see page431)
(This is not displayed on the instrument)
Message number and longer description
new indicates a message generated
since messages were last viewed.
The annunciator indicates an unviewed message.
Error messages appear in the lower left corner
of the display as they occur.
Reading Error Messages
Reading Error Messages
If an error condition occurs, the signal generator reports it to both the front panel display error
queue and the SCPI (remote interface) error queue. These two queues are viewed and managed
separately; for information on the SCPI error queue, refer to the Programming Guide.
CharacteristicFront Panel Display Error Queue
Capacity (#errors)30
Overflow HandlingDrops the oldest error as each new error comes in.
Viewing Entries
Clearing the Queue
Unresolved Errors
Press:
Press: Error > Clear Error Queue(s)
a
Re–reported after queue is cleared.
When the queue is empty (every error in the queue has been read, or the queue is cleared), the
No Errors
following message appears in the queue:
No Error Message(s) in Queue 0 of 0
a. Errors that must be resolved. For example, unlock.
Error Message Format
Error > View Next (or Previous) Error Page
68 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Keysight X-Series Signal Generators
N5171B/72B/73B EXG and N5181B/82B/83B MXG
User’s Guide
4Optimize Performance
Before using this information, you should be familiar with the basic operation of the signal
generator. If you are not comfortable with functions such as setting the power level and frequency,
refer to Chapter 3, “Basic Operation”, on page 39 and familiarize yourself with the information in
that chapter.
— Using the Dual Power Meter Display on page 70
— Using the USB Pass Through Commands on page 77
— Using the Power Meter Servo on page 79
— Using Flatness Correction on page 82
— Using Internal Channel Correction (N5172B/82B Only) on page 91
— Using External Leveling (N5173B/83B Only) on page 95
— Using Unleveled Operating Modes on page 102
— Using an Output Offset, Reference, or Multiplier on page 106
— Using Free Run, Step Dwell, and Timer Trigger on page 110
— Using a USB Keyboard on page 112
69
Optimize Performance
Once turned on, the
power meter
readings are always
visible, even if
another instrument
feature is selected
(i.e. Sweep, AM,
etc.).
This area is the main
display for the dual
power meter
calibration frequency
and current power
measured.
If channel A or
channel B power
sensors or both are
displayed, and the
AMPTD or the FREQ
softkeys are pressed,
the large power
meter measurements
remains displayed.
After the first
connection, if the
connection is
successful, the
model number and
serial number of the
power sensor is
displayed.
For details on each key, use key help as described on page40.
See page71
Enables the power sensor on
channel A.
Enables the power sensor on
channel B.
Channel B is configured similarly to
channel A. See page71
For details on each key, use key help
as described on page40.
Using the Dual Power Meter Display
Using the Dual Power Meter Display
The dual power meter display can be used to display the current frequency and power of either one
or two power sensors. The display outputs the current frequency and power measured by the power
sensors in the larger center display and in the upper right corner of the display. Refer to Figure 4-2,
Figure 4-2, and Figure 4-3.
Figure 4-1 Dual Power Meter Display with Power Sensors A and B Calibrated
Figure 4-2 Dual Power Meter Display Menu
70 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Figure 4-3 Configuring the Power Sensor Channels
AUX Fctn > Power Meter
Measurements
Sockets LAN: Sets the
IP port to 5025 (stand ard)
or 5023 (telnet)
programming.
Sets the power meter’s IP
address or LAN–GPIB
gateway’s IP address (Sockets
LAN and VXI–11 LAN only).
Whether a softkey is available
depends on the selected
Connection Type.
VXI–11 LAN: Opens a menu for
entering a device name for the
power meter being used.
Enables the power meter
connection type: Sockets
LAN, VXI–11 LAN, or USB.
Note: The VXI–11 softkey is
used to communicate
remotely with a power meter
that has a GPIB connector
via LAN–GPIB gateway.
USB U2000A S er i es Pow e r
Sensors do not require the
sensor to be calibrated.
The MXG/EXG use the
USB U2000A S er i es Pow e r
Sensor’s “Internal” zero.
Refer to the USB U2000A
Series Power Sensor’s
documentation.
Zero Sensor and Calibrate
Sensor softkeys are
inactive during a
measurement.
Selects the frequency that channel A has
been calibrated to.
Selects the channel on the external
power meter that the power sensor
is connected to.
Enables a power offset in dB for
the current channel.
Enables averaging. Instrument
default is Auto with 1024 points.
Note: This figure illustrates channel A. Channel B is similar.
For details on each key , us e key help as
described on page40.
Optimize Performance
Using the Dual Power Meter Display
Example: Dual Power Meter Calibration
In the following example a U2004A USB Power Sensor is connected to channel A and a N1912A P–
Series Power Meter and 8482A Power Sensor are connected to channel B and are zeroed and
calibrated, as required.
On the signal generator:
Keysight EXG and MXG X-Series Signal Generators User’s Guide 71
Optimize Performance
Using the Dual Power Meter Display
1. Setup for Step Sweep. “Configuring a Swept Output” on page 46.
Verify RF Output power is off before continuing.
2. Connecting the Channel A power sensor: Connect USB sensor to the signal generator. The
MXG/EXG should display a message across the bottom that reads similar to:
USB TMC488 device (USB POWER SENSOR,MY47400143) connected
Figure 4-4 MXG/EXG Displays Connection to U2000 USB Power Sensor
3. Press Aux Fctn > Power Meter Measurements > Channel A Setup > Connection Settings >
Connection Type > USB Device (None) > USB POWER SENSOR (MY47400143)
4. Press Return > Zero Sensor
A diagnostic dialog box appears the initial time that a U2000 Series power sensor with a
different serial number is connected to the signal generator (refer to Figure 4-5). After the
U2000 has been recognized by the signal generator, the U2000 power sensor is saved as a
softkey in the instrument and the dialog box in Figure 4-5 won’t be displayed (press DONE, if
you see this message).
Figure 4-5 Diagnostic Dialog Box for USB Sensor
A Running Calibration(s) bar is displayed on the signal generator. Refer to Figure 4-6 on
page 73.
72 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Figure 4-6 Running Calibration(s) Bar (Zeroing Sensor)
For details on
each key, use key
help as described
on page40.
For details on
each key, use key
help as described
on page40.
For details on
each key, use key
help as described
on page40.
Optimize Performance
Using the Dual Power Meter Display
The U2000 Series USB Power Sensor, does
not require a 50 MHz
calibration. If a calibration is attempted with the U2000 Series Power
Sensors, the signal generator displays a message reading:
The U2000 series power sensor does not require a 50 MHz
calibration. Refer to Figure 4-7 on page 73.
Figure 4-7 Diagnostic Dialog Box Displayed for U2000 Power Sensor
5. Press Return > Return > Channel A to On
The current power meter sensor reading should be displayed in the ChA portion of the
instrument's display and in the upper right portion of the display under Power Meter. Refer to
Figure 4-8.
Figure 4-8 Channel A Power Sensor Displayed on MXG/EXG
Keysight EXG and MXG X-Series Signal Generators User’s Guide 73
Optimize Performance
For details on
each key, use key
help as described
on page40.
Using the Dual Power Meter Display
6. On the N1912A P–Series Power Meter (Channel B power sensor): Connect the N1912A P–
Series Power Meter to the LAN.
7. Connect the power meter sensor to channel B of the power meter.
It is recommended, but not required to use the channel B on the N1912A.
This provides continuity with the signal generator’s dual display. For this
example, the U2004A has already used up the channel A position on the
signal generator.
8. Connect the power sensor input to the 50 MHz reference of the power meter.
9. Press Channel B Setup
10.Press Connection Settings > Sockets
11.Press IP Address > IP address > Enter
The IP address of the power meter should be displayed in the ChB section
of the display.
Figure 4-9 Channel B Power Sensor with IP Address Entered
12.Press Return > Channel Settings > External Power Meter Channel to B.
74 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Optimize Performance
For details on
each key, use key
help as described
on page40.
For details on
each key, use key
help as described
on page40.
Using the Dual Power Meter Display
13.On the signal generator: Press Channel B to On and then back to Off again. This initializes the
signal generator to the external power meter.
14.Press Return > Zero Sensor
A diagnostic dialog box is displayed each time an external power meter is being used and the
Zero Sensor or Calibrate Sensor softkey is pressed (refer to Figure 4-10 on page 75).
Verify the power sensor is connected to the 50 MHz reference of the power meter.
Figure 4-10 Diagnostic Dialog Box for Channel B
15.Press Done
The Running Calibration(s) bar is displayed: Zeroing Sensor Please wait....
16.After Running Calibration(s) bar disappears: Press Calibrate Sensor
Diagnostic Dialog box is displayed that prompts for verifying the connection of the power
sensor to the power meter 50 MHz reference (refer to Figure 4-11 on page 75).
Figure 4-11 Diagnostic Dialog Box for Calibration
Keysight EXG and MXG X-Series Signal Generators User’s Guide 75
Optimize Performance
For details on
each key, use key
help as described
on page40.
For details on
each key, use key
help as described
on page40.
Using the Dual Power Meter Display
17.Press Done
Calibration progress bar is displayed. Refer to Figure 4-12 on page 76.
Figure 4-12 Running Calibration(s) Bar (Calibrating Sensor)
18.Press Return > Channel B to On
19.The current power meter sensor reading should be displayed on the signal generator in the
ChB portion of the display and in the upper right corner of the display under Power Meter and
to the left of the Power Meter power sensor reading.
Figure 4-13 Channel B Power Sensor Displayed on MXG/EXG
20.The power sensors are now ready to be connected in a measurement setup.
76 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Optimize Performance
Using the USB Pass Through Commands
Using the USB Pass Through Commands
The USB pass through SCPI commands are used remotely and or to program your system setup
and power meter sensor setup.
This section applies to the following USB power sensors:
— U2040 X-Series
— U2020 X-Series
— U8480 Series
— U2000 Series
—
If the power sensor is configured to return peak power, then the MXG
displays peak power (i.e., Use the
SCPI commands, which returns a value based on the configuration). Refer
to the X-Series Signal Generators SCPI Command Reference.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 77
a. Enter:
:SYSTem:PMETer:PAS
Sthrough:ENABle?
The instrument should return a 1, indicating that the
pass through feature has been enabled.
Optimize Performance
Using the USB Pass Through Commands
StepSubstepsResults/Notes
4. Sending additional
SCPI commands.
If you are sending a
query, go to step 5.
Else, repeat steps 4 and
or 5 as needed.
5. Sending additional
SCPI queries.
If you are sending a
command, go to step 4.
Else, repeat steps 4 and
or 5as needed.
a. Enter:
:SYSTem:PMETer:PAS
Sthrough "SCPI
command"
a. Enter:
:SYSTem:PMETer:PAS
Sthrough? “*IDN?”
Where "SCPI command" is any SCPI pass through
command. Refer to Table 4-1 on page 77.
The power sensor model and serial number should be
returned.
78 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Optimize Performance
For details on each key, use key help as
described on see page40.
See page71
Sets the allowed
settling error.
The system automatically adjusts
power level according to the power
meter reading.
Once performs the adjustment only at
the end of any transition (amplitude or
frequency change). After the
adjustment is performed and the
power error is corrected, no further
adjustments are performed until the
next transition.
Continuous performs the adjustment
as in Once mode, and continues to
adjust the power periodically if the
value differs by more than the
specified Settling Error.
N5173B/83B only
N5173B/83B
only
Sets the relative initial power
when leveling with the Power
Meter Servo mode. The
default of 0 dB means that the
system starts with the power
it thinks is needed (based on
output power, offset and
reference) but this might lead
to power overshoots in certain
cases (especially when the
offset is not set up correctly).
This value allows to specify
an additional safe-margin to
start the measure/adjust
cycles lower than the target
power to protect the device
from power overshoots.
Using the Power Meter Servo
Using the Power Meter Servo
The Power Meter Servo mode uses power meter readings to adjust the output power of the source,
maintaining a constant DUT output power.
The servo loop measures the output power of the DUT, compares it to the user-provided reference
power, and adjusts the output of the source to achieve the user-provided power level within the
settling error. The servo loop will abort after twenty unsuccessful attempts to achieve the
user-provided power level.
Figure 4-14 Power Meter Servo Menus
Keysight EXG and MXG X-Series Signal Generators User’s Guide 79
Optimize Performance
For details on each key,
use key help as described
on page40.
Power Sensor
Directional Coupler
Using the Power Meter Servo
Power Meter Servo Configuration
The following procedure is a basic configuration for using the signal generator’s Power Meter Servo
mode.
The configuration described below is one possible setup example.
Consider the limits of your DUT and use caution to protect the DUT from
being exposed to too much power.
1. Connect the equipment as shown in Figure 4-15.
Figure 4-15 Power Meter Servo Configuration
2. Set the signal generator frequency and amplitude to the desired settings for your device.
To get the best results and reduce stress on the DUT, the amplitude offset
should be set to match the approximate gain of the device. For example, if
the DUT is an amplifier with 20 dB of gain, set the amplitude offset to 20
dB.
Press AMPTD > More > Amptd Offset, then enter a value.
3. On the signal generator, press AMPTD > Leveling Control > Power Meter Servo Configuration.
4. Press Channel A Setup and configure the power meter. Refer to Configuring the Power Sensor
Channels on page 71.
5. Press Return > Return > Settling Error. Set the Settling Error value.
Power Meter Once performs the adjustment only at the end of any transition (amplitude or
frequency change). After the adjustment is performed and the power is corrected, no further
adjustments are performed until the next transition.
80 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Optimize Performance
Using the Power Meter Servo
Power Meter Continuous performs the adjustment as in Once mode, and continues to adjust
the power periodically if the value differs by more than the specified Settling Error.
Once these parameters are set, the servo loop engages and levels the DUT’s output power.
Example
The following example emphasizes the importance of setting the amplitude offset, as it protects the
DUT from being exposed to too much power.
For this example, the source amplitude offset is 20 dB and the source amplitude is programmed to
25 dBm. The offset is subtracted from the programmed level, making the actual source output
power 5 dBm. If the power meter measures 24.5 dBm, for example, the output power will be
adjusted by 0.5 dBm since the power meter measurement is 0.5 dBm lower than the desired 25
dBm. The new source output power is 5.5 dBm. If the power meter then measures 24.97 dBm, no
further adjustments will be made since the measured value is within half of the settling error of 0.2
dB.
If using the continuous leveling mode, the power meter readings are monitored and the output
power of the source will be adjusted if the measurement drifts outside the specified settling error.
Continuing with this example, if the amplitude offset remained at the default of 0 dB, the output
power would have been 25 dBm, resulting in a power meter measurement of 40 dBm (assuming the
amplifier already went into compression). The algorithm would have reduced the source power by
15 dB, thus outputting 10 dBm and resulting in a measurement of maybe 29.5 dBm. The additional
reduction of another 4.5 dB would have lead to the same outcome as with the 20 dB amplitude
offset (source output power of 5.5 dBm). However, in-between, the DUT was stressed possibly past
its specified operating range.
Keysight EXG and MXG X-Series Signal Generators User’s Guide 81
Optimize Performance
Using Flatness Correction
Using Flatness Correction
User flatness correction allows the digital adjustment of RF output amplitude for up to 1601
sequential linearly or arbitrarily spaced frequency points to compensate for external losses in
cables, switches, or other devices. Using an Keysight N1911A/12A, E4419A/B, or U2000 Series
power meter/sensor to calibrate the measurement system, a table of power level corrections can
automatically be created for frequencies where power level variations or losses occur. Supported
connection types to the power meter/sensor are Sockets LAN, VXI–11 LAN, USB, and GPIB via VXI–
11 LAN using a LAN–GPIB gateway (e.g. E5810A Gateway or equivalent).
A power meter with GPIB requires using the Connection Type VXI–11
softkey, as well as a LAN–GPIB gateway, to control a power meter. Refer to
the Keysight Connectivity Guide USB/LAN/GPIB Connectivity Guide
(E2094–90009), Keysight X-Series FAQs “How do I connect to the LAN?”,
and to the E5810A User’s Guide or equivalent, LAN/GPIB gateway device.
If you do not have an Keysight N1911A/12A or E4419A/B power meter, or U2000A/01A/02A/04A
power sensor, or if your power meter does not have a LAN, GPIB, or USB interface, the correction
values can be manually entered into the signal generator.
To allow different correction arrays for different test setups or different frequency ranges, you may
save individual user flatness correction tables to the signal generator’s memory catalog and recall
them on demand.
Follow the steps in the next sections to create and apply user flatness correction to the signal
generator’s RF output (see page 86).
Afterward, follow the steps in “Recalling and Applying a User Flatness Correction Array” on
page 90 to recall a user flatness file from the memory catalog and apply it to the signal generator’s
RF output.
82 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Figure 4-16 User Flatness Correction Softkeys
Confirm
For details on each key, use key help
as described on page40.
See
page85
Starts the
user
flatness
calibration.
Optimize Performance
Using Flatness Correction
Keysight EXG and MXG X-Series Signal Generators User’s Guide 83
Optimize Performance
Using Flatness Correction
Creating a User Flatness Correction Array
In this example, you will create a user flatness correction array. The flatness correction array
contains ten frequency correction pairs (amplitude correction values for each specified frequency),
from 500 MHz to 1 GHz.
An Keysight N1911A/12A or E4419A/B power meter and E4413A power sensor are used to
measure the RF output amplitude at the specified correction frequencies and transfer the results to
the signal generator. (A U2000 Series power meter/sensor could be used in lieu of the power meter
and E4413A power sensor.) The signal generator reads the power level data from the power meter,
calculates the correction values, and stores the correction pairs in the user flatness correction
array.
If you do not have the required Keysight power meter, or if your power meter does not have a LAN,
GPIB, or USB interface, you can enter correction values manually.
Required Equipment
— Keysight N1911A/12A or E4419A/B power meter (a power meter is not required with the
U2000A/01A/02A/04A Power Sensor)
— Keysight E4413A E Series CW power sensor or U2000A/01A/02A/04A Power Sensor
— GPIB, LAN, or USB interface cables, as required
— adapters and cables, as required
For operating information on a particular power meter/sensor, refer to its
operating guide.
Connect the Equipment
Connect the equipment as shown in “Connect the Equipment” on page 85.
During the process of creating the user flatness correction array, the power
meter is remotely controlled by the signal generator.
84 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.