duced in any form or by any means
(including electronic storage and
retrieval or translation into a foreign
language) without prior agreement and
written consent from Keysight Technologies as governed by United States and
international copyright laws.
Manual Part Number
N4985-90001
Edition
Edition 3, March 17, 2017
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies
Bayan Lepas Free Industrial Zone,
11900 Penang, Malaysia
Technology Licenses
The hard ware and/or software
described in this document are furnished under a license and may be
used or copied only in accordance with
the terms of such license.
Declaration of Conformity
Declarations of Conformity for this
product and for other Keysight products may be downloaded from the
Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest Declaration of Conformity.
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and
Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S.
government acquires commercial computer software under the same terms
by which the software is customarily
provided to the public. Accordingly,
Keysight provides the Software to U.S.
government customers under its standard commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of which can
be found at http://www.keysight.com/
find/sweula. The license set forth in the
EULA represents the exclusive authority
by which the U.S. government may use,
modify, distribute, or disclose the Software. The EULA and the license set
forth therein, does not require or permit, among other things, that Keysight:
(1) Furnish technical information
related to commercial computer software or commercial computer software
documentation that is not customarily
provided to the public; or (2) Relinquish
to, or otherwise provide, the government rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release, perform, display, or disclose commercial
computer software or commercial computer software documentation. No
additional government requirements
beyond those set forth in the EULA
shall apply, except to the extent that
those terms, rights, or licenses are
explicitly required from all providers of
commercial computer software pursuant to the FAR and the DFARS and are
set forth specifically in writing elsewhere in the EULA. Keysight shall be
under no obligation to update, revise or
otherwise modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government
acquires no greater than Limited Rights
as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any
technical data.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER
EXPRESS OR IMPLIED, WITH REGARD
TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT
OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT
AND THE USER HAVE A SEPARATE
WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not proceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
2Keysight N4985A Operating and Service Manual
Compliance Notices and Regulatory Information
Compliance with Electromagnetic Compatibility (EMC)
This product conforms with the protection requirements of EMC Directive 2004/
108/EC for Electromagnetic Compatibility.
This product complies to the EMC Directive by assessment according to the IEC/
EN61326-1:2005 EMC standard.
In order to preserve the EMC performance of this product, any cable which
becomes worn or damaged must be replaced with the same type and
specifications.
Keysight N4985A Operating and Service Manual3
Regulatory markings
The CE mark is a registered trademark of the
European Community. This CE mark shows that
the product complies with all the relevant
European Legal Directives.
The RCM mark is a registered trademark of the
Australian Communications and Media Authority.
The CE mark is a registered
trademark of the European
Community. This CE mark shows
that the product complies with
all the relevant European Legal
Directives.
ICES/NMB-001 indicates that
this ISM device complies with
the
Canadian ICES-001.
Cet appareil ISM est conforme a
la norme NMB-001 du Canada.
ISM GRP.1 Class A indicates that
this is an Industrial Scientific and
Medical Group 1 Class A
product.
This symbol indicates the time
period during which no
hazardous or toxic substance
elements are expected to leak or
deteriorate during normal use.
Forty years is the expected useful
life of the product.
South Korean Class A EMC declaration:
This equipment is Class A suitable for
professional use and is for use in electromagnetic
environments outside of the home.
This instrument complies with
the WEEE Directive (2002/96/
EC) marking requirement. This
affixed product label indicates
that you must not discard this
electrical or electronic product in
domestic household waste.
4Keysight N4985A Operating and Service Manual
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/
96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking
requirement. This affixed product label indicates that you must not discard this
electrical or electronic product in domestic household waste.
Product category
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center,
or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml
for more information.
Sales and Technical Support
To contact Keysight for sales and technical support, refer to the support links on
the following Keysight websites:
– www.keysight.com/find/mta
(product-specific information and support, software and
documentation updates)
– www.keysight.com/find/assist
(worldwide contact information for repair and service)
Table 3-5Supplementary specifications for the N4985A amplifiers
41
Table 3-6Physical specifications for the N4985A amplifiers . . .41
Keysight N4985A Operating and Service Manual11
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
12Keysight N4985A Operating and Service Manual
Keysight N4985A System Amplifiers
Operating and Service Manual
1Introduction
Product Overview 14
This chapter provides an overview of the N4985A system amplifiers.
13
1Introduction
N4985A-S30 amplifier
N4985A-S50 amplifier
Product Overview
High performance broadband RF amplifier
The N4985A-S30 and N4985A-S50 system amplifiers are high performance,
broadband amplifiers featuring baseband RF (<100 kHz) through millimeter wave
frequency coverage. These amplifiers, which have many uses in research,
development, or production-line environment, are a useful addition to the
laboratory where RF, microwave, or millimeter wave measurements are made.
Typical applications include:
– RF source amplifier.
– Mixer LO amplifier.
– Noise figure LNA and noise figure system amplifier.
The N4985A-P15 and N4985A-P25 are high performance, broadband amplifiers
with excellent power and gain up to 50 GHz, which have many uses in research,
development, or production-line environment. These amplifiers are a useful
addition to the laboratory where RF, microwave, or millimeter wave
measurements are made. Typical applications include:
– Driver for high power amplifiers such as TWT or linear power amplifiers.
– Overcome system losses such as long transmission lines, power dividers, etc.
– Antenna research and development; antenna characterization systems.
The N4985A is a high performance amplifier featuring broadband frequency
coverage.
ModelFrequency rangeGain
N4985A-S30100 kHz to 30 GHz30 dB
N4985A-S50100 kHz to 50 GHz30 dB
N4985A-P1510 MHz to 50 GHz25 dB
N4985A-P252 GHz to 50 GHz27 dB
N4985A-S30 and N4985A-S50
– The amplifier is designed to be a multi-use laboratory RF amplifier as a gain
block for frequency domain applications or as a time domain pulse amplifier.
– The amplifier’s small size and versatile performance make it an excellent
selection as a general purpose gain block with moderate power output in a
single package potentially replacing two or three narrower band amplifiers.
N4985A-P15 and N4985A-P25
– The amplifier is designed to be easily used in laboratory and test applications.
– It is small in size and has built-in bias power supplies allowing close
placement to the measurement or source reference plane. Its high gain helps
to compensate for system losses such as long transmission lines and system
components.
16Keysight N4985A Operating and Service Manual
Applications
The N4985A-S30 and N4985A-S50 system amplifiers are an often needed
companion to synthesized signal generators and microwave sources. The small
size of the amplifiers allows close placement to the DUT or test fixture. High gain
and greater than 20 dBm output power will overcome long cable losses from the
signal source and provide the additional output power needed in many
development and test applications.
The N4985A-P15 and N4985A-P25 are test amplifiers designed for broadband
power and gain applications. It is ideally suited for saturated RF amplifier testing,
driving TWT amplifiers, or as an antenna system amplifier. High gain and typically
22 dBm output power at 50 GHz will overcome long cable losses from the signal
source, and will provide additional power required in many development and test
applications.
Introduction1
Keysight N4985A Operating and Service Manual17
1Introduction
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
18Keysight N4985A Operating and Service Manual
Keysight N4985A System Amplifiers
Operating and Service Manual
2Installation
Initial Inspection 20
Service and Recalibration 21
Related Documentation 21
Operating and Safety Precautions 22
This chapter provides you important information on how to check and prepare
your instrument for operation.
19
2Installation
Initial Inspection
1 Unpack and inspect the shipping container and its contents thoroughly to
ensure that nothing was damaged during shipment. If the shipping container
or cushioning material is damaged, the contents should be checked both
mechanically and electrically. Check for mechanical damage such as
scratches or dents. Procedures for checking electrical performance are given
under “Performance verification” on page 46.
2 If the contents are damage or defective, contact your nearest Keysight
Technologies Service and Support office. Refer to “Sales and Technical
Support” on page 5 of this manual. Keysight Technologies will arrange for
repair or replacement of the damaged or defective equipment. Keep the
shipping materials for the carrier’s inspection.
3 If you are returning the instrument under warranty or for service, repackaging
the instrument requires original shipping containers and materials or their
equivalents. Keysight Technologies can provide packaging materials identical
to the original materials. Refer to “Sales and Technical Support” on page 5 of
this manual for the Keysight Technologies nearest to you. Attach a tag
indicating the type of service required, return address, model number, and
serial number. Mark the container FRAGILE to insure careful handling. In any
correspondence, refer to the instrument by model number and serial number.
20Keysight N4985A Operating and Service Manual
Service and Recalibration
If your N4985A amplifiers require service or repair, contact the nearest Keysight
office for information on where to send it. Refer to “Sales and Technical Support”
on page 5 of this manual. The performance of the N4985A can only be verified by
specially-manufactured equipment and calibration standards from Keysight. The
recommended interval for recalibration is 12 months.
Related Documentation
This Operating and Service Manual is shipped with the product. It is also available
at www.keysight.com/find/mta.
Installation2
Keysight N4985A Operating and Service Manual21
2Installation
Operating and Safety
Observe the following guidelines before connecting or operating the N4985A
system amplifiers.
ESD damage
Protection against electrostatic discharge (ESD) is important while handling and
operating the N4985A.
Static electricity can build up on your body and can easily damage sensitive
components when discharged.
Static discharges too small to be felt can cause permanent damage to the unit.
To prevent damage from ESD:
– Use a grounded antistatic mat in front of your test equipment and wear a
grounded wrist strap attached to it when handling or operating the N4985A.
– Wear a heel strap when working in an area with a conductive floor.
– Ground yourself before you clean, inspect, or make a connection to the
N4985A. You can, for example, grasp the grounded outer shell of the analyzer
test port or cable connector briefly.
Precautions
– Avoid touching the exposed connector pins.
Connector care
Because connectors can become defective due to wear during normal use, all
connectors should be inspected and maintained to maximize their service life.
– Inspect the mating surface each time a connection is made. Metal particles
from connector threads often find their way onto the mating surface when a
connection is made or disconnected.
– Clean dirt and contamination from the connector mating surface and threads.
This simple step can extend the service life of the connector and improve the
quality of your calibration and measurements.
– Gage connectors periodically. This not only provides assurance of proper
mechanical tolerances and thus connector performance, but can also indicate
situations where the potential for damage to another connector may exist.
22Keysight N4985A Operating and Service Manual
Installation2
CAUTION
The N4985A can be damaged if excessive torque is applied to the
connectors.
The recommended torque value is 8 lb-in torque for 2.4 mm and 2.92 mm,
5 Ib-in torque for 1.85 mm DCA sampling head.
Keysight N4985A Operating and Service Manual23
2Installation
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
24Keysight N4985A Operating and Service Manual
Keysight N4985A System Amplifiers
Operating and Service Manual
This chapter provides the specifications of the N4985A system amplifiers.
25
3Specifications
General Specifications
Specifications
Specifications refer to the performance standards or limits against which the
N4985A is tested.
Typical characteristics are included for additional information only and they are
not specifications. These are denoted as “typical”, “nominal”, or “approximate” and
are printed in italics.
Table 3-1N4985A-S30 and N4985A-S50 specifications
SpecificationN4985A-S30N4985A-S50
Frequency100 kHz to 30 GHz100 kHz to 50 GHz
Gain (dB)30 dB30 dB
Small signal gain27 (min), 30 (typ) 1 GHz to 26 GHz
Output power (at P
Input match S
Output match S
) (dB)
sat
(dB)
11
(dB)
22
20 (min), 22 (typ)
100 kHz to 26 GHz
–10 (typ) 1 GHz to 26 GHz
–10 (typ) 1 GHz to 26 GHz
Noise figure (dBm)5 (typ) 2 GHz to 30 GHz
2nd harmonics H
(dBc) @P
2
1
–30 (typ) 2 GHz to 25 GHz–30 (typ)2 GHz to 25 GHz
27 (min), 30 (typ)
24 (min), 27 (typ)
20 (min), 22 (typ)
20 (typ)
17 (typ)
–10 (typ)
–8 (typ)
–10 (typ)
–8 (typ)
5 (typ)
6 (typ)
1 GHz to 26 GHz
26 GHz to 45 GHz
100 kHz to 26 GHz
26 GHz to 40 GHz
40 GHz to 50 GHz
1 GHz to 26 GHz
26 GHz to 45 GHz
1 GHz to 26 GHz
26 GHz to 45 GHz
2 GHz to 3 GHz
30 GHz to 40 GHz
26Keysight N4985A Operating and Service Manual
Specifications3
Table 3-2Additional specifications for N4985A-OA3
DescriptionMinimumTypicalMaximum
Output eye amplitude77.5—
S21 (V)
Input eye amplitude—0.5—
S21 (ps)Jitter additive (rms)—< 0.5< 1
Tr/Tf (ps)Rise/fall time—< 8< 10
Table 3-3Additional specifications for N4985A-OA5
DescriptionMinimumTypicalMaximum
Output eye amplitude77.5—
S21 (V)
Input eye amplitude—0.5—
S21 (ps)Jitter additive (rms)—< 0.5< 1
Tr/Tf (ps)Rise/fall time—< 8< 10
Figure 3-1Small signal parameters versus frequency (for N4985A-S30)
Keysight N4985A Operating and Service Manual27
3Specifications
Figure 3-2Gain versus output power (P
indicated with *) [for
1dB
N4985A-S30]
Figure 3-3Saturated output power (for N4985A-S30)
28Keysight N4985A Operating and Service Manual
Figure 3-4N4985A-OA3 (2
Specifications3
31
– 1) EYE: >7 V output (1.5 V/div)
Figure 3-5N4985A-OA3 (2
Keysight N4985A Operating and Service Manual29
31
– 1) EYE: <500 fs added rms jitter (5 ps/div)
3Specifications
Figure 3-6N4985A-OA3 (2
31
– 1) EYE: <10 ps Tf/Tf (5 ps/div)
Figure 3-7N4985A-OA3 (2
30Keysight N4985A Operating and Service Manual
31
– 1) EYE: Input signal from PRBS generator
Specifications3
Figure 3-8Small signal parameters versus frequency (for N4985A-S50)
Figure 3-9Gain versus output power (P
Keysight N4985A Operating and Service Manual31
N4985A-S50]
indicated with *) [for
1dB
3Specifications
Figure 3-10Saturated output power (for N4985A-S50)
Figure 3-11Input pulse = Ch3, 420 m Vppk, 10.4 pS rise time, Output pulse
32Keysight N4985A Operating and Service Manual
= Ch4, 6.5 Vppk, 11.0 pS rise time (for N4985A-S50)
Figure 3-12N4985A-OA5 (2
Specifications3
31
– 1) EYE: >7 V output (1.5 V/div)
Figure 3-13N4985A-OA5 (2
Keysight N4985A Operating and Service Manual33
31
– 1) EYE: <500 fs added rms jitter (5 ps/div)
3Specifications
Figure 3-14N4985A-OA5 (2
31
– 1) EYE: <10 ps Tf/Tf (5 ps/div)
Figure 3-15N4985A-OA5 (2
34Keysight N4985A Operating and Service Manual
31
– 1) EYE: Input signal from PRBS generator
Specifications3
Tab le 3-4N4985A-P15 and N4985A-P25 specifications
Table 3-5Supplementary specifications for the N4985A amplifiers
ModelN4985A-S30N4985A-S50N4985A-P15N4985A-P25
Bias voltage and current9 Vdc at 500 mA9 Vdc at 500 mA9 Vdc at 1.5 A9 Vdc at 1.5 A
RF connectors2.92 mm (f)2.4 mm (f)2.4 mm (f)2.4 mm (f)
Survival input power20 dBm20 dBm26 dBm26 dBm
Operating temperature0 to +55 °C0 to +55 °C0 to +55 °C0 to +55 °C
[a] Including the power supply and the DC power supply cable.
Physical Specifications
Table 3-6Physical specifications for the N4985A amplifiers
ModelN4985A-S30N4985A-S50N4985A-P15N4985A-P25
Net weight0.26 kg0.26 kg1.03 kg1.03 kg
Shipping weight1.8 kg1.8 kg3.2 kg3.2 kg
[a]
Shipping dimensions
– Length90 mm90 mm114 mm114 mm
– Width90 mm90 mm140 mm140 mm
– Height25 mm25 mm70 mm70 mm
Keysight N4985A Operating and Service Manual41
3Specifications
Mechanical Dimensions
Figure 3-24Mechanical dimensions of the N4985A-S30/N4985A-S50
[1]
amplifier
[1] Dimensions in millimeters.
42Keysight N4985A Operating and Service Manual
Specifications3
Figure 3-25Mechanical dimensions of the N4985A-P15/N4985A-P25
amplifier
Keysight N4985A Operating and Service Manual43
3Specifications
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
44Keysight N4985A Operating and Service Manual
Keysight N4985A System Amplifiers
Operating and Service Manual
4Operating Guide
Operating Instructions 46
Service and Maintenance 48
This chapter provides the operating instructions for the N4985A system amplifiers.
45
4Operating Guide
Operating Instructions
Performance verification
The figures below show the setup used to verify the performance of the N4985A
system amplifiers.
A calibrated power splitter (or calibrated coupler) connected to the amplifier input
reference plane and to power sensor A, as shown, establishes the input power
level. The output power is measured on power sensor B. Measurements of gain
versus output power at frequencies of interest establish the amplifier
performance.
Figure 4-1Verification setup with a power splitter
Figure 4-2Verification setup with a calibrated coupler
46Keysight N4985A Operating and Service Manual
Example Application
The N4985A amplifies the RF Source output power to the level needed to
characterize DUT performance.
The small size of the N4985A is easy to place close to the DUT input “Reference Plane”. The high gain of the amplifier easily overcomes the cable losses from the
remote RF Source.
The figure below shows an amplified RF power measurement system using the
N4985A.
Operating Guide4
Figure 4-3Example of an amplified RF power measurement system
The DUT is characterized by setting the power level at the DUT input “Reference
Plane” and measuring the output power at various input power levels for each test
frequency.
The RF Source, used in this example, is a laboratory grade Synthesized Source or
Signal Generator with precise amplitude control (e.g. Keysight 83650L RF
Source). The power levels are measured using a power meter with two
measurement port sensors (A and B) (e.g. Keysight E4419B).
An RF power coupler provides a small, proportional amount of the amplified
power to power sensor A as shown in the above figure.
Keysight N4985A Operating and Service Manual47
4Operating Guide
Service and Maintenance
Service
The N4985A does not have internal adjustments and should not be opened; it
should only be repaired by service-trained personnel. Should it become necessary
to return the N4985A for repair or service, contact your nearest Keysight Sales
and Service Center.
Maintenance
The connectors of the N4985A, particularly the connector faces, must be kept
clean. Keysight recommends that the connectors be periodically inspected and
cleaned if necessary. For instructions on the connection and maintenance of your
connectors, refer to the Connector Care Quick Reference Card (08510- 90360).
48Keysight N4985A Operating and Service Manual
This information is subject to change
without notice. Always refer to the
Keysight website for the latest
revision.