Keysight Technologies M8000 Series, M8020A, M8030A, M9536A, M9537A Installation Manual

...
Keysight M8000 Series of BER Test Solutions
J-BERT M8020A High-Performance BERT M8030A Multi-Channel BERT M8040A High-Performance BERT
Installation Guide
Notices
CAUTION
WARNING
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla­tion into a foreign language) without prior agreement and written consent from Keysight Technologies as governed by United States and international copyright laws.
Trademarks
PCI Express® and PCIe® are registered trademarks of PCI-SIG.
Manual Part Number
M8000-91B05
Edition
Edition 2.0, May 2019 Keysight Technologies Deutschland GmbH Herrenberger Strasse 130, 71034 Böblingen, Germany
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement
(“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the soft­ware is customarily provided to the public. Accordingly, Keysight provides the Soft­ware to U.S. government customers under its standard commercial license, which is embodied in its End User License Agree-
ment (EULA), a copy of which can be found at http://www.keysight.com/find/sweula. The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Key­sight: (1) Furnish technical information related to commercial computer software or commercial computer software docu­mentation that is not customarily provided to the public; or (2) Relinquish to, or other­wise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose com­mercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as appli­cable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS DOCU­MENT IS PROVIDED "AS IS," AND IS SUB­JECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MER­CHANTABILITY AND FITNESS FOR A PAR­TICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDEN­TAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCU­MENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREE­MENT WITH WARRANTY TERMS COVER­ING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating pro­cedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating pro­cedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight M8000 Series of BER Test Solutions Installation Guide

Safety Summary

General This product is a Safety Class 1 instrument (provided with a protective earth terminal).
The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings or operating instructions in the product manuals violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer's failure to comply with these requirements. Product manuals are provided with your instrument on CD-ROM and/or in printed form. Printed manuals are an option for many products. Manuals may also be available on the Web. Go to
www.keysight.com and type in your product number in the Search field at the top of the
page.
The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.
All Light Emitting Diodes (LEDs) used in this product are Class 1 LEDs as per IEC 60825-1.
Environment Conditions
Before Applying Power
Ground the Instrument
Do Not Operate in an
Explosive Atmosphere
Do Not Remove the
Instrument Cover
This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is designed to operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters.
Refer to the specifications tables for the ac mains voltage requirements and ambient operating temperature range.
Verify that all safety precautions are taken. The power cable inlet of the instrument serves as a device to disconnect from the mains in case of hazard. The instrument must be positioned so that the operator can easily access the power cable inlet. When the instrument is rack mounted the rack must be provided with an easily accessible mains switch.
To minimize shock hazard, the instrument chassis and cover must be connected to an electrical protective earth ground. The instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
Do not operate the instrument in the presence of flammable gases or fumes.
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made only by qualified personnel.
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
Keysight M8000 Series of BER Test Solutions Installation Guide 3

Safety Symbols

Table 1 Safety Symbol
Symbol Description
Indicates warning or caution. If you see this symbol on a product, you must refer to the manuals for specific Warning or Caution information to avoid personal injury or damage to the product.
Frame or chassis ground terminal. Typically connects to the equipment’s metal frame.
KC is the Korean certification mark to demonstrate that the equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.
Indicates that anti-static precautions should be taken.
Indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
The RCM Mark is a compliance mark to the ACMA (Australian Spectrum Management Agency). This indicates compliance with all Australian EMC regulatory information.
Indicates that the product was tested and has met the certification requirements for electrical, plumbing and/or mechanical products.
4 Keysight M8000 Series of BER Test Solutions Installation Guide
Symbol Description
CE compliance marking to the EU Safety and EMC Directives. ISM GRP-1A classification according to the international EMC standard. ICES/NMB-001 compliance marking to the Canadian EMC standard.
This symbol on all primary and secondary packaging indicates compliance to China standard GB 18455-2001.
Keysight M8000 Series of BER Test Solutions Installation Guide 5

Compliance and Environmental Information

Table 2 Compliance and Environmental Information
Safety Symbol Description
This product complies with WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste.
Product Category: With reference to the equipment types in WEEE Directive Annex I, this product is classed as a “Monitoring and Control instrumentation” product.
Do not dispose in domestic household waste.
To return unwanted products, contact your local Keysight office, or see
http://about.keysight.com/en/companyinfo/environment/takeback.shtml
for more information.
6 Keysight M8000 Series of BER Test Solutions Installation Guide

About This Guide

This guide provides detailed information for installing Keysight M8000 modules, including the Keysight M9536A or M9537A AXIe Embedded Controller, AXIe System Module (ASM) in the Keysight M9505A/M9514A AXIe chassis. The procedures in this guide are not required for “bundled” systems such as the M8020A-BU1, M8020A-BU2, M8030A-BU1, M8030A-BU2, M8040A-BU1, M8040A-BU2 or M8040A-BU3. For these systems, refer to the Keysight M8000 Series BER Test Solutions Getting Started Guide and Keysight M8040A Getting Started Guide.
After performing the procedures in this document, you are directed to procedures in the
Keysight M8000 Series BER Test Solutions Getting Started Guide and Keysight M8040A Getting Started Guide to complete the installation.
Keysight M8000 Series of BER Test Solutions Installation Guide 7

Contents

1 System Requirements
Safety Summary 3
Compliance and Environmental Information 6
About This Guide 7
Hardware and Software Requirements 14
Hardware Configurations for M8020A 15
M8020A Base System Configuration 15 M8020A Four Channel System Configuration 16 M8020A 32 Gb/s BERT Configuration (Pattern Generator Only) 17 M8062A 32Gb/s Front-end for J-BERT M8020A High-Performance
BERT 18 Single M8020A LAN Network Configuration 19
Hardware Configurations for M8030A 20
Hardware Configurations for M8040A 22
M8045A Pattern Generator Module 22 M8046A Analyzer Module 23 M8057A/B Remote Head 23 M8040A LAN Network Configuration 24
Keysight M8000 Series of BER Test Solutions Installation Guide 9
Contents
2 Installing Modules
AXIe Instrument Modules 26
Installing M8020A Module(s) 27
To install the M8041A module 28 To install the M8041A and M8051A modules 30 To install the M8041A and M8061A modules 33 To install the M8041A and M8062A modules 35 To remove an M8020A module 37
Setting up the M8020A 39
Benchtop Configuration 39 Benchtop Configuration with Support Assembly 40 Rack Mounted Configuration 40 Installing the AXIe Embedded Controller Module 41
Setting up the M8030A 43
Removing Filler Panels 46 Installing the AXIe Embedded Controller Module 46 Installing a Module 49 Removing a Module 52 Synchronization Cable Configuration 52
Installing M8040A Module(s) 55
To install the M8045A module 56 To install the M8045A and M8046A modules 58 To Connect the M8057A/B Remote Head 60 To remove an M8040A module 61
Setting up the M8040A 63
Benchtop Configuration 63 Rack Mounted Configuration 63 Installing the AXIe Embedded Controller Module 64
Completing the Installation and Setup Process 66
10 Keysight M8000 Series of BER Test Solutions Installation Guide
3 Controlling Multiple Instruments
Controlling Multiple Instruments by Multiple Instances of M8070B System Software 70
Remote Control of Multiple M8070B Software Instances 74
Command Line Options 74
Creating M8070B Software Shortcut on the Desktop 75
Index
Contents
Keysight M8000 Series of BER Test Solutions Installation Guide 11
Keysight M8000 Series of BER Test Solutions
Installation Guide

1 System Requirements

Hardware and Software Requirements / 14 Hardware Configurations for M8020A / 15 Hardware Configurations for M8030A / 20 Hardware Configurations for M8040A / 22
This chapter provides information about possible configurations for Keysight M8020A, M8030A and M8040A systems.
1 System Requirements
NOTE
NOTE

Hardware and Software Requirements

The following are the hardware and software requirements that should be met before the installation of software components on the controller (host computer):
Minimum hardware requirements
• Pentium® processor 1 GHz or equivalent
• 16 GB available RAM
• USB 3.0 connection
• PCIe 2.0/8x (only for highest data throughput and desktop PC)
• VGA resolution 1024 x 768
• 1.5 GB or more free hard disc space
Software requirements
• The following operating systems are supported:
Windows 7 (64 bit) SP1
Windows 8 (64 bit)
Windows 8.1 (64 bit)
Windows 10 (64 bit)
• Keysight I/O libraries version 17.2
The M8070B software is required to control the M8020A, M8030A and M8040A systems.
In case of M8020A, PCIe connectivity between the M9505A AXIe Chassis and an external desktop PC controller is recommended when full channel plus large patterns need to be downloaded.
14 Keysight M8000 Series of BER Test Solutions Installation Guide

Hardware Configurations for M8020A

NOTE
The following section describes and illustrates various setup combinations in which you can install the M8020A modules.
The M8041A module must be installed in slots 1 through 3 of the M9505A AXIe Chassis for proper local bus communication unless the M9536A or M9537A AXIe Embedded Controller is installed (must be in slot 1).
M8020A Base System Configuration
The base configuration is a single channel system (a second channel can be added with license) consisting of the 5-slot M9505A AXIe Chassis and an M8041A module. The M8041A occupies three slots. A maximum of two M8020A modules can be installed in a 5-slot chassis.
System Requirements 1
Figure 1 M8020A configuration for 16 Gb/s BERT with 1 to 2 channel
Keysight M8000 Series of BER Test Solutions Installation Guide 15
1 System Requirements
NOTE
M8020A Four Channel System Configuration
The four channel configuration consists of the 5-slot M9505A AXIe Chassis, an M8041A module, and an M8051A module. The M8041A occupies three slots and the M8051A occupies two slots.
This configuration requires a cable (provided with the M8051A) that connects the M8041A SYNC OUT to the M8051A SYNC IN to synchronize the two modules to a common system clock.
Figure 2 M8020A configuration for 16 Gb/s BERT for up to 4 channels
16 Keysight M8000 Series of BER Test Solutions Installation Guide
M8020A 32 Gb/s BERT Configuration (Pattern Generator Only)
A typical configuration using the M8061A 32 Gb/s multiplexer with de-emphasis consists of the 5-slot M9505A AXIe Chassis, an M8041A module, and an M8061A module. The M8041A occupies three slots and the M8061A occupies two slots.
System Requirements 1
Figure 3 M8020A configuration for 32 Gb/s BERT (external demultiplexer is recom-
mended)
Keysight M8000 Series of BER Test Solutions Installation Guide 17
1 System Requirements
M8062A 32Gb/s Front-end for J-BERT M8020A High-Performance BERT
A typical configuration for an M8020A 32 Gb/s full BERT consists of the 5-slot M9505A AXIe Chassis, an M8041A module, and an M8062A module. The M8041A occupies three slots and the M8062A occupies two slots.
Figure 4 M8020A configuration for 32 Gb/s BERT
18 Keysight M8000 Series of BER Test Solutions Installation Guide
Single M8020A LAN Network Configuration
*
Host PC with
M8070B
J-BERT M8020A with M8041A/ M8051A, M8061A, or M8062A Modules
USB Connection
LAN Network
Client PC
Client PC
Client PC
(remote desktop or VNC)
*
Host PC can also be M9536A or M9537A
AXIe Embedded Controller
Multiple engineers can all be connected to a single M8020A via a LAN network and controlled using the M8070B software running on a host PC. The host PC tracks the number of licenses checked out and the number of licenses available for use. In addition, the host PC can be a dedicated computer running the license server or it can also run the M8070B software concurrently. The M9536A or M9537A AXIe Embedded Controller can also be used as the host PC in this configuration.
System Requirements 1
Keysight M8000 Series of BER Test Solutions Installation Guide 19
Figure 5 Single M8020A LAN network configuration
1 System Requirements

Hardware Configurations for M8030A

The M8030A is a modular test solution which can be tailored to your specific needs from two channels with one M8041A to up to 10 channels.
The M8030A supports the following modules.
• M8041A high-performance BERT generator-analyzer-clock 8/16 Gb/s
• M8051A high-performance BERT generator-analyzer 8/16 Gb/s
• M8192A Multi-channel synchronization module
The modules must be installed in the M9514A AXIe 14-slot chassis as shown in Table 3 on page -20:
Table 3 M8030A modules’ arrangement in the M9514A AXIe chassis
Slot Number Module
# 1 For M8030A-BU1, M9536A or M9537A AXIe embedded controller. For
# 2-4 M8041A module
# 5-6 M8051A module
# 7 M9521A AXIe system module, always included in M8030A-BU1 or
# 8-9 M8051A module
# 10-11 M8051A module
# 12-13 M8051A module
# 14 M8192A multi-channel synchronization module, mandatory
M8030A-BU2, this slot is empty and covered with filler front-panel
M8030A-BU2, mandatory
20 Keysight M8000 Series of BER Test Solutions Installation Guide
System Requirements 1
Figure 6 on page -21 shows an example of modules arrangement in the
M9514A AXIe 14-slot chassis.
Figure 6 Example of M8030A module arrangement
Keysight M8000 Series of BER Test Solutions Installation Guide 21
1 System Requirements
M8045A Module (2 Channels)
NOTE

Hardware Configurations for M8040A

The M8040A is a modular test solution which simplifies accurate receiver characterization of devices operating up to 32 and 64 GBaud with NRZ and PAM-4. It supports the following modules.
• M8045A pattern generator
• M8046A analyzer
• M8057A/B remote head
The following section describes and illustrates various setup combinations in which you can install the M8040A modules.
M8045A Pattern Generator Module
The M8045A module can be a one or two data channel system (a second channel can be added with license). A one channel instrument has to be returned to the factory for installing the second channel (hardware) and license. It occupies three slots of the 5-slot M9505A AXIe chassis. The following figure illustrates an M8045A module (two data channel system) installed in an M9505A AXI chassis.
22 Keysight M8000 Series of BER Test Solutions Installation Guide
Figure 7 M8045A configuration for 2 channels
The M8045A module must be installed in slots 1 through 3 of the M9505A AXIe chassis for proper local bus communication unless the M9537A AXIe Embedded Controller is installed (must be in slot 1).
M8046A Analyzer Module
M8046A Module
M8057A/B Remote Head
The M8046A module occupies single slot of the 5-slot M9505A AXIe chassis. The following figure illustrates an M8045A module with M8046A module installed in an M9505A AXI chassis.
Figure 8 M8045A and M8046A configuration
M8057A/B Remote Head
System Requirements 1
The M8057A/B remote head is an external box which can be connected to each channel of M8045A module, using the matched pair of cables. It helps in minimizing signal degradations caused by lossy channels. The following figure illustrates an M8057A/B remote head connected with one channel of M8045A module.
Figure 9 M8045A, M8046A and M8057A/B configuration
Keysight M8000 Series of BER Test Solutions Installation Guide 23
1 System Requirements
*
Host PC with
M8070B
M8040A with M8045A, M8046A and M8057A/B Modules
USB Connection
LAN Network
Client PC
Client PC
Client PC
(remote desktop or VNC)
*
Host PC can also be M9537A
AXIe Embedded Controller
M8040A LAN Network Configuration
Multiple engineers can all be connected to a single M8040A via a LAN network and controlled using the M8070B software running on a host PC. The host PC tracks the number of licenses checked out and the number of licenses available for use. In addition, the host PC can be a dedicated computer running the license server or it can also run the M8070B software concurrently. The M9537A AXIe Embedded Controller can also be used as the host PC in this configuration.
24 Keysight M8000 Series of BER Test Solutions Installation Guide
Figure 10 M8040A LAN network configuration