No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Manual Part Number
E5092-90000
Edition
Edition 3, May 2019
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this
product and for other Keysight
products may be downloaded from
the Web. Go to
http://www.keysight.com/go/confor
mity. You can then search by product
number to find the latest Declaration
of Conformity.
perform, display, or disclose
commercial computer software or
commercial computer software
documentation. No additional
government requirements beyond
those set forth in the EULA shall
apply, except to the extent that those
terms, rights, or licenses are
explicitly required from all providers
of commercial computer software
pursuant to the FAR and the DFARS
and are set forth specifically in
writing elsewhere in the EULA.
Keysight shall be under no obligation
to update, revise or otherwise modify
the Software. With respect to any
technical data as defined by FAR
2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the
U.S. government acquires no greater
than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as
applicable in any technical data.
Printed in Malaysia
Published by:
Keysight Technologies International
Japan G.K,
1-3-3 Higashikawasaki-cho
Chuo-ku
Kobe-shi, Hyogo, Japan
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
U.S. Government Rights
The Software is “commercial
computer software,” as defined by
Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202, the
U.S. government acquires
commercial computer software
under the same terms by which the
software is customarily provided to
the public. Accordingly, Keysight
provides the Software to U.S.
government customers under its
standard commercial license, which
is embodied in its End User License
Agreement (EULA), a copy of which
can be found at
http://www.keysight.com/find/sweu
laThe license set forth in the EULA
represents the exclusive authority by
which the U.S. government may use,
modify, distribute, or disclose the
Software. The EULA and the license
set forth therein, does not require or
permit, among other things, that
Keysight: (1) Furnish technical
information related to commercial
computer software or commercial
computer software documentation
that is not customarily provided to
the public; or (2) Relinquish to, or
otherwise provide, the government
rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release,
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
Caution
Keysight E5092A Configurable Multiport Test Set
Installation Guide
Do not exceed the operating input power, voltage, and current level and signal
type appropriate for the instrument being used, refer to your instrument’s
Function Reference.
Electrostatic discharge (ESD) can damage the highly sensitive microcircuits in
your instrument. ESD damage is most likely to occur as the test fixtures are
being connected or disconnected. Protect them from ESD damage by wearing
a grounding strap that provides a high resistance path to ground. Alternatively,
ground yourself to discharge any static charge built-up by touching the outer
shell of any grounded instrument chassis before touching the test port
connectors.
Safety Summary
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific WARNINGS elsewhere in this manual may impair
the protection provided by the equipment. Such noncompliance would also
violate safety standards of design, manufacture, and intended use of the
instrument. Keysight Technologies assumes no liability for the customer’s
failure to comply with these precautions.
The E5092A complies with INSTALLATION CATEGORY II as well as
POLLUTION DEGREE 2 in IEC61010-1. The E5092A is an INDOOR USE
product.
The LEDs in the E5092A are Class 1 in accordance with IEC60825-1,
CLASS 1 LED PRODUCT
This equipment is MEASUREMENT CATEGORY I (CAT I). Do not use for CAT
II, III, or IV.
3
-
Caution
Safety Summary
This equipment is tested with stand-alone condition or with the
combination with the accessories supplied by Keysight Technologies
against the requirement of the standards described in the Declaration of
Conformity. If it is used as a system component, compliance of related
regulations and safety requirements are to be confirmed by the builder of
the system.
•Ground the Instrument
To avoid electric shock, the instrument chassis and cabinet must be
grounded with the supplied power cable’s grounding prong.
•Do NOT Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of inflammable gases or
fumes. Operation of any electrical instrument in such an environment
clearly constitutes a safety hazard.
•Keep Away from Live Circuits
Operators must not remove instrument covers. Component replacement
and internal adjustments must be made by qualified maintenance
personnel. Do not replace components with the power cable connected.
Under certain conditions, dangerous voltage levels may remain even after
the power cable has been disconnected. To avoid injuries, always
disconnect the power and discharge circuits before touching them.
•DO NOT Service or Adjust the Instrument Alone
Do not attempt internal service or adjustment unless another person,
capable of rendering first aid and resuscitation, is present.
•Do NOT Substitute Parts or Modify the Instrument
To avoid the danger of introducing additional hazards, do not install
substitute parts or perform unauthorized modifications to the instrument.
Return the instrument to a Keysight Technologies Sales and Service Office
for service and repair to ensure that safety features are maintained in
operational condition.
•Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous
procedures throughout this manual. Instructions contained in the warnings
must be followed.
Dangerous voltage levels, capable of causing death, are present in this
instrument. Use extreme caution when handling, testing, and adjusting
this instrument.
4 Keysight E5092A Configurable Multiport Test Set
Safety Symbols
Caution
Safety Symbols
General definitions of safety symbols used on the instrument or in manuals are
listed below.
Instruction Manual symbol: the product is marked with this
symbol when it is necessary for the user to refer to the instrument
manual.
Alternating current.
Direct current.
On (Supply).
Off (Supply).
Certification
Keysight Technologies certifies that this product met its published
specifications at the time of shipment from the factory. Keysight Technologies
further certifies that its calibration measurements are traceable to the United
States National Institute of Standards and Technology, to the extent allowed
by the Institution’s calibration facility or by the calibration facilities of other
International Standards Organization members.
Exclusive Remedies
The remedies provided herein are Buyer’s sole and exclusive remedies.
Keysight Technologies shall not be liable for any direct, indirect, special,
incidental, or consequential damages, whether based on contract, tort, or any
other legal theory.
A chassis terminal; a connection to the instrument’s chassis,
which includes all exposed metal structure.
Standby.
Assistance
Product maintenance agreements and other customer assistance agreements
are available for Keysight Technologies products.
Keysight E5092A Configurable Multiport Test Set 5
-
Caution
Safety Symbols
For any assistance, contact your nearest Keysight Technologies Sales and
Service Office. Addresses are provided at the back of this manual.
This chapter introduces the documentation for Keysight E5092A and the
overview of this manual to use this manual efficiently.
Manuals for E5092A
Keysight provides the following three manuals for E5092A.
o Installation Guide
The installaton guide (this manual) provides start up setup information when
you use the E5092A for the first time and maintenance information.
o User Guide
The user guide (Online Help) provides information about connecting the
E5071C ENA Series Network Analyzer and the E5092A configurable multiport
test set, setting, measurement operation and controlling the configurable
multiport test set. For more information, refer to E5071C ENA Network
Analyzer Online Help.
o Service Guide
The service manual provides information about the parts, troubleshooting,
performance test, adjustment and service menu.
9
1-
Introduction
Contents of This Manual
Contents of This Manual
The following shows the contents of this manual.
o Chapter 1, “Introduction.”
This chapter introduces the documentation for Keysight E5092A and the
overview of this manual to use this manual efficiently.
o Chapter 2, “Installation.”
This chapter describes the setup of E5092A.
o Chapter 3, “Supplementary Information.”
This chapter describes information related to E5092A.
o Chapter 4, “Information on Maintenance.”
This chapter explains the measures you should take to maintain the Keysight
E5092A.
o Appendix A , “Manual Changes,”
This appendix contains the information required to adapt this manual to
versions or configurations of the E5092A manufactured earlier than the current
printing date of this manual.
10 Keysight E5092A Configurable Multiport Test Set
Keysight E5092A Configurable Multiport Test Set
Installation Guide
2Installation
This chapter describes the setup of E5092A.
Incoming inspection
When you receive the product, follow the below steps to perform the incoming
inspection.
When unpacking the configurable multiport test set, if the external face of
the device (such as, the cover, front/rear panel, power switch, and port
connectors) appear to be damaged during transport, do not turn on the
power. Otherwise, you may get an electrical shock.
Step 1. Check that the packing box or shock-absorbing material used to package the
E5092A has not been damaged.
If the packing box or shock-absorbing material has been damaged, leave
the packing box or shock-absorbing material as is until other inspections
items are checked as follows.
Step 2. Check the packaged items supplied with the configurable multiport test set
for any damage or defects.
Step 3. By referring to the box contents list that is separately appended, check that
all packaged items supplied with the configurable multiport test set have
been provided as per the specified options.
Step 4. After checking, if one of the following applies, contact your nearest Keysight
Technologies sales and service office.
1. If the packing box or shock-absorbing material is damaged or the
shock-absorbing material has traces where extreme force has been applied.
2. A packaged item supplied with the configurable multiport test set has
mechanical damage or defects.
11
2-
Installation
Incoming inspection
3. A packaged item supplied with the configurable multiport test set is
missing.
4. A fault has been detected in the subsequent operation check of the
configurable multiport test set.
If an abnormality is detected in step 1, contact the company that transported
the configurable multiport test set as well as your nearest Keysight
Technologies sales and service office. For inspection by the transport
company, save the packing box, shock-absorbing material, and packaged
items as you received them.
Figure 2-1 Front Panel
Figure 2-2 Rear Panel
12 Keysight E5092A Configurable Multiport Test Set
Installation
Requirements for installation environment
Requirements for installation environment
Install the E5092A under an environment that satisfies the following
conditions.
Operating environment
Use the E5092A under the following environment.
Temperature5 ºC to 40 ºC
Temperature range during
calibration
Humidity20-80% at wet bulb temperature < +29 ºC (no
Altitude0 m to 2,000 m (0 feet to 6,561 feet)
Vibration0.21 G max, 5 Hz to 500 Hz
The above operating environment does not mean the conditions under
which all the specifications and measurement accuracy of the E5092A are
satisfied but those under which the E5092A operates normally.
23 ºC 5 ºC (90 minutes after the instrument has
turned ON)
condensation)
Keysight E5092A Configurable Multiport Test Set 13
2-
Installation
Requirements for installation environment
Heat dissipation clearance
To satisfy the safety, specifications, and measurement accuracy of the product,
you need to keep the temperature around the product within a specific range
by providing proper heat dissipation clearance around the product or by
air-cooling the inside of the rack forcefully when you mount the product in the
rack. For information on ambient temperature to satisfy the specifications and
measurement accuracy of the product, refer to the “Specifications” Chapter in
E5071C ENA Network Analyzers Online Help.
When the temperature around the product is kept within the temperature
range in the operating environmental specifications (see “Operating
environment” on page 13), the product complies with the safety standards.
Furthermore, when the product is installed providing the following heat
dissipation clearance under the above temperature environment, the product
complies with the safety standards.
Condition
Rear 180 mm
Sides 0 mm (each for right and left)
Figure 2-3 Providing heat dissipation clearance
14 Keysight E5092A Configurable Multiport Test Set
Installation
Requirements for installation environment
Antistatic measures
To protect electronic components from static damage, take antistatic measures
as shown in Figure 2-4.
Figure 2-4 Example of antistatic measures
Providing space to disconnect the plug of the power cable
As described in “Cutting supply of power” on page 2-21, the disconnecting
device (the device to cut the supply of power) of the E5092A is the plug of the
power cable. When installing the E5092A, provide sufficient clearance not to
hinder the operation of disconnecting the plug of the power cable (on the
power outlet side or the E5092A side) in an emergency to cut the supply of
power.
Keysight E5092A Configurable Multiport Test Set 15
2-
Installation
Mounting front handles/flanges for rackmount
Mounting front handles/flanges for rackmount
You can use the E5092A on a workbench or mounted in a rack. This section
describes how to mount the front handles used to move/transport it (Option
1CN) and how to mount it into a rack as part of a measurement system (Option
1CM: without handles/Option 1CP: with handles).
Table 2-1 Handle/rack options for the E5092A
OptionName Keysight part number
1CMRackmount kit5063-9214
1CNHandle kit5063-9227
1CPRackmount/handle kit5063-9221
Table 2-2 Contents of each option kit
OptionContentsQuantity
1CNFront handle2
Screw4
Trim strip2
1CMFlange for rackmount (side plate for securing)2
Screw4
1CPFlange for rackmount (side plate for securing)2
Front handle2
Screw4
16 Keysight E5092A Configurable Multiport Test Set
Installation
Mounting front handles/flanges for rackmount
Figure 2-5 Mounting the handle/rackmount kit
Mounting handle kit (Option 1CN)
The handle kit is a kit that contains a pair of front handles used to
move/transport the E5092A. Follow these steps to mount it, referring to the
Figure 2-5.
Step 1. Remove the trim strips (1) stuck to both the sides of the front panel (stuck to
the frame).
Step 2. Mount the front handles (3) to both the sides of the front panel of the E5092A
with the attached screws.
Step 3. Attach the trim strips for the handle (4) so that they cover the screws
securing the front handles.
If a front handle you mounted is damaged, replace it with new one
immediately. If you use a damaged front handle to move/transport the
instrument, the handle may break, which may injure the operator or break
the instrument.
Keysight E5092A Configurable Multiport Test Set 17
2-
Installation
Mounting front handles/flanges for rackmount
Mounting rackmount kit (Option 1CM)
The rackmount kit is a kit that contains 2 flanges (side plates for securing) to
mount the E5092A into the standard rack (width: 482.6 mm) of the EIA
standard. Follow these steps to mount it, referring to the Figure 2-5.
Step 1. Remove the trim strips (1) stuck to both the sides of the front panel (stuck to
the frame).
Step 2. Mount the flanges for rackmount (2) to both the sides of the front panel of the
E5092A with the attached screws.
Step 3. Remove the four legs on the bottom of the E5092A (pull up the part with the
(TAB indication and slide it to the arrow direction).
Step 4. Mount the E5092A into the rack.
Mounting rackmount/handle kit (Option 1CP)
The rackmount/handle kit is a kit that contains both the flanges for rackmount
and the front handles for the E5092A. Follow these steps to mount it, referring
to the Figure 2-5.
Step 1. Remove the trim strips (1) stuck to both the sides of the front panel (stuck to
the frame).
Step 2. Mount the front handles (3) and the flanges for rackmount (5) to both the
sides of the front panel of the E5092A with the attached screws.
Step 3. Remove the 4 legs on the bottom of the E5092A (pull up the part with the
(TAB indication and slide it to the arrow direction).
Step 4. Mount the E5092A into the rack.
18 Keysight E5092A Configurable Multiport Test Set
Installation
Power supply and fuse
Power supply and fuse
Before turn ON the E5092A, check the following.
Checking the supply of power
Check that the power supplied to the E5092A satisfies the following
conditions.
Condition
Voltage
90 to 132 Vac or 198 to 264 Vac
Frequency47 to 63 Hz
Power consumptionMax. 150 VA
1. The E5092A automatically switches between them depending on the voltage.
Setting up fuse
Use the fuse that meets the following specifications.
UL/CSA type, Slo-Blo, 5¥20 mm miniature fuse, 3A 250V (Keysight part number:
2110-1017)
Spare fuses are available from Keysight Technologies sales office. To check or
replace the fuse, disconnect the power cable and pull out the fuse holder.
Figure 2-6 Fuse holder and power cable inlet
1
Keysight E5092A Configurable Multiport Test Set 19
2-
Installation
Power supply and fuse
Checking and connecting power cable
The power cable that comes with the E5092A has 3-wire structure and one of
them is the grounding wire. This power cable grounds the E5092A through an
outlet, protecting the operator from shock hazards.
Step 1. Check that the power cable you use for damage.
Never use a power cable that may be damaged. You may get an electrical
shock.
Step 2. Connect the attached power cable between the power cable receptacle on
the rear panel of the E5092A and a 3-pole outlet whose grounding terminal is
grounded securely.
Securely ground the E5092A with the attached 3-wire power cable with
the grounding wire.
Power cord information is available on
https://www.keysight.com/find/powercords.
20 Keysight E5092A Configurable Multiport Test Set
Installation
Starting up E5092A
Starting up E5092A
This section describes how to turn ON/OFF the E5092A and how to cut the
supply of power in an emergency.
Turning ON/OFF the power
Turning ON power
Step 1. Check that the line switch in the lower left part of the front panel is pulled up
(). If it is pushed in () which means the power is OFF, push the
standby switch to pull it up ().
Step 2. Push the line switch to push it in ().
Turning OFF power
To turn OFF the E5092A, do the following.
•Push the line switch in the lower left part of the front panel so that the
switch pushed in () is pulled up ().
To turn OFF the E5092A in normal times, press the line switch. In normal
times, never disconnect the power cable to cut the supply of power to the
power cable receptacle on the rear panel.
Cutting supply of power
For the E5092A, the disconnecting device (the device to cut the supply of
power) is the plug of the power cable (on the power outlet side or the E5092A
side).
If you need to cut the supply of power to avoid danger of electric shocks,
disconnect the plug of the power cable (on the power outlet side or the
E5092A side).
Follow the description in “Providing space to disconnect the plug of the
power cable” on page 15 so that you can perform this operation reliably.
To turn OFF the power in normal times, be sure to follow the procedure in
“Turning OFF power”.
Keysight E5092A Configurable Multiport Test Set 21
2-
Installation
Starting up E5092A
22 Keysight E5092A Configurable Multiport Test Set
Keysight E5092A Configurable Multiport Test Set
Installation Guide
3Supplementary Information
This chapter describes information related to E5092A.
Product Overview
For more information on the overview and setting of E5092A Configurable
multiport test set, refer to "Controlling the Multiport Test Set" in E5071C ENA
Network Analyzer Online Help.
Specifications and Supplemental Information
For more information on Specifications and Supplemental of E5092A
Configurable multiport test set, refer to E5071C Data Sheet in E5071C ENA
Network Analyzer Online Help.
23
3-
Supplementary Information
Specifications and Supplemental Information
24 Keysight E5092A Configurable Multiport Test Set
Keysight E5092A Configurable Multiport Test Set
Installation Guide
4Information on Maintenance
This chapter explains the measures you should take to maintain the Keysight
E5092A.
Cleaning this Instrument
This section describes how to clean the instrument.
To protect yourself from electrical shock, be sure to unplug the power
cable from the outlet before cleaning the instrument.
Never clean the internal components of the instrument.
Maintenance of Test Ports
Test ports on the front panel of the E5092A are fitted with 3.5mm SMA type
connectors (f). Stains or other damage to these connectors would significantly
affect the accuracy in measurements in the RF range. Always pay attention to
the following precautions.
•Always keep the connectors free from stains and dust.
•Do not touch the contact surface on the connectors.
•Do not plug damaged or scratched connectors into the test ports.
•Use compressed air for cleaning connectors. Do not use abrasives under
any circumstance.
Cleaning a Display Other than Test Ports
To remove stains on parts other than test ports of the instrument, wipe them
gently with a soft cloth that is dry or wetted with a small amount of water and
wrung tightly.
25
4-
Information on Maintenance
Cautions Applicable to Requesting Repair, Replacement, Regular Calibration, etc.
Cautions Applicable to Requesting Repair, Replacement, Regular
Calibration, etc.
Devices to be Sent Back for Repair or Regular Calibration
If it is necessary to send the unit to the Service Center of Keysight
Technologies for repair or regular calibration, please follow the instructions
below.
Equipment to be Sent
When requesting repair or regular calibration of the unit by our Service Center,
send only the E5092A main unit without any installed option you may have
ordered. Unless specifically instructed, it is not necessary to send accessories
and calibration kits.
Packing
Use the original package and shock absorbers, or equivalent antistatic packing
materials, when sending the unit.
Shipping Address
For the location of the nearest Keysight Technologies Service Center, contact
the Customer Contact listed at the end of this brochure.
Recommended Calibration Period
The recommended calibration period for this instrument is one year. The user is
recommended to request the Company’s Service Center to perform regular
calibration every year.
26 Keysight E5092A Configurable Multiport Test Set
Keysight E5092A
Microwave Downconverter
Installation Guide
A Manual Changes
This appendix contains the information required to adapt this manual to
versions or configurations of the E5092A manufactured earlier than the current
printing date of this manual.
Manual Changes
To adapt this manual to your E5092A, refer to Table A-1.
Table A-1 Manual Changes by Serial Number
Serial Prefix or NumberMake Manual Changes
Keysight Technologies uses a two-part, ten-character serial number that is
stamped on the serial number plate. The first five characters are the serial
prefix and the last five digits are the suffix.
27
-
Manual Changes
Manual Changes
28 Keysight E5092A Installation Guide
Index
Index
A
assistance
5
C
Cleaning
25
Connectors
Maintenance
25
E
Exclusive Remedies
5
F
flange for rackmount
mounting
16
front handle
mounting
17
H
handle kit
mounting
18
I
ID
setting
23
inspection
incoming inspection
Interconnection Ports
Maintenance
25
11
R
rackmount kit
mounting
18
rackmount/handle kit
mounting
18
S
safety
symbols
serial number plate
Setting the Instrument
5
27
23
space to disconnect the plug
starting up E5091A
21
T
Test Ports
Maintenance
25
15
M
manual
contents
manual changes
9, 10
27
O
operating environment
13
P
power supply
checking cable
checking the supply
connecting cable
cutting power
turning on/off
20
19
202121
Q
Quick Start Guide
contents
Keysight E5092A Installation Guide
9, 10
29
Index
30
Keysight E5092A Installation Guide
This information is subject to change without notice.