Keysight Technologies E5080B Installation Manual

Page 1
Keysight E5080B ENA Series RF Network Analyzers

Installation Guide

Page 2
Notices
© Keysight Technologies 2019
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Trademark Acknowledgments
Manual Part Number
E5080-90010
Edition
Edition 1, March 2019
DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight products may be downloaded from the Web. Go to
http://www.keysight.com/go/confor mity. You can then search by product
number to find the latest Declaration of Conformity.
commercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR
2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Printed in Malaysia
Published by:
Keysight Technologies International Japan G.K,
1-3-3 Higashikawasaki-cho Chuo-ku Kobe-shi, Hyogo, Japan
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
http://www.keysight.com/find/sweul a The license set forth in the EULA
represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
This NOTE sign denotes important information. It calls attention to a procedure, practice, or condition that is essential for the user to understand.
Page 3

Caution

Keysight E5080B ENA Series RF Network Analyzers
Installation Guide
Do not exceed the operating input power, voltage, and current level and signal type appropriate for the instrument being used, refer to your instrument’s Function Reference.
Electrostatic discharge (ESD) can damage the highly sensitive microcircuits in your instrument. ESD damage is most likely to occur as the test fixtures are being connected or disconnected. Protect them from ESD damage by wearing a grounding strap that provides a high resistance path to ground. Alternatively, ground yourself to discharge any static charge built-up by touching the outer shell of any grounded instrument chassis before touching the test port connectors.

Safety Summary

When you notice any of the unusual conditions listed below, immediately terminate operation and disconnect the power cable.
Contact your local Keysight Technologies sales representative or authorized service company for repair of the instrument. If you continue to operate without repairing the instrument, there is a potential fire or shock hazard to the operator.
Instrument operates abnormally.
Instrument emits abnormal noise, smell, smoke or a spark-like light during
Instrument generates high temperature or electrical shock during
Power cable, plug, or receptacle on instrument is damaged.
Foreign substance or liquid has fallen into the instrument.
operation.
operation.
3
Page 4
-
Caution Manufacturer’s Declaration

Manufacturer’s Declaration

Herstellerbescheinigung
GERA- USCHEMISSION
LpA < 70 dB
am Arbeitsplatz
normaler Betrieb
nach DIN 45635 T. 19
Manufacturer’s Declaration
ACOUSTIC NOISE EMISSION
LpA < 70 dB
operator position
normal operation
per ISO 7779

Regulatory Compliance Information

This product complies with the essential requirements of the following applicable European Directives, and carries the CE marking accordingly:
The Low Voltage Directive 2014/35/EU
The EMC Directive 2014/30/EU
To obtain Declaration of Conformity, please contact your local Keysight Technologies sales office, agent or distributor.
4 Keysight E5080B ENA Series RF Network Analyzers
Page 5
Caution Safety Notice Supplement

Safety Notice Supplement

This equipment complies with EN/IEC61010-1:2010.
This equipment is of MEASUREMENT CATEGORY I (CAT I). Do not use for CAT II, III, or IV.
Do not connect the measuring terminals to mains.
This equipment is a POLLUTION DEGREE 2, INDOOR USE product.
This equipment is tested in stand-alone condition and in combination with the accessories supplied by Keysight Technologies against the requirement of the standards described in the Declaration of Conformity. If it is used as a system component, compliance of related regulations and safety requirements are to be confirmed by the builder of the system.
Do not connect the measuring terminals to mains.
Keysight E5080B ENA Series RF Network Analyzers 5
Page 6
-
Caution General Safety Precautions

General Safety Precautions

The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific WARNINGS elsewhere in this manual may impair the protection provided by the equipment. Such noncompliance would also violate safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these precautions.
The E5080B complies with INSTALLATION CATEGORY II as well as POLLUTION DEGREE 2 in IEC61010-1. The E5080B is an INDOOR USE product.
The LEDs in the E5080B are Class 1 in accordance with IEC60825-1, CLASS 1 LED PRODUCT.
Ground the Instrument
To avoid electric shock, the instrument chassis and cabinet must be grounded with the supplied power cable’s grounding prong.
Do NOT Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of inflammable gases or fumes. Operation of any electrical instrument in such an environment clearly constitutes a safety hazard.
Keep Away from Live Circuits
Operators must not remove instrument covers. Component replacement and internal adjustments must be made by qualified maintenance personnel. Do not replace components with the power cable connected. Under certain conditions, dangerous voltage levels may remain even after the power cable has been disconnected. To avoid injuries, always disconnect the power and discharge circuits before touching them.
DO NOT Service or Adjust the Instrument Alone
Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
Do NOT Substitute Parts or Modify the Instrument
To avoid the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument. Return the instrument to a Keysight Technologies Sales and Service Office for service and repair to ensure that safety features are maintained in operational condition.
6 Keysight E5080B ENA Series RF Network Analyzers
Page 7

Safety Symbols

Caution Safety Symbols
Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed.
Dangerous voltage levels, capable of causing death, are present in this instrument. Use extreme caution when handling, testing, and adjusting this instrument.
General definitions of safety symbols used on the instrument or in manuals are listed below.
Instruction Manual symbol: the product is marked with this symbol when it is necessary for the user to refer to the instrument manual.

Certification

Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institution’s calibration facility or by the calibration facilities of other International Standards Organization members.
Alternating current.
Direct current.
On (Supply).
Off (Supply).
A chassis terminal; a connection to the instrument’s chassis, which includes all exposed metal structure.
Standby.
Keysight E5080B ENA Series RF Network Analyzers 7
Page 8
-
Caution Safety Symbols

Documentation Warranty

The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or any information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement will control.

Exclusive Remedies

The remedies provided herein are Buyer’s sole and exclusive remedies. Keysight Technologies shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory.

Assistance

Product maintenance agreements and other customer assistance agreements are available for Keysight Technologies products.
For any assistance, contact your nearest Keysight Technologies Sales and Service Office. Addresses are provided at the back of this manual.
8 Keysight E5080B ENA Series RF Network Analyzers
Page 9
Caution Manuals for E5080B

Manuals for E5080B

Keysight provides the following three manuals for E5080B. The latest version of all documentations can be downloaded from
http://www.keysight.com/manual/e5080b.

Installation Guide

The installation guide (this manual) provides start up setup information when you use the E5080B for the first time and troubleshooting information when the Windows cannot be boot up. See this manual first when you use the E5080B for the first time.

Online Help

The online help provides the information about the quick start, measurement operation, programming, built-in VBA, I/O interface This is pre-installed in the E5080B. Press [Help] hard key on the front panel to open. Quick Start helps in understanding the E5080B operation quickly.
The latest version of online help is available at:
http://www.keysight.com/manuals/e5080b
The online help has context sensitive help, which is a great feature of the E5080B help. It allows you to get information about the selected softkey by pressing the Help key in the E5080B or by pressing F1 in a keyboard attached to the E5080B or by clicking the help button in a dialog box. With context sensitive help, users can receive information quickly about the area the user is working in the firmware of the E5080B. It provides information relevant to the task that needs to be accomplished and reduces the time to search relevant information required to complete a task.

Service Guide

The service manual provides information about the parts, troubleshooting, performance test, adjustment and service menu.
Keysight E5080B ENA Series RF Network Analyzers 9
Page 10
-
Caution Manuals for E5080B
10 Keysight E5080B ENA Series RF Network Analyzers
Page 11
Table of Contents
Caution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Manufacturer’s Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Regulatory Compliance Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Safety Notice Supplement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
General Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Documentation Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Exclusive Remedies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Assistance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Manuals for E5080B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Installation Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Online Help. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Service Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Contents
1. Installation
Checking the Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Operating Environments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Ventilation Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Protection Against Electrostatic Discharge (ESD). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of Power
Cable in Case of Emergency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Verification of the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Verification and Connection of Power Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Blown Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Connecting the BNC Adaptor (Option 1E5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Starting the E5080B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Turning the Power ON and OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Disconnection from Supply Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Initial Registration of E5080B. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2. Troubleshooting
Troubleshooting during Startup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
System Recovery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Notes on executing factory recovery function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Keysight E5080B Installation Guide 11
Page 12
Contents
Procedure to execute the factory recovery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
12 Keysight E5080B Installation Guide
Page 13
Keysight E5080B ENA Series RF Network Analyzers
Installation Guide

1 Installation

This chapter provides information about how to set up the Keysight E5080B.

Checking the Shipment

After you receive the analyzer, carry out checks during unpacking according to the following procedure.
When unpacking the analyzer, if the exterior of the analyzer (such as the cover, front/rear panel, LCD screen, power switch, and port connectors) appear to be damaged during transport, do not turn on the power switch. Otherwise, you may get an electric shock.
Step 1. Check that the packing box or shock-absorbing material used to package the
analyzer has not been damaged.
If the packing box or shock-absorbing material has been damaged, leave the packing box and shock-absorbing material as is until other inspection items are checked as follows.
Step 2. Check the packaged items supplied with the analyzer for any damage or
defects.
Step 3. By referring to the furnished contents list, check that all packaged items
supplied with the analyzer have been received as per the specified options.
Step 4. After checking, if one of the following applies, contact your nearest Keysight
Technologies sales and service office.
1. The packing box or shock-absorbing material used to package the analyzer has been damaged or the shock-absorbing material has traces where extreme force has been applied.
2. A packaged item supplied with the analyzer has mechanical damage or defects.
13
Page 14
1-
Installation Environmental Requirements
3. A packaged item supplied with the analyzer is missing.
4. A fault has been detected in the subsequent operation check of the analyzer.
If an abnormality is detected in step 1, contact the company that transported the analyzer as well as your nearest Keysight Technologies sales and service office.

Environmental Requirements

Set up the E5080B where the following environmental requirements are met.

Operating Environments

Ensure that the operating environment meets the following requirements.
Temperature 0˚C to 40˚C
Temperature range at the error-correction
Humidity 20% to 80% at wet bulb temperature < +29˚C
Altitude 0 to 2,000 m (0 to 6,561 feet)
Vibration 0.21 G maximum, 5 Hz to 500 Hz
Above environmental requirements are NOT for the specifications and measurement accuracy of the analyzer, but for the operating environment of the analyzer.

Ventilation Requirements

To ensure safety requirements, the specifications and measurement accuracy of the analyzer, you must keep environmental temperature within the specified range by providing appropriate cooling clearance around the analyzer or, for the rack-mount type, by forcefully air-cooling inside the rack housing. For more information on environmental temperature to satisfy the specifications and measurement accuracy of the analyzer, see the specification in the E5080B Online Help.
23˚C 3˚C (< 1˚C deviation from the temperature when performing the error-correction)
(non-condensation)
When the environmental temperature around the analyzer is kept within the temperature range of the operating environment specification (See Section
“Operating Environments” on page 14), the analyzer conforms to the
14 Keysight E5080B ENA Series RF Network Analyzers
Page 15
Installation Environmental Requirements
requirements of the safety standard. Furthermore, under that temperature requirement, the analyzer still conforms to the requirements of the safety standard even when placing the analyzer with cooling clearance as follows:
Requirements
Back 180 mm
Sides 60 mm (both right and left)

Protection Against Electrostatic Discharge (ESD)

Set up a static-safe work-station to protect the electronic components from the damage by the electrostatic discharge (ESD) as shown in Figure 1-1.
Figure 1-1 Example of the static-safe work station

Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of Power Cable in Case of Emergency

As described in “Disconnection from Supply Source” on page 19, the power supply is disconnected by removing the power cable’s connector plug from either the AC outlet or the E5080B unit. When installing the E5080B, ensure that there is sufficient free space around the unit to permit quick disconnection of the plug (from AC outlet or E5080B unit) in case of emergency.
Keysight E5080B ENA Series RF Network Analyzers 15
Page 16
1-
Installation Power Supply

Power Supply

Before turning on the E5080B power, check the following.

Verification of the Power Supply

Confirm that the power supplied to the E5080B meets the following requirements:
Requirements
Voltage 100 to 240 VAC
Frequency 50/60 Hz
Maximum power consumption 350 W

Verification and Connection of Power Cable

The three-wire power cable attached to the E5080B has one wire serving as a ground. Using this power cable allows the E5080B to be grounded, thereby protecting you against electrical shock from the power outlet.
Step 1. Confirm that the power cable is not damaged.
NEVER use a power cable showing any sign of damage. Faulty cables can cause electrical shock.
Step 2. Use the supplied cable to connect between the power cable receptacle
(Figure 1-2 on page 19) on the rear panel of the E5080B and a three-wire power outlet with the grounding prong firmly connected in the ground slot.
Use the supplied power cable with grounding wire to securely ground the E5080B.
For power cord list and options, refer to www.keysight.com/find/powercords.
16 Keysight E5080B ENA Series RF Network Analyzers
Page 17
Installation Blown Fuses

Blown Fuses

If the fuse appears to have blown during operation, this instrument may be subject to failure and must be repaired. For any assistance, contact Keysight Technologies Customer contact center listed at the end of this guide.
The E5080B uses the following fuse types:
UL/CSA Type, Slow-Blo, 8 A-250 Vac.
DO NOT replace the fuse yoursel f; doing this may expose you to electrical shock.

Connecting the BNC Adaptor (Option 1E5)

When the E5080B is equipped with Option 1E5 (high-stability frequency standard), connect the BNC adapter between the 10 MHz REF OVEN and the 10 MHz REF IN connectors on the rear panel. The BNC adapter is included in Option 1E5. When Option 1E5 is installed, the frequency accuracy and stability of the E5080B will further improve.
Keysight E5080B ENA Series RF Network Analyzers 17
Page 18
1-
Installation Starting the E5080B

Starting the E5080B

This section describes how to turn ON/OFF the E5080B power and how to cut off the power supply in the case of emergency.

Turning the Power ON and OFF

The standby switch can turn ON/OFF the E5080B. The color on the button shows the status, as indicated below:
Indicator Color Description
Green Normal power on status
Orange Standby status
Orange + Green Illegal power on status
Perform the following steps to turn the power ON or OFF.
Turning the Power ON
Step 1. Press the standby switch once. It changes to green color.
This operation turns ON the power, and the E5080B starts the self-test.
Step 2. Confirm that the self-test indicates normal operation.
Normal operation is confirmed by the self-test if no error message appears.
Turning the Power OFF
Step 1. To turn OFF the power of the E5080B, first, press this standby switch or send
a shutdown command from the external controller to activate the shutdown process (the process of software and hardware necessary to turn OFF the power supply). This puts the E5080B into the standby state and the button changes to orange color.
Under normal use, never directly interrupt the power supply to the power cable receptacle on the rear panel when the power supply is turned ON.
If you directly interrupt the power supply to the power cable receptacle when the power supply is turned ON, the shutdown process will not work. This could damage the software and hard ware of the E5080B and lead to device failure.
Turning ON the power supply after a faulty shutdown may cause the system to start up in a condition called "safe mode." If this occurs, first shut down the system to set it to the standby state and then turn ON the power supply again to start up the system in normal mode.
18 Keysight E5080B ENA Series RF Network Analyzers
Page 19
Installation Starting the E5080B
Figure 1-2 Power cable receptacle

Disconnection from Supply Source

The power supply of the E5080B is cut off by disconnecting the plug of the power cable (on either AC outlet side or E5080B side). When it is necessary to disconnect the power supply in order to avoid shock hazards, etc., pull out the power cable plug from either the AC outlet side or the E5080B side.
To allow this operation to be performed smoothly, be sure to follow the guidelines in “Ensuring Adequate Free Space around Analyzer for
Immediate Disconnection of Power Cable in Case of Emergency” on page 15.
When turning the power OFF under normal circumstances, always follow the methods described in “Turning the Power OFF” on page 18.
Keysight E5080B ENA Series RF Network Analyzers 19
Page 20
1-
Installation Initial Registration of E5080B

Initial Registration of E5080B

When you start up the E5080B for the first time, you need to perform the initial registration of the Windows operating system of the E5080B.
When the Windows agreement message box appears, click Agree. The rest of the procedure runs automatically.
The Touch Screen does not require any calibration and it works from the very first time you use the E5080B.
20 Keysight E5080B ENA Series RF Network Analyzers
Page 21
Keysight E5080B ENA Series RF Network Analyzers
Installation Guide

2 Troubleshooting

This chapter describes the troubleshooting during start up and the procedure of the operating system (OS) recovery when the Windows OS has been damaged.

Troubleshooting during Startup

When you encounter the problem during start up, see Table 2-1. System recovery saves most of problems.
Table 2-1 Troubleshooting during startup
Symptom Solution
Turning on the standby switch does not start up the system.
The system starts up, but it automatically shuts down immediately.
The system starts up, but it enters the service mode (The instrument status bar in the lower right part of the screen displays SVC in red).
The measurement screen appears after startup, but the power-on test is failed, with Error Message 241 appearing against a red background in the instrument message/warning area in the lower left part of the screen.
Confirm that the power cable is properly plugged in.
Execute the system recovery.
21
Page 22
2-
Troubleshooting System Recovery

System Recovery

By executing system recovery, you can return the system of the E5080B (the Windows operating system and the firmware) to the factory state (at the time
1
of purchase
Strictly follow the steps described below. If you do any operation other than the following steps, the system may not be recovered.

Notes on executing factory recovery function

Executing the factory recovery function causes the following conditions:
o In addition to the Windows operating system and the firmware, the
following settings of the E5080B are returned to the factory state.
•Network setting
•GPIB setting
•Printer setting
).
o The driver for the supported printer installed after purchase is deleted.
o You need to execute initial registration again.
Files that you have created using the save function (files in the D drive) are not affected, but we recommend backing them up before executing system recovery for precautionary purposes. For more information on backup, refer to “Backing Up the Data” as described in E5080B Online Help.
1. The default setting of the storage is recovered during the system recovery. For the storage that had failed earlier and had been replaced (after purchase), its default setting may not be the same as the factory state.
22 Keysight E5080B ENA Series RF Network Analyzers
Page 23
Troubleshooting System Recovery

Procedure to execute the factory recovery

This section describes how to return the contents of the C drive to the factory state.
You need a keyboard for this operation.
Step 1. Shut down the E5080B.
Step 2. Disconnect all of the USB device from the USB ports.
Step 3. Connect the keyboard to the E5080B.
Step 4. Press the standby switch of the E5080B to turn it ON.
Step 5. When the screen as shown in the figure below appears, select Instrument
Image Recovery System and press [Enter].
Figure 2-1 Instrument Image Recovery System Selection Screen
After several seconds, the next screen appears automatically even if you do not press any key.
Keysight E5080B ENA Series RF Network Analyzers 23
Page 24
2-
Troubleshooting System Recovery
Step 6. Software End User License Agreement appears. Press [Agree].
Figure 2-2 Software End User License Agreement
Step 7. Type 2 to select Recover the original factory system image option and press
[Enter].
Figure 2-3 System recovery selection screen
24 Keysight E5080B ENA Series RF Network Analyzers
Page 25
Troubleshooting System Recovery
Step 8. A confirmation message appears. Click OK to continue the process.
Figure 2-4 System recovery confirmation screen
Step 9. The progress of the system configuration is displayed on the screen. The
recovery takes a few minutes depending on the amount of data.
Figure 2-5 System configuration progress screen
Step 10. Once the recovery process is completed, completion message prompts. Click
OK. The E5080B restarts automatically.
Step 11. Software End User License Agreement appears. Press [Agree]. Refer to
Figure 2-2.
Step 12. After the restart, execute initial registration. For information on the execution
procedure, refer to “Initial Registration of E5080B” on page 20
Never turn OFF the power during the system recovery because doing so may cause serious damage to the E5080B.
Keysight E5080B ENA Series RF Network Analyzers 25
Page 26
2-
Troubleshooting System Recovery
26 Keysight E5080B ENA Series RF Network Analyzers
Page 27
Index
A
accessories assistance
5
8
C
cable
power cable options
16
D
documentation warranty
8
E
environmental requirements
operation environment ventilation space
14
14
error-correction
temperature range
Exclusive Remedies
14
8
Index
I
Inspection
checking the shipment
13
P
power
check the power supply disconnection from supply source power cable options
16
19
16
verification and connection of power cable
S
safety
symbols
7
shutdown
turning the power OFF
18
System Recovery
How to Execute
22
16
Keysight E5080B Installation Guide
27
Page 28
Index
28
Keysight E5080B Installation Guide
Page 29
This information is subject to change without notice.
© Keysight Technologies 2019
Edition 1, March 2019
*E5080-90010*
E5080-90010
www.keysight.com