Page 1
Keysight Technologies
M9170A PXI Attenuator/Switch
Driver Module
Conguration Guide
Page 2
Overview
Key Features and Benets
This conguration guide will help you congure your
PXI attenuator/switch driver module with the expansive
portfolio of Keysight Technologies, Inc., RF and
microwave step attenuators and electromechanical
switches.
For more detailed product information and specications,
please refer to the following:
– M9170A Data Sheet (literature no. 5991-0130EN)
– M9170A Flyer (literature no. 5991-0053EN)
The M9170A is a one-slot PXI attenuator/switch driver
module that provides drive control for the Keysight
Technologies, Inc. RF and
and electromechanical switches.
solution offers limitless topologies for creating various
congurations for signal routing and signal conditioning,
driving external switches and attenuators through its dual
20-pin header.
– Drive up to 12 external SPDT switches, or 4 external
SP4T/6T switches, or 12 transfer switches, or 2
external attenuators. Improve testing efciency while
maximizing the PXI chassis slot utilization, which
ultimately improves testing efciency.
– Dual voltage supply of 5V and 24V and dual driving
mode for pulse and continuous, ensures biasing
compatibility for most switches on the market, hence
increasing system exibility.
– Soft front panel provides a dynamic dashboard view
and control of the connected attenuators or switches.
Indicated on the front panel is the selected device
model and the subsequent switch path or attenuation
levels.
– Point-to-point interface cable options are available for
connecting driver to switching device. Intuitive and
direct conguration for all Keysight switches and
attenuators.
microwave step attenuators
This cost effective
Page 3
3 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Specications
Drive Power Supply* M9170A
Voltage 23 ± 10%
4.5 ± 10%
Current 1.0A for 24V supply
3.85A for 5V supply
*This refers to the output of the driver module.
TTL drive is not supported
Note 1: Total current limit per bank = 2A
Note 2: Current limit per channel = 0.5A
Note 3 : Drive mode -pulsed (20 ms)
-continuous
Product congurations
This document will serve as a step-by-step guide to congure the point-to-point connection between the M9170A
with Keysight’s attenuators and electromechanical switches. Please refer to the various gures detailing the type of
customized interface cables, switch/attenuator options, and most importantly, the cross-reference tables that provide all
the necessary details (i.e. pin numbers, wires colors, etc.) to connect the varied combinations of switches or attenuators
to the M9170A module.
Cable option Part number Description
Option 001 M9170-20005 20 pin to 6x10 pin interface cable assembly
Option 002 M9170-20008 20 pin to 10 pin interface cable assembly
Option 003 M9170-20009 20 pin to 12 pin Viking interface cable assembly
Option 501 M9170-20007 20 pin to 6x9 pin D-Sub interface cable assembly
Option 601 M9170-20004 20 pin to 2x16 pin interface cable assembly
Option 201 M9170-20006 20 pin to bare wire interface cable assembly
Page 4
4 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Five Simple Steps to Congure Your Switching System
Step 1. Determine the device’s model and option (DC connector)
Example Model: 87104D (SP4T switch)
Option: 100 (solder terminal)
Step 2. Determine the M9170A’s interface cable option
Example Model: M9170
Option 201 (20 pin to bare wire interface cable assembly)
Step 3. Use the selection guide, Table 1 for switches and Table 2 for attenuators to
determine which conguration table to use for further reference.
Example Selection Guide: Table 1 (for switches)
Conguration Table: Table 5-A
Table 1: Selection guide for switches
Switch
family
Bypa ss
SPDT
SP3T 8766K 060 Tabl e 9
SP4T
Switch model number Switch
option
8763A, 8763B, 8763C 011/ 024 Table 3-B
8764A, 8764B, 8764C 011/ 024 Table 3-C
N18 11TL
N1812UL
8762A, 876 2B, 8762 C, 8762F 011/ 024 Table 3-A
8765A, 8765B, 8765C, 8765D, 8765F 30 5/324 Tabl e 4
N1810 UL
N1810 TL
87104A, 87104B, 87104C, 87104D
87204A, 87204B, 8720 4C 10 0 Tabl e 6
L7104A, L7104B, L7104C
202 Table 12-G
201 Tabl e 12-H
202 Tabl e 12-E
201 Tabl e 12-F
202 Tabl e 12-A
201 Tabl e 12-B
202 Table 12-C
201 Tabl e 12-D
100 Tabl e 5-A
161 Tabl e 5-B
100 Tabl e 5-A
161 Tabl e 5-B
M9170A
002 001 003 501 601 201
Step 4. Congure your switching system using Table 5-A (page 9) as a reference
Table 5-A. Conguration of M9170A (Option 201) to 87104A/B/C/D, L7104A/B/C and
L7204A/B/C SP4T (Option 100)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow 1 Vcc for D UT 1
4
2 Green 5 2 to C clos ed
3 Red 7 3 to C clos ed
5 Black/ Pink 11 5 to C clo sed
6
7
10 Violet 15 GND for DU T 2
Orange
White
Black/Green
To 87104A/B/C/D, L7104A/B/C and L7204A/B/C SP4T (Option 100)
15 GND for DU T 1
DUT 1
13 6 to C clos ed
1 Vcc for D UT 2
Page 5
5 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Step 5. System ready for operation.
Table 1: Selection guide for switches
Switch
family
Bypass
SPDT
SP3T 8766K 060 Table 9
SP4T
SP5T
SP6T
Matrix
Transfer
Switch model number Switch
option
8763A, 8763B, 8763C 011/024 Table 3-B
8764A, 8764B, 8764C 011/024 Table 3-C
N18 11T L
N1812UL
8762A, 8762B, 8762C, 8762F 011/024 Table 3-A
8765A, 8765B, 8765C, 8765D,
8765F
N1810UL
N1810T L
87104A, 87104B, 87104C, 87104D
87204A, 87204B, 87204C 100 Table 6
L7104A, L7104B, L7104C
L7204A, L7204B, L7204C
8767K 060 Table 9
8767M No Option Table 10
8768K 060 Table 9
8768M No Option Table 10
87106A, 87106B, 87106C, 87106D
87206A, 87206B, 87206C 100 Table 8
L7106A, L7106B, L7106C
L7206A, L7206B, L7206C
874 0 6B
87606B 100 Table 8
87222C, 87222D, 87222E
L7222C
202 Table 12-G
201 Table 12-H
202 Table 12-E
201 Table 12-F
305/324 Table 4
202 Table 12-A
201 Table 12-B
202 Table 12-C
201 Table 12-D
100 Table 5-A
161 Table 5-B
100 Table 5-A
161 Table 5-B
100 Table 5-A
161 Table 5-B
100 Table 7-A
161 Table 7-B
100 Table 7-A
161 Table 7-B
100 Table 7-A
161 Table 7-B
100 Table 7-A
161 Table 7-B
100 Table 11-A
161 Table 11-B
100 Table 11-A
161 Table 11-B
002 001 003 501 601 201
M9170A
Page 6
6 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 1-A. Switch option descriptions
Option 011 5 Vdc
Option 024 24 Vdc
Option 201 D-submini 9 pin (f)
Option 202 Solder Lug
Option 305 5 Vdc with solder terminals
Option 324 24 Vdc with solder terminals
Option 060 Viking cable connector
Option 100 Solder terminals
Option 161 Ribbon receptacle
Table 2. Selection guide for attenuators
Attenuator model number Connector type
8494G, 8494H
8495G, 8495H Table 13
8496G, 8496H Table 13
8495K Table 13
8497K Table 13
84904K, 84904L, 84904M
84905M Table 14
84906K, 84906L Table 14
84907K, 84907L Table 14
84908M Table 14
12 pin Viking connector (Option 060)
Table 14
10 pin DIP connector
M9170A
002 003
Table 13
Page 7
7 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Conguration information for switches
Table 3-A. Conguration of M9170A (Option 201) to 8762A/B/C/F SPDT (Option 011/024)
From M9170A (Option 201)
Interface cable
To 8762A/B/C/F SPDT (Option 011/024)
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow C VCC DUT 1
DUT 1 2 Green 1 1 to C closed, 2 terminated
3 Red 2 2 to C closed, 1 terminated
1 Yellow C VCC DUT 2
DUT 2 5 Black/Pink 1 1 to C closed, 2 terminated
6 White 2 2 to C closed, 1 terminated
7 Black/Green C VCC DUT 3
DUT 3 8 Black/Yellow 1 1 to C closed, 2 terminated
9 Black 2 2 to C closed, 1 terminated
11 Pink C VCC DUT 4
DUT 4 12 Black/Gray 1 1 to C closed, 2 terminated
13 Black/White 2 2 to C closed, 1 terminated
11 Pink C VCC DUT 5
DUT 5 15 Gray 1 1 to C closed, 2 terminated
16 Black/Blue 2 2 to C closed, 1 terminated
17 Black/Violet C VCC DUT 6
DUT 6 18 Blue 1 1 to C closed, 2 terminated
19 Black/Red 2 2 to C closed, 1 terminated
Table 3-B. Conguration of M9170A (Option 201) to 8763A/B/C bypass (Option 011/024)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow C VCC DUT 1
2 Green 1 1 to 2 closed, 3 to 4 closed
3 Red 2 1 terminated, 2 to 3 closed, 4 open
1 Yellow C VCC DUT 2
5 Black/Pink 1 1 to 2 closed, 3 to 4 closed
6 White 2 1 terminated, 2 to 3 closed, 4 open
7 Black/Green C VCC DUT 3
8 Black/Yellow 1 1 to 2 closed, 3 to 4 closed
9 Black 2 1 terminated, 2 to 3 closed, 4 open
11 Pink C VCC DUT 4
12 Black/Gray 1 1 to 2 closed, 3 to 4 closed
13 Black/White 2 1 terminated, 2 to 3 closed, 4 open
11 Pink C VCC DUT 5
15 Gray 1 1 to 2 closed, 3 to 4 closed
16 Black/Blue 2 1 terminated, 2 to 3 closed, 4 open
17 Black/Violet C VCC DUT 6
18 Blue 1 1 to 2 closed, 3 to 4 closed
19 Black/Red 2 1 terminated, 2 to 3 closed, 4 open
To 8763A/B/C bypass (Option 011/024)
DUT 1
DUT 2
DUT 3
DUT 4
DUT 5
DUT 6
Page 8
8 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 3-C. Conguration of M9170 (Option 201) to 8764A/B/C bypass (Option 011/024)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow C VCC DUT 1
3 Red 2 1 to 2 closed, 3 to 4 closed, 5 open
1 Yellow C VCC DUT 2
6 White 2 1 to 2 closed, 3 to 4 closed, 5 open
7 Black/Green C VCC DUT 3
9 Black 2 1 to 2 closed, 3 to 4 closed, 5 open
11 Pink C VCC DUT 4
13 Black/White 2 1 to 2 closed, 3 to 4 closed, 5 open
11 Pink C VCC DUT 5
16 Black/Blue 2 1 to 2 closed, 3 to 4 closed, 5 open
17 Black/Violet C VCC DUT 6
19 Black/Red 2 1 to 2 closed, 3 to 4 closed, 5 open
To 8764A/B/C bypass (Option 011/024)
DUT 1 2 Green 1 1 open, 2 to 3 closed, 4 to 5 closed
DUT 2 5 Black/Pink 1 1 open, 2 to 3 closed, 4 to 5 closed
DUT 3 8 Black/Yellow 1 1 open, 2 to 3 closed, 4 to 5 closed
DUT 4 12 Black/Gray 1 1 open, 2 to 3 closed, 4 to 5 closed
DUT 5 15 Gray 1 1 open, 2 to 3 closed, 4 to 5 closed
DUT 6 18 Blue 1 1 open, 2 to 3 closed, 4 to 5 closed
Table 4. Conguration of M9170A (Option 201) to 8765A/B/C/F SPDT (Option 305/324)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow 2 and 3 VCC DUT 1
3 Red 4 1 to C closed, 2 open
1 Yellow 2 and 3 VCC DUT 2
6 White 4 1 to C closed, 2 open
7 Black/Green 2 and 3 VCC DUT 3
9 Black 4 1 to C closed, 2 open
11 Pink 2 and 3 VCC DUT 4
13 Black/White 4 1 to C closed, 2 open
11 Pink 2 and 3 VCC DUT 5
16 Black/Blue 4 1 to C closed, 2 open
17 Black/Violet 2 and 3 VCC DUT 6
19 Black/Red 4 1 to C closed, 2 open
To 8765A/B/C/F SPDT (Option 305/324)
DUT 1 2 Green 1 2 to C closed, 1 open
DUT 2 5 Black/Pink 1 2 to C closed, 1 open
DUT 3 8 Black/Yellow 1 2 to C closed, 1 open
DUT 4 12 Black/Gray 1 2 to C closed, 1 open
DUT 5 15 Gray 1 2 to C closed, 1 open
DUT 6 18 Blue 1 2 to C closed, 1 open
Page 9
9 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 5-A. Conguration of M9170A (Option 201) to 87104A/B/C/D, L7104A/B/C and
L7204A/B/C SP4T (Option 100)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow 1 Vcc for DUT 1
4 Orange 15 GND for DUT 1
2 Green 5 2 to C closed
3 Red 7 3 to C closed
5 Black/Pink 11 5 to C closed
6 White 13 6 to C closed
7 Black/Green 1 Vcc for DUT 2
10 Violet 15 GND for DUT 2
8 Black/Yellow 5 2 to C closed
9 Black 7 3 to C closed
12 Black/Gray 11 5 to C closed
13 Black/White 13 6 to C closed
17 Black /Violet 1 Vcc for DUT 3
20 Brown 15 GND for DUT 3
15 Gray 5 2 to C closed
16 Black/Blue 7 3 to C closed
18 Blue 11 5 to C closed
19 Black/Red 13 6 to C closed
To 87104A/B/C/D, L7104A/B/C and L7204A/B/C SP4T (Option 100)
DUT 1
DUT 2
DUT 3
Table 5-B. Conguration of M9170A (Option 601) to 87104A/B/C/D, L7104A/B/C and
L7204A/B/C SP4T (Option 161)
From M9170A (Option 601)
Interface cable
20 PIN 16-Pin number RF path Device under test (DUT)
1 1 Vcc for DUT 1
10 15 GND for DUT 1
3 5 2 to C closed
5 7 3 to C closed
8 11 5 to C closed
9 13 6 to C closed
11 1 Vcc for DUT 1
20 15 GND for DUT 1
13 5 2 to C closed
15 7 3 to C closed
18 11 5 to C closed
19 13 6 to C closed
To 87104A/B/C/D, L7104A/B/C and L7204A/B/C SP4T (Option 161)
DUT 1
DUT 2
Page 10
10 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 6. Conguration of M9170A (Option 201) to 87204A/B/C SP4T (Option 100)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow 1 Vcc for DUT 1
4 Orange 15 GND for DUT 1
2 Green 5 2 to C closed
3 Red 6 2 to C open
5 Black/Pink 7 3 to C closed
6 White 8 3 to C open
8 Black/Yellow 11 5 to C closed
9 Black 12 5 to C open
12 Black/Gray 13 6 to C closed
13 Black/White 14 6 to C open
To 87204A/B/C SP4T (Option 100)
DUT 1
Page 11
11 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 7-A. Conguration of M9170A (Option 201) to 87106A/B/C/D, L7106A/B/C and
L7206A/B/C SP6T (Option 100) and 87406B matrix (Option 100)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF Path Device under test (DUT)
1 Yellow 1 Vcc for DUT 1
10 Violet 15 GND for DUT 1
2 Green 3 1 to C closed
3 Red 5 2 to C closed
5 Black/Pink 7 3 to C closed
6 White 9 4 to C closed
8 Black/Yellow 11 5 to C closed
9 Black 13 6 to C closed
11 Pink 1 Vcc for DUT
20 Brown 15 GND for DUT 1
12 Black/Gray 3 1 to C closed
13 Black/White 5 2 to C closed
15 Gray 7 3 to C closed
16 Black/Blue 9 4 to C closed
18 Blue 11 5 to C closed
19 Black/Red 13 6 to C closed
To 87106A/B/C/D, L7106A/B/C and L7206A/B/C SP6T (Option 100) and
87406B matrix (Option 100)
DUT 1
DUT 2
Table 7-B. Conguration of M9170A (Option 601) to 87106A/B/C/D, L7106A/B/C and
L7206A/B/C SP6T (Option 161) and 87406B matrix (Option 161)
From M9170A (Option 601)
Interface cable
20 PIN 16-Pin Number RF path Device under test (DUT)
1 1 Vcc for DUT 1
10 15 GND for DUT 1
2 3 1 to C closed
3 5 2 to C closed
5 7 3 to C closed
6 9 4 to C closed
8 11 5 to C closed
9 13 6 to C closed
11 1 Vcc for DUT 1
20 15 GND for DUT 1
12 3 1 to C closed
13 5 2 to C closed
15 7 3 to C closed
16 9 4 to C closed
18 11 5 to C closed
19 13 6 to C closed
To 87106A/B/C/D, L7106A/B/C and L7206A/B/C SP6T (Option 161) and
87406B matrix (Option 161)
DUT 1
DUT 2
Page 12
12 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 8. Conguration of M9170A (Option 201) to 87206A/B/C SP6T (Option 100) and
87606B matrix (Option 100)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow 1 Vcc to DUT 1
4 Orange 15 GND to DUT 1
2 Green 3 1 to C closed
3 Red 4 1 to C open
5 Black/Pink 5 2 to C closed
6 White 6 2 to C open
8 Black/Yellow 7 3 to C closed
9 Black 8 3 to C open
12 Black/Gray 9 4 to C closed
13 Black/White 10 4 to C open
15 Gray 11 5 to C closed
16 Black/Blue 12 5 to C open
18 Blue 13 6 to C closed
19 Black /Red 14 6 to C open
To 87206A/B/C SP6T (Option 100) and 87606B matrix (Option 100)
DUT 1
Table 9. Conguration of M9170A (Option 003) to 8766K, 8767K and 8768K (Option 060)
From M9170A (Option 003) To 8766K, 8767K and 8768K (Option 060)
Interface cable 8766K 8767K 8768K
20 PIN 12 pin Viking RF path RF path RF path
1 1 Vcc Vcc Vcc
2 5 Bypass 1 Bypass 3 Bypass 4
3 6 1 to C Closed 3 to C closed 4 to C closed
5 7 Bypass 2 Bypass 1 Bypass 2
6 8 2 to Closed 1 to C Closed 2 to Closed
8 9 Bypass 2 Bypass 3
9 10 2 to Closed 3 to C closed
12 11 Bypass 1
13 12 1 to C Closed
Page 13
13 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 10. Conguration of M9170 (Option 002) to 8767M/8768M
From M9170A (Option 002) To 8767M/8768M
Interface cable
20 PIN 10-Pin Dsub pin number RF path RF path
17 10 Vcc Vcc
2 1 Bypass 3 Bypass 4
3 2 3 to C Closed 4 to C closed
6 5 Bypass 1 Bypass 2
12 8 1 to Closed 2 to C Closed
5 4 Bypass 2 Bypass 3
13 9 2 to Closed 3 to Closed
8 6 — Bypass 1
9 7 — 1 to Closed
8767M 8768M
Table 11-A. Conguration of M9170A (Option 201) to L7222C and 87222C/D/E transfer
(Option 100)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow 1 Vcc for DUT 1
4 Orange 9 GND for DUT 1
2 Green 3 1 to 2 closed, 3 to 4 closed
3 Red 5 1 to 4 closed, 2 to 3 closed
1 Yellow 1 Vcc for DUT 2
4 Orange 9 GND for DUT 2
5 Green 3 1 to 2 closed, 3 to 4 closed
6 Red 5 1 to 4 closed, 2 to 3 closed
7 Black/Green 1 Vcc for DUT 3
10 Violet 9 GND for DUT 3
8 Black/Yellow 3 1 to 2 closed, 3 to 4 closed
9 Black 5 1 to 4 closed, 2 to 3 closed
11 Pink 1 Vcc for DUT 4
14 Black/Orange 9 GND for DUT 4
12 Black/Gray 3 1 to 2 closed, 3 to 4 closed
13 Black/White 5 1 to 4 closed, 2 to 3 closed
11 Pink 1 Vcc for DUT 5
14 Black/Orange 9 GND for DUT 5
15 Gray 3 1 to 4 closed, 2 to 3 closed
16 Black/Blue 5 1 to 4 closed, 2 to 3 closed
17 Black/ Violet 1 Vcc for DUT 6
20 Brown 9 GND for DUT 6
18 Blue 3 1 to 4 closed, 2 to 3 closed
19 Black/Red 5 1 to 4 closed, 2 to 3 closed
To L7222C and 87222C/D/E transfer (Option 100)
DUT 1
DUT 2
DUT 3
DUT 4
DUT 5
DUT 6
Page 14
14 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 11-B. Conguration of M9170 (Option 001) to L7222C and 87222C/D/E transfer (Option
161)
From M9170A (Option 001)
Interface cable
20 PIN 10-Pin Dsub pin number RF path Device under test (DUT)
1 1 Vcc for DUT 1
4 9 GND for DUT 1
2 3 1 to 2 closed, 3 to 4 closed
3 5 1 to 4 closed, 2 to 3 closed
1 1 Vcc for DUT 2
4 9 GND for DUT 2
5 3 1 to 2 closed, 3 to 4 closed
6 5 1 to 4 closed, 2 to 3 closed
7 1 Vcc for DUT 3
10 9 GND for DUT 3
8 3 1 to 2 closed, 3 to 4 closed
9 5 1 to 4 closed, 2 to 3 closed
11 1 Vcc for DUT 4
14 9 GND for DUT 4
12 3 1 to 2 closed, 3 to 4 closed
13 5 1 to 4 closed, 2 to 3 closed
11 1 Vcc for DUT 5
14 9 GND for DUT 5
15 3 1 to 2 closed, 3 to 4 closed
16 5 1 to 4 closed, 2 to 3 closed
17 1 Vcc for DUT 6
20 9 GND for DUT 6
18 3 1 to 2 closed, 3 to 4 closed
19 5 1 to 4 closed, 2 to 3 closed
To L7222C and 87222C/D/E transfer (Option 161)
DUT 1
DUT 2
DUT 3
DUT 4
DUT 5
DUT 6
Page 15
15 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 12-A. Conguration of M9170A (Option 201) to N1810UL SPDT (Option 202)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow +V Vcc for DUT 1
4 Orange GND GND for DUT 1
2 Green A 1 to C closed, 2 Open
3 Red B 2 to C closed, 1 Open
1 Yellow +V Vcc for DUT 2
4 Orange GND GND for DUT 2
5 Black/Pink A 1 to C closed, 2 Open
6 White B 2 to C closed, 1 Open
7 Black/Green +V Vcc for DUT 2
10 Voilet GND GND for DUT 2
8 Black/Yellow A 1 to C closed, 2 Open
9 Black B 2 to C closed, 1 Open
11 Pink +V Vcc for DUT 3
14 Black/Orange GND GND for DUT 3
12 Black/Gray A 1 to C closed, 2 Open
13 Black/White B 2 to C closed, 1 Open
11 Pink +V Vcc for DUT 4
14 Black/Orange GND GND for DUT 4
15 Gray A 1 to C closed, 2 Open
16 Black/Blue B 2 to C closed, 1 Open
17 Black/Violet +V Vcc for DUT 4
20 Brown GND GND for DUT 4
18 Blue A 1 to C closed, 2 Open
19 Black/Red B 2 to C closed, 1 Open
To N1810UL SPDT (Option 202)
DUT 1
DUT 2
DUT 3
DUT 4
DUT 5
DUT 6
Page 16
16 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 12-B. Conguration of M9170A (Option 501) to N1810UL SPDT (Option 201)
From M9170A (Option 501)
Interface cable
20 PIN 9-Pin Dsub pin number RF path Device under test (DUT)
1 5 Vcc for DUT 1
4 1 GND for DUT 1
2 4 1 to C close, 2 Open
3 3 2 to C close, 1 Open
1 5 Vcc for DUT 2
4 1 GND for DUT 2
5 4 1 to C close, 2 Open
6 3 2 to C close, 1 Open
7 5 Vcc for DUT 2
10 1 GND for DUT 2
8 4 1 to C close, 2 Open
9 3 2 to C close, 1 Open
11 5 Vcc for DUT 3
14 1 GND for DUT 3
12 4 1 to C close, 2 Open
13 3 2 to C close, 1 Open
11 5 Vcc for DUT 4
14 1 GND for DUT 4
15 4 1 to C close, 2 Open
16 3 2 to C close, 1 Open
17 5 Vcc for DUT 4
20 1 GND for DUT 4
18 4 1 to C close, 2 Open
19 3 2 to C close, 1 Open
To N1810UL SPDT (Option 201)
DUT 1
DUT 2
DUT 3
DUT 4
DUT 5
DUT 6
Page 17
17 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 12-C. Conguration of M9170A (Option 201) to N1810TL SPDT (Option 202)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow +V Vcc for DUT 1
4 Orange GND GND for DUT 1
2 Green A 1 to C close, 2 terminated
3 Red B 2 to C close, 1 terminated
1 Yellow +V Vcc for DUT 2
4 Orange GND GND for DUT 2
5 Black/Pink A 1 to C close, 2 terminated
6 White B 2 to C close, 1 terminated
7 Black/Green +V Vcc for DUT 3
10 Voilet GND GND for DUT 3
8 Black/Yellow A 1 to C close, 2 terminated
9 Black B 2 to C close, 1 terminated
11 Pink +V Vcc for DUT 4
14 Black/Orange GND GND for DUT 4
12 Black/Gray A 1 to C close, 2 terminated
13 Black/White B 2 to C close, 1 terminated
11 Pink +V Vcc for DUT 5
14 Black/Orange GND GND for DUT 5
15 Gray A 1 to C close, 2 terminated
16 Black/Blue B 2 to C close, 1 terminated
17 Black/Violet +V Vcc for DUT 6
20 Brown GND GND for DUT 6
18 Blue A 1 to C close, 2 terminated
19 Black/Red B 2 to C close, 1 terminated
To N1810TL SPDT (Option 202)
DUT 1
DUT 2
DUT 3
DUT 4
DUT 5
DUT 6
Page 18
18 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 12-D. Conguration of M9170A (Option 501) to N1810TL SPDT (Option 201)
From M9170A (Option 501)
Interface cable
20 PIN 9-Pin Dsub pin number RF path Device under test (DUT)
1 5 Vcc for DUT 1
4 1 GND for DUT 1
2 4 1 to C close, 2 terminated
3 3 2 to C close, 1 terminated
1 5 Vcc for DUT 2
4 1 GND for DUT 2
5 4 1 to C close, 2 terminated
6 3 2 to C close, 1 terminated
7 5 Vcc for DUT 3
10 1 GND for DUT 3
8 4 1 to C close, 2 terminated
9 3 2 to C close, 1 terminated
11 5 Vcc for DUT 4
14 1 GND for DUT 4
12 4 1 to C close, 2 terminated
13 3 2 to C close, 1 terminated
11 5 Vcc for DUT 5
14 1 GND for DUT 5
15 4 1 to C close, 2 terminated
16 3 2 to C close, 1 terminated
17 5 Vcc for DUT 6
20 1 GND for DUT 6
18 4 1 to C close, 2 terminated
19 3 2 to C close, 1 terminated
To N1810TL SPDT (Option 201)
DUT 1
DUT 2
DUT 3
DUT 4
DUT 5
DUT 6
Page 19
19 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 12-E. Conguration of M9170A (Option 201) to N1812UL bypass (Option 202)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow +V Vcc for DUT 1
4 Orange GND GND for DUT 1
2 Green A 1 to open, 2 to 3, 4 to 5
3 Red B 1 to 2, 3 to 4, 5 to open
1 Yellow +V Vcc for DUT 2
4 Orange GND GND for DUT 2
5 Black/Pink A 1 to open, 2 to 3, 4 to 5
6 White B 1 to 2, 3 to 4, 5 to open
7 Black/Green +V Vcc for DUT 3
10 Voilet GND GND for DUT 3
8 Black/Yellow A 1 to open, 2 to 3, 4 to 5
9 Black B 1 to 2, 3 to 4, 5 to open
11 Pink +V Vcc for DUT 4
14 Black/Orange GND GND for DUT 4
12 Black/Gray A 1 to open, 2 to 3, 4 to 5
13 Black/White B 1 to 2, 3 to 4, 5 to open
11 Pink +V Vcc for DUT 5
14 Black/Orange GND GND for DUT 5
15 Gray A 1 to open, 2 to 3, 4 to 5
16 Black/Blue B 1 to 2, 3 to 4, 5 to open
17 Black/Violet +V Vcc for DUT 6
20 Brown GND GND for DUT 6
18 Blue A 1 to open, 2 to 3, 4 to 5
19 Black/Red B 1 to 2, 3 to 4, 5 to open
To N1812UL bypass (Option 202)
DUT 1
DUT 2
DUT 3
DUT 4
DUT 5
DUT 6
Page 20
20 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 12-F. Conguration of M9170A (Option 501) to N1812UL bypass (Option 201)
From M9170A (Option 501)
Interface cable
20 PIN 9-Pin Dsub pin number RF path Device under test (DUT)
1 5 Vcc for DUT 1
4 1 GND for DUT 1
2 4 1 to open, 2 to 3, 4 to 5
3 3 1 to 2, 3 to 4, 5 to open
1 5 Vcc for DUT 2
4 1 GND for DUT 2
5 4 1 to open, 2 to 3, 4 to 5
6 3 1 to 2, 3 to 4, 5 to open
7 5 Vcc for DUT 3
10 1 GND for DUT 3
8 4 1 to open, 2 to 3, 4 to 5
9 3 1 to 2, 3 to 4, 5 to open
11 5 Vcc for DUT 4
14 1 GND for DUT 4
12 4 1 to open, 2 to 3, 4 to 5
13 3 1 to 2, 3 to 4, 5 to open
11 5 Vcc for DUT 5
14 1 GND for DUT 5
15 4 1 to open, 2 to 3, 4 to 5
16 3 1 to 2, 3 to 4, 5 to open
17 5 Vcc for DUT 6
20 1 GND for DUT 6
18 4 1 to open, 2 to 3, 4 to 5
19 3 1 to 2, 3 to 4, 5 to open
To N1812UL bypass (Option 201)
DUT 1
DUT 2
DUT 3
DUT 4
DUT 5
DUT 6
Page 21
21 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 12-G. Conguration of M9170A (Option 201) to N1811TL bypass (Option 202)
From M9170A (Option 201)
Interface cable
20 PIN Bare wire cable Solder terminal number RF path Device under test (DUT)
1 Yellow +V Vcc for DUT 1
4 Orange GND GND for DUT 1
2 Green A 1 to 2, 3 to 4
3 Red B 1 terminated, 2 to 3, 4 to open
1 Yellow +V Vcc for DUT 2
4 Orange GND GND for DUT 2
5 Black/Pink A 1 to 2, 3 to 4
6 White B 1 terminated, 2 to 3, 4 to open
7 Black/Green +V Vcc for DUT 3
10 Voilet GND GND for DUT 3
8 Black/Yellow A 1 to 2, 3 to 4
9 Black B 1 terminated, 2 to 3, 4 to open
11 Pink +V Vcc for DUT 4
14 Black/Orange GND GND for DUT 4
12 Black/Gray A 1 to 2, 3 to 4
13 Black/White B 1 terminated, 2 to 3, 4 to open
11 Pink +V Vcc for DUT 5
14 Black/Orange GND GND for DUT 5
15 Gray A 1 to 2, 3 to 4
16 Black/Blue B 1 terminated, 2 to 3, 4 to open
17 Black/Violet +V Vcc for DUT 6
20 Brown GND GND for DUT 6
18 Blue A 1 to 2, 3 to 4
19 Black/Red B 1 terminated, 2 to 3, 4 to open
To N1811TL bypass (Option 202)
DUT 1
DUT 2
DUT 3
DUT 4
DUT 5
DUT 6
Page 22
22 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 12-H. Configuration of M9170A (Option 501) to N1811TL bypass (Option 201)
From M9170A (Option 501)
Interface cable
20 PIN 9-Pin Dsub pin number RF path Device under test (DUT)
1 5 Vcc for DUT 1
4 1 GND for DUT 1
2 4 1 to 2, 3 to 4
3 3 1 terminated, 2 to 3, 4 to open
1 5 Vcc for DUT 2
4 1 GND for DUT 2
5 4 1 to 2, 3 to 4
6 3 1 terminated, 2 to 3, 4 to open
7 5 Vcc for DUT 3
10 1 GND for DUT 3
8 4 1 to 2, 3 to 4
9 3 1 terminated, 2 to 3, 4 to open
11 5 Vcc for DUT 4
14 1 GND for DUT 4
12 4 1 to 2, 3 to 4
13 3 1 terminated, 2 to 3, 4 to open
11 5 Vcc for DUT 5
14 1 GND for DUT 5
15 4 1 to 2, 3 to 4
16 3 1 terminated, 2 to 3, 4 to open
17 5 Vcc for DUT 6
20 1 GND for DUT 6
18 4 1 to 2, 3 to 4
19 3 1 terminated, 2 to 3, 4 to open
To N1811TL bypass (Option 201)
DUT 1
DUT 2
DUT 3
DUT 4
DUT 5
DUT 6
Page 23
23 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Conguration information for attenuators
Table 13. Conguration of M9170A (Option 003) to 8494G/H, 8495G/H, 8496G/H, 8495K
and 8497K programmable attenuators (Option 060)
From M9170A (Option 003)
Interface cable 8494G/H 8495G/H 8496G/H 8495K 8497K
20 PIN 12 PIN Viking connector pin Attenuation (dB)
1 1 Vcc Vcc Vcc Vcc Vcc
2 5 0 0 0 0 0
3 6 1 10 10 10 10
5 7 0 0 0 0 0
6 8 2 20 20 20 20
8 9 0 0 0 0 0
9 10 4 40 40 20 30
12 11 0 — 0 0 0
13 12 4 — 40 20 30
To 8494G/H,8495G/H, 8496G/H,8495K and 8497K
programmable attenuators (Option 060)
Table 14. Conguration of M9170A (Option 002) to 84904K/L/M, 84906K/L, 84907K/L,
84905M, 84908M programmable attenuators
From M9170A (Option 002) To 84904K/L/M, 84906K/L, 84907K/L, 84905M, 84908M programmable attenuators
Interface cable 84904K/L/M 84906K/L 84907K/L 84905M 84908M
20 PIN 10 Pin Attenuation (dB)
17 10 Vcc Vcc Vcc Vcc Vcc
2 1 0 0 0 0 0
3 2 1 10 10 10 5
6 5 0 0 0 0 0
12 8 2 20 20 20 10
5 4 0 0 0 0 0
13 9 4 30 40 30 20
8 6 0 0 — — 0
9 7 4 30 — — 30
Page 24
24 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Table 15. Generic bare wire connection
Bare wire can be used to drive the switch when no connector option available.
From
Coaxial wire color Drive control
Yellow
Green
Red
Orange
Black/Pink
White
Black/Green
Black/Yellow
Black
Violet
Pink
Black/Gray
Black/White
Black/Orange
Gray
Black/Blue
Black/Violet
Blue
Black/Red
Brown
Vcc 1 1
Drive 1 2
Drive 2 3
GND 1 4
Drive 3 5
Drive 4 6
Vcc 2 7
Drive 5 8
Drive 6 9
GND 2 10
Vcc 3 11
Drive 7 12
Drive 8 13
GND 3 14
Drive 9 15
Drive 10 16
Vcc 4 17
Dr ive 11 18
Drive 12 19
GND 4 20
P1
Page 25
25 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Interface Cable Drawings
Figure 1. Option 001 - 2 0 pin to 6x10 pin interface cable assembly
Figure 2. Option 0 02 - 20 pin to 10 pin inter face cable assembly
Page 26
26 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Interface Cable Drawings (CONT)
Figure 3. Option 0 03 - 20 pin to 12 Pin Viking inter face cable assembly
Figure 4. Option 5 01 - 20 pin to 6x9 pin-D-Sub inter face cable assembly
Page 27
27 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Interface Cable Drawings (CONT)
Figure 5. Option 6 01 - 20 pin to 2x16 pin inter face cable assembly
Figure 6. Option 201 - 20 pin to bare wire inter face cable assembly
Page 28
28 | Keysight | M9170A PXI Attenuator/Switch Driver Module - Conguration Guide
Evolving Since 1939
Our unique combination of hardware, software, services, and people can help you
reach your next breakthrough. We are unlocking the future of technology.
From Hewlett-Packard to Agilent to Keysight.
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
http://www.keysight.com/find/emt_product_registration
Register your products to get up-to-date product information and find
warranty information.
www.axiestandard.org
AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an
open standard that extends the AdvancedTCA for general purpose and
semiconductor test. The business that became Keysight was a founding
member of the AXIe consortium. ATCA®, AdvancedTCA®, and the ATCA logo
are registered US trademarks of the PCI Industrial Computer Manufacturers
Group.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the Web
inside your test systems. The business that became Keysight was a founding
member of the LXI consortium.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a
rugged, PC-based high-performance measurement and automation system.
Keysight Services
www.keysight.com/find/service
Keysight Services can help from acquisition to renewal across your
instrument’s lifecycle. Our comprehensive service offerings—one-stop
calibration, repair, asset management, technology refresh, consulting,
training and more—helps you improve product quality and lower costs.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure your
instruments are operating to specification, so you can rely on accurate
measurements.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada (877) 894 4414
Brazil 55 11 3351 7010
Mexico 001 800 254 2440
United States (800) 829 4444
Asia Pacic
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 11 2626
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 6375 8100
Europe & Middle East
Austria 0800 001122
Belgium 0800 58580
Finland 0800 523252
France 0805 980333
Germany 0800 6270999
Ireland 1800 832700
Israel 1 809 343051
Italy 800 599100
Luxembourg +32 800 58580
Netherlands 0800 0233200
Russia 8800 5009286
Spain 800 000154
Sweden 0200 882255
Switzerland 0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
United Kingdom 0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-9-7-17)
DEKRA Certified
ISO9001 Quality Management System
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2015
Quality Management System
www.keysight.com/find/modular
www.keysight.com/find/PXIdriver
www.keysight.com/find/PXIattenuator
This information is subject to change without notice.
© Keysight Technologies, 2017
Published in USA, January 10, 2018
5991-0052EN
www.keysight.com