Keysight Technologies InfiniVision 6000L Series Service Manual

Page 1
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide
Page 2
Notices
CAUTION
WARNING
© Keysight Technologies, Inc. 2005-2019
Manual Part Number
54684-97035, October 2019
54684-97001, February 2005 54684-97003, April 2005 54684-97008, January 2006 54684-97010, June 2006 54684-97014, November 2006 54684-97018, March 2007 54684-97021, April 2007 54684-97026, March 2008 54684-97029, August 2008 54684-97033, January 2013 54684-97034, January 2017 54684-97035, October 2019
Published by: Keysight Technologies, Inc. 1900 Garden of the Gods Road Colorado Springs, CO 80907 USA
A newer version of this manual may be available at
www.keysight.com/find/ 6000manual
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Key­sight shall not be liable for errors or for inci­dental or consequential damages in
connection with the furnishing, use, or per­formance of this document or of any infor­mation contained herein. Should Keysight and the user have a separate written agree­ment with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sepa­rate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
U.S. Government Rights
The Software is "commercial computer software," as defined by Federal Acquisition Regulation ("FAR") 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement ("DFARS")
227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
www.keysight.com/find/sweula. The
license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software
documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR
2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 6000L Series Oscilloscopes Service Guide
Page 3
In This Service Guide
This book provides the service information for the Keysight 6000L Series oscilloscopes. This manual is divided into these chapters:
1 Characteristics and Specifications
This chapter lists characteristics and specifications for the Keysight 6000L Series oscilloscopes.
2 Testing Performance
This chapter explains how to verify correct oscilloscope operation and perform tests to ensure that the oscilloscope meets the performance specifications.
3 Calibrating and Adjusting
This chapter explains how to adjust the oscilloscope for optimum operating performance.
4 Troubleshooting
This chapter begins with suggestions for solving general problems that you may encounter with the oscilloscope. Procedures for troubleshooting the oscilloscope follow the problem solving suggestions.
5 Replacing 6000L Assemblies
This chapter describes how to remove assemblies from the 6000L Series oscilloscope.
6Replaceable Parts
This chapter describes how to order replaceable assemblies and parts for the Keysight 6000L Series oscilloscopes. It includes diagrams and parts lists for hardware that you can order.
7Safety Notices
At the front of the book you will find safety notice descriptions and document warranties.
6000L Series Oscilloscopes Service Guide 3
Page 4
Using this book with the 6000L Series oscilloscopes
The 6000L Series oscilloscopes do not have a built-in display or front panel control keys. If you are using a 6000L Series oscilloscope, and this book refers to using front panel controls, you can use the built-in Web control feature described in the 6000L Series Oscilloscopes User’s Guide to complete the instructions. If you do not have the 6000L Series Oscilloscopes User’s Guide, you may obtain a printable electronic copy at www.Keysight.com/find/6000manual.
Digital Channels
Because all of the oscilloscopes in the Keysight 6000L Series have analog channels, the analog channel topics in this book apply to all instruments. Whenever a topic discusses the digital channels, that information applies to DSO models that have been upgraded to an MSO only.
Abbreviated instructions for pressing a series of keys
Instructions for pressing a series of keys are written in an abbreviated manner. Instructions for pressing Key1, then pressing Softkey2, then pressing Softkey3 are abbreviated as follows: Press [Key1] & Softkey2 & Softkey3. The keys may be front panel keys, or softkeys, which are located directly below the oscilloscope display.
Warranty Status
To find the warranty status of your oscilloscope:
1 Direct your web browser to: www.keysight.com/find/warrantystatus 2 Enter your product’s model number and serial number. The system will search
for the warranty status of your product and display the results. If the system cannot find your product’s warranty status, select Contact Us and speak with an Keysight Technologies representative.
4 6000L Series Oscilloscopes Service Guide
Page 5

Contents

Figures / 9
Tab les / 11
1 Characteristics and Specifications
Power Requirements / 14
6000L Models / 14
Measurement Category / 15
Measurement Category / 15 Measurement Category Definitions / 15 6000L Series Maximum Input Voltages / 16
6000L Environmental Conditions / 17
Specifications / 18
Contact Us / 18
2 Testing Performance
Overview / 20
Conventions / 20
List of Test Equipment / 21
To construct the test connector (for use with MSO models
only) / 22
To test digital channels (MSO models only) / 24
To verify digital channel threshold accuracy (MSO models
only) / 25 When to Test / 25
6000L Series Oscilloscopes Service Guide 5
Page 6
Contents
What to Test / 25 Verifying Test Results / 25
To verify voltage measurement accuracy / 29
To verify bandwid th / 33
To verify time scale accuracy / 37
To verify horizontal delta t accuracy / 39
To verify trigger sensitivity / 41
Test Internal Trigger Sensitivity / 42 Test External Trigger Sensitivity / 43
Keysight 6000L Series Oscilloscopes Performance Test Record / 46
3 Calibrating and Adjusting
User Calibration / 51
To perform User C al / 51 User Cal Status / 53
4 Troubleshooting
Solving General Problems with the Oscilloscope / 56
Troubleshooting the Oscilloscope / 59
Equipment required for troubleshooting / 59 To check out the oscilloscope / 60 To verify basic oscilloscope operation / 62 To compensate the analog probes / 63 Troubleshooting Flowchart / 64 To check the 6000L Series oscilloscope power supply / 65 To check the 6000L Series system board / 67 To check the 6000L Series display output / 68 To check the fan / 68 To run the internal self-tests / 68 To verify default setup / 69
6 6000L Series Oscilloscopes Service Guide
Page 7
5 Replacing 6000L Assemblies
To remove the bottom cover / 73
To remove the front panel assembly / 74
To remove the system board / 76
To remove the power supply / 79
To remove the fan / 81
To remove the AC power input assembly / 83
To remove the power shaft / 85
6 Replaceable Parts
Ordering Replaceable Parts / 88
Listed Parts / 88 Unlisted Parts / 88 Direct Mail Order System / 88 Exchange Assemblies / 89
Contents
Replaceable Parts for 6000L Series Oscilloscopes / 90
6000L Series Oscilloscope Exploded Views / 90 6000L Series Oscilloscope Replaceable Parts List / 92
7 Safety Notices
Warnings / 96
To clean the instr ument / 97
Safety Symbols / 98
Index
6000L Series Oscilloscopes Service Guide 7
Page 8
Contents
8 6000L Series Oscilloscopes Service Guide
Page 9

Figures

Figure 1. Constructing the 8-by-2 Connector / 23 Figure 2. Setting Up Equipment and Test Connector for the Threshold Test / 26 Figure 3. Connect equipment for voltage measurement accuracy test / 31 Figure 4. Using a Blocking Capacitor to Reduce Noise / 32 Figure 5. Connect equipment for bandwidth test / 34 Figure 6. Connect equipment for internal trigger sensitivity test / 42 Figure 7. Connect equipment for external trigger sensitivity test / 44 Figure 8. User Calibration cable for 4-channel oscilloscope / 52 Figure 9. 6000L Series LED indicators / 61 Figure 10. Example pulses / 63 Figure 11. System Board Test Points / 66 Figure 12. Default setup screen / 70 Figure 13. Removing the cover / 73 Figure 14. Removing the intensity knob / 74 Figure 15. Disconnecting ribbon cable and releasing tab retainers / 75 Figure 16. Preparing to remove the system board / 76 Figure 17. Avoid damage to thermal pads (500 MHz models only) / 77 Figure 18. Removing the power supply. / 79 Figure 19. Removing the fan. / 81 Figure 20. Installing the fan. / 82 Figure 21. Removing the AC power input board. / 84 Figure 22. Removing the power shaft. / 85 Figure 23. 6000L Series Oscilloscope Exploded View / 91
6000L Series Oscilloscopes Service Guide 9
Page 10
Figures
10 6000L Series Oscilloscopes Service Guide
Page 11

Tables

Table 1. Conventions / 20 Table 2. List of test equipment / 21 Table 3. Materials required to construct the test connectors / 22 Table 4. Equipment Required to Test Digital Channel Threshold Accuracy / 25 Table 5. Threshold Accuracy Voltage Test Settings / 27 Table 6. Equipment Required to Verify Voltage Measurement Accuracy / 29 Table 7. Settings Used to Verify Voltage Measurement Accuracy / 30 Table 8. Equipment Required to Verify Bandwidth / 33 Table 9. Equipment Required to Verify Time Scale Accuracy / 37 Table 10. Equipment Required to Verify Horizontal Dt Accuracy / 39 Table 11. Equipment Required to Verify Trigger Sensitivity / 41 Table 12. Equipment Required to Troubleshoot the Oscilloscope / 60 Table 13. System Board Test Points / 67 Table 14. 6000L Series Oscilloscope Replaceable Parts / 92
6000L Series Oscilloscopes Service Guide 11
Page 12
Tables
12 6000L Series Oscilloscopes Service Guide
Page 13
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide
1 Characteristics and
Specifications
Power Requirements / 14 Measurement Category / 15 6000L Environmental Conditions / 17 Specifications / 18 Contact Us / 18
This chapter contains a partial list of characteristics and specifications for the Keysight InfiniiVision 6000L Series Oscilloscopes.
For a full list of Keysight InfiniiVision 6000L Series Oscilloscopes characteristics and specifications see the data sheet.
The data sheet is available at www.keysight.com/find/6000.
13
Page 14
1 Characteristics and Specifications

Power Requirements

6000L Models

Line voltage, frequency, and power:
• ~Line 100-240 Vac,
• 50/60 Hz
•80W max
14 6000L Series Oscilloscopes Service Guide
Page 15

Measurement Category

WARNING

Measurement Category

The InfiniiVision oscilloscopes are not intended to be used for measurements in Measurement Category II, III, or IV.
Use this instrument only for measurements within its specified measurement category (not rated for CAT II, III, IV). No transient overvoltages allowed.

Measurement Category Definitions

The "Not rated for CAT II, III, IV" measurement category is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits. In the latter case, transient stresses are variable; for that reason, the transient withstand capability of the equipment is made known to the user.
Measurement category II is for measurements performed on circuits directly connected to the low voltage installation. Examples are measurements on household appliances, portable tools and similar equipment.
Characteristics and Specifications 1
6000L Series Oscilloscopes Service Guide 15
Measurement category III is for measurements performed in the building installation. Examples are measurements on distribution boards, circuit-breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to the fixed installation.
Measurement category IV is for measurements performed at the source of the low-voltage installation. Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.
Page 16
1 Characteristics and Specifications
CAUTION
CAUTION

6000L Series Maximum Input Voltages

Maximum input voltage at analog inputs
Do not exceed the maximum input voltage rating.
The maximum input voltage for the 50 Ω input impedance setting is ±5 V.
Input protection is enabled in 50 mode and the 50 load will disconnect if greater than 5 Vrms is detected. However the inputs could still be damaged, depending on the time constant of the signal. The 50 input protection functions only when the oscilloscope is powered on.
The maximum input voltage for the 1 M input impedance setting is 30 Vrms or ±40 Vmax (DC+Vpeak)
Probing technology allows for testing of higher voltages; the included 10073D 10:1 probe supports 300 Vrms or ±400 Vmax (DC+Vpeak). No transient overvoltage allowed.
Maximum input voltage at digital channels
±40 V peak
16 6000L Series Oscilloscopes Service Guide
Page 17

6000L Environmental Conditions

Environment Indoor use only.
Ambient temperature Operating –10 °C to +40 °C; non-operating –40 °C to
+70 °C
Humidity Operating 95% RH at 40 °C for 24 hr; non-operating 90% RH at 65 °C
for 24 hr
Altitude Operating to 2,000 m (6,561 ft); non-operating to 15,244 m (50,000 ft)
Overvoltage Category This product is intended to be powered by MAINS that
comply to Overvoltage Category II, which is typical of cord-and-plug connected equipment.
Pollution Degree The InfiniiVision 6000L Series oscilloscopes may be operated
in environments of Pollution Degree 2 (or Pollution Degree 1).
Characteristics and Specifications 1
Pollution Degree Definitions Pollution Degree 1: No pollution or only dry,
non-conductive pollution occurs. The pollution has no influence. Example: A clean room or climate controlled office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs. Occasionally a temporary conductivity caused by condensation may occur. Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive pollution occurs which becomes conductive due to condensation which is expected. Example: Sheltered outdoor environment.
6000L Series Oscilloscopes Service Guide 17
Page 18
1 Characteristics and Specifications

Specifications

Please see the InfiniiVision 6000L Series Oscilloscopes Data Sheet for complete, up-to-date specifications and characteristics.
To download a copy of the data sheet please visit:
www.keysight.com/find/6000.
Or go to the Keysight home page at www.keysight.com and search for 6000L series oscilloscopes data sheet.
To order a data sheet by phone, please contact your local Keysight office.

Contact Us

To contact Keysight, see: www.keysight.com/find/contactus
18 6000L Series Oscilloscopes Service Guide
Page 19
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide

2 Testing Performance

Overview / 20 List of Test Equipment / 21 To construct the test connector (for use with MSO models only) / 22 To test digital channels (MSO models only) / 24 To verify digital channel threshold accuracy (MSO models only) / 25 To verify voltage measurement accuracy / 29 To verify bandwidth / 33 To verify time scale accuracy / 37 To verify horizontal delta t accuracy / 39 To verify trigger sensitivity / 41 Keysight 6000L Series Oscilloscopes Performance Test Record / 46
This chapter explains how to verify correct oscilloscope operation and perform tests to ensure that the oscilloscope meets the performance specifications.
19
Page 20
2 Testing Performance

Overview

To completely test and troubleshoot the mixed-signal oscilloscope, you will create and use a test connector accessory, as described in this chapter.
• The test connector is only required for oscilloscopes that have the MSO option licensed (enabled).
• The connector is used in the digital channel threshold accuracy test.
• The test connector makes it easy for you to connect the oscilloscope probes to function generators and measurement equipment with minimum electrical distortion.
Let the Equipment Warm Up Before Testing
For accurate test results, let the test equipment and the oscilloscope warm up 30 minutes before testing.
Verifying Test Results
During the tests, record the readings in the Performance Test Record on page 46. To verify whether a test passes, verify that the reading is within the limits in the Performance Test Record.
If a performance test fails
If a performance test fails, first perform the User Cal procedure

Conventions

The following conventions will be used when referring to oscilloscope models throughout this chapter.
Table 1 Conventions
Models Referred to as:
DSO6014L 100 MHz Models
DSO6054L 500 MHz Models
DSO6104L 1 GHz Models
20 6000L Series Oscilloscopes Service Guide
Page 21
Testing Performance 2

List of Test Equipment

Below is a list of test equipment and accessories required to perform the performance test verification procedures.
Table 2 List of test equipment
Recommended Model/
Equipment Critical Specifications
Tes t conn ector , 8-by- 2* Se e page 22 for instructions on building test connector. n/a
Digital Multimeter 0.1 mV resolution, 0.005% accuracy Keysight 34401A
Power Splitter Outputs differ by 0.15 dB Keysight 11667B
Oscilloscope Calibrator DC offset voltage of -5.5 V to 35.5 V, 0.1 V resolution Fluke 5820A
Signal Generator 25 MHz, 100 MHz, 500 MHz, and 1 GHz sine waves Keysight N5181A, or
Power Meter/Sensor 1 GHz
Oscilloscope Calibrator 25 MHz—500 MHz sine wave, 5 ppm Fluke 5820A
BNC banana cable Keysight 11001-66001
BNC cable (qty 3) Keysight 10503A
Cable Type N (m) 609.6 mm (24 in.) Keysight 11500B
Probe cable* Keysight 01650-61607
Shorting Cap BNC Keysight 1250-0774
Adapter BNC(f) to banana(m) Keysight 1251-2277
Adapter BNC Tee (m) (f) (f) Keysight 1250-0781
Adapter Type N (m) to BNC (m) Keysight 1250-0082
Blocking capacitor Keysight 10240-60001
Adapter (qty 3) N(m) to BNC(f) Keysight 1250-0780
Feedthrough (qty 2) 50
* Required only for testing digital channels of oscilloscopes that have the MSO option.
Some parts and equipment are available at www.keysight.com or www.parts.keysight.com.
±3% accuracy Keysight E4418B/8482A
Ω BNC (f) to BNC (m) Keysight 0960-0301
Part Number
E4400B, or 8648A
or Pomona 3288 with Pomona 3533
6000L Series Oscilloscopes Service Guide 21
Page 22
2 Testing Performance

To construct the test connector (for use with MSO models only)

Keysight 6000L Series Oscilloscopes that have digital channels enabled require the test connector described below. Follow the steps to build the test connector.
Table 3 Materials required to construct the test connectors
Description Recommended Part Qty
BNC (f) Connector Keysight 1250-1032 or
Pomona 4578
Berg Strip, 8-by-2 1
Jumper wire
1 Obtain a BNC connector and an 8-by-2 section of Berg strip. 2 On one side of the Berg strip, solder a jumper wire to all of the pins (shown in
Figure 1 on page 23).
3 On the other side of the Berg strip, solder another jumper wire to all of the pins. 4 Solder the center of the BNC connector to a center pin on one of the rows on
the Berg strip.
5 Solder the ground tab of the BNC connector to a center pin on the other row on
the Berg strip.
1
22 6000L Series Oscilloscopes Service Guide
Page 23

Figure 1 Constructing the 8-by-2 Connector

Jumper (2)
Signal Lead (from scope’s MSO cable)
8 x 2 Berg Strip
BNC Panel Mount Connector
Ground Lead (from scope’s MSO cable)
Testing Performance 2
6000L Series Oscilloscopes Service Guide 23
Page 24
2 Testing Performance

To test digital channels (MSO models only)

The acquisition system testing provides confidence that the acquisition system is functioning correctly. It does not, however, check a particular specification.
1 Disconnect all probes from the circuit under test and from any other input
source.
2 Using probe leads and grabbers, connect digital channels D0, D1, D2, and D3
to the Probe Comp signal on the center of the front panel.
3 Press the AutoScale key.
If four square waves appear, the acquisition system is functioning correctly.
If the square waves do not appear, go to the “Troubleshooting” chapter. Then return here to finish testing the digital channels.
4 Disconnect the digital channels from the calibration point. 5 Use steps 2 and 3 to test the following sets of digital channels. After you test
one set of digital channels, remove them before connecting the next set.
D4, D5, D6, D7
D8, D9, D10, D11
D12, D13, D14, D15
24 6000L Series Oscilloscopes Service Guide
Page 25
Testing Performance 2

To verify digital channel threshold accuracy (MSO models only)

This test verifies the digital channel threshold accuracy specification of the Keysight 6000L Series Oscilloscopes.
Threshold accuracy test limits= ±(100 mV + 3% of threshold setting)

When to Test

You should perform this test every 12 months or after 2000 hours of operation, whichever comes first.

What to Test

Use these instructions to test the threshold settings of digital channels D7-D0. Then, use the same instructions to test digital channels D15-D8.

Verifying Test Results

After each threshold test, record the voltage reading in the Performance Test Record on page 46. To verify whether a test passes, verify that the voltage reading is within the limits in the Performance Test Record.
Table 4 Equipment Required to Test Digital Channel Threshold Accuracy
Equipment Critical Specifications Recommended Model/Part
Digital Multimeter 0.1 mV resolution,
0.005% accuracy
Oscilloscope Calibrator DC offset voltage 6.3 V Fluke 5820A
BNC-Banana Cable Keysight 11001-60001
BNC Tee Keysight 1250-0781
BNC Cable Fluke 50
BNC Test Connector, 8-by-2 User-built (See page 22.)
Probe Cable Keysight 01650-61607
6000L Series Oscilloscopes Service Guide 25
Keysight 34401A
Ω cable, P/N 686318
Page 26
2 Testing Performance
Test Connector
BNC-Banana cable
Oscilloscope Calibrator
Digital Multimeter
BNC Tee
Probe Cables
Channels 8 - 15
Channels 0 - 7
6000L Series Oscilloscope
1 Turn on the test equipment and the oscilloscope. Let them warm up for 30
2 Set up the oscilloscope calibrator.
3 Use the 8-by-2 test connector and the BNC cable assembly to connect digital
minutes before starting the test.
a Set the oscilloscope calibrator to provide a DC offset voltage at the
Channel 1 output.
b Use the multimeter to monitor the oscilloscope calibrator DC output
voltage.
channels D0-D7 to one side of the BNC Tee. Then connect the D0-D7 ground lead to the ground side of the 8-by-2 connector. See Figure 2.
Figure 2 Setting Up Equipment and Test Connector for the Threshold Test
26 6000L Series Oscilloscopes Service Guide
Page 27
Testing Performance 2
4 Use a BNC-banana cable to connect the multimeter to the other side of the
BNC Tee.
5 Connect the BNC Tee to the Channel 1 output of the calibrator as shown in
Figure 2.
6 On the oscilloscope, press the D15 Thru D0 key, then press the Thresholds
softkey, then press the D15 Thru D0 softkey repeatedly until the check mark is next to User.
7 Press the oscilloscope User softkey, then turn the Entry knob ( ) on the front
panel on the oscilloscope to set the threshold test settings as shown in Table 5.
Table 5 Threshold Accuracy Voltage Test Settings
Threshold voltage setting (in oscilloscope User softkey)
+5.00 V +5.250 V ±1 mV dc Lower limit = +4.750 V
–5.00 V –4.750 V ±1 mV dc Lower limit = –5.250 V
0.00 V +100m V ±1 mV dc Upper limit = +100 mV
8 Do the following steps for each of the threshold voltage levels shown in Table 5.
a Set the threshold voltage shown in the User softkey using the Entry knob on
the oscilloscope.
b Enter the corresponding DC offset voltage on the oscilloscope calibrator
front panel. Then use the multimeter to verify the voltage.
Digital channel activity indicators are displayed on the status line at the top of the oscilloscope display. The activity indicators for D7-D0 should show all of the channels at digital high levels.
c Use the knob on the oscilloscope calibrator to decrease the offset voltage, in
increments of 10 mV, until the activity indicators for digital channels D7-D0 are all at digital low levels. Record the oscilloscope calibrator voltage in the Performance Test Record (see page 46).
d Use the knob on the oscilloscope calibrator to increase the offset voltage, in
increments of 10 mV, until the activity indicators for digital channels D7-D0
DC offset voltage setting (on oscilloscope calibrator) Limits
Upper limit = +5.250 V
Upper limit = –4.750 V
Lower limit = –100 mV
6000L Series Oscilloscopes Service Guide 27
Page 28
2 Testing Performance
9 Use the 8-by-2 test connector to connect digital channels D15-D8 to the
10 Repeat this procedure (steps 7 and 8) for digital channels D15-D8 to verify
are all at digital high levels. Record the oscilloscope calibrator voltage in the Performance Test Record (see page 46).
Before proceeding to the next step, make sure that you have recorded the oscilloscope calibrator voltage levels for each of the threshold settings shown in Tab le 5.
output of the oscilloscope calibrator. Then connect the D15-D8 ground lead to the ground side of the 8-by-2 connector.
threshold accuracy and record the threshold levels in the Performance Test Record (see page 46).
28 6000L Series Oscilloscopes Service Guide
Page 29

To verify voltage measurement accuracy

This test verifies the accuracy of the analog channel voltage measurement for each channel (DC Vertical Gain Accuracy and Dual Cursor Accuracy specifications). In this test, you will measure the dc voltage output of an oscilloscope calibrator using dual cursors on the oscilloscope, and compare the results with the multimeter reading.
Test limits: ±2.0% of full scale ±1 LSB*
• On 500 MHz and 1 GHz models, full scale is defined as 32 mV on the 2 mV/div range.
• On 100 MHz models full scale is defined as 16 mV on the 1 mV/div range.
• Full scale on all other ranges is defined as 8 divisions times the V/div setting.
*1 LSB = 0.4% of full scale
Table 6 Equipment Required to Verify Voltage Measurement Accuracy
Equipment Critical Specifications Recommended Model/Part
Testing Performance 2
Oscilloscope Calibrator 14 mV to 35 Vdc,
0.1 V resolution
Digital multimeter Better than 0.01% accuracy Keysight 34401A
Cable BNC, Qty 2 Keysight 10503A
Shorting cap BNC Keysight 1250-0774
Adapter BNC (f) to banana (m) Keysight 1251-2277
Adapter BNC tee (m) (f) (f ) Keysight 1250-0781
Blocking capacitor Keysight 10240B
1 Set up the oscilloscope.
a Adjust the channel 1 position knob to place the baseline at 0.5 major division
from the bottom of the display.
b Set the Volts/Div setting to the value in the first line in Table 7.
6000L Series Oscilloscopes Service Guide 29
Fluke 5820A
Page 30
2 Testing Performance
Table 7 Settings Used to Verify Voltage Measurement Accuracy
Volts/Div Setting Oscilloscope Calibrator
Setting
5 V/Div 35 V 34.04 V to 35.96 V
2 V/Div 14 V 13.616 V to 14.384 V
1 V/Div 7 V 6.808 V to 7.192 V
500 mV/Div 3.5 V 3.404 V to 3.596 V
200 mV/Div 1.4 V 1.3616 V to 1.4384 V
100 mV/Div 700 mV 680.8 mV to 719.2 mV
50 mV/Div 350 mV 340.4 mV to 359.6 mV
20 mV/Div 140 mV 136.16 mV to 143.84 mV
10 mV/Div 70 mV 68.08 mV to 71.92 mV
5 mV/Div 35 mV 34.04 mV to 35.96 mV
2mV/Div1 (for 1 GHz and 500 MHz models)
2mV/Div (for 100 MHz models)
1mV/Div
1
2
Full scale is defined as 32 mV on the 2 mV/div range for 500 MHz and 1 GHz models.
14 mV 13.232 mV to 14.768 mV
14 mV 13.616 mV to 14.384 mV
7 mV 6.616 mV to 7.384 mV
Test Limits
2
Full scale is defined as 16 mV on the 1 mV/div range for 100 MHz models.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
c Press the Acquire key. Then press the Averaging softkey and set #Avgs to 64.
Wait a few seconds for the measurement to settle.
2 Press the Cursors key, set the Mode softkey to Normal, then press the XY softkey
and select Y. Press the Y1 softkey, then use the Entry knob (labeled on the front panel) to set the Y1 cursor on the baseline of the signal.
3 Use the BNC tee and cables to connect the oscilloscope calibrator /power
supply to both the oscilloscope and the multimeter (see Figure 3).
30 6000L Series Oscilloscopes Service Guide
Page 31
Testing Performance 2
Oscilloscope
Oscilloscope Calibrator
Digital Multimeter
BNC Tee
BNC (f) to dual banana

Figure 3 Connect equipment for voltage measurement accuracy test

4 Adjust the output so that the multimeter reading displays the first Volts/div
calibrator setting value in Tab le 7.
Wait a few seconds for the measurement to settle.
5 Press the Y2 softkey, then position the Y2 cursor to the center of the voltage
trace using the Entry knob.
The ΔY value on the lower line of the display should be within the test limits of Tab le 7. If a result is not within the test limits, go to the “Troubleshooting” chapter. Then return here.
6000L Series Oscilloscopes Service Guide 31
6 Continue to check the voltage measurement accuracy with the remaining
Volts/div setting values in Table 7 .
Page 32
2 Testing Performance
To oscilloscope input
BNC shorting cap
Blocking Capacitor
7 When you are finished checking all of the voltage values, disconnect the
8 Repeat this procedure for the remaining channels to be tested.
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 2 mV/div and 5 mV/div, noise may be a factor. To eliminate the noise, add a BNC Tee, blocking capacitor, and BNC shorting cap at the oscilloscope channel input to shunt the noise to ground. See Figure 4.
oscilloscope calibrator from the oscilloscope.

Figure 4 Using a Blocking Capacitor to Reduce Noise

32 6000L Series Oscilloscopes Service Guide
Page 33

To verify bandwidth

This test checks the bandwidth of the oscilloscope. In this test you will use a signal generator and a power meter.
1GHz Models
Test limits at 2 mV/div to 5 V/div
• All channels (-3 dB)
dc to 1 GHz
500 MHz Models
Test limits at 2 mV/div to 5 V/div
• All channels (-3 dB)
dc to 500 MHz
100 MHz Models
Testing Performance 2
Test limits at 1 mV/div to 5 V/div
• All channels (-3 dB)
dc to 100 MHz
Table 8 Equipment Required to Verify Bandwidth
Equipment Critical Specifications Recommended Model/Part
Signal Generator 100 kHz - 1 GHz at 200 mVrms Keysight N5181A, or E4400B,
or 8648A
Power Meter/Sensor 1 MHz - 1 GHz ±3% accuracy Keysight E4418B/8482A
Power Splitter outputs differ by < 0.15 dB Keysight 11667A
Cable Type N (m) 24 inch Keysight 11500B
Adapter Type N (m) to BNC (m) Keysight 1250-0082 Feedthrough 50Ω BNC (f) to BNC (m) Keysight 0960-0301
6000L Series Oscilloscopes Service Guide 33
Page 34
2 Testing Performance
Oscilloscope
Signal Generator
Power Meter
Power Splitter
Power Sensor
N to BNC adapter
N Cable
50Ω Feedthrough*
* 100 MHz models only.
1 Connect the equipment (see Figure 5).
a Use the N cable to connect the signal generator to the input of the power
splitter input.
b Connect the power sensor to one output of the power splitter. c For 500 MHz and 1 GHz oscilloscope models use an N-to-BNC adapter to
connect the other splitter output to the channel 1 input.
d For 100 MHz oscilloscopes, use an N-to-BNC adapter and 50Ω feedthrough
termination to connect the other splitter output to the channel 1 input on the oscilloscope.

Figure 5 Connect equipment for bandwidth test

2 Set up the power meter.
34 6000L Series Oscilloscopes Service Guide
Page 35
Testing Performance 2
Vin
1MHz
Pmeas
1MHz
50Ω×=
Set the power meter to display measurements in units of watts.
3 Set up the oscilloscope.
a Press the Save/Recall key, then press the Default Setup softkey. b Press the Acquire key, then press the Realtime softkey to unselect Realtime. c Set channel 1 Coupling to DC. d On 500 MHz and 1 GHz models only, set channel 1 Imped to 50 Ohm. e Set the time base to 500 ns/div. f Set the Volts/Div for channel 1 to 200 mV/div. g Press the Acquire key, then press the Averaging softkey. h Turn the Entry knob to set # Avgs to 8 averages.
4 Set the signal generator for 1 MHz and six divisions of amplitude.
The signal on the oscilloscope screen should be about five cycles at six divisions amplitude.
5 Set up the Amplitude measurement
a Press the Quick Meas key. b Press the Clear Meas softkey. c Press the Select: softkey and use the Entry knob to select Std Deviation within
the select menu.
d Press the Measure Std Dev softkey.
6 Note the oscilloscope Std Dev(1) reading at the bottom of the screen. (This is
the RMS value with any dc offset removed.)
7 Set the power meter Cal Factor % to the 1 MHz value on the calibration chart
on the power sensor.
8 Note the reading on the power meter and covert to Vrms using the expression:
For example, if the power meter reading is 892 uW, then Vin
1/2
50Ω)
9 Change the signal generator output frequency according to the maximum
frequency for the oscilloscope using the following:
1GHz Models: 1GHz
500 MHz Models: 500 MHz
6000L Series Oscilloscopes Service Guide 35
= 211.2 mV
rms
.
= (892*10-6 *
1MHz
Page 36
2 Testing Performance
Vin
maxfreq
Pmeas
maxfreq
50Ω×=
response(dB) = 20 log
10
freqmax freqmax
Vin / Vout
Vin / Vout
response(dB) = 20 log
10
mV 185.3 / mV 161.6
= -1.16 dB
Example If:
100 MHz Models: 100 MHz
10 Referencing the frequency from step 9, set the power meter Cal Factor % to the
frequency value on the calibration chart on the power sensor.
11 Set the oscilloscope sweep speed according to the following:
1 GHz Models: 500 ps/div
500 MHz Models: 1 ns/div
100 MHz Models: 5 ns/div
12 Note the oscilloscope Std Dev(1) reading at the bottom of the screen. 13 Note the reading on the power meter and covert to Vrms using the expression:
14 Calculate the response using the expression:
MHz1MHz1
Pmeas Std Dev(n) Pmeas Std Dev(n)
Then after converting the values from the power meter to Vrms:
1_MHz
1MHz
max_freq
max freq
= 892 uW
= 210.4 mV
= 687 uW
= 161.6 mV
36 6000L Series Oscilloscopes Service Guide
mV211.2mV/ 210.4
15 The result from step 14 should be within -3.0 dB. Record the result in the
Performance Test Record (see page 46).
16 Move the power splitter from the channel 1 to the channel 2 input. 17 Turn off the current channel and turn on the next channel using the channel
keys.
18 Repeat steps 3 through 17 for the remaining channels, setting the parameters
of the channel being tested where appropriate.
Page 37

To verify time scale accuracy

Date Code: 0747 07 = Year 47 = Week
Time Scale Accuracy is a warranted specification for 6000L Series oscilloscopes manufactured after January 1, 2008.
If your oscilloscope’s serial tag has old-style (linear) barcodes instead of the 3-D barcodes shown above, or if the date indicated on the serial tag is older than January 1, 2008, then disregard this test.
This test verifies the accuracy of the time scale. In this test you will measure the absolute error of the timebase oscillator and compare the results to the specification.
Testing Performance 2
Table 9 Equipment Required to Verify Time Scale Accuracy
Equipment Critical Specifications Recommended Model/Part
Signal Generator 100 kHz - 1 GHz, 0.01 Hz frequency
resolution, jitter: < 2ps
Cable BNC, 3 feet Keysight 10503A
1 Set up the signal generator.
a Set the output to 10 MHz, approximately 1 V
2 Connect the output of the signal generator to oscilloscope channel 1 using the
BNC cable.
6000L Series Oscilloscopes Service Guide 37
Keysight N5181A, or E4400B, or 8648A
sine wave.
pp
Page 38
2 Testing Performance
Date Code: 0747 07 = Year 47 = Week
3 Set up the oscilloscope:
a Press AutoScale. b Set the oscilloscope Channel 1 vertical sensitivity to 200 mv/div. c Set the oscilloscope horizontal sweep speed control to 5 ns/div. d Adjust the intensity to get a sharp, clear trace. e Adjust the oscilloscope’s trigger level so that the rising edge of the
waveform at the center of the screen is located where the center horizontal and vertical grid lines cross (center screen).
f Ensure the horizontal position control is set to 0.0 seconds.
4 Make the measurement.
a Set oscilloscope horizontal sweep speed control to 1 ms/div. b Set horizontal position control to +1 ms (rotate control CCW). c Set the oscilloscope horizontal sweep speed control to 5 ns/div. d Record the number of nanoseconds from where the rising edge crosses the
center horizontal grid line to the center vertical grid line. The number of nanoseconds is equivalent to the time scale error in ppm.
e Use the date code on the oscilloscope’s serial tag to calculate the number of
years since manufacture. Include any fractional portion of a year.
f Calculate the test limits using the following formula:
TSA Limit: ± (15 + 2 * YrsSinceMfr) ppm
g Record the result and compare it to the limits in the Performance Test
Record (see page 46).
38 6000L Series Oscilloscopes Service Guide
Page 39

To verify horizontal delta t accuracy

The following test is optional because horizontal Δt accuracy is a characteristic, not a specification.
This test verifies the horizontal Δt accuracy. In this test, you will use the oscilloscope to measure the output of a time mark generator.
Test limits: ±0.0015% of reading ±0.1% of full scale ±20 ps (same channel)
Table 10 Equipment Required to Verify Horizontal Δt Accuracy
Equipment Critical Specifications Recommended Model/Part
Oscilloscope Calibrator Stability 5 ppm after 1/2 hour Fluke 5820A
Cable BNC, 3 feet Keysight 10503A Feedthrough 50Ω BNC (f) to BNC (m) Keysight 0960-0301
1 Connect the equipment:
a For 500 MHz and 1 GHz oscilloscope models, connect the calibrator output
to the oscilloscope channel 1 input.
b For 100 MHz oscilloscopes, use a 50Ω feedthrough termination to connect
the calibrator output to the oscilloscope channel 1 input.
2 Set up the signal source.
a Select Marker on the oscilloscope calibrator. b Set the calibrator for 100 μs markers (period = 100 μs).
3 Set up the oscilloscope.
a Set channel 1 Coupling to DC. b On 1 GHz and 500 MHz models only, set channel 1 Imped to 50 Ohm. c Press the Display key, then set the Vectors softkey to off. d Press the AutoScale key. e Set the time base to 20 μs/div. f Press the Main/Delayed (or Zoom) key, then set the Time Ref softkey to Left. g Adjust the Trigger Level knob to obtain a stable display.
Testing Performance 2
6000L Series Oscilloscopes Service Guide 39
Page 40
2 Testing Performance
4 Press the Quick Meas softkey, set the Source softkey to 1, then press Select and
choose Period. Press the Measure softkey and measure the following: Period 100 μs — The test limits are 99.8 μs to 100.2 μs.
If the measurements are not within the test limits, go to the “Troubleshooting” chapter. Then return here.
5 Change the calibrator to 100 ns markers. Change the time base to 20 ns/div.
Adjust the trigger level to obtain a stable display.
6 Measure the following. If the measurements are not within the test limits, go to
the “Troubleshooting” chapter. Then return here.
Period 100 ns— The test limits are 99.8 ns to 100.2 ns.
7 Change the time base and calibrator markers as follows:
a On 500 MHz and 1 GHz models, change time base to 2 ns/div and the
calibrator to 5 ns markers
b On 100 MHz models, change time base to 5 ns/div and the calibrator to
10 ns markers
8 Make the following measurements. If the measurements are not within the test
limits, go to the “Troubleshooting” chapter. Then return here. a For 500 MHz, and 1 GHz models, period 5 ns — the test limits are 4.96 ns to
5.04 ns.
b For 100 MHz models, period 10 ns — the test limits are 9.93 ns to 10.07 ns.
40 6000L Series Oscilloscopes Service Guide
Page 41

To verify trigger sensitivity

This test verifies the trigger sensitivity. In this test, you will apply a sine wave to the oscilloscope at the upper bandwidth limit. You will then decrease the amplitude of the signal to the specified levels, and check to see if the oscilloscope is still triggered.
The internal trigger sensitivity test is mandatory because it is a specification. The external trigger test is optional because it is a characteristic, not a specification.
Test limits for:
• Internal trigger:
< 10 mV/div: greater of 1 div or 5 mV
>=10 mV/div: 0.6 div
• External trigger:
1GHz and 500MHz models: (DSO6104L/6054L) DC to 500 MHz: < 500 mV
100 MHz models: (DSO6014L) DC to 100 MHz: < 500 mV
pp
pp
Testing Performance 2
pp
Table 11 Equipment Required to Verify Trigger Sensitivity
Equipment Critical Specifications Recommended Model/Part
Signal Generator 25-MHz, 100-MHz, 500-MHz, and
1GHz sine waves
Power splitter Outputs differ < 0.15 dB Keysight 11667A
Cable BNC, Qty 3 Keysight 10503A
Adapter N (m) to BNC (f), Qty 3 Keysight 1250-0780 Feedthrough 50Ω BNC (f) to BNC (m) Qty 2 Keysight 0960-0301
6000L Series Oscilloscopes Service Guide 41
Keysight N5181A, or E4400B, or 8648A
Page 42
2 Testing Performance
Signal Generator
N to BNC adapter
N Cable
50Ω Feedthrough*
* 100 MHz models only.
Oscilloscope

Test Internal Trigger Sensitivity

1 Connect the equipment (see Figure 6).
a For 500 MHz and 1 GHz models, connect the signal generator output to the
oscilloscope channel 1 input.
b For 100 MHz models, use a 50Ω feedthrough termination to connect the
signal generator output to the oscilloscope channel 1 input.
Figure 6 Connect equipment for internal trigger sensitivity test
42 6000L Series Oscilloscopes Service Guide
Page 43
Testing Performance 2
2 Verify the trigger sensitivity at maximum bandwidth.
1 GHz models: 1 GHz 500 MHz models: 500 MHz 100 MHz models: 100 MHz
a Press the Save/Recall key, then press the Default Setup softkey. b On 500 MHz and 1 GHz models, set channel 1 Imped to 50 Ohm. c Set the output frequency of the signal generator to the maximum bandwidth
of the oscilloscope and set the amplitude to about 10 mV
d Press the AutoScale key. e Set the time base to 10 ns/div. f Set channel 1 to 5 mV/div. g Decrease the amplitude from the signal generator until 1 vertical division of
the signal is displayed.
The trigger is stable when the displayed waveform is stable. If the trigger is not stable, try adjusting the trigger level. If adjusting the trigger level makes the trigger stable, the test still passes. If adjusting the trigger does not help, see the “Troubleshooting” chapter. Then return here.
h Record the result as Pass or Fail in the Performance Test Record (see
page 46).
3 Repeat this procedure for the remaining oscilloscope channels.
pp
.

Test External Trigger Sensitivity

The following test is optional because External Trigger Sensitivity is a characteristic, not a specification.
Verify the external trigger sensitivity at these settings:
1 GHz and 500 MHz bandwidth models:
500 MHz, 500 mV
100 MHz bandwidth models:
100 MHz, 500 mV
1 Connect the equipment (see Figure 7).
a Use the N cable to connect the signal generator to the power splitter input. b Connect one output of the power splitter to the Aux Trig input through a 50Ω
feedthrough termination.
c Connect the power sensor to the other output of the power splitter.
6000L Series Oscilloscopes Service Guide 43
pp
pp
Page 44
2 Testing Performance
Oscilloscope
Signal Generator
Power Meter
Power Splitter
Power Sensor
N to BNC adapter
N Cable
50Ω Feedthrough*
* Required for all models.
44 6000L Series Oscilloscopes Service Guide
Figure 7 Connect equipment for external trigger sensitivity test
2 Set up the oscilloscope.
a Press the Save/Recall key, then press the Default Setup softkey.
3 Change the signal generator output frequency as follows:
For 500 MHz and 1 GHz models, set the signal generator output frequency
to 500 MHz
For 100 MHz models, set the signal generator output frequency to 100 MHz
Page 45
Testing Performance 2
4 Set the power meter Cal Factor % to the appropriate value (500 MHz or
100 MHz) on the calibration chart on the power sensor. If necessary, do a linear interpolation if a 500 MHz or 100 MHz factor is not included in the power meter’s calibration chart.
5 Adjust the signal generator output for reading on the power meter of 625μW.
(500mV Power = Vin
6 Press the Trigger Edge key, then press the Source softkey to set the trigger
source to External.
7 Check for stable triggering and adjust the trigger level if necessary. Triggering
is indicated by the Trig’d indicator at the top of the display. When it is flashing, the oscilloscope is not triggered. When it is not flashing, the oscilloscope is triggered.
8 Record the results as Pass or Fail in the Performance Test Record (see
page 46).
If the test fails, see the “Troubleshooting” chapter. Then return here.
= 176.78mV rms,
pp
2
/50Ω = 176.78 mV2/50Ω = 625μW.)
6000L Series Oscilloscopes Service Guide 45
Page 46
2 Testing Performance

Keysight 6000L Series Oscilloscopes Performance Test Record

Serial No. ______________________________________ Test by _____________________________ Test Interval ____________________________________ Work Order No. ______________________ Recommended Next Testing ________________________ Temperature ____________
Threshold Specification Limits Ch D7-D0 Ch D15-D8 Accuracy Test 5 V - 250 mV 4.750 V ________ ________ (100 mV + 3% of
threshold setting)
Voltage Measurement Accuracy Range Power Supply Setting Test Limits Channel 1 Channel 2 Channel 3* Channel 4*
5 V/Div 35 V 34.04 V to 35.96 V ________ ________ ________ ________ 2 V/Div 14 V 13.616 V to 14.384 V ________ ________ ________ ________ 1 V/Div 7 V 6.808 V to 7.192 V ________ ________ ________ ________ 500 mV/Div 3.5 V 3.404 V to 3.596 V ________ ________ ________ ________ 200 mV/Div 1.4 V 1.3616 V to 1.4384 V ________ ________ ________ ________ 100 mV/Div 700 mV 680.8 mV to 719.2 mV ________ ________ ________ ________ 50 mV/Div 350 mV 340.4 mV to 359.6 mV ________ ________ ________ ________ 20 mV/Div 140 mV 136.16 mV to 143.84 mV ________ ________ ________ ________ 10 mV/Div 70 mV 68.08 mV to 71.92 mV ________ ________ ________ ________ 5 mV/Div 35 mV 34.04 mV to 35.96 mV ________ ________ ________ ________ 2mV/Div
2mV/Div
1mV/Div
(500 MHz and 1 GHz models)
(100 MHz models)
(MSO/DSO6012A)
5 V + 250 mV 5.250 V ________ ________
-5 V - 250 mV -5.250 V ________ ________
-5 V + 250 mV -4.750 V ________ ________ 0 V - 100 mV -100 mV ________ ________ 0 V + 100 mV 100 mV ________ ________
14 mV 13.232 mV to 14.768 mV ________ ________ ________ ________
14 mV 13.616 mV to 14.384 mV ________ ________ ________ ________
7 mV 6.616 mV to 7.384 mV ________ ________ ________ ________
Bandwidth Model Test Limits Channel 1 Channel 2 Channel 3* Channel 4*
Time Scale Accuracy
610x -3 dB at 1 GHz ________ ________ ________ ________ 605x -3 dB at 500 MHz ________ ________ ________ ________ 601x -3 dB at 100 MHz ________ ________ ________ ________
Limits
Time Scale Accuracy is a warranted specification for 6000L Series instruments manufactured after January 1, 2008.
TSA Limit: ± (15 + 2 * YrsSinceMfr) ppm ___________ ___________ ___________
Calculated TSA Limit (ppm)
Measured time scale error (ppm)
Pass/Fail
46 6000L Series Oscilloscopes Service Guide
Page 47
Testing Performance 2
Horizontal Δt Accuracy
Period 100 μs 99.8 μs to 100.2 μs ________ Period 100 ns 99.8 ns to 100.2 ns ________ Period 10 ns Period 5 ns
1
100 MHz models
2
500 MHz and 1 GHz models
Trigger Sensitivity Test Limits Channel 1 Channel 2 Channel 3* Channel 4*
Trigger Sensitivity Test Limits Channel 1 Channel 2 Channel 3* Channel 4*
* Where applicable
Generator Setting Test Limits Results
1
2
Internal trigger
1 GHz models 0.6 division at 1 GHz ________ ________ ________ _______ 500 MHz models 0.6 division at 500 MHz ________ ________ ________ _______ 100 MHz models 0.6 division at 100 MHz ________ ________ ________ _______
External trigger (6014, 6054, 6104) 6104, 6054 500 mV at 500 MHz ________ ________ ________ ________ 6014 500 mV at 100 MHz ________ ________ ________ ________
9.93 ns to 10.07 ns ________
4.96 ns to 5.04 ns ________
1 division at 25 MHz ________ ________ ________ ________
6000L Series Oscilloscopes Service Guide 47
Page 48
2 Testing Performance
48 6000L Series Oscilloscopes Service Guide
Page 49
Keysight InfiniVision 6000L Series Oscilloscopes
WARNING
CAUTION
Service Guide

3 Calibrating and Adjusting

This chapter explains how to adjust the oscilloscope for optimum operating performance. You should perform self-calibration according to the following recommendations:
• Every 12 months or after 2000 hours of operation
• If the ambient temperature is >10 °C from the calibration temperature
• If you want to maximize the measurement accuracy
The amount of use, environmental conditions, and experience with other instruments help determine if you need shorter adjustment intervals.
Let the Equipment Warm Up Before Adjusting
Before you start the adjustments, let the oscilloscope and test equipment warm up for at least 30 minutes.
Read All Cautions and Warnings
Read the following cautions and warning before making adjustments or performing self-calibration.
HAZARDOUS VOLTAGES!
Read the safety notice at the front of this book before proceeding. Maintenance is performed with power supplied to the oscilloscope and with the protective covers removed. Only trained service personnel who are aware of the hazards involved should perform the maintenance. Whenever possible, perform the procedures with the power cord removed from the oscilloscope.
REMOVE POWER TO AVOID DAMAGE!
Do not disconnect any cables or remove any assemblies with power applied to the oscilloscope. Otherwise, damage to the oscilloscope can occur.
49
Page 50
3 Calibrating and Adjusting
CAUTION
CAUTION
USE EXTERNAL FAN TO REDUCE TEMPERATURE!
When you must operate the oscilloscope with its cover and main shield removed, use an external fan to provide continuous air flow over the samplers (the ICs with heat sinks on them). Air flow over the samplers is reduced when the cover and main shield is removed, which leads to higher than normal operating temperatures. Have the fan blow air across the system board where the heat sinks are located. If the cover is removed but the main shield remains installed and the bottom holes are not blocked, the instrument will cool properly.
AVOID DAMAGE TO ELECTRONIC COMPONENTS!
Electrostatic discharge (ESD) can damage electronic components. When you use any of the procedures in this chapter, use proper ESD precautions. As a minimum, place the oscilloscope on a properly grounded ESD mat and wear a properly grounded ESD strap.
50 6000L Series Oscilloscopes Service Guide
Page 51

User Calibration

Perform user-calibration:
• Each year or after 2000 hours of operation.
• If the ambient temperature is >10° C from the calibration temperature.
• If you want to maximize the measurement accuracy.
The amount of use, environmental conditions, and experience with other instruments help determine if you need shorter User Cal intervals.
User Cal performs an internal self-alignment routine to optimize the signal path in the oscilloscope. The routine uses internally generated signals to optimize circuits that affect channel sensitivity, offset, and trigger parameters. Disconnect all inputs and allow the oscilloscope to warm up before performing this procedure.
Performing User Cal will invalidate your Certificate of Calibration. If NIST (National Institute of Standards and Technology) traceability is required perform the procedures in Chapter 2 in this book using traceable sources.
Calibrating and Adjusting 3

To perform User C al

1 Set the rear-panel CALIBRATION switch to UNPROTECTED. 2 Connect short (12 inch maximum) equal length cables to each analog channel’s
BNC connector on the front of the oscilloscope. You will need four equal-length cables for a 4-channel oscilloscope.
6000L Series Oscilloscopes Service Guide 51
Page 52
3 Calibrating and Adjusting
Longer cable to TRIG OUT
To Channel 3
To Channel 2
To Channel 1
To Channel 4
Use 50Ω RG58AU or equivalent BNC cables when performing User Cal. a Connect BNC tees to the equal-length cables as shown below. Then connect
a BNC(f)-to-BNC(f) (barrel connector) to the tee as shown below.
Figure 8 User Calibration cable for 4-channel oscilloscope
3 Connect a BNC cable (40 inches maximum) from the TRIG OUT connector on
the rear panel to the BNC barrel connector.
4 Press the Utility key, then press the Service softkey.
52 6000L Series Oscilloscopes Service Guide
5 Begin the Self Cal by pressing the Start User Cal softkey. 6 When the User Cal is completed, set the rear-panel CALIBRATION switch to
PROTECTED.
Page 53

User Cal Status

Pressing the User Cal Status softkey displays the following summary results of the previous User Cal, and the status of probe calibrations for probes that can be calibrated. Note that AutoProbes do not need to be calibrated, but InfiniiMax can be calibrated.
Results: User Cal date: Change in temperature since last User Cal: Failure: Comments: Probe Cal Status:
Calibrating and Adjusting 3
6000L Series Oscilloscopes Service Guide 53
Page 54
3 Calibrating and Adjusting
54 6000L Series Oscilloscopes Service Guide
Page 55
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide

4 Troubleshooting

Solving General Problems with the Oscilloscope / 56 Troubleshooting the Oscilloscope / 59
This chapter begins with suggestions for solving general problems that you may encounter with the oscilloscope. It tells you what to do in these cases:
• If there is no trace display
• If the trace display is unusual or unexpected
• If you cannot see a channel
• If you cannot get any response from the oscilloscope
Procedures for troubleshooting the oscilloscope follow the problem solving suggestions. The troubleshooting section shows you how to:
• Check out the oscilloscope
• Check power supply
• Check the system board
• Check the display
• Check the fan
• Run internal self-tests
• Verify default setup
Read All Cautions and Warnings
Before you begin any troubleshooting, read all Warning and Cautions in the “Troubleshooting” section.
55
Page 56
4 Troubleshooting

Solving General Problems with the Oscilloscope

This section describes how to solve general problems that you may encounter while using the Keysight 6000L Series Oscilloscopes to make measurements.
After troubleshooting the oscilloscope, if you need to replace parts, refer to the “Replaceable Parts” chapter.
On 6000L Series oscilloscopes, connect an XGA display to the display port on the rear panel before you begin troubleshooting. Then connect the oscilloscope to the network, establish communication, and use the Remote Front Panel feature described in the 6000L Series Oscilloscopes User’s Guide to control the oscilloscope. If you do not have the user’s guide, you can download it from www.keysight.com/find/6000.
If there is no display
Check that the power cord is firmly seated in the oscilloscope power
receptacle.
Check that the power source is live.
Check that the front-panel power switch is on.
If there is still no display, go to the troubleshooting procedures in this chapter.
If there is no trace display
Check that the INTENSITY knob on the front panel is adjusted correctly.
Recall the default setup by pressing Save/Recall then Default Setup. This will
ensure that the trigger mode is Auto.
Check that the probe clips are securely connected to points in the circuit under
test, and that the ground is connected.
Check that the circuit under test is powered on.
Press the AutoScale key.
Obtain service from Keysight Technologies, if necessary.
56 6000L Series Oscilloscopes Service Guide
Page 57
Troubleshooting 4
If the trace display is unusual or unexpected
Check that the horizontal sweep speed control setting is correct for the
expected frequency range of the input signals.
The sampling speed of the oscilloscope depends on the time/division setting. It
may be that when time/division is set to slower speeds, the oscilloscope is sampling too slowly to capture all of the transitions on the waveform. Use peak detect mode.
Check that all oscilloscope probes are connected to the correct signals in the
circuit under test.
Ensure that the probe’s ground lead is securely connected to a ground point in
the circuit under test. For high-speed measurements, each probe’s individual ground lead should also be connected to a ground point closest to the signal point in the circuit under test.
Check that the trigger setup is correct.
A correct trigger setup is the most important factor in helping you capture the
data you desire. See the User’s Guide for information about triggering.
Check that infinite persistence in the Display menu is turned off, then press the
Clear Display softkey.
Press the AutoScale key.
If you cannot see a channel
Recall the default setup by pressing Save/Recall then Default Setup. This will
ensure that the trigger mode is Auto.
Check that the oscilloscope probe’s BNC connector is securely attached to the
oscilloscope’s input connector.
Check that the probe clips are securely connected to points in the circuit under
test.
Check that the circuit under test is powered on.
You may have pressed the AutoScale key before an input signal was available.
Performing the checks listed here ensures that the signals from the circuit under test will be seen by the oscilloscope. Perform the remaining checks in this topic to make sure the oscilloscope channels are on, and to obtain an automatic setup.
6000L Series Oscilloscopes Service Guide 57
Page 58
4 Troubleshooting
Check that the desired oscilloscope channels are turned on.
a Switch on the analog channel by selecting Analog from the menu bar of the
Remote Front Panel (6000L Series models).
b Switch on the digital channels by selecting Digital from the menu bar (6000L
Series models with the MSO option). Ensure that the desired channels are turned on.
Press the AutoScale key to automatically set up all channels.
58 6000L Series Oscilloscopes Service Guide
Page 59

Troubleshooting the Oscilloscope

WARNING
WARNING
CAUTION
CAUTION
The service policy for the Keysight 6000L Series Oscilloscopes is assembly level replacement. If you need parts or assistance from Keysight Technologies to repair your instrument, go to www.keysight.com and locate the service facility for your area.
HAZARDOUS VOLTAGES EXIST — REMOVE POWER FIRST !
The procedures described in this section are performed with power supplied to the oscilloscope and with the protective covers removed. Only trained service personnel who are aware of the hazards involved should perform the procedures. Whenever possible, perform the procedures with the power cord removed from the oscilloscope. Read the safety notice at the back of this book before proceeding.
HAZARDOUS VOLTAGES EXIST — HIGH VOLTAGE IS PRESENT ON DSO6000L SERIES POWER SUPPLY HEAT SINKS !
The power supply heat sinks of the 6000L Series oscilloscopes are at a high potential. This presents an electric shock hazard. Protect yourself from electric shock by keeping this area covered or by not coming in contact with the heat sinks when the power cord is attached to the oscilloscope!
Troubleshooting 4
REMOVE POWER TO AVOID DAMAGE !
Do not disconnect any cables or remove any assemblies while power is applied to the oscilloscope, or damage to the oscilloscope can occur.
AVOID ESD DAMAGE TO COMPONENTS !
ELECTROSTATIC DISCHARGE (ESD) can damage electronic components. Use proper ESD precautions when doing any of the procedures in this chapter. As a minimum, place the oscilloscope on a properly grounded ESD mat and wear a properly grounded ESD strap.

Equipment required for troubleshooting

The equipment listed in this table is required to troubleshoot the oscilloscope.
6000L Series Oscilloscopes Service Guide 59
Page 60
4 Troubleshooting
CAUTION
WARNING
CAUTION
Table 12 Equipment Required to Troubleshoot the Oscilloscope
Equipment Critical Specifications Recommended Model/Part
Digital multimeter Accuracy ±0.05,% 1 mV resolution Keysight 34401A Oscilloscope Capable of measuring 500 MHz signal. 1
MΩ input impedance.
Keysight DSO6102A, MSO6102A, DSO6102A, or MSO6104A

To check out the oscilloscope

1 Disconnect any external cables from the front panel. 2 Disconnect the power cord, then remove the cabinet following the instructions
on page 73 for Keysight 6000L Series models.
USE AN EXTERNAL FAN TO AVOID OVERHEATING COMPONENTS !
When you remove the oscilloscope cover and main shield, use an external fan to provide continuous air flow over the heat sinks. Air flow over the heat sinks is reduced when the cover and main shield are removed, which leads to higher than normal operating temperatures. Have the fan blow air across the system board where the heat sinks are located. Otherwise, damage to the components can occur.
HAZARDOUS VOLTAGES EXIST — HIGH VOLTAGE IS PRESENT ON DSO6000L SERIES POWER SUPPLY HEAT SINKS !
The power supply heat sinks of the 6000L Series oscilloscopes are at a high potential. This presents an electric shock hazard. Protect yourself from electric shock by keeping this area covered or by not coming in contact with the heat sinks when the power cord is attached to the oscilloscope!
60 6000L Series Oscilloscopes Service Guide
3 Connect the power cord to the rear of the oscilloscope, then to a suitable ac
voltage source.
The oscilloscope power supply automatically adjusts for input line voltages in the range of 100 to 240 VAC. Ensure that you have the correct line cord. The power cord provided is matched to the country of origin.
T
Use only the power cord that came with the oscilloscope
The power cord provided is matched to the country of origin of the order.
Page 61
Troubleshooting 4
WARNING
LAN Status
Indicator
Power
Switch
Power
Indicator
Reset
Switch
Always use a grounded power cord. Do not defeat the power cord ground.
• On 6000L Series models, press the power switch.
The Power indicator LED is green. The LAN indicator LED is bi-color (green and/or red).
Shortly after pressing the power switch, the power indicator and LAN
indicator will illuminate.
Both will extinguish.
Both will illuminate again. If a LAN is connected, the LAN indicator will
illuminate red. It will turn green when the LAN is connected as it is configured. The amount of time it takes the indicator to turn green depends on the network and the oscilloscope’s LAN configuration setup.
6000L Series Oscilloscopes Service Guide 61
Figure 9 6000L Series LED indicators
4 If the lights illuminate, then the power supply is probably working correctly. 5 Proceed to “To verify basic oscilloscope operation" on page 62.
Page 62
4 Troubleshooting

To verify basic oscilloscope operation

For 6000L Series oscilloscopes, connect the oscilloscope to your network or establish a point-to-point connection, and use the Remote Front Panel feature described in the 6000L Series Oscilloscopes User’s Guide to control the oscilloscope. If you do not have the user’s guide, you can download it from www.keysight.com/find/6000.
1 Press the Save/Recall key on the front panel, then press the Default Setup
softkey under the display. The oscilloscope is now configured to its default settings.
2 Connect an oscilloscope probe from channel 1 to the Probe Comp signal
terminal on the front panel.
3 Connect the probe’s ground lead to the ground terminal that is next to the
Probe Comp terminal.
4 Press AutoScale. 5 You should see a waveform on the oscilloscope’s display similar to this:
If you see the waveform, but the square wave is not shaped correctly as shown above, perform the procedure “To compensate the analog probes" on page 63.
If you do not see the waveform, ensure your power source is adequate, the oscilloscope is properly powered-on, and the probe is connected securely to the front-panel analog channel input BNC and to the Probe Comp terminal.
62 6000L Series Oscilloscopes Service Guide
Page 63
6 If you still do not see the waveform, use the troubleshooting flowchart in this
comp.cdr
Perfectly compensated
Over compensated
Under compensated
chapter to isolate the problem.

To compensate the analog probes

You should compensate your analog probes to match their characteristics to the oscilloscope’s channels. A poorly compensated probe can introduce measurement errors.
1 Perform the procedure “To verify basic oscilloscope operation" on page 62 2 Use a nonmetallic tool to adjust the trimmer capacitor on the probe for the
flattest pulse possible. The trimmer capacitor is located on the probe BNC connector.
Troubleshooting 4
Figure 10 Example pulses
3 Connect probes to all other analog channels (channels 2, 3, and 4 of a
4-channel oscilloscope). Repeat the procedure for each channel. This matches each probe to each channel.
The process of compensating the probes serves as a basic test to verify that the oscilloscope is functional.
6000L Series Oscilloscopes Service Guide 63
Page 64
4 Troubleshooting
Start
Test Power Supply
Okay?
Test System
Board
Okay?
Test Display
Okay?
End
Replace Power
Supply
Replace System
Board
Replace Display
Yes
Yes
Yes
No
No
No

Troubleshooting Flowchart

The following flowchart describes how to troubleshoot 6000L Series models.
64 6000L Series Oscilloscopes Service Guide
Page 65

To check the 6000L Series oscilloscope power supply

WARNING
CAUTION
NOTE
HAZARDOUS VOLTAGES EXIST — HIGH VOLTAGE IS PRESENT ON DSO6000L SERIES POWER SUPPLY HEAT SINKS !
The power supply heat sinks of the 6000L Series oscilloscopes are at a high potential. This presents an electric shock hazard. Protect yourself from electric shock by keeping this area covered or by not coming in contact with the heat sinks when the power cord is attached to the oscilloscope!
USE AN EXTERNAL FAN TO AVOID OVERHEATING COMPONENTS !
When you remove the oscilloscope cover and main shield, use an external fan to provide continuous air flow over the heat sinks. Air flow over the heat sinks is reduced when the cover and main shield are removed, which leads to higher than normal operating temperatures. Have the fan blow air across the system board where the heat sinks are located. Otherwise, damage to the components can occur.
If the power supply fuse is blown, the power supply is defective, and you must replace it. See the “Replaceable Parts” chapter for information about removing the power supply.
Troubleshooting 4
1 Disconnect the power cord from the oscilloscope. Then remove the oscilloscope
cover.
2 Connect the negative lead of the multimeter to a ground point on the
oscilloscope.
3 Connect the power cord and turn on the oscilloscope. 4 Measure the power supply voltage at J3200, pin 9 on the system board. See
Figure 11 on page 66. The voltage should be 15 V ±10%. If the voltage is
correct, the power supply is working correctly.
6000L Series Oscilloscopes Service Guide 65
Page 66
4 Troubleshooting
Pin 1
Pin 9
J3200
Figure 11 System Board Test Points
5 If the voltage is not correct:
a Disconnect the power cord. b Remove the fan. See “To remove the fan" on page 81. c Disconnect plug J3 from the power supply and from the system board. d Perform a continuity check on the “DC cable” (item W1 on page 93). If the
cable is shorted or open, replace it. If not, continue to the next step.
e Reconnect the “DC cable” to the power supply, but not to the system board.
66 6000L Series Oscilloscopes Service Guide
Page 67
WARNING
HAZARDOUS VOLTAGES EXIST — HIGH VOLTAGE IS PRESENT ON DSO6000L SERIES POWER SUPPLY HEAT SINKS !
The power supply heat sinks of the 6000L Series oscilloscopes are at a high potential. This presents an electric shock hazard. Protect yourself from electric shock by keeping this area covered or by not coming in contact with the heat sinks when the power cord is attached to the oscilloscope!
f Connect the power cord and check the power supply voltage between pins 1
and 6 of the system board side of the “DC cable”. The voltage should be 15 V ±10%.
If the voltage is correct, then test the system board (see page 67).
If the voltage is not correct, then replace the power supply (see page 79).

To check the 6000L Series system board

1 Remove the cabinet. 2 Check that all cable connections are securely connected from the system board
to:
Power supply
Fan
3 Verify the voltages at the system board test points listed in the table below.
Refer to Figure 11 on page 66 to locate the test points.
Troubleshooting 4
6000L Series Oscilloscopes Service Guide 67
Table 13 System Board Test Points
Test point Voltage Output from
regulator:
L3204 3.3 (+- 0.1) U3202 L3201 5.0 (+- 0.1) U3202 L3301 1.5 (+- 0.1) U3300 L3302 -5.2 (+- 0.1) U3301
4 If the voltage at test point L3301 and/or L3302 is not within the specified
range, replace the system board.
5 If all cables are properly connected and none of the previous tests confirm a
failure on another assembly, replace the system board.
Page 68
4 Troubleshooting

To check the 6000L Series display output

To check the fan

1 Connect a known good external XGA display and power-up the display and the
oscilloscope.
2 If the known good display does show an output from the XGA port of the
oscilloscope, replace the system board.
The fan speed is controlled by a circuit on the system board. 1 If the fan is running, perform the internal self-tests. Go to “To run the internal
self-tests" on page 68.
2 If the fan is not running, it may be defective. Follow these steps:
a Disconnect the fan cable from the system board. b Measure the fan voltage at the connector on the system board.
See the figure below for the location of the fan connector.
c If the fan voltage is approximately +8.5 Vdc at room temperature, replace the
fan. If the fan voltage is not approximately +8.5 Vdc, replace the system board.
The proper voltage range depending on temperature is between +6.0 Vdc to +11.5 Vdc.

To run the internal self-tests

Self Test performs a series of internal procedures to verify that the oscilloscope is operating properly.
It is recommended that you run the Self Test:
• after experiencing abnormal operation
• for additional information to better describe an oscilloscope failure
• to verify proper operation after the oscilloscope has been repaired
Successfully passing Self Test does not guarantee 100% of the oscilloscope’s functionality. Self Test is designed to provide an 80% confidence level that the oscilloscope is operating properly.
1 Press the Utility key, then press the Service softkey. Use the Remote Front Panel
feature if you are testing a 6000L Series model.
68 6000L Series Oscilloscopes Service Guide
Page 69
2 Begin the internal self tests by pressing the Start Self Test softkey.

To verify default setup

The oscilloscope is designed to turn on with the setup from the last turn on or previous setup. However, if the Secure Environment option is installed, the oscilloscope will always execute a Default Setup upon power-up.
To recall the factory default setup:
1 Press the Save/Recall key. 2 Press the Default Setup softkey.
This returns the oscilloscope to its factory default settings and places the oscilloscope in a known operating condition. The major default settings are:
Horizontal - main mode, 100 us/div scale, 0 s delay, center time reference
Vertical - Channel 1 on, 5 V/div scale, dc coupling, 0 V position, probe factor
to 1.0 if an AutoProbe probe is not connected to the channel
Trigger - Edge trigger, Auto sweep mode, 0 V level, channel 1 source, dc
coupling, rising edge slope, 60 ns holdoff time
Display - Vectors on, 20% grid intensity, infinite persistence off
Other - Acquire mode normal, Run/Stop to Run, cursor measurements off
Troubleshooting 4
6000L Series Oscilloscopes Service Guide 69
Page 70
4 Troubleshooting
Figure 12 Default setup screen
3 If your screen looks substantially different, replace the system board.
70 6000L Series Oscilloscopes Service Guide
Page 71
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide
5 Replacing 6000L
Assemblies
To remove the bottom cover / 73 To remove the front panel assembly / 74 To remove the system board / 76 To remove the power supply / 79 To remove the fan / 81 To remove the AC power input assembly / 83 To remove the powe r shaft / 85
This chapter describes how to remove assemblies from the Keysight 6000L Series Oscilloscopes. After you have removed an assembly, to install the replacement assembly, follow the instructions in reverse order.
The parts shown in the following figures are representative and may look different than what you have in your oscilloscope.
Tools Used for Disassembly
Use these tools to remove and replace the oscilloscope assemblies:
• T10 TORX drivers
• 5/8-inch and 9/32-inch socket drivers
See how the Oscilloscope Parts Fit Together
An exploded view of the oscilloscope is included in the Chapter 6, “Replaceable Parts,” starting on page 87. It shows the individual part numbers used in the assemblies, and shows you how all of the assemblies fit together.
71
Page 72
5 Replacing 6000L Assemblies
WARNING
WARNING
CAUTION
CAUTION
Read All Warnings and Cautions
Read the following warnings and cautions before removing and replacing any assemblies in the oscilloscope.
HAZARDOUS VOLTAGES !
Read the safety summary at the back of this book before proceeding. Maintenance is performed with power supplied to the oscilloscope and with the protective covers removed. Only trained service personnel who are aware of the hazards involved should perform the maintenance. Whenever possible, perform the procedures with the power cord removed from the oscilloscope.
AVOID ELECTRICAL SHOCK !
Hazardous voltages exist on the LCD assembly and power supply. To avoid electrical shock:
1 Disconnect the power cord from the oscilloscope. 2 Wait at least three minutes for the capacitors in the oscilloscope to discharge before you begin
disassembly.
Read the Safety Summary at the back of this manual before you begin.
72 6000L Series Oscilloscopes Service Guide
REMOVE POWER TO AVOID DAMAGE !
Remove power before you begin to remove and replace assemblies. Do not remove or replace assemblies while the oscilloscope is turned on, or damage to the components can occur.
AVOID DAMAGE TO ELECTRONIC COMPONENTS !
ELECTROSTATIC DISCHARGE (ESD) can damage electronic components. When doing any of the procedures in this chapter, use proper ESD precautions. As a minimum, you should place the instrument on a properly grounded ESD mat and wear a properly grounded ESD strap.
Page 73

To remove the bottom cover

T-10 Screws
T-10 Screw
T-10 Screw
1 Turn off the oscilloscope and disconnect the power cable. 2 Turn the oscilloscope so the bottom is facing up. 3 Using a T10 screwdriver, remove 5 screws from the rear and sides of the
cabinet.
4 Slide the bottom cover back and up to remove.
Replacing 6000L Assemblies 5

Figure 13 Removing the cover

6000L Series Oscilloscopes Service Guide 73
Page 74
5 Replacing 6000L Assemblies
Intensity Knob

To remove the front panel assembly

1 Perform the following procedures:
“To remove the bottom cover" on page 73
2 Insert a flat-blade screwdriver under the center of the intensity knob and gently
twist the screwdriver as you pull the knob off.
Using a twisting motion on the screwdriver rather than prying prevents marking or damaging the front panel.

Figure 14 Removing the intensity knob

3 Disconnect the front panel ribbon cable from the system board by pressing the
release tabs and then disconnect it from the front panel PC board.
4 Use a flat-blade screwdriver to release the center retainer tab and push it
slightly out of the slot.
5 Use a flat-blade screwdriver to release the 2 side retainer tabs and remove
them from their slots.
74 6000L Series Oscilloscopes Service Guide
Page 75
Replacing 6000L Assemblies 5
Front Panel Ribbon Cable
Retainer Tab
System Board
Front Panel
Retainer Tab Slot
Retainer Tab
Center Retainer Tab
100 MHz Auto Probe Label
500 MHz Auto Probe Label
Front Panel PC Board

Figure 15 Disconnecting ribbon cable and releasing tab retainers

6 Pull the front panel off of the oscilloscope. 7 Lift the front panel PC board to remove it from the front panel. 8 Reverse this procedure to replace the front panel.
6000L Series Oscilloscopes Service Guide 75
Page 76
5 Replacing 6000L Assemblies
T10 Screws
Fan Cable
System Board
Backlight Inverter Cable
Power Supply Cable
Front Panel Cable
Power Supply Cable
Hex Standoffs

To remove the system board

1 Perform the previous procedures:
“To remove the bottom cover" on page 73
2 Remove the intensity knob as described in step 2 on page 74. 3 Disconnect the front panel ribbon cable from the system board by pressing the
release tabs.
4 Disconnect the fan cable. 5 Remove the 3 hex nuts and washers from the rear BNC connectors using a
5/8-inch socket driver.

Figure 16 Preparing to remove the system board

76 6000L Series Oscilloscopes Service Guide
Page 77
Replacing 6000L Assemblies 5
CAUTION
ICs
Thermal Pads
NOTE
6 Using a 9/23-inch hex driver, remove the two hex standoffs and washers from
the GPIB connector
7 Using a T10 Torx driver, remove the 5 screws that hold the system board to the
deck.
8 Remove the system board
a Lift the back of the board to clear the chassis. b Gently pull the board out until you can disconnect the power supply cable. c Continue to remove the board from the chassis.
To avoid tearing the thermal pads on a 500 MHz oscilloscope when removing or installing the system board, hold the board up, away from the thermal pads until all components are clear.

Figure 17 Avoid damage to thermal pads (500 MHz models only)

If the battery on the system board needs to be replaced, use a CR2477/1HF or CR2477/HFN 3V manganese dioxide lithium battery.
6000L Series Oscilloscopes Service Guide 77
Page 78
5 Replacing 6000L Assemblies
9 Reverse this procedure to reinstall the system board:
78 6000L Series Oscilloscopes Service Guide
Page 79

To remove the power supply

Power Input Board Cable
T10 Screws
Power Input Board
Ground Connector
Hole in Divider
AC Connector
Snap Fasteners
Power to System Board
1 Perform the previous procedures:
“To remove the bottom cover" on page 73
“To remove the system board" on page 76
2 Disconnect the ground cable and AC cables from the power supply using a
needle nose pliers if necessary.
3 Press the release latch to disconnect the power input board cable from the
power input board.
Replacing 6000L Assemblies 5

Figure 18 Removing the power supply.

4 Using a T10 Torx driver, remove the 2 screws securing the power supply to the
chassis.
6000L Series Oscilloscopes Service Guide 79
Page 80
5 Replacing 6000L Assemblies
5 Using a flat-blade screwdriver through the hole in the divider, gently pry the
power supply board up enough to snap the power supply board off the rear snap fastener.
6 Using a twisting motion, with a flat-blade screwdriver gently snap the power
supply board off the front snap fastener.
7 Reverse this procedure to install the power supply.
80 6000L Series Oscilloscopes Service Guide
Page 81

To remove the fan

Fan
Flexible Fan Mount
Chassis Divider
1 Perform the previous procedures:
“To remove the bottom cover" on page 73
“To remove the system board" on page 76
“To remove the power supply" on page 79
2 Slide the fan and fan mount slightly up and out of the chassis. 3 Remove the fan from the flexible fan mount.
Replacing 6000L Assemblies 5
Though the fan can be removed without performing all of the above, it is difficult to reinstall it.

Figure 19 Removing the fan.

6000L Series Oscilloscopes Service Guide 81
Page 82
5 Replacing 6000L Assemblies
Pegs Fit in Fan Holes
Holes in Bottom Fit Over Studs
Align Notch and Chassis
Chassis Divider
4 Replace the fan.
a Place the rectangular part of the flexible fan mount over the fan. b Place the pegs in the flexible fan mount into the holes in the top and bottom
of the fan.
c Align the holes in the bottom of the flexible fan mount over the studs on the
chassis as you slide the fan over the divider in the chassis.
d Ensure that the notch in the rectangular portion of the fan mount aligns with
the front of the chassis.
82 6000L Series Oscilloscopes Service Guide

Figure 20 Installing the fan.

Page 83

To remove the AC power input assembly

CAUTION
1 Perform the previous procedures:
“To remove the bottom cover" on page 73
“To remove the power supply" on page 79
“To remove the fan" on page 81
2 Using a TORX T10 screwdriver, remove the screw securing the power
receptacle ground wire to the chassis.
3 Using a TORX T10 screwdriver, remove the screw securing the input board to
the chassis.
4 Lift and rotate the grooved input receptacle just enough to free it from the
opening in the chassis.
5 Slide the assembly toward the front of the chassis to remove the posts from the
keyholes.
6 Lift the whole assembly up and back to free the power shaft from the chassis.
Avoid breaking the power shaft. Use one hand on each of the 2 pieces of the assembly to keep it in alignment as you remove it.
Replacing 6000L Assemblies 5
6000L Series Oscilloscopes Service Guide 83
Page 84
5 Replacing 6000L Assemblies
Grooved Receptacle
Power Shaft
Posts (2) & Keyholes (2)
T10 Screws
AC Input Board
Ground Wire

Figure 21 Removing the AC power input board.

7 Reverse this procedure to install the AC power input board.
84 6000L Series Oscilloscopes Service Guide
Page 85

To remove the power shaft

CAUTION
1 Perform the previous procedures:
“To remove the bottom cover" on page 73
“To remove the power supply" on page 79
“To remove the fan" on page 81
2 Use a flat-blade screwdriver to gently spread the latch while pushing the power
shaft forward.
Twisting the latch too much could cause it to break!
Replacing 6000L Assemblies 5

Figure 22 Removing the power shaft.

3 Lift and remove it from the deck. 4 When reinserting the power shaft, push the shaft into the power switch until the
shaft snaps onto the switch.
6000L Series Oscilloscopes Service Guide 85
Page 86
5 Replacing 6000L Assemblies
86 6000L Series Oscilloscopes Service Guide
Page 87
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide

6 Replaceable Parts

Ordering Replaceable Parts / 88 Replaceable Parts for 6000L Series Oscilloscopes / 90
This chapter describes how to order replaceable assemblies and parts for the Keysight 6000L Series Oscilloscopes.
Diagrams and parts lists are included for assemblies and hardware that you can order.
Before working on the oscilloscope, read the safety summary at the back of this book.
87
Page 88
6 Replaceable Parts

Ordering Replaceable Parts

Listed Parts

To order a part in the parts list, quote the Keysight Technologies part number, indicate the quantity desired, and address the order to the nearest Keysight Technologies Sales Office. To find your nearest sales office go to
www.keysight.com.

Unlisted Parts

To order a part not listed in the parts list, include the instrument part number, instrument serial number, a description of the part (including its function), and the number of parts required. Address the order to the nearest Keysight Technologies Sales Office.

Direct Mail Order System

Within the USA, Keysight Technologies can supply parts through a direct mail order system. There are several advantages to this system:
• Direct ordering and shipping from the Keysight Technologies parts center in California, USA.
• No maximum or minimum on any mail order. (There is a minimum amount for parts ordered through a local Keysight Technologies Sales Office when the orders require billing and invoicing.)
• Prepaid transportation. (There is a small handling charge for each order.)
• No invoices.
In order for Keysight Technologies to provide these advantages, please send a check or money order with each order.
Mail order forms and specific ordering information are available through your local Keysight Technologies Sales Office. Addresses and telephone numbers are located in a separate document shipped with the manuals.
88 6000L Series Oscilloscopes Service Guide
Page 89

Exchange Assemblies

Some parts used in this instrument have been set up for an exchange program. This program allows the customer to exchange a faulty assembly with one that has been repaired, calibrated, and performance-verified by the factory. The cost is significantly less than that of a new part. The exchange parts have a part number in the form XXXXX-695XX.
After receiving the repaired exchange part from Keysight Technologies, a United States customer has 30 days to return the faulty assembly. For orders not originating in the United States, contact the local Keysight Technologies service organization. If the faulty assembly is not returned within 30 days, the customer will be charged an additional amount. The additional amount will be the difference in price between a new assembly and that of an exchange assembly.
Replaceable Parts 6
6000L Series Oscilloscopes Service Guide 89
Page 90
6 Replaceable Parts

Replaceable Parts for 6000L Series Oscilloscopes

6000L Series Oscilloscope Exploded Views

The following exploded views provide a graphical representation of the oscilloscope at the time this manual was released. Not all parts are shown. Your parts may be slightly different thank those shown. These views provide reference designator numbers that map to those used in the replaceable parts list table that follows.
90 6000L Series Oscilloscopes Service Guide
Page 91
MP2
W1
MP5
A3
MP4
A1
A4
MP11
W2
MP12
A2
MP6
MP7
M82
MP10
MP9
MP2
MP3
Replaceable Parts 6
Figure 23 6000L Series Oscilloscope Exploded View
6000L Series Oscilloscopes Service Guide 91
Page 92
6 Replaceable Parts

6000L Series Oscilloscope Replaceable Parts List

The information given for each part consists of the following:
• Reference designation from the exploded view.
• Keysight Technologies part number.
• Total quantity (QTY) in the instrument or on assembly.
• Description of the part.
Table 14 6000L Series Oscilloscope Replaceable Parts
Ref Des Keysight Part
Number
A1 D6104-66422 1 AC line filter/power switch/PC board assembly
A2 D6104-00101 1 Deck assembly
A3 D6104-68501 1 Fan assembly
A4 D6104-60001 1 Rackmount kit
MP1 D6104-66421 1 Autoprobe board
MP2 D6104-00104 1 Cover assembly
MP3 D6104-85401 1 Dielectric insulator panel
MP4 D6104-43901 1 Extender - power switch
MP5 D6104-44701 1 Fan mount
MP6 D6104-42201 1 Front panel bezel
MP7 D6104-94301 1 ID Label - 4 Ch, DSO 1 GHz
MP7 D6104-94302 1 ID Label - 4 Ch, DSO 500 MHz
MP7 D6104-94303 1 ID Label - 4 Ch, DSO 100 MHz
Qty. Description
MP8 54801-47401 1 Knob - 12 mm flint gray
MP9 54684-94314 1 Label - auto probe
92 6000L Series Oscilloscopes Service Guide
Page 93
Table 14 6000L Series Oscilloscope Replaceable Parts (continued)
Replaceable Parts 6
Ref Des Keysight Part
Number
MP10 D6104-94307 1 Label - main 4 Ch, 100 MHz
MP10 D6104-94308 1 Label - main 4 Ch, 500 MHz/1 GHz
MP11 54684-66506 1 PC board assembly - 4 Ch, 1 GHz
MP11 54634-66501 1 PC board assembly - 4 Ch, 100 MHz
MP11 54674-66506 1 PC board assembly - 4 Ch, 500 MHz
MP12 0950-4853 1 Power supply
W1 D6104-61607 1 DC cable
W2 D6104-61601 1 Keyboard cable
Qty. Description
6000L Series Oscilloscopes Service Guide 93
Page 94
6 Replaceable Parts
94 6000L Series Oscilloscopes Service Guide
Page 95
Keysight InfiniVision 6000L Series Oscilloscopes
Service Guide

7 Safety Notices

This apparatus has been designed and tested in accordance with IEC Publication 61010-1, Safety Requirements for Measuring Apparatus, and has been supplied in a safe condition. This is a Safety Class I instrument (provided with terminal for protective earthing). Before applying power, verify that the correct safety precautions are taken (see the following warnings). In addition, note the external markings on the instrument that are described under "Safety Symbols."
95
Page 96
7 Safety Notices

Warnings

Before turning on the instrument, you must connect the protective earth terminal of the instrument to the protective conductor of the (mains) power cord. The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. You must not negate the protective action by using an extension cord (power cable) without a protective conductor (grounding). Grounding one conductor of a two-conductor outlet is not sufficient protection.
Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited fuseholders. To do so could cause a shock or fire hazard.
If you energize this instrument by an auto transformer (for voltage reduction or mains isolation), the common terminal must be connected to the earth terminal of the power source.
Whenever it is likely that the ground protection is impaired, you must make the instrument inoperative and secure it against any unintended operation.
Service instructions are for trained service personnel. To avoid dangerous electric shock, do not perform any service unless qualified to do so. Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
Do not install substitute parts or perform any unauthorized modification to the instrument.
Capacitors inside the instrument may retain a charge even if the instrument is disconnected from its source of supply.
Do not operate the instrument in the presence of flammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
Do not use the instrument in a manner not specified by the manufacturer.
96 6000L Series Oscilloscopes Service Guide
Page 97

To clean the instrument

If the instrument requires cleaning: (1) Remove power from the instrument. (2) Clean the external surfaces of the instrument with a soft cloth dampened with a mixture of mild detergent and water. (3) Make sure that the instrument is completely dry before reconnecting it to a power source.
Safety Notices 7
6000L Series Oscilloscopes Service Guide 97
Page 98
7 Safety Notices
!

Safety Symbols

Instruction manual symbol: the product is marked with this symbol when it is necessary for you to refer to the instruction manual in order to protect against damage to the product and preserve safe conditions.
Hazardous voltage symbol.
Earth terminal symbol: Used to indicate a circuit common connected to grounded chassis.
98 6000L Series Oscilloscopes Service Guide
Page 99

Index

Numerics
6000L, 4
A
AC input board removal, 83 assemblies
exchange, 89 replacing, 71
B
bandwidth
verification, 33
battery on system board, 77
C
cabinet, removing, 73 calibration, 51 cautions, 59, 60, 65, 72, 85 certificate of calibration, 51 channels
problem solving, 57 cleaning the instrument, 97 compensating your probe, 63 connecting for threshold test, 26 contact Keysight, 18
D
digital channel
testing, 24
verification, 25 digital channels, 4 display
problem solving, 56
E
equipment, test, 21
exchange assemblies, 89 exploded view
front panel, 91
F
fan removal, 81 front panel, 4
exploded view, 91 removal, 74
fuse, 65
H
horizontal
accuracy verification, 39
I
internal self-tests, 68
K
Keysight
contact information, 18
L
LED, startup sequence, 60
M
making test connector, 22 measurement accuracy, voltage, 29 measurement category, 15
definitions, 15
N
NIST, 51
O
ordering parts, 88 overvoltage category, 17
P
parts
list, 92
ordering, 88 perform user cal, 51 performance
test record, 46 pollution degree, 17
definitions, 17 power
supply removal, 79
switch removal, 85
test points, 66 power requirements, 14 probes
compensating, 63 problem solving, 55
R
record tests, 46 remove
AC input board, 83
cabinet, 73
fan, 81
front panel, 74
power supply, 79
power switch, 85
system board, 76 replaceable parts list, 92 replacing assemblies, 71 run self-tests, 68
S
safety notice, 49, 59, 60, 65, 67
6000L Series Oscilloscopes Service Guide 99
Page 100
Index
self-alignment, user cal, 51 self-tests, 68 specifications, 18 startup sequence, 60 status, User Cal, 53 status, user cal, 51 system board
removal, 76
T
test
connector, constructing, 22 digital channels, 24 equipment, 21 points, power, 66 record, 46
threshold
accuracy, digital channel, 25 test diagram, 26
trace display
problem solving, 56, 57 transient withstand capability, 16 trigger
sensitivity, 41
sensitivity, external, 43
sensitivity, internal, 42 troubleshooting, 55
display, 68
equipment required, 60
fan, 68
flowchart, 64
power supply, 65
W
warnings, 72 warranted specifications, 18 warranty
status, 4
Web control, 4
U
user calibration, 51
V
verify
bandwidth, 33
digital channel, 25
horizontal accuracy, 39
trigger, 41 voltage measurement accuracy, 29
100 6000L Series Oscilloscopes Service Guide
Loading...