No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation into
a foreign language) without prior agreement
and written consent from Keysight Technologies, Inc. as governed by United States and
international copyright laws.
Manual Part Number
54684-97035, October 2019
54684-97001, February 2005
54684-97003, April 2005
54684-97008, January 2006
54684-97010, June 2006
54684-97014, November 2006
54684-97018, March 2007
54684-97021, April 2007
54684-97026, March 2008
54684-97029, August 2008
54684-97033, January 2013
54684-97034, January 2017
54684-97035, October 2019
Published by:
Keysight Technologies, Inc.
1900 Garden of the Gods Road
Colorado Springs, CO 80907 USA
A newer version of this manual may
be available at
www.keysight.com/find/
6000manual
Warranty
The material contained in this document is
provided “as is,” and is subject to being
changed, without notice, in future editions.
Further, to the maximum extent permitted
by applicable law, Keysight disclaims all
warranties, either express or implied, with
regard to this manual and any information
contained herein, including but not limited
to the implied warranties of merchantability
and fitness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in
connection with the furnishing, use, or performance of this document or of any information contained herein. Should Keysight
and the user have a separate written agreement with warranty terms covering the
material in this document that conflict with
these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is "commercial computer
software," as defined by Federal Acquisition
Regulation ("FAR") 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of
Defense FAR Supplement ("DFARS")
227.7202, the U.S. government acquires
commercial computer software under the
same terms by which the software is
customarily provided to the public.
Accordingly, Keysight provides the Software
to U.S. government customers under its
standard commercial license, which is
embodied in its End User License Agreement
(EULA), a copy of which can be found at
www.keysight.com/find/sweula. The
license set forth in the EULA represents the
exclusive authority by which the U.S.
government may use, modify, distribute, or
disclose the Software. The EULA and the
license set forth therein, does not require or
permit, among other things, that Keysight: (1)
Furnish technical information related to
commercial computer software or
commercial computer software
documentation that is not customarily
provided to the public; or (2) Relinquish to, or
otherwise provide, the government rights in
excess of these rights customarily provided
to the public to use, modify, reproduce,
release, perform, display, or disclose
commercial computer software or
commercial computer software
documentation. No additional government
requirements beyond those set forth in the
EULA shall apply, except to the extent that
those terms, rights, or licenses are explicitly
required from all providers of commercial
computer software pursuant to the FAR and
the DFARS and are set forth specifically in
writing elsewhere in the EULA. Keysight shall
be under no obligation to update, revise or
otherwise modify the Software. With respect
to any technical data as defined by FAR
2.101, pursuant to FAR 12.211 and 27.404.2
and DFARS 227.7102, the U.S. government
acquires no greater than Limited Rights as
defined in FAR 27.401 or DFAR 227.7103-5
(c), as applicable in any technical data.
A CAUTION notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in personal
injury or death. Do not proceed
beyond a WARNING notice until the
indicated conditions are fully
understood and met.
26000L Series Oscilloscopes Service Guide
Page 3
In This Service Guide
This book provides the service information for the Keysight 6000L Series
oscilloscopes. This manual is divided into these chapters:
1Characteristics and Specifications
This chapter lists characteristics and specifications for the Keysight 6000L Series
oscilloscopes.
2Testing Performance
This chapter explains how to verify correct oscilloscope operation and perform
tests to ensure that the oscilloscope meets the performance specifications.
3Calibrating and Adjusting
This chapter explains how to adjust the oscilloscope for optimum operating
performance.
4Troubleshooting
This chapter begins with suggestions for solving general problems that you may
encounter with the oscilloscope. Procedures for troubleshooting the oscilloscope
follow the problem solving suggestions.
5Replacing 6000L Assemblies
This chapter describes how to remove assemblies from the 6000L Series
oscilloscope.
6Replaceable Parts
This chapter describes how to order replaceable assemblies and parts for the
Keysight 6000L Series oscilloscopes. It includes diagrams and parts lists for
hardware that you can order.
7Safety Notices
At the front of the book you will find safety notice descriptions and document
warranties.
6000L Series Oscilloscopes Service Guide3
Page 4
Using this book with the 6000L Series oscilloscopes
The 6000L Series oscilloscopes do not have a built-in display or front panel control keys. If you are
using a 6000L Series oscilloscope, and this book refers to using front panel controls, you can use the
built-in Web control feature described in the 6000L Series Oscilloscopes User’s Guide to complete the
instructions. If you do not have the 6000L Series Oscilloscopes User’s Guide, you may obtain a
printable electronic copy at www.Keysight.com/find/6000manual.
Digital Channels
Because all of the oscilloscopes in the Keysight 6000L Series have analog channels, the analog channel
topics in this book apply to all instruments. Whenever a topic discusses the digital channels, that
information applies to DSO models that have been upgraded to an MSO only.
Abbreviated instructions for pressing a series of keys
Instructions for pressing a series of keys are written in an abbreviated manner. Instructions for pressing
Key1, then pressing Softkey2, then pressing Softkey3 are abbreviated as follows:
Press [Key1] & Softkey2 & Softkey3.
The keys may be front panel keys, or softkeys, which are located directly below the oscilloscope
display.
Warranty Status
To find the warranty status of your oscilloscope:
1 Direct your web browser to: www.keysight.com/find/warrantystatus
2 Enter your product’s model number and serial number. The system will search
for the warranty status of your product and display the results. If the system
cannot find your product’s warranty status, select Contact Us and speak with an
Keysight Technologies representative.
46000L Series Oscilloscopes Service Guide
Page 5
Contents
Figures / 9
Tab les / 11
1Characteristics and Specifications
Power Requirements / 14
6000L Models / 14
Measurement Category / 15
Measurement Category / 15
Measurement Category Definitions / 15
6000L Series Maximum Input Voltages / 16
6000L Environmental Conditions / 17
Specifications / 18
Contact Us / 18
2Testing Performance
Overview / 20
Conventions / 20
List of Test Equipment / 21
To construct the test connector (for use with MSO models
only) / 22
To test digital channels (MSO models only) / 24
To verify digital channel threshold accuracy (MSO models
only) / 25
When to Test / 25
6000L Series Oscilloscopes Service Guide5
Page 6
Contents
What to Test / 25
Verifying Test Results / 25
To verify voltage measurement accuracy / 29
To verify bandwid th / 33
To verify time scale accuracy / 37
To verify horizontal delta t accuracy / 39
To verify trigger sensitivity / 41
Test Internal Trigger Sensitivity / 42
Test External Trigger Sensitivity / 43
Keysight 6000L Series Oscilloscopes Performance Test Record / 46
3Calibrating and Adjusting
User Calibration / 51
To perform User C al / 51
User Cal Status / 53
4Troubleshooting
Solving General Problems with the Oscilloscope / 56
Troubleshooting the Oscilloscope / 59
Equipment required for troubleshooting / 59
To check out the oscilloscope / 60
To verify basic oscilloscope operation / 62
To compensate the analog probes / 63
Troubleshooting Flowchart / 64
To check the 6000L Series oscilloscope power supply / 65
To check the 6000L Series system board / 67
To check the 6000L Series display output / 68
To check the fan / 68
To run the internal self-tests / 68
To verify default setup / 69
66000L Series Oscilloscopes Service Guide
Page 7
5Replacing 6000L Assemblies
To remove the bottom cover / 73
To remove the front panel assembly / 74
To remove the system board / 76
To remove the power supply / 79
To remove the fan / 81
To remove the AC power input assembly / 83
To remove the power shaft / 85
6Replaceable Parts
Ordering Replaceable Parts / 88
Listed Parts / 88
Unlisted Parts / 88
Direct Mail Order System / 88
Exchange Assemblies / 89
Contents
Replaceable Parts for 6000L Series Oscilloscopes / 90
6000L Series Oscilloscope Exploded Views / 90
6000L Series Oscilloscope Replaceable Parts List / 92
7Safety Notices
Warnings / 96
To clean the instr ument / 97
Safety Symbols / 98
Index
6000L Series Oscilloscopes Service Guide7
Page 8
Contents
86000L Series Oscilloscopes Service Guide
Page 9
Figures
Figure 1. Constructing the 8-by-2 Connector / 23
Figure 2. Setting Up Equipment and Test Connector for the Threshold Test / 26
Figure 3. Connect equipment for voltage measurement accuracy test / 31
Figure 4. Using a Blocking Capacitor to Reduce Noise / 32
Figure 5. Connect equipment for bandwidth test / 34
Figure 6. Connect equipment for internal trigger sensitivity test / 42
Figure 7. Connect equipment for external trigger sensitivity test / 44
Figure 8. User Calibration cable for 4-channel oscilloscope / 52
Figure 9. 6000L Series LED indicators / 61
Figure 10. Example pulses / 63
Figure 11. System Board Test Points / 66
Figure 12. Default setup screen / 70
Figure 13. Removing the cover / 73
Figure 14. Removing the intensity knob / 74
Figure 15. Disconnecting ribbon cable and releasing tab retainers / 75
Figure 16. Preparing to remove the system board / 76
Figure 17. Avoid damage to thermal pads (500 MHz models only) / 77
Figure 18. Removing the power supply. / 79
Figure 19. Removing the fan. / 81
Figure 20. Installing the fan. / 82
Figure 21. Removing the AC power input board. / 84
Figure 22. Removing the power shaft. / 85
Figure 23. 6000L Series Oscilloscope Exploded View / 91
6000L Series Oscilloscopes Service Guide9
Page 10
Figures
106000L Series Oscilloscopes Service Guide
Page 11
Tables
Table 1. Conventions / 20
Table 2. List of test equipment / 21
Table 3. Materials required to construct the test connectors / 22
Table 4. Equipment Required to Test Digital Channel Threshold Accuracy / 25
Table 5. Threshold Accuracy Voltage Test Settings / 27
Table 6. Equipment Required to Verify Voltage Measurement Accuracy / 29
Table 7. Settings Used to Verify Voltage Measurement Accuracy / 30
Table 8. Equipment Required to Verify Bandwidth / 33
Table 9. Equipment Required to Verify Time Scale Accuracy / 37
Table 10. Equipment Required to Verify Horizontal Dt Accuracy / 39
Table 11. Equipment Required to Verify Trigger Sensitivity / 41
Table 12. Equipment Required to Troubleshoot the Oscilloscope / 60
Table 13. System Board Test Points / 67
Table 14. 6000L Series Oscilloscope Replaceable Parts / 92
6000L Series Oscilloscopes Service Guide11
Page 12
Tables
126000L Series Oscilloscopes Service Guide
Page 13
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide
1Characteristics and
Specifications
Power Requirements / 14
Measurement Category / 15
6000L Environmental Conditions / 17
Specifications / 18
Contact Us / 18
This chapter contains a partial list of characteristics and specifications for the
Keysight InfiniiVision 6000L Series Oscilloscopes.
For a full list of Keysight InfiniiVision 6000L Series Oscilloscopes characteristics
and specifications see the data sheet.
The data sheet is available at www.keysight.com/find/6000.
13
Page 14
1Characteristics and Specifications
Power Requirements
6000L Models
Line voltage, frequency, and power:
• ~Line 100-240 Vac,
• 50/60 Hz
•80W max
146000L Series Oscilloscopes Service Guide
Page 15
Measurement Category
WARNING
Measurement Category
The InfiniiVision oscilloscopes are not intended to be used for measurements in
Measurement Category II, III, or IV.
Use this instrument only for measurements within its specified measurement category
(not rated for CAT II, III, IV). No transient overvoltages allowed.
Measurement Category Definitions
The "Not rated for CAT II, III, IV" measurement category is for measurements
performed on circuits not directly connected to MAINS. Examples are
measurements on circuits not derived from MAINS, and specially protected
(internal) MAINS derived circuits. In the latter case, transient stresses are variable;
for that reason, the transient withstand capability of the equipment is made known
to the user.
Measurement category II is for measurements performed on circuits directly
connected to the low voltage installation. Examples are measurements on
household appliances, portable tools and similar equipment.
Characteristics and Specifications1
6000L Series Oscilloscopes Service Guide15
Measurement category III is for measurements performed in the building
installation. Examples are measurements on distribution boards, circuit-breakers,
wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the
fixed installation, and equipment for industrial use and some other equipment, for
example, stationary motors with permanent connection to the fixed installation.
Measurement category IV is for measurements performed at the source of the
low-voltage installation. Examples are electricity meters and measurements on
primary overcurrent protection devices and ripple control units.
Page 16
1Characteristics and Specifications
CAUTION
CAUTION
6000L Series Maximum Input Voltages
Maximum input voltage at analog inputs
Do not exceed the maximum input voltage rating.
The maximum input voltage for the 50 Ω input impedance setting is ±5 V.
Input protection is enabled in 50 Ω mode and the 50 Ω load will disconnect if greater
than 5 Vrms is detected. However the inputs could still be damaged, depending on the
time constant of the signal. The 50 Ω input protection functions only when the
oscilloscope is powered on.
The maximum input voltage for the 1 MΩ input impedance setting is 30 Vrms or
±40 Vmax (DC+Vpeak)
Probing technology allows for testing of higher voltages; the included 10073D 10:1 probe
supports 300 Vrms or ±400 Vmax (DC+Vpeak). No transient overvoltage allowed.
Maximum input voltage at digital channels
±40 V peak
166000L Series Oscilloscopes Service Guide
Page 17
6000L Environmental Conditions
EnvironmentIndoor use only.
Ambient temperature Operating –10 °C to +40 °C; non-operating –40 °C to
+70 °C
Humidity Operating 95% RH at 40 °C for 24 hr; non-operating 90% RH at 65 °C
for 24 hr
Altitude Operating to 2,000 m (6,561 ft); non-operating to 15,244 m (50,000 ft)
Overvoltage CategoryThis product is intended to be powered by MAINS that
comply to Overvoltage Category II, which is typical of cord-and-plug connected
equipment.
Pollution Degree The InfiniiVision 6000L Series oscilloscopes may be operated
in environments of Pollution Degree 2 (or Pollution Degree 1).
Characteristics and Specifications1
Pollution Degree DefinitionsPollution Degree 1: No pollution or only dry,
non-conductive pollution occurs. The pollution has no influence. Example: A clean
room or climate controlled office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation may occur.
Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive pollution
occurs which becomes conductive due to condensation which is expected.
Example: Sheltered outdoor environment.
6000L Series Oscilloscopes Service Guide17
Page 18
1Characteristics and Specifications
Specifications
Please see the InfiniiVision 6000L Series Oscilloscopes Data Sheet for complete,
up-to-date specifications and characteristics.
To download a copy of the data sheet please visit:
www.keysight.com/find/6000.
Or go to the Keysight home page at www.keysight.com and search for 6000L series
oscilloscopes data sheet.
To order a data sheet by phone, please contact your local Keysight office.
Contact Us
To contact Keysight, see: www.keysight.com/find/contactus
186000L Series Oscilloscopes Service Guide
Page 19
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide
2Testing Performance
Overview / 20
List of Test Equipment / 21
To construct the test connector (for use with MSO models only) / 22
To test digital channels (MSO models only) / 24
To verify digital channel threshold accuracy (MSO models only) / 25
To verify voltage measurement accuracy / 29
To verify bandwidth / 33
To verify time scale accuracy / 37
To verify horizontal delta t accuracy / 39
To verify trigger sensitivity / 41
Keysight 6000L Series Oscilloscopes Performance Test Record / 46
This chapter explains how to verify correct oscilloscope operation and perform
tests to ensure that the oscilloscope meets the performance specifications.
19
Page 20
2Testing Performance
Overview
To completely test and troubleshoot the mixed-signal oscilloscope, you will create
and use a test connector accessory, as described in this chapter.
• The test connector is only required for oscilloscopes that have the MSO option
licensed (enabled).
• The connector is used in the digital channel threshold accuracy test.
• The test connector makes it easy for you to connect the oscilloscope probes to
function generators and measurement equipment with minimum electrical
distortion.
Let the Equipment Warm Up Before Testing
For accurate test results, let the test equipment and the oscilloscope warm up 30
minutes before testing.
Verifying Test Results
During the tests, record the readings in the Performance Test Record on page 46.
To verify whether a test passes, verify that the reading is within the limits in the
Performance Test Record.
If a performance test fails
If a performance test fails, first perform the User Cal procedure
Conventions
The following conventions will be used when referring to oscilloscope models
throughout this chapter.
Table 1Conventions
Models Referred to as:
DSO6014L100 MHz Models
DSO6054L500 MHz Models
DSO6104L1 GHz Models
206000L Series Oscilloscopes Service Guide
Page 21
Testing Performance2
List of Test Equipment
Below is a list of test equipment and accessories required to perform the
performance test verification procedures.
Table 2List of test equipment
Recommended Model/
EquipmentCritical Specifications
Tes t conn ector , 8-by- 2*Se e page 22 for instructions on building test connector.n/a
Digital Multimeter0.1 mV resolution, 0.005% accuracyKeysight 34401A
Power SplitterOutputs differ by 0.15 dBKeysight 11667B
Oscilloscope CalibratorDC offset voltage of -5.5 V to 35.5 V, 0.1 V resolutionFluke 5820A
Signal Generator25 MHz, 100 MHz, 500 MHz, and 1 GHz sine wavesKeysight N5181A, or
Power Meter/Sensor1 GHz
Oscilloscope Calibrator25 MHz—500 MHz sine wave, 5 ppmFluke 5820A
BNC banana cableKeysight 11001-66001
BNC cable (qty 3)Keysight 10503A
CableType N (m) 609.6 mm (24 in.)Keysight 11500B
Probe cable*Keysight 01650-61607
Shorting Cap BNCKeysight 1250-0774
AdapterBNC(f) to banana(m)Keysight 1251-2277
AdapterBNC Tee (m) (f) (f)Keysight 1250-0781
AdapterType N (m) to BNC (m)Keysight 1250-0082
Blocking capacitorKeysight 10240-60001
Adapter (qty 3)N(m) to BNC(f)Keysight 1250-0780
Feedthrough (qty 2)50
* Required only for testing digital channels of oscilloscopes that have the MSO option.
Some parts and equipment are available at www.keysight.com or www.parts.keysight.com.
±3% accuracyKeysight E4418B/8482A
Ω BNC (f) to BNC (m)Keysight 0960-0301
Part Number
E4400B, or 8648A
or Pomona 3288 with
Pomona 3533
6000L Series Oscilloscopes Service Guide21
Page 22
2Testing Performance
To construct the test connector (for use with MSO models only)
Keysight 6000L Series Oscilloscopes that have digital channels enabled require
the test connector described below. Follow the steps to build the test connector.
Table 3Materials required to construct the test connectors
DescriptionRecommended PartQty
BNC (f) ConnectorKeysight 1250-1032 or
Pomona 4578
Berg Strip, 8-by-21
Jumper wire
1 Obtain a BNC connector and an 8-by-2 section of Berg strip.
2 On one side of the Berg strip, solder a jumper wire to all of the pins (shown in
Figure 1 on page 23).
3 On the other side of the Berg strip, solder another jumper wire to all of the pins.
4 Solder the center of the BNC connector to a center pin on one of the rows on
the Berg strip.
5 Solder the ground tab of the BNC connector to a center pin on the other row on
the Berg strip.
1
226000L Series Oscilloscopes Service Guide
Page 23
Figure 1Constructing the 8-by-2 Connector
Jumper (2)
Signal Lead
(from scope’s MSO cable)
8 x 2 Berg Strip
BNC Panel Mount Connector
Ground Lead
(from scope’s MSO cable)
Testing Performance2
6000L Series Oscilloscopes Service Guide23
Page 24
2Testing Performance
To test digital channels (MSO models only)
The acquisition system testing provides confidence that the acquisition system is
functioning correctly. It does not, however, check a particular specification.
1 Disconnect all probes from the circuit under test and from any other input
source.
2 Using probe leads and grabbers, connect digital channels D0, D1, D2, and D3
to the Probe Comp signal on the center of the front panel.
3 Press the AutoScale key.
If four square waves appear, the acquisition system is functioning correctly.
If the square waves do not appear, go to the “Troubleshooting” chapter. Then
return here to finish testing the digital channels.
4 Disconnect the digital channels from the calibration point.
5 Use steps 2 and 3 to test the following sets of digital channels. After you test
one set of digital channels, remove them before connecting the next set.
• D4, D5, D6, D7
• D8, D9, D10, D11
• D12, D13, D14, D15
246000L Series Oscilloscopes Service Guide
Page 25
Testing Performance2
To verify digital channel threshold accuracy (MSO models only)
This test verifies the digital channel threshold accuracy specification of the
Keysight 6000L Series Oscilloscopes.
Threshold accuracy test limits= ±(100 mV + 3% of threshold setting)
When to Test
You should perform this test every 12 months or after 2000 hours of operation,
whichever comes first.
What to Test
Use these instructions to test the threshold settings of digital channels D7-D0.
Then, use the same instructions to test digital channels D15-D8.
Verifying Test Results
After each threshold test, record the voltage reading in the Performance Test
Record on page 46. To verify whether a test passes, verify that the voltage reading
is within the limits in the Performance Test Record.
Table 4Equipment Required to Test Digital Channel Threshold Accuracy
Oscilloscope CalibratorDC offset voltage 6.3 VFluke 5820A
BNC-Banana CableKeysight 11001-60001
BNC TeeKeysight 1250-0781
BNC CableFluke 50
BNC Test Connector, 8-by-2User-built (See page 22.)
Probe CableKeysight 01650-61607
6000L Series Oscilloscopes Service Guide25
Keysight 34401A
Ω cable, P/N 686318
Page 26
2Testing Performance
Test
Connector
BNC-Banana
cable
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
Probe
Cables
Channels
8 - 15
Channels
0 - 7
6000L Series Oscilloscope
1 Turn on the test equipment and the oscilloscope. Let them warm up for 30
2 Set up the oscilloscope calibrator.
3 Use the 8-by-2 test connector and the BNC cable assembly to connect digital
minutes before starting the test.
a Set the oscilloscope calibrator to provide a DC offset voltage at the
Channel 1 output.
b Use the multimeter to monitor the oscilloscope calibrator DC output
voltage.
channels D0-D7 to one side of the BNC Tee. Then connect the D0-D7 ground
lead to the ground side of the 8-by-2 connector. See Figure 2.
Figure 2Setting Up Equipment and Test Connector for the Threshold Test
266000L Series Oscilloscopes Service Guide
Page 27
Testing Performance2
4 Use a BNC-banana cable to connect the multimeter to the other side of the
BNC Tee.
5 Connect the BNC Tee to the Channel 1 output of the calibrator as shown in
Figure 2.
6 On the oscilloscope, press the D15 Thru D0 key, then press the Thresholds
softkey, then press the D15 Thru D0 softkey repeatedly until the check mark is
next to User.
7 Press the oscilloscope User softkey, then turn the Entry knob () on the front
panel on the oscilloscope to set the threshold test settings as shown in Table 5.
Table 5Threshold Accuracy Voltage Test Settings
Threshold voltage setting
(in oscilloscope User softkey)
+5.00 V+5.250 V ±1 mV dc Lower limit = +4.750 V
–5.00 V–4.750 V ±1 mV dc Lower limit = –5.250 V
0.00 V+100m V ±1 mV dc Upper limit = +100 mV
8 Do the following steps for each of the threshold voltage levels shown in Table 5.
a Set the threshold voltage shown in the User softkey using the Entry knob on
the oscilloscope.
b Enter the corresponding DC offset voltage on the oscilloscope calibrator
front panel. Then use the multimeter to verify the voltage.
Digital channel activity indicators are displayed on the status line at the top
of the oscilloscope display. The activity indicators for D7-D0 should show all
of the channels at digital high levels.
c Use the knob on the oscilloscope calibrator to decrease the offset voltage, in
increments of 10 mV, until the activity indicators for digital channels D7-D0
are all at digital low levels. Record the oscilloscope calibrator voltage in the
Performance Test Record (see page 46).
d Use the knob on the oscilloscope calibrator to increase the offset voltage, in
increments of 10 mV, until the activity indicators for digital channels D7-D0
DC offset voltage setting
(on oscilloscope calibrator)Limits
Upper limit = +5.250 V
Upper limit = –4.750 V
Lower limit = –100 mV
6000L Series Oscilloscopes Service Guide27
Page 28
2Testing Performance
9 Use the 8-by-2 test connector to connect digital channels D15-D8 to the
10 Repeat this procedure (steps 7 and 8) for digital channels D15-D8 to verify
are all at digital high levels. Record the oscilloscope calibrator voltage in the
Performance Test Record (see page 46).
Before proceeding to the next step, make sure that you have recorded the
oscilloscope calibrator voltage levels for each of the threshold settings shown
in Tab le 5.
output of the oscilloscope calibrator. Then connect the D15-D8 ground lead to
the ground side of the 8-by-2 connector.
threshold accuracy and record the threshold levels in the Performance Test
Record (see page 46).
286000L Series Oscilloscopes Service Guide
Page 29
To verify voltage measurement accuracy
This test verifies the accuracy of the analog channel voltage measurement for
each channel (DC Vertical Gain Accuracy and Dual Cursor Accuracy
specifications). In this test, you will measure the dc voltage output of an
oscilloscope calibrator using dual cursors on the oscilloscope, and compare the
results with the multimeter reading.
Test limits: ±2.0% of full scale ±1 LSB*
• On 500 MHz and 1 GHz models, full scale is defined as 32 mV on the 2 mV/div
range.
• On 100 MHz models full scale is defined as 16 mV on the 1 mV/div range.
• Full scale on all other ranges is defined as 8 divisions times the V/div setting.
*1 LSB = 0.4% of full scale
Table 6Equipment Required to Verify Voltage Measurement Accuracy
Digital multimeterBetter than 0.01% accuracyKeysight 34401A
CableBNC, Qty 2Keysight 10503A
Shorting capBNCKeysight 1250-0774
AdapterBNC (f) to banana (m)Keysight 1251-2277
Adapter BNC tee (m) (f) (f )Keysight 1250-0781
Blocking capacitorKeysight 10240B
1 Set up the oscilloscope.
a Adjust the channel 1 position knob to place the baseline at 0.5 major division
from the bottom of the display.
b Set the Volts/Div setting to the value in the first line in Table 7.
6000L Series Oscilloscopes Service Guide29
Fluke 5820A
Page 30
2Testing Performance
Table 7Settings Used to Verify Voltage Measurement Accuracy
Volts/Div Setting Oscilloscope Calibrator
Setting
5 V/Div35 V34.04 Vto35.96 V
2 V/Div14 V13.616 Vto14.384 V
1 V/Div7 V6.808 Vto7.192 V
500 mV/Div3.5 V3.404 Vto3.596 V
200 mV/Div1.4 V1.3616 Vto1.4384 V
100 mV/Div700 mV680.8 mVto719.2 mV
50 mV/Div350 mV340.4 mVto359.6 mV
20 mV/Div140 mV136.16 mVto143.84 mV
10 mV/Div70 mV68.08 mVto71.92 mV
5 mV/Div35 mV34.04 mVto35.96 mV
2mV/Div1 (for 1 GHz and 500 MHz models)
2mV/Div (for 100 MHz models)
1mV/Div
1
2
Full scale is defined as 32 mV on the 2 mV/div range for 500 MHz and 1 GHz models.
14 mV13.232 mVto14.768 mV
14 mV13.616 mVto14.384 mV
7 mV6.616 mVto7.384 mV
Test Limits
2
Full scale is defined as 16 mV on the 1 mV/div range for 100 MHz models.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
c Press the Acquire key. Then press the Averaging softkey and set #Avgs to 64.
Wait a few seconds for the measurement to settle.
2 Press the Cursors key, set the Mode softkey to Normal, then press the XY softkey
and select Y. Press the Y1 softkey, then use the Entry knob (labeled on the
front panel) to set the Y1 cursor on the baseline of the signal.
3 Use the BNC tee and cables to connect the oscilloscope calibrator /power
supply to both the oscilloscope and the multimeter (see Figure 3).
306000L Series Oscilloscopes Service Guide
Page 31
Testing Performance2
Oscilloscope
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
BNC (f) to dual
banana
Figure 3Connect equipment for voltage measurement accuracy test
4 Adjust the output so that the multimeter reading displays the first Volts/div
calibrator setting value in Tab le 7.
Wait a few seconds for the measurement to settle.
5 Press the Y2 softkey, then position the Y2 cursor to the center of the voltage
trace using the Entry knob.
The ΔY value on the lower line of the display should be within the test limits
of Tab le 7. If a result is not within the test limits, go to the “Troubleshooting”
chapter. Then return here.
6000L Series Oscilloscopes Service Guide31
6 Continue to check the voltage measurement accuracy with the remaining
Volts/div setting values in Table 7 .
Page 32
2Testing Performance
To oscilloscope input
BNC shorting
cap
Blocking
Capacitor
7 When you are finished checking all of the voltage values, disconnect the
8 Repeat this procedure for the remaining channels to be tested.
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 2 mV/div and 5 mV/div, noise may be a factor. To eliminate the
noise, add a BNC Tee, blocking capacitor, and BNC shorting cap at the oscilloscope channel input to
shunt the noise to ground. See Figure 4.
oscilloscope calibrator from the oscilloscope.
Figure 4Using a Blocking Capacitor to Reduce Noise
326000L Series Oscilloscopes Service Guide
Page 33
To verify bandwidth
This test checks the bandwidth of the oscilloscope. In this test you will use a
signal generator and a power meter.
Signal Generator 100 kHz - 1 GHz at 200 mVrmsKeysight N5181A, or E4400B,
or 8648A
Power Meter/Sensor1 MHz - 1 GHz ±3% accuracyKeysight E4418B/8482A
Power Splitteroutputs differ by < 0.15 dB Keysight 11667A
Cable Type N (m) 24 inch Keysight 11500B
AdapterType N (m) to BNC (m)Keysight 1250-0082
Feedthrough50Ω BNC (f) to BNC (m)Keysight 0960-0301
6000L Series Oscilloscopes Service Guide33
Page 34
2Testing Performance
Oscilloscope
Signal
Generator
Power Meter
Power Splitter
Power Sensor
N to BNC adapter
N Cable
50Ω Feedthrough*
* 100 MHz models only.
1 Connect the equipment (see Figure 5).
a Use the N cable to connect the signal generator to the input of the power
splitter input.
b Connect the power sensor to one output of the power splitter.
c For 500 MHz and 1 GHz oscilloscope models use an N-to-BNC adapter to
connect the other splitter output to the channel 1 input.
d For 100 MHz oscilloscopes, use an N-to-BNC adapter and 50Ω feedthrough
termination to connect the other splitter output to the channel 1 input on
the oscilloscope.
Figure 5Connect equipment for bandwidth test
2 Set up the power meter.
346000L Series Oscilloscopes Service Guide
Page 35
Testing Performance2
Vin
1MHz
Pmeas
1MHz
50Ω×=
Set the power meter to display measurements in units of watts.
3 Set up the oscilloscope.
a Press the Save/Recall key, then press the Default Setup softkey.
b Press the Acquire key, then press the Realtime softkey to unselect Realtime.
c Set channel 1 Coupling to DC.
d On 500 MHz and 1 GHz models only, set channel 1 Imped to 50 Ohm.
e Set the time base to 500 ns/div.
f Set the Volts/Div for channel 1 to 200 mV/div.
g Press the Acquire key, then press the Averaging softkey.
h Turn the Entry knob to set # Avgs to 8 averages.
4 Set the signal generator for 1 MHz and six divisions of amplitude.
The signal on the oscilloscope screen should be about five cycles at six
divisions amplitude.
5 Set up the Amplitude measurement
a Press the Quick Meas key.
b Press the Clear Meas softkey.
c Press the Select: softkey and use the Entry knob to select Std Deviation within
the select menu.
d Press the Measure Std Dev softkey.
6 Note the oscilloscope Std Dev(1) reading at the bottom of the screen. (This is
the RMS value with any dc offset removed.)
7 Set the power meter Cal Factor % to the 1 MHz value on the calibration chart
on the power sensor.
8 Note the reading on the power meter and covert to Vrms using the expression:
For example, if the power meter reading is 892 uW, then Vin
1/2
50Ω)
9 Change the signal generator output frequency according to the maximum
frequency for the oscilloscope using the following:
• 1GHz Models: 1GHz
• 500 MHz Models: 500 MHz
6000L Series Oscilloscopes Service Guide35
= 211.2 mV
rms
.
= (892*10-6 *
1MHz
Page 36
2Testing Performance
Vin
maxfreq
Pmeas
maxfreq
50Ω×=
response(dB) = 20 log
10
freqmax freqmax
Vin / Vout
Vin / Vout
response(dB) = 20 log
10
mV 185.3 / mV 161.6
= -1.16 dB
ExampleIf:
• 100 MHz Models: 100 MHz
10 Referencing the frequency from step 9, set the power meter Cal Factor % to the
frequency value on the calibration chart on the power sensor.
11 Set the oscilloscope sweep speed according to the following:
• 1 GHz Models: 500 ps/div
• 500 MHz Models: 1 ns/div
• 100 MHz Models: 5 ns/div
12 Note the oscilloscope Std Dev(1) reading at the bottom of the screen.
13 Note the reading on the power meter and covert to Vrms using the expression:
14 Calculate the response using the expression:
MHz1MHz1
Pmeas
Std Dev(n)
Pmeas
Std Dev(n)
Then after converting the values from the power meter to Vrms:
1_MHz
1MHz
max_freq
max freq
= 892 uW
= 210.4 mV
= 687 uW
= 161.6 mV
366000L Series Oscilloscopes Service Guide
mV211.2mV/ 210.4
15 The result from step 14 should be within -3.0 dB. Record the result in the
Performance Test Record (see page 46).
16 Move the power splitter from the channel 1 to the channel 2 input.
17 Turn off the current channel and turn on the next channel using the channel
keys.
18 Repeat steps 3 through 17 for the remaining channels, setting the parameters
of the channel being tested where appropriate.
Page 37
To verify time scale accuracy
Date Code:
0747
07 = Year
47 = Week
Time Scale Accuracy is a warranted specification for 6000L Series oscilloscopes
manufactured after January 1, 2008.
If your oscilloscope’s serial tag has old-style (linear) barcodes instead of the 3-D
barcodes shown above, or if the date indicated on the serial tag is older than
January 1, 2008, then disregard this test.
This test verifies the accuracy of the time scale. In this test you will measure the
absolute error of the timebase oscillator and compare the results to the
specification.
Testing Performance2
Table 9Equipment Required to Verify Time Scale Accuracy
Signal Generator 100 kHz - 1 GHz, 0.01 Hz frequency
resolution,
jitter: < 2ps
Cable BNC, 3 feetKeysight 10503A
1 Set up the signal generator.
a Set the output to 10 MHz, approximately 1 V
2 Connect the output of the signal generator to oscilloscope channel 1 using the
BNC cable.
6000L Series Oscilloscopes Service Guide37
Keysight N5181A, or E4400B,
or 8648A
sine wave.
pp
Page 38
2Testing Performance
Date Code:
0747
07 = Year
47 = Week
3 Set up the oscilloscope:
a Press AutoScale.
b Set the oscilloscope Channel 1 vertical sensitivity to 200 mv/div.
c Set the oscilloscope horizontal sweep speed control to 5 ns/div.
d Adjust the intensity to get a sharp, clear trace.
e Adjust the oscilloscope’s trigger level so that the rising edge of the
waveform at the center of the screen is located where the center horizontal
and vertical grid lines cross (center screen).
f Ensure the horizontal position control is set to 0.0 seconds.
4 Make the measurement.
a Set oscilloscope horizontal sweep speed control to 1 ms/div.
b Set horizontal position control to +1 ms (rotate control CCW).
c Set the oscilloscope horizontal sweep speed control to 5 ns/div.
d Record the number of nanoseconds from where the rising edge crosses the
center horizontal grid line to the center vertical grid line. The number of
nanoseconds is equivalent to the time scale error in ppm.
e Use the date code on the oscilloscope’s serial tag to calculate the number of
years since manufacture. Include any fractional portion of a year.
f Calculate the test limits using the following formula:
TSA Limit: ± (15 + 2 * YrsSinceMfr) ppm
g Record the result and compare it to the limits in the Performance Test
Record (see page 46).
386000L Series Oscilloscopes Service Guide
Page 39
To verify horizontal delta t accuracy
The following test is optional because horizontal Δt accuracy is a characteristic,
not a specification.
This test verifies the horizontal Δt accuracy. In this test, you will use the
oscilloscope to measure the output of a time mark generator.
Test limits: ±0.0015% of reading ±0.1% of full scale ±20 ps (same channel)
Table 10Equipment Required to Verify Horizontal Δt Accuracy
a For 500 MHz and 1 GHz oscilloscope models, connect the calibrator output
to the oscilloscope channel 1 input.
b For 100 MHz oscilloscopes, use a 50Ω feedthrough termination to connect
the calibrator output to the oscilloscope channel 1 input.
2 Set up the signal source.
a Select Marker on the oscilloscope calibrator.
b Set the calibrator for 100 μs markers (period = 100 μs).
3 Set up the oscilloscope.
a Set channel 1 Coupling to DC.
b On 1 GHz and 500 MHz models only, set channel 1 Imped to 50 Ohm.
c Press the Display key, then set the Vectors softkey to off.
d Press the AutoScale key.
e Set the time base to 20 μs/div.
f Press the Main/Delayed (or Zoom) key, then set the Time Ref softkey to Left.
g Adjust the Trigger Level knob to obtain a stable display.
Testing Performance2
6000L Series Oscilloscopes Service Guide39
Page 40
2Testing Performance
4 Press the Quick Meas softkey, set the Source softkey to 1, then press Select and
choose Period. Press the Measure softkey and measure the following:
Period 100 μs — The test limits are 99.8 μs to 100.2 μs.
If the measurements are not within the test limits, go to the “Troubleshooting”
chapter. Then return here.
5 Change the calibrator to 100 ns markers. Change the time base to 20 ns/div.
Adjust the trigger level to obtain a stable display.
6 Measure the following. If the measurements are not within the test limits, go to
the “Troubleshooting” chapter. Then return here.
Period 100 ns— The test limits are 99.8 ns to 100.2 ns.
7 Change the time base and calibrator markers as follows:
a On 500 MHz and 1 GHz models, change time base to 2 ns/div and the
calibrator to 5 ns markers
b On 100 MHz models, change time base to 5 ns/div and the calibrator to
10 ns markers
8 Make the following measurements. If the measurements are not within the test
limits, go to the “Troubleshooting” chapter. Then return here.
a For 500 MHz, and 1 GHz models, period 5 ns — the test limits are 4.96 ns to
5.04 ns.
b For 100 MHz models, period 10 ns — the test limits are 9.93 ns to 10.07 ns.
406000L Series Oscilloscopes Service Guide
Page 41
To verify trigger sensitivity
This test verifies the trigger sensitivity. In this test, you will apply a sine wave to
the oscilloscope at the upper bandwidth limit. You will then decrease the
amplitude of the signal to the specified levels, and check to see if the oscilloscope
is still triggered.
The internal trigger sensitivity test is mandatory because it is a specification. The
external trigger test is optional because it is a characteristic, not a specification.
Test limits for:
• Internal trigger:
• < 10 mV/div: greater of 1 div or 5 mV
>=10 mV/div: 0.6 div
• External trigger:
1GHz and 500MHz models:
(DSO6104L/6054L)
DC to 500 MHz: < 500 mV
100 MHz models:
(DSO6014L)
DC to 100 MHz: < 500 mV
pp
pp
Testing Performance2
pp
Table 11Equipment Required to Verify Trigger Sensitivity
a Press the Save/Recall key, then press the Default Setup softkey.
b On 500 MHz and 1 GHz models, set channel 1 Imped to 50 Ohm.
c Set the output frequency of the signal generator to the maximum bandwidth
of the oscilloscope and set the amplitude to about 10 mV
d Press the AutoScale key.
e Set the time base to 10 ns/div.
f Set channel 1 to 5 mV/div.
g Decrease the amplitude from the signal generator until 1 vertical division of
the signal is displayed.
The trigger is stable when the displayed waveform is stable. If the trigger is
not stable, try adjusting the trigger level. If adjusting the trigger level makes
the trigger stable, the test still passes. If adjusting the trigger does not help,
see the “Troubleshooting” chapter. Then return here.
h Record the result as Pass or Fail in the Performance Test Record (see
page 46).
3 Repeat this procedure for the remaining oscilloscope channels.
pp
.
Test External Trigger Sensitivity
The following test is optional because External Trigger Sensitivity is a
characteristic, not a specification.
Verify the external trigger sensitivity at these settings:
• 1 GHz and 500 MHz bandwidth models:
500 MHz, 500 mV
• 100 MHz bandwidth models:
100 MHz, 500 mV
1 Connect the equipment (see Figure 7).
a Use the N cable to connect the signal generator to the power splitter input.
b Connect one output of the power splitter to the Aux Trig input through a 50Ω
feedthrough termination.
c Connect the power sensor to the other output of the power splitter.
6000L Series Oscilloscopes Service Guide43
pp
pp
Page 44
2Testing Performance
Oscilloscope
Signal
Generator
Power Meter
Power Splitter
Power Sensor
N to BNC adapter
N Cable
50Ω Feedthrough*
* Required for all models.
446000L Series Oscilloscopes Service Guide
Figure 7Connect equipment for external trigger sensitivity test
2 Set up the oscilloscope.
a Press the Save/Recall key, then press the Default Setup softkey.
3 Change the signal generator output frequency as follows:
• For 500 MHz and 1 GHz models, set the signal generator output frequency
to 500 MHz
• For 100 MHz models, set the signal generator output frequency to 100 MHz
Page 45
Testing Performance2
4 Set the power meter Cal Factor % to the appropriate value (500 MHz or
100 MHz) on the calibration chart on the power sensor. If necessary, do a linear
interpolation if a 500 MHz or 100 MHz factor is not included in the power
meter’s calibration chart.
5 Adjust the signal generator output for reading on the power meter of 625μW.
(500mV
Power = Vin
6 Press the Trigger Edge key, then press the Source softkey to set the trigger
source to External.
7 Check for stable triggering and adjust the trigger level if necessary. Triggering
is indicated by the Trig’d indicator at the top of the display. When it is flashing,
the oscilloscope is not triggered. When it is not flashing, the oscilloscope is
triggered.
8 Record the results as Pass or Fail in the Performance Test Record (see
page 46).
If the test fails, see the “Troubleshooting” chapter. Then return here.
= 176.78mV rms,
pp
2
/50Ω = 176.78 mV2/50Ω = 625μW.)
6000L Series Oscilloscopes Service Guide45
Page 46
2Testing Performance
Keysight 6000L Series Oscilloscopes Performance Test Record
Serial No. ______________________________________Test by _____________________________
Test Interval ____________________________________Work Order No. ______________________
Recommended Next Testing ________________________Temperature ____________
Threshold SpecificationLimitsCh D7-D0 Ch D15-D8
Accuracy Test5 V - 250 mV4.750 V________________
(100 mV + 3% of
threshold setting)
Voltage Measurement Accuracy
RangePower Supply Setting Test LimitsChannel 1Channel 2 Channel 3*Channel 4*
5 V/Div35 V34.04 V to 35.96 V________________________________
2 V/Div14 V13.616 V to 14.384 V________________________________
1 V/Div7 V6.808 V to 7.192 V________________________________
500 mV/Div3.5 V3.404 V to 3.596 V________________________________
200 mV/Div1.4 V1.3616 V to 1.4384 V________________________________
100 mV/Div700 mV680.8 mV to 719.2 mV________________________________
50 mV/Div350 mV340.4 mV to 359.6 mV________________________________
20 mV/Div140 mV136.16 mV to 143.84 mV ________________________________
10 mV/Div70 mV68.08 mV to 71.92 mV________________________________
5 mV/Div35 mV34.04 mV to 35.96 mV________________________________
2mV/Div
2mV/Div
1mV/Div
(500 MHz and 1 GHz models)
(100 MHz models)
(MSO/DSO6012A)
5 V + 250 mV5.250 V________________
-5 V - 250 mV-5.250 V________________
-5 V + 250 mV-4.750 V________________
0 V - 100 mV-100 mV________________
0 V + 100 mV100 mV________________
14 mV13.232 mV to 14.768 mV ________________________________
14 mV13.616 mV to 14.384 mV ________________________________
7 mV6.616 mV to 7.384 mV________________________________
610x -3 dB at 1 GHz________________________________
605x -3 dB at 500 MHz________________________________
601x -3 dB at 100 MHz________________________________
Limits
Time Scale Accuracy is a warranted specification
for 6000L Series instruments manufactured after
January 1, 2008.
1 GHz models0.6 division at 1 GHz_______________________________
500 MHz models0.6 division at 500 MHz _______________________________
100 MHz models0.6 division at 100 MHz _______________________________
External trigger (6014, 6054, 6104)
6104, 6054500 mV at 500 MHz________________________________
6014500 mV at 100 MHz________________________________
9.93 ns to 10.07 ns________
4.96 ns to 5.04 ns________
1 division at 25 MHz________________________________
6000L Series Oscilloscopes Service Guide47
Page 48
2Testing Performance
486000L Series Oscilloscopes Service Guide
Page 49
Keysight InfiniVision 6000L Series Oscilloscopes
WARNING
CAUTION
Service Guide
3Calibrating and Adjusting
This chapter explains how to adjust the oscilloscope for optimum operating
performance. You should perform self-calibration according to the following
recommendations:
• Every 12 months or after 2000 hours of operation
• If the ambient temperature is >10 °C from the calibration temperature
• If you want to maximize the measurement accuracy
The amount of use, environmental conditions, and experience with other
instruments help determine if you need shorter adjustment intervals.
Let the Equipment Warm Up Before Adjusting
Before you start the adjustments, let the oscilloscope and test equipment warm
up for at least 30 minutes.
Read All Cautions and Warnings
Read the following cautions and warning before making adjustments or
performing self-calibration.
HAZARDOUS VOLTAGES!
Read the safety notice at the front of this book before proceeding.
Maintenance is performed with power supplied to the oscilloscope and with the protective covers
removed. Only trained service personnel who are aware of the hazards involved should perform the
maintenance. Whenever possible, perform the procedures with the power cord removed from the
oscilloscope.
REMOVE POWER TO AVOID DAMAGE!
Do not disconnect any cables or remove any assemblies with power applied to the oscilloscope.
Otherwise, damage to the oscilloscope can occur.
49
Page 50
3Calibrating and Adjusting
CAUTION
CAUTION
USE EXTERNAL FAN TO REDUCE TEMPERATURE!
When you must operate the oscilloscope with its cover and main shield removed, use an external fan to
provide continuous air flow over the samplers (the ICs with heat sinks on them). Air flow over the
samplers is reduced when the cover and main shield is removed, which leads to higher than normal
operating temperatures. Have the fan blow air across the system board where the heat sinks are located.
If the cover is removed but the main shield remains installed and the bottom holes are not blocked, the
instrument will cool properly.
AVOID DAMAGE TO ELECTRONIC COMPONENTS!
Electrostatic discharge (ESD) can damage electronic components. When you use any of the procedures
in this chapter, use proper ESD precautions. As a minimum, place the oscilloscope on a properly
grounded ESD mat and wear a properly grounded ESD strap.
506000L Series Oscilloscopes Service Guide
Page 51
User Calibration
Perform user-calibration:
• Each year or after 2000 hours of operation.
• If the ambient temperature is >10° C from the calibration temperature.
• If you want to maximize the measurement accuracy.
The amount of use, environmental conditions, and experience with other
instruments help determine if you need shorter User Cal intervals.
User Cal performs an internal self-alignment routine to optimize the signal path in
the oscilloscope. The routine uses internally generated signals to optimize circuits
that affect channel sensitivity, offset, and trigger parameters. Disconnect all
inputs and allow the oscilloscope to warm up before performing this procedure.
Performing User Cal will invalidate your Certificate of Calibration. If NIST (National
Institute of Standards and Technology) traceability is required perform the
procedures in Chapter 2 in this book using traceable sources.
Calibrating and Adjusting3
To perform User C al
1 Set the rear-panel CALIBRATION switch to UNPROTECTED.
2 Connect short (12 inch maximum) equal length cables to each analog channel’s
BNC connector on the front of the oscilloscope. You will need four equal-length
cables for a 4-channel oscilloscope.
6000L Series Oscilloscopes Service Guide51
Page 52
3Calibrating and Adjusting
Longer cable
to TRIG OUT
To Channel 3
To Channel 2
To Channel 1
To Channel 4
Use 50Ω RG58AU or equivalent BNC cables when performing User Cal.
a Connect BNC tees to the equal-length cables as shown below. Then connect
a BNC(f)-to-BNC(f) (barrel connector) to the tee as shown below.
Figure 8User Calibration cable for 4-channel oscilloscope
3 Connect a BNC cable (40 inches maximum) from the TRIG OUT connector on
the rear panel to the BNC barrel connector.
4 Press the Utility key, then press the Service softkey.
526000L Series Oscilloscopes Service Guide
5 Begin the Self Cal by pressing the Start User Cal softkey.
6 When the User Cal is completed, set the rear-panel CALIBRATION switch to
PROTECTED.
Page 53
User Cal Status
Pressing the User Cal Status softkey displays the following summary results of the
previous User Cal, and the status of probe calibrations for probes that can be
calibrated. Note that AutoProbes do not need to be calibrated, but InfiniiMax can
be calibrated.
Results:
User Cal date:
Change in temperature since last User Cal:
Failure:
Comments:
Probe Cal Status:
Calibrating and Adjusting3
6000L Series Oscilloscopes Service Guide53
Page 54
3Calibrating and Adjusting
546000L Series Oscilloscopes Service Guide
Page 55
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide
4Troubleshooting
Solving General Problems with the Oscilloscope / 56
Troubleshooting the Oscilloscope / 59
This chapter begins with suggestions for solving general problems that you may
encounter with the oscilloscope. It tells you what to do in these cases:
• If there is no trace display
• If the trace display is unusual or unexpected
• If you cannot see a channel
• If you cannot get any response from the oscilloscope
Procedures for troubleshooting the oscilloscope follow the problem solving
suggestions. The troubleshooting section shows you how to:
• Check out the oscilloscope
• Check power supply
• Check the system board
• Check the display
• Check the fan
• Run internal self-tests
• Verify default setup
Read All Cautions and Warnings
Before you begin any troubleshooting, read all Warning and Cautions in the “Troubleshooting” section.
55
Page 56
4Troubleshooting
Solving General Problems with the Oscilloscope
This section describes how to solve general problems that you may encounter
while using the Keysight 6000L Series Oscilloscopes to make measurements.
After troubleshooting the oscilloscope, if you need to replace parts, refer to the
“Replaceable Parts” chapter.
On 6000L Series oscilloscopes, connect an XGA display to the display port on the
rear panel before you begin troubleshooting. Then connect the oscilloscope to the
network, establish communication, and use the Remote Front Panel feature
described in the 6000L Series Oscilloscopes User’s Guide to control the
oscilloscope. If you do not have the user’s guide, you can download it from
www.keysight.com/find/6000.
If there is no display
✔ Check that the power cord is firmly seated in the oscilloscope power
receptacle.
✔ Check that the power source is live.
✔ Check that the front-panel power switch is on.
✔ If there is still no display, go to the troubleshooting procedures in this chapter.
If there is no trace display
✔ Check that the INTENSITY knob on the front panel is adjusted correctly.
✔ Recall the default setup by pressing Save/Recall then Default Setup. This will
ensure that the trigger mode is Auto.
✔ Check that the probe clips are securely connected to points in the circuit under
test, and that the ground is connected.
✔ Check that the circuit under test is powered on.
✔ Press the AutoScale key.
✔ Obtain service from Keysight Technologies, if necessary.
566000L Series Oscilloscopes Service Guide
Page 57
Troubleshooting4
If the trace display is unusual or unexpected
✔ Check that the horizontal sweep speed control setting is correct for the
expected frequency range of the input signals.
✔ The sampling speed of the oscilloscope depends on the time/division setting. It
may be that when time/division is set to slower speeds, the oscilloscope is
sampling too slowly to capture all of the transitions on the waveform. Use peak
detect mode.
✔ Check that all oscilloscope probes are connected to the correct signals in the
circuit under test.
✔ Ensure that the probe’s ground lead is securely connected to a ground point in
the circuit under test. For high-speed measurements, each probe’s individual
ground lead should also be connected to a ground point closest to the signal
point in the circuit under test.
✔ Check that the trigger setup is correct.
✔ A correct trigger setup is the most important factor in helping you capture the
data you desire. See the User’s Guide for information about triggering.
✔ Check that infinite persistence in the Display menu is turned off, then press the
Clear Display softkey.
✔ Press the AutoScale key.
If you cannot see a channel
✔ Recall the default setup by pressing Save/Recall then Default Setup. This will
ensure that the trigger mode is Auto.
✔ Check that the oscilloscope probe’s BNC connector is securely attached to the
oscilloscope’s input connector.
✔ Check that the probe clips are securely connected to points in the circuit under
test.
✔ Check that the circuit under test is powered on.
You may have pressed the AutoScale key before an input signal was available.
Performing the checks listed here ensures that the signals from the circuit
under test will be seen by the oscilloscope. Perform the remaining checks in
this topic to make sure the oscilloscope channels are on, and to obtain an
automatic setup.
6000L Series Oscilloscopes Service Guide57
Page 58
4Troubleshooting
✔ Check that the desired oscilloscope channels are turned on.
a Switch on the analog channel by selecting Analog from the menu bar of the
Remote Front Panel (6000L Series models).
b Switch on the digital channels by selecting Digital from the menu bar (6000L
Series models with the MSO option). Ensure that the desired channels are
turned on.
✔ Press the AutoScale key to automatically set up all channels.
586000L Series Oscilloscopes Service Guide
Page 59
Troubleshooting the Oscilloscope
WARNING
WARNING
CAUTION
CAUTION
The service policy for the Keysight 6000L Series Oscilloscopes is assembly level
replacement. If you need parts or assistance from Keysight Technologies to repair
your instrument, go to www.keysight.com and locate the service facility for your
area.
HAZARDOUS VOLTAGES EXIST — REMOVE POWER FIRST !
The procedures described in this section are performed with power supplied to the oscilloscope and
with the protective covers removed. Only trained service personnel who are aware of the hazards
involved should perform the procedures. Whenever possible, perform the procedures with the power
cord removed from the oscilloscope. Read the safety notice at the back of this book before proceeding.
HAZARDOUS VOLTAGES EXIST — HIGH VOLTAGE IS PRESENT ON DSO6000L SERIES POWER SUPPLY
HEAT SINKS !
The power supply heat sinks of the 6000L Series oscilloscopes are at a high potential. This presents an
electric shock hazard. Protect yourself from electric shock by keeping this area covered or by not coming
in contact with the heat sinks when the power cord is attached to the oscilloscope!
Troubleshooting4
REMOVE POWER TO AVOID DAMAGE !
Do not disconnect any cables or remove any assemblies while power is applied to the oscilloscope, or
damage to the oscilloscope can occur.
AVOID ESD DAMAGE TO COMPONENTS !
ELECTROSTATIC DISCHARGE (ESD) can damage electronic components. Use proper ESD precautions
when doing any of the procedures in this chapter. As a minimum, place the oscilloscope on a properly
grounded ESD mat and wear a properly grounded ESD strap.
Equipment required for troubleshooting
The equipment listed in this table is required to troubleshoot the oscilloscope.
6000L Series Oscilloscopes Service Guide59
Page 60
4Troubleshooting
CAUTION
WARNING
CAUTION
Table 12Equipment Required to Troubleshoot the Oscilloscope
Digital multimeterAccuracy ±0.05,% 1 mV resolutionKeysight 34401A
OscilloscopeCapable of measuring ≥ 500 MHz signal. 1
MΩ input impedance.
Keysight
DSO6102A, MSO6102A,
DSO6102A, or MSO6104A
To check out the oscilloscope
1 Disconnect any external cables from the front panel.
2 Disconnect the power cord, then remove the cabinet following the instructions
on page 73 for Keysight 6000L Series models.
USE AN EXTERNAL FAN TO AVOID OVERHEATING COMPONENTS !
When you remove the oscilloscope cover and main shield, use an external fan to provide continuous air
flow over the heat sinks. Air flow over the heat sinks is reduced when the cover and main shield are
removed, which leads to higher than normal operating temperatures. Have the fan blow air across the
system board where the heat sinks are located. Otherwise, damage to the components can occur.
HAZARDOUS VOLTAGES EXIST — HIGH VOLTAGE IS PRESENT ON DSO6000L SERIES POWER SUPPLY
HEAT SINKS !
The power supply heat sinks of the 6000L Series oscilloscopes are at a high potential. This presents an
electric shock hazard. Protect yourself from electric shock by keeping this area covered or by not coming
in contact with the heat sinks when the power cord is attached to the oscilloscope!
606000L Series Oscilloscopes Service Guide
3 Connect the power cord to the rear of the oscilloscope, then to a suitable ac
voltage source.
The oscilloscope power supply automatically adjusts for input line voltages in
the range of 100 to 240 VAC. Ensure that you have the correct line cord. The
power cord provided is matched to the country of origin.
T
Use only the power cord that came with the oscilloscope
The power cord provided is matched to the country of origin of the order.
Page 61
Troubleshooting4
WARNING
LAN Status
Indicator
Power
Switch
Power
Indicator
Reset
Switch
Always use a grounded power cord. Do not defeat the power cord ground.
• On 6000L Series models, press the power switch.
The Power indicator LED is green. The LAN indicator LED is bi-color (green
and/or red).
• Shortly after pressing the power switch, the power indicator and LAN
indicator will illuminate.
• Both will extinguish.
• Both will illuminate again. If a LAN is connected, the LAN indicator will
illuminate red. It will turn green when the LAN is connected as it is
configured. The amount of time it takes the indicator to turn green depends
on the network and the oscilloscope’s LAN configuration setup.
6000L Series Oscilloscopes Service Guide61
Figure 96000L Series LED indicators
4 If the lights illuminate, then the power supply is probably working correctly.
5 Proceed to “To verify basic oscilloscope operation" on page 62.
Page 62
4Troubleshooting
To verify basic oscilloscope operation
For 6000L Series oscilloscopes, connect the oscilloscope to your network or
establish a point-to-point connection, and use the Remote Front Panel feature
described in the 6000L Series Oscilloscopes User’s Guide to control the
oscilloscope. If you do not have the user’s guide, you can download it from
www.keysight.com/find/6000.
1 Press the Save/Recall key on the front panel, then press the Default Setup
softkey under the display. The oscilloscope is now configured to its default
settings.
2 Connect an oscilloscope probe from channel 1 to the Probe Comp signal
terminal on the front panel.
3 Connect the probe’s ground lead to the ground terminal that is next to the
Probe Comp terminal.
4 Press AutoScale.
5 You should see a waveform on the oscilloscope’s display similar to this:
If you see the waveform, but the square wave is not shaped correctly as shown
above, perform the procedure “To compensate the analog probes" on page 63.
If you do not see the waveform, ensure your power source is adequate, the
oscilloscope is properly powered-on, and the probe is connected securely to
the front-panel analog channel input BNC and to the Probe Comp terminal.
626000L Series Oscilloscopes Service Guide
Page 63
6 If you still do not see the waveform, use the troubleshooting flowchart in this
comp.cdr
Perfectly compensated
Over compensated
Under compensated
chapter to isolate the problem.
To compensate the analog probes
You should compensate your analog probes to match their characteristics to the
oscilloscope’s channels. A poorly compensated probe can introduce measurement
errors.
1 Perform the procedure “To verify basic oscilloscope operation" on page 62
2 Use a nonmetallic tool to adjust the trimmer capacitor on the probe for the
flattest pulse possible. The trimmer capacitor is located on the probe BNC
connector.
Troubleshooting4
Figure 10Example pulses
3 Connect probes to all other analog channels (channels 2, 3, and 4 of a
4-channel oscilloscope). Repeat the procedure for each channel. This matches
each probe to each channel.
The process of compensating the probes serves as a basic test to verify that the
oscilloscope is functional.
6000L Series Oscilloscopes Service Guide63
Page 64
4Troubleshooting
Start
Test Power Supply
Okay?
Test System
Board
Okay?
Test Display
Okay?
End
Replace Power
Supply
Replace System
Board
Replace Display
Yes
Yes
Yes
No
No
No
Troubleshooting Flowchart
The following flowchart describes how to troubleshoot 6000L Series models.
646000L Series Oscilloscopes Service Guide
Page 65
To check the 6000L Series oscilloscope power supply
WARNING
CAUTION
NOTE
HAZARDOUS VOLTAGES EXIST — HIGH VOLTAGE IS PRESENT ON DSO6000L SERIES POWER SUPPLY
HEAT SINKS !
The power supply heat sinks of the 6000L Series oscilloscopes are at a high potential. This presents an
electric shock hazard. Protect yourself from electric shock by keeping this area covered or by not coming
in contact with the heat sinks when the power cord is attached to the oscilloscope!
USE AN EXTERNAL FAN TO AVOID OVERHEATING COMPONENTS !
When you remove the oscilloscope cover and main shield, use an external fan to provide continuous air
flow over the heat sinks. Air flow over the heat sinks is reduced when the cover and main shield are
removed, which leads to higher than normal operating temperatures. Have the fan blow air across the
system board where the heat sinks are located. Otherwise, damage to the components can occur.
If the power supply fuse is blown, the power supply is defective, and you must replace it. See
the “Replaceable Parts” chapter for information about removing the power supply.
Troubleshooting4
1 Disconnect the power cord from the oscilloscope. Then remove the oscilloscope
cover.
2 Connect the negative lead of the multimeter to a ground point on the
oscilloscope.
3 Connect the power cord and turn on the oscilloscope.
4 Measure the power supply voltage at J3200, pin 9 on the system board. See
Figure 11 on page 66. The voltage should be 15 V ±10%. If the voltage is
correct, the power supply is working correctly.
6000L Series Oscilloscopes Service Guide65
Page 66
4Troubleshooting
Pin 1
Pin 9
J3200
Figure 11System Board Test Points
5 If the voltage is not correct:
a Disconnect the power cord.
b Remove the fan. See “To remove the fan" on page 81.
c Disconnect plug J3 from the power supply and from the system board.
d Perform a continuity check on the “DC cable” (item W1 on page 93). If the
cable is shorted or open, replace it. If not, continue to the next step.
e Reconnect the “DC cable” to the power supply, but not to the system board.
666000L Series Oscilloscopes Service Guide
Page 67
WARNING
HAZARDOUS VOLTAGES EXIST — HIGH VOLTAGE IS PRESENT ON DSO6000L SERIES POWER SUPPLY
HEAT SINKS !
The power supply heat sinks of the 6000L Series oscilloscopes are at a high potential. This presents an
electric shock hazard. Protect yourself from electric shock by keeping this area covered or by not coming
in contact with the heat sinks when the power cord is attached to the oscilloscope!
f Connect the power cord and check the power supply voltage between pins 1
and 6 of the system board side of the “DC cable”. The voltage should be 15 V
±10%.
• If the voltage is correct, then test the system board (see page 67).
• If the voltage is not correct, then replace the power supply (see page 79).
To check the 6000L Series system board
1 Remove the cabinet.
2 Check that all cable connections are securely connected from the system board
to:
• Power supply
• Fan
3 Verify the voltages at the system board test points listed in the table below.
Refer to Figure 11 on page 66 to locate the test points.
4 If the voltage at test point L3301 and/or L3302 is not within the specified
range, replace the system board.
5 If all cables are properly connected and none of the previous tests confirm a
failure on another assembly, replace the system board.
Page 68
4Troubleshooting
To check the 6000L Series display output
To check the fan
1 Connect a known good external XGA display and power-up the display and the
oscilloscope.
2 If the known good display does show an output from the XGA port of the
oscilloscope, replace the system board.
The fan speed is controlled by a circuit on the system board.
1 If the fan is running, perform the internal self-tests. Go to “To run the internal
self-tests" on page 68.
2 If the fan is not running, it may be defective. Follow these steps:
a Disconnect the fan cable from the system board.
b Measure the fan voltage at the connector on the system board.
See the figure below for the location of the fan connector.
c If the fan voltage is approximately +8.5 Vdc at room temperature, replace the
fan. If the fan voltage is not approximately +8.5 Vdc, replace the system
board.
The proper voltage range depending on temperature is between +6.0 Vdc to
+11.5 Vdc.
To run the internal self-tests
Self Test performs a series of internal procedures to verify that the oscilloscope is
operating properly.
It is recommended that you run the Self Test:
• after experiencing abnormal operation
• for additional information to better describe an oscilloscope failure
• to verify proper operation after the oscilloscope has been repaired
Successfully passing Self Test does not guarantee 100% of the oscilloscope’s
functionality. Self Test is designed to provide an 80% confidence level that the
oscilloscope is operating properly.
1 Press the Utility key, then press the Service softkey. Use the Remote Front Panel
feature if you are testing a 6000L Series model.
686000L Series Oscilloscopes Service Guide
Page 69
2 Begin the internal self tests by pressing the Start Self Test softkey.
To verify default setup
The oscilloscope is designed to turn on with the setup from the last turn on or
previous setup. However, if the Secure Environment option is installed, the
oscilloscope will always execute a Default Setup upon power-up.
To recall the factory default setup:
1 Press the Save/Recall key.
2 Press the Default Setup softkey.
This returns the oscilloscope to its factory default settings and places the
oscilloscope in a known operating condition. The major default settings are:
• Horizontal - main mode, 100 us/div scale, 0 s delay, center time reference
• Vertical - Channel 1 on, 5 V/div scale, dc coupling, 0 V position, probe factor
to 1.0 if an AutoProbe probe is not connected to the channel
• Trigger - Edge trigger, Auto sweep mode, 0 V level, channel 1 source, dc
coupling, rising edge slope, 60 ns holdoff time
• Display - Vectors on, 20% grid intensity, infinite persistence off
• Other - Acquire mode normal, Run/Stop to Run, cursor measurements off
Troubleshooting4
6000L Series Oscilloscopes Service Guide69
Page 70
4Troubleshooting
Figure 12Default setup screen
3 If your screen looks substantially different, replace the system board.
706000L Series Oscilloscopes Service Guide
Page 71
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide
5Replacing 6000L
Assemblies
To remove the bottom cover / 73
To remove the front panel assembly / 74
To remove the system board / 76
To remove the power supply / 79
To remove the fan / 81
To remove the AC power input assembly / 83
To remove the powe r shaft / 85
This chapter describes how to remove assemblies from the Keysight 6000L Series
Oscilloscopes. After you have removed an assembly, to install the replacement
assembly, follow the instructions in reverse order.
The parts shown in the following figures are representative and may look different
than what you have in your oscilloscope.
Tools Used for Disassembly
Use these tools to remove and replace the oscilloscope assemblies:
• T10 TORX drivers
• 5/8-inch and 9/32-inch socket drivers
See how the Oscilloscope Parts Fit Together
An exploded view of the oscilloscope is included in the Chapter 6, “Replaceable
Parts,” starting on page 87. It shows the individual part numbers used in the
assemblies, and shows you how all of the assemblies fit together.
71
Page 72
5Replacing 6000L Assemblies
WARNING
WARNING
CAUTION
CAUTION
Read All Warnings and Cautions
Read the following warnings and cautions before removing and replacing any
assemblies in the oscilloscope.
HAZARDOUS VOLTAGES !
Read the safety summary at the back of this book before proceeding.
Maintenance is performed with power supplied to the oscilloscope and with the protective covers
removed. Only trained service personnel who are aware of the hazards involved should perform the
maintenance. Whenever possible, perform the procedures with the power cord removed from the
oscilloscope.
AVOID ELECTRICAL SHOCK !
Hazardous voltages exist on the LCD assembly and power supply. To avoid electrical shock:
1 Disconnect the power cord from the oscilloscope.
2 Wait at least three minutes for the capacitors in the oscilloscope to discharge before you begin
disassembly.
Read the Safety Summary at the back of this manual before you begin.
726000L Series Oscilloscopes Service Guide
REMOVE POWER TO AVOID DAMAGE !
Remove power before you begin to remove and replace assemblies. Do not remove or replace
assemblies while the oscilloscope is turned on, or damage to the components can occur.
AVOID DAMAGE TO ELECTRONIC COMPONENTS !
ELECTROSTATIC DISCHARGE (ESD) can damage electronic components. When doing any of the
procedures in this chapter, use proper ESD precautions. As a minimum, you should place the instrument
on a properly grounded ESD mat and wear a properly grounded ESD strap.
Page 73
To remove the bottom cover
T-10 Screws
T-10 Screw
T-10 Screw
1 Turn off the oscilloscope and disconnect the power cable.
2 Turn the oscilloscope so the bottom is facing up.
3 Using a T10 screwdriver, remove 5 screws from the rear and sides of the
cabinet.
4 Slide the bottom cover back and up to remove.
Replacing 6000L Assemblies5
Figure 13Removing the cover
6000L Series Oscilloscopes Service Guide73
Page 74
5Replacing 6000L Assemblies
Intensity
Knob
To remove the front panel assembly
1 Perform the following procedures:
• “To remove the bottom cover" on page 73
2 Insert a flat-blade screwdriver under the center of the intensity knob and gently
twist the screwdriver as you pull the knob off.
Using a twisting motion on the screwdriver rather than prying prevents marking
or damaging the front panel.
Figure 14Removing the intensity knob
3 Disconnect the front panel ribbon cable from the system board by pressing the
release tabs and then disconnect it from the front panel PC board.
4 Use a flat-blade screwdriver to release the center retainer tab and push it
slightly out of the slot.
5 Use a flat-blade screwdriver to release the 2 side retainer tabs and remove
them from their slots.
746000L Series Oscilloscopes Service Guide
Page 75
Replacing 6000L Assemblies5
Front Panel
Ribbon Cable
Retainer Tab
System Board
Front
Panel
Retainer Tab Slot
Retainer Tab
Center Retainer Tab
100 MHz Auto
Probe Label
500 MHz Auto
Probe Label
Front
Panel PC
Board
Figure 15Disconnecting ribbon cable and releasing tab retainers
6 Pull the front panel off of the oscilloscope.
7 Lift the front panel PC board to remove it from the front panel.
8 Reverse this procedure to replace the front panel.
6000L Series Oscilloscopes Service Guide75
Page 76
5Replacing 6000L Assemblies
T10 Screws
Fan Cable
System Board
Backlight Inverter Cable
Power Supply
Cable
Front Panel
Cable
Power
Supply
Cable
Hex
Standoffs
To remove the system board
1 Perform the previous procedures:
• “To remove the bottom cover" on page 73
2 Remove the intensity knob as described in step 2 on page 74.
3 Disconnect the front panel ribbon cable from the system board by pressing the
release tabs.
4 Disconnect the fan cable.
5 Remove the 3 hex nuts and washers from the rear BNC connectors using a
5/8-inch socket driver.
Figure 16Preparing to remove the system board
766000L Series Oscilloscopes Service Guide
Page 77
Replacing 6000L Assemblies5
CAUTION
ICs
Thermal
Pads
NOTE
6 Using a 9/23-inch hex driver, remove the two hex standoffs and washers from
the GPIB connector
7 Using a T10 Torx driver, remove the 5 screws that hold the system board to the
deck.
8 Remove the system board
a Lift the back of the board to clear the chassis.
b Gently pull the board out until you can disconnect the power supply cable.
c Continue to remove the board from the chassis.
To avoid tearing the thermal pads on a 500 MHz oscilloscope when removing or installing
the system board, hold the board up, away from the thermal pads until all components are
clear.
Figure 17Avoid damage to thermal pads (500 MHz models only)
If the battery on the system board needs to be replaced, use a CR2477/1HF or CR2477/HFN
3V manganese dioxide lithium battery.
6000L Series Oscilloscopes Service Guide77
Page 78
5Replacing 6000L Assemblies
9 Reverse this procedure to reinstall the system board:
786000L Series Oscilloscopes Service Guide
Page 79
To remove the power supply
Power Input Board
Cable
T10 Screws
Power Input Board
Ground
Connector
Hole in
Divider
AC
Connector
Snap
Fasteners
Power to System
Board
1 Perform the previous procedures:
• “To remove the bottom cover" on page 73
• “To remove the system board" on page 76
2 Disconnect the ground cable and AC cables from the power supply using a
needle nose pliers if necessary.
3 Press the release latch to disconnect the power input board cable from the
power input board.
Replacing 6000L Assemblies5
Figure 18Removing the power supply.
4 Using a T10 Torx driver, remove the 2 screws securing the power supply to the
chassis.
6000L Series Oscilloscopes Service Guide79
Page 80
5Replacing 6000L Assemblies
5 Using a flat-blade screwdriver through the hole in the divider, gently pry the
power supply board up enough to snap the power supply board off the rear
snap fastener.
6 Using a twisting motion, with a flat-blade screwdriver gently snap the power
supply board off the front snap fastener.
7 Reverse this procedure to install the power supply.
806000L Series Oscilloscopes Service Guide
Page 81
To remove the fan
Fan
Flexible Fan
Mount
Chassis Divider
1 Perform the previous procedures:
• “To remove the bottom cover" on page 73
• “To remove the system board" on page 76
• “To remove the power supply" on page 79
2 Slide the fan and fan mount slightly up and out of the chassis.
3 Remove the fan from the flexible fan mount.
Replacing 6000L Assemblies5
Though the fan can be removed without performing all of the above, it is
difficult to reinstall it.
Figure 19Removing the fan.
6000L Series Oscilloscopes Service Guide81
Page 82
5Replacing 6000L Assemblies
Pegs Fit in
Fan Holes
Holes in
Bottom Fit
Over Studs
Align
Notch
and
Chassis
Chassis
Divider
4 Replace the fan.
a Place the rectangular part of the flexible fan mount over the fan.
b Place the pegs in the flexible fan mount into the holes in the top and bottom
of the fan.
c Align the holes in the bottom of the flexible fan mount over the studs on the
chassis as you slide the fan over the divider in the chassis.
d Ensure that the notch in the rectangular portion of the fan mount aligns with
the front of the chassis.
826000L Series Oscilloscopes Service Guide
Figure 20Installing the fan.
Page 83
To remove the AC power input assembly
CAUTION
1 Perform the previous procedures:
• “To remove the bottom cover" on page 73
• “To remove the power supply" on page 79
• “To remove the fan" on page 81
2 Using a TORX T10 screwdriver, remove the screw securing the power
receptacle ground wire to the chassis.
3 Using a TORX T10 screwdriver, remove the screw securing the input board to
the chassis.
4 Lift and rotate the grooved input receptacle just enough to free it from the
opening in the chassis.
5 Slide the assembly toward the front of the chassis to remove the posts from the
keyholes.
6 Lift the whole assembly up and back to free the power shaft from the chassis.
Avoid breaking the power shaft. Use one hand on each of the 2 pieces of the assembly to
keep it in alignment as you remove it.
Replacing 6000L Assemblies5
6000L Series Oscilloscopes Service Guide83
Page 84
5Replacing 6000L Assemblies
Grooved
Receptacle
Power
Shaft
Posts (2) &
Keyholes (2)
T10 Screws
AC Input
Board
Ground Wire
Figure 21Removing the AC power input board.
7 Reverse this procedure to install the AC power input board.
846000L Series Oscilloscopes Service Guide
Page 85
To remove the power shaft
CAUTION
1 Perform the previous procedures:
• “To remove the bottom cover" on page 73
• “To remove the power supply" on page 79
• “To remove the fan" on page 81
2 Use a flat-blade screwdriver to gently spread the latch while pushing the power
shaft forward.
Twisting the latch too much could cause it to break!
Replacing 6000L Assemblies5
Figure 22Removing the power shaft.
3 Lift and remove it from the deck.
4 When reinserting the power shaft, push the shaft into the power switch until the
shaft snaps onto the switch.
6000L Series Oscilloscopes Service Guide85
Page 86
5Replacing 6000L Assemblies
866000L Series Oscilloscopes Service Guide
Page 87
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide
6Replaceable Parts
Ordering Replaceable Parts / 88
Replaceable Parts for 6000L Series Oscilloscopes / 90
This chapter describes how to order replaceable assemblies and parts for the
Keysight 6000L Series Oscilloscopes.
Diagrams and parts lists are included for assemblies and hardware that you can
order.
Before working on the oscilloscope, read the safety summary at the back of this
book.
87
Page 88
6Replaceable Parts
Ordering Replaceable Parts
Listed Parts
To order a part in the parts list, quote the Keysight Technologies part number,
indicate the quantity desired, and address the order to the nearest Keysight
Technologies Sales Office. To find your nearest sales office go to
www.keysight.com.
Unlisted Parts
To order a part not listed in the parts list, include the instrument part number,
instrument serial number, a description of the part (including its function), and the
number of parts required. Address the order to the nearest Keysight Technologies
Sales Office.
Direct Mail Order System
Within the USA, Keysight Technologies can supply parts through a direct mail
order system. There are several advantages to this system:
• Direct ordering and shipping from the Keysight Technologies parts center in
California, USA.
• No maximum or minimum on any mail order. (There is a minimum amount for
parts ordered through a local Keysight Technologies Sales Office when the
orders require billing and invoicing.)
• Prepaid transportation. (There is a small handling charge for each order.)
• No invoices.
In order for Keysight Technologies to provide these advantages, please send a
check or money order with each order.
Mail order forms and specific ordering information are available through your local
Keysight Technologies Sales Office. Addresses and telephone numbers are located
in a separate document shipped with the manuals.
886000L Series Oscilloscopes Service Guide
Page 89
Exchange Assemblies
Some parts used in this instrument have been set up for an exchange program.
This program allows the customer to exchange a faulty assembly with one that
has been repaired, calibrated, and performance-verified by the factory. The cost is
significantly less than that of a new part. The exchange parts have a part number
in the form XXXXX-695XX.
After receiving the repaired exchange part from Keysight Technologies, a United
States customer has 30 days to return the faulty assembly. For orders not
originating in the United States, contact the local Keysight Technologies service
organization. If the faulty assembly is not returned within 30 days, the customer
will be charged an additional amount. The additional amount will be the difference
in price between a new assembly and that of an exchange assembly.
Replaceable Parts6
6000L Series Oscilloscopes Service Guide89
Page 90
6Replaceable Parts
Replaceable Parts for 6000L Series Oscilloscopes
6000L Series Oscilloscope Exploded Views
The following exploded views provide a graphical representation of the
oscilloscope at the time this manual was released. Not all parts are shown. Your
parts may be slightly different thank those shown. These views provide reference
designator numbers that map to those used in the replaceable parts list table that
follows.
906000L Series Oscilloscopes Service Guide
Page 91
MP2
W1
MP5
A3
MP4
A1
A4
MP11
W2
MP12
A2
MP6
MP7
M82
MP10
MP9
MP2
MP3
Replaceable Parts6
Figure 236000L Series Oscilloscope Exploded View
6000L Series Oscilloscopes Service Guide91
Page 92
6Replaceable Parts
6000L Series Oscilloscope Replaceable Parts List
The information given for each part consists of the following:
• Reference designation from the exploded view.
• Keysight Technologies part number.
• Total quantity (QTY) in the instrument or on assembly.
• Description of the part.
Table 146000L Series Oscilloscope Replaceable Parts
Ref Des Keysight Part
Number
A1D6104-664221AC line filter/power switch/PC board assembly
A2D6104-001011Deck assembly
A3D6104-685011Fan assembly
A4D6104-600011Rackmount kit
MP1D6104-664211Autoprobe board
MP2D6104-001041Cover assembly
MP3D6104-854011Dielectric insulator panel
MP4D6104-439011Extender - power switch
MP5D6104-447011Fan mount
MP6D6104-422011Front panel bezel
MP7D6104-943011ID Label - 4 Ch, DSO 1 GHz
MP7D6104-94302 1ID Label - 4 Ch, DSO 500 MHz
MP7D6104-943031ID Label - 4 Ch, DSO 100 MHz
Qty.Description
MP854801-474011Knob - 12 mm flint gray
MP954684-943141Label - auto probe
926000L Series Oscilloscopes Service Guide
Page 93
Table 146000L Series Oscilloscope Replaceable Parts (continued)
Replaceable Parts6
Ref Des Keysight Part
Number
MP10D6104-943071Label - main 4 Ch, 100 MHz
MP10D6104-943081Label - main 4 Ch, 500 MHz/1 GHz
MP1154684-665061PC board assembly - 4 Ch, 1 GHz
MP1154634-665011PC board assembly - 4 Ch, 100 MHz
MP1154674-665061PC board assembly - 4 Ch, 500 MHz
MP120950-48531Power supply
W1D6104-616071DC cable
W2D6104-616011Keyboard cable
Qty.Description
6000L Series Oscilloscopes Service Guide93
Page 94
6Replaceable Parts
946000L Series Oscilloscopes Service Guide
Page 95
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide
7Safety Notices
This apparatus has been designed and tested in accordance with IEC Publication
61010-1, Safety Requirements for Measuring Apparatus, and has been supplied in
a safe condition. This is a Safety Class I instrument (provided with terminal for
protective earthing). Before applying power, verify that the correct safety
precautions are taken (see the following warnings). In addition, note the external
markings on the instrument that are described under "Safety Symbols."
95
Page 96
7Safety Notices
Warnings
Before turning on the instrument, you must connect the protective earth terminal
of the instrument to the protective conductor of the (mains) power cord. The
mains plug shall only be inserted in a socket outlet provided with a protective
earth contact. You must not negate the protective action by using an extension
cord (power cable) without a protective conductor (grounding). Grounding one
conductor of a two-conductor outlet is not sufficient protection.
Only fuses with the required rated current, voltage, and specified type (normal
blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited
fuseholders. To do so could cause a shock or fire hazard.
If you energize this instrument by an auto transformer (for voltage reduction or
mains isolation), the common terminal must be connected to the earth terminal of
the power source.
Whenever it is likely that the ground protection is impaired, you must make the
instrument inoperative and secure it against any unintended operation.
Service instructions are for trained service personnel. To avoid dangerous electric
shock, do not perform any service unless qualified to do so. Do not attempt
internal service or adjustment unless another person, capable of rendering first aid
and resuscitation, is present.
Do not install substitute parts or perform any unauthorized modification to the
instrument.
Capacitors inside the instrument may retain a charge even if the instrument is
disconnected from its source of supply.
Do not operate the instrument in the presence of flammable gasses or fumes.
Operation of any electrical instrument in such an environment constitutes a
definite safety hazard.
Do not use the instrument in a manner not specified by the manufacturer.
966000L Series Oscilloscopes Service Guide
Page 97
To clean the instrument
If the instrument requires cleaning: (1) Remove power from the instrument. (2)
Clean the external surfaces of the instrument with a soft cloth dampened with a
mixture of mild detergent and water. (3) Make sure that the instrument is
completely dry before reconnecting it to a power source.
Safety Notices7
6000L Series Oscilloscopes Service Guide97
Page 98
7Safety Notices
!
Safety Symbols
Instruction manual symbol: the product is marked with this symbol when it is
necessary for you to refer to the instruction manual in order to protect against
damage to the product and preserve safe conditions.
Hazardous voltage symbol.
Earth terminal symbol: Used to indicate a circuit common connected to grounded
chassis.