Keysight Technologies Keysight VXT PXIe Series, Keysight M9410A VXT, Keysight M9411A VXT User Manual

Keysight VXT PXIe Vector Transceiver
This manual provides documentation for: Keysight M9410A VXT Vector Transceiver
Keysight M9411A VXT Vector Transceiver
Source User’s Guide

Notices

© Keysight Technologies, Inc. 2018
Trademark Acknowledgments
Manual Part Number
M9410-90010
Edition
Edition 1, December 2018
Published in China
Published by: Keysight Technologies
No 116 Tianfu 4th Street Chengdu, China 610041
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government requirements
beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

Where to Find the Latest Information

Documentation is updated periodically. For the latest information about these products, including instrument software upgrades, application information, and product information, browse to one of the following URLs, according to the name of your product:
http://www.keysight.com/find/m9410a
http://www.keysight.com/find/m9411a
To receive the latest updates by email, subscribe to Keysight Email Updates at the following URL:
http://www.keysight.com/find/MyKeysight
Information on preventing instrument damage can be found at:
www.keysight.com/find/PreventingInstrumentRepair
Is your product software up-to-date?
Periodically, Keysight releases software updates to fix known defects and incorporate product enhancements. To search for software updates for your product, go to the Keysight Technical Support website at:
http://www.keysight.com/find/techsupport
iii
iv
Table of Contents
1 Source Overview
Source Features 6
Modes of Operation 7
Continuous Wave 7
Digital Modulation 7
2 Basic Operation
Viewing Key Descriptions 10
Setting RF Source parameters 10
Presetting the Source 11

Contents

Saving and Recalling States 12
Selecting Source Output 12
Setting Frequency and Power (Amplitude) 13
Example: Configuring a 700 MHz, 20 dBm Continuous Wave Output 13
Working with Files 14
Viewing a List of Stored Files 14
Viewing a List of Files Stored on the transceiver 14
Storing a File 15
Loading (Recalling) a Stored File 15
3 Optimizing Performance
Using an Output Offset and Reference 18
Setting an Output Offset 18
Setting an Output Reference 19
Using Amplitude Corrections 20
Selecting a port 20
Setting up a corrections table 21
M9410A/M9411A Source User’s Guide 1
Contents
4 Basic Digital Operation
Waveform File Basics 24
Instrument Memory 24
Dual ARB Player 24
Storing, Loading, and Playing a Waveform Segment 25
Loading a Waveform Segment into ARB memory 25
Deleting a Waveform Segment on Internal Disk 26
Deleting a Waveform Segment from ARB memory 26
Playing a Waveform Segment 26
Saving a Waveform Settings & Parameters 27
Viewing and Modifying Header Information 27
Using Waveform Markers 30
Waveform Marker Concepts 30
Marker Signal Response 31
Marker File Generation 31
Marker Point Edit Requirements 31
Saving Marker Polarity and Routing Settings 31
Selecting which Marker to output 31
Viewing a Marker Pulse 31
Using the Pulse/RF Blanking Marker Function 32
Setting Marker Polarity 33
Using the TRIGGER 2 Signal as a Transceiver Trigger 34
Triggering a Waveform 35
Trigger Type 35
Trigger Source 36
Source Alignment 37
Licensing 38
Signal Studio 38
Waveform multi-pack licensing 38
Installing an Option N7650B–22x/25x Waveform 5/50–Pack Licensing 39
Licensing a Source Waveform File 40
Waveform Licensing Interface 41
Waveform Multi–Pack Licensing Status Messages 43
Other menu functions under Multi–Pack Licenses 44
2 M9410A/M9411A Source User’s Guide
5 Troubleshooting
RF Output 46
No RF Output 46
Power Supply Shuts Down 46
No Modulation at the RF Output 46
RF Output Power too Low 46
Error Messages 47
Licenses 47
A time–based license stops working 47
Cannot load a time–based license 47
Contacting Keysight Technologies 48
Contents
M9410A/M9411A Source User’s Guide 3
Contents
4 M9410A/M9411A Source User’s Guide
Keysight M9410A/M9411A PXIe Vector Transceiver
Source User’s Guide

1 Source Overview

The source in this instrument can be operated independently of the transceiver to provide arbitrary waveform generation to test wireless mobile devices. The source can also be used together with the transceiver to simultaneously test both the transmitter and receiver of a wireless device.
“Source Features” on page 6
“Modes of Operation” on page 7
5
Source Overview Source Features

Source Features

— 380 MHz to 6.0 GHz
— arbitrary I/Q waveform playback up to 375 MSa/s
— SCPI and IVI–COM driver
— user flatness correction
— two output connectors
— with Signal Studio Software, the transceiver can generate 802.11 WLAN,
For more details on hardware, firmware, software, and documentation features and options, refer to the data sheet shipped with the transceiver and available from the Keysight Technologies website at:
http://www.Keysight.com/find/m9410a
http://www.Keysight.com/find/m9411a
W–CDMA, cdma2000, 1xEV–DO, GSM, EDGE, and more.
6 Source User’s Guide
Source Overview Modes of Operation

Modes of Operation

The instrument source provides two basic modes of operation: continuous wave (CW) and digital modulation mode.

Continuous Wave

In this mode, the transceiver produces a CW signal. The transceiver is set to a single frequency and power level, which is set up using the RF Source tab menu. To learn more, refer to Chapter 2, Basic Operation.

Digital Modulation

In this mode, an internal baseband generator in the transceiver modulates a CW signal with an arbitrary I/Q waveform (ARB). The ARB waveforms are downloaded through a remote interface into the ARB memory. The ARB waveform setup is accessed using the RF Source tab, followed by the Modulation Setup key. To learn more, refer to Chapter 4, Basic Digital
Operation.
Source User’s Guide 7
Source Overview Modes of Operation
8 Source User’s Guide
Keysight M9410A/M9411A PXIe Vector Transceiver
Source User’s Guide

2 Basic Operation

This chapter introduces fundamental front panel operation. For information about remote operation, refer to the Online Help or Programmer’s Guide
“Viewing Key Descriptions” on page 10
“Setting RF Source parameters” on page 10
“Presetting the Source” on page 11
“Saving and Recalling States” on page 12
“Selecting Source Output” on page 12
“Setting Frequency and Power (Amplitude)” on page 13
“Working with Files” on page 14
9
Basic Operation Viewing Key Descriptions

Viewing Key Descriptions

Table 2-1 Online Help Access
Press the question mark key for comprehensive help information. The online help is context-sensitive (that is, the question mark key opens to a page that is relevant to the last key you pressed). Also, while the help screen is open, press any key, and a page relevant to that key is displayed; the normal key function does not execute. Press ESC to exit the online help.

Setting RF Source parameters

The source functionality is accessed by selecting Input/Output key as below.
Selecting the Input/Output key opens up a softkeys menu under the RF Source tab on the right side of the application window. These keys allow complete configuration of the corresponding tab’s parameters.
10 Source User’s Guide
Basic Operation Presetting the Source

Presetting the Source

Presetting the source is a method of restoring all source settings to its default values, after some or all of them have been changed. If you want to be able to return the source settings to the values they had before the preset, save the current state before you preset the source (see “Saving and Recalling States below).
1. Select the Input/Output key.
2. Press the RF Source Tab (viewed on right of softkey menu).
3. Press Source Preset softkey.
Source User’s Guide 11
Basic Operation Saving and Recalling States

Saving and Recalling States

The current state (including both source and analyzer settings) can be saved and later recalled.
1. Click file icon in the bottom bar, and click Save key to access
the file operation window. Select State in the left side of this window.
2. Choose a register in which to save the state by pressing one of the Register keys. (A register with no state saved in it is in blank; when you
save a state in it, the register is labeled with the date and time.)
3. Alternatively, you can save the state to a file rather than to a register. Click the Save To File key in the upper-right corner of the window, and use the file-save window to chose a name and location for the file.
4. Recall a saved state by clicking Recall key, followed by the State key.
5. Choose a register from which to retrieve a state by Selecting one of the Register keys.
6. Alternatively, you can retrieve the state from a file rather than from a
register. Click the Recall From File key, and chose a name and location for the file.

Selecting Source Output

To select the output port for the source, access the Input/Output menu.
7. Click Input/Output, RF Source tab, RF Output Port, RF Output.
8. Select from RF Output, and RFIO HD. If you select RF Output, this means that the RF Output designated as an output will be used.
12 Source User’s Guide
Basic Operation Setting Frequency and Power (Amplitude)

Setting Frequency and Power (Amplitude)

The M9410A/M9411A transceiver does not have an Automatic Level Control feature; amplitude accuracy is regulated by means of calibration for frequency and temperature variations.
To set the frequency and power of the source output proceed as follows:
1. Select Input/Output, RF Source, Amplitude, Source Amplitude.
2. Use the numeric keypad to set a value for power. Press the appropriate
units key to complete the entry.
3. Select Frequency.
4. Use the numeric keypad to set a value for frequency. Press the appropriate
units key to complete the entry.
5. Turn the RF Output key to On.
Example: Configuring a 700 MHz, 20 dBm Continuous Wave Output
1. Preset the source by Input/Output, RF Source, Source Preset.
2. Select Input/Output, RF Source, Frequency key. Press 700 on the keypad and select the MHz key.
3. Click Amplitude key. Press 20 on keypad and select the -dBm key.
4. Turn the RF Output to On.
5. Press Input/Output, RF Output Port, RF Output.
The 700 MHz, 20 dBm CW signal is available at the RF OUTPUT port (to select the output port see “Selecting Source Output” on page 12).
Source User’s Guide 13
Basic Operation Working with Files

Working with Files

“Viewing a List of Stored Files” on page 14
“Loading (Recalling) a Stored File” on page 15
The transceiver recognizes several types of files, such as state files, license files, and correction files. Files are stored in the controller internal storage. This section provides an overview of how to navigate the file menus, and how to view, store, load, and move files.
See also: “Storing, Loading, and Playing a Waveform Segment” on page 25.

Viewing a List of Stored Files

The arbitrary waveform generator (ARB) files that have been created on the transceiver can be viewed using the Waveform Selection screen.
Viewing a List of Files Stored on the transceiver
1. Select Input/Output key and RF Source tab.
2. Press Modulation Setup, ARB Setup.
3. Press Select Waveform.
The files listed are identified as either waveform sequences or waveform segments. A sequence is made up of two or more segments.
Use mouse to highlight a waveform for selection. Once the desired file is highlighted, use the Select Waveform key to select the waveform to be output.
14 Source User’s Guide
Basic Operation Working with Files

Storing a File

To store a file that you have set up, use the Save key as described in
“Storing, Loading, and Playing a Waveform Segment” on page 25.
To store a transceiver state that you have set up and may want to re-use, use the Save key and select State from the sub-menu and select a register in which to store the state.
For more information about the Save function see the online help.

Loading (Recalling) a Stored File

To recall a file that you have previously saved, use the Recall key. For
transceiver state recalls, select State from the sub-menu and select the required register.
For more information about the Recall function see the online help.
Source User’s Guide 15
Basic Operation Working with Files
16 Source User’s Guide
Keysight M9410A/M9411A PXIe Vector Transceiver
Source User’s Guide

3 Optimizing Performance

Before using this information, you should be familiar with the basic operation of the transceiver source. If you are not comfortable with functions such as setting the power level and frequency, refer to Chapter 2, Basic Operation and familiarize yourself with the information in that chapter.
“Using an Output Offset and Reference” on page 18
“Using Amplitude Corrections” on page 20
17
Optimizing Performance
IF Amplifier
RF Amplifier
Mixer
Filter
Source (local oscillator)
Output Frequency = 1000 MHz
IF = 321 MHz
Antenna tuned to 1321 MHz
Selected Offset
321 MHz 679 MHz
Source Display 1321 MHz (Antenna Frequency)
321 MHz (IF Output)
IF Output 321 MHz

Using an Output Offset and Reference

Using an Output Offset and Reference

Setting an Output Offset

Using an output offset, the source can output a frequency or amplitude that is offset (positive or negative) from the entered value.
RF Output = entered value - offset value
Displayed Value = output frequency (or power) + offset value
To set an offset:
Frequency: Click Input/Output, RF Source, Frequency Setup, Freq Offset >
offset value > frequency unit keys.
Amplitude: Click Input/Output, RF Source, Amplitude Setup, Amptd
Offset > offset value > dB keys.
Examples
Parameter
Entered (and
Example#1Example#2Example
#3
Comments
300 MHz 300 MHz 2 GHz The entered value must be positive.
displayed) Value:
Offset: 50 MHz 50 MHz 1 GHz An offset value can be positive or negative.
The source alerts you if the
Output Frequency: 250 MHz 350 MHz 3 GHz
output frequency or amplitude is out of range.
When using the source as a local oscillator (LO), you can use the offset to display the frequency of interest, as illustrated below:
18 Source User’s Guide
Optimizing Performance Using an Output Offset and Reference

Setting an Output Reference

Using an output reference, the source can output a frequency or amplitude that is offset (positive or negative) by the entered value from a chosen reference value.
RF Output = reference value + entered value
To set a reference:
1. Set the frequency or amplitude to the value you want as the output reference level.
2. Frequency: Press Input/Output, RF Source, Frequency Setup, Set Reference Frequency. The frequency displays 0.00 Hz, indicating that this
is the RF output frequency “zero level.” All frequencies entered are interpreted as being relative to this reference frequency.
Amplitude: Press Input/Output, RF Source, Amplitude Setup, Set Reference Power.
The amplitude displays 0.00 dB, indicating that this is the RF output amplitude “zero level.” All amplitudes entered are interpreted as being relative to this reference amplitude.
Examples
Parameter
Reference: 50 MHz 50 MHz 2 GHz A reference value must be positive. Entered (and
displayed) Value:
Output Frequency: 52 MHz 48 MHz 1 GHz
Example#1Example#2Example
#3
2 MHz 2 MHz 1 GHz The entered value can be positive or negative.
The source alerts you if the output frequency or amplitude is out of range.
Comments
To set a new frequency or amplitude reference, turn the frequency reference off, and then follow the steps above.
Source User’s Guide 19
Optimizing Performance
RF Amplifier
Attenuator
transceiver
Source Output
Cable Loss
Receiving Device

Using Amplitude Corrections

Using Amplitude Corrections
Amplitude corrections can be applied to both source and analyzer ports on the transceiver. The Amplitude Corrections arrays can be entered by the user, sent over SCPI or loaded from a file. They allow a user to correct the response of the transceiver for variations in signal paths external to the transceiver. For example if you have a test system with a frequency dependent path due to cable, amplifier or attenuator non-linearities, similar to path illustrated below, then provided you can quantify the frequency variations, you can apply a frequency dependent correction to the transceiver source output. In this way the signal appearing at the receiving device can appear to have the spectral characteristics associated with a flat path.
The corrections are applied as a table of x and y parameters with the x specifying the frequency and y specifying the correction to be applied at the frequency, x. There are four sets of corrections and each may be individually enabled or disabled. All of the enabled corrections are added together to give the resultant correction. In the instrument, each port can have four different corrections applied to it, but two different ports cannot have the same set of corrections applied to both.
To access the Amplitude Corrections menu press the Input/Output key. Select the Corrections Tab.

Selecting a port

You can select whether you want to apply the corrections to one of the ports. In the case of the dual purpose ports, the corrections can be applied to the input, output or RFIO HD.
To select the port, from the Corrections tab, select Correction Port. You can select one of the three ports to which the selected corrections apply.
20 Source User’s Guide
Optimizing Performance Using Amplitude Corrections

Setting up a corrections table

To set up the corrections table using the front panel, from the Corrections tab, select Edit Correction. Click Insert Row Below to start the editing. The Frequency and Amplitude keys allow you to enter a frequency and the correction in dB that you require at that frequency. As you enter values the actual entries are displayed in tabular format on the left of the display and also, the entered values appear on the graticule as a graph. Because of the large number of possible entries in the corrections table, the Navigate key allows you to select an entry by row number to quickly access and edit a particular row.
For more detailed information about the correction data keys and the remote control commands see the online help, accessible by pressing the Help button on the front panel (or the User’s and Programmer’s Reference).
Source User’s Guide 21
Optimizing Performance Using Amplitude Corrections
22 Source User’s Guide
Keysight M9410A/M9411A PXIe Vector Transceiver
Source User’s Guide

4 Basic Digital Operation

Before using this information, you should be familiar with the basic operation of the transceiver source. If you are not comfortable with functions such as setting power level and frequency, refer to Chapter 2 and familiarize yourself with the information in that chapter.
“Waveform File Basics” on page 24
“Storing, Loading, and Playing a Waveform Segment” on page 25
“Saving a Waveform Settings & Parameters” on page 27
“Using Waveform Markers” on page 30
“Triggering a Waveform” on page 35
“Licensing” on page 38
23
Basic Digital Operation Waveform File Basics

Waveform File Basics

There are two types of waveform files:
—A segment is a waveform file that you download to the transceiver.
—A sequence is a file you create in the transceiver source that contains
pointers to one or more waveform files. The current M9410A/M9411A software only supports waveform segment operation and the waveform sequence operation will be supported in future.

Instrument Memory

The instrument has two types of memory:
Volatile memory, baseband generator (ARB) media, where waveform files
are played from or edited.
Non–volatile memory, the controller internal hard disk, where waveform
files are stored.

Dual ARB Player

The dual ARB waveform player enables you to play, rename, delete, store, and load (external or internal) waveform files. The dual ARB waveform player also provides markers (page 30) and triggering (page 35) capabilities.
24 Source User’s Guide
Basic Digital Operation Storing, Loading, and Playing a Waveform Segment

Storing, Loading, and Playing a Waveform Segment

Before using this information, you should be familiar with the source file menus. If you are not, refer to “Working with Files” on page 14 and familiarize yourself with that information.
The source has two types of waveform storage. The waveforms are stored on the controller hard disk which is non-volatile storage. In order to play the files out they must be moved into ARB memory which is volatile storage. ARB memory is also called “working” media, because before you can play, edit, or include a waveform file, the waveform file must be loaded into ARB memory.

Loading a Waveform Segment into ARB memory

Waveforms must reside in ARB memory before they can be played, or edited. Cycling power or rebooting the transceiver deletes the files in ARB memory.
1. Click file icon in the bottom bar, and click Recall key to access
the recall sub-menu.
2. Select Waveform in the left column, and click Recall From File to display files on the hard disk in C:\NVARB and then, highlight the desired waveform segment.
3. Press Recall to load this waveform segment from hard disk to ARB memory.
Or you could double click on the file name of desired waveform segment to load it to ARB memory directly.
Source User’s Guide 25
Basic Digital Operation Storing, Loading, and Playing a Waveform Segment

Deleting a Waveform Segment on Internal Disk

Use the following steps to delete a file on the hard disk.
1. Click file icon in the bottom bar, and click Recall key to access
the recall sub-menu.
2. Select Waveform in the left column, and click Recall From File to display files on the hard disk in C:\NVARB and then, right-click on the file name which you want to delete. A window named "Delete xxx.wfm" will pop up.
3. Press Delete xxx.wfm to delete the waveform file immediately.

Deleting a Waveform Segment from ARB memory

1. Click file icon in the bottom bar, and click Recall key to access
the recall sub-menu.
2. Select Waveform in the left column, and click Recall From File to display files on the hard disk in C:\NVARB and then, the waveform segments in ARB memory is listed in the left column of the window.
3. highlight the desired waveform segment which you want to delete.
4. Press Delete Segment From ARB Memory.
Alternatively, if you want to clear all ARB memory select Delete All From ARB Memory.

Playing a Waveform Segment

Waveforms must reside in ARB memory before they can be played, or edited. Cycling power or rebooting the transceiver deletes the files in ARB memory.
1. Select Input/Output key and RF Source tab.
2. Press Modulation Setup, ARB Setup, Select Waveform.
3. Select a waveform from the list in the display.
4. Press OK to execute the waveform.
5. In the Basic Control menu, turn ARB State to On to enable the modulation.
Then the waveform modulates the RF carrier.
6. Configure the RF Output:
Set the RF carrier frequency and amplitude, and turn on the RF output in Input/Output, RF Source menu.
The waveform segment is now available at the selected RF Output connector.
26 Source User’s Guide
Basic Digital Operation Saving a Waveform Settings & Parameters

Saving a Waveform Settings & Parameters

This section describes how to edit and save a file header. When you download only a waveform file (I/Q data, which the source treats as a waveform segment), the source automatically generates a file header and a marker file with the same name as the waveform file. Initially the file header has no source settings saved to it, and the marker file consists of all zeros. For a given waveform, you can save source settings and parameters in its file header and marker settings in its marker file (page 30); when you load a stored waveform file into ARB memory, the file header and marker file settings automatically apply to the source so that the dual ARB player sets up the same way each time the waveform file plays.
Some of the current source settings shown in the file header appear as part of the key labels, and others appear in the dual ARB summary display, shown in the following example. To view the file header:
1. Select Input/Output key and RF Source tab.
2. Press Modulation Setup, ARB Setup keys. Ensure that a waveform has been selected under the Select Waveform key.
3. Press Header Utilities key in the ARB Setup column. The header
information is shown as below.
Sample Rate The waveform playback rate. This is the ARB sample clock rate, set in the Arb Setup menu.
Runtime Scaling The Runtime scaling value is applied in real–time while the waveform is playing. This setting can be
changed only for files playing in the dual ARB player.
RMS When the modulator attenuation setting is set to Auto, this value is used to calculate the I/Q modulator
attenuation setting to optimize ACPR. Value: 0 to 1.414213562
Marker 1...4 Polarity
RF Blank Routing
Over Range Protect Indicated whether DAC Over–Range Protection is on (1) or off (0).
Unique Waveform Id 0 = no Id; once an Id is assigned, it cannot be changed.
License Required Indicates whether a license is required to play the waveform.
Marker polarity can be positive or negative (described on
Which marker, if any, implements the Pulse/RF Blanking function (described on marker signal is low. When the marker signal goes high, Pulse/RF Blanking discontinues.
page 33).
page 32) when the

Viewing and Modifying Header Information

1. From ARB memory, select the desired waveform:
a.Select Input/Output key and RF Source tab..
b.Press Modulation Setup, ARB Setup, Select Waveform.
c.In the display, select the desired waveform.
d.Press Select Waveform.
Source User’s Guide 27
Basic Digital Operation Saving a Waveform Settings & Parameters
2. Open the Header Utilities menu:
In the ARB Setup menu, select Header Utilities
The Figure below shows the default file header for the desired waveform.
The column with text in gray shows that most of the settings are Unspecified. Unspecified means that there is no setting saved for that particular parameter.
The column with text in white shows the current source settings. These are the settings that are saved to the file header.
Figure 4-1 Example File Header
If a setting is unspecified in the file header, the source uses its current value for that setting when you select and play the waveform.
3. Save the information in the white text column to the file header:
Press Save Header key.
Both the gray and white text column now display the same values; the gray text column lists the settings saved in the file header.
4. Edit and update settings
a.Set the ARB sample rate to 5 MHz:
Press Sample Rate, 5, MHz.
b.Set waveform runtime scaling to 60%:
Press Run-time Scaling, 60, Enter.
The white text column now reflects the changes to the current source setup, but gray text column have not changed.
28 Source User’s Guide
Basic Digital Operation Saving a Waveform Settings & Parameters
c.Save the current settings to the file header:
Press the Save Header key.
The settings from the white text column now appear in the gray text column. This saves the new current transceiver settings to the file header.
If you change any of the settings listed in the file header after you select the waveform file, the changed setting(s) appear in the file header white text column and are used instead of the saved header settings. To reapply the saved header settings, reselect the waveform for playback.
Source User’s Guide 29
Basic Digital Operation
Marker
File
Bit N
Marker
Polarity
Marker N
RF Blank Off On
Marker N
Blanks RF when
Marker is Low
EVENT N
Negative
Positive
Set Marker
On Off
Marker N
ALC Hold Off On
Marker N
Holds ALC when
Marker is Low
When the source encounters an enabled marker (described on page 24), an auxiliary output signal is generated and routed to the rear panel.
Events 1-4 are available at the TRIGGER 1 connector and at the TRIGGER 2 connector.
RF Blank Only: includes ALC Hold

Using Waveform Markers

Using Waveform Markers
The source provides four waveform markers to mark specific points on a waveform segment. When the source encounters an enabled marker, an auxiliary signal is routed to a rear panel trigger output. The marker that is routed to the TRIGGER 2 connector is selected using the Input/Output menu.
You can use the output signal to synchronize another transceiver with the waveform, or as a trigger signal to start a measurement at a given point on a waveform.
You can also configure markers to initiate Pulse/RF Blanking.
When you download a waveform file that does not have a marker file associated with it, the source creates a marker file without any marker points.
The following procedures demonstrate how to use markers while working with the ARB player. These procedures also discuss two types of points: a marker point and a sample point. A marker point is a point at which a given marker is set on a waveform; you can set one or more marker points for each marker. A sample point is one of the many points that compose a waveform.
This section also provides the following information:
“Waveform Marker Concepts
“Viewing a Marker Pulse” on page 31
“Using the Pulse/RF Blanking Marker Function” on page 32
“Setting Marker Polarity” on page 33

Waveform Marker Concepts

The source dual ARB provides four waveform markers for use on a waveform segment. You can set each marker polarity and marker points (on a single sample point or over a range of sample points). Each marker can also perform Pulse/RF Blanking.
30 Source User’s Guide
Basic Digital Operation Using Waveform Markers
Marker Signal Response
The source aligns the marker signals with the I and Q signals at the baseband generator. However some settings such as amplitude, filters, and so forth within the RF output path can create delays between the marker TRIGGER 2 output signal and the modulated RF output. When using the marker TRIGGER 2 output signal, observe the signals (marker relative to modulated RF) for any latency, and if needed, reset the marker point positions, include delay or both.
Marker File Generation
Downloading a waveform file that does not have a marker file associated with it causes the source to automatically create a marker file, but does not place any marker points.
Marker Point Edit Requirements
Before you can modify a waveform segment marker points, the segment must reside in ARB memory (see “Loading a Waveform Segment into ARB memory”
on page 25).
Saving Marker Polarity and Routing Settings
Marker polarity and routing settings remain until you reconfigure them, preset the source, or cycle power. To ensure that a waveform uses the correct settings when it is played, set the marker polarities or routing (Pulse/RF Blanking) and save the information to the file header (page 27).
When you use a waveform that does not have marker routings and polarity settings stored in the file header, and the previously played waveform used Pulse/RF Blanking, ensure that you set Pulse/RF Blanking to result in no RF output or a distorted waveform.

Selecting which Marker to output

To select which one of the four markers is output at the TRIGGER 2 connector:
1. Select Input/Output key and Trigger Ouput tab.
2. Press Trig 2 Out keys.
3. Press Trig 2 Out key, click Source Marker 1, Source Marker 2, Source Marker 3, or Source Marker 4 key as required to choose which marker to
output.
None. Failure to do so can

Viewing a Marker Pulse

When a waveform plays, you can detect an enabled marker pulse at the rear panel TRIGGER 2 connector. This example demonstrates how to view a marker pulse generated by a waveform segment that has at least one marker point set. The process is the same for a waveform sequence.
Source User’s Guide 31
Basic Digital Operation
RF OUT
Marker pulse on the TRIGGER 2.
Using Waveform Markers
This example uses a sine waveform segment in the dual ARB Player. Factory–supplied segments have a marker point on the first sample point for all four markers.
1. follow the procedure on page 25 to select and load the waveform segment file you need.
2. In the ARB Setup menu, Set ARB State to On.
3. Connect the instrument source output to the oscilloscope channel 1 input.
4. Connect the instrument TRIG 2 output to the oscilloscope channel 2
input. When marker 1 is present, the instrument outputs a signal through TRIG
2 as shown in the following example.

Using the Pulse/RF Blanking Marker Function

While you can set a marker function (described as Marker Routing on the key label in the Marker Utilities menu) either before or after setting the marker points, setting a marker function before you set marker points may change the RF output. The source blanks the RF output when the marker signal goes low. This example is a continuation of the previous example, “Viewing a Marker
Pulse” on page 31.
1. Using a segment, assume Marker 1 is set across points 1180.
2. From the Marker Routing key menu, assign Pulse/RF Blanking to Marker
1: In the ARB Setup menu, Click Basic Control, Pulse/RF Blank, Marker 1.
32 Source User’s Guide
Basic Digital Operation
Marker Polarity = Positive When marker polarity is positive (the default setting), the RF output is blanked during the off marker points.
200
180
Marker
Segment
RF Signal
RF Signal
200180
Marker
Segment
Marker Polarity = Negative When marker polarity is negative, the RF output is blanked during the on marker points
3.3
0 V
3.3
0 V
RF Signal
RF Signal
Using Waveform Markers

Setting Marker Polarity

Setting a negative marker polarity inverts the marker signal.
1. In ARB Setup menu, Click Marker Polarity to choose Negative or Positive.
2. For each marker, set the marker polarity as desired.
The default marker polarity is positive.
Each marker polarity is set independently.
See also, “Saving Marker Polarity and Routing Settings” on page 31.
As shown on page 32:
Positive Polarity: On marker points are high (3.3 V).
Negative Polarity: On marker points are low (0 V).
Source User’s Guide 33
Basic Digital Operation
The jitter is gone with an optimal sample rate
Waveform appears to exhibit jitter when triggered using EVENT signal with jitter.
Oscilloscope triggering on waveform
Oscilloscope triggering on EVENT signal
Oscilloscope triggering on EVENT signal
TRIGGER 2 signal exhibits jitter due to a non-optimal sample rate
Using Waveform Markers
Pulse/RF Blanking always occurs on the low part of the signal regardless of the polarity setting.

Using the TRIGGER 2 Signal as a Transceiver Trigger

One of the uses for the TRIGGER 2 signal (marker signal) is to trigger a measurement instrument. You can set up the markers to start the measurement at the beginning of the waveform, at any single point in the waveform, or on multiple points in the waveform. To optimize the use of the TRIGGER 2 OUT signal for measurements, you may also need to adjust the sample rate. To change the sample rate setting click Input/Output, RF Source, Modulation Setup, ARB Setup, Sample Rate keys.
The TRIGGER 2 output signal can exhibit jitter of up to ±4 ns on the rising and falling edge. This jitter can be minimized in either of two ways.
Method 1: Use a sample clock of 200 MHz/N where N is a positive integer and where 200 MHz/N can be represented exactly on the display. For example: 200 MHz, 100 MHz, 50 MHz, 25 MHz, and so on.
If the result cannot be represented exactly on the display, jitter will be present. For example: N = 6 will result in jitter, because 200 MHz/6 = 33.333 Mhz, which is truncated when displayed.
Method 2: Select a sample clock and waveform length that spaces the markers by a multiple of 8 ns. For example: A 200 point waveform with a marker on the first point and a sample clock of 50 MHz provides a marker every 4 µs. Because 4 µs is a multiple of 8 ns, the jitter is minimized.
When the TRIGGER 2 signal exhibits jitter and it is used as a measurement trigger, it can cause the waveform to falsely appear as having jitter. If this condition occurs, you can adjust the sample rate to a value (see above) that does not cause the jitter appearance. To maintain the integrity of the original waveform with a sample rate change, you will have to also recalculate the sample values. The following figures illustrate the marker signal jitter and its effect on the waveform.
34 Source User’s Guide
Basic Digital Operation Triggering a Waveform

Triggering a Waveform

Triggers control data transmission by controlling when the transceiver transmits the modulating signal. You can configure trigger settings so that data transmission occurs once (Single mode), or continuously (Continuous mode).
A trigger signal contains both positive and negative states; you can use either for triggering.
When you initially select a trigger mode or when you change from one triggering mode to another, you may lose the carrier signal at the RF output until the modulating signal is triggered. This is because the source sets the I and Q signals to zero volts prior to the first trigger event. To maintain the carrier signal at the RF output, create a data pattern with the initial I and Q voltages set to values other than zero.
When you initially turn the Arb State ON or select a trigger mode or when you change from one triggering mode to another, you may temporarily lose the carrier signal for a few tens of milliseconds at the RF output. The Arb will present the idle IQrms value of the next Arb waveform to the IQ modulator. This ensures that the RF carrier output is at the correct amplitude level while the Arb waits for a trigger. When that trigger is received, the Arb begins playing the waveform and the modulated RF carrier exhibits no undesirable transients.
There are two parts to configuring a waveform trigger:
Type determines the behavior of the waveform when it plays (see “Trigger
Type” on page 35).
Source determines how the source receives the trigger that starts the
modulating waveform playing (see “Trigger Source” on page 36).

Trigger Type

Type defines the trigger mode: how the waveform plays when triggered. Trigger Type is selected by clicking the Input/Output key, followed by the Modulation Setup, ARB Setup, Basic Control, Trig Type keys.
Continuous mode repeats the waveform until you turn the signal off or
select a different waveform, trigger mode, or response
Single mode plays the waveform once.
Free Run immediately triggers and plays the waveform; triggers
received while the waveform is playing are ignored.
Trigger and Run plays the waveform when a trigger is received;
subsequent triggers are ignored.
No Retrigger ignores triggers received while a waveform is playing. If a
trigger is received early it will be ignored. The gap in your playback is dependent on the trigger period, after which time the RF will start up again where it is expected.
Source User’s Guide 35
Basic Digital Operation Triggering a Waveform

Trigger Source

Trigger Source is selected by pressing the Input/Output key, followed by the Modulation Setup, ARB Setup, Basic Control, Trigger Source keys.
Key - the waveform is triggered when the Single key is pressed
Bus - the waveform is triggered over LAN or USB using the
:SOURce:RADio:ARB:TRIGger :INITitate SCPI command.
External 2 - the waveform is triggered by a trigger signal on the TRIGGER 1
port or the TRIGGER 2 port on the transceiver rear panel.
36 Source User’s Guide
Basic Digital Operation Source Alignment

Source Alignment

Use the Alignments menu to provide a manual signal alignment of the source. When you execute an alignment of the source. The transceiver stops any sequence of the source, presets the source, and performs the alignment.
Run the source alignment by pressing the System front-panel key, followed by Alignments, Align Now, Align Source. A message box appears advising you of the alignment process.
To abort the alignment, click on Cancel or press the Cancel (Esc) key.
Source User’s Guide 37

Licensing

Basic Digital Operation Licensing
You can create your own signal waveforms to output from the transceiver ARB using a number of methods. One of them is to use Keysight Signal Studio which is a suite of PC-based signal creation software that allows you to create customized signals for multiple radio standards. You can also create your own signals using other signal creation software (such as MATLAB or ADS) and download them to the transceiver hard disk. You can use Signal Studio Toolkit to process your customized files for use on the transceiver, if necessary. The instrument also provides a multi-pack licensing solution, a cost-effective solution that allows you to license individual Signal Studio waveforms rather than purchase the Signal Studio software.
Note that the Signal Studio files are encrypted waveforms; they are stored on the hard disk on the transceiver and are moved to the volatile ARB memory for playout. Other waveforms that you create may not be encrypted.

Signal Studio

Signal Studio software allows you to create technology-specific waveforms. It can be downloaded from the Keysight website at:
www.keysight.com/find/signalstudio
To use Signal Studio waveforms on the transceiver, you can either purchase a Signal Studio license or use waveform multi-pack licensing (see the next section). The Signal Studio website allows you to purchase the Signal Studio license and also provides trial licenses. The Signal Studio license is transceiver specific and therefore cannot be used on a transceiver other than the one specified in the purchase agreement. Once you have licensed the required Signal Studio software package, any waveforms that you create using this package can be used on the transceiver.

Waveform multi-pack licensing

Each license gives you certain number of slots for waveforms. For 45 slots, buy nine 5-pack licenses. For 500 slots, buy ten 50-pack licenses. For 545 slots, buy nine 5-pack licenses and ten 50-pack licenses.
Waveform multi-pack licensing (Option 2xx) enables you to create, generate and permanently license multiple Signal Studio waveforms without the need to purchase the full Signal Studio license.
A multi-pack license includes a number of license slots which are used to license waveform segments. The number of license slots varies depending on the multi-pack license option that has been purchased. When a waveform segment is licensed, it is assigned to one of the license slots. Waveform segments licensed with a multi-pack license are perpetual and cannot be exchanged. Once a waveform is licensed, that license is permanent and cannot be revoked or replaced.
38 Source User’s Guide
Basic Digital Operation Licensing
Signal Studio waveforms can be downloaded to the instrument and placed into a waveform slot for a 48 hour trial period. This allows you to edit and reuse this slot for 48 hours. After the 48 hour time period the STATUS of this slot will be changed to LOCK REQUIRED. This waveform slot now has been consumed and is no longer available for editing or reuse.
Option 2xx waveform licenses are transceiver specific (i.e. transceiver Host ID / PCSERNO). If the licensed waveform segment is transferred from one instrument to another the segment file must be re-licensed on the new instrument by a multi-pack license on that instrument before it can be played back.
To license additional waveforms that exceed the number permitted by a multi-pack option, you must purchase an additional multi-pack option that you do not already own. For example, if you have Option 221 and need additional waveforms, you should purchase Option 222. Re-purchasing Option 221 for the same transceiver provides no additional multi-pack licenses.
Waveforms licensed with Options 2xx cannot be exchanged for different waveforms. Once a waveform is licensed, that license is permanent and cannot be revoked or replaced.
After licensing a waveform, you can make copies of the waveform using different file names for use on the same transceiver and even rename the original file without affecting the waveform license.
You can also use the Option 2xx to license waveforms from N76xxB Signal Studio software downloaded during its 14–day free trial license. All of the N76xxB Signal Studio software products provide a 14–day trial period (trial license). This 14–day trial license lets you download and play back waveforms during the trial period. These waveforms are denoted by the TRL in the status message area of the waveform segment catalog. After the trial period expires, the TRL message is removed but the waveform remains. You can license these waveforms after the TRL message is gone.
To redeem an additional Option 2xx upgrade, refer to the Entitlement Certificate that comes with the instrument order.

Installing an Option N7650B–22x/25x Waveform 5/50–Pack Licensing

1. Load a Waveform 5–Pack license, Option N7650B–22x/25x, into the transceiver using License Manager or a USB media. For more information on loading the Waveform 5/50–Pack License, refer to the N7650B–22x/25x Entitlement Certificate.
Source User’s Guide 39
Basic Digital Operation Licensing

Licensing a Source Waveform File

1. Create the waveform:
a. Download any of the N76xxB Signal Stud io software that interests
you. For downloading N76xxB Signal Studio software, refer to the N76xxB Entitlement Certificate or trial website.
b. Create and download a waveform to the transceiver using any of the
N76xxB Signal Studio software. Refer to your Signal Studio software Help. You can add the waveform to a license slot for a 48-hour trial period. During the trial period, the waveform can be played and replaced any number of times. When the trial time expires, the slot can no longer be used for playback until the slot is locked for permanent playback capability. The slot status will
changed from "Remaining Trial Time" to "LOCK REQUIRED".
40 Source User’s Guide
Basic Digital Operation Licensing

Waveform Licensing Interface

Click file icon in the bottom bar, and click Save key to access the
file type selection menu.
Select Waveform in the left column, and click Recall From File to display files on the hard disk in C:\NVARB and then, highlight the desired waveform segment. Press Recall to load this waveform segment from hard disk to ARB memory.
The selected file name appears in the left column once it is loaded.
Go back to the ARB Setup menu. Click Waveform Utilities, Ad d Waveform, select the waveform you want to assign to a slot, and press OK.
Figure 4-2 Assigning the waveform to a slot
Source User’s Guide 41
Basic Digital Operation Licensing
The loaded waveform will be displayed with a status message of "Remaining Trial Time 48 Hours". Select the desired waveform and press Lock Waveform In Slot. A popup window asks you to confirm the lock operation; selecting OK will change the status of the waveform to "LOCKED [date]". This indicates that you have successfully licensed the waveform permanently.
Figure 4-3 Locking the waveform to the slot
42 Source User’s Guide
Basic Digital Operation Licensing
Waveform Multi–Pack Licensing Status Messages
The following messages appear in the Waveform Multi-Pack license display in the Status column. Table shows the meaning of the entries in the Status column.
Table 4-1 Waveform Multi–Pack Licensing Status Messages
Status Message Meaning Notes
Remaining Trial Time 48 Hours
Multi-Pack Licensed
Multi-Pack License Not Required
The waveform can be replaced or modified for the next 48 hours.
This waveform is licensed by Option 2xx.
This status message applies to:
Any free waveforms provided with the instrument
Any customer created waveform
Any waveforms that have a valid license (e.g. Trial (TRL) licenses, Advanced Design System (ADS), etc.).
Once the 48 hour period has expired, the waveform status message changes to "LOCK REQUIRED".
Once a Trial (TRL) license expires, the waveform becomes licensable (i.e. the status message for the trial waveform becomes "LOCK REQUIRED") .
Source User’s Guide 43
Basic Digital Operation Licensing
Other menu functions under Multi–Pack Licenses
Replace Waveform: This key allows you to replace the waveform in the
selected slot during the 48 hour trial period.
Clear Waveform from Slot: This key clears the waveform name from a used
slot so that the name can be used in another slot. Note that this does not make the original waveform slot reusable, it merely makes the waveform name reusable.
44 Source User’s Guide
Keysight M9410A/M9411A PXIe Vector Transceiver
Source User’s Guide

5 Troubleshooting

This chapter identifies some checks you can make if the source in the transceiver is not operating as expected. For information about non-source related problems see the Getting Started Guide.
“RF Output” on page 46
“No RF Output“Power Supply Shuts Down“No Modulation at the RF Output“RF Output Power too Low
“Error Messages” on page 47
“Licenses” on page 47
“Contacting Keysight Technologies” on page 48
45

RF Output

No RF Output

Power Supply Shuts Down

Troubleshooting RF Output
—Check the RF Output On/Off key (In RF Source tab). If it is off, turn the RF
Output On to turn the output on (selection is underlined).
— Ensure that the amplitude is set within the source range.
— If the transceiver is playing a waveform, ensure that marker polarity and
routing settings are correct (see “Saving Marker Polarity and Routing
Settings” on page 31).
If the power supply does not work, it requires repair or replacement. If you are unable to service the transceiver, send the transceiver to an Keysight service center for repair (see “Contacting Keysight Technologies” on page 48).

No Modulation at the RF Output

Check the Modulation On/Off key, and ensure it is set to on.

RF Output Power too Low

— Check to see if the amplitude offset is set and if necessary, reset to 0 dB:
Choose Input/Output from the drop-down panel followed by Amplitude Setup key of RF Source tab. Set Amptd Offset to 0 dB. See also “Setting an
Output Offset” on page 18.
46 Source User’s Guide
Troubleshooting Error Messages
— Check to see if power reference is set and turned on and if necessary, turn

Error Messages

A complete list of error messages is provided in PDF format on the transceiver (accessible from the Additional Documentation page of the online help) and also on the Keysight website (www.keysight.com/find/m9410a).

Licenses

off the reference mode
1.Choose Input/Output from the drop-down panel followed by Amplitude Setup key of RF Source tab. Press Power Ref to set to Off.
2.Reset the output power to the desired level.
See also “Setting an Output Reference” on page 19.

A time–based license stops working

— The transceiver time or date may have been reset forward causing the
time–based license to expire.
— The transceiver time or date may have been reset backward more than
approximately 25 hours, causing the transceiver to ignore time–based licenses.

Cannot load a time–based license

The transceiver time or date may have been reset backward more than approximately 25 hours, causing the transceiver to ignore time–based licenses.
Source User’s Guide 47
Troubleshooting Contacting Keysight Technologies

Contacting Keysight Technologies

— assistance with test and measurements needs, and information on finding a
local Keysight office:
http://www.Keysight.com/find/assist
— accessories, documentation, or new firmware releases:
www.keysight.com/find/m9410a
If you do not have access to the Internet, please contact your field engineer.
In any correspondence or telephone conversation, refer to the transceiver by its model number and full serial number. With this information, the Keysight representative can determine whether your unit is still within its warranty period.
48 Source User’s Guide
This information is subject to change without notice. © Keysight Technologies 2018
Edition 1, December 2018
M9410-90010
www.keysight.com
Loading...