No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation into
a foreign language) without prior agreement
and written consent from Keysight Technologies, Inc. as governed by United States and
international copyright laws.
Manual Part Number
N2797-97000
Edition
Second edition, October 2014
Published by:
Keysight Technologies, Inc.
1400 Fountaingrove Parkway
Santa Rosa, CA, 95403
Warranty
The material contained in this document is
provided “as is,” and is subject to being
changed, without notice, in future ed itions.
Further, to the maximum extent permitted
by applicable law, Keysight disclaims all
warranties, either express or implied, with
regard to this manual and any information
contained herein, incl uding but not l imited
to the implied warranties of merchantability
and fitness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in
connection with the furnishing, use, or performance of this document or of any information contained herein. Should Keysight
and the user have a separate written agreement with warranty terms covering the
material in this document that conflict with
these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
“Commercial computer software” as defined
in DFAR 252.227-7014 (June 1995), or as a
“commercial item” as defined in FAR 2.101(a)
or as “Restricted computer software” as
defined in FAR 52.227-19 (June 1987) or any
equivalent agency regulation or contract
clause. Use, duplication or d isclosure of
Software is subject to Keysight Technologies’
standard commercial license terms, and
non-DOD Departments and Agencies of the
U.S. Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could resul t in personal
injury or death. Do not proceed
beyond a WARNING notice until the
indicated conditions are fully
understood and met.
2N2797A User’s Guide
Contents
1Using the Probe/5
Probe Features / 7
Using the Accessory Tips and Grounds / 14
Dynamic Range and Offset Voltage Limitations / 20
Functional Test / 21
Inspecting the Probe / 22
Cleaning the Probe / 22
Returning the Probe for Service / 23
Safety Information / 24
2Probe Calibration/27
Calibrating the Probe on Infiniium Oscilloscopes / 28
Calibrating the Probe on InfiniiVision Oscilloscopes / 30
3Characteristics and Specifications / 31
Dimensions / 33
4Performance Data Plots/35
5Performance Verification / 37
Test 1. DC Input Resistance / 39
Test 2. Bandwidth / 40
Performance Test Record / 46
Index
N2797A User’s Guide3
Contents
4N2797A User’s Guide
Keysight N2797A
WARNING
User’s Guide
1Using the Probe
Probe Features 7
Using the Accessory Tips and Grounds 14
Dynamic Range and Offset Voltage Limitations 20
Functional Test 21
Inspecting the Probe 22
Cleaning the Probe 22
Returning the Probe for Service 23
Safety Information 24
The N2797A single-ended active probe is designed to be used in extreme temperatures
environments. With a 2 meter cable and ability to operate in temperatures from –40°C to
+85°C, the N2797A probe is ideal for environmental labs. The probe’s bandwidth is 1.5 GHz.
The N2797A integrates many of the characteristics needed for today’s general-purpose,
high-speed probing, especially in digital system design, component
design/characterization, and educational research applications. Its 1 M
and extremely low input capacitance (1 pF) provide ultra low loading of the device under
test. This, accompanied with superior signal fidelity, makes this probe useful for most of
today’s digital logic voltages. With a wide ±8V dynamic range and ±12V offset range, this
probe can be used in a wide variety of applications.
Ω input resistance
For high signal integrity probing, the N2797A 1.5 GHz active probe is a perfect
complements to Keysight’s 500 MHz – 600 MHz and 1 GHz bandwidth scopes.
The probe is powered directly by the InfiniiVision and Infiniium AutoProbe interface,
eliminating the need for an additional power supply. The probe also come with a number of
accessories that allow for easy connections to the circuit under test. Refer to “Supplied
Accessories" on page 8.
Handle the probe with care to avoid injury, especially when it is fitted with the sharp spring or rigid tip.
5
1Using the Probe
CAUTION
CAUTION
CAUTION
This probe is an ESD sensitive devices, particularly at the probe tips. Follow standard ESD precautions
when handling. Remove tip accessories when storing the probe.
Before using the probe, refer to “Safety Information" on page 24.
Handle the probe cable carefully to avoid damaging it through excessive bending or pulling. Avoid any
mechanical shocks to the probe in order to guarantee accurate performance and protection.
6N2797A User’s Guide
Probe Features
NOTE
Oscilloscope Compatibility
The N2797A probe is compatible with the Keysight oscilloscopes shown in Tab le 1 . Up to
four probes can be connected to the oscilloscope. The table also lists the minimum required
firmware version for the oscilloscope.
The N2797A probe is designed for oscilloscopes with 1 MW AutoProbe-interface channel
inputs.
Is Your Oscilloscope Software Up-to-Date? Keysight periodically releases software updates
to support your probe, fix known defects, and incorporate product enhancements. To
download the latest firmware, go to www.keysight.com and search for your oscilloscope’s
topic. Click on the “Drivers, Firmware & Software” tab.
Table 1Compatible Oscilloscopes and Support
Using the Probe1
OscilloscopeAdapter Required
Infiniium Oscilloscopes (firmware version ¡3.50 or above)
90000 Q-SeriesN5442A
90000 X-SeriesN5442A
90000A—
9000 H-Series—
9000A-Series—
InfiniiVision Oscilloscopes (firmware version ¡2.11 or above)
4000 X-Series—
3000 X-Series—
N2797A User’s Guide7
1Using the Probe
CAUTION
Probe Headlight
The probe is equipped with a white LED headlights to illuminate the area around the probe
tip. This headlight has a fixed intensity and is turned on and off by pressing the button that
is located on the probe’s body.
Figure 1Headlights and Button
Channel-Identification Rings
When multiple probes are connected to the oscilloscope, the channel identification rings
(yellow, green, blue, and purple) allow you to quickly determine which channel input is
associated with each probe. On the probe, place one colored ring near the probe’s channel
connector and place an identical color ring near the probe head.
Supplied Accessories
The probe comes with the accessories shown in Figure 2.
All of the supplied accessories can be used at the probe’s specified extreme temperature range of –40°C
to +85°C (–40°F to +185°F) except the flex nose clip and pico hook tips. These adapters have a slightly
reduced temperature range of –20°C to +80°C (–4°F to +176°F).
8N2797A User’s Guide
Using the Probe1
Figure 2Accessories Supplied With the Probe (Not to Scale)
N2797A User’s Guide9
1Using the Probe
Table 2Supplied Accessories
Accessory
Ambient Operating
TemperatureQuantity
Flex Nose Clip (red)–20°C to +80°C
Flex Nose Clip (black)2
Pico Hook (red)1
Pico Hook (black)1
Rigid Probe TipSame as probe
Solderable Tip (0.97 mm square)10
Ground Blade2
Y-Lead Adapter, long with damping resistor (9 cm)2
Y-Lead Adapter, short with damping resistor (6 cm)2
Right-Angle Ground (5 cm)2
(–4°F to +176°F)
–20°C to +85°C
(–4°F to +185°F)
2
5
Available Accessories
Tabl e 3 shows the replacement accessories available in the N2798A replacement accessory
kit. The table lists the ambient operating temperatures for each accessory. Those
accessories that have the same temperature range as the probe are marked
“extreme temp”.
Tabl e 4 shows additional accessories that work with the probe, but are not provided with
the probe. Notice that most of these accessories can not be used at extreme temperature
range of the probe.
10N2797A User’s Guide
Table 3Replacement Accessory Kit (N2798A)
Using the Probe1
ItemDescription
Solderable Tip, extreme tempSame as probe
Ground Blade, extreme temp25061-7393
Rigid Probe Tip, extreme temp55061-7389
Flex Nose Clips
This adapter has a slightly reduced
temperature range compared to the probe.
Pico Hook Nose Clips
This adapter has a slightly reduced
temperature range compared to the probe.
Long Y-Lead Adapter with damping resistor
(10 cm), extreme temp
Short Y-Lead Adapter with damping resistor
(5 cm), extreme temp
Ambient Operating
Temperature
–40°C to +85°C
(–40°F to +185°F)
–20°C to +80°C
(–4°F to +176°F)
–20°C to +80°C
(–4°F to +176°F)
Same as probe
–20°C to +85°C
(–4°F to +185°F)
Kit Part
Qty
105061-7385
2 (red)
2 (black)
1 (red)
1 (black)
25061-7386
25061-7387
Number
5061-7390 (red)
5061-7391 (black)
—
Right Angle Ground (5 cm), extreme tempSame as probe
–40°C to +85°C
(–40°F to +185°F)
25061-7384
N2797A User’s Guide11
1Using the Probe
Table 4Additional Available Accessories
ItemDescription
Spring Probe Tip, extreme temp25061-7388
Long Y-Lead Adapter(10 cm)0°C to 50°C
Right Angle Ground (10 cm)0°C to 50°C
Right Angle Ground (5 cm)25061-7399
Ground Lead (6 cm)0°C to 50°C
Ground Lead (12 cm)25061-7396
Offset Ground0°C to 50°C
N2787A 3D Probe PositionerNot Specified——
Ambient Operating
TemperatureQty
15061-7392
(32°F to 122°F)
25061-7400
(32°F to 122°F)
25061-7395
(32°F to 122°F)
25061-7394
(32°F to 122°F)
Kit Part
Number
12N2797A User’s Guide
Attaching Probe Accessories
It’s easy to change the probe tip and ground accessories. Figure 3 shows locations for
inserting the tip and the ground accessory. To change an accessory, gently pull it straight
out of its contact socket along the axis of the probe. Once the accessory is removed, insert
the new accessory until it is seated in the contact socket. In order to insert the probe tip
completely into the housing, carefully press the probe tip against a hard surface.
Figure 3Tip and Grounding Connections
Using the Probe1
N2797A User’s Guide13
1Using the Probe
Using the Accessory Tips and Grounds
This section presents the various probe accessory configurations listed in order of available
bandwidth for your measurements. Select the configuration to match your measurement
needs, including ease of use.
Table 5Probe Configuration in Order of Decreasing Bandwidth
Recommended
Order of UseProbe Configuration Topic
1“Rigid or Solderable Tip with Ground Blade" on page 151.5 GHz
2“6 cm Y-Leads with 0.1” Headers" on page 161.0 GHz
3“Solderable Tip with Right-Angle Ground" on page 17800 MHz
4 “9 cm Y-Leads with 0.1” Headers" on page 18800 MHz
5“Y-Lead with Flexible Nose Clips and
Pico Hooks" on page 19
Typical
Band wid th
500 MHz
14N2797A User’s Guide
Rigid or Solderable Tip with Ground Blade
WARNING
The combination of rigid or solderable tip and ground blade provide the highest
bandwidth of any combination of accessory tips and grounds. Use this
configuration at room temperatures where hand probing is possible or at extreme
temperatures when using a probe positioner that can withstand the required
temperature. The bandwidth with accessory is 1.5 GHz.
Handle the probe with care to avoid injury, especially when it is fitted with the sharp rigid or spring tip.
Using the Probe1
N2797A User’s Guide15
When inserting the tip, insert the end with the ridge into the probe head as shown in
Figure 4.
Figure 4Inserting the Rigid Tip
1Using the Probe
6 cm Y-Leads with 0.1” Headers
The short 6 cm long Y-leads offer a convenient and reliable
method to connect both the probe signal and ground to probe
points on the board via or the solderable pins that are provided
with each probe. The sockets accept 0.56 to 0.86 mm round and
0.64 mm square (0.022 to 0.034 inch round and 0.025 inch
square) diameter pins. The maximum insertion depth is 7.5 mm
(0.295 inch) and the outer insulation of the socket is less than
2.54 mm. You can use it for 2.54 mm pitch (0.1 inch) terminals. The bandwidth with
accessory is 1.0 GHz.
Figure 5Probe with Y-Leads Attached
16N2797A User’s Guide
Solderable Tip with Right-Angle Ground
NOTE
The solderable pin and right-angle ground provide hands free probing at
extreme temperatures. Solder the tip on the printed-circuit board as shown in
Figure 6 and then slip the probe onto the free end of the solderable pin. The
bandwidth with accessory is 800 MHz.
The longer right-angle ground lead introduces increased inductance in the ground path.
Therefore, keep the right-angle ground lead as straight as possible for maximum
measurement bandwidth. Loops and bends in the ground lead, as shown in this picture, can
reduce the measurement band wid th.
Using the Probe1
Figure 6Placing the Probe Onto the Solderable Tip
N2797A User’s Guide17
1Using the Probe
9 cm Y-Leads with 0.1” Headers
Like the 6 cm Y-leads, the 9 cm Y-leads offer a convenient and
reliable method to connect both the probe signal and ground to
probe points on the board via or the solderable pins that are provided
with each probe. The added convenience of the longer leads results
in a slightly reduced bandwidth. The bandwidth with accessory is
800 MHz.
The sockets accept 0.56 to 0.86 mm round and 0.64 mm square (0.022 to 0.034 inch round
and 0.025 inch square) diameter pins. The maximum insertion depth is 7.5 mm (0.295 inch)
and the outer insulation of the socket is less than 2.54 mm. You can use it for 2.54 mm pitch
(0.1 inch) terminals.
Figure 7Probe with Y-Leads Attached
18N2797A User’s Guide
Y-Lead with Flexible Nose Clips and Pico Hooks
CAUTION
You can use flexible nose clips and Pico hook clips to make connections to components or
wires with leads that are 1.01 mm (0.04 inches) in diameter or smaller. With today’s
miniature IC- and component-packaging techniques, these clips can make probing
challenging devices much easier. The bandwidth with accessory is 500 MHz.
Figure 8Flexible Nose and Pico Hook Clips
Most users prefer to attach the clips to the probe via the Y-lead adapter as shown in the
picture below. Press the back of the clip to extend the grasping hook and then release to
tighten the hook around the component you are testing.
Using the Probe1
The clips have an ambient operating temperature range of –20°C to +80°C (–4°F to +176°F). This is a
slightly reduced range compared to the probe.
Figure 9Probe with Flexible Nose Clips Attached
N2797A User’s Guide19
1Using the Probe
Dynamic Range and Offset Voltage Limitations
With the 16Vpp (±8V) of dynamic range and ±12V of offset range, the probes have a wide
measuring range of ±20V, which allows them to be used for a wide variety of applications.
For example, if you wanted to measure small AC noise riding on top of a +10V DC signal,
you could use DC offset to position the signal at the center of the screen to keep the probe
within its dynamic range. You could then zoom in to see the detail at a higher resolution.
Trying to measure a signal out of the dynamic range may result in a clipped or distorted
waveform.
Figure 10Dynamic Range and Offset Limits
20N2797A User’s Guide
Functional Test
Use the following generic measurement procedure to ensure that your probe is functioning
properly.
1Connect the probe to an oscilloscope channel input and ensure the input impedance of the
oscilloscope matches the output impedance of the probe (50W).
2Connect a grounding accessory to the probe and connect the ground to the ground
terminal on the oscilloscope.
3Connect the probe tip to the oscilloscope’s probe compensation output.
4Enable autoscale on the oscilloscope.
5A square wave should now be displayed on the oscilloscope, if the probe is functioning
properly.
Using the Probe1
N2797A User’s Guide21
1Using the Probe
Inspecting the Probe
• Inspect the shipping container for damage.
Keep the damaged shipping container or cushioning material until the contents of the
shipment have been checked for completeness and the probe has been checked
mechanically and electrically.
• Check the accessories.
• If the contents are incomplete or damaged, notify your Keysight Technologies Sales
Office.
• Inspect the probe.
If there is mechanical damage or defect, or if the probe does not operate properly or
pass calibration tests, notify your Keysight Technologies Sales Office.
If the shipping container is damaged, or the cushioning materials show signs of stress,
notify the carrier as well as your Keysight Technologies Sales Office. Keep the shipping
materials for the carrier’s inspection. The Keysight Technologies office will arrange for
repair or replacement at Keysight Technologies’ option without waiting for claim
settlement.
Cleaning the Probe
Disconnect the probe from the oscilloscope and clean the probe with a soft cloth dampened
with a mild soap and water solution. Make sure the probe is completely dry before
reconnecting it to an oscilloscope.
22N2797A User’s Guide
Returning the Probe for Service
NOTE
NOTE
If the probe is found to be defective we recommend sending it to an authorized service
center for all repair and calibration needs. Perform the following steps before shipping the
probe back to Keysight Technologies for service.
1Contact your nearest Keysight sales office for information on obtaining an RMA number and
return address.
2Write the following information on a tag and attach it to the malfunctioning equipment.
• Name and address of owner
• Product model number (for example, N2797A)
• Product Serial Number (for example, MYXXXXXXXX)
• Description of failure or service required
Include probing and browsing tips if you feel the probe is not meeting performance
specifications or a yearly calibration is requested.
3Protect the probe by wrapping in plastic or heavy paper.
4Pack the probe in the original carrying case or if not available use bubble wrap or packing
peanuts.
5Place securely in sealed shipping container and mark container as "FRAGILE".
Using the Probe1
If any correspondence is required, refer to the product by serial number and model number.
Contacting Keysight Technologies
For technical assistance, contact your local Keysight Call Center.
• In the Americas, call 1 (800) 829-4444
• In other regions, visit http://www.keysight.com/find/assist
Before returning an instrument for service, you must first call the Call Center at 1 (800)
829-4444.
N2797A User’s Guide23
1Using the Probe
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
Safety Information
To avoid personal injury and to prevent fire or damage to this product or products
connected to it, review and comply with the following safety precautions. Be aware that if
you use this probe assembly in a manner not specified, the protection this product provides
may be impaired.
Handle the probe with care to avoid injury, especially when it is fitted with the sharp spring or rigid tip.
Handle Probe Tips and Accessories Carefully. Some of the probe tips and accessories are very sharp
(the browser tips, for example). You should handle these with care to avoid personal injury.
Use Only Grounded Instruments. Do not connect the probe’s ground lead to a potential other than earth
ground. Always make sure the probe and the oscilloscope are grounded properly.
24N2797A User’s Guide
Connect and Disconnect Properly. Connect the probe to the oscilloscope and connect the ground lead
to earth ground before connecting the probe to the circuit under test. Disconnect the probe input and
the probe ground lead from the circuit under test before disconnecting the probe from the
oscilloscope.
Observe Probe Ratings. Do not apply any electrical potential to the probe input which exceeds the
maximum rating of the probe. Make sure to comply with the voltage versus frequency derating curve
found in this manual.
Keep Away From Live Circuits. Avoid open circuitry. Do not touch connections or components when
power is present.
Using the Probe1
WARNING
WARNING
CAUTION
Indoor Use Only. Do not operate in wet/damp environments. Keep product surfaces dry and clean.
Do Not Operate With Suspected Failures. Refer to qualified service personnel.
The probe cable is a sensitive part of the probe and, therefore, you should be careful not to damage it
through excessive bending or pulling. Avoid any mechanical shocks to this product in order to guarantee
accurate performance and protection.
N2797A User’s Guide25
1Using the Probe
26N2797A User’s Guide
Keysight N2797A
CAUTION
User’s Guide
2Probe Calibration
Calibrating the Probe on Infiniium Oscilloscopes 28
Calibrating the Probe on InfiniiVision Oscilloscopes 30
Always calibrate the probe before making any critical measurements. A probe calibration
removes attenuation errors, offset errors, and timing delays that are introduced by the
probe. This chapter contains basic calibration procedures for both Infiniium and InfiniiVision
oscilloscopes. For additional information on the probe calibration refer to the oscilloscope’s
user documentation.
Always wear an ESD wrist strap when working with active probes. Not doing so can result in the probe
becoming permanently damaged.
27
2Probe Calibration
Calibrating the Probe on Infiniium Oscilloscopes
Probe calibrations vary slightly between series 9000, 90000, and 90000 X oscilloscopes. For
9000 series oscilloscopes, the following calibrations are used:
• Skew Calibration
• Attenuation/Offset Calibration
For 90000 series oscilloscopes, the following calibrations are used:
• DC Attenuation/Offset Calibration
• Skew Calibration
When performing a the skew calibration, the E2655B Performance Verification (PV) and
deskew fixture is used. The PV fixture is included with the Deskew Kit which is shipped with
Infiniium oscilloscopes. The fixture is also used in the performance verification procedure
that is documented in Chapter 5, “Performance Verification".
An optional N2787A 3D Probe Positioner or equivalent can be used to hold
the probe during the calibration.
Procedure
1Turn on the oscilloscope. Allow the oscilloscope and probe to warm up for 20 minutes
before performing the probe calibrations.
2If the oscilloscope needs calibration, perform a user calibration before the probe
calibration. On the oscilloscope, click Utilities > Calibration.
3On the oscilloscope, click Setup > Probes.
4In the Probe Calibration dialog box, select the tab representing the channel that has the
probe attached.
5In the dialog box, select the probe head and the type of calibration. Click Start and follow
the instructions shown on the oscilloscope.
6When performing the skew calibration, you will be instructed to use the PV fixture. Observe
these points:
• Connect the PV fixture’s input to the oscilloscope’s calibration output as shown in
Figure 11 on page 29. Connect a 50 ohm terminator to the PV’s output.
28N2797A User’s Guide
Probe Calibration2
CAUTION
CAUTION
NOTE
NOTE
Figure 11E2655B PV Fixture with 50 Ohm Termination
• Position the probe so that the signal lead touches the PV fixture’s center conductor next
to the dialectric area. The ground lead should touch the PV fixture’s ground plane. On
90000 X-series oscilloscopes, use an N5442A adapter when instructed to connect the
probe to one of the oscilloscope’s channel inputs.
To avoid damaging the oscilloscope’s Cal Out connector, do not apply force to the PV fixture. Light probe
contact is all that is needed for the calibration.
N2797A User’s Guide29
NEVER solder a probe tip to the thick-film gold. The gold will immediately dissolve into the solder and
disappear.
If you are using the browser tip, it is recommended that you use the N2787A 3D Probe
Positioner to hold the probe in place.
You can check that the probe leads are correctly connected by pressing the front panel
autoscale button. A stable step should be shown on the screen. Pressing autoscale will close
the Probe Calibration dialog box.
2Probe Calibration
NOTE
Calibrating the Probe on InfiniiVision Oscilloscopes
Allow the oscilloscope and probe to warm up for 20 minutes before performing the probe
calibration.
Procedure
1Connect the probe to one of the oscilloscope channels.
When you connect the probe, the Calibrate Probe softkey in the Channel Probe Menu
becomes active.
2Connect the probe to the calibration terminals. On 3000 X-series oscilloscopes, connect
the probe’s signal tip to the front-panel Demo 2 / Probe Comp terminal and the ground tip to
the ground terminal.
Figure 12Probe Calibration Terminals on 3000 X-Series Oscilloscope
3Press the Channel on/off key to turn the channel on (if the channel is off).
4In the Channel Menu, press the Probe softkey.
5Press the Calibrate Probe softkey and follow the instructions on the display.
30N2797A User’s Guide
Keysight N2797A
User’s Guide
3Characteristics and
Specifications
This chapter provides the characteristics and specifications for the N2797A active probe.
The probe and oscilloscope should be warmed up for at least 20 minutes before any testing
and the environmental conditions should not exceed the probe’s specified limits.
Table 6Mechanical Characteristics
ParameterCharacteristic
Probe Weight108g (approximate)
Cable Length2m
Table 7Environmental Specifications
ParameterCharacteristic
Ambient Operating Temperature–40°C to +85°C (–40°F to +185°F)
Ambient Non-Operating Temperature–40°C to +85°C (–40°F to +185°F)
Operating Humidity95% RH at +40°C
Non-Operating Humidity95% RH at +65°C
Operating Altitude4,000m (13,000 ft)
ESD8 kV HBM
31
3Characteristics and Specifications
Table 8Electrical Characteristics and Specifications
ParameterCharacteristic or Specification (where indicated)
Bandwidth (–3 dB)
Rise Time (10% – 90%, calculated)233 ps
System Bandwidth (with Keysight oscilloscope)1 GHz (with Keysight’s 1 GHz InfiniiVision or Infiniium
Flatness Over Temperature (–40° C to +85° C)Typical 0.3 dB (10 Hz – 100 kHz)
Input Resistance
a
1.5 GHz (specification)
oscilloscope)
Typical 0.8 dB (100 MHz – 1 GHz)
Typical 2.0 dB (1 GHz – 1.5 GHz)
Typical 0.6 dB (100 kHz – 100 MHz)
Typical 0.8 dB (100 MHz – 500 MHz)
Typical 2.0 dB (500 MHz – 1 GHz)
Typical 2.5 dB (1 GHz – 1.5 GHz)
a
1 MΩ ± 3% (specification)
Input Capacitance1 pF
Probe Noise< 2.5 mV
(referred to input)
rms
Output Impedance50Ω
Internal PowerKeysight AutoProbe interface from oscilloscope
(InfiniiVision and Infiniium)
a Denotes warranted electrical specification at 25 °C room temperature after 20 minute warm-up. All others are typical.
32N2797A User’s Guide
WEEE Compliance
Dimensions
Characteristics and Specifications3
This symbol indicates the Environmental Protection Use Period (EPUP) for the
product’s toxic substances for the China RoHS requirements.
The CE mark is a registered trademark of the European Community. ISM GRP 1-A
denotes the instrument is an Industrial Scientific and Medical Group 1 Class A
product. ICES/NMB-001 indicates product compliance with the Canadian
Interference-Causing Equipment Standard.
This product complies with the WEEE Directive (2002/96/EC) marking
requirements. The affixed label indicates that you must not discard this
electrical/electronic product in domestic household waste.
Product Category: With reference to the equipment types in the WEEE
Directive Annex I, this product is classed as a "Monitoring and Control
Instrumentation" product. Do not dispose in domestic household. To
return unwanted products, contact your local Keysight office, or refer to
www.keysight.com for more information.
N2797A User’s Guide33
3Characteristics and Specifications
34N2797A User’s Guide
Keysight N2797A
User’s Guide
4Performance Data Plots
Figure 13 shows the frequency response for the spring or rigid probe tip and an offset
ground or ground blade.
Figure 13Frequency Response (Vout/Vin)
35
4Performance Data Plots
Figure 14Input Impedance Equivalent Model
Figure 15Typical Input Impedance Plot
36N2797A User’s Guide
Keysight N2797A
CAUTION
User’s Guide
5Performance Verification
Test 1. DC Input Resistance 39
Test 2. Bandwidth 40
Performance Test Record 46
These procedures are used to test the warranted specifications for the N2797A. The
recommended calibration test interval is once a year or as required. Use the equipment
listed in Tabl e 9 on page 38.
Electrostatic discharge (ESD) can quickly and imperceptibly damage or destroy high performance
probes, resulting in costly repairs. Always wear a wrist strap when handling probe components and
insure that cables are discharged before being connected.
37
5Performance Verification
Table 9Required Test Equipment
Test EquipmentCritical SpecificationModel Number
Digital Multimeter (DMM)Resistance ±1%Keysight 34401A
Vector Network Analyzer (VNA)13 GHz sweep range full 2 port cal Option 1D5Keysight 8720ES
Outside thread 3.5 mm (male) to 3.5 mm
(female) adapter
Cable (2)3.5 mil; SMA; High QualityKeysight 8120-4948
Cable1.5 mil Probe Power Extension No SubstituteKeysight 01143-61602
PV/DS Test BoardNo Substitute (In E2655B Kit)Keysight E2655-66503
No SubstituteKeysight 5062-1247
38N2797A User’s Guide
Test 1. DC Input Resistance
Procedure
1Connect the DMM probes between the probe tip and ground at the tip of the probe.
2Set up the DMM to measure resistance. The resistance should read 1 MW ± 3%.
3Record the resistance in Table 10 on page 46.
Performance Verification5
N2797A User’s Guide39
5Performance Verification
NOTE
Test 2. Bandwidth
Allow the probe to warm up for at least 20 minutes.
Using the 8720ES VNA successfully
Observe these simple guidelines when working with the 8720ES VNA during this procedure.
• Sometimes it may take a few seconds for the waveforms to settle completely. Allow time
for waveforms to settle before continuing.
• Make sure all connections are tight and secure. If needed, use a vise to hold the cables
and test board stable while making measurements.
• Be careful not to cross thread or force any connectors. This could be a very costly error
to correct.
Procedure
Initial Setup
1Turn on the 8720ES VNA and let warm up for 20 minutes.
2Press the green [Preset] key on the 8720ES VNA.
3On the VNA, press the [Power] key and set the power to 0 dBm.
4On the VNA, press the [AVG] key and then select the Averaging Factor screen key. Set
averaging to 4.
5On the VNA, press the [Sweep Setup] key and then press the sweep type menu screen key.
Select the log freq screen key.
6Connect the probe under test to the AutoProbe Adapter and power the probe using the
1143A power supply as shown in Figure 16 on page 41. Install the outside thread adapter to
the AutoProbe Adapter.
40N2797A User’s Guide
Figure 16Probe Connected to Probe Adapter (not to scale)
Calibrating a Reference Plane
Performance Verification5
To get a reliable measurement from the VNA you must calibrate a reference plane so that
the VNA knows where the probe under test is located along the transmission line.
7On the VNA, press the [Cal] key.
8Press the cal menu screen key.
9Press the full 2 port screen key.
10Connect the PV fixture to the Port 1 on the VNA using a high quality SMA cables. Connect
the cable to the pincher side of PV fixture as shown in Figure 17 on page 42.
11The calibration reference plane is at the other end of PV fixture.
N2797A User’s Guide41
5Performance Verification
Figure 17PV Fixture Connected to VNA
12Perform a Calibration for the PORT 1 side of the Reference plane.
a Press the reflection screen key.
b Connect the open end of 85052D Calibration Standard to the non-pincher side of the
PV/DS test board.
c Select the open screen key under the Forward group.
d Wait until the VNA beeps indicating that it has completed the task.
e Connect short end of Calibration Standard to the non-pincher side of the PV/DS test
board.
f Select short screen key under the Forward group.
g Wait until the VNA beeps indicating that it has completed the task.
h Connect load end of Calibration Standard to the non-pincher side of the PV/DS test
board.
iSelect the loads screen key under the Forward group.
jPress broadband screen key selection.
k Wait until the VNA beeps indicating that it has completed the task.
l Press the done loads screen key.
m You have just calibrated one side of the reference plane.
13Connect the other high quality SMA cable to the VNA’s PORT 2 as shown in Figure 18.
42N2797A User’s Guide
Performance Verification5
Figure 18SMA Cable Connected to Port 2
14Get the opposite sex of the Calibration Standards for the next step.
15Perform Calibration for the PORT 2 side of the Reference plane.
a Press the reflection screen key.
b Connect the open end of Calibration Standard to the available end of the PORT 2 SMA
cable.
c Select the open screen key under the Reverse group.
d Wait until the VNA beeps indicating that it has completed the task.
e Connect short end of Calibration Standard to the available end of the PORT 2 SMA cable.
f Select short screen key the Reverse group.
g Wait until the VNA beeps indicating that it has completed the task.
h Connect load end of Calibration Standard to the available end of the PORT 2 SMA cable.
iSelect the loads screen key the Reverse group.
jPress broadband screen key selection.
k Wait until the VNA beeps indicating that it has completed the task.
l Press the done loads screen key.
m You have just calibrated the other side of the reference plane.
16Press standards done key.
17Connect the PORT 2 SMA cable to the non-pincher side of PV fixture as shown in Figure 19.
N2797A User’s Guide43
5Performance Verification
Figure 19Forward and Reverse Setup
18Press the transmission screen key.
19Press the do both fwd and reverse screen key.
20Wait until the VNA beeps four times indicating that it has completed the task.
21Press the isolation screen key.
22Press the omit isolation screen key.
23Press done 2 port cal screen key.
24Set the VNA's averaging to off.
25Save the reference plane cal by pressing the [save recall] key then the [save state] key.
26You may change name if you wish.
27Press the [scale reference] key. Then set the scale to 1 dB per division and the reference
position for 7 divisions.
28Set reference value for 0 dB.
29Press the [measure] key.
30Press the s21 screen key.
31Ensure s21 response on screen is flat out to 1.5 GHz.
44N2797A User’s Guide
Performance Verification5
Measuring Vin Response
32Position the probe conveniently to allow the probe tip to be normal to the PV/DS board.
33Hold the probe in position, or use a positioner, so that the signal and ground are making
contact and are perpendicular to the fixture.
34Press the [Sweep Setup] key on the VNA. Then press the trigger menu screen key. Select the
continuous screen key.
35You should now have the V
36Select [display] key then data->memory screen key.
37You have now saved Vin waveform into the VNA's memory for future use.
waveform on screen.
in
Measuring Vout Response
38Disconnect the PORT 2 cable from PV/DS test board and attach to probe output on the
AutoProbe Adapter.
39Connect the Calibration Standard load to PV/DS test board (non-pincher side).
40Press [scale reference] key on the VNA.
41Set reference value to –20 dB.
42Hold probe in place as described previously.
43The display on screen is V
out
.
Displaying Vout/Vin Response on the VNA
44Press the [Display] key.
45Then select the Data/Memory screen key. You may need to adjust the Reference Value, located
under the Scale Ref key, slightly to position the waveform at center screen.
46Press marker key and position the marker to the first point that the signal is –3 dB below
center screen.
47Read the marker frequency measurement and record it in Tab le 10 on page 46.
48The bandwidth test passes if the frequency measurement is greater than or equal to the
probe's bandwidth limit, which is 1.5 GHz.
N2797A User’s Guide45
5Performance Verification
Performance Test Record
The recommended Test Interval: 1 year / 2000 hours.
Table 10Performance Test Record
Probe AmplifierTest LimitsResultPass/Fail
Test 1. DC Input Resistance1 MΩ ± 3%
Test 2. Bandwidth≥ 1.5 GHz
46N2797A User’s Guide
Index
Index
Numerics
3D probe positioner, 28
A
accessories
available, 10
supplied, 8
attenuation ratio, 32
available accessories, 10